Evaluation of Dental Implant Placement with Stereolithographic Handpiece


Guided Surgical Stent : In-Vivo & In-Vitro study /
Ali, Abd el Raouf Mohamed.
617

617.605
.Ali, Abd el Raouf Mohamed
Evaluation of Dental Implant Placement with Stereolithographic
: Handpiece Guided Surgical Stent
/ In-Vivo & In-Vitro study
Abd el Raouf Mohamed Ali; Supervised by Mohamed Diaa Zein AlAbdeen Ismaiel, Salah Abd- El Fattah Metwalli, Heba Abd el Wahed
.Sleem

( )
.2015
: .p 156
.cm 30
Thesis (Ph.D.)- Ain Shams University. Faculty of Dentistry. Oral and
.Maxillofacial Surgery Department
The aim of this study was to evaluate the accuracy of the computer aided
stereolithographic handpiece guided surgical stent in comparison to the
ordinary computer aided stereolithographic surgical stent both in-vivo
and in-vitro using CBCT, CAM software and additive manufacturing in
.the form of 3D printing
Nine patients either totally or partially edentulous were included in the
current study received 41 dental implants through the use of standard or
.handpiece supported surgical guide
The handpiece was scanned and the handpiece guiding apparatus was
designed using 3D Max then the guiding apparatus was sent for 3D
printing
The patients and their dental casts or denture were scanned using CBCT
machine in a dual scan technique, following the DICOM files were sent
.for specialised softwares
Simplant pro 11 and Mimics 15 softwares were used to plan the implants
in the patients jaws and to design the surgical guides which were either
tooth - mucosa supported for partially edentulous or bone supported for
completely edentulous patients then the surgical guides were sent for 3D
.printing
Under local anaesthesia, the patients received their dental implants using
the surgical guide either standard or handpiece supported and after a
healing period a postoperative CBCT was done for every patient in the
.study
The patients jaws were extracted from the software project and sent for
3D printing and the same surgery was repeated on the printed jaws that
resemble patients jaws using the same surgical guide used in the patient
.mouth, then another CBCT was done for the printed jaw
Postoperative CBCT images was superimposed over preoperative CBCT
images for the clinical and In-vitro study to determine the accuracy of
.the implant placement and amount of error of deviation
The results of the study showed that the standard surgical guide was
more accurate than the handpiece supported surgical guide and the in.vivo study was more accurate than the in-vitro study
The clinical study results for both groups are considered of moderate
.deviation according to Valente classification
:

.


.

:
:
:
:
:

:
:
:
:
:
: