You are on page 1of 2

Standing Type STEAM STERILIZER

Vacuum Drying

HS-3460VD
THE MOST EFFECTIVE STEAM STERILIZER. VACUUM DRYING SYSTEM MINIMIZE THE WET PACK PROBLEM AFTER THE COMPLETION OF STERILIZATION.

HS-3460VD

HANSHIN 26

Vacuum Drying
Standing Type STEAM STERILIZER
HANSHIN MEDICAL

HS-3460VD
FEATURE
1. Five (5) types of cycle program that is composed of frequently using typical sterilization parameters are embedded in the controls of a sterilizer. Therefore the operator can select the needed cycle and change the programs with ease. In case that the liquid cycle would not exhaust the residual steam after sterilization, but cool down and condense into water naturally, consequently, the innovative method has been used which enables to prevent liquid from boiling up or breaking the bottle. In addition, the redrying cycle program is embedded for drying function only. 2. User doesn't need to open the door when drying phase due to its use of vacuum pump, hence moisture retained by the fabric package shall not exceed 2% than original weight after drying. 3. Steam generator is separated from the chamber that prevents any possible damage of sterilizing goods might be generated by the heater. Also it is effectively manageable that prevents any accidents might be occurred by the user's carelessness. 4. If door is not locked, the cycle would not start in any case. It has a door cover which enables to protect operator from thermal hazards and open/close mechanism which enables to operate the door with ease. The safety valve opens to protect the chamber from exceeding its rated pressure. 5. The front panel of the sterilizer has a display unit which shows the process of cycle such setting values as pressure, temperature, time and the phase such as ready, water fill, heating, sterilize, vent, vacuum drying and complete.
PRESSURE TEMPERATURE

CYCLE GRAPH

CYCLE No.1 UNWRAPPED

CYCLE No.2 WRAPPED

CYCLE No.3 UNWRAPPED

ALARM
When an error takes place,Erris displayed on the [TIME] and a number (1~6) correspond to the error is displayed on [TEMP.] with the audible signal by the self-diagnosis function of the microprocessor. Each of the error number means the following situation:
CYCLE No.4 WRAPPED CYCLE No.5 LIQUID

Error Code
Err Err Err Err Err Err 1 2 3 4 5 6

Problem
Low water level in reservoir Overheating of heater Door opens during cycle Deviation of exposure temperature Excess of water filling time Excess of heating time

CYCLE PROGRAM
CYCLE NO.
LOADS TYPE

Unwrapped Wrapped Unwrapped Wrapped Liquids 5 min 15 min 15 min 30 min 20 min 30 min 30 min 30 min 30 min No Exhaust

EXPOSURE TEMPERATURE 134.0 134.0 121.0 121.0 121.0 EXPOSURE TIME DRY TIME AIR REMOVAL

Fast Exhaust

SPECIFICATIONS
MODEL
OVERALL SIZE / CABINET CHAMBER RESERVOIR SHELF POWER SOURCE POWER CONSUMPTION STERILIZING AGENT CONTROLS EXPOSURE TEMPERATURE VACUUM PUMP DRYING ABILITY AIR FILTER PRESSURE GAUGE PRINTER NEW WEIGHT ENVIRONMENT CONDITIONS
340

HS-3460VD
W632 X H1,246 X D955 mm / W632 X H1,106 X D808 mm X D723 mm / SUS 316 Volume 60 ,Cylindrical Type

PRESSURE VESSEL : Druckbehlterverordnung Gruppe I Max. Working Pressure : 2.80 bar / Min. Working Pressure : -0.90 bar Max. Working Temperature : 138.0 / Test Pressure : 3.00 bar W238 X H233 X D117 mm, Water cap. 4 W236 X D635 mm, 1ea, W300 X D635 mm, 1ea AC230 V, 50/60 Hz 2,950 W Saturated Steam of max. 2.10 bar Pressure Microprocessor 121.0 ~ 134.0 Flow Rate : 40 / min Less than 2.0% of the amount of moisture within porous loads Min. filtration efficiency : 99.97% for 0.3 particles Dial Type, 760mmHg ~ 4 bar Optional : Model STP-103, Thermal Printing, 200 x 200 dpi, W58 x 60 mm Paper Roll 152 kg Altitudes up to 2,000 meters. Temperature range of +5 to +40 . Maximum relative humidity of 80 %. Main supply voltage fluctuation of 10 % of normal. Installation Category (Overvoltage Category) II, Pollution Degree 2.

27 HANSHIN