When the heart beats, a pressure wave moves out along the arteries at a few metres per second
(appreciably faster than the blood actually flows). This pressure wave can be felt at the wrist, but it also
causes an increase in the blood volume in the tissues, which can be detected by a plethysmograph. This
word comes from the Greek "plethysmos" for increase and is a term for a "fullness" (ie change in volume)
measuring device. Over the years, all sorts of HeathRobinson devices have been used but described
here is a photoelectric pulse plethysmograph, which is robust and easy to make and which will allow the
beating of the heart to be recorded without the need to make direct electrical connections to the body.
Equipment required
PC with PicoLog datalogging software(http://www.picotech.com/data-logging-software.html) installed
PC oscilloscope(http://www.picotech.com/oscilloscope.html) or data
logger(http://www.picotech.com/data.html)
Sensor
Amplifier
Experiment setup
Sensor
The sensor consists of a light source and photodetector; light is shone through the tissues and variation in blood volume alters
the amount of light falling on the detector. The source and detector can be mounted side by side to look at changes in reflected
light or on either side of a finger or earlobe to detect changes in transmitted light. The particular arrangement here uses a wooden
clothes peg to hold an infra red light emitting diode and a matched phototransistor (see figures 1 and 2). The infra red filter of the
phototransistor reduces interference from fluorescent lights, which have a large AC component in their output.
The peg is drilled with 3mm holes to take the led, the phototransistor, the
pair of wires linking the two and the twocore screened output cable. The
holes for the led and phototransistor are drilled in one go so that they line
up. The ends of each side of the peg are filed on the inside to enlarge the
gap and pieces of black closedcell foam (cannibalized from a mouse mat
and punched with 3mm holes) are stuck in place (Super Glue / Crazy Glue)
to improve grip and make a (more or less) lighttight seal against the skin.
At this point, the spring should be adjusted so that the peg will grip an ear
lobe while at the same time not being so tight that it excludes blood from a
finger. Pieces of stripboard glued to the peg are used to make connections
to the wires; two copper strips wide for the led (anode and cathode
connections) and three for the phototransistor side (collector, emitter and
LED cathode, and LED anode LED wires coming through the peg from
the other side). The lightemitting diode (Siemens SFH487) and the
phototransistor (Siemens SFH309FA) are wedged in their holes and
soldered to their respective pieces of strip board. Neither component is
critical and many other types will work. The wires are then soldered in
place; the screen of the connecting lead is soldered to the emitter and
cathode copper pad. Once everything has been checked and proved to
work, the connections and the backs of the components should be covered with a bead of an opaque silicone rubber caulk which
will insulate and keep out extraneous light. All bare wires should be covered.
Cart(http://www.picotech.com/cart?ACTION=viewCart)
Calculating the Heart Rate With a Pulse Plethysmograph
23/04/13 Calculating the heart rate - data logging experiment
www.picotech.com/experiments/calculating_heart_rate/ 2/2
Figure 2: Constructing the sensor
Amplifier
The amplifier (see figure 3) uses an LM358 dual op amp to provide two identical broadlytuned band pass stages with gains of
100. Again, the type of op amp is not particularly critical, as long as it will work at 6V and drive the output rail to rail. The signal
frequencies are boxed in by movement artefacts at the low end (generated by the peg moving and distorting the underlying
tissues; light pegs are better) and at the top end by mainshum interference. The circuit runs from a single 6 Volt battery and the
output zero is offset by about 1 Volt by referring everything to an internal common line at a voltage set by a pair of forward
biased silicon diodes. This is convenient for interfaces with a 05Volt input. The potentiometer allows the overall gain to be
adjusted so as to prevent clipping on large signals. Components are not critical but the two 2.2 F capacitors must be able to
stand some reverse bias so they should be nonpolarized or tantalum. The circuit can easily be made up on a small piece of strip
board.
Figure 3: Pulse plethysmograph amplifier circuit
Carrying out the experiment
Connect the output of the amplifier to the input of the interface and run PicoLog set to record 500 samples at 20 ms intervals.
Switch on and clip the peg to an earlobe or index finger; wait for about 10 seconds to allow the circuit to settle and then start
recording. Adjust the potentiometer so that the output is about 2 Volts peak, the trace should then look like that in figure 4, with
each pulse wave clearly visible. This was acquired using an ADC200 but with its capability of 100Ms/s it is overkill for this circuit,
which should work with DrDAQ, all of the PC oscilloscopes and all but the slowest voltage-input data loggers.
Two problems are common when using a pulse plethysmograph: movement will cause the trace to swing around wildly, so
persuade the subject not to move as much and if the subject is very cold (pale, pinched looking) the circulation at the extremities
may be reduced to the point where there is very little signal.
A useful feature of Picolog is that, if the alarm is set to trigger at about 1 Volt, there will be a beep in time with the heart beat.
Having proved that everything is working, the device can be used for such investigations as comparing individual resting heart
rates, following heart rate changes during and after exercise and looking at the changes in heart rate that occur as the subject
breaths in and out. For example, figure 5 shows 100 seconds of deep, slow breathing. Two effects, synchronised with ventilation,
are visible: variations in heart rate can be seen as changes in the spacing between the pulse waves and variations in the stroke
volume of the heart can be seen as changes in the amplitudes of the pulse waves.
Calculating Heart Rate
Unlike many heart rate monitors, this set up will allow the calculation of beatbybeat (instantaneous) heart rate. Using the
cursor, note the times of two adjacent peaks, subtract the first from the second to give the time between the two beats (beat to
beat interval) and express this in seconds. Divide this into 60 to obtain the instantaneous heart rate in beats per minute. Thus, for
the trace in figure 4, the first peak is at 1052 ms and the second is at 2190 ms. The difference is 1138 ms or 1.138 s, which gives a
rate of 52.7 beats per minute. The last peak is at 9206 ms, or 8.154 s after the first and as there are seven intervals between, so the
average interval is 1.165 s and the average rate is 51.5 beats per minute.
Credits, comments and further info
This experiment was written by Dr G J Compton of the School of Science, Nottingham Trent University, UK.