Professional Documents
Culture Documents
The DC Power Analyzer provides an easy way for an R&D or design validation engin
eer to recreate some of the
automotive power waveforms such as slow decreasing/increasing of operating volta
ge, quick charges, cranking
profiles and voltage dips all from the front panel. While the design has been o
ptimized for use on the bench,
the N6705A DC Power Analyzer is also an LXI Class C instrument with LAN, GPIB an
d USB interfaces.
There are over twenty different power modules ranging in performance (basic, hi
gh-performance and precision)
and power (50 W, 100 W and 300 W).
There are arbitrary waveform controls built into the N6705A that allow users to
create nine different
waveforms: sine, step, pulse, ramp, trapezoid, staircase, exponential, user def
ined voltage and user
defined current waveforms. These waveforms are all configurable from the front
panel without having
to write a single line of code!
The N6705A also allows users to save user-defined waveform setups and scope dat
a to the 64 MB of internal
memory or to an external USB memory device.
-Biasing Multiple Input Voltage Devices in R&D
During R&D and design validation stages, some tests begin with multiple, seque
nced voltages to power the
device under test (DUT). For example, an ATX PC motherboard requires a specific
power-on sequence to power
the board. It is important to power the board properly to avoid sub-assemblies
drawing excessive current.
An incorrect power start-up sequence may cause damage to integrated circuits. I
n turn, this can compromise
the reliability and quality of the circuit board. R&D engineers must generate
the correct power-up sequence
to the board to prevent harming circuits and causing additional problems.
This test requires multiple power outputs for the 12 V, 5 V and 3.3 V lines wit
h precise timing accuracy
within the millisecond range. In addition, these power outputs need to have pro
grammable slew rates to
simulate the rate of change for the specified timing conditions.
The Agilent N6705A DC Power Analyzer can precisely and repeatedly synchronize
outputs.
The N6705A has built-in output delay controls that are accessible from the fron
t panel.
As shown in Figure 2a, users can program output on and off delays by setting th
e time delay
before the output turns on and off. This delay is applied after the All Outputs
On or Off key is pressed.
Users can enter delays from 0 ms to 1023 ms in 1 ms increments.
In addition to having multiple programmable output delays, the N6705A has progr
ammable slew rates which
control the voltage rate of change. Slew rates can be programmed as slow as 4.7
6 V/s.
The maximum slew rate is limited by the up and down programming time of the mod
ule and the load created
by the DUT. As a best case scenario, the N6751A and N6752A modules have a maxim
um up-programming slew
rate of 50 kV/s with a full resistive load (10% to 90% of total voltage). Each
module has its own
controllable slew rate.
-Simulating Power Interruptions for DC Input Devices
Simulating power interrupts for DC input devices is important to verify the desi
gn of the device under test
(DUT). Interruptions originate from disturbances on the power line such as sudde
n power dropouts or voltage
sags. DC devices have speci cations that describe how wide of an interrupt the de
vice can tolerate.
The purpose of the power interrupt test is to
nd the point at which the device f
ails by controlling the
length of the interrupt pulse.
The N6705A allows users to program pulse waveforms and other arbitrary waveforms
without having to write a
single line of code. From the example described earlier, users can con gure the dr
op out pulse on the 5 V
line directly from the front panel.
Beyond sourcing arbitrary waveforms, the N6705A is also capable of providing use
ful measurements.
Users can characterize the device under test by measuring the voltage and curren
t sourced into the DUT.
The N6705A measures and displays current waveforms without the need for a curren
t transducer such as a
current shunt or probe. The N6705A also has a built-in datalogger which enables
you to capture voltage
and current waveforms for seconds, minutes, hours, or even days.
-Avoid DUT Damage by Sequencing Multiple Power Inputs Off Upon a Fault Event
There are many testing applications in which a device under test (DUT) is powere
d by multiple DC input
voltages and the DUT is sensitive to the order in which those multiple power sou
rces turn off.
For example, individual assemblies used in satellites are especially susceptible
to damage during
uncontrolled multiple power source turn-off events, and these assemblies are ver
y costly.
One unexpected cause of power source turn-off occurs when a fault condition such
as an over-voltage
or over-current condition is detected on one of the sources that causes it to sh
ut down. Having the
ability to control the power supply shut-down sequence built into the power supp
General Differences
There are a number of differences between the low-profile and dc power analyzer
mainframes.
The low-profile mainframe is flat and optimal for Automated Test Equipment (ATE)
in manufacturing
and design validation systems where instruments are rack mounted and space is at
a premium.
It has a robust command set and is typically controlled by a computer program th
at sends commands
it against expect
to correlate with
device when it is
device performa
The N6781A has a current measurement only mode that sets the power supply to behav
e like a zero-ohm shunt.
When its output is connected in series with the battery and the batterypowered d
evice, as in Figure 1,
the N6781A emulates a zeroburden ammeter. The voltage is regulated where the rem
ote sense lines are connected.
In Figure 2, you can see a picture of an actual setup. After you con gure the set
up, select Current Measure
Only as the emulating mode in the 14585A source settings screen, as shown in Fig
ure 3. The voltage priority
mode is set by default and the +/- current limits are set to their maximum allow
able value. The output is
programmed to zero volts, hence zero-burden.
The N6781A SMU power module has an auxiliary voltage metering input that can be
used to measure battery
voltage in battery drain applications. Measurements from the auxiliary voltmeter
are used to validate
battery run-time and performance. Refer to DVM+ and DVM- connections in Figure 1
for setup con guration
of the auxiliary voltmeter. To enable auxiliary voltage measurements using the 1
4585A, expand the Instrument
Control tab and select Meter then Properties.
Using the Scope Mode
There is a scope mode in 14585A control and analysis software. The mode allows y
ou to
monitor the current drain as it happens. In the scope mode, the graphical user i
nterface allows you to
control the choices of voltage/current measurements, measurement ranges, and the
scaling of the display much
like an oscilloscope. You can also adjust the number of sample points per trace
to a maximum of up to 256 K
points for a single trace.
These features are also available from the front panel using Scope View.
The scope acquisition can be started or stopped by pressing the circular button
on the lower right-hand
corner under the word Scope. You can pull
up markers to further analyze the trace, as shown in Figure 5. Marker can be use
d to narrow in on a
specific time interval and extract information speci c to that time interval such
as min, max, and average
value.
Using the Data Logging Mode
The data logging mode in the N6705B and 14585A software permit long-term data co
llection. The settings allow
you to define the duration and the period of the data logging. The duration can
be up to several hours or
even days at a time. Set it to run longer than the expected actual battery rundo
wn duration to ensure that
you capture the entire event. You can define the integration period for the data
logging. For each
integration period, a set of min, max, and average values is generated and logge
d. In the data log display,
the min, max, and average values are all plotted together in the same graph. Fig
ure 6 shows a data log plot
of an actual battery rundown. These features are also available from the front p
anel using Data Logger.
Markers are available to use with data logging. Turn on the vertical measurement
markers and place them at
the start and shut-down points. This will set up the software to base all of its
numerical calculations just
over the enclosed time interval rather than the entire display.
You can set the data logging function to measure current as well as the voltage
reading from the auxiliary
DVM. Once the data is captured, it is stored in a binary format. There is an opt
ion to export the binary file
into a CSV (comma-separated variable)
le format. A CCDF (Complimentary cumulativ
e distribution function)
is another display feature. It is a cumulative form of a histogram that provides
a concise display of shortand long-term battery drain measurement. It is a distribution plot of the curren
t amplitude versus its
relative frequency of occurrence.