TN-011-2, Dec00.
Calibration Procedure
A hands-on guide for SenseAir IR gas sensors
ASSOCIATED DOCUMENTS:
SCOPE:
INTRODUCTION
SenseAir infrared (IR) technology for gas sensing is
well proven, stable, and reliable. All sensors are
tested and fully calibrated at the factory before
delivery, where each individual sensor, based on its
serial number, has a data file of its own with all
relevant parameters stored.
TN-010
TN-001
ieF0005
ieVT-UIP
Theory of Calibration
Function test of sensor/controllers
model family 2001VT
Service bag model F0005
(Users manual)
User interface program for VT
(Users manual)
Calibration Procedure
1 of 3
TN-011
Technical Note
1.
2.
3.
4.
Diagram 1:
This test demonstrates that even
very poorly calibrated sensors will automatically
adjust themselves as times goes by. Therefore,
verification tests shall not be performed directly
after installation.
1 ,4
1 ,2
1
0 ,8
0 ,6
0 ,4
0 ,2
13
50
:2
1:
23
0:
:2
15
15
00
15
:2
0:
38
15
:2
0:
16
15
:1
9:
54
15
:1
9:
33
15
:1
8:
09
:1
8:
47
8:
:1
15
15
23
15
:1
7:
02
15
:1
7:
38
15
:1
7:
14
15
:1
6:
50
15
:1
6:
28
15
:1
5:
06
15
:1
5:
45
15
:1
5:
23
15
:1
4:
01
15
:1
4:
37
15
:1
4:
16
:1
3:
54
3:
:1
15
15
32
15
:1
2:
10
15
:1
2:
49
15
:1
2:
27
15
:1
1:
05
15
:1
1:
44
:1
1:
22
0:
:1
15
15
00
15
:1
0:
38
0:
15
:1
9:
:0
15
15
:0
9:
15
tim e
Diagram 2: A ZERO and SPAN test gas sequence illustrating the sensor response time at 0.2 liters/minute
flow rate, as well as the back diffusion rates seen after the test gas valves were closed.
Calibration Procedure
2 of 3
TN-011
Technical Note
PRECAUTIONS
When applying gas to the sensor there are
issues one has to consider pressure build-up
back diffusion from surrounding air via
membrane covered ventilation holes at the
bottom (not visible).
two
and
the
cell
Figure 2:
Prepare a 2 mm inner diameter
plastic tube for the test gas by A) cutting the edge
to be connected to the sample cell at 45 degrees,
and B) cut away the resulting top edge. This will
make the tube fit nicely on the sample cell test gas
inlet nipple. C) cut open a small hole about 1 mm2
that will prevent an over pressure build-up.
L-type sensor
80
60
35 ppm
40
18 ppm
20
9 ppm
0
0
1000
2000
3000
100
300
78 ppm
200
39 ppm
19 ppm
100
0
0
0,5
1,5
Diagram 3:
Correlation curves for zero
constant errors 1%, 0.5%, and 0.25%, with the
respective zero point error indicated to the left.
Calibration Procedure
3 of 3