ADVIA Centaur and ADVIA Centaur XP Immunoassay Systems

Parts and Repair Service Guide

078D0080-02 Rev. A, 2006-07

2006 Bayer HealthCare LLC. All rights reserved. No part of this service guide or the products it describes may be reproduced by any means or in any form without prior consent in writing from Bayer HealthCare LLC. Bayer, the Bayer cross, ADVIA Centaur, and ReadyPack are trademarks of Bayer. BayerCare Connect is a servicemark of Bayer. Corvac and Monoject are trademarks of Sherwood Medical Company. Courier and U.S. Robotics are trademarks of 3COM Corporation. GlidePoint is a trademark of Cirque Corporation. Lexmark is a trademark of Lexmark International, Inc. Microsoft Excel is a trademark of Microsoft Corporation. Microtainer and Vacutainer are trademarks of Becton, Dickinson and Company.

Monovette is a trademark of Walter Sarstedt Kunststoff-Spritzgusswerk. Sparc is a trademark of Sparc International, Inc. Origin: Ireland

The information in this Parts and Repairs guide was correct at the time of printing. However, Bayer HealthCare LLC continues to improve products and reserves the right to change specifications, equipment, and maintenance procedures at any time without notice. If the system is used in a manner differently than specified by Bayer HealthCare LLC, the protection provided by the equipment may be impaired. See warning and hazard statements.

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Contents 1 Parts, Parts Lists, Replacement Procedures


Cabinetry for the ADVIA Centaur System
System Overview - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 System Overview Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 A/C Input Line Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4 Monitor Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5 Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6 Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7 Electronics Module Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8 Sample Tip Waste Area Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9 Cuvette Waste Area Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10 Lab Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12 Parts List - Cabinetry for ADVIA Centaur System . . . . . . . . . . 1-13

Cabinetry for the ADVIA Centaur XP System


System Overview - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17 System Overview - Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18 A/C Input Line Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18 Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18 Monitor Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19 Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19 Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20 Electronics Module Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20 Corner Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21 Lab Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23 Parts List - Cabinetry for ADVIA Centaur XP System . . . . . . . 1-29

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Cuvette Waste Area


Cuvette Waste Area Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-33 Parts List Cuvette Waste Area . . . . . . . . . . . . . . . . . . . . . . . . 1-35

Cuvette Loader 113339


Cuvette Loader Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-37 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-38 Parts List Cuvette Loader 113339 . . . . . . . . . . . . . . . . . . . . . 1-41 Replacing the Cuvette Chute Solenoid. . . . . . . . . . . . . . . . . . . 1-43
Removing the Cuvette Chute Solenoid . . . . . . . . . . . . . . . . . . . . . . . . 1-43 Installing the New Cuvette Chute Solenoid Assembly . . . . . . . . . . . . . 1-44 Verifying the Operation of the Cuvette Chute Solenoid Using Diagnostics Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-44 Verifying the Operation of the Cuvette Chute Solenoid Using Controller Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-45

Elevator 571023B
Cuvette Elevator PC Board Location . . . . . . . . . . . . . . . . . . . . 1-47 Belt Clamp Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-48 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-49 Parts List Elevator 571023B . . . . . . . . . . . . . . . . . . . . . . . . . . 1-51

Incubation Chamber 571021B


Incubation Chamber Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-53 Wash Station Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56 TED Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56 Parts List Incubation Chamber 571021B . . . . . . . . . . . . . . . . 1-59 Replacing the Incubation Chamber Bearings . . . . . . . . . . . . . 1-63 Replacing the Index Pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-67 Replacing the Incubation Chamber TEDs . . . . . . . . . . . . . . . . 1-69
Servo Motor (Quad B) or Wash Station Area (Quad A) . . . . . . . . . . . . 1-69 Primary Reagent Compartment (Quad D) or Back Area (Quad C) . . . 1-71

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Index Mechanism 571024B


Index Mechanism Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-75 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-75 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-76 Parts List Index Mechanism 571024B . . . . . . . . . . . . . . . . . . 1-79

Luminometer 078-K122-01
Luminometer Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-81 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-82 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-83 Parts List Luminometer 078-K122-01. . . . . . . . . . . . . . . . . . . 1-85 Replacing the PMT TED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-89 Replacing the PMT Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 1-95

Preheater 078-B059-01
Preheater Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-99 Cuvette Pusher Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-100 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-101 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-101 TED Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-102 Parts List Preheater 078-B059-01. . . . . . . . . . . . . . . . . . . . . 1-105 Replacing the Preheater TEDs . . . . . . . . . . . . . . . . . . . . . . . . 1-107

Ring Loader 572400B


Ring Loader Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-111 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-112 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-113 Parts List Ring Loader 572400B. . . . . . . . . . . . . . . . . . . . . . 1-115

Servo Motor 571730B (Old)


Servo Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-117 Latch Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-118

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Servo Motor (Old) Detail 571730B . . . . . . . . . . . . . . . . . . . . . . 1-118

Servo Motor 571022B (New)


Servo Motor (New) Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-121 Latch Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-122 Servo Motor (New) Detail 571022B . . . . . . . . . . . . . . . . . . . . . 1-123 Parts List Servo Motor 571730B and 571022B . . . . . . . . . . 1-125

Wash Station 571025D


Wash Station Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-127 Acid Probe Assembly Detail (572889). . . . . . . . . . . . . . . . . . . 1-128 Friction Pad Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-129 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-129 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-130 Parts List Wash Station 571025D . . . . . . . . . . . . . . . . . . . . . 1-133

Electronics Module (fixed configuration) 121911D


Electronics Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-135 Backplane PC Board 78-B047-01 (Fixed Configuration) . . . . 1-136 RT SPARC I/O PC Board 575267 . . . . . . . . . . . . . . . . . . . . . . . 1-137 Fixed Electronics Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-138

Electronics Module (slide out configuration) 571115B


Electronics Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-139 BackPlane PC Board 575252F (Slide Out) . . . . . . . . . . . . . . . 1-140 Slide Out Electronics Module . . . . . . . . . . . . . . . . . . . . . . . . . 1-141 Parts List Electronics Module 571115B and 121911D . . . . 1-143 Replacing the Real Time SPARC PC Board . . . . . . . . . . . . . . 1-145 Replacing the Real Time SPARC PC Board . . . . . . . . . . . . . . 1-147 Replacing the User Interface SPARC PC Board (pre FCO 15) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-149

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

User Interface Module 078-0103-XX for the ADVIA Centaur System


User Interface Module Assembly . . . . . . . . . . . . . . . . . . . . . . 1-151 Parts List - User Interface Module 078-0103-XX* for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-153

User Interface Module 078-0330-XX for the ADVIA Centaur XP System


User Interface Module Assembly . . . . . . . . . . . . . . . . . . . . . . 1-155 Parts List - User Interface Module 078-0330-XX* for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-157 Replacing the Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . 1-159
Removing the UI Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-159 Removing the Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-159 Installing the New Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-161 Installing the UI Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-162

Application Module 078-0331-XX for the ADVIA Centaur System


Application Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-163 Parts List Application Module 078-0331-XX* for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-165

Power Supply
Power supply 571258D (slide out electronics module). . . . . 1-167 Power Supply 078B04701 (fixed electronics module). . . . 1-168 Parts List Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-169 Replacing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . 1-171

Thermal Control PC Module for the ADVIA Centaur System S575257E


Thermal Control PC Module Diagram . . . . . . . . . . . . . . . . . . . 1-173 Parts List Thermal Control PC Module for the ADVIA Centaur System 575257E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-175

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Thermal Control PC Module for the ADVIA Centaur XP System XP 078-B109-XX


Thermal Control PC Module Centaur XP Diagram. . . . . . . . . 1-177 Parts List Thermal Control PC Module for the ADVIA Centaur XP System 078-B109-XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-179 Replacing the Thermal Control Module or the PCB . . . . . . . 1-181

Reagent Fluidics Module for the ADVIA Centaur System 078-B028-01


Reagent Fluidics Module Assembly . . . . . . . . . . . . . . . . . . . . 1-183 Diluter 2 mL 078-B061-01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-184 TriContinent Diluter 078-B061-01 . . . . . . . . . . . . . . . . . . . . . 1-185 Parts List Reagent Fluidics Module 078-B028-01 . . . . . . . . 1-187

Reagent Fluidics Module for the ADVIA Centaur XP System 078-B028-01


Reagent Fluidics Module Assembly . . . . . . . . . . . . . . . . . . . . 1-189 Diluter 2 mL 115401C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-190 Tri-Continent Diluter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-191 Parts List Reagent Fluidics Module 078-B028-01 . . . . . . . . 1-193

System Fluids Drawer 078-B055-01 for the ADVIA Centaur System


System Fluids Drawer Assembly . . . . . . . . . . . . . . . . . . . . . . 1-195 Wash Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-198 Diluter Detail 115401C (2 mL) 115402C (5 mL). . . . . . . . . . . . 1-199 Tri-Continent Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-199 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-201 Parts List System Fluids Drawer 078-B055-01 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-203

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

System Fluids Drawer 078-081-01 for the ADVIA Centaur XP System


System Fluids Drawer Assembly . . . . . . . . . . . . . . . . . . . . . . 1-207
Wash Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-210

Diluter Detail 115401C (2 mL) 115402C (5 mL). . . . . . . . . . . . 1-211 Tri-Continent Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-211 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-213 Parts List System Fluids Drawer 078-B081-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-215

Vacuum Pump 572703B


Vacuum Pump Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-219

Vacuum Pump 078-B050-02


Vacuum Pump Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-221 Parts List Vacuum Pump 572703B. . . . . . . . . . . . . . . . . . . . 1-223

Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Waste and Water Drawer Assembly . . . . . . . . . . . . . . . . . . . . 1-225 Water Reservoir Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-226 Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-227 Water Pressure Pump Detail . . . . . . . . . . . . . . . . . . . . . . . . . . 1-228 Caster Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-229 Parts List Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-231

Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Waste and Water Drawer Assembly . . . . . . . . . . . . . . . . . . . . 1-235 Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-236 Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-237 Water Pressure Pump Detail . . . . . . . . . . . . . . . . . . . . . . . . . . 1-238

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Caster Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-239 Parts List Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-241

Primary Reagent Compartment 571111C


Primary Reagent Compartment Assembly. . . . . . . . . . . . . . . 1-245 Lead Screw Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-251 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-252 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-252 TED Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-253 Parts List Primary Reagent Compartment 571111C. . . . . . 1-255 Replacing the Primary Reagent Compartment Leadscrew. . 1-259 Replacing the Primary Reagent Compartment TEDs . . . . . . 1-269

Primary Reagent Compartment Door 571104B


Primary Reagent Compartment Door Assembly . . . . . . . . . . 1-279 Parts List Primary Reagent Compartment Door 571104B . 1-281

Reagent Barcode Scanner 571103


Reagent Barcode Scanner Assembly . . . . . . . . . . . . . . . . . . . 1-283 Belt Clamp Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-284 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-284 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-285 Parts List Reagent Barcode Scanner 571103C . . . . . . . . . . 1-287

Reagent Probe Guide 571110B


Reagent Probe Guide Assembly . . . . . . . . . . . . . . . . . . . . . . . 1-289 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-291 Parts List Reagent Probe Guide 571110B . . . . . . . . . . . . . . 1-293

Reagent Probe 1 571107C


Reagent Probe 1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-295

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Reagent Probe 1 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-297 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-297 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-297 Parts List Reagent Probe 1 571107C . . . . . . . . . . . . . . . . . . 1-301

Reagent Probe 2 571108C


Reagent Probe 2 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-303 Reagent Probe 2 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-304 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-305 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-305 Parts List Reagent Probe 2 571108C . . . . . . . . . . . . . . . . . . 1-309

Reagent Probe 3 571109C


Reagent Probe 3 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-311 Reagent Probe 3 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-312 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-313 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-313 Parts List Reagent Probe 3 571109C . . . . . . . . . . . . . . . . . . 1-317

Ancillary Queue for the ADVIA Centaur System 078-B035-01


Ancillary Queue Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-319 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-321 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-323 Parts List Ancillary Queue 078-B035-01 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-325

Ancillary Queue for the ADVIA Centaur XP System 078-B035-01


Ancillary Queue Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-327 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-329 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-331 Parts List Ancillary Queue 078-B069-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-333

078D0080-02, Rev. A

10

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Replacing the Ancillary Queue TEDs . . . . . . . . . . . . . . . . . . . 1-335

Ancillary Entry 078-B036-01


Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-339 Ancillary Entry Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-340 Leadscrew Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-340 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-341 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-341 Parts List Ancillary Entry 078-B036-01 . . . . . . . . . . . . . . . . 1-345

Inprocess and Exit Queue 571040B


Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-347 Motor Gap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-348 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-349 Parts List Inprocess and Exit Queue 571040B . . . . . . . . . . 1-351

Sample and Ancillary Probe for the ADVIA Centaur System 078-B045-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-353 Sample Motor and Belt Location. . . . . . . . . . . . . . . . . . . . . . . 1-355 Sample Probe Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . 1-356
Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-356

Ancillary Probe Assembly Detail. . . . . . . . . . . . . . . . . . . . . . . 1-357


Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-357

Sample Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-358 Ancillary Diluter Sample Air Pump Assembly Detail. . . . . . . 1-358 Diluter 2 mL 115401 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-359 TriContinent Diluter 572888 . . . . . . . . . . . . . . . . . . . . . . . . . . 1-361 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-361

Sample and Ancillary Probe for the ADVIA Centaur XP System 078-B123-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-363

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

11

Sample Motor and Belt Location. . . . . . . . . . . . . . . . . . . . . . . 1-365 Sample Probe Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . 1-366
Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-366

Ancillary Probe Assembly Detail. . . . . . . . . . . . . . . . . . . . . . . 1-366


Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-366

Sample Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-368 Ancillary Diluter Sample Air Pump Assembly Detail. . . . . . . 1-368 Diluter 2 mL 115401 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-369 TriContinent Diluter 572888 . . . . . . . . . . . . . . . . . . . . . . . . . . 1-371 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-371 Parts List Sample and Ancillary Probe 078-B123-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-373 Parts List Sample and Ancillary Probe 078-B045-01 . . . . . 1-377

Sample Entry Queue for the ADVIA Centaur System 078-B002-01


Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-381 Belt Tensioner Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-382 Belt Drive Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-382 Magnet Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-383 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-383 Parts List Sample Entry Queue for the ADVIA Centaur System 078-B002-01. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-387

Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01


Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-389 Belt Tensioner Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-389 Belt Drive Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-390 Magnet Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-391 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-391 Parts List Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-395

078D0080-02, Rev. A

12

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Sample Exit Pusher 571037B


Sample Exit Pusher Assembly . . . . . . . . . . . . . . . . . . . . . . . . 1-397 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-397 Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-397 Parts List Sample Exit Pusher 571037B . . . . . . . . . . . . . . . 1-401

Sample Tip Loader 571036C for the ADVIA Centaur System


Sample Tip Loader Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-403 Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-404 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-405 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-405

Sample Tip Loader 078-B136-01 for the ADVIA Centaur XP System


Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-408 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-409 Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-409 Parts List Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . 1-411

Sample Tip Waste Area for the ADVIA Centaur System 078-B0009-02
Sample Tip Waste Area Assembly . . . . . . . . . . . . . . . . . . . . . 1-413 Parts List Sample Tip Waste Area 078-B009-02 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-417

Sample Tip Waste Area for the ADVIA Centaur XP System 078-B088-01
Parts List Sample Tip Waste Area 078-B088-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-423

Stat Sample Pusher 571035


Stat Sample Pusher Assembly . . . . . . . . . . . . . . . . . . . . . . . . 1-425 Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-426

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

13

Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-427 Parts List Stat Sample Pusher 078-B026-01 . . . . . . . . . . . . 1-429

Universal Rack Option


Sample Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-431 Stat Sample Pusher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-432 Sample Exit Pusher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-433 Inprocess Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-434 Rack Sensor Pusher Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-435 Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-436 Sample Exit Queue Rack Present Sensor . . . . . . . . . . . . . . . 1-437 Sample Exit Queue Full Sensor. . . . . . . . . . . . . . . . . . . . . . . . 1-438 Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-439 Sample Entry Queue Work Surface . . . . . . . . . . . . . . . . . . . . 1-440 Sample Entry Queue Tensioner . . . . . . . . . . . . . . . . . . . . . . . 1-441 Parts List Universal Rack Option . . . . . . . . . . . . . . . . . . . . . 1-443

Miscellaneous Parts
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-445

2 Kits and Tools


Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

3 Centaur PC Boards
Sample and Ancillary Probes Horizontal PC Board 575212C . 3-1 Inprocess Queue Tip Sensor PC Board 575217B . . . . . . . . . . . 3-2 Ancillary Queue Sensor PC Board 575222B . . . . . . . . . . . . . . . 3-3 Sample Probe Vertical Sensor PC Board for the ADVIA Centaur System 575227 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 Sample Probe Vertical Sensor PC Board for the ADVIA Centaur XP System 118365 . . . . . . . . . . . . . . . . . . . . . . 3-5 Ancillary Entry PC Board 575232D . . . . . . . . . . . . . . . . . . . . . . 3-6

078D0080-02, Rev. A

14

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

Rack Loaded Sensor PC Board 575237B . . . . . . . . . . . . . . . . . 3-8 Entry Queue Sensor PC Board 575242B . . . . . . . . . . . . . . . . . . 3-9 Low Level Controller PC Board 078-B042S01 . . . . . . . . . . . . . 3-10 Backplane PC Board 575252F (slide out electronics module) 3-16 Backplane PC Board S117937D (fixed electronics module). . 3-21 Thermal Control Module PC Board for the ADVIA Centaur System S575257E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26 Thermal Control Module PC Board for the ADVIA Centaur XP System 078-B111SXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29 Servo Low Level Controller PC Board S575262C . . . . . . . . . . 3-32 Fluid GIOB PC Board S575272E . . . . . . . . . . . . . . . . . . . . . . . . 3-34 Wash Fluid Interconnect PC Board 575277B . . . . . . . . . . . . . . 3-36 Ancillary Probe Vertical PC Board S575282C . . . . . . . . . . . . . 3-37 Reagent Handling GIOB PC Board S575287C . . . . . . . . . . . . . 3-38 Rack Present PC Board S575292B . . . . . . . . . . . . . . . . . . . . . . 3-40 Reagent Fluid GIOB PC Board 078-B027S02 . . . . . . . . . . . . . . 3-41 Exit Queue Interconnect PC Board 575302C . . . . . . . . . . . . . . 3-43 Stat Area PC Board 575307C . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-44 Sample and Ancillary Probe GIOB PC Board S575322H. . . . . 3-45 Sample and Ancillary Queue GIOB PC Board S575327D . . . . 3-47 Cuvette Elevator PC Board 575337 . . . . . . . . . . . . . . . . . . . . . . 3-49 Top Luminometer PC Board 575342C . . . . . . . . . . . . . . . . . . . 3-50 Reagent Probe Pack-to-Pack PC Board 575347 . . . . . . . . . . . 3-51 Sample Tip Loader PC Board 575352B. . . . . . . . . . . . . . . . . . . 3-52 Servo Motor Interface PC Board 575357C . . . . . . . . . . . . . . . . 3-53 Aspirate Probe PC Board 575362 . . . . . . . . . . . . . . . . . . . . . . . 3-54 Bottom Luminometer PC Board 575367 . . . . . . . . . . . . . . . . . . 3-56 Preheater PC Board 575372. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-57 Cuvette GIOB PC Board S575377F . . . . . . . . . . . . . . . . . . . . . . 3-58 Reagent Pack LED Array PC Board 575387 . . . . . . . . . . . . . . . 3-61 Ancillary Entry LEDs PC Board 575392B . . . . . . . . . . . . . . . . . 3-62 Tip Tray and Band PC Board S575397C . . . . . . . . . . . . . . . . . . 3-63

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1

15

Reagent Probe 1 Heater Control and Volume Detect Probe Interconnect PC Board 575402B . . . . . . . . . . . . . . . . . . . . . . . . 3-64 Reagent Probe 2 Heater Control and Volume Detect Probe Interconnect PC Board 575407C . . . . . . . . . . . . . . . . . . . . . . . . 3-65 Reagent Probe 3 Heater Control and Volume Detect Probe Interconnect PC Board 575412C . . . . . . . . . . . . . . . . . . . . . . . . 3-66 Mixer Interconnect PC Board 575422C. . . . . . . . . . . . . . . . . . . 3-67 Reagent Probe 1 Horizontal Interconnect PC Board 575432B 3-68 Reagent Probe 2 Horizontal Interconnect PC Board 575437 . 3-69 Reagent Probe 3 Horizontal Interconnect PC Board 575442B 3-70 Reflective Chute Sensor Assembly PC Board 575447B . . . . . 3-71 System Fluid Detect PC Board 575452C . . . . . . . . . . . . . . . . . 3-72 Ancillary Pack Sensor PC Board 575457B . . . . . . . . . . . . . . . . 3-74 Sample Pressure Level Sense PC Board 575462E . . . . . . . . . 3-75

Appendix A: Safety Instructions


Protecting Yourself from Biohazards . . . . . . . . . . . . . . . . . . . . . A-1
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2

Protecting Yourself from Barcode Scanner Lasers . . . . . . . . . A-2


Laser Safety Classification of the ADVIA Centaur System. . . . . . . . . . .A-2 Laser Safety Classification of the Barcode Scanners. . . . . . . . . . . . . . .A-3 Reagent and Sample Barcode Scanners . . . . . . . . . . . . . . . . . . . . . . . .A-3 Hand-held Barcode Scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5

Appendix B: Service, Ordering, and Warranty


Bayer Authorized Representative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1 Limited Instrument Warranty and Service Delivery Policy . . . . . . . . . . .B-1 Warranty Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1

Information for Technical Assistance. . . . . . . . . . . . . . . . . . . . . B-4


Printing the System Serial Number and Versions. . . . . . . . . . . . . . . . . .B-5 Printing the Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5 Printing Windows on the Workspace . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6 Printing a Runtime Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6 Printing Calibration Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6

Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
Bayer Authorized Representative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6 Addresses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6

078D0080-02, Rev. A

1 Parts, Parts Lists, Replacement Procedures Cabinetry for the ADVIA Centaur System
System Overview - Front

078D0080-02, Rev. A

1-2

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Overview Back

System Overview Back

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 A/C Input Line Assembly

1-3

A/C Input Line Assembly

078D0080-02, Rev. A

1-4

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Top Cover

Top Cover

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Monitor Stand

1-5

Monitor Stand

078D0080-02, Rev. A

1-6

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue

Sample Entry Queue

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue

1-7

Sample Exit Queue

078D0080-02, Rev. A

1-8

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Electronics Module Door

Electronics Module Door

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Door

1-9

Sample Tip Waste Area Door

078D0080-02, Rev. A

1-10

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Door

Cuvette Waste Area Door

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Door

1-11

078D0080-02, Rev. A

1-12

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation

Lab Automation

078D0080-02, Rev. A

Parts List - Cabinetry for ADVIA Centaur System


Item Description System Overview - Front REF Part Number

System Overview - Back

5c 13

Cover, Cable UI Corner, User Interface Cap Box, Stop Switch Cover, Cable Access
Top Cover

121575 572534 078-0111-01 572381

Monitor Stand Assembly

1 2 3 4 5 6

Clamp, Monitor Holder, Barcode Scanner Screw, Flat M4 x 0.7 x 12 mm Handle, UI Clamp, Keyboard Palmrest Assembly Glidepoint and Cable External Glidepoint and Cable

5513292

572761 115670 571781 572757 078010001

1109578 8957558 478537

078B05801 078-0236-01 111215 112864 572851

7 8 33 34 35 36 37 38 12 13 14

Screw, Thumb M6 x 1.0 x 30 mm Nut, Knurled M6 x 1.0 x 15 mm


Entry Queue

Cover, Entry Queue Housing, Stat Area PC Board, Ancillary Entry LEDs Tray, Entry Queue Work Surface, Entry Queue Panel, Entry Queue, Back Cover Cover, Entry Queue Back
Electronic Module Door

5585684 1765254 4253513

572541 572535C 575392B 078-0006-01 078-B001-01 110214 571996B 572525 72426B 572524

Latch, Adjustable Grip Retainer Bracket, Door Hinge

078D0080-02, Rev. A

1-14

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List - Cabinetry for ADVIA

Item

Description

REF

Part Number

15 16 17 18 19 20 21 22

Hinge, Door Door, Electronics Compartment Adapter, Monitor Clamp Screw, M4x.8 Phillips Access Panel, Reagent Probes Switch, Emergency Stop (red) Switch, Computer Startup (green) Filter, Air Retainer, Filter, Air
Lab Automation

572370 572365D 078007501 9843467 9132080 60007x55 572155 572158 572516 3918457 572165 572164

23 24 25 26

Screw, Truss M5 x 12 mm Cover, Rear Sample, Lab Automation Panel, Upper Back, Lab Automation Cover, Window, Lab Automation
Exit Queue

117101 115867 115866 115868

27 28 29 30 31 32

Work Surface, Exit Queue Cover, Exit Queue Rack Cover, Exit Queue Tray, Exit Queue Bracket, Midlevel Front Block, Queue Cover Hinge Cover, Back
Sample Tip Waste Area

571985 571963C 5625414 572542 571004 110249B 572537B 078-0118-01

39 40 41 42 43 44 45

Door, Sample Tip Waste Handle, Pocket Pull Hinge, Upper Leaf Hinge, Single Bolt Hinge, Single Bolt Hinge, Lower Leaf Catch, Magnetic

572366 572374 572376 572377 572375 572368

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List - Cabinetry for ADVIA Centaur

Item

Description

REF

Part Number

Bracket, Magnet Mounting 46 Housing, Interior Tip Waste


Cuvette Waste Area Door

110611 572178

47 48 49 50 51 52 53 54 55 56 57 58 59 60

Housing, Cuvette Waste Bracket, Retaining Handle, Pocket Pull Catch, Twin Ball Support, Sample Entry Queue Door, Cuvette Waste Line Assy, A/C Input Not Used Hinge, Upper Leaf Hinge, Single Bolt Hinge, Single Bolt Hinge, Lower Leaf Bin, Tip Tray Waste Cover, Fluid Module Access Bracket, Fluid Module Access Cover, Right Bracket, Fluid Module Access Cover, Left Fluids Bracket, Access Panel

572363B 110408B 572369 572530 572079 572364 572271B 572372 572377 572376 572373 572425 572408 572409 572410 572160 121470 572605 572604B 572608B 572607 078B03801 572619 7072218 117159 846-1021-09 572531

61 62 63 64 65 66 67 9 10 11

Top Cover Window, Diluter Window, Front Cover, Sample Tip Loader Window, Reagent Probes Window, Top Handle, Top Window Shield, Air Deflector, Lab Automation Washer, Flat M6 Cover, Ancillary Queue

078D0080-02, Rev. A

Cabinetry for the ADVIA Centaur XP System


System Overview - Front

078D0080-02, Rev. A

1-18

header Melias Operating Manual

System Overview - Back

A/C Input Line Assembly

Top Cover

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Monitor Stand

1-19

Monitor Stand

Sample Entry Queue

078D0080-02, Rev. A

1-20

header Melias Operating Manual

Sample Exit Queue

Electronics Module Door

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Corner Door

1-21

Corner Door

078D0080-02, Rev. A

1-22

header Melias Operating Manual

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation

1-23

Lab Automation

078D0080-02, Rev. A

1-24

header Melias Operating Manual

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation

1-25

078D0080-02, Rev. A

1-26

header Melias Operating Manual

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation

1-27

078D0080-02, Rev. A

Parts List Cabinetry for ADVIA Centaur XP System


Item Description System Overview - Front REF Part Number

1 2 3

Cover, Ancillary Queue Cover, Fluid Module Access Box, Stop Switch Bracket, Fluid Module Access Cover, Right Bracket, Fluid Module Access Cover, Left Fluids Bracket, Access Panel

078-B091-01 572408 078-0371-01 572409 572410 572160

System Overview - Back

4 5 6 7 8 9 10 10 12 13 14 15 16

Switch, Computer Startup (green) Cover, Service Back access Screw Filter, Air Retainer, Air Filter Line Assy, A/C Input Cover, Right Back Cover, Back Access, Electronics Module Cover, Cable UI Cover, Cable Access Shield, Air Deflector, Lab Automation Corner, User Interface Cap Switch, Emergency Stop (red) 7072218 3918457

572516 078-0313-01 60012x2 572165 078-0315-01 572271B 078-0312-01 078-0314-01 078-0422-01 078-0372-01 117159 078-0369-01 572158

Top Cover

8 9

Cover Top, Rear Cover Top, Front Window, Top Access

078-0347-01 078-0348-01 078-0349-01

Monitor Stand Assembly

078-B122-01

078D0080-02, Rev. A

Item

Description

REF

Part Number

10 11 12 13 14 15

Assembly, Arm Clamp, Cable Holder, Barcode Scanner Nut, M4 Acorn Hex Tray, Keyboard Touchscreen Keyboard

not orderable separately 509-0118-01 078-0457-01 543-0173-08 078-0407-01 515-7247-01 See Miscellaneous Section

Entry Queue

16 17 18 19 20 21 22

Cover, Entry Queue Housing, Stat Area PC Board, Ancillary Entry LEDs Tray, Entry Queue Work Surface, Entry Queue Panel, Entry Queue, Back Cover Cover, Entry Queue Back

5585684 1765254 9169294

572541 078-0368-01 575392B 078-0006-01 078-0302-01 110214 078-0384-02

Electronic Module Door

078-0382-01

23 24 25 26 20

Latch, Adjustable Grip Bracket, Door Hinge Hinge, Door Door, Electronics Compartment Screw, M4x.8 Phillips 9843467

572525 572524 572370 078-0382-01 60007x55

Lab Automation

24 25 26 27

Screw, Truss M5 x 12 mm Cover, Rear Sample, Lab Automation Panel, Upper Back, Lab Automation Cover, Window, Lab Automation
Exit Queue

117101 115867 115866 115868

28

Work Surface, Exit Queue

078-0366-01

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Cabinetry for

Item

Description

REF

Part Number

29 30 31 32

Cover, Exit Queue Tray, Exit Queue Bracket, Midlevel Front Block, Queue Cover Hinge Cover, Back

5625414

572542 571004 110249B

572537B

572537B 078-0365-01

Sample Tip Waste Area

39 40 41 42 43 44 45 46

Door, Sample Tip Waste Handle, Pocket Pull Hinge, Upper Leaf Hinge, Single Bolt Hinge, Single Bolt Hinge, Lower Leaf Catch, Magnetic Bracket, Magnet Mounting Housing, Interior Tip Waste
Cuvette Waste Area Door

572366 572374 572376 572377 572375 572368 110611 572178


078-B094-01

46 47 48 49 50 51

Cover, Corner Door Magnet Mount, door Catch Plate, Corner Door Magnet, Shielded Screw, Countersink, Flat Bin, Tip Tray Waste

078-0308-01 078-0453-01 078-0452-01 539-0010-01 846-1022-34 572425

I THINK THE FOLLOWING ARE NOT NEEDED 62 63 64 65 67 Window, Window, Front Cover, Sample Tip Loader Window, Reagent Probes Handle, Top Window 572605 572604B 572608B 572607 572619

078D0080-02, Rev. A

Cuvette Waste Area


Cuvette Waste Area Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Assembly

1-34

078D0080-02, Rev. A

Parts List Cuvette Waste Area


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14

Collar, Cuvette Waste Bin Screw, Retractable, M4 x 0.7 x 5.7 mm Shutter, Waste Reservoir Chute, Cuvette Waste Sensor, Optical Switch Bin, Cuvette Waste Liner, Cuvette Waste Clamp, Spring Screw, Pan-Head M4 x 0.7 x 10 mm Spring, Waste Bin Interlock Not Used Screw, Pan-Head, M4 x 0.7 x 6 mm Washer, Flat, M4 Reservoir, Waste Chute

06907707 00220165 156351 00962374 06412376 0764029 04324275 02968655 04215735

110106 571964 572622 572544 111674 572615 078-0067-01 110242 60000x65 115023 60000x35

04563393 09580326

60028x5 572547

078D0080-02, Rev. A

Cuvette Loader 113339


Cuvette Loader Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-38

Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-39

078D0080-02, Rev. A

Parts List Cuvette Loader 113339


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Clamp, Torque Transfer Cam, Wheel Bearing, Ball, 0.250 x 0.500 x 0.187 Ring, Retaining, E-ring, 0.250D Support, Cuvette Loader Spacer, 0.253 x 0.373 x 0.010 Bearing, Journal, Flanged 0.250 x 0.375 x 0.250 Cam, Arm Screw, Cap-Head M4 x 0.7 x 55 mm Screw, Cap-Head M4 x 0.7 x 9 mm Washer, Lock Sensor, PC Board Chute Full Kit, Cuvette Chute Solenoid Tube, Drop Screw, Cap-Head M4 x 0.7 x 12 mm Chute, Orientation Fastener Gate, Lift, Cuvette Wedge, Cuvette Loader Screw, Button-Head M4 x 0.7 x 6 mm Ionizer, Air Bracket, Ionizer Support Bin, Cuvette Loader Block, Mixer Screw, Cap-Head M4 x 0.7 x 16 mm Block, Slotted Shaft, Idler Sprocket Support, Cuvette Loader Screw, Cap-Head M4 x 0.7 x 50 mm Washer, Flat, M4 Motor, Stepper, 6VDC

396069 911338 5785519 03674086 06373893 6068713 7015400 00593158 06913146 04162038 05384913 01353061 01876943 09678857 01662811 03536260 09725693 7904256 09864405

26586 26490 20573 26107 571727B 12861 26082x7 112701 60012x107 60012x9 60026x5 575447B 120390 572069 60012x23 27145 27186x2 27187B 27325 572500 572268 572413 572071B 112702

02345038

60012x37 112703 26373

03674086 04563393 03828563

571727B 60012x100 60028x5 118498

078D0080-02, Rev. A

1-42 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Cuvette Loader 113339

Item

Description

REF

Part Number

32 33 34 35 36 37 38 39 40 41 42 43 44 45 46

Bushing, Sprocket, Mounting Screw, Set M4 x 0.7 x 04 mm Sprocket, 15T, 0.500P, Single Strand Screw, Set 1/4-20 x 0.25 Spacer, Stepped Block, Chain, Tension Adjuster Screw, Pan-Head M4 x 0.7 x 20 mm, Slotted Spacer, Stepped Bearing, Flanged, Journal 0.251 x 0.377 x 0.375, Bronze Chain, Cuvette, Lift Guide, Chain Support, Hopper, Bin, Back Support, Feeder, Angled Bearing, Clutch, One-Way 0.250 x 0.438 x 0.500 Partition, Cuvette Bin 9759016

26444 60023x40 26376 26573 26509 116684 60003x95 26507 26572 26746 26495 26570 26505 19556 123281

078D0080-02, Rev. A

Replacing the Cuvette Chute Solenoid


1. At the workspace, select System Status. 2. Select Turn Mechanics Off, and then wait while the system mechanics power down.

Removing the Cuvette Chute Solenoid


CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components. 1. Disconnect the chute full sensor cable from J4 on the Preheater PC board. 2. Disconnect the cuvette chute solenoid cable from the cable extending through the opening at the back of the luminometer.
NOTE: Push the waste probe arm down to allow easier access to the cuvette chute solenoid assembly.

3. Remove the four screws and eight washers that secure the cuvette chute solenoid assembly to the orientation chute. Save the screws and washers. 4. Remove the cuvette chute solenoid assembly.

Figure 1-1

Removing the Cuvette Solenoid Assembly

5. Remove the two screws that secure the chute full sensor to the cuvette chute solenoid assembly. Save the screws and sensor, and then discard the cuvette chute solenoid assembly.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid

Installing the New Cuvette Chute Solenoid Assembly


1. Install the chute full sensor to the new cuvette chute solenoid assembly using the two screws removed previously. 2. Install the new cuvette chute solenoid assembly to the orientation chute using the four screws and eight washers removed previously.

Figure 1-2

Installing the Cuvette Chute Solenoid

3. Connect the cable from the chute full sensor to J4 on the Preheater PC board.

Figure 1-3

Preheater PC Board

4. Connect the cable from the cuvette chute solenoid to the cable extending through the opening at the back of the luminometer. Route these cables so they do not interfere with the cuvette loader cam. 5. Select Turn Mechanics On.

Verifying the Operation of the Cuvette Chute Solenoid Using Diagnostics Tools
Use this procedure to cycle the cuvette chute solenoid in diagnostics tools. 1. At the workspace, select System Status.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid

2. Select Diagnostics Tools. 3. At the System-Diagnostics Tools window, select Empty & Fill Ring. 4. Select Perform.
WARNING

Before selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury. 5. As the system performs the empty & fill ring, watch the cuvettes as they slide down the orientation chute and enter the vertical cuvette chute. Ensure that the cuvette chute solenoid cycles as a cuvette loads into the incubation ring. This process takes approximately 12 minutes.

WARNING

Before closing the System-Diagnostic Tools window, ensure that you are clear of subassemblies that can move. The system returns subassemblies to their home positions and can cause injury. 6. Close the System-Diagnostics Tools window.

Verifying the Operation of the Cuvette Chute Solenoid Using Controller Diagnostics
NOTE: Use this procedure if the Empty & Fill Ring procedure in Diagnostics Tools fails.

Use this procedure to cycle the cuvette chute solenoid in controller diagnostics. 1. At the workspace, select System Status. 2. Select Controller Diagnostics. 3. Insert the Controller Diagnostics diskette into the diskette drive.
CAUTION

Wait for the system to access the controller diagnostics sequences on the diskette. If you select the diskette drive twice, damage can occur to the diskette and cause an RT error. 4. Select Diskette in Drive. 5. If your Controller Diagnostics diskette has the cuv_sol sequence, use the following procedure to run the cuv_sol sequence: a. Select the cuv_sol sequence. b. Select Sequence Definition. c. Select Run Sequence.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid

Verify that the cuvette chute solenoid cycles five times.

078D0080-02, Rev. A

Elevator 571023B
Cuvette Elevator PC Board Location

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Clamp Detail

1-48

Belt Clamp Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-49

Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-50

078D0080-02, Rev. A

Parts List Elevator 571023B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19

Screw, Pan-Head M4 x 0.7 x 12 mm Phillips PC Board, Cuvette Elevator Screw, Cap-Head M4 x 0.7 x 16 mm Motor, Stepper Screw, Retractable M5 x 0.8 x 10.5 mm Washer, Flat M3 Screw, Cap-Head #4-40 x 0.375 Guide, Linear Screw, Cap-Head M2 x 0.4 x 8 mm Pulley, Idler, 0.080P 25T 0.290W 0.432B Screw, Shoulder M4 x 0.7 x 14 mm Belt 0.08P 0.25W 190T Carriage, Cuvette Elevator Clamp, Belt Screw, Cap-Head M4 x 0.7 x 8 mm Phillips Pulley, 0.08P 20T 0.290W 0.188B Screw, Cap-Head M2 x 0.4 x 10 mm Pusher, Cuvette Elevator Flag, Elevator EOT 371015 9864200

60013x23 575337 60013x37 572915 572052 60056x3 118447 572064 572418 572046 27014x7 572044 571887 572045 60013x9 572049 572419 572050 572051

078D0080-02, Rev. A

Incubation Chamber 571021B


Incubation Chamber Assembly

078D0080-02, Rev. A

1-54

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Incubation Chamber Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Incubation Chamber Assembly

1-55

078D0080-02, Rev. A

1-56

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Wash Station Cover

Wash Station Cover

Sensor Detail

TED Assembly Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail

1-57

078D0080-02, Rev. A

1-58

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail

078D0080-02, Rev. A

Parts List Incubation Chamber 571021B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23

Cover, Chamber Lumo Area Cover, Chamber Back Area Screw, Cap-Head M4 x 0.7 x 30 mm Cover, Chamber Servo Area Cover, Chamber Sample Area Chute, Luminometer Entrance Cover, Luminometer Entrance Cover, Chamber Wash Separation Area Screw, Cap Head M4 x 70 mm Magnet, Resuspension Window, Incubation Chamber Screw, Flat-Head, Phillips M4 x 0.7 x 5 mm Cover, Chamber Index Area Cover, Chamber R3 Area Cover, Reagent Dispense Access Screw, Retractable M4 x 0.7 x 8 mm Screw, Cap-Head M4 x 0.7 x 65 mm Screw, Cap-Head M4 x 0.7 x 16 mm Screw, Flat-Head M3 x 0.5 x 20 mm Phillips Cover, Cuvette Ring Sensor Housing Sensor, Optical Switch, Ring Tooth Housing, Cuvette Ring Sensor Screw, Hex M4 x 0.7 x 10 mm Ring, Cuvette Assembly Insulation Activator, Loctite Quick Metal, Loctite Plate, Retainer Bracket, Belt Plate, Magnet Mounting 771641 4450912 7205765 5319631

571197B 571198B 60013x72 571192 571193B 571852B 571854B 571194 60013x128 115463 571195 60007x35 571200 571196 571201 571533 60013x128 60013x37 60001x93 572380 571256C 572379 60013x16 111676 571251 111291 111290 571253 571187 115464B

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Incubation Chamber

Item

Description

REF

Part Number

Magnet, Resuspension Screw, Flat-Head, Phillips M4 x 0.7 x 5 mm Strip, Non-Reflective Spring, Compression Screw, Cap-Head M4 x 0.7 x 8 mm Screw, Phillips M2.5 x 0.4 x 5 mm Screw, Set Pulley, Belt, Timing Screw, Cap-Head M3 x 0.5 x 8 mm 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 Pin, Index Detent Spring, Compression Screw, Phillips M2.5 x 0.4 x 5 mm Bearing, Cuvette Ring Screw, Set M5 x 0.8 x 8 mm Screw, Cap-Head M3 x 0.5 x 8 mm Plate, Retainer Pulley, Timing Belt Screw, Flat-Head Screw, Button Not Used Not Used Plate, Magnet Mounting Screw, Cap-Head M4 x 0.7 x 8 mm Bracket, Belt Washer, Flat M5 Screw, M3 x 0.5 x 12 mm Sensor, Cuvette Present Bracket, Cuvette Present Sensor Screw, Cap-Head M3 x 0.5 x 8 mm Gasket, Seal, Cuvette Elevator Gasket, O-ring Screw, Cap-Head M3 x 0.5 x 10 mm 7333968 771641 1453686 7285122 8232340 1453686

115463 60007x35 111604 571511 60013x9 571532 571531 571186 60013x8 571188 571511 571532 571175 571531 60013x8 571253 571186 60007x35 572500 N/A N/A 115464B 60013x9 571187 60056x6 60013x22 111672 113070 60007x53 571202 571521 60013x15

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Incubation Chamber

Item

Description

REF

Part Number

47 48 49 50 51 52 53 54 55

Sensor, Optical Switch Spacer 3.2 mm x 6.0 mm x 3.0 mm Screw, Cap-Head M5 x 0.8 x 25 mm Stop, Magnet Ring O-ring, Metric 4 mm D x 15 mm wide Fitting, M6 to 0.125 Barb Block, TED Mounting TED, Assembly Thermal Grease Thermistor, Threaded Body Shim

7713183

111671 571529 60013x66 571177 571520 111970 571182

5483539 1849083

111677 120491 571184C 557-7282-01

078D0080-02, Rev. A

Replacing the Incubation Chamber Bearings


1. Install the set screws: a. Locate the bearing holes on the magnet ring. b. Screw the set screw until the tip of the screw is flush with the outside of the magnet ring. c. Repeat steps a and b for the five remaining bearing holes.

Figure 1-1

Incubation Ring Set Screw

2. Place the 6 bearings on the magnet ring.

Figure 1-2

Incubation Ring Bearings

3. Tilt the cuvette ring to place it on top of the bearings and the magnet ring.

078D0080-02, Rev. A

1-64 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber

Figure 1-3

Incubation Ring

4. Using a small screw driver, position the six bearings on the magnet ring until they align with the six set screws installed previously. 5. Install the screws into the bearings: a. Position the cuvette ring on top of the bearings. b. Partially install two screws into the bearing. Ensure the cuvette ring moves without interference from the top of the screw. c. Repeat steps a and b for each bearing.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber

Figure 1-4

Incubation Ring Screws

CAUTION

To ensure consistent bearing tension, adjust all of the set screws equally each time you perform this procedure. 6. Adjust the set screws: a. Locate the six set screws that were installed previously. b. Adjust each set screw 1/4 turn clockwise. c. Rotate the cuvette ring and check for excessive sidetoside movement or binding. d. If you feel sidetoside movements, adjust the set screws clockwise. If you feel binding, adjust the set screws counterclockwise.

078D0080-02, Rev. A

Replacing the Index Pin


1. Replace the index pin: a. Locate the index pin hole on the cuvette ring assembly. You can view the index pin hole from the top of the incubation chamber through the inside ring opening. The hole is located on the magnet ring approximately 180 degrees from the incubation chamber drive belt clamp.

Figure 1-1

Index Pin Hole on Magnet Ring

b. Insert the compression spring and index pin into the index pin hole ensuring the pin opening is facing the outside of the ring.

Figure 1-2

Compression Spring

c. While pressing down on the index pin, insert the retainer through the outside of the magnet ring into the index pin.

078D0080-02, Rev. A

1-68

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Index Pin

d. Using the two screws, secure the retainer to the magnet ring.

Figure 1-3

Index Pin

078D0080-02, Rev. A

Replacing the Incubation Chamber TEDs


Servo Motor (Quad B) or Wash Station Area (Quad A)
1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. c. At the back of the system, turn off the power and unplug the power cord. 2. Remove the servo motor or the wash station assembly. 3. Remove the cuvette loader. 4. Remove the defective TEDs: a. Remove the 2 screws that secure the heatsink to the TED mounting block.

Figure 1-1

Heatsink Screws

b. Remove the 2 screws that secure the TED mounting block to the bottom of the incubation ring.

078D0080-02, Rev. A

1-70 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber

Figure 1-2

TED Mounting Block

c. Remove the TED mounting block. d. Remove the defective TEDs. 5. Install the TEDs: a. Apply a thin layer of thermal grease to both sides of the new TEDs. b. Install the new set of TEDs on the bottom of the incubation chamber ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
CAUTION

Ensure the TEDs are installed flush with the bottom of the incubation ring or damage can occur to the TEDs. c. Install the TED mounting block and finger tighten the 2 screws. d. Install the 2 screws that secure the heatsink to the TED mounting block.
CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. e. Tighten the 2 screws 0.9 Nm (8 inlb) on the TED mounting block. f. Connect the cable for the TEDs.

6. Install the servo motor or the wash station assembly. 7. Calibrate the incubation chamber thermals.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber TEDs

Primary Reagent Compartment (Quad D) or Back Area (Quad C)


1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. c. At the back of the system, turn off the power switch and unplug the power cord. 2. Remove the luminometer and the preheater assemblies. 3. Remove the cuvette loader. 4. Remove the defective TEDs: a. Remove the 2 screws that secure the heatsink to the TED mounting block.

Figure 1-3

Removing the Heatsink

b. Remove the 2 screws that secure the heatsink to the base of the system, and then remove the heatsink. c. Remove the 2 screws that secure the TED mounting block.

078D0080-02, Rev. A

1-72 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber

Figure 1-4

Removing the TED Mounting Block

d. Remove the defective TEDs. 5. Install the TEDs:


CAUTION

Ensure that no thermal grease is inside the TED or heating problems will occur. a. Apply a thin layer of thermal grease to both sides of the new TEDs. b. Install the new set of TEDs on the bottom of the incubation chamber with the red (R) and black (B) wires in the same position as the old TEDs.

CAUTION

Ensure the TEDs are installed flush with the bottom of the incubation ring or damage can occur to the TEDs. c. Install the TED mounting block and finger tighten the 2 screws. d. Connect the cable for the TEDs. 6. Install the heat sink: a. Install the heatsink and tighten the 2 screws that secure the heatsink to the base of the system. b. Install the 2 screws that secure the heatsink to the TED mounting block. c. Remove the 2 screws that secure the heatsink to the TED mounting block.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber TEDs

d. Remove the 2 screws that secure the heatsink to the base of the system, and then remove the heatsink.
CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. e. Tighten the 2 screws 0.9 Nm (8 inlb) that secure the TED mounting block. f. Install the heat sink and tighten the 2 screws that secure the heatsink to the Centaur baseplate.

g. Install the 2 screws that secure the heatsink to the TED mounting block. 7. Install the luminometer and preheater assemblies. 8. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 9. Calibrate the incubation chamber thermals.

078D0080-02, Rev. A

Index Mechanism 571024B


Index Mechanism Assembly

Sensor Detail

078D0080-02, Rev. A

1-76

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-77

078D0080-02, Rev. A

Parts List Index Mechanism 571024B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9

Screw, Cap-Head M4 x 0.7 x 10 mm Motor, Stepper 4VDC Shaft, Index Mechanism Flag, EOT Sensor Screw, Pan-Head M2.5 x 0.45 x 5 mm Foot, Index Plunger Housing, Index Mechanism Sensor, Optical Switch Screw, Pan-Head, Self-Tap 962374 9706931 1184235

60013x16 572923 571743 571006 60001x32 23404 571005 111674 571745

078D0080-02, Rev. A

Luminometer 078-K122-01
Luminometer Assembly

078D0080-02, Rev. A

1-82

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-83

Motor Detail

078D0080-02, Rev. A

1-84

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

078D0080-02, Rev. A

Parts List Luminometer 078-K122-01


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11

Screw, Cap-Head M4 x 0.7 x 16 mm Belt, 0.080P .25W 175T Neoprene Clip, Probe Retainer Pulley, Luminometer Rotor Probe, Waste, Aspirate Wheel, Sensor Screw, Pan-Head M2.5 x 0.4 x 5 mm Phillips Flag, EOT Sensor PC Board, Luminometer, Top Screw, Cap-Head M4 x 0.7 x 6 mm Housing, Luminometer Chute, Lumo Exit Screw, Cap-Head 9140830 1190170 8107910 4740554

60013x37 571822 571812B 571548 571813B 571173 60001x32 571006 575342C 60013x2 571009C 571818B 60013x9 571544B 571010 27014x10 8947137 117109B 571821B 571821B 571817 572417 60056x3 572915 571815B 60013x9 571814 572417 572415 962374 111674 571807 571808

12

Rotor, Lumo Segment, Lumo Rotor Screw, Shoulder

13 14 15 16

Guide, Waste Probe Pulley, Timing, 0.080P 15T 0.290W 0.188 Motor, Stepper 4VDC Latch, Over-Center Draw Screw, Cap- Head Washer

17 18 19 20 21 22 23 24 25

Motor, Stepper 4VDC Gear, Spur 32P 24T 0.188W 0.187B Delrin 1116124 Screw, Cap-Head M4 x 0.7 x 8 mm Phillips Plate, Motor Mounting Screw, Cap-Head M3 x 0.5 x 5 mm Hex Screw, Flat-Head #4-40 x 0.25 Phillips Sensor, Optical Switch Spring, Comp 0.240 x 0.022 x 0.250 Stnl Pad, Friction, Short

078D0080-02, Rev. A

1-86

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Luminometer

Item

Description

REF

Part Number

26 27 28 29 30 31 32 33 34

Housing Assy, Waste Housing, Waste Probe Fitting, M6 to 3/16 OD Tube, 0.250 x 0.375 Black Tygon Nut, Knurled M6 Cover, Top Luminometer Clamp Screw, Flat M4 x 0.7 x 20 mm PMT, Assembly ORing, 5.987x.103 BunaN Cover, Luminometer Top Ball Bearing Ring, Internal Retaining Screw, Cap-Head

4805834

571030B 571805 571534 571524 571719B 12811x1 60055x95 571029B 571527 6482145 571546 9854647 60013x1 117108 111678 120491 26740 60013x86 60013x86 572385 575367 60013x23 571804

35 36 37 38 39 40 41 42 43 44

Retainer, Waste Probe Guide TED, Assembly Thermal Grease Gasket, Ted Mounting Screw, Cap-Head M4 x 0.7 x 40 mm Screw, Cap-Head M3 x 0.5 x 30 mm Screw, Set M3 x 0.5 x 4 mm PC Board, Luminometer Bottom Screw, Pan-Head M4 x 0.7 x 12 mm Phillips Heatsink, PMT Ted Kit, Base Probe Assembly Base Probe Tubing, 75 in Tubing, Heatshrink 5.5 in Fitting, arbed Fitting, 1/4-28 O-Ring Teflon Tape 6102075

078-K081-01 078-0267-01 572689 26781 518-0883-01 518-7076-01 116828 19720

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Luminometer 078-K122-01

Item

Description

REF

Part Number

45 46 47 48

Not used Arm, Waste Probe Retainer, E-ring Rod, Pusher 571811B 571672 113744

078D0080-02, Rev. A

Replacing the PMT TED


1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. c. At the back of the system, turn off the power and unplug the power cord. 2. Disconnect the PMT thermistor cable from connector J6 on the Luminometer Bottom PC board. 3. Disconnect the PMT cable from connector J1 on the Luminometer Bottom PC board.
CAUTION

Do not scratch or soil the PMT lens. Scratching or soiling the lens will interfere with the light rays and cause erroneous results.

CAUTION

Ensure that you select Turn System Mechanics Off and cover the PMT lens when exposing the inside of the luminometer to light. Low levels of light can saturate the PMT and it can take several hours for the PMT to recover. High levels of direct light can damage the PMT. 4. Remove the luminometer. 5. Place the lens cover (part number 27127) over the lens of the PMT to protect it from light, scratches, and contamination. 6. Remove the defective TED:

078D0080-02, Rev. A

1-90

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED

a. Remove the two screws that secure the heatsink to the back of the luminometer.

Figure 1-1

Luminometer Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED

1-91

b. Remove the two screws that secure the heatsink to the PMT housing.

Figure 1-2

PMT Housing Assembly

c. Remove the TED gasket. d. Remove the defective TED. 7. Install the TED:
CAUTION

Ensure that no thermal grease is inside the TED or heating problems will occur. a. Apply a thin layer of thermal grease to both sides of the new TED.
CAUTION

Ensure the TED is installed flush with the bottom of the PMT housing or damage can occur to the TED. b. Install the new TED on the bottom of the PMT housing, ensuring the red (R) and black (B) wires are in the same position as the old TED.

078D0080-02, Rev. A

1-92

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED

c. Install the TED gasket.


CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. d. Install the TED heatsink and tighten the two screws. Alternate the tightening of each screw on the TED mounting block to 0.9 N m (8 inlb). e. Remove the lens cover (part number 27127) from the lens of the PMT.
CAUTION

Do not allow silicone oil on the PMT lens. Oil on the lens will Interfere with the light rays and cause erroneous results. f. Apply silicone oil (part number 27295) to a lint free tissue and then apply the oil to the Oring.

g. Install the two screws that secure the heatsink to the back of the luminometer. 8. Install the luminometer. 9. Connect the PMT thermistor cable to connector J6 on the Luminometer Bottom PC board. 10. Connect the PMT cable to connector J1 on the Luminometer Bottom PC board. 11. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 12. Verify PMT performance: a. At the workspace, select System Status. b. Select Diagnostic Tools. c. At the System-Diagnostic Tools window, select Dark Count without Cuvette.
WARNING

Before selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury. d. Select Perform. The System-Dark Count window opens. e. Ensure that the RLU 1 values for Total Mean are less than 600. RLUs higher than 600 may indicate a defective PMT, electronic noise, or light leaks.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED

1-93

f.

Refer to the chart below for the Total SD limit. Match the RLU1 Total Mean with the SD Limit curve. Verify the Total SD value is at or below the SD Limit curve.

Figure 1-3

Total SD Limit Chart

13. Check the dark counts with a cuvette: a. At the System-Diagnostic Tools window, select Dark Count with Cuvette.
WARNING

Before selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury. b. Select Perform. The System-Dark Count window opens. c. Ensure that the difference between RLU 1 values for Total Mean with cuvettes and without cuvettes are less than 25. RLUs higher than 25 may indicate light leaks. 14. If the dark count tests do not meet specifications, refer to the ADVIA Centaur Service Troubleshooting Manual. 15. Calibrate the luminometer thermals.

078D0080-02, Rev. A

Replacing the PMT Assembly


1. Remove the two screws securing the PMT assembly to the luminometer.
NOTE: You may need to pry the PMT away from the luminometer.

2. Remove the PMT assembly from the luminometer.

Figure 1-1

Luminometer Assembly

CAUTION

Do not scratch or soil the PMT lens. Scratching or soiling the lens will interfere with the light rays and cause erroneous results.

CAUTION

Ensure that you select Turn System Mechanics Off and cover the PMT lens when exposing the inside of the luminometer to light. Low levels of light can saturate the PMT and it can take several hours for the PMT to recover. High levels of direct light can damage the PMT. 3. Place the lens cover (part number 27127) over the lens of the PMT to protect it from scratches and contamination.

078D0080-02, Rev. A

1-96

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly

4. Disconnect the PMT thermistor cable from connector J6 on the Luminometer Bottom PC board. 5. Disconnect the PMT cable from connector J1 on the Luminometer Bottom PC board. 6. Replace the PMT assembly:
CAUTION

Do not allow silicone oil on the PMT lens. Oil on the lens will Interfere with the light rays and cause erroneous results. a. Remove the lens cover. b. Apply silicone oil (part number 27295) to a lintfree tissue and then apply the oil to the Oring.
CAUTION

Ensure the PMT is seated onto the housing of the luminometer. Incorrect placement of the PMT will cause light leakage and erroneous results. c. Press the PMT assembly into the luminometer so that no light can be seen between the heatsink and the PMT. d. Ensure that the notch on the back of the PMT housing aligns with the driver motor. 7. Secure the PMT assembly to the luminometer using two screws. 8. Connect the PMT thermistor cable to connector J6 on the Luminometer Bottom PC board. 9. Connect the PMT cable to connector J1 on the Luminometer Bottom PC board. 10. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer start button. 11. Verify PMT performance: 12. Check the dark counts without a cuvette: a. At the workspace, select System Status. b. Select Diagnostic Tools. c. At the System-Diagnostic Tools window, select Dark Count without Cuvette.
WARNING

Before selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly

1-97

d. Select Perform. The System-Dark Count window opens. e. Ensure that the RLU 1 values for Total Mean are less than 600. RLUs higher than 600 may indicate a defective PMT, electronic noise, or light leaks. f. Refer to the chart below for the Total SD limit. Match the RLU1 Total Mean with the SD Limit curve. Verify the Total SD value is at or below the SD Limit curve.

13. Check the dark counts with a cuvette: a. At the System-Diagnostic Tools window, select Dark Count with Cuvette.
WARNING

Before selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury. b. Select Perform. The System-Dark Count window opens. c. Ensure that the difference between RLU 1 values for Total Mean with cuvettes and without cuvettes are less than 25. RLUs higher than 25 may indicate light leaks.
NOTE: For both tests (Dark Counts With Cuvettes and Dark Count Without Cuvettes), record and plot the Total Mean and Total SD values. Find the point of intersection between the two plots using the graph above. If the two plots intersect above the curve, there is noise in the electronic circuitry (a defective PMT or a cable picking up interference).

14. If the dark count tests do not meet specifications, refer to the ADVIA Centaur Service Troubleshooting Manual.
NOTE: : Reset test (assay) calibrations to ensure that the system uses the calibration ranges that were defined by the manufacturer until the system calculates new observed ranges.

078D0080-02, Rev. A

1-98

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly

15. Reset test calibrations: a. At the workspace, select Calibration. b. Select Reset Test Calibrations. c. Select Yes. d. Select Continue.

078D0080-02, Rev. A

Preheater 078-B059-01
Preheater Assembly

078D0080-02, Rev. A

1-100

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Pusher Detail

Cuvette Pusher Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-101

Sensor Detail

Motor Detail

078D0080-02, Rev. A

1-102

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail

TED Assembly Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail

1-103

078D0080-02, Rev. A

Parts List Preheater 078-B059-01


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29

Cover, Preheater Channel Screw, Cap-Head M4 x 0.7 x 16 mm Bracket, Tube Connection Cover, Preheater Exit Cover, Cuvette Entrance Screw, Retractable M4 x 0.7 x 8.9 mm PC Board, Reflective Chute Sensor Screw, Cap-Head M4 x 0.7 x 5 mm Rack, Preheater Pusher Screw, Flat-Head #4-40 x 0.25 mm Phillips Plate, Motor Mounting Screw, Cap-Head M4 x 0.7 x 8 mm Gear, Spur 48P 28T 0.187W 0.187B Motor, Stepper 4VDC PC Board, Preheater Thermistor, Threaded Body Thermal Grease TED, Assembly Thermal Grease Heatsink, Preheater Washer, Flat M5 Screw, Cap-Head M5 x 0.8 x 30 mm Screw, Cap-Head M3 x 0.5 x 6 mm Sensor, Optical Switch Bracket, Sensor Screw, Flat-Head M3 x 0.5 x 6 mm Phillips Pusher, Block Screw, Cap-Head M2 x 0.4 x 10 mm Screw, Cap-Head M5 x 0.8 x 16 mm Screw, Cap-Head M4 x 0.7 x 40 mm Screw, Cap-Head M4 x 0.7 x 70 mm

7127594

078-B067-01 60013x37 571858 571855

8966808 9867382 4162038 3062064

571853B 572422 575447B 572416 572643 572415 571841 60013x9 522030901

9864200 5051035 1849083 5483539

572915 575372B 571184C 121055 111677 121055 571856 60056x6 60013x73

7882579 3048878

60013x1 111675 571860 60007x43 571859 572419 60013x38 60013x86 60013x128

078D0080-02, Rev. A

Replacing the Preheater TEDs


1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. c. At the back of the system, turn off the main power. 2. Remove the luminometer. 3. Remove the preheater. 4. Removing the defective TEDs: a. Remove the bottom screw and washer that secure the TED mounting block to the preheater.

Figure 1-1

Preheater Assembly

b. Remove the top screw and the washer that secure the TED mounting block to the preheater.

078D0080-02, Rev. A

1-108

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Preheater TEDs

c. Remove the TEDs.

Figure 1-2

Preheater/TEDs Assembly

5. Install the TEDs:


CAUTION

Ensure that no thermal grease is inside the TED or heating problems will occur. a. Apply a thin layer of thermal grease to both sides of the new TEDs.
CAUTION

Ensure the TEDs are installed flush with the bottom of the preheater or damage can occur to the TEDs.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Preheater TEDs

1-109

b. Install the new set of TEDs on the bottom of the preheater ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. c. Install the TED mounting block and tighten the two screws. Alternate tightening each screw to 0.9 Nm (8 inlb). 6. Install the preheater. 7. Install the luminometer. 8. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 9. Calibrate the preheater thermals.

078D0080-02, Rev. A

Ring Loader 572400B


Ring Loader Assembly

Figure 1-1

Ring Loader Assembly

078D0080-02, Rev. A

1-112

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-113

Motor Detail

078D0080-02, Rev. A

Parts List Ring Loader 572400B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Screw, Flat-Head M5 x 0.8 x 12 mm Arm, Ring Loader Screw, Cap-Head M4 x 0.7 x 8 mm Screw, Retractable M4 x 0.7 x 8.9 mm Plate, Ring Loader Mounting Housing, Waste Probe Spring, Compression 0.240 x 0.022 x 0.250 Pad, Friction Screw, Flat-Head #4-40 x 0.25 Phillips Plate, Mounting, Waste Probe Motor Gear, Spur 32P 24P 0.188W 0.187B Motor, Stepper 4VDC Bipolar Sensor, Optical Switch Screw, Cap-Head M3 x 0.5 x 5 mm Screw, Cap-Head M4 x 0.7 x 8 mm 1116124 9864200 962374

572423 572057B 60013x9 846-0039-10 078-0061-01 571805 571807 571808 572415 571814 571815B 572915 111674 572417 60013x9

078D0080-02, Rev. A

Servo Motor 571730B (Old)


Servo Motor Assembly

078D0080-02, Rev. A

1-118

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Latch Detail

Latch Detail

Servo Motor (Old) Detail 571730B

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Servo Motor (Old) Detail 571730B 1-119

078D0080-02, Rev. A

Servo Motor 571022B (New)


Servo Motor (New) Assembly

078D0080-02, Rev. A

1-122

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Latch Detail

Latch Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Servo Motor (New) Detail 571022B

Servo Motor (New) Detail 571022B

Figure 1-1

Servo Motor Assembly Detail

078D0080-02, Rev. A

Parts List Servo Motor 571730B and 571022B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10

Screw, Cap-Head M5 x 0.8 x 16 mm Screw, Cap-Head M4 x 0.7 x 12 mm Bracket, Servo Motor Plate, Servo Motor Mounting Latch, Compression Motor, Servo (old) PC Board, Servo Motor Pulley, Timing Motor, Servo (new) Adjuster, Belt 4611886 1592244 7617141

60013x38 571753 571725 078-0060-01 571746 571730B 575357C 571748 571022B 571752

078D0080-02, Rev. A

Wash Station 571025D


Wash Station Assembly

Figure 1-1

Wash Station Assembly

078D0080-02, Rev. A

1-128

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Acid Probe Assembly Detail

Acid Probe Assembly Detail (572889)

Figure 1-2

Acid Probe Assembly Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Friction Pad Detail

1-129

Friction Pad Detail

Figure 1-3

Friction Pad Detail

Sensor Detail

Figure 1-4

Wash Station Sensor Detail

078D0080-02, Rev. A

1-130

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

Motor Detail

Figure 1-5

Wash Station Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-131

078D0080-02, Rev. A

Parts List Wash Station 571025D


Item Description REF Part Number

1 2 3

Wash Manifold Tube, Connection, Arm Assembly, Aspirate Probe Connector Plate Screw, Captive Fitting Tubing

01080626 03310289 05061677 03649057 02280653 00123534 06291994 06291994 03040281 02873484 04740554 08373815 07960792 03944946 03548633 04483748 02026323 03944946 03204667 09571793 0189043 962374 06892408

078-0353-01 117116 078-B076-01 571871 571533 518-0891-01 562-3070-03 60013x9 571843 125669 571812B 60013x51 21233 116699B 112405B 571518 112404 572417 116699B 110607C 575362 60013x2 111674 60013x1 572415 571841

4 5 6 7 8 9

Screw, Cap-Head M4 x 0.7 x 8 mm Probe, Wash Aspirate Grounded Probe, Wash Aspirate Clip, Probe Retainer Screw, Cap-Head M4 x 0.7 x 20 mm Clamp, Cable Block, Wash Separation Assembly Guide, Probe (Quantity 4) O-ring (Quantity 8) Retainer, Probe Guide (Quantity 4) Block, Wash Separation

Screw, Cap-Head M3 x 0.5 x 5 mm 08102390 10 11 12 13 14 15 16 17 18 19 20 21 22 Probe, Acid Dispense PC Board, Aspirate Probe Screw, Cap-Head M4 x 0.7 x 6 mm Sensor, Optical Switch0 Screw, Cap-Head M3 x 0.5 x 6 mm Screw, Flat-Head #4-40 x 0.25 Phillips Plate, Motor Mounting, Wash Probe Gear, Spur Motor, Stepper 4VDC Screw, Cap-Head M4 x 0.7 x 5 mm Arm, Wash Probe Cover, Fixed Probe Base, Fixed Probe 3490449 03329850 01116124 09864200

571815B 572915 572416 571212 26286 26285

078D0080-02, Rev. A

134 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Wash Station 571025D

Item

Description

REF

Part Number

23 24 25 26 27 28 29

Spring, Comp. 0.240 x 0.022 x 0.250 Pad, Friction, Long Gasket, Junction Block Retainer, Probe Guide Screw, Cap-Head M3 x 0.5 x 5 mm Housing, Wash Probe Assembly, Wash Tubing Block, Junction Tubing Retainer, Tubing Clamp Fitting Plate, Manifold (Quantity 1) Plate Manifold (Quantity 2) Fitting, (Quantity 14) Screw, Captive (Quantity 3) Screw, M4 x 0.7 x 16 mm (Quantity 2) O-rings, Junction Block (Quantity 20)

01112021 03058539 05150238 02026323 08102390 08242214 05097795 05247469 01296947

571807 571829 078-0275-02 112404 572417 571213 121587 078-0505-01 572689 111972 26699

05271787 03649057 05271787 02280653 02960484 07737546 02960484 04913823 07254626

571826 571871 571827 571826 571533 60013x37 116828 60013x37 078008801 078K11001

30 31

Screw, Cap M4 x 16 mm Guide, Aspirate Probe (single guide) Kit, Guide, Aspirate Probe (4 guides)

078D0080-02, Rev. A

Electronics Module (fixed configuration) 121911D


Electronics Module Assembly

Figure 1-1

Electronics Module Assembly

078D0080-02, Rev. A

1-136 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Backplane PC Board 78-B047-01

Backplane PC Board 78-B047-01 (Fixed Configuration)

Figure 1-2

Backplane PC Board

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 RT SPARC I/O PC Board 575267 1-137

RT SPARC I/O PC Board 575267

Figure 1-3

RT SPARC I/O PC Board

078D0080-02, Rev. A

1-138

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Fixed Electronics Module

Fixed Electronics Module

Figure 1-4

Fixed Electronics Module

078D0080-02, Rev. A

Electronics Module (slide out configuration) 571115B


Electronics Module Assembly

Figure 1-1

Electronics Module (slide out configuration) Assembly

078D0080-02, Rev. A

BackPlane PC Board 575252F (Slide Out)

Figure 1-2

Back Plane PC Board (slide out configuration)

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Slide Out Electronics Module

1-141

Slide Out Electronics Module

Figure 1-3

Slide Out Electronics Module

078D0080-02, Rev. A

Parts List Electronics Module 571115B and 121911D


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13

PC Board, Servo, Low Level Controller Fuse, 2AMP PC Board, Sequence Controller PC Board, Low Level Controller Fuse, 2AMP PC Board, RT SPARC IC, Memory 16MB Dram Simm PC Board, Multi Master Bus PC Board, Backplane (slide out module) Standoff, #4-40 x 0.188 Assembly, PC Photon Counter PC Board, RT SPARC I/O PC Board, Backplane (fixed module) PC Board, Power On/Off Control (fixed module) Power Supply (fixed module) RT SPARC Service Assembly 9136329

S575262C 572808 117154B 078-B042S01 572808 572345C 575191 117149B 575252F 571889 575317 575267 078B04701 117947 119079 078-B060R-01

078D0080-02, Rev. A

Replacing the Real Time SPARC PC Board


1. Power down the system: a. At the Workspace, select the System Status button. b. Select Turn System Off. c. At the back of the system, turn the power off. 2. Pull out the electronics module.
CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components. 3. Remove the Real Time (RT) SPARC PC board from the Backplane PC board. 4. Disassemble the RT SPARC PC board: a. Separate the multiport S-BUS PC board (575207) from the SPARC PC board (572345). b. Remove the six memory SIMMs (575191) plugged into the SPARC PC board (572345). c. Remove the two screws securing the SPARC PC board to its metal bracket. d. Remove the SPARC PC board from the bracket and from the Interface PC board. e. Remove the three screws securing the Interface PC board to the metal bracket and remove the PC board from the bracket. 5. Remove NVRAM the chip from the old board. If the chip is not damaged, save it for installation into the new board. 6. Assemble a new RT SPARC PC board: a. Obtain a new SPARC PC board. b. Mount the multiport S-BUS board on the SPARC PC board. c. Install the six memory SIMMs into the SPARC PC board. d. Install the NVRAM chip into the board. 7. If the old NVRAM chip was damaged and you installed a new chip, place the system ethernet address on the new chip: a. Obtain the evaporating password. b. Turn the system on. c. At the Welcome window, log in as root. d. Enter the evaporating password. e. Move the cursor off the Command Tool window.

078D0080-02, Rev. A

1-146 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Real Time SPARC

f.

While holding down the right glidepoint button, move the cursor to the right arrow at Programs.

g. Select Texteditor and release the right glidepoint button. h. Select File. i. j. Open the etc folder. Double-click the ethers file.

k. You will see a number like 8:0:20:80:5e:2b m0rt. The underlined portion is the NVRAM ID. l. Delete these numbers and enter the new address found at the U0707 on the new Sparc board. Ensure that you do not delete the colons. Verify that you entered the numbers correctly. File changes to Save for 1 second. n. Move the cursor to the top bar of the text editor window, hold down the right glidepoint button and select Quit. o. In the cmdtool window, enter init 5. p. Reboot the system. q. To confirm the RT NVRAM ID, open a new xterm window and enter cat /etc/ ethers. The system will display the address you entered. r. Enter exit. If you entered the NVRAM ID incorrectly, the system will have a 700 00 07 error.

m. Select File for save.

078D0080-02, Rev. A

Replacing the Real Time SPARC PC Board


1. Power down the system: a. At the Workspace, select the System Status button. b. Select Turn System Off. c. At the back of the system, turn the power off. 2. Pull out the electronics module.
CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components. 3. Remove the Real Time (RT) SPARC PC board from the Backplane PC board. 4. Disassemble the RT SPARC PC board: a. Separate the multiport S-BUS PC board (575207) from the SPARC PC board (572345). b. Remove the six memory SIMMs (575191) plugged into the SPARC PC board (572345). c. Remove the two screws securing the SPARC PC board to its metal bracket. d. Remove the SPARC PC board from the bracket and from the Interface PC board. e. Remove the three screws securing the Interface PC board to the metal bracket and remove the PC board from the bracket. 5. Remove NVRAM the chip from the old board. If the chip is not damaged, save it for installation into the new board. 6. Assemble a new RT SPARC PC board: a. Obtain a new SPARC PC board. b. Mount the multiport S-BUS board on the SPARC PC board. c. Install the six memory SIMMs into the SPARC PC board. d. Install the NVRAM chip into the board. 7. If the old NVRAM chip was damaged and you installed a new chip, place the system ethernet address on the new chip: a. Obtain the evaporating password. b. Turn the system on. c. At the Welcome window, log in as root. d. Enter the evaporating password. e. Move the cursor off the Command Tool window.

078D0080-02, Rev. A

1-148 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Real Time SPARC

f.

While holding down the right glidepoint button, move the cursor to the right arrow at Programs.

g. Select Texteditor and release the right glidepoint button. h. Select File. i. j. Open the etc folder. Double-click the ethers file.

k. You will see a number like 8:0:20:80:5e:2b m0rt. The underlined portion is the NVRAM ID. l. Delete these numbers and enter the new address found at the U0707 on the new Sparc board. Ensure that you do not delete the colons. Verify that you entered the numbers correctly. File changes to Save for 1 second. n. Move the cursor to the top bar of the text editor window, hold down the right glidepoint button and select Quit. o. In the cmdtool window, enter init 5. p. Reboot the system. q. To confirm the RT NVRAM ID, open a new xterm window and enter cat /etc/ ethers. The system will display the address you entered. r. Enter exit. If you entered the NVRAM ID incorrectly, the system will have a 700 00 07 error.

m. Select File for save.

078D0080-02, Rev. A

Replacing the User Interface SPARC PC Board (pre FCO 15)


1. Power down the system: a. At the Workspace, select the System Status button. b. Select Turn System Off. c. At the back of the system, turn the power off. 2. Pull out the electronics module. 3. Remove the User Interface (UI) SPARC PC board. a. Remove the two small pan head screws that secure the UI SPARC PC board connector to the backplate. b. Unplug the UI SPARC PC board from the I/O PC board that is secured to the backplane. Do not remove the UI SPARC PC board from the electronics module. c. Remove the three cables from J27, J30, and J32 on the Backplane PC board that are connected to the UI SPARC PC board. d. Partially remove the UI SPARC PC board. e. Disconnect the remaining cable from the UI SPARC PC board. f. Remove the UI SPARC PC board.

CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components. 4. Disassemble the UI SPARC PC board: a. Separate the SBUS to MMBUS PC board (571263) from the SPARC PC board (572345) and from the Interface PC board (575267). 5. Assemble a new UI SPARC PC board: a. Secure the Interface PC board to the metal bracket with three screws. b. Obtain a new SPARC PC board. c. Insert the SPARC PC board into the Interface PC board. d. Secure the SPARC PC board to the metal bracket with two screws. e. Mount the SBUS to MMBUS PC board on the SPARC PC board and the Interface PC board. f. Remove the six memory SIMMs (575191) plugged into the old SPARC PC board. g. Install the six memory SIMMs on the new SPARC PC board. 6. Reconnect the User Interface (UI) SPARC PC board. a. Partially insert the UI SPARC PC board in the electronics module.

078D0080-02, Rev. A

1-150

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the User Interface

b. Reconnect the cable to the UI SPARC PC board that was removed in step 3e. c. Plug the UI SPARC PC board into the I/O PC board that is secured to the backplane. d. Secure the UI SPARC PC board connector to the backplate with two small pan head screws. e. Reconnect the three cables removed from the UI SPARC PC board in step 3c.

078D0080-02, Rev. A

User Interface Module 078-0103-XX for the ADVIA Centaur System


User Interface Module Assembly

078D0080-02, Rev. A

152

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 User Interface Module Assembly

CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.

078D0080-02, Rev. A

Parts List - User Interface Module 078-0103-XX* for the ADVIA Centaur System
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Power Supply, 180W Medical Kit, Hard Disk Drive, 32 GB, SCSI Drive, Diskette Drive, DVD ROM PC Board, Audio Card PC Board, Video Card PC Board, Power Control Fan, Ultra Axi CPU Fan, Enclosure DIMM, Memory, 64 MB Speaker Cable, SCSI Interface, with Terminator Cable, Diskette Drive Cable, Serial Splitter A-modem (B-LIS) Adapter, Serial Port 724929 1758746 4887059 7326511 6811254 9469166 8876019 9211398 5578939 2378602

123422 078-K016-02 515-7173-12 078K18601 123426 078-K157-01 123428 123429 123430 123431 123432 078-0016-01 123434 123435 123014

**Last 2 digits of part number correspond to module version.

078D0080-02, Rev. A

User Interface Module 078-0330-XX for the ADVIA Centaur XP System


User Interface Module Assembly

CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.

078D0080-02, Rev. A

Parts List - User Interface Module 078-0330-XX* for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Power Supply, 180W Medical Kit, Hard Disk Drive, 32 GB, SCSI Drive, Diskette Drive, DVD ROM PC Board, Audio Card PC Board, Video Card PC Board, Power Control Fan, Ultra Axi CPU Fan, Enclosure DIMM, Memory, 64 MB Speaker Cable, SCSI Interface, with Terminator Cable, Diskette Drive Cable, Serial Splitter A-modem (B-LIS) Adapter, Serial Port 724929 1758746 4887059 7326511 9469166 8876019 9211398 5578939

123422 078-K016-02 515-7173-12 078021102 123426 515731602 123428 123429 123430 123431 123432 078-0016-01 123434 123435 123014

**Last 2 digits of part number correspond to software version.

078D0080-02, Rev. A

Replacing the Hard Disk Drive


CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components. Use these procedure to replace the hard disk drive in the User Interface (UI) Module.

Removing the UI Module


1. Turn the system off. a. At the workspace, select System Status. b. Select Turn System Off. c. Select Yes. d. At the back of the system, turn the power switch off. e. Disconnect the power cord from the system. 2. Loosen the two captive screws that secure the UI Module to the support rails. 3. Partially slide the UI Module forward allowing you access to the cable connections. 4. Disconnect the cables from the back of the UI Module. 5. Remove the UI Module from the support rails.

Removing the Hard Disk Drive


1. Remove the cover from the UI Module. 2. Remove the CD-Rom drive. a. Disconnect the two cables from the back of the CD-Rom drive.
NOTE: If you cannot remove the two inside mounting screws on the CD-Rom, remove the diskette drive module. The diskette drive module is secured to the bottom of the UI module with four screws.

b. Remove the four screws the secure the CD-Rom drive.


CAUTION

Do not pull the CD-Rom drive to the inside of the UI module or damage will occur to the EMI shielding strips. c. Push and remove the CD-Rom drive through the front of the UI Module. 3. Remove the SCSI interface cable from the back of the hard disk drive.

078D0080-02, Rev. A

1-160

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive

NOTE: Perform step 4 to remove the SCSI interface cable from the UI module only if the old cable does not include an SCSI terminator.

4. If required, remove the cable for the hard drive and SCSI:
NOTE: The ferrite block is attached to the base of the UI module with double-sided

tape. Remove the ferrite block from the base if you cannot open the ferrite block. a. Remove the ferrite block from the base of the UI module. b. Open the ferrite block surrounding the cable. Save the ferrite block. c. Disconnect the SCSI interface cable from the CPU board, CD-Rom drive, and hard disk drive. d. Remove the cable adapter for the CD-Rom drive. Save the cable adapter and discard the cable. 5. Remove the hard disk drive: a. Disconnect the two cables from the back of the hard disk drive. b. Disconnect the 2-pin connector from the hard disk drive. c. Remove the five screws the secure the mounting bracket and hard drive to the bottom of the UI Module. d. Remove the mounting bracket and hard disk drive from the UI Module.

Figure 1-1

Hard Disk Drive Assembly

e. Remove the four screws the secure the hard drive to the mounting bracket. f. Remove the hard disk drive.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive

1-161

Installing the New Hard Disk Drive


1. Install the Hard disk drive (part number 078-0017-01).
NOTE: Some hard disk drives have a jumper next to the LED OUT pins. You may need to trim the side of the LED connector to ensure that the LED connector attaches to the pins securely. NOTE: Check the label on the hard drive for the LED OUT position.

a. Connect the 2-pin connector to the LED OUT pins on the hard disk drive. b. Secure the hard disk drive to the mounting bracket using four screws. c. Secure the mounting bracket and hard disk drive to the bottom of the UI Module using five screws. d. Connect the power cable to the hard disk drive. e. If you are not replacing the SCSI interface cable, connect the old cable to the hard disk drive and continue with step 3.
NOTE: Perform step 2 to install the SCSI interface cable from the UI module only

if the old cable does not include an SCSI terminator. 2. If required, install the new cable for the hard disk and SCSI: a. Connect the SCSI interface cable (part number 078-0016-01) to the CPU PC Board.
NOTE: The ferrite block is attached to the base of the UI module with

double-sided tape. Remove the ferrite block from the base if is not positioned as shown above.

Figure 1-2

User Interface Drive Component

b. Route the cable through the ferrite block. c. Close the ferrite block. d. Position the ferrite block in the UI module base.
CAUTION

To prevent damage to the connectors, do not twist the ribbon cable. e. Remove the backing from the tape on the bottom of the terminator. f. Secure the terminator to the UI module base close to the hard drive.

078D0080-02, Rev. A

1-162

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive

g. Connect the SCSI interface cable to the back of the hard disk drive.
NOTE: Ensure that you install the CD-Rom drive upside down. This position properly orients the drive on the system. The CD-Rom drive faces the same direction as the diskette drive.

3. Install the CD-Rom drive.


CAUTION

Be careful that you do not damage the EMI shielding strip when you install the CD-Rom drive. The strip reduces EMI emissions from the UI Module. a. Gently push the CD-Rom drive into position through the front of the UI Module. b. Install the CD-Rom drive using the four screws that secure it. c. Connect the power cable to the back of the CD-Rom drive. d. Using the cable adapter removed previously, connect the SCSI interface cable to the back of the CD-Rom drive. 4. Install the cover to the UI Module.

Installing the UI Module


1. Partially slide the UI Module forward onto the two support rails allowing you access to the cable connections. 2. Reconnect and secure the cables to the back of the UI Module. 3. Slide the UI Module to the end of the support rails. 4. Secure the UI Module using the two captive screws. 5. Turn the system on. a. Connect the power cord to the back of the system and then turn the power switch on. b. Turn the system on by pressing the green computer startup button located at the back of the system. 6. Load the correct version of software on the system. For more information, refer to the current version of software installation instructions.

078D0080-02, Rev. A

Application Module 078-0331-XX for the ADVIA Centaur System


Application Module Assembly

CAUTION

Observe precautions for handling electrostaticsensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.

078D0080-02, Rev. A

1-164

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Application Module Assembly

078D0080-02, Rev. A

Parts List Application Module 078-0331-XX* for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 2 3 3 4 5

Application Module Isolation Transformer, ADVIA Centaur System Isolation Transformer, ADVIA Centaur XP System Right Side Panel Assembly, ADVIA Centaur System Right Side Panel Assembly, ADVIA Centaur XP System Centaur XP System Application PC Software Cable, Ethernet, Shielded 01654657 01654657 07580108 08203286 04297030 6203920

078-0331-** 078-0402-01 078-0402-01 122241 078-B087-01 078-S046-01 527-5153-03

**Last 2 digits of part number correspond to software version.

078D00??-01, Rev. A

ADVIA Centaur XP Operators Guide: 1 Parts List Application Module 078-0331-XX* for the ADVIA Centaur XP

078D00??-01, Rev. A

Power Supply
Power supply 571258D (slide out electronics module)

Figure 1-1

Power Supply

078D0080-02, Rev. A

1-168ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Power Supply 078B04701 (fixed

Power Supply 078B04701 (fixed electronics module)

Figure 1-2

Power Supply

078D0080-02, Rev. A

Parts List Power Supply


Item Description REF Part Number

Power Supply (slide out electronics module) Fuse, 10A SB Power Supply (fixed electronics module) Transformer, Isolation (fixed electronics module) AC Input Line Assembly

3628599

571258D 60044x28

2 3 4

8335220 9312976 7820265

078-B047-01 078-B02001 078-B048-01

078D0080-02, Rev. A

Replacing the Power Supply


ELECTRICAL HAZARD

Instrument operation with the power supply cover removed or open presents a potentially lethal electrical hazard. Do not operate the system with the power supply exposed. 1. If the power is on, power down the system: a. At the workspace, select System Status. b. Select Turn System Off. 2. At the back of the system, turn the power off and unplug the power cord. 3. Remove the front cover of the electronics module. 4. Disconnect the AC power cable from the left side of the power supply module. 5. Disconnect the green (top) wire and white (bottom) wire from the DC Filter.

6. Remove the two Phillips head screws that secure the front of the power supply module to the electronic module base.

078D0080-02, Rev. A

1-172

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Power Supply

7. Disconnect the two power supply cables J26 and J32 from the Backplane PC board.

8. Pull the power supply module out of the electronics module. 9. Install the new power supply into the electronics module. 10. Install the two Phillips head screws that secure the front of the power supply module to the electronic module base. 11. Connect the two power supply cables J26 and J32 to the Backplane PC board. 12. Connect the green (top) wire and white (bottom) wire to the DC Filter. 13. Connect the AC power cable to the left side of the power supply module. 14. Install the front cover of the electronics module. 15. Power up the system: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the green computer startup switch.

078D0080-02, Rev. A

Thermal Control PC Module for the ADVIA Centaur System S575257E


Thermal Control PC Module Diagram

F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 F11

Primary Reagent Compartment Loop A Primary Reagent Compartment Loop C Incubation Chamber Loop B Incubation Chamber Loop A Incubation Chamber Loop C Incubation Chamber Loop D Primary Reagent Compartment Loop B Ancillary Queue Preheater PMT1 PMT2

078D0080-02, Rev. A

Parts List Thermal Control PC Module for the ADVIA Centaur System 575257E
Item Description REF Part Number

PC Board, Thermal Control Fuse, 5ASB (F3, F4, F5, F6, F9, F10, F11) Fuse, 6.3ASB (F1, F2, F7, F8) 03841330

575256C 60044x14 575194

078D0080-02, Rev. A

Thermal Control PC Module for the ADVIA Centaur XP System XP 078-B109-XX


Thermal Control PC Module Centaur XP Diagram

F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 F11

Primary Reagent Compartment Loop A Primary Reagent Compartment Loop C Incubation Chamber Loop B Incubation Chamber Loop A Incubation Chamber Loop C Incubation Chamber Loop D Primary Reagent Compartment Loop B Ancillary Queue Preheater PMT1 PMT2

078D0080-02, Rev. A

Parts List Thermal Control PC Module for the ADVIA Centaur XP System 078-B109-XX
Item Description REF Part Number

PC Board, Thermal Control Fuse, 5ASB (F3, F4, F5, F6, F9, F10, F11) Fuse, 6.3ASB (F1, F2, F7, F8)

05229977 03841330

078-B111SXX 60044x14 575194

078D0080-02, Rev. A

Replacing the Thermal Control Module or the PCB


1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. 2. At the back of the system, turn off the power and unplug the power cord. 3. Open the Electronics Module door. 4. Remove all cables from the Module. 5. Remove the Module from the system. If the PCB needs to be replaced, open the Module.

6. Remove the PCB. 7. Install the new Thermal Control Module PCB. 8. Reassemble the Module. 9. Reinstall the Module into the Centaur. 10. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 11. Sign in as an FSE.
NOTE: The following steps must be completed in order to restore the thermal calibration settings.

12. Go to the Thermals Window and select Perform Calibration. 13. At the bottom of the Thermal Calibration window you will see three buttons: First select Restore Calibration When the Restore Calibration button is selected, the system will go to the hard drive and retrieve the thermal calibration values and store them in ram on the Thermal PCB. Now select the Save Calibration button When the Save Calibration button is selected, the calibration values are Flashed to the EEProm on the Temperature Control Module and saved to the hard drive. Thermal Calibrations have now been restored to the Temperature Control Module.

078D0080-02, Rev. A

Reagent Fluidics Module for the ADVIA Centaur System 078-B028-01


Reagent Fluidics Module Assembly

1-184

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 078-B061-01

Diluter 2 mL 078-B061-01

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TriContinent Diluter 078-B061-01 1-185

TriContinent Diluter 078-B061-01

078D0080-02, Rev. A

Parts List Reagent Fluidics Module 078-B028-01


Item Description REF Part Number

1 2* 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

Syringe, Tri-Continent, 2 mL Plunger Plate, Mounting Valve, Pinch, V37 Gauge, Vacuum Indicator Regulator, Vacuum Fitting, Barb, #10-32_5/32 UNF PC Board, Reagent Fluid GIOB Manifold, Reagent 2386559 2637217 9691438 2508468 1939228 6561061 7121359

078-0502-01 112248 115808 112581 19395 571617 116637 078-B027S02 572827 571567B 572886 078-B061-01 7325388 3275688 1926711 7026771 110080 115398 115945 115395 114830D 60030x403 1718884 6601055 4169504 7737546 6971510 6509019 6845434 115633B 114679 566-0911-01 116828 60013x9 26766 26899x6 122166

Valve, Two-Way NC Manifold, 264987 Mounted,V1922, V2831, V47, V63, V64 Transducer, Pressure Diluter, 2 mL, Fitting (two piece insert with barb) Chamber, Fluid O-Ring 0.614ID x 070W Ethylene Propylene Seal, 2.0 mL, 0.500 ID PC Board, Diluter Screw #4-40 x 187 Nut, Chamber Clamp, Tubing Valve, Relief O-ring, Fitting Screw, Cap M4 x 8 mm Tubing, Tygon Connector Tubing, Silicon
*May not be present

1846351

078D0080-02, Rev. A

Reagent Fluidics Module for the ADVIA Centaur XP System 078-B028-01


Reagent Fluidics Module Assembly

078D0080-02, Rev. A

Diluter 2 mL 115401C

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter

1-191

Tri-Continent Diluter

078D0080-02, Rev. A

Parts List Reagent Fluidics Module 078-B028-01


Item Description REF Part Number

1 2* 3 4 5 6 7 8 9 10 11

Syringe, Tri-Continent, 2 mL Plunger Plate, Mounting Valve, Pinch, V37 Gauge, Vacuum Indicator Regulator, Vacuum Fitting, Barb, #10-32_5/32 UNF PC Board, Reagent Fluid GIOB Manifold, Reagent 2386559 2637217 9691438 2508468 1939228 6561061 7121359

078-0502-01 112248 115808 112581 19395 571617 116637 078-B027S02 572827 571567B 572886 115401C

Valve, Two-Way NC Manifold, 264987 Mounted,V1922, V2831, V47, V63, V64 Transducer, Pressure Diluter, 2 mL, Note: The new style diluter is interchangeable with the Tri-Continent diluter. 1846351

12 13 14 15 16 17 18 19 20 21 22 23 24 25

Fitting (two piece insert with barb) Chamber, Fluid O-Ring 0.614ID x 070W Ethylene Propylene Seal, 2.0 mL, 0.500 ID PC Board, Diluter Screw #4-40 x 187 Nut, Chamber Clamp, Tubing Valve, Relief O-ring, Fitting Screw, Cap M4 x 8 mm Tubing, Tygon Connector Tubing, Silicon
*May not be present

7325388 3275688 1926711 7026771

110080 115398 115945 115395 114830D 60030x403

1718884 6601055 4169504 7737546 6971510 6509019 6845434

115633B 114679 566-0911-01 116828 60013x9 26766 26899x6 122166

078D0080-02, Rev. A

System Fluids Drawer 078-B055-01 for the ADVIA Centaur System


System Fluids Drawer Assembly

078D0080-02, Rev. A

196

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly

197

078D0080-02, Rev. A

198

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Wash Manifold

Wash Manifold

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter Detail 115401C (2 mL) 115402C

Diluter Detail 115401C (2 mL) 115402C (5 mL)

Tri-Continent Diluter Detail

078D0080-02, Rev. A

200

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

201

Sensor Detail

078D0080-02, Rev. A

Parts List System Fluids Drawer 078-B055-01 for the ADVIA Centaur System
Item Description REF Part Number

1 2 3

Screw, Flat M3 x 0.5 x 6 mm Tube, Formed Lid Cover Connector Kit Lid, Bulk Bottle Connector, Fluid Bottle 7526731 4594191

60007x43 571957 078-K027-01 571017 571600B 60000x43 111674 60000x63 27265 575277B 110348 113263 112839 571011 572830 078-0273-01 572777 571595 113747 1458x69 110080 572889 110607 26286 27341 21347 26519x6 112581 S575272E 113746 1458x252

4 5 6 7 8 9 10 11 12 13 14 15 16 17

Screw, Pan-Head M3 x 0.5 x 6 mm Sensor, Optical Switch Screw, M3 x 0.5 x 10 mm Valve, 3-Way, V4346, V61 PC Board, Wash Fluid Interconnect Guide, Tube Sensor, Capacitance Stand, Tube Reservoir Gasket, System Fluid Reservoir Connector, System Fluid Reservoir Retainer, Holder Reservoir Drip Tray, Lower Module Pump, Acid, Kit Pump, Acid Fitting Tubing, Acid Kit Probe, Acid Dispense Cover, Fixed Probe Tubing, 3/8 OD x 3/16 ID Tubing, Teflon, 0.035 ID Fitting, Quick Disconnect

18 19 20

Valve, Pinch, V3841 PC Board, Fluid GIOB Pump, Base, Kit Pump, Base

078D0080-02, Rev. A

204

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List System Fluids Drawer

Item

Description

REF

Part Number

Clamp, Wire Tubing, Silicon, 0.194 OD Tubing, LDPE, 0.058 ID Fitting, Y-, Barbed Fitting, Checkvalve Fitting Tubing, Base, Kit Tubing, Base Probe Tubing, 30.4 cm (12 in) Tubing, 190.5 cm (75 in) Tubing, 15.2 cm (6 in) Cover, Fixed Probe Probe, Dispense Fitting, Quick Disconnect 21 22 23 24 25 26 Syringe, Tri-Continent, 5 mL Plunger, 5 mL Syringe, Tri-Continent, 2 mL Plunger, 2 mL Valve, 2-way, normally closed, V70 andV71 Valve, 2-Way, NC Manifold, Mounted,V14, V617, V32, V59 60, V68 69,and V72 76 Fitting, With Insert, Barb, 2-Piece O-ring, Fitting Valve, 3-Way, Manifold Mounted, 4 inch, V48 57 Wash Manifold Valve, 3-Way, V66 67, and V77 78 Clip, Plastic, Snap-In Connector, Tubing Male Fitting (Quantity 4) Plate Screw, Captive 7121359 5224002 264987 7256777

12811x1 26861 26601 26599x6 26860 110080 078-K081-01 12286 116573 572689 572489 26286 110608 110080 078-0503-01 112247 078-0502-01 112248 115870 571567B

27 28 29 30 31 32 33

7325388 7737546 3352291 3096287 7841092

110080 116828 571568 118506 078-K185-01 27135 113390 571826 571827 571533

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List System Fluids Drawer

Item

Description

REF

Part Number

34

Connector, Tubing, Male Fitting (Quantity 7) Screw, Captive Plate

3273316

113919 571826 571533 571871 078-B061-01 078-B062-01

35

Diluter Diluter, 2 mL Diluter, 5 mL

36 barb) 37

Fitting (two piece insert with Chamber, Fluid Chamber, Fluid, 2 mL Chamber, Fluid, 5 mL

7325388

110080

3275688 2409877 1926711 1129145 7026771 1571786

115398 115399 115945 115946 115395 115396 114830D 60030x403

38

O-Ring, Ethylene propylene O-Ring, 2 mL, 0.614ID x 070W O-Ring, 5 mL, 0.864ID x 070W

39

Seal Seal, 2.0 mL, 0.500 ID Seal, 5.0 mL, 0.750 ID

40 41 42 43 44 45 46 47

PC Board, Diluter Screw #4-0 x 187 Nut, Chamber Clip, Valve Plate, Mounting Screw, Cap M4 x 4 mm Not used Guard, Diluter Thumbwheel Handle, System Fluid Door, Diluter Tray, System Fluids Spacer, Lid Slides 03948003 1718884 9654265

115633B 114308 115808 60013x9 078-0485-01 078-0443-01 078-0178-03 571123 078-0322-01 571688

48

Reservoir, Large

08411342

078-0270-01

078D0080-02, Rev. A

System Fluids Drawer 078-081-01 for the ADVIA Centaur XP System


System Fluids Drawer Assembly

078D0080-02, Rev. A

208

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly

209

078D0080-02, Rev. A

210

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly

Wash Manifold

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter Detail 115401C (2 mL) 115402C

Diluter Detail 115401C (2 mL) 115402C (5 mL)

Tri-Continent Diluter Detail

078D0080-02, Rev. A

212

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

213

Sensor Detail

078D0080-02, Rev. A

Parts List System Fluids Drawer 078-B081-01 for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 3

Screw, Flat M3 x 0.5 x 6 mm Tube, Formed Lid Cover Connector Kit Lid, Bulk Bottle Connector, Fluid Bottle 7395645 5559055

60007x43 078-0316-01 078-K192-01 078-0317-01 078-B082-01 60000x43 111674 60000x63 27265 575277B 110348 113263 112839 08411342 07395645 078-0270-01 572830 078-B082-01 572777 078-0321-01 113747 1458x69 110080 572889 110607 26286 27341 21347 26519x6 112581 S575272E 113746 1458x252

4 5 6 7 8 9 10 11 12 13 14 15 16 17

Screw, Pan-Head M3 x 0.5 x 6 mm Sensor, Optical Switch Screw, M3 x 0.5 x 10 mm Valve, 3-Way, V4346, V61 PC Board, Wash Fluid Interconnect Guide, Tube Sensor, Capacitance Stand, Tube Reservoir Gasket, System Fluid Reservoir Connector, System Fluid Reservoir Retainer, Holder Reservoir Drip Tray, Lower Module Pump, Acid, Kit Pump, Acid Fitting Tubing, Acid Kit Probe, Acid Dispense Cover, Fixed Probe Tubing, 3/8 OD x 3/16 ID Tubing, Teflon, 0.035 ID Fitting, Quick Disconnect

18 19 20

Valve, Pinch, V3841 PC Board, Fluid GIOB Pump, Base, Kit Pump, Base

078D0080-02, Rev. A

216

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List System Fluids Drawer

Item

Description

REF

Part Number

Clamp, Wire Tubing, Silicon, 0.194 OD Tubing, LDPE, 0.058 ID Fitting, Y-, Barbed Fitting, Checkvalve Fitting Tubing, Base, Kit Tubing, Base Probe Tubing, 30.4 cm (12 in) Tubing, 190.5 cm (75 in) Tubing, 15.2 cm (6 in) Cover, Fixed Probe Probe, Dispense Fitting, Quick Disconnect 21 22 23 24 25 26 Syringe, Tri-Continent, 5 mL Plunger, 5 mL Syringe, Tri-Continent, 2 mL Plunger, 2 mL Valve, 2-way, normally closed, V70 andV71 Valve, 2-Way, NC Manifold, Mounted,V1 4, V617, V32, V59 60, V68 69,and V72 76 Fitting, With Insert, Barb, 2-Piece O-ring, Fitting Valve, 3-Way, Manifold Mounted, 4 inch, V48 57 Wash Manifold Valve, 3-Way, V66 67, and V77 78 Clip, Plastic, Snap-In Connector, Tubing Male Fitting (Quantity 4) Plate Screw, Captive 7121359 5224002 264987 7256777

12811x1 26861 26601 26599x6 26860 110080 078-K081-01 12286 116573 572689 572489 26286 110608 110080 078-0503-01 112247 078-0502-01 112248 115870 571567B

27 28 29 30 31 32 33

7325388 7737546 3352291 3096287 7841092

110080 116828 571568 118506 078-K185-01 27135 113390 571826 571827 571533

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List System Fluids Drawer

Item

Description

REF

Part Number

34

Connector, Tubing, Male Fitting Screw, Captive Plate

3273316

113919 571826 571533 571871 078-B061-01 078-B062-01

35

Diluter Diluter, 2 mL Diluter, 5 mL

36 37

Fitting (two piece insert with barb) Chamber, Fluid Chamber, Fluid, 2 mL Chamber, Fluid, 5 mL

7325388 3275688 2409877 1926711 1129145 7026771 1571786

110080 115398 115399 115945 115946 115395 115396 114830D 60030x403

38

O-Ring, Ethylene propylene O-Ring, 2 mL, 0.614ID x 070W O-Ring, 5 mL, 0.864ID x 070W

39

Seal Seal, 2.0 mL, 0.500 ID Seal, 5.0 mL, 0.750 ID

40 41 42 43 44 45 46 47

PC Board, Diluter Screw #4-0 x 187 Nut, Chamber Clip, Valve Plate, Mounting Screw, Cap M4 x 4 mm Not used Guard, Diluter Thumbwheel Handle, System Fluid Door, Diluter Tray, System Fluids Spacer, Lid Slides Valve, Ancillary Queue Drain Bracket, 3-Way Valve 01449808 03948003 1718884 9654265

115633B 114308 115808 60013x9 078-0485-01 078-0318-01 078-0319-01 078-0320-01 078-0322-01 078-0323-01 078-0363-02 078-0360-01

078D0080-02, Rev. A

Vacuum Pump 572703B


Vacuum Pump Assembly

078D0080-02, Rev. A

Vacuum Pump 078-B050-02


Vacuum Pump Assembly

078D0080-02, Rev. A

Parts List Vacuum Pump 572703B


Item Description REF Part Number

1 2

Screw, Cap-Head M4 x 0.7 x 30 mm Fan, 24V Connector Lug 2586418

60013x72 23287 572317 111288 7904256 7861530 572500 60013x9 571616D 25042x92 20069 4586245 116636 572065 572712B 572068 572067 60035X5 846-1021-07 558-0867-16 653-0034-01 078-0036-01 078-0035-01 108488 124209 25042x92 572711B 571420 60013x9 556-0383-01 3387753 117121 60044x14 25049x92 60038x5

3 4 5 6 7 8 9

Screw, Button M4 x 0.7 x 6 mm Screw, Cap-Head M4 x 0.7 x 8 mm Pump, Vacuum and Controller Screw, Pan-Head #8-32 x 0.625 Phillips Guard, Fan Fitting Latch, Compression Spacer Spring, Compression Pawl Nut

10 11 12 13 14 15 16 17 18 19 20 21 22

Washer, Flat Standoff Stem, Adjustment Cover, Terminal Block Cover, Vacuum Pump Nut, Hex Mount, Vibration Washer, Lock Plate, Vacuum Mount Not Used Block, Terminal Screw, Cap-Head M4 x 0.7 x 6 mm Grommet Holder, Fuse Fuse, 5A Washer, Lock Nut, Hex

078D0080-02, Rev. A

Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Waste and Water Drawer Assembly

078D0080-02, Rev. A

226

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Reservoir Detail

Water Reservoir Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail

227

Waste Reservoir Detail

078D0080-02, Rev. A

228

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Pressure Pump Detail

Water Pressure Pump Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail

229

Caster Detail

078D0080-02, Rev. A

230

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail

078D0080-02, Rev. A

Parts List Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Item Description REF Part Number

Lid, Water, ADVIA Centaur System Pin Ring, Retaining 3/16 inch Shaft

571128F 571625A 572439 571127F 571625A 572439 5511044 1295053 7940767 3024294 2757433 3842272 575452C 60013x9 549-8072-01 078-0042-01 078-0065-01 590-0042-01 518-0876-01 26970 114306 078-K002-01 571615 19720 6601055 5317728 2341512 962374 114679 114424 121253 60013x9 111674 60023x40 846-0021-11 846-1021-08 8686287 7542990 4150501 572870 572863 078-0026-01 078-0028-01

Lid, Waste Pin Ring, Retaining 3/16 inch Shaft

3 4

PC Board, System Fluid Detect Screw, Cap M4 x 8 SST Trap, Water Assembly Bracket, Water Trap Gasket Tape, Teflon

5 6 7

Fitting, Elbow, 1/4 NPT-3 Flow Filter, Water Strap Kit, Water Pressure Pump Assembly Pump, Water Pressure Tape, Teflon Clamp

Pump, Bellows Bracket, Mounting Screw, Cap M4 x 8 SST

9 10 11 12 13 14 15

Sensor, Optical Switch Screw, Set M4 x 0.7 x 4 mm Hex Screw, Shoulder 5 mm x 12 SST Washer, M5, SST Spring, Comp 0.390 x 0.043 x 0.500 Elbow, 1/4 inch Flow, 1/4 Barb, Shutoff Fitting, Waste Transfer Tube Tube, Waste Transfer

078D0080-02, Rev. A

232ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Waste and Water Drawer

Item

Description

REF

Part Number

16 17 18 19 20 21 22 23 24

Sensor, Optical, Water Reservoir Full, 5.08 cm (2 in) Reservoir, Water Cap, Water Bottle Clip, Optical Sensor Elbow, 1/8 inch Flow, 1/8 Barb, Thru Elbow, 1/4 inch Flow, 1/4 Barb, Thru Coupling, 1/8 inch Flow, 1/8 NPT, Shutoff Reservoir, Waste Cap, Waste Cap Baffle Seal Fitting O-Ring

4935924 3514445 9962296 7471368 4255656

078-B128-01 571101 572596 078-0024-01 571771 571773 571767 078-0025-01

1712924 29058 306833

118359 114218 114630 114303 571766 545-0196-01

25

Waste Bottle Fluid Sensor Operator Kit Sensor, Optical, Waste Bottle, 7.62 cm (3 in) Language Pack, Replacing the Waste Bottle Fluid Sensor

0946549

078-K074-02 078-B029-01 078D0271-01

25

Waste Bottle Fluid Sensor and Cable Kit (Service) Sensor, Optical, Waste Bottle, 7.62 cm (3 in) Cable, sensor Clamp P Screw, S.H.C.S, M4 12, SST

8768658

078-K075-01 078-B029-01 078-B030-01

26 27 28 29 30 31 32

Coupling, 1/4 inch Flow, Thru Bottle, Waste Spigot Bottle, Water Filter, Air Vent Sensor, Optical, Waste Reservoir Low, 10.16 cm (4 in) Screw, Pan-Head M3 x 0.5 x 6 mm 6298441

571772 124365 571618 113850B 116319 078-B130-01 60000x43

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Waste and Water Drawer

Item

Description

REF

Part Number

33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58

Sensor, Optical, Waste Reservoir Full, 7.62 cm (3 in) Elbow, 1/8 inch Flow, 1/8 Barb, Shutoff Coupling, 1/8 inch Flow, 1/8 Barb, Thru Valve, Check Clamp, Tubing Nut, M3 Isolator, Vibration Coupling, 1/4 inch Flow Screw, M4 x 0.7 x 20mm Stainless Socket Plate, Caster Mounting Caster Washer, Flat M4 Screw, Pan M4 x .08 Phillips Kit, Degasser Assembly Clamp, Tubing, 7/16 OD Manifold, Water Reservoir Bracket Slide Clamp, Tubing Fitting, Pressure Pump Output Gasket, Optical Sensor Clip, Water Trap Screw, Cap M4 x 8 SST Drawer, Waste Water Fitting, Waste Reservoir Screw, Pan M4 x 6 Wall, Back Screw, Cap M4 x 14 Hex. Clamp Assembly, Sensor Cable

8134047

078-B129-01 114591 115022 116151 114679 60024x4 117266 112075 60055x95 115486B 115487 60028x5 60000x55 078-K049-02 19947x7 117115 571621C 571689 114679 518-7075-01 571937 509-0331-01 60013x9 571596 518-0818-05 60003x45 571129D 60013x30 23491 078-B030-01

078D0080-02, Rev. A

Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Waste and Water Drawer Assembly

078D0080-02, Rev. A

236

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail

Waste Reservoir Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail

237

Waste Reservoir Detail

078D0080-02, Rev. A

238

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Pressure Pump Detail

Water Pressure Pump Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail

239

Caster Detail

078D0080-02, Rev. A

240

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail

078D0080-02, Rev. A

Parts List Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Item Description REF Part Number

1 2

Lid, Water Lid, Waste Pin Ring, Retaining 3/16 inch Shaft

078-0358-01 078-0359-01 571625A 572439 5511044 1295053 0831660 7940767 3024294 2757433 0803047 3842272 575452C 60013x9 549-8072-01 078-0042-02 078-0065-01 590-0042-01 518-0876-01 26970 114306 571714B 078-K002-01 571615 19720 6601055 5317728 2341512 962374 114679 114424 121253 60013x9 111674 60023x40 846-0021-11 846-1021-08 8686287 7542990 4150501 4935924 572870 572863 078-0026-01 078-0028-01 078-B128-01

3 4

PC Board, System Fluid Detect Screw, Cap M4 x 8 SST Trap, Water Assembly Bracket, Water Trap Gasket Tape, Teflon

5 6

Fitting, Elbow, 1/4 NPT-3 Flow Filter, Water Strap Clip

Kit, Water Pressure Pump Assembly Pump, Water Pressure Tape, Teflon Clamp

Pump, Bellows Bracket, Mounting Screw, Cap M4 x 8 SST

9 10 11 12 13 14 15 16

Sensor, Optical Switch Screw, Set M4 x 0.7 x 4 mm Hex Screw, Shoulder 5 mm x 12 SST Washer, M5, SST Spring, Comp 0.390 x 0.043 x 0.500 Elbow, 1/4 inch Flow, 1/4 Barb, Shutoff Fitting, Waste Transfer Tube Tube, Waste Transfer Sensor, Optical, Water Reservoir Full, 5.08 cm (2 in)

078D0080-02, Rev. A

242ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Waste and Water Drawer

Item

Description

REF

Part Number

17 18 19 20 21 22 23 24

Reservoir, Water Cap, Water Bottle Clip, Optical Sensor Elbow, 1/8 inch Flow, 1/8 Barb, Thru Elbow, 1/4 inch Flow, 1/4 Barb, Thru Coupling, 1/8 inch Flow, 1/8 NPT, Shutoff Reservoir, Waste Cap, Waste Cap Baffle Seal Fitting O-Ring

3514445 9962296 7471368 4255656 04329595 1712924 29058 306833

571101 572596 078-0024-01 571771 571773 571767 078-B090-01 118359 114218 114630 114303 571766 545-0196-01

25

Waste Bottle Fluid Sensor Operator Kit Sensor, Optical, Waste Bottle, 7.62 cm (3 in) Language Pack, Replacing the Waste Bottle Fluid Sensor

0946549

078-K074-02 078-B029-01 078D0271-01

25

Waste Bottle Fluid Sensor and Cable Kit (Service) Sensor, Optical, Waste Bottle, 7.62 cm (3 in) Cable, sensor Clamp P Screw, S.H.C.S, M4 12, SST

8768658

078-K075-01 078-B029-01 078-B030-01

26 27 28 29 30 31 32 33

Coupling, 1/4 inch Flow, Thru Bottle, Waste Spigot Bottle, Water Filter, Air Vent Sensor, Optical, Waste Reservoir Low, 10.16 cm (4 in) Screw, Pan-Head M3 x 0.5 x 6 mm Sensor, Optical, Waste Reservoir Full, 7.62 cm (3 in) 8134047 6298441 00004926

571772 078-0166-01 571618 078-B046-01 116319 078-B130-01 60000x43 078-B129-01

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Waste and Water Drawer

Item

Description

REF

Part Number

34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61

Elbow, 1/8 inch Flow, 1/8 Barb, Shutoff Coupling, 1/8 inch Flow, 1/8 Barb, Thru Valve, Check Clamp, Tubing Nut, M3 Isolator, Vibration Coupling, 1/4 inch Flow Screw, M4 x 0.7 x 20mm Stainless Socket Plate, Caster Mounting Caster Washer, Flat M4 Screw, Pan M4 x .08 Phillips Kit, Degasser Assembly Clamp, Tubing, 7/16 OD Manifold, Water Reservoir Bracket Slide Clamp, Tubing Fitting, Pressure Pump Output Gasket, Optical Sensor Clip, Water Trap Screw, Cap M4 x 8 SST Drawer, Waste Water Fitting, Waste Reservoir Screw, Pan M4 x 6 Wall, Back, ADVIA Clamp Assembly, Sensor Cable Valve, Monthly Cleaning Diaphragm, Water Pump Kit, Water Pump Poppet Valve 04457976 00732506 1338682

114591 115022 116151 114679 60024x4 117266 112075 60055x95 115486B 115487 60028x5 60000x55 078-K049-02 19947x7 117115 571621C 571689 114679 518-7075-01 571937 509-0331-01 60013x9 078-0356-01 518-0818-05 60003x45 078-0356-01 23491 078-B030-01 078-B014-02 549-B073-01 566-1237-01

078D0080-02, Rev. A

Primary Reagent Compartment 571111C


Primary Reagent Compartment Assembly

Figure 1-1

Reagent Compartment Assembly

078D0080-02, Rev. A

1-246

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-2

Reagent Compartment Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-3

Reagent Compartment Assembly

078D0080-02, Rev. A

1-248

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-4

Reagent Compartment Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-5

Reagent Compartment Assembly

078D0080-02, Rev. A

1-250

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-6

Reagent Compartment Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lead Screw Detail

1-251

Lead Screw Detail

Figure 1-7

Reagent Compartment Lead Screw Detail

078D0080-02, Rev. A

1-252

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

Sensor Detail

Figure 1-8

Reagent Compartment Sensor Detail

Motor Detail

Figure 1-9

Reagent Compartment Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Detail

1-253

TED Detail

Figure 1-10

Reagent Compartment TED Detail

078D0080-02, Rev. A

Parts List Primary Reagent Compartment 571111C


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

Screw, Button M4 x 0.7 x 12 mm Socket Holder, Rinse Station 2 & 3 PC Board, Mixer Interconnect Screw, Cap-Head M4 x0.7 x 05 mm Holder, Rinse Station 1 Shutter, Reagent Probe Screw, Cap-Head M4 x 0.7 x 16 mm Washer, Flat M4 Stnl Cover, Reagent Box, ADVIA Centaur System Cover, Reagent Box, ADVIA Centaur XP System Assembly, Crossflow Blower, ADVIA Centaur System Screw, Cap-Head M4 x 0.7 x 10 mm Motor, Stepper 4VDC Screw, Flat-Head #4-40 x .38 Phillips Sensor, Optical Switch Screw, Cap-Head M3 x 0.5 x 06 mm Plate, Interrupter Screw, Set M4 x 0.7 x 4 mm Cam Plate Adapter, Probe Shutter Screw, Flat-Head M4 x 0.7 x 10 mm Socket Screw, Flat-Head M4 x 0.7 x 12 mm Socket Plate, Motor Mounting BC Shutter Screw, Flat-Head M3 x 0.5 x 10 mm Phillips Plate, Top Reagent Tray Plate, Mixer Slide, Reagent Pack (1-30) Slide #1 Slide #2 0009995 7713183 7882579 9864200 3360812 2825080 8653699 8523078 2841884 7483250

109903 572657B 575422C 572416 572656B 571785 60013x37 60056x5 571636C 078-0377-01 572934 60013x16 572915 60031x406 111671 60013x1 572461 60022x39 572450 572472 60055x65 571781 572451 60007x63 571506C 571482C 124356 571892B 571893B

078D0080-02, Rev. A

1-256

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Primary Reagent

Item

Description

REF

Part Number

Slide #3 Slide #4 Slide #5 Slide #6 Slide #7 Slide #8 Slide #9 Slide #10 Slide #11 Slide #12 Slide #13 Slide #14 Slide #15 Slide #16 Slide #17 Slide #18 Slide #19 Slide #20 Slide #21 Slide #22 Slide #23 Slide #24 Slide #25 Slide #26 Slide #27 Slide #28 Slide #29 Slide #30 27 28 29 30 Shaft, Reagent Tray Retainer, Shaft, Reagent Tray Grease, Lithium Screw, Cap-Head M4 x 0.7 x 12 mm Screw, Flat-Head M4 x 0.7 x 16 mm Socket

571894B 571895B 571896B 571897B 571898B 571899B 571900B 571901B 571902B 571903B 571904B 571905B 571906B 571907B 571908B 571909B 571910B 571911B 571912B 571913B 571914B 571915B 571916B 571917B 571918B 571919B 571920B 571921B 571495B 571847B 572662 60013x23 60055x85

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Primary Reagent

Item

Description

REF

Part Number

31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Connector, Drive, Reagent Tray Screw, Cap-Head M4 x 0.7 x 14 mm Screw, Cap-Head M4 x 0.7 x 22 mm Washer, Shoulder, #8 Nylon Spacer, TED, Reagent Box Thermistor, Threaded Body Thermal Grease Coupling, Flexible 0.25 x 0.75 x 1.025 Motor Mount, Reagent Drive Screw, Cap-Head M4 x 0.7 x 8 mm Motor, Stepper 6VDC Sensor, Slotted Clamp, Split-Hub 0.251D x 0.187 Spacer, 0.250 x 0.375 x 0.063 Stainless Bearing, Ball Flanged 0.250 x 0.500 x 0.188 Support, Leadscrew Assembly Ring, Retaining 0.168 x 0.300 x 0.15 Bearing, Journal 0.19 x 0.315 x0.375-Delrin Leadscrew, Assembly Shaft, Drive Reagent Mixer Flag, Reagent Drive Shutter, Bar Code Retainer, Bearing TED, Assembly Thermal Grease Insulation Insulation Kit, Reagent Door Seal Bracket, Reagent Drive Guide, Reagent Shaft Washer, Lock Nut, Hex 5518057 8570068 5518057 5266031 4308474 2124384 962374 1849083 5286512

571501 60013x30 60013x58 572106 571498B 571184C 121055 571509B 571488B 60013x9 572913 111674 572228 26538 26541 572449 571474 571473 572661B 571472 571475 571786C 571480 112586 121055 572232 572477 572460B 078-0097-01 078-0098-01 25042x92 108488

078D0080-02, Rev. A

1-258

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Primary Reagent

Item

Description

REF

Part Number

61 62 63 64

Mount, Isolator Magnet Screw, Shoulder Seal, Gasket 9408892 2793456

124209 571991 572237 572235

078D0080-02, Rev. A

Replacing the Primary Reagent Compartment Leadscrew


1. Remove the primary reagent packs and store them in a refrigerator. 2. At the workspace, select System Status. 3. Select Turn Mechanics Off, and then wait while the system mechanics power down. 4. Remove the three screws that secure the back left cover to the frame. 5. Remove the back left cover.

Figure 1-1

Primary Reagent Compartment ADVIA Centaur System

078D0080-02, Rev. A

1-260

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

6. Access the leadscrew assembly through the back of the system.

Figure 1-2

ADVIA Centaur System Back View

7. Move the reagent probe 1 and reagent barcode scanner to the right allowing access to the back left corner of the primary reagent compartment.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

8. Remove the reagent mixer drive shaft from the leadscrew housing.

Figure 1-3

Primary Reagent Compartment Mixer Assembly

078D0080-02, Rev. A

1-262

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

9. Using a 7/64 inch ball driver, loosen the two set screws that secure the mixer leadscrew to the mixer motor coupling.

Figure 1-4

Reagent Compartment Mixer Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

10. Using a 3-mm hex driver, remove the two screws that secure the leadscrew mounting block to the back of the primary reagent compartment. Save the screws.

Figure 1-5

Primary Reagent Compartment Mixer Assembly

078D0080-02, Rev. A

1-264

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

11. Remove the leadscrew assembly from the primary reagent compartment.

Figure 1-6

Primary Reagent Compartment Mixer Assembly

12. Remove the clamp, the spacer, the leadscrew mounting block, and the reagent drive flag from the leadscrew assembly and save the parts.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

13. Discard the leadscrew housing and the leadscrew.

Figure 1-7

Primary Reagent Compartment Leadscrew Assembly

078D0080-02, Rev. A

1-266

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

14. Install the clamp, the spacer, the leadscrew mounting block, and the reagent drive flag on the new leadscrew assembly.

Figure 1-8

Primary Reagent Compartment Leadscrew Assembly

15. Press up on the bottom of the leadscrew until the top edge of the leadscrew is flush against the bearing.
CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. 16. Press down and tighten the screw on the clamp. 17. Apply NSK Grease No. 1 to the middle of the leadscrew and leadscrew mounting block arm. 18. Move the probe mechanism manually until a thin layer of grease spreads to the entire leadscrew. Remove any excess grease.
NOTE: Do not tighten the set screws on the motor mixer coupling at this time.

19. Place the leadscrew assembly in the motor mixer coupling that is in the primary reagent compartment.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

20. Ensure that the motor mixer coupling moves freely on the shaft. 21. Using a 3-mm hex driver, install the two screws that secure the leadscrew mounting block to the back of the primary reagent compartment. 22. Use a small flatblade screwdriver to lift the bottom of the rubber coupling from the mixer motor, and then insert an edge of a standard piece of paper (the width is approximately 0.102 mm [0.004 in]). Remove the screwdriver.

Figure 1-9

Primary Reagent Compartment Leadscrew Assembly

23. Push down on the coupling and the sheet of paper. 24. Ensure that the upper and lower set screws on the rubber coupling are aligned, and then use a 7/64 inch ball driver to tighten the bottom set screw. 25. Tighten the top set screw, and then remove the sheet of paper.
CAUTION

Ensure that the rubber coupling moves freely on the motor shaft or system damage can occur.

CAUTION

Ensure that the reagent drive flag does not hit the sensors or system damage can occur. 26. Manually move the leadscrew:

078D0080-02, Rev. A

1-268

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

a. Ensure that the bottom of the rubber coupling is not rubbing against the motor. b. Ensure that the set screws on the clamps are not rubbing against the mounting plate. c. Ensure that the reagent drive flag is not hitting the mixer movement sensors. 27. Install the reagent mixer drive shaft: a. Tilt the reagent mixer to the back of the primary reagent compartment. b. Rotate the leadscrew and lower the leadscrew housing to the bottom of travel.
CAUTION

Do not apply Loctite to the new reagent mixer drive shaft. If you apply Loctite, the leadscrew housing can crack.

CAUTION

Do not force the new reagent mixer drive shaft into the leadscrew housing. If you force the drive shaft into the leadscrew housing, you can damage the threads. c. Install the new reagent mixer drive shaft into the leadscrew housing. If the drive shaft is not going into the leadscrew housing smoothly, remove the drive shaft and install it again. Ensure that the drive shaft engages the reagent mixer plate.
NOTE: Ensure that the rinse station tubing does not interfere with the movement of

the leadscrew. 28. Install the back left cover to the frame. 29. Select Turn System Mechanics On. 30. Calibrate the reagent mixer. 31. Verify that the reagent probes bottom calibrations are correct. 32. Install the primary reagent packs.

078D0080-02, Rev. A

Replacing the Primary Reagent Compartment TEDs


1. Turn off the system: a. At the workspace, select System Status. b. Select Turn System Off. c. At the back of the system, turn off the main power and unplug the power cord. d. Remove the primary reagent packs from the primary reagent compartment and store them in a refrigerator. 2. Remove the three screws that secure the back left cover to the frame. 3. Remove the back left cover.

Figure 1-1

Primary Reagent Compartment

078D0080-02, Rev. A

1-270

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

4. Remove the primary reagent compartment cables:

Figure 1-2

Primary Reagent Compartment Assembly

NOTE: Label all cable connections before removing the cables.

a. Disconnect the sensor motor interconnect cable (P10) and the thermistors cable (P14) from the primary reagent compartment PC board. b. Disconnect the fan cable. c. Disconnect the LED array upper cable (P19) and the LED array lower cable (P22) from the Reagent GIOB PC board. d. Move the cables away from the primary reagent compartment. 5. Remove the five screws and washers that secure the top cover and reagent probe shutter to the primary reagent compartment. 6. Remove the primary reagent compartment:

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

a. Remove the two rinse stations. b. Remove the four screws that secure the primary reagent compartment to the frame. c. Move the reagent compartment out and away from the Incubation Chamber approximately 152 mm (6 in). d. Remove the tubing from the two drain ports located at the bottom left and right side of the primary reagent compartment. e. Disconnect the TED power cable from the connector. f. Remove the primary reagent compartment and place it on a flat surface.

7. Remove the two screws that secure the TED wire cover to the primary reagent compartment. 8. Pull the TED wire cover away from the reagent compartment. 9. Remove the TED connector from the cover. 10. Remove the top cover and reagent probe shutter.

078D0080-02, Rev. A

1-272

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

11. Loosen the two screws and retainers that secure the reagent mixer to the mixer bearing.

Figure 1-3

Primary Reagent Compartment Assembly

12. Remove the reagent mixer. 13. Remove the two screws that secure the coldsink cover to the coldsink. 14. Remove the coldsink cover. 15. Remove the two screws that secure the fan to the reagent compartment. 16. Remove the fan.
CAUTION

Be careful not to kink the thermistor wires or damage can occur to the wires. 17. Disconnect the three thermistor cables from the primary reagent compartment PC board and route the cables through the left side of the reagent compartment.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

18. Remove the bottom plate to the primary reagent compartment: a. Remove the 11 screws and washers that secure the bottom plate to the reagent compartment.

Figure 1-4

Primary Reagent Compartment TEDS Assembly

CAUTION

Do not pry the bottom plate from the reagent compartment or damage will occur to the reagent compartment. b. Simultaneously pull and twist the mixer bearing until you break the tight seal between the bottom plate and the reagent compartment. c. Guide the TED wires through the reagent compartment and remove the bottom plate. 19. Remove the TED heatsinks and insulation: a. Remove the 12 screws that secure the six TED spacer blocks to the bottom plate. b. Remove the two pieces of insulation surrounding the TED mounting blocks.

078D0080-02, Rev. A

1-274

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

20. Remove the TEDs:

Figure 1-5

Primary Reagent Compartment TEDS Assembly

a. Remove the two screws that secure each TED mounting block to the heatsink. b. Remove the six TEDs. c. Clean the thermal grease from the heatsinks, TED mounting blocks, and bottom plate. 21. Install the TEDs:
CAUTION

Apply only enough thermal grease to fill any gaps or holes on the metal surface of the TED mounting blocks. Excess thermal grease acts as a insulator and reduces the performance. Use only 40 to 60 L of thermal grease. Thermal grease on the edges of the TEDs can cause cooling problems.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

a. Apply a thin layer (20 30 L) of thermal grease to the TED side of each TED spacer block. b. Apply a thin layer (20 30 L) of thermal grease to the two TED areas of each heatsink. c. Remove any excess thermal grease from the TED spacer blocks and heatsinks with a straight edge.
CAUTION

Ensure the TEDs are installed flush with the bottom of the heatsink or damage can occur to the TEDs. d. Install the TEDs on the bottom of each heatsink ensuring the red (R) and black (B) wires are in the same position as the old TEDs.

Figure 1-6

Primary Reagent Compartment Connector Back View

CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. e. Install the TED spacer blocks and torque the two screws. Alternate tightening each screw, a little at a time to 0.68 Nm (6 inlb) on the TED spacer block. f. Wait three minutes and then tighten each screw to 0.68 Nm (6 inlb).

078D0080-02, Rev. A

1-276

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

22. Install the heatsinks: a. Install the two pieces of insulation over the TED mounting blocks. b. Apply a thin layer of thermal grease to each TED spacer block. c. Apply a thin layer of thermal grease to each TED spacer area of the bottom plate. d. Remove any excess thermal grease from the blocks with a straight edge. e. Place the bottom plate over the TED mounting blocks.
CAUTION

Ensure the TED mounting blocks are installed flush with the bottom plate or cooling problems will occur. f. Install the 12 screws that secure the six TED spacer blocks to the bottom plate. Alternate tightening each screw, a little at a time to 1.7 Nm (15 inlb) on the TED spacer block.

g. Wait three minutes and then tighten the 12 screws to 1.7 Nm (15 inlb). 23. Assemble the reagent compartment: a. Guide the TED wires through the reagent compartment. b. Pull the TED wires away from the reagent compartment. c. Ensure the TED wires are routed inside the reagent compartment and install the bottom plate into the reagent compartment. Tighten the 11 screws to 0.68 Nm (6 inlb) on the TED spacer block. d. Install the TED connector into the TED wire cover. e. Fold the TED wires inside the cover and secure the TED wire cover to the reagent compartment. f. Route the thermistor cables through the left side of the reagent compartment and connect them to the primary reagent compartment PC board.

g. Install the fan into the primary reagent compartment. h. Install the coldsink cover onto the cold sink. i. j. l. Clean the grease from the mixer bearing. Apply a thin layer of lithium grease (part number 572662) on the mixer bearing. Install the top cover and the probe shutter to the primary reagent compartment.

k. Install the primary reagent mixer to the mixer bearing.

24. Install the primary reagent compartment: a. Place the primary reagent compartment on the frame leaving approximately 152 mm (6 in) behind the primary reagent compartment. b. Install the two rinse stations. c. Install the tubing from the two drain ports. d. Connect the TED power cable from the connector.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent

e. Install the four screws that secure the primary reagent compartment to the frame. f. Connect the sensor motor interconnect cable, thermistor cable, fan cable, LED upper cable, and LED lower cable.

g. Install the back left cover. 25. Turn the system on: a. Plug in the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 26. Calibrate the reagent probe shutter. 27. Calibrate the reagent mixer. 28. Calibrate the reagent probes. 29. Calibrate the primary reagent compartment thermals.

078D0080-02, Rev. A

Primary Reagent Compartment Door 571104B


Primary Reagent Compartment Door Assembly

Figure 1-1

Primary Reagent Compartment Door Assembly

078D0080-02, Rev. A

1-280

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment

Figure 1-2

Primary Reagent Compartment Door Assembly

078D0080-02, Rev. A

Parts List Primary Reagent Compartment Door 571104B


Item Description REF Part Number

1 2 2 3 4 5 5 6 7 8 9 10 10 11 12 13

Handle, Reagent Door Door, Reagent Hinged, ADVIA Centaur System Door, Reagent Hinged, ADVIA Centaur XP System Screw, Cap-Head M4 x 0.7 x 20 mm Cover, LED Display Plate, Reagent Door Closer, ADVIA Centaur System Plate, Reagent Door Closer, ADVIA Centaur XP System Screw, Button M4 x 0.7 x 6 mm Screw, Cap-Head M4 x 0.7 x 6 mm Hex Switch, Tilt Bracket, Tilt Switch Panel, Rear Reagent Door, ADVIA Centaur System Panel, Rear Reagent Door, ADVIA Centaur XP System Screw, Button-Head M4 x 0.7 x 12 mm Seal, Hinge Housing Display, LED PC Board, Reagent Pack LED Array Panel, LED Display Screw, Cap-Head M3 x 0.5 x 6 mm 5255676 7835548

572457 571564B 078-0380-01 60012x51 571735 572256B 078-0379-01 572500 60013x2 26770 078-0029-01 572253 078-0378-01 572499 572480B 571955B 575387 571943 60012x1 572456 572479 60055x85 120198 12811x1

14 15 16 17 18

Fastener, Loop Fastener, Hook Screw, Flat-Head M3 x 0.5 x 16 mm Thermometer Clamp

078D0080-02, Rev. A

Reagent Barcode Scanner 571103


Reagent Barcode Scanner Assembly

078D0080-02, Rev. A

1-284

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Clamp Detail

Belt Clamp Detail

Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-285

Motor Detail

078D0080-02, Rev. A

1-286

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

078D0080-02, Rev. A

Parts List Reagent Barcode Scanner 571103C


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11

Motor, Stepper 4VDC Plate, Barcode Idler Scanner, Barcode Head Bracket, Barcode Head Screw, Cap-Head 6/32 x 0.31 Slotted Steel Screw, Flat-Head M3 x 0.5 x 8 mm Screw, Cap-Head M3 x 0.5 x 8 mm Screw, Cap-Head M3 x 0.5 x 6 mm Clamp, Belt Screw, Shoulder #10-0.25 x 0.375 x 0.125 Pulley, Idler Pulley Bearing (Quantity 2)

9864200 6912824

572915 078-0252-01 571259E 078-0085-01 800-0178-01

300150

60006x43 60012x8 60012x155 571790 572686

7316923 6335851 7713183

114473 571683 23254 111671 60012x1 571797 60028x5 60014x79 078-0084-01 571791 571792 572502 110989

12 13 14 15 16 17 18 19 20 21 22

Sensor, Optical Switch Screw, Cap-Head M3 x 0.5 x 6 mm Plate, Motor Mount Washer, Flat M4 Zinc Screw, Cap-Head M4 x 0.7 x 35 mm Guide, Linear Barcode Assembly Belt, Timing 3 mm Pulley, Timing Belt Screw, Pan-Head #4-40 x 0.438 Phillips Guide, Chain Screw, Thumb

9867382

572422

078D0080-02, Rev. A

Reagent Probe Guide 571110B


Reagent Probe Guide Assembly

078D0080-02, Rev. A

1-290

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe Guide Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-291

Motor Detail

078D0080-02, Rev. A

Parts List Reagent Probe Guide 571110B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Screw, Cap-Head M4 x 0.7 x 12 mm Motor, Stepper 6VDC Pulley, Timing Chain Guide, Probe 1, Upper Chain Guide, Probe 3, Upper Chain Guide, Probe 2, Upper Screw, Cap-Head M4 x 0.7 x 18 mm Not Used Not Used Not Used Pulley, Idler Not Used Not Used Guide, Linear PC Board, Reagent Probe, Pack-to-Pack Plate, Belt Mounting Screw, Flat-Head M3 x 0.5 x 10 mm Phillips Screw, Cap-Head M3 x 0.5 x 8 mm Plate, Adapter Screw, Cap-Head M4 x 0.7 x 6 mm Guide, Linear Plate, Adapter, Probe 2 Belt, 0.200P x 0.375W 2387431 8274116 5175982 7102842 2124384 6398047 2511825 9804062 1056989

60012x23 572913 571588 110979 110982 110981 60006x43 60012x44

Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150

078-B006-01

571574 575347 119028 60006x63 60012x8 571575 60012x2 571573 571589 571576

078D0080-02, Rev. A

Reagent Probe 1 571107C


Reagent Probe 1 Assembly

078D0080-02, Rev. A

1-296

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 1 Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 1 Tubing Routing 1-297

Reagent Probe 1 Tubing Routing

Motor Detail

Sensor Detail

078D0080-02, Rev. A

1-298

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-299

078D0080-02, Rev. A

Parts List Reagent Probe 1 571107C


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27

Screw, Pan-Head M3 x 0.5 x 6 mm Phillips Spacer PC Board, Probe 1 Horizontal Interconnect Screw, Cap-Head M4 x 0.7 x 12 mm Motor, Stepper 4VDC Plate, Motor Mounting Screw, Flat-Head #4-40 Sensor, Optical Switch PC Board, Reagent Probe 1 Heater Control and Volume Detect Screw, Pan-Head #4-40 x 0.438 Phillips Gear, Spur 24P 16T 0.250W 0.187B Pulley, 40DP 10T Screw, Special Hex Spacer, 0.125 x 0.187 x 0.010 Stainless Bearing, Ball Flanged 0.124 x 0.312 x 0.1406 Not Used Not Used Shaft, Guide Leadscrew & Follower Guide, Block Follower Probe, Reagent Sleeve, Probe Bracket, Cable Carrier Grommet Strip Block, Cable Bracket Mounting Guide, Linear Reagent Probe Washer, Lock M4 Zinc (only use with item 29 on older systems with the reagent probe assembly mounted with two screws) Washer, Flat M4 Zinc 1873480 5483601 1096417 4042237 4692398 7817930 5566949 9841251 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 6300918 6362506 9489841

60000x43 26626 575432B 60012x23 572905 571224 60031x406 112585 60006x43 575402B 572502 571226 571230 571236 26625 571222

571219 571218 571217 571233 078-0087-02 571239B 12298 571240 571214B 60026x5

28

60028x5

078D0080-02, Rev. A

1-302

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Reagent Probe 1

Item

Description

REF

Part Number

29

See note below: NOTE: Screw, Cap-Head M4 x 0.7 x 5 mm (only use on older systems with the reagent probe assembly mounted with two screws) NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1) 60012x2

572417

30 31 32 33 34 35 36 37 38 39

Screw, Pan-Head M3 x 0.5 x 8 mm Phillips Belt, 40DP 0.188W Pulley, 40DP 36T 0.188W 0.188B Plate, R1 Pulley Guard Flags-Sensor, Side/Side Probe 1 Rack, 24DP Mount, Rail and Rack Bracket, Upper Chain Guide, Chain, Probe 1 Lower Kit, Heater Coil w/ Bubble Detector Sensor, Volume Detector Insulator Gasket, Volume Detection Loctite, Adhesive Bracket, Reagent Bubble Detector 9051595 2368984 6034835 4806784

60000x53 571223 26083 078-0063-01 078-0019-01 571215B 571216 571231B 110983 078-K015-03 113692 571238 116314 18006 078-0051-01 115269 518-0877-01 9991636 615801 562-3070-01 25029x85

40 41 42

Casting, Machined Probe 1 Carriage Fitting, Barbed Tubing, Silicon Fastener, Ty-wrap NOTE: For replacement of the reagent probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.

078D0080-02, Rev. A

Reagent Probe 2 571108C


Reagent Probe 2 Assembly

078D0080-02, Rev. A

1-304 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 2 Tubing Routing

Reagent Probe 2 Tubing Routing

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-305

Motor Detail

Sensor Detail

078D0080-02, Rev. A

1-306

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-307

078D0080-02, Rev. A

Parts List Reagent Probe 2 571108C


Item Description REF Part Number

1 2

Screw, Cap-Head M4 x 0.7 x 6 mm Washer, Lock M4 (only use with item 15 on older systems with the reagent probe assembly mounted with two screws) Washer, Flat M4 Bracket, Cable Carrier, Probe 2 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 Guide, Chain, Probe 2 Lower Screw, Pan-Head M3 x 0.5 x 6 mm Guide, Linear Bracket, Upper Chain Mount, Rail & Rack Clamp, Cable Loctite, Adhesive Rack 24 DP Flags, Sensor Side/Side Not Used See note below: NOTE: Screw, Cap-Head M4 x 0.7 x 6 mm (only use on older systems with the reagent probe assembly mounted with two screws) NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1) 7295993 1873480

60012x2 60026x5

3 4 5 6 7 8 9 10 11 12 13 14 15

60028x5 571244B 60006x43 110984 60000x43 571214B 571247B 571216 26870 18006x1 571215B 078-0020-01

60012x2

572417

16 17 18 19 20 21 22 23

Spacer PC Board, Probe 2 Horizontal Interconnect Gear, Spur 24P 16T 0.250W 0.187B Screw, Pan-Head #4-40 x 0.438 Phillips PC Board, Reagent Probe 2 Heater Control and Volume Detect Sensor Screw, Flat-Head #4-40 x 0.38 Phillips Plate, Motor Mounting Motor, Stepper 4VDC

6362506 1363407

26626 575437 571226B 572502

2565887

575407C 60031x406 571224

6300918

572905

078D0080-02, Rev. A

1-310

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Reagent Probe 2

Item

Description

REF

Part Number

24 25 26 27 28 29 30 31 32 33 34 35 36 37 38

Pulley, 40DP 10T Screw, Special Hex Spacer, 0.125 x 0.187 x 0.010 Stainless Bearing, Ball Flanged, 0.124 x 0.312 x 0.1406 Not Used Not Used Shaft, Guide Leadscrew and Follower Guide, Block Follower Probe, Reagent Sleeve, Probe Belt Pulley, 40 DP Sensor, Optical Switch 112585 Kit, Heater Coil w/ Bubble Detector Sensor, Volume Detector Insulator Gasket, Volume Detection Loctite, Adhesive Bracket, Reagent Bubble Detector

4692398 7817930

571230 571236 26625 571222

571219 5483601 6362506 571218 26626 571233 078-0087-02 6034835 4806784 571223 26083 078-K015-03 113692 571238 9051595 116314 18006 078-0051-01 115272 518-0877-01 9991636 615801 562-3070-01 25029x85

39 40 41 42

Not Used Casting, Machined Probe 2 Carriage Fitting, Barbed Tubing, Silicon Fastener, Tywrap NOTE: For replacement of the reagent probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.

078D0080-02, Rev. A

Reagent Probe 3 571109C


Reagent Probe 3 Assembly

078D0080-02, Rev. A

1-312 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 3 Tubing Routing

Reagent Probe 3 Tubing Routing

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

1-313

Motor Detail

Sensor Detail

078D0080-02, Rev. A

1-314

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-315

078D0080-02, Rev. A

Parts List Reagent Probe 3 571109C


Item Description REF Part Number

1 2

Screw, Cap-Head M4 x 0.7 x 12 mm Washer, Lock M4 Zinc (only use with item 16 on older systems with the reagent probe assembly mounted with two screws) Washer, Flat M4 Zinc 60028x5 Screw, Pan-Head M3 x 0.5 x 6 mm Phillips Plate, R3 Pulley Guard Bracket, Cable Carrier Probe 3 Grommet Strips Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 Bracket, Upper Chain Mount, Rail & Rack Rack, 24DP Guide, Linear Reagent Probe Flags, Sensor Side/Side Pulley, 40DP 36T 0.188W 0.188B Belt, 40DP 0.188W Screw, Pan-Head M3 x 0.5 x 8 mm Phillips See note below: NOTE: Screw, Cap-Head M4 x 0.7 x 6 mm (only use on older systems with the reagent probe assembly mounted with two screws) NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1) Spacer PC Board, Probe 3 Horizontal Interconnect Gear, Spur 24P 16T 0.250W 0.187B Screw, Pan-Head #4-40 x 0.438 Phillips PC Board, Reagent Probe 3 Heater Control Volume Detect Screw, Flat-Head #4-40 x 0.38 Phillips Plate, Motor Mounting 1153232 6362506 1591744 4806784 6034835 1873480

60012x23 60026x5

3 4 5 6 7 8 9 10 11 12 13 14 15 16

60000x43 078-0064-01 571250 12298 60006x43 571247B 571216 571215B 571214B 078-0021-01 26083 571223 60000x53 60012x2

572417

17 18 19 20 21 22 23

26626 575442B 571226B 572502 575412C 60031x406 571224

078D0080-02, Rev. A

1-318

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Reagent Probe 3

Item

Description

REF

Part Number

24 25 26 27 28 29 30 31 32 33 34 35 36 37 38

Motor, Stepper 4VDC Screw, Special Hex Head #8-32 Nylon Not Used Not Used Leadscrew & Follower Probe, Reagent Sleeve, Probe Bearing, Ball Flanged 0.125 x 0.3125 x 0.1406 Guide, Block Follower Shaft, Guide Spacer, Inner Race 0.125 x 0.187 x 0.010 Pulley, 40DP 10T 0.188W 0.125B Guide, Chain, Probe 3 Lower Sensor, Optical Switch Kit, Heater Coil w/ Bubble Detector Sensor, Volume Detector Insulator Gasket, Volume Detection Loctite, Adhesive Bracket, Reagent Bubble Detector

6300918 7817930

572905 571236

5483601 4042237

571218 571233 078-0087-02 571222

1096417

571217 571219 26625

4692398 5566949 2368984

571230 110986 112585 078-K015-03 113692 571238

9051595

116314 18006 078-0051-01 115275 518-0877-01

39 40 41 42

Not Used Casting, Machined Probe 3 Carriage Fitting, Barbed Tubing, Silicon Fastener, Ty-wrap
NOTE: For replacement of the reagent

9991636 615801

562-3070-01 25029x85

probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.

078D0080-02, Rev. A

Ancillary Queue for the ADVIA Centaur System 078-B035-01


Ancillary Queue Assembly

078D0080-02, Rev. A

1-320

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Queue Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-321

Sensor Detail

078D0080-02, Rev. A

1-322

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

1-323

Motor and Pulley Detail

078D0080-02, Rev. A

Parts List Ancillary Queue 078-B035-01 for the ADVIA Centaur System
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28

Screw, Pan-Head M5 x 0.8 x 16 mm Washer, Flat, M5 Cover, Ancillary Queue Screw, Flat-Head M5 x 0.8 x 16 mm Screw, Cap-Head M4 x 0.7 x 10 mm Washer, Flat M4 Block, Mounting Screw, Cap-Head M4 x 0.7 x 16 mm Mount, Sensor Sensor, Optical Switch Not used Spring, Ancillary Clamp Clamp, Bearing Holder, Ancillary Pack Gear, Spur Base, Ring Thermistor, Threaded Body Thermal Grease Spacer, TED TED, Assembly Clamp Insulator, Ancillary Queue, ADVIA Centuar System Sink, Heat, ADVIA Centaur System Washer, Shoulder Screw, Cap-Head M4 x 0.7 x 20 mm Motor, Stepper 4VDC Plate, Motor Mounting Screw #4-40 x .31 Screw, Cap M4 x 06 Hex Screw, Pan-Head M4 x 0.7 x 6 mm 01689043 04380558 05286512 08373815 01943640 07583492 08823500 02052758 1849083 09770826 09397904 00962374 02960484 06903884 09269183

60004x86 60056x6 571924B 60055x86 60013x23 60056x5 572208 60013x2 572115B 111674 078-0038-01 078-0077-01 078-0079-01 572108 078-0080-01 571184C 121055 572118B 111679 18312 572124C 572110 572106 60013x51 572924 078-0038-01 835-2127-01 60013x2 60003x45

078D0080-02, Rev. A

Item

Description

REF

Part Number

29 30 31 32 33

Scanner, Barcode PC Board, Ancillary Queue Sensor Not used Screw, Self-Tapping, #4-20 x 0.38 Bracket, Barcode Tube, Drainage

03369739 01195296

571259E 575222B 572773 572202D 572206B

Ancillary Queue for the ADVIA Centaur XP System 078-B035-01


Ancillary Queue Assembly

078D0080-02, Rev. A

1-328

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Queue Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

1-329

Sensor Detail

078D0080-02, Rev. A

1-330

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

1-331

Motor and Pulley Detail

078D0080-02, Rev. A

Parts List Ancillary Queue 078-B069-01 for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

Screw, Pan-Head M5 x 0.8 x 16 mm Washer, Flat, M5 Cover, Ancillary Queue Screw, Flat-Head M5 x 0.8 x 16 mm Block, Mounting Screw, Cap-Head M4 x 0.7 x 20 mm Washer, Flat M4 Mount, Sensor Sensor, Optical Switch Screw, Self-Tapping, #4-20 x 0.38 Screw, Button M3 x 0.5 x 6 mm Bracket, Flag Spring, Ancillary Pack Clamp Standoff, M3 x 25 mm Clamp, Bearing Holder, Ancillary Pack Gear, Spur Motor, Stepper 4VDC Screw, Cap-Head M4 x 0.7 x 10 mm Screw, Flat-Head, M 6 x 20 Base, Ring TED, Assembly Clamp, Cable, Flat Sink, Heat Screw, Cap M4 x 06 Hex Scanner, Barcode Bracket, Ancillary Queue Motor Bracket, Barcode Bracket, PC Board PC Board, Ancillary Queue Sensor Screw, Pan-Head M4 x 0.7 x 6 mm 01195296 03545693 08497077 08827956 08823500 03714916 01689043 3369739 03159785 06903884 01166571 06088064 02683391 01654495 07015613 08373815 09269183 01988466 03818185 00474949

60004x86 60056x6 078-0279-01 60055x86 572208 60013x51 60056x5 078-0278-01 21470x1 572773 572496 078-0277-01 078-0283-01 558-0869-06 078-0291-01 078-0281-01 078-0387-01 078-0388-01 60013X16 846-1032-62 078-0276-01 571505B 18312 078-0280-01 60013x2 571259E 078-0430-01 572202D 572123 575222B 60000x45

078D0080-02, Rev. A

1-334

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Ancillary Queue

Item

Description

REF

Part Number

31 32 33 34

Standoff, Ring Base Thermistor, Threaded Body Thermal Grease Support, Thermistor, Insulation Washer, Shoulder #8 Tube, Drainage

00576148 01849083 00217709 05286512 03214786

078-0284-01 571184C 120491 078-0285-01 572106 078-0286-01

078D0080-02, Rev. A

Replacing the Ancillary Queue TEDs


CAUTION

Failure to remove the ancillary reagent packs before replacing the TEDs will damage the system. 1. Remove the ancillary reagent packs: a. At the workspace, select System Status. b. Select Diagnostic Tools. c. Select Empty Ancillary Queue. d. Remove the ancillary reagent packs and store them in a refrigerator. 2. Turn off the system: a. At the Workspace, select the system symbol and then select Turn System Off. b. At the back of the system, turn off the main power and unplug the power cord. 3. Remove the ancillary reagent queue.
NOTE: Label all cable connections before removing the cables.

4. Remove the four screws that secure the ancillary entry to the baseplate.
CAUTION

Failure to remove the ancillary drive motor before removing the baseplate will damage the system. 5. Remove the ancillary queue drive motor. 6. Remove the ancillary entry. 7. Remove the defective TEDs:
CAUTION

Disconnect ancillary queue motor and sensor cables before removing the bottom plate or damage can occur to the system.

078D0080-02, Rev. A

336ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs

a. Remove the three screws that secure the bottom plate to the standoffs.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs

b. Remove the four screws and nylon washers that secure the heatsink to the bottom of the ancillary queue.

c. Remove the two TEDs. d. Remove the two TED spacer plates. e. Clean the old thermal grease from each spacer plate. f. Apply new thermal grease to the chamfered side of each spacer plate. g. Install the spacer plates with the chamfered side flush with the bottom of the ancillary queue. 8. Loosen the eight screws that secure the ducting to the heatsink.

078D0080-02, Rev. A

338ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs

9. Install the TEDs:


CAUTION

Ensure that no thermal grease is inside the TED or heating problems will occur. a. Apply a thin layer of thermal grease to both sides of the new TEDs.
CAUTION

Ensure the TEDs are installed flush with the bottom of the ancillary queue or damage can occur to the TEDs. b. Install the new set of TEDs on the bottom of the preheater ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
CAUTION

Failure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs. c. Install the TED heatsink and tighten the four screws. Tighten the two inside screws first and then the two outside screws to 0.68 N-m (6 in-lb) on the TED heat sink. d. Wait three minutes and then tighten the two inside screws first and then the two outside screws to 0.68 N-m (6 in-lb). 10. Install the three screws that secure the baseplate to the standoffs. 11. Place the ducting flat against the baseplate. 12. Tighten the eight screws that secure the ducting to the heatsink. 13. Align and install the ancillary entry. 14. Connect the motor and sensor cables. 15. Align and install the ancillary queue motor. 16. Install the ancillary queue. 17. Turn the system on: a. Connect the power cord. b. At the back of the system, turn the power on. c. At the back of the system, press the computer startup switch. 18. Calibrate the ancillary probe. 19. Calibrate the ancillary queue thermals.

078D0080-02, Rev. A

Ancillary Entry 078-B036-01


Housing

078D0080-02, Rev. A

340

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Entry Motor

Ancillary Entry Motor

Leadscrew Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

341

Sensor Detail

Motor and Pulley Detail

078D0080-02, Rev. A

342

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

343

078D0080-02, Rev. A

Parts List Ancillary Entry 078-B036-01


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

Screw, #4-20 Self-Tap Top Guide, Sliding Door Door, Ancillary Infeed Insulation Gear, Spur Screw, #4-40 x 0.625 Washer, Flat 0.125 x 0.280 x 0.032 Screw, Cap-Head M4 x 0.7 x 6 mm Washer, Flat M4 Screw, Cap-Head M3 x 0.5 x 10 mm PC Board, Pack Sensor Bracket, Pack Sensor Board Rod, Anti-pivot Housing, Infeed Spacer, 0.194 x 0.500 x 0.031 Pulley, Timing Belt, 40DP Pulley, Timing Belt, 40P Belt Motor, Stepper 4VDC Foam, Strip Bearing, Ball 0.188 x 0.375 x 0.125 Sensor, Optical Switch Screw, Pan-Head M3 x 0.5 x 5 mm PC Board, Ancillary Entry Screw, Flat-Head M3 x 0.5 x 10 mm Leadscrew Follower Assembly Leadscrew follower bar nut Arm Washer, #4 Screw, #4-40X Bracket, Infeed 05094672 848-0128-01 01487173 02792883 07236865 00962374 6301817 05762977 6043184 05981962 01689043 04563393

572773 572189B 078-0082-01 078-0171-01 572192 572774 25042x63 60013x2 60028x5 60013x15 575457B 572200D 572196 572188B 572187 571628B 572197 572198 572196 572201B 572019 111674 60001x33 575232D 60007x63 078-B019-01 078-0052-01 818-0053-01 848-0128-01 078-0385-01

078D0080-02, Rev. A

Inprocess and Exit Queue 571040B


Motor Location

078D0080-02, Rev. A

348

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Gap

Motor Gap

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

349

Motor and Pulley Detail

078D0080-02, Rev. A

Parts List Inprocess and Exit Queue 571040B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23

Screw, Cap-Head M5 x 0.8 x 8 mm Support, Tip Tray Screw, Flat-Head M3 x 0.5 x 8 mm Plate, Positioning, Sample Rack Screw, Flat-Head M4 x 0.7 x 10 mm Spring, Positioning Screw, Button M3 x 0.5 x 6 mm Socket Steel Bracket, Flag Strip Washer, Lock Int Tooth M3 Zinc Screw, Cap-Head M3 x 0.5 x 6 mm Screw, Pan-Head M4 x 0.7 x 12 mm PC Board, Inprocess Queue Tip Sensor Washer, Lock Int Tooth M4 Zinc Screw, Cap-Head M4 x 0.7 x 16 mm Washer, Flat M4 Zinc Screw, Cap-Head M4 x 0.7 x 8 mm Guide, Linear, Inprocess Queue Mounting, Support Ring, Retaining Spacer, 0.2525 x 0.310 x 0.051 Clamp, Belt Support Clamp, Timing Belt Pulley, Timing, 0.200P 30G 0.438W 0.500B Pulley Bearing (Quantity 2) 4308474 6811300 916127 7937995 8744694 7882579 7946811 7871447

60013x10 571699B 60055x53 571968B 60055x65 571976 572496 571971B 60027x3 60013x1 60003x75 575217B 60027x5 60013x37 60056x5 60013x9 571965 571969B 11774 23277 571970 571974 114477 571967 26541 571973 571977 6506508 571966 571978

24 25 26 27

Pulley, Timing, 0.200P 20G 0.438W 0.250B Plate, Motor Mounting Belt, Timing, 0.200P 0.38W 296G Neoprene Spacer, 0.250 x 0.375 x 0.375 Stainless

078D0080-02, Rev. A

352

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Inprocess and Exit

Item

Description

REF

Part Number

28 29 30 31 32 33 34

PC Board, Rack Present Latch, Draw Adjustable Non-corrosive PC Board, Exit Queue Interconnect Motor, Stepper 6VDC Screw, Cap-Head M4 x 0.7 x 10 mm Cover, Exit Queue Back Tray, Exit Queue

6438634 1346863 4616756

S575292B 572101 575302C 572909 60013x16 571963 571004

078D0080-02, Rev. A

Sample and Ancillary Probe for the ADVIA Centaur System 078-B045-01
Ancillary Motor and Belt Location

078D0080-02, Rev. A

354

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Motor and Belt Location

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Motor and Belt Location

355

Sample Motor and Belt Location

078D0080-02, Rev. A

356

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Probe Assembly Detail

Sample Probe Assembly Detail


Vertical Motor Location

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Probe Assembly Detail

357

Ancillary Probe Assembly Detail


Vertical Motor Location

078D0080-02, Rev. A

358

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Diluter Detail

Sample Diluter Detail

Ancillary Diluter Sample Air Pump Assembly Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401

359

Diluter 2 mL 115401

078D0080-02, Rev. A

360

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TriContinent Diluter 572888

361

TriContinent Diluter 572888

Motor and Pulley Detail

078D0080-02, Rev. A

362

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

078D0080-02, Rev. A

Sample and Ancillary Probe for the ADVIA Centaur XP System 078-B123-01
Ancillary Motor and Belt Location

078D0080-02, Rev. A

364

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Motor and Belt Location

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Motor and Belt Location

365

Sample Motor and Belt Location

078D0080-02, Rev. A

366

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Probe Assembly Detail

Sample Probe Assembly Detail


Vertical Motor Location

Ancillary Probe Assembly Detail


Vertical Motor Location

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Probe Assembly Detail

367

078D0080-02, Rev. A

368

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Diluter Detail

Sample Diluter Detail

Ancillary Diluter Sample Air Pump Assembly Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401

369

Diluter 2 mL 115401

078D0080-02, Rev. A

370

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TriContinent Diluter 572888

371

TriContinent Diluter 572888

Motor and Pulley Detail

078D0080-02, Rev. A

372

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

078D0080-02, Rev. A

Parts List Sample and Ancillary Probe 078-B123-01 for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12

Pulley, Timing Belt, 0.080P 15T 0.41W 0.188B Motor, Stepper 4VDC PC Board, Sample and Ancillary Probe GIOB Belt, Timing, 0.080P 0.375W 487T Neoprene Ring, Retaining Spacer, 0.2525 x 0.310 x 0.051 Plate, Sample Ancillary Probe Pulley PC Board, Sample and Ancillary Probe Horizontal Travel Motor, Stepper 4VDC Belt, Timing, 0.080P 0.375W 235T Neoprene Bearing, Ball, Flanged 0.312 x 0.500 x 0.156 Shielded Leadscrew and Follower Flag, Follower Washer, Flat M4 Screw, Cap M4 x 10

5232986 5192771 9822435 4513760 7937995 7879650 2070780 6300918 6902853 4811788

572209B 572914 S575322H 572224 11774 23277 078-B039-01 575212C 572905 572223 572242 078-0039-01 571464B 60056x5 60013x16 60013x17

13 14 15 16 17 18 19 20 21

Screw, Cap-Head M5 x 0.8 x 10 mm Hex Sensor, Bubble Detector Support, Ancillary Probe Probe, Ancillary Sleeve, Ancillary Probe Pulley, Timing Belt, 40DP 16G 0.312W 0.188B Belt, Timing, 40P 0.312W 105T Polyurethane Pulley, Timing, 40DP 48G 0.35W 0.188B Motor, Stepper 4VDC 572924 6043184 9327949 4737391 9457664 9580040 4042237

571954C 078-0040-01 571233 078-0087-02 571628B 572215B 571598

078D0080-02, Rev. A

374

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary

Item

Description

REF

Part Number

22 23 24 25 26 27 28 29

Screw, Flat-Head #4-40 x 0.25 Slotted 25040x45 PC Board, Ancillary Probe Vertical S575282C Bearing, Ball, Flanged 0.188 x 0.375 x 0.125 Shielded Screw, Pan-Head M2.5 x 0.4 x 10 mm Phillips PC Board, Sample Pressure Level Sense Plate, Mounting Not Used Plunger, Sample, Assembly Body, Plunger Spring, Comp SS Lee LC-029C-14 Plunger Ring, Internal Retaining E-RIX-7 Grease 3570019 6663735 8999323 1337473 5047542 3398496 6821152 117735 118345 572463B 572217C 572218 20920 118344 118365 23281 23280 8793466 1776434 5341874 2447566 9129802 7121359 8675129 3544034 8255995 1630340 11570 26134 572032B 572030B 572031B 112248 112095 572220 111669 078-K013-01 571042 5545666 7236865 572019 60000x62 575462E 115808

30 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46

Pulley, 40DP 24T 0.312W 0.188B PC Board, Sample Probe Vertical Screw, Pan-Head #6-32 x 0.20 Spacer 0.380 OD x 0.252 ID Gasket, O-ring, 0.380 x 0.060 Diluter, Sample 250 L1458x17 Syringe, Sample 250 L Guide, Linear, Probe Upper Guide, Linear, Ancillary Probe Guide, Linear, Probe Lower Plunger, 2 mL, Tri-Continent Not Used Filter, Level Sense Connector, T 20GA Pump, Sample Probe Fluid Sense Kit Kit, Valve, Solenoid, Two-Way, V23, V24 Damper, Pulsation

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary Probe

Item

Description

REF

Part Number

47 48 49 50 51 52 53

Plate, Motor Mounting Not Used Carriage Fitting, Quick Disconnect Fitting, Luer-20GA Spacer 0.187 x 0.250 x 0.063 Carriage, Sample and Ancillary Frame, Carriage Bearing Bearing (Quantity 2) Leadscrew and Follower Rod 4811788 7236865 7041363 1179975 1879357

571465B 572120B 23559 110268 110267 113764 572129C 572242 572019 572130 571570B 2807554 110265 60013x9 078-0007-01 13002x1 60055x55 60007x73 571466 572138B 572136B 572137

54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71

Rinse Station, Ancillary Probe Screw, Soc. Head Cap M4 x 8 SST Bracket, Motor Mounting Screw, Binding #4-40x 0.31 Screw, Flat M4 x 8 mm Screw, Flat M3 x 0.5 x 12 mm Clamp, Belt Bracket Clamp, Belt Bracket Screw, Cap M3 x 0.5 x 10 mm 60013x15 Screw, M3 x 0.5 x 8 mm 60013x8 Screw, Cap M4 x 0.7 x 10 mm 60013x16 Washer, Flat M4 60056x5 Motor, Diluter 1458x295 Clamp, Rinse Station 26699 Washer, Flat M4 60056x5 Washer, Lock M4 60027x5 Screw, Cap M4 x 0.7 x 16 mm 60013x37 Washer, Flat M3 60056x3

078D0080-02, Rev. A

376

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary

Item

Description

REF

Part Number

72

Diluter, 2 mL, Note: The new style diluter is interchangeable with the Tri-Continent diluter. Fitting (two piece insert with barb) 110080 Chamber, Fluid O-Ring 0.614ID 070W Ethylene propylene Seal, 2.0 mL, 0.500 ID PC Board, Diluter Screw #4-40 x 187 Nut, Chamber Screw, Cap-Head M4 x 12 1718884 3275688 1926711 7026771

115401

73 74 75 76 77 78 79 80

115398 115945 115395 114830 60030x403 115633 60013x23

078D0080-02, Rev. A

Parts List Sample and Ancillary Probe 078-B045-01


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12

Pulley, Timing Belt, 0.080P 15T 0.41W 0.188B Motor, Stepper 4VDC PC Board, Sample and Ancillary Probe GIOB Belt, Timing, 0.080P 0.375W 487T Neoprene Ring, Retaining Spacer, 0.2525 x 0.310 x 0.051 Plate, Sample Ancillary Probe Pulley PC Board, Sample and Ancillary Probe Horizontal Travel Motor, Stepper 4VDC Belt, Timing, 0.080P 0.375W 235T Neoprene Bearing, Ball, Flanged 0.312 x 0.500 x 0.156 Shielded Leadscrew and Follower Flag, Follower Washer, Flat M4 Screw, Cap M4 x 10

5232986 5192771 9822435 4513760 7937995 7879650 2070780 6300918 6902853 4811788

572209B 572914 S575322H 572224 11774 23277 078-B039-01 575212C 572905 572223 572242 078-0039-01 571464B 60056x5 60013x16 60013x17

13 14 15 16 17 18 19 20 21 22

Screw, Cap-Head M5 x 0.8 x 10 mm Hex Sensor, Bubble Detector Support, Ancillary Probe Probe, Ancillary Sleeve, Ancillary Probe Pulley, Timing Belt, 40DP 16G 0.312W 0.188B Belt, Timing, 40P 0.312W 105T Polyurethane Pulley, Timing, 40DP 48G 0.35W 0.188B Motor, Stepper 4VDC 572924 Screw, Flat-Head #4-40 x 0.25 Slotted 25040x45 6043184 9327949 4737391 9457664 9580040 4042237

571954C 078-0040-01 571233 078-0087-02 571628B 572215B 571598

078D0080-02, Rev. A

378

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary

Item

Description

REF

Part Number

23 24 25 26 27 28 29

PC Board, Ancillary Probe Vertical S575282C Bearing, Ball, Flanged 0.188 x 0.375 x 0.125 Shielded Screw, Pan-Head M2.5 x 0.4 x 10 mm Phillips PC Board, Sample Pressure Level Sense Plate, Mounting Not Used Plunger, Sample, Assembly Body, Plunger, ADVIA Centaur System Spring, Comp SS Lee LC-029C-14 Plunger Ring, Internal Retaining E-RIX-7 Grease 8999323 1337473 5047542 5977175 539781 572463B 572217C 572218 20920 572205 575227B 23281 23280 8793466 1776434 5341874 2447566 9129802 7121359 8675129 3544034 8255995 1630340 11570 26134 572032B 572030B 572031B 112248 112095 572220 111669 078-K013-01 571042 571465B 3570019 6661600 117735 572133B 5545666 7236865 572019 60000x62 575462E 115808

30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47

Pulley, 40DP 24T 0.312W 0.188B PC Board, Sample Probe Vertical Screw, Pan-Head #6-32 x 0.20 Spacer 0.380 OD x 0.252 ID Gasket, O-ring, 0.380 x 0.060 Diluter, Sample 250 L1458x17 Syringe, Sample 250 L Guide, Linear, Probe Upper Guide, Linear, Ancillary Probe Guide, Linear, Probe Lower Plunger, 2 mL, Tri-Continent Not Used Filter, Level Sense Connector, T 20GA Pump, Sample Probe Fluid Sense Kit Kit, Valve, Solenoid, Two-Way, V23, V24 Damper, Pulsation Plate, Motor Mounting

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary Probe

Item

Description

REF

Part Number

48 49 50 51 52 53

Not Used Carriage Fitting, Quick Disconnect Fitting, Luer-20GA Spacer 0.187 x 0.250 x 0.063 Carriage, Sample and Ancillary Frame, Carriage Bearing Bearing (Quantity 2) Leadscrew and Follower Rod 4811788 7236865 7041363 1179975 1879357 572120B 23559 110268 110267 113764 572129C 572242 572019 572130 571570B 2807554 110265 60013x9 078-0007-01 13002x1 60055x55 60007x73 571466 572138B 572136B 572137

54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72

Rinse Station, Ancillary Probe Screw, Soc. Head Cap M4 x 8 SST Bracket, Motor Mounting Screw, Binding #4-40x 0.31 Screw, Flat M4 x 8 mm Screw, Flat M3 x 0.5 x 12 mm Clamp, Belt Bracket Clamp, Belt Bracket Screw, Cap M3 x 0.5 x 10 mm 60013x15 Screw, M3 x 0.5 x 8 mm 60013x8 Screw, Cap M4 x 0.7 x 10 mm 60013x16 Washer, Flat M4 60056x5 Motor, Diluter 1458x295 Clamp, Rinse Station 26699 Washer, Flat M4 60056x5 Washer, Lock M4 60027x5 Screw, Cap M4 x 0.7 x 16 mm 60013x37 Washer, Flat M3 60056x3 Diluter, 2 mL, Note: The new style diluter is interchangeable with the Tri-Continent diluter.

115401

078D0080-02, Rev. A

380

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample and Ancillary

Item

Description

REF

Part Number

73 74 75 76 77 78 79 80

Fitting (two piece insert with barb) 110080 Chamber, Fluid O-Ring 0.614ID 070W Ethylene propylene Seal, 2.0 mL, 0.500 ID PC Board, Diluter Screw #4-40 x 187 Nut, Chamber Screw, Cap-Head M4 x 12 1718884 3275688 1926711 7026771 115398 115945 115395 114830 60030x403 115633 60013x23

078D0080-02, Rev. A

Sample Entry Queue for the ADVIA Centaur System 078-B002-01


Ancillary Motor and Belt Location

078D0080-02, Rev. A

382

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Tensioner Detail

Belt Tensioner Detail

Belt Drive Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Magnet Detail

383

Magnet Detail

Motor Detail

078D0080-02, Rev. A

384

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

385

078D0080-02, Rev. A

Parts List Sample Entry Queue for the ADVIA Centaur System 078-B002-01
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Screw, Cap-Head M4 x 0.7 X 8 mm Cover, Entry Queue, Back Screw, Button-Head M3 x 0.5 X 5 mm Phillips PC Board, Rack Present PC Board, Sample Entry Queue Sensor Motor, Stepper 4VDC Bearing, Ball 0.250 x 0.375 x 0.125 Bearing Gear, Spur Gear, Spur Loctite, 222 Shaft, Input Queue Drive Loctite, Moly-paste Ring, Retaining E-ring 0.250D Screw, Cap-Head M3 x 12 mm Washer, Flat Screw, Cap-Head #4-40 x 0.375 Belt, Entry Queue Panel, Entry Queue, Back Cover Spring Block, Tensioner Mounting Block, Tensioner Bearing Shaft, Tensioner Pulley, Tensioner Loctite, Thread Locker Washer, Flat M3 Queue, Magnet Input Cover, Magnet Magnet

02784961 05561920 02683391 06438634 08726947 05192771 06068713 06776297 09157865 03120951 03904871 00995779 03338817 05785519 08057921 02073054 09697398 05581174 00155614 03105235 00307384 00188148 05314990 03131961 00237246 08249375 03931585

60013x9 078-0117-01 572496 S575292B 575242B 572914 26082x7 116258 571988 572770 17414 078-0023-01 590-0331-04 26107 60013x22 846-1000-06 572421 571001 110214 116259 571995 571993 572003 571994 590-0014-02 60056x3 078-B011-01 571990 571991 078-K071-01

25

Kit, Entry Queue Belt Guide

078D0080-02, Rev. A

388 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample Entry Queue for

Item

Description

REF

Part Number

Plate, Base

571999

078D0080-02, Rev. A

Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01


Ancillary Motor and Belt Location

Belt Tensioner Detail

078D0080-02, Rev. A

390

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Drive Detail

Belt Drive Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Magnet Detail

391

Magnet Detail

Motor Detail

078D0080-02, Rev. A

392

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

393

078D0080-02, Rev. A

Parts List Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01
Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Not used Not used Screw, Button-Head M3 x 0.5 X 5 mm Phillips PC Board, Rack Present PC Board, Sample Entry Queue Sensor Motor, Stepper 4VDC Bearing, Ball 0.250 x 0.375 x 0.125 Bearing Gear, Spur Gear, Spur Loctite, 222 Shaft, Input Queue Drive Loctite, Moly-paste Ring, Retaining E-ring 0.250D Screw, Cap-Head M3 x 12 mm Washer, Flat Screw, Cap-Head #4-40 x 0.375 Belt, Entry Queue Not used Spring Block, Tensioner Mounting Block, Tensioner Bearing Shaft, Tensioner Pulley, Tensioner Loctite, Thread Locker Washer, Flat M3 Queue, Magnet Input Cover, Magnet Magnet 237246 307384 188148 5314990 155614 116259 571995 571993 572003 571994 590-0014-02 60056x3 078-B011-01 571990 571991 078-K071-01 9697398 2073054 5785519 995779 9157865 3120951 5192771 6438634 572496 S575292B 575242B 572914 26082x7 116258 571988 572770 17414 078-0023-01 590-0331-04 26107 60013x22 846-1000-06 572421 571001

25

Kit, Entry Queue Belt Guide

078D0080-02, Rev. A

396 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample Entry Queue for

Item

Description

REF

Part Number

Plate, Base

078-0327-01

078D0080-02, Rev. A

Sample Exit Pusher 571037B


Sample Exit Pusher Assembly

Sensor Detail

Motor Detail

078D0080-02, Rev. A

398

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

399

078D0080-02, Rev. A

Parts List Sample Exit Pusher 571037B


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10

Motor, Stepper 4VDC Gear, Spur 48P 28T 0.187W 0.187B Loctite, Adhesive Housing, Pusher Screw, Cap-Head #4-40 x 0.375 Sensor, Optical Switch, Wide Gap Screw, Cap-Head M3 x 0.5 x 6 mm Gear, Rack, Exit Pusher Washer, Pusher End Screw, Flat-Head M3 x 0.5 x 6 mm Washer, Flat 3048878 7882579

572914 522-0309-01 590-0014-02 572009B 572421 111675 60013x1 572007 572595 60055x43 60056x3

078D0080-02, Rev. A

Sample Tip Loader 571036C for the ADVIA Centaur System


Sample Tip Loader Assembly

078D0080-02, Rev. A

404

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Loader

Sample Tip Loader

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

405

Sensor Detail

Motor and Pulley Detail

078D0080-02, Rev. A

Sample Tip Loader 078-B136-01 for the ADVIA Centaur XP System

078D0080-02, Rev. A

408

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Loader

Sample Tip Loader

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

409

Sensor Detail

Motor and Pulley Detail

078D0080-02, Rev. A

410

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail

078D0080-02, Rev. A

Parts List Sample Tip Loader


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Screw, Cap-Head M3 x 0.5 X 10 mm Washer, Flat M3 PC Board, Tip Tray and Band Sensor Cover, Sensor Screw, Set M3 x 0.5 x 4 mm Guide, Tip Tray Cover Guide, Tip Tray, Indexing Left Bearing, Plastic Flanged 0.125 x 0.172 x 0.078 Ring, Retaining, E-Ring 0.095 x 0.240 x 0.017 Stainless Spring, Extension 0.18 x 1.12 3.08 lb/In Screw, Pan M3 x 6 mm Block, Bearing Screw, Cap-Head M4 x 0.7 X 10 mm Washer, Flat M4 SST Loader, Sample Tip Tray Assembly Shaft, Square Shaft, Arm Pivot Ring, Retaining (Quantity 2) Arm (Quantity 2) Block (Quantity 2) Spring Screw, Set M3 x 8 (Quantity 2) Loctite, Adhesive 222 1794238 856027 4179763

60013x15 60056x3 S575397C 572434 60022x39 571653 572213 571646 571672 572682B 6004x43 078000901 846101030 846100004 113762C 571676C 571675E 14458x4 571644C 572765B 571693B 572771B 17414 571693B 60055x43 5144580 571692B 571695 571696 962374 3909210 111674 575352B

16 17 18 19 20 21 22

Spring, Tip Tray Latch Screw, Flat-Head M3 x 0.5 x 6 mm Socket Spring, Guide, Front Support, Spring Guide, Back Support, Spring Guide, Front Sensor, Optical Switch PC Board, Tip Loader

078D0080-02, Rev. A

412

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample Tip Loader

Item

Description

REF

Part Number

23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40

Screw, Pan-Head #4-40 x 0.187 Phillips Screw, Cap-Head M4 x 0.7 x 8 mm Hex Gear, Spur 32P 20T 0.187W 0.187B Motor, Stepper 4VDC Spring, Guide, Back Linkage, Tip Tray, Guide Lock Guide, Tip Tray, Indexing Right Plate, Vertical Base Plate, Bearing Bracket, Tip Loader Support Plate, Front, Vertical Plate, Back, Vertical Guide, Tip Tray Orientation Kit, Sample Tip Loader or Guide, Tip Tray Exit Rod, Support Screw, Pan M3 x 6 mm Linkage, Tip Tray Indexing Bar, Tray Guide Right 6455776 204046 8676982 5192771

60030x403 60013x9 572022B 572914 571691B 078-0010-01 572214 571669B 078-0008-01 572433A 571670B 571668B 571657 078-K001-01 571655B 571673A 60004x43 571660 571666

078D0080-02, Rev. A

Sample Tip Waste Area for the ADVIA Centaur System 078-B0009-02
Sample Tip Waste Area Assembly

078D0080-02, Rev. A

414

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Assembly

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Assembly

415

078D0080-02, Rev. A

Parts List Sample Tip Waste Area 078-B009-02 for the ADVIA Centaur System
Item Description REF Part Number

1 2 3 4 5 6 6 7 8 8 9 10 10 11 12 13 14 14 15 16 17 18 19 20

Kit, Cover, Connector (Includes lid and connector) Nut, Hex-Head, M8 Sensor, Optical Switch Screw, Flat-Head M3 x 0.5 x 5 mm Screw, Cap-Head M3 x 0.5 x 6 mm Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur System Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur XP System Sensor, Proximity Slide, Ball Bearing, 11 18 lbs., ADVIA Centaur System Slide, Ball Bearing, 11 - 18 lbs., ADVIA XP Centaur System Screw, Pan-Head, M4 x 0.7 x 6 mm Liner, Tip Waste, ADVIA Centaur System Liner, Tip Waste, ADVIA Centaur XP System Bin, Tip Waste Collar, Tip Waste Bin Screw, Cap-Head M4 x 0.7 x 40 mm Carrier, Cleaning Solution Reservoir, ADVIA Centaur System Carrier, Cleaning Solution Reservoir ADVIA Centaur XP System Bottle, Cleaning Solution Reservoir Tube, Formed Screw, Pan-Head M8 x 90 mm Connector Shutter, Waste Reservoir Screw, Retractable M4 x 0.7 x 5.7 mm, Knurled, Snap Fit 220165 2908954 7526731 0878489 113261 8848643 3828105 111674

078-K02701 60024x9 21470x1 60006x33 60012x1 110228 078-0340-01572432 572523 078-0344-01 60003x35 078-0066-01 078-0343-01 572616 110107 60013x86 572519 078-0342-01 572520 571957 60003x218 See Item 1 572622 571964

078D0080-02, Rev. A

418 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample Tip Waste Area

Item

Description

REF

Part Number

21 22 23 24 25 26 27 28 29 30 31

Clamp, Spring Screw, Pan-Head M4 x 0.7 x 10 mm Spring, Waste Bin Interlock Detent, Waste Bin Interlock Screw, Pan-Head M4 x 0.7 x 6 mm Washer, Flat, M4 Chute, Tip Waste Clamp, Cable Clamp, Cable Screw, Cap-Head M3 x 0.5 x 6 mm Reservoir, Waste Chute

4324275

110242 60000x65 572638 572639

3545693 7109448 2362447 8461889 9580326

60000x45 60028x5 572548 12811x1 23491 60012x1 572547

078D0080-02, Rev. A

Sample Tip Waste Area for the ADVIA Centaur XP System 078-B088-01

078D0080-02, Rev. A

420

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area for the

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area for the

078D0080-02, Rev. A

Parts List Sample Tip Waste Area 078-B088-01 for the ADVIA Centaur XP System
Item Description REF Part Number

1 2 3 4 5 6 6 7 8 8 9 10 10 11 12 13 14 14 15 16 17 18 19 20

Kit, Cover, Connector (Includes lid and connector) Nut, Hex-Head, M8 Sensor, Optical Switch Screw, Flat-Head M3 x 0.5 x 5 mm Screw, Cap-Head M3 x 0.5 x 6 mm Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur System Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur XP System Sensor, Proximity Slide, Ball Bearing, 11 18 lbs., ADVIA Centaur System Slide, Ball Bearing, 11 - 18 lbs., ADVIA XP Centaur System Screw, Pan-Head, M4 x 0.7 x 6 mm Liner, Tip Waste, ADVIA Centaur System Liner, Tip Waste, ADVIA Centaur XP System Bin, Tip Waste Collar, Tip Waste Bin Screw, Cap-Head M4 x 0.7 x 40 mm Carrier, Cleaning Solution Reservoir, ADVIA Centaur System Carrier, Cleaning Solution Reservoir ADVIA Centaur XP System Bottle, Cleaning Solution Reservoir Tube, Formed Screw, Pan-Head M8 x 90 mm Connector Shutter, Waste Reservoir Screw, Retractable M4 x 0.7 x 5.7 mm, Knurled, Snap Fit 220165 2908954 7526731 113261 8848643 3828105 111674

078-K02701 60024x9 21470x1 60006x33 60012x1 110228 078-0340-01572432 572523 078-0344-01 60003x35 572177 078-0343-01 572616 110107 60013x86 572519 078-0342-01 572520 571957 60003x218 See Item 1 572622 571964

078D0080-02, Rev. A

424 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Sample Tip Waste Area

Item

Description

REF

Part Number

21 22 23 24 25 26 27 28 29 30 31

Clamp, Spring Screw, Pan-Head M4 x 0.7 x 10 mm Spring, Waste Bin Interlock Detent, Waste Bin Interlock Screw, Pan-Head M4 x 0.7 x 6 mm Washer, Flat, M4 Chute, Tip Waste Clamp, Cable Clamp, Cable Screw, Cap-Head M3 x 0.5 x 6 mm Reservoir, Waste Chute

4324275

110242 60000x65 572638 572639

3545693 7109448 2362447 8461889 9580326

60000x45 60028x5 572548 12811x1 23491 60012x1 572547

078D0080-02, Rev. A

Stat Sample Pusher 571035


Stat Sample Pusher Assembly

078D0080-02, Rev. A

426

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail

Sensor Detail

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail

427

Motor Detail

078D0080-02, Rev. A

Parts List Stat Sample Pusher 078-B026-01


Item Description REF Part Number

1 2 3 4

Screw, Cap-Head M4 x 0.7 x 6 mm PC Board, Rack Loader Belt, Rack, Infeed Block, Infeed Block Magnet Loctite 8575213 4904972 3048541

60013x2 575237B 571002 113765 572020 572025 572501 572021 7882579 4965793 7343254 60013x1 115027C 078-0390-01 572016 572494 572992 60013x66 5520752 113766 572024 572012 7236865 924782 572019 113767 572024 572013 7236865 572019 60013x64 60013x71 572024 572015 572764 111674 21470x1 572493

5 6 7 7 8 9 10 11 12

Bracket, Infeed, Block, Support Screw, Cap-Head M3 x 0.5 x 6 mm Scanner, Barcode, ADVIA Centaur System Scanner, Barcode, ADVIA Centaur XP System Bracket, Barcode Scanner Screw, Cap-Head #6-32 x 0.250 Cable Tie, Ty-Wrap Screw, Cap-Head M5 x 0.8 x 25 mm Pulley, Infeed Idler Assembly Housing Pulley Bearing (Quantity 2)

13

Pulley, Infeed Drive Assembly Housing Pulley Bearing

14 15 16 17 18 19 20

Screw, Cap-Head M4 x 0.5 x 25 mm Screw, Cap-Head M3 x 0.5 x 30 mm Housing, Infeed Mechanism Block, Pusher, Stat Exit Bracket, Rack Stop Not Used Sensor, Slotted Optical Switch Screw, Cap-Head #4-40 x 0.250

078D0080-02, Rev. A

430

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Stat Sample Pusher

Item

Description

REF

Part Number

21 22 23 24 25 26 27 28 29 30 31 32

Plate, Stat Top Screw, Captive M4 x 0.7 x 15.75 mm PC Board, Rack Present PC Board, Stat Area Gear, Spur 32P 20T 0.187W 0.187B Acetal Motor, Step 4VDC BP 44.4 Ht .80 Shaft Washer, Flat M4 Washer, Flat M3 Housing, Stat Exit Pusher Label, Laser Warning 6438634 8924552 8676982 572914

572026 572027 S575292B 575307C 572022B 571634 60056x5 60056x3 078-0043-01 078-0003-01

078D0080-02, Rev. A

Universal Rack Option


Sample Probe

078D0080-02, Rev. A

432

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Stat Sample Pusher

Stat Sample Pusher

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Pusher

433

Sample Exit Pusher

078D0080-02, Rev. A

434

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Inprocess Queue

Inprocess Queue

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Rack Sensor Pusher Cable

435

Rack Sensor Pusher Cable

078D0080-02, Rev. A

436

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue

Sample Exit Queue

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue Rack Present

Sample Exit Queue Rack Present Sensor

078D0080-02, Rev. A

438

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue Full Sensor

Sample Exit Queue Full Sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue

439

Sample Entry Queue

078D0080-02, Rev. A

440 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue Work Surface

Sample Entry Queue Work Surface

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue Tensioner

441

Sample Entry Queue Tensioner

078D0080-02, Rev. A

Parts List Universal Rack Option


Item Description REF Part Number

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Body, Sample Probe Plunger Pulley, Leadscrew PC Board, Universal Normal Rack Present Standoff Belt, Rack Loader PC Board, Stat Rack Present Housing, Stat Exit Pusher Standoff, Pusher Extension Bar, Spring Mounting Screw, Cap M4 x 20 Stainless Spring, Rack Positioning Screw, Button Head M3 x 6 Stainless Plate, Positioning Cable, Rack Pusher Sensor Cable, Hitachi Exit Queue PC Board, Hitachi Exit Queue Full Sensor, Rack Present Screw, Pan M3 x 0.5 x 6mm Tray, Exit Queue Screw, Flat M3 x 0.5 x 6mm Stainless Steel Washer, Lock M3 Sensor, Exit Queue Full Washer, Flat M3 Nut, M3 x 0.5 Screw, Pan M3 x 0.5 x 5mm Stainless Steel

06663735 03398496 08176629 02062737 0118368C 03538387

118345 118344 118362 118524 118773 118368 118547 118503 118400

02683391 04406301 02683391 03873089 09255077 01603491 08990202 05672579 07908944 00843898 03617287 07311808 05551992 06552356

572496 118402 572496 118401 119728 118479 118375 119729 60000x43 118780 60007x43 60026x3 118481 60028x3 60024x4 60001x33 119668 60013x1 119657 118788 60007x85

PC Board, Rack Pusher/Entry Queue Home 9644596 Screw, M3 x 0.5 x 6mm Stainless Steel Bracket Bracket Screw, Flat M4 x 0.7 x 12mm Stainless Steel Block, Sample Rack Pusher, Assembly 01691579 06892408

119825

078D0080-02, Rev. A

444

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List Universal Rack Option

Item

Description

REF

Part Number

32 32 33 34 35 36

Tray, Entry Queue, ADVIA Centaur System Tray, Entry Queue, ADVIA Centaur XP System Work Surface, Entry Queue Washer, Flat #8 Screw, Cap M4 x 0.7 x 12mm Pulley

06848190 06848190 02254393 06734691 01876943

118781 078-0441-01 118789 25042x67 60012x23 119656

078D0080-02, Rev. A

Miscellaneous Parts
Parts List
Item Description REF Part Number

Blower, Ancillary, Assembly Blower, Primary Reagent Compartment, Assembly Cord, Power Line Fan, Electronics Module Keyboard, English, ADVIA Centaur System Keyboard, English, ADVIA Centaur XP System Keyboard, German, ADVIA Centaur System Keyboard, German, ADVIA Centaur XP System Keyboard, Italian, ADVIA Centaur System Keyboard, Italian, ADVIA Centaur XP System Keyboard, French, ADVIA Centaur XP System Keyboard, Spanish, ADVIA Centaur System Keyboard, Spanish, ADVIA Centaur XP System Modem, Kit Monitor,Kit Printer, North America Scanner, Barcode Assembly, Hand Held Switch, Fast S top (red) Cord, Power Line, European Switch, Computer Startup (green)

09694747 07318179 05430311 01103618 02000952 03010811

572925 572931 572334 571878 571266D 078-0393-01 109955D 078-0394-01

00814111 04719040

109956C 078-0395-01 109958C 078-0397-01

Keyboard, French, ADVIA Centaur System 03036438

07012150 02992998 04227660 03376425

109957C 078-0396-01 115519 078-K070-01 Check eBRITE for latest information

02865821

118099 572158 572866 572516

078D0080-02, Rev. A

1-446

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List

Item

Description

REF

Part Number

Labels, Rack 1-100, ADVIA Centaur System Labels, Rack 101-200, ADVIA Centaur System Labels, Rack 201-300, ADVIA Centaur System Labels, Rack 301-400, ADVIA Centaur System Labels, Rack 401-500, ADVIA Centaur System Labels, Rack 1-100, ADVIA Centaur XP System Labels, Rack 101-200, ADVIA Centaur XP System Labels, Rack 201-300, ADVIA Centaur XP System Labels, Rack 301-400, ADVIA Centaur XP System Labels, Rack 401-500, ADVIA Centaur XP System Kit, Mouse, 3-button SUN

00424364 09270017 00294533 07421808 01135889 09391302 03828105 08579502 01160808 02115415 05749490

112249B 112250B 112251B 112252B 112253B 078-0333-01 078-0334-01 078-0335-01 078-0336-01 078-0337-01 078-K080-01

078D0080-02, Rev. A

2 Kits and Tools


Kits
Description Part REF

Kit, Routine Certification Kit, Extended Certification Kit, Direct Plumbing Kit, Lab Automation Kit, Maintenance Kit, Labcell/Workcell

115835 115836 078-K087-01 115869 112712 078-K077-01

078D0080-02, Rev. A

2-2

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 2 Tools

Tools
Description Part Number REF

Connector, GIOB Bypass Diskette, Controller Diagnostics Sequences Flowmeter Tool, Reagent Pack Tool, Elevator Up Tool, Flow Adapter Tool, Alignment, Reagent/Anc Probe to Cuvette Tool, Ancillary Pack Calibration Tool, Aspirate Probe Home Tool, Barcode Shutter Tool, Reagent Bottom Tool, Sample Probe Calibration Tool, Sample Tip Tray Assembly Alignment Tool, Sample Tip Tray Alignment Tool, Vacuum Calibration Wrench Set, Bondhus Balldriver Hex 1.5 to 10mm Terminator, GIOB YSI M4610 Temperature Meter YSI M4611 Probe Pliers, Groove Joint

116325 078S01901 1458x144 112749 113907 078B02401 117112 111010 113906 114046 114565 117883 114348 114347 117736 110864 111009 110367 110368 113958

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probes Horizontal

3 Centaur PC Boards
Sample and Ancillary Probes Horizontal PC Board 575212C

Connector

Cable

Connected to . . .

J1

571280

J6 of Sample and Ancillary Probe GIOB PC board S575322H

Sensor

Function

U1 U2 U3 U4 U5 U6 U7 U8 U9 U10 U11 U12 U13 U14 U15 U16 U17

Tip 1 Position Tip 2/Tube 1 Position Tip 3 Position Tip 4/Tube 2 Position Tip 5 Position Tip 6/Tube 3 Position Tip 7 Position Tip 8/Tube 4 Position Tip 9 Position Tip 10/Tube 5 Position Tip Eject Position Sample Cuvette Position Lab Auto Position Ancillary Horizontal Home Position Ancillary Cuvette Position Ancillary Wash Position Ancillary Pack Position

078D0080-02, Rev. A

3-2

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Inprocess Queue Tip Sensor PC

Inprocess Queue Tip Sensor PC Board 575217B

Connector

Cable

Connected to . . .

J1

571296

J6 of Sample and Ancillary GIOB PC board S575327D

Sensor

Function

U1 U2 U3 U4 U5

Inprocess Queue Home Rack Position Rack Direction Tip Position Tip Direction

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Queue Sensor PC Board

Ancillary Queue Sensor PC Board 575222B

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8

571292

J4 of Exit Queue Interconnect PC board 575302C Ancillary Queue Home sensor Ancillary Queue Position sensor Ancillary Queue Direction sensor Ancillary Queue motor Ancillary Queue thermistor 1 Not used Not used

078D0080-02, Rev. A

3-4

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Probe Vertical Sensor PC

Sample Probe Vertical Sensor PC Board for the ADVIA Centaur System 575227

Connector

Cable

Connected to . . .

J1 J2

571284

J7 of Sample and Ancillary Probe GIOB S575322H Sample Vertical motor

Sensor

Function

U1 U2 U3

Vertical Home Above Tip Eject Below Tip Eject

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Probe Vertical Sensor PC Board

Sample Probe Vertical Sensor PC Board for the ADVIA Centaur XP System 118365

Connector

Cable

Connected to . . .

J1 J2

571284

J7 of Sample and Ancillary Probe GIOB S575322H Sample Vertical motor

Sensor

Function

U1 U2 U3

Vertical Home Above Tip Eject Below Tip Eject

078D0080-02, Rev. A

3-6 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry PC Board 575232D

Ancillary Entry PC Board 575232D

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7

571287

J5 of Exit Queue Interconnect PC board 575302C Ancillary Entry motor Ancillary Door motor Ancillary Door Home sensor Ancillary Door Position sensor Ancillary LED PC board 575392B Ancillary Pack PC board 575457B

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry PC Board 575232D

3-7

Sensor

Function

U1 U2

Door Home Door Extend

078D0080-02, Rev. A

3-8

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Rack Loaded Sensor PC Board

Rack Loaded Sensor PC Board 575237B

Connector

Cable

Connected to . . .

J1

571342

J8 of Stat Area PC board 575307C

Sensor

Function

U1 U2

Stat Load Position Normal Load Position

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Entry Queue Sensor PC Board 575242B

Entry Queue Sensor PC Board 575242B

Connector

Cable

Connected to . . .

J1

571291

J3 of Stat Area PC board 575307C

Sensor

Function

U1

Entry Queue Home

078D0080-02, Rev. A

3-10

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

Low Level Controller PC Board 078-B042S01

There are four fuses on each PC board; their functions are described below:
Volts Label Rating Location Protects

+24

F0

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 57280

Pumps 2 LLC PC board 078-B042S01

resuspend dispense, motor 0 (P1)

+24

F1

Pumps 2 LLC PC board wash 1, motor 1 (P1)

+24

F2

Pumps 2 LLC PC board wash 2, motor 2 (P1)

+24

F3

Pumps 2 LLC PC board acid, motor 3 (P1)

+24

F0

Pumps 1 LLC PC board wash displacement, motor 0 (P1)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

Volts

Label

Rating

Location

Protects

+24

F1

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Pumps 1 LLC PC board wash 3, motor 1 (P1)

+24

F2

Pumps 1 LLC PC board not used

+24

F3

Pumps 1 LLC PC board base, motor 3 (P1)

+24

F0

Aspirate Probes LLC PC board Aspirate Probes LLC PC board Aspirate Probes LLC PC board Aspirate Probes LLC PC board Preheater & Lumo LLC PC board Preheater & Lumo LLC PC board Preheater & Lumo LLC PC board Preheater & Lumo LLC PC board Reagent Pack LLC PC board

probe 1, motor 0 (P1)

+24

F1

probe 2, motor 1 (P1)

+24

F2

probe 3, motor 2 (P1)

+24

F3

probe 4, motor 3 (P1)

+24

F0

cuvette conveyor, motor 0 (P1) cuvette pusher, motor 1 (P1) luminometer rotor, motor 2 (P1) waste probe, motor 3 (P1) reagent shelf, motor 0 (P1)

+24

F1

+24

F2

+24

F3

+24

F0

078D0080-02, Rev. A

3-12

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

Volts

Label

Rating

Location

Protects

+24

F1

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Reagent Pack LLC PC board Reagent Pack LLC PC board Reagent Pack LLC PC board Reagent Probe 3 LLC PC board Reagent Probe 3 LLC PC board Reagent Probe 3 LLC PC board Reagent Probe 3 LLC PC board Reagent Probe 2 LLC PC board Reagent Probe 2 LLC PC board Reagent Probe 2 LLC PC board Reagent Probe 2 LLC PC board Reagent Probe 1 LLC PC board

reagent barcode, motor 1 (P1) reagent barcode shutter, motor 2 (P1) reagent probe shutter, motor 3 (P1) pack to pack, motor 0 (P1) pack to cuvette, motor 1 (P1) vertical, motor 2(P1)

+24

F2

+24

F3

+24

F0

+24

F1

+24

F2

+24

F3

diluter, motor 3 (P1)

+24

F0

pack to pack, motor 0 (P1) pack to cuvette, motor 1 (P1) vertical, motor 2 (P1)

+24

F1

+24

F2

+24

F3

diluter, motor 3 (P1)

+24

F0

pack to pack, motor 0 (P1)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

Volts

Label

Rating

Location

Protects

+24

F1

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 FB mm 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Reagent Probe 1 LLC PC board Reagent Probe 1 LLC PC board Reagent Probe 1 LLC PC board Sample Tip LLC PC board Sample Tip LLC PC board Sample Tip LLC PC board Sample Tip LLC PC board Ancillary LLC PC board Ancillary LLC PC board Ancillary LLC PC board Ancillary LLC PC board Sample Racks LLC PC board

pack to cuvette, motor 1 (P1) vertical, motor 2 (P1)

+24

F2

+24

F3

diluter, motor 3 (P1)

+24

F0

tip loader, motor 0 (P1) entrance queue, motor 1 (P1) not used

+24

F1

+24

F2

+24

F3

not used

+24

F0

ancillary entry, motor 0 (P1) ancillary queue, motor 1 (P1) ring loader, motor 2 (P1) ancillary door, motor 3 (P1) rack loader, motor 0 (P1)

+24

F1

+24

F2

+24

F3

+24

F0

078D0080-02, Rev. A

3-14

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

Volts

Label

Rating

Location

Protects

+24

F1

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Sample RLC PC board

inprocess queue, motor 1 (P1) sample exit pusher, motor 2 (P1) stat sample pusher, motor 3 (P1) pack to cuvette, motor 0 (P1) not used

+24

F2

Sample Racks LLC PC board Sample Racks LLC PC board Ancillary Probe LLC PC board Ancillary Probe LLC PC board Ancillary Probe LLC PC board Ancillary Probe LLC PC board Sample Probe LLC PC board Sample Probe LLC PC board Sample Probe LLC PC board Sample Probe LLC PC board

+24

F3

+24

F0

+24

F1

+24

F2

vertical, motor 2 (P1)

+24

F3

diluter, motor 3 (P1)

+24

F0

Sample to cuvette, motor 0 (P1) Not used

+24

F1

+24

F2

vertical, motor 2 (P1)

+24

F3

diluter, motor 3 (P1)

LED

Function

CR1 CR2

On PC board red LED On PC board, green LED

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board

NOTE: Green LED lit and red LED not lit indicates normal operation of PC board. Red

LED lit and green LED not lit indicates fault condition on PC board. Red LED lit and green LED flashing indicates reset condition.

078D0080-02, Rev. A

3-16ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide

Backplane PC Board 575252F (slide out electronics module)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide out

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 J14 J15 J16 J17 J18 J19 J20 J21 J22 J23 J24 J25 J26 J27 J28 J29 572301 121151 572298 571411 572302

Sample Probe low level controller PC board 078-B042S01 Ancillary Probe low level controller PC board Sample Racks low level controller PC board Ancillary low level controller PC board Sample Tip low level controller PC board Not Used Reagent Probe 1 low level controller PC board Reagent Probe 2 low level controller PC board Reagent Probe 3 low level controller PC board Reagent Pack low level controller PC board Not Used J101 Fluid GIOB PC board 078-B027S01 Real Time SPARC PC board Sample Handling SC1 sequencer PC board Reagent Handling SC2 sequencer PC board Cuvette Handling SC3 sequencer PC board J101 Sample & Ancillary Queue GIOB PC board S575327D Preheater and Luminometer low level controller PC board Aspirate Probes low level controller PC board Pumps 1 low level controller PC board Pumps 2 low level controller PC board Not Used Not used J101 Sample & Ancillary Probe GIOB PC board S575322H Ring and Elevator Servo low level controller PC board Power Supply Power Connector Not Used J101 Cuvette GIOB PC board S575377F Ethernet

078D0080-02, Rev. A

3-18ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide

Connector

Cable

Connected to . . .

J30 J31 J32 J33 J34 J35 J36 J37 J38 571344 571320 571319 or 119161 572315 572300 121152

Not Used J101 Reagent Handling GIOB PC board S575287C User Interface Module J5 Thermal Control Module SS575257EE External barcode scanner J102 Sample & Ancillary Queue GIOB PC board S575327D J3 Bottom Luminometer Assembly 575367 J3 Servo Motor 575357C NOTE: Use cable part number 119161 only on systems that have white connectors for both J38 on the Backplane PC board and J3 on the Servo Motor Interconnect PC board. J5 Sample & Ancillary Probe GIOB PC board S575322H J1 Sample & Ancillary Queue GIOB S575327D J7 Reagent Handling GIOB PC board S575287C J13 Reagent Handling PC board S575287C Fluid GIOB PC board 078-B027S01 J3 Cuvette GIOB PC board S575377F J101 Reagent Fluid GIOB PC board 078-B027S02 J9 Exit Queue Interconnection PC board 575302C Reagent Pack Barcode Scanner Lab Automation RS-232

J39 J40 J41 J42 J43 J44 J45 J46 J47 J48

571313 571312 571314 571316 571318 571317 572299 571450 571449

Test Point

Sparc

Specifications

TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8

RT RT RT RT RT UI UI UI

12.00 VDC +/ 2% (11.76 to 12.24) 24.00 VDC +/ 3% (23.28 to 24.72) 12.10 VDC +/ 2% (11.86 to 12.34) 5.10 VDC +/ 2% (4.99 to 5.20) Ground 12.30 VDC +/ 2% (12.05 to 12.55) 12.10 VDC +/ 2% (11.86 to 12.34) 5.30 VDC +/ 2% (5.19 to 5.41)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide out

Volts

Label

Rating

Location

Protects

+5V +24V +5V +5V +24V +5V -12V +24V

PTC1 PTC2 PTC3 PTC4 PTC5 PTC6 PTC7 PTC8

0.5A 0.5A 2.5A 2.5A 0.5A 2.5A 2.5A 7A

Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F Backplane PC board 575252F

encoder (J38) luminometer solenoid (J44) sample and ancillary barcode reader (J46) reagent barcode reader (J47) Ionizer (J44) disk enclosure module (J30) not used (J45) Sample and Ancillary Queue GIOB PC board S575327D (J24) Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (J45) Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J17) Fluid GIOB PC board 078-B027S01 (J12) not used bottom luminometer assembly 575367 (J37)

+5V

PTC9

7A

Backplane PC board 575252F

+12V +5V

PTC10 PTC11

2.5A 2.5A

Backplane PC board 575252F Backplane PC board 575252F

078D0080-02, Rev. A

3-20ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide

Volts

Label

Rating

Location

Protects

+5V

PTC12

7A

Backplane PC board 575252F

Sample and Ancillary Queue GIOB PC board S575327D (J24) Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (JJ45) disk enclosure module (J30) Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J31) Fluid GIOB PC board 078-B027S01 (J12) handheld barcode reader (J35) PMT (J37)

+12V +24V

PTC13 PTC14

2.5A 7A

Backplane PC board 575252F Backplane PC board 575252F

+5V +12V

PTC15 PTC16

2.5A 2.5A

Backplane PC board 575252F Backplane PC board 575252F

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed

Backplane PC Board S117937D (fixed electronics module)

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9

Sample Probe low level controller PC board Ancillary Probe low level controller PC board Sample Racks low level controller PC board Ancillary low level controller PC board Sample Tip low level controller PC board Servo Decoder Reagent Probe 1 low level controller PC board Reagent Probe 2 low level controller PC board Reagent Probe 3 low level controller PC board

078D0080-02, Rev. A

3-22

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D

Connector

Cable

Connected to . . .

J10 J11 J12 J13 J14 J15 J16 J17 J18 J19 J20 J21 J22 J23 J24 J25 J26 J27 J28 J29 J30 J31 J32 J33 J34 J35 J36 J37 J38 J39 571313 572315 572328 575344 571320 572300 121151 121152 572298 571411 121152 572302

Reagent Pack low level controller PC board User Interface Module J101 Fluid GIOB PC board 078-B027S01 Real Time SPARC PC board Sample Handling SC1 sequencer PC board Reagent Handling SC2 sequencer PC board Cuvette Handling SC3 sequencer PC board J101 Sample & Ancillary Queue GIOB PC board S575327D Preheater and Luminometer low level controller PC board Aspirate Probes low level controller PC board Pumps 1 low level controller PC board Pumps 2 low level controller PC board Not Used User Interface Module J101 Sample & Ancillary Probe GIOB PC board S575322H Ring and Elevator Servo low level controller PC board Power Supply Power Connector Not Used J101 Cuvette GIOB PC board S575377F Ethernet Not Used J101 Reagent Handling GIOB PC board S575287C Power Supply Sense and Control Connector Not Used J5 Thermal Control Module S575257E External barcode scanner J102 Sample & Ancillary Queue GIOB PC board S575327D J3 Bottom Luminometer Assembly 575367 Servo Motor J5 Sample & Ancillary Probe GIOB PC board S575322H

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed

Connector

Cable

Connected to . . .

J40 J41 J42 J43 J44 J45 J46 J47 J48

571312 571314 571316 571318 571317 572299 57145 571449 572327

J1 Sample & Ancillary Queue GIOB S575327D J7 Reagent Handling GIOB PC board S575287C J13 Reagent Handling PC board S575287C Fluid GIOB PC board 078-B027S01 J3 Cuvette GIOB PC board S575377F J101 Reagent Fluid GIOB PC board 078-B027S02 J9 Exit Queue Interconnection PC board 575302C Reagent Pack Barcode Scanner Lab Automation RS-232

Test Point

Sparc

Specifications

TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8

RT RT RT RT RT UI UI UI

-12.13 VDC +/- 4% (-11.65 to -12.61) 24.20 VDC +/- 3% (23.47 to 24.93) 12.12 VDC +/- 3% (11.76 to 12.48) 5.18 VDC +/- 1.5% (5.10 to 5.25) Ground -12.16 VDC +/- 2% (-11.91 to -12.40) 12.10 VDC +/- 2% (11.85 to 12.34) 5.075 VDC +/- 2% (4.97 to 5.18)

Volts

Label

Rating

Location

Protects

+5V +24V +5V +5V +24V -12V

PTC1 PTC2 PTC3 PTC4 PTC5 PTC7

0.5A 0.5A 2.5A 2.5A 0.9A 2.5A

Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D

encoder (J38) luminometer solenoid (J44) sample and ancillary barcode reader (J46) reagent barcode reader (J47) Ionizer (J44) not used (J45)

078D0080-02, Rev. A

3-24

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D

Volts

Label

Rating

Location

Protects

+24V

PTC8

7A

Backplane PC board S117937D

Sample and Ancillary Queue GIOB PC board S575327D (J24) Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (J45) Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J17) Fluid GIOB PC board 078-B027S01 (J12) not used bottom luminometer assembly 575367 (J37) Sample and Ancillary Queue GIOB PC board S575327D (J24) Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (JJ45) Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J31) Fluid GIOB PC board 078-B027S01 (J12) handheld barcode reader (J35) PMT (J37) J50 Cardcage Fan J50 Cardcage Fan

+5V

PTC9

7A

Backplane PC board S117937D

+12V +5V +5V

PTC10 PTC11 PTC12

2.5A 2.5A 7A

Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D

+24V

PTC14

7A

Backplane PC board S117937D

+5V +12V +24V +24V

PTC15 PTC16 PTC17 PTC18

2.5A 2.5A 2.5A 2.5A

Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D Backplane PC board S117937D

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed

Volts

Label

Rating

Location

Protects

+24V

F1

10ASB

Backplane PC board S117937D

Vacuum Pump

078D0080-02, Rev. A

3-26 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board

Thermal Control Module PC Board for the ADVIA Centaur System S575257E

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J7 J8

571374 572311 572312 572313 572315 571360 572265

Primary Reagent Compartment TEDs, heat sink fan, circulation fan Ancillary Queue TEDs, heat sink fan Preheater TEDs and PMT TEDs Incubation Chamber TEDs J34 of Backplane PC board 575252F J2 of Cuvette GIOB PC board S575377F J14 of Mixer Interconnect PC board 575422C and J12 of Sample and Ancillary Queue GIOB PC board S575327D Power Supply 571258

J9

571352

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for

LED

Function

CR42 CR43 CR44 CR45 CR65 CR67 CR68 CR69 CR70 CR71 CR72

On PC board, red LED (lit when ancillary compartment starts cooling) On PC board, green LED (blinking is normal during running) On PC board, red LED (inhibit, power supply is out of tolerance low, watchdog trip) On PC board, red LED (lit when reagent compartment starts cooling) On PC board, LED, P2 sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Preheater sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring D sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring C sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, PMT1 sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring A sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring B sign bit for bi-polar loops (lit = heating; not lit = cooling)

Volts

Label

Rating

Location

Protects

+24V

F1

6.3A 250V 5x20 mm SB 575194

Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

primary reagent compartment (front of instrument) thermal electric device (J1) primary reagent compartment (middle of instrument) thermal electric device (J1) ring B thermistor via Cuvette GIOB PC board S575377F (J7) ring A thermistor via Cuvette GIOB PC board S575377F (J7) ring C thermistor via Cuvette GIOB PC board S575377F (J7)

+24V

F2

6.3A 250V 5x20 mm SB 575194

+24V

F3

5A 250V 5x20 mm SB 60044x14 5A 250V 5x20 mm SB 60044x14 5A 250V 5x20 mm SB 60044x14

+24V

F4

+24V

F5

078D0080-02, Rev. A

3-28 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board

Volts

Label

Rating

Location

Protects

+24V

F6

5A 250V 5x20 mm SB 60044x14 6.3A 250V 5x20 mm SB 575194

Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

ring D thermistor via Cuvette GIOB PC board S575377F (J7) primary reagent compartment (middle of instrument) thermal electric device (J1) ancillary refrigerator thermal electric device (J2) preheater thermal electric devices (J3) PMT 1 thermal electric device (J3) PMT 2 thermal electric device (J3) ancillary heatsink fan and fan wiring (J2) reagent heatsink fan and fan wiring (J1) reagent circulation fan and fan wiring (J1)

+24V

F7

+24V

F8

6.3A 250V 5x20 mm SB 575194

+24V

F9

5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 0.9A Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

+24V

F10

+24V

F11

+24V

PTC2

+24V

PTC

525 1.35A

+24V

PTC1

0.9A

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for

Thermal Control Module PC Board for the ADVIA Centaur XP System 078-B111SXX

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J7 J8

571374 572311 572312 572313 572315 571360 572265

Primary Reagent Compartment TEDs, heat sink fan, circulation fan Ancillary Queue TEDs, heat sink fan Preheater TEDs and PMT TEDs Incubation Chamber TEDs J34 of Backplane PC board 575252F J2 of Cuvette GIOB PC board S575377F J14 of Mixer Interconnect PC board 575422C and J12 of Sample and Ancillary Queue GIOB PC board S575327D Power Supply 571258

J9

571352

078D0080-02, Rev. A

3-30 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board

LED

Function

CR42 CR43 CR44 CR45 CR65 CR67 CR68 CR69 CR70 CR71 CR72

On PC board, red LED (lit when ancillary compartment starts cooling) On PC board, green LED (blinking is normal during running) On PC board, red LED (inhibit, power supply is out of tolerance low, watchdog trip) On PC board, red LED (lit when reagent compartment starts cooling) On PC board, LED, P2 sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Preheater sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring D sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring C sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, PMT1 sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring A sign bit for bi-polar loops (lit = heating; not lit = cooling) On PC board, LED, Ring B sign bit for bi-polar loops (lit = heating; not lit = cooling)

Volts

Label

Rating

Location

Protects

+24V

F1

6.3A 250V 5x20 mm SB 575194

Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

primary reagent compartment (front of instrument) thermal electric device (J1) primary reagent compartment (middle of instrument) thermal electric device (J1) ring B thermistor via Cuvette GIOB PC board S575377F (J7) ring A thermistor via Cuvette GIOB PC board S575377F (J7) ring C thermistor via Cuvette GIOB PC board S575377F (J7)

+24V

F2

6.3A 250V 5x20 mm SB 575194

+24V

F3

5A 250V 5x20 mm SB 60044x14 5A 250V 5x20 mm SB 60044x14 5A 250V 5x20 mm SB 60044x14

+24V

F4

+24V

F5

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for

Volts

Label

Rating

Location

Protects

+24V

F6

5A 250V 5x20 mm SB 60044x14 6.3A 250V 5x20 mm SB 575194

Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

ring D thermistor via Cuvette GIOB PC board S575377F (J7) primary reagent compartment (middle of instrument) thermal electric device (J1) ancillary refrigerator thermal electric device (J2) preheater thermal electric devices (J3) PMT 1 thermal electric device (J3) PMT 2 thermal electric device (J3) ancillary heatsink fan and fan wiring (J2) reagent heatsink fan and fan wiring (J1) reagent circulation fan and fan wiring (J1)

+24V

F7

+24V

F8

6.3A 250V 5x20 mm SB 575194

+24V

F9

5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 5A 250V 5x20 Thermal Control mm SB 60044x14 Module S575257E 0.9A Thermal Control Module S575257E Thermal Control Module S575257E Thermal Control Module S575257E

+24V

F10

+24V

F11

+24V

PTC2

+24V

PTC

525 1.35A

+24V

PTC1

0.9A

078D0080-02, Rev. A

3-32

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Low Level Controller PC

Servo Low Level Controller PC Board S575262C

There are three fuses on the PC board; their functions are described below:
Volts Label Rating Location Protects

+24V

F0

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Ring and Elevator Servo LLC PC board S575262C Ring and Elevator Servo LLC PC board S575262C Ring and Elevator Servo LLC PC board S575262C

index pin, motor 0 (P1) cuvette elevator, motor 1 (P1) ring, servo 0 (P1)

+24V

F1

+24V

F2

Connector

Cable

Connected to . . .

P1 J3 J5

Plugs into J25 of Backplane PC board 575252F Expansion PC board I/O Expansion PC board I/O

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Low Level Controller PC Board

LED

Function

CR1 CR2

On PC board, red LED On PC board, green LED LED lit and green LED not lit indicates fault condition on PC board. Red LED lit and green LED flashing indicates reset condition.

NOTE: Green LED lit and red LED not lit indicates normal operation of PC board. Red

078D0080-02, Rev. A

3-34

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Fluid GIOB PC Board S575272E

Fluid GIOB PC Board S575272E

Connector

Cable

Connected to . . .

J10 J11 J12 J13 J14 J15 J16 J17 J18 J19 J101 J102

571318 571328 571327 571321 571322 571351 571350 571349 572314 571304 572302 571457

J43 Backplane PC board 575252F Diluter and FMI Pump Motors and Switches Diluter and FMI Pump Motors and Switches J11 Wash Fluid Interconnect PC board 575277B J12 Wash Fluid Interconnect PC board 575277B Manifold Valves Manifold Valves Pinch Valves Slotted sensor (cleaning solution Lid) Capacitive Level sensor (cleaning solution) J12 Backplane PC board 575252F Reagent Fluid GIOB PC board 078-B027S02

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Fluid GIOB PC Board S575272E

3-35

Connector

Cable

Connected to . . .

J103

Not Used

Volts

Label

Rating

Location

Protects

+24V

F1

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 0.9A

Fluid GIOB PC board

PROT +24A to valve drive circuitry (J13 and J15) PROT +24B to valve drive circuitry (J16 and J17) PROT VCC to diluters (J11 and J12) and to wash fluid interconnect PC board (J13 and J14)

+24V

F2

Fluid GIOB PC board

+24V

PTC1

Fluid GIOB PC board

078D0080-02, Rev. A

3-36 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Wash Fluid Interconnect PC Board

Wash Fluid Interconnect PC Board 575277B

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 J14 J15 J16 J17 571321 571322

Wash 2 Lid sensor Wash 1 Lid sensor Acid Level sensor Base Lid sensor Base Level sensor Acid Lid sensor Wash 1 Level sensor Wash 3 Lid sensor Wash 2 Level sensor Wash 3 Level sensor J13 of Fluid GIOB PC board J14 of Fluid GIOB PC board Wash 2 Vent Acid Vent Wash 3 Vent Wash 1 Vent Base Vent

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Probe Vertical PC Board

Ancillary Probe Vertical PC Board S575282C

Connector

Cable

Connected to . . .

J1 J2 J3
Sensor

571283

J8 Sample and Ancillary Probe GIOB PC board S575322H Ancillary Probe Vertical motor Ancillary Probe Volume Detector 571954

Function

U4

Ancillary Vertical Home

078D0080-02, Rev. A

3-38 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Handling GIOB PC Board

Reagent Handling GIOB PC Board S575287C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J12 J13

571297 571300 571299

J1 of Reagent Probe 1 Pack-to-Pack PC board 575347 J1 of Reagent Probe 2 Pack-to-Pack PC board 575347 J1 of Reagent Probe 3 Pack-to-Pack PC board 575347 Reagent Probe 1 Pack-to-Pack motor Reagent Probe 2 Pack-to-Pack motor Reagent Probe 3 Pack-to-Pack motor

571314 571298 571301 571302 571311 571316

J41 of Backplane PC board 575252F J3 of Reagent Probe 1 Heater Control/Volume Detect/ Probe Interconnect PC board 575402B J3 of Reagent Probe 2 Heater Control/Volume Detect/ Probe Interconnect PC board 575407C J3 of Reagent Probe 3 Heater Control/Volume Detect/ Probe Interconnect PC board 575412C J1 Reagent Probe 1 Diluter, J1 Reagent Probe 2 Diluter, and J1 Reagent Probe 3 Diluter, J42 of Backplane PC board 575252F

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Handling GIOB PC Board

Connector

Cable

Connected to . . .

J14 J15 J16 J18 J19 J22 J101 J102 J103 571341 571366 571367 571447 572300 571455 571456

Reagent Barcode Scanner motor Reagent Barcode Scanner Home sensor Reagent Barcode Scanner End of Travel sensor J10 of Reagent Mixer Interconnect PC board J1 Upper Reagent Pack LED Indicators PC board 575387 J1 Lower Reagent Pack LED Indicators PC board 575387 J28 of Backplane PC board 575252F (Reagent Power) J103 of Cuvette GIOB PC board S575377F J102 of Fluid GIOB PC board

LED

Function

CR1 CR2 CR3 CR4 CR5 CR6

Reagent Probe 1 Heater Status (Green LED indicates the temperature of the probe is in tolerance) Reagent Probe 2 Heater Status (Green LED indicates the temperature of the probe is in tolerance) Reagent Probe 3 Heater Status (Green LED indicates the temperature of the probe is in tolerance) Reagent Probe 1 Volume Check (Amber LED indicates that air is in front of the volume detector) Reagent Probe 2 Volume Check (Amber LED indicates that air is in front of the volume detector) Reagent Probe 3 Volume Check (Amber LED indicates that air is in front of the volume detector)

Volts

Label

Rating

Location

Protects

+24V

F2

2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808 2A 250V 5x20 mm FB 572808

Reagent Handling GIOB PC board S575287C Reagent Handling GIOB PC board S575287C Reagent Handling GIOB PC board S575287C

reagent probe 1 heater drive (J8) reagent probe 2 heater drive (J9) reagent probe 3 heater

+24V

F3

+24V

F4

078D0080-02, Rev. A

3-40 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Rack Present PC Board S575292B

Rack Present PC Board S575292B

Connector

Cable

Connected to . . .

Direct Direct Direct


Sensor Function

J6 of Exit Queue PC board 575302C J4 of Stat Area PC board 575307C J5 of Stat Area PC board 575307C

U1 U1 U1

External Queue Full (J6 of 575302C) Normal Rack Present (J4 of 575307C) Stat Rack Present (J5 of 575307C)

NOTE: Same PC board used three places.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Fluid GIOB PC Board

Reagent Fluid GIOB PC Board 078-B027S02

Table 3-1

Connector

Cable

Connected to . . .

J2 J7 J8 J15 J101 J102 J103 P1

572305 572306 572303 571348 572299 571456 571457 571764

+24V from power block Water pump J5 of System fluid sensor PC board 575452C Reagent Fluids Manifold valves Power from J45 on Backplane PC board 575252F J103 on Reagent Handling GIOB PC board S575287C J102 on Fluid GIOB PC board Water pressure transducer

078D0080-02, Rev. A

3-42

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Fluid GIOB PC Board

Connector

Cable

Connected to . . .

P8
Volts Label

572304
Rating

Vacuum pump enable


Location Protects

+24V

F1

2A 250V 5x20 mm FB 572808

Reagent Fluid GIOB PC board 078-B027S02 Reagent Fluid GIOB PC board J8 078-B027S02 Reagent Fluid GIOB PC board J8 078-B027S02

valve drive circuitry for all pressurized manifold valves and pinch valves (J15) System Fluid Sensor PC board 575452C (see F1 on this PC board) (J8) water pump (J7)

+24V

PTC1

1.35A

+24V

PTC2

1.35A

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Exit Queue Interconnect PC Board

Exit Queue Interconnect PC Board 575302C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10

571285 571286 575012 575011 571292 571287

J3 Sample and Ancillary Queue GIOB PC board S575327D J4 Sample and Ancillary Queue GIOB PC board S575327D Female on 575302C (J3) Male on 575307C (J2) J1 Ancillary Queue PC board 575222B J1 Ancillary Entry PC board 575232D Exit Queue Full sensor PC board S575292B Spare motor Spare sensor

571450

J46 of Backplane PC board 575252F Ancillary Barcode Head 571259

078D0080-02, Rev. A

3-44

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Stat Area PC Board 575307C

Stat Area PC Board 575307C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12

575012 575011 571291

J3 of Interconnect Exit Queue 575302C (Female Connector) Male connector J1 on this PC board Sample Barcode Reader 571259 Entry Queue PC board 575242B Normal Rack Present PC board S575292B Stat Rack Present PC board S575292B Stat Pusher Home sensor Stat Pusher Position sensor

571342 110334 110333

Rack loader PC board 575237B Start Stop PC board 575312 Stat Pusher Motor Rack Loader Motor Entry Queue Motor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probe GIOB PC

Sample and Ancillary Probe GIOB PC Board S575322H

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11

572947

Sample Level Sense PC board 575462E Sample Pump 3-way valve Bleed 3-way valve Sample Pump Drive

571313 571280 571284 571283 571281

J39 of Backplane PC board 575252F J1 Sample and Ancillary Horizontal PC board 575212C J1 Sample Vertical PC board 575227 J1 Ancillary Vertical PC board S575282C J1 Ancillary Diluter and J1 Sample Diluter Sample Probe Horizontal motor Ancillary Probe Horizontal motor

078D0080-02, Rev. A

3-46 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probe GIOB

Connector

Cable

Connected to . . .

J12 J13 J14 J101 J102 J103

571290

Sample Diluter motor Spare Sample motor Spare Ancillary motor

572298 571453 571454

J17 of Backplane PC board 575252F J103 of Sample/Ancillary Queue PC Board S575327D J102 Cuvette GIOB S575377F

Volts

Label

Rating

Location

Protects

VCC

PTC1

1.35A

Sample and Ancillary Probe GIOB PC board S575322H

Sample Level Sense PC board 575462E (J1) Sample Probe Vertical Sensor PC board 575227 (J7) Auxiliary Probe Vertical PC board S575282C (J8) ancillary diluter, sample diluter (J9) Sample Level Sense PC board 575462E (J1) pump valve (J2) bleed valve (J3) level sense pump (J4) Sample and Ancillary Probes Horizontal PC board 575212C (J6)

+24V

PTC2

1.85A

Sample and Ancillary Probe GIOB PC board S575322H

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Queue GIOB PC

Sample and Ancillary Queue GIOB PC Board S575327D

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 J14

571312 571285 571286 571330 571296 571294

J40 Backplane PC board 575252F Spare sensor J1 of Sample Interconnect PC board 575302C J2 of Sample Interconnect PC board 575302C J1 of Sample Tip Loader PC board 575352B J1 of Sample Queue PC board 575217B Inprocess Queue motor Exit Pusher Motor Spare motor Exit Pusher Home sensor Exit Pusher Position sensor

572265 571282 572297

J8 of Thermal Control PC board S575257E Cuvette Loader motor Stop Action switch

078D0080-02, Rev. A

3-48 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Queue GIOB

Connector

Cable

Connected to . . .

J15 J16 J101 J102 J103

571288 571288 571411 571453 571344

Cuvette Loader Up/Home sensor Cuvette Loader Down sensor J24 of Backplane PC board 575252F J102 Sample and Ancillary Probe GIOB PC board S575322H J36 of Backplane PC board 575252F

Volts

Label

Rating

Location

Protects

VCC

PTC1

1.85A

Sample and Ancillary Queue GIOB PC board S575327D

spare sensor (J2) Exit Queue Interconnect PC board 575302C (J3and J4) Sample Tip Loader PC board 575352B (J5) Inprocess Queue Tip Sensor PC board 575217B (J6) exit pusher home sensor (J10) exit pusher position sensor (J11) ring loader motor (J13) ring loader home sensor (J15) ring loader sensor (J16) Exit Queue Interconnect PC board 575302C (J3) Inprocess Queue Tip Sensor PC board 575217B (J6)

+24V

PTC2

1.35A

Sample and Ancillary Queue GIOB PC board S575327D

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette Elevator PC Board 575337 3-49

Cuvette Elevator PC Board 575337

Connector

Cable

Connected to . . .

J1 J2 571308

Cuvette Motor of Cuvette GIOB PC board S575377F

Sensor

Function

U1 U2

Elevator lower End of Travel sensor Elevator upper End of Travel sensor

078D0080-02, Rev. A

3-50

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Top Luminometer PC Board

Top Luminometer PC Board 575342C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 571307

Luminometer motor Waste Probe motor Waste Probe Home sensor J5 Bottom Luminometer PC board 575367

Sensor

Function

U1 U2 U3

Luminometer Direction sensor Luminometer Position sensor Luminometer Home sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe Pack-to-Pack PC Board

Reagent Probe Pack-to-Pack PC Board 575347

Connector

Cable

Connected to . . .

J1
Sensor

571297
Function

J1 Reagent Handling GIOB PC board S575287C

U1 U2 U3

End of Travel sensor Position sensor Direction sensor

078D0080-02, Rev. A

3-52

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Tip Loader PC Board

Sample Tip Loader PC Board 575352B

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9

571330 110854

J5 Sample and Ancillary GIOB PC board S575327D Tip Tray and Band PC board S575397C (Tray Present) Sample Tip Loader Door sensor Sample Tip Loader Home sensor Sample Tip Loader Position sensor Sample Tip Tray High sensor Tip Tray and Band PC board S575397C (Band Present) Sample Tip Loader motor

572296

Stop Action switch

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Motor Interface PC Board

Servo Motor Interface PC Board 575357C

Connector

Cable

Connected to . . .

J1 J2 J3 571319 or 119161

Servo motor Encoder J38 of Backplane PC board 575252F NOTE: Use cable part number 119161 only on systems that have white connectors for both J38 on the Backplane PC board and J3 on the Servo Motor Interconnect PC board

078D0080-02, Rev. A

3-54

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Aspirate Probe PC Board 575362

Aspirate Probe PC Board 575362

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 571286

Aspirate Probe 1 motor Aspirate Probe 2 motor J13 of Cuvette GIOB PC board S575377F Aspirate Probe 3 motor Aspirate Probe 4 motor Aspirate Probe 1 Home sensor Aspirate Probe 2 Home sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Aspirate Probe PC Board 575362

3-55

Connector

Cable

Connected to . . .

J8 J9 J10 J11 J12 J13

Aspirate Probe 3 Home sensor Aspirate Probe 4 Home sensor Index Mechanism motor Index Mechanism Pin Up sensor Index Mechanism Pin Down sensor Incubation Chamber Ring Tooth sensor

078D0080-02, Rev. A

3-56

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Bottom Luminometer PC Board

Bottom Luminometer PC Board 575367

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7

112254 112254 571320 571309 571307

PMT 1 PMT 2 J37 Backplane PC board 575252F J6 of Cuvette GIOB PC board S575377F J5 Top Luminometer PC board 575342C Thermistor PMT 1 Thermistor PMT 2

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Preheater PC Board 575372

3-57

Preheater PC Board 575372

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9 571310

Preheater Pusher motor Cuvette Conveyor motor (cuvette loader) Reflective Chute Empty sensor Reflective Chute Full sensor Cuvette Present sensor Preheater thermistor J5 Cuvette GIOB PC board S575377F Preheat Pusher In sensor (home) Preheat Pusher Out sensor (end of travel)

078D0080-02, Rev. A

3-58 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F

Cuvette GIOB PC Board S575377F

Connector

Cable

Connected to . . .

J1 J2

116810 571360

J5 of Fluids Heater Control PC board J7 of Thermal Module PC board S575257E

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F 3-59

Connector

Cable

Connected to . . .

J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 J14 J15 J16 J101 J102 J103

571317 571308 571310 571309 571362

J44 of Backplane PC board 575252F J2 of Cuvette Elevator PC board 575337 J7 of Preheater PC board 575372 J4 of Bottom Luminometer PC board 575367 Ring A Thermistor 1 Incubation Chamber Ring B Thermistor 2 Incubation Chamber Ring C Thermistor 3 Incubation Chamber Ring D Thermistor 4 Incubation Chamber Home sensor not used

571286 571357 571358 572267 572301 571454 571455

J3 of Wash Station and Index Mechanism PC board Tip Waste Bin sensor Cuvette Waste Bin sensor Ionizer J31 of Backplane PC board 575252F (GIOB Power) J103 Sample Probe GIOB PC board S575322H J102 Reagent Handling GIOB PC board S575287C

Volts

Label

Rating

Location

Protects

+5V

PTC1

1.85A

Cuvette GIOB PC board S575377F

PRT-VCC to: bottom luminometer (J6) cuvette lift (J4) preheater (J5) ring home sensor (J11) ring EOT sensor (J12) Wash Station and Index Mechanism PC board (J13) tip waste bin (J14) cuvette waste bin (J15)

LED

Function

CR1 CR2 CR3

Green LED indicates the acid probe heater is working Green LED indicates the base probe heater is working Green LED indicates the system fluid heater is working

078D0080-02, Rev. A

3-60 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F

LED

Function

CR4 CR5 CR101

Not Used Not Used Red LED indicates a parity error or the system is in the reset mode

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Pack LED Array PC Board

Reagent Pack LED Array PC Board 575387

Connector

Cable

Connected to . . .

J1 J1

571367 571447

J19 on Reagent Handling GIOB S575287C PC board (Upper LED Display) J22 on Reagent Handling GIOB PC Board S575287C (Lower LED Display)

NOTE: Same PC board used two places. NOTE: LEDs are numbered from left to right, CR1 through CR30.

LED

Function

CR1 CR2 CR3 CR4 CR5 CR6 CR7 CR8

Green LED, column 0, row o Green LED, column 0, row 1 Green LED, column 0, row 2 Green LED, column 0, row 3 Green LED, column 0, row 4 Green LED, column 0, row 5 Green LED, column 0, row 6 Green LED, column 0, row 7

078D0080-02, Rev. A

3-62

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry LEDs PC Board

Ancillary Entry LEDs PC Board 575392B

Connector

Cable

Connected to . . .

Hard Wired

J6 Ancillary Entry PC board 57523

LED

Function

CR1 CR2

Ancillary Pack Yellow LED (from J6 of 575232D) Ancillary Pack Green LED (from J6 of 575232D)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Tip Tray and Band PC Board S575397C

Tip Tray and Band PC Board S575397C

Connector

Cable

Connected to . . .

Direct Wired Direct Wired

27146 27146

J2 Tip Loader PC board 575352B (Tray Present) J7 Tip Loader PC board 575352B (Band Present)

NOTE: Same PC board used twice. Sensor Function

U1 U1

Tip Tray Present (J2 on 575352B) Tip Band Present (J7 of 575352B)

078D0080-02, Rev. A

3-64ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 1 Heater Control and

Reagent Probe 1 Heater Control and Volume Detect Probe Interconnect PC Board 575402B

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 571372 571298 571954

Reagent Probe 1 Horizontal motor Reagent Probe 1 Vertical motor J8 Reagent Handling GIOB PC board S575287C Probe 1 Volume Detector Probe 1 Heater Coil Horizontal Slotted Sensors PC board 575432B Vertical Slotted Sensors

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 2 Heater Control and

Reagent Probe 2 Heater Control and Volume Detect Probe Interconnect PC Board 575407C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 571372 571301 571954

Probe 2 Horizontal motor Probe 2 Vertical motor J9 Reagent Handling GIOB PC board S575287C Probe 2 Volume Detector Probe 2 Heater coil Horizontal Slotted Sensors PC board 575437 Vertical Slotted sensors

078D0080-02, Rev. A

3-66ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 3 Heater Control and

Reagent Probe 3 Heater Control and Volume Detect Probe Interconnect PC Board 575412C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 571372 571302 571954

Probe 3 Horizontal Motor Probe 3 Vertical Motor J10 Reagent Handling GIOB PC board S575287C Volume Detector 571954 Heater Coil 571235 Horizontal Slotted Sensor PC board 575442B Vertical Slotted sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Mixer Interconnect PC Board 575422C

Mixer Interconnect PC Board 575422C

Figure 3-1 Connector

Mixer Interconnect PC Board 575422C Cable Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J10 J11 J12 J13 J14 J15 J16 J17 J18 J19 572265 571366 571326

Reagent Probe Shutter motor Reagent Probe Shutter Open sensor Reagent Probe Shutter Closed sensor Primary Compartment Door tilt switch Reagent Mixer motor Reagent Mixer Upper sensor Reagent Mixer Middle sensor Reagent Mixer Lower sensor J18 Reagent Handling GIOB PC board S575287C Bar Code Scanner Shutter Closed sensor Bar Code Scanner Shutter Open sensor Bar Code Scanner Shutter Motor sensor J8 Temperature Control Module PC board S575257E Primary Reagent Compartment, Loop A thermistor Primary Reagent Compartment, Loop B thermistor Primary Reagent Compartment, Loop C thermistor Spare Spare

078D0080-02, Rev. A

3-68

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 1 Horizontal

Reagent Probe 1 Horizontal Interconnect PC Board 575432B

Connector

Cable

Connected to . . .

J1

571372

J6 Heater Control and Volume Detect Probe Interconnect PC board 575402B

Sensor

Function

U1 U2 U3

Horizontal Position sensor Horizontal Direction sensor Horizontal End of Travel sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 2 Horizontal Interconnect

Reagent Probe 2 Horizontal Interconnect PC Board 575437

Connector

Cable

Connected to . . .

J1

571372

J6 Heater Control and Volume Detect Probe Interconnect PC board 575407C

Sensor

Function

U1 U2 U3

Horizontal Position sensor Horizontal Direction sensor Horizontal End of Travel sensor

078D0080-02, Rev. A

3-70

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 3 Horizontal

Reagent Probe 3 Horizontal Interconnect PC Board 575442B

Connector

Cable

Connected to . . .

J1

571372

Heater Control and Volume Detect Probe Interconnect PC board 575412C

Sensor

Function

U1 U2 U3

Horizontal Position sensor Horizontal Direction sensor Horizontal End of Travel sensor

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reflective Chute Sensor Assembly PC

Reflective Chute Sensor Assembly PC Board 575447B

Connector

Cable

Connected to . . .

Direct wired Direct wired Same assembly used twice.


Sensor Function

J3 Preheater Assembly 575372 J4 Preheater Assembly 575372

U1 U1

Chute empty (J3 on 575372) Chute full (J4 on 575372)

078D0080-02, Rev. A

3-72

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 System Fluid Detect PC Board

System Fluid Detect PC Board 575452C

Connector

Cable

Connected to . . .

J1 J2 J3 J4 J5 J6 J7 J8 J9
LED Function

Water Transfer Pump motor Water Overflow sensor Waste Working sensor Waste Bulk Overflow sensor 572303 J8 Reagent Fluid GIOB PC board 078-B027S02 Water Lid sensor Waste Lid sensor Water Low sensor Waste Transfer Pump motor

CR2 CR3 CR4 CR5

Water Overflow Water Low Waste Working Waste Overflow

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 System Fluid Detect PC Board 575452C

Volts

Label

Rating

Location

Protects

+24V

F1

2A 250V 5x20 mm FB 572808

System Fluid Sensor/Pump PC board 575452C

Reagent Fluid GIOB PC board 078-B027S02 (J5) water transfer pump (J1) and waste transfer pump circuitry (J9)

078D0080-02, Rev. A

3-74

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Pack Sensor PC Board

Ancillary Pack Sensor PC Board 575457B

Connector

Cable

Connected to . . .

Direct Wired

27146 J7

Ancillary Entry PC board 575232D (Ancillary Pack Present)

Sensor

Function

U1

Ancillary pack present (J7 on 575232D)

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Pressure Level Sense PC Board

Sample Pressure Level Sense PC Board 575462E

Connector

Cable

Connected to . . .

J1

572947

J1 Sample and Ancillary Probe GIOB PC board S575322H

Sensor

Function

U4

Pressure sensor

078D0080-02, Rev. A

Appendix A: Safety Instructions


This information summarizes the established guidelines for handling laboratory biohazards. This summary is based on the guidelines developed by the Centers for Disease Control, the Clinical and Laboratory Standards Institute Document M29-A3, Protection of Laboratory Workers from Occupationally Acquired Infections, and the Occupational Safety and Health Administrations Bloodborne Pathogens Standard.13

Protecting Yourself from Biohazards


Use this summary for general information only. It is not intended to replace or supplement your laboratory or hospital biohazard control procedures. By definition, a biohazardous condition is a situation involving infectious agents biological in nature, such as the hepatitis B virus, the human immunodeficiency virus, and the tuberculosis bacterium. These infectious agents may be present in human blood and blood products and in other body fluids. The following are the major sources of contamination when handling potentially infectious agents: needlesticks hand-to-mouth contact hand-to-eye contact direct contact with superficial cuts, open wounds, and other skin conditions that may permit absorption into subcutaneous skin layers splashes or aerosol contact with skin and eyes

To prevent accidental contamination in a clinical laboratory, strictly adhere to the following procedures: Wear gloves while servicing parts of the instrument that have contact with body fluids such as serum, plasma, urine, or whole blood. Wash your hands before going from a contaminated area to a noncontaminated area, or when you remove or change gloves. Perform procedures carefully to minimize aerosol formation. Wear facial protection when splatter or aerosol formation are possible. Wear personal protective equipment such as safety glasses, gloves, lab coats or aprons when working with possible biohazard contaminants. Keep your hands away from your face.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode

Cover all superficial cuts and wounds before starting any work. Dispose of contaminated materials according to your laboratorys biohazard control procedures. Keep your work area disinfected. Disinfect tools and other items that have been near any part of the instrument sample path or waste area with 10% v/v bleach. Do not eat, drink, smoke, or apply cosmetics or contact lenses while in the laboratory. Do not mouth pipet any liquid, including water. Do not place tools or any other items in your mouth. Do not use the biohazard sink for personal cleaning such as rinsing coffee cups or washing hands.

To prevent needlestick injuries, needles should not be recapped, purposely bent, cut, broken, removed from disposable syringes, or otherwise manipulated by hand.

References
1. Centers for Disease Control. 1988. Update: Universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus and other bloodborne pathogens in healthcare settings. MMWR, 37:377382, 387, 388. 2. Clinical and Laboratory Standards Institute (formerly NCCLS). Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline Third Edition. CLSI Document M29-A3.[ISBN 1-56238-567-4]. Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2005). 3. Federal Occupational Safety and Health Administration. Bloodborne Pathogens Standard. 29 CFR 1910. 1030.

Protecting Yourself from Barcode Scanner Lasers


To avoid damage to the eyes, never look directly at the laser beam or at its reflection from a shiny surface. Never point a hand-held barcode scanner at anyone.

Laser Safety Classification of the ADVIA Centaur System


During normal operation with all of the protective housings in place, the ADVIA Centaur system is classified as CDRH Class I and EN 60825-1 Class 1. No direct exposure to laser hazard exists for persons in the immediate area.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode

.
Figure A-1 CDRH Class 1 Label

Some field service procedures require the removal of the protective housings that prevent human access to the laser radiation. The removal of the protective housings may change the classification of the system to CDRH Class 2 and EN60825-1 Class 2. All field service procedures must be followed precisely. Only Bayer HealthCare-trained field service personnel should perform procedures related to laser assemblies.

Figure A-2

CDRH Class 2 Label

The laser labels are positioned on the instrument as shown below.

Laser Safety Classification of the Barcode Scanners


The laser safety classification of the reagent and sample barcode scanners when they are unprotected by the system housings is CDRH Class 2 and EN 60825-1 Class 2. The laser safety classification of the hand-held barcode scanner used with the system is CDRH Class 2 and EN 60825-1 Class 2.

Reagent and Sample Barcode Scanners


ADVIA Centaur XP Operators Guide procedures for testing the reagent or sample barcode scanners contain the following laser warning:

LASER WARNING: To avoid damage to the eyes, never look directly at the laser beam or

at its reflection from a shiny surface. Only Bayer-trained field service personnel should perform procedures related to laser assemblies. Refer to Appendix A, Protecting Yourself from Barcode Scanner Lasers, for more information.

078D0080-02, Rev. A

The specifications for the laser optical assemblies in the ADVIA Centaur ancillary reagent, primary reagent, and sample barcode scanners are summarized in the following table:
Characteristic Specification

Maximum Power Output Wavelength Pulse Duration Units of Beam Divergence

1.0 mW 670 nm Continuous Wave (cw) 0.7 mr

The locations of the reagent and sample barcode scanners and their associated laser safety labels are shown below. The laser apertures for the sample and ancillary reagent barcode scanners are directed into the system away from the operator. During normal operation, reflections from the sample barcode scanner laser beam may be visible to persons in the immediate area, but no laser safety hazard is associated with this exposure. The laser aperture for the primary reagent barcode scanner is directed toward the barcode labels at the end of the reagent packs inside the primary reagent compartment.

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode

During normal operation, the barcode scanner does not scan the reagent packs until the primary reagent compartment door is closed. When the primary reagent compartment door is closed, no operator exposure to the laser exists.

1 2 3

Primary Reagent Barcode Scanner Sample Barcode Scanner Ancillary Reagent Barcode Scanner

Figure A-3

Class 1 and Class 2 Laser Hazard Labels

Hand-held Barcode Scanner


ADVIA Centaur XP Operators Guide procedures that use the hand-held barcode scanner

are introduced with the following laser warning:

LASER WARNING: The hand-held barcode scanner emits low-power visible laser

radiation. To avoid damage to the eyes, never look directly at the laser beam or at its reflection from a shiny surface. Also, never point the scanner at anyone. Only Bayer-trained field service personnel should perform procedures related to laser assemblies. Refer to Appendix A, Protecting Yourself from Barcode Scanner Lasers, for more information.

078D0080-02, Rev. A

A-6

header Melias Operating Manual

This table contains a summary of the specifications for the laser optical assembly in the ADVIA Centaur hand-held barcode scanner:
Characteristic Specification

Maximum Power Output Wavelength Pulse Duration Units of Beam Divergence

1.0 mW 650 nm 7.4 ms 2mr

The laser aperture and laser safety label for the hand-held barcode scanner are on the other side of the barcode scanner window:

Barcode scanner window.

Figure A-4

Handheld Barcode Scanner Laser Warning Label

footer We never used this.....

Appendix B: Service, Ordering, and Warranty


This section provides the following information: address of the Bayer authorized representative, which is the Bayer contact within the European community addresses for obtaining service and technical information and for ordering supplies system warranty and service delivery policy information

Bayer Authorized Representative


Bayer Diagnostics Europe Limited Chapel Lane, Swords, Co. Dublin, Ireland

Limited Instrument Warranty and Service Delivery Policy


Bayer HealthCare and its authorized distributors may provide customers who acquire new Bayer HealthCare instruments with a limited warranty either in a specific agreement or in standard language on their invoices. This limited warranty is designed to protect customers from the cost associated with repairing instruments that exhibit malfunctions due to defects in materials and/or workmanship during the warranty period. Bayer Diagnostics, at its election, provides warranty service either by providing repair service of the instrument on site, or by exchanging the defective instrument or component, subject to the limitations and exclusions set forth in Replacement of Parts and Warranty and Service Exclusions below. Repairs, replacements or exchanges of instruments or components provided during the warranty or any additional service period, does not extend the warranty or service period beyond the initially agreed upon period. When the customer calls for service, the Bayer HealthCare representative or authorized distributor informs the customer of the type of service available for the customers instrument, and instructs the customer as to how to obtain that service.

Warranty Period
The limited warranty period generally commences upon installation of the original instrument at the customers location and extends for a period of one year thereafter, unless otherwise specifically agreed to by and between Bayer HealthCare (or its authorized distributors) and customer in a writing signed by duly authorized representatives of both parties (sales representatives are generally not authorized representatives of Bayer HealthCare for these purposes).

078D0080-02, Rev. A

B-2

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B

Additional Service Period


The customers, with some exceptions, may purchase additional service coverage beyond any initial warranty period as part of the original instrument acquisition for second or subsequent years beyond the original installation date. The customers original Purchase Invoice or appropriate Agreement Addendum must indicate the term in months for additional service coverage.

Service During Normal Hours


The customer may obtain service for instruments during normal business hours by contacting the nearest Bayer HealthCare location or authorized distributor. Refer to the list of Bayer HealthCare locations in this section.

Extent of a Service Call


Warranty or service calls generally include onsite repair or exchange of instruments or components, travel to the location of the instrument, and onsite labor during normal business hours. A warranty or service call is initiated by the customer by following the instructions on how to obtain service for the customers instrument. The service call is considered complete when any defects in material or workmanship have been corrected by repair or replacement and the instrument conforms to the applicable specifications. When service is complete, the customer receives a copy of the documentation detailing all work performed by the Bayer HealthCare representative or authorized distributor.

Service Outside Normal Hours


Customers, with some exceptions, may also request service to be delivered or an exchange to be initiated outside normal business hours, including evenings, weekend days, or nationally observed holidays, by contacting the nearest Bayer HealthCare location or authorized distributor. Service performed outside normal hours is subject to a surcharge unless the customer has in place a service product option that provides service at the time requested.

Replacement of Parts
In performing service, Bayer HealthCare or its authorized distributors provide appropriate parts to repair the instrument, or arranges for the exchange of the instrument or affected parts, at no charge with the exception of certain parts or subassemblies that are considered Customer Maintenance Items. Customer Maintenance Items include, but are not limited to, the following items: lamps, electrodes or sensors (which are covered by a separate warranty), reagents, calibrators, controls, paper, and pens. Consult the appropriate system operators manuals for a complete list of Customer Maintenance Items for any specific model of instrument.

078D0080-02, Rev. A

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B

B-3

Design Changes and Retrofitting of Instruments


Bayer HealthCare reserves the right to change the design or construction of specific models of instruments at any time without incurring any obligation to make such changes available to individual customers or instruments. If Bayer HealthCare notifies customers of a change that improves the performance or reliability of their instrument, and requests to retrofit that instrument, the customer must agree to allow Bayer HealthCare or an authorized distributor, at Bayer HealthCare expense, to retrofit components or make design changes, which does not adversely affect the instruments performance characteristics.

Key Operator Designation


Each customer designates a key operator who is available to Bayer HealthCare representatives to describe instrument malfunctions by telephone and/or to perform simple adjustments and corrections as requested. If a key operator is not designated or is unavailable when the customer requests service, the delivery of service may be delayed.

OSHA Requirements (US only)


When service is required at a customer location, the customer must provide the Bayer HealthCare representative with adequate facilities that comply with the regulations of the Secretary of Labor under the Occupational Safety and Health Act (OSHA) of 1970, as amended.

Warranty and Service Exclusions


The following exclusions are in addition to any exclusions provided for in any written warranty or service agreement. IF ANY OF THE FOLLOWING EVENTS OCCUR, THE WARRANTY OR SERVICE PROVISIONS DO NOT APPLY: 1. Repairs or modifications have been made to the instrument by someone other than an authorized Bayer HealthCare representative. 2. The instrument has been operated using accessories and supplies other than Bayer HealthCare brand accessories, or consumable supplies and/or reagents not having the same grade, quality, and composition as defined in the system operators manuals. 3. Bayer HealthCare has notified customers of a change that improves the performance or reliability of their instrument and customer has not agreed to retrofit or make design changes to the instrument. 4. Customer did not purchase the instrument from Bayer HealthCare or one of its authorized distributors. 5. The instrument has not been installed within 90 days of shipment to the customers facility unless otherwise specified.

078D0080-02, Rev. A

6. The customer has not performed appropriate customer maintenance procedures, as outlined in the system operators manuals. 7. The instrument has been misused or used for a purpose for which it was not intended. 8. The instrument has been damaged in transit to the customer or damaged by the customer while moving or relocating it without supervision by a Bayer HealthCare representative. 9. Damage was caused by floods, earthquakes, tornados, hurricanes, or other natural or man-made disasters. 10. Damage was caused by Acts of War, vandalism, sabotage, arson, or civil commotion. 11. Damage was caused by electrical surges or voltages exceeding the tolerances outlined in the system operators manuals. 12. Damage was caused by water from any source external to the instrument. 13. The customer has purchased an alternative agreement whose terms of warranty or service supersede these provisions. Bayer HealthCare or its authorized distributors can invoice customers, at current standard labor and parts rates, for instruments repaired to correct damage or malfunctions due to any of the reasons listed above. OTHER THAN AS STATED ABOVE, THERE ARE NO OTHER WARRANTIES, EXPRESS OR IMPLIED, WITH RESPECT TO THE INSTRUMENT, ITS SALE TO THE CUSTOMER, ITS LEASE TO THE CUSTOMER, OR THE SALE OF THE INSTRUMENT TO THE CUSTOMER AT THE EXPIRATION OR TERMINATION OF THE LEASE AGREEMENT. BAYER HEALTHCARE SPECIFICALLY DISCLAIMS ANY AND ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE OR PURPOSE. BAYER HEALTHCARES LIABILITY FOR BREACH OF ANY WARRANTY OR SERVICE AGREEMENT SHALL BE LIMITED ONLY TO THE REPAIR OR REPLACEMENT OF DEFECTIVE EQUIPMENT AND SHALL NOT INCLUDE ANY DAMAGES OF ANY KIND, WHETHER DIRECT, INDIRECT, INCIDENTAL, CONTINGENT, OR CONSEQUENTIAL. BAYER DIAGNOSTICS SHALL NOT BE LIABLE FOR DELAY FROM ANY CAUSE IN PROVIDING REPAIR OR EXCHANGE SERVICE. ANY LIMITATIONS OR OTHER PROVISIONS NOT CONSISTENT WITH APPLICABLE LAW IN PARTICULAR JURISDICTIONS OR A SPECIFIC WRITTEN AGREEMENT DO NOT APPLY TO CUSTOMERS IN THOSE JURISDICTIONS OR SUBJECT TO THOSE AGREEMENTS.

Information for Technical Assistance


Refer to the procedures in this appendix to provide system information that you may need when you call for technical assistance.

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Information for Technical Assistance B-5

Printing the System Serial Number and Versions


Use this procedure to print information about the system, such as the serial number and the current version of software used on your ADVIA Centaur XP system. 1. At the workspace, select System Status. 2. Select About the System. 3. At the workspace, select Print. 4. Select Print Workspace. The serial number is displayed on the System Status button. The version information is displayed at the About the System window.

Printing the Event Log


Use this procedure to print the list of system events displayed at the System Event Log window. You can display specific information at the System Event Log window by selecting the appropriate View and Sort combinations: The View that you select determines the information that the system displays. For example, if you select Errors, the system displays only the events that are errors. The Sort that you select determines the order in which the system displays the events. For example, if you select Severity, the system displays the events in order of most severe to least severe.

The following table describes the View and Sort combinations you can use to perform some routine tasks:
Task View Sort

Display the events in reverse chronological in order of when the events occurred Display the critical and warning events in order of most severe to least severe Display the events by subsystem 1. At the workspace, select Event Log.

Events Errors Events

Date Severity Subsystem

2. Select the appropriate view and identify the problem you want to solve. To search for a specific event code in the event log, select Event Code in Search for and then enter the event code. 3. Select Print. 4. Select Print Workspace.

078D0080-02, Rev. A

B-6

header Melias Operating Manual

Printing Windows on the Workspace


Use this procedure to print the workspace, including any windows that are open. Windows that can be helpful in resolving problems include Quality Control Data, Calibration Summary, Calibration Data, and Worklist Summary. Open the appropriate windows before you print the workspace. 1. At the workspace, select Print. 2. Select Print Workspace.

Printing a Runtime Report


Use this procedure to print the report of all results generated (including replicates, ratios, and means) for all samples run. 1. At the workspace, select Print. 2. Select Report Options. 3. At the Print Report Options window, select Runtime Report. 4. Select Print Report.

Printing Calibration Data


You can print detailed information on specific calibration data as described in Section 5, Calibration.

Contacts
This section provides the following information: the address of the Bayer authorized representative, which is the Bayer contact within the European community the Bayer addresses for obtaining service and technical information and for ordering supplies

Bayer Authorized Representative


Bayer Diagnostics Europe Limited Chapel Lane, Swords, Co. Dublin, Ireland

Addresses
For technical assistance contact your local technical support provider. For customer service or additional information contact your local technical support distributor.

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Contacts

B-7

Bayer S.A. Divisin Diagnsticos Ricardo Gutirrez 3652 B1605EHD Munro - Buenos Aires Argentina +54 (0) 11-4762-7000 Bayer Australia Limited ABN 22 000 138 714 Diagnostics Division 2 Keith Campbell Court Scoresby Victoria 3179 Australia +61 (0) 3-9212-8444 Bayer Austria GesmbH Geschftsbereich Diagnostika Lerchenfelder Grtel 9-11 A-1164 Wien, Austria +43 (0) 1-711-46-2424 Bayer s.a.-n.v. Division Diagnostics Avenue Louise 143 Louizalaan 1050 Bruxelles-Brussel, Belgium +32 (0) 2-535-66-81

Bayer S.A. Produtos Diagnsticos Rua Domingos Jorge 1100 04779-900 - So Paulo - SP Brazil +55-11-5694-5166 Bayer Inc. 77 Belfield Road Toronto, Ontario Canada M9W 1G6 416-248-0771 Bayer s.r.o. Oddlen Diagnostika Zastoupen pro eskou a Slovenskou republiku Litvnovsk 609/3 CZ-190 21 Praha 9 esk republika +420 266101442 Bayer S.A. Divisin Diagnstica A.A. 80387 Av. De las Amricas No 57-52 Santaf de Bogot, D.C. - Colombia +571 (09) 423-4199 Bayer HealthCare - Diagnostics Nrgaardsvej 32 DK-2800 Kgs. Lyngby, Denmark +45-45-23-50-00

078D0080-02, Rev. A

B-8

header Melias Operating Manual

Bayer OY Suomalaistentie 7 FIN 02270 Espoo, Finland +358-9-887-887 Bayer Diagnostics Tour Horizon 52, quai de Dion Bouton 92807 Puteaux Cedex, France +33 (0) 1-49-06-56-00 Bayer Vital GmbH Diagnostika Siemensstrae 3 D-35463 Fernwald, Germany +49 (0) 641 4003-0 Bayer Hellas AG 18-22 Sorou str. GR-151 25 Marousi Athens, Greece +30 (0) 210 6187 500 Bayer HealthCare Co. Ltd. 20/F Gee Chang Hong Centre 65 Wong Chuk Hang Road Hong Kong +852-28147337 Bayer Hungria Kft. Alkots Point H-1123 Budapest, Hungary Alkots u. 50 +36 (06) 1 4874 100 Bayer Diagnostics India Limited 589, Sayajipura Ajwa Road Baroda - 390 019 Gujarat, India +91 (0) 26-5562720 Bayer S.p.A. Divisione Diagnostici Via Grosio 10/4 20151 Milano, Italia +39-2-3978-1

Bayer Medical Ltd. Marunouchi Kitaguchi Bldg. 1-6-5, Marunouchi, Chiyoda-ku Tokyo 100-8263, Japan Bayer Diagnostics Korea Limited Kye Myung Bldg. 4F Myungil-Don, 48-7 Kangdong-Ku Seoul 134070 Korea +82 (0) 2-428-5987 Bayer (Malaysia) Sdn. Bhd 11th Floor Wisma Damansara Jalan Semantan 50490 Kuala Lumpur Malaysia +603 (0) 62093088 Bayer de Mxico, S.A. de C.V. Divisin Diagnsticos Va Morelos 330-E Col. Santa Clara C.P. 55540 Ecatepec, Estado de Mxico Centro de Atencin Telefnica (CAT) +52 55 57 28 33 06 Health Care Division Diagnostics Energieweg 1 3641 RT Mijdrecht The Netherlands +31 (0) 297-280666 Bayer New Zealand Ltd Diagnostics Business Group 3 Argus Place Glenfield Auckland, New Zealand +64 9-441-8549

footer We never used this.....

ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Contacts

B-9

Bayer B.V. Bayer AS Drammensvn. 147 B NO-0277 Oslo, Norway +47 24 11 18 00 Bayer Sp. Z o.o. Al. Jerozolimskie 158 02-326 Warszawa, Polska +48 (0) 22-572-3500 Bayer Diagnostics Europe Ltd (sucursal em Portugal) Rua Quinta do Pinheiro, 5 2795-653 Carnaxide, Portugal +351-21-416-4227 Bayer Puerto Rico Inc. Diagnostics Division Victoria Industrial Park Building 1 Carolina, Puerto Rico 787-752-8989 Bayer (South East Asia) Pte Ltd. No. 9 Benoi Sector Singapore 629844 +65 6261 3389 Bayer (Pty) Ltd. Healthcare Division 27 Wrench Road Isando 1600, South Africa +27 (0) 11-921-5048

Qumica Farmacutica Bayer, S.A. Divisin Diagnsticos c/ Pau Claris, 196 08037 Barcelona, Espaa +34-93 4956500 Bayer AB Drakegatan 1 S-402 24 Gteborg, Sweden +46-31-83-98-00 Bayer (Schweiz) AG GB DS Grubenstrasse 6 CH-8045 Zrich Switzerland +41 (0) 1-465-81-11 Bayer plc Diagnostics Division Bayer House Strawberry Hill Newbury, RG14 1JA United Kingdom +44 (0) 1635-563000 Bayer HealthCare LLC Subsidiary of Bayer Corporation 511 Benedict Avenue Tarrytown, NY 10591-5097 USA 914-631-8000

078D0080-02, Rev. A