Hospital: dedicated modules

To secure power distribution and monitoring solution for operating theatres Operating rooms require extremely high availability and quality of electric power to offer patients maximum safety. For this reason, the standards lay down very strict rules to ensure the continuity of service of electrical installations. What do the standards say? > In group 2 rooms for medical use, the > For the satisfactory operation of medical

> An audible and visual alarm must be


electric power supply.

medical IT system should be used for the circuits powering medical electrical equipment and systems for survival and surgical applications, and the other equipment located in the environment of the patient. provided in the room in question to alert medical personnel.

> Monitoring of overload and high temperature > An alarm should take place if the earth
for the medical IT transformer is required. connection or the connection to the system is disconnected.

equipment, prevention of electromagnetic disturbances may be necessary.

IEC
Our solution complies with international standard IEC 60364-7-710 and national standards and regulations

> Operating activities must have continuity of

IM10-H and HRP (Hospital Remote Panel) for the "Classic" solution > Graphic display > Bargraph > Smart HMI > 8 languages > Earth and injection
connection monitoring

Simple and efficent  udible and visual alarm for an insulation or >A

> Testing of the insulation monitoring system > Audible alarm stoppage > 24 V DC power supply > Antibacterial > Tested with Anios products (disinfection products) > Complies with IEC 60601-1 (medical equipment)

electrical fault (transformer overload or circuit breaker tripping)

IM20-H as a brick of the advanced solution > Modbus Communication > Datalog with timestamping of all events > Transformer management:

- secondary load current display - alarm on threshold (in % of nominal current) - alarm on temperature through the sensor (bimetal)

Vigilohm IM10-H and IM20-H


Insulation monitoring devices for hospitals
PB106375_55

Vigilohm IM10-H and IM20-H


LV AC / DC IT systems Frequency

Type of installation to be monitored


Phase-to-phase voltage y 230 V AC +15 % 50/60 Hz 1 k to 10 M 1 (password protected) 50 k to 500 k y1s Yes As standard At 50 Hz Number Breaking Capacity Input contact 1 Type AC 250 V DC 12 to 24 V Voltage supplied Minimum load 50/60 Hz DC Cable size Maximum device consumption Measurement voltage Measurement current Dielectric strength 110 k 5% Standard or failsafe Changeover 6A 6A 24 V 5 mA 110/230 V AC 15 % 125/250 V DC 15 % 0.2 to 2.5 mm2 12 VA 25 V max 0.9 mA 4000 V AC / 5500 V DC 0.25 kg Mounting Front Panel mount or DIN rail IP52

Electrical characteristics
Range for insulation resistance readings Fault signalling Response time Device operating test Failsafe feature (1) Internal impedance Accuracy Output contact Number of thresholds Thresholds

Use

Transformer bimetal Auxiliary supply voltage

These modules are dedicated to hospital IT networks.

Operation

b Injection of a low-frequency AC voltage between the network and the earth.

Measurement Indication

b Insulation measurement through the earth leakage current in the IMD. b Satisfactory insulation resistance (green light). b Drop in insulation resistance below the fault threshold (orange light). b Earth or injection connection lost.

Mechanical characteristics
Weight Thermoplastic case Degree of protection Installation

Display (8 languages (3))

Cat. III, pollution 2, moulded case, disconnectable assembly, symetrical or embedded For operation For storage -25 C to +55 C -40 C to +70 C Product Safety Installation IEC 61557-8 IEC 61010-1 IEC 60364-7-710

b Insulation resistance. b Thresholds. b Alarms with dedicated pop-up windows.

Other characteristics
Temperature range Climatic conditions (2) Standards

(3) French, English, Spanish, Italian, Portuguese, German, Russian, Chinese.

Additional functions with IM20-H


b b b v v v

Modbus communication. Datalog with timestamping of all events. Transformer management: secondary load current display (in %) alarm on a threshold (in % of nominal current) alarm on temperature through the sensor (bimetal).

(1) Failsafe: the relay is deactivated on occurrence of a fault or if the auxiliary supply voltage accidentally fails. (2) Suitable for use in all climates: b damp heat, equipment not operating (IEC 60068-2-30) b damp heat, equipment operating (IEC 60068-2-56) b salt mist (IEC 60068-2-52). Commercial references: b IMD-IM10-H b IMD-IM20-H

Accessories

Hospital Remote Panel HRP (ref 50168) see pages 36 and 37. Or as part of the solution for operating theatres.

Vigilohm_Catalogue_Part_2.indd

version: 1.0

17

Vigilohm

HRP

Hospital Remote Panel

PB105943_58

Presentation

This accessory provides the interface for an electrical installation monitoring system in hospitals. Installed in the operating theatre, the Vigilohm HRP informs hospital personnel, in real time, of correct operation or the presence of a fault: b faulty insulation of operating theatre equipment b electrical fault following a circuit breaker trip or transformer overload.

Use
b b b b b

Insulation fault signal light (orange). Electrical fault signal light (red). Correct operation signal light (green). Push button to test installation insulation. Push button to stop alarm buzzer.

Vigilohm HRP
Mechanical characteristics
Weight Case Degree of protection Dimensions Height Width Depth Buzzer Factory setting 24 V DC Plastic 0.5 kg Vertical mounting IP54 IK08 170 mm 170 mm 20 mm 80 db 65 mA 0 C to 40 C -25 C to +70 C 90 % 2000 m IEC 60364-7-710 IEC 61557-8 IEC 60601-1 Medical locations Electrical safety Medical electrical equipment

Electrical characteristics
Auxiliary supply voltage

Environment
Operating temperature Storage temperature Maximum relative humidity Altitude

Standards
Complies with

Tested with Anios products (disinfection, antiseptis and detergency products). Commercial reference: 50168.
PB105945

Functions identification
Visual alarm on insulation fault Right operation lamp  isual alarm on electrical fault (overload or overheating transformer or circuitV breaker fault tripping)  T est button of the insulation monitoring system (daily test in compliance with IEC 60364-7-710 standard)  udible alarm stoppage, in case of insulation fault or electrical fault. The alarm A volume setting is made at the mounting phase (available in the rear)

36

version: 1.0

Vigilohm_Catalogue_Part_5.indd

HRP and IM10-H or IM20-H

CPI Cabinet 1

CPI Cabinet 2

Example architecture 1 using IM10-H and HRP


This monitoring system makes it possible, with the Vigilohm HRP in the operating room, to gather and display information concerning the alarms generated by the electrical installation.
Vigilohm HRP

DB127897

Electrical fault Insulation fault

Electrical fault Vigilohm Insulation HRP fault

IM10-H

1 wire electrical fault (CB fault tripping or transformer overload) 2 wires (NC Contact) 2 wires (N, Earth) 2 wires (+24 V 0 V)

HRP

ABL8MEM24003

Existing or specific power supply Ex: ABL8MEM24003

Example architecture 2 using IM20-H and Supervision


The Modbus communication provides the operating room, and/or the maintenance personnel: b insulation value b transformer load level b alarms b timestamping of events.
Electrical fault Insulation fault Electrical fault Insulation fault Magelis HRP

DB127898

This enables event traceability.

Sensors

Sensors Magelis HRP

IM20-H 2 wires (NC contact) 4 wires (Modbus) PLC Modbus TCP-IP

Magelis HRP

Note: It is also possible to have a variant of the architecture 1, with an IM20-H linked through Modbus to a supervisor outside the operating room.

Vigilohm_Catalogue_Part_5.indd

version: 1.0

37

Vigilohm

Installation and connection


Product dimensions

IM9, IM9-OL
Mounting on symmetrical DIN rail
DB127905

Cutout
2 3.85 =

45

90

85

46.50.3 =

660.3

44 67.5

71

610.3 = 72.50.4

IM10, IM20, IM10-H, IM20-H


Mounting on symmetrical DIN rail
DB127906

Cutout

110 93.7

45

96

92.40.4

46.3 59

96 92.40.4

XM200,XM300C, XL308, XL316


Flush mounting
DB127899 DB127938

Cutout

129

153 9.5 192 12

XD301, XD312
Flush mounting
DB127936

Cutout
XD301 2 4.2 66 XD312 2 4.2 66
DB127937

Surface mounting
XD301 2 4.2 XD312 2 4.2

65

45

46

46 10 7 93 110

6.5 37.5 49.5 73.5 XD301 XD312 5 57 73

66

81

66

81

4.5

57 72

1 6.5

93 106

7.5

38

version: 1.0

Vigilohm_Catalogue_Part_6.indd

IM10-H, IM20-H
PB106375_32 DB127957

L U 110....230 Va N

IT transformer bimetal Contact min. y 24 V c 5 mA

6A 1 2 3

6A 4 5

Shorting block

6 7 8

9 10 1112 13141516

Failsafe wiring (recommended)

Standard wiring IM20-H


6 7 8

IM20-H
6 7 8

0V D0 = A / Rx-, A / TxD1 = B / Rx+, B / Tx+

Modbus RS 485

u 12 Vc 100 mA y 250 Va 24 Vc6 A

6A

u 12 Vc 100 mA y 250 Va/ 24 Vc6 A

6A

XM200
044147_SE_32

10 9 8 7 6 5

prevention (falsafe feature) fault

XM200

MT/BT

3 2 1 N

DB127923

DB127924

4 cardew C 11

XM200

Tunnel terminals for 1.5 mm wiring


r aa
4 3 2 1 11 4 11 MT/BT 3 2 1

cardew C Legend aa: AC auxiliary supply a 15 %, +10 %, 50-60 Hz ut: application r: installation terminal 13: earth connection by round lug (4 mm diameter) to be crimped.

XM200

4 11

XM200

Vigilohm_Catalogue_Part_6.indd

version: 1.0

43

Vigilohm

Installation and connection


Product connection

XLI300, XTU300
DB127949

18 V17 B not used 16 V+ 15 S

b Never connect more than one interface to the Vigilohm System bus. b To connect devices to the communication bus (shielded, double twisted pair) it is recommended to interconnect their respective terminals S, B, V+ and V-.
RS485

1 5
RS482

XLI300 Sub D9

6 9

1 signal ground 4 received data 5 transmitted data 8 received data +

1 5
RS485

6 9

1 : GND 4 : B'(RD-) 5 : B(TD-) 8 : A'(RD+) 9 : A(TD+) 13

9 transmitted data +

2 1

aa

Example of line connection


DB127935

VXM300c B V+ S XL308 XL316

VB V+ S XD308c

VB V+ S XML308 XML316

VB V+ S XLI300 XTU300 XAS

VB V+ S

HRP
PB105943_42 DB127909

2.5 mm max.

1 2 3 4 5 6 7

34 V c 14.5 mA

Electrical fault Common

34 V c 14.5 mA +24 V c 65 mA 0V c

Insulation fault Auxiliary power supply

48

version: 1.0

Vigilohm_Catalogue_Part_6.indd

Vigilohm

Vigilohm insulation monitoring


Commercial references

Vigilohm commercial references


Product
PB106370_15

Voltage range
110/415 V AC 50/60 Hz 110/415 V AC 50/60 Hz 110/415 V AC 50/60 Hz 110/415 V AC 50/60 Hz 110/240 V AC 50/60 Hz 110/240 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz

Reference
IMD-IM9 IMD-IM9-OL IMD-IM10 IMD-IM20 IMD-IM10-H IMD-IM20-H 50727 50728 50729

IM9 IM9-OL IM10 & IM20

Insulation monitoring device IM9 IM9-OL Off-line insulation monitoring device Insulation monitoring devices IM10 IM20

PB106374_15

IM10-H & IM20-H

Insulation monitoring devices for hospitals IM10-H IM20-H

044147_SE_20

XM200

Insulation monitoring device XM200 XM200 XM200

044148_SE_20

XM300C

Insulation monitoring device XM300C XM300C XM300C 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 50540 50541 50542

044149_SE_30

XML308 & XML316

Monitoring-locating devices XML308 XML308 XML308 XML316 XML316 XML316 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 50490 50491 50492 50322 50323 50324 50506 50507 50508 50535 50536 50537 50723 50724 50725

044144_SE_20

XD301 & XD312

Automatic insulation fault detectors XD301 XD301 XD301 XD312 XD312 XD312

044143_SE_12

XD308C

Communicating automatic insulation fault detector XD308C XD308C XD308C

044145_SE_22

XL308 & XL316

Insulation fault locators XL308 XL308 XL308 XL316 XL316 XL316 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 50606 50607 50608 50615 50616 50617

Vigilohm_Catalogue_Part_7.indd

version: 1.0

49

Vigilohm commercial references


Product
051350_SE_10

Voltage range

Reference
50310 50278

XGR & XRM & probes

Mobile fault locating Kit case: XGR (50282) + XRM + XP15 + XP50 + XP100 XRM XGR 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz

50281 50282 50283 50494 50498 50499 50285

PB100471_SE_25

XGR XGR XP15 probe XP50 probe XP100 probe Empty case

044141_SE_15

XLI300 & XTU300

Communication interfaces XLI300 XLI300 XLI300 XTU300 XTU300 XTU300 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 115/127 V AC 50/60 Hz 220/240 V AC 50/60 Hz 380/415 V AC 50/60 Hz 50515 50516 50517 50545 (1) 50546 (1) 50547 (1) 50437 50438 50439 50440 50441 50442 50845 50486 50136 50170 50171 50172 50183 50169 IMD-HV-IM20-1.7 50159 50248 50168

051352_SE_25

Toroids

Earth leakage current detectors TA30 PA50 IA80 MA120 SA200 GA300 Open POA Open GOA 100 m cable (shielded)

042589_SE_20

041857_SE_10

Auxiliaries(2)

Accessories for various installations 250 V Cardew C 440 V Cardew C 660 V Cardew C 1000 V Cardew C Cardew C base

044139_SE_18

HV-IM20-1.7 subassembly ZX limiting impedance PHT1000 additional subassembly HRP

PB105943_22

(1) Please consult Schneider Electric for configuration. (2) See choice of Cardew on page 33 and auxiliary selection table on page 35.

50

version: 1.0

Vigilohm_Catalogue_Part_7.indd