You are on page 1of 14

TECHNICAL EVALUATION REPORTS AGAINST OPEN TENDER NOTICE NO.

SPC-1/2010

Item No.: CEE-1


Name of the indenter: Dr. Mamta Awasthi
S.No Name of firms 1/6 Scientifi c and Allied Industrie s, Agra Model Quoted /Not quoted Model MZ-6-T Qualified/ Not qualified Not qualified Remarks

1. Stereo Zoom Trinocular Microscope Objectives for providing continuous variable magnification from 10x to 20x depending upon the eye piece and supplementary lenses used. Wide magnification range and long working distance (86mm). Supplied with a built in Latest pole type microscope stand having both incident light and base illumination arrangement 6V-20W halogen lamp controlled through solid stage variable transformer. Coarse focusing by piller clamp and fine focusing by rock and pinion. Microphotography equipment: Digital Camera compatible with stereo zoom microscope (latest and advanced). Twin zoom objectives (zooming ration -1:5 and one pair of 10 high eye point wide-field eye pieces present a large filed of view and crisp, sharp and erect image.

Not mentioned about built in Latest pole type microscope stand having both incident light and base illumination arrangement (only adjustable height mentioned) 12v-10W instead of 6V-20W halogen lamp Not mentioned about Coarse focusing by piller clamp and fine focusing by rock and pinion. Not mentioned about microphotography equipment

2/6

Scientifi c India, Ambala

Model ASZ66TR

Not qualified

3/6

Jain Scientifi c, Baddi

Model SM series

Not qualified

Not mentioned about Twin zoom objectives (zooming ration -1:5 and one pair of 10 high eye point wide-field eye pieces) in the literature provided. Not mentioned about working distance (not mentioned/clear from the literature provided Not mentioned about built in Latest pole type microscope stand having both incident light and base illumination arrangement (only adjustable height mentioned) Working distance 80mm (required 86 mm) Zooming ratio not mentioned

4/6

Lab India chandiga rh Raj Ji Scientifi c Palampu r Radicals Ambala

Leica S6 D

Qualified

5/6

MAGN US

Qualified provisiona lly Not qualified

Subject to the production of Authorized dealership certificate from principals/Manufacturer

6/6

Model RSM-9

Not mentioned about Latest pole type microscope stand having both incident light and base illumination arrangement

S.No 2. Binocular Microscope with Photographic attachment 3 Mega pixels CMOS sensor, USB output connecting camera to computer and to provide power to the camera (LAS) EZ Software for easy camera control, image capture, annotation, measurement and documentation, variety of trinocular viewing tubes and C mount adapters 1/6

Name of firms Scientific and Allied Industries, Agra

Model Quoted/Not quoted CXR-3

Qualified/Not qualified Not as per specifications

Remarks Microprocessor based Not mentioned CMOS sensor, USB output connecting camera to computer Not mentioned EZ Software for easy camera control Subject to the production of Authorized dealership certificate from principals/Manufacturer

2/6

Scientific India, Ambala

Microscan-20

Qualified provisionally

4/6

Lab India chandigarh

Leica DM 500

Qualified As per specifications

5/6

Raj Ji Scientific Palampur Radicals Ambala

6/6

Magnus MSXi/Olympus binocular research microscope Magnus MSXi/Olympus binocular research microscope

Qualified provisionally Not qualified

Subject to the production of Authorized dealership certificate from principals/Manufacturer Not mentioned about EZ software, trinocular viewing tubes and C Mount adapters in the literature

S.No 3. Digital Imaging System Digital Imaging System Latest high resolution Digital camera with resolution of 4.5 million pixels and is supplied with accessories for computer compatibility. The high resolution digital camera, provides finest ever printouts of highly magnified microscopic images. In addition it is a multipurpose equipments for routine photography, video recording and wall projection of date through LCD projector. Technical features digital camera, Digital zoom with 4.5cm/1.8" colour TFT-LCD monitor. Y2 CCD with 4.5 million pixels. 5/6 4/6

Name of firms Lab India chandigar h

Model Quoted/No t quoted Leica EC3

Qualified/Not qualified Qualified

Remarks

Raj Ji Scientific Palampur Radicals Ambala

Digital Camera Model CT3 Model Coolpix S22

Qualified provisionally Not Qualified

6/6

Subject to the production of Authorized dealership certificate from principals/Manufacturer Specifications not clear as literature is not provided

Item No.: CEE-2


Name of the indenter: Dr. Mamta Awasthi
S.No 1.Double beam UV-VIS Spectrophotometer Wavelength range : 190-1100 nm, spectral bandwidth :1 nm(1901100 nm) wavelength display 0.1 nm increments, wavelength accuracy : 00.1 nm at D2 peak 656.1 nm 0.3nm for entire range, wavelength repeatability : 0.1 nm, wavelength slew rate about 6000 nm /minute , wavelength scanning speed: 3000-2 nm/minute, lamp interchange wavelength, automatic interchange linked to wavelength, Photometric system: Double beam optics, photometric range: absorbance -4 to 4 Abs. Transmittance: 0% to 400%, photometric accuracy: 0.002 abs. at 0.5 Abs, 0.004 Abs. at 1.0 Abs., 0.006 Abs. at 2.0 Abs. Photometric repeatability: < 0.001 Abs. at 0.5 Abs. Baseline stability < 0.0003Abs./hr; Light source: 20 Watt Halogen lamp and Deuterium lamp built in light source autoposition adjustment. Monochromator: Blazed Holographic grating CZERLY Tourner mounting, Silicon photodiode detector. Quote for Pentium Core 2 Duo PC branded with LCD Monitor and Laser color printer. Required stabilizer/UPS should be provided. Quote for 10mm quartz cell for routine and micro sampling. Holmium oxide filter to be quoted. S.No 1/6 Name of firms ECIL New Delhi Model Quoted Model: 5704SS Qualified/ Not qualified Not qualified Remarks Slew speed, Wavelength accuracy at D2 peak not mentioned, Photometric accuracy not mentioned, Transmittance range not clear, stability, Monochromator not mentioned (according to the literature provided), PC model and printer doesnt comply with specifications The specifications not mentioned/Clear from the literature are as follows Model 1901: wavelength accuracy : 00.1 nm at D2 peak 656.1 nm, wavelength slew rate about 6000 nm /minute wavelength scanning speed: 30002 nm/minute, lamp interchange wavelength, automatic interchange linked to wavelength, Photometric system: Transmittance: 0% to 400%, photometric accuracy: 0.006 Abs. at 2.0 Abs Monochromator: Blazed Holographic grating CZERNY Turner mounting, Silicon photodiode detector Subject to the production of Authorized dealership certificate from principals/Manufacturer

2/6

LD Analab Instrume nt,

Model 1901

Not qualified

3/6

4/6

Jain Scientific Agencies , Baddi Perkin Elmer India Shiva Glo bal Tec hno log y

UV 1800

Qualified Provisiona lly qualified

Lambda 35/45

5/6

Analytic Jena Model SPECOR D 200 PLUS B

qualified

Name of firms

Model Quoted/Not quoted

Qualified/N ot qualified

Remarks

2.pH meter pH Meter (microprocessor based) Microprocessor based PH meters, Range 0.00 to 14.00, Soft touch key, Accessories pH electrode, Temperature Probe, Dust Cover, Electrode Stand, Buffer Tablets, manual etc.

2/6 3/6

LD Analab Instrument, New Delhi Jain Scientific Agencies, Baddi Janvi Enterprises, Panchkula Haryana

6/6

Model Sx711 and MP512 Microproces sor based pH system 1010 Cyber Scan pH tutor

qualified qualified

qualified

S.No 3.Nephlometer (Conductivity) meter Conductivity Range: 0.0 to 2.00E6 S/cm Accuracy: 0.5000 % Full Scale Process Temp: -4 to 572 F Number of Control Relays: 3.00 Operating Temperature: -4 to 140 F 3/6 6/6 2/6

Name of firms LD Analab Instrument, New Delhi Jain Scientific Agencies, Baddi Janvi Enterprises, Panchkula Haryana

Model Quoted/Not quoted Model HK 338 Model 331 COND 610

Qualified/No t qualified

Remarks

qualified qualified qualified

S.No 4.Digital Oxygen analyzer with probe Dissolved O2 Range: 0 to 25 ppm, Accuracy: 0.1000 ppm Number of Control Relays: 3.00, Operating Temperature: -4 to 149 F 6/6 3/6 2/6

Name of firms LD Analab Instrument, New Delhi Jain Scientific Agencies, Baddi Janvi Enterprises, Panchkula Haryana

Model Quoted/Not quoted HK 318 Digital Dissolved Oxygen Meter 801 DO 600

Qualified/No t qualified

Remarks

qualified qualified

qualified

Item No.: CEE-3


Name of the indenter: Dr. Mamta Awasthi 4

S.No 1. Table Top Centrifuge Temp. Range 9 to + 400C, Short spin key, Fast cool option & stand by cooling. Facility to put at set rpm function, 10 acceleration & 10 deceleration rates. Automatic rotor recognition and Imbalance sensor. Motorized lid latch automatic lid opening, Ability to spin 15 ml and 50 ml falcon tubes and oak-ridge tubes at 12000 rpm and 18000g with fixed angle rotor. Adaptor should be available for volumes from 1.5 ml to 85 ml in the fixed angle rotor. The fixed angle and swing out rotor should accommodate both round bottom tubes as well as falcon tubes Swing Out Rotor with provision for 10MTP Plate, and adaptors for 384 & 96 well PCR plates. 1/4

Name of firms LD Analab, Delhi

Model Quoted/Not quoted K 241 R

Qualified/N ot qualified Not qualified

Remarks Detailed specification is not clear from the literature provided No mention of Motorized lid latch automatic lid opening, Ability to spin 15 ml and 50 ml falcon tubes and oak-ridge tubes at 12000 rpm and 18000g with fixed angle rotor. Adaptors for volumes from 1.5 ml to 85 ml in the fixed angle rotor, The fixed angle and swing out rotor should accommodate both round bottom tubes as well as falcon tubes Swing Out Rotor with provision for 10MTP Plate, and adaptors for 384 & 96 well PCR plates Subject to the production of Authorized dealership certificate from principals/Manufacturer

2/4

Jain Scientific Agencies, Baddi

CPR23/CPR-24

Qualified provisionall y

3/4

Eppendorf, India New Delhi

5804R:

Qualified

4/4

RajJi Scientific, Palampur LD Analab, Delhi

Multifuge X3R Model K 2045

Qualified provisionall y Not Qualified

2.

General Purpose Laboratory Centrifuge Upto 10000 rpm, 4/6 tube holders, with speed regulator, 0-60 minutes digital countdown timer. Dynamic brake for quick deceleration & Imbalance detector with cutoff in case of uneven loads. Presetting of speeds

1/4

2/4

Jain Scientific Agencies, Baddi Eppendorf, India New Delhi LD Analab, Delhi

Model R 24 Yes Eppendorf Centrifuge Minispin Yes Homogeniz er H500

Qualified provisionall y Qualified

Subject to the production of Authorized dealership certificate from principals/Manufacturer Not mentioned about Dynamic brake for quick deceleration & Imbalance detector with cutoff in case of uneven loads. Presetting of speeds It is not clear from the literature. Subject to the production of Authorized dealership certificate from principals/Manufacturer

3/4

3. Homogenizer Volume: 0-250 ml

1/4

Qualified

Flow rate:0.1-05 L/min

4/4

RajJi Scientific, Palampur

Homogeniz er: IKA germany Yes

Qualified provisionall y

Subject to the production of Authorized dealership certificate from principals/Manufacturer

Item No.: CEE-4


Name of the indenter: Dr. Mamta Awasth
S.No 1/5 Name of firms ECIL Hyderabad, Model AAS 4141 Lab India AA 7000 Model Quoted/Not quoted
Not quoted

Qualified/No t qualified N.A

Remarks

1.Vacuum Pump Portable

2/5

Not quoted

N.A

3/5

Perkin Elmer India Model AA400 GBC, Australia Advance Scientific Equipments Analytik Jenas Atomic Absorptio n Spectroph otometer Name of firms ECIL Hyderabad, Model AAS 4141 Lab India AA 7000

Not quoted

N.A

4/5

Not quoted

N.A

5/5

Not quoted

N.A

S.No 1/5

Model Quoted/Not quoted


Not quoted

Qualified/No t qualified N.A

Remarks

2. Vacuum Desicator 2/5 with pump

Not quoted

N.A

3/5

Perkin Elmer India Model AA400 GBC, Australia Advance Scientific Equipments

Not quoted

N.A

4/5

Not quoted

N.A

5/5

Analytik Jenas Atomic Absorptio n Spectroph otometer

Not quoted

N.A

S.No 1/5 3. Atomic Absorption Spectrophotometer Fully automated PC controlled true double beam Atomic Absorption Spectrophotometer system with absorption and emission capability with deuterium background correction and should have the following specifications: The system should be capable to analyze all mounted element in a Single aspiration without any turret movement Optics: Optical double beam designs, fully sealed optics against dust and vapour Automatic wavelength and slit selection with automatic peaking. Monochromator: Czerny turner with 250mm focal length with holographic grating having 1200 lines/mm or more. Wavelength range: 185-900nm. Slits of 0.2, 0.5 and 1.0nm full height and wide range PMT to cover entire wavelength Background Correction: correction up to 2.5 abs at 2ms response time with electronic modulation and automatic attenuation. Safety system: safety interlocks for burner, liquid trap, pressure relief bung, flame shield, flame operation, mains Power, gas and oxidant pressure, safety reservoir, spray Chamber. Gas Control: Instrument should have fully automatic gas control, and automatic change over between Air Acetylene and Nitrous oxide flame. Sensitivity: Better than 0.9 Abs with 0.5% RSD for 5ppm Cu solution. Burner & Air Compressor: Air Acetylene Burner, Nitrous Oxide acetylene burner with Acetylene cylinders, N20 Cylinder, and their regulators ( one for each) with pre heaters, Oil free Air compressor with filter, Fume hood with Exhaust fan. Coded Hollow Cathode Lamps for Fe, Co, Ni, Mn, Cu, Cr, Ca, Mg, Na, K and Zn Software & Computer/Printer: IBM Pentium IV, 2.2 GHz, 256 MB RAM, 48x CDROM Drive, 40 GB Hard Disk, 15 color monitor, 104 key board, mouse, Window 98/ 2000, Serial and parallel ports, and HP desk jet color printer.

Name of firms ECIL Hyderabad, Model AAS 4141

Qualified/ Not qualified Not qualified

Remarks Not mentioned Czerny turner monochromator Not clear about safety interlocks for pressure relief bung, flame shield, gas and oxidant pressure, safety reservoir, spray chamber NO The system not capable to analyze all mounted element in a Single aspiration without any turret movement. No automatic change over between Air Acetylene and Nitrous oxide flame. Not fully equipped with safety system Not mentioned Czerny turner monochromator Not clear about safety interlocks for pressure relief bung, flame shield, gas and oxidant pressure, safety reservoir, spray chamber NO The system not capable to analyze all mounted element in a Single aspiration without any turret movement. No automatic change over between Air Acetylene and Nitrous oxide flame. Not fully equipped with safety system NO The system not capable to analyze all mounted element in a Single aspiration without any turret movement Fume hood with Exhaust fan quoted NO The system not capable to analyze all mounted element in a Single aspiration without any turret movement. No automatic change over between Air Acetylene and Nitrous oxide flame. Microprocessor controlled system NO The system not capable to analyze all mounted element in a Single aspiration without any turret movement No automatic change over between Air Acetylene and Nitrous oxide flame

2/5

Lab India AA 7000

Not qualified

3/5

4/5

Perkin Elmer India Model AA400 GBC, Australia Advance Scientific Equipments Analytik Jenas Atomi c Absor ption Spectr ophoto meter

Not qualified

Not qualified

5/5

Not qualified

S.no 1 and 2 is not quoted by any of the vendor. S.NO 3: No system is capable to analyze all mounted element in a Single aspiration without any turret movement. Hence I dont recommend any of the vendors who qualify fully to the specifications mentioned.

Item No.: CEE-5


Name of Indenter: Dr. Mamta Awasthi
S.No Name of firm 2/9 BioAge, Mohali If Quoted Model LabPur-Series Analytica 6UV Technically qualified or disqualified Qualified provisional Remarks

1. Analytical Grade Water Purification System Capacity should be around 20 to 50 litres /day. Product water should have resistivity of 10-15 Megohm. cm, Conductivity < 0.2 semens, TOC levels less than 30 ppb, Flow rate 3 lit/hr, should have 3 stage purification systems. With auto diagnostic facility with Error No and Alarm Code for easy maintenance along with auto rinse facility. Unit temperature feed back pump with temperature compensation to 0.1C for consistent flow rate irrespective of temperature changes. In built display to ensure system operating parameters are displayed at all times

Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Manuf acturer Specification not clear as Literature not attached Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Manuf acturer

9/9

Janvi Enterprise, Panchkula

ZRX00031N-ELIX 3 S. KIT (IN)

Qualified provisionall y

2. Hot plate maximum temperature around 370 Deg. C with Electronic digital temperature controller cum indicator

1/9

3/9

Scientific and Allied Industries, Agra Jain Scientific Agencies

Navyug Model not mentioned MAC MSW-421

Not qualified

Qualified provisionall y

8/9

RajJi Scientific Surgico, Palampur

MAC MSW421/423

Qualified provisionall y

3.

Microwave Digestion system

4/9

Bash Chemo Agencies, Chandigarh Lab India, New Delhi

Model Mars 5

Qualified

5/9

Model Mars 5

Qualified

4.

Orbital Shaking Incubator

6/9

Eppendorf

Triple walled Orbital Shaking Incubator cum B.O.D. Incubator specifically designed to controlled incubation with orbital shaking arrangement with powerful geared motor for many laboratory procedures such as tissue cultures, fermentation studies, enzyme reactions, controlled purification and ageing tests, growth studies etc. Inside chamber is made of Stainless Steel 304 Qlty. having insulation between two walls by mineral wool. The acrylic glass door makes it easier to inspect specimen without disturbing the temperature. Temperature is controlled by Electronic Digital Temperature Controller Cum Indicator having range 5 Deg. C. to 60 Deg.C. The Evaporator coil, heating elements are provided in the air duct at the back near the top wall in the working space. Inside Air circulating system is provided by F.H. Motor with fan blade which keeps uniform temperature through out the chamber. A quiet running sealed compressor with all its protective devices Alongwith a fan cooled condensing coil are placed below the working chamber. The standard shaking platform shall accommodate 16 flasks of 500ml capacity. It has got lighting system from top of the chamber by 4 nos. 60 cms. Fluorescent tubes. Shaking speed is Controlled by Speed controller from 10 R.P.M. to 160 R.P.M. and indication of R.P.M. of shaker can be watched on Electronic R.P.M. Meter fitted in control panel. It has been provided with 7 days/24 hours programmed Timer. This shaker is supplied with 3 step Voltage Stabliser to work on 200/220 Volt A.C. 50 cycles. SIZE OF SHAKING INCUBATOR:660 W X 533 D X 533H (MM) 8/9 RajJi Scientific Surgico, Palampur

NBS Stackable Refrigerated Incubator Shaker Model Innova 42R Alternative model: New Brunswick Scientifics excella Shaker Modes E24R Iccubated Refrigerated MAC MSW-132 MSW-132(DX)

Qualified

Qualified provisionall y

Subject to the production of Authorized dealership certificate from principals/Manuf acturer

9/9

Janvi Enterprise, Panchkula

ACM-22064-I

Qualified

5. Water bath Doubled wall. Lid should be provided with either 6 holes or 12 holes to accommodate various capacity of flask with concentric ring. Equipment should be fitted with heater & thermostat with Digital Temperature Controller.

1/9

3/9

Scientific and Allied Industries, Agra Jain Scientific Agencies, Baddi

Navyug Model not mentioned Literature not attached

Not qualified

Not qualified

Specification not clear as Literature not attached Literature not attached

8/9

RajJi Scientific Surgico, Palampur

MAC MSW-272 Thermostatically controlled Digitally controlled ACM-54001

Both model qualified provisionall y

9/9

Janvi Enterprise, Panchkula

Qualified provisionall y

6. Thermometer 7. Vertical Autoclave double wall construction with boiler made of Stainless Steel 16 gauge sheet provided with air insulation. Lid should be provided with tightened alround radial locking . Joint less gaskets with Neoprene rubber tested upto 40 PSI. Sterilizers should be fitted with water level arrangement to indicate water position inside the boiler. Pressure gauge, Air/Steam release cock, safety valve should be provided. 220 Volts 50 cycles A.C. supply STAINLESS STEEL Capacity around 98 Ltrs. Fully automatic by PID Controller having facility for :Auto Purging: - Set Temp., Purging stop: - Set Temp., Auto Exhaust: - Cool, Elapse time, To Set Temperature., Facility for temp. Control, Auto Cycle Stop, Buzzer At Cycle over.

Not Quoted by any of the vendors 3/9 Jain Scientific Agencies, Baddi MAC MSW-101 (FAD) Qualified provisionall y

Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Man ufacturer Subject to the production of Authorized dealership certificate from principals/Man ufacturer Subject to the production of Authorized dealership certificate from principals/Man ufacturer Specification not clear as Literature not attached Specifications not clear as Literature not attached

8/9

RajJi Scientific Surgico, Palampur

MAC MSW-101 MAC MSW-101 (SAD) MAC MSW-101 (FAD)

Qualified provisionall y

9/9

Janvi Enterprise, Panchkula

Economy Model Double walled Cylindrical

Qualified provisionall y

7. Oven Temperature upto 2500C with digital temperature controller Double walled inside anodized stainless steel. Outside mild steel painted in epoxy powder coating. Perforated adjustable shelves. Inside one circulation fan should be provided for uniform temperature maintenance. The system should work on 220-240 V, 50 Hz. The inner size of the oven should be 60x60x60 cm approximately having overall minimum space of 0.2 cubic meters.

1/9

3/9

7/9

Scientific and Allied Industries, Agra Jain Scientific Agencies, Baddi Shiva Global Technologi es, Delhi

Navyug Model not mentioned

Not qualified

Literature not attached

Not qualified

Yamatos Drying Oven

Qualified

10

Variation in temperature should be within 50C

8/9

RajJi Scientific Surgico, Palampur

MSW211(Thermostaticall y controlled) MSW -211 (Digitally controlled) MSW-211 (SAD) Navyug Model not mentioned

Qualified provisionall y

Subject to the production of Authorized dealership certificate from principals/Manuf acturer Specification not clear as Literature not attached Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Manuf acturer Model not mentioned. Specification not clear as Literature not attached Model not mentioned. Specification not clear as Literature not attached Subject to the production of Authorized dealership certificate from principals/Manuf acturer Model not mentioned. Temp: 7000C (Mentioned in the main text however not as per specification) Specification not

8. Vertical Laminar Flow Cabinets An atmosphere free from air borne particles with Laminar air flowing downward the work table. An exhaust system may be fitted at bottom and back to throw the air outside the room. Complete Made Of Stainless Steel (304Qlty) With acrylic glass door & U.V. Light Complete structure made of Stainless Steel 304 Qlty). With cock for gas or vaccum line & exhaust fan. 10. 1800 X 600 X 600MM

1/9

8/9

Scientific and Allied Industries, Agra RajJi Scientific Surgico, Palampur

Not qualified

MSW-162

Qualified provisionall y

9/9

Janvi Enterprise, Panchkula

ACM-42102-L

Qualified provisionall y

11. 9. Automatic Pipette Washer Stainless steel washing cylinder 56 cm. Height upto 15cm in diameter mounted in a base 26.2cm in diameter, and a removable pipette holder, which is 13.75cm dia. Holder of stainless steel with a heavy wire-mesh bottom, rolled in and spotwelded for strength and durability, the cylinder should have 1.25cm inlet with a removable hose nipple for attaching to permanent dram if desired. Can accommodate pipettes up to 37.5cm in length.

1/9

Scientific and Allied Industries, Agra

Navyug Model not mentioned

Not qualified

3/9

Jain Scientific Agencies, Baddi

Literature not attached

Not qualified

8/9

RajJi Scientific Surgico

MSW-405

Qualified provisionall y

12. 10. Muffle Furnace Internal chamber size 5 x 5 x 10 13. Max. Temp. 1200C

1/9

Scientific and Allied Industries, Agra

Navyug Model not mentioned

Not qualified

11

clear as Literature not attached 3/9 Jain Scientific Agencies, Baddi Shiva Global Technologi es, Delhi RajJi Scientific Surgico Literature not attached Not qualified Temp upto 9000C Required 12000C (Literature not attached)

7/9

Yamatos muffle furnace model FO610 MSW-251

Qualified

8/9

Qualified provisionall y

Subject to the production of Authorized dealership certificate from principals/Manuf acturer Subject to the production of Authorized dealership certificate from principals/Manuf acturer

9/9

Janvi Enterprise, Panchkula

ACM-82301

Qualified provisionall y

Sd/Head(CEE)

Sd/Mamta Awasthi Indenter

Item No. VI (MED-1) Tender No. SPC-1/2010 DATED 08/04/10

Name of Item: Computerized single cylinder, 4 stroke variable compression ratio diesel engine(water cooled) test rig with eddy current dynamometer diesel 12

S.No. 1

Nameoffirm

Technically Remarks/reasonsof qualified/disquallified disqualification Thefirmmaybeasked forprovidingcopiesof supplyordersof ComputerisedVariable compressionratio dieselenginessupplied toIITs/NITs/Technical Institutions

M/STechnicalTeaching(D)equipment, Technicallyqualified 73/13,DodannaIndustrialestate, Bangalore

M/SEduTekequipment(India)Pvt. Ltd.,162VigyanViharDelhi

Technicallyqualified

Sd/Sd/Signature of indenter Signature of HOD

Technical Bid Evaluation Summary Item No.- VII (MED-2)


Tender No: SPC-1/2010(Item No. VII, Tender ref MED 2)

S.

Name of

Name of the firm

Technically

Remarks/ Reason
13

No.

the Item KAPTL INSTRUMENTATION 351 Mohit Nagar, Lane/13, Dehradun 248006 (UA) SHAILRON TECHNOLOGY PVT. Ltd. E-21 Suryakunj, Near CRPF, New Delhi - 110072 SHAILRON TECHNOLOGY PVT. Ltd. E-21 Suryakunj, Near CRPF, New Delhi - 110072

qualified/disqualifi ed

for disqualification 1. Not USB supported 2. Measurement Sensitivity of the system is not mentioned Along compatible logger/read unit with data out

1.

Thermocou ple Data Logger

Disqualified

2.

Pyranomet er

Qualified

3.

Digital Anemomet er

Disqualified

Note:The Financial Bids of QUALIFIED VENDERS shall be opened on 20.7.2010 in the conference room of Institute. Interested parties may attend the opening at the own expenses on the scheduled date please. Sd/- (25/6/2010)

Sd/Signature Head (MED)

Not compatible with Flow Rate measurement range Sd/Signature of Indenter

Faculty In-charge Purchase NIT, Hamirpur (H.P)


No.NIT/HMR/SPC/Tender-1/2010/- 1225 -27 Dated:25/6/2010 Copy to:1.HOD Centre for Energy & Environment, NIT, Hamirpur for necessary action. 2.Head Mech. Engg. Deptt., NIT, Hamirpur for necessary action. 3. Head (Computer Centre) for placing the same in Tender Portal immediately. Sd/- (25/6/2010) Faculty In-charge(Purchase) NIT, Hamirpur(HP) 14