You are on page 1of 9

Dental Materials Journal 27(1):7‐15, 2008

Mechanical and Thermal Cycling Effects on the Flexural Strength of Glass Ceramics
Fused to Titanium
Vanessa VÁSQUEZ1, Mutlu ÖZCAN2, Renato NISHIOKA1, Rodrigo SOUZA1, Alfredo MESQUITA1 and
Carlos PAVANELLI1
1
São Jose dos Campos Dental School, Department of Dental Materials and Prosthodontics, São Paulo State University, São
Jose dos Campos, Brazil
2
University Medical Center Groningen, University of Groningen, Department of Dentistry and Dental Hygiene, Clinical
Dental Biomaterials, Antonius Deusinglaan 1, 9713 AV, Groningen, The Netherlands
Corresponding author, Mutlu Özcan; E-mail: mutluozcan@hotmail.com

Received April 4, 2007/Accepted July 20, 2007

                                                         
This study evaluated the effects of mechanical and thermal cycling on the Áexural strength (ISO 9693) of three brands of
ceramics fused to commercially pure titanium (cpTi). Metallic frameworks of 25 3 0.5 mm dimensions (N=84) were cast
in cpTi, followed by 150-μm aluminum oxide airborne particle abrasion at a designated area of the frameworks (8 3 mm).
Bonder and opaque ceramic were applied on the frameworks, and then the corresponding ceramic (Triceram, Super Porce-
lain Ti-22, Vita Titankeramik) was Àred onto them (thickness: 1 mm). Half of the specimens from each ceramic-metal com-
bination were randomly tested without aging (only water storage at 37ºC for 24 hours), while the other half were mechani-
cally loaded (20,000 cycles under 10 N load, immersion in distilled water at 37℃) and thermocycled (3,000 cycles, between
5‐55℃, dwell time of 13 seconds). After the Áexural strength test, failure types were noted. Mechanical and thermal
cycling decreased the mean Áexural strength values signiÀcantly (p<0.05) for all the three ceramic-cpTi combinations tested
when compared to the control group. In all the three groups, failure type was exclusively adhesive at the opaque ceramic-
cpTi interfacial zone with no presence of ceramic on the substrate surface except for a visible oxide layer.

Keywords: Flexural strength, Aging, Titanium


                                                         

oxygen during casting when conventional casting and


INTRODUCTION
investing methods are used5). Furthermore, tempera-
Although there is a growing trend for metal-free res- tures above 800℃ increase the oxygen-rich titanium
torations in the dental profession, failures associated oxide layer on the surface, so called α-case, which
with such materials are also being reported ― indi- impairs the mechanical compatibility of the titanium-
cating that there is still place for metal-ceramic Àxed ceramic system6-9). It should also be mentioned that
partial dentures (FPD). Furthermore, extensive oral cpTi has a low thermal expansion coefÀcient that
rehabilitation could be achieved with metal-ceramic makes it a difÀcult substrate for ceramics to bond
FPDs only. However, the high cost of noble alloys onto.
and the potential biological hazards of base metal   To the end of better compatibility with cpTi,
alloys pose some concerns for practical application. ceramics to meet this speciÀc purpose have been
To circumvent these concerns, commercially pure developed1). These ceramics, so called low-fusing
titanium (cpTi) and some of its alloys were intro- ceramics (LFC), melt at temperatures lower than
duced for the construction of dental prostheses1). 800℃. Moreover, they have thermal expansion coef-
  According to the American Society for Test- Àcients close to that of titanium, thereby reducing
ing and Materials (ASTM), cpTi is available in four the thermo-mechanical stresses at the interface and
different grades (grade I to IV). It is based on the allowing satisfactory bonding of the two elements10).
incorporation of small amounts of oxygen, nitrogen, By virtue of the advances in dental technology, the
hydrogen, iron, and carbon during puriÀcation pro- α-case layer can also be controlled almost com-
cedures, whereby each grade has different physi- pletely through improved casting techniques. This
cal and mechanical properties. Grades I and II are is done by induction in an inert atmosphere of argon
the most commonly used types in the production of or helium gas and with the use of refractory invest-
metal-ceramic FPDs1). In implants and implant-sup- ments that contain oxides such as magnesium,
ported FPDs, cpTi and its alloys exhibit remarkable yttrium, and zirconium11-13).
advantages due to their excellent biocompatibility,   The success of metal-ceramic FPDs hinges on
corrosion resistance, high strength, and low modulus a durable adhesion between ceramic and the metal
of elasticity1-4). substructure9,14,15). In today s dentistry, cpTi compat-
  However, cpTi has a high melting point and it ibility with ceramics, refractory materials, and spe-
is active at high temperatures. For these reasons, cial casting systems are not widely studied. These
it may become fragile if it reacts with atmospheric studies should indeed be undertaken with a view to
8 Ceramic-titanium Áexural strength

improving the success rate of metal-ceramic FPDs1). used for the fabrication of frameworks. A wax sprue
The cpTi-ceramic adhesion could be tested in sev- (Horus, Herpo Produtos Dentários Ltd., São Paulo,
eral ways. These tests can be classiÀed according to Brazil) was attached perpendicular at one end of the
the nature of the applied stress, such as shear, ten- template and connected to a central wax rod of 5 mm
sile, Áexural strength, or torsion test. However, to diameter (Wax Wire for Casting Sprues, Dentaurum,
date, there is still no consensus as to which method Pforzheim, Germany). The assembly was mounted in
is more valid for clinical applications16). In addition, a silicone ring and poured with investment material
these methods often do not involve the fatigue com- (Rematitan® Ultra, Dentaurum JP Winkelstroeter
ponent of mechanical tests, thereby rendering the KG, Pforzheim, Germany) that was mixed at a ratio
results unrealistically optimistic to a certain extent. of 100 g of powder to 14 ml of liquid. After the
  To better predict the clinical behaviors of dif- investment material set, the silicone ring and sprue
ferent materials and material combinations, their former were separated from the investment mold.
mechanical fatigue tests should be carried out in a Metallic frameworks were cast in cpTi (N=84) in an
wet environment17,18). In this connection, thermo- electrical induction furnace (Rematitan® Autocast,
cycling that is based on temperature alternations Dentaurum) under argon gas. Elimination of sprues
induces repeated stress to the metal-ceramic inter- and separation of metallic strips were performed
face18,19). During thermocycling, differences between with the aid of carbide disks at low speed.
the thermal expansion coefÀcients of the two com-   After removal from the investment mold, mar-
ponents could affect adhesive strength. However, a gins of the frameworks were trimmed to the Ànal
combination of both mechanical and thermal cycling dimensions of 25 3 0.5 mm. Surfaces of the speci-
could be considered a more aggressive way of aging mens that would receive the ceramic layer were
the ceramic-metal interface. treated with airborne particle abrasion with 150-μm
  Undisputedly, cpTi and its alloys have many aluminum oxide (Korox, Bego, Bremen, Germany).
advantages and favorable properties over the other This was done at an angle of 45 for 10 seconds,
dental alloys. However, controversies exist in the lit- and from a distance of approximately 2 cm under 2
erature regarding the adhesion of ceramics to these bar pressure. Frameworks were then ultrasonically
alloys7,20-22). Therefore, the objectives of this study cleaned in isopropyl alcohol (Vitasonic II, Vita
were twofold: (1) to evaluate the effects of mechanical Zahnfabrik, Bad Säckingen, Germany) for Àve
and thermal cycling on the Áexural strength of three minutes and allowed to dry at room temperature.
brands of ceramics fused to cpTi; and (2) to asses the
type of failure at the ceramic-alloy interface. Application of ceramic layer
An area of 8 3 mm was marked on the cpTi frame-
works with a graphite pencil. Then, the bonder
MATERIALS AND METHODS
of each ceramic was applied in a thin layer with a
Materials used brush. After Àring, opaque ceramic paste ― con-
Three brands of ceramics ― namely, Triceram, Super sisting of opaque ceramic powder mixed and homog-
Porcelain Ti-22, and Vita Titankeramik ― were enized with liquid in a container connected to a dis-
used in combination with cpTi in this study. Brand penser ― was applied on the bonder. Thickness of
names, manufacturers, and batch numbers of the the ceramic layer corresponding to dentin ceramic
ceramic types and titanium are presented in Table 1. (1 mm) was standardized by positioning the frame-
works in a metallic template. After removal from
Fabrication of metallic frameworks the assembly, ceramic was Àred. Due to shrinkage,
Rectangular acrylic templates (25 3 0.7 mm) were a second layer was applied and then the specimens

Table 1 Brand names, indications, compositions, and manufacturers of dental ceramics and titanium used in this study
Brand name Ceramic Type Manufacturer Batch number

Triceram Low-fusing ceramic Dentaurum, Ispringen, Germany 003

Super Porcelain Ti-22 Low-fusing ceramic Noritake, Nagoya, Japan 60506

Vita Titankeramik Low-fusing ceramic Vita Zahnfabrik, Bad Säckingen, 2370


Germany
Commercially pure titanium (Tritan) Dentaurum J.P. Winkelstroeter KG, 098
Pforzheim, Germany
VÁSQUEZ et al. 9

Table 2 Firing procedures of the dental ceramics tested


Ceramics Starting Drying Time Final Temperature Holding time
Temperature (ºC) (min) Temperature (℃) rate of increase (min)
(℃/min)

Triceram

Bonder 500 4 795 65 1

Opaque 500 4 795 65 1

First dentine layer 500 6 755 40 1

Second dentine layer 500 4 755 40 1

Super Porcelain Ti-22

Oxidation 500 3 800 50 3

Bonder 500 5 800 50 1

Opaque 500 5 780 50 1

First dentine layer 500 7 760 40 1

Second dentine layer 500 7 760 40 1

Vita Titankeramik

Bonder 400 6 800 60 1

Opaque 400 2 790 110 1

First dentine layer 400 6 770 50 1

Second dentine layer 400 6 770 50 1

were submitted for Ànal glaze Àring (Table 2). water (Nova Etica, São Paulo, Brazil). Dwelling time
  Twenty-eight specimens for each ceramic-cpTi at each temperature was 10 seconds, and the trans-
combination were randomly divided into two sub- fer time from one bath to the other was Àve seconds.
groups: mechanical and thermal cycling versus only   The process of subjecting the specimens to
stored in distilled water for 24 hours at 37℃ (control 20,000 times of mechanical cycling followed by 3,000
group) prior to Áexural strength test. times of thermal cycling is regarded as the aging
procedure.
Mechanical and thermal cycling
Mechanical cycling of the specimens was carried Flexural strength test
out in a mechanical cycling machine (custom made, Flexural tests were performed in a universal test-
Paulista State University, Dental School, UNESP, ing machine (Instron 4301, Instron Corp., Norwood,
Sao Jose dos Campos, Brazil). The latter was devel- MA, USA). The load was applied at a constant speed
oped to simulate the mechanical forces generated of 1.5 mm per minute until fracture occurred (Figs.
during the chewing cycle. The device employed for 1(a)‐(d)). The formula according to the guidelines
this test was composed of two bases, 2 cm apart from of ISO 969323) was adopted for the calculation of
each other, on which cylinders (radius: 1.0 mm) were data obtained from the Áexural strength test. The
placed to allow specimens to be positioned parallel to load that led to initial separation of materials was
the ground and perpendicular to the axial load. An obtained in kilogram force (kgf). It was converted to
upper rod with a 1 mm diameter tip was Àxed on a Newton (N) for the calculation of Áexural strength
plier to apply a 10 N load for 20,000 times at a fre- according to the following equation:
quency of 1 cycle per second. The device for testing
was placed on the machine base that contained a
thermostat to allow testing in an aqueous medium at
a constant temperature of 37℃.
  Subsequently, the specimens were thermocycled where P is the maximum load upon fracture (N), I
for 3,000 cycles between 4℃ and 55℃ in deionized the distance between two supports (mm), and b the
10 Ceramic-titanium Áexural strength

Fig. 1 (a) Final shape and dimensions of ceramic-alloy specimen; (b) cross-section dimensions of
specimen according to ISO 9693; (c) application of axial force; (d) separation of ceramic from
titanium surface.

width (mm) and d the thickness (mm) of the speci- analysis of variance (ANOVA), with Áexural strength
men. as the dependent variable and the ceramic-metal
  Specimens were analyzed under a stereomi- combination and aging condition as the independent
croscope (Stemi 2000-C, Carl Zeiss, Gottingen, Ger- factors. P values less than 0.05 were considered to
many). Images were digitally recorded with a cam- be statistically signiÀcant in all the tests. Multiple
era (Cybershot, Model DSC-S85, Sony, Tokyo, Japan), comparisons were made by Tukey s adjustment.
which was connected to the microscope, for metal
surface characterization and assessment of failure
RESULTS
modes. In this study, failure types were classiÀed
as follows: (a) adhesive failure along the interfacial Table 3 presents the two-way analysis of variance
region between opaque ceramic and ceramic-metal (ANOVA) results for the experimental conditions.
interaction zone; (b) inside the interaction zone; Interaction between ceramic type and aging factor
and (c) cohesive failure along the interfacial region was not statistically signiÀcant (p=0.168) (ANOVA,
between metal and the interaction zone9). Tukey s test). In Table 4, it was shown that mechan-
ical and thermal cycling decreased the mean Áex-
Statistical analysis ural strength values signiÀcantly (p<0.05) for all the
Statistical analysis was performed using STATIS- three ceramic-cpTi combinations tested (27.4 4.1‐
TICA for Windows (version 5.5, StatSoft Inc., Tulsa, 28.5 5.4 MPa) when compared to the control group.
OK, USA) and Statistix for Windows (version 8.0, In the latter group, tests were performed after 24-
Analytical Software Inc., Tallahassee, FL, USA). hour water storage at 37℃ (32.5 4.3‐38.5 2.3
The means of each group were analyzed by two-way MPa).
VÁSQUEZ et al. 11

  Results of Tukey s multiple comparison test systems). There was no presence of ceramic on the
established that only Vita Titankeramik-cpTi system substrate surface, except for a visible layer of dark
showed a signiÀcantly higher value than those of titanium oxide. Representative images of original
other ceramic-cpTi combinations in the control group cpTi-ceramic substrate and cpTi surfaces after Áex-
(p<0.05) (Fig. 2). ural strength test are illustrated in Figs. 3(a)‐(g).
  Stereomicroscope images at 25 magniÀca-
tion showed exclusively adhesive failure mode at the
opaque ceramic-cpTi interfacial zone in all the exper- Table 4 Mean ( standard deviation) Áexural strength
imental groups (Triceram-cpTi, Super Porcelain Ti values (MPa) of ceramic-cpTi combinations with
22-cpTi, and Vita Titankeramik-cpTi metal-ceramic and without mechanical and thermal cycling.
Same superscript letters indicate no signiÀcant
differences (Tukey s test, α=0.05)
Table 3 Results of two-way analysis of variance (ANOVA)
Mechanical- and
for ceramic type, cycling fatigue conditions, and Experimental thermal-cycling Mean (SD)
their interaction according to Áexural strength Groups
data (*: p<0.05) Without With
Triceram-cpTi 32.5 4.3a,b,c 28.4 5.3b,c 30.5 5.1
Effect DF SS MS F P (Group 1)
Cycling 1 594.00 594.00 22.78 0.001* Super 35.4 7.6a,b 28.5 5.4b,c 32 7.3
Groups 2 38.21 19.11 0.73 0.488 Porcelain Ti-22-
(metal-ceramic) cpTi (Group 2)

Vita 38.5 2.3a 27.4 4.1c 32.8 6.8


Interaction 2 97.76 48.88 1.87 0.168 Titankeramik-
Residue 36 938.74 26.08 cpTi (Group 3)
Total
41 1668.72 Mean (SD) 35.5 5.5 35.5 5.5
* Statistically signiÀcant difference at the level of 5%. * Means followed by equal letters do not differ statistically

Fig. 2 Mean Áexural strength values according to the experimental conditions established by the variables of ceramic
type and aging (20,000 times of mechanical cycling followed by 3,000 times of thermal cycling).
12 Ceramic-titanium Áexural strength

Fig. 3 (a) Representative image of ceramic specimen Àred onto cpTi and optical microscopic images of specimens ( 25)
after Áexural strength test: (b)‐(c) Triceram-cpTi; (d)‐(e) Super Porcelain Ti-22-cpTi; (f)‐(g) Vita Titankeramik-
cpTi. Clear appearance of cpTi metal surface is seen as well as blur appearance of oxide layer on the ceramic
surface. Metal and oxide surfaces are indicated by an arrow for each ceramic-cpTi combination
VÁSQUEZ et al. 13

this study presented low Àring temperatures for each


DISCUSSION
layer ― between 750 and 800℃. Lautenschlager
The clinical performance of metal-ceramic FPDs and Monaghan2), Wang and Fenton1), and Esquivel
is usually estimated by mechanical strength tests, et al.10) proposed improved adhesion between ceram-
which assess the adhesion between a ceramic and the ics and cpTi with so-called ultra-low fusing ceramics,
metal substrate. The nature of metal-ceramic bond- whereby the Àring temperatures range between 650
ing has been studied extensively, and it is funda- and 850℃9). These ceramics decrease the transfor-
mentally based on three mechanisms: micromechani- mation phase of cpTi, resulting in a stable oxide layer
cal retention, compressive adaptation, and chemical and good adhesion of the ceramics. Failure analysis
bonding9,15,24). also indicated a visible dark zone of oxide layer.
  In dental materials research, Áexural strength   Differences in thermal expansion coefÀcient
test is considered as the most appropriate method between the metal and ceramic can produce resid-
to measure bond strength between two materi- ual stresses along the interface, which can result in
als4,10,14,18,19,21,25-28). Some authors would recommend debonding or fracture of the ceramic9,21,22). Thermal
three-point bending test29,30), while others would expansion coefÀcients of the ceramics used in this
advocate four-point or biaxial Áexure strength test to research varied between 8.4 and 8.9 10‐6K‐1, being
evaluate metal-ceramic adhesion31). According to ISO very close to each other. Thus, non-signiÀcant differ-
969323), Áexural strength test should be preferred ences among the ceramics tested could be explained
over the other methods because the stresses it simu- on this ground. In the present study, results of the
lates closely represent the stresses that occur in den- control group without mechanical and thermal aging
tal prostheses with multiples elements. (32.5‐38.5 MPa) corroborated with the Àndings of
  The α-case layer that formed on the surface of Yilmaz and Dincer14) and White et al.17) ― where low-
melted cpTi alloys provided good chemical stability. fusing ceramics were also used. However, between
However, at the same time, this intermediate oxide the studies of Pröbster et al.19) and Yilmaz and
layer could impair the chemical reactions between Dincer14), the Áexural strength values showed great
cpTi and the bonder ceramic components, thereby variations ranging from 14 MPa to 37 MPa. None-
weakening the cpTi-ceramic bonding. It is note- theless, results of the current study ― both the con-
worthy that this oxide layer is often not strongly trol group without aging and even after aging ― were
adhered to the metal surface and that it is porous, higher than the recommended minimum value of 25
brittle, and incompatible with ceramics. It was MPa as established by DIN 13.92734).
therefore little wonder that the oxide layer itself or   In terms of specimen dimensions, Yoda et al.4)
the interface between the oxide layer and the alloy suggested that the speciÀcations of ISO 969323) did
has been shown to be responsible for metal-ceramic not entirely represent those of metal-ceramic dental
fractures3,6,14,28). In a bid to circumvent the incompat- restorative systems and Troia et al.28) attributed this
ibility problem between the α-case layer and ceram- to non-uniformity of ceramic thickness. For this rea-
ics, a ceramic bonder corresponding to each ceramic son, ceramic thickness was standardized at 1 mm in
material was thus used in this study. this study.
  Besides, composition of the investment mate-   The effects of mechanical and thermal cycling
rial and casting procedure also inÁuence the forma- simulate ― to a certain extent ― the effects of the
tion of α-case layer. For this reason, some authors clinical service which dental materials are subjected
advocated the use of Mg-based investment materi- to in the oral environment. In dental materials
als32), while others suggested Zr-modiÀed Mg-based research, most in vitro experiments are performed
investment materials33). Addition of these elements using static mechanical tests that do not address
increases the thermal expansion coefÀcient of the the aggressive oral environment. It is known that
investment material and limits its interfacial reactiv- the oral environment is able to induce physicochemi-
ity with cpTi at high temperatures. In the present cal alterations in dental materials. Temperature
study, the investment material used was in compli- changes provide conditions for occurrence of deg-
ance with the manufacturer s recommendations. radation in an aqueous environment35). They also
  The pressure at which liquid titanium was encourage mechanical fatigue of the materials them-
injected into the investment32), and the temperature selves or their interfaces, which is triggered by the
at which this procedure was performed, could also repeated chewing action36,37). As for the water stor-
inÁuence the consequential adhesion of ceramics to age of ceramic materials, it decreases the latter s
cpTi22). Within this context, ceramics have been spe- mechanical properties38). The reduction may be
ciÀcally developed to make their Àring onto cpTi pos- related to the solubility of different oxides. This
sible at temperatures below 800℃, so as to prevent process may be higher in ceramics designed for use
excessive formation of oxides on the metallic frame- with cpTi because of the presence of alkaline metallic
work surface2). The ceramic systems employed in oxides35) ― which was also evident from the failure
14 Ceramic-titanium Áexural strength

analysis of this study.


  Previous studies which investigated the adhesion
CONCLUSIONS
between metal and laboratory resin or between metal
and ceramics disclosed reduced adhesion after ther- Mechanical cycling for 20,000 times followed by ther-
mal cycling18,19,39-41). In the study of Tróia Jr et al.28), mal cycling for 3,000 times decreased the mean Áex-
the same form of thermal variation and metallic sub- ural strength values signiÀcantly for all the three
strate (cpTi) as the present study were investigated. ceramic-cpTi combinations tested when compared to
However, it was found that thermal cycling did not the control group ― where tests were performed after
exert any inÁuence on adhesion of ceramic with this mere 24 hours water storage at 37℃. All tested
metal. Nonetheless, this factor became more sig- ceramic-cpTi combinations showed adhesive fail-
niÀcant and pronounced in more drastic conditions ure with a visible oxide layer on the cpTi substrate,
as those employed in the study of Shimoe et al.42), indicating that the weakest link was still between
where extended thermal cycling (100,000 cycles) was ceramic and the titanium oxide assembly.
performed. As a result, a 30% reduction in mean
bond strength was observed. In this study, the chief
ACKNOWLEDGEMENTS
focus was on the combined effects of mechanical and
thermal cycling. On this ground, extended thermal This investigation was Ànancially supported by a
cycling was not performed. grant, Grant No. 01166/04, from FUNDUNESP
  According to Scherrer et al.36), all materials and (Fundação para Desenvolvimento da Universidade
their combinations should be subjected to fatigue Estadual Paulista).
conditioning before mechanical testing is performed.
Mechanical fatigue is a helpful means to predict the
REFERENCES
clinical behavior of dental restorative systems in
order to avoid catastrophic failures in vivo37). Some 1) Wang RR, Fenton A. Titanium for prosthodontic
factors inherent to mechanical cycling may also applications: a review of the literature. Quintes-
inÁuence the outcome. In this study, the applied sence Int 1996; 27: 401-408.
2) Lautenschlager EP, Monaghan P. Titanium and
load was 10 N and the number of cycles was 20,000
titanium alloys as dental materials. Int Dent J
times at a frequency of 1 Hz as described by Itinochi 1993; 43: 245-253.
et al.43). As for thermal cycling, the number of cycles 3) Könönen M, Kivilahti J. Bonding of low-fusing
adopted and the temperature alterations employed dental porcelain to commercially pure titanium. J
were based on previous reports18,19,28). Thermal Biomed Mater Res 1994; 28: 1027-1035.
cycling induces repeated stress at the metal-ceramic 4) Yoda M, Konno T, Takada Y, Iijima K, Griggs J,
interface, thereby weakening the bond between the Okuno O, Kimura K, Okabe T. Bond strength of
two components28,44,45). In this connection, Tróia et binary titanium alloys to porcelain. Biomater 2001;
22: 1675-1681.
al.28) suggested that extended immersion times in
5) Akagi K, Okamoto Y, Matsuura T, Horibe T. Prop-
each bath might produce higher tension at the metal- erties of test metal ceramic titanium alloys. J Pros-
ceramic interface. On this note, Pröbster et al.19) and thet Dent 1992; 68: 462-467.
Leibrock et al.45) suggested that 6,000 thermal cycles 6) Kimura H, Horng CJ, Okazaki M, Takahashi J.
would correspond to Àve years in physiological condi- Oxidation effects on porcelain-titanium interface
tions. reactions and bond strength. Dent Mater J 1990; 9:
  In this study, reduction in mean Áexural 91-99.
strength was evident after aging. Although previous 7) Adachi M, Mackert JR Jr, Parry EE, Fairhurst CW.
Oxide adherence and porcelain bonding to titanium
studies that investigated ceramic-cpTi bond strength
and Ti-6Al-4V alloy. J Dent Res 1990; 69: 1230-
did not incorporate the aging conditions19,28), other 1235.
studies sought to achieve the same research objective 8) Könönen M, Kivilahti J. Fusing of dental ceramics
by modifying fatigue variables such as load magni- to titanium. J Dent Res 2001; 80: 840-854.
tude and type of load (such as repetitive or dynamic 9) Anusavice KJ. Phillips: Materiais dentários, Trad
load)18). Therefore, with a view to making study Dourado et al., 11th ed, Elsevier, Rio de Janeiro,
results comparable, there seems to be a need to stan- 2005, p.763.
dardize the aging parameters. 10) Esquivel JF, Chai J, Wozniak WT. The physical
properties of low-fusing porcelains for titanium. Int
  In clinical situations, many metal-ceramic FPD
J Prosthodont 1996; 9: 563-571.
failures are in the form of ceramic fractures with
11) Herφ H, Syverud M, Waarli M. Mold Àlling and
metal exposure. While the search is still on for a porosity in castings of titanium. Dent Mater 1993;
better ceramic-compatible metal/alloy with improved 19: 15-18.
clinical performance, clinicians should in the mean- 12) King AW, Lautenschlager E, Chai J, Gilbert J. A
while bear in mind that ceramic-metal bonding is comparison of the hardness of different types of tita-
susceptible to mechanical and thermal fatigue. nium and conventional metal ceramics. J Prosthet
VÁSQUEZ et al. 15

Dent 1994; 72: 314-319. to titanium and a titanium alloy. J Dent 2003; 31:
13) Syverud M, Herφ H. Mold Àlling of Ti casting using 509-518.
investments with different gas permeability. Dent 30) Suansuwan N, Swain MV. New approach for eval-
Mater 1995; 11: 14-18. uating metal-porcelain interfacial bonding. Int J
14) Yilmaz H, Dinçer C. Comparison of the bond com- Prosthodont 1999; 12: 547-552.
patibility of titanium and an NiCr alloy to dental 31) Cai Z, Bunce N, Nunn ME, Okabe T. Porcelain
porcelain. J Dent 1999; 27: 215-222. adherence to dental cast CP titanium: effects of sur-
15) Powers JM, Sakaguchi RL. Craig s restorative den- face modiÀcations. Biomater 2001; 22: 979-986.
tal materials, 12th ed, Mosby, St. Louis, 2006, p.632. 32) Ida K, Togaya T, Tsutsumi S, Takeuchi M. Effect of
16) Hammad IA, Talic YF. Designs of bond strength magnesia investments in the dental casting of pure
tests for metal ceramic complexes: review of the lit- titanium or titanium alloys. Dent Mater J 1982; 1:
erature. J Prosthet Dent 1996; 75: 602-608. 8-21.
17) White SN, Li ZC, Yu Z, Kipnis V. Relationship 33) Hung CC, Hou GL, Tsai CC, Huang CC. Pure tita-
between static chemical and cyclic mechanical nium casting into zirconia-modiÀed magnesia-based
fatigue in a feldspathic porcelain. Dent Mater 1997; investment molds. Dent Mater 2004; 20: 846-851.
13: 103-110. 34) German Standard Institution. DIN 13.927 metal-
18) Poljak-Guberina R, Catovic A, Jerolimov V, Franz ceramic system, Beuth, Berlin, 1990.
M, Bergman V. The fatigue strength of the inter- 35) Kvam K, Derand T, Austrheim EK. Fracture tough-
face between Ag-Pd alloy and hydrothermal ceramic. ness and Áexural strength of dental ceramics for
Dent Mater 1999; 15: 417-420. titanium. Biomaterials 1995; 16: 73-76.
19) Pröbster L, Maiwald U, Weber H. Three-point bend- 36) Scherrer SS, Wiskott AH, Coto-Hunziker V, Belser
ing strength of ceramics fused to cast titanium. Eur UC. Monotonic Áexure and fatigue strength of com-
J Oral Sci 1996; 104: 313-319. posites for provisional and deÀnitive restorations. J
20) Wang RR, Welsch GE, Castro-Cedeno M. Interfacial Prosthet Dent 2003; 89: 579-588.
reactions of cast titanium with mold materials. Int 37) Teoh SH. Fatigue of biomaterials: a review. Int
J Prosthodont 1998; 11: 33-43. Biomat Fatigue 2000; 22: 825-837.
21) Garbelini WJ, Henriques GEP, Tróia Júnior M, 38) Morena R, Beaudreau GM, Lockwood PE, Evans AL,
Mesquita MF, Dezan CC. Evaluation of low-fusing Fairhurst CW. Fatigue of dental ceramics in a sim-
ceramic systems combined with titanium grades II ulated oral environment. J Dent Res 1986; 65: 993-
and V by bending test and scanning electron micros- 997.
copy. J Appl Oral Sci 2003; 11: 354-360. 39) Taira M, Moser JB, Greener EH. Studies of Ti
22) Bondioli IR, Bottino MR. Evaluation of shear bond alloys for dental castings. Dent Mater 1989; 5: 45-
strength at the interface of two porcelains and pure 50.
titanium injected into the casting mold at three dif- 40) Özcan M, Pfeiffer P, Nergiz . A brief history and
ferent temperatures. J Prosthet Dent 2004; 91: 541- current status of metal/ceramic surface condition-
547. ing concepts for resin bonding in dentistry. Quintes-
23) International Organization for Standardization. sence Int 1998; 29: 713-724.
ISO 9693: Metal-ceramic dental restorative systems, 41) Kim JY, Pfeiffer P, Niedermeier W. Effect of labora-
Geneva, 1999, pp.1-14. tory procedures and thermocycling on the shear bond
24) Riley EJ. Ceramo-metal restorations. Dent Clin strength of resin-metal bonding systems. J Prosthet
North Am 1977; 21: 669-682. Dent 2003; 90: 184-189.
25) Gilbert JL, Covey DA, Lautenschlager EP. Bond 42) Shimoe S, Tanoue N, Yanagida H, Atsuta M,
characteristics of porcelain fused to milled titanium. Koizumi H, Matsumura H. Comparative strength
Dent Mater 1994; 10: 134-140. of metal-ceramic and metal-composite bonds after
26) Pang IC, Gilbert JL, Chai J, Lautenschlager EP. extended thermocycling. J Oral Rehabil 2004; 31:
Bonding characteristic of low fusing porcelain 689-694.
bonded to pure titanium and palladium-copper alloy. 43) Itinoche KM, Özcan M, Bottino MA, Oyafuso D.
J Prosthet Dent 1995; 73: 17-25. Effect of mechanical cycling on the Áexural strength
27) White SN, Ho L, Caputo AA, Goo E. Strength of of densely sintered ceramics. Dent Mater 2006; 22:
porcelain fused to titanium beams. J Prosthet Dent 1029-1034.
1996; 75: 640-648. 44) Gale MS, Darwell BW. Thermal cycling procedures
28) Tróia Jr MG, Henriques GEP, Nóbilo MAA, for laboratory testing of dental restorations. J Dent
Mesquita MF. The effect of thermal cycling on the 1999; 27: 89-99.
bond strength of low-fusing porcelain to commer- 45) Leibrock A, Degenhart M, Behr M, Rosentritt M,
cially pure titanium and titanium-aluminum- Handel G. In vitro study of the effect of thermo- and
vanadium alloy. Dent Mater 2003; 19: 790-796. load-cycling on the bond strength of porcelain repair
29) Suansuwan N, Swain MV. Adhesion of porcelain systems. J Oral Rehabil 1999; 26: 130-137.

You might also like