Product Information
Macro extensometer TC-EXMACRO
All data at ambient temperature. All rights reserved.
Description
Drive unit for sensor arms (required for use with temperature chamber)
Automatic L
0
adjustment (connection to CAN-BUS measurement module)
Order item TC-EXMACRO.001 TC-EXMACRO.002
TC-EXMACRO.H01 TC-EXMACRO.H02
Unit
Resolution
300mm sensor arms for ambient temperature 0.3 0.12 0.3 0.15 m
450mm sensor arms for small temperature chamber 0.45 0.18 0.45 0.225 m
600mm sensor arms for wide temperature chamber 0.6 0.24 0.6 0.3 m
Measurement range
300mm sensor arms for ambient temperature 80 80 75 (60)
1
75 (60)
1
mm
450mm sensor arms for small temperature chamber 120 120 112.5 (90)
1
112.5 (90)
1
mm
600mm sensor arms for wide temperature chamber 160 160 150 (120)
1
150 (120)
1
mm
L
0
Range 10 to 100 10 to 100 10 to 205 10 to 205 mm
Size
Width x Height x Depth 125 x 360 x 340 125 x 540 x 340 mm
Drag force < 0.05 < 0.05 < 0.05 < 0.05 N
Accuracy: Accuracy grade 1 according to EN ISO 9513
Here is required:
Measurement module: Incremental or
CAN-BUS and controller
2
For use with a temperature chamber: drive unit (see below)
Drive unit for sensor arms and L
0
adjustment
Order item
TC-EXMACRO.DU2/
TC-EXMACRO.DU3
TC-EXMACRO.DU4/
TC-EXMACRO.DU5
Reduction-in-width monitor
1
The value in brackets indicates the measurement path in the compression/flexure direction which differs only here from the one in tensile
direction
2
Only in connection with a materials testing machine of the Allround-Line with option TC-OPPLUS
Order item
TC-MOINCRC
TC-MOCANBU and TC-EXMACRO.CC1
TC-EXMACRO.DU2 or TC-EXMACRO.DU3
Order item
TC-EXMACWD.001
TC-EXMACWD.H01
TC-EXMACWD.002
TC-EXMACWD.H02
TC-EXMACWD.012
TC-EXMACWD.H12
TC-EXMACWD.003
TC-EXMACWD.H03
TC-EXMACWD.004
TC-EXMACWD.H04
3
Initial gauge length of the macro extensomter
Drive unit for reduction-in-width monitor with connection to CAN-BUS measurement module
and control unit
TC-EXMACWD.DU1/
TC-EXMACWD.DU2
Drive unit for reduction-in-width monitor
Meas. Clamping Specimen Meas. Init. gauge Resolution
lines pressure width thickness range length
(1
With 4 0.1 m
mechanical 1.5 ... 2 10...15/ 0.01 m
switching 2 N 20...25 0...4 4...9 50 0.1 m
for different mm mm mm mm 0.01 m
specimen 1 0.1 m
widths 0.01 m
Swivelled-out 4 0.1 m
up to 90 0.5 ... 0.7 10...25 0...4 6 10 0.01 m
1 N mm mm mm mm 0.1 m
0.01 m