Heating Bath Circulators


CW-05G/10G/20G/30G (Amb. +5 to 150)

External and internal circulation


with powerful pumping capability

Certification without CW-30G

Model
Bath Volume (L / cu ft)
Working Temperature Range2) ( / )
Temperature Stability1)( / )
Max. Pressure, Max. Flow Rate
Pump
Max. Suction Flow Rate

CW-05G
5 / 0.2
0.05 / 0.09

CW-10G

CW-20G

CW-30G

10 / 0.4
20 / 0.7
Amb. +5 to 150 / Amb. +9 to 302
0.05 / 0.09
0.1 / 0.18
5.9 PSI (406.8 millibar), 7.4 gallons / minute (28 liters / minute)
2.6 gallons / minute (10 liters / minute)

30 / 1.1
0.1 / 0.18

15099, 160
/ 5.93.9, 6.3

211154, 180
/ 8.36.1, 7.1

290214, 200
/ 11.48.4, 7.9

300264, 230
/ 11.810.4, 9

Overall

220395424
/ 8.715.616.7

281450444
/ 11.117.717.5

360510464
/ 14.220.118.3

370560494
/ 14.62219.4

Net Weight (kg / lbs)


Electrical Requirements (230V, 50 / 60 Hz)
Cat. No.
Electrical Requirements (120V, 50 / 60 Hz)
Cat. No.

12.5 / 27.6
4.5A
AAH52305K
8.6A
AAH52306U

15 / 33
6.7A
AAH52315K
8.6A
AAH52316U

17.5 / 38.6
8.8A
AAH52325K
8.6A
AAH52326U

19.5 / 43
8.8A
AAH52335K
8.6A
AAH52336U

Bath Opening / Depth

(WL, D) (mm / inch)

Dimension

(WLH) (mm / inch)

1) Technical data according to DIN 12876.


2) If you operate the unit in 1 to 3 steps of pump, the temperature range can be from Amb. +5 but in 4 and 5 steps, the range will be from Amb. +15.

Amb. +5 to 150

Protection against low fluid


level and dry running by
automatic shut down of the
heater together with audible
and visual alarms

Microprocessor PID control/


Automatic tuning/ Calibration
Digital timer for delayed
ON / OFF (from 1 min to 99 hr 59 min)
Protection against low fluid
level and dry-running
Convenient memory function
of up to three frequently
used temperature settings
Digital LED display (0.1 resolution)
Splash-proof keypad
RS-232 interface

Accurate, stable, and efficient


temperature management by
using a powerful variablespeed pump (1 to 3 steps of pump

Corrosion-resistant,
seamless, leak-proof, roundcornered, and easy-to-clean
stainless steel bath with a
drain valve

Powerful pressure and


suction pumps provide
contrast pressure and flow
rate throughout the whole
temperature range

Compact footprints ideal for


cramped lab space

for internal circulation and 4 and 5 steps for


external circulation)

Isolated design of agitator,


heater, and sensor by a
baffle plate to minimize the
turbulence in the bath fluid
and also to protect users
and samples against direct
contact

STANDARD
Accessories
detailedSteel
information
of Cover
accessories, please refer to the page 14-17.
For
Stainless
Flat Type
One 30cm (1 ft) Length and 12mm of Silicone Rubber Tubing

Refrigerated & Heating Bath Circulators


RW-0525G/1025G/2025G/3025G (Amb. -25 to 150)

Rapid heating and cooling


throughout the entire temp.
range

Model

RW-0525G

Bath Volume (L / cu ft)


Working Temperature Range2) ( / )
Temperature Stability1)( / )
at -20 (-4), W
at 0 (+32), W
Cooling
Capacity
at +20 (68), W
Refrigerator
Max. Pressure, Max. Flow rate
Pump
Max. Suction Flow Rate3)

0.05 / 0.09
30
230
320
1 / 3 LBP, R-507

Bath Opening / Depth

(WL, D) (mm / inch)

Dimension

RW-1025G

5 / 0.2

Overall

(WLH) (mm / inch)

Net Weight (kg / lbs)


Electrical Requirements (230V, 60Hz)
Cat. No.
Electrical Requirements (230V, 50Hz)
Cat. No.
Electrical Requirements (120V, 60Hz)
Cat. No.

10 / 0.4

RW-2025G

RW-3025G

20 / 0.7

30 / 1.1

-25 to 150 / -13 to 302


0.05 / 0.09
0.05 / 0.09
50
260
250
410
380
560
1 / 3 LBP, R-507
1 / 2 LBP, R-404A
5.9 PSI (406.8 millibar), 7.4 gallons / minute (28 liters / minute)
2.6 gallons / minute (10 liters / minute)

0.08 / 0.14
390
570
680
3 / 4 LBP, R-404A

15099, 160
/ 5.93.9, 6.3

211154, 180
/ 8.36.1, 7.1

290214, 200
/ 11.48.4, 7.9

300264, 230
/ 11.810.4, 9

302438690
/ 11.917.227.2

360490780
/ 14.219.330.7

380560825
/ 152232.5

440620895
/ 17.324.435.2

37 / 81.6
7A
AAH57001K
7A
AAH57002K
12.6A
AAH57003U

44 / 97
9.2A
AAH57011K
9.2A
AAH57012K
12.6A
AAH57013U

60 / 132.2
12.8A
AAH57021K
12.8A
AAH57022K

70 / 147.7
13.8A
AAH57031K
13.8A
AAH57032K

* RW-2025G and RW-3025G Models are available ONLY for 230V.


1) Technical data according to DIN 12876.
2) The cooling capacity is measured on 230V / 60Hz electrical condition.
3) The additional suction function is optional.

Temp. range from


-25 to 150
Microprocessor PID control/
Auto-tuning/ Calibration
Digital timer for delayed
ON / OFF (from 1 min to 99 hr 59 min)
Protection against overtemperature and low fluid
level
Convenient memory function
of up to three frequently
used temperature settings
Digital LED display (0.1 resolution)
Splash-proof keypad
RS-232 interface

STANDARD
Accessories
detailedsteel
information
of accessories, please refer to the page 14-17.
For
Stainless
Flat Cover
30 cm (1 ft) long and 12 mm Silicone Rubber Tubing

10

Suitable for both internal and


external bath environment
due to reduced turbulence
by using five-speed duplex
pump and built-in cooling coil
(1 to 3 steps of the pump for internal
circulation and 4 and 5 steps for external
circulation)

Isolated design of agitator,


heater, and sensor by a
baffle plate to minimize the
turbulence in the bath fluid
and also to protect users
and samples against direct
contact
Protection against low fluid
level and dry running by
automatic shut down of the
heater together with audible
and visual alarms

Corrosion-resistant,
seamless, leak-proof, roundcornered, and easy-to-clean
stainless steel reservoir with
a drain valve
Usage of CFC-free efrigerant
(R-507, R-404A) and a double
service valve for leakage
prevention
Quick-release front grille for
easy cleaning of the
condenser
Compact footprints ideal for
cramped lab space

Refrigerated & Heating Bath Circulators


RW-0540G/1040G/2040G/3040G (Amb. -35 to 150)

Stable temp. control capability


at very low temp. by adopting a
powerful pump for both external
and interial circulation

Model
Bath Volume (L / cu ft)
Working Temperature Range2) ( / )
Temperature Stability1)( / )
at -20 (-4), W
at 0 (+32), W
Cooling
Capacity
at +20 (68), W
Refrigerator
Max. Pressure, Max. Flow rate
Pump
Max. Suction Flow Rate3)
Bath Opening /Depth

(WL, D) (mm / inch)

Dimension

Overall

(WLH) (mm / inch)

Net Weight (kg / lbs)


Electrical Requirements (230V, 60Hz)
Cat. No.
Electrical Requirements (230V, 50Hz)
Cat. No.

RW-0540G
5 / 0.2
0.05 / 0.09
80
150
280
1 / 2 LBP, R-404A

RW-1040G
10 / 0.4

RW-2040G

RW-3040G

20 / 0.7

30 / 1.1

-35 to 150 / -31 to 302


0.05 / 0.09
0.05 / 0.09
150
400
330
660
450
800
1 / 2 LBP, R-404A
3 / 4 LBP, R-404A
5.9 PSI (406.8 millibar), 7.4 gallons / minute (28 liters / minute)
2.6 gallons / minute (10 liters / minute)

0.08 / 0.14
440
700
880
1 LBP, R-404A

15099, 160
/ 5.93.9, 6.3

211154, 180
/ 8.36.1, 7.1

290214, 200
/ 11.48.4, 7.9

300264, 230
/ 11.810.4, 9

302438690
/ 11.917.227.2

360490780
/ 14.219.330.7

380560825
/ 152232.5

440620895
/ 17.324.435.2

38 / 83.8
8.5A
AAH57101K
8.5A
AAH57102K

44 / 97
10.7A
AAH57111K
10.7A
AAH57112K

60 / 132.2
13.8A
AAH57121K
13.8A
AAH57122K

67 / 147.7
14.8A
AAH57131K
14.8A
AAH57132K

* RW-40G Models are available ONLY for 230V.


1) Technical data according to DIN 12876.
2) The cooling capacity is measured on 230V / 60Hz electrical condition.
3) Additional suction pump is optional.

Temp. range from


-35 to 150
Microprocessor PID control/
Auto-tuning/ Calibration
Digital timer for delayed
ON / OFF (from 1 min to 99 hr 59 min)
Protection against overtemperature and low fluid
level
Convenient memory function
of up to three frequently
used temperature settings
Digital LED display (0.1 resolution)
Splash-proof keypad
RS-232 interface

Suitable for both internal and


external bath environment
due to reduced turbulence
by using five-speed duplex
pump and built-in cooling coil
(1 to 3 steps of the pump for internal
circulation and 4 and 5 steps for external
circulation)

Isolated design of agitator,


heater, and sensor by a
baffle plate to minimize the
turbulence in the bath fluid
and also to protect users
and samples against direct
contact

Corrosion-resistant,
seamless, leak-proof, roundcornered, and easy-to-clean
stainless steel reservoir with
a drain valve
Usage of CFC-free efrigerant
(R-134A, R-404A) and a double
service valve for leakage
prevention
Quick-release front grille for
easy cleaning of the
condenser
Compact footprints ideal for
cramped lab space

Protection against low fluid


level and dry running by
automatic shut down of the
heater together with audible
and visual alarms

STANDARD
Accessories
detailedsteel
information
of accessories, please refer to the page 14-17.
For
Stainless
Flat Cover
30 cm (1 ft) long and 12 mm Silicone Rubber Tubing

11

Accessories

Spring Wire Racks

Half-shelf Adjusters

Additional Suction Pumps

Spring Wire Racks

Additional Suction Pumps*

Ideal for holding a wide variety of tubes, flasks, and other


glassware. All components are made of stainless steel.

 dditional suction device for more powerful external and


A
internal circulation. Maximum suction flow rate is 10.0
liters per minute. (only for CW-G and RW-G Models)

(for all models)

Cat. No.
AAA45502
AAA45503
AAA41502
AAA41503
AAA51522
AAA51523

Suitable for
BW-10H / BW-10B
BW-20H / BW-20B
BW-10G
BW-20G
CW-20G / RW-2025G, 2040G
CW-30G / RW-3025G, 3040G

Cat. No.
52503
52513
52523
52533

Suitable for
CW-05G / RW-0525G, 0540G
CW-10G / RW-1025G, 1040G
CW-20G / RW-2025G, 2040G
CW-30G / RW-3025G, 3040G

* Adding this additional suction function after your purchase is timeconsuming and costly. Prior request is highly recommended.

Half-shelf Adjusters
Stainless steel perforated sample tray shelves with adjustable height support prevent leaks and guard against rust and contamination. (only for BW-H, BW-B, and BW-G Models)
Cat. No.
AAA45511
AAA45512
AAA45513
AAA41551
AAA41552
AAA41553

Suitable for
BW-05H / BW-05B
BW-10H / BW-10B
BW-20H / BW-20B
BW-05G
BW-10G
BW-20G

Viscometer Tube Holders


S pecially designed to hold a two-opening viscometer
glass tube and made of Bakelite, renowned for its high
thermal endurance. (only for VB Models)
Cat. No.
AAA43501

Suitable for
VB-25G (need Max. 5 holders)
VB-40G (need Max. 8 holders)

Viscometer Tube Holders

Cooling Circulators
T he temperature range of viscometer heating baths can be lowered up to +10 (+50) when used with an optional cooling
circulator. (only for VB Models)
Cat. No.
AAH57011K
AAH57012K
AAH57021K
AAH57022K

Suitable for
VB-25G (230V, 60Hz)
VB-25G (230V, 50Hz)
VB-40G (230V, 60Hz)
VB-40G (230V, 50Hz)

Cooling Circulators *
RW-1025G (230V, 60Hz)
RW-1025G (230V, 50Hz)
RW-2025G (230V, 60Hz)
RW-2025G (230V, 50Hz)

* Refer to p.16-17 for more information on the cooling circulators.

15

/ Autoclaves
/ Baths & Circulators
/ Desiccators
/ Furniture
/ Hoods
/ Incubators
/ Ovens
/ Pumps
/ Recirculating Coolers (Chillers)
/ Refrigerators & Freezers
/ Shakers
/ Heating & Cooling Blocks
/ Hotplates & Stirrers
/ Vortex mixers
/ Temperature Chambers
/ Temperature & Humidity Chambers
/ Pipettes
/ Plasticware
/ Ultrasonic Cleaners
/ Other Products

You might also like