Manometers are used to provide a visible pressure measurement and to perform acc
urate pressure measurement of pressure sensing devices.
Manometers are used as an indicating device connected to pressure sensing device
s with actual pressure being applied to both the pressure sensing device and the
manometer to verify accurate sensing capabilities to the device being calibrate
d.
Bellows type elements are constructed of tubular membranes that are convoluted a
round the circumference.
The membrane is attached at one end to the source and at the other end to an ind
icating device or instrument. The bellows element can provide a long range of mo
tion (stroke) in the direction of the arrow when input pressure is applied.
Capsular There are two different devices that are referred to as capsule.
The first is pressure is applied to the inside of the capsule and if it is fixed
only at the air inlet it can expand like a balloon. This arrangement is not muc
h different from the diaphragm except that it expands both ways. The capsule con
sists of two circular shaped, convoluted membranes (usually stainless steel) sea
led tight around the circumference. The pressure acts on the inside of the capsu
le and the generated stroke movement is shown by the direction of the arrow.
The second type of capsule, the capsule in the bottom is constructed with two di
aphragms forming an outer case and the inter-space is filled with viscous oil. P
ressure is applied to both side of the diaphragm and it will deflect towards the
lower pressure. To provide over-pressurized protection, a solid plate with diap
hragm-matching convolutions is usually mounted in the centre of the capsule. Sil
icone oil is then used to fill the cavity between the diaphragms for even pressu
re transmission.
II.
1.
Semiconductor Strain Gauge, The function of a semiconductor strain gauge
essentially consists in a very considerable and accurate change in its electric
al resistance with applied mechanical strain. Sensors with these elements give t
o all mechanisms the ability to respond sensitively to mechanic impulses. They a
re usually used for force, mass, pressure, torque, acceleration and deformation
electric measurements. The most common use of strain gauges is for half-bridge o
r full-bridge, because during statical measurements it compensates the temperatu
re effects. Measurement of electric resistance of one strain gauge can be used o
nly there, where we don t need to compensate the temperature effects.
2.
Variable Capsule Pressure Detector is strain-based, wherein a magnetic c
ircuit is formed, and the parameter input causes mechanical deflection of the sp