You are on page 1of 19

Test Report issued under the responsibility of:

TEST REPORT IEC 62471

Photobiological safety of lamps and lamp systems


Report Reference No. .................... : Date of issue .................................... : Total number of pages ................... : CB Testing Laboratory .................. : Address ............................................ : Applicants name ........................... : Address ............................................ : Test specification: Standard .......................................... : Test procedure ................................. : Non-standard test method..: Test Report Form No. .................... : TRF Originator ................................. : Master TRF ...................................... : IEC 62471:2006 (First Edition) Type test N/A IEC62471A VDE Testing and Certification Institute Dated 2009-05 3127206.50A 2012-06-26 19 DEKRA Testing and Certification China Ltd. 10/F, #250, Jiangchangsan Road building 16 Headquater Economy Park Shibei Hi-Tech Park, Zhabei District, Shanghai, P.R.C 200436 Sheenly Lighting Co.,Ltd. No.88 Zhuyuan Road, Minhang District, Shanghai 201112, China.

Copyright 2009 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE), Geneva, Switzerland. All rights reserved.
This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo and the reference to the CB Scheme procedure shall be removed.

This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory and appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02. Test item description ..................... : Trade Mark....................................... : Manufacturer .................................... : Model/Type reference : Ratings ............................................. : Panel light Sheenly Sheenly Lighting Co.,Ltd. No.88 Zhuyuan Road, Minhang District, Shanghai 201112, China. Please refer to appendix 3 100 - 240 V~; 50/60 Hz For details please refer to appendix 3

Page 2 of 19 Testing procedure and testing location: CB Testing Laboratory:

Report No.:3127206.50A

DEKRA Testing and Certification China Ltd.

Testing location/ address .......................... : 10/F, #250, Jiangchangsan Road building 16 Headquater Economy Park Shibei Hi-Tech Park, Zhabei District, Shanghai, P.R.C 200436 Associated CB Laboratory: Testing location/ address ..........................: Tested by (name + signature) ........................................................ : Abby Yang Approved by (+ signature) ........................................................ : Hanson Zhang Testing procedure: TMP Tested by (name + signature) ......... : Approved by (+ signature) ............... : Testing location/ address ..........................: Testing procedure: WMT Tested by (name + signature) ......... : Witnessed by (+ signature) ............. : Approved by (+ signature) ............... : Testing location/ address ..........................: Testing procedure: SMT Tested by (name + signature) ......... : Approved by (+ signature) ............... : Supervised by (+ signature) ............ : Testing location/ address ..........................: Testing procedure: RMT Tested by (name + signature) ......... : Approved by (+ signature) ............... : Supervised by (+ signature) ............ : Testing location/ address ..........................:

TRF No. IEC62471A

Page 3 of 19 Summary of testing: Tests performed (name of test and test clause): These tests Fulfil the requirements of standard ISO/IEC 17025. When determining the test conclusion, the Measurement Uncertainty of test has been considered. The tested samples of panel light list as below Please refer to appendix 3 Have been tested according to the IEC 62471(first edition, 2006-07) and been classified as Exempt group. Testing location:

Report No.:3127206.50A

DEKRA Testing and Certification China Ltd.10/F, #250, Jiangchangsan Road building 16 Headquater Economy Park Shibei Hi-Tech Park, Zhabei District, Shanghai, P.R.C 200436

Summary of compliance with National Differences: PASS

Copy of marking plate: N/A

TRF No. IEC62471A

Page 4 of 19 Test item particulars ..................................................: Tested lamp ................................................................ : Lamp classification group .............................................: continuous wave lamps exempt risk 1

Report No.:3127206.50A

pulsed lamps risk 2 risk 3

Tested lamp system .....................................................: N/A Lamp cap .................................................................... : N/A Bulb ............................................................................. : LEDs Rated of the lamp .........................................................: 100 - 240 V~; 50/60 Hz For details please refer to appendix 3 Furthermore marking on the lamp................................: N/A Seasoning of lamps according IEC standard ..............: N/A Used measurement instrument ....................................: spectroradiometer Temperature by measurement .....................................: 23-28 C Information for safety use .............................................: -Possible test case verdicts: test case does not apply to the test object ............ : N/A test object does meet the requirement .................. : P (Pass) test object does not meet the requirement ............ : F (Fail)

Testing: Date of receipt of test item ........................................... : 2012-06-20 Date (s) of performance of tests................................... : 2012-06-25 General remarks: The test results presented in this report relate only to the object tested. This report shall not be reproduced, except in full, without the written approval of the Issuing testing laboratory. "(See Enclosure #)" refers to additional information appended to the report. "(See appended table)" refers to a table appended to the report. Throughout this report a comma (point) is used as the decimal separator. List of test equipment must be kept on file and available for review.

The product complied with the following standards: IEC 62471:2006 IEC/TR 62471-2:2009 EN 62471:2008
This report should be read in conjunction with the attached pages concerned with the European group differences and national differences of the standards EN 62471:2008 with the reference number of 3127206.50B. (2 pages)

Factory: Sheenly Lighting Co.,Ltd. No.88 Zhuyuan Road, Minhang District, Shanghai 201112, China.

TRF No. IEC62471A

Page 5 of 19

Report No.:3127206.50A

General product information: The products are LED panel lights. They have different CCT; rated power and demensions. For details please refer to appendix 3. Finally, SL-PAN-40W-PW-AN was selected to perform the test. The product considered as GLS which should be evaluated at the distance of 500lx but not less than 200mm. The sample of SL-PAN-40W-PW-AN was tested at 1650 mm from the light source. CCT of the spectral irradiance was found at 5776 K.

Type test was performed according to IEC 62471:2006 procedure.

TRF No. IEC62471A

Page 6 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

4 4.1

EXPOSURE LIMITS General The exposure limits in this standard is not less than 0,01 ms and not more than any 8-hour period and should be used as guides in the control of exposure Detailed spectral data of a light source are generally required only if the luminance of the source exceeds see clause 4.3 104 cd.m-2

P P P

4.3 4.3.1

Hazard exposure limits Actinic UV hazard exposure limit for the skin and eye The exposure limit for effective radiant exposure is 30 J.m-2 within any 8-hour period To protect against injury of the eye or skin from ultraviolet radiation exposure produced by a broadband source, the effective integrated spectral irradiance , ES, of the light source shall not exceed the levels defined by:

P P P P

The permissible time for exposure to ultraviolet radiation incident upon the unprotected eye or skin shall be computed by:

P 4.3.2 Near-UV hazard exposure limit for eye For the spectral region 315 nm to 400 nm (UV-A) the total radiant exposure to the eye shall not exceed 10000 J.m-2 for exposure times less than 1000 s. For exposure times greater than 1000 s (approximately 16 minutes) the UV-A irradiance for the unprotected eye, EUVA, shall not exceed 10 W.m-2. The permissible time for exposure to ultraviolet radiation incident upon the unprotected eye for time less than 1000 s, shall be computed by: P P

P 4.3.3 Retinal blue light hazard exposure limit To protect against retinal photochemical injury from chronic blue-light exposure, the integrated spectral radiance of the light source weighted against the blue-light hazard function, B(), i.e., the blue-light weighted radiance , LB, shall not exceed the levels defined by: for t 104 s P P

TRF No. IEC62471A

Page 7 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

for t > 104 s 4.3.4 Retinal blue light hazard exposure limit - small source Thus the spectral irradiance at the eye E, weighted against the blue-light hazard function B() shall not see table 4.2 exceed the levels defined by: for t 100 s for t > 100 s 4.3.5 Retinal thermal hazard exposure limit To protect against retinal thermal injury, the integrated spectral radiance of the light source, L, weighted by the burn hazard weighting function R() (from Figure 4.2 and Table 4.2), i.e., the burn hazard weighted radiance, shall not exceed the levels defined by: (10 s t 10 s) 4.3.6 Retinal thermal hazard exposure limit weak visual stimulus For an infrared heat lamp or any near-infrared source where a weak visual stimulus is inadequate to activate the aversion response, the near infrared (780 nm to 1400 nm) radiance, LIR, as viewed by the eye for exposure times greater than 10 s shall be limited to: t > 10 s 4.3.7 Infrared radiation hazard exposure limits for the eye The avoid thermal injury of the cornea and possible delayed effects upon the lens of the eye (cataractogenesis), ocular exposure to infrared radiation, EIR, over the wavelength range 780 nm to 3000 nm, for times less than 1000 s, shall not exceed: t 1000 s For times greater than 1000 s the limit becomes: t > 1000 s 4.3.8 Thermal hazard exposure limit for the skin Visible and infrared radiant exposure (380 nm to 3000 nm) of the skin shall be limited to: TRF No. IEC62471A

P N/A N/A

N/A N/A P P

N/A N/A

N/A

P P

P P

P P

Page 8 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

5 5.1

MEASUREMENT OF LAMPS AND LAMP SYSTEMS Measurement conditions Measurement conditions shall be reported as part of the evaluation against the exposure limits and the assignment of risk classification.

P P P

5.1.1

Lamp ageing (seasoning) Seasoning of lamps shall be done as stated in the appropriate IEC lamp standard.

N/A N/A P P

5.1.2

Test environment For specific test conditions, see the appropriate IEC lamp standard or in absence of such standards, the appropriate national standards or manufacturers recommendations.

5.1.3

Extraneous radiation Careful checks should be made to ensure that extraneous sources of radiation and reflections do not add significantly to the measurement results.

P P

5.1.4

Lamp operation Operation of the test lamp shall be provided in accordance with: the appropriate IEC lamp standard, or the manufacturer s recommendation

P N/A N/A P N/A N/A N/A N/A P P P P P P P P P

5.1.5

Lamp system operation The power source for operation of the test lamp shall be provided in accordance with: the appropriate IEC standard, or the manufacturer s recommendation

5.2 5.2.1

Measurement procedure Irradiance measurements Minimum aperture diameter 7mm. Maximum aperture diameter 50 mm. The measurement shall be made in that position of the beam giving the maximum reading. The measurement instrument is adequate calibrated.

5.2.2 5.2.2.1

Radiance measurements Standard method The measurements made with an optical system.

TRF No. IEC62471A

Page 9 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

The instrument shall be calibrated to read in absolute radiant power per unit receiving area and per unit solid angle to acceptance averaged over the field of view of the instrument. 5.2.2.2 Alternative method Alternatively to an imaging radiance set-up, an irradiance measurement set-up with a circular field stop placed at the source can be used to perform radiance measurements. 5.2.3 Measurement of source size The determination of , the angle subtended by a source, requires the determination of the 50% emission points of the source. 5.2.4 Pulse width measurement for pulsed sources The determination of t, the nominal pulse duration of a source, requires the determination of the time during which the emission is > 50% of its peak value. 5.3 5.3.1 Analysis methods Weighting curve interpolations To standardize interpolated values, use linear interpolation on the log of given values to obtain intermediate points at the wavelength intervals desired. 5.3.2 Calculations The calculation of source hazard values shall be performed by weighting the spectral scan by the appropriate function and calculating the total weighted energy. 5.3.3 Measurement uncertainty The quality of all measurement results must be quantified by an analysis of the uncertainty. see Annex C in the norm see table 4.1

P P

P P

N/A N/A

P P P

P P

P P

LAMP CLASSIFICATION For the purposes of this standard it was decided that see table 6.1 the values shall be reported as follows: for lamps intended for general lighting service, the hazard values shall be reported as either irradiance or radiance values at a distance which produces an illuminance of 500 lux, but not at a distance less than 200 mm for all other light sources, including pulsed lamp sources, the hazard values shall be reported at a distance of 200 mm

P P P

N/A

6.1 6.1.1

Continuous wave lamps Exempt Group

P P

TRF No. IEC62471A

Page 10 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

In the exempt group are lamps, which does not pose any photobiological hazard. The requirement is met by any lamp that does not pose: 6.1.2 an actinic ultraviolet hazard (ES) within 8-hours exposure (30000 s), nor a near-UV hazard (EUVA) within 1000 s, (about 16 min), nor a retinal blue-light hazard (LB) within 10000 s (about 2,8 h), nor a retinal thermal hazard (LR) within 10 s, nor an infrared radiation hazard for the eye (EIR) within 1000 s

P P P P P N/A N/A N/A N/A N/A N/A N/A N/A

Risk Group 1 (Low-Risk) In this group are lamps, which exceeds the limits for the except group but that does not pose: an actinic ultraviolet hazard (ES) within 10000 s, nor a near ultraviolet hazard (EUVA) within 300 s, nor a retinal blue-light hazard (LB) within 100 s, nor a retinal thermal hazard (LR) within 10 s, nor an infrared radiation hazard for the eye (EIR) within 100 s

Lamps that emit infrared radiation without a strong visual stimulus and do not pose a near-infrared retinal hazard (LIR), within 100 s are in Risk Group 1. 6.1.3 Risk Group 2 (Moderate-Risk) This requirement is met by any lamp that exceeds the limits for Risk Group 1, but that does not pose: an actinic ultraviolet hazard (ES) within 1000 s exposure, nor a near ultraviolet hazard (EUVA) within 100 s, nor a retinal blue-light hazard (LB) within 0,25 s (aversion response), nor a retinal thermal hazard (LR) within 0,25 s (aversion response), nor an infrared radiation hazard for the eye (EIR) within 10 s

N/A N/A N/A N/A N/A N/A N/A N/A

Lamps that emit infrared radiation without a strong visual stimulus and do not pose a near-infrared retinal hazard (LIR), within 10 s are in Risk Group 2. 6.1.4 Risk Group 3 (High-Risk) Lamps which exceed the limits for Risk Group 2 are in Group 3.

N/A N/A

TRF No. IEC62471A

Page 11 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

6.2

Pulsed lamps Pulse lamp criteria shall apply to a single pulse and to any group of pulses within 0,25 s. A pulsed lamp shall be evaluated at the highest nominal energy loading as specified by the manufacturer. The risk group determination of the lamp being tested shall be made as follows: a lamp that exceeds the exposure limit shall be classified as belonging to Risk Group 3 (High-Risk) for single pulsed lamps, a lamp whose weighted radiant exposure or weighted radiance does is below the EL shall be classified as belonging to the Exempt Group for repetitively pulsed lamps, a lamp whose weighted radiant exposure or weighted radiance dose is below the EL, shall be evaluated using the continuous wave risk criteria discussed in clause 6.1, using time averaged values of the pulsed emission

N/A N/A N/A

N/A N/A

N/A

N/A

TRF No. IEC62471A

Page 12 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

Table 4.1

Spectral weighting function for assessing ultraviolet hazards for skin and eye UV hazard function SUV() 0,030 0,051 0,075 0,095 0,120 0,150 0,190 0,240 0,300 0,360 0,430 0,500 0,520 0,650 0,810 1,000 0,960 0,880 0,770 0,640 0,540 0,460 0,300 0,120 0,060 0,026 0,015 Wavelength , nm 313* 315 316 317 318 319 320 322 323 325 328 330 333* 335 340 345 350 355 360 365* 370 375 380 385 390 395 400 UV hazard function SUV() 0,006 0,003 0,0024 0,0020 0,0016 0,0012 0,0010 0,00067 0,00054 0,00050 0,00044 0,00041 0,00037 0,00034 0,00028 0,00024 0,00020 0,00016 0,00013 0,00011 0,000093 0,000077 0,000064 0,000053 0,000044 0,000036 0,000030

Wavelength1 , nm 200 205 210 215 220 225 230 235 240 245 250 254* 255 260 265 270 275 280* 285 290 295 297* 300 303* 305 308 310
1

Wavelengths chosen are representative: other values should be obtained by logarithmic interpolation at intermediate wavelengths. * Emission lines of a mercury discharge spectrum.

TRF No. IEC62471A

Page 13 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

Table 4.2

Spectral weighting functions for assessing retinal hazards from broadband optical sources Wavelength nm 300 305 310 315 320 325 330 335 340 345 350 355 360 365 370 375 380 385 390 395 400 405 410 415 420 425 430 435 440 445 450 455 460 465 470 475 480 485 490 495 500-600 600-700 700-1050 1050-1150 1150-1200 1200-1400 Blue-light hazard function B () 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,01 0,013 0,025 0,05 0,10 0,20 0,40 0,80 0,90 0,95 0,98 1,00 1,00 0,97 0,94 0,90 0,80 0,70 0,62 0,55 0,45 0,40 0,22 0,16 10[(450-)/50] 0,001 Burn hazard function R ()

10[(700-)/500] 0,2 . 0,2 100,02(1150-) 0,02

0,1 0,13 0,25 0,5 1,0 2,0 4,0 8,0 9,0 9,5 9,8 10,0 10,0 9,7 9,4 9,0 8,0 7,0 6,2 5,5 4,5 4,0 2,2 1,6 1,0 1,0

TRF No. IEC62471A

Page 14 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.:3127206.50A

Verdict

Table 5.4 Hazard Name Actinic UV skin & eye Eye UV-A Blue-light small source Eye IR Skin thermal

Summary of the ELs for the surface of the skin or cornea (irradiance based values) Relevant equation ES = E S() EUVA = E EB = E B() Wavelength range nm 200 400 315 400 300 700 780 3000 380 3000 Exposure duration sec < 30000 1000 >1000 100 >100 1000 >1000 < 10 Limiting aperture rad (deg) 1,4 (80) 1,4 (80) < 0,011 1,4 (80) 2 sr EL in terms of constant irradiance Wm-2 30/t 10000/t 10 100/t 1,0 18000/t 0,75 100 20000/t 0,75

EIR = E EH = E

Table 5.5

Summary of the ELs for the retina (radiance based values) Relevant equation Wavelength range nm Exposure duration sec 0,25 10 10-100 100-10000 10000 < 0,25 0,25 10 > 10 Field of view radians 0,011(t/10) 0,011 0,0011t 0,1 0,0017 0,011(t/10) 0,011 EL in terms of constant radiance Wm-2sr-1) 106/t 106/t 106/t 100 50000/(t 0,25) 50000/(t 0,25) 6000/

Hazard Name

Blue light

LB = L B()

300 700

Retinal thermal Retinal thermal (weak visual stimulus)

LR = L R()

380 1400

LIR = L R()

780 1400

TRF No. IEC62471A

Page 15 of 19 IEC 62471 Clause Requirement + Test Result Remark

Report No.3127206.50A

Verdict

Table 6.1

Emission limits for risk groups of continuous wave lamps (SL-PAN-40W-PW-AN, =100 mrad ) Emission Measurement Action spectrum SUV()

Risk

Symbol

Units Limit

Exempt Result 0,0000 0,0000 50,55 -68,70

Low risk Limit 0,003 33 10000 1,0 28000/ Result

Mod risk Limit 0,03 100 4000000 400 71000/ Result

Actinic UV Near UV Blue light Blue light, small source Retinal thermal Retinal thermal, weak visual stimulus** IR radiation, eye * **

Es EUVA

Wm-2 Wm-2 Wm-2sr-1 Wm-2 Wm-2sr-1

0,001 10 100 1,0* 28000/

B() B() R()

LB EB LR

R()

LIR

Wm-2sr-1

6000/

--

6000/

6000/

EIR

Wm-2

100

0,01

570

3200

Small source defined as one with < 0,011 radian. Averaging field of view at 10000 s is 0,1 radian. Involves evaluation of non-GLS source

TRF No. IEC62471A

Page 16 of 19

Report No.:3127206.50A

Furthermore remarks: Appenix 1: List of test equipment used: Clause 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 Measurement/ testing Irradiance measurements Radiance measurements Radiance measurements Irradiance measurements Irradiance measurements Radiance measurements Radiance measurements Irradiance measurements Irradiance measurements Irradiance measurements Irradiance measurements Radiance measurements Irradiance measurements Radiance measurements Irradiance measurements Radiance measurements Irradiance measurements Irradiance measurements Irradiance measurements Radiance measurements Irradiance measurements Irradiance measurements Irradiance measurements Radiance measurements Registration Number SH 344 SH 345 SH 346 SH 347 SH 348 SH 349 SH 350 SH 351 SH 352 SH 353 SH 354 SH 355 SH 356 SH 357 SH 358 SH 359 SH070 Testing/measuring equipment/material used MONOCHROMATOR S009 TELESCOPE S400_417 DETECTION ELECTRONIC 608 CONSTANT CURRENT SRS12 RADIANCE 705 DEUTERIUM SUPPLY CL6 STANDARD CL7 STANDARD PHOTOMULTIPLIER INGAAS DETECTOR SILICON DETECTOR PBS-TE DETECTOR RELAY OPTIC D8 INTEGRATING SPHER D7 COSINE DIFFUSER PHOTOMETRIC DETECTOR WATTMETER

TRF No. IEC62471A

Page 17 of 19

Report No.:3127206.50A

Appendix 2: Photo documentation

Overview of SL-PAN-40W-PW-AN

TRF No. IEC62471A

Page 18 of 19

Report No.:3127206.50A

Appendix 3: Model list Lum en Out put (lm) SL-PAN-06W-WW-SN SL-PAN-06W-PW-SN SL-PAN-12W-WW-SN SL-PAN-12W-PW-SN SL-PAN-12W-WW-SS SL-PAN-12W-PW-SS SL-PAN-12W-WW-AN SL-PAN-12W-PW-AN SL-PAN-20W-WW-AN SL-PAN-20W-PW-AN SL-PAN-20W-BW-AN SL-PAN-26W-WW-BN SL-PAN-26W-PW-BN SL-PAN-26W-WW-AN SL-PAN-26W-PW-AN SL-PAN-40W-WW-AN SL-PAN-40W-PW-AN SL-PAN-50W-WW-AN SL-PAN-50W-PW-AN SL-PAN-50W-BW-AN SL-PAN-20W-WW-SN SL-PAN-20W-PW-SN SL-PAN-45W-WW-SN SL-PAN-45W-PW-SN WW PW WW PW WW PW WW PW WW PW WW/ PW WW PW WW PW WW PW WW PW WW/ PW WW PW WW PW 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 280 360 560 730 560 12W 730 560 730 1120 1460 1300 1900 26W 2100 1900 2100 2900 3100 3800 4200 4000 1400 20W 1600 2800 3200 45W 100240 192 120 150*1200*12 376 376 50W 40W 100240 20W 12W 100240 120 300*300*12

Picture

Part Number

CCT

LED

Powe r (W)

Input Volta ge (V)

Bea m

Dimension

LED Q'ty

(mm) 48

6W 100240 12W 120 150*150*12

48 96 96

100240

120

170*170*12 CutOut: 150*150

96 96 96 96 192 192 192 240

100240

120

300*600*12 240 240 240 376

26W

120

600*600*12

376 480 480 480 192

TRF No. IEC62471A

Page 19 of 19

Report No.:3127206.50A

SL-PAN-45W-WW-AN SL-PAN-45W-PW-AN SL-PAN-26W-WW-LS SL-PAN-26W-PW-LS SL-PAN-50W-WW-LS SL-PAN-50W-PW-LS SL-PAN-50W-BW-LS SL-PAN-45W-WW-BN SL-PAN-45W-PW-BN SL-PAN-70W-WW-BN SL-PAN-70W-PW-BN

WW PW WW PW WW PW WW/ PW WW PW WW PW

3014 3014 3014 3014 3014 3014 3014 3014 3014 3014 3014

2800 45W 3200 1900 26W 2100 3800 4200 4000 2800 45W 3200 4700 70W 5300 100240 50W 100240 120 620*620*12 100240 120 300*1200*12

376 376 240 240 480 480 480 376 376 120 600*1200*12 576 576

TRF No. IEC62471A

You might also like