MODEL CV-86 SERIES PIEZOELECTRIC VELOCITY TRANSDUCER INSTRUCTION MANUAL

FOREWORD
For Use in Safety.....
Thank you for your using our CV-86 Series Piezoelectric Velocity Transducer.

CV-86

SHINKAWA Sensor Technology applies strict quality control and inspections to ensure the high reliability of its products. This instruction manual contains descriptive information , specifications, principle of operation, installation procedures and field wiring with monitor. Please study contents of this manual and related manuals thoroughly before installing or operating the equipment, and keep it handy for future reference. (This manual is applied to Model CV-861, -862 version.)

! CAUTION
1. Do not drop the transducer and otherwise mishandle. 2. Be sure to install the transducer at environmental (case) temperature within -50 to +120C. 3. Field wiring should be performed correctly by reading this manual. 4. Install the transducer at the place where is most sensitive for vibration. the mounting surface must be clean, smooth, flat, and large enough to accommodate the entire base of transducer. 5. Use the specified connector and cable in the specifications. Keep cabling away from power and control lines to prevent noise interference. 6. Do not measure insulation resistance and dielectric strength other than those at places specified. If measured, transducer damage may result. 7. Do not remodel this unit without permission. Otherwise the guarantee can not be made. 8. This unit is designed for use by specialists or persons thoroughly familiar with the field. 9. Make sure that the end user receives the Instruction Manual delivered with this unit.

Before Use.....
When the unit is received, inspect it for damage suffered in transport and check whether it is the item you ordered. In the unlikely event that it was damaged in transport or does not function according to specifications, please contact the SHINKAWA Office or dealer nearest you. Store the unit under the ambient conditions given in the specification. Avoid places where it is exposed to high humidity or corrosive gases.

Follow closely all instructions written in the manual.

CONTENTS

CV-86

1. GENERAL INFORMATION
1-1 1-2 GENERAL SYSTEM CONFIGURATION EXAMPLE 1 1

2. SPECIFICATIONS
2 SPECIFICATIONS 2

3. RECEIVING INSPECTION AND STORING


3-1 3-2 RECEIVING INSPECTION STORING CONSIDERATIONS 4 4

4. INSTALLATION
4-1 4-2 4-3 4-4 GENERAL INSTALLATION PROCEDURE CONNECTOR CONNECTION PROCEDURE CABLE 5 5 6 7

5.

WIRING DIAGRAM
5-1 WIRING PROCEDURE 8

6. MAINTENANCE
6-1 MAINTENANCE 9

7. INTRINSIC SAFETY
7-1 7-2 TIIS 10 NEPSI 11

ii

1.

GENERAL INFORMATION

CV-86

1-1 GENERAL
Model CV-86 Sereis Piezoelectric Velocity Transducer is piezoelectric accelerometer with internal integration to velocity. The case configuration of the CV-86 is of the permanent mount type making it an ideal choice for permanent machine monitoring. The CV-86 offers a frequency response and phase stability that is superior to conventional velocity transducer.

1-2 SYSTEM CONFIGURATION EXAMPLE


Monitoring Dvice
CW Cable Recorder Output

Signal

Waveform Analyzing +24V power Vibration Monitor

Alarm Output

CV-86 Serise Piezoelectric Velocity Transducer

Machine Case

-1-

2.

SPECIFICATIONS

CV-86

Model Code

CV-86
Cable Type 1 2

/EX
Intrinsically Safety 1 7 TIIS (Ex ia A T4 X) NEPSI (Ex ia A T4)

Connector Type With integral cable type *1 (Coax. cable 3m standard)

Standard type Drip-proof type

CWCWCable length 050 100 150 5m 10m 15m 0 R Y

E-EFF-FFLug type*2 Cut off Ring lug Spade lug 0 1

00Armor Without With *1 Cable length 3m (end cut off), CW cable not required. *2 M3.5 crimping terminals at the cable ends are standard.

SPECIFICATIONS
Sensitivity Max. Velocity Vibration Limit Shock Limit Max. Shock Energy Natural Frequency Frequency Response Max.Transverse Sensitivity Output Impedance Amplitude Nonlinearity Temperature Response Power Requirement Grounding Operating Temperature range Range of Temperature at Explosion proof construction Relative Humidity Case Material Protection Rating Weight Accessories Supplied Output Connector*3 Matching Connector*3 Recommended Extension Cable*3 Optional Cable*3 3.94mV/mm/s (100mV/in/s REF.)pk 5% at 100Hz, 25C (77F REF.) 1,270mm/s (50in/s REF.)pk 2,450m/s (250g REF.)pk 24,500m/s (2,500g REF.)pk (non-Intrinsically Safe) 2 23,520m/s (2,400g REF.)pk (Intrinsically Safe) 4J (Intrinsically safe) 15kHz 2.5 to 3,500Hz 10%, 2 to 7,000Hz 3dB 5% 200(typical) 1% of F.S. 10% (around the operating temperature range) 18 to 30VDC, 2 to 10mADC (constant current) Case isolated, internally shielded -50C to +120C (-58 to +248F REF.) EX1,7 : -20 to +60C (Transducer, Cable) 100%RH Stainless Steel IP67(CV-861 & CW-F-FF, CV-862) Approx.145g(CV-861) Approx.250g(CV-862(including cable)) M6 mounting stud MIL-C-5015 2-pin MS3106A-10SL-4S 2-conductor shielded cable 2 2 AWG No.20(Approx.0.52mm to 0.75mm ) CW-E-EF-0- CW-F-FF-0- Not required *3 CV-861 TYPE SPECIFICATIONS
2 2

-2-

39348E3.3

2.

SPECIFICATIONS
Standard type cable CW-E-EF-0-
White Black Yellow L10%

CV-86

Connector type CV-861


2-pin connector

Signal,Power 0V Shield

(53)

(62) Hex. 24

Drip-proof type cable CW-F-FF-0-

White Black Yellow

Signal,Power 0V Shield

(4)

(25.4)

(80 ) M61 (mounting stud)

L10%

With integral cable type CV-862

White Signal,Power 0V, Shield

(5)

L10%

Black

(55)

Hex. 24

(50)

(25.4)

(4) M61 (mounting stud)

Dimension : mm

-3-

3.

RECEIVING INSPECTION AND STORING

CV-86

3-1 RECEIVING INSPECTION


Visually inspect the transducer for obvious shipping damage. If shippng damage is apparent, file a claim with the carrier and submit a copy to SHINKAWA Sensor Technology, LTD.

3-2 STORING CONSIDERATIONS


If storing is required temporarily, observe the following notes. (1) Store the transducers in the shipping box. (2) Do not store in the following places. a) with extreme temperature. b) with much moisture. c) with magnetic field. d) with shock or vibration. e) with corrosive materials. f) with water or dust.

-4-

4.

INSTALLATION
Model CV-86 can be mounted to the measured point by using attached mounting stud. ! CAUTION

CV-86

4-1 GENERAL

The mounting surface must be clean, smooth, flat, and large enough to accommodate the entire base of the transducer.

4.2 INSTALLATION PROCEDURE


Mount the CV-86 transducer by performing the following steps. (1) Drill and tap the mounting hole (M6, 5mm deep min.) perpendicular (901) to the mounting surface. (2) Clean the mounting surface. Dirt, dust, paint and oils must be wiped out from the mounting surface. Ensure the mounting hole is free off foreign material and the threads are clean. (3) Install the mounting stud into the transducer base and finger tighten until snug. (4) Apply some silicon grease to the mounting surface. (5) Install the transducer into tapped mounting hole and tighten the transducer with a torque wrench until snug. The transducer is torqued to about 320Ncm(33kgcm REF.). ! CAUTION Do not over-torque. Excessive torque can result in degraded performance, internal damage and inaccurate readings.

Connecto

Mounting stud Mounting hole Tapped mounting hole (M6, 5mm deep min.) Mounting surface

-5-

4.

INSTALLATION
Connect connector by performing the following steps for Model CV-861 connector type version. (1) Securely connect the connector of the cable to the connector receptacle of the transducer.

CV-86

4-3 CONNECTOR CONNECTION PROCEDURE

(2) To prevent loosening of the connector due to vibration, secure the connector with stainless wire as shown below.

Lock wire hole Stainless wire Lock wire hole

Connector

CAUTION 1. If used in the place with much moisture, encapsulate the connector with silicon compound that does not have an acetic base. 2. If used in outdoor environment, drip-proof type connector cable should be used. 3. Be ware not to apply the excessive force to the connector, which may cause the damage.

-6-

4.

INSTALLATION
Observe the following notes for cabling. (1) Use the CW-050 standard cable for connection with the CV-861 connector type version.

CV-86

4-4 CABLE

Use the twisted pair, shielded cable (AWG No.16 to 20) for connection between the junction terminal and monitoring device. Select suitable cable for operating environments. (2) The cable should not exceed 300 meters. Excessive length may limit the systems frequency response. (3) Keep cabling away from power and control lines, to prevent noise interference. (4) Cable should be securely fixed. Prevent the cable from bending sharply, twisting, kinking, knotting, and straining. The recommended minimum bend radius is 80mm. SHINKAWA recommends that the cable should be routed, through conduit to prevent physical damage and noise interference. ! CAUTION Be ware not to apply the excessive force to the connector cable, which may cause the damage.

-7-

5.

WIRING DIAGRAM

CV-86

5-1 WIRING PROCEDURE


The CV-86 requires an external DC power source to operate. Proper operation can be obtained by applying a DC power to the appropriate lead of transducer through a current regulating (CR) diode, as shown below. ! CAUTION 1. Never omit the current regulating diode. 2. Do not reverse polarity. 3. If the transducer is connected to the monitoring device (e.g. vibration monitor) that can supply DC power output to the transducer, refer to the instruction of the monitoring device.
May be provided as part of monitoring device. Power +18 to +30VDC (Recommended+24VDC) * CR diode Power, signal 2 to 10mA Signal output 0V(return)

0V(return) CV-86

Typical wiring

e.g. Motorola 1N5312 B+ B+ (power side) (power side)

Power, signal 0V

IN COM Fig.1 or

CV-86 Monitor Power, signal 0V IN COM Fig.2

Diode connection (Bottom view)

CV-86 Monitor

WARNING Beware of electric shock from high-voltage parts.

CAUTION 1. A blocking capacitor (10F) is required in series with the output signal if the readout instrument is DC coupled. 2. When the CW shielded cables connected to the shell, do not connect the shield cable to common terminal. (Refer to Fig.2)

-8-

6.

MAINTENANCE
The following items should be periodically checked to operate the system satisfactory. 1) Check the transducer body is properly mounted and secured. 2) Check there is no contact fault and water entered at connector. 3) Check the cable is connected securely to monitoring device. 4) Check the cable is in good condition.

CV-86

6-1 MAINTENANCE

-9-

7.

INTRINSIC SAFETY

CV-86

7-1 TIIS (Japanese Only)

-10-

7.

INTRINSIC SAFETY

CV-86

7-2 NEPSI (Chinese Only)

-11-

7.

INTRINSIC SAFETY

CV-86

-12-

Warranty

The warranty period for the delivered product shall be 3 years from the date of delivery from our factory. But the warranty period for the special order product and software shall be 1 year, and for the repairing parts shall be 6 months. In the event that the delivered product develops any defects within the warranty period for which Shinkawa Electric is liable, the defective part will be replaced or repaired in our factory at our own expense. In the event that an engineer needs to be dispatched for repair purposes during the warranty period, travel expenses for the said engineer will be changed in accordance with the above item. The cost of repair will be borne by Shinkawa Electric. However, the following are excluded from our warranty: (1) Any defects due to improper handling or operation by the user. (2) Any defects, the case of which is not attributable to Shinkawa Electric. (3) Any modifications or repairs made by others then Shinkawa Electric or persons commissioned by Shinkawa Electric. (4) Any handling, storage and operation under severe environmental conditions that exceed the design specifications. (5) Any defects due to fire, flood, earthquake, lightning and other Acts of God. (6) The consumables (7) Other defects considered not to be attributable to Shinkawa Electric. This warranty applies exclusively to the delivered product units. Shinkawa Electric will not be held liable for consequential damage caused either directly or indirectly through a defect of the delivered product. The warranty period and scope of machinery and equipment made by manufacturers other than Shinkawa Electric shall be in accordance with the conditions of the respective manufacturer, regardless of the provisions made above.