You are on page 1of 124

MULTICONT WEIGHFEEDER Service Manual

FH 345 GB

Copyright 1999 SCHENCK PROCESS GmbH, Landwehrstrasse 55, D-64293 Darmstadt All rights reserved. Any reproduction of the manual, regardless of method, without prior SCHENCK PROCESS GmbH in writing, even by excerpt, is prohibited. Sales:Tel.: +49 (0)6151 32 - 10 28 Email: sales2.process@csd.de

App lication Segments Service: Heavy Duty Weighing & Feeding Service Tel.: +49 (0) 61 51 32 - 26 23 Service Fax: +49 (0) 61 51 32 - 32 70 Light Duty Weighing & Feeding Service Tel.: +49 (0) 61 51 32 - 25 72 Service Fax: +49 (0) 61 51 32 - 17 49 Screening / Drying / Cooling Service Tel.: +49 (0) 61 51 32 - 10 75 Service Fax: +49 (0) 61 51 32 - 30 96

Weighing in Transport & Dispatch Automation Service Tel.: +49 (0) 61 51 32 - 24 48 Service Fax: +49 (0) 61 51 32 - 13 69

2 4h Emergency hotline: +49 (0) 172 - 650 1700 Email: service3.process@csd.de

Spare Parts: Electrical: +49 (0) 6151 32 - 17 58 Mechanical: +49 (0) 6151 32 - 31 22 Fax.: +49 (0) 6151 32 - 36 32 Hotline:+49 (0) 171 - 225 1195 http:/ /www.schenck-process.com

MULTICONT
Contens
1.

WEIGHFEEDER

General Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1. 1.2. 1.3. The Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The MULTICONT System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overview Weighfeeder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.1. 1.3.2. 1.3.3. Gravimetric / Volumetric Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Check System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Batching Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-1 1-1 1-1 1-3 1-4 1-4 1-5 2-1 2-1 2-3 2-3 2-3 2-3 2-5 2-5 2-5 2-6 2-7 2-7 2-8 2-8 2-8 2-9 2-9 2-9 2-9 2-10 2-10 2-11 2-11

2.

Modes of Operation and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1. 2.2. Survey of Modes of Operation and Functions . . . . . . . . . . . . . . . . . . . . . . . . . Main Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2.1. 2.2.2. 2.2.3. 2.3. Batching Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Display Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Check System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Calibration Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3.1. 2.3.2. 2.3.3. 2.3.4. 2.3.5. 2.3.6. 2.3.7. 2.3.8. 2.3.9. Acquire Belt Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Taring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weight Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Taring Second Weighbridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Taring of Supply Bin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Controller Adaptation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Time Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Service Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.4.

Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4.1. 2.4.2. 2.4.3. 2.4.4. 2.4.5. Display Parameters / Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . Enter Parameters /Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . Print Parameters / Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . MULTIKEY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Data Lost . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9947

2.4.6.

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmgH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

CONTENTS FH 345 GB/ I-1

WEIGHFEEDER
2.4.7. 2.5. 3.

MULTICONT
2-12 2-12 3-1 3-1 3-1 3-1 3-1 3-1 3-1 3-1 3-1 3-2 4-1 4-1 4-4 4-8 5-1 5-1 5-2 5-6 6-1 6-1 6-1 6-2 6-2 6-2 6-3 6-4 6-4 6-4
9947

Multi-Memory Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Restoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Commissioning and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1. 3.2. 3.3. 3.4. 3.5. 3.6. 3.7. 3.8. 3.9. Power ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System Adaptation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Belt Circuit Code Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Taring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weight Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Controller Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Control Signals Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting of Belt Influence Compensation and Control to Discharge Point . . . . . Local Control Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 4.2 4.3 General Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Survey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . List of Configuration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.

Parametrization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1. 5.2. 5.3. Parameter Designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Survey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . List of Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.

Event Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1. General Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1.1. 6.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

List of Events Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2.1. 6.2.2. 6.2.3. 6.2.4. 6.2.5. 6.2.6. System Faults SY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety Indications SC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety Indications WE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety Indications WM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety Indications MF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Interlocking IL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

CONTENTS FH 345 GB/ I - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
6.2.7. 6.2.8. 6.2.9. 6.2.10 6.2.11

WEIGHFEEDER
Controller CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Check Measurement CH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calibration CA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MAX Limit values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MIN Limit values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5 6-5 6-6 6-7 6-7 A-1 B-1 B-1 B-5 B-8 B-8 C-1

Annex A: Alphabetic Search List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Annex B: List of Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B.1 B.2 B.3 B.4 Digital Inputs DE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital Outputs DA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analog and Measurement Inputs AE, GA, WZ, . . . . . . . . . . . . . . . . . . . . . . . . Analog and Pulse Output , AA, IM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Annex C: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

This manual applies to software:

FDB 1201-09

Edition: 9947
9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmgH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

CONTENTS FH 345 GB/ I-3

MULTICONT
1General Comments
1.1The Manual
The Service Manual FH 345 for Weighfeeders completes the Operating Manual Fh 344 with additional information on commissioning, calibration, configuration, and parametrization. The manual centers around Chapter 4 Configuration and Chapter 5 Parametrization. The presentation is arranged such that all requisite information is conveyed at a glance with no need for explanatory texts. The general comments at the beginning of each chapter should be read. A survey in each chapter as well as a general key word search list in Annex A facilitate location of desired data.

WEIGHFEEDER

Annex B comprises a list of interfaces complete with all signals to which an interface can be assigned. Settings deviating from the default values can be entered into this list. Chapter 6 contains a list of all event messages with comments. The modes of operation and functions accessible to the operator as well as the relevant procedures are described in Chapter 2 Modes of Operation and Functions. Chapter 3 Calibration and Commissioning conveys information on commissioning and calibration. Section 1.2 gives a general survey of the MULTICONT system.

1.2The MULTICONT System


The MULTICONT system is a micoprocessor-controlled decentralized data acquisition and evaluation system for continous weighers and feeders. Each weighing system basically comprises at least one system unit FSE, one or multiple input/output units FEA, and measurement data acquisition units FME (version RC), or one or multiple input/output units FEA (version NC). The weighing systems are individually controlled and monitored via one or two machine control stations FLB or collectively via a group control station FZB. A group control station enables group rate control functions to be effected. A host system can be connected to the group control station.

The MULTICONT system is programmed for the application by software. The specific settings (adaptation to hardware, interface assignment, controller setting) are effected by configuration and parametrization. Access to these functions is protected.

Fig.1.1: Version NC (Normal Cabling)


9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

General Comments FH 345 GB/ 1 - 1

WEIGHFEEDER
System unit FSE monitors and controls all sequences of operation. The exchangeable multi-memory module comprising the specific software plugs into the system unit. Feeder drive and additional units are connected to input/output unit FEA, in version NC also the load cell. Equipped with six digital inputs, one analog input, measurement data acquisition unit FME (version RC) servers for local signal acquisition. In addition, the FME is complete with a plug connection for linking a machine control station, which can be connected, for instance, for effecting maintenance work. Machine control station FLB servers for operation, configuration, and parametrization of one weigher each time. Maximum two machine control stations can be connected to one weigher.

MULTICONT

Group control station FZB is designed for operating and controlling serveral weighers, as individual machines or as a group of weighing and feeding units Fig.1.2: Version RC (Reduced Cabling) enabling group rate control functions to be effected. The belt weigher in connected with the batch processor. Power supply unit FNT L00x provides the supply voltage, in version RC exclusively for the electronic For commissioning and calibration weigher can be assemblies, in version NC also for the machine control controlled by means of a local control box in Local station. Mode. Visual contact to weigher is required. Local mode overrides all other modes of operation. Coupling module FNT 005x supplies the measurement data acquisition unit and the control stations. Drive unit FAE servers for varying the feed drive.

9947

General Comments FH 345 GB/ 1 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
1.3 Overview Weighfeeder

WEIGHFEEDER

Fig. 1.3: Weighfeeder Basic Diagram

The weighfeeder serves for continuous weighing and feeding of bluk solides. The belt load is acquired through a weighbridge located under the conveyor belt. A control loop adjusts feed rate p to a preset setpoint via belt speed v. The weightbridge is limited by two carrying idlers. By means of a centrally positioned weighed idler and a lever system, the load on platform exerts a force, platform load Q, on load cell (L/C). The output voltage of the load cell is proportional to the platform load and is transformed into the digital information through an A/D converter.

The weight force picked up by the weighed idler is

Q=
> Q = Platform load > q = Belt load

qL 2
[kg], [kg/m], [m]

> L/2 = Effective belt length The formula resulting for belt load is

q=
Fig. 1.4 shows the distribution of load an platform for ideal conditions. Feed rate P is

Q Leff

P=qv
> P = Feed rate > q = Belt load > v = Belt speed [kg/s], [kg/m], [m/s].

Fig. 1.4: Load Distribution at Weighbridge


9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

General Comments FH 345 GB/ 1 - 3

WEIGHFEEDER
1.3.1 Gravimetric / Volumetric Mode
The gravimetric mode of operation is the mode for feeding. Depending on belt load, the feed rate is controlled to the active setpoint via the belt speed with belt load q being determined by the prefeeder. In volumetric mode, the acquired belt load is not used for control purposes. The belt speed is controlled in dependence of the permanently set nominal belt load and the active feed rate setpoint. The volumetric mode is activated > in start-up mode (P 10.03), LG > in clearance mode (P 10.04) > by connection of local mode > by call via function distributor, SE bus, or contact. KMK = =

MULTICONT
N= KMK - KMS LG
KMS KMK

B=
with N =

Zero point error of continuous feeder in kg/m Span error of continuous feeder Belt lenght passed over weighing platform during check in m Quantity acquired by continuous feeder in kg Quantity acquired by static feeder in kg.

KMS =

1.3.2 Check System


The check system as such is a hopper scale prepared to check and correct the actual feed rate values acquired by the weightfeeder. During check and correction, the material flow is not interrupted. The check system can neither be started in batching mode nor while a set-up program is running. The following check system versions are availabel: CHECK SYSTEM TYPE EEIN: Check system and weighfeeder form a mechanical unit.

The duration of check measurement is either determined by the number of belt circuits, or by check quantity. The check system is configured ans parametrized in Block K22 (TYP), Block K23 (EEIN, EAUS), Block K24 (EXTERN) and Block P 20. In addition, an automatic weight check can be activated starting a new check cycle after elapse of a check measurement with subsequent cycle time (P 20.17). Activate automatic acquisition of check result with Parameter P 20.01. In the course of the checking operation, check amount (P20.07/08, K23.10), check fill weight (K23.09/10, P20.09/10) and correction (K23.11 - K23.14, P20.11 P20.14) are monitored for excess of limit values.

CHECK SYSTEM Check system and weighfeeder do TYPE EAUS: not form a mechanical unit. CHECK SYSTEM Check system is not controlled by TYPE EXTERN MULTICONT but externally. The check result consists of the measurement results of both feeders for the material amount actully fed. As static weighing is much more accurate than the dynamic weighing, it is taken for correct. The correction value is computed from the difference, or ratio, of both quantities. Zero point correction compensates additive errors caused, for instance, by material deposits on weighing platform. Span correction servers for compensation of multiplying errors. Span corrections are principally required after commissioning, zero point corrections during operation as well.

9947

General Comments FH 345 GB/ 1 - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
1.3.3 Batching Mode

Check Measurement Sequence

KME TYPE EEI

KME TYPE EAUS

Start check measurement with preset check setpoint aiming at - zero point correction - span correction.

Release signal is expected.

Check amount and bin are checked and refilled

Prefeeder is cut off (K21.07), feedback signal and stabilization time elapse.

Fig. 1.5: Batching Mode

Measurement of continuous feeder is started.

Fill weight F1 is measured.

In batching mode the discharge unit is automatically cut off after discharge of the set amount. A) Start of batch with set amount and set feed rate.

Wait for blind distance LB to elapse. B) Pre-contact residual amount is reached, set feed rate is limited to pre-contact set feed rate (creep mode with constant setpoint). Main contact residual amount is reached. Feed unit is cut off. Clearance mode, if preselected, becomes active when main contact resedual amount is reached. Drive cuts of after elapse of set time. When feed unit is in no-motion, the set amount has been discharged.

Fill weight F1 is measured.

Measurement of continuous feeder is started.

C)

If check setpoint is reached measurement stops.

If check setpoint is reached second fill weight F2 is measured.

D)

Fill weight F2 is measured, KMS = F1 - F2.

Measurement of continuous feeder stops.

Result is evaluated.

Enter pre-act contact residual amount , pre-act contact set feed rate and main contact residual amount as parameters. Main contacts is self-optimizing, i.e., after a batch is terminated the difference between batch setpoint and actually fed batch is acquired and, if deviation is too high, main contact residual amount is calculated anew.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

General Comments FH 345 GB/ 1 - 5

MULTICONT
2 Modes of Operation and Functions
2.1Survey of Modes of Operation and Functions

WEIGHFEEDER

Main Modes Functions Mode 0 / Mode 1 Preselect Main Mode Weigher ON/OFF Enter abs. setpoint Enter rel. setpoint Reset Totalizing Counter Stop Totalizing Counter DE / SE / TT DE / SE / TT AE / SE / TT .0 / SE / TT DE / SE / TT Test Mode SE / TT DE / SE / TT AE / SE / TT .0 / SE / TT DE / SE TT DE / SE DE Local Mode DE Mode-Independent

Acknowledge Event Message Independent of configuration, SE bus and keyboard act as sources with equal priority. Volumetric Mode Gravimetric Mode Belt Influence Compensation BIC Control Delay for Discharge Point Automatic Weight Check Weight Check Prefeeder Batching Mode Open loop / closed loop fill weight control Check system Moisture correction 2nd platform Dead time element Dead time element
9947

SE bus and keyboard act as sources of squal priority. In local mode, volumetric mode is permanenty active.

DE

.0 / .1

.0 / .1

.0 / .1 DE / SE / TT See configuration Block K 17, Block K 18, Block K 19. See configuration Block K 20. See configuration Block K 21.

See configuration Block K 21, Block K 22, Block K 23. See configuration Block K 25. See configuration Block K 26. See configuration Block K 27. See configuration Block K 25.

Tab.2.1: Main Modes, Functions, and Sources

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 1

WEIGHFEEDER
Table 2.1 shows all configurable functions with their admissible sources determining the access to the individual functions. Depending on function, access can be set for each main mode, or independent of the individual main mode, per configuration (see Chapter 4). Main modes Mode 0 and Mode 1 are provided for normal operation. In most cases, the weigher is operated via a group control station in Mode 0 and via a machine control station in Mode 1. Test Mode and Local Mode serve for commissioning and service, the access is protected. In Local Mode, operation is via the local control box. In this mode, only a limited number of weigher functions is accessible and active. Some of the functions listed in Table 2.1 are operated by function keys, others via the function distributor. Main Function. . . . . . . . . . . . . Display Preselect batching mode . . BATCH MODE ON Abort batch mode . . . . . . . . BATCH MODE OFF Abort batching mode. . . . . . BATCH INIT. Start zero point calibration . CALIB. ZERO Start span calibration . . . . . CALIB. SPAN Abort calibration . . . . . . . . . STOP CALIB. Display check. . . . . . . . . . . . . . DISPLAY CHECK Calibration functions . . . . . . CALIBRATION Programming . . . . . . . . . . . PROGRAMMING Restoring . . . . . . . . . . . . . . RECOVER Subfunction . . . . . . . . . . . . . . Display Calibration Functions Acquire belt length . . . . . . . . . . Taring . . . . . . . . . . . . . . . . . Weight check . . . . . . . . . . . Taring supply bin. . . . . . . . . Taring second weighbridge . Controller optimization . . . . . . . 2nd weighbridge active . . . . 2nd weighbridge not active . Time entry . . . . . . . . . . . . . . . . Display service values . . . . . . . Test mode . . . . . . . . . . . . . . . . Volumetric mode . . . . . . . . . . . Gravimetric mode. . . . . . . . . . . Prefeeder ON . . . . . . . . . . . Prefeeder OFF . . . . . . . . . . LB: PULSE/REV. TW: TARE CW: CHECK Q TF: SUPPLY BIN T2: TARE WB2 CONTR. ADAPT. WB2 ACTIVE WB2 INACTIVE SET TIME SUPPL. DISPLAY TEST-MODE VOLUMETRIC GRAVIMETRIC PREFEEDER ON PREFEEDER OFF

MULTICONT
Table 2.2 shows the sub-distributors and functions accessed via function distributor. Access to the functions identified by o can be protected by fuse-link contact (key-operated switch) and password. Password (7274) cannot be changed, interrogation may be switched ON, or OFF, by means of Parameter P 00.01. Upon selection of function distributor, or sub-distributors, the sub-modes and functions active last are displayed. The fuse-link contact being closed, subdistributor Calibration Function is selected first. The functions identified by are displayed only if they have been configured actively and if the control station in the active main mode has access privilege for these functions.

Sub-function . . . . . . . . . . . . . Display Programming Display parameters. . . . . . . . . . Display configuration data . . . . Enter parameters . . . . . . . . . . . Configurate . . . . . . . . . . . . . . . . Print parameters . . . . . . . . . . . . Print configuration data. . . . . . . Load parameters from multi-memory module . . . . . . . . Load configuration from multi-memory module . . . . . . . . Store parameters in multi-memory module . . . . . . . . Store configuration in multi-memory module . . . . . . . . Load default parameters from EPROM . . . . . . . . . . . . . . . . . . Load default configuration from EPROM . . . . . . . . . . . . . . . . . . Load parameters from MULTIKEY. . . . . . . . . . . . . . Load configuration from Multikey . . . . . . . . . . . . . . . . Store parameters in MULTIKEY. . . . . . . . . . . . . . Store configuration in MULTIKEY. . . . . . . . . . . . . .

READ PARAM. READ CONF. ENTER PARAM. ENTER CONF. DUMP PARAM. DUMP CONFIG. PARAM. MMM-SE CONFIG. MMM-SE PARAM. SE-MMM CONFIG. SE-MMM LOAD INIT. PAR. LOAD INIT. CONF. PARAM. MK-SE CONFIG. MK-SE PARAM. SE-MK CONFIG. SE-MK

Tab.2.2: Functions Distributor


9947

Modes of Operation and Functions FH 345 GB/ 2 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
2.2 Main Functions 2.2.1 Batching Mode
Sequence: FUNCTION PRESELECT BATCH Preset batching mode (feeder requires to be OFF), enter batch setpoint.

WEIGHFEEDER

2.2.2 Display Check


Display is checked by means of this function. Display elements flash for approx. 5 s. Then version number of software is displayed. The version number indicated in the table of contents must correspond to FDB XXXX-YY. FUNCTION DISPLAY CHECK Display in function distributor.

P H8 JP I8 A B

Zt = xxxx.xx kg

Enter batch setpoint which has to appear in the setpoint field. Open input field and enter value. Enter pre-contact and main contact residual amounts as parameters in Block P 15. In the input/output field, amounts fed, or residual amounts, can be displayed. Batch is started, amount fed and residual value are initialized. A running batch is stopped by operating key OFF or by an ALARM message. Drive is cut off, all values remain stored. When feeder is restarted, batch continues being processed. A running batch is aborted, drive is cut off. When feeder is switched on, a new batch is started. Drive is switched off, submode batching mode is reset and the batch setpoint cleared. Feeder returns to its previous mode of operation. Function ABORT BATCH is called automatically if mode is changed during batching.

P
MULTICONT FSE:FDB XXXX-YY ZZZ

Start display check. Display after termination of check.

Zb xxx.xx kg Zd xxx.xx kg

2.2.3 Check System


The configuration is effected in Blocks K 22, K 23 and K 24. Setting of check system is in Block P 20. Serving for zero point or span correction, the system is started via function distributor. If necessary, check measurement can be aborted. The active system is represented in the submode field with CH, the event messages are displayed in the event field with the symbol only. The results of the latest measurement can be rolled into the input and output field. Depending on configuration, the correction value is accepted automatically, or by manual intervention. If an error occurs, the result has to be accepted, or rejected, in manual anyway. Start-up and Operating Conditions: No batching mode, No set-up program active.

FUNCTION ABORT BATCH

P P

FUNCTION ABORT BATCH

Program End: Accept correction value as Parameter P 03.04 in accordance with configuration.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 3

WEIGHFEEDER
Messages: For running- and error messages, see Chapter 6, 6.2.8 Check Measurement CH.

MULTICONT

N
FUNCTION STOP CALIB.

Reject result and terminate check by operating key ESC.

Sequence: FUNCTION CALIB. ZERO Display in sub-function distributor.

Abort check measurement.

P
CH 09: NO PERMISSION CH 10: PREPARATION ACTIVE

Start check measurement for zero point correction (KME T) or span correction (KME B). Check is logged on, system expects release. Release was given, no error messages are available. Bin fill weight is checked, if necessary, refilling is initiated. End of filling must be accepted by a feedback signal. Message CH 08 is displayed until acknowledgement is effected and a damping time has elapsed. This is when measurement starts.. Result of check measurement for zero point correction (F/Z B for span correction). Check result can be displayed in input/output field. The relation of static measurement to continuous measurement is displayed. The correction value is accepted with ENTER if manual acceptance has been preselected (configuration), or if an error in automatic acceptance has occurred.

CH 08: FEEDBACK PREFEEDER

F/Z T

9947

Modes of Operation and Functions FH 345 GB/ 2 - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
2.3 Calibration Functions
Subdistributor Calibration Functions comprises all set-up and check programs as well as main mode Test Mode. Before calibration effect parametrization and configuration in accordance with data sheets. Main modes and submodes are not represented on display. Turn off automatic belt influence compensation as long as calibrating program is running, then restart. LB: xxxx sec xxxx Imp

WEIGHFEEDER

Running display. Belt with no marking; Display shows residual run time and totalized number of pulses.

LB: xxxxx I/B

2.3.1 Acquire Belt Length


Program LB acquires the number of speed generator pulses for one belt circuit with the result being stored as belt circuit code number IB. Conveyor belts with and without marking can be used. If belts with marking area are used, belt start and belt circuit time are determined with the aid of a proximity initiator. Belts with no marking require the time for one belt circuit to be entered as Parameter P 02.02. Display during running program varies according to belt type. v xxxxx m/s xxxxx I/B

End display. Belt with marking; Display shows number of pulses measured per belt circuit. End display. Belt with no marking; Display shows mean belt speed and number of pulses per belt circuit. Accept value on display.

P N
2.3.2 Taring

Abort program.

Start-up and Operating Conditions Weigher ON, Volumetric mode, No material on belt. The acquired value is stored as tare N and deducted from gross value in normal mode. Program run time is set with Parameter P 02.01. Start-up and Operating Conditions: FUNCTION LB: PULSE/REV. Weigher ON, Volumetric mode, No material on belt.

Program end: Belt circuit code number IB is stored as Parameter P 03.07. Program Sequence: Display in sub-function distributor. Start program. Running display. Belt with marking.

Program End: Value tare N is stored as Parameter P 03.03. Program Sequence: FUNCTION TW: TARE Display in subfunction distributor.

LB: CALIB. ACTIVE

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 5

WEIGHFEEDER

MULTICONT
Program Sequence:

P
TW: xx.x % Tara: x.xx % Abw. x.xx % Tara x.xx %

Start program. Running display showing residual run time of program [% total time] and deviation from zero point [% qNENN]. End display showing tare correction N [% qNENN] and the deviation from previous value [% qNENN]. FUNCTION CHECK Q Display in sub-function distributor.

P
CW: xx.x % SOLL/IST x.xx

Start program. Running display, with upper line showing the residual run time of program [% total time], and lower line indicating the relation of setpoint to measured value. Upon program end running message stops at 0 %, the lower line shows span correction KOR. If deviations exceed 3 % (KOR > 1.03, or KOR < 0.97), check set parameters and mechanical equipment. Abort program

P N
TW: E-Prog. aktiv. TW: Remove PRG

Store value. Abort program. CW: 0% KOR x.xxxx

Automatic Weight Check If automatic weight check has been selected, additional messages are displayed. Program start is acknowledged after approx. 7 s. Check weight is positioned on feeder. Program is not started. Remove check weight.

P N

Automatic Weight Check:

2.3.3 Weight Check


Programm CW acquires mean value of platform load for the check weight applied. The weight should not exceed the range of 2/3 to 4/3 of nominal platform load. Start-Up and Operating Conditions Weigher ON, Volumetric mode, No material on weigher, Check weight (P 02.09) in position.

If automatic weight check has been selected, additional messages are displayed. CW: E-Prog. active CW: Position PRG Program start is acknowledged after approx. 7 s. Check weight is not in position. Program is not started. Position check weight.

Program End: Span correction KOR is stored as Parameter P 03.06.


9947

Modes of Operation and Functions FH 345 GB/ 2 - 6

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
2.3.4 Taring Second Weighbridge
Program T2" acquires the mean value of platform load over the preset number of belt circuits. For program processing, see program TW". Start-up and Operating Conditions: Weigher ON Volumetric mode

WEIGHFEEDER
2.3.5 Taring of Supply Bin
With program TF: Supply Bin, the load cell amplifier is optimized and tare N of supply bin is acquired. Start-up and Operating Conditions Weigher OFF, No material in supply bin.

Program End: No material on feeder Tare N is stored as Parameter 13.01. Program End: Program Sequence: Tare N is stored as parameter P 23.01. Program Sequence: FUNCTION TW: TARE Display in subfunction distributor. FUNCTION: TF: SUPPLY BIN TF: CALIB. ACITVE Display in sub-function distributor. Acknowledge program start. Program is running twice. During first run, load cell amplifier is optimized, in second run tare N is acquired. Running display. The upper line shows remaining run time of program [% total time], in the lower line, the acquired tare [% nominal fill weight FGNENN] is displayed. End display. Display shows deviation from previous value [% FGNENN] and newly acquired tare N [% FGNENN].

P
TW: xx.x % Tara: x.xx %

Start program. Running display showing residual run time of program [% total time] and deviation from zero point [% qNENN]. TF: xxxx % Tara xxxx %

Abw. x.xx % Tara x.xx %

End display showing tare correction N [% qNENN] and deviation from previous value [% qNENN].

Abw. xxxx % Tara xxxx %

P N

Store value.

Abort program.

P N

Accept values. Abort program. No values are stored.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 7

WEIGHFEEDER
2.3.6 Controller Adaptation
With program Contr. Adapt. belt load controller (with prefeeder and second weighbridge) can be optimized. Start-up and Operating Conditions: EAx_DE None. Program End: Controller parameters are stored automatically. EAx_DA 000000 000000 frq [Hz]

MULTICONT
Raw measured value of speed generator; Control magnitude Feed unit duty cycle; Status digital inputs, input/output unit x, measurement data acquisition unit x; Status digital outputs, input/output unit x; Slip, variation of belt length; Current load cell utilization, bin; Current load cell utilization, Second Weighbridge; Prefeeder control magnibude Relative span error, check system Relative zero point error, check system; Control magnitude, fill weight control; Belt drift; Belt drift left Belt drift right. Result of weighbridge check (K 26.06)

Y ED

[V] [h]

s Program Processing: ab FUNCTION CONTR. ADAPT Display in subfunction distributor. ap Program start. YZ

[%] [%]

P
2.3.7 Time Entry

[%]

[V]

ch B

[%]

ch T

[%]

Function Time Entry serves for entry and interrogation of system time. After call of function, year, month, hour, minute, and second are interrogated one after the other. Entry is in 2-digit form.

yf

[V]

tr xx % <-

[cm]

2.3.8 Service Values


The following values can be displayed: EAx.AEx [V] Voltage analog input x, input/output unit x, measurement data acquisition unit x; Voltage analog output x, input/output unit x; Current load cell utilization;

xx %-> dQ [%]

The values identified by are displayed only if the respective functions have been activated.

EAx.AAx

[V]

aw

[%]

9947

Modes of Operation and Functions FH 345 GB/ 2 - 8

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
2.3.9 Test Mode
Main mode Test Mode enables the load cell and speed generator signals to be simulated and thus all weigher control signals to be checked with no load. Sequence: FUNCTION TEST MODE Display in function distributor.

WEIGHFEEDER

WZ SIMULATION J = 1; N= 0 _

Interrogate load cell simulation.

1P
GA SIMULATION J = 1; N = 0 _

Load cell signal is simulated. Interrogate speed simulation.

Select main mode.

BP
P00.01 Passakt * YES

Speed signal is not simulated, Test Mode becomes active.

2.4 Programming
By means of the functions comprised in Programming parameters and configuration data can be entered, stored internally or externally, or loaded. Setpoint field shows designation, input field value of parameter. Fields whose values can be changed by cursor operation flash. Parameter, Block, Number, Symbol, No default value, Current parameter value. Configuration data are displayed accordingly. For details, see Chapter 4. Use keys 4 and 6 to denote block, or number, keys 2 and 8 for increasing, or decreasing, the value of the denoted field.

2.4.1 Display Parameters / Configuration


With these functions, parameters or configuration data are exclusively displayed, they cannot be modified. For sequence of operation, see parameter entry.

P 00 01 Passakt * YES

2.4.2 Enter Parameters / Configuration


These functions serve for entry. Parameters can be entered with weigher switched ON, configuration data with weigher switched OFF only. Operating procedures are the same. Configuration data are exclusively entered by cursor operation. Program Sequence: FUNCTION ENTER PARAM. Display in subfunction distributor.

46 28 J

P
9947

Open entry field by operating key ENTER. Previous display is replaced by dashes. Entry is left-flush. Some parameters can only be changed by cursor operation (see parameter list). Select and change configuration data by cursor operation. Only denote fields whose values can be changed.

Subfunction is selected, first parameter is displayed.

K 03.02 T_MAX1 DE -.NO W1

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 9

WEIGHFEEDER

MULTICONT
2.4.4 MULTIKEY

P O N N

Accept value by operating key ENTER. Next parameter is displayed.

Delete faulty value by operating key DELETE.

Abort entry by operating key ESCAPE.

The MULTIKEY is an external memory module connected to the FSE system unit for transmission of data (central interface, activation per configuration K 00.10). With functions Param. MK-SE and Conf. MK-SE data are loaded from MULTIKEY, with functions Param. SE-MK and Conf. SE-MK they are stored in MULTIKEY. Data exchange is monitored, errors are reported instantaneously. Program sequence is the same with all memory functions. Program Sequence:

Terminate parametrization by operating key ESCAPE once again. Subfunction distributor is left.

FUNCTION PARAM. SE-MK

Display in subfunction distributor.

2.4.3 Print Parameters / Configuration


A commercial printer can be connected to the central interface of the system unit by using connecting cable FSC 0006. (See Hardware Manual FH 369 E 13/14.) Printer setting: Transmission rate 9600 bps No parity 1 stop bit 7 data bits.

P
MK CODE X J=1, N=0

Call function. Configuration data can be loaded only with weigher being OFF. The attempt to load data with weigher running entails abort of program and information SWITCH OFF. Display shows module code.

FUNCTION DUMP PARAM.

1P
FUNCTION OK!! OVERWRITE? OK? J=1, N=0

Enter 1 for acknowledgement. Display appears after faultless transmission of all data. Errors in data transmission are reported at once. If errors are detected upon storing into MULTIKEY, error messages are displayed. After that it is possible to overwrite the data in MULTIKEY. Enter 1 for overwriting, 0 for abort, or abort with key ESCAPE.

Upon call of function, print-out is produced showing version number of software, print date, SE address as well as all valid parameters.

FUNCTION DUMP CONF.

Print-out shows version number of software, print date, address of SE, all valid configuration data as well as the assignment of all hardware inputs and outputs.

Deviations from default values are crossmarked (*).


9947

Modes of Operation and Functions FH 345 GB/ 2 - 10

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Error Messages: NO CONNECTION Memory module not properly connected. Memory module has no code. Function is aborted. Error in data transmission, repeat function. If message is repeated, exchange memory.

WEIGHFEEDER
2.4.5 Default Values
With functions Load Init. Par. and Load Init. Conf. the default values indicated in Chapters 4 and 5 are loaded. For program sequence, see Item 2.4. For acknowledgement, message Data Accepted is output.

NO CODE WRITE ERROR READ ERROR STATUS ERROR QUANTITY ERROR WR. LENGTH CHECKSUM FAULT

2.4.6 Data Lost


MULTIKEY contains data of a different type of weigher than the active store (e.g. weighfeeder). MULTIKEY and MULTICONT active store contain different version numbers of software. Upon change of program version or data loss after power-on, parameters can be loaded from various sources. Commissioning is in the dialog mode as under: Upper display Lower display : : Data lost Restau. 0/1 ?

WRONG TYPE

WRONG VERS.

SY 09

WRONG REV

Error upon check of current program position.

After Entry 1 (ENT) data are loaded from the multi-memory module. After Entry 0 (ENT), system prompts loading of default values. Upper display Lower display : : Data lost Default values 0/1 ?

WRONG P,K

MULTIKEY and MULTICONT active store contain different parameters or configuration data.

SY 09

After Entry 1, default values are loaded; after 0, system prompts: REFUSED Data could not be loaded. Upper display Lower display MK NOT PROG! The MULTIKEY from which data are to be loaded is empty. : : SY 09 Data lost Multikey 0/1 ?

After 1, data are loaded from multikey; after 0, system returns to first prompt.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Modes of Operation and Functions FH 345 GB/ 2 - 11

WEIGHFEEDER
2.4.7 Multi-Memory Module
All programmed and calibrated system data can be permanently stored in the backup memory of the multi-memory module. With functions Param. SE-MMM and Conf. SE-MMM data are stored in module, with Param.MMM-SE and Conf. MMM-SE data are loaded from module. FUNCTION PARAM. SE-MMM Display subfunction distributor. Error Messages:

MULTICONT

CHECKSUM FAULT Data transmission was faulty. WRITE ERROR Repeat function. Exchange READ ERROR module if message is repeated.

P
J=1, N=0 PARAM. SE-MMM

Start program by operating ENTER. Interrogation for acknowledgement.

1P

Data are stored.

2.5 Restoring
This function enables simultaneous loading of the configuration data and parameters stored in the multi memory module. Function becomes active only with weigher being switched OFF. Program Sequence: FUNCTION RECOVER Display in sub-function distributor and start of function. CHECKSUM FAULT Error Messages: SWITCH OFF RECOVER Feeder is not switched off. Only parameters have been loaded. Switch off feeder and restart function. Data transmission was faulty, repeat function. Exchange memory if message is repeated.

P
J=1, N=0_ RECOVER Interrogation for acknowledgement.

1P

Data are loaded. At first parameters are checked and loaded, then configuration data. Errors entail abort of functions.

9947

Modes of Operation and Functions FH 345 GB/ 2 - 12

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
3 Commissioning and Calibration
This chapter all operations for commissioning or calibration of a weighfeeder which are to be performed at the Multicont measuring and control system. For details of the individual operator steps such as, for instance, marginal conditions of set-up programs, see relevant items of Chapter 3. Calibration is largely theoretical; form entered characteristic values (Block P 01, P 02), device constant and nominal belt load are calculated. The norminal best load serves as reference value for display and limit values. All calibrating operations are effected by set-up program.

WEIGHFEEDER

3.4 Taring
As may configured, feeder is tared with taring programs TW (one weighbridge). T2 (second weighbridge) and TF (supply bin).

3.5 Weight Check


The weight check enables mechanical impacts not measured by theoretical calibration to be balanced. For the first weighbridge, program CW is provided.

3.1 Power ON
Before operating the power switch, check mechanical installation of weighfeeder in accordance with manual, and electrical connections in accordance with circuit diagrams. After start, feeder data are loaded from EPROM into active store. This is when system message Schenck Multicont is displayed. If system is ready to operate, message effaces and the current operating state is displayed. For systematic suppression of event messages,occurring e.g. due to a missing release signal, see instructions in Chapter 6.

3.6 Controller Setting


The feeder rate controller (P 04) is preset with standard values. Controllers of prefeeders, second weighbridge, and belt load controller can be set by program Controler Adaptation or in manual.

3.7 Control Signals Check


In test mode, load cell and speed signals can be simulated enabling feeders to be operated in all sub-modes with no load. Thus control signals (e.g. fill control) and operation or peripherals can be checked. With the aid of funcion Service Values signals on measurement data acquisition units and input/output units can be checked.

3.2 System Adaptation


Feeder is factory-configured in accordance with specification. Deviations form specification have to be realized upon commissioning and configuration data and parameters require to be entered accordingly. For rated data of feeder, see data sheets. Analog and pulse outputs as well as analog input of feed rate setpoints can be standardized and adapted by parameters.

3.8 Setting of Belt Influence Compensation and Control to Discharge Point


Belt Influence Compensation (BIC) and control delay for discharge point are set in Blocks K 14 and P 09. Limit value Belt Slip P 06.07 should be set to 1%. For entry of rate data, BIC system requires to be switched off. After acquisition of belt circuit code mumber and setting of belt slip (K03.03) to internal, BIC can be switched on. If parameters P 01.04 VNenn, P 02.07 VSKenn or P 03.07 IB are changed or set-up program LB is active, BIC has to be stopped and restarted.

3.3 Belt Circuit Code Number


The belt circuit code number indicates the number of speed generator pulses per belt circuit. The code number is acquired by program LB before further calibrating operations are effected.
9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Commissioning and Calibration FH 345 GB/ 3 - 1

WEIGHFEEDER
3.9 Local Control Box
The local control box (see also Hardware Manual FH369) is provided for local control effected in manual in Local Mode, preceding all other main modes. The local control box is available in the following versions: FLG 0050 for weighfeeders,

MULTICONT
In Local Mode following functions are performed: Configuration Preselect LocalK 06.01 Mode Start Belt DriveK 06.02 Stop Belt DriveK 06.03 SetpointK 06.01 Prefeeder OPENK 19.04 Prefeeder ShUTK 19.05 Prefeeder ONK 18.03 Prefeeder SetpointK 18.10 Local Control Box Switch FLG 0050 S1 FLG 0050 S2 FLG 0050 S2 FLG 0050 R1 FLG 0053 S2 FLG 0053 S2 FLG 0054 S3 FLG 0054 R1

FLG 0053 for weighfeeders with position-controlled prefeeder, FLG 0054 for weighfeeders with speed-controlled prefeeder.

In its neutral state, Switch S1 can be secured by a lockable bracket.

S2

R1 S2

S1

Fig.3.1: Local Control Box FLG 0050

Fig.3.2: Local Control Box FLG 0053

S3

R1

Fig.3.3: Local Control Box FLG 0054

9947

Commissioning and Calibration FH 345 GB/ 3 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
4 Configuration
4.1 General Comments
The MULTICONT system is adapted to users requirements by configuration. From the scope of available functions only those assemblies and functions required for the application are activated. Analog inputs and outputs are assigned to logical signals and data. The event messages are classified. Entry is via function Configuration (s. 2.4.2). The values for source type, source, and event class can be selected from the admissible range by cursor operation. The system is factory-configured in accordance with the specification. Modifications should exclusively be effected by persons with comprehensive system knowledge. Section 4.2 Survey comprises a list of configuration blocks complete with available signals and functions which enables prompt survey and facilitates the location of individual data. In Section 4.3 Index, all configuration data including default values and setting ranges are listed. The representation corresponds to the display of a machine control station; additional symbols and block names largely reduce the need for explanatory texts.

WEIGHFEEDER

As form of representation a bordered three-line block has been selected. The first line contains the name and (with inputs and outputs) a symbol. The indications in lines 2 and 3 correspond to the display in the setpoint field and in the input/output field of a machine control station. If the lines are dashed, no entry is possible.

Name Number Source Type Source Symbol Event Class

Range: Settable source types / Event classes Reference: To parameters To event messages Note: The note completes representation by additional information.

Together with symbol, source type and event message, the name of the signal describes the function. This name is used in the alphabetic search list. The symbols beside the names refer to the source type.

m n o p q
9947
Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Input

Output

N/C contact

N/O contact

Flank-controlled

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 1

WEIGHFEEDER
The number serves for identification and is shown in the setpoint field together with the symbol. It is composed of block number, position of signal within the block and the letter K. The symbol serves for explanation of signal on display. The source type defines the type of interface, the signal can be assigned to. The signal is either assigned to an input, or output, or generated internally. The following types are possible: DE DA AE AA WZ GA IM SE TT IN .0/.1 Digital Input Digital Output Analog Input Analog Output Load Cell Input Speed Input GA Pulse Output SE Bus IM Keyboard (Machine Control Station) Internally Generated Signal Not Active/Active. TT SE IN The source defines unit and channel, or dummy sources, with logical YES/NO in accordance with type. Depending on stage of extension, the following units can be selected: EA1 - EA3 3 Input/output units FEA .1 (YES) .0 (NO) WZ AA DE (NE) DA (NA) AE

MULTICONT
The configurations listed in the table below are possible: Source Type Unit EA1 -EA3 ME1-ME3 EA1 - EA3 EA1 - EA3 ME1 - ME3 EA1 - EA3 EA1 - EA3 ME1 - ME3 EA1 - EA3 ME1 - ME3 EA1 EA3 LB1 - LB2 ZBS, IPC, PLS Source Channel* DE1 - DE6 DE1 - DE6 YES / NO DA1 -DA7 AE1 AE1 AA1 -AA2 WZ1 WZ1 GA1 GA1 IM1

* The number of available channels depends on the units used. All digital inputs and outputs can be inverted if NE is scrolled into display field in place of DE (NE), or NA in place of DA.

ME1 - EA3 3 Measurement data acquisition units FME LB1, LB2 ZBS PLS IPC YES/NO 2 Machine control stations FLB Group control station FZB Interface to process control system Diagnosis interface (planned) Logical YES/NO

9947

Configuration FH 345 GB/ 4 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
The event class defines the type of event message (see Chapter 6) assigned to the signal. The following message types are available: A W1 W2 IG ALARM WARNING 1 WARNING 2 IGNORE

WEIGHFEEDER

The range indicates the eligible source types, or event classes. If DE (NE) may be chosen as source type, the permissible logical states are indicated with YES/NO .

Example K 00.00 DE(NE) EA2.DE1 BSP A

Range: DE (NE)/SE YES / NO A/W1 See: P 00.00 beispiel E 00 Example Note: To signal Example K 00.00 BSP, a digital input (N/O contact) is assigned as source type. As source, digital input 1 of input/output unit 2 is determined. A missing signal is reported as ALARM. As an alternative, source type SE bus and event class Warning 1 can be assigned to the signal. As sources, all active assemblies with digital inputs, or interfaces, can be configured via SE bus. To source type DE (NE) solely logical state NO can be assigned. If no source is assigned to the signal, -.NO has to be configurated as source, otherwise event message Example would be permanently available.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 3

WEIGHFEEDER
4.2 Survey
Block K 00. Hardware Configuration Block K 04. < > < >

MULTICONT

Safety Indications: Material K 04.00 ST_MAT K 04.01MAT_TK K 04.02 QMIN_AN K 04.03 MAT_MAX

Measurement Data Acquisition Unit 1 Measurement Data Acquisition Unit 2 Measurement Data Acquisition Unit 3 Input/Output Unit 1 Input/Output Unit 2 Input/Output Unit 3 Machine Control Station 1 Machine Control Station 2 Batch Processor Memory SE Multikey No SI Units

K 00.01 ME1_AKT K 00.02 ME2_AKT K 00.03 ME3_AKT K 00.04 EA1_AKT K 00.05 EA2_AKT K 00.06 EA3_AKT K 00.07 LB1_AKT K 00.08 LB2_AKT K 00.09 SE*_AKT K 00.10 SE_MEM K 00.11 MK_AKT K 00.12 NON-SI

Wrong Material Flow Sensor Flap Belt Load q < qMIN Material > MAX

Block K 05. > > > > > < < < < < < < < Emergency Stop Safety Disconnector Release 1 Release 2 Release 3 No Release Electrical Equipment Ready Externally Ready Ready in Mode 0 Ready in Mode 1 Ready to Start in Mode 0 Ready to Start in Mode 1 No EMERGENCY-OFF

External Interlocking K 05.01 NOTAUS K 05.02 FREISCH K 05.03 FREI_1 K 05.04 FREI_2 K 05.05 FREI_3 K 05.06 NO_FREI K 05.07 EL_BER K 05.08 EX_BER K 05.09 M0_BER K 05.10 M1_BER K 05.11 M0_EBER K 05.12 M1_EBER K 05.13 NO-NOT

Block K 01.

Safety Indications: General

Feeder Blocked K 01.01 BLOCKSC Source Local Mode Maintenance Interval Reached K01.01 WARTUNG Block K 06. Standby K 01.03 STANDBY Feeder START K 06.00 LO_EIN Block K 02. Safety Indications: Electrical > Preselect Local Mode K 06.01 LO_VORW Equipment of Feeder > Start Belt Drive K 06.02 LO_STAR > Stop Belt Drive5 K 06.03 LO_STOP < Drives OK K 02.00 ANTR_OK < Preselect Local Mode Active K 06.04 L0_AKT Power Failure K 02.01 NETZAUS > Drive Signal Speed K 06.10 LO_STEL > Error: Feed Drive K 02.02 MS_DAN > Error: Controller Feed Drive K 02.03 DAN_REG Block K 07. Test Mode Sources < Cable Breakage/Short Circuit K 02.04 F_KABEL Setpoint Analog Input Feeder START K 07.00 TM_EIN Range < MIN K 02.05 FP_ANAL Preselect Test Mode K 07.01 TM_VORW > Feed Drive Temperature > MAXK 02.06 TEMP_DO > Feeder START/STOP K 07.02 TM_I/O > Reset Totalizing Counter 1 K 07.03 TM_RES1 Block K 03. Safety Indications: Mechanical > Reset Totalizing Counter 2 K 07.04 TM_RES2 Equipment of Feeder < Test Mode Preselection Active K 07.05 TM_AKT > Absolute Setpoint K 07.10 TM_PABS < Feed Unit OK K 03.00 FOERDER > Belt Tracking K 03.01 SCHIEF Block K 08. Mode 0 Sources > Belt Tension K 03.02 BAND_SP > Belt Slip K 03.03 SCHLUPF Feeder START K 08.00 M0_EIN Warning: Belt Tracking Left K 03.04 SCHIEFL > Preselect Mode 0 K 08.01 M0_VORW Warning: Belt Tracking Right K 03.05 SCHIEFR > Feeder START/STOP K 08.02 M0_I/O > Slip Monitoring Transmitter 1 K 03.06 SCHLU_1 > Reset Totalizing Counter 1 K 08.03 M0_RES1 > Slip Monitoring Transmitter 2 K 03.07 SCHLU_2 > Reset Totalizing Counter 2 K 08.04 M0_RES2 < Mode 0 Preselection Active K 08.05 M0_AKT

9947

Configuration FH 345 GB/ 4 - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
> Absolute Setpoint Relative Setpoint Block K 09. Feeder START Preselect Mode 1 Feeder START/STOP Reset Totalizing Counter 1 Reset Totalizing Counter 2 Mode 1 Preselection Active Absolute Setpoint Relative Setpoint K 08.10 M0_PABS K 08.11 M0_PREL Mode 1 Sources Block K 13.

WEIGHFEEDER
Calibration Event Messages K 13.01 WZ_MIN K 13.02 WZ_MAX K 13.03 GA_ERR K 13.04 SIM_AKT

K 09.00 M1_EIN K 09.01 M1_VORW K 09.02 M1_I/O K 09.03 M1_RES1 K 09.04 M1_RES2 Block K 14. Belt Influence Compensation (BIC), K 09.05 M1_AKT Control to Point of Discharge K 09.10 M1_PABS K 09.11 M1_PREL > Synchronous Pulse K 14.01 SYNCHRO BIC ON K 14.02 BIC_EIN Control Delay ON K 14.03 VAP_EIN Block K 10. Mode-Independent Sources K 14.04 BIC_ERR and Destinations < BIC Asynchronous Belt Tracking Measurement K 14.05 BPM_EIN Belt Drive START K 10.00 ANTR_ON > Acknowledge Event Message K 10.01 STCLR Block K 15. Electrical Equipment of Feeder, > Volumetric Mode K 10.02 VOLUM Extended Version < Message: Feeder Started K 10.03 W_EIN > General Alarm K 10.04 S_ALARM > Feed Drive Overload K 15.01 MS_ZAN < General Warning K 10.05 S_WARN > Aux. Drive 1 Overload K 15.02 MS_H1AN Message: Auxiliary Drive > Aux. Drive 2 Overload K 15.03 MS_H2AN started K 10.06 H_EIN > Controller Feeder Drive K 15.04 ZAN_REG < Totalizing Counter 1 K 10.08 FM1_IA > Prefeeder Temperature > MAX K 15.05 TEMP_ZA < Totalizing Counter 2 K 10.09 FM2_IA > Belt Load K 10.10 QU_q > Belt Speed K 10.11 QU_v < Belt Drive Control Magnitude K 10.12 STG_DO < Actual Feed Rate 1 K 10.13 P_IST1 Block K 16. Mode-Independent Sources < Actual Feed Rate 2 K 10.14 P_IST2 and Destinations, Extended Version < Belt Load K 10.15 Q_BAND > > > > < > Automatic Weight Check Check Weight in Position Stop Totalizing Counter 1 Stop Totalizing Counter 2 Weight Check Prefeeder START/STOP Apply Check Weight Volumetric Belt Speed, Analog Output Setpoint > Taring > > > > > < < < K 16.01 PRG_AUT K 16.02 PRG_LA K 16.03 QU_1ST K 16.04 QU_2ST K 16.05 G_KONTR K 16.06 Z_RIEG K 16.07 PRG_AUF K 16.08 VOL_AKT K 16.10 V_BAND K 16.11 P_SOLL K 16.12 DE_TARA

L/C Input < MIN L/C Input > MAX PLL Overflow Simulator Active

Block K 11. > Activate Emergency Setpoint < Message: Deviation < Message: Emergency Setpoint Active < Controller Out Of Limit < Deviation

Controller K 11.01 QU_NOTS K 11.02 DA_REGA K 11.03 NOTSOL K 11.04 REG_ERR K 11.10 AA_REGA

Block K 12. < < < < < <


9947

Limit Values K 12.01 K_MIN K 12.02 K_MAX K 12.03 Q_MIN K 12.04 Q_MAX K 12.05 V_MIN K 12.06 V_MAX Block K 17. Prefeeder, Intermittent K 17.00 ZI_AON K 17.01 ZI_AKT K 17.02 ZI_NON K 17.03 ZI_LON K 17.04 ZI_BON

Actual Feed Rate < PMIN Actual Feed Rate > PMAX Belt Load < qMIN Belt Load > qMAX Belt Speed < vMIN Belt Speed > vMAX

< Prefeeder Drive START Prefeeder Type Active > Prefeeder START > Prefeeder START Local > Prefeeder START, Bin Monitor

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 5

WEIGHFEEDER
Block K 18. Prefeeder, Speed-Controlled K 18.00 ZD_AON K 18.01 ZD_AKT K 18.02 ZD_NON K 18.03 ZD_LON K 18.10 ZD_LST K 18.11 ZD_NST Block K 22.

MULTICONT
Types of Check System

< Prefeeder Drive START Prefeeder Type Active < Prefeeder START > Prefeeder START Local > Drive Signal Prefeeder, Local < Control Magnitude Prefeeder Block K 19.

Prefeeder, Position-Controlled K 19.00 ZS_AON K 19.01 ZS_AKT K 19.02 ZS_MAX K 19.03 ZS_MIN K 19.04 ZS_LAUF K 19.05 ZS_LZU K 19.06 ZS_AAUF K 19.07 ZS_AZU K 19.15 ZS_STM K 19.16 ZS_ADO

Check System Type EEIN Active Check System Type EAUS Active Check System Type EXTERN Active Cyclic Auto Start Tare Cyclic Auto Start Span

K 22.01 KME_EIN K 22.02 KME_AUS K 22.03 KME_EXT K 22.04 KME_TAU K 22.05 KME_BAU

< Prefeeder Drive ON Prefeeder Type Active > Prefeeder Limit Position MAX > Prefeeder Limit Position MIN > Prefeeder OPEN, Local > Prefeeder CLOSE Local < Prefeeder Drive OPEN < Prefeeder Drive CLOSE > Prefeeder Position Signal < Prefeeder Position Signal

Block K 23. > > > > > > <

Internal Check System K 23.01 KME_FR K 23.02 KME_SBI K 23.03 KME_STI K 23.04 KME_KUI K 23.05 KME_KNO K 23.06 KME_FEA K 23.07 KME_AKT K 23.08 KMEBMIN K 23.09 KMEBMAX K 23.10 KME_KVM K 23.11 KME_KTA K 23.12 KME_KTR K 23.13 KME_KBA K 23.14 KME_KBR K 23.15 KME_KAN K 23.16 KME_KVO K 23.17 KME_KON K23.18KME_WQU

Block K 20. > Preselect Batching Mode 0 > Preselect Batching Mode 1 > Preselect Batching Mode Test Mode > Abort Batch, Mode 0 > Abort Batch, Mode 1 > Abort Batch, Test Mode < Pre-contact < Main Contact Batch Setpoint Mode 0 Batch Setpoint Mode 1 Batch Setpoint Test Mode

Batching Mode K 20.01 M0_BATQ K 20.02 M1_BATQ K 20.03 TM_BATQ K 20.04 M0_BATA K 20.05 M1_BATA K 20.06 TM_BATA K 20.07 VORKON K 20.08 HAUPTK K 20.10 M0_PBAT K 20.11 M1_PBAT K 20.12 TM_PBAT < < < < < < < <

Release Check System Start Span Correction Start Zero Point Correction Accept Correction Do Not Accept Correction Fill/Discharge Unit Stopped Message: Check Measurement Started Minimum Check Fill Weight Maximum Check Fill Weight Check Quantity < MIN Corr. Range Exceeded, Abs. Zero Point Corr. Range Exceeded, Rel. Zero Point Corr. Range Exceeded, Abs. Span Corr. Range Exceeded, Rel. Span Check Logged On Prepare Check Measurement Check Measurement Active Check System Expects Acceptance

Block K 21. > Fill Weight > Fill Weight Control START > Initialize Controller Discharge/Fill Control < Fill Weight < BMIN < Fill Weight > BMAX < Bin Filling START Load Cell Input Fill Weight Underamplified Load Cell Input Fill Weight Overamplified < Fill Weight < Fill Weight Control Magnitude

Fill Weight, Bin Block K 24. K 21.01 QU_FGVB K 21.02 B_RON K 21.03 B_INIT K 21.04 B_TYP_A K 21.05 FVB_MIN K 21.06 FVB_MAX K 21.07 FILORVB K 21.08 WZVB_MI K 21.09 WZVB_MA K 21.10 FGVB K 21.11 B_STG > Moisture Correction ON > Material Moisture < Actual Feed Rate, Dry Counter 1, Pulse Output Corrected K 25.01 FEU_EIN K 25.02 MATFEU K 25.03 PTROCK K 25.04 FM1_TR
9947

External Check System K 24.01 KME_KUB K 24.02 KME_KUT K 24.03 KME_STE K 24.04 KME_KGB K 24.05 KMT_KGT Moisture Correction

Accept Span Correction Accept Zero Point Correction Start External Check System Correction Value: Span Correction Value: Zero Point Block K 25.

Configuration FH 345 GB/ 4 - 6

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Counter 2, Pulse Output Corrected < Moisture MAX < Moisture Correction Active Material Moisture, Wrong Analog Input K 28. K 25.05 FM2_TR K 25.06 FEU_MAX K 25.07 FEU_AKT K 25.08 FEU_FAN

WEIGHFEEDER
Interpulse Period Controlled Prefeeder K 28.00 ZP_AON K 28.01 ZP_AKT K 28.02 ZP_NON K 28.03 ZP_LON K 28.04 ZP_LST

< START Prefeeder Drive Prefeeder Type Active > START Prefeeder > START Local Prefeeder > Local Prefeeder Setpoint

Block 26.

Second Weighbridge

Second Weighbridge K 26.01 WB2_WZ Load Cell Input 2nd Weighbridge under amplified K 26.02 WB2_MIN Load Cell Input 2nd Weighbridge over amplified K 26.03 WB2_MAX Belt Load 2nd Weighbridge Too Small K 26.04 wb2_qmin Belt Load 2nd Weighbridge Too High K 26.05 wb2_qmax Check Weighbridge K 26.06 Wb2_ctl Check Weighbridge Event Message K 26.07 wb2_dif 2nd Weighbridge K 26.08 WB2_AA 2nd Weighbridge Only K 26.09 WB2_AUS Block K 27. > Dead Time Element Setpoint Preselect M0 > Dead Time Element Setpoint Preselect M1 > Dead Time Element Setpoint Clock ON M0 > Dead Time Element Setpoint Clock ON M1 > Preassign Setpoint Dead Time Element > Initialize Setpoint Dead Time Element > Dead Time Element START/STOP Preselet M0 > Dead Time Element START/STOP Preselect M1 > Dead Time Element START/STOP, Clock ON M0 > Dead Time Element START/STOP, Clock ON M1 < Dead Time Elements Preselected < Dead Time Element Setpoint Active < Dead Time Element START/STOP Active
9947

Dead Time Element

K 27.01 TOT_VM0 K 27.02 TOT_VM1 K 27.03 TOT_TM0 K 27.04 TOT_TM1 K 27.05 TOT_VOR K 27.06 TOT_INI K 27.07 TOT_IM0 K 27.08 TOT_IM1 K 27.09 TOT_TI0 K 27.10 TOT_TI1 K 27.11 TOT_GEW K 27.12 TOT_PON K 27.13 TOT_ON

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 7

WEIGHFEEDER
4.3 List of Configuration Data

MULTICONT

K 00. Hardware Configuration

In this block the assemblies used require to be activated, assemblies not used must be de-activated. Measurement Data Acquisition Unit 1 K 00.01 .1 Range: .0 / .1 See: SY 01 ME1 DATA LINK Measurement Data Acquisition Unit 2 Input/Output Unit 3 K 00.02 .0 Range: .0/.1 See: SY 02 ME2 DATA LINK ME2_AKT K 00.06 A .0 A Range: .0 / .1 See: SY 06 EA3 DATA LINK Machine Control Station 1 K 00.07 .1 LB1_AKT A A A EA3_AKT ME1_AKT A A K 00.05 .0 Range: .0 / .1 See: SY 05 EA2 DATA LINK Input/Output Unit 2 EA2_AKT A A

Measurement Data Acquisition Unit 3 K 00.03 .0 Range: .0/.1 See: SY 03 ME3 DATA LINK Input/Output Unit 1 K 00.04 .1 Range: .0 / .1 See: SY 04 EA1 DATA LINK EA1_AKT A / IG K 00.08 A .0 ME3_AKT A A

Range: .1 A /I G See: SY 07 LBS DATA LINK Note: Machine control station 1 cannot be de-activated. Configurate the event message as IG. Machine Control Station 2 LB2_AKT A A / IG
9947

Range: .0 / .1 See: SY 08 LBS2 DATA LINK

Configuration FH 345 GB/ 4 - 8

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Batch Processor K 00.09 .0 Range: .0 / .1 See: SY 09 SE* DATA LINK SE*_AKT A A / W2 K 00.11 .0 Range: .0 /.1

WEIGHFEEDER
Multikey MK_AKT

No SI Units Memory SE K 00.10 IN See: SY 10 MEMORY FAIL SE_MEM A K 00.12 .0 Range: .0 / .1 Reference: Chapter 5.1 Parameter designation Note: The international units (SI) are converted into American units. Related parameters, current readings and setpoints are converted upon activation. The SE bus message continues transmitting data in SI untis. An active batch is aborted, a running check is terminated. NON-SI

K01. Safety Indications: General

Feeder Blocked K 01.01 IN Range: A / IG See: SC 01 FEEDER BLOCKED Parameter P 04.10 Maintenance Interval Reached K 01.02 IN See: P 02.10 wartint SC 02 MAINTENANCE INTERVAL WARTUNG W1 BLOCKSC A K 01.03 IN

Stand-By standby W1

Range: W1 / W2 / IG See: P 04.11 SC 07 Standby

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 9

WEIGHFEEDER

MULTICONT

K 02. Safety Indications: Electrical Equipment of Feeder

m o
K 02.00 DA(NA) Range: DA(NA)

Drives OK ANTR_OK -

mo
K 02.04 DA(NA)

Cable Breakage/Short Circuit F_KABEL A

Power Failure K 02.01 IN Range: See: WE 01 POWER DOWN NETZAUS A A / W1 / IG

Range: DA (NA) A / IG See: WE 04 CABLE SHORT OR INT. Note: Only applies to NAMUR inputs. Special case: IG Contact connection possible. Open : Level 1 Closed : Level 0

Setpoint Analog Input Range < MIN K 02.05 IN FP_ANAL A A / IG

np
K 02.02 DE(NE)

Error: Feed Drive MS_DAN EA1.DE2 A A

Range: See: P 08.01 adan1 P 08.02 adbe1 WE 10 ANALOG INPUT P

Range: DE (NE) NO See: WE 02 MS: FEED DRIVE OVERLOAD

no
K 02.06 DE(NE)

Feed Drive Temperature > MAX TEMP_DO -.NO A A

np
K 02.03 DE(NE)

Error: Controller Feed Drive DAN_REG -.NO A A Range: DE (NE) NO See: WE 12 FEED DRIVE MAX TEMP

Range: DE (NE) NO See: WE 03 MS: FEED DRIVE SCR FAIL

9947

Configuration FH 345 GB/ 4 - 10

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

K 03. Safety Indications: Mechanical Equipment of Feeder

mo
K 03.00 DA(NA) Range: DA (NA)

Feed Unit OK FOERDER -

np
K 03.03 DE(NE)

Belt Slip SCHLUPF -.NO A

np
K 03.01 DE(NE)

Belt Tracking SCHIEF ME1.DE6 A

Range: DE (NE) / IN NO A / W1 / W2 See: WM 03 BELT SLIP K 14.01 SYNCHRO P 02.11 s_lng Note: DE Monitoring for belt slip by limit switches. IN Monitoring for belt slip by sensor. Message remains stored until acknowledgement. Warning: Belt Tracking Left K 03.04 IN SCHIEFL W1

Range: DE (NE) / IN NO A / W1 See: WM 01 BELT TRACKING K 03.04 SCHIEFL K 03.05 SCHIEFR K 14.01 SYNCHRO K 14.05 BPM_EIN p 02.11 ml_sfl Note: DE Belt tracking monitoring by limit switches. IN Belt tracking monitoring by sensor. In addition K 14.01, K 14.05 and P 02.11 require to be set accordingly. Warning Belt Tracking Right/Left is output only in case of sensor monitoring.

Range: IN W1 / W2 / IG See: K 03.01 SCHIEF WM 04 BELT TRACKING LEFT Note: Only with sensor monitoring. Message remains stored until acknowledgement.

np
K 03.02 DE(NE)

Belt Tension BAND_SP -.NO A K 03.05 IN

Warning: Belt Tracking Right SCHIEFR W1

Range: DE (NE) NO A / W1 See: WM 02 BELT TENSION Note: Belt tension can only be monitored by limit switches.

Range: IN W1 / W2 / IG See: K 03.01 SCHIEF WM 05 BELT TRACKING RIGHT Note: Only with sensor monitoring. Message remains stored until acknowledgement.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 11

WEIGHFEEDER

MULTICONT

no
K 03.06 DE (NE)

Slip Monitoring Transmitter 1 SCHLU_1 .NO

no
K 03.07 DE (NE)

Slip Monitoring Transmitter 2 SCHLU-2 -.NO

Range: DE (NE) NO See: K 03.07, P 06.08, Message WM03, K 03.03 For details, see P 06.08.

Range: DE (NE) NO See: K 03.06, P 06.08, Message WM03, K 03.03 For details, see P 06.08.

K 04. Safety Indications: Material

mo
K 04.00 DA(NA)

Wrong Material Flow ST_MAT -

mo
K 04.02 DA(NA)

Belt Load q < qMIN QMIN_AN W1 A / W1 / W2

Range: DA (NA)

Range DA See: P 06.03 qmin P 10.02 fahrtim P 10.03 anfahr MF 02 NO PREFEED

no
K 04.01 DE (NE)

Sensor Flap MAT_TK ME1.DE4 W1

no
K 04.03 DE(NE)

Material > MAX MAT_MAX -.NO W1 A / W1 / W2

Range: DE (NE) NO A / W1 / W2 See MF 01 BULK SOILD SENSOR

Range: DE (NE) NO See: MF 03 MATERIALS > MAX

9947

Configuration FH 345 GB/ 4 - 12

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
K 05. External Interlocking

WEIGHFEEDER

np
K 05.01 DE(NE)

Emergency Stop NOTAUS EA1.DE1 A A

mo
K 05.06 DA(NA)

No Release NO_FREI W2 W2

Range: DE (NE) NO See: IL 01 EMERGENCY STOP

Range: DA (NA) See: IL 03 NO RELEASE

np
K 05.02 DE(NE)

Safety Disconnector FREISCH EA1.DE1 A A

mo
K 05.07 DA(NA)

Electrical Equipment Ready EL_BER -

Range: DE (NE) NO See: IL 02 LOCAL MAIN SWITCH

Range: DA (NA)

no
K 05.03 DE(NE)

Release 1 FREI_1 EA1.DE6 YES / NO

mo
K 05.08 DA(NA) Range: DA (NA)

Externally Ready EX_BER -

Range: DE (NE) / SE

no
K 05.04 DE(NE)

Release 2 FREI_2 -.YES YES / NO

mo
K 05.09 DA(DA) Range: DA (NA)

Ready in Mode 0 M0_BER EA1.DA5

Range: DE (NE) / SE

no
K 05.05 DE(NE)
9947

Release 3 FREI_3 -.NO YES / NO

mo
K 05.10 DA(NA) Range: DA (NA)

Ready in Mode 1 M1_BER -

Range: DE (NE) / SE

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 13

WEIGHFEEDER

MULTICONT

mo
K 05.11 DA(NA)

Ready to Start in Mode 0 M0_EBER -

mo
K 05.13 DA(NA)

No EMERGENCY-OFF NO-NOT -

Range: DA (NA)

mo
K 05.12 DA(NA)

Ready to Start in Mode 1 M1_EBER -

Range: DA (NA) Message No EMERGENCY STOP = NO is not cleared before event message IL01 is reset and input signal EMERGENCY STOP inactive.

Range: DA (NA)

K 06. Source Local Mode

Feeder START K 06.00 IN LO_EIN

n oq
K 06.02 DE(NE) Range: DE (NE)

Start Belt Drive LO_STAR ME1.DE2 YES / NO

no
K 06.01 DE(NE)

Preselect Local Mode LO_VORW ME1.DE1 YES / NO

n oq
K 06.03 DE(NE) Range: DE (NE)

Stop Belt Drive LO_STOP ME1.DE3 YES / NO

Range: DE (NE)

Note: Contact precedes all other main modes. If Local Mode is cut off, feeder automatically stops (without clearance) and returns to its previous mode. After power failure, feeder is in Mode 0.

9947

Configuration FH 345 GB/ 4 - 14

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

mo
K 06.04 DA(NA)

Preselect Local Mode Active L0_AKT -

n
K 06.10 AE

Drive Signal Speed LO_STEL ME2.AE1

Range: DA (NA)

Range: AE See: P 08.03 LDr_Anh

K 07. Test Mode Sources

Feeder START K 07.00 IN TM_EIN

n oq
K 07.03 TT

Reset Totalizing Counter 1 TM_RES1 LB1

Range: DE (NE) / SE / TT

no
Preselect Test Mode K 07.01 TT Range: SE / TT LB1 TM_VORW K 07.04 TT

Reset Totalizing Counter 2 TM_RES2 LB1

Range: DE (NE) / SE / TT Test Mode Preselection Active K 07.05 TM_AKT -

n oq
K 07.02 TT

Feeder START/STOP TM_I/O LB1

DA Range: DA (NA)

Range: DE (NE) / SE / TT K 07.10 TT

Absolute Setpoint TM_PABS LB1

Range: AE / SE / TT
9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 15

WEIGHFEEDER
K 08. Mode 0 Sources

MULTICONT

Feeder START K 08.00 IN M0_EIN

no
K 08.04 SE

Reset Totalizing Counter 2 M0_RES2 ZBS

Range: DE (NE) / SE / TT

n oq
K 08.01 TT

Preselect Mode 0 M0_VORW LB1

mo
K 08.05 DA(NA)

Mode 0 Preselection Active M0_AKT -

Range: DE (NE) / SE / TT

n oq
K 08.02 SE

Feeder START/STOP M0_I/O ZBS

n
K 08.10 SE

Absolute Setpoint M0_PABS ZBS

Range: DE (NE) / SE / TT

Range: AE / SE / TT

n oq
K 08.03 SE

Reset Totalizing Counter 1 M0_RES1 ZBS K 08.11 .0

Relative Setpoint M0_PREL

Range: DE (NE) / SE / TT

Range: SE / TT / .0

9947

Configuration FH 345 GB/ 4 - 16

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
K 09. Mode 1 Sources

WEIGHFEEDER

Feeder START K 09.00 IN M1_EIN

no
K 09.04 TT

Reset Totalizing Counter 2 M1_RES2 LB1

Range: DE (NE) / SE / TT

n oq
K 09.01 TT

Preselect Mode 1 M1_VORW LB1

mo
K 09.05 DA(NA)

Mode 1 Preselection Active M1_AKT -

Range: DE (NE) / SE / TT

n oq
K 09.02 TT

Feeder START/STOP M1_I/O LB1

n
K 09.10 TT

Absolute Setpoint M1_PABS LB1

Range: DE (NE) / SE / TT

Range: AE / SE / TT

no
K 09.03 TT

Reset Totalizing Counter 1 M1_RES1 LB1 K 09.11 M1_PREL

Relative Setpoint

Range: DE (NE) / SE / TT .0 Range: SE / TT / .0

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 17

WEIGHFEEDER

MULTICONT

K 10. Mode-Independent Sources and Destinations

mo
K 10.00 DA(NA) Range: DE (NE)

Belt Drive START ANTR_ON EA1.DA1

mp
K 10.04 DA(NA) Range: DA (NA)

General Alarm S_ALARM EA1.DA2

no
K 10.01 DE(NE)

Acknowledge Event Message STCLR EA1.DE4

mp
K 10.05 DA(NA) Range: DA (NA)

General Warning S_WARN EA1.DA7

Range: DE (NE) YES / NO Note: Independent of configuration, SE bus and keyboard of a machine control station act as sources with equal priority.

no
K 10.02 DE(NE)

Volumetric Mode VOLUM -.NO

mo
K 10.06 DA(NA)

Message: Auxiliary Drive started H_EIN -

Range: DE (NE) YES / NO Note: Independent of configuration, SE bus and keyboard of a machine control station act as sources with equality of access.

Range: DA (NA) Note: Signal is identical with K 10.03 W_EIN.

mo
K 10.03 DA(NA)

Message: Feeder Started W_EIN -

m
K 10.08 IM

Totalizing Counter 1 FM1_IA EA1.IM1

Range: DA (NA)

Range: IM See: P 01.05 fm1dim P 01.06 fm1puls

9947

Configuration FH 345 GB/ 4 - 18

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

m
K 10.09 IM

Totalizing Counter 2 FM2_IA EA2.IM1

m
K 10.13 AA

Actual Feed Rate 1 P_IST1 EA1.AA2

Range: IM See: P 01.07 fm2dim P 01.08 fm2puls

Range: AA See: P 07.01 P1_Anh P 07.02 P1_Ber

n
K 10.10 WZ

Belt Load QU_q ME1.WZ1 A

m
K 10.14 AA

Actual Feed Rate 2 P_IST2 -

Range: WZ See: SY 11 q WRONG UNIT

Range: AA See: P 07.03 P2_Anh P 07.04 P2_Ber

n
K 10.11 GA

Belt Speed QU_v ME1.GA1 A

m
K 10.15 AA

Belt Load Q_BAND EA2.AA2

Range: GA / IN A See: SY 12 v WRONG UNIT Note: With configuration IN speed measurement is cut off. Belt Drive Control Magnitude STG_DO EA1.AA1

Range: AA See: P 07.05 Q_Anh P 07.06 Q_Ber

K 10.12 AA Range: AA

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 19

WEIGHFEEDER
K 11. Controller

MULTICONT

no
K 11.01 DE(NE)

Activate Emergency Setpoint QU_NOTS -.NO YES / NO

mo
K 11.04 DA(NA)

Controller Out Of Limit REG_ERR W1

Range: DE (NE) / SE See: P 04.05 notsol

Range: W1 / W2 / IG See: CO 03 CONTROLLER OUT LIMIT

mo
K 11.02 DA(NA)

Message: Deviation DA_REGA EA2.DA2 W1 A / W1 / W2

m
K 11.10 AA

Deviation AA_REGA -

Range:DA(NA) See: P 04.03 difmax P 04.04 regtim CO 01 DEVIATION

Range: AA See: P 04.05 rtau P 07.09 r_Anh P 07.10 r_Ber

mo
K 11.03 DA(NA)

Message: Emergency Setpoint Active NOTSOL W1

Range:DA(NA) W1 / W2 / IG See: P 01.02 pnenn P 04.12 notsol CO 02 SETPOINT LIMITED

9947

Configuration FH 345 GB/ 4 - 20

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
K 12. Limit Values

WEIGHFEEDER

mo
K 12.01 DA(NA)

Actual Feed Rate < PMIN K_MIN EA1.DA3 W2

mo
K 12.05 DA(NA)

Belt Speed < vMIN V_MIN EA2.DA6 W2 A / W1 / W2 / IG

Range: DA (NA) A / W1 / W2 / IG See: P 06.01 pmin LO 01 FEED RATE I LO

Range: DA (NA) See: P 06.05 vmin LO 03 SPEED v LO

mo
K 12.02 DA(NA)

Actual Feed Rate > PMAX K_MAX EA2.DA4 W2 A / W1 / W2 / IG

mo
K 12.06 DA(NA)

Belt Speed > vMAX V_MAX EA2.DA7 W2 A / W1 / W2 / IG

Range: DA (NA) See: P 06.02 Pmax HI 01 FEED RATE I HI

Range: DA (NA) See: P 06.06 vmax HI 03: SPEED v HI

mo
K 12.03 DA(NA)

Belt Load < qMIN Q_MIN W2 A / W1 / W2 / IG

Range: DA (NA) See: P 06.03 qmin LO 02 LOAD Q LO

mo
K 12.04 DA(NA)

Belt Load > qMAX Q_MAX W2 A / W1 / W2 / IG

Range: DA (NA) See: P 06.04 qmax HI 02 LOAD Q HI


9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 21

WEIGHFEEDER
K 13. Calibration Event Messages

MULTICONT

L/C Input < MIN K 13.01 IN Range: IN See: CA 01 LC OUT < MIN WZ_MIN A A / W1

L/C Input > MAX K 13.02 IN Range: IN See: CA 02 LC OUT > MAX WZ_MAX A A / W1

PLL Overflow K 13.03 IN Range: IN See: K 10.11 Qu_v CA 03: PLL OVERFLOW Note: Belt speed input is overamplified. GA_ERR A

Simulator Active K 13.04 IN Range: IN See: CA 04 SIMULATOR ACTIVE SIM_AKT W2 W2

9947

Configuration FH 345 GB/ 4 - 22

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

K 14. Belt Influence Compensation (BIC), Control to Point of Discharge

no
K 14.01 DE(NE)

Synchronous Pulse SYNCHRO -.NO

mo
K 14.04 DA(NA)

BIC Asynchronous BIC_ERR W1

Range: ME1.DE6/EA1.DE6/NO See: P 02.02 tband P 02.11 s_lng Note: Input for signal of marking area in conveyor belt. Assignment of input actuates belt length measurement rendering Parameter P 02.02 inactive.

Range: DA (NA) W1/W2/IG See: CA 06 BIC ASYNCHRONOUS

BIC ON K 14.02 .0 Range: .0 / .1 See: K 14.01 SYNCHRO P 09.01 BAT_Fil Note: Only in conjunction with K 14.01. Filter setting with P 09.01. BIC_EIN K 14.05 .0

Belt Tracking Measurement BPM_EIN

Range: .0 / .1 See: K 03.01 SCHIEF K 14.01 SYNCHRO Note: Belt tracking measurement is possible only if belts are equipped with sensors.

Control Delay ON K 14.03 .0 Range: .0 / .1 See: K 14.01 SYNCHRO P 09.02 DAP_LNG Note: Control delay for discharge synchronous pulse only. VAP_EIN

point

with

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 23

WEIGHFEEDER

MULTICONT

K 15. Electrical Equipment of Feeder, Extended Version

np
K 15.01 DE(NE)

Feed Drive Overload MS_ZAN EA1.DE5 A

no
K 15.05 DE(NE)

Prefeeder Temperature > MAX TEMP_ZA -.NO A A / W1

Range: DE (NE) NO A / W1 See: WE 05 MS: PREFEEDER OVERLOAD

Range: DE (NE) NO See: WE 09 PREFEEDER TEMP. HI

np
K 15.02 DE(NE)

Aux. Drive 1 Overload MS_H1AN -.NO W1 @BLOCK 6 =

Range: DE (NE) NO A / W1/ W2 / IG See WE 06 MS: DRIVE 1 OVERLOAD

np
K 15.03 DE(NE)

Aux. Drive 2 Overload MS_H2AN -.NO W1

Range: DE (NE) NO A / W1/ W2 / IG See: WE 07 MS: DRIVE 2 OVERLOAD

np
K 15.04 DE(NE)

Controller Feeder Drive ZAN_REG -.NO A A / W1

Range: DE (NE) NO See: WE 08 PREFEEDER SCR FAIL

9947

Configuration FH 345 GB/ 4 - 24

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

K 16. Mode-Independent Sources and Destinations, Extended Version

Automatic Weight Check K 16.01 .0 PRG_AUT

no
K 16.05 TT

Weight Check G_KONTR LB1

Range: .0 / .1 Range: DE (NE) / SE / TTNO 0: Call of calibrating program independent of operating source K 16.05 via keypad or DE (NE) Call of calibrating program from preselected source K 16.05 only

1:

no
K 16.06 TT

Prefeeder START/STOP Z_RIEG LB1

no
K 16.02 DE(NE)

Check Weight in Position PRG_LA EA2.DE4 W1

Range: DE (NE) / SE / TTNO See: K 17.02 ZI_NON K 18.02 ZD_NON

Range: DE (NE) YES / NO W1 / W2 / IG See: CA 05 CHECK WEIGHT IN POS.

mo
K 16.07 DA(NA)

Apply Check Weight PRG_AUF EA2.DA5

no
K 16.03 DE(NE)

Stop Totalizing Counter 1 QU_Z1ST -.NO YES / NO

Range: DA (NA)

Range: DE (NE) / SE

mo
K 16.08 DA(NA) QZ_2ST Range: DA (NA)

Volumetric VOL_AKT -

no
K 16.04 DE(NE)
9947

Stop Totalizing Counter 2

-.NO YES / NO

Reports Volumetric Mode of start-up and clearance period.

Range: DE (NE) / SE

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 25

WEIGHFEEDER
In stop state of weigher, contact is open.

MULTICONT

m
K 16.10 AA

noq
K 16.12 DE (NE) V_BAND

P 07.12 Ps_Ber Taring DE_TARA -,NO NO

Belt Speed, Analog Output

Range: DE (NE) -

Range: AA See: P 07.07 v_Anh. P 07.08 v_Ber. @BLOCK 1 = Setpoint K 16.11 AA Range: AA See : P 07.11 Ps_Anh P_SOLL

Positive edge starts, negative edge aborts taring program. After elapse of taring program, result is automatically acquired. Contact acts in parallel to residual operating sources.

K 17. Prefeeder, Intermittent

mo
K 17.00 DA(NA)

Prefeeder Drive START ZI_AON EA1.DA6

no
K 17.02 DE(NE)

Prefeeder START ZI_NON -.YES

Range: DA (NA) See: P 26.01

Range: DE (NE) / SE YES / NO See: K 16.06 Z_RIEG Note: Applies to M0, M1, and Test Mode. Prefeeder requires to be additionally started with K 16.06.

m oq
K 17.01 .0 Range: .0 / .1

Prefeeder Type Active ZI_AKT

no
K 17.03 DE(NE)

Prefeeder START Local ZI_LON ME1.DE5 YES / NO


9947

Range: DE (NE)

Configuration FH 345 GB/ 4 - 26

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

no
K 17.04 DE(NE)

Prefeeder START, Bin Monitor ZI_BON ME1.DE4 YES / NO

Range: DE (NE)

K 18. Prefeeder, Speed-Controlled

mo
K 18.00 DA(NA)

Prefeeder Drive START ZD_AON EA1.DA6

n oq
K 18.03 DE(NE)

Prefeeder START Local ZD_LON ME1.DE5 YES / NO

Range: DA (NA)

n
ZD_AKT

Range: DE (NE)

Drive Signal Prefeeder, Local ZD_LST ME2.AE1

Prefeeder Type Active K 18.01 .0 Range: .0 / .1

K 18.10 AE

Range: AE See : P 13.01 zd_lanh P 13.02 zd_lber

mo
K 18.02 DE(NE)

Prefeeder START ZD_NON -.YES

m
K 18.11 AA

Control Magnitude Prefeeder ZD_NST EA2.AA1

Range: DE (NE) YES / NO See: K 16.06 Z_RIEG Note: Aplies to M0, M1, and Test Mode Prefeeder requires to be additionally started with K 16.06.

Range: AA

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 27

WEIGHFEEDER

MULTICONT

K 19. Prefeeder, Position-Controlled

mo
K 19.00 DA(NA)

Prefeeder Drive ON ZS_AON EA1.DA6

no
K 19.05 DE(NE)

Prefeeder CLOSE Local ZS_LZU ME2.DE2 YES / NO

Range: DA (NA)

Range: DE (NE)

Prefeeder Type Active K 19.01 .0 Range: .0/.1 ZS_AKT

mo
K 19.06 DA(NA)

Prefeeder Drive OPEN ZS_AAUF EA2.DA3

Range: DA (NA)

no
K 19.02 DE(NE)

Prefeeder Limit Position MAX ZS_MAX ME2.DE4 YES / NO

mo
K 19.07 DA(NA)

Prefeeder Drive CLOSE ZS_AZU EA2.DA4

Range: DE (NE)

Range: DA (NA)

no
K 19.03 DE(NE)

Prefeeder Limit Position MIN ZS_MIN ME2.DE5 YES / NO

n
K 19.15 AE

Prefeeder Position Signal ZS_STM ME2.AE1

Range: DE (NE)

no
K 19.04 DE(NE)

Range: AE See: P 14.01 zs_anh P 14.02 zs_ber

Prefeeder OPEN, Local ZS_LAUF ME2.DE1 YES / NO


9947

Range: DE (NE)

Configuration FH 345 GB/ 4 - 28

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

m
K 19.16 AA

Prefeeder Position Signal ZS_ADO -

Range: AA See: P 14.07 zs_aanh P 14.08 zs_aber

For Pulse Period Controlled Prefeeder, see K 28.

K 20. Batching Mode

See: Block 15.

no
K 20.01 SE

Preselect Batching Mode 0 M0_BATQ

no
K 20.03 TT

Preselect Batching Mode Test Mode TM_BATQ LB1

ZBS

Range: DE (NE) / SE / TTYES / NO Range: DE (NE) / SE / TTYES / NO

no
K 20.02 TT

Preselect Batching Mode 1 M1_BATQ

no
K 20.04 SE

Abort Batch, Mode 0 M0_BATA ZBS

LB1 Range: DE (NE) / SE / TTNO

Range: DE (NE) / SE / TTYES / NO

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 29

WEIGHFEEDER

MULTICONT
Batch Setpoint Mode 0 M1_BATA K 20.10 SE Range: SE / TT ZBS M0_PBAT

no
K 20.05 TT

Abort Batch, Mode 1

LB1

Range: DE (NE) / SE / TTNO

no
K 20.06 TT

Abort Batch, Test Mode TM_BATA LB1 K 20.11 TT

Batch Setpoint Mode 1 M1_PBAT LB1

Range: DE (NE) / SE / TTNO

Range: SE / TT

mo
K 20.07 DA(NA) Range: DA (NA)

Pre-Contact VORKON K 20.12 TT

Batch Setpoint Test Mode TM_PBAT LB1

Range: SE / TT

mo
K 20.08 DA(NA) Range: DA (NA)

Main-Contact HAUPTK -

K 20.09

Not dedicated

9947

Configuration FH 345 GB/ 4 - 30

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
K 21. Fill Weight, Bin

WEIGHFEEDER

n
K 21.01 WZ

Fill Weight QU_FGVB A A

mo
K 21.05 DA(NA)

Fill Weight < BMIN FVB_MIN W2 W1/ W2 / IG

Range: WZ / SE See: SY 12 qb WRONG UNIT

Range: DA (NA) See: P 18.04 vbmin LO 04 BIN LEVEL LO

no
K 21.02 DE(NE)

Fill Weight Control START B_RON -.NO YES / NO

mo
K 21.06 DA(NA)

Fill Weight > BMAX FVB_MAX W2 W1 / W2 / IG

Range: DE (NE)

Range: DA (NA) See: P 18.05 vbmax HI 04 BIN LEVEL HI

no
K 21.03 DE(NE)

Initialize Controller B_INIT -.NO YES / NO

mo
K 21.07 DA(NA) Range: DA (NA)

Bin Filling START FILORVB

Range: DE (NE)

Discharge/Fill Control K 21.04 .1 Range: .0 /.1 See: .1 = Discharge control .0 = Fill control B_TYP_A

Load Cell Input Fill Weight Underamplified K 21.08 IN Range: IN See: CA 05: LC BIN < MIN WZVB_MI A A / W1

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 31

WEIGHFEEDER
Load Cell Fill Weight Overamplified K 21.09 IN Range: IN See: CA 08 LC BIN > MAX Fill Weight FGVB WZVB_MA A A / W1 K 21.11 AA

MULTICONT
Fill Weight Control Magnitude B_STG -

Range: AA See: P 19.12 vbanh P 19.13 B_go P 19.14 B_gu

K 21.10 AA

Range: AA See: P 16.06 vbanh P 16.07 vbdakorm

22. Types of Check System

Check System Type EEIN Active K 22.01 .0 Range: .0 / .1 KME_EIN K 22.04 .0 Range: .0 / .1

Cyclic Auto Start Tare KME_TAU

Check System Type EAUS Active K 22.02 .0 Range: .0 /.1 Check System Type EXTERN Active K 22.03 .0 Range: .0 / .1 KME_EXT KME_AUS K 22.05 .0 Range: .0 / .1

Cyclic Auto Start Span KME_BAU

9947

Configuration FH 345 GB/ 4 - 32

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
K 23. Internal Check System

no
K 23.01 DE (NE)

Release Check System KME_FR -.YES YES / NO

no
K 23.06 DE(NE)

Fill/Discharge Unit Stopped KME_FEA -.NO W2

Range: DE (NE) / SE

n oq
K 23.02 TT

Range: DE (NE) YES / NO W2 / IG See: CH 08 FEEDBACK PREFEEDER

Start Span Correction KME_SBI LB1 NO

mo
K 23.07 DA(NA)

Message: Check Measurement Started KME_AKT -

Range: DE (NE) / SE / TT

n oq
K 23.03 TT

Range: DA (NA) Start Zero Point Correction KME_STI LB1 NO K 23.08 IN See: KME_KUI Maximum Check Fill Weight LB1 NO K 23.09 IN KMEBMAX W1 CH 03 BIN LEVEL HI CH 02 BIN LEVEL LO Minimum Check Fill Weight KMEBMIN W1

Range: DE (NE) / SE / TT

no
K 23.04 TT

Accept Correction

Range: DE (NE) / SE / TT

no
K 23.05 TT

Do Not Accept Correction KME_KNO LB1 NO

See:

Check Quantity < MIN K 23.10 IN KME_KVM W1 CH 01 CHECK QUANTITY < MIN

Range: DE (NE) / SE / TT

9947

See:

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 33

WEIGHFEEDER

MULTICONT

mo
K 23.11 DA(NA)

Corr. Range Exceeded, Abs. Zero Point KME_KTA

mo
K 23.16 DA(NA)

Prepare Check Measurement KME_KVO W2 W2 / IG

W1 Range: DA (NA) See: CH 10 PREPARATION ACTIVE

Range: DA (NA) See: CH 04 CORRECTION 0 ABS HI

mo
K 23.12 DA(NA)

Corr. Range Exceeded, Rel. Zero Point KME_KTR W1

mo
K 23.17 DA(NA)

Check Measurement Active KME_KON -

Range: DA (NA)

Range: DA (NA) See: CH 05 CORRECTION 0 REL HI

mo
K 23.13 DA(NA)

mo
K 23.18 KME_KBA DA(NA) W1

Check System Expects Acceptance KME_WQU W2 W2 / IG

Corr. Range Exceeded, Abs. Span

Range: DA (NA) See: CH 06 CORR. SPAN ABS. HI

Range: DA (NA) See: CH 11 ACCEPTANCE REQ.

mo
K 23.14 DA(NA)

Corr. Range Exceeded, Rel. Span KME_KBR

mo
K 23.19 DA(NA) Range: AA

KME Result KME_WQU W2

W1

Range: DA (NA) See: CH 07 CORR. SPAN REL. HI

Output of check system measuring result F/Z. F = Static scale measuring result Z = Solids flow feeder measuring result KME_KAN Analog output calibration cannot be set. F/Z = 0,9 : Output signal = 1,0 : Output signal = 1,1 : Output signal 4 mA 12 mA 20 mA (-10%) ( 0%) (+10%)

mo
K 23.15 DA(NA)

Check Logged On

W2 W2 / IG

Range: DA (NA) See: CH 09 NO PERMISSION

9947

Configuration FH 345 GB/ 4 - 34

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
K 24. External Check System

Accept Span Correction K 24.01 SE Range: SE ZBS KME_KUB K 24.04 SE Range: SE

Correction Value: Span KME_KGB ZBS

Correction Value: Zero Point Accept Zero Point Correction K 24.05 K 24.02 SE Range: SE ZBS Range: SE KME_KUT SE ZBS KMT_KGT

Start External Check System K 24.03 SE Range: SE ZBS KME_STE

K 25. Moisture Correction

no
K 25.01 DE(NE)

Moisture Correction ON FEU_EIN -.NO YES / NO

n
K 25.02 AE

Material Moisture MATFEU -

Range: DE (NE) / SE

Range: AE See: P 21.02 FEU_ANH P 21.03 FEU_BER

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 35

WEIGHFEEDER

MULTICONT

m
K 25.03 AA

Actual Feed Rate, Dry PTROCK -

mo
K 25.06 DA(NA)

Moisture MAX FEU_MAX W1 W1 / W2 / IG

Range: AA See: P 21.04 PTR_ANH P 21.05 PTR_BER Counter 1, Pulse Output Corrected K 25.04 .0 Range: .0 / .1 FM1_TR

Range: DA (NA) See: HI 05 Moisture HI

mo
K 25.07 DA(NA)

Moisture Correction Active FEU_AKT -

Range: DA (NA)

Counter 2, Pulse Output Corrected K 25.05 .0 Range: .0 / .1 FM2_TR Material Moisture, Wrong Analog Input K 25.08 IN FEU_FAN A

Range: IN A / IG See: WE 08 ANALOG INPUT MOISTURE

9947

Configuration FH 345 GB/ 4 - 36

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
K 26.2. Second Weighbridge

The second weighbridge has to be switched on in function distributor, too. Second Weighbridge K 26.01 WZ WB2_WZ A A

m
DA

Belt Load 2nd Weighbridge Too High WB2_QMAX W1 A / W1 / W2 / IG

K 26.05

Range: WZ / AE See: Calibration AE, P 22.08

Range: DA (NA) See: Message HI 06 Parameter P 22.11

Load Cell Input 2nd Weighbridge Underamplified K 26.02 IN Range: WB2_MIN A A / W1

Check Weighbridge K 26.06 NO Range: NO, YES See: K 26.07, Message HI 07, P 22.12, P 22.13, Service Value dQ Acquires meaned difference between belt loads of first and second weighbridge. If value exceeds set threshold, message HI 07 is output. Monitoring is active only in Gravimetric Mode. WB2_CTL

Load Cell Input 2nd Weighbridge Overamplified K 26.03 IN Range: WB2_MAX

Check Weighbridge Event Message A A / W1 K 26.07 IN Belt Load 2nd Weighbridge Too Small WB2_QMIN W1 A / W1 / W2 / IG W1 WB2_DIF

mp
K 26.04 DA

Range: W1 / W2 / A / IG See: K 26.06, Message HI 07

Range: DA (NA) See: Message LO 06 Parameter P 22.10


9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 37

WEIGHFEEDER
2nd Weighbridge K 26.08 AA WB2_AA K 26.09 .0

MULTICONT
2nd Weighbridge Only WB2_AUS

Range: AA See: P 22.14 WB2_ANH P 22.15 WB2_BER

Range: .0/.1 .0 : First and second weighbridge controllers are active. .1 : Only second weighbridge controller is active.

K 27. Dead Time Element

One dead time element is provided for the setpoint and one for the ON/OFF command. The dead time element for the setpoint can be initialized externally, pre-assigned with the available setpoint and held. Both dead time elements can be selected separately in Mode 0 and Mode 1. The inputs are not monitored for double assignment so that with one input both dead time elements can be selected at the same time. The dead time for both elements is entered with Parameter P 22.01.

no
K 27.02 DE(NE)

Dead Time Element Setpoint Preselect M1 TOT_VM1 -.NO YES / NO

Range: DE (NE) / SE

no
K 27.01 DE(NE)

Dead Time Element Setpoint Preselect M0 TOT_VM0 -.NO YES / NO

no
K 27.03 DE(NE)

Dead Time Element Setpoint Clock ON M0 TOT_TM0 -.NO YES / NO

Range: DE (NE) / SE

Range: DE (NE) / SE

9947

Configuration FH 345 GB/ 4 - 38

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

no
K 27.04 DE(NE)

Dead Time Element Setpoint Clock ON M1 TOT_TM1 -.NO YES / NO

no
K 27.09 DE(NE)

Dead Time Element START/STOP, Clock ON M0 TOT_TI0 -.NO YES / NO

Range: DE (NE) / SE

Range: DE (NE) / SE

no
K 27.05 DE(NE)

Preassign Setpoint Dead Time Element TOT_VOR -.NO YES / NO

no
K 27.10 DE(NE)

Dead Time Element START/STOP, Clock ON M1 TOT_TI1 -.NO YES / NO

Range: DE (NE) / SE

no
K 27.06 DE(NE)

mo
K 27.11 DA(NA) TOT_INI

Range: DE (NE) / SE

Dead Time Elements Preselected TOT_GEW -

Initialize Setpoint Dead Time Element

Range: DA (NA) -.NO YES / NO

Range: DE (NE) / SE

no
K 27.07 DE(NE)

mo
K 27.12 DA(NA) TOT_IM0

Dead Time Element Setpoint Active TOT_PON -

Dead Time Element START/STOP Preselet M0

Range: DA (NA) -.NO YES / NO

Range: DE (NE) / SE

no
K 27.08 DE(NE)

Dead Time Element START/STOP Preselect M1 TOT_IM1

mo
K 27.13 DA(NA)

Dead Time Element START/STOP Active TOT_ON -

-.NO YES / NO

Range: DE (NE) / SE
9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Configuration FH 345 GB/ 4 - 39

WEIGHFEEDER

MULTICONT

K 28. Interpulse Period Controlled Prefeeder

For other prefeeder types, see K 17...19.

Prefeeder has to be turned on with K 16.06.

mo
K 28.00 DA (NA)

START Prefeeder Drive ZP_AON EA1.DA6

noq
K 28.03 DE (NE)

START Local Prefeeder ZP_LON ME1.DE5 YES / NO

Range: DA (NA) See: P 27.01...

Range: DE (NE)

n
K 28.04 Prefeeder Type Active K 28.01 .0 Range: .0 / .1 ZP_AKT AE

Local Prefeeder Setpoint ZP_LST ME2.AE1

Range: AE See: P 27.02 zp_lanh P 27.03 zp_lber

no
K 28.02 DE (NE)

Start Prefeeder ZP_NON -.YES

Range: DE (NE) / SE YES / NO See: K 16.06 Z_RIEG Note: Holds for M0, M1 and Test Mode.

9947

Configuration FH 345 GB/ 4 - 40

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
5 Parametrization
5.1 Parameter Designation
Parameters complete configuration and serve for total system definition. All rated data, calibration and evaluation data as well as controller settings, limit values, and standard interfaces are entered as parameters. All functions required for parameter setting can be accessed via the function distributor. The operating functions are described in Chapter 2 Functions and Operating Modes. For rated data, calibration and evaluation data, see relevant data sheets. Section 5.2 Survey comprises a list of all parameter blocks which enables instantaneous survey and facilitates location of individual parameters. In Section 5.3 Index, all parameter blocks including default values and eligible settings are listed. The parameter units can be converted from the international (SI) to the American system of units by configuration (K 00.11 Nonsi). Conversion Intl. Units => American Units g kg t => => => oz lb t inch f 1 oz 1 lb 1t 1 inch 1f = 28.3703 g = 0.4535925 kg = 2000 lb = 25.4 mm = 0.3048 m

WEIGHFEEDER

The representation in Section 5.3 follows the display of a machine control station. All parameters are represented in a bordered three-line block. The first line contains the parameter name. The indications in lines 2 and 3 correspond to the display in the setpoint field and the input/output field of a machine control station.

Name Number Value Range (Setting range, value can be continuously changed.) 0 / 1 / 2 (Setting range, select from values displayed, select by cursor operation) Reference: To configuration, see Chapter 4. To event messages, see Chapter 6. The note completes representation by additional information. A...Z Symbol

Note:

The parameter name refers to the function and serves as keyword in the alphabetic search list. The parameter number is composed of block number, position of parameter within the block and the letter P. This number serves for identification and is shown in the setpoint field together with the symbol. The symbol as such serves for identification of parameter on display. The value shows the current parameter setting. The index comprises the default values of the standard version. They can be loaded any time with sub-function Load Init. Par.. Deviating values are marked on display by a * at the beginning of the line.

mm => m =>

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 1

WEIGHFEEDER
5.2 Survey
Block P 00. Password Active Language Dialog Behaviour P 00.01 passakt P 00.02 spr

MULTICONT

Device Constant GK Span Correction KOR Belt Circuit Code Number IB Diagnosis Tool

P 03.05 GK P 03.06 KOR P 03.07 Imp/Uml P 03.08 dt_kenn

Block P 01.

Rated Data of Feeder

Block P 04.

Feed Rate Controller

Feed Rate, Dimension and Display Format Nominal Feed Rate PNENN Effective Weighbridge Length LEFF Nominal Speed VNENN Display and Output Format Totalizing Counter 1 Pulse Length, Totalizing Counter 1 Display and Output Format Totalizing Counter 2 Pulse Length, Totalizing Counter 2

P 01.01 pdim P 01.02 pnenn P 01.03 blaenge P 01.04 vnenn P 01.05 fm1dim P 01.06 fm1puls P 01.07 fm2dim P 01.08 fm2puls

P Portion KP P 04.01 P_Ant Reset Time TN P 04.02 tnast Deviation Threshold P 04.03 difmax Reaction Time Deviation P 04.04 regtim Filter Deviation P 04.05 rtau Filter Controller Input Actual Feed Rate P 04.06 reintau Acceleration Limit P 04.07 bgrenz Control Magnitude Elevation ANH P 04.08 anh Lower Control Magnitude Limit Gu P 04.09 r_gu Feeder Blocked Response TimeP 04.10 tblksch Standby P 04.11 standby Emergency Setpoint Limit P 04.12 notsol Deviation Threshold P 04.13 Xd_Fak Block P 05. Display and Output Filter P 05.01 plbstau P 05.02 pdatau P 05.03 psetau P 05.04 qtau P 05.05 vtau

Block P 02.

Calibration and Evaluation Data

Number of Belt Circuits NBAND P 02.01 nband Belt Circuit Time TBAND P 02.02 tband Load Cell Characteristic Value E P 02.03 wz_kenn Load Cell Rated Capacities P 02.04 wz_nenn WZNENN Lever Ratio P 02.05 Hebel Angle a P 02.06 winkel VS Characteristic Value P 02.07 vs_kenn Load Cell Filter P 02.08 wz_filt Check Weight P 02.09 qprf Maintenance Interval P 02.10 Sensor Length P 02.11 s_lng Sensor Width P 12.12 s_br Sensor Offset P 12.13 s_offs SPC Time P 12.14 spczeit SPC Filter P 12.15 spcfilt XD Filter P 12.16 xdfilt

Actual Feed Rate Display Actual Feed Rate Analog Output Actual Feed Rate SE Bus Belt Load q Belt Speed v

Block P 06. Actual Feed Rate, Limit Value PMIN Actual Feed Rate, Limit Value PMAX Belt Load Limit Value qMIN Belt Load Limit Value qMAX Belt Speed Limit Value vMIN Belt Speed Limit Value vMAX Belt Slip, Belt Length Change SMAX Pulley Slip Block P 07.

Limit Values wartint P 06.01 pmin P 06.02 pmax P 06.03 qmin P 06.04 qmax P 06.05 vmin P 06.06 vmax P 06.07 Schlupf P 06.08 TRSCHLU

Block P 03. Nominal Belt Load qNENN Effective Tare N + T Tare N Tare Correction T

Calibration Results P 03.01 qnenn P03.02 qtara P 03.03 Tara_N P 03.04 Tarkor

Standardized Analog Outputs P 07.01 P1_Anh.


9947

Actual Feed Rate 1, Elevation

Parametrization FH 345 GB/ 5 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Actual Feed Rate 1, Range Actual Feed Rate 2, Elevation Actual Feed Rate 2, Range Belt Load, Elevation Belt Load, Range Speed Elevation Speed Span Deviation, Zero Point Deviation, Range Setpoint, Zero Point Setpoint, Span Block P 08. P 07.02 P1_Ber. P 07.03 P2_Anh. P 07.04 P2_Ber. P 07.05 Q_Anh. P 07.06 Q_Ber P 07.07 v_Anh P 07.08 v_Ber P 07.09 r_Anh. P 07.10 r_Ber. P 07.11 Ps_Anh. P 07.12 PS_Ber.

WEIGHFEEDER
Storage I Portion Elevation Control Magnitude ANH Bypass BYP Lower Control Magnitude Limits Gu Upper Control Magnitude Limits Go Adaptation Controller Input Adaptation Controller Output Block P 13. P 12.05 I_Save P 12.06 qm_anh P 12.07 qm_byp P 12.08 qm_min P 12.09 qm_max P 12.10 qm_xe P 12.11 qm_xm

Prefeeder, Speed-Controlled

Standardization Analog Inputs P 08.01 adan1 P 08.02 adbe1 P 08.03 LDr_Anh P 08.04 LDr_Ber Drive Signal Prefeeder Local, Elevation Drive Signal Prefeeder Local, Range P 13.01 zd_lanh P 13.02 zd_Idber

Setpoint, Zero Point Setpoint, Span Setpoint, Zero Point, Local Setpoint, Span, Local

Block P 14.

Prefeeder, Position-Controlled

Block P 09.

Belt Influence Compensation, Control Delay for Point of Discharge

Belt Influence Compensation Filter Belt Length Weighbridge Discharge

P 09.01 BAT_Fil P 09.02 DAP_LNG

Block P 10. Afterflow Totalization Start-Up/Clearance Time Start-up Mode Clearance Mode Emergency Setpoint Limit

Preset Times P 10.01 tnach P 10.02 fahrtim P 10.03 anfahr P 10.04 leerfa P 10.05 Not_Sol

Position Signal Prefeeder, Elevation Position Signal Prefeeder, Range P Portion KP I Portion Reset Time TN D Portion Rate Time TV Response Threshold Deviation Prefeeder Position Signal Output, Zero Point Prefeeder Position Signal Output, Span

P 14.01 zs_anh P 14.02 zs_ber P 14.03 zs_kp P 14.04 zs_tn P 14.05 zs_tv P 14.06 zs_e P 14.07 zs_aanh P 14.08 zs_aber

Block P 15. Feed Rate Pre-Act Contact Residual Amount Pre-Act Contact Residual Amount Main Contact Adaptation Factor

Batching Mode P 15.01 dpvor P 15.02 vbatch P 15.03 hbatch P 15.04 hkadap

Block P 11.

Prefeeder, Rated Data

Dead Time Prefeeder Weighbridge P 11.01 ZUT_Tot Prefeeder Constant P 11.02 ZUT_Kon Nominal Transit Time PositionerP 11.03 ZUT_LZ Dynamic Factor P 11.04 ZUT_Dyn

Block P 16.

Fill Weight Measurement, Rated Data and Calibration

Block P 12. P Portion KP I Portion: Reset Time TN D- Portion Rate Time TV Rate Filter TF

Belt Load Controller P 12.01 qm_kp P 12.02 qm_tn P 12.03 qm_tv P 12.04 qm_tf

9947

Fill Weight, Display Format and Dimension Nominal Fill Weight Load Cell Rated Capacity BWZNENN Load Cell Characteristic E Load Cell Filter Fill Weight, Analog Output,

P 16.01 vbdim P 16.02 vbnenn P 16.03 vwz_nen P 16.04 vwz_ken P 16.05 vb_filt

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 3

WEIGHFEEDER
Elevation Fill Weight, Analog Output, Range Block P 17. P 16.06 vbanh P 16.07 vbdakor

MULTICONT
Circuits Check Quantity Minimum Check Quantity Check Storage Quantity Minimum Check Fill Weight Maximum Check Fill Weight Relative Zero Point KP Absolute Zero Point NA Relative Range BR Absolute Range BA Correction Evaluation Correction Speed Cycle Time P 20.05 KMEBAND P 20.06 KME_SOL P 20.07 KME_MIN P 20.08 KME_PMV P 20.09 KMEBMIN P 20.10 KMEBMAX P 20.11 KMETREL P 20.12 KMETABS P 20.13 KMEBREL P 20.14 KMEBABS P 20.15 KME_BEW P 20.16 KME_KV P 20.17 KME_ZYK

Fill Weight Measurement, Calibration Results P 17.01 vb_tara P 17.02 vb_kor

Tare N Span Correction KOR

Block P 18.

Fill Weight Control

Fill Weight Filter P 18.01 vbtau Fill Control, Limit Value STMIN P 18.02 vbmin Fill Control, Limit Value STMAX P 18.03 vhmax Fill Weight, Limit Value BMIN P 18.04 vbmin Fill Weight, Limit Value BMAX P 18.05 vbmax

Block P 21.

Moisture Correction P 21.01 FEU_MAX P 21.02 FEU_ANH P 21.03 FEU_BER P 21.04 PTR_ANH P 21.05 PTR_BER

Block P 19.

Fill Weight Controller P 19.01 B_soll P 19.02 B_fil P 19.03 B_rtau P 19.04 B_schw P 19.05 B_absch P 19.06 B_kp P 19.07 B_tn P 19.08 B_tv P 19.09 B_tf P 19.10 B_stkp P 19.11 B_sttv P 19.12 vbanh P 19.13 B_go P 19.14 B_gu P 19.15 B_Isave P 15.16 B_kost P 19.17 B_kon

Fill Weight Setpoint Setpoint Filter Actual Value Filter Response Threshold Deviation Attenuation P Portion KP I Portion: Setback Time TN D Portion: Rate Time TV Rate Filter TF Interference Quantity Effect STKP Interference Quantity Effect STTV Control Magnitude Elevation ANH Upper Control Magnitude Limit FGo Lower Control Magnitude Limit FGu Preselect Memory Emergency Control Check Status Normal Control Check Status

Moisture Limit Value FEUMAX Material Moisture, Elevation Material Moisture, Range Actual Feed Rate Corrected, Elevation Actual Feed Rate Corrected, Range

Block P 22.

Second Weighbridge, Rated Data, and Calibration

Block P 20. Correction Modes Manual/Automatic Damping Time Blind Distance Factor Circuits/Quantity Check Setpoint

Check System

P 20.01 KME_KOR P 20.02 KME_TB P 20.03 KME_BL P20.04KME_VOR

Effective Weighbridge Length L2EFF P 22.01 WB2_LNG Load Cell Characteristic Value E P 22.02 WM2_KEN Load Cell Rated Capacity WZNenn P 22.03 WB2_F0 Lever Ratio P 22.04 WB2_HEB Angle a P 22.05 WB2_WIN Load Cell Filter P 22.06 WB2_FIL Dead Time Prefeeder Second Weighbridge P 22.07 ZUT_T2B Analog Output Second Weighbridge P 22.08 WB2_ANH Analog Input Second Weighbridge P 22.09 WB"_BER MIN Value Second WeighbridgeP 22.10 wb2_qmin MAX Value Second Weighbridge P 22.11 wb2_qmax Check Weigher Response Threshold P 22.12 dif_max Check Weigher Filter P 22.13 dif_tau 2nd Weighbridge Analog OutputP 22.14 WB2_Anh 2nd Weighbridge Analog OutputP 22.15 WB2_BER

9947

Parametrization FH 345 GB/ 5 - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Block P 23. Second Weighbridge, Calibration Results P 23.01 WB2_TAR P 23.02 WB2_GK P 23.03 WB2_KOR Block P 25. Dead Time Tare N Device Constant GK Correction Factor KOR Block P 26.

WEIGHFEEDER
Dead Time P 25.01 Totzeit Intermittierender Zuteiler P 26.01 zi_ed

Prefeeder Connection Period Block P 24. Second Weighbridge, Controller Block P 27. P Portion KP I Portion Reset Time TN D Portion Rate Time TV Rate Filter TF Adaptation Controller Output P 24.01 qa_kp P 24.02 qa_tn P 24.03 qa_tv P 24.04 qa_tf P 24.05 qa_xm

Pulse-Controlled Prefeeder

Prefeeder Pulse Period K 27.01 zp_il Local Prefeeder Speed Control Signal, Elevation P 27.02 zp_lanh Local Prefeeder Speed Control Signal, Range P 27.03 zp_idber

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 5

WEIGHFEEDER
5.3 List of Parameters

MULTICONT

P 00. Dialog Behaviour

Password Active P 00.01 N Range: J = YES / N = N(O) Note: Password = 7274. passakt P 00.02 DEUTSCH

Language spr

Range: DEUTSCH / ENGLISH / FRANCAIS / ITALIANO / ESPAGNOL

P 01. Rated Data of Feeder

Feed Rate, Dimension and Display Format P 01.01 kg/h Range: .- kg/h ... t/h Note: Inclusive decimal point, 6 places with a maximum of 3 decimal places are displayed. Format and dimension are selected by operating cursors UP and DOWN. pdim

Effective Weighbridge Length LEFF P 01.03 0.300 m Range: 0.100 m ... 50.000 m blaenge

Nominal Feed Rate PNENN P 01.02 10000 kg/h Range: 10 kg/h ... 229999000 kg/h pnenn P 01.04 0.1000 m/s

Nominal Speed VNENN vnenn

Range: 0,099 m/s ... 50 m/s

9947

Parametrization FH 345 GB/ 5 - 6

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Display and Output Format Totalizing Counter 1 P 01.05 -t Range: -.- kg ... -t Note: Inclusive decimal point, display shows 7 places with maximum 3 decimal places. Display size and dimension are selected by operating cursors UP and DOWN. One pulse corresponds to the smallest place displayed. fm1dim

WEIGHFEEDER
Display and Output Format Totalizing Counter 2 P 01.07 -t Range: -.- kg ... -t Note: See totalizing counter 1. fm2dim

Pulse Length, Totalizing Counter 1 Pulse Length, Totalizing Counter 2 P 01.06 50 ms 50 ms Range: 50 ms ... 500 ms See: K 10.08 FM1_IA Range: 50 ms ... 500 ms See: K 10.09 FM2_IA fm1puls P 01.08 fm2puls

P 02. Calibration and Evaluation Data

Number of Belt Circuits NBAND P 02.01 1.00 Range: 1 ... 100 Note: Applies to set-up programs. nband P 02.02 30.0 s

Belt Circuit Time TBAND tband

Range: 1 s ... 9999 s See: K 14.01 SYNCHRO Note: TBand is the time required by the conveyor belt for one total circuit. Enter this time only if automatic belt length measurement (K 11.01) is cut off. With belt length measurement connected the measured value is automatically read in by program Belt Length LB.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 7

WEIGHFEEDER
Load Cell Characteristic Value E P02.03 2.0000 mV/V Range: 0.2 mV/V; 9.9999 mV/V wz_kenn P 02.07 10000 l/m

MULTICONT
VS Characteristic Value vs_kenn

Range: 10 l/m ... 1000000 l/m Note: The characteristic value indicates the number of speed transducer pulses per belt length meter.

Load Cell Rated Capacities WZNENN P 02.04 50.000 kg Range: 0.5 kg ... 22000 kg Note: Enter total of all load cell rated capacities. wz_nenn P 02.08 1

Load Cell Filter wz_filt

Range: 0 / 1 / 2 / 3 Note: 0 is the smallest filter time.

Lever Ratio P 02.05 0.8636 Range: 0.01 ... 2.0 Note: This parameter considers the constructional transmission ratio between force transducing weighed idler and load cell. WZ = . Q WZ = Load on load cell Q = Weighbridge load Hebel P 02.09 7.000 kg

Check Weight qprf

Range: 1.0 kg ... 22000 kg Note: The check weight can be used for additional calibration check. It should not exceed the range of 2/3 to 4/3 of nominal weighbridge load. Enter effective check weight.

Angle a P 02.06 0.00 gr Range: 0 degr. ... 20 degr. Note: This parameter indicates the position of the longitudinal axis of feeder. winkel P 02.10 3000 h

Maintenance Interval wartint

Range: 1 h ... 10000 h See: SC 02 Maintenance Interval

9947

Parametrization FH 345 GB/ 5 - 8

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Sensor Length P 02.11 8.00 cm Range: 0.01 ... 400.00 cm See: K 14.01 Length of sensor triangle in conveyor belt in direction of travel. Serves for detection of belt drift, belt slip, etc. s_lng P 02.14 8.0 h

WEIGHFEEDER
SPC Time spczeit

Range: 0.1...24.0 h Monitoring time for diagnostic and quality tool. Maximum belt load acquisition cycle.

SPC Filter Sensor Width P 02.15 P 02.12 6.00 cm Range: 0.00...24.00 h Range: 0.01...400.00 cm See: K 14.01, P 02.11 Filter time for diagnostic and quality tool for acquisition of variance and belt load mean value. XD Filter Sensor Offset P 02.16 P 02.13 0.00 cm Range: 0.00...24.00 h Range: -400.00...+400.00 cm See: K 14.01, P 02.11 Corrects zero point of belt drift detection. Limits maximum display range. Bigger differences should be compensated by sensor displacement. New display = old display - offset Filter time for diagnostic and quality tool for deviation variance acquisition. s_offs 0.10 h xdfilt s_br 1.00 h spcfilt

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 9

WEIGHFEEDER
P 03. Calibration Results

MULTICONT

Nominal Belt Load qNENN P 03.01 27.78 kg/m Range: 0.1 kg/m ... 10 000 kg/m Note: No entry possible. qnenn P 03.05 0.12436

Device Constant GK GK

Range: 0.0001 mV/(kg/m) ... 1000 mV/(kg/m) Note: The device constant is computed internally.

Effective Tare N + T P03.02 0.00 kg/m Range: 0 kg/m ... 10000 kg/m Note: No entry possible. qtara P 03.06 1.000

Span Correction KOR KOR

Tare N P 03.03 0.00 kg/m Tara_N

Range: 0.5 ... 2.0 Note: This parameter serves for correction of calibration by means of check weight with the value being acquired by setting program Weight Check CW.

Belt Circuit Code Number IB P 03.07 Imp/Uml

Range: 0 kg/m ... 10000 kg/m Note: Result from taring program TW.

10000 I/U Range: 0 I/U ... 900000 I/U Note: The code number states the number of pulses per belt circuit computed by setting program LB Pulse/Rev.. Tarkor Diagnosis Tool P 03.08 0 Code number of diagnostic tool. Parameter cannot be changed. dt_kenn

Tare Correction T P 03.04 0.00 kg/m Range: -1000 kg/m ... 1000 kg/m Note: Result from zero setting program or check system.

9947

Parametrization FH 345 GB/ 5 - 10

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
P04. Feed Rate Controller

WEIGHFEEDER

The controller is designed as P-I controller. Special designs P controller and I controller can be provoded as well. P Portion KP P 04.01 3.00 mV/% Range: 0.0 mV/% ... 10000.0 mV/% Note: KP = 0 mV% I- controller P_Ant P 04.04 20.0 s Range: 0 s ... 300 s See: CO 01 DEVIATION Reaction Time Deviation regtim

Reset Time TN P 04.02 0.3 s Range: 0.0 s ... 60000 s Note: TN is the time in which with abrupt deviations the I portion causes the same control magnitude change as the P portion. TN < 0.15s P-controller tnast P 04.05 3.0 s

Filter Deviation rtau

Range: 0 s ... 600 s See: The deviation value meaned by the filter is used for displays and outputs.

Deviation Threshold P 04.03 5.0 % p Range: 0 %; 15 % PNenn See: CO 01 DEVIATION. This parameter determines the response threshold of event message CO 01 DEVIATION. difmax

Filter Controller Input Actual Feed Rate P 04.06 0.0 s Range: 0 s ... 600 s See: The filter filters the value of the actual feed rate used in the feed rate controller. reintau

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 11

WEIGHFEEDER
Acceleration Limit P 04.07 100000 mV/s Range: 0 mV/s ... 100000 mV/s See: This parameter determines the maximum admissible speed of control magnitude change for the weighfeeder drive. This is to prevent that weighfeeder reacts quicker than connected prefeeder, if any. If no prefeeder is connected, the default value remains set and the acceleration limit becomes ineffective. Control Magnitude Elevation ANH P 04.08 0.00 m/V Range: 0 .... 10000 mV Lower Control Magnitude Limit Gu P 04.09 0 mV r_gu anh P 04.11 2% Range: 0...100% bgrenz P 04.10 10 Range: 0..1000s See:

MULTICONT
Feeder Blocked Response Time tblksch

Message SC 01 Configurator K 01.01

If, with 0 control magnitude, drive stands still longer than set in Parameter P 04.10, Message SC 01 is output.

Standby standby

See: K 01.03, Meldung SC 07 Cuts of feed drive (not feeder) if setpoint falls below P 04.11 after 3 s. Switch-back is without delay.

Range: 0 mV / roffset_0 / roffset_1 See: K 11.04 REG_ERR CO 03 CONTROLLER OUT LIMIT Note: The lower limit can be set either to 0 mV or roffset (P 04.08) Control Magnitude at STOP 0 mV 0 mV roffset P 04.12 0.00 kg/h

Emergency Setpoint Limit notsol

r_gu 0 mV roffset_0 roffset_1

Limit 0 mV roffset roffset

Range: 0 kg/h ... 2200000 kg/h See: K 11.01 QU_NOTS Note: The active setpoint can be limited to the set emergency setpoint threshold. The limitation is activated via K 11.01.

9947

Parametrization FH 345 GB/ 5 - 12

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Deviation Threshold P 04.13 1 Range: 0.0...1.0 See: P 04.03, P 04.04 Xd_Fak =0

WEIGHFEEDER
The response threshold is proportional to setpoint. Response threshold i= P 04.03 * setpoint / nominal value = 0,5 With 0 setpoint, the response threshold is 0.5 * P 04.03; at 100% setpoint, equal to P 04.03.

Deviation monitoring response threshold can be set with Parameter P 04.13 as a function of setpoint. P 04.13 = 1 Independent of setpoint, the response threshold always corresponds to the value of Parameter P 04.03.

P 05. Display and Output Filter

Actual Feed Rate Display P 05.01 3.0 s Range: 0 s ... 600 s plbstau P 05.03 3.0 s

Actual Feed Rate SE Bus psetau

Range: 0 s ... 600 s Belt Load q

Actual Feed Rate Analog Output P 05.02 3.0 s Range: 0 s ... 600 s See: K 10.14 P_IST1 K 10.15 P_IST2 pdatau

P 05.04 3.0 s Range: 0 s ... 600 s Belt Speed v P 05.05 3.0 s Range: 0 s ... 600 s

qtau

vtau

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 13

WEIGHFEEDER
P 06. Limit Values

MULTICONT

Actual Feed Rate, Limit Value PMIN P 06.01 5.0 % p Range: -10% ... 200 % PNENN See: LO 01 FEED RATE I LO Actual Feed Rate, Limit Value PMAX P 06.02 110.0 % p Range: -10% ... 200 % PNENN See: HI 01 FEED RATE I HI pmax P 06.06 150 % v pmin P 06.05 3.0 % v

Belt Speed Limit Value vMIN vmin

Range: -10 % ... 200 % VNENN See: LO 03 SPEED V LO Belt Speed Limit Value vMAX vmax

Range: -10 % ... 200 % VNENN See: HI 03 SPEED V HI

Belt Load Limit Value qMIN P 06.03 60.0 % q Range: -10% ... 200 %qNENN See: LO 02 LOAD Q LO qmin

Belt Slip, Belt Length Change SMAX P 06.07 2.00 % IB Range: 0 % ... 10 % IB See: WM 03 BELT SLIP Schlupf

Belt Load Limit Value qMAX P 06.04 140.0 % q Range: -10 % ... 200 % qNENN See: HI 02 LOAD Q HI qmax P 06.08 3.0%

Pulley Slip TRSCHLU

Range: 0.5...100.0% See: Message WM03, K 03.03, K 03.06, K 03.07 Belt slip can be monitored by 1. 2. 3. External contact Sensor on each of the two tail pulleys Sensor on conveyor belt.

9947

Parametrization FH 345 GB/ 5 - 14

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Pulley slip monitoring (Case 2) acts parallel to other possibilities: Prerequisites: f <1,25 Hz Pulse width of period > 0.35s Sets smallest slip between the two pulleys that can still be acquired. Message is output after N pulses.

WEIGHFEEDER
SCHLU N = 5 100 S
SCHLU = S = Parameter P 06.08 Slip between two pulleys in %.

Monitoring starts upon turn-on of weighfeeder.

P 07. Standardized Analog Outputs

Actual Feed Rate 1, Elevation P 07.01 2.00 V Range: 0 kg/h = 0 V ... 10 V See: K 10.13 P_IST1 P1_Anh. P 07.04 10.00 V

Actual Feed Rate 2, Range P2_Ber.

Range: PNENN = 0 V ... 100 V

Actual Feed Rate 1, Range P 07.02 10.00 V Range: PNENN = 0 V ... 100 V P1_Ber. P 07.05 2.00 V

Belt Load, Elevation Q_Anh.

Range: 0 % = V ... 10 V See: K 10.15 Q_BAND Belt Load, Range

Actual Feed Rate 2, Elevation P 07.06 P 07.03 2.00 V Range: qNENN = 0 V ... 100 V Range: 0 kg/h = 0 V ... 10 V See: K 10.14 P_IST2 P2_Anh. 6.00 V Q_Ber

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 15

WEIGHFEEDER
Speed Elevation P 07.07 2.00 V Range: 0 m/s = 0 V ... 10 V See: K 16.10 V_BAND v_Anh P 07.10 20.00 V

MULTICONT
Deviation, Range r_Ber.

Range: 100 % = 0 V ... 100 V Setpoint, Zero Point

Speed Span P 07.08 6.00 V Range: VNENN = 0 V ... 100 V v_Ber

P 07.11 2.00 V Range: 0 kg/h = 0 V ... 10 V

Ps_Anh.

Setpoint, Span Deviation, Zero Point P 07.09 5.00 V Range: 0 % = 0 V ... 10 V r_Anh. P 07.12 10.00 V Range: PNENN = 0 V ... 100 V PS_Ber.

9947

Parametrization FH 345 GB/ 5 - 16

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
P 08. Standardization Analog Inputs

Setpoint, Zero Point P 08.01 2.00 V Range: 0 kg/h = 0 V ... 10 V See: K 02.05 P_ANAL Setpoint, Span P 08.02 10.00 V Range: PNENN = 0 V ... 100 V adbe1 P 08.04 5.00 V adan1 P 08.03 0.00 V

Setpoint, Zero Point, Local LDr_Anh

Range: 0 kg = 0 V ... 10 V See: K 06.10 LO_STEL Setpoint, Span, Local LDr_Ber

Range: 0 V ... 100 V qNENN

P 09. Belt Influence Compensation, Control Delay for Point of Discharge

Belt Influence Compensation Filter P 09.01 1 Range: 0 / 1 / 2 / 3 See: K 14.02 BIC_EIN Note: Stage 0 means highest compensation speed. BAT_Fil

Belt Length Weighbridge Discharge P 09.02 0.00 % LB Range: 0 % ... 50 % belt length See: K 14.03 DAP_EIN Note: Enter parameter only if control delay at discharge point is activated. DAP_LNG

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 17

WEIGHFEEDER
P 10. Preset Times

MULTICONT

Afterflow Totalization P 10.01 3.00 s Range: 0.1 s ... 100 s Note: After feeder cut off, totalization continues for time tnach. tnach P 10.04 0

Clearance Mode leerfa

Start-Up/Clearance Time P 10.02 0.0 s Range: 0 s ... 1000 s See: P 10.03 anfahr P 10.04 leerfa Start-up Mode P 10.03 Qmin Range: Qmin / fahrtim See: P 05.03 qmin P 10.02 fahrtim Note: After start up, weighfeeder operates in volumetric mode. fahrtim For the time set with P 10.02. Qmin Until first excess of QMIN, however, not longer than for fahrtim. anfahr fahrtim

Range: 0 / fahrtim See: P 10.02 fahrtim Note: After STOP, weighfeeder operations in volumetric mode. fahrtim For the time set with P 10.02. 0 Feeder stops immediately after cut-off.

9947

Parametrization FH 345 GB/ 5 - 18

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER
P 11. Prefeeder, Rated Data

The rated data apply to all types of prefeeders. Dead Time Prefeeder Weighbridge P 11.01 10.00 s Range: 0 s ... 1000 s Note: This parameter indicates the dead time from transfer point of prefeeder to center of weighbridge for nominal belt speed VNenn. ZUT_Tot

Dynamic Factor P 11.04 1.00 Range: 0.5 ... 2 Note: The dynamic factor is active only in conjunction with setting program Contr. Adapt. for prefeeder control. With setting 2 controller speed is highest. ZUT_Dyn

Prefeeder Constant P 11.02 0.02 %/mV Range: 0 % / mV ... 1000 % / mV ZUT_Kon

Nominal Transit Time Positioner P 11.03 30.00 s Range: 1 s ... 1000 s Note: This parameter only applies to the position-controlled prefeeder and indicates the transit time of prefeeder from position 0 to the position generating nominal belt load on weighfeeder at nominal conditions (mean bulk density, nominal speed of weighfeeder). Dynamic Factor ZUT_LZ

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 19

WEIGHFEEDER
P 12. Belt Load Controller

MULTICONT

Designed as P-I-D type, the belt load controller controls the belt load via the feed rate of the prefeeder for nominal belt load. It is active only in conjunction with the position-controlled prefeeder (Block P14, K19) or the speed-controlled prefeeder (Blocks P13, K 18). The controller can be set by program Contr. Adapt. The controller is controlled adaptively via the belt speed. P Portion KP P 12.01 0.0000 mV/% qm_kp

Rate Filter TF P 12.04 0.00 s Range: 0 s ... 60000 s Note: Filter setting should not exceed the rate time. qm_tf

Storage I Component P 12.05 I_Save

Range: 0 mV/% ... 10000 mV/% Note P = 0 s I controller

N Range: NO / YES / YES-A

I Portion: Reset Time TN P 12.02 5.00 s qm_tn

NO

Controller I component is reset upon feeder cut-off. I component remains stored. I component remains stored only if no alarm has occurred.

YES

YES-A : Range: 0 s ... 60000 s Note: TN is the time in which with abrupt deviations the I portion causes the same control magnitude change as the P portion. TN < 0,5 s P-D controller

Elevation Control Magnitude ANH P 12.06 0.00 mV qm_anh

D- Portion Rate Time TV P 12.03 0.00 s Range: 0 s ... 60000 s qm_tv Range: 0 mV ... 10000 mV Note: The output signal of the control magnitude is raised by the entered value.

9947

Parametrization FH 345 GB/ 5 - 20

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Bypass BYP P 12.07 5000.00 mV Range: 0 mV ... 10000 mV Note: Set bypass such that prefeeder generates nominal belt load at nominal belt speed. qm_byp P 12.10 LIN

WEIGHFEEDER
Adaptation Controller Input qm_xe

Range: AUS / LIN / ADAP Note: This parameter determines the adaptation type. Setting is automatic by program Contr. Adapt..

Lower Control Magnitude Limits Gu Adaptation Controller Output P 12.08 0 mV LIN Range: 0 mV/ qm_anh Note: The control magnitude limits are only active with speed-controlled prefeeder. The lower limit can be set to 0 mV or to value qm_anh (P 12.06). Control Magnitude at STOP 0 mV 0 mV qm_min Range: AUS / LIN Note: See Adaptation Controller Input. qm_min P 12.11 qm_xm

qm_min 0 mV qm_min_0 qm_min_1

Limit 0 mV qm_min qm_min

Upper Control Magnitude Limits Go P 12.09 10000.00 mV Range: 0 mV ... 10000 mV Note: For determination of the upper control magnitude limit the maximum admissible control magnitude for setpoint 0 % of feed rate is entered. The maximum admissible control magnitude for setpoint 100 % is set to 10V. Example: Enter 5V; control magnitude for setpoint 50 % is 7.5 V; control magnitude for 100 % is 10 V. qm_max

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 21

WEIGHFEEDER
P 13. Prefeeder, Speed-Controlled

MULTICONT

Drive Signal Prefeeder Local, Elevation P 13.01 0.00 V Range: 0 m/s = 0 V ... 10 V See: K 18.10 ZD_LST zd_lanh

Drive Signal Prefeeder Local, Range P 13.02 5.00 V Range: VNENN = 0 V ... 100 V zd_Idber

P 14. Prefeeder, Position-Controlled

Position Signal Prefeeder, Elevation P 14.01 0.00 V Range: 0 % open = - 10V ... 10 V See: K 19.15 ZS_STM Note: Set parameter such that 0 % open is displayed with prefeeder closed. zs_anh P 14.03 0.0040 s/%

P Portion KP zs_kp

Range: 0 s/% ... 1000 s/%

Position Signal Prefeeder, Range P 14.02 5.00 V Range: 100 % open = 0 V ... 100 V Note: Set range such that 100 % open is displayed at nominal opening angle of prefeeder (= nominal belt load). zs_ber P 14.04 3.0 s

I Portion Reset Time TN zs_tn

Range: 0 s; ... 60000 s

9947

Parametrization FH 345 GB/ 5 - 22

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
D Portion Rate Time TV P 14.05 0.00 s Range: 0 s; 60000 s zs_tv

WEIGHFEEDER
Prefeeder Position Signal Output, Zero Point P 14.07 2.00 V Range: 0 % = -10 V ... + 10 V See: K 19.16 ZS_ADO zs_aanh

Response Threshold Deviation P 14.06 0.0000 mV Range: 0 mV ... 10000 mV Note: 5000 mV correspond to 100 % deviation. zs_e

Prefeeder Position Signal Output, Span P 14.08 10.00 V Range: 100 % = 0 V ... 100 V zs_aber

P 15. Batching Mode

See Note

P : 01.09 fm1dim B : atching counter display format = Format of totalizing counter 1. Feed Rate Pre-Act Contact Residual Amount Pre-Act Contact pvor P 15.02 10.0000 kg Range: 0 kg ... 10000 kg vbat

P 15.01 50.0000 kg/h Range: 0 kg/h ... 10 000 kg/h

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 23

WEIGHFEEDER
Residual Amount Main Contact P 15.03 0.0000 kg Range : 0 kg ... 10000 kg hbat P 15.04 1.0 Range: 0.0 ... 1.0

MULTICONT
Adaptation Factor hkadap

Residual quantities P 15.02 and P 15.03 are self-optimizing. The two parameters are adapted by value Batch Error P 15.04". P 15.04 =0: No adaptation =1: Full adaptation

P 16. Fill Weight Measurement, Rated Data and Calibration

Fill Weight, Display Format and Dimension P 16.01 -kg Range: -.- kg ... t Note: Inclusive decimal point display shows 7 places with maximum 3 decimal places. Select format and dimension with cursors UP and DOWN. One pulse corresponds to the smallest place displayed. vbdim

Load Cell Rated Capacity BWZNENN P 16.03 10000.00 kg Range: 0.1 kg ... 2200000 kg Note: Enter total of all load cell rated capacities. vwz_nen

Nominal Fill Weight P 16.02 10000.00 kg Range: 1 kg ... 2200000 kg vbnenn P 16.04

Load Cell Characteristic E vwz_ken

2.8500 mV/V Range: 0.5 mV/V ... 9.9999 mV/V

9947

Parametrization FH 345 GB/ 5 - 24

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Load Cell Filter P 16.05 1 Range: 0 / 1 / 2 / 3 vb_filt

WEIGHFEEDER

Fill Weight, Analog Output, Elevation P 16.06 2.00 V Range: 0 kg = 0 ... 10 V See: K 21.10 FGVB vbanh

Fill Weight, Analog Output, Range P 16.07 10.00 V Range: FGNENN = 0 V ... 10 V vbdakor

P 17. Fill Weight Measurement, Calibration Results

Tare N P 17.01 0.00 kg Range: 0 kg ... 2200000 kg Note: The active tare is the total from tare N and preload compensation. The parameters are acquired by setting program TF: SUPPLY BIN.
9947

Span Correction KOR vb_tara P 17.02 1.000 Range: 0.5 ... 2 Note: This parameter acts on display in proportion. Setting is in manual. After taring, display can be corrected with the aid of a check weight. vb_kor

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 25

WEIGHFEEDER
P 18. Fill Weight Control

MULTICONT

See:

Block K21. Fill Weight Filter Fill Weight, Limit Value BMAX vbttau P 18.05 100.00 % B Range: 0 % ... 200 % FGNENN vbmax

P 18.01 3.0 s Range: 0 s ... 600 s Note: Holds for display only.

Fill Control, Limit Value STMIN P 18.02 20.00 % B Range: -10 % ... 200 % FGNENN Note: The MIN and MAX values of fill control serve as limit values for fill control of supply bin. vhmin

Fill Control, Limit Value STMAX P 18.03 80.00 % B Range: 0 % ... 200 % FGNENN vhmax

Fill Weight, Limit Value BMIN P 18.04 0.00 % B Range: -10 % ... 200 % FGNENN Note: The MIN and MAX values of the fill weight serve for bin level monitoring and output of event messages.
9947

vbmin

Parametrization FH 345 GB/ 5 - 26

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
P 19. Fill Weight Controller

WEIGHFEEDER

Fill Weight Setpoint P 19.01 50.00 % B Range: 0 % ... 200 % FGNENN B_soll P 19.05 0.000

Attenuation B_absch

Range: 0 ... 1 Note: The value is a factor for attenuation of deviation in the range of the response threshold.

Setpoint Filter P Portion KP P 19.02 0.5 s 0.00000 V/kg Range: 0.5 s ... 200 s Note: The filter filters setpoint changes caused by P 19.01 or the check system. Range: 0 V/kg ...10 V/kg Note: KP = 0 V/kg I Controller B_fil P 19.06 B_kp

Actual Value Filter P 19.03 0.5 s Range: 0 s ... 200 s Note: The value is used for display and output. B_rtau P 19.07 0.00 s

I Portion: Setback Time TN B_tn

Range: 0 s ... 60000 s Note: Special case I < 0.5 s P-D controller

Response Threshold Deviation P 19.04 0.00 kg Range: 0 kg ... 10000 kg B_schw P 19.08 0.00 s

D Portion: Rate Time TV B_tv

Range: 0 s ... 60000 s

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 27

WEIGHFEEDER
Rate Filter TF P 19.09 0.00 s Range: 0 s ... 60000 s B_tf

MULTICONT
Upper Control Magnitude Limit FGo P 19.13 10000.00 mV Range: 0 mV ... 10000 mV Note: This parameter determines the maximum value of the control magnitude. B_go

Interference Quantity Effect STKP Lower Control Magnitude Limit FGu P 19.10 5000.00 mV 0 mV Range: 0 mV ... 10 000 mV Note: Set the voltage that generates nominal feed rate of feeder at prefeeder. Range: 0 mV/ b_anh See: P 19.12 vbanh Note: This parameter determines the lower limit of the control magnitude. B_stkp P 19.14 B_gu

Interference Quantity Effect STTV P 19.11 0.00 s Range: 0 s ... 60000 s Note: Superimposes D portion to P 15.10. B_sttv P 19.15 N Range: J = YES / N = NO Note: N Controller is cleared after filling period of check system. J Controller is not cleared. vbanh Preselect Memory B_Isave

Control Magnitude Elevation ANH P 19.12 5000.00 mV Range: 0 mV; 10000 mV Note: This parameter serves for elevating the output signal of the control magnitude, e.g. for compensating the base speed of a vibrating feeder.

9947

Parametrization FH 345 GB/ 5 - 28

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Emergency Control Check Status P 19.16 0 Range: 0 / 1 / 2 See: P 18.02, P 18.03 Note: 0 No influence 1 If MAX is exceeded = > Prefeeder for supply bin OFF, control magnitude set to 0. If MIN is exceeded down = > Prefeeder for supply bin ON, controller is activated from its initial position. 2 If MAX is exceeded = > Prefeeder for supply bin OFF, control magnitude set to 0. If MIN is exceeded down = > Prefeeder for supply bin ON, controller is activated from its previous status. Unlike in sequence 2, controller is cleared in sequence 1. B_kost P 19.17 0

WEIGHFEEDER
Normal Control Check Status B_kon

Range: 0 / 1 See: P 18.04, P 18.05 Note: 0 Normal control has no influence. 1 If contact MIN is exceeded down = > active setpoint of feeder is limited to the emergency setpoint. If contact MAX is exceeded up = > active setpoint of feeder is not limited.

P 20. Check System

Correction Modes Manual/Automatic P 20.01 MAN Range: MAN / AUTO KME_KOR P 20.03 0.00 %

Blind Distance Factor KME_BL

Range: 0 % ... 50 %

Damping Time P 20.02 3s Range: 0 s ... 300 s KME_TB P 20.04

Circuits/Quantity Check Setpoint KME_VOR

DIM KMESOLL n Range: DIM KMESOLL n / DIM KMWSOLL kg

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 29

WEIGHFEEDER
Circuits P 20.05 1.0 Uml Range: 1 ... 100 KMEBAND P 20.10 1000.00 kg

MULTICONT
Maximum Check Fill Weight KMEBMAX

Range: 0 kg ... 2200000 kg

Check Quantity P 20.06 0.00 kg Range: 0 kg ... 2200000 kg KME_SOL P 20.11 10.0 %

Relative Zero Point KP KMETREL

Range: 0 % ... 100 % qNENN

Minimum Check Quantity P 20.07 100.00 kg Range: 0 kg ... 2200000 kg KME_MIN P 20.12 20.0 %

Absolute Zero Point NA KMETABS

Range: 0 % ... 1000 % qNENN

Check Storage Quantity P 20.08 500.00 kg Range: 0.1 kg ... 2200000 kg KME_PMV P 20.13 3.0 %

Relative Range BR KMEBREL

Range: 0 % ... 100 % qNENN Absolute Range BA

Minimum Check Fill Weight P 20.09 0.00 kg Range: 0 kg ... 2200000 kg KMEBMIN

P 20.14 10.0 % Range: 0 % ... 100 % qNENN

KMEBABS

9947

Parametrization FH 345 GB/ 5 - 30

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Correction Evaluation P 20.15 1.000 Range: 0.100 ... 1.000 (Factor) Correction Speed P 20.16 0.0 Range: 0 s ... 3600 s KME_KV KME_BEW P 20.17 30.0 min

WEIGHFEEDER
Cycle Time KME_ZYK

Range: 0 .. 1500 min

P 21. Moisture Correction

Moisture Limit Value FEUMAX P 21.01 15.00 % Range: 0 % ... 100 % Material Moisture, Elevation P 21.02 2.00 V Range: 0 % = 0 V; 10 V Material Moisture, Range P 21.03 10.00 V Range: FEUMAX = 0 V ... 100 V FEU_BER FEU_ANH FEU_MAX

Actual Feed Rate Corrected, Elevation P 21.04 2.00 V Range: 0 kg/h = 2 V ... 10 V Actual Feed Rate Corrected, Range P 21.05 10.00 V Range: PNENN = 0 V ... 100 V PTR_BER PTR_ANH

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 31

WEIGHFEEDER

MULTICONT

P 22. Second Weighbridge, Rated Data, and Calibration

Effective Weighbridge Length L2EFF P 22.01 0.5 m Range: 0.100 m ... 10.000 m Note: Enter effective weighbridge length of second weighbridge. WB2_LNG P 22.05 0.00 gr

Angle a WB2_WIN

Range: 0 degr ...20 degr.

Load Cell Filter Load Cell Characteristic Value E P 22.06 P 22.02 2 mV/V Range: 0.5000 mV/V ... 9.9999 mV/V Range: 0 / 1 / 2 / 3 Note: 0 is the lowest setting. WM2_KEN 1 WB2_FIL

Dead Time Prefeeder Second Weighbridge Load Cell Rated Capacity WZNenn P 22.07 P 22.03 50.000 kg Range: 1.000 kg ... 22000 kg Note: Enter total of all load cell rated capacities. Range: 0 s; 1000 s Note: Enter dead time from prefeeder transfer point to second weighbridge center at nominal belt speed. WB2_F0 10.00 s ZUT_T2B

Lever Ratio Analog Output Second Weighbridge P 22.04 1.0000 2.00 V Range: 0.0100 ... 2.0000 Range: 0...10.00 V Note: WZ = x Q WZ= Load on load cell Q= Weighbridge load See: K 26.01 Parameters P 23.01/03 and taring program also act on analog input. Input voltage for 0% belt load.
9947

WB2_HEB P 22.08 WB2_ANH

Parametrization FH 345 GB/ 5 - 32

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Analog Input Second Weighbridge P 22.09 10.00 Range: 0...100.00 V Input voltage for 100% belt load. MIN Value Second Weighbridge P 22.10 60.0 % Range: -10.0....200.0% Ref: See: wb2_qmin P 22.14 2.00V WB2_BER P 22.13 3.0s

WEIGHFEEDER
Check Weighbridge Filter dif_tau

Range: 0.0...3600.0s See: K 26.06, K 26.07, P 22.12, Message HI 07

Filtered value is monitored and displayed. 2nd Weighbridge Analog Output WB2_Anh

FO P22.03 = P22.01 P22.04 cos P22.05 L cosa


Message LO 06, K 26.04 MAX Value Second Weighbridge

Range : 0% = 0...10V See: K 26.08, P 22.15

2nd Weighbridge Analog Output P 22.15 10.00V Range: QNENN = 0...100V See: K 26.08, P 22.14 WB2_BER

P 22.11 140.0% Range: -10.0...200.0 % Ref: See: See P 22.10 Message HI 06, K 26.05

wb2_qmax

Check Weigher Response Threshold P 22.12 10.0% Range: 0.0...200.0% of QNENN See: K 26.06, K 26.07, P 22.13, Message HI 07 dif_max

For details, see K 26.06.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 33

WEIGHFEEDER

MULTICONT

P 23. Second Weighbridge, Calibration Results

Tare N P 23.01 0.00 kg/m Range: 0 kg/m ... 100000 kg/m Note: Tare N is acquired by setting program T2: Tare WB2". WB2_TAR P 23.03 1.000

Correction Factor KOR WB2_KOR

Range: 0.5 ... 2 Note: The correction factor effects on display proportionally. Enter by hand.

Device Constant GK P 23.02 0.171 Range: 0,0001mV/(kg/m) ... 1000 mV(kg/m) Note: The device constant is calculated internally. WB2_GK

24. Second Weighbridge, Controller

Controller is designed as P-I-D controller, special desgns type P-D and I are possible. The controller is set by setting program Contr. Adapt. P Portion KP P 24.01 20.0000 Range: 0 ... 10000 Note: KP = 0 = > I controller qa_kp P 24.02 5.00 s Range: 0 s ... 60000 s Note: l TN < 0,15 s = > P-D controller I Portion Reset Time TN qa_tn

9947

Parametrization FH 345 GB/ 5 - 34

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
D Portion Rate Time TV P 24.03 0.00 s Range: 0 s ... 60000 s Note: The rate time determines the D portion. qa_tv P 24.05 LIN

WEIGHFEEDER
Adaption Controller Output qa_xm

Range: AUS / LIN Note: Setting is by program Contr. Adapt..

Rate Filter TF P 24.04 0.00 s Range: 0 s ... 60000 s qa_tf

P 25. Dead Time

See: Note:

Block K 27 The dead time applies to all dead time elements. Dead Time

P 25.01 0.00 s Range: 0 s ... 600 s

Totzeit

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Parametrization FH 345 GB/ 5 - 35

WEIGHFEEDER
P 26. Intermittent Prefeeder

MULTICONT

Prefeeder Connection Period P 26.01 1.00s Range: 0 s... 100 s See: Note: K 17 Determines intermittent prefeeder minimum connection time. zi_ed

P 27. Pulse Period Controlled Prefeeder

Prefeeder Pulse Period P 27.01 0s Range: 0...1000 s See: K 18.... Pulse Period for Prefeeder Control Modulator Local Prefeeder Speed Control Signal, Elevation P 27.02 0.00 V Range: 0 V ... 10 V See: K 28.04 ZD_LST zp_lanh zp_il

Local Prefeeder Speed Control Signal, Range P 27.03 5.00 V Range: 0 V ... 100 V zp_ldber

9947

Parametrization FH 345 GB/ 5 - 36

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
6 Even t Messag es
6.1 General Comments
The MULTICONT system monitors the hardware, as well as the signals and parameters decisive for operating safety, feeding operation and feeder functions. Malfunctions or deviations from setpoints are reported as events. To each event a class and thus a reaction are assigned per configuration. There are four event classes with the following reactions: ALARM Warning 1 Warning 2 IGNORE Stop feeder, Display W1, Display W2, Suppress display.

WEIGHFEEDER

For identification on display at control station and for data exchange each message has a code resulting from group index and number of position within the group.

6.1.1 Display
The event messages are displayed in the event field as follows: ALARM W1 W2 underlined and flashing, underlined and permanent, permanent.

If several events occur at the same time, the following ranking applies: ALARM before W1 before W2. Within an event class in accordance with respective group as per group sequence in the event list. Within a group as per position number. Displays of classes Alarm and W1" require to be acknowledged. Display as W2 event follows, however, with the correct status being internally maintained. After remedy of cause display effaces. Displays of class W2 must not be acknowledged. They disappear after remedy of cause. The acknowledgement has a general effect on all event messages available at the same time. After start of feeder, all event messages which would occur in actual operation are suppressed during start-up phase. In Local Mode, any avaible controller fault is reset.

The event messages are assembled in groups so as to facilitate troubleshooting. SY SC WE System Faults, Safety Indications: General, Safety Indications: Electrical Equipment of Feeder, Safety Indications: Mechanical Equipment of Feeder, Safety Indications: Material Flow Interlocking, Controller, Check Measurement, Calibration, MAX Limit Values, MIN Limit Values.

WM

MF IL CO CH CA HI LO

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Event Messages FH 345 GB/ 6 - 1

WEIGHFEEDER
6.2 List of Event Messages 6.2.1 System Faults SY
This group comprises the messages of the configuration check. The following faults are acquired: (a) Double assignment of digital and analog outputs, Double assignment of L/C inputs, Plausibility of assignment of channel to channel type, Actual existence of activated units, Existence of channels assigned on unit.

MULTICONT

SY 08: LBS 2 DATA LINK Machine control station K 00.08 LB2_AKT FLB2. SY 09: SE* DATA LINK Batch processor. K 00.09 SE*_AKT SY 10: MEMORY FAIL SE memory. K 00.10 SE_MEM SY 11: q WRONG UNIT Belt load configuration. K 10.10 QU_q SY 12: v WRONG UNIT Speed configuration. K 10.11 QU_v SY 13: qb WRONG UNIT Fill weight configuration. K 21.01 B_WZ SY 14: qp WRONG UNIT 2nd weighbridge K 26.01 WB2_WZ configuration

(b) (c)

(d) (e)

If the fault is recognized at once, entry is aborted. Certain faults, however, cannot be detected before first communication with the respective unit. This is when the message is output. Check configuration upon commissioning. Check cabling and connections; check assemblies for correct installation. Assembly faults are reported by LEDs. Exchange defective components. SY 01: ME1 DATA LINK Measurement data acquisiti- K 00.01 ME1_AKT on unit FME1. SY 02: ME2 DATA LINK Measurement data acquisiti- K 00.02 ME2_AKT on unit FME2. SY 03: ME3 DATA LINK Measurement data acquisiti- K 00.03 ME3_AKT on unit FME2. SY 04: EA 1 DATA LINK Input/output unit FEA1. K 00.04 EA1_AKT SY 05: EA 2 DATA LINK Input/output unit FEA2. K 00.05 EA2_AKT SY 06: EA 3 DATA LINK Input/output unit FEA3. K 00.06 EA3_AKT SY 07: LBS 1 DATA LINK Machine control station K 00.07 LB1_AKT FLB1.

6.2.2 Safety Indications SC


SC 01: FEEDER BLOCKED Drive blocked K 01.01 BLOCKSC Feeder drive is blocked for more than set time, although drive signal is available.

SC 02: MAINTENANCE Maintenance interval K 01.02 WARTUNG reached. P 02.10 wartint System has reached the set number of operating hours. Effect requisite maintenance work.

SC 07:

Stand-By K 01.03 STANDBY

Feeder is in stand-by mode, e.g. setpoint is smaller than preset limit value. Drive is cut off, feeder continues reporting START.

9947

Event Messages FH 345 GB/ 6 - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
6.2.3 Safety Indications WE
WE 01: POWER DOWN Power failure. K 02.01 NETZAUS Power switch turned off or power failure.

WEIGHFEEDER
WE 08: PREFEEDER SCR FAIL Controller prefeeder drive. K 15.04 ZAN_REG Controller built-in in drive unit is faulty. For troubleshooting, see manufacturers instructions.

WE 02: FEED DRIVE OVERLOAD Feed drive overload K 02.02 MS_DAN Protective motor switch triggered. Check motor and load on conveyor belt. When starting feeder, observe process-specific starting conditions.

WE 09: PREFEEDER TEMP. HI Prefeeder drive temperature K 15.05 TEMP_ZA too high. Temperature of prefeeder drive is too high.

WE 03: FEED DRIVE SCR FAIL Controller feed drive K 02.03 DAN_REG Failure of external controller. For trouble shooting, see manufacturers instructions.

WE 10: ANALOG INPUT P Error analog input Set K 02.05 FP_ANAL Feed Rate K 07.10 TM_PABS K 08.10 M0_PABS K 09.11 M1_PABS P 08.01 adan1 P 08.02 adbe1 Signal at analog input Set Feed Rate is out of admissible range. Check connection, cabling, and setpoint transducer.

WE 04: CABLE SHORT OR INT. Cable breakage/short circuit. K 02.04 F_KABEL Cable breakage or short circuit in one of the lines with NAMUR limit switch. WE 11: ANALOG INPUT MOISTURE Error analog input material P 25.08 FEU_FAN moisture. P 21.02 FEU_ANH P 21.03 FEU_BER See: WE 10

WE 05: MS: PREFEEDER OVERLOAD Prefeeder drive motor proK 15.01 MS_ZAN tection. Motor protective switch of prefeeder drive has broken circuit.

WE 12: FEED DRIVE MAX TEMP Prefeeder drive temperature K 02.06 DA_TEMP too high. Prefeeder drive temperature too high. Protective switch has broken circuit.

WE 06: MS: DRIVE 1 OVERLOAD Motor protective switch K 15.02 MS_H1AN auxiliary drive 1. Motor protective switch of auxiliary drive 1 has broken circuit.

WE 07: MS: DRIVE 2 OVERLOAD Motor protective switch K 15.03 MS_H2AN auxiliary drive 2. Motor protective switch of auxiliary drive 2 has broken circuit.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Event Messages FH 345 GB/ 6 - 3

WEIGHFEEDER
6.2.4 Safety Indications WM
WM 01: BELT TRACKING Belt tracking. K 03.01 SCHIEF Belt drifts extremely right or left, correct functioning cannot be granted. Belt tracking can be monitored by limit switches or sensors (Block K 11).

MULTICONT
6.2.5 Safety Indications MF
MF 01: BULK SOLID SENSOR Material sensor flap. K 04.01 ST_MAT A sensor flap for monitoring feed of material flow can be mounted at the prefeeder. The message is output if feeder is connected and no material is fed.

WM 02: BELT TENSION Wrong belt tension. K 03.02 BAND_SP Belt tension is monitored by additional limit switches. Check tension.

MF 02: NO PREFEED No material after start-up.

K 04.02 QMIN_AN P 10.02 fahrtim P 10.03 anfahr P 06.03 qmin

WM 03: BELT SLIP Belt slip out of tolerance.

K 03.03 SLIP P 06.07 Slip

Material flow during start-up phase is monitored via belt load. After the set time (minimum 10 s) belt load must have exceeded limit value qmin.

Belt slip can be monitored by limit switches, two pulley sensors (K 3) or one belt sensor.

MF 03: MATERIALS > MAX. Excess of material. K 04.03 MAT_MAX The external control (sensor flap) is closed. There is too much material on belt.

WM 04: BELT TRACKING LEFT Belt tracking left. K 03.04 SCHIEFL Belt travels left of correct position. Position can be displayed as service value.

6.2.6 Interlocking IL
IL 01: EMERGENCY STOP Emergency stop. K 05.01 NOTAUS An emergency stop switch was operated.

WM 05: BELT TRACKING RIGHT Belt tracking right. K 03.05 SCHIEFR Belt travels right of correct position. Position can be displayed as service value.

IL 02: LOCAL MAIN SWITCH Safety disconnector. K 05.02 FREISCH Safety disconnector signal is not available.

IL 03: NO RELEASE No release.

K 05.06 NO_FREI K 05.03 FREI_1 K 05.04 FREI_2 K 05.05 FREI_3

Release 1 and 2 signals or Release 3 signal are not available or Local Mode is not active.
9947

Event Messages FH 345 GB/ 6 - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
6.2.7 Controller CO
CO 01: Deviation. DEVIATION K 11.02 DA_REGA P 04.03 difmax P 04.04 regtim

WEIGHFEEDER
CH 04. CORRECTION 0 ABS HI Correction range zero K 23.11 KME_KTA point absolutely exceeded. P 20.12 KMETABS Absolute zero point error is higher than set limit value.

Deviation exceeds the limit set for the preset time.

CH 05: CORRECTION 0 REL HI Correction range zero K 23.12 KME_KTR point relatively exceeded. P 20.11 KMETREL Relative zero point error is higher than set limit value.

CO 02: SETPOINT LIMITED Set feed rate limited. K 11.03 NOTSOL P 04.12 Notsol Entered feed rate setpoint is too high and was reduced to nominal feed rate, or preset emergency setpoint is active.

CH 06: CORR. SPAN ABS HI Correction range span K 23.13 KME_KBA absolutely exceeded. P 20.14 KMEBABS Absolute span error is higher than set limit value.

CO 03: CONTROLLER OUT LIMIT Control magnitude of conK 11.04 REG_ERR troller out of limit. The controller control magnitude is limited to 10 V, or 0 V. Frequent occurance can effect on feeding accuracy. Check material discharge and controller setting. Check controller adaptation in the FAE unit.

CH 07: CORR. SPAN REL HI Correction range span K 23.14 KME_KBR relatively exceeded. P 20.13 KMEBREL Relative span error higher than set limit value.

6.2.8 Check Measurement CH


CH 01: CHECK QUANTITY < MIN Comparative check amount K 23.10 KME_KVM too small. P 20.07 KME_MIN Minimum check amount not reached.

CH 08: FEEDBACK PREFEEDER Fill/discharge unit feedback K 23.06 KME_FEA signal missing. Check system feed back signal is missing, fill/discharge unit of system is not OFF.

CH 09: NO PERMISSION Check logged on. K 23.15 KME_KAN K 23.01 KME_FR Check system logged on, release signal not available.

CH 02: BIN LEVEL LO Minimum check fill weight. K 23.08 KMEBMIN P 20.09 KMEBMIN Minimum check fill weight exceeded down.

CH 10: PREPARATION ACTIVE Prepare check system. K 23.16 KME_KVO CH 03: BIN LEVEL HI Maximum check fill weight. K 23.09 KMEBMAX P 20.10 KMEBMAX Check system correctly prepared.

9947

Check supply quantity higher than set maximum check fill weight.

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Event Messages FH 345 GB/ 6 - 5

WEIGHFEEDER
CH 11: ACCEPTANCE REQ. Check system expects K 23.18 KME_WQU acceptance. After computation phase check system expects acceptance. Accept correction by operating key ENTER or reject with ESCAPE.

MULTICONT
CA 06: BIC ASYNCHRONOUS BIC asynchronous. K 14.04 BIC_ERR Belt marking pulse not received in expected time. Cut off BIC and check belt tension.

CA 07: LC BIN < MIN Load cell input fill weight underamplified.

K 21.08 WZVB_MI

6.2.9 Calibration CA
CA 01: LC OUT < MIN Load cell input underampli- K 13.01 WZ_MIN fied. Load cell input is underamplified. In case of frequent occurance check mechanical and electrical equipment.

CA 08: LC BIN > MAX Load cell input fill weight K 21.09 WZVB_MA overamplified. CA 09: LC 2. WB <MIN Load cell input 2nd weigh- K 26.02 WB2_MIN bridge underamplified. CA 10: LC 2. WB >MAX Load cell input 2nd weigh- K 26.03 KWB2_MAX bridge overamplified.

CA 02: LC INP > MAX Load cell input overamplified.

K 13.02 WZ_MAX

Input of load cell measuring amplifier is overamplified.

CA 03: PLL OVERFLOW PLL overflow. K 13.03 GA_ERR Conveyor belt speed generator does not supply usable signal. Check speed generator and cabling.

CA 04: SIMULATION ACTIVE Simulator is active. K 13.04 SIM_AKT Load cell signal, or speed simulator, in test mode is ON.

CA 05: CHECK WEIGHT IN POSITION Check weight in position K 16.02 PRG_LA during normal operation. Check weight of automatic check system in position.
9947

Event Messages FH 345 GB/ 6 - 6

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
6.2.10 MAX Limit Values HI
HI 01: FEEDRATE I HI Actual feed rate higher K 12.02 K_MAX PMAX. P 06.02 pmax Actual feed rate is higher than set maximum feed rate.

WEIGHFEEDER
6.2.11 MIN Limit Values LO
LO 01: FEEDRATE I LO Actual feed rate smaller K 12.01 K_MIN PMIN. P 06.01 pmin Actual feed rate is smaller than set limit feed rate.

HI 02: LOAD Q HI Belt load higher q_MAX.

LO 02: LOAD Q LO Belt load smaller qmin. K 12.04 P 06.04 Q_MAX qmax

K 12.03 Q_MIN P 06.03 qmin

Belt load is smaller than set limit belt load. Belt load is higher than set maximum load. LO 03: SPEED V LO Belt speed smaller vMIN: K 12.06 V_MAX P 06.06 vmax

HI 03: SPEED V HI Belt speed higher vMAX.

K 12.05 V_MIN P 06.05 vmin

Belt speed is smaller than set limit belt speed. Belt speed is higher than set limit speed. LO 04: BIN LEVEL LO Fill weight smaller BMIN. K 21.05 FVB_MIN P 18.04 vbmax Fill weight is smaller than limit value BMIN. Fill weight of supply bin is higher than set limit value. LO 06: Load 2nd WB<MIN 2nd Weighbridge K 26.04 WB2QMIN P22.10 wb2qmin K 25.06 FEU_MAX P 21.01 FEU_MAX

HI 04: BIN LEVEL HI Fill weight higher BMAX.

K 21.06 FVB_MAX P 18.05 vbmax

HI 05: MOISTURE HI Moisture higher MAX.

Moisture is higher than set limit value.

HI 06 Load 2nd WB>MAX 2nd Weighbridge K 26.05 WB2QMAX P 22.11 wb2qmax HI 07: |Q2-Q1| >MAX K 26.07 WB2_DIF P 22.12 dif_max Difference between the two weighbridges is too big.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Event Messages FH 345 GB/ 6 - 7

MULTICONT
Annex A: Alphabetic Search List
Abort Batch, Mode 0 Abort Batch, Mode 1 Abort Batch, Test Mode Absolute Range BA Absolute Setpoint Absolute Setpoint Absolute Setpoint Absolute Zero Point NA Acceleration Limit Accept Correction Accept Span Correction Accept Zero Point Correction Acknowledge Event Message Activate Emergency Setpoint Actual Feed Rate < PMIN Actual Feed Rate > PMAX Actual Feed Rate 1 Actual Feed Rate 1, Elevation Actual Feed Rate 1, Range Actual Feed Rate 2 Actual Feed Rate 2, Elevation Actual Feed Rate 2, Range Actual Feed Rate Analog Output Actual Feed Rate Corrected, Elevation Actual Feed Rate Corrected, Range Actual Feed Rate Display Actual Feed Rate, Dry Actual Feed Rate, Limit Value PMAX Actual Feed Rate, Limit Value PMIN Actual Feed Rate SE Bus Actual Value Filter Adaptation Controller Input Adaptation Controller Output Adaptation Controller Output Adaptation Factor Afterflow Totalization Analog Input Second Weighbridge Analog Output Second Weighbridge Angle a Angle a Apply Check Weight Attenuation Automatic Weight Check Aux. Drive 1 Overload K 20.04 K 20.05 K 20.06 P 20.14 K 09.10 K 07.10 K 08.10 P 20.12 P 04.07 K 23.04 K 24.01 K 24.02 K 10.01 K 11.01 K 12.01 K 12.02 K 10.13 P 07.01 P 07.02 K 10.14 P 07.03 P 07.04 M0_BATA M1_BATA TM_BATA KMEBABS M1_PABS TM_PABS M0_PABS KMETABS bgrenz KME_KUI KME_KUB KME_KUT STCLR QU_NOTS K_MIN K_MAX P_IST1 P1_Anh. P1_Ber. P_IST2 P2_Anh. P2_Ber.

WEIGHFEEDER

Aux. Drive 2 Overload Batch Processor Batch Setpoint Mode 0 Batch Setpoint Mode 1 Batch Setpoint Test Mode Belt Circuit Code Number IB Belt Circuit Time TBAND Belt Drive Control Magnitude Belt Drive START Belt Influence Compensation Filter Belt Length Platform Discharge Belt Load Belt Load Belt Load < qMIN Belt Load > qMAX Belt Load 2nd Weighbridge Too High Belt Load 2nd Weighbridge Too Small Belt Load, Elevation Belt Load Limit Value qMAX Belt Load Limit Value qMIN Belt Load q Belt Load q < qMIN Belt Load, Range Belt Slip Belt Slip, Belt Length Change SMAX Belt Speed Belt Speed < vMIN Belt Speed > vMAX Belt Speed, Analog Output Belt Speed Limit Value vMAX Belt Speed Limit Value vMIN Belt Speed v Belt Tension Belt Tracking Belt Tracking Measurement BIC Asynchronous BIC ON Bin Filling START Blind Distance Factor Bypass BYP Cable Breakage/Short Circuit Check Logged On Check Measurement Active Check Quantity Check Quantity < MIN

K 15.03 MS_H2AN K 00.09 K 20.10 K 20.11 K 20.12 P 03.07 P 02.02 K 10.12 K 10.00 P 09.01 P 09.02 K 10.15 K 10.10 K 12.03 K 12.04 SE*_AKT M0_PBAT M1_PBAT TM_PBAT Imp/Uml tband STG_DO ANTR_ON BAT_Fil DAP_LNG Q_BAND QU_q Q_MIN Q_MAX

K 26.05 wb2_qmax K 26.04 P 07.05 P 06.04 P 06.03 P 05.04 K 04.02 P 07.06 K 03.03 P 06.07 K 10.11 K 12.05 K 12.06 K 16.10 P 06.06 P 06.05 P 05.05 K 03.02 K 03.01 K 14.05 K 14.04 K 14.02 K 21.07 P 20.03 P 12.07 K 02.04 K 23.15 K 23.17 P 20.06 K 23.10 wb2_qmin Q_Anh. qmax qmin qtau QMIN_AN Q_Ber SCHLUPF Schlupf QU_v V_MIN V_MAX V_BAND vmax vmin vtau BAND_SP SCHIEF BPM_EIN BIC_ERR BIC_EIN FILORVB KME_BL qm_byp F_KABEL KME_KAN KME_KON KME_SOL KME_KVM

P 05.02 pdatau P 21.04 PTR_ANH P 21.05 PTR_BER P 05.01 plbstau K 25.03 PTROCK P 06.02 pmax P 06.01 P 05.03 P 19.03 P 12.10 P 12.11 P 24.05 P 15.04 P 10.01 pmin psetau B_rtau qm_xe qm_xm qa_xm hkadap tnach

P 22.09 WB"_BER P 22.08 P 22.05 P 02.06 K 16.07 P 19.05 K 16.01 K 15.02 WB2_ANH WB2_WIN winkel PRG_AUF B_absch PRG_AUT MS_H1AN

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

A: Alphabetic Search List FH 345 GB/ A - 1

WEIGHFEEDER
Check Storage Quantity Check System Expects Acceptance Check System Type EAUS Active Check System Type EEINAcive Check System Type EXTERN Active Check Weighbridge Check Weighbridge Event Message Check Weigher Filter Check Weigher Response Threshold Check Weight Check Weight in Position Circuits Circuits/Quantity Check Setpoint Clearance Mode Control Delay ON Control Magnitude Elevation ANH Control Magnitude Elevation ANH Control Magnitude Prefeeder Controller Feeder Drive Controller Out Of Limit Corr. Range Exceeded, Rel. Span Corr. Range Exceeded, Abs. Span Corr. Range Exceeded, Abs. Zero Point Corr. Range Exceeded, Rel. Zero Point Correction Evaluation Correction Factor KOR Correction Modes Manual/Automatic Correction Speed Correction Value: Span Correction Value: Zero Point Counter 1, Pulse Output Corrected Counter 2, Pulse Output Corrected Cycle Time Cyclic Auto Start Span Cyclic Auto Start Tare D Portion Rate Time TV D- Portion Rate Time TV D Portion Rate Time TV D Portion: Rate Time TV P 20.08 KME_PMV

MULTICONT
P 20.02 KME_TB P 25.01 Totzeit K 27.12 TOT_PON K 27.04 TOT_TM1 K 27.03 TOT_TM0 K 27.01 TOT_VM0 K 27.02 TOT_VM1 K 27.13 TOT_ON K 27.10 TOT_TI1 K 27.09 TOT_TI0 K 27.08 TOT_IM1 K 27.07 TOT_IM0 K 27.11 TOT_GEW P 22.07 ZUT_T2B P 11.01 K 11.10 P 07.10 P 04.03 P 07.09 P 23.02 P 03.05 P 03.08 K 21.04 ZUT_Tot AA_REGA r_Ber. difmax r_Anh. WB2_GK GK dt_kenn B_TYP_A

Damping Time Dead Time K23.18KME_WQU Dead Time Element Setpoint Active K 22.02 KME_AUS Dead Time Element Setpoint K 22.01 KME_EIN Clock ON M1 Dead Time Element Setpoint K 22.03 KME_EXT Clock ON M0 K 26.06 Wb2_ctl Dead Time Element Setpoint Preselect M0 K 26.07 wb2_dif Dead Time Element Setpoint P 22.13 dif_tau Preselect M1 Dead Time Element P 22.12 dif_max START/STOP Active P 02.09 qprf Dead Time Element K 16.02 PRG_LA START/STOP, Clock ON M1 P 20.05 KMEBAND Dead Time Element START/STOP, Clock ON M0 P 20.04 KME_VOR Dead Time Element P 10.04 leerfa START/STOP Preselect M1 K 14.03 VAP_EIN Dead Time Element START/STOP Preselet M0 P 19.12 vbanh Dead Time Elements Preselected P 04.08 anh Dead Time Prefeeder Second K 18.11 ZD_NST Weighbridge K 15.04 ZAN_REG Dead Time Prefeeder K 11.04 REG_ERR Weighbridge Deviation K 23.14 KME_KBR Deviation, Range Deviation Threshold K 23.13 KME_KBA Deviation, Zero Point Device Constant GK K 23.11 KME_KTA Device Constant GK Diagnosis Tool K 23.12 KME_KTR Discharge/Fill Control P 20.15 KME_BEW Display and Output Format P 23.03 WB2_KOR Totalizing Counter 1 Display and Output Format P 20.01 KME_KOR Totalizing Counter 2 P 20.16 KME_KV Do Not Accept Correction K 24.04 KME_KGB Drive Signal Prefeeder, Local K 24.05 KMT_KGT Drive Signal Prefeeder Local, Elevation K 25.04 FM1_TR Drive Signal Prefeeder Local, Range K 25.05 FM2_TR Drive Signal Speed P 20.17 KME_ZYK Drives OK K 22.05 KME_BAU Dynamic Factor K 22.04 KME_TAU Deviation Threshold P 24.03 P 12.03 P 14.05 P 19.08 qa_tv qm_tv zs_tv B_tv

P 01.05 fm1dim P 01.07 fm2dim K 23.05 KME_KNO K 18.10 ZD_LST P 13.01 zd_lanh P 13.02 K 06.10 K 02.00 P 11.04 P 04.13 zd_Idber LO_STEL ANTR_OK ZUT_Dyn Xd_Fak

Effective Platform LengthL2EFF P 22.01 WB2_LNG Effective Platform Length LEFF P 01.03 blaenge Effective Tare N + T P03.02 qtara
9947

A: Alphabetic Search List FH 345 GB/ A - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Electrical Equipment Ready Elevation Control Magnitude ANH Emergency Control Check Status Emergency Setpoint Limit Emergency Setpoint Limit Emergency Stop Error: Controller Feed Drive Error: Feed Drive Externally Ready Feed Drive Overload Feed Drive Temperature > MAX Feed Rate, Dimension and Display Format Feed Rate Pre-Act Contact Feed Unit OK Feeder Blocked Feeder Blocked Response Time Feeder START Feeder START Feeder START Feeder START Feeder START/STOP Feeder START/STOP Feeder START/STOP Fill Control, Limit Value STMAX Fill Control, Limit Value STMIN Fill/Discharge Unit Stopped Fill Weight Fill Weight Fill Weight < BMIN Fill Weight > BMAX Fill Weight, Analog Output, Elevation Fill Weight, Analog Output, Range Fill Weight Control Magnitude Fill Weight Control START Fill Weight, Display Format and Dimension Fill Weight Filter Fill Weight, Limit Value BMAX Fill Weight, Limit Value BMIN Fill Weight Setpoint Filter Controller Input Actual Feed Rate Filter Deviation General Alarm General Warning I Portion Reset Time TN I Portion Reset Time TN K 05.07 EL_BER P 12.06 qm_anh P 15.16 P 04.12 P 10.05 K 05.01 K 02.03 K 02.02 K 05.08 B_kost notsol Not_Sol NOTAUS DAN_REG MS_DAN EX_BER

WEIGHFEEDER
I Portion: Reset Time TN I Portion: Setback Time TN Initialize Controller Initialize Setpoint Dead Time Element Input/Output Unit 1 Input/Output Unit 2 Input/Output Unit 3 Interference Quantity Effect STKP Interference Quantity Effect STTV L/C Input < MIN L/C Input > MAX Language Lever Ratio Lever Ratio Load Cell Characteristic E Load Cell Characteristic ValueE Load Cell Characteristic ValueE Load Cell Filter Load Cell Filter Load Cell Filter Load Cell Input Fill Weight Overamplified Load Cell Input Fill Weight Underamplified Load Cell Input 2nd Weighbridge over amplified Load Cell Input 2nd Weighbridge under amplified Load Cell Rated Capacities WZNENN Load Cell Rated Capacity BWZNENN Load Cell Rated Capacity WZNenn Local Prefeeder Setpoint Local Prefeeder Speed Control Signal, Elevation Local Prefeeder Speed Control Signal, Range Lower Control Magnitude Limit FGu Lower Control Magnitude Limit Gu Lower Control Magnitude Limits Gu Machine Control Station 1 Machine Control Station 2 Main Contact Maintenance Interval Maintenance Interval Reached P 12.02 qm_tn P 19.07 B_TN K 21.03 B_INIT K 27.06 K 00.04 K 00.05 K 00.06 TOT_INI EA1_AKT EA2_AKT EA3_AKT

P 19.10 B_stkp P 19.11 B_sttv K 13.01 K 13.02 P 00.02 P 22.04 P 02.05 P 16.04 P 02.03 P 22.02 P 22.06 P 02.08 P 16.05 WZ_MIN WZ_MAX spr WB2_HEB Hebel vwz_ken wz_kenn WM2_KEN WB2_FIL wz_filt vb_filt

K 15.01 MS_ZAN K 02.06 TEMP_DO P 01.01 P 15.01 K 03.00 K 01.01 P 04.10 K 07.00 K 09.00 K 08.00 K 06.00 K 07.02 K 09.02 K 08.02 P 18.03 P 18.02 K 23.06 K 21.10 K 21.01 K 21.05 K 21.06 pdim dpvor FOERDER BLOCKSC tblksch TM_EIN M1_EIN M0_EIN LO_EIN TM_I/O M1_I/O M0_I/O vhmax vbmin KME_FEA FGVB QU_FGVB FVB_MIN FVB_MAX

K 21.09 WZVB_MA K 21.08 WZVB_MI K 26.03 WB2_MAX K 26.02 WB2_MIN P 02.04 wz_nenn P 16.03 vwz_nen P 22.03 WB2_F0 K 28.04 ZP_LST P 27.02 zp_lanh P 27.03 zp_idber P 19.14 B_gu P 04.09 r_gu P 12.08 qm_min K 00.07 K 00.08 K 20.08 P 02.10 K 01.01 LB1_AKT LB2_AKT HAUPTK wartint WARTUNG

P 16.06 vbanh P 16.07 vbdakor K 21.11 B_STG K 21.02 B_RON P 16.01 P 18.01 P 18.05 P 18.04 P 19.01 vbdim vbtau vbmax vbmin B_soll

P 04.06 reintau P 04.05 rtau K 10.04 S_ALARM K 10.05 S_WARN P 24.02 qa_tn P 14.04 zs_tn

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

A: Alphabetic Search List FH 345 GB/ A - 3

WEIGHFEEDER
Material > MAX Material Moisture Material Moisture, Elevation Material Moisture, Range Material Moisture, Wrong Analog Input MAX Value Second Weighbridge Maximum Check Fill Weight Maximum Check Fill Weight Measurement Data Acquisition Unit 1 Measurement Data Acquisition Unit 2 Measurement Data Acquisition Unit 3 Memory SE Message: Auxiliary Drive started Message: Check Measurement Started Message: Deviation Active Emergency Setpoint Message: Feeder Started MIN Value Second Weighbridge Minimum Check Fill Weight Minimum Check Fill Weight Minimum Check Quantity Mode 0 Preselection Active Mode 1 Preselection Active Moisture Correction Active Moisture Correction ON Moisture Limit Value FEUMAX Moisture MAX Multikey K 04.03 K 25.02 P 21.02 P 21.03 MAT_MAX MATFEU FEU_ANH FEU_BER

MULTICONT
P 14.01 zs_anh P 14.02 zs_ber K 02.01 NETZAUS K 20.07 VORKON K 27.05 K 19.05 P 26.01 P 11.02 K 19.07 K 19.00 K 19.06 K 17.00 K 18.00 K 19.02 K 19.03 K 19.04 K 19.15 K 19.16 TOT_VOR ZS_LZU zi_ed ZUT_Kon ZS_AZU ZS_AON ZS_AAUF ZI_AON ZD_AON ZS_MAX ZS_MIN ZS_LAUF ZS_STM ZS_ADO

Position Signal Prefeeder, Elevation Position Signal Prefeeder, Range Power Failure K 25.08 FEU_FAN Pre-contact Preassign Setpoint Dead Time Element P 22.11 wb2_qmax P 20.10 KMEBMAX Prefeeder CLOSE Local K 23.09 KMEBMAX Prefeeder Connection Period Prefeeder Constant K 00.01 ME1_AKT Prefeeder Drive CLOSE Prefeeder Drive ON K 00.02 ME2_AKT Prefeeder Drive OPEN Prefeeder Drive START K 00.03 ME3_AKT Prefeeder Drive START Prefeeder Limit Position MAX K 00.10 SE_MEM Prefeeder Limit Position MIN Prefeeder OPEN, Local K 10.06 H_EIN Prefeeder Position Signal K 23.07 KME_AKT Prefeeder Position Signal K 11.02 DA_REGA Prefeeder Position Signal Output, Span K 11.03 NOTSOL Prefeeder Position Signal K 10.03 W_EIN Output, Zero Point P 22.10 wb2_qmin P 20.09 KMEBMIN Prefeeder Pulse Period K 23.08 KMEBMIN Prefeeder START P 20.07 KME_MIN Prefeeder START Prefeeder START, Bin Monitor K 08.05 M0_AKT Prefeeder START Local K 09.05 M1_AKT Prefeeder START Local K 25.07 FEU_AKT Prefeeder START/STOP K 25.01 FEU_EIN P 21.01 FEU_MAX Prefeeder Temperature > MAX K 25.06 FEU_MAX Prefeeder Type Active Prefeeder Type Active K 00.11 MK_AKT Prefeeder Type Active Prefeeder Type Active No EMERGENCY-OFF K 05.13 NO-NOT Prepare Check Measurement No Release K 05.06 NO_FREI Preselect Batching Mode Test No SI Units K 00.12 NON-SI P 03.01 qnenn Mode Nominal Belt Load qNENN P 01.02 pnenn Nominal Feed Rate PNENN Preselect Batching Mode 0 Nominal Fill Weight P 16.02 vbnenn Preselect Batching Mode 1 Nominal Speed VNENN P 01.04 vnenn Preselect Local Mode Nominal Transit Time Positioner P 11.03 ZUT_LZ Preselect Local Mode Active Normal Control Check Status P 19.17 B_kon Preselect Memory Number of Belt Circuits NBAND P 02.01 nband Preselect Mode 0 Preselect Mode 1 P Portion KP P 24.01 qa_kp Preselect Test Mode P Portion KP P 12.01 qm_kp Pulley Slip P Portion KP P 04.01 P_Ant Pulse Length, Totalizing P Portion KP P 19.06 B_kp Counter 1 P Portion KP P 14.03 zs_kp Pulse Length, Totalizing Password Active P 00.01 passakt Counter 2 PLL Overflow K 13.03 GA_ERR Rate Filter TF

P 14.08 zs_aber P 14.07 K 27.01 K 18.02 K 17.02 K 17.04 K 18.03 K 17.03 K 16.06 K 15.05 K 19.01 K 28.01 K 17.01 K 18.01 K 23.16 K 20.03 K 20.01 K 20.02 K 06.01 K 06.04 P 19.15 K 08.01 K 09.01 K 07.01 P 06.08 zs_aanh zp_il ZD_NON ZI_NON ZI_BON ZD_LON ZI_LON Z_RIEG TEMP_ZA ZS_AKT ZP_AKT ZI_AKT ZD_AKT KME_KVO TM_BATQ M0_BATQ M1_BATQ LO_VORW L0_AKT B_Isave M0_VORW M1_VORW TM_VORW TRSCHLU

P 01.06 fm1puls P 01.08 fm2puls P 24.04 qa_tf


9947

A: Alphabetic Search List FH 345 GB/ A - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Rate Filter TF Rate Filter TF Ready in Mode 1 Reaction Time Deviation Ready in Mode 0 Ready to Start in Mode 0 Ready to Start in Mode 1 Relative Range BR Relative Setpoint Relative Setpoint Relative Zero Point KP Release 1 Release 2 Release 3 Release Check System Reset Time TN Reset Totalizing Counter 1 Reset Totalizing Counter 1 Reset Totalizing Counter 1 Reset Totalizing Counter 2 Reset Totalizing Counter 2 Reset Totalizing Counter 2 Residual Amount Main Contact Residual Amount Pre-Act Contact Response Threshold Deviation Response Threshold Deviation Safety Disconnector Second Weighbridge Sensor Flap Sensor Length Sensor Offset Sensor Width Setpoint Setpoint Analog Input Range < MIN Setpoint Filter Setpoint, Span Setpoint, Span Setpoint, Span, Local Setpoint, Zero Point Setpoint, Zero Point Setpoint, Zero Point, Local Simulator Active Slip Monitoring Transmitter 1 Slip Monitoring Transmitter 2 Span Correction KOR Span Correction KOR SPC Filter P 12.04 P 19.09 K 05.10 P 04.04 K 05.09 K 05.11 K 05.12 P 20.13 K 08.11 K 09.11 P 20.11 K 05.03 K 05.04 K 05.05 K 23.01 P 04.02 K 08.03 K 09.03 K 07.03 K 07.04 K 09.04 K 08.04 P 15.03 qm_tf B_tf M1_BER regtim M0_BER M0_EBER M1_EBER KMEBREL M0_PREL M1_PREL KMETREL FREI_1 FREI_2 FREI_3 KME_FR tnast M0_RES1 M1_RES1 TM_RES1 TM_RES2 M1_RES2 M0_RES2 hbatch

WEIGHFEEDER
SPC Time Speed Elevation Speed Span Standby Standby Start Belt Drive Start External Check System Start Span Correction Start-up Mode Start-Up/Clearance Time Start Zero Point Correction START Local Prefeeder START Prefeeder START Prefeeder Drive Stop Belt Drive5 Stop Totalizing Counter 1 Stop Totalizing Counter 2 Storage I Portion Synchronous Pulse Tare Correction T Tare N Tare N Tare N Taring Test Mode Preselection Active Totalizing Counter 1 Totalizing Counter 2 Upper Control Magnitude Limit FGo Upper Control Magnitude Limits Go Volumetric Volumetric Mode VS Characteristic Value Warning: Belt Tracking Left Warning: Belt Tracking Right Weight Check Wrong Material Flow XD Filter 2nd Weighbridge 2nd Weighbridge Only 2nd Weighbridge Analog Output 2nd Weighbridge Analog Output P 12.14 P 07.07 P 07.08 P 04.11 K 01.03 K 06.02 K 24.03 K 23.02 P 10.03 P 10.02 K 23.03 K 28.03 K 28.02 K 28.00 K 06.03 K 16.03 K 16.04 P 12.05 K 14.01 spczeit v_Anh v_Ber standby STANDBY LO_STAR KME_STE KME_SBI anfahr fahrtim KME_STI ZP_LON ZP_NON ZP_AON LO_STOP QU_1ST QU_2ST I_Save SYNCHRO

P 15.02 vbatch P 14.06 zs_e P 19.04 B_schw K 05.02 K 26.01 K 04.01 P 02.11 P 12.13 P 12.12 K 16.11 K 02.05 P 19.02 P 08.02 P 07.12 P 08.04 P 08.01 P 07.11 P 08.03 K 13.04 K 03.06 K 03.07 P 17.02 P 03.06 P 12.15 FREISCH WB2_WZ MAT_TK s_lng s_offs s_br P_SOLL FP_ANAL B_fil adbe1 PS_Ber. LDr_Ber adan1 Ps_Anh. LDr_Anh SIM_AKT SCHLU_1 SCHLU_2 vb_kor KOR spcfilt

P 03.04 Tarkor P 17.01 vb_tara P 23.01 WB2_TAR P 03.03 Tara_N K 16.12 DE_TARA K 07.05 TM_AKT K 10.08 FM1_IA K 10.09 FM2_IA

P 19.13 B_go P 12.09 K 16.08 K 10.02 P 02.07 K 03.04 K 03.05 K 16.05 K 04.00 qm_max VOL_AKT VOLUM vs_kenn SCHIEFL SCHIEFR G_KONTR ST_MAT

P 12.16 xdfilt K 26.08 K 26.09 P 22.14 P 22.15 WB2_AA WB2_AUS WB2_Anh WB2_BER

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

A: Alphabetic Search List FH 345 GB/ A - 5

MULTICONT
Annex B: List of Interfaces
B.1 Digital Inputs DE
NAME Error: Feed Drive Error: Controller Feed Drive Feed Drive Temperature > MAX Belt Tracking Belt Tension Belt Slip Slip Monitoring Transmitter 1 Slip Monitoring Transmitter 2 Sensor Flap Material > MAX EMERGENCY STOP Safety Disconnector Release 1 Release 2 Release 3 Preselect Local Mode Start Belt Drive Stop Belt Drive5 Feeder START/STOP Reset Totalizing Counter 1
9947

WEIGHFEEDER

NUMBER K 02.02 K 02.03

CODE MS_DAN DAN_REG

DEFAULT VALUE EA1.DE2 -.NO A A

VALUE

SWITCH

K 02.06

DA_TEMP

-.NO

K 03.01 K 03.02 K 03.03 K 03.06

SCHIEF BAND_SP SCHLUPF SCHLU_1

ME1.DE -.NO

A A A

-.NO -.NO

K 03.07

SCHLU_2

-.NO

K 04.01 K 04.03 K 05.01 K 05.02 K 05.03 K 05.04 K 05.05 K 06.01 K 06.02 K 06.03 K 07.02 K 07.03

MAT_TK MAT_MAX NOTAUS FREISCH FREI_1 FREI_2 FREI_3 LO_VORW LO_STAR LO_STOP TM_I/O TM_RES1

ME1.DE4 -.NO EA1.DE1 EA1.DE1 EA1.DE6 -.YES -.NO ME1.DE1 ME1.DE2 ME1.DE3 LB1 LB1

W1 W1 A A A

p p p p p p o o o o o o o o o o o o oq oq
List of Interfaces FH 345 GB/ B - 1

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

WEIGHFEEDER
NAME Reset Totalizing Counter 2 Preselect Mode 0 Feeder START/STOP Reset Totalizing Counter 1 Reset Totalizing Counter 2 Presect Mode 1 Feeder START/STOP Reset Totalizing Counter 1 Reset Totalizing Counter 2 Acknowledge Event Message Volumetric Mode Activate Emergency Setpoint Synchronous Pulse Feed Drive Loverload Aux. Drive 1 Overload Aux. Drive 2 Overload Controller Feeder Drive Prefeeder Drive Temperaure > MAX Check Weight in Position Stop Totalizing Counter 1 Stop Totalizing Counter 2 Weight Check Prefeeder START/STOP Taring NUMBER K 07.04 K 08.01 K 08.02 K 08.03 K 08.04 K 09.01 K 09.02 K 09.03 K 09.04 K 10.01 CODE TM_RES2 DEFAULT VALUE LB1 LB1 ZBS ZBS ZBS LB1 LB1 LB1 LB1

MULTICONT
VALUE SWITCH

M0_VORW M0_I/O M0_RES1 M0_RES2 M1_VORW M1_I/O M1_RES1 M1_RES2 STCLR

EA1.DE4

K 10.02 K 11.01

VOLUM QU_NOTS

-.NO -.NO -.NO EA1.DE5 -.NO -.NO -.NO -.NO A A A A A

K 14.01 K 15.01 K 15.02 K 15.03 K 15.04 K 15.05

SYNCHRO MS_ZAN MS_H1AN MS_H2AN ZAN_REG TEMP_ZA

K 16.02 K 16.03 K 16.04 K 16.05 K 16.06 K 16.12

PRG_LA QU_Z1ST QU_Z2ST G_KONTR Z_RIEG DE_TARA

EA2.DE4 -.NO -.NO LB1 LB1 -.NO

W1

o oq oq oq o oq oq o o o o o o p p p p o o o o o o oq
9947
Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

List of Interfaces FH 345 GB/ B - 2

Technische Redaktion BVED

MULTICONT
NAME Prefeeder START Prefeeder START Local Prefeeder START , Bin Monitor Prefeeder START Prefeeder START Local Prefeeder Limit Position MAX Prefeeder Limit Position MIN Prefeeder OPEN, Local Prefeeder CLOSE Local Preselect Batching Mode 0 Preselect Batching Mode 1 Preselect Batching Mode Test Mode Abort Batch Mode 0 Abort Batch Mode 1 Abort Batch Test Mode Fill Weight Control START Initialize Controller Release Check System Start Span Correction Start Zero Point Correction Accept Correction Do Not Accept Correction Fill/Discharge Unit Stopped
9947

WEIGHFEEDER
NUMBER K 17.02 K 17.03 K 17.04 CODE ZI_NON ZI_LON ZI_BON DEFAULT VALUE YES VALUE SWITCH

ME1.DE6 ME1.DE4

K 18.02 K 18.03 K 19.02

ZD_NON ZD_LON ZS_MAX

-.YES ME2.DE1 ME2.DE4

K 19.03

ZS_MIN

ME2.DE5

K 19.04 K 19.05 K 20.01 K 20.02 K 20.03

ZS_LAUF ZS_LZU M0_BATQ M1_BATQ TM_BATQ

ME2.DE1 ME2.DE2 ZBS LB1 LB1 ZBS LB1 LB1 -.NO -.NO -.YES LB1 LB1 LB1 LB1 -.NO W2

K 20.04 K 20.05 K 20.06 K 21.02 K 21.03 K 23.01 K 23.02 K 23.03 K 23.04 K 23.05 K 23.06

M0_BATA M1_BATA TM_BATA B_RON B_INIT KME_FR KME-SBI KME_STI KME_KUI KME_KNO KME_FEA

o o o o o o o o o o o o o o o o o o o o o o o
List of Interfaces FH 345 GB/ B - 3

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

WEIGHFEEDER
NAME Moisture Correction START Dead Time Element Setpoint Preselect Mode 0 Dead Time Element Setpoint Preselect Mode 1 Dead Time Element Setpoint Clock ON M0 Dead Time Element Setpoint Clock ON M1 Preassign Dead Time Element Setpoint Initialize Dead Time Element Setpoint Dead Time Element START/STOP Preselect M0 Dead Time Element START/STOP Preselect M1 Dead Time Element START/STOP Clock ON M0 Dead Time Element START/STOP Clock ON M1 START Prefeeder START Local Prefeeder NUMBER K 25.01 K 27.01 CODE FEU_EIN TOT_VM0 K 27.02 TOT_VM1 K 27.03 TOT_TM0 K 27.04 TOT_TM1 K 27.05 TOT_VOR -.NO -.NO -.NO -,NO DEFAULT VALUE -.NO

MULTICONT
VALUE SWITCH

o o o o o oq oq o o o o o oq

-.NO

K 27.06

TOT_INI

-.NO

K 27.07

TOT_IM0 -.NO

K 27.08

TOT_IM1 -.NO

K 27.09

TOT_TI0 -.NO

K 27.10

TOT_TI1 -.NO -.YES ME1.DE5

K 28.02 K 28.03

ZP_NON ZP_LON

Other source types are possible as well.

9947

List of Interfaces FH 345 GB/ B - 4

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT

WEIGHFEEDER

B.2 Digital Outputs DA

NAME Drives OK Cable Breakage/Short Circuit Feed Unit OK Wrong Material Flow Q < PMIN After Start-up No Release Electrical Equipment Ready Externally Ready Ready In Mode 0 Ready In Mode 1 Ready to Start in Mode 0 Ready to Start in Mode 1 No EMERGENCY-OFF Preselect Local Mode Active Preselect Test Mode Active Preselect Mode 0 Active Preselect Mode 1 Active Belt Drive START Message: Feeder Started General Alarm
9947

NUMBER K 02.00 K 02.04

CODE ANTR_OK F_KABEL

DEFAULT VALUE A

VALUE

SWITCH

K 03.00 K 04.00 K 04.02 K 05.06 K 05.07

MECH_OK ST_MAT QMIN_AN NO_FREI EL_BER

W1 W2

K 05.08 K 05.09 K 05.10 K 05.11 K 05.12 K 05.13 K 06.04

EX_BER M0_BER M1_BER M0_EBER M1_EBER NO_NOT LO_AKT

EA1.DA5 -

K 07.05

TM_AKT

K 08.05 K 09.05 K 10.00 K 10.03 K 10.04

M0_AKT M1_AKT ANTR_ON W_EIN S_ALARM

EA1.DA1 -

o o o o o o o o o o o o o o o o o o o o
List of Interfaces FH 345 GB/ B - 5

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

WEIGHFEEDER
NAME General Warning Message: Auxiliary Drive started Message: Deviation Activate Emergency Setpoint Controller Out Of Limit Actual Feed Rate < PMIN Actual Feed Rate > PMAX Belt Load < qMIN Belt Load > qMAX Belt Speed < vMIN Belt Speed > vMAX BIC Asynchronous Apply Check Weight Volumetric Prefeeder Drive START Prefeeder Drive ON Prefeeder Drive OPEN Prefeeder Drive CLOSE Pre-contact Fill Weight < BMIN Fill Weight > BMAX Bin Filling: START Message: Check Measurement Started Corr. Range Exceeded, Abs. Zero Point NUMBER K 10.05 K 10.06 CODE S_WARN H_EIN DEFAULT VALUE EA1.DA2 EA1.DA7

MULTICONT
VALUE SWITCH

K 11.02 K 11.03

DA_REGA NOTSOL

EA2.DA2 -

W1 W1

K 11.04 K 12.01 K 12.02 K 12.03 K 12.04 K 12.05 K 12.06 K 14.04 K 16.07 K 16.08 K 18.00 K 19.00 K 19.06 K 19.07 K 20.07 K 21.05 K 21.06 K 21.07 K 23.07

REG_ERR K_PMIN K_PMAX Q_MIN Q_MAX V_MIN V_MAX BIC_ERR PRG_ERR VOL_AKT ZD_AON ZS_AON ZS_AAUF ZS_AZU VORKON FVB_MIN FVB_MAX FILORVB KME_AKT

EA1.DA3 EA1.DA4 EA2.DA6 EA2.DA7 EA2.DA5 EA2.DA1 EA1.DA6 EA1.DA3 EA1.DA4 -

W1 W2 W2 W2 W2 W2 W2 W1 W1

W2 W2

K 23.11

KME_KTA

W1

p o o o o o o o o o o o o o o o o o o o o o o
Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

9947

List of Interfaces FH 345 GB/ B - 6

Technische Redaktion BVED

MULTICONT
NAME Corr. Range Exceeded, Rel. Zero Point Corr. Range Exceeded, Abs. Span Corr. Range Exceeded, Rel. Span Check Logged On Prepare Check Measurement Check Measurement Active Check System Expects Acceptance Humidity Limit Value FEUMAX Moisture Correction Active Preselect Dead Time Elements Dead Time Element Setpoint Active Dead Time Element START/STOP Active START Prefeeder Drive NUMBER K 23.12 CODE KME_KTR

WEIGHFEEDER
DEFAULT VALUE W1 VALUE SWITCH

K 23.13

KME_KBA

W1

K 23.14

KME_KBR

W1

K 23.15 K 23.16

KME_KAN KME_KVO

W2 W2

K 23.17

KME_KON

K 23.18

KME_WQU

W2

K 25.06

FEU_MAX

W1

K 25.07 K 27.11

FEU_AKT TOT_GEW

K 27.12

TOT_PON

K 27.13

TOT_ON

K 28.00

ZP_AON

EA1.DA6

o o o o o o o o o o o o o

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

List of Interfaces FH 345 GB/ B - 7

WEIGHFEEDER

MULTICONT

B.3 Analog and Measurement Inputs AE, GA, WZ


NAME Drive Signal Speed Absolute Setpoint Absolute Setpoint Absolute Setpoint Belt Load Belt Speed Drive Signal Prefeeder, Local Prefeeder Position Signal Fill Weight Material Moisture Second Weighbridge NUMBER K 06.10 K 07.10 K 08.10 K 09.10 K 10.10 K 10.11 K 18.10 K 19.15 K 21.01 K 25.02 K 26.01 CODE LO_STEL TM_PABS M0_PABS M1_PABS QU_q QU_v ZD_LST ZS_STM QU_FGVB MATFEU WB2_WZ DEFAULT VALUE ME1.AE1 LB1 ZBS LB1 ME1.WZ1 ME1.GA1 ME2.AE1 ME2.AE1 A A A VALUE

B.4 Analog and Pulse Outputs AA, IM


NAME Totalizing Counter 1 Totalizing Counter 2 Belt Drive Control Magnitude Actual Feed Rate 1 Actual Feed Rate 2 Belt Load Deviation Belt Speed, Analog Output Setpoint NUMBER K 10.08 K 10.09 K 10.12 K 10.13 K 10.14 K 10.15 K 11.10 K 16.10 K 16.11 CODE FM1_IA FM2_IA STG_DO P_IST1 P_IST2 Q_BAND AA_REGA V_BAND P_SOLL DEFAULT VALUE EA1.IM1 EA2.IM1 EA1.AA1 EA1.AA2 9947

VALUE

List of Interfaces FH 345 GB/ B - 8

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
NAME Control Magnitude Prefeeder Prefeeder Position Signal Fill Weight Fill Weight Control Magnitude Actual Feed Rate, Dry NUMBER K 18.11 K 19.16 CODE ZD_NST ZS_ADO

WEIGHFEEDER
DEFAULT VALUE EA1.AA1 VALUE

K 21.10 K 21.11 K 25.03

FGVB B_STG PTROCK

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

List of Interfaces FH 345 GB/ B - 9

MULTICONT
Annex : C
From Version: New Configurers:
K 00.12 K 02.06 K 05.13 K 10.06 NON_SI DA_TEMP NO_NOT H_EIN Conversion of units SI/US Feed drive temperature Output: No Emergency Stop Auxiliary Drive started

WEIGHFEEDER

FDB 1001-01(02)

Function Distributor:
Fading out of control functions selected out in current mode of operation.

From Version: Configurers With Modified Function:

FDB 1001-04

Configurers with Changed Function:


K 22.02 Start range correction with pos. flank K 22.03 Start zero point correction with pos. flank K 02.01 K 02.04 NETZAUS Default value from W1 to A KURZSCH Cable breakage/short circuit. Connection of contact switches to Namur inputs possible.

Operation:
After operation of HELP key, the 10 most significant event messages are displayed in plain language, previously only the message with highest priority.

Parameter with Changed Function:


P 20.15 KME_MIT KME_MIT replaced by KME_BEW. Correction filter for check system replaced by evaluation factor.

From Version:

FDB 1001-03

Configurers With Modified Function:


K 06.01 LO_VORW Automatic feeder cut-off, if Local Mode is aborted. KMW: In the measuring and filling phase of a check measurement (KME), bin level is controlled by the check system.

New Parameters:
P 04.08 P 04.09 P 11.06 roffset r_gu zi_il Feed rate controller control magnitude elevation Lower limit Intermittent prefeeder pulse time

From Version:

FDB 1001-05

While a setting program is running, analog outputs are not to 0 but to the preset elevation value.

Parameters With Modified Function: From Version:


P 06.05 zd_anh Stop of belt load controller at 0V or zd_min (elevation). Control magnitude at STOP, either 0V or zd_min.

FDB 1001-06

New configurer:
K 16.08 VOL_AKT Report of volumetric mode.

9947

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Annex C FH 345 GB/ C - 1

WEIGHFEEDER
From Version: New Configurers:
K 01.03 K 16.12 K 26.04 K 26.05 STANDBY Event Message SC7 (Stand-By) DE_TARA Taring command via contact WB2QMIN Event message LO 06 (2nd weighbridge) WB2QMAX Event message HI 06 (2nd weighbridge)

MULTICONT
From Version: 1201-01 New Configurers:
K 03.06 K 03.07 K 26.06 K 26.07 K 28.00 K 28.01 SCHLU_1 SCHLU_2 WB2_CTL WB2_DIF ZP_AON ZO_AKT ZP_NON ZP_LON ZP_LST Pulley Slip Monitoring Source Pulley Slip Monitoring Source Check Weighbridge Check Weighbridge (Message HI 07) Configuration Pulse Period Controlled Prefeeder Pulse Period Controlled Prefeeder Pulse Period Controlled Prefeeder Pulse Period Controlled Prefeeder

FDB 1101-00

Configurers with Changed Functions:


K 28.02 K 16.01 K 26.01 PRG_AUT Check Weight Automatics Source WB2_WZ 2nd weighbridge measuring input extended by analog input Omitted (Message CA 11) K 28.03 K 28.04

K 20.09

New Parameters:
In addition, all configurers for digital inputs/outputs are now capable of inverting switching levels. P 02.11 P 02.12 P 02.13 P 02.14 P 02.15 P 02.16 P 03.08 P 06.08 P 22.12 P 22.13 P 27.01 P 27.02 s_lng s_br s_offs spczeit spcfilter xdfilter dt_kenn TRSCHLU diff_max dif_tau zp_lanh zp_ldber New entry for belt sensor New entry for belt sensor New entry for belt sensor SPC values SPC values SPC values Diagnostic tool Pulley slip monitoring Check weighbridge Check weighbridge Prefeeder calibration local mode Prefeeder calibration local mode

New Parameters:
P 04.10 P 04.11 P 15.04 P 22.08 P 22.09 P 22.10 P 22.11 tblksch standby hkadap Feeder Blocked Time Stand-by Adaptation factor for Batch Mode WB2_ANH 2nd weighbridge analog input WB2_BER 2nd weighbridge analog input wb2qmin 2nd weighbridge limit value wb2qmax 2nd weighbridge limit value

New or Changed Functions:


1. Installation procedure with data loss after powerup 2. Volumetric Mode display in turned-of state of feeder 3. Load Cell Utilization service display corrected 4. Message CA 11 omitted (Pre-act contact > main contact)

9947

Anhang C FH 345 GB/ C - 2

Technische Redaktion BVED

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

MULTICONT
Changed Parameters:
P 02.03 P 03.08 P 03.09 P 23.04 P 17.03 P 02.11 P 10.03 P 10.05 P 11.05 P 11.06 P 12.05 WZ_kenn Extended to 0.2 mV/V Omitted (analog tare) Omitted (analog tare) Omitted (analog tare) Omitted (analog tare) See New Parameters Selection 0 omitted (no Start-Up Mode) Shifted to P 04.11 Shifted to P 26.01 Shifted to P 27.01 Extended by possibility No storage of I component after alarm

WEIGHFEEDER

New configurers:
K23.19 Analog output of check weighing (KME) result.

New function
1. K23.19 2. Deviation output available for min. 0.5s. 3. BIC turned off during calibrating program run.

New or Changed Functions:


1. Software is suitable for diagnostic tool.

From Version: New Parameter:

FDB 1201-08

2. Below minimum check amount, check measurement is aborted. 3. Tracking and slip messages remain stored in turned-off state of feeder. 4. Result of check weight setting program is no longer acquired automatically. 5. New service values (SPC) 6. Parametrization via MULTIKEY no longer possible.

P 04.012 Xd_Fak

Deviation Threshold

From version: New configurers:


K 26.08 WB2_AA K 26.09 WB2_AA

FDB 1201-09

From Version: 1201-02..04 Changed parameter


Changes in SE-bus protocol

2nd Weighbridge Analog Output 1st weighbridge control can be cut off.

New Parameters:
P 22.14 WB2_ANH P 22.15 WB2_BER 2nd Weighbridge Analog Output 2nd Weighbridge Analog Output

From Version: 1201-05 Changed parameter


P01.06 P01.08 P24.03 P24.04
9947

New Functions:
Max. pulse length for totalizing counters Increased to 500 ms Default changed to 0 Default changed to 0 MIN/MAX contact outputs2nd weighbridge (K 26.04, K 26.05)

Transmission to third parties and reproduction of this documentation are not permitted. SCHENCK PROCESS GmbH reserves all rights of ownership and copyrights.

Technische Redaktion BVED

Annex C FH 345 GB/ C - 3