CENTRA-R 200

ENGINeeRING DATA SHeeT

Product Overview The CENTRA R200 range of water purication systems has been specically designed to produce, distribute and maintain puried water for laboratory, medical and industrial applications. The CENTRA R200 process links up to ve purication technologies, Reverse Osmosis, adsorption, ionexchange, photo oxidation and 0.2 m particulate ltration and also incorporates a recirculation pump. The system is designed to operate from a good quality potable water supply, at 15C and produces up to 300 liters per hour of puried Type III reverse osmosis grade water up to Type I ultrapure water (18.2 M-cm) which is stored in its integral 350 liter reservoir. Puried water may be distributed around an installation under pressure to single or multiple points of use.
Left side view Front view
730.0 (28.74)

Right side view


891.0 (35.08)

460.5 (18.13) 416.0 (16.38) 372.5 (14.66) 329.0 (12.95)

746.0 (29.37)

573.5 (22.58)

Top view

Hydraulic Connections and Specications


Ref/Item A B C D E F G H Description Inlet feed water Ion Exchange inlet Ion Exchange outlet Reservoir drain (manual) Ringmain return Ringmain outlet Sanitization drain Concentrate drain

Details

CNXXXLDS_115 CNXXXLDS_230

BSP female thread BSP female thread BSP female thread BSP female thread BSP female thread BSP female thread BSP female thread BSP female thread

329.0 (12.95) 372.5 (14.66) 416.0 (16.38) 460.5 (18.13)

A B C D

E F G H

1820.0 (71.65)

CENTRA-R 200

ENGINeeRING DATA SHeeT

Treated Water Specications


Model Flowrate (exit from unit) l/min (USG/min) Daily usage typical Daily usage maximum Inorganics TOC typical Bacteria typical Particles
1

Voltage/freq. 230/50 115/60

Ion-exchange cylinder installed (Nuclear or Hypex grade). 2 System to be installed following ELGA LabWater installation guidelines and regularly sanitized.

R 200 16 (4.2) 18 (4.8) Up to 5000 liters (1321 USG) Up to 6000 liters (1585 USG) Up to 18.2M-cm 1 <10 ppb 2 <5 CFU/ml 2 0.2 m ltration

R 200 HFV 30 (8) 38 (10) Up to 5000 liters (1321 USG) Up to 6000 liters (1585 USG) Up to 18.2M-cm 1 <10 ppb 2 <5 CFU/ml 2 0.2 m ltration

R 200 HFR 30 (8) 38 (10) Up to 5000 liters (1321 USG) Up to 6000 liters (1585 USG) Up to 18.2M-cm 1 <10 ppb 2 <5 CFU/ml 2 0.2 m ltration

Dimensions and weights


Dimensions Shipping weight Operational weight Height 1820mm (71.7in), Width 730mm (28.75in), Depth 890mm (35in) 178kg (392lb) 527kg (1162lb)

Feedwater Requirements
Source

Potable tap water as detailed below.


Note: It is essential that the feedwater be suitably conditioned. Please refer to the contaminants listed below and ensure that suitable pretreatment is included in the installation.

Contaminant Calcium Total chlorine Silica Fouling Index Iron/manganese Organics Temperature Flowrate (maximum requirement @ 15OC) Drain requirements (gravity fall with air gap) Feedwater Pressure

Measure Ca ppm as CaCO3 Cl ppm SiO2 ppm FI Fe/Mn ppm TOC ppm C

Range <250 >250 <0.1 0.1 to 0.5 >0.5 <30 >30 <10 10 - 20 >20 <0.05 >0.05 >0.1 <2 2-3 >3

Pretreatment None 3 Softener or use very low RO recovery 3 None 20 inch carbon block Cylinder of carbon sized correctly to obtain <0.1ppm None 20 inch cartridge depth lter None 20 inch cartridge depth lter Back washable media lter with a minimum ow rate of 20 l/min None 20 inch cartridge depth lter 5 Back-washable pre-lter 5 None 20 inch carbon block 4 Cylinder of carbon sized correctly for TOC demand 4 1 - 40C (Recommended 15 - 25C) 20 l/min (5.3 GPM) 45 l/min (12 GPM)

Check LSI, increase frequency of acid cleaning. 4 Increase frequency of alkaline cleaning. 5 Increase frequency of acid cleaning.

4 bar (60 psi) maximum, 2 bar (30 psi) minimum.

Electrical Requirements
Mains input System control voltage Power consumption during recirculation Electrical protection rating Reservoir level connection for control of additional ELGA products Noise level during recirculation 230V ac, 50Hz, 115V ac, 60Hz 24V dc 2000VA 20 amps Jack Plug 3.5mm DIN plug 6 way <70dBA

ELGA LabWater Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505 Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. VWS (UK) Ltd. 2011 All rights reserved. ELGA and CENTRA are registered trademarks of VWS (UK) Ltd.