Anritsu Master series OTDR
Presents:
The next generation in
OTDRs
The ACCESS Master
MT9083A / B / C
The Network Master
MT9090A
Slide 2
Anritsu Master series OTDR
OTDR 101
Before we start, lets review key
OTDR principles and terminology
2
Slide 3
Anritsu Master series OTDR
OTDR Operation
Block Diagram

Slide 4
Anritsu Master series OTDR
d = distance
t = elapsed time
c = speed of light in a vacuum
n = Index of Refraction
Test pulse fired

Test pulse returned
d
d =
t C
2 n
If n is incorrect, then
the distance measured
will also be wrong!!
Must divide by
two to get one
way elapsed time.

Distance Measurements
3
Slide 5
Anritsu Master series OTDR
Fiber Phenomena
Backscattering
Transmitted Light
Random scattering
Backscattering
Fresnel Reflection
Slide 6
Anritsu Master series OTDR
Fiber Type
Multimode Fiber
Wavelength: 850nm and 1300nm
Short distance application such as in-building Ethernet
connection
Graded-Index support higher data speed than Step-index
Singlemode Fiber
Wavelength: 1310 and 1625nm
High data speed over long distance application. Metro
Network, Long-haul network.
4
Slide 7
Anritsu Master series OTDR
Fiber Type
Slide 8
Anritsu Master series OTDR
Fiber Parameter
IOR (Index Of Refraction)
obtain from fiber manufacturer, enter into OTDR
Affecting the measured cable distance

Backscatter coefficient
obtain from fiber manufacturer, enter into OTDR
Affecting the measured reflectance.
5
Slide 9
Anritsu Master series OTDR
OTDR Basics
Typical OTDR trace with events
Slide 10
Anritsu Master series OTDR
OTDR Basics
Deadzones
Why event deadzone matters
Verification of two close reflections
Provides ability to verify which
connector is bad (high reflectance)
With 80 cm deadzone, MT9083A can
detect two connectors at 1 m apart.

Why backscatter deadzone matters
Verification of a second event after a
reflective event
Provides ability to test vault splices
Average MT9083A spec
is 5 m (15 ft)
EDZ
ADZ
Mechanical connectors
Mechanical connectors
Fusion
splice
6
Slide 11
Anritsu Master series OTDR
OTDR Basics
Event Deadzone

Connectors less than 80cm apart Connectors more than 1m apart
Slide 12
Anritsu Master series OTDR
OTDR Basics
Backscatter Deadzone

Connector Splice
5m backscatter deadzone
Undetectable splice

Connector Splice
5m backscatter deadzone
Detectable splice

7
Slide 13
Anritsu Master series OTDR
OTDR Basics
10m (33)
Actual Splice Location
Measured Splice Location
Actual End Location
Measured End Location
Low Density (wider Data Point Sampling)
Resolution

Slide 14
Anritsu Master series OTDR
OTDR Basics
3m (10)
Actual Splice Location
Measured Splice Location
Actual End Location
Measured End Location
High Density (narrow Data Point Sampling)
Resolution

8
Slide 15
Anritsu Master series OTDR
OTDR Basics
Dynamic Range: How far can you measure.
Dynamic Range OK to measure
entire fiber.
Insufficient Dynamic Range
events near the end of the fiber
cannot be measured.
Slide 16
Anritsu Master series OTDR
OTDR Basics
How far can our modules test?
Converting Dynamic Range into Distance
Can only be done on a straight piece of glass (no events)
Need to know the db/Km loss for each wavelength
Need to know the dynamic range at each wavelength.
Distance = Dynamic Range = 38dB/0.20dB/km = 190km
dB/km

Wavelength Typical Loss
850nm multimode 3.0dB/km
1300nm multimode 1.0 dB/km
1310nm single mode 0.35 dB/km
1550nm single mode 0.20 dB/km
1625nm single mode 0.25dB/km

9
Slide 17
Anritsu Master series OTDR
OTDR Basics
Pulse Width
The single most important parameter
Use the pulsewidth which still provides sufficient dynamic range
Too short
(short deadzones
but insufficient DR)
Too long
(good DR
but large
deadzones)
Just right
(good DR, short
deadzones)
Slide 18
Anritsu Master series OTDR
Pulse Width (PW), what does it effect,
Smaller PW = Smaller Dead Zones,
Able to see close events,
Not able to see longer distances.
OTDRs specification deadzone is tested at the smallest PW.
Larger PW = Larger Dynamic Range,
Able so large distances,
Not able to see close events.
OTDRs specification Dynamic Range is tested at the largest PW.
Its not possible to test close events with a large PW or large
distances with a small PW.
Using the Auto function on your OTDR sets the PW to the cable
length.
Pulse width Dead Zone Dynamic Range
10ns 2 - 3m 6 - 9 dB
100ns 10 - 15m 11 - 14dB
500ns 50m 18 21dB
1us 100m 22 25dB
4us 400m 29 - 32 dB
OTDR Basics
Pulse Width
10
Slide 19
Anritsu Master series OTDR
Presents:
The next generation in
OTDRs
The ACCESS Master
MT9083A / B / C
Slide 20
Anritsu Master series OTDR
Historical Fiber Optic System
Basic point to point deployment
OP-AMP/Regeneration
Stations

City A
City B

City A
City B
Splice points
Yesterdays
networks were
very simple and
most test
equipment was
adequate.
11
Slide 21
Anritsu Master series OTDR
But networks are evolving
Current Situation with Optical Networks

Increased complexity

Shift from long-haul to access installations

Increase in Metro and FTTx installations

Triple play (including sensitive video applications)

And

Budget reductions have forced service providers
to get the job done with less people, equipment,
knowledge and time

Shouldnt your test equipment evolve too?
Slide 22
Anritsu Master series OTDR
Introduce the all-in-one concept
Power Meter
and Light Source
High Performance OTDR
MT9083 Series
ACCESS Master
Visible LD
IP Test function
Less equipment to carry and maintain!
12
Slide 23
Anritsu Master series OTDR
Introducing the MT9083 seriesdesigned
with the features that matter most.
Dual mode design (HR/ER mode)
Compact
Lightweight
Fool-proof ease of use
All day battery operation
Quick boot-up
High resolution, but also good
range
Quick testing time
Reliable
Rugged
Complete testing tool
Inexpensive
Slide 24
Anritsu Master series OTDR
MT9083B/C: Dual Circuit Design
Standard, high resolution (HR)
Current operation
All pulsewidths
Use for best deadzone

Enhanced Range (ER)
Optimized for PON
Optimized for range
50ns-2us

User selectable
13
Slide 25
Anritsu Master series OTDR
MT9083B/C: FTTx/PON Measurement Using ER
In ER mode, MT9083B/C allows
user to measure entire span
Up to a 64/128-branched splitter!
Section A Section C
Section
B
MT9083B/C can detect
fault location beyond
splitter very accurately.
Slide 26
Anritsu Master series OTDR
Best Technical specs where it counts
Small Dead Zone to allow
measurement of near end
connectors.
Within CO
In riser cables
Premise

5cm Resolution
Event plotted accurately
Up to 150,001 sampling
points

dB Range also allows for over
100km measurement range.
Up to 45dB (1310/1550nm)
Most Telco long haul links
are from 70km to 100km.

14
Slide 27
Anritsu Master series OTDR
Best Technical specs where it counts
One touch Fault Locate mode,
Automatically sets all the Optical
parameters after a single touch
the user has the results.
Will offer a result with the
minimum required averages.


When in Standard OTDR mode,
By pushing the Start (Green)
button all optical parameters will
be configured.


Lost Test Mode,
Allow for simple interface to
Power Meter and Light Source
measurements.
High power meter option of
+30dBm
Slide 28
Anritsu Master series OTDR
Other Technical specs
Battery life
8hours of use battery life.

Boot-up time
Ready to use in 15sec
Engineers dont waste time on the
job.

Easy file transfer
Simple drag and drop file transfer via
USB.

Latest OTDR file format
SR-4731 (issue 2).

PC Emulation software available,
Multiple version of PC software
available included reporting software
downloadable from the web.

System upgrade software available from
the web.
Download the latest software from the
Anritsu web site.
Your OTDR is never out of date.
15
Slide 29
Anritsu Master series OTDR
High Performance, Field Tested Hardware
Dedicated keys
Setups
Save/recall
Contrast
Number pad
Rotary encoder
Precise marker placement
Easy menu navigation
START button
True one-button fault
location
All parameters selected
Perfect for novice users
Dedicated softkeys
Quick, easy operation
Parameters
Wavelength
Display modes
Rugged Protector Option
Side bumpers
Rigid display cover
Shoulder strap
Easy to read, TFT color
display
Standard display or enhanced
indoor/outdoor display
LRUD keypad (left,
right, up, down) for
easy operation
User Friendly
Rugged
Small
Lightweight
Easy to use

Impressive Battery
operation
8 hour typical
6 hour continuous
LiIon no memory
Slide 30
Anritsu Master series OTDR
Dedicated Testing Interfaces
Clearly labeled
Integrated dust caps
A true all-in-one for optical fiber
construction and maintenance
OTDR ports
Low cost, interchangeable
adapters
Built-in light source (std)
Integrated power meter (std)
SM/MM
Optional power meters
Interchangeable adapters
For loop-back testing
+3 to -70 dBm (SM/MM)
+23 to -50 dBm (SM)
+30 to 43dBm (SM)
Visible laser diode
option
~0 dBm output
2.5mm universal
~5 km (3 mile) range
CW or modulated
Dual USB ports
Plug & play
mass storage

IP testing Option
RJ45 electrical
10/100MB
1000MB upgrade
Ping, download test,
throughput, frame loss
16
Slide 31
Anritsu Master series OTDR
MT9083B/C: Waveform Comparison Using ER Mode
MT9083B/C can capture
waveform accurately using
short pulse for entire fiber
path through PON splitter.
Electrical noise
Bad
Bad
Bad
Good!
Competitor
A
Competitor
B
Competitor
C
Slide 32
Anritsu Master series OTDR
MT9083B/C: Waveform Comparison Using ER Mode
Competitor may be able to
see waveform beyond
splitter by using long
pulse. However
Zoom in
Very sharp waveform even
between close splitters
OK
?
Undefined waveform
due to poor dead zone
performance
Not
OK
OK!
Competitor
A
Competitor
B
17
Slide 33
Anritsu Master series OTDR
MT9083B/C: Longest Distance Using ER Mode
MT9083B/C provides
superior range (100km) at
100ns with excellent dead
zone performance.
100km
100km
100km
100km
(with 100ns
Pulse)
Competitor
A
Competitor
B
Competitor
C
Slide 34
Anritsu Master series OTDR
MT9083B/C: Best Dead Zone Using HR Mode
There are faults our competitors can not see.
MT9083B quickly identifies location of faults at the installation
and reduces risk of future trouble.
ANRITSU
15m
25m
30m
MT9083B/C is the only
OTDR that can detect a
fault in this zone.
Competitor
B
Competitor
C
18
Slide 35
Anritsu Master series OTDR
Access Master MT9083[ ] Lineup
Opt Wavelength MT9083A MT9083B MT9083C
053 1310/1550 42/41 45/45
055 1310/1550/1650 37.5/36/33.5 42/41/35
056 1310/1490/1550 36/34.5/34.5 40/39/39
057 1310/1550/1625 36/34.5/31.5 40/39/38 45/45/43
058 1310/1490/1550/1625 38/37/37/36
059 1310/1383/1550/1625 38/37/37/36
063 M850/1300/S1310/1550 28/27/38/36.5 29/28/42/41
064 MMF850/1300 28/27
073 1310/1550 38/36.5
A = 35dB class
B = 40dB class
C = 45dB class

Slide 36
Anritsu Master series OTDR
Easy to use, task-oriented top menu

Dedicated testing
modes simplify
operation
1. Select the testing
to be performed
2. The GUI is then
optimized to
perform it
3. Saves time,
increases
productivity
Upon boot-up or when
the Top Menu key is
pushed, the Top Menu
will appear. A testing
function to be performed
can be chosen
immediately.
19
Slide 37
Anritsu Master series OTDR
Fault Locate Mode for novice users
True one-button fault location

Quick test for cable damage
Perfect for CO fault location
Reduces MTTR
Ideal for inexperienced users
Launch condition verified
Test parameters automatically selected
Quick data average
Typically less than 10 seconds
Easy to read results for even the most
inexperienced user
Includes break/end location, span loss
Full trace view if desired

Slide 38
Anritsu Master series OTDR
Perform a Fault Locate test (1/2)
1. Clean and connect fiber
2. Highlight fault locate using the
arrow keys & press enter
3. Connection check immediately
verifies connection
1. Press continue to begin test if
status bar is black
2. Clean & reconnect if red
3. User can not start test if fiber is
connected incorrectly no
chance of incorrect results
4. Automatically test sequence begins
1. longest wavelength is selected
2. testing parameters selected
3. Fiber test begins
1
SM OTDR
MM OTDR
2
20
Slide 39
Anritsu Master series OTDR
Perform a Fault Locate test (2/2)
1. Results displayed in seconds
1. End/fault location
2. Event/loss
3. Total span loss
4. Full event table with end fault
highlighted
2. Select view trace for more
information (if desired)
1. Full event table
2. Full trace
3. Or connect next fiber and
press
Slide 40
Anritsu Master series OTDR
Standard OTDR Modefor any task
Flexible operation for any skill level or task

Customize the unit
Units of measure
File save format
File save location
Reflectance calculation
Complete event table control
Break/end threshold
Loss specification
Reflectance specification
End to end loss
Full functionality OTDR testing
Multiple trace comparison
Automated or manual parameter
selection your choice
Selectable display modes - from A, B,
Origin or Anywhere (zoom box)
21
Slide 41
Anritsu Master series OTDR
Run a quick test
Press f1 and
All parameters are automatically selected
+
=
Complete analysis location, type,
loss, pass/fail in seconds!
Slide 42
Anritsu Master series OTDR
Customize Test parameter(1/2)
From the top menu, select
Press f4
Select 1km (3.3kft), low density, 3 ns
Press
Press Averaging Time, enter 10 sec, then
22
Slide 43
Anritsu Master series OTDR
Customize Test parameter (2/2)
Press , then f1

Slide 44
Anritsu Master series OTDR
Real Time Test
Select 1550 or 1625 nm
Select 2.5 km, low density, 10 ns & press
Contract as shown using the arrow keys
Comment on the screen update rate (0.15 sec)
While still in real time, bend the fiber near the OTDR
The trace will drop out
Release the bend
Trace will restore instantly


23
Slide 45
Anritsu Master series OTDR
All fiber information is on a single screen no
need to flip through tabs
Battery status (%
and time remaining)
Current
testing mode
Axis scaling
A and B
measurement
cursors
Softkeys
Loss mode
type and
results
Complete event
table:
Event #
Location
Type
Loss
Reflectance
Segment dB/km
Cumulative loss
Key testing parameters:
Wavelength
Distance range
Pulsewidth
Index of refraction
Resolution
Number or averages
Slide 46
Anritsu Master series OTDR
Interpreting Event Table
1
2
3
24
Slide 47
Anritsu Master series OTDR
Construction OTDR Mode for cable installation
Automated testing of multiple fibers at multiple wavelength
Enter the information once saves times
Each fiber tested, named and saved automatically
Once the single page set-up is complete, the software becomes
the project manager
Identifies which fiber to connect
Verifies connection
Ensures consistency in filename and parameters
Eliminates user errors such as
missing files
Inconsistent naming
Inconsistent parameter selection
Slide 48
Anritsu Master series OTDR
Perform a Construction Mode Test (1/2)

1. Select file storage location
2. Determine base file order
(example shown in filename
field)
3. Enter test site locations
4. Use other field for cable
ID, project, customer, etc.
5. Select wavelengths
6. Select testing options
7. Enter number of fibers to
test
1
2
3
4
5
6
7
25
Slide 49
Anritsu Master series OTDR
Perform a Construction Mode Test (2/2)
Once set-up is complete, connect the
fiber as instructed and press
continue
Connection level is verified
Optimize and press continue if
bar is black
Clean and reconnect if red
Sit back and watch up to four
wavelengths tested, analyzed and
saved without pressing a button
Connect next fiber and press
continue
Slide 50
Anritsu Master series OTDR
Software Overview
Set-ups always available from front panel 5
Neatly arranged into categories
Notes explain each setting
Simply one-button
software upgrades via
USB!
26
Slide 51
Anritsu Master series OTDR
Loss Test Set Modefor optical budget verification
Stabilized light source standard
More cost effective than buying stand-alone handheld
sources
CW, 1kHz or 2kHz output modes
Universal connector easily adapts to all connector
styles
same port as the OTDR
High (-8 dBm) output provides exceptional dynamic
range for testing even the longest fibers
(-8 dBm output) - (-65 dB sensitivity) = 57 dB of range
Integrated power meter standard
-5 to -50 dBm
High power CATV range (+20 to 55 dBm) optional
Data table for easy storage of results
Slide 52
Anritsu Master series OTDR
Perform Optical Loss Test Set (OLTS) function
Select same wavelength for
Power Meter and Light source
Set Reference
Connect patch cord between power
meter and light source through an
adapter.
Turn on the light source and record
the power in the Reference field
Connect patch cord to cable
under test
Measure loss


27
Slide 53
Anritsu Master series OTDR
Reference Procedure for Loss Measurement
Use two short patch cable
for referencing


Connect the short patch
cable to both end of the fiber
under test.


Light Source
Access Master in
Power Meter Mode
Slide 54
Anritsu Master series OTDR
USB connection
Can be used on PCs with USB
port
Fiberscope for x200/400
magnification
Adaptors for 1.25 & 2.5mm
patch cord.
Universe connectors including
APC
Image saving function


Video Inspection Probe (VIP)
28
Slide 55
Anritsu Master series OTDR
Visible Fault Locator Option
Integrated visible red light for
troubleshooting within the OTDR dead
zone

Provides easy identification of a
particular fiber

Select output mode: CW or 2 Hz
modulated

Universal 2.5 mm connector fits all
standard size connectors

0dBm (1mW) max output
~5km (3 mile) range
Slide 56
Anritsu Master series OTDR
Dedicated field test solution for installation and
maintenance of optical links



OTDR Tester.
29
Slide 57
Anritsu Master series OTDR
Introducing the MT9090A with
MU909014x/15x OTDR Module
A truly revolutionary OTDR!
Pocket-sized
Easy Operation
Excellent OTDR Performance
Unique Battery Operation
Priced Right
Slide 58
Anritsu Master series OTDR
Overview.
Product overview,
Operation, Performance and Options,
Specs,
Comparison to MT9083x,





30
Slide 59
Anritsu Master series OTDR
Value without compromise.
OCA
(Optical Channel
Analyzer for CWDM)
DCFL
(Drop Cable Fault Locator)
GiGE
(Gigabit Ethernet test)
Part of an ever-
expanding platform!
???
Modular platform ensures
maximum return on investment!
OTDR
Slide 60
Anritsu Master series OTDR
For Highly Portable Field Use.
Easily held with one hand,
Palm-size (190mm x 96mm x 48mm),
Approximately 700g (1.54 lbs).
Battery operation,
Up to 8 hour operation
(with STD Battery pack),
4 hour recharge
(with STD Battery pack),
NiMH battery pack or AA Alkaline Dry Battery
operation.
Rugged/sealed design,
No vents or fans,
Wrist strap to prevent accidental drops.
Color indoor/outdoor display standard,
4.3 high resolution,
Landscape format naturally fits OTDR trace.
31
Slide 61
Anritsu Master series OTDR
The Foolproof testing device.
Easy to use,
True one-button fault location,
Automatic testing parameters selection in
Full-automatic mode (Full-manual
operation also available),
Quick fiber evaluation,
Event analysis table and full Trace view.

Quick boot-up,
Ready to test in <15 seconds.
Truly
Easy!!
Slide 62
Anritsu Master series OTDR
Simple
One-button
Operation.
Push to Turn on power
Push to test
Self testing
ok
Self Test
Automatically moves to
OTDR Menu from Top menu
(Can be bypassed )
Automatically checks fiber connection,
proceeds if the result is good
(With Poor or Fair result check
connection or start test manually)
During testing,
Trace (waveform) is displayed
Upon completion, the Trace analysis
(Event table) is displayed
If you want to view the trace then;
- Select F4 key
Or
- Select an event then select Autozoom
User selectable test parameters
(or select full automode)
32
Slide 63
Anritsu Master series OTDR
Complete Optical Fiber Testing.
Excellent OTDR Performance,
Capable of,
Full parameter selection (5ns-20us),
125,000 data points,
Up to 37dB Dynamic range,
High resolution accuracy,
Dual wave 1310/1550nm config,
Complete testing from Access to Core.
Complete PON testing tool,
High resolution and extremely short
dead zones ensure entire PON (1X32, 1X64 branches)
network evaluation,
ITU-T L.41 compliant - 1625nm/1650nm
wavelength for in-service maintenance
of PONs without external filters.
Slide 64
Anritsu Master series OTDR
1:32 Splitter
2km Fiber
4km Fiber
2.2km Fiber
Complete PON Measurement for I&M.
33
Slide 65
Anritsu Master series OTDR
Long Battery Life,
Up to 8 hours / 4 hour recharge,
with Standard NiMH Battery pack (Supplied with the original
unit),
Alkaline battery (AA) operation also available.
Batteries.
Available 4 x AA Type
NiMH
Or
Alkaline Dry Batteries
Slide 66
Anritsu Master series OTDR
Additional Highlights.
Built-in 10m (30ft) launch fiber,
Measure front-end connection
without additional patchcords,
Can be enabled or disabled.

Complete data management,
Up to 1,000 traces internal/10,000+ with USB,
Easy drag and drop transfers to PC,
Reporting and printing with NETWORKS or
TraceView PC software.

34
Slide 67
Anritsu Master series OTDR
Optional Features.
Visual Fault Locator Option,
Integrated 650nm laser,
Can also be used to identify a
particular fiber (in the splicing tray),
Can be used simultaneously with power meter or OTDR,
Up to 5km (3 mile) range.

Connector Inspection Option,
400x image displayed,
All software pre-loaded,
Eye-safe (Often an OH&S requirement),
Image save/recall,
Hardware includes probe/tips.
Slide 68
Anritsu Master series OTDR
Different Applications.
FTTx,
Full network testing: ONT to OLT (incl. splitter(s)),
In service maintenance (filters installed).
General fiber
installation/certification,
Premise/CO,
Access,
Metro,
Core/Long Haul.
Backhaul,
FTTc (fiber to the cell site),
DAS,
RRH.
Optical Fiber
(when used Ethernet)
Base
Terminal
Station :
NodeB
Antenna
Optical
Fiber
B
T
S

R
R
H

Remote
Radio
Head
Optical Drop Cable
Fault Locator,
OTDR
Optical Fiber Test
35
Slide 69
Anritsu Master series OTDR
MU909015B/B1-056/066 MU909014B/B1-056/066 MU909014A/A1-053/063 MU909014A/A1-054/064
1310/1550 20 nm * 1310/1550 20 nm * 1625 15 nm 1650 15 nm
PW=20sec 37/36dB 32.5/31dB 32.5dB *
(1490/1550 cut filter included)
32.5dB *
(1490/1550 cut filter included)
PW=500nsec 28/26dB 24.5/23dB 24.5dB *
(1490/1550 cut filter included)
24dB*
(1490/1550 cut filter included)
Modulation: CW, 1Hz
* Please see detailed on Product brochure
Fiber Type 10m/125m SMF (ITU-T G.652)
Loss Measurement mode 2 point loss, Splice loss, dB/km Loss LSA, ORL, Event
Environment Operating temperature and humidity: -5 to +40C*10, <80%(no condensation)
Size (WxHxD, including mainframe and module): 190mm x 96mm x 48mm (7.5" x 3.8" x 1.9")
Weight (including mainframe, module and Standard battery): <700g (1.54 lbs.)
Automatic, Displayed in table format based on user defined PASS/FAIL thresholds
User Selectable (English, Simplified Chinese, Traditional Chinese, Danish, Finish, French, German, Italian,
Japanese, Korean, Polish, Portuguese, Russian, Spanish, Spanish (L.America) and Swedish)
Connection check: Automatic check of OTDR to FUT connection quality
Operating time (Standard Battery): 8 hours (typ.) *
Battery Type: NiMH (Standard battery), NiMH (AA Type), Alkaline Dry Battery (AA Type) *
Language
Other Functions
<125,001 pts (Course: <6,251 pts, Medium: <25,001 pts, Fine: <125,001 pts) Number of Sampling Points *
Full automatic, Manual, Real time, Optical Power Meter,
Visible Fault Locator (Option), Video Inspection Prove (Option)
Integrated launch fiber: 10m (30ft)
Testing Modes
Live Fiber detect (Option 056/066 only): Verifies presence of communication light in fiber
Fresnel: <1 m, Backscatter: <5m
5ns/10ns/20ns/50ns/100ns/200ns/500ns/1s/2s/5s/10s/20s
MT9090A CONTROLLER SPECIFICATIONS
MU909014X/15X OTDR MODULE SPECIFICATIONS
Model name
Center Wavelength *
Dead zone *
Pulse width
Dynamic range *
Dimension and Mass
Visible Laser Diode
(only for MU909014A1/B1/15B1)
Fiber Event Analysis
Output Power: 0 3dBm (CW)
Not applicable Optical Power meter
(same port as OTDR port)
Wavelength: 1310/1490/1550nm
Measurement Range: -50 to -5dBm
Accuracy: 0.5dB (at -20dBm, CW)
Real time sweep: <1sec (Typ.)
Specifications.
Slide 70
Anritsu Master series OTDR
MT9090A/MU909011A
Drop Cable/Premise Fault Locator
Customer Presentation
36
Slide 71
Anritsu Master series OTDR
MU909011A Drop Cable Fault Locator Module
Slide 72
Anritsu Master series OTDR
PURPOSE-BUILT FOR TESTING SHORT FIBERS
Unique only true short-fiber tester available
Perfect solution for FTTX drop cables, CO,
MDU risers, wireless towers & premise
Integrated 10m launch cord standard
Dynamic range of >7dB (10dB typ)
For testing up to 10km (~6 miles)
Unit will display transmission power levels,
then fault locate with the press of a single
button
Testing parameters are fixed
Operation simplified
Ensures proper settings and valid results
658 feet!
37
Slide 73
Anritsu Master series OTDR
Additional Features
Integrated power meter option for FTTx
maintenance
For measuring power to ONT
Connect to MT9090A/MU909011A
Verify power level and classify
PASS fiber OK, ONT fault
FAIL fiber damaged, launch fault locate

Power meter and fault locate from same port

Complete breakdown as to failure and responsibility
ONT, drop cable or feed fiber

General purpose power meter
-5 to -45dB range
Slide 74
Anritsu Master series OTDR
74
MT9090A Specifications
MT9090A CONTROLLER SPECIFICATIONS
Display 4.3 inches TFT Color LCD (480 x 272 pixels, transmissive)
Interface USB (v1.1)
Battery type NiMH
Battery life time 3.5 hours (typ)
Battery charging time <3 hours
Size (WxHxD) 7.5x3.8x1.9 (190mm x 96mm x 48mm)
Weight 2 lbs. (~800g)
Power supply 9VDC
100 to 240 VAC, 50/60Hz
Standard Accessories Includes controller, battery pack, AC charger, users guide, soft carry case
MU909011A DROP CABLE FAULT LOCATOR MODULE SPECIFICATIONS
Center Wavelength 1550 30 nm
Fiber Type 10/125um SMF (ITU-T G.652)
Optical Connector APC or UPC (factory option) with SC, FC, LC, ST adapters
Pulse width <15 ns
Dynamic Range 7 dB (10 dB typical)
Deadzone (Fresnel) <1 m
Deadzone (Backscatter) <5 m
Sampling 5 cm
Distance range 1.0km (~3,000ft), 2.5km (~8,000ft) or 10.0km (~32,000ft) set automatically

Very high
resolution for
complete
short fiber
analysis
38
Slide 75
Anritsu Master series OTDR
CMA5000a OTDR
Product Introduction
Slide 76
Anritsu Master series OTDR
CMA5000a
The all-in-one, portable CMA5000a supports network
transmission and optical physical layer tests for
installation and maintenance.
Transmission
SDH/SONET
Ethernet

Optical physical layer
Fiber end test
Loss test
OTDR
Chromatic dispersion
Polarization mode dispersion
Spectrum analysis


39
Slide 77
Anritsu Master series OTDR
Modular Structure
One mainframe supports full range of versatile
measurements by changing modules

Transmission systems (modules)
SDH/SONET (UTA, XTA)
Ethernet (UTA, GigE II)
Optical physical layer (modules)
Fiber end face check (VIP)
Loss test (OTDR)
Optical Time Domain
Reflectometer (OTDR)
Chromatic dispersion
(CD-OTDR)
Polarization mode dispersion (PMD)
Spectrum analyzer (OSA)

Slide 78
Anritsu Master series OTDR
Mainframe Types and Modules
Two mainframes models allowing greater flexibility
SBA: Two bays up for up two modules
MBA: Four bays for up to four modules
Multiple measurement modules can be
set and operated simultaneously in
one mainframe.
1 bay used
GigE II
OTDR
CD-OTDR
SBA
2 bays used
UTA
XTA
Bays empty
USB Connection
VIP
MBA
4 bays used
OSA
PMD
40
Slide 79
Anritsu Master series OTDR
CMA 5000a OTDR Available Modules






Model
Number
Wavelengths Dynamic Range
5261 QUAD SM/MM:
850/1300nm (62.5um)
/1310/1550nm
24/26dB
35/35dB
5262 QUAD SM/MM:
850/1300nm (62.5um)
/1310/1550nm
24/26dB
40/40dB
5266 850/1300nm 62.5um 24/26dB
5269 850/1300nm 50um 24/26dB
5283 1310/1383/1550/1625nm 35dB
5491 1310/1490/1550nm 40/40/40dB
5493 1310/1490/1550/1625nm 37dB
Model
Number
Wavelengths Dynamic Range
5225 1310/1550nm 37/36dB
5235 1310/1550nm 40/40dB
5236 1310/1550/1625nm 40/40/40dB
5245 1310/1550nm 43/45dB
5246 1310/1550/1625nm 43/45/43dB
5254 1550nm 50dB
5260 QUAD SM/MM:
850/1300nm (50um)
/1310/1550nm
24/26dB
35/35dB
Slide 80
Anritsu Master series OTDR
Key Features and Benefits
Unparalleled data sampling - Up to
256,000 data points
Twice as many as the nearest
competitor!
Unparalleled waveform creation
Ultra-fast real time screen update - 0.1
sec
Great for optimizing
splices/connectors
Gives perception of fast testing
Fully configurable to your customers
needs
Over 20 different optics choices
single mode or multimode, 25 to
50dB, dual, tri- or quad wavelengths
State of the art platform
Powerful PC, all common I/O
Active product continued
development
Designed for simplicity and ease of use
Dedicated testing modes no macros
One button testing (AUTO-TEST)
Minimal set-up time, maximum testing time
A strength for all skill levels
Superior trace linearity 0.04 dB/dB
Provides greater measurement accuracy
and consistency
Overlay up to 8 traces simultaneous
Compare between wavelengths to see
stresses effecting the cable
Compare between fibers in a cable to see
inconsistencies

41
Slide 81
Anritsu Master series OTDR
Testing Modes
Fault Locate
One Button
Automated Testing
Standard
Traditional OTDR
Testing
Construction
Automated Multi-
Wavelength Multi-
Fiber Testing
Loss Test Set
Integrated Power
Meter and Light
Source
Visual Fault Locator
Visible Red Light
Video Inspection Probe
Connector
inspection
microscope
Slide 82
Anritsu Master series OTDR
Fault Locate Mode One Button Automated Testing
One-button break/end locator
Quick test for cable damage
Ideal testing mode for
inexperienced users
Launch condition verified
Test parameters automatically
selected
Data averaged just long enough
to produce valid results
Easy to read results for even the
most inexperienced user
Includes break/end location,
span loss and distance past
prior event
42
Slide 83
Anritsu Master series OTDR
Standard OTDR Mode Traditional OTDR Testing
Full functionality OTDR testing
Multiple trace comparison
Automated or manual parameter
selection your choice
Selectable display modes - from A, B,
Origin or Anywhere (zoom box)
Customize the unit
Units of measure
File save format
File save location
Reflectance calculation
Complete event table control
Break/end threshold
Loss specification
Reflectance specification
End to end loss
Slide 84
Anritsu Master series OTDR
Construct OTDR Mode
Automated Multi-Wavelength Testing
Automated testing of multiple fiber
cables
Test multiple fibers at multiple
wavelength quickly and easily
Enter the information once and
test/save each fiber with a single key
press
Ensures consistency in filename and
parameters
Combines the best of CMA 4000
Construction and Multiple
Wavelength modes
43
Slide 85
Anritsu Master series OTDR
Loss Test Set Mode
Optional Power Meter and Light Source
Adds a stabilized mode to the OTDR
laser for point to point testing
More cost effective than buying
stand-alone handheld sources

CW, 1kHz or 2kHz output modes

Universal connector easily adapts to
all connector styles
Uses the same UC-130-XX
adapters as the OTDR

High (-8 dBm) output provides
exceptional dynamic range for
testing even the longest fibers
(8 dBm output) (65 dB
sensitivity) = 57 dB of range

High power CATV range
+20 to 55 dBm
Data table for easy storage of results
Slide 86
Anritsu Master series OTDR
Visual Fault Locator Mode
Optional Red Light Test
Visible red light for
troubleshooting within
the OTDR dead zone

Provides easy
identification of a
particular fiber

Select output mode: CW
or 2 KHz modulated

Universal 2.5 mm
connector fits all
standard size
connectors

0dBm (1mW) max output
~5km (3 mile) range
44
Slide 87
Anritsu Master series OTDR
NetWorks Data Processing Utility
The ultimate data processing tool
Integrated software utility for viewing,
printing and analyzing OTDR data
Analyze data on any PC
Complete report generating tool
Uni- and Bi-directional splice loss
Exception
Fiber Acceptance
Trace Summary
Batch processing for high count fiber
cables
Full analysis and Bellcore/Telcordia file
format processing
Anritsu Master series OTDR
Slide 88
Anritsu Master series OTDR
The New Anritsu
Why Choose Us?
Now includes Nettest
The portable OTDR market share leader
Over 30 years in the optical industry
Over 25 years OTDR experience
Stable and debt-free: your partner for the
long haul
Focused on core competency not a
jack of all trades
Solutions and support that enable you to
be more competitive and profitable
Superior sales and technical support:
anytime, anywhere
Anritsu Master series OTDR
45
Slide 89
Anritsu Master series OTDR
The right productplus the support you deserve

Easy and always FREE software upgrades
Simply, one time registration
https://www1.anritsu.co.jp/Download/MService/Login.asp

Dedicated, knowledgeable technical support
Via our worldwide network.
Email : ACSG.techsupport@anritsu.com

Local service centers
Via our worldwide network.
Email : ACSG.service@anritsu.com

Anritsu Master series OTDR
Slide 90
Anritsu Master series OTDR
Anritsu Master series OTDR