Professional Documents
Culture Documents
Scan area calibration is done with the x and y selection buttons. An appropriate sample such as a
TEM grid should be loaded. A calibrated image of the sample from another source should be
available in printed form for comparison.
X-calibration
Select the x-scan button and start image acquisition. Adjust offset to match the left side of the
live preview image to the same position as the calibrated image. Adjust gain to match the
horizontal view.
Y-calibration
Select the y-scan button and start image acquisition. Adjust offset to match the top of the live
preview image to the same position as the calibrated image. Adjust gain to match the vertical
view.
Halt acquisition and save settings to disk. A floppy will be required for this operation. Exit the
image setup program, start AutoBeam, and collect a larger image (512 or larger) to check the
calibration.
Gray calibration
This uses the same SEM setup as the PCI brightness calibration, i.e., TEM grid on carbon, Tetra
detector, waveform just inside the minimum and maximum range.
Select the Image-1 button and start image acquisition. Adjust the offset so that a small number of
the dark pixels are colored blue as being in the darkest range. Adjust the gain so that a small
number of the light pixels appear reddish pink as being in the lightest range.
Halt acquisition and save settings to disk. A floppy will be required for this operation. Exit the
image setup program, start AutoBeam, and collect a larger image (512 or larger) to check the
calibration. Click the histogram area measurement button to display a histogram of gray scales.
Proper calibration should result in two peaks in the histogram, one near each end of the gray
scale. Neither peak should be clipped by the end of the gray scale.
Calibrations were performed on.
December 15, 1999
The spatial calibrations of ____ were still correct. A 3.05-mm TEM grid was measured as
2.96x2.98 mm on Polaroid film. It was measured as 2.97/3.00 mm by PCI, and it was measured
as 2.97x3.01 mm by AutoBeam. This represents less than a 2% error at 30x magnification.
The grayscale calibration was in error and required adjusting both PCI and AutoBeam systems to
match. Polaroid Type 53 film requires setting the f-stop almost one stop larger than indicated to
produce a decent gray scale image.
December 16, 1999
The AutoBeam setup was de-calibrated to match the calibration in effect during April 1999 when
images were collected for the DOTs 92-97 project comparison. At first, only the Y calibration
appeared to need changing. It was matched to the first images collected April __. The Y view
matched at 100%, but the X-view was 104% of what was viewed earlier. (This was later
corrected and used for collection of the first batch of maps.)
Vertical span
Horizontal span
(% of original)
(% of original)
4-6-99
100
104
4-20-99
92
96
4-21-99
87
92
5-3-99
88
91
It appears that the same calibration was in effect on 4-21 and 5-3. There appears to have two
changes in calibration. One sometime after the start and before 4-20, and one between the runs of
4-20 and 4-21. Since most SEMs do not guarantee accuracy closer than 5%, so the changes
between 4-20 and 4-21 may be negligible and might be due to a steady drift. This yet needs to be
confirmed.
Parameter
9-99
4-99 match
X-gain
1322
1280
X-offset
-1359
-1388
Y gain
1669
1910
Y-offset
-1531
-1669
SE gain
256
231
SE offset
-719
-478