HMB CRANE LOAD CELL CONTROLLER CALIBRATION PROCEDURE 1

Enter in the setup mode by pressing <G> & <T> Key simultaneously, display will show P 00. Enter code number 10 and press <PRNT> Key. Enter the password using key < > < > ; < > < >, .Press <PRNT> Key, if the password is correct system will display Good for 3 seconds, and then will show P 11 Enter Code 24, Press <PRNT>, Display will Show 10.000 by default, here enter the full scale weight and press <PRNT> Key The indicator will Display P 0, here if the you want to calibrate the Zero of the system, change this 0 to 1 , Press <PRNT> Key. Now the indicator will Display Unload Here Press <-0> Key to Start Zeroing the System. The message will get displayed as Wait After completing Zeroing operation the system will display the calibration weight value which is 10% of the full Scale Entered by the user by default. Here the you can change the value as desired and press <PRNT> Key. Now the indicator will Display Put Wt, Put the weight and press <PRNT> Key, Now the indicator will Display Wait till the calibration completes. After completion of calibration the indicator will show P 27. Press <EXIT> key.

Display will Show Store? for 3 seconds & then will Show P 0, change 0 to 1 and press <PRNT> this operation, calibration process is complete and display will start showing current weight.

10 Display will Show SAVING. After completion of