Digital Peak Holder

EN294-1240-G

Forwards
Thank you very much for your purchasing Minebeas Digital Peak Holder CSD819C.
This manual explains installation procedures and connecting method and also operating
method for the Digital Peak Holder CSD819C. Make use of it properly after reading
through the manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the
manual at hand after reading it over.
This manual is intended for the technical experts to read.

The contents of the manual may subject to change for improvement without notice.
I

Marks and arrangements used in this manual


The following marks are attached to the explanation on the matters that indicate Dont do
this., Take care. and For reference.
Be sure to read these items where these marks are attached.

Warning

Warning may cause injury or accident that may harm to the operator.
Dont do these things described here.

Caution during operation and working.


Be sure to read the item to prevent malfunction.

Mark during operation.


Press the switch.

II

For safe operation


Be sure to read this instruction manual before use.
1. Installation place

Use the instrument where the temperature/humidity specifies with the range as
follows

Environmental temperature
Environmental humidity

10 to 50
Less than 85 %R.H. (Non condensing)

(1) Location where installation is not allowed.

Warning

Dont locate the instrument on the places as follows :


It may cause an unexpected faulty in the instrument.

D Dont locate the instrument in direct and/or high temperature area.


D Dont use the instrument in a high humid area.
D Dont install the instrument where there are vibrations and shocks.
D Dont use the instrument where there is excess of dusts and fine particles.
D Dont use the instrument where there are corrosive gas and salt and like
that.
D Dont install the instrument where there is rapid change of temperature
and humidity.
D Dont install the instrument near the devices that are magnetized or
generate an electromagnetic field.
D Dont install the instrument where the instrument may be affected by
radioactivity or radial rays.
D Avoid the location where chemical reaction may take place such as in a
laboratory, or like that.

III

(2) Installation

When installing the instrument, install as referring to the following figures and
secure the space around the instrument.

Each dimensions of the instrument and required dimensions for the environmental
spaces are as follows:
Outline dimensions
Side

Front

Rear

Upper

Panel cut size

Unitmm

IV

2. Power supply

Warning

Be sure to check that the power supply is off in connecting each cable. If the work is
done while the power is on, there may have the case that electric shock to the
operator or even may have damage to the instrument.

Before supplying the power, check that the indication of power supply voltage
/specifications for the instrument and the power going to supply should be the
same. If they are not equal, contact with Minebea.
If you use the instrument without checking them, it may cause a damage in the
instrument or electric shock to the operator.

Earth wire should be grounded securely.


When earth wire is not connected, it may cause a malfunction of the instrument or
electric shock to the operator.

3. Application note

Warning

Warning
Warning
Warning
Warning

Before using a new instrument or exchanging the strain gage applied transducer
for a new one, be sure to make calibration. If calibration will not be made, the
correct measuring results may not be obtained nor which may cause malfunction in
the instrument and there may exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the similar case
when the results are not correct, so make calibration, again.

In case of using the instrument, check that the connections are executed properly.
If not connected properly, the correct measuring result will not be obtained, nor it
may cause malfunctions of the instrument, damage to the peripheral equipments or
even more serious accidents.

When change of setting is made carelessly on the instrument during measurement,


correct measured results may not be obtained and it may cause malfunction in the
instrument and even have the possibility of damage in peripheral instruments.

Do not shock the instrument such as throwing something on it.


If neglected, it may cause destruction of the parts and damage to the electrical
circuits.

Do not push the panel sheet on the instrument with the excessive strong force nor
push it with sharp edge object such as a driver.
If neglected, it may cause a damage to the panel switch and even have the
possibility of damage to resist to environments or operational performances.

Warning

Dont remove the cover of the case of the instrument, nor peel off the panel sheet
nor take the instrument into pieces.
If neglected, it may cause a damage to the case and the panel sheet and even have
the possibility of damage to resist to environments or operational performances.

At the time of shipment from the factory, the instrument has been plated with a
clear sheet on the panel sheet for protective purpose.
In case of application, use the instrument after removing the clearsheet first.

4. CE conformity standard
This instrument has suited the following standard.
EN6132612006
Electrical equipment for measurement, control, and laboratory use EMC requirements
Immunity test requirements for equipment intended for use in

industrial locations
EN610101 2001
Safety requirements for electrical equipment for measurement, control and laboratory use
Part 1 : General requirement

Warning

Please observe the following conditions strictly when this instrument suits the
abovementioned standard.
If neglected, there is a possibility of not suiting the abovementioned standard.

(1). Wiring
1 Shield processing

Cables other than power cable must use all shielded cables.
Please connect all shielded cables with F.G. terminal on No.2 terminal block.
2 Grounding

The ground of this instrument shall apply the individual ground by using the protective ground
terminal

VI

History of revision
Date
July. 2005

Instruction Manual No.

Details of revised point

DRW.NO.EN2941240

First Version ROM Ver. 1.000 or later

DRW.NO.EN2941240A

Due to ECN No.FN0602119


Addition
82.F51:00002=Display interlock
935.(2).Reading out the display interlocked data
945.(1).Reading out the display interlocked data

June 2009

DRW.NO.EN2941240B

Due to ECN.FN0902035
Changed CE conformity standard EN613261:2006
Addition
Immunity test requirements for equipment intended for
use in industrial locations

Feb. 2010

DRW.NO.EN2941240C

Due to ECN.FN1002026
Change
Front covers logo is changed.

April 2007

May, 2010

DRW.NO.EN2941240D

Due to ECN.FN1001019A
Addition
34. [Applicable environment]is added.
1110.Accessories Panel monting gasket is added.

DRW.NO.EN2941240E

Due to ECN.FN1002140
Change
Minebea logo is changed

Jan. 2011

DRW.NO.EN2941240F

Due to ECN NO.FN1102018


Correction
934. DE9SN (JAE)
to DE9SNR by JAE or equivalent.

May 2012

DRW.NO.EN2941240G

Due to ECN.FN1002140D
Change
Minebea logo is changed

Oct. 2010

VII

Index
Forwards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Marks and arrangements used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


For safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. Installation place . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3. Application note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. CE conformity standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(1). Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
History of revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11. Features

..........................................................

2. Name and function of each point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

21. Front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


22. Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2
3

3. Installation procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

31.
32.
33.
34.

Installation place . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Location where installation is not allowed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Applicable environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4
4
5
6

4. Connecting method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

41. Layout of the terminal block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


42. Note on connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
43. Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
431. Connection with strain gage applied transducers . . . . . . . . . . . . . . . . . .
432. Connection with external control inputs . . . . . . . . . . . . . . . . . . . . . . . . . .
433. Connection with open collector outputs . . . . . . . . . . . . . . . . . . . . . . . . . . .
434. Connection with the power supply and the earth . . . . . . . . . . . . . . . . . . .
435. Connection with analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7
8
9
9
14
15
16
17

5. Calibration procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

51. Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

52. Calibration procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


521. Calibration method to register the output of strain gage applied transducer
at the time of maximum display after setting the loa . . . . . . . . . . . . . . .
522. Calibration procedures to register the output of strain gage applied transducer
at the time of zero and the maximum display . . . . . . . . . . . . . . . . . . . . . .
523. Zero fine adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
524. Span fine adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
525. Calibration procedure to apply registration again for zero point only
526. Compensation of symmetric property
(Fine adjustment of minus side span) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
53. Selection of calibration methods on each condition . . . . . . . . . . . . . . . . . . . . . .
531. In case of executing the calibration on the instrument newly. . . . . . . .
532. When the calibration is executed again. . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

37
39
39
42

54. Setting the prohibition against calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


55. Changeover of calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

42
43

6. Operation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

44

20
25
31
33
35

61.
key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
611. Operated in the Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

62.
key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
621. Operated in the Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
622. Operation is made in another mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
63.
631. Operated in the Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
632. Operation is made in another mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

64.
key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
641. When operated in the Measurement mode. . . . . . . . . . . . . . . . . . . . . . . . .
642. When operated in another modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

65.
key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
651. When operated in the Measurement mode. . . . . . . . . . . . . . . . . . . . . . . . .

66.
key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
661. When operated in the Measurement mode . . . . . . . . . . . . . . . . . . . . . . . .
662. When the A/Z function is selected. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
67.
671. When operated in the Measurement mode. . . . . . . . . . . . . . . . . . . . . . . . .
68.

key

44
44
44
44
44
44
44
45
45
45
45
45
45
46
46
46
46
46

............................................................

46

7. Function and operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

47

71. External control input signal and open collector output signal . . . . . . . . . . . .
711. Input signal for external control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
712. Open collector output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
713. Equivalent circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
72. Comparator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
721. ON/OFF for the Comparator S0, S1, S2, S3 and S4. . . . . . . . . . . . . . . . .
722. Changeover of comparator code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
723. Change of set value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
724. Operation on comparator S1, S2, S3 and S4 . . . . . . . . . . . . . . . . . . . . . . .
725. Comparative target for comparator S1, S2, S3 and S4 . . . . . . . . . . . . . .
726. Operation of comparator SO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
727. Hysteresis on comparator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
728. Light on/off setting of comparator value display section . . . . . . . . . . . .
73. How to use the filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
731. Analog filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
732. Digital filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

47
47
48
48
49
49
49
50
51
52
53
56
57
58
58
58

74. Selection of A/D sampling rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


75. How to use various kinds of peak holds. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
751. Related function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
752. Peak hold function by all zone specification . . . . . . . . . . . . . . . . . . . . . . .
753. Peak hold function by zone specified . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
754. Peak holding function by specifying time zone . . . . . . . . . . . . . . . . . . . . .
755. Peak holding function by range specification at automatic start time
756. Setting of detecting condition of the local maximal value
and the local minimal value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
757. Detection condition setting in inflection point . . . . . . . . . . . . . . . . . . . . .

58
59
59
60
69
78
85
91
92

76. Various kinds of functions related with the display . . . . . . . . . . . . . . . . . . . . . .


761. Selection of display rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
762. Selection of position of decimal point display . . . . . . . . . . . . . . . . . . . . . .
763. Load display range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

93
93
93
93

77. Selection the target for HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


78. Changeover of bridge power supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
79. Zero set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
710. Key lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
711. CHECK value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
712. How to use the analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7121. Scaling of analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7122. Fine adjustment 1 on analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7123. Fine adjustment 2 on analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7124. Selection of the target of analog output . . . . . . . . . . . . . . . . . . . . . . . . . . .
713. Memory location for setting data and so on . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

93
93
94
94
94
95
95
97
99
102
103

714. Prohibition of calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


715. Check mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7151. Operating procedure for the check mode . . . . . . . . . . . . . . . . . . . . . . . . . .

103
104
104

716. Monitor mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

110

8. Function mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

112

81. Setting method for function mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


82. Function of function data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

112
114

9. Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

122

91. Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


911. Related functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
912. Specifications for the current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

122
122
122

92. BCD output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


921. Related function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
922. Specifications of BCD output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
923. Pin configurations for the BCD output connector . . . . . . . . . . . . . . . . . .
924. Equivalent circuit for input/output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
925. Timing chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
926. Output condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
927. Selection of output logic for P.C.(Print command), and of its width . . .
93. RS232C interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
931. Related function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
932. Specifications for interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
933. Procedures of data transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
934. Pin configurations for connector pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
935. Data format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
936. Communication error processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

123
123
123
124
124
125
126
126
127
127
127
128
129
130
135

94. RS422/485 interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


941. Related functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
942. Specifications on interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
943. Procedure of data transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
944. Pin layout and wiring of Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
945. Data format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
946. Process of communication error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

136
136
137
137
138
140
144

10. Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

145

101. Execute trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

146

102. Optional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


103. Error display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

155
161

11. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

162

111.
112.
113.
114.

Specifications for analog section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


Specifications for digital section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Front panel sheet key function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
External control function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

162
162
163
163

115.
116.
117.
118.
119.
1110.

Comparator function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Open collector output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Various kinds of functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Standard specifications at the shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

164
164
164
165
165
165

1111. Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11111. Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11112. BCD output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11113. RS232C interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11114. RS422/485 interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1112. Outline dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

165
165
166
166
167
168

12. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

169

121. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
122. Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

169
169

13. Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

170

131. Replacement of fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


132. Characters pattern for display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

170
172

133. Setting table for functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

173

1. General
The instrument is a digital peak holder for the application of strain gage applied transducer.

11. Features
Main features for CSD819C are as follows :
(1) Compact size and light weight
96 mm96 mm130 mm Approx.500 g

(Without any options)

(2) 24 kinds of peak hold modes can be selectable.


The peak mode can be used by selecting from Peak hold, Bottom hold, Peak/bottom hold, The
local minimal value hold, The local maximal value hold, Inflecting point hold.
(3) 4 kinds of Calibration data and Comparator code are memorized, and can be used by
changeover of them.
(4) Nonlinearity

0.025 %F.S.

2. Name and function of each point


21. Front panel

Load display section


The load data is shown in the Measurement mode.
The condition is shown in various kinds of Calibration mode.

Comparison value selection display section


The selecting number(S0S4) of comparison value is displayed.

Comparison value setting display section


The comparison value selected by is displayed.
The condition is displayed in various kind of Setting mode.

Judgement display section (S0S4)


The comparison result (S0S4) by the comparator function is displayed.

Condition display section


SEL.1SEL.2 Lights on when the shortcircuit is made between SEL.1, SEL.2 and
COM.1 of external control input.
HOLD
Lights on when the shortcircuit is made between HOLD and COM.1 of
external control input.
CHECK
PEAK
MEAS.
END

key.
Lights on when CHECK is ON by pressing
Lights on when any peak mode is selected.
Lights on when any peak mode measurement is proceeded.
Lights on when any peak mode measurement is completed.

key
Used when shifting to the Function mode.

key
Executes zero set. (Onetouched zero point adjustment)

key
Used when the changing mode of S0, S1, S2, S3 and S4 for comparison value selecting number
is called, and used for the numeric carryingover at the time of various kind of setting.
2

key
Used for the changeover of comparison value selecting display, and the numeric increment at
the time of various kind of setting.

10

key
Used for the changeover of peak and track.

11

key
Used for reset of hold value.

12

key
Used for registering set values at the time of various kinds of settings.

13

key
Used when ON/OFF of CHECK value is required.

14

Position of pasting the Unit seal


As necessity requires, paste the Unit seal attached.

22. Rear panel

Terminal block 1
Connects with external control input, open collector output, various kinds of strain gage
applied transducer such as load cell and analog output.

Terminal block 2
Connects with AC power supply, shielded wire and a grounding wire. (When it is out of CE
conformity standard.)

Installing space for options


Whichever one can be installed from the optional BCDOUT, RS232C and RS422/485.
When there is no option, the cover is attached instead.

Protective ground terminal


Connect the grounding wire when the instrument suits to CE conformity standard.
Dont connect except the grounding wire.

3. Installation procedures
31. Installation place

Use the instrument where the temperature/humidity specifies within the range as
follows:
Environmental temperature
Environmental humidity

10 to 50
85 %RH or less(Non condensing.)

32. Location where installation is not allowed.

Warning

Dont locate the instrument on the places such as follows:


It may cause an unexpected faulty in the instrument.

D Do not expose the instrument in direct sunlight and/or high temperature area.
D Do not use the instrument in a high humid area.
D Do not install the instrument where there is high mechanical vibrations and shock.
D Do not use the instrument where there are excess of dusts and fine particles.
D Do not install the instrument where there include any corrosive gas or any salty
atmosphere.
D Do not install the instrument where there is rapid change of temperature and humidity.
D Do not install the instrument near the devices that are magnetized or generate an
electromagnetic field.
D Do not install the instrument where there may suffer radioactivity or radioactive rays.
D Avoid the location where chemical reaction may take place such as in a laboratory, or like
that.

33. Installation

When installing the instrument, install as the following figures and secure the
space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental spaces are as
follows:
Outline dimensions
Side

Front

Rear

Upper

Panel cut size

Unitmm

34. Applicable environment


Warning: The instrument may subject to use in a highly humid area or in full of powder dust.
In such a case, use the instrument by inserting the panel mount gasket attached between the control panel (cabinet) and the main body.

By inserting the panel mount gasket, the front panel section becomes IP65 (International Protection Code) or equivalent in dustproof and waterproof construction.

4. Connecting method
41. Layout of the terminal block
There are 2 pieces of terminal block and M4 thread, one has 21 points of terminals and the other
has 3 points of terminals.
Layout of terminal blocks and thread are shown in the following figure.
1

Terminal block 1 (21P)


Terminal
Nos.

Descriptions

AOUT

AOUT

ZERO

HOLD

10

PEAK/
TRACK

Strain gage
applied
t
transducer
d

Analog output

External control
i
input

Terminal
Nos.

Descriptions

11

RESET

12

SEL.1

13

SEL.2

14

COM.1

15

END

16

S0

17

S1

18

S2

19

S3

20

S4

21

COM.2

Applications

External control
input

Open
p collector
output

Terminal block 2 (3P)


Name
AC
AC
F.G.

Applications

Application
Power supply
Frame ground
(Functional ground)

Protective ground terminal (M4 thread)


Name

Application
Protective ground

The COM.1(Terminal No.14) and COM.2(Terminal No.21) are isolated.

42. Note on connection

Warning

In case of connection with the instrument, keep strictly to the


following items. If neglected, it may cause an unexpected failure
or a damage to the instrument.

D Please go after turning off the power supply without fail when you connect the wires.
D Since the terminal blocks at rear side of the instrument is made of resin, take care not to
drop it down or not to apply strong impact.
D Recommended torque to tighten the terminal screws for terminal block is as follows
Terminal block

Torque to tighten the terminal screws

Terminal block 1

0.6 Nm

Terminal block 2

1.4 Nm

Terminal block for


protective ground

1.4 Nm

D The suitable crimp type terminal lugs for the terminal block are as follows:
Terminal block

Width of crimp
type terminal lugs

Terminal block 1

6.2 mm or less

1.253 or Y type 1.253.5

Terminal block 2

9.0 mm or less

1.254, 24 or Y type 1.254, 24

Suitable crimp type terminal lugs

Terminal block for


protective ground
D Please separate the cable connected with this unit from the noise sources such as power
supply line and/or I/O line for control as much as possible.
D Please use the exclusive conduit wiring on this instrument, and avoid using with another
line together.
D Please execute all connecting wires surely according to this book.

43. Connection
431. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so the
connection with another type of strain gage applied transducers is proceeded in the same way.

1 When tension is applied with the application of tension type or universal


(compression/tension) type of load cell, and display of + direction is required,
connect Green with Terminal No.2 and Blue with Terminal No.4
individually. As there is a case which standard wiring color is different, please
confirm the inspection data sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may have the case
that the accuracy is out of warranty because the resistance of cable makes the
input voltage of the instrument decreased.

1
WHT

BLU

RED

Shield

Connection with 1 piece of load cell and CSD819C

GRN

3m

When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

1 When tension is applied with the application of tension type or universal


(compression/tension) type of load cell, and display of + direction is required,
connect Green with Terminal No.2 and Blue with Terminal No.4
individually. As there is a case which standard wiring color is different, please
confirm the inspection data sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may have the case
that the accuracy is out of warranty because the resistance of cable makes the
input voltage of the instrument decreased.
2

Connection with 1 piece of load cell and Junction box for extension use(B304) and
CSD819C

GRN

1
WHT

BLU

B304
RED

3m

Shield

Junction box

CAB502

2 (Total length of CAB502 is within 30 m)

Internal wiring diagram of B304


to load cell
RED BLU WHT GRN YEL(Shield)

Suitable crimp type terminal lugs


1.254 or 24
YEL(Shield)

Terminal pitch 9.5 mm


RED BLU WHT GRN

to CSD819C

When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

10

1 When tension is applied with the application of tension type or universal


(compression/tension) type of load cell, and display of + direction is required,
connect Green with Terminal No.2 and Blue with Terminal No.4
individually. As there is a case which standard wiring color is different, please
confirm the inspection data sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may have the case
that the accuracy is out of warranty because the resistance of cable makes the
input voltage of the instrument decreased.
3

Connection with 2 to 4 pieces of load cells, Summing type junction box(B307) and CSD819C

CAB502

2 (Total length of CAB502 is within 30 m)

11

GRN

WHT

B307

BLU

RED

Shield

Junction box

Internal wiring diagram of B307


WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN

1W
1R
1G
1B
1Y
RD
OR
BK
WH
GN
BL
YE

2W
2R
2G
2B
2Y
3W
3R
3G
3B
3Y

BLU

4W
4R
4G
4B

YEL(Shield)

4Y

RED

WHT
GRN
BLU

to CSD819C

YEL(Shield)

When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

12

1 When tension is applied with the application of tension type or


universal(compression/tension) type of load cell, and display of + direction is
required, connect Green with Terminal No.2 and Blue with Terminal No.4
individually. As there is a case which standard wiring color is different, please
confirm the inspection data sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may have the case
that the accuracy is out of warranty because the resistance of cable makes the
input voltage of the instrument decreased.
4

Connection with 2 to 4 pieces of load cells, Summing type junction box(SB310) and

GRN

1
WHT

BLU

RED

Shield

CSD819C

SB310

CAB502

2 (Total length of CAB502 is within 30 m)


Internal wiring diagram of SB310
Short bar

Output port
Input port No.1

to CSD819C
RD

RD

WH GR BL YE

RD

WH

WH GR BL YE

Input port No.2

GR

Input port
No.4

BL
YE

RD

WHGR BL YE

Input port No.3

or
Y type

WH GR BL YE

B
A

Suitable crimp type


terminal lugs

B
A

RD

A3.2 mm or less
B7 mm or less

Round type

When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).
13

432. Connection with external control inputs


The connections with ZERO,HOLD, PEAK/TRACK, RESET, SEL.1 and SEL.2 of
external control input is connected between each terminal and COM.1 at terminal No.14 with
the contact point or the open collector as shown in the figure below.
Refer to the paragraph 71. for the function of each input.

or

Shield

Warning

Connect the external control input surely as shown in the above figure.
If neglected, it may cause an unexpected failure and/or malfunction to the
instrument.

For the connections with external control inputs, be sure to apply shielded cable,
and the shielded cable is connected with E terminal of the instrument. (Terminals
No.5) If not connected, it may cause malfunction due to the effects from external
noises and so on.
When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

14

433. Connection with open collector outputs


Connections with END, S0, S1, S2, S3 and S4 of open collector outputs and
external loading is connected between each terminal and COM.2 of terminal No. 21 as shown
in the below figure.
At the same time, take care that the load will not exceed the value of specified open collector
output.
Specified open collector output VCEDC30 V, IC30 mA MAX.

Shield
Load
(Please prepare an external power
supply for the load separately. )
Surge preventive element

Warning

Connections with open collector outputs is made securely according to the figures
and also within the specified open collector output. If neglected, it may cause an
unexpected failure and/or malfunction to the instrument.
For the protection of the open collector output of the instrument, connect the surge
preventive element that satisfies the characteristics of external load to connect. If
neglected, it may cause unexpected failure and/or malfunction on the open collector
output of this instrument.

For the connections with open collector outputs, be sure to apply shielded cable,
and the shielded cable is connected with E terminal of the instrument. (Terminal
No.5) If not connected, it may cause malfunction due to the effects from external
noises and so on.
When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

15

434. Connection with the power supply and the earth


Connections with the power supply and the earth is made as the following figure.
The D class with single earth is required for grounding.
Power supply voltage
Frequency for power supply
Power consumption

AC100 V to AC240 V
(Permissible variation rangeAC85 V to AC264 V)
50/60 Hz
Approx.15 VA at maximum. (at AC100 V)

D class with single earth

Warning

AC100 V to AC240 V
(Permissible variation range
AC85 V to AC264 V)

Connections with the power supply and the earth is made securely according to the
figures and also within the rated capacity of the instrument. If neglected, it may
cause an unexpected cause of failure.

The D class with single earth is required for grounding.


If neglected, it may cause an unexpected malfunction due to the effects of noise
from other equipments.
When this instrument suits with the CE conformity standard, please make the
single earth with a protective ground terminal.

16

435. Connection with analog outputs


(1) In case of standard instrument (voltage output)
The instrument prepares output voltage for analog outputs as a standard.
Please connect the voltage output as shown in the figure below.
Voltage output
DC10 V to 10 V
Overrange
At OL display
Approx.11 V
At OL display
Approx.11 V
Load resistance
5 k or more

V
Shield

Warning

Please make the connection of voltage output surely as showing above, and use its
connection in the specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to the
instrument.

For the connections with voltage outputs, be sure to apply shielded cable, and the
shielded cable is connected with the E terminal of the instrument (Terminal No.5).
If not connected, it may cause malfunction due to the effects from external noises
and so on.
When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

17

(2) In case of option (current output)


Please connect the electric current output as shown in the figure below.
Current output
DC4 mA to 20 mA
Overrange
At OL display
Approx.2.4 mA
At OL display
Approx.21.6 mA
Load resistance
260 or less
A
Shield

Warning

Please make the connection of current output surely as showing above, and also use
its connection in the specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to the
instrument.

For the connections with current outputs, be sure to apply shielded cable, and the
shielded cable is connected with the E terminal of the instrument (Terminal No.5).
If not connected, it may cause malfunction due to the effects from external noises
and so on.
When this instrument suits with the CE conformity standard, please connect the
shielded wire with F.G. terminal (Terminal block 2 Terminal No.3).

18

5. Calibration procedures

Warning

Before using the new instrument or after exchanging the strain gage applied
transducer with a new one, be sure to make calibration.
If calibration is not made, correct measured results may not be obtained, or it may
cause malfunction to the instrument and it may damage the peripheral equipment.
Moreover, even if calibration has made, there may occur the similar case as above
when the result is not correct. So make precise calibration again.
The calibration for the instrument and Display value at the time of minimum
analog output(F21) and Display value at the time of maximum analog output
(F22) are not interlocked. In due course, make check on the setting for F21 and
F22 securely. If neglected, correct outputs may not be obtained, or it may cause
malfunction to the instrument and it may damage the peripheral equipment.

51. Preparations
According to the Chapter 4. Connecting method, connect the instrument and the strain gage
applied transducer properly, then supply the power.

52. Calibration procedures


Load calibration procedures for the instrument are as follows
1

Calibration method to register the output (conversion with mV/V) of strain gage applied
transducer at the time of maximum display (weighing capacity) after setting the load to zero
(Initial load condition with tare weight).

Calibration method (Actual load calibration) to register by the reading output of strain gage
applied transducer, when setting in the condition of zero load applied (Initial load application
with tare) and in the condition of actual load applied individually.

Fine adjustment on Zero

Fine adjustment on Span

Calibration procedures to apply registration again for zero point only(Tare weight
cancellation).

Compensation of symmetric property(Fine adjustment of minus side span)

The accuracy of calibration obtained from is 1/1 000 or so. If the accuracy more
than 1/1 000 is required, make calibration of type.
In the following paragraphs, we will describe each calibration procedure by showing the
examples with load cell applied.

19

521. Calibration method to register the output of strain gage applied transducer at the time
of maximum display after setting the load to zero.

Warning

Before using a new instrument or exchanging the strain gage applied transducer
for a new one, be sure to make calibration.
If calibration will not be made, correct measured results may not be obtained nor it
may cause malfunction in the instrument and there may exist damage to the
peripheral equipment.
Besides, even though the calibration has been made, there may occur the similar
case when the result is not correct, so make calibration again.
During the calibration is executing, be sure to make cancellation (Execution of
F98) for compensated data on zero set
When the changeover target of the calibration data is set to the changeover by
external control input (Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During calibration procedures, press the


key in case of interrupting the
calibration is required. The calibration data will be kept as they are before entering
the calibration and then returns to the Measurement mode.

key is pressed with the setting display section of FUNC, the


display will change as the following arrow marks. However, every time the key

Every time the

is pressed, the display will change as the reverse direction of the following arrow
marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)

Procedures

Press the
key for approx. one second.
The setting display section shows FUNC.
The load display value continuously shows the
present load value.

Press the
key once.
The setting display section shows CCAL.

20

Procedures

Press the
key.
CCAL mode can be entered, then the load
display section shows SCAL.

Press the
key.
The load display section shows D01.
When the calibration has completed already, the
set value of minimum scale registered at that
time will be displayed.
Set the minimum scale with the right keys.
Setting value for the minimum scale are 4 (four)
kind as follows:
1, 2, 5, 10

key Set value increment key

Press the
key.
The load display section shows DISP.

21

Procedures

Press the
key.
The load display section shows 2000, and the
minimum display digit flashes on and off.
When the calibration has completed already, the
maximum display value registered at that time is
displayed.

By the setting of minimum scale, the


minimum display digit that flashes on and
off is as follows
The minimum scale
1, 2, 5 100 digit
The minimum scale
10
101 digit

key Set value carry key


key Set value increment key
key Set value initialization key

Set the maximum display value with the right


keys. Setting range for the maximum display
value is as follows:
In order to make effective use of the performance,
set within the following ranges.
Setting range for the
maximum display value The minimum scale
10010 000

20020 000

50050 000

1 00099 990

10

Press the
key.
The load display section shows S MV.

22

Procedures

Press the
key.
The load display section shows 0.5000, and the
digit of 100 flashes on and off.
In case that calibration has completed already,
the registered output value of load cell at that
time will be displayed. After the ACAL mode is
over, the load display section shows the present
load.
Set the given value with the right keys, which is
subtracted the output value of load cell at the
time of initial load application from the output
value of load cell corresponding to the maximum
display value set in the step 6.

Though the number of digits has not


8

prepared in the Inspection data for load


cell so many as shown in the right figure,
extra digits are necessary for the
compensation with the standard point at
internal of the instrument.
In case of actual setting, insert 0, into the
extra digits.
As for the value for extra digit, when tare
compensation and fine adjustment on load
are applied, it will be rewritten as a
compensated value automatically.
Setting range for the output of load cell is
from 0.400 0 mV/V to 3.100 0 mV/V.

By pressing the key continuously,


increment can be provided continuously.

Press the
key.
The load display section shows ZERO.
Here, set the instrument with initial load
application.

23

key Set value carry key


key Set value increment key
key Set value initialization key

Procedures

Press the
key.
The load display section shows ZERO with
lighting display on and off, then zero adjustment
can be started.

Warning At the same time, take care not


10

to apply load variation due to vibration and


so on. When load variation is applied, there
will be possibilities that zero point is
unstable, and precise reading of zero will
not be obtained.

After completed, the load display section


becomes END.
However, when the initial load is not entered
within the range from 2.0 mV/V to 2.0 mV/V,
the error code shown in the right figure will
show for about 2 seconds, then load display
section shows ZERO and return to step 9.
TEL Zero point
TEH Zero point

Error code

Error code

OVER
OVER

Press the
key.
After CCAL mode is over, the load display
section shows the present load.

11

24

522. Calibration procedures to register the output of strain gage applied transducer at the
time of zero and the maximum display
(1) Procedure by key operation

Warning

Before using a new instrument or exchanging the strain gage applied transducer
for a new one, be sure to make calibration.
If calibration will not be made, correct measured results may not be obtained nor it
may cause malfunction in the instrument and there may exist damage to the
peripheral equipment.
Besides, even though the calibration has been made, there may occur the similar
case when the result is not correct, so make calibration again.
During the calibration is executing, be sure to make cancellation (Execution of
F98) for compensated data on zero set
When the changeover target of the calibration data is set to the changeover by
external control input(Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During calibration procedures, press the


key in case of interrupting the
calibration is required. The calibration data will be kept as they are before entering
the calibration and then returns to the Measurement mode.

key is pressed with the setting display section of FUNC, the


display will change as the following arrow marks. However, every time the key

Every time the

is pressed, the display will change as the reverse direction of the following arrow
marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)

Procedures

Press the
key for about one second
The setting display section shows FUNC.
The load display value continuously shows the
present load value.

25

Procedures

Press the
key twice.
It makes the setting display section proceed as
FUNCCCALLCAL.

Press the
key.
LCAL mode can be entered, then the load
display section shows SCAL.

Press the
key.
The load display section shows D01.
When the calibration has completed already, the
set value of minimum scale which has registered
at that time is displayed.
Set the minimum scale with the right keys.
Setting values for the minimum scale are (four)
kinds as follows
1, 2, 5, 10

key Set value increment key

Press the
key.
The load display section shows DISP.

26

Procedures

Press the
key.
The load display section shows 2000 and the
minimum display digit flashes on and off.
When the calibration has completed already, the
maximum display value which has registered at
that time is displayed.

By the setting of minimum scale, the


minimum display digit that flashes on and
off is as follows
The minimum scale
1, 2, 5 100 digit
The minimum scale
10
101 digit

Set the maximum display value with the right


keys. Setting range for the maximum display
value is as follows:
In order to make effective use of the performance,
set within the following ranges.
Setting range for the
maximum display value The minimum scale
10010 000

20020 000

50050 000

1 00099 990

10

By pressing the key continuously,


increment can be provided continuously.

Press the
key.
The load display section shows LOAD.

27

key Set value carry key


key Set value increment key
key Set value initialization key

Procedures

Press the
key.
The load display section shows 2000 and the
minimum display digit flashes on and off.
When the calibration has completed already, the
maximum display value which has registered at
that time is displayed.

By the
h setting off minimum scale,
l the
h
8

minimum display digit that flashes on and


off is as follows
The minimum scale
1, 2, 5 100 digit
The minimum scale
10
101 digit

key Set value carry key


key Set value increment key
key Set value initialization key

Set the actual load value with the right keys.


The load value applied on the load cell actually is
the weight less than the maximum display value
set in step 6, and it will be the maximum load
that can apply on the load cell with the range of
(the minimum scale 100)
99 990 as well.

By pressing the key continuously,


increment can be provided continuously.

Press the
key.
The load display section shows ZERO.
Here, set the initial load application.

28

Procedures

Press the
key.
The ZERO on load display section flashes on
and off, and zero adjustment can be started.

Warning Take care not to apply load


10

variations due to vibrations and so on.


If load variation is applied, the zero point
will not stabilized, in due course there is a
possibility that correct reading of zero
wont be obtained.

When completed, the display on the load display Error code


section shows SPAN.
However, when the initial load is not entered the
range of 2.0
2.0 mV/V to 2.0 mV/V, the right Error
code will be shown for about 2 seconds, then the Error code
display on the load display section will be shown
as ZERO, and then the step 9 can be entered.
TELZero point OVER
TEHZero point OVER

11

Apply the same load on the load cell as set in the


step 8.

29

Procedures

Press the
key.
The SPAN on the load display section flashes
on and off, and span adjustment can be started.

Warning Take care not to apply load


12

variations due to vibrations and so on.


If load variation is applied, the span will
not stabilized, in due course there is a
possibility that correct reading of span
wont be obtained.

When completed, the display on the load display Error code


section shows END.
However, when the value corresponding to the
maximum display value does not satisfy the
range from 0.4
0 4 mV/V to 3.1
3 1 mV/V,
mV/V the right
Error code will light up for about 2 seconds, then Error code
the display on the load display section shows
SPAN, and then returns to the step 10.
SPLSpan point OVER
SPHSpan point OVER

Press the
key.
After LCAL mode is over, the load display
section shows the present load.

13

30

523. Zero fine adjustment


(1) Procedures by key operation

During zero set is executed and also during effective for Peak ON, Zero fine
adjustment mode cant be entered into. (Displays ER5.)
The Zero fine adjustment can be entered after the cancellation of the Compensated
data at Zero set (Execution of F98) and set the peak OFF.
When the changeover target of the calibration data is set to the changeover by
external control input(Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During the calibration procedure, press the


key to interrupt the calibration.
The calibration data will keep the same condition as it is entered before, then
returns to the Measurement mode.

When the
key is pressed with the setting display section of FUNC, the
display will change as the following arrow marks indicate at every time the key is

key is pressed, the display will change as the


pressed. However, every time the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)

Procedures

Press the
key for about one second.
The setting display section shows FUNC.
The load display value continuously shows the
present load value.

Press the
key three times.
It makes display in the load display section
proceeded as FUNCCCALLCAL
ZERO.
Here, set the initial load condition.

31

Procedures

Press the
key.
Zero fine adjustment mode can be entered, then
the display on load display section shows the
present load value and lights on and off. At the
same time, set the present load value to 0 with
the right keys.

By pressing the key continuously,


increment can be provided continuously.

The variation of load value for one push of

key Zero fine adjustment

the right key is less than 1 digit of display.


Therefore, a few pushes of these keys are
required to get the change of 1 digit of
display value.

display decreasing key

key Zero fine adjustment


display increasing key

Press the
key.
The indication of load display section shows
END.

Press the
key.
After quitting from zero fine adjustment mode,
the load display section shows the present load
value.

32

524. Span fine adjustment


(1) Procedures by key operation

During zero set is executed and also during effective for Peak ON, Span fine
adjustment mode cant be entered into. (Displays ER5.)
The Zero fine adjustment can be entered after the cancellation of the Compensated
data at Zero set (Execution of F98) and set the peak OFF.
When the changeover target of the calibration data is set to the changeover by
external control input(Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During the calibration procedure, press the


key to interrupt the calibration.
The calibration data will keep the same condition as it is entered before, then
returns to the Measurement mode.

When the
key is pressed with the setting display section of FUNC, the
display will change as the following arrow marks indicate at every time the key is

is pressed, the display will change as the


pressed. However, every time the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)

Procedures

Press the
key for about one second
The setting display section shows FUNC.
The load display value continuously shows the
present load value.

Press the
key four times.
It makes the setting display section proceeded as
FUNCCCALLCALZERO
SPAN.
Here, set the maximum load that can be applied
within the maximum value on the load cell.

33

Procedures

Press the
key.
Span fine adjustment mode can be entered, then
the display on load display section shows the
present load value and lights on and off. At the
same time, adjust the present load value to be the
same load applied on the load cell with the right
keys.

By pressing the key continuously,


increment can be provided continuously.

The variation of load value for one push of

key Span fine adjustment

the right key is less than 1 digit of display.


Therefore, a few pushes of these keys are
required to get the change of 1 digit of
display value.

key Span fine adjustment

display decreasing key


display increasing key

Press the
key.
The indication of load display section shows
END.

Press the
key.
After quitting from zero Span fine adjustment
mode, the load display section shows the present
load value.

34

525. Calibration procedure to apply registration again for zero point only
(1) Procedures by key operation

Warning

During the execution of calibration, be sure to set cancellation of the Compensated


data at Zero set (Execution of F98)
When the changeover target of the calibration data is set to the changeover by
external control input(Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During the calibration procedure, press the


key to interrupt the calibration.
The calibration data will keep the same condition as it is entered before, then
returns to the Measurement mode.

key is pressed with the setting display section of FUNC, the


When the
display will change as the following arrow marks indicate at every time the key is

is pressed, the display will change as the


pressed. However, every time the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)

Procedures

Press the
key for about one second
The load display section shows FUNC.
The load display value continuously shows the
present load value.

Press the
key five times.
It makes the setting display section proceeded as
FUNCCCALLCALZERO
SPANTARE.
Here, set the initial load condition.

35

Procedures

Press the
key.
The TARE mode can be entered.
The display on the load display section shows
ZERO.

Press the
key.
The display of ZERO on the load display
section flashes on and off, and Tare weight
cancellation is entered.

Warning At the same time, care is taken


4

not to apply load variation due to vibration


and so on.
If load variation is applied, zero point
becomes unstable, so there is a possibility
that correct zero cant be read.

When completed, the indication of load display


section shows END.
However, when the initial load isnt entered
within the range of 2.0 mV/V and 2.0 mV/V,
the error code in the right is shown for about 2
seconds, then the display on the load display
section shows ZERO, and returns to the step 3.

Error code

Error code

TELZero point OVER


TEHZero point OVER

Press the
key.
After quitting from the TARE mode, the load
display section shows the present load value.

36

526. Compensation of symmetric property(Fine adjustment of minus side span)


(1) Procedures by key operation

During the execution of zero set, or Peak ON, compensation of symmetric property
mode cant be entered into(Display ER5).
The Compensation of symmetric property can be entered after the cancel of the
compensation data at Zero set(Execution of F98), and set the peak OFF.
When the changeover target of the calibration data is set to the changeover by
external control input(Setting of F80 00001 or 00011), execute the
calibration after the changeover of the calibration data changed by the external
control output(SEL.1 and SEL.2).

During the calibration procedure, press the


key to interrupt the calibration.
The calibration data will keep the same condition as it is entered before, then
returns to the Measurement mode.

key is pressed with the setting display section of FUNC, the


When the
display will change as the following arrow marks indicate at every time the key is

key is pressed, the display will change as the


pressed. However, every time the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
(Hereinafter, it will repeat.)
As for the compensation of symmetric property, the polarity becomes effective only
for the indicated value on a minus side.
This instrument can make the compensation of approx.0.5 % to the maximum
display value setting.
When the compensation of symmetric property is executed, the zero point might be
changed.

Procedures

Press the
key for about one second
The setting display section shows FUNC.
The load display value continuously shows the
present load value.

37

Procedures

Press the
key ten times.
It makes the setting display section proceeded as
FUNCCCALLCALZERO
SPANTARECHEKMONT
VCALVADJSYMM.
Here, set the maximum weight on the load cell,
which initial load condition.
Here, please give the load cell the maximum load
which can be applied to the load cell below the
maximum indicated value on a minus side.

Press the
key.
Enters to the symmetric property compensation
mode, and the display of load display section
blinks. At the same time, adjust the present load
value to be the same load applied on the load cell
with the right keys.

By pressing the key continuously,


increment can be provided continuously.

The variation of load value for one push of

key Span fine adjustment

the right key is less than 1 digit of display.


Therefore, a few pushes of these keys are
required to get the change of 1 digit of
display value.

display decreasing key

key Span fine adjustment


display increasing key

Press the
key.
The indication of load display section shows
END.

Press the
key.
After quitting from the compensation mode of
symmetric property, the load display section
shows the present load value.

38

53. Selection of calibration methods on each condition


The instrument prepares calibration methods shown in the paragraph in 52. Calibration
procedures, well explain some conditions to execute actual calibration here.
(1) When executing calibration on the new instrument.
(In case that Combined Inspection at Minebea has not executed.)
D When load condition and output condition of load cell are clarified.
(Required accuracy is less than 1/1 000 or so.)
Proceed to the paragraph 531(1)
D When load condition and output condition of load cell are clarified.
(Required accuracy is more than 1/1 000 or so.)
Proceed to the paragraph 531(2)
D When load condition is clarified, but output condition of load cell is unclear.
Proceed to the paragraph 531(3)
D When exchanging with existing CSD819C is required.
Proceed to the paragraph 531(4)
(2) When making calibration again.
D When calibration only for tare weight is required.
(In case that the combined Inspection at Minebea has already executed, and the calibration
only for tare weight is required.)
D When fine adjustment on zero and span is required.
D When executing the compensation of symmetric property is required.

531. In case of executing the calibration on the instrument newly.


When the new instrument is purchased or reuse is desired with the new specific conditions,
execute the calibration with whichever procedure as follows
(1) When the load condition and the output condition of load cell are clarified.
(In case of desired accuracy is less than 1/1000 or so.)

Warning

The calibration accuracy obtained in this procedure is less than 1/1 000 or so. When
precise accuracy more than 1/1 000 is necessary, make calibration with actual load
according to the paragraph 531(2). Besides, the accuracy described here is a
combined accuracy of the instrument and the strain gage applied transducer
connected. If there may exist another factors of error such as mechanical elements
and so on, it will become out of warranty, so care is taken fully.
The rated output value for load cell applicable by the calculation is assumed as the
value described on the Inspection data individually.

For example, we will show the calibration procedures as follows, that is, 3 points of load cells
with 3 mV/V of rated output and 5 t of rated capacity.
Tare weight
Weighing capacity
Maximum display

1.5 t
5t
5 000

39

Calculate the output of load cell at maximum display from the above conditions. Check that

the calculated value is within the range from 0.4 mV/V to 3.1 mV/V. If the value is out of the
range, calibration cant be executed.
Output of load cell at maximum display

(Rated output)(Rated output)(Rated output)


Number of load cells
3 mV/V3 mV/V3 mV/V
3 points

Weighing capacity
(No. of load cells)(Rated load)

5t
3 points5 t

1 mV/V
2

After making the load cell to the initial load condition (tare weight), execute the calibration
according to the paragraph 521. In this case, input 5000 in the step 6, and input 1.0000
in the step 8 individually.

If necessity requires, apply zero/span fine adjustment according the paragraph 523, and
524.

(2) When the both conditions of load and the output of load cell are clarified.
(In case that required accuracy is more than 1/1 000 or so.)

Warning

The accuracy obtained through the procedures of this calibration consists from
combined accuracy with the instrument and combined strain gage applied
transducer, the accuracy of weight used during the calibration, error factors on
mechanical and also error factors on calibration works, that is, the total accuracy of
these. If high accuracy is required, full considerations is made on each factor. If
neglected, there will be a case that desired accuracy may not be obtained, so care is
taken fully.
When high accuracy is required, actual load calibration by using the weight and so on are
required.
For example, well show the calibration procedures in the following conditions, that is, 3 points
of load cell with 3 mV/V of rated output and 5 t of rated capacity.
Tare weight
1.5 t
Weighing capacity
5t
Maximum display
5 000

5231. Calculate the output of load cell at the maximum display from the above conditions.

Check that the calculated value at this point is within the range from 0.4 mV/V to 3.1 mV/V.
If the value is out of the range, calibration cant be executed.
Output of load cell at maximum display

(Rated output)(Rated output)(Rated output)


Number of load cells
3 mV/V3 mV/V3 mV/V
3 points

Weighing capacity
(No. of load cells)(Rated load)

5t
3 points5 t

1 mV/V
2

After making the load cell to the initial load condition (tare weight), execute the calibration
according to the paragraph 522. In this case, input 5000 in the step 6, and input the load
value applied on the load cell in the step 8 individually.

If necessity requires, apply zero/span fine adjustment according the paragraph 523, and
524.
40

(3) When the load condition is clarified but the output condition of load cell is not clarified.
In the case of using the existing load detecting section, and adopting the new digital indicator
only, it is necessary to execute calibration after checking the output of load cell when its
output is not clarified.
For example, followings are calibration procedures when weighing capacity is 5 t and the
others are not clarified.
1

Set the instrument in the monitor mode according to the paragraph 716. In this condition,
the output level of load cell connecting with the instrument can be monitored up to approx.
3.100 0 with the unit of mV/V.

After making the load cell section to the initial load condition (tare weight), record the display
value on load display. This value is the output of load cell at the time of initial load condition.

Record the display value on load display after applying 5 t load on the load cell section. This
value is the output of load cell with weighing capacity applied.

4. From the load cell output at the time of initial load application recorded at , and load cell
output recorded at at the time of application of weighing capacity, output of load cell at the
time of maximum display can be calculated according to the below formula.
Check that the calculated value is within the range from 0.4 mV/V to 3.1 mV/V. The
calibration cant be executed if the value is out of the range.
(Output of load cell at the time of maximum display)
(Output of load cell at weighing capacity)(Output of load cell at initial load application)

Quit the monitor mode of the instrument.

After making the load cell to the initial load condition (Tare weight), execute calibration
according to the paragraph 521. In this case, the accuracy is less than 1/1 000 or so.
At this moment, input each value, 5000 in the step 5 and another input is the value of
Output of load cell at the maximum display calculated from the in the step 8. If the
accuracy more than of 1/1 000 or so is required, execute calibration according to the paragraph
522. And at the same time, input 5000 in the step 6 and also input Load value going to
apply on load cell in the step 8 individually.

As necessity requires, make fine adjustment on Zero and Span according to the paragraph
523 and 524.

(4) When replacing the existing CSD819C with a new one.

Warning

The accuracy in this procedure is less than 1/1 000 or so.


If higher accuracy is required, make calibration by using the actual load according
to the paragraph 531(2).
Moreover, the accuracy described here is a combined accuracy with the instrument
and strain gage applied transducer connected.
When another error factors may exist, such as constructional error factors or so, it
will become out of warranty for accuracy, so care is taken fully.

When the load at the section of load cell cant make it with initial load application due to
failure on the existing CSD819C, execute calibration by referring to the procedures as below.
However, in case that the initial load condition can be obtained, make calibration according to
the procedures of (1) and (2).
1

According to the paragraph 81, read out and write down the function F90 Increment
value, the F91 Maximum display value, the F93 Zero calibration value, and the F94
Span calibration value in the existing CSD819C.

41

According to the paragraph 4, replace the exiting CSD819C with a normal instrument and
make connections.

After turning ON the normal instrument, make calibration according to the paragraph
521. In case of this, make the weight on the load cell to the initial load condition and input
the Increment value recorded at in the step 4, the Maximum display value in the step 6
and input the value subtracted Zero calibration value from Span calibration value in the
step 9.

532. When the calibration is executed again.


When purchasing a new instrument and the combined inspection has executed at Minebea,
however, the tare weight has changed, fine adjustment on zero and span, or the compensation of
symmetric property(fine adjustment for span at the minus side) are required, make calibration
with whichever the following methods.
(1) Calculation on tare weight only
(When combined inspection has completed at Minebea and calibration on only tare weight is
required.)
When the initial load (tare weight) has changed after completing the calibration, or when the
combined inspection at Minebea has been made and the calibration only for the initial
load(tare weight) is required after the installation, proceed the calibration in the following
steps.
1

After setting the initial load (tare weight) on load cell section, execute calibration according to
the paragraph 525.

(2) In case of executing fine adjustment on zero and span


Make adjustment according to the paragraph 523(Fine adjustment on zero), and 524
(Fine adjustment on span).
(3) In case of executing the compensation of symmetric property(fine adjustment for span at the
minus side)
To compensate the symmetric property of strain gage applied transducer which outputs the
polarity, the fine adjustment can be made only for the display value of minus side polarity.
Please execute according to the paragraph 526.

Warning

The accuracy obtained through the calibration procedures consists from combined
accuracy with the instrument and strain gage applied transducer, the accuracy of
weight used during the calibration, error factors on mechanical and also error
factors on calibration works, that is , total accuracy of these.
If high accuracy is required, full considerations is made on each factor.
If neglected, there will be a case that the desired accuracy is not obtained.

54. Setting the prohibition against calibration


After completing all of the calibration procedures, setting can be made to prohibit any more
calibration again by setting the funciton(Related function F97).
For details, refer to the paragraph 714.

42

55. Changeover of calibration data


This unit can memorize four kinds of calibration data, and can use by changeover them. That
changeover is made by the external control input, or function F81. The changeover method is
selected by function F80. (Related functionF80 and F81)
Default has selected Calibration data changeover by function F81.
(1) Calibration data changeover by external control input
When Changeover by the external control input is selected by function F80, the
changeover of calibration data is executed by combining the SEL.1 input and the SEL.2 input.
Calibration data No.
0
1
2
3

SEL.1
Open
Short
Open
Short

SEL.2
Open
Open
Short
Short

(2) Calibration data changeover by function F81


When Changeover by function (F81) is selected by function F80, the changeover of the
calibration data is executed by setting function F81. Default has selected calibration data 0.
Calibration data No.
0
1
2
3

Warning

Setting of function F81


00000
00001
00002
00003

When the analog output is made against four kinds of calibration data, the function
setting by Display value at the minimum analog output(F21), Display value at
the maximum analog output (F22), or the execution of Fine adjustment of
Analog output is executed on each calibration data. If neglected, the correct output
may not be obtained.
The set content displayed in the function mode is displayed against the calibration
data which has been selected by External control input and Function F81.

When you execute the calibration, please execute it after setting the calibration
data by using the external control input or function F81.
SEL.1 and SEL.2 of the external control input signal are used combinedly with the
input signal of Comparator code changeover. When the both setting of function
F80 is selected to the target of External control input, Calibration data and
Comparator code changeover at the same time.

43

6. Operation procedure
Well show the operating procedures with keys located on the front panel.

Warning

61.

Each key operation is made after interrupting the measurement. If it is made


during measurement, it may cause an unexpected malfunction.

Key operation in Measurement mode can be effective by pressing it for about one
second or so.

key

611. Operated in the Measurement mode


When the display value on load display section is within 10 % against the maximum display
value, (Refer to the Chapter 5.), the 0 display will be shown compulsively due to the zero set
function of this operation . As for details, refer to the paragraph 79.

62.

key

621. Operated in the Measurement mode


When the display value on load display section is within 10 % against the maximum display
value, (Refer to the Chapter 5.), the 0 display will be shown compulsively due to the zero set
function of this operation . As for details, refer to the paragraph 79.

622. Operation is made in another mode


(1) Calibration
In each procedure of calibration procedures, setting value can be initialized.
(2) Function mode
In function mode, function number and setting value for the function can be made 0
compulsively.
(3) Comparator setting
The set value can be made 0 compulsively, with the condition of possible to set for
comparator set value.

63.

key

631. Operated in the Measurement mode


Entering in the status where the set value of the comparator can be set, LED of comparison
value selection display panel blinks. The comparison value number which wants to change from
this status is selected. If you keep the instrument untouched for approx. 20 seconds, the
Measurement mode can be reentered automatically.

44

632. Operation is made in another mode


(1) Carrying up the set value.

key with the condition of displaying various kinds of set values, the
By pressing the
flashing digits for set value will carry up from 100, 101, 102, 103 to 104 in order.
(However, the range of carry differs depending on the number of digits of set values and yes/no
of polarity sign.)
(2) Change of function and so on.
Executes change of various kinds of functions in Function mode, and also change of function
during Check mode.
(3) Decrement at the time of fine adjustment.

key is pressed at the time of making fine adjustment on ZERO, SPAN,


When the
correction of symmetric property and analog output, the targeted value will decrease.

64.

In the operation method of the paragraph 632(3), when the key is pressed for
more than about 1 second, each operation will be executed continuously at a
constant interval without ON/OFF operation of the key.

key

641. When operated in the Measurement mode.

By pressing the
key, the comparison value selecting display will increase per one count
from 0, 1, 2, 3, 4 and 0 again in order, and then the comparison set value of selected number is
displayed.

642. When operated in another modes


(1) Increment of set value

key with the condition of displaying various kinds of set values, the set
By pressing the
value will increase per one count from 0, 1, 2, 3, 4, 5, 6, 7, 8, 9 and 0 again in order.
(2) Increment at the time of fine adjustment

key is pressed at the time of making fine adjustment on ZERO, SPAN,


When the
Correction of symmetric property and analog output, the targeted value will increase.

65.

In the operation method of the paragraph 642(2), when the key is pressed for
more than about 1 second, each operation will be executed continuously at a
constant interval without ON/OFF operation of the key.

key

651. When operated in the Measurement mode.


(1) When the peak function is selected.
Changeover the display of peak value/track value.

45

66.

key

661. When operated in the Measurement mode


(1) When the peak function is selected.
Set the display 0, and reset the hold value.

662. When the A/Z function is selected.

key is pressed while entered in various kinds of modes, you can shift to the
When the
Measurement mode.

67.

key

671. When operated in the Measurement mode.


At the same time the CHECK value ON and the CHECK on the status display will light up,
the load value equal to the set value with the F11 on the load display value will be added.
By pressing the key again, the CHECK value will be OFF and return to the condition as it was.

68.

key

Pressing the
key makes to register the set values internally, then the condition of Possible to
set will be over.

46

7. Function and operation


71. External control input signal and open collector output signal
The instrument is available to the external control through various kinds of input signals.

711. Input signal for external control


Activates by shortening with COM.1 (Terminal No. 14).
Terminal No.
8

Name
ZERO

HOLD

10

PEAK
/TRACK

11

12

RESET
SEL.1

13

SEL.2

14

COM.1

Operation
When the indicated value on load display section is within 10 %
against the maximum display value, zero set function will activate
by the operation and make the display 0 compulsively. (Same
key operation in the paragraph 62.) As for the operational
details, refer to the paragraph 79.
While inputting the signal, the target selected with Function
F10 among display, open collector output, analog output and
options will be frozen. As for the operational details, refer to the
paragraph 77.
Changeover of track/peak hold, bottom hold, peak/bottom hold,
the local maximum hold, the local minimum hold and inflection
point hold.
Open Track
ShortPeak hold, Bottom hold, Peak/bottom hold, The local
maximal value hold, The local minimal value hold, The inflection
point hold
During the various kinds of peak hold operation, reset condition is
made by short circuit.
Changeover of comparator code and calibration data by the
combination of SEL.1
SEL 1 and SEL.2.
SEL 2
(The target selected by the function F80 is changed.)
The common for the input signal (Terminal No.8 to 13)

Operation of input signal will be executed after shortening for more than 50 ms
approximately. (The changeover of level and pulse width of 1, 2, 5, 10 and 20 ms is
available. Related function F72)
As for the ZERO signal input, the operation of each calibration data which has
been selected is executed.

key on the front


During the input of HOLD signal, when ZERO signal (or the
panel) is input, operation of ZERO will be executed at the same time of cancellation
of HOLD signal.
After inputting the HOLD signal in powerOFF condition, turn ON the power,
then the HOLD lights on the load display section. The load value is shown
simultaneous with cancellation of HOLD signal.
The COM. 1 (terminal No. 14) and the COM.2 (terminal No. 21) are isolated.

47

712. Open collector output signal

Terminal No.

Name

15

END

16

S0

17
18
19
20

S1
S2
S3
S4

21

COM.2

Operation
Output when the completion of various kinds of peak mode
measurement. (Outputs with END LED interlock.)
Operated with whichever condition as follows by the setting F33.
a) ON in the condition of Load display valueThe maximum
display value
b) ON when both open collector outputs S1 and S2 are OFF
c) ON when both open collector outputs S1 and S3 are OFF
d) ON when both open collector outputs S1 and S4 are OFF
e) ON when both open collector outputs S2 and S3 are OFF
f) ON when both open collector outputs S2 and S4 are OFF
g) ON when both open collector outputs S3 and S4 are OFF
h)Operation whichever Open collector ON at more than the set
value, or Open collector ON at less than the set value
i) Interlock with HOLD LED
j) Interlock with PEAK LED
k)Interlock with MEAS. LED
Open collector output for S1 comparator.
Open collector output for S2 comparator.
Open collector output for S3 comparator.
Open collector output for S4 comparator.
The common of open collector output(Terminal No.1520).

COM. 1(terminal No.14) and COM.2(terminal No.21) are isolated.


The comparator in the instrument executes comparative operations synchronous
with The setting of A/D sampling rate (F02).
The delay time starting from preparing the output conditions to the contact output
ON, will be 1 ms approximately.

713. Equivalent circuit


(1) External control input section
12 V

10 k

2.2 k

12 V

1 k

5 V

INPUT
3 300 pF

(2) Contact output section

3.3 k

OUTPUT

COM.2
VCE=DC30 V, IC=30 mA MAX.

48

COM.1

72. Comparator
The instrument prepares comparators that consist of 4 kinds of set values S1, S2, S3 and S4 and
comparator S0 that can change the operation by the setting function F33. Additionally, this can
memorize four kinds of comparator code, and changeover and use them

The comparator of the instrument executes comparative operation synchronous


with the setting of A/D sampling rate by F02.
In the measurement mode, the change of the comparator setting value is available.

721. ON/OFF for the Comparator S0, S1, S2, S3 and S4.
Operational selection of ON/OFF can be made for each comparator S0, S1, S2, S3 and S4. These
selection can be made with the function (Related function F30).
As for default, all of the S0, S1, S2, S3 and S4 are selected ON.

722. Changeover of comparator code


The changeover is executed by the external control input or the function F82. The selection of
the changeover method is executed by function F80.(Related function F80 and F82)
As for default, Changeover of comparator by function(F82) is selected.
(1) Changeover of comparator code by the external control input.
When Changeover by the external control input is selected by function F80, the
changeover of the comparator code is executed by combining the input of SEL.1 and SEL.2.
Code No.

SEL.1 input

SEL.2 input

Open

Open

Short

Open

Open

Short

Short

Short

(2) Changeover of comparator code by function F82


When Changeover by function (F82) is selected by function F80, the changeover of the
comparator code by setting function F82 is executed. As for default, code 0 is selected.

Code No.

Setting by function F82

00000

00001

00002

00003

When you change the comparator set value, change the comparator set value after
adjusting the changing code by the external control input or setting by function
F82.
SEL.1 and SEL.2 of external control input signal is used combinedly with the input
signal of Calibration data changeover. When the all setting of function F80 is
selected to the target of External control input, the changeover of Calibration
data and Comparator code is made at the same time.

49

723. Change of set value

Warning

When the set value for the comparator is set wrong, or set in the wrong procedures,
it may not obtain the correct results from the comparator, and it may cause
malfunctions in peripheral equipments and also cause a damage as well.

Procedures
Selects the code which wants to set by external
control input signal or function F82. In the

key is
measurement mode, when the
pressed, the comparison value selecting display
section blinks.
Selects the number which wants to change by

using the
key and the
key. The set
value corresponding to the number is displayed
in setting display section.

To make effective for the S0 set value, set the

F33 as 00007=Comparator operation


(Normal operation).

Press the key to interrupt the change of

Decrement key of setting section


Increment key of setting section
A setting section is adjusted to 0.

set value. It will return to the measurement


mode.

Press the
key. The minimum display
effective digit of setting display section blinks.
Set the comparator set value of the selected
number by using the right keys.

Press the key to interrupt the change


of set value. It will return to the
measurement mode.

If you keep the instrument untouched for

approx. 20 seconds in the condition of


change of setting is available, the
Measurement mode can be reentered
automatically.

Press the
key upon completing the change.
The set value becomes effective and the
selecting display section of the comparative
value blinks. Return to step 2 to change the
other set value, or return to step 1 to set the
value of the other code.

Press the
mode.

Warning

Set value carry key


Set value increment key
Set value initializing key
(Adjust the set value to 0)

key to return to the measurement


When you set the changeover of comparator code in the external control input, the
input signal is always effective. So, be much careful to change in the measurement
mode.
50

724. Operation on comparator S1, S2, S3 and S4


The comparator in the instrument, S1, S2, S3 and S4 can select the operation whichever Open
collector output ON at more than the set value, or Open collector output ON at less than the
set value. This selection can be made in the Function mode. (Related function F32)
As for default, the Open collector output ON at more than the set value is selected for both of
S1, S2, S3 and S4.

Warning

Depending on the operational selection for comparator, ON/OFF condition for each
open collector output may differ. If wrong operation is selected, ON/OFF condition
for open collector output becomes inadequate and it may cause an unexpected
accident due to malfunctions on peripheral instruments, so care is taken fully.

When the load display is OL or OL, the display value for the comparison of
comparator is assumed as (infinity) and (infinity) individually.
Operation on judgement display section and open collector output will be shown as follows for
the S1 as a sample. S2, S3 and S4 are also in the same operation.
(1) When the operation of open collector ON at more than the set value is selected.
S1 judgement display, S1 open collector output ON at (S1 set value)(Display value)
Display (large)
Display (small)
S1 set value
ON
S1 judgement display,
S1 open collector output

OFF

(2) When the operation of open collector ON at less than the set value is selected.
S1 judgement display, S1 open collector output ON at (S1 set value)(Display value)
Display (large)
Display (small)
S1 set value
ON
S1 judgement display,
S1 open collector output

OFF

51

725. Comparative target for comparator S1, S2, S3 and S4


The comparator in the instrument, S1, S2 individually can select the comparative target from
the two, that is, PEAK, TRACK. This selection can be made in the Function mode. (Related
function FUNC31) As for default, the TRACK is selected for both of S1, S2, S3 and S4.

Warning

Depending on the selection of comparative target for the comparator, ON/OFF


condition for each open collector output may differ. If wrong operation is selected,
ON/OFF condition for open collector output becomes inadequate and it may cause
an unexpected accident due to malfunctions on peripheral instruments, so care is
taken fully.

Operation on judgement display section each comparative target and the operation of open
collector output will be shown as follows when the operation of open collector ON at more
than the set value is selected by the S1, for an example. The same is the operation of S2, S3 and
S4.
(1) When the operation of PEAK is selected.
S1 judgement display, S1 open collector output ON at (S1 set value)(Display load)
Load display (small)
Load display (large)
S1 set value
ON
S1 judgement display,
S1 open collector output

OFF

(2) When the TRACK is selected.


S1 judgement display, S1 open collector output ON at (S1 set value)(TRACK)
TRACK (large)
TRACK (small)
S1 set value
ON
S1 judgement display,
S1 open collector output

OFF

52

726. Operation of comparator SO


The comparator S0 in the instrument can select one among 11 operations from 00000 to
00010. These selections can be made in the Function mode (Related function F33)
As for default, the 00000 has selected.

Warning

Depending on the selection of S0 operation, ON/OFF condition for each S0 open


collector output may differ. If wrong operation is selected, ON/OFF condition for S0
open collector output becomes inadequate and it may cause an unexpected accident
due to malfunctions on peripheral instruments, so care is taken fully.

As example, the operation of S0 judgement display and S0 open collector output at the time of
whichever is selected from the 00000 to 00010 with the function F33, are shown in the
following table.
Setting of F33
00000
00001
00002
00003

Operation
ON at (Load display value)(Maximum display value)
ON when both open collector output S1 and S2 are OFF.
ON when both open collector output S1 and S3 are OFF.
ON when both open collector output S1 and S4 are OFF.

00004
00005

ON when both open collector output S2 and S3 are OFF.


ON when both open collector output S2 and S4 are OFF.

00006
00007
00008
00009

ON when both open collector output S3 and S4 are OFF.


Operation of whichever open collector ON at more than the
set value, or open collector ON at less than the set value.
Interlock with HOLD LED
Interlock with PEAK LED

00010

Interlock with MEAS. LED

In the next, as for the sample, the operation at S0 judgement display and S0 open collector
output selected 00001 with the Function F33 will be shown as follows. The same operation
will be made when 00002 to 00006 are selected. The operation for S0 comparator when the
00007 is selected will be the same as the operation of S1, S2, S3 and S4 written in the
paragraph 724, and 725.
(1) When the both of S1 and S2 select more than with the Function F32.
S1 judgement display, S1 open collector output
ON at (S1 set value)(Display value)
S2 judgement display, S2 open collector output
ON at (S2 set value)(Display value)
S0 judgement display, S0 open collector output
ON at (S1 set value)(Display value) and (S2 set value)(Display value)
Display (large)
Display (small)
S1 set value S2 set value
ON
S2 judgement display,
S2 open collector output

OFF
ON

S1 judgement display,
S1 open collector output

OFF
ON

S0 judgement display,
S0 open collector output

OFF

53

(2) When the S1 selects less than and the S2 selects more than with the Function F32.
S1 judgement display, S1 open collector output
S2 judgement display, S2 open collector output
S0 judgement display, S0 open collector output
Display (small)
S1 set value

ON at (S1 set value)(display value)


ON at (S2 set value)(display value)
ON at (S1 set value)(display value)
(S2 set value)
Display (large)
S2 set value
ON

S2 judgement display,
S2 open collector output

OFF
ON

S1 judgement display,
S1 open collector output

OFF
ON

S0 judgement display,
S0 open collector output

OFF

Above figure indicates the case of (S1 set value)<(S2 set value).
In the case of (S1 set value)(S2 set value), the S0 judgement display and the S0
open collector output will be normally OFF.
(3) When the S1 selects more than, and the S2 selects less than at the function F32.
S1 judgement display, S1 open collector output
S2 judgement display, S2 open collector output
S0 judgement display, S0 open collector output

ON at (S1 set value)(display value)


ON at (S2 set value)(display value)
Normally OFF
Display (large)

Display (small)
S1 set value

S2 set value
ON

S2 judgement display,
S2 open collector output

OFF
ON

S1 judgement display,
S1 open collector output

OFF
ON

S0 judgement display,
S0 open collector output

OFF

Above figure indicates the case of (S1 set value)(S2 set value).
In the case of (S1 set value)(S2 set value), the S0 judgement display and the S0
open collector output will ON in the condition of (S2 set value)(display
value)(S1 set value)

54

(4) When both of the S1 and S2 select less than at the function F32
S1 judgement display, S1 open collector output
ON at (S1 set value)(display value)
S2 judgement display, S2 open collector output
ON at (S2 set value)(display value)
S0 judgement display, S0 open collector output
ON at (S1 set value)(display value) and also (S2 set value)
(display value) at the same time.
Display (large)
Display (small)
S1 set value

S2 set value
ON

S2 judgement display,
S2 open collector output

OFF
ON

S1 judgement display,
S1 open collector output

OFF
ON

S0 judgement display,
S0 open collector output

OFF

55

727. Hysteresis on comparator


The comparator S1, S2, S3 and S4 and normal mode for S0 (Function F3300007 setting)
can set hysteresis
Hysteresis can be used by the combined setting of data width and time width. Moreover,
effective direction for hysteresis can be selected from either Off delay or On delay.
These selections can be made in Function mode. (Related function F34, F35, F36) As for
default, hysteresis OFF is set.

Warning

Depending on the setting of comparator hysteresis, ON/OFF condition for each


open collector output may differ. If wrong mode is selected, ON/OFF condition for
open collector output becomes inadequate and it may cause an unexpected accident
due to malfunctions on peripheral instruments, so care is taken fully.

As for the example of S1, the operation of judgement display section and open collector output
when the hysteresis on comparator is set, will be shown as follows
The same operation will be obtained in the case of S2, S3, S4 and Normal mode at S0.
(1) When the operation of open collector ON at more than the set value is selected at S1 and
also effective direction for hysteresis is set as On delay.

Display

Hysteresis time width

Hysteresis
data width

S1 set value

Time
S1 judgement
ON
display
S1 open
OFF
collector output
(2) When the operation of open collector ON at more than the set value is selected at S1 and
also effective direction for hysteresis is set as Off delay.
Hysteresis time width
Display

S1 set value
Hysteresis
data width

Time
ON

S1 judgement display
S1 open collector output

OFF

56

(3) When the operation of open collector ON at less than the set value is selected at S1 and also
effective direction for hysteresis is set as On delay.

Display

Hysteresis time width


S1 set value
Hysteresis
data width

Time
ON

S1 judgement display
S1 open collector output

OFF
(4) When the operation of open collector ON at less than the set value is selected at S1 and also
effective direction for hysteresis is set as Off delay.

Display

Hysteresis time width

Hysteresis
data width

S1 set value

Time
ON

S1 judgement display
S1 open collector output

OFF

728. Light on/off setting of comparator value display section


The changeover of on/off for comparator value selecting display section and comparator value
setting display section is set by the function. (Related function F38)

57

73. How to use the filter


The instrument prepares the analog filter consists of the low pass filter incorporated on the analog
circuit, and the digital filter that stabilizes data converted into digital through calculation process.

Warning

In the measurement mode, the changeover of filter is available.


When setting filter is not suitable, correct measurement cant be made and it may
cause an unexpected accident, so care is taken fully.

731. Analog filter


The instrument can change the pass band for the analog filter into 3 steps, such as 10 Hz,
100 Hz and 1 kHz. (Related function F05) As for default, 1 kHz is selected.
The tendency of characteristics by the frequency are listed as below
Frequency
Resist to noise

10 Hz
stable

Response speed

slow

100 Hz

1 kHz
rapid
quick

732. Digital filter


The digital filter for the instrument can be set from 00000 to 00088.
The averagedout times of digital filter can be decided from the set value (Related function
F04)
As for default, 00055 is selected.
The relations between the setting and the averagedout times are as follows
msetting value of 101 digit
(Averagedout times) 2m2n
nsetting value of 100 digit
For example, when 00055 is selected, averaged out times is as follows
(Averagedout times) 2525
64(Times)
The tendency of characteristics by the Averagedout times are listed as below
Averaged out times
Resist to noise
Response speed

low
rapid
quick

high
stable
slow

74. Selection of A/D sampling rate


This instrument can select the A/D sampling rate from 100 times/s, 500 times/s, 1 000 times/s and
2 000 times/s.(Related function F02). As for default, 2 000 times/s is selected.

The comparator comparison operation and the analog output of this unit have
synchronized with the A/D sampling. Therefore, please execute the change in the
A/D sampling rate when you change the conversion rate of the comparator and
analog output.

58

75. How to use various kinds of peak holds.


The instrument prepares 6 kinds of Peak holding functions, such as peak hold, bottom hold, peak
bottom hold, the local minimal value hold, the local maximal value hold and inflection point hold,
and 4 kinds of Peak bottom holding detection mode, such as all zone, specified zone, time specified
zone, and automatic start time specified zone. These selections can be made in the function mode.

751. Related function


F60

Setting of hold mode

00000Peak hold
00001Bottom hold
00002Peak bottom hold
00003The local minimal value hold
00004The local maximal value hold
00005Inflection point hold

F61

Setting kind of detecting mode

F62

Setting of reference point (Effective in


peak bottom hold)
Setting data width of local maximum
/local minimum detection
(Effective in the local maximal value
hold and the local minimal value hold)
Setting of local maximum and local
minimum detection magnification
(Effective in the local maximal value
hold and the local minimal value hold)
Setting data width of inflection point
detection (Effective in inflection point
hold.)

00000All zone specification


00001Zone specification
00002time zone specification
00003Auto start time zone specification
9999999999
Unit1D
0000099999
Unit1D

F63

F64

F65

0000000049
Unit0.1 times

9999999999
Unit1D

F66

Setting time width A of inflection point


detection (Effective in inflection point
hold)

0000000450
Unit1 ms

F67

Setting time width B of inflection point


detection (Effective in inflection point
hold)

0000000450
Unit1 ms

F68

Setting time of hold detection


(Effective in time zone mode and
automatic start time zone mode.)

0000009999
Unit1 ms

F69

Setting of automatic start level


(Effective in automatic start time zone
mode)

9999999999
Unit1D

The table below shows the combination of various holding functions and related functions.
F60 F61 F62 F63 F64 F65 F66 F67 F68 F69

Peak hold
Bottom hold

Peak/bottom hold
Local minimal value hold

Local maximal value hold


Inflection point hold

59

752. Peak hold function by all zone specification


1

Peak hold in all zone


The maximum value of the load is held. (Related function F6000000 and F6100000)
The load display becomes 0 against the RESET input during the detection, the detection will
be executed again with RESET OFF.
Load

Load display

Time

RESET
zone

Detecting zone

Detecting
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

60

Bottom hold in all zone


The minimum value of load is held. (Related function F6000001 and F6100000)
The load display becomes 0 against the RESET input during detecting, and detected again by
RESET input OFF.

Load
Load display

Time

Detecting zone

RESET
zone

Detecting
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input

PEAK LED
2

MEAS. LED
2

END LED and


END output

1 Whichever

key input or PEAK/TRACK input becomes effective. At the time

of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

61

Peak/bottom hold in all zone


The maximum (minimum) value of the difference between set value (F62) of the reference
point registered beforehand and the load is held from the point by which the PEAK/TRACK
signal is input. (Related function F6000002, F6100000 and F629999999999)
At this time, the reference point set value becomes a standard (load display =0) on the load
display. The load display becomes 0 against the RESET input during detecting, and detected
again by RESET input OFF.
Load

Load display
Load display0

Set value of reference point


Time

RESET
zone

Detecting zone

Detecting
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2

MEAS. LED
2

END LED and


END output

1 Whichever

key input or PEAK/TRACK input becomes effective. At the time

of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

62

The local minimal value hold in all zone


The local minimal value of the load is detected and held. (Related function F6000003,
F6100000, F630000099999 and F640000000049)
The load display becomes 0 against the RESET input during detecting, and detected again by
RESET input OFF. As for the detecting condition, refer to the paragraph 756. (Set of
detecting condition for the local maximal value and the local minimal value.)
Load

Load display

Time
Detecting
zone

Detecting
zone
zone

RESET

Holding zone

Holding zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2
PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the local maximal (minimal)


value held during the PEAK/TRACK ON is reset, and the measuring is begun
again. The load display becomes 0 while inputting RESET.

key.
Press the
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
63

The local maximal value hold in all zone


The local maximal value of load is held. (Related function F6000004, F6100000,
F630000099999 and F640000000049)
The load display becomes 0 against the RESET input during detecting, and detected again by
RESET input OFF. As for the detecting condition, refer to the paragraph 756.
(Set of detecting condition for the maximal value and the minimal value.)
Load

Load display

Time
Detecting
zone

Holding zone

RESET
zone

Detecting
zone

Holding zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2
PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the local maximal (minimal)


value held during the PEAK/TRACK ON is reset, and the measuring is begun
again. The load display becomes 0 while inputting RESET.

key.
Press the
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
64

Inflection point hold in all zone


The value just before there is a rapid change in the load is detected and held.
(Related function F6000005, F6100000, F659999999999, F6600000
000450 and F6700000000450)
The load display becomes 0 against the RESET input during detecting, and detected again by
RESET input OFF. As for the detecting condition, refer to the paragraph 757.
(The set of the inflection point detecting condition).
Load

Load display

Time
Detecting
zone

RESET

Holding zone

zone

Detecting Holding
zone
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2
PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the inflection point held during
the PEAK/TRACK ON is reset, and the measuring is begun again.
The load display becomes 0 while inputting RESET.

Press the
key.
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
65

753. Peak hold function by zone specified


1

Peak hold in the specified zone


Upon specifying the detecting zone from the outside externally, the maximum value of load is
detected and held. (Related function F6000000F6100001)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
Load

Load display

Time

Detecting
zone

Holding
zone

Detecting RESET
zone
zone

Detecting
zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2

MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

66

Bottom hold in the specified zone


Upon specifying the detecting zone from the outside externally, the minimum value of load is
detected and held. (Related function F6000000F6100001)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
Load

Load display

Time

Detecting
zone

Holding Detecting
zone
zone

RESET
zone

Detecting Holding
zone
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input

PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

67

Peak/bottom hold in the specified zone


The maximum (minimum) value of the difference between set value (F62) of the reference
point registered beforehand and the load is detected and held from the point PEAK ON
specifying the detected zone from the outside optionally.
(Related function F6000002, F6100001 and F629999999999)
The reference point set value becomes a standard (load display =0) on the load display this
time. It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
Load display
0

Load
Load display

Set value of reference point


Time

Detecting
Detecting
zone
Holding zone
zone

RESET
zone

Detecting Holding
zone
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

68

The local minimal value hold in the specified zone


Upon specifying the detecting zone from the outside externally, the local minimal value of load
is detected and held. (Related function F6000004, F6100001, F630000099999
and F640000000049)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input in the detecting, and detected again by RESET input OFF. As for the
detecting condition, refer to the paragraph 756. (The set of the local maximal and minimal
point detecting condition).
Load

Load display

Time
Detecting
zone

Holding zone

RESET Detecting
zone
zone

Holding zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2

PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the local maximal (minimal)


value held during the PEAK/TRACK ON is reset, and the measuring is begun
again. The load display becomes 0 while inputting RESET.

Press the
key.
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
69

The local maximal value hold in the specified zone


Upon specifying the detecting zone from the outside externally, the local maximal value of load
is detected and held. (Related function F6000004, F6100001, F630000099999
and F640000000049)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input in the detecting, and detected again by RESET input OFF. As for the
detecting condition, refer to the paragraph 756. (The set of the local maximal and minimal
point detecting condition).
Load

Load display

Time
Detecting
zone

Detecting
zone
zone

RESET

Holding zone

Holding zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2
PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the local maximal (minimal)


value held during the PEAK/TRACK ON is reset, and the measuring is begun
again. The load display becomes 0 while inputting RESET.

Press the
key.
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
70

Inflecting point hold in the specified zone


The value immediately before there is a rapid change in the load specifying the detected zone
from the outside optionally is detected and held. (Related function F6000005, F61
00001, F659999999999, F660000000450 and F670000000450)
The input of RESET signal to release holding is necessary. The load display becomes 0 against
RESET input in the detecting, and detected again by RESET input OFF. As for the detecting
condition, refer to the paragraph 757. (The set of inflecting point detecting condition).
Load

Load display

Time
Detecting
zone

RESET

Holding zone

zone

Detecting
zone

Holding zone

key input 1

PEAK/TRACK
input 1

key or
RESET input 2
PEAK LED
3
MEAS. LED
3
END LED and
END output 4

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Proceed either of undermentioned operation when the local maximal (minimal)


value held during the PEAK/TRACK ON is reset, and the measuring is begun
again. The load display becomes 0 while inputting RESET.

Press the
key.
RESET of the external control input signal (terminal No.11) and COM.1
(terminal No.14) are shorted.
The provided command/data is transmitted from the host (personal computer
and sequencer, etc.) by the RS232C interface or the RS422/485 interface.
3 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
4 The END output and END LED are turned on when the holding section.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.
71

754. Peak holding function by specifying time zone


1

Peak hold in the specified time


The maximum value of the load is detected and held from point by which the PEAK/TRACK
signal is input during detection time set by the function (F68). (Related function F60
000000, F6100002 and F680000009999)
It is necessary to input the RESET signal to release holding. At this time, the reference point
set value becomes a standard (load display =0) on the load display. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
Load

Load display

Time

Detecting zone(F68)

Holding
zone

Detecting RESET Detecting


zone
zone zone

Detecting zone(F68)
Detecting zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

72

Bottom hold in the specified time


The minimum value of the load is detected and held from point by which the PEAK/TRACK
signal is input during detection time set by the function (F68). (Related function F60
000000, F6100002 and F680000009999)
It is necessary to input the RESET signal to release holding. At this time, the reference point
set value becomes a standard (load display =0) on the load display. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
Load

Load display

Time

Holding
zone
Detecting RESET Detecting
zone
zone zone

Detecting zone(F68)

Detecting zone(F68)
Detecting zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input

PEAK LED
2

MEAS. LED
2
END LED and
END output 3

1 Whichever

key input or PEAK/TRACK input becomes effective. At the time

of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

73

Peak/bottom hold in the specified time


The maximum (minimum) value of the difference between set value of the reference point
registered beforehand (F62) and the load is detected and held during detection time set by
the function (F68) from point by which the PEAK/TRACK signal is input. (Related function
F6000002, F6100002, F629999999999 and F680000009999)
At this time, the reference point set value becomes a standard (load display =0) on the load
display. It is necessary to input the RESET signal to release holding. The load display becomes
0 against the RESET input during detecting, and detected again by RESET input OFF.
Load display
0

Load
Load display

Set value of reference point

Reference line

Time

Detecting time(F68) Holding Detecting time(F68)


zone
Detecting RESET Detecting
Detecting zone
zone
zone
zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

74

The local minimal value hold in the specified time

The minimal value of the load is detected and held during detection time set by the function
(F68) from point by which the PEAK/TRACK signal is input. (Related function F6000003,
F6100002, F630000099999, F640000000049 and F680000009999)
It is necessary to input the RESET signal to release holding. The load display becomes 0 against
the RESET input during detecting, and detected again by RESET input OFF.
As for the detecting condition, refer to the paragraph 756. (The set of the local maximal and
minimal point detecting condition).
Load
Load display

Time
Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Detecting time(F68)
Detecting
zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever

key input or PEAK/TRACK input becomes effective. At the time

of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

When the local maximal (minimal) value cannot be detected during the detection
time (set of function F68), the TRACK operation will be made.

key or
key during the various kinds of hold function is
The input of
effective only in the measurement mode.
75

The local maximal value hold in the specified time


The local maximal value of the load is detected and held during the detection time set by the
function (F68) from point by which the PEAK/TRACK signal is input.
(Related function F6000004, F6100002, F630000099999, F640000000049
and F680000009999)
The input of RESET signal to release holding is necessary. The load display becomes 0 against
the RESET input during detecting, and detected again by RESET input OFF.
As for the detecting condition, refer to the paragraph 756. (The set of the local maximal
and minimal point detecting condition).
Load

Load display

Time
Detecting time(F68)

Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Detecting
zone

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2
END LED and
END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

When the local maximal (minimal) value cannot be detected during the detection
time (set of function F68), the TRACK operation will be made.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

76

Inflection point hold in the specified time


The value just before a rapid change in the load is detected and held during detection time set
by the function (F68) from point by which the PEAK/TRACK signal is input.
(Related function F6000005, F6100002, F659999999999, F6600000
00450, F670000000450 and F680000009999)
The input of RESET signal to release holding is necessary. The load display becomes 0 against
RESET input in the detecting, and detected again by RESET input OFF. As for the detecting
condition, refer to the paragraph 757. (The set of the inflecting point detecting condition).
Load

Load display

Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Detecting time(F68)
Detecting
zone

Time

Holding
zone

key input 1

PEAK/TRACK
input 1

key or
RESET input
PEAK LED
2
MEAS. LED
2

END LED and


END output 3

1 Whichever
key input or PEAK/TRACK input becomes effective. At the time
of short of PEAK/TRACK input, the peak hold condition will not change even if
the

key is pressed ON.

2 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
3 END output and END LED are turned on during the holding zone.

When the inflection point value cannot be detected during the detection time (set of
function F68), the TRACK operation will be made.

key or
key during the various kinds of hold function is
The input of
effective only in the measurement mode.
77

755. Peak holding function by range specification at automatic start time


1

Peak hold in the specified zone from automatic start time


The maximum value of the load is detected and held during detection time set by the function
(F68) from the time when the load indicated value passed automatic start level (F69).
(Related function F6000000, F6100003, F680000009999, F6999999
99999)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input in the detecting, and detected again by RESET input OFF.
Load

Load display

Automatic start level

Time

Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Holding
zone

key or
RESET input
PEAK LED
1
MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

78

Bottom hold in the specified zone from automatic start time


The minimum value of the load is detected and held during detection time set by the function
(F68) at time when the load indicated value passed automatic start level (F69). (Related
function F6000001, F6100003, F680000009999 and F699999999999)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input in the detecting, and detected again by RESET input OFF.
Load

Load display

Time
Automatic start level

Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Holding
zone

key or
RESET input

PEAK LED
1

MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

79

Peak/bottom hold in the specified zone from automatic start time


The maximum (minimum) value of the difference between set value of the reference point
registered beforehand (F62) and the load is detected and held during detection time set by
the function (F68) from the time when the load indicated value passed automatic start level
(F69). (Related function F6000002, F6100003, F629999999999, F68
0000009999 and F699999999999)
At this time, the reference point set value becomes a standard (load display =0) on the load
display. It is necessary to input the RESET signal to release holding. The load display becomes
0 against the RESET input in the detecting, and detected again by RESET input OFF.
Load display

Load

0
Load display

Set value of reference point

Automatic start level

Time

Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Holding zone

key or
RESET input
PEAK LED
1
MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

The input of
key or
key during the various kinds of hold function is
effective only in the measurement mode.

80

The local minimal value hold in the specified zone from automatic start time
The local minimal value of the load is detected and held during set time from the point by
which the load indicated value crossed automatic start level (F69).
(Related function F6000003, F6100004, F630000099999, F640000000049,
F680000009999 and F699999999999)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
As for the detecting condition, refer to the paragraph 756 (The set of the local maximal and
minimal point detecting condition).
Load

Load display

Automatic start level

Time
Detecting time(F68)
Detecting
zone

RESET
zone

Detecting
zone

Detecting time(F68)
Detecting
zone

Holding zone

key or
RESET input
PEAK LED
1

MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

When the local maximal (minimal) value cannot be detected in the detection time
(set of function F68), TRACK operation is proceeded, and it detects again every
time the load change passes the automatic start level afterwards (Regardless of the
direction).

81

The local maximal value hold in the specified zone from automatic start time
The local maximal value of the load is detected and held during detection time set by the
function (F68) from the time when the load indicated value passed automatic start level
(F69). (Related function F6000004, F6100004, F630000099999, F6400000
00049, F680000009999 and F699999999999)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input during detecting, and detected again by RESET input OFF.
As for the detecting condition, refer to the paragraph 756 (The set of the local maximal and
minimal point detecting condition).
Load

Load display

Automatic start level

Time
Detecting time(F68)
Detecting
zone

RESET
zone

Detecting time(F68)
Detecting
zone

Detecting
zone

Holding zone

key or
RESET input
PEAK LED
1

MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from S0 terminal
is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

When the local maximal (minimal) value cannot be detected in the detection time
(set of function F68), TRACK operation is proceeded, and it detects again every
time the load change passes the automatic start level afterwards (Regardless of the
direction).

82

Inflection point hold in the specified zone from automatic start time
The value just before a rapid change in the load is detected and held during detection time set
by the function (F68) from the time when the load indicated value passed automatic start
level (F69). (Related function F6000005, F6100003, F659999999999, F66
0000000450, F670000000450, F680000009999 and F699999999999)
It is necessary to input the RESET signal to release holding. The load display becomes 0
against the RESET input in the detecting, and detected again by RESET input OFF. As for the
detecting condition, refer to the paragraph 757 (The set of the inflecting point detecting
condition).
Load

Load display

Automatic start level

Time
Detecting time(F68)
Detecting RESET
Detecting
zone
zone
zone

Detecting time(F68)
Detecting
zone

Holding zone

key or
RESET input
PEAK LED
1

MEAS. LED
1

END LED and


END output 2

1 Interlocked to PEAK LED or MEAS. LED, open collector output from SO


terminal is made. (by function F33)
2 END output and END LED are turned on during the holding zone.

When the inflecting point value cannot be detected in the detection time (set of
function F68), TRACK operation is proceeded, and it detects again every time the
load change passes the automatic start level afterwards (Regardless of the
direction).

83

756. Setting of detecting condition of the local maximal value and the local minimal value
As a peak holding function of this instrument, Local maximal value holding function by
which the first peak magnitude (prime maximum value) is held most for the load change and
The local mimimal value holding function by which the first bottom value (prime minimum
value) is held most after the first peak magnitude (prime maximum value) is passed most are
provided. In the The local maximal value holding function and The local mimimal value
holding function, the parameter to detect the local maximal value and the local minimal value
is set in the function mode. The width of data to judge the local maximal value and the local
minimal value by function F63 is set. Magnification a to detect the local maximal value and
the local minimal value by function F64 is set.
D Method of detecting the local maximal value and the local minimal value

A
X
B
In case of the load change like the abovementioned, make the first peak magnitude as A
point and the first bottom value as B point. Difference X between A point and B point is
calculated, and judge the local maximal value as A point and the the local minimal value as
B point when the value of this X is more than function F63 set value.

C
A

(YaXa0.14.9)

X
B
The value by which the difference X between A as the local maximal value and B as the
local minimal value judged, and set value a (0.14.9) of function F64 is multiplied
becomes level Y by which the local maximal value and the local minimal value are detected.
When the load change passes B point and exceeds over the C point(BY), the load
display section displays A point (B point) as the local maximal value (the local minimal
value), and becomes a hold status.

In the environment with load change by vibration and load change by exogenous
noise etc., when the width of the local maximal/minimal value judgement data (set
of F63) is too small, the load change might be detected as local maximal/minimal
value and a correct measurement result may not be obtained.
The detection standard of the local maximal/minimal value is different according to
the measured sample. Please adjust referring to the detection method to the
optimum value.
84

757. Detection condition setting in inflection point


There is Inflection point holding function by which the value just before the first rapid
change is suitable most is held for the load change. In Inflection point holding function, the
parameter to detect the inflection point is set in the function mode.
Set the data width to detect the inflection point by function F65.
Set the time width A of the detecting data to judge the inflection point by F66.
Set the time width B of the detecting data to judge the inflection point by F66
D Detecting method of inflection point

Z
Inflection point

YX=Z

X
a
A

X
B

When there is a load change like the abovementioned, the amount of the load change
during the time width A set by function F66 is assumed to be X. In addition, the load
change during the time width B set by function F67 is assumed to be Y.
When you assume the change amount of the inclination to have pulled X of the change
amount of time width A from Y of the change amount of time width B to be Z, and when
the value of change amount Z is more than the data width set value of function F65 (set
value of YXF65), a point is displayed in load display section as inflection point and
becomes a hold condition.

The detection standard of the inflection point is different according to the


measured sample. Adjust the optimum value referring to the detection method.
When the inclination of the load change is gradual, the inflection point becomes
easy by assuming as detecting time width A (F66) detecting time width
B(F67).

85

76. Various kinds of functions related with the display


761. Selection of display rate
The instrument can select the display rate from the 4 times/s, 20 times/s 50 times/s and 100
times/s. This selection is available in the Function mode. (Related function F03) As for the
default, 4 times/s has selected.

762. Selection of position of decimal point display


The instrument can display the decimal point at the Load display section of the instrument.
The selection of display can be made in the Function mode. (Related function F01) As for the
default, No decimal point display has selected.

763. Load display range


The load display range for the instrument is fixed from the 110 % to 110 % of the maximum
display value at the time of setting during calibration. When less than the range, OL
displays and over the range OL displays. For example, when the maximum display value is
1 000, the load display range will be from 1 100 to 1 100.
Besides, when under 1 100, the OL displays, and over 1 100, the OL will display.

77. Selection the target for HOLD


The instrument can select the target for HOLD function with the combination from Load
display, Comparator S0 open collector output, LED display, Comparator S1, S2, S3 and S4
open collector output, LED display, Analog output and Options of BCD output.
This selection can be made in the Function mode(Related function F10).
As for the default, All selected has selected. Layout of setting for the F10 are as follows
100 digitLoad display
101 digitComparator S0 open collector output, LED display
102 digitComparator S1, S2 open collector output, LED display
103 digitAnalog output
104 digitOptional BCD output
With the 0 setting, out of the target, and with 1 setting, target of HOLD.

78. Changeover of bridge power supply voltage


The instrument can select the bridge power supply from 10 V, 5 V and 2.5 V. This selection
can be made in the Function mode(Related function F12). As for the default, 10 V has selected.

When the bridge power supply voltage is changed, make calibration again.
It is common to all of four kinds of calibration data and executed.

86

79. Zero set


The instrument prepares zero set function.
When the display value on load display is within 10 % against the maximum display value (Refer to

key makes zero set function operated and the display will show
the chapter 5.), pressing the
0 compulsively.
However, zero set will not be accepted when zero compensation for total 10 % is executed with
zero set and zero tracking until that time. (ERR0 display)
Also, the same operation can be made with the operation of ZERO at the input signal of external
control. Cancellation for data applied zero compensation by zero set can be executed with Function
F98.

710. Key lock


The instrument prepares key lock function.
With the Function F06, execute key lock OFF by setting each digit=0, and executes key lock ON
by setting each digit=1. As for default, all is key lock OFF has selected. Besides, the
correspondence between the target of key lock and setting digits are as follows:

10 digit,
10 digit
10 digit
10 digit
Moreover, when the key is locked , lock for the key will be cancelled only once after the
key is pressed together with the key for more than 3 seconds.
100 digit
1

3
4

711. CHECK value

When the
key is pressed in the Measurement mode, CHECK value equivalent to the set value
with the Function F11 will be ON. And at the same time when the CHECK on status display
lights up, the load value equivalent to the set value with Function F11 will be added to the load
display value. As for default, 0.3 mV/V has set.
By pressing the
as it is.

key again, the CHECK value will be OFF and returns to the former condition

When the setting of Function F11 is 00000, load display value will not vary
even if the CHECK lights up on the status display by pressing the
Because CHECK is 0.0 mV/V.)
It is common to all of four kinds of calibration data and executed.

87

key.

712. How to use the analog output

Warning

When the power supply is turned on with shorting the external HOLD signal, the
analog output will be made as follows:
When the target of HOLD by F10 is the analog output, the analog output will
become approx.0 V in voltage output, and approx.12 mA in current output..

Be careful on the following point when


key is in use.
The OL error display(Analog output at OVR.) will be displayed by setting the
CHECK ON.
Please execute function setting of Indicated value with a minimum analog output
(F21), Indicated value with the maximum analog output (F22), Fine
adjustment of the analog output of each calibration data when you use the analog
output for four kinds of calibration data.
The setting details displayed in the function mode is displayed to the calibration
data has been selected by External control input and Function F81.

The analog outputs for the instrument are 2 type of Voltage output (Standard)
and Current output(Optional).
The analog output of the instrument executes rewriting the outputs synchronous
with the display.
The analog output has possibility of the output variation when the power is turned
on. To take the stable condition, use the instrument about one hour after the power
has been turned on.

7121. Scaling of analog output


The analog output for standard specifications is set between the minimum value and the
maximum value with the output of 0 to 2 000.
By changing the F21 and F22, optional value can be decided.
Approx.21.6 mA
Approx.11 V

Analog output

The maximum 20 mA
10 V
value
110 % of the
maximum display

The minimum
value

4 mA
0V

Display
The maximum
display
110 % of the
maximum display

0 display
Approx.2.4 mA
Approx.11 V

88

F21 sets the display when the minimum value(standard voltage output0 V, optional current
output4 mA) is desired to output. F20 sets the display when the maximum value(standard
voltage output10 V, optional current output20 mA) is desired to output.
EX) F21Set as 1 000
F22Set as 5 000
When the display is 5 000, the maximum value outputs.
When the display is 1 000, the minimum value outputs.
Analog output
Approx.21.6 mA
Approx.11 V
The maximum value 20 mA
10 V

The minimum value 4 mA


0V

Warning

1 000
Approx.2.4 mA
Approx.11 V

5 000

Display

5 500

Take care that the setting on the F22 doesnt exceed the maximum display value
that has set in the Chapter 5. Meantime, please put the set value of F21 into the
condition of (the absolute value of F21 setting value) the set value of F22. If
neglected the correct scaling cant be obtained.

89

7122. Fine adjustment 1 on analog output


Fine adjustment described here, is the one to arrange each the minimum value and the
maximum value without applying the actual load during the procedures. Refer to the
paragraph 7123, for the fine adjustment with actual load applied.

During the application on fine adjustment, if you want to suspend, press the
key. The data of the minimum value and the maximum value are kept as they were
before entering the fine adjustment, and the Measurement mode can be
reentered.
Make fine adjustment one (1) hour or so after feeding the power.
You can make fine adjustment with safer conditions.

Procedures

Press the
key for about one second.
The setting display section shows FUNC.
Load display section continuously displays a
present load value.

Press the
key 8 times.
The setting display section changes as FUNC
CCALLCALZEROSPAN
TARECHEKMONTVCAL.

Press the
key.
The load display section shows ALOW.

90

Procedures

Press the
key.
The load display section flashes on and off
showing ALOW. At the same time, the analog
output equivalent to the minimum value of
analog output will be obtained.
Adjust so the output will become the minimum
value with the right keys.

Voltage output
(Standard)
Minimum
value

0V

Current output
(Optional)
4 mA

By pressing the key continuously,


increment can be provided continuously

key Decrease the analog output


key Increase the analog output

Fine adjustment on the maximum value of


analog output

Press the
key.
The load display section flashes display showing
A_HI.

Press the
key.
The load display section flashes on and off
showing A_HI.
At the same time, the analog output equivalent
to the maximum value of analog output will be
obtained.
Adjust so the output will become the maximum
value with the right keys.
Voltage output
(Standard)
Maximum
value

10 V

Current output
(Optional)
20 mA

By pressing the key continuously,


increment can be provided continuously

91

key Decrease the analog output


key Increase the analog output

Procedures

Press the
key.
The load display section shows END.

By pressing the
key again, the
Measurement mode can be returned through the
VCAL mode, then the present load is displayed.
At this moment, the result of fine adjustment on
the minimum/maximum output of analog output
can be renewed.

92

7123. Fine adjustment 2 on analog output


The fine adjustment explained here is the procedures with applying the actual weight.

Warning

Before making the fine adjustment, be sure to make scaling for the analog output
by referring to the paragraph 7121. If neglected, deviation of output cant be
adjusted during the fine adjustment.

During the application on fine adjustment, if you want to suspend the fine

adjustment, press the


key. The zero data and the span data are kept as they
were before entering the fine adjustment, and the Measurement mode can be
reentered.
Make fine adjustment in one (1) hour or so after feeding the power.
You can make fine adjustment with safer conditions.

Procedures

Press the
key for about one second.
The setting display section shows FUNC.
Load display section continuously displays a
present load value.

Press the
key 9 times.
The setting display section changes as FUNC
CCALLCALZEROSPAN
TARECHEKMONTVCAL
VADJ.

Press the
key.
The load display section shows ALOW.
At the same time, set the weight equivalent to
the minimum output value(standard voltage
output0 V, optional current output4 mA) of
the analog output.

93

Procedures

Press the
key.
The load display section flashes on and off
showing the current load value.
At the same time, adjust the deviation with the
right keys so the analog output meets with the
actual weight.

By pressing the key continuously,


increment can be provided continuously.

key Decrease the analog output


key Increase the analog output
Fine adjustment on the maximum value of
analog output

key.
Press the
The load display section shows A_HI.
At the same time, set the weight equivalent to
the maximum output value(standard voltage
output10 V, optional current output20 mA)
of the analog output.

Press the
key.
The load display section shows the current load
value and flashes on and off.
At the same time, adjust the deviation of analog
output against the load with the right keys.

By pressing the key continuously,


increment can be provided continuously.

key Decrease the analog output


key Increase the analog output

94

Procedures

Press the
key.
The load display section shows END.

By pressing the
key again, the
Measurement mode can be returned through the
VADJ mode, then the present load will be shown.
At this moment, the result of fine adjustment on
the minimum/maximum output of analog output
can be renewed.

7124. Selection of the target of analog output


The instrument can select the target of analog output from the TRACK and the PEAK.
This selection can be made in the Function mode(Related function F20).

95

713. Memory location for setting data and so on


The instrument can memorize each data at the RAM and the EEPROM as follows:
Since the EEPROM is nonvolatile, it will be stored almost indefinitely.
Besides, the RAM is not a backup type, so the RAM data will be disappeared by turning OFF the
power supply.
1

Data memorized at the RAM


ZERO tracking data

Data clear is possible with the execution of F98.

Data stored at the EEPROM


FUNC data
Possible to initialize data with the execution of F99
Calibration data
Possible to write by recalibration
Fine adjustment data for the analog output
Possible to write by the fine adjustment again
Each set value for the comparator S0, S1, S2, S3 and S4
Possible to clear with the change of each set value

The memory locations for the ZERO data can be changed to whichever the RAM
or the EEPROM with the setting Function F71.

714. Prohibition of calibration


The instrument prepares the setting for prohibition of calibration to prevent from excessive
calibrations. This setting can be made in the Function mode.(Related Function F97.)
As for the default, Possible to calibrate has selected. The targets of prohibition are each
calibration described in the Chapter 5, and each fine adjustment on the analog output described in
the paragraph 7122 and the paragraph 7123.
When executing the calibration with the setting of Prohibition of calibration, the ER6 will be
displayed.

96

715. Check mode


The following confirmations can be made in the Check mode.
Check on ROM version
Check on the option installed.
Check on bridge voltage
Check on the external control input
Check on the open collector output
Check on the analog output
Check on the BCD output (option)

The check on the BCD output can be operated only when the BCD output is
installed.
The instrument can return to the Measurement mode by pressing the
even in the halfway of the Check mode.

key

7151. Operating procedure for the check mode

When the
key is pressed with the setting display section of FUNC, the
display will change as the following arrow marks indicate at every time the key is

pressed. However, every time the


key is pressed, the display will change as the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJSYMMFUNCCCAL
Hereinafter over and over again.

Procedures

Press the
key for one second.
The setting display section shows FUNC.
Load display section continuously displays a
present load value.

Press the
key 6 times.
The setting display section changes as FUNC
CCALLCALZEROSPAN
TARECHEK.

97

Procedures
Check on the ROM version

Press the
key.
The Check mode can be entered, and the
display on the load display section shows

key again, the


ROM. By pressing the
ROM version is shown on the load display
section.

In the Check mode, when the display on


the load display section will be the one in
the below, the display can be changed by
the

key operation.
ROM

OP

BV

IN

SOUT

AOUT

BCD

END

ROM

key

98

Procedures
Check on the options

Press the
key.
The load display section shows OP.
Check on bridge power supply voltage

key again, the load


By pressing the
display section shows the optional number of
the option installed at present on the load
display. When no option is installed, the
P is appeared.

Display

Installed options

07

Current output

15

BCD output

74

RS232C interface

76

RS422/485 interface

Check on the bridge power supply voltage

Press the
key.
The load display section shows BV.

key again, the bridge


By pressing the
power supply voltage selected at present will
be seen on the load display section.

99

Procedures
Check on the external control input

Press the
key.
The load display section shows IN.

key again, the load


By pressing the
display section changes into the IN flashing
display. At the same time, the ON/OFF
condition of external control input signal can
be monitored on the Status display LED.

MEAS. display
HOLD display
PEAK display
END display
SEL.1 display
SEL.2 display

ZERO input
HOLD input
PEAK/TRACK input
RESET input
SEL.1 input
SEL.2 input

Check on the contact output

Press the
key.
The load display section show SOUT.

key again, the load


By pressing the
display section flashes S0.
At the same time, by the operations of right
keys, each judgement display for each open
collector output and the load display section
changes as below:

S0

S1

S2

S3

S4

END

key
key

Set ON the each contact output


from the above to the below in
order on the left figure.

Set ON the each contact output


from the below to the above in
order on the left figure.
S0 display
S0 output
S1 display
S1 output
S2 display
S2 output
S3 display
S3 output
S4 display
S4 output
END display END output

100

Procedures
Check on the analog output

Press the
key.
The load display section shows AOUT.

key again, the display


By pressing the
shows LOW and also flashes on and off.
At this moment the analog output will output
the value equal to the LOW value shown in
the below table.

key is pressed, the display


Every time the
will change as the arrow marks indicate, and
g output
p will change
g as shown in
also the analog
the below table.
LOW

MID

HIGH

Display

Standard
Option
(Voltage output) (Current output)

LOW

0V

4 mA

MID

5V

12 mA

HIGH

10 V

20 mA

The output values here reflect the values


adjusted in the paragraph 7122, and
7123.

key Change the analog output from


up to down in the left column.

key Change the analog output from


down to up in the left column.

Check on the BCD output

In case that the BCD output isnt installed,


shift to the step 10.

Press the
key.
The load display section shows BCD.

By pressing the
key again, the display shows
00000 and the digit of 100 flashes on and off.
At this moment, the same value as the display
outputs from the BCD output. When changing
the display with the right keys, the BCD output
changes corresponding to it.

101

key Set value carry key


key Set value increment key.

Procedures

Press the
key.
The load display section shows END.

By pressing the
key again, it will over the
Check mode and returns to the Measurement
mode to show the load value.

10

102

716. Monitor mode


In the Monitor mode, the applied load on the strain gage applied transducer at present can be
displayed with the converted unit of mV/V.
For example, in case that the load cell is used, and its output value is unclear, apply actual load in
order to read the output value at the time of initial load application and also at the time of the
maximum load application by using the function and then make calibration with the obtained
value as a base.

The display value in the Monitor mode is a reference value.


The accuracy of display is 0.5 % approximately.
In the Monitor mode, the range which can be monitored is from 3.1 mV/V to
3.1 mV/V approximately.
A minimum display in the monitor mode is five count increment.
Therefore, the value of minimum display digit becomes either of 0 or 5.

key is pressed with the setting display section of FUNC, the


When the
display will change as the following arrow marks indicate at every time the key is

is pressed, the display will change as the


pressed. However, every time the
reverse direction of the following arrow marks.
FUNCCCALLCALZEROSPANTARECHEK
MONTVCALVADJFUNCSYMMCCAL
(Hereinafter, over and over again.)

Procedures

Press the
key for one second.
The setting display section will show FUNC.
The load display section continuously displays a
present load value.

Press the
key 7 times.
The setting display section changes as CCAL
LCALZEROSPANTARE
CHEKMONT.

103

Procedures

Press the
key.
The Monitor mode can be entered, and the
converted value into mV/V for the present input
value for the transducer will flash on and off on
the load display section.

Press the
key.
The load display section shows END.

Press the
key.
The Monitor mode can be over, and the present
load is shown on the load display section.

104

8. Function mode

A set change by the function mode is possible in the measurement mode.


When the display is treated as a target of HOLD (F10), the Function mode cant
be entered with the condition of inputting the HOLD signal. Before entering
Function mode, be sure to cancel the input of HOLD.

81. Setting method for function mode


Procedures

Press the
key for one second.
The setting display section shows FUNC.
The load display section continuously displays a
present load value.

Press the
key. Function mode can be
entered, then the setting display section shows
F**, and the digit of 100 will flash on and off.
The last called Function No. is shown at **.
Suspend the setting of Function mode, then press

key when returning to the Measurement


the
mode is required.

Select the Function No. desired to set with the


right keys.
Suspend the setting of Function mode, then
Measurement mode can be reentered by

pressing the

key.

By pressing the key continuously,


continuous increase will be provided.

105

key Set value carry key


key Set value increment key
key Set value initialization key

Procedures

Press the
key.
Content of setting Function that has selected is
displayed and the digit of 100 will flash on and off
in the load display section.
Change the setting with the right keys.

key to suspend the setting of


Press the
Function mode, then return to the measurement
mode.

By pressing the key continuously,

key Set value carry key


key Set value increment key
key Set value initialization key

continuous increase will be provided.

Press the
key.
The set contents are registered, then the load
display section returns to the registered Function
No., and the 100 digit will flash on and off.

Press the
key.
Setting another Function No. is desired, return
to step 3.

Press the
key.
Quitting the Function mode, the Measurement
mode can be returned.

106

82. Function of function data


D F01
Selection of decimal point at display position
Initial value00000 00000Non
00001101
00002102
00003103
00004104
D F02
Selection of A/D sampling rate
Initial value00003 00000100 times/s
00001500 times/s
000021 000 times/s
000032 000 times/s
D F03
Selection of display rate
Initial value00000 000004 times/s
0000120 times/s
0000250 times/s
00003100 times/s
D F04
Setting digital filter
Initial value00055 Setting range0000000088, Average of 2(m)2(n) times
100 digitAverage of 2n times
101 digitAverage of 2m times
Select the average times for digital filter. When the figure grows larger, the filter will
become stronger, then effects from vibrations and so on are shown scarcely on the
display. However, if too large figure is selected, the response to variation of input
sensor will become worse.
Using the digital filter where vibrations and so on may exist is effective for removing
the deflection on the display.
D F05
Setting analog filter
Initial value00002 0000010 Hz
00001100 Hz
000021 kHz
D F06

Setting key lock

Initial value00000

key
digit key
digit key
digit key
digit key

100 digit
101
102
103

104
Key lock cancellation is made by setting 0, and key lock is made by setting 1.

key is locked, and the key is cancelled only once


immediately after by pressing the key for more than 3 seconds with the key
Moreover, when the

pressed together.

107

D F10
Setting of the target of HOLD
Initial value11111 100 digitLoad display
101 digitComparator S0 open collector output, LED display
102 digitComparator S1, S2, S3 and S4 open collector output,
LED display
3
10 digitAnalog output
104 digitOptional BCD output
Out of the target at the setting 0, and target of HOLD is available at the setting 1.
D F11
Setting the CHECK value
Initial value00003 Setting range0000000020
Unit0.1 mV/V
Approx. 1.5 mV/V CHECK value at the setting 00015.
It is common to all of four kinds of calibration data and executed.
D F12
Setting bridge power supply voltage
Initial value00000 0000010 V
000015 V
000022.5 V
It is common to all of four kinds of calibration data and executed.
D F20
Setting the target of analog output
Initial value00000 00000TRACK
00001PEAK
It is common to all of four kinds of calibration data and executed.
D F21

Display value at the time of the minimum analog output value


(standard voltage output0 V, optional current output4 mA)
Initial value00000 Setting rage 9999999999
The selected each calibration data can be set.

D F22

Display value at the time of the maximum analog output value


(standard voltage output10 V, optional current output20 mA)
Initial value02000 Setting rage 9999999999
The selected each calibration data can be set.

D F30
Setting comparator operation
Initial value11111 0OFF
1ON
100 digitComparator S0
101 digitComparator S1
102 digitComparator S2
103 digitComparator S3
104 digitComparator S4
It is common to all of four kinds of calibration data and executed.

108

D F31
Setting the target of comparator
Initial value00000 0TRACK
1Synchronous with display
100 digitComparator S0
101 digitComparator S1
102 digitComparator S2
103 digitComparator S3
104 digitComparator S4
It is common to all of four kinds of calibration data and executed.
D F32
Setting the direction of comparator
Initial value00000 0Greater or equal
1less or equal
100 digitComparator S0
101 digitComparator S1
102 digitComparator S2
103 digitComparator S3
104 digitComparator S4
It is common to all of four kinds of calibration data and executed.
D F33
Setting applicable condition for the comparator S0
Initial value00000 00000ON when comparator S0 is in the condition of
Load display value(The maximum display value)
00001ON when both of S1 and S2 open collector output are off.
00002ON when both of S1 and S3 open collector output are off.
00003ON when both of S1 and S4 open collector output are off.
00004ON when both of S2 and S3 open collector output are off.
00005ON when both of S2 and S4 open collector output are off.
00006ON when both of S3 and S4 open collector output are off.
00007Operate whichever Open collector ON more than the set
value or Open collector ON below the set value.
00008Interlock with HOLD LED
00008Interlock with PEAK LED
00008Interlock with MEAS. LED
It is common to all of four kinds of calibration data and executed.
D F34
Setting the condition of Hysteresis operation for comparator
Initial value00000 00000ON delay
00001OFF delay
It is common to all of four kinds of calibration data and executed.
D F35
Setting Hysteresis data for comparator
Initial value00000 00000Hysteresis data width OFF
Setting rage0000000099
Unit1D
Data width of 99D at the setting of 00099.
It is common to all of four kinds of calibration data and executed.
D F36
Setting Hysteresis time width for comparator
Initial value00000 00000Hysteresis time width OFF
Setting rage0000000099
Unit0.1 s
Data width of 9.99 s at the setting of 00999
It is common to all of four kinds of calibration data and executed.

109

D F38
Setting light ON/OFF of comparative value display section
Initial value00000 00000Light ON
00001Light OFF
D F40
Setting the target of BCD output(Effective when the option is installed.)
Initial value00000 00000TRACK
00001PEAK
D F41
Setting the logic of BCD output (Effective when the option is installed.)
Initial value00000 00000Negative logic
00001Positive logic
D F42
Setting the polarity of BCD output (Effective when the option is installed.)
Initial value00000 00000Negative logic
00001Positive logic
D F43
Setting the logic of BCD flag output (Effective when the option is installed.)
Initial value00000 00000Negative logic
00001Positive logic
D F44
Setting the logic of BCD P.C. output (Effective when the option is installed.)
Initial value00000 00000Negative logic
00001Positive logic
D F45
Setting the BCD P.C. width(Effective when the option is installed.)
Initial value00004 00000125 ms
0000125 ms
0000210 ms
000035 ms
000042.5 ms
D F46
Setting BCD output times(Effective when the option is installed.)
Initial value00004 000004 times/s
0000120 times/s
0000250 times/s
00003100 times/s
00004200 times/s
D F50

Setting the operation mode of RS232C


(Effective when the option is installed.)
Initial value00001 00000Stream mode
00001Command mode
In the Command mode, the data are transferred from the host by the command.
In the Stream mode, the latest data is output continuously.

D F51

Setting the target of output at the time of stream mode of RS232C


(Effective when the option is installed.)
Initial value00000 00000TRACK
00001PEAK
00002Display interlock

D F52

Setting the baud rate of RS232C/422/485


(Effective when the option is installed.)
Initial value00003 000001 200 bps
000012 400 bps
000024 800 bps
000039 600 bps
0000419 200 bps
0000538 400 bps
110

D F53

Setting the data bit length and parity of RS232C/422/485


(Effective when the option is installed.)
Initial value00021 100 digitSetting data bit length
08 bit
17 bit
101 digitSetting parity
0No parity
1Even number parity
2Odd number parity

D F54

Setting the stop bit of RS232C/422/485


(Effective when the option is installed.)
Initial value00000 000001 bit
000012 bit

D F55

Setting the terminator of RS232C/422/485


(Effective when the option is installed.)
Initial value00001 00000CR
00001CR+LF

D F56

Setting the decimal point of sending code of RS232C/422/485


(Effective when the option is installed.)
Initial value00000 00000No decimal point
00001Yes of Decimal point

D F57
Setting the ID number of RS422/485(Effective when the option is installed.)
Initial value00000 Setting range0000000031
D F58
Changing RS422/485(Effective when the option is installed.)
Initial value00000 00000RS422
00001RS485
D F59
Setting the delay time of returning RS485
Initial value00005 Setting range0000000999
Unit
1 ms
The delay is 999 ms at the 00999 setting.
D F60
Setting the Peak, Bottom and Peak bottom hold
Initial value00000 00000Peak hold
00001Bottom hold
00002Peak bottom hold
00003Local minimal value hold
00004Local maximal value hold
00005Inflection point hold
D F61
Setting the kinds of detecting mode
Initial value00000 00000All zone specification
00001Zone specification
00002Time zone specification
00003Automatic start time zone specification
D F62
Setting referring point(Effective when peak/bottom holding condition)
Initial value00000 Setting range9999999999
D F63

Setting the detecting data width of local maximal/minimal value


(Effective when local maximal/minimal value holding condition)
Initial value00200 Setting range0000099999

111

D F64

Setting the detecting magnification of local maximal/minimal value


(Effective when local maximal/minimal value holding condition)
Initial value00015 Setting range0000000049
Unit0.1 times

D F65

Setting the detecting data width of inflection point


(Effective when inflection point detecting condition)
Initial value00100 Setting range9999999999

D F66

Setting the detecting time width A of inflection point


(Effective when inflection point detecting condition)
Initial value00100 Setting range0000000450
Unit1 ms

D F67

Setting the detecting time width B of inflection point


(Effective when inflection point detecting condition)
Initial value00100 Setting range0000000450
Unit1 ms

D F68

Setting the detecting time of holding


(Effective when the holding mode of time zone or automatic start time zone)
Initial value01000 Setting range0000009999
Unit1 ms

D F69

Setting the start level of automatic start


(Effective when automatic start time zone mode.)
Initial value01000 Setting range9999999999

D F71
Setting the storing area of ZERO data
Initial value00001 0RAM
1EEPROM
Selects whether the data at the time of ZERO is recorded at the EEPROM or
not. In case that the high frequency of use of ZERO, the durability times will be over,
so set 0. The durability times for the EEPROM will be approx. one million times.
It is common to all of four kinds of calibration data and executed.
D F72
Setting the effective time for external control input
Initial value00000 0000050 ms
0000120 ms
0000210 ms
000035 ms
000042 ms
000051 ms
D F75
Symmetric property compensation value(for reference)
The compensation value set when the symmetric property compensation is executed
can be referred.
A setting here cannot be changed.
The reference value here is a value to the calibration data which has been selected.
D F80
Setting to the changeover target of calibration data and comparator code
Initial value00000 0Changeover by the function(F81 or F82)
1Changeover by external control input
100 digitCalibration data
101 digitCode

112

D F81
Setting of the calibration data selection
Initial value00000 00000Calibration data 0
00001Calibration data 1
00002Calibration data 2
00003Calibration data 3
D F82
Setting of the comparator code selection
Initial value00000 00000Code 0
00001Code 1
00002Code 2
00003Code 3
D F90
Increment value (For reference)
The increment value set when the calibration is applied can be referred to.
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F91
The maximum display value (For reference)
The maximum display value set when the calibration is applied can be referred to.
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F92
The actual load value (For reference)
The actual load value set when the calibration is applied(LCAL) can be referred to.
When the calibration except LCAL is made, this value will not change.
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F93
Zero calibration value (For reference)
The input voltage value that has read as the initial load value at the time of executing
calibration can be referred to.
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F94
Span calibration value (For reference)
The input voltage value at the time of the maximum display can be referred to.
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F95
Internal calibration data value (for our maintenance)
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
Internal calibration data value (for our maintenance)
D F96
The setting cant be changed in the function.
The reference value here is the value for the selecting calibration data.
D F97
Prohibition of calibration
Initial value00000 00000Possible to calibrate
00001Prohibition from calibration
It is common to all of four kinds of calibration data and executed.

113

D F98
ZERO clear
Zero compensated data by zero set function can be cancelled.

key is pressed with F98 displayed, ZCLR can be displayed.


When the
(At the same time, the display lights on and off.)

At this point, press the


key when suspending the ZERO clear is desired.
Measurement mode can be returned and Zero clear will not be executed.

key is pressed while ZCLR display lights on and off, F98 display can
When the
be returned. Now, ZERO clear has completed.
The selected each calibration data is executed.
D F99
Memory clear
Setting from F01 to F97 recorded at EEPROM will return to the default value.

key is pressed with the display of F99, then FCLR display can be
obtained. (At the same time, the display lights on and off.) At this point, press the

When the

key when suspending memory clear is desired. Measurement mode can be returned and
Memory clear will not be executed.

When the
key is pressed during FCLR load display lights on and off, and after
about 1 second, it will become FUNC display and the operation of Memory clear has
completed.

Warning

Never use the following functions because they may destroy the functions at
internal of the instrument.
F00

F07

F08

F09

F13

F14

F15

F16

F17

F18

F19

F23

F24

F25

F26

F27

F28

F29

F37

F39

F47

F48

F49

F70

F73

F74

F76

F77

F78

F79

F83

F84

F85

F86

F87

F88

F89

114

9. Options
The options for the instrument are as follows:
1

Current output[4 mA to 20 mA]

P/N.CSD819CP07

BCD output

P/N.CSD819CP15

RS232C interface

P/N.CSD819CP74

RS422/485 interface

P/N.CSD819CP76

For the options from to , there are some limitations among their combinations as follows
P07

P15

P74

P76

P07

P15

P74

P76

Possible, Impossible

91. Current output

Refer to the paragraph 435, for the connecting method, and refer to the
paragraph 712, for their applications.

911. Related functions


F20

Selection of analog output target

TRACK and PEAK

F21

Display value when the analog


output is the minimum value.

Display value at approx. 4 mA output is


applied.

F22

Display value when the analog


output is the maximum value.

Display value at approx. 20 mA output is


applied.

912. Specifications for the current output


P/N.

CSD819CP07

(1) Output

Output

4 mA to 20 mA

Over range Approx. 2.4 mA with the display OL


Approx. 21.6 mA with the display OL
(2) Load resistance

260 or less

(3) Nonlinearity

0.025 %F.S.

(4) Resolution

As same as display resolution

(5) Output rate

Synchronous with the A/D sampling rate


(100 times/s, 500 times/s, 1 000 times/s, 2 000 times/s)

115

92. BCD output

Warning

When power is ON for the instrument with the external HOLD signal shorted, the
BCD output will be as follows:
1

Even when the Display is targeted for the HOLD with the F10, BCD outputs
00000 if the target for BCD output is set as Display.

When the BCD output is assumed to be the target of HOLD with the F10, BCD
output outputs 00000.
Other than the case above and , the present load value will output after
00000 has output.

In other than the Measurement mode, the BCD output will not be renewed. In due
course, the ERROR for the BCD output wont be ON in other than the mode of
the Measurement mode, so care is taken fully.

When
key is applied, pay attention to the following point. By the ON operation
of CHECK, the OL error display (BCD output is OVR.) might be shown.

921. Related function


F40

Setting the target of BCD output

TRACK, PEAK

F41

Setting output logic for BCD data

Negative logic, Positive logic

F42

Setting output logic for BCD polarity

Negative logic, Positive logic

F43

Setting output logic for BCD flag

Negative logic, Positive logic

F44

Setting output logic for BCD P.C.

Negative logic, Positive logic

F45

Setting the width of BCD P.C.

2.5 ms, 5 ms, 10 ms, 25 ms, 125 ms

F46

Setting the output rate of BCD

4 times/s, 20 times/s, 50 times/s,


100 times/s and 200 times/s

922. Specifications of BCD output


(1) Output logic
Relative function

Negative logic, Positive logic can be changeable by the related


functions F41, F42 and F43.

(2) Output data


BCD
POL.(Polarity)
P.C.(Print command)
ERROR
OVR.(Over)

5 digits parallel output


output ON at minus, and output OFF at plus.

(3) Input data


ZERO
PEAK/TRACK
RESET/A/Z OFF
HOLD
BCDENABLE
SEL.1 and SEL.2

key
Same as key
Same as key
Same as

Hold of display and BCD output


Compulsive OFF for related output with BCD (Highimpedance)
As same as SEL.1 and SEL.2 for external control input signal.
116

(4) Output target

Changeover of TRACK and PEAK can be available with


the related function F40.

(5) Output times

Changeover of 4 times/s, 20 times/s, 50 times/s, 100 times/s and


200times/s can be available with the related function F46.

923. Pin configurations for the BCD output connector


1
2
3
4
5

COM.
1100
2100
4100
8100

13
14
15
16
17

8102
1103
2103
4103
8103

25
26
27
28
29

ERROR
P.C.
HOLD
N.C.
SEL.1

6
7
8
9

1101
2101
4101
8101

18
19
20
21

1104
COM.
2104
4104

30
31
32
33

SEL.2
ZERO
PEAK/TRACK
RESET

10
11
12

1102
2102
4102

22
23
24

8104
POL.
OVR.

34
35
36

N.C.
BCDENABLE
N.C.

Suitable plug : 5730360 made by DDK

Never connects with the N.C. pins.


An internal circuit and photocoupler are insulated.

924. Equivalent circuit for input/output


(1) Equivalent circuit for input

430

2.2 k

Approx.12 V

0.022 F
INPUT
COM.

(2) Equivalent circuit for output


OUTPUT

COM.
VCEDC30 VIc20 mA MAX.

117

925. Timing chart

Warning

In case of reading each output data of the BCD output through the sequencer and
so on, read them at the timing of The edge from ON to OFF for the P.C. (Print
command). Moreover, read the Input response delay time for the reading
instrument such as a sequencer and so on, with full considerations. If neglected,
there may have the possibility that correct reading of data wont be obtained.

(1) Normal
200 times/sApprox.5 ms
100 times/sApprox.10 ms
50 times/sApprox.20 ms
20 times/sApprox.50 ms
4 times/sApprox.250 ms
ON

DATA

ON

ON

ON

POL.
Changeable from approx.2.5 ms, 5 ms, 10 ms, 25 ms and 125 ms by function

ON

ON

ON

ON

P.C.

At the time of data output of each P.C., DATA and POL., output transistor will
become ON(Negative logic electrically).
(2) When the data is overranged
DATA
POL.
ON

ON

ON

ON

ON

ON

P.C.
ON
OVR.

At the time of OVR output, output transistor will at the OVR signal will become
ON(Negative logic electrically). Moreover, for all of the DATA, output transistor
will become OFF (Positive logic electrically) at the time of OVR output. (However,
for the POL., normal OFF at the OL, and normal ON at the OL.
(3) When ERROR is occurred
DATA
POL.
ON

ON

ON

ON

ON

ON

P.C.
ON
ERROR

At the time of ERROR output, output transistor at ERROR signal will become
ON(Negative logic electrically). Moreover, for each DATA, POL., all of the output
transistor will become OFF at the time of ERROR output(Positive logic
electrically).
118

(4) When the HOLD signal is input


200 times/sApprox.5 ms at max.
200 times/sApprox.5 ms at max.
100 times/sApprox.10 ms at max. 100 times/sApprox.10 ms at max.
50 times/sApprox.20 ms at max.
50 times/sApprox.20 ms at max.
20 times/sApprox.50 ms at max.
20 times/sApprox. 50 ms at max.
4 times/sApprox.250 ms at max.
4 times/sApprox. 250 ms at max.
ON
HOLD
DATA
POL
P.C.

At the time of HOLD signal input, output transistor for the P.C. will be OFF
condition. (Positive logic electrically.)
After inputting the HOLD signal, it takes the following response times to freeze the
DATA, POL or cancellation of HOLD is executed.
At 200 times/sApprox.5 ms at maximumInput response time
At 100 times/sApprox.10 ms at maximumInput response time
At 50 times/sApprox.20 ms at maximumInput response time
At 20 times/sApprox.50 ms at maximumInput response time
At 4 times/sApprox.250 ms at maximumInput response time

926. Output condition

Setting output logic


Negative logic
Positive logic

Output data

Condition of
transistor

Yes

ON

PinCOM level
at the time of voltage
supply externally.
L

No

OFF

Yes

OFF

No

ON

927. Selection of output logic for P.C.(Print command), and of its width
Selection of the P.C. logic for BCD output, and its width can be made by setting
function. (Related functionF44, F45)
Select the P.C. width to meet with the display rate set by F03, and execute the suitable
selection according to the following list.
Display
rate(F46)

P.C. width
Approx.125 ms Approx.25 ms Approx.10 ms

Approx.5 ms

Approx.2.5 ms

4 times/s
20 times/s
50 times/s
100 times/s

200 times/s

Possible to use, Impossible to use.(P.C. output doesnt operate correctly.)

Warning

The selection of P.C. width is made adequately according to the display times set
with the F03. If neglected, the P.C. output will not operate correctly.
119

93. RS232C interface

Warning

In other than the mode of Measurement mode, such as in Function mode,


Calibration mode, Fine adjustment mode and Setting mode, the error command
will be send against the command from the Host.
When the CHECK switch is applied, pay attention to the following points.
1

By ON operation of the CHECK, the instrument isnt in the Measurement mode


anymore, however RS232C interface responses to the command from the host.

By ON operation of the CHECK, OL error display might be shown. At this


time, when reading command for load is executed, the OL will be transferred
to the host.

931. Related function


F50

Setting of operation mode


Setting of output target at the
time of stream mode
Setting of baud rate

F51
F52

Setting of data bit length &


parity
Setting of the stop bit
Setting of terminator
Setting of decimal point for
sending code

F53
F54
F55
F56

Stream mode, Command mode


TRACK, PEAK
1 200, 2 400, 4 800, 9 600, 19 200 or 38 400 bps
Parity bit Non, Even parity or Odd parity
Data length 7 bit or 8 bit
1 bit or 2 bit
CR or CR+LF
No or Yes

Setting of this function makes the setting activated just after setting.

932. Specifications for interface


(1) Method

Corresponds to RS232C

(2) Communication methodHalf duplex


(3) Specifications for signal
Baud rate

1 200, 2 400, 4 800, 9 600, 19 200 or 38 400 bps

Data length

7 bit or 8 bit

Parity bit

Non, Even parity or Odd parity

Stop bit

1 bit or 2 bit

Terminator

CR or CRLF

Synchronous method
Transmission data

Startstop synchronous method


ASCII code

(4) Cable length

within 15 m

(5) Input/Output monitor with LED

120

933. Procedures of data transfer


There are two kinds of data transfers in the instrument, that is, stream mode and command
mode.
(1) Stream mode
The latest targeted data targeted/selected with the F51 keep on outputting. However, output
times will change depending on the setting of display times and baud rate.
(2) Command mode
By sending the determined command/data from the host (Personal computer, sequencer and so
on) to the CSD819C, the data will be send back to the host side from the CSD819C
corresponding to the command/data.
Be sure to execute communication according to the below procedures.
Start

Command/Data

Data

HostCSD819C

CSD819CHost

The communication operation can be made in all of the modes. However, in the
Calibration mode, Fine adjustment mode and Setting mode, the Error command
will be sent.
The flow control is not executed in the CSD819C.
The CTS/RTS signal isnt applied.
The X flow control isnt performed.
The operating communication is a conversational dialogue type.

121

934. Pin configurations for connector pin


(1) Pin configuration
Pin No.

Signal Name

CD

TXD

RXD

N.C.

S.G.

N.C.

RTS

CTS

N.C.

Suitable plugDE9SNR by JAE or equivalent. Not attached.

The screws for the fixing base of plug at the connector of RS232C interface is inch
type thread.
An internal circuit and photocoupler are insulated.
(2) Example of connection at the RS232C interface
1

Example 1
Host(25 pin

CSD819C
1

CD

F.G.

TXD

TXD

RXD

RXD

N.C.

3
4

S.G.

CTS

N.C.

DSR

RTS

S.G.

CTS

DCD

N.C.

RTS

19
20

21
25

122

DTR

Example 2
Host(9 pins)

CSD819C
1

CD

DCD

TXD

RXD

3
4

RXD

TXD

N.C.

3
4

S.G.

S.G.

N.C.

DSR

RTS

RTS

CTS

CTS

N.C.

RI

DTR

935. Data format


(1) Stream mode
2
@

4
0

10

12

14

Terminator

Sign of
load data

Command No.

Load data

D The setting with the F55 can be entered to the terminator.


D Load weighing data enters from the right end.
D When the data is minus, sign, and when plus, sign is added.
D Load weighing data performs zero suppress.
D In case that the decimal point is set with the F56, and at the same time the
decimal point is specified with the F01, it will be added to the specified
position.
D The message will be output at the time of overloaded.
2
@

D The empty sections are all spaces.


4
6
8
10
0

2
@

10

12
O

123

12
O

14
3

12

10

6
9

10
0

4
0

2
@

2
@

12

Terminator

14
3

Terminator

14
Terminator

14
Terminator

(2) Command mode


1

Reading out the load data(HostCSD819C)


2
@

4
0

6
2

Terminator

20

Reading out the display


interlocked data

Command No.

21

Reading out TRACK

22

Reading out PEAK

Return(CSD819CHost)
2
4
6
@

Operation

Command No.

10

12

14

Terminator

Sign of load data


Command No.

Load data

D The load data enters from the right end.


D When the data is minus, sign and when plus, sign is added.
D Load weighing data performs zero suppress.
D In case that the decimal point is set with the F56, and at the same time the
decimal point is specified with the F01, it will be added to the specified
position.
D The message will be output at the time of overloaded.
2
@

D The empty sections are all spaces.


4
6
8
10
0

2
@

10
0

10

12
O

124

12
O

14
3

12

10

6
2

4
0

2
@

2
@

12

Terminator

14
3

Terminator

14
Terminator

14
Terminator

Reading out comparative data (HostCSD819C)


2
@

6
0

Command No.

Operation

30

Reading out S0 data

31

Reading out S1 data

32

Reading out S2 data

33

Reading out S3 data

34

Reading out S4 data

Terminator
Command No.

Return(CSD819CHost)
2
4
6
@

10

12

14

Command No.

Terminator
Sign

Set value

D The load data enters from the right end.


D When the data is minus, sign and when plus, sign is added.
D The sign data performs zero suppress.
D In case that the decimal point is set with the F56, and at the same time the
decimal point is specified with the F01, it will be added to the specified
position.
D The empty sections are all spaces.
3

Readingout the condition (HostCSD819C)


2
@

4
0

6
0

Command No.

Operation

40

Reading out the condition

Terminator
Command No.

Return(CSD819CHost)
2
4
@

6
a

8
b

10
d

Command No.
aSEL.1
bSEL.2
cCHECK
dHOLD
ePEAK
fMEAS.
gEND

LED display
LED display
LED display
LED display
LED display
LED display
LED display

12
f

14
Terminator
Space
Condition

1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF

125

Readingout the comparative results (HostCSD819C)


2
@

4
0

Command No.

6
1

41

Terminator

Operation
Reading out the comparative
result

Command No.
Return(CSD819CHost)
2
4
6
@

S0

8
S1

S2

10
S3

12

S4

14
Terminator

Command No.

Space
Condition

S01ON, 0OFF
S11ON, 0OFF
S21ON, 0OFF
S31ON, 0OFF
S41ON, 0OFF
5

Change of condition(HostCSD819C)
2
4
6
@

Command No.

Terminator

50

Operation
Zero set

Command No.

51
52

PEAK/TRACK
RESET

Return(CSD819CHost)
2
4
6
@

Terminator
Command No.

Writing comparative data(HostCSD819C)


2
@

4
0

10

12

Command No.

14
Terminator
Set value

Sign
Command No.

Return(CSD819CHost)
2
4
6
@

Command No.

Operation

60
61

Writing the S0 data


Writing the S1 data

62
63
64

Writing the S2 data


Writing the S3 data
Writing the S4 data

10

12

14
Terminator
Set value

Sign

D The set value enters from the right end.


D Setting range is from 99 000 to 99 999.
D Never add the decimal point.

126

Writing the function data of peak function section(HostCSD819C)


2
@

4
0

6
0

12

Command No.
Function No.

14
Terminator
Setting value

Sign

Operation
Setting of holding mode (Change contents of F60)
00000 Peak hold
00001 Bottom hold
00002 Peak/bottom hold
00003 Local minimal value hold
00004 Local maximal value hold
00005 Inflection point hold
Setting of kinds for detecting mode
(Change contents of F61)
00000 All zone specification
00001 Zone specification
00002 Time zone specification
00003 Automatic start time specification
Setting of standard point (Change contents of F62)
Setting range9999999999 Unit1D
Setting of local maximal/minimal data width
(Change contents of F63)
Setting range0000099999 Unit1D

Setting value
0000000005

64

Setting of local maximal/minimal detection


magnification (Change contents of F64)
Setting range049 Unit0.1 times

0000000049

65

Setting of inflection point detection data width


(Change contents of F65)
Setting range9999999999 Unit1D

9999999999

66

Setting of inflection point detection time width A


(Change contents of F66)
Setting range0450 Unit1 ms

0000000450

67

Setting of inflection point detection time width B


(Change contents of F67)
Setting range0450 Unit1 ms

0000000450

68

Setting of detection time of holding


(Change contents of F68)
Setting range09999 Unit1 ms

0000009999

69

Setting automatic start level


(Change contents of F69)
Setting range9999999999

9999999999

Function No.
60

61

62
63

Return(CSD819CHost)
2
4
6
@

10

Command No.
Function No.

10

0000000003

9999999999
0000099999

Unit1D
12

14
Terminator

Sign

D The set value enters from the right end.


D Never add the decimal point.

127

Setting value

Changeover of calibration data and comparator code(HostCSD819C)


2
@

4
0

6
0

81

82

14
Terminator
Setting value

Space

Operation
Setting of calibration data selection
00000Calibration data 0
00001Calibration data 1
00002Calibration data 2
00003Calibration data 3

Setting value
0000000003

Setting of comparator code selection


00000Code 0
00001Code 1
00002Code 2
00003Code 3

0000000003

Return (CSD819CHost)
2
4
6
@

12

Command No.
Function No.
Function No.

10

10

12

Command No.
Function No.

14
Terminator
Setting value

Space

936. Communication error processing


This instrument replies error command to the host side when the communication error or the
execution is occurred.
2
@

4
0

6
R

8
1

Terminator
Error command No.

Error
Command No.
01
02
10
11
12
13
14

Contents
Nonexecutable error
Additionally, error by
this unit convenience
Parity error
Framing error
Overrunning error
Data code and data
length error
No applicable command

Remarks
For the calibration mode, fine adjustment mode,
check mode and monitor mode.
When the reception command cannot be
executed.
At the parity detection error
At the stop bit detection error
At the reception command reading error
When neither the received data code nor data
length are corresponding
When reception command isnt corresponding

Error code is not replied when the completion code (terminator) cannot be
detected.
Please consider dealing on the host side when the communication error command
is replied from this unit.

128

94. RS422/485 interface

Warning

The error code is sent against the command from the host during other than the
mode of Measurement mode, such as in the Calibration mode, Fine adjustment
mode, Check mode and Monitor mode.
When CHECK switch is ON status, be careful on the following points.
1

When CHECK switch will be ON status, the instrument will be out of the
Measurement mode, but RS422/485 interface will make response to the
command from the Host.

By ON operation of CHECK, there may be a case that display will show as OL


error. At the same time, executing the reading command for load makes the
OL transmitted to the Host.

The sending terminal is low impedance for 1 ms after finishing sending in the
instrument. Therefore, to prevent from competing with other instruments, apply
the sending wire (SDA, SDB) at the intervals of 1ms or more.

941. Related functions

F52

Setting baud rates

1 200, 2 400, 4 800, 9 600, 19 200 or 38 400 bps

F53

Setting a data bit length and


a parity bit

Parity bit Non, Even or Odd parity


Data length 7 bit or 8 bit

F54

Setting a stop bit

1 bit or 2 bit

F55

Setting the terminator

CR or CR+LF

F56

Setting the decimal point in


transmitting code

Non or Exist

F57

Setting the ID No.

0 to 31

F58

Change of RS421/485

Operation of RS422 or RS485

F59

Setting the delay time for


sending back the RS485

0 to 999

every 1 ms (unit)

Setting this function becomes effective immediately after setting.


When RS485 is operated, function F59 becomes effective.
After completing the transmission at the host side by the function, set the time
until the transmitting terminal of the instrument becomes high impedance.

129

942. Specifications on interface


(1) Method

Based on RS422/485

(2) Communication methodHalfduplex


(3) Specifications on Signal
Baud rate

1 200, 2 400, 4 800, 9 600, 19 200 or 38 400 bps

Data bit length

7 bit or 8 bit

Parity bit

None, Even or Odd parity

Stop bit

1 bit or 2 bit

Terminator

CR or CRLF

Synchronous method
Communication data

Startstop synchronous method


ASCII code

Address

00 to 31

(4) Cable length

Approx.1 km

(5) Number of connections 32 sets at maximum (RS422: 10 sets)


(6) Termination

Builtin
(Yes/No can be selectable by the connection of terminal blocks.)

(7) Change of RS422/485 Setting by the function.


(8) Input/output monitor

With LED
(The layout is shown on the P.C. board at rear side of terminal block.)

943. Procedure of data transmission


By sending the determined command/data from the host (personal computer, sequencer and so
on) to the CSD819C, data will be sent back to the host side form the CSD819C
corresponding to the command/data.
Be sure to execute communication according to the below procedures.
Start

Command/Data

Data

HostCSD819C

CSD819CHost

The communicating operation can be made in all of the modes. However, the error
command is transmitted in the Calibration mode, Fine adjustment mode, Check
mode and Monitor mode.
The flow control is not executed in the CSD819C.
The CTS/RTS signal isnt applied.
The X flow control isnt performed.
The communication operation is a conversational type.
130

944. Pin layout and wiring of Connector


(1) Pin configuration
SDA
SDB

Differential output
Differential output

RDA
RDB
TRM.
S.G.

Differential input
Differential input
Terminator
Signal ground

The TRM. is a terminal resistance. Connect the terminal resistance by shorting


between the TRM. and RDM. at the last end of the host looking from the
host(personal computer, sequencer and so on).
For the connection, we recommend to apply twisted pair wires.
An internal circuit and photocoupler are insulated.
(2) Example of connection
1

1 to 1
RS422
Host

CSD819C
RDA
RDB

SDA
SDB

SDA
SDB

RDA
RDB

RS485
CSD819C

Host

RDA
RDB

SDA
SDB

SDA
SDB

RDA
RDB

Connect the terminal resistance at the

131

section.

1 to n

RDA
RDB

CSD819C
SDA
SDB

RDA
RDB

SDA
SDB

RDA
RDB

CSD819C

Host

Host

Connect the terminal resistance at the

section

132

RDA
RDB

CSD819C
SDA
SDB

RDA
RDB

CSD819C
SDA
SDB

CSD819C
RDA
RDB

RDA
RDB
SDA
SDB

SDA
SDB

RDA
RDB
SDA
SDB

SDA
SDB

RS485
CSD819C

CSD819C

SDA
SDB

SDA
SDB
RDA
RDB

RS422
CSD819C

RDA
RDB

945. Data format


(1) Command mode
1

Reading out the load data(HostCSD819C)


2
@

4
0

6
2

Terminator
Command No.

ID No.
0031

Return(CSD819CHost)
2
4
6
@

Operation

Command No.

20

Reading out the display


interlocked data

21

Reading out the TRACK

22

Reading out the PEAK

10

12

14

ID No.
0031

Terminator

Sign of load data


Command No.

Load data

D The comparative data enters from the right end.


D When the data is minus, sign and when plus, sign is added.
D The sign data perform zero suppress.
D When the decimal point is set with the F56 and at the same time the decimal
point is determined with the F01, it will be added to the specified position.
D The message will be output at the time of overload.
2
@

D The empty sections are all spaces.


4
6
8
10
0

ID No.
0031 2
@

ID No.
0031
@

10
O

ID No.
0031

133

14
3

12

10

6
2

10
0

4
0

2
0

ID No.
0031 2
@

12

Terminator

14
3

Terminator

12

14

Terminator

12

14

Terminator

Readingout the comparative data (HostCSD819C)


2
@

6
0

Operation

30

Reading out S0 data

31

Reading out S1 data

32

Reading out S2 data

33

Reading out S3 data

34

Reading out S4 data

Terminator
Command No.

ID No.
0031

Return(CSD819CHost)
2
4
6
@

Command No.

10

12

14

ID No.
0031

Terminator
Sign
Command No.

Setting value

D The comparative data enter from the right end.


D When the data is minus, sign and when plus, sign is added.
D The sign data perform zero suppress.
D When the decimal point is set with the F56 and at the same time the decimal
point is determined with the F01, it will be added to the specified position.
D The empty sections are all spaces.
3

Readingout the status (HostCSD819C)


2
@

4
0

6
0

40

Reading out the status

10
d

12
f

14
Terminator
Space
Condition

ID No.
0031
aSEL.1
bSEL.2
cCHECK
dHOLD
ePEAK
fMEAS.
gEND

Operation

Terminator

Command No.
ID No.
0031
Return(CSD819CHost)
2
4
6
8
@

Command No.

LED display
LED display
LED display
LED display
LED display
LED display
LED display

Command No.
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF
1ON, 0OFF

134

Readingout the comparative result(HostCSD819C)


2
4
6
Command No.
@

Terminator

41

Command No.
ID No.
0031
Return(CSD819CHost)
2
4
6
8
@

S0

S1

S2

10
S3

Terminator
Space

Command No.
50
51
52

Terminator

Command No.
ID No.
0031
Return(CSD819CHost)
2
4
6
@

Sign
Command No.

ID No.
0031

14
Terminator

Return(CSD819CHost)
2
4
6
@

12

ID No.
0031

Operation
Zero set
PEAK/TRACK
RESET

Terminator

Command No.
ID No.
0031
6 Writing comparative data (HostCSD819C)
2
4
6
8
10
@

14

Condition

Change of status(HostCSD819C)
2
4
6
@

12

S4

ID No.
0031
Command No.
S01ON, 0OFF
S11ON, 0OFF
S21ON, 0OFF
S31ON, 0OFF
S41ON, 0OFF
5

Operation
Read out comparative result

Set value
Command No.

10

60

Operation
Writing S0 data

61
62
63
64

Writing S1 data
Writing S2 data
Writing S3 data
Writing S4 data
12

14
Terminator

Sign
Command No.
D The comparative data enters from the right end.
D The setting range is within 99 999 to 99 999.
D Dont add a decimal point.
135

Set value

Writing function data of peak function part(HostCSD819C)


2
@

6
0

10

12

ID No.
0031
Command No.
Function No.

14
Terminator
Set value

Sign

Operation
Setting hold mode (Change contents of F60)
00000Peak hold
00001Bottom hold
00002Peak/bottom hold
00003Local minimal value hold
00004Local maximal value hold
00005Inflection point hold
Setting the kind of detecting mode
(Change contents of F61)
00000All zone specification
00001Zone specification
00002Time zone specification
00003Automatic start time
Setting of standard point (Change contents of F62)
Setting range9999999999 Unit1D
Setting data width of local maximal/minimal value
(Change contents of F63)
Setting range0000099999 Unit1D

Setting value
0000000005

64

Setting of local maximal/minimal detection magnification


(Change contents of F64)
Setting range049 Unit0.1 times

0000000049

65

Setting data width for inflection point


(Change contents of F65)
Setting range9999999999 Unit1D

9999999999

66

Setting detecting time width A of inflecting point


(Change contents of F66)
Setting range0450 Unit1 ms

0000000450

67

Setting detecting time width B of inflecting point


(Change contents of F67)
Setting range0450 Unit1 ms

0000000450

68

Setting hold detecting time(Change contents of F68)


Setting range09999 Unit1 ms
Setting automatic start level (Change contents of F69)
Setting range9999999999 Unit1D

0000009999

Function No.
60

61

62
63

69

Return(CSD819CHost)
2
4
6
@

ID No.
0031
Command No.
Function No.

10

12

0000000003

9999999999
0000099999

9999999999

14
Terminator
Set value

Sign

D The comparative data enters from the right end.


D Dont add a decimal point.

136

Changeover of calibration code(HostCSD819C)


2
@

4
0

6
0

10

12

14
Terminator

ID No.
0031
Command No.

Set value

Space
Function No.

Function No.
81

Operation
Setting selection of calibration data
00000Calibration data 0
00001Calibration data 1
00002Calibration data 2
00003Calibration data 3

Setting value
0000000003

82

Setting selection of comparator code


00000Code 0
00001Code 1
00002Code 2
00003Code 3

0000000003

Return(CSD819CHost)
2
4
6
@

10

12

14
Terminator

ID No.
0031
Command No.

Space
Function No.

Set value

946. Process of communication error


This instrument returns error command to the host when the communication error or
execution error occurs.
2
@

4
0

6
R

8
1

Terminator

ID No.
0031

Error command No.

Error
Command No.
01
02
10
11
12
13
14

Contents
Nonexecutable error
Additionally, error by
this unit convenience
Parity error
Framing error
Overrunning error
Data code and data
length error
No applicable command

Remarks
For the calibration mode, fine adjustment mode,
check mode and monitor mode.
When the reception command cannot be
executed.
At the parity detection error
At the stop bit detection error
At the reception command reading error
When neither the received data code nor data
length are corresponding
When reception command isnt corresponding

Error code is not replied when the completion code (terminator) cannot be
detected.
Please consider dealing on the host side when the communication error command
is replied from this unit.

137

10. Trouble shooting


When abnormal point(s) is/are found during the operation of the instrument, check by the following
procedures. However, when you cant find applicable item nor solve the symptom of trouble even after you
have taken some measures, contact with Minebea.

Supply power, again.

Operation

NG

OK

Abnormal operation

Normal Operation
Execute trouble
shooting.
Start measurement

138

101. Execute trouble shooting

Execute trouble shooting.

NO

Does supply power voltage


satisfy the specifications?

Supply power source suitable for


the specifications.

YES

YES

Display is wrong.

NO

Operation of
comparator is wrong.

YES

NO

YES

External control input


doesnt operate.
NO

Connecting method
is good.
YES

Key operation
doesnt work.
NO

YES
4

YES

Analog output
is wrong.
NO

Operation of
options is wrong.

NO

YES
Inform Minebea about the contents of
failure and situation at site in details.
Check the options.

139

NO

Displays the load.

NO

YES
Load display
doesnt vary.

YES

NO
Display fluctuates
abnormally.

Displays
optional values.

YES

NO

NO

8
OL display

Execute the calibration again


according to the procedures in
the paragraph 5.

NO

YES

YES
7

NO

OL display

YES

Same situation
NO

YES

Start measurement.
Fuse is broken.
NO

Inform Minebea about the contents of


failure and situation at site in details.

140

YES

Replace fuse by
referring to the
paragraph 131.

Set value is good.

NO

Confirmation/change for set value is made


according to the paragraph 723.

NO

Confirmation/change for operation of


comparator, comparative target for
comparator is made according to the
paragraph 724, 725 and 726.

YES

Operation of
comparator is the same as
its intention.
YES

YES
Contact output chatters.

Apply setting for hysteresis according


to the paragraph 727.

NO

YES
Contact output isnt ON.

Following causes can be considered.


1 Failure in contact by using other than the
rated capacity of contact.
Capacity of contactVCEDC30 V,
IC30 mA at max.
2

NO

NO
Connecting method is good.
YES

Inform Minebea about the contents of


failure and situation at site in details.

141

Failure in customers input circuit.


Confirm the contact output signal by
referring to the paragraph 715.

Connect securely according to the paragraph


433 and 712.

ZERO input
doesnt operate.

YES

NO

YES Other than the range of Zero


set.

Displays ER0.

Range of zero set is within

NO

PEAK/TRACK
input doesnt work.

10 % of the maximum
display value.

YES

NO
RESET input
doesnt work.
NO

HOLD input
doesnt work.

YES
Either the HOLD
of external input or BCD
output is used.

YES

YES

Set the HOLD input open.

NO

NO
Connecting method
is good.
YES

Inform Minebea about the


contents of failure and
situation at site in details.

142

NO

Connect securely according to


the paragraph 432 and
711.

YES

Key lock is set with the


function F06.

Cancel the key lock with the function


F06.

NO

YES

HOLD of external
control input is used.
NO

Inform Minebea about the contents of


failure and situation at site in details.

143

Set the HOLD input open.

Option of
current output

YES

NO

10
NO

Load resistance is
5 k or more.

Load resistance is 5 k or more.

YES
1
2

Remove the connecting cable between the AOUT terminals.


Measure the voltage between the AOUT terminal and .

Set the connecting range to DCV range for the measuring instrument
3

such as tester and so on.


Apply operational check on voltage output in the Check mode.
(Refer to the paragraph 715.)

NO

Varies to 0 V, 5 V
and 10 V.

Contact with Minebea.

YES
1
2

Put the connecting cable for the AOUT terminal where it was.
Check the setting related with voltage output.
F21 Display value when analog output is the minimum.
F22 Display value when analog output is the maximum.
(Refer to the paragraph 8.)

Setting is correct.

NO

Set correctly.
Refer to the chapter 8.

YES

HOLD for external


control input is applied.

YES
Set the HOLD input open.

NO
Inform Minebea about the
contents of failure and
situation at site in details.

144

1
2

Remove the connecting cable for strain gage applied transducer from terminal block.
Measure the voltage between the AC terminals on the terminal block.

Set the connecting range to DCV range for the measuring instrument such as
tester and so on.

The voltage between


AC is the same as the setting
with F12.

NO

Contact with Minebea.

YES
1

Connect the connecting cable for strain gage applied transducer to the terminal
block again. (Refer to the chapter 4)

By checking the voltage between the terminal B and D on the terminal block, apply
load on the strain gage applied transducer.

Set the connecting range to DCmV range for the measuring instrument such as
tester and so on.

YES

The voltage between B


and D varies.

Make recalibration according


to the procedures in chapter 5.

NO

The same condition


YES

There may have fears that the


strain gage applied transducer
connected has broken or signal
route (signal wire) has cut off.

Inform Minebea about the


contents of failure and
situation at site in details.

145

NO

Start Measurement.

Remove the connecting cable for strain gage applied transducer from the terminal block.
Measure the voltage between the AC terminals on the terminal block.

Set the connecting range to DCV for the measuring instrument such as tester and so on.

The voltage between


AC is the same as the setting
with F12.

NO

Contact with Minebea.

YES
1
2

Connect the connecting cable for strain gage applied transducer to the terminal block again.
(Refer to the chapter 4.)
Set the load display section to be the monitoring condition for strain gage applied transducer
according to the paragraph 716.

Monitoring display
exceeds the range from 2.0 mV/V
to 2.0 mV/V.

YES

Short between BD on
the terminal block.

NO

Make recalibration
according to the procedures
of chapter 5.

Monitoring display
exceeds the range from 2.0 mV/V
to 2.0 mV/V.
YES

Same condition

Following causes can be considered


1

The input of strain gage applied


transducer exceeds the range
from 2.0 mV/V to 2.0 mV/V.

The strain gage applied


transducer has broken due to
overload and so on.
Signal route (signal wire) of the
strain gage applied transducer
has cut off.

YES

NO

Start measurement.

Inform Minebea about the


contents of failure and
situation at site in details.

146

NO

1
2

Remove the connecting cable for strain gage applied transducer from the terminal block.
Measure the voltage between the AC terminals on the terminal block.

Set the connecting range to DCV for the measuring instrument such as tester and so
on.

The voltage between


AC is the same as the setting
with F12 and stable
as well.

NO

Inform Minebea about the contents of


failure and situation at site in details.

YES

1
2

NO

Short the B, D and COM.1 on the terminal block.


Set the load display section to be the monitoring condition for strain gage
applied transducer according to the paragraph 715.

The display stabilizes


with the optional
value.
YES

Check the various kinds of connecting conditions


according to the procedures in the chapter 4.
Especially, check that there may have effects from
the noise source such as an inverter or like that.

NO

Same condition
YES

Start measurement.

Inform Minebea about the contents of


failure and situation at site in details.

147

102. Optional check


Check on options.

Current output is wrong.

YES

NO

BCD output is wrong.

10

YES

NO

RS232C is wrong.

11

YES

NO

RS422 is wrong.

12

YES

NO

RS485 is wrong.

13

YES

NO

14

Inform Minebea about the contents of


failure and situation at site in details.

148

10

NO

Load resistance is
260 or less.

Load resistance is 260 or less.

YES
1
2

Remove the connecting cable at the current output terminal.


Measure the voltage between the current output terminal + and .

Set the connecting range to DCmA range for the measuring instrument such as
3

tester and so on.


Apply operational check on current output in the Check mode.
(Refer to the paragraph 715.)

NO

Varies to 4 mA, 12 mA
and 20 mA.

Contact with Minebea.

YES

1
2

Put the connecting cable for the current output terminal where it was.
Check the setting related with current output.
F20 Selection of target for the analog output.
F21 Display value at the time of the minimum analog output.
F22 Display value at the time of the maximum analog output.
(Refer to the paragraph 712)

NO

Setting is correct.

Set correctly. Refer to the


paragraph 712, 8 and 91.

YES

External control
input or HOLD at BCD
output is applied.

YES

NO

Inform Minebea about the contents of


failure and situation at site in details.

149

Set the HOLD input open.

11

Set correctly.

NO

Connect with the external instrument by


referring to the paragraph 92

YES
Adjust the setting related with BCD output to the
specifications for external connecting instruments.
F41 Setting the output logic for BCD data
Other than the BCD
is output.

YES

F42 Setting the output logic for BCD


polarity
F43 Setting the output logic for BCD flag
F44 Setting the output logic for P.C.

NO

F45 Setting the P.C. width


F46 Setting BCD output rate
(Refer to the chapter 8.)
Intended value
isnt output.

YES

NO

BCD output will


not vary.

YES

Select the target of BCD output.


F40 Setting the target of BCD output
(Refer to the chapter 8.)

Set the HOLD input open.

NO

Inform Minebea about the contents of


failure and situation at site in details.

150

12

Connecting is good.

NO

Connect correctly by referring to the paragraph 93.

YES

Baud rate and so on


are set correctly.

Set the following items correctly by conforming to the


specifications for the host.
F52 Setting the baud rate

NO

F53 Setting the data bit length and the parity


F54 Setting the stop bit

YES

Setting for
the command/data are
correct.

F55 Setting terminator


(Refer to the paragraph 8 and 93.)

NO

Use the command/data correctly by referring to the


paragraph 93.

YES

Flow control is
specified by the host.

YES

The host doesnt make the flow control

YES

Select the output target in the stream mode.


F50 Setting in the Operation mode

NO

Intended value
is not transmitted in the
stream mode.

F51 Setting the output target in the Stream


mode
(Refer to the paragraph 8 and 93.)

NO

Inform Minebea about the contents of


failure and situation at site in details.

151

13

Connected correctly.

NO

Connect correctly by referring to the paragraph 94.

YES

ID Nos. are set


correctly.

NO

Set the ID Nos. correctly by conforming to the


paragraph 8 and 94.

YES

Baud rate and so on


are set correctly.

NO

F53 Setting the data bit length and the parity


F54 Setting the stop bit

YES

Setting for the


command/data are correct.

Set the following items correctly by conforming to the


specifications for the host.
F52 Setting the baud rate

F55 Setting terminator


(Refer to the paragraph 8 and 94.)
NO

Use the command/data correctly by referring to


the paragraph 94.

YES

Flow control
is specified by the host.

YES

The host doesnt make the flow control

NO

Change of RS422/485
is set correctly.

NO

Perform the change at RS422/485 correctly.


F58 Change of RS422/485

YES

Inform Minebea about the contents of


failure and situation at site in details.

152

14

Connected correctly.

NO

Connect correctly by referring to the paragraph 94.

NO

Set the ID Nos. correctly by conforming to the


paragraph 8 and 94.

YES
ID Nos. are set
correctly.
YES

Baud rate and so on


are set correctly.

Set the following items correctly by conforming to the


specifications for the host.
F52 Setting the baud rate

NO

F53 Setting the data bit length and the parity


F54 Setting the stop bit
F55 Setting terminator
(Refer to the paragraph 8 and 94.)

YES
Setting for the
command/data are correct.

NO
Use the command/data correctly by referring to
the paragraph 94.

YES
Flow control
is specified by the host.

YES

The host doesnt make the flow control

NO
Change of RS422/485
is set correctly.

NO

Perform the change RS422/485 correctly


F58 Change of RS422/485

YES

There is 1 ms
or more of interval between
receiving the command and
sending for the
host.

NO

Set the interval for more than 1 ms between receiving


the command and sending for the host.

YES

Against the
response from the host,
the delay time setting for
response is adequate
or not.

There is a possibility that the response is not obtained


due to the delay of response from the host. Set the
delay time for response larger.
F59 Setting delay time for response for RS485

NO

YES
Inform Minebea about the contents of
failure and situation at site in details.

153

103. Error display


Error
code

Contents of error

Remedy

Zero set has executed with more than


10 % of the maximum display value.

Apply zero set after making it within


10 % of maximum display value.

When setting the CHECK ON, zero setting is


executed.

After setting the CHECK OFF, execute


the zero setting.

ER1

Setting mistake

Set correctly.

ER2

Mistake in setting during calibration

Set correctly.

ER4

Displays when the calibration(CCAL, LCAL,


TARE) is executed with the CHECK ON.

After setting the CHECK OFF, execute


the calibration.

ER5

At the time of fine adjustment on zero or span,


it is shown when zero set, and peak are ON.

Set OFF while the peak is ON.


Apply ZERO clear(F98).

ER6

When prohibiting the calibration is set,


calibration or fine adjustment for analog is
proceeded.

Remove the prohibition of calibration


(F97).

HOLD

Powered ON with the HOLD input is shorted.

Set the HOLD input open.

TEL

Displays when the initial value at the time of


calibration is less than 2.0 mV/V

TEH

Displays when the initial load is more than 2.0


m V/V, or the total value with the initial value
and the load equal to the maximum display
value exceeds 3.1 mV/V during calibration.

SPL

The value equal to the maximum display value


at the time of calibration (LCAL) is less than
0.4 mV/V(0.8 mV/V at B.V. 2.5 V), or the
difference between the initial value and the
load equal to the maximum display value is
less than 0.4 mV/V(0.8 mV/V at B.V. 2.5 V).

ER0

SPH

The value equal to the maximum display value


at the time of calibration (LCAL) exceeds 3.1
mV/V, or the difference between the initial
value and the load equal to the maximum
display value exceeds 3.1 mV/V.

ERE

EEPROM error

ERR

EEPROM error

OL

OL

Adjust so that the initial load is within


the range from 2.0 mV/V to 2.0 mV/V.

Adjust so that the value equal to the


p y value is within the
maximum display
range from 0.4 mV/V to 3.1 mV/V.

Contact with Minebea.


Minebea

Displays when 110 % of the maximum display


value is exceeded at the time of measurement
or in calibration.
Displays when 110 % of the maximum
display value is exceeded at the time of
measurement or in calibration.

154

Set so that the load display will be within


the range from 110 % to 110 %.

11. Specifications
111. Specifications for analog section
DC10 V0.5 V

Applicable transducers

Bridge power supply Connectable strain gage applied transducers


DC10 V
2 pieces
DC5 V
4 pieces
DC2.5 V
8 pieces

Input range

F.S. setting is available at the input range from 0.4 mV/V


to 3.1 mV/V.(When bridge power supply is DC10 V or DC5 V.)
F.S. setting is available at the input range from 0.8 mV/V
to 3.1 mV/V.(When bridge power supply is DC2.5 V.)

Output range

DC10 V Load resistance is 5 k or more.


(F.S. setting is available in Function.)

Output times

Changeable to 100 times/s, 500 times/s, 1 000 times/s


or 2 000 times/s.(Synchronous with A/D sampling rate.)

Output resolution

As same as display resolution

Zero adjustment range

2.0 mV/V

Nonlinearity
Display
Output

0.025 %F.S.
0.025 %F.S.

Temperature coefficient
Zero point
Sensitivity

within 60 mA

(Changeable to DC5 V or DC2.5 V

Bridge power supply

0.5 V/(Input conversion, at the time of F.S. setting


at the input from 0.5 mV/V to 3.1 mV/V)
0.01 %F.S./(Input conversion, at the time of F.S. setting
at the input from 0.5 mV/V to 3.1 mV/V)

Input noise

0.6 Vpp or less


(With the default setting of digital filter and stabilized filter)

Input filter

10 Hz, 100 Hz or 1 kHz

A/D sampling

2 000 times/s
(changeable to 100 times/s, 500 times/s or 1 000 times/s)

CHECK

Approx.0.3 mV/V
(Setting can be available at the approx. 0.1 mV/V interval within the
range from approx.0.1 mV/V to 2.0 mV/V.)
The extension cable is applied to Minebeas standard cable
CAB-502(4 cores) within the length of 30 m.
Not applicable when the zener barrier is used.

112. Specifications for digital section


Load display
Display range
Display increment
Display

11 000 to 11 000
1 (changeable to 2, 5 or 10)
7 segment red LED, with 17 mm characters height

Comparator display
Display range
Display
Number of display

99 999 to 99 999
7 segment green LED, with 8 mm characters height
1 point(Selectable from S0, S1, S2, S3 and S4)

Status display

SEL.1, SEL.2, CHECK, HOLD, PEAK, MEAS. and END

Judgement display

S0, S1, S2, S3, and S4

Display rate

20 times/schangeable to 4 times/s, 50 times/s or 100 times/s.)

Decimal point display

Changeable to Non, ,101, 102, 103 or 104.


155

113. Front panel sheet key function

Change of Function mode


Zero set
S set value displayCarry on set value
Increment of set value
Changeover of Track and Peak hold, Bottom hold, Peak bottom hold,
Local maximal value hold, Local minimal value hold or Inflection point hold
(When all zone specified mode)
Or control of holding zone in zone specification mode or time zone
specification mode.
Reset of peak value During ON, the display is fixed to zero(0).
ON/OFF for check value
Enter key/Shift key

114. External control function

ZERO

Same as the
key
Above is pulse input, and effective once at the pulse input and the pulse
width is 50 ms or more. (Pulse width is changeable to 1 ms, 2 ms, 5 ms,
10 ms or 20 ms.)

PEAK/TRACK

Change of Track and Peak hold, Bottom hold, Peak bottom hold, Local
maximal value hold, Local minimal value hold or Inflection point hold
(It can be changed by the function.)
Or control of holding zone in zone specification mode or time zone
specification mode
Open
Track
Short Peak hold, Bottom hold, Peak bottom hold, Local maximal value
hold, Local minimal value hold or Inflection point hold
(It can be changed by the function.)

HOLD

Hold of Display, Comparative output, Analog output and BCD output

RESET

Same as the

SEL.1, SEL.2

Changeover of 4 kinds for Comparator code or Calibration data


combined with SEL.1 and SEL.2.
(Select the changeover target from Comparator code or Calibration data
by the function.)
Above is level input, and effective during the short input more than 50 ms.
(Level width Changeable to 1 ms, 2 ms, 5 ms, 10 ms or 20 ms)

key, reset condition can be made by short.

12 V

12 V

100 k
100 k

330 pF

156

2.2 k

1k

Equivalent circuit of external control input.


5 V

INPUT
COM.1

115. Comparator function


Set value

99 999 to 99 999

Numbers of setting

5 points (S0, S1, S2, S3 and S4)


S0 is set by function.

Set value for hysteresis data

0 to 99 digits

Setting hysteresis time width 0 to 9.9 s


Hysteresis direction

Can be selected whichever On delay or Off delay.

Conversion times for comparator


100 times/s, 500 times/s, 1 000 times/s and 2 000 times/s changeable.

116. Open collector output signal


S1, S2, S3, S4

Open collector ON when reached under/over the comparator set value.

S0

Open collector ON with either condition in below by function setting.


FULL condition (100 % of rated load).
When the both two of S1, S2, S3 and S4 are OFF condition.
Operates when reached under/over the S0 set value.
(Same as the comparative operation of S1, S2, S3 and S4.)
ON interlocked to HOLD LED of condition display
ON interlocked to PEAK LED of condition display
ON interlocked to MEAS. LED of condition display

Open collector specifications


VCEDC30 V, IC30 mA MAX.
Equivalent circuit of contact output section
OUTPUT

COM.2

117. Various kinds of functions


Digital filter

Stabilizes the data by the computing process through CPU.

Change of target of HOLD

With the combination of Display,Comparative output, Analog


output, BCD output(Option), target of HOLD can be made.

Sheet key lock

Prohibition of operation of optional key.

Change of target of analog output


The target of analog output can be changed either TRACK value
and PEAK value.
Change of calibration data

Four kinds of calibration data can be memorized, and they be


selected by the function or the external control input. (The
changeover target of Calibration data and Comparator code is
set by the function)

Change of comparator code

The set value of the comparator of S0, S1, S2, S3 and S4 can be
memorized up to four kinds, and do changeover of them by the
function or the external control input. (The changeover target of
Calibration data and Comparator code is set by the function)

Selection of Peak mode

Selectable from six mode of Peak hold, Bottom hold, Peak/bottom


hold, Local maximal value hold, Local minimal value hold and
Inflection point hold by 24 modes by which four kinds of zone mode
of All zone specification, Zone specification, Time zone specification
and Automatic start time zone specification are combined.
157

118. General specifications


Operating temperature/humidity range
Temperature
10 to 50
Humidity
Less than 85 %RH (Non condensing.)
Power supply
Power supply voltage

AC100 V to AC240 V
(Permissible variation rangeAC85 V to AC264 V)
Power supply frequency 50/60 Hz
Power consumption
Approx. 8 VA(Without option, at AC100 V)
Approx. 17 VA at max. (With option, at AC100 V to AC240 V)

Outline dimensions(WHD) 96 mm96 mm129.5 mm (Excludes protruding parts.)


Weight

Approx.500 g (Without any options.)

119. Standard specifications at the shipment


Bridge power supply

DC10 V

Span adjustment

2 000 display at the input of 0.5 mV/V.

The minimum scale

Analog output

0 V to 10. 000 V at 0 to 2 000 display

1110. Accessories
Instruction manual

1 piece

Timelug fuse

1 piece (1 A)

Unit seal

1 piece

Panel mounting attachment

2 pieces

Panel mounting gasket

1 piece

BCD output plug

1 piece(Attached only when optional BCD output is installed.)

1111. Options
11111. Current output
P/N
Specifications
Output
Nonlinearity
Resolution
Over range
Output rate

CSD819C-P07
DC4 mA to 20 mA Load resistance 260 or less
0.025 %F.S.
As same as display resolution
OL display at approx.DC2.4 mA and OL display at approx.DC21.6 mA.
Changeable to 100 times/s, 500 times/s, 1 000 times/s and 2 000 times/s
(Synchronous with A/D sampling rate.)
Voltage output isnt available when this option is installed.

158

11112. BCD output


P/N

CSD819CP15

Specifications
Output

BCD

Input

5 digits Parallel output, with polarity(POL.) applied


(Output ON with minus, and output OFF with plus.)
P.C.(Print command)
ON for a fixed time after conversion of BCD output is
completed.
ERROR
ON at the time of various errors are occurred.
OVR(over)
Above are open collector outputs.
VCEDC30 V, ICDC20 mA MAX
Output rate
Changeable to 4 times/s, 20 times/s, 50 times/s, 100
times/s and 200 times/s

ZERO
Same as the
key.
Above pulse input and pulse width is 50 ms or more is effective once.
(Pulse width is changeable to 1 ms, 2 ms, 5 ms, 10 ms and 20 ms.)

key
As same as the key

PEAK/TRACK As same as the


RESET
HOLD
BCDENABLE

Hold of display and BCD output


Compulsive OFF for the related output of BCD
(Hi impedance)
Above are level inputs, and effective during short input of 50 ms or more.
(Level widthChangeable to 1 ms, 2 ms, 5 ms, 10 ms or 20 ms.)

11113. RS232C interface


P/N

CSD819CP74

Specifications
Baud rate

Select from 1 200, 2 400, 4 800, 9 600, 19 200


or 38 400 bps.
Data bit length
Select from 7 bits or 8 bits.
Parity bit
Select from Non, Even or Odd.
Stop bit
Select from 1 bit or 2 bits.
Terminator
Select from CRLF or CR.
Communication method Halfduplex
Synchronous method Startstop synchronous method
Communication data ASCII code
Cable length
Within approx. 15 m
Input/output monitor with LED

159

11114. RS422/485 interface


P/N

CSD819CP76

Specifications
Baud rate

Select from 1 200, 2 400, 4 800, 9 600, 19 200


or 38 400 bps.
Data bit length
Select from 7 bits or 8 bits.
Parity bit
Select from Non, Even or Odd.
Stop bit
Select from 1 bit or 2 bits.
Terminator
Select from CRLF or CR.
Communication method Halfduplex
Synchronous method Startstop synchronous method
Address
Select one among 0 to 31.
Communication data ASCII code
Cable length
Approx. 1 km
No. of connections
32 sets at max.(RS42210 sets)
Termination
Builtin (Yes/No can be selected by the connection with
terminal block.)
Input/output monitor with LED.
Change of RS422/485 Can be set in Function.

160

1112. Outline dimensions


Side

Front

Rear

Upper

Panel cut size

Unitmm

161

12. Warranty
121. Warranty
D The instrument is covered by a warranty for a period of one year from the date of delivery.
D As for repairs and/or after service is required during the period of warranty, contact with
Minebeas sales office or sales agent from which you have purchased.

122. Repair
Before asking repairs, make checks once again that the connection, setting and adjustment for the
instrument have finished properly by referring to 10. Trouble shooting.
Especially, make checks whether the connections of sensors are disconnected or cut off.
After that, still there may be found some defects in the instrument, contact with Minebeas sales
office or sales agency from which you have purchased.

162

13. Appendix
131. Replacement of fuse

Warning

When installation method for the fuse is wrong and/or the capacity of installed fuse
is inadequate, it causes and unexpected faulty of the instrument.

Turn OFF the power supply for the instrument.

Remove the 2 pieces of setscrews on the rear panel.

Holding the rear panel assembly, draw out it slowly to the arrow marked direction.

Case
Rear panel assembly

Setscrew for rear panel


2M36

Fuse installed section

163

Replace the fuse installed on the POWER CARD.

Fuse

Small type minus driver


Fuse holder

Fuse

Fuse holder

After replacing the fuse, install the rear panel assembly into the case.

Setscrew for rear panel


2M36

Set the 2 pieces of setscrews on the rear panel. Check that the rear panel and the case is
attached closely.

Warning

When the rear panel is not attached closely, you cant feel the click touch and also
there may have the possibility that key operation is not made.

164

132. Characters pattern for display


The followings are the table to show the display pattern used at 7 segments display on the
instrument.

165

133. Setting table for functions


Make use of them in case that the customer has changed setting for the function.
Function No.

Initial value

F01

Customers setting

Function No.

Initial value

00000

F55

00001

F02

00003

F56

00000

F03

00001

F57

00000

F04

00055

F58

00000

F05

00002

F59

00005

F06

00000

F60

00000

F10

11111

F61

00000

F11

00003

F62

00000

F12

00000

F63

00200

F20

00000

F64

00015

F21

00000

F65

00100

F22

02000

F66

00100

F30

11111

F67

00100

F31

00000

F68

01000

F32

00000

F69

01000

F33

00000

F71

00000

F34

00000

F72

00000

F35

00000

F75

F36

00000

F80

00000

F38

00000

F81

00000

F40

00000

F82

00000

F41

00000

F90

F42

00000

F91

F43

00000

F92

F44

00000

F93

F45

00004

F94

F46

00004

F95

F50

00001

F96

F51

00000

F97

00000

F52

00003

F98

F53

00021

F99

F54

00000

Memo

166

Customers setting

D The contents of this manual may subject to change without notice.


HEAD QUARTER :

MINEBEA CO., LTD.


410673 Miyota, Miyotamachi, Kitasakugun, Naganoken 3890293, Japan