Some Definitions
A calorimeter measures the heat into or out of a sample. A differential calorimeter measures the heat of a sample relative to a reference. A differential scanning calorimeterdoes all of the above and heats the sample with a linear temperature ramp. Endothermic heat flows into the sample. Exothermic heat flows out of the sample. Melting- The endothermic transition upon heating from a crystalline solid to the liquid state Crystallization- The exothermic transition upon cooling from liquid to crystalline solid
DSC Components
DSC Accessories
= Cp
thermal events
where: Cp = specific heat (J/gC) T = temperature (C) H = heat (J) dH/dt = heat flow (J/min.) dT/dt = heating rate (C/min.)
Heat-flux DSC
Measurement of HF and T
Furnace Temp.
Sample Temp.
Furnace Temp.
Temperature (oC)
Exothermic up
Sample Preparation
Reference Pan Lid O-ring Pan
Sample Pan
Operating of TA Q20
Turning on in order of: DSC Cell Run Computer Open gas flow Heating Method (NOTE: No equilibrate segment necessary if starting at or near ambient temperature) Ramp 10C/min. to 300C Cooling Method Equilibrate at 300C Ramp 10C/min. to 25C
Computer Analysis
Provides an interface between you and the analysis instruments Enables you to set up experiments and enter constants Stores experimental data Runs data analysis programs.