Professional Documents
Culture Documents
AI 1200
Atomic Absorption Spectrometer
AI 1200 Cookbook
Table of Contents
AI 1200 COOKBOOK
Table of Contents
BOOK ONE- FAAS
Chapter 1: Theory of AAS
Introduction
Flame Atomic Absorption Spectrometry (FAAS)
Graphite Furnace Atomic Absorption Spectrometry (GFAAS)
Vapor Hydride Generation Atomic Absorption Spectrometry (VG AAS)
Chapter 2: AAS Instrumentation
Fundamentals
Light Source
Atomizer
Optics
Detector
Optics
Lenses
Mirrors
Monochromator
Diffraction Grating
Slit Width
Atomizer
Flame
Graphite Furnace
Detector
2
3
4
5
9
10
15
18
22
24
27
28
29
AI 1200 Cookbook
Table of Contents
Things to Consider
Applications
Expected Concentration Ranges
Elements
Atomization Efficiency
Interferences
Spectral
Background
Matrix
Detection Limits
Sensitivity
Precision
Linear Working Range
Minimum Sample Volume
Sample Throughput
Sample Usage
Total Dissolved Solids
Method Development
Ease of Use
Automation/Unattended Operation
Costs
Initial Investment
Running Costs
35
35
35
35
36
36
36
36
36
36
36
37
ii
41
41
41
42
42
43
43
43
AI 1200 Cookbook
Table of Contents
iii
47
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
AI 1200 Cookbook
Table of Contents
Scandium, Sc
Selenium, Se
Silicon, Si
Silver, Ag
Sodium, Na
Strontium, Sr
Tantalum, Ta
Tellurium, Te
Thallium, Tl
Tin, Sn
Titanium, Ti
Tungsten, W
Uranium, U
Vanadium, V
Ytterbium, Yb
Yttrium, Y
Zinc, Zn
Zirconium, Zr
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
120
121
122
123
124
127
129
References
132
iv
1 - Theory of AAS
Chapter 1:
Theory of AAS
Introduction
Flame Atomic Absorption
Spectrometry (FAAS)
Graphite Furnace Atomic Absorption
Spectrometry (GFAAS)
Vapor Hydride Generation Atomic
Absorption Spectrometry (VG AAS)
1 - Theory of AAS
Introduction
The essential elements of the theory behind the analytical technique of atomic
absorption spectroscopy (AAS) are compacted into the following paragraphs.
Atomic absorption spectroscopy (AAS) relies on the fact that the light
absorption of free atoms [1-7]. All atoms can absorb light, but only at discrete
wavelengths corresponding to the energy requirements of the particular atom. In
other words, each element absorbs light at some specific and unique wavelength and
does not absorb light at all on other wavelengths. For example, in a sample with
multiple elements (say, copper, lead, iron, and nickel), only copper will absorb light
that is at the characteristic wavelength for copper. Furthermore, the amount of light
absorbed depends on the number of absorbing atoms that are present in the light
path. All these factors enable AAS to be used as a tool for quantitative analysis.
In practice, measuring the amount of light absorbed by several known
standards allows a calibration curve to be constructed. Then, the unknown
concentration of a sample can easily be determined based on the amount of light it
absorbs.
The amount of light energy absorbed at this wavelength depends on the
concentration of the atoms in the medium (as dictated by Lamberts law and Beers
law). Lamberts law states that the portion of light absorbed by a transparent
medium is independent of the intensity of the incident light and each successive unit
layer of the medium absorbs an equal fraction of the light passing through it. Beers
law states that the light absorbed is proportional to the number of absorbing atoms
in the medium. Mathematically, when light of intensity Io passes through a medium
of length x with atom concentration of C, the intensity I of the light beam emerging
from the medium is given by:
I = Io e-kCx
where k is a proportionality constant (the absorption coefficient). The
absorption of the medium, A, is defined to be:
A = lg (Io/I) = kCx
This equation states that the absorbance, A, of the medium is linearly
proportional to the concentration of the absorbing atoms. The absorption coefficient
(or absorptivity), k, can be determined by constructing a calibration curve (i.e.
plotting the observed absorbance versus the known sample concentration). The
slope of the calibration curve is kx, and x is easily measurable or already known.
Unknown sample concentrations may be determined from the calibration curve based
on their measured absorbances.
Any way that you look at, every AAS experiment can be broken down to the
following procedure:
1 - Theory of AAS
A blank solution must be prepared. This blank must contain none of the
element of interest.
A series of standard solutions must be prepared. These standards have
known (but varied) concentrations of the element of interest. These
standards are used to prepare a calibration curve.
Analyze the blank solution to determine the "blank" absorbance value. This is
the absorbance value for a sample with a zero concentration of the element of
interest.
Individually analyze all of the standard solutions.
Construct a calibration graph. For the blank and each standard solution, plot
its absorbance value against its concentration.
Analyze the unknown sample. The concentration of the unknown sample,
based on its measured absorbance value, can be determined from the
calibration curve.
1 - Theory of AAS
M (g)
M (g) + CO (g)
MO (g)
M (g) + O (g)
CO2 (g).
Analysis times for GFAAS are longer than those for FAAS, and fewer elements
can be determined with this technique. Nonetheless, GFAASs enhanced sensitivity,
ability to analyze very small sample sizes, and ability to directly analyze certain
types of solid samples significantly expand the capability of atomic absorption
spectrometry.
1 - Theory of AAS
1 - Theory of AAS
1 - Theory of AAS
2 - Instrumentation
Chapter 2:
AAS Instrumentation
Fundamentals
Light Source
Atomizer
Optics
Detector
Optics
Lenses
Mirrors
Monochromator
Diffraction Grating
Slit Width
Atomizer
Flame
Graphite Furnace
Detector
2 - Instrumentation
Fundamentals
There are four components that are essential to every AAS instrument: a
light source, an atomizer, an optics system, and a detector.
Light Source
Usually a hollow cathode lamp (HCL) is used as the light source in AAS. A
less common light source is the electrodeless discharge lamp (EDL). An HCL
produces an intense, narrow line emission of light at a wavelength that is specific
to the element that the HCL cathode is coated with. Most elements emit at
multiple wavelengths, but all emissions are intense, sharp lines (called resonance
lines). For example, a copper HCL emits at 324.75 nm, 327.40 nm, 222.6 nm,
249.2 nm, and 244.2 nm. Generally, one line is more intense than the others
and is therefore the most sensitive and useful line for AAS analysis.
HCLs are coated with the element of interest to produce light of the
resonance wavelength(s) that is/are specific to element. When this light is
passed through a medium that contains atoms of the same element, the light will
be partially absorbed.
Atomizer
The purpose of an atomizer is to create a population of free atoms that is
suitable for absorption of light. The atomizer must have an energy source in
order to do this. Usually the energy comes from heat, and the most common
source of the heat is a flame (either an air/acetylene or a nitrous oxide/acetylene
flame). An AAS instrument with a flame atomizer is called a flame atomic
absorption spectrometer (FAAS). With a flame atomizer, the sample is
introduced into the flame as an aerosol (a mist of tiny droplets). The flame
burner head [8] is designed to be long, thin, and aligned with the light path.
Such a design causes the aerosol atoms to be atomized in the flame while they
are in the light path so that they can absorb the light.
A very important part of the atomizer system is the nebulizer. The
nebulizer is responsible for nebulizing a liquid sample into an aerosol. The
sensitivity of a FAAS instrument depends heavily on how efficiently the nebulizer
can convert a sample to an aerosol.
Optics
The optics system of an AAS instrument is responsible for getting the light
from the light source to the detector. Along the way, the light must be passed
through the atomized sample and through a monochromator. A monochromator
is used to isolate specific wavelengths from the bulk light that it receives. For
example, it may be necessary to isolate the analytical wavelength of interest
from light that was emitted from the fill gas of the HCL, or from stray room light
that entered the spectrometer.
Detector
The detector part of an AAS instrument measures how much light is
transmitted through the spectrometer. Most commonly a photomultiplier tube
(PMT) is employed for this purpose.
While the above four components are the essential ones of an AAS
instrument, there are still others that play important roles.
2 - Instrumentation
There inevitably are electronic devices that convert the signals from the
detector into something that is useful for the human researcher. Older
instruments used to employ signal meters and plotters that would chart the
strength of the absorbance signal on a moving strip of paper. Instruments of
today have replaced the meter and chart recorder with computer software that
has many more capabilities. Modern software provides real-time plots of
absorbance versus time, constructs calibration curves, and calculates statistics
such as RSD values.
On the starting end of the sample analysis spectrum, computer software
can keep track of samples that are running, that you will run, and that you did
run. It can also be used to setup and run the instrument without operator
intervention at all.
Optics
The ideal optics system will have the following characteristics:
Unfortunately, there is no such thing as an ideal optics system. The best that
you can hope for is to have a system that is optimized for your needs.
There are two ways to control the path of light within an optics system: with
mirrors or with lenses. Most optics systems make use of both.
Lenses
Good quality lenses are made from silica glass and have good light
transmission over a broad wavelength range (~190 - 900 nm). Transmission
losses occur at both interfaces of the lenses (i.e. at the lens surface where the
light enters and at the lens surface where the light exits). The losses are
typically between 10-14% for each lens in the optical path.
A feature of lenses that must be kept in mind is that the refractive index
of the lens is dependent on the wavelength of the light being refracted. This
means that the focal length of the lens will be different for every wavelength.
Rather than moving a lens to keep the focal point in the same position for
different wavelengths, optics systems will keep their lenses fixed and tolerate the
relatively minor losses associated with the changing focal lengths.
Most lenses used are designed for wavelengths in the UV region. This is
because most analytical wavelengths are in this range and the median refractive
index is about 250 nm.
For an air/acetylene flame, where the path length is around 10 cm, the
losses due to focal length differences are negligible. For a nitrous
oxide/acetylene flame, however, the path length is only 5 cm and the losses can
start to become noticeable. In GF AAS, where the sample atomization occurs
10
2 - Instrumentation
only at a small point in the center of the graphite tube, the focal length from the
light source is absolutely critical. As the light source wavelength is increased, the
focal point will move away from its original position at the center of the graphite
tube. Because of this, optics systems for GF AAS systems usually employ only
mirrors, and not lenses.
Mirrors
Mirrors perform much better than lenses, both in terms of reflecting
efficiency and focal length change. A mirror has a very thin (e.g. 2 m) top
coating of aluminum that reflects more than 90% of the light that strikes it (for
the range 190 - 900 nm). Also, when light reflects off a mirror there is no
change in the focal length. That is, the focal length of a mirror depends only on
its shape and is independent of wavelength.
Plane mirrors are used to fold light and curved mirrors (also called
collimating mirrors) are used to focus light. For example, a plane mirror is
needed to fold light around a 90 corner, and a curved mirror is needed to focus
that light onto an entrance slit. While the use of mirrors does solve the problem
of focal length differences, it raises the challenge of designing and manufacturing
focusing mirrors that are free from other optical aberrations, such as
astigmatism.
Because of the thinness of the mirror coating, mirrors are
extremely fragile and must be handled with care. Finger prints, and even soft
tissues, can irreversibly damage a mirrors coating. Reactive liquids and gases
can even cause harm. To increase their longevity, most optics mirrors are
further coated with a transparent silica or magnesium fluoride film for protection.
Even still, one should avoid any kind of direct contact with mirror surfaces.
Monochromator
There are several different designs of monochromators available. No
matter which design is used in an AAS instrument, however, some fundamental
principles remain the same. Light enters the monochromator through an
entrance slit. The light is folded and focused in the monochromator by use of
mirrors. The light is dispersed into its component wavelengths by some sort of
diffracting element. The diffracted light then leaves the monochromator through
an exit slit.
The most common monochromator design is the Czerny-Turner design,
which is used in the AI 1200. A schematic of this type of monochromator is
shown in Figure 2.1.
The Czerny-Turner monochromator uses two separate mirrors to collimate
and focus light. Mirror #1 receives the light that was focused through the
entrance slit. The light that reflects off this mirror is collimated into parallel
beams and then strikes the diffraction grating, which diffracts the light into a
spectrum of wavelengths. This spectrum is dispersed at a variety of angles,
depending on the wavelength of each component of the spectrum. The light then
strikes Mirror #2, which focuses the light through the exit slit and into the
detector. Some monochromators make use of only one mirror, but there is a
definite advantage to using two mirrors. The two mirrors will each be smaller
than a single mirror, so they are easier to manufacture. This means that there is
less chance for surface aberrations and therefore allows optimum light
throughput and resolution.
11
2 - Instrumentation
12
2 - Instrumentation
13
2 - Instrumentation
spaced from each other. The closer the grooves are spaced, the better the
resolving capability of the grating. Gratings are produced with groove densities
from 500 to 6000 grooves/mm.
The blaze wavelength of a diffraction grating is the wavelength of light
that will be most efficiently diffracted by the grating. Generally, gratings can be
used to diffract light that is 2/3 to 3/2 of the blaze wavelength. For example,
consider a grating blazed at 400 nm. The grating can be used for wavelengths
from 270 nm to 600 nm, but will diffract most efficiently wavelengths of 400 nm.
Ruled grating are physically etched, groove-by-groove, by a machine.
Basically, a mirror is mounted on a grooving machine, a diamond bit etches a
straight groove, the mirror is moved a short distance and then the bit etches
another straight groove parallel to the previous one.
Holographic gratings, on the other hand, are manufactured with light, not
a physical machine. A piece of glass is coated with a light-sensitive material,
which is then exposed to two parallel beams of coherent light that produce an
interference pattern on the coated glass. The bright areas of the interference
pattern (where the light beams add constructively) form the grooves in the
developed photoresist. A thin layer of aluminum is then applied onto the
etched glass piece to form a mirrored, grooved surface a diffraction grating.
The advantage of the holographic technique over the machine etching technique
is that the holographic technique produces no systematic errors, since the
grooves are the result of a perfect optical phenomenon. The machined
technique, on the other hand, is only as good as the quality of the etching
machine itself (which, for many purposes, is excellent, but will never be truly
perfect).
Gratings that are used in monochromators are always copies of master
gratings. A master grating is the original grating that was manufactured (either
holographically or physically). Subsequent gratings can be made from the
master by a process that essentially makes a molded copy of the master.
There are advantages and disadvantages to both types of gratings. The
appropriate one to use in a monochromator depends on several factors. Ruled
gratings produce significant more stray light than holographic gratings, and this
is especially true when groove density increases. For this reason, the maximum
groove density of rules gratings is around 3600 grooves/mm. Holographic
gratings can have up to 6000 grooves/mm. So, based on this factor, a
holographic grating is better to use if a higher groove density is required to
achieve a higher resolution and maintain a high signal to noise ratio.
Ruled gratings do exhibit significantly better efficiency than holographic
gratings. So, based on this factor, a ruled grating is more appropriate to use if
light throughput is critical (for example, if the light source is very weak). If the
light source is intense (as is the case in AAS instruments), then using a
holographic grating is more advantageous than using a ruled grating.
Slit Width
The slit width affects how much light enters and exits the monochromator,
and so is very important for light throughput. A wide slit width will allow more
light to reach the detector and will improve the signal strength. But, if nonanalytical lines also reach the detector, then this will increase the noise and
decrease the signal to noise ratio. Conversely, a narrow slit width will block out
all non-analytical wavelengths, but may reduce the light throughput so much to
make the signal to noise ratio unsatisfactory. So, the best slit width is the one
that allows the most light to reach the detector and blocks out most of the noise.
14
2 - Instrumentation
Atomizer
The atomizer is arguably the most important part of an AAS instrument,
since this is where the sample is converted into atoms that can absorb light.
Having an efficient atomizer is essential. The absorbance signal is completely
dependent on how many atoms there are to absorb the source light. So, a good
atomizer will display good sensitivity to the sample being analyzed, and a poor
atomizer will display poor sensitivity.
The ideal atomizer has the following characteristics:
Of course, no real atomizer is ideal, and the degree to which any sample
is atomized depends to a large extent on the element being analyzed.
Atomization is achieved by heating the sample to an extent where free
ground state atoms are formed. In FAAS, atomization is done with a flame. In
GFAAS, atomization is done with an electrically heated graphite tube furnace. In
VG-AAS, atomization is done with an electrothermally heated furnace or a flame.
Flame
The atomizer in a FAAS instrument uses a nebulizer to convert a liquid
sample into an aerosol. The sample is introduced into the nebulizer by aspiration
though capillary tubing. The aspiration occurs pneumatically from the flow of fuel
and oxidant gases through the nebulizer chamber. After the sample has traveled
through the capillary tubing, it strikes a glass impact bead. This impact bead is
designed so that when a stream of liquid strikes it, the liquid breaks apart into a
mist of drops (an aerosol). This aerosol invariably contains drops of many sizes.
The larger drops fall out, but the smaller drops remain suspended and are
thoroughly mixed with the fuel and oxidant gases as the mixture is carried into
the spray chamber. As the sample mist gets mixed with the gases, it moves
along through the spray chamber and up towards the burner head. The sample
entering the burner head is a uniform mixture of fuel gas, oxidant gas, and tiny
sample droplets. Once the mixture enters the flame, the process of atomization
by heat begins. The heat from the flame is usually sufficient to desolvate the
sample droplets. Then, the solid particles that were formed (e.g. salts) are
broken down, melted, or volatilized into gases. Finally, the molecules are
thermally dissociated into atoms that are capable of absorbing their characteristic
wavelength of light. Of course, the heat from the flame may be excessive and
15
2 - Instrumentation
may ionize some atoms (remove an outer electron from the atom), thus
decreasing the absorption signal. As well, the heat from the flame may be
insufficient and not atomize enough of the sample molecules. This also
decreases the absorption signal. The process of thermally atomizing the sample
in the flame is a complex equilibrium, and the actual chemistry inside the flame
at any stage (especially the atomization stage) is not clear. There are often
numerous side reactions that occur simultaneously.
The degree to which a sample may be ionized in a flame depends on the
element, since each element has different energy requirements for ionization. To
reduce ionization of atoms, easily ionizable elements (EIEs), such as the Group I
elements (Li, Na, K, Cs) can be added to the original sample solution. If the EIEs
are much more easily ionized than the sample atoms, then they will create a
large population of electrons in the flame and shift the atomization/ionization
equilibrium of the sample atoms towards the atomization side.
There are two types of flames that are commonly used for FAAS:
air/acetylene and nitrous oxide/acetylene. The air/acetylene flame (air being the
oxidant, acetylene being the reductant, or fuel) burns at around 2300 C. The
nitrous oxide/acetylene flame burns much hotter at around 3000 C. So, the
flame temperature is a factor when determining which type of flame is best
suited for atomization of a given element. Because of its cooler burning
temperature, the air/acetylene flame works well for elements that are relatively
easily atomized, such as copper, iron, nickel, and gold. The nitrous
oxide/acetylene flame, with its higher burning temperature, is needed to atomize
elements that require more energy to atomize, such as aluminum, silicon,
titanium, and tungsten.
Another important factor in optimizing the atomization of an element in a
flame is the stoichiometry of the oxidant and reductant gases. A lean flame is
fuel poor, and is therefore an oxidizing flame. A rich flame has excess fuel, and
is therefore a reducing flame. For each type of flame, certain elements are
atomized best in reducing flames, and certain elements are atomized best in
oxidizing flames. There is extensive data on the absorbance characteristics of all
the elements in flames. For example, in a reducing flame there are excess
carbon and hydrogen atoms present in the flame (from the acetylene molecules).
These excess atoms help break down the strong oxide bonds that form with some
elements, such as chromium. Other elements that are best atomized in a
reducing flame are tin and molybdenum. On the other end of the spectrum,
elements like silver, cadmium, gold, and nickel are best atomized in an oxidizing
flame. Some elements, like iron and gallium, are best atomized in a
stoichiometric flame (i.e. neither rich nor lean). Furthermore, some elements are
satisfactorily atomized over a wide range of flame gas mixtures. Copper, for
example, is atomized in both rich and lean flames. For this reason, copper is
often used to test or validate the sensitivity of an AAS instrument.
The major disadvantage of the flame atomization technique is that it is
very inefficient at converting the original sample to atoms. Overall, the atomizer
system of a FAAS instrument can convert less than 0.1% of the original sample
to absorbing atoms. The nebulizer component usually transports only less than
10% of the aspirated sample into the aerosol, and the other 90% is lost as waste
in the spray chamber and nebulizer chamber. Furthermore, the sample that does
make it into the flame (in aerosol form) is already greatly diluted from its mixing
with the flame gases. And once the sample gets atomized in the flame, the
atoms residence times in the light path are extremely short. Atoms travel
through the light path at great speeds (at least 1 cm/10 ms) as they exit the slit
in the burner head and travel up the flame.
16
2 - Instrumentation
Graphite Furnace
The graphite furnace (GF) atomizer solves the two major problems of the
flame atomizer: poor atomization efficiency and short residence times. A
graphite furnace is a tube that is connected to two low voltage electrodes. When
a current is forced through the tube, the tube heats up. The amount of heating
caused by the current flow can be accurately controlled, and the atomization of
samples in a graphite furnace is usually performed over several heating steps.
The graphite furnace is mounted in the electrodes so that one open end of the
tube faces the light source and the other open end faces the entrance to the
optics. This allows the light to pass freely through the graphite furnace along the
axis of the tube. The tube is aligned so that the light path travels down the
graphite tube axis and right through the center of the tube.
With a GF atomizer, a very small amount of liquid sample (between 5 and
100 L) is placed inside the center of the graphite tube. A heating program to
atomize the sample consists of at least three principal steps:
1. Drying Step
The graphite tube is quickly heated to a temperature just below the
boiling point of the solvent. Then the temperature is slowly ramped past
the boiling point. This step gently evaporates the solvent (without
causing splattering or sample ejection) and leaves the dried sample
inside the tube.
2. Ashing Step (or Charring)
This step removes any dry or semi-dry matrix that is left over from
the drying step. Matrix modifiers can be added to the sample before the
heating program to stabilize the sample during the ashing step. Modifier
gases, such as oxygen or hydrogen, can be added to the graphite
furnace workhead during the ashing step to help remove the matrix.
Ashing temperatures depend on the element being analyzed in its
matrix. For example, cadmium ashes at 300 C, arsenic at 1400 C, iron
at 600 C, and lead at 480 C.
3. Atomization Step
At this stage, the sample is a dry solid at the bottom of the graphite
tube. The sample is atomized by rapidly increasing the temperature of
the tube. The AI 1200 can heat at a rate upto 3800 K/s. The required
atomization temperature depends on the element being analyzed. For
example, cadmium atomizes at 1250 C, arsenic at 2250 C, nickel at
2250 C, and lead at 1400 C.
There are distinct advantages to the graphite furnace atomizer when its
performance is compared to the flame atomizer. The graphite furnace atomizes
100% of the sample (compared to less than 0.1% for the flame). Also, the
residence time of the atomized sample in the light path is much longer in the
graphite furnace than in the flame. The residence time in the graphite tube can
range from 0.2 to 0.5 second, whereas in the flame it is only milliseconds. Both
of these factors increase the sensitivity of the graphite furnace atomizer, and
detection limits with GFAAS are typically one to two orders of magnitude better
than with FAAS.
17
2 - Instrumentation
Detector
The detector is the part of the AAS instrument that receives the light
output from the monochromator. The detector quantifies how much light it
receives and creates an electrical current. That current is then amplified and
converted into a digital signal that is recorded by a data acquisition system. By
far the most common detector used in AAS instruments is the photomultiplier
tube (PMT).
Essentially, a PMT is a photon counter. Light from the monochromator
enters the PMT through a quartz window. Photons (the quantum packets that
make up light) strike the photocathode of the PMT. This converts the photon to a
photoelectron via the photoelectric effect. However, the production of a single
electron from a single photon wont generate a very strong signal, so an
amplification 5 or 6 orders of magnitude is required. This amplification is
achieved through the use of a series of 8 to 12 dynode plates. A voltage is
applied between the photocathode and first dynode (on the order of 100 V). This
voltage difference causes the photoelectron to accelerate from the photocathode
to the dynode. When it strikes the dynode, several more electrons are produced.
These electrons are in turn accelerated towards the second dynode, since there is
also a voltage difference across the first and second dynodes. Each of the
electrons striking the second dynode creates several more electrons. For
example, if the electron collision into the first dynode created 5 electrons, then
the collision of those 5 electrons into the second dynode will create 25 electrons.
This chain reaction of electron production continues along the series of dynodes.
If there are 12 dynodes in the chain, then the original single photon will produce
244,140,625 electrons from the final twelfth dynode. These final electrons are
then collected by the PMT anode. The current that can result from the collection
due to a single photon by the PMT can be as high as 100 mA! Clearly, PMTs are
extremely sensitive detectors. Furthermore, the large amplification of the signal
that is achieved by the PMT is achieved with very little increase in noise.
18
2 - Instrumentation
19
2 - Instrumentation
20
Chapter 3:
Background Correction
Methods
Fundamentals
The Frequency of Measurement
The Interval between
Measurements
The Function used to Calculate
Net Absorption
Spectral and Structure
Backgrounds
The Effect on Linear Working
Range
Deuterium (D2) Background
Correction
Smith-Hieftje (S-H) Background
Correction
Zeeman Background Correction
Comparison of Background
Correction Methods
21
Fundamentals
Background correction is a necessary part of any good AAS instrument. Basically,
there are two types of backgrounds that need to be corrected for: non-specific radiation
and non-specific absorption.
Non-specific radiation was dealt with in Chapter 2. Non-specific radiation is the
extra light (non-analytical wavelengths) that can pass through the optics of an AAS
instrument, enter the monochromator, and have the chance of reaching the detector. If
this non-specific radiation reaches the detector, it will result in a falsely signal. Nonspecific radiation can come from many sources, including the HCL fill gas, sunlight, room
light, and the light emitted by the flame itself. As was discussed in Chapter 2
(Monochromator section), it is the job of the monochromator to effectively filter all entering
light and allow only a specific, variable wavelength to exit and reach the detector. So, the
monochromator ensures that the only light that reaches the detector is the wavelength of
light that is being absorbed by the sample being analyzed.
The other type of background that must be corrected for is non-specific absorption
(and will be discussed in this chapter). This correction cannot be accomplished by the
monochromator, since it involves the same analytical wavelength that is selected by the
monochromator. Non-specific absorption (also called background absorption) has a
broadband effect and occurs when the source light is prevented from reaching the detector
by means other than absorption by analyte atoms, such as scattering and blocking of light
by other species in the light path. Molecular species and solid particles present in the flame
are the major causes of non-specific absorption. When the heat of the flame is not
sufficient to fully break down all molecular species (for example, matrix compounds), there
can be sufficient remaining molecules to absorb, block, or scatter the source light. These
molecules can be thought of as causing the same interference as putting ones hand in the
light path: the light gets blocked, less light reaches the detector, and a falsely high
absorbance signal results (because the signal analysis software thinks that less light is
reaching the detector because more light is being absorbed by the analyte atoms).
In FAAS, the background absorption is relatively minor (usually less than 0.05
absorbance units). In GFAAS, on the other hand, background absorption is severe and can
reach levels of 2.0 absorbance units. Therefore, effective background correction methods
become essential for accurate GF analyses.
There are three background correction methods currently being used by AAS
instrument manufacturers: Deuterium (D2), Smith-Hieftje (S-H), and Zeeman. For all
three methods, there are several common factors that determine the effectiveness of the
method:
The
The
The
The
The
23
24
25
Figure 3-2 The principle behind the Smith-Hieftje background correction technique
When the S-H method is used, the HCL is alternately pulsed between normal
operating currents (where no significant self absorption occurs) and high intensity currents.
The duration of the HI pulses is very short compared to the normal current pulses. During
the period of the normal pulses, the analyte absorbance signal is measured. During the
period of the HI pulse, the background absorbance signal is measured.
With the S-H method, the profile of the emission from a HCL is semi-broadband (in
that the emission has been broadened to an extent) and the intensity of the analytical
wavelength (the valley between the shoulders) has been reduced. So, during the HI pulse,
the wavelengths above and below the analytical wavelength (the shoulder wavelengths)
are used to measure the background absorbance. The attenuation of the alreadydiminished analytical wavelength by analyte atoms is negligible compared to the
attenuation of the shoulder wavelengths by the non-specific absorbing species. So, the
signal measured during the HI pulse is the background signal (BAS)
Of course, during the normal pulses, the HCL emits the same sharp line as for the
D2 background correction method. So, during the normal pulse of the HCL, the total
absorption signal (TAS) is measured.
A key advantage of S-H background correction method, in terms of accuracy, is that
only one lamp is needed. Since a single lamp acts as both the continuum source and the
sharp line emitter, there will never be any problems with the two optical paths not being
exactly the same.
27
However, there is also a disadvantage of the method related to accuracy. Since the
shoulder wavelengths of the HI pulse are used to measure the background (and not the
analytical wavelength itself), there is some degree of approximation going on when that
background measurement is subtracted from the total absorbance signal (for which the
analytical wavelength is used). The assumption is that the background doesnt change
much in the short spectral width between the two shoulder wavelengths. In practice, this
assumption is adequate.
There are several other disadvantages to the S-H background correction method.
While the requirement of a special HCL (designed for S-H background correction) cant
really be called a disadvantage in itself (since all background correction methods require
some kind of additional equipment), the fact that special HCLs are not available for all
elements limits the methods usefulness. Also, it is required that the high and low currents
of the HCL be stabilized before any measurements can be made. This results in a typical
sensitivity loss of 50%. Calibration curves can sometimes exhibit roll over points at higher
sample concentrations, so this decreases the linear dynamic range of analysis. Also, the SH method cannot correct for any structured or spectral interferences. Finally, the method
is limited to relatively low modulation frequencies (normal 10 Hz), so is incompatible with
the rapid transient signals common to GFAAS experiments.
element being determined). Also, it frequently results in roll over of calibration curves at
higher concentrations, so the linear dynamic range is significantly reduced. Another
important point relates to the cost of the Zeeman instrumentation, which requires a
relatively large initial investment. This is a result of all the additional equipment that is
required on an AAS instrument equipped with Zeeman background correction (a stable
electromagnet, a power supply, and a polarizer).
29
Zeeman
Moderate, 20% loss
GFAAS
180-900 nm
All
Normal
Roll over occurs
Decreased
Good (single optical
path)
Yes
60-120 Hz
4.5 ms
D2 seems to be the overall best background correction method of all the three,
since it offers the broadest range of advantages and does not suffer in the really key areas
of sensitivity and calibration linearity.
There seems to be an impression in the analytical chemistry community that
Zeeman background correction is the superior method available. However, once the
factors have all been weighed, it should become apparent that this impression is flawed.
As was mentioned before, the fact that Zeeman method is the only one that can correct for
spectral and structured background interferences is of no consequence in over 99% of the
samples that will be encountered in a laboratory. Even if it is determined that this feature
is necessary and Zeeman background correction is required, there will be other important
factors that will suffer. Initially, the cost of the Zeeman instrumentation is very high
compared to D2 and S-H. Also, Zeeman has significantly poorer sensitivity, calibration
linearity, and linear dynamic range than the D2 method offers. Another misconception is
that D2 background correction is inadequate for samples with rapidly changing background
signals. The frequency of measurement and interval between measurements
characteristics show that D2 is on par or better than Zeeman is in these areas.
Based on all these considerations, the decision between the D2 and Zeeman
background correction methods can be boiled down to the following:
Choose D2
if you need:
Choose Zeeman
if:
maximum sensitivity
calibration linearity
maximum linear dynamic range
rapid measurement capability
low cost
30
31
32
Chapter 4:
Comparison of AAS
Techniques
Things to Consider
Applications
Expected Concentration Ranges
Elements
Atomization Efficiency
Interferences
Spectral
Background
Matrix
Detection Limits
Sensitivity
Precision
Linear Working Range
Minimum Sample Volume
Sample Throughput
Sample Usage
Total Dissolved Solids
Method Development
Ease of Use
Automation/Unattended Operation
Costs
Initial Investment
Running Costs
33
Things to Consider
When deciding whether to purchase a flame AAS or a graphite furnace AAS, one should
first make an assessment of ones analytical needs:
Applications
The applications that you will be running (e.g. environmental, marine, petroleum,
mining) will play a part in determining the AAS technique that is right for you. However,
many of the techniques overlap when they are compared on the basis of applications.
Therefore, your particular application will be a relatively minor factor in the decision. That
aside, there are some applications that one technique is much better suited for than the
other (this is usually based on the element(s) being analyzed).
Its also important when deciding on a AAS instrument to think of the future. Will
your anticipated long-term needs be satisfied by the instrument that is appropriate to
purchase for your needs today?
Elements
The FAAS technique can analyze a total of 61 elements with its air/acetylene and
nitrous oxide/acetylene flames. The GFAAS technique can analyze a total of 41 elements.
Also, any element that GFAAS can analyze can also be analyzed with FAAS.
Atomization Efficiency
The atomization efficiency is arguably the most important performance
characteristic of any AAS technique. The air/acetylene flame burns at around 2300 C, and
the nitrous oxide/acetylene flame burns significantly hotter at around 3000 C. For some
elements, the cooler air flame is not hot enough to atomize the sample, and the nitrous
oxide flame is required instead. However, the excess energy of the hotter nitrous oxide
flame can ionize other elements. So, its often a trade-off with air flame atomizers. Some
elements have low atomization energies, and so are better suited to the cooler flame
34
(copper, iron, nickel, and gold). Others have high atomization energies, and so a nitrous
oxide flame is needed (aluminum, silicon, titanium, and tungsten).
And for some elements, no flame is hot enough to efficiently atomize the sample.
In these cases, the GFAAS technique is needed, because the furnace can reach
temperatures up to 3000 C.
Another consideration is the nebulization efficiency of the FAAS nebulizer. Only less
10% of original sample introduced into the spray chamber gets converted in an aerosol and
makes it into the flame for atomization. In this sense, the FAAS technique is very
wasteful. The GFAAS technique, on the other hand, is very efficient. All of the sample
deposited in the graphite furnace remains inside the furnace and therefore has a good
chance of being atomized. As well, atomized sample have relatively long residence times
inside the furnace tube (0.2 0.5 s)
Interferences
Spectral
Spectral interferences do occur, but are rare, in both FAAS and GFAAS techniques.
Background
Background interferences are much more common, and both techniques are
vulnerable. Background correction techniques are usually employed to compensate for the
interference.
In GFAAS, background interference typically arises from vaporized and atomized
matrix components.
Matrix
Matrix interferences can be serious for both FAAS and GFAAS analyses. For GFAAS,
the effects can be countered by adding modifiers to the samples. For FAAS, the ionization
buffers are usually employed to control matrix effects related to ionization.
Detection Limits
The detection limits of the GFAAS technique (ppb range) are generally 100-1000
times better than the FAAS technique (ppm range).
Sensitivity
This factor is closely linked to detection limits. GFAAS is much more sensitive than
FAAS, since the graphite furnace atomizer has much longer residence times and much
higher atomization efficiency than the flame atomizer.
Precision
The flame atomizer has better short term precision (0.5%) than the graphite
furnace atomizer (2-5%). Long term precision is usually better with the flame atomizer
too, since the graphite furnace invariably degrades and gets worn out as it is used.
Sample Throughput
The sample throughput is the rate at which individual samples can be analyzed. It
is often expressed in the format minutes per sample. FAAS has a higher sample
throughput rate than GFAAS. The reason for this is that the graphite furnace atomization
sequence has multiple steps, each with hold times, and a cool-down time is required at the
end of each analysis.
Typical sample throughput values are 10 seconds per sample for FAAS analyses and
4 minutes per sample for GFAAS analyses.
Sample Usage
Sample usage refers to how much sample is consumed by the instrument while it
performs an analysis. FAAS uses a lot of sample solution for each analysis, and much of it
is wasted and never even makes it into the atomizer for an absorption measurement.
GFAAS, on the other hand, uses very little sample and wastes none. The entire volume of
sample deposited in a graphite furnace remains in the graphite furnace during the
atomization steps.
Method Development
The FAAS is easier to develop methods for than the GFAAS instrument. This is
mostly because of the strict constraints set on the optical alignment of the furnace atomizer
and the requirement of accurate and consistent sample injections. As well, the operation of
the FAAS doesnt require as much experience as the operation of the GFAAS. Nonetheless,
for both instruments, a skilled operator is required to develop a good method for a specific
application. Fortunately, most instruments come equipped with extensive libraries of premade, common methods that are ready to use.
Ease of use
The FAAS instrument is very easy to use, and an expert operator is not required for
routine analyses. The GFAAS instrument is also relatively simple and straight-forward to
use, but a skilled operator is still required because accurate and consistent sample
injections are essential for good results, even for the most routine analyses.
36
Automation/Unattended operation
GFAAS lends itself well to automation. The use of inert gases does not pose any
dangers, so unattended operation is also feasible. Overnight, automated operation of
GFAAS is commonly employed to achieve a higher sample throughputs. FAAS is equally
suited to automation, but unattended operation is unadvisable due to the use of
combustible gases. An operator should always be present to ensure the safe operation of
the instrument and to be able to deal with gas leaks or other emergencies as they arise.
Costs
Initial Investment
There is no way to specifically say how much an instrument will cost, since there are
many factors involved. The cost depends on the configuration, accessories, options,
background correction method, and, of course, the vendor. Roughly speaking, FAAS
instruments can cost from $20,000 to $35,000. GFAAS instruments are more expensive,
and usually cost about 2 times as much ($40,000 to $70,000).
Running Costs
Both techniques are comparable in terms of the costs of running the instruments on
a day to day basis, although FAAS is generally a cheaper instrument to run. There is a
constant expense for the fuel and oxidant gases in FAAS (although, air is free), and the
inert gas used in GFAAS (usually argon) isnt cheap. In addition, a periodic expense of the
GFAAS technique is the replacement cost of the graphite furnace atomization tube, which
has a limited life and needs to be replaced from time to time to ensure efficient
atomization.
Another consideration is whether your specific applications with GFAAS will require
clean room conditions. Maintaining clean room conditions is no small task and adds
considerably to the operating costs of the instrument.
The following table is a summary of the comparisons between FAAS and GFAAS.
Table 4.1: A Summary Comparison of Flame AAS and Graphite Furnace AAS
FAAS
GFAAS
Detection Limit
sub ppm
sub ppb
Linear Working Range
103
102
Precision
0.5%
2 - 5%
Sensitivity
good
excellent
Spectral Interferences
virtually none
occasionally
Chemical/Matrix Interferences
many
many
Ionization
occasionally
minimal
Minimum Sample Volume
0.5 mL
20 L
Sample Throughput
~10 s/sample
~4 min/sample
Sample Usage
high
low
Number of Elements
61
41
Maximum Dissolved Solids
~ 3%
> 20%
Ease of routine use
easy
requires expertise
Method Development/Optimization
easy
difficult
Easily Automated?
moderately
yes
Combustible Gases?
yes
no
Initial Investment
low
high
Running Costs
low
moderate
37
38
39
Chapter 5:
Standards and Sample
Preparation
Apparatus
Water
Standard and Blank
Solutions
Sample Solutions
Storage of Solutions
Calibrations
Matrix Effects
Chemical Interferences
Incomplete Dissociation
of Analyte Compounds
Ionization
40
Apparatus
All apparatus to be used during the process of an analysis, from the sample and
standard preparation to the introduction into the instrument, must be scrupulously clean.
Even glassware that had been cleaned previously and stored should be cleaned again
immediately before use.
As well, the instrument itself should be periodically cleaned to ensure that no
contamination errors or memory effects are introduced. In FAAS instruments, the sample
aspiration tubing is easily and inexpensively replaced. The nebulizer chamber should also
be cleaned and washed regularly. Residue buildup can be a problem, especially when flow
spoilers or baffles are employed. Corrosion resistant nebulizer should be used when
corrosive acids or organics are being analyzed (e.g. use a Teflon constructed spray
chamber and glass impact bead).
Volumetric flasks should be Class A (accurate to 0.01%), since Class B glassware
will introduce systematic measurement errors into your analysis.
Analytical balances should be high-accuracy and high-precision. Less than 1% error
is needed when weighing out samples to ensure minimal error in prepared solutions.
Pipettes and micropipettes should always be calibrated to ensure accuracy of
delivery volumes.
Water
All water used for AAS analyses should be deionized. Distilled water alone is likely
to have significant amounts of dissolved metals and gases, and can introduce significant
error into analyses. Using distilled water that has also been deionized is acceptable.
For some GF applications, particularly when trace levels are being determined,
simple deionized water is not good enough. In these cases, an ultra-pure water treatment
is needed (eg. the Millipore system).
In general, one can expect most acids to contain trace amounts of metallic
impurities. While the levels are trace, they can still introduce error in some applications
(particularly in trace-analyses and GF analyses). Fortunately, specially purified acids are
commercially available to alleviate this source of error.
is the same, but the chance for error is much less, since a small volume measurement
error has a relatively greater effect on a small volume than on a larger volume. However,
the error in volume measurements that do occur in the two step method are doubled, since
the dilution is performed twice. However, the relative effect of the volume measurement
error is of far greater concern. In this example, the two step dilution uses volumes of the
bulk standard that are 100 times greater than volumes used with the single dilution
method. Therefore, the two step dilution method, in theory, has 50 times less error.
Ideal standard solutions are solutions that are identical in chemical and physical
composition to the sample solutions, except that they have known concentrations of the
analytical component. Ideal blank solutions are solutions that are identical in chemical and
physical composition to the sample solution, except that they dont have any of the
component that is of analytical interest. Proper standards and blanks should be prepared
by the same procedure that the sample is prepared by. For example, if the sample is
microwave digested by nitric acid, then the standards and blanks should also be microwave
digested by nitric acid (expect that the standards will have a known amount of the
analytical component added, and the blanks will have none added).
Matching the physical properties of sample with standards and blanks is important
because the amount of solution that is aspirated by the nebulizer and into the flame
depends, to an extent, on the physical properties of the solution. The viscosity, density,
surface tension, and solvent vapor pressure will all affect the aspiration rate.
Sample Solutions
There are many different kinds of sample solutions. Sometimes the sample is
obtained from a third party, and is already in ready-to-analyze form. Often, however,
digestion and dissolution steps are required. In either case, it is important that sample
solutions be homogenous and free of solids. This is especially important for the GFAAS
analyses, since the sample volumes analyzed are much smaller than for FAAS. If 50 L of
a heterogeneous sample is deposited in the graphite furnace and analyzed, it is very likely
that its signal will be different from the next 50 L sample analyzed, producing inconsistent
results. In other words, it is very difficult to get a representative sample aliquot from a
heterogeneous sample solution.
Storage of Solutions
It is recommended that all working standard solutions to be used in an analysis be
prepared (from the bulk standards) fresh on the same day as the analysis. All solutions
should be labeled and dated to keep track of them. This is to prevent any errors due to the
standard solution ions absorbing into the container walls, and also molecules leaching from
the container walls into the sample solution. The same applies to sample solutions. Bulk
solutions stored on the shelf for long periods should have high concentrations (at least
1000 ppm). There is significant degradation solutions with low concentrations (less than
10 ppm), mostly from adsorption onto the container walls, so it is imperative that they be
prepared fresh and utilized promptly.
Additionally, when storing solutions, it is good practice to use polyethylene
containers instead of glass vessels. This is because some metal ions have an affinity for
glass walls and will become adsorbed onto glass surfaces.
A common and convenient way to prolong the shelf life of working samples and
standards is to acidify the solutions with mineral acids to a pH of about 2. Nitric acid
(HNO3) is most frequently used.
42
Calibrations
The method of calibration used in an analysis depends almost entirely on the
chemical and physical properties of the unknown sample. For example, if the sample
solution has a very simple aqueous matrix, and its properties and behavior in the atomizer
are well known, then a normal calibration will be sufficient. On the opposite end of the
spectrum, if the sample solution has a complex matrix and the chemical and physical
properties are not all known, then a standard addition calibration method will be required.
The former case can be divided into two sub categories: samples with simple,
known matrices, and samples with complex, known matrices. In both cases, the matrices
are reproducible in the standards and samples, and so a normal calibration is sufficient. An
example of a simple, known matrix is a sample that has an unknown amount of copper
dissolved in nitric acid and water. An example of a complex, known matrix is a sample that
has an unknown amount of copper from a steel sample. This sample has a complex matrix
because there are other elements present (eg. iron, nickel, aluminum, and chromium).
The matrix is known because the amount of other elements present is roughly known.
The latter case can also be divided into two sub categories: complex, unknown
matrices, and organic matrices. In both cases, the matrix is not reproducible in the
standards and blanks, so the differences between the signals of the sample and standards
would be due to more factors than just the analytical component. An example of a
complex, unknown sample matrix is a sample with copper that was derived from raw oyster
meat by mechanical grinding, two-stage acid dissolution, and then furnace heating and
drying. While all the digestion steps could be reproduced in the standards and blanks, the
added matrix components from the oyster meat could not.
To accurately analyze samples with unknown matrices, a standard addition
calibration method is required. This usually means preparing a series of standards that
are prepared from samples of the unknown that have different-sized aliquots of the analyte
element added to them. The standards prepared in this way all have a matrix that is
approximately the same as the unknown sample, since all the standards were prepared
from the unknown sample.
The calibration curve is generated by first plotting absorbance versus amount of
standard added to unknown. The amount can be expressed as volume, mass, moles, or
whatever unit is appropriate. So, when the unknown is analyzed without any standard
added, the absorbance value is plotted at x = 0 mL. When the unknown is analyzed with,
for example, 0.5 mL of standard added, the absorbance is plotted at x = 0.5 mL. A
calibration curve is generated by continuing on in this manner. Extrapolation of the curve
to intersect the x-axis gives the concentration of analyte element in the original unknown
sample.
Matrix Effects
The precipitation of the element of interest from working solutions is a common
cause of inaccurate analyses. This is common, for example, with tin and silicon solutions.
Chemical Interferences
The technique of AAS is based on the formation of ground state atoms being formed
in the atomizer, thereby making them capable of absorbing light. The usefulness of AAS
can be severely compromised when the formation of ground state atoms is inhibited in any
way. Chemical interference is one cause of such inhibition. Chemical interference is a
result of either of two phenomena:
(a) incomplete dissociation of analyte compounds
(b) ionization of analyte atoms
43
44
45
46
Chapter 6:
FAAS Analytical Data Sheets
Introduction:
How to Use These Analytical Data Sheets
The following Analytical Data section is the heart of this cookbook. It has guideline
methods for each individual element. It is important to realize that the parameters listed
are only meant as guidelines. The parameter values listed will not necessarily lead to the
best performance on every instrument. Rather, they are to serve as a starting point for
developing optimized methods on an instrument-by-instrument basis.
For example, on page 56 is the analytical data for analyzing cadmium by FAAS. The
parameter PMT voltage is listed as 375 V. This does not mean that 375 V will be the right
voltage for every instrument. In fact, its very unlikely that it ever will be. Rather, the 375
V value is meant as a ballpark figure; it indicates that the PMT voltage of an optimized
method for measuring cadmium with FAAS should be around 375 V. This information is
valuable to the operators creating the method because they have a general idea before
hand of what the PMT voltage should be. When the operator creates a method for
cadmium, they can start at 375 V and make adjustments from there. This is much easier
and quicker than guessing at a starting point. The operator could guess to start at 100 V,
which would definitely be too low. Or the operator could guess to start at 600 V, which
would definitely be too high. In summary, the parameter values listed in the Analytical
Data sheets are meant as guidelines, or starting points, for creating optimized methods.
They are not meant as exact values that could be cut-and-paste into blank methods to
produce perfect, optimized methods every time.
There are a few things that the operator should know when looking at the Guideline
Parameters table of the data sheets. An oxidizing flame is fuel lean (relatively low ratio of
fuel to air) and is blue in color. A reducing flame is fuel rich (relatively high ratio of fuel to
air) and is orange or red in color. A stoichiometric (or neutral) flame has a fuel to air ratio
of one.
The characteristic concentration of an element is the concentration of analyzed
sample that would produce an absorption signal of 0.0044 Abs, or 1% absorption.
The working range is the concentration range within which a reasonably linear
calibration curve can be constructed. Excessive curvature of the calibration curve occurs
outside of this range.
Some graphs are included on the element data sheets if the absorption sensitivity of
the element is sensitive to the parameter in the graph. For example, a graph of
absorbance vs. lamp current is included on the data sheet for silver because the absorption
signal depends heavily on the lamp current. The graph gives the operator a general idea of
what kind of absorption changes to expect when the parameter is varied.
Information is provided on how to prepare standard solutions for each element.
Alternatively, certified standards can be purchased from chemical suppliers.
47
48
13
Al
Analytical lines:
Aluminum
309.3 nm
396.2 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
0
Interferences
Aluminum is prone to ionization in the nitrous oxide
flame. Add an easily ionizable element (K or La,
0.1%) to alleviate this interference.
Absorption is depressed by silicon, calcium, and
phosphate.
Absorption is increased by iron, titanium, nickel, and
manganese.
49
25
50
75
Concentration (mg/L)
100
51
Sb
Analytical lines:
Antimony
217.6 nm
231.2 nm
Absorption Sensitivity
(217.6nm upper; 231.2nm lower line)
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.15 mg/L
0.1-30 mg/L
Standard Solutions
0.6
0.4
0.2
0
0
10
20
30
C oncentration (mg/L)
40
OR
Dissolve 2.743 g of K(SbO)C4H4O6 H2O
(potassium antimonyl tartrate) in H2O. Dilute to 1
L to give a 1000 ppm Sb solution.
0.4
0.3
0.2
0.1
0
0.5
Interferences
High levels of acids cause a depression of the
signal. Usually, matrix matching of solutions can
compensate for this interference.
Cu and Ni decrease the signal. This interference is
more pronounced in reducing flames than in
oxidizing ones.
50
1.5
2
2.5
Fuel Flow Rate (L/min)
3.5
33
As
Analytical lines:
Arsenic
193.7 nm
197.2 nm
A bsorption S ensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6 mg/L
2 - 100 mg/L
0.6
0.4
0.2
0
0
40
80
C oncentration (m g/L)
Standard Solutions
120
Absorbance
0.4
Interferences
The flame itself absorbs or scatters more than half
of the source light at 193.7nm, so background
correction is highly advised.
The 189nm line is generally not used because of
heavy atmospheric absorption.
Total salt content greater than 1% can also cause
non-specific absorption.
51
0.3
0.2
0.1
0
1
1.5
2
Fuel Flow Rate (L/min)
2.5
56
Ba
Analytical lines:
Barium
553.6 nm
350.1 nm
Parameter
0.8
Absorbance
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
10 mA
PMT Voltage:
381 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.9 mg/L
Working Range:
0.4-150mg/L
Data collected with potassium chloride added
(final concentration of 2g/L) to suppress
ionization.
0.6
0.4
0.2
0
0
50
100
Concentration (mg/L)
Standard Solutions
Dissolve 1.7785 g of BaCl2 2H2O (barium chloride)
in H2O. Dilute to 1 L to give a 1000 ppm Ba
solution.
OR
Dissolve 1.437 g of BaCO3 (barium carbonate) in a
minimum volume of 1:1 HCl. Dilute to 1 L with
1% HCl to give a 1000 ppm Ba solution.
Interferences
A b s o rb a n c e
0 .3
0 .2
0 .1
52
be fore
a fte r
A d d itio n o f 2 g / L K C L
150
Be
Analytical lines:
Beryllium
234.9 nm
Absorption Sensitivity
1.2
1
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.8
0.6
0.4
0.2
Standard Solutions
0
0
1
2
3
C oncentration (mg/L)
53
Interferences
0.4
0.3
0.2
0.1
3.5
4.5
5.5
Fuel Flow Rate (L/min)
6.5
83
Bi
Analytical lines:
Bismuth
222.8
306.8
206.2
227.7
nm
nm
nm
nm
Flame:
air/acetylene
Flame Stoichiometry:
oxidizing
HCL Current:
10 mA
PMT Voltage:
409 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.19 mg/L
Working Range:
0.4-35 mg/L
Standard Solutions
0.6
0.4
0.2
0
0
Interferences
10
20
30
Concentration (mg/L)
40
Absorbance
0.2
0.1
0
0
54
0.2
0.4
0.6
Slit Width (nm)
0.8
B
Analytical lines:
Boron
249.7 nm
208.9 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
Standard Solutions
0
0
0.4
55
Absorbance
3000
Interferences
1000
2000
C oncentration (mg/L)
0.3
0.2
0.1
0
0
0.2
0.4
0.6
Slit Width (nm)
0.8
48
Cd
Analytical lines:
Cadmium
228.8 nm
326.1 nm
air/acetylene
oxidizing
5.0 mA
375 V
0.2 nm
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.8
0.6
0.4
0.2
0
0.01 mg/L
0.02 - 2.5 mg/L
1
2
Concentration (ppm)
Standard Solutions
Interferences
No major interferences to note.
56
0.44
Absorbance
0.43
0.42
0.41
0.4
0.4
0.6
0.8
1
1.2
1.4
Fuel Flow Rate (L/min)
1.6
20
Ca
Analytical lines:
Calcium
422.7 nm
239.9 nm
Flame:
air/acetylene
Flame Stoichiometry:
reducing
HCL Current:
10 mA
PMT Voltage:
254 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.066 mg/L
Working Range:
0.005 - 4 mg/L
0 .6
0 .4
0 .2
0
0
6
9
12
C o n c e n tr a tio n (m g / L)
15
Absorbance
Standard Solutions
0.1
0.05
0
0.5
Interferences
In the air/acetylene flame, interferences such as
refractory oxide depress the calcium absorbance.
Addition of a releasing agent such as strontium or
lanthanum is used to release calcium atoms for
determination.
57
1
1.5
2
Fuel Flow Rate (L/min)
2.5
20
Ca
Analytical lines:
Calcium
422.7 nm
239.9 nm
Standard Solutions
Add 2.497 g of dried CaCO3 (calcium carbonate) to 50
mL of H2O. Dissolve with a minimum volume of 1:4
HNO3, adding the acid dropwise. Dilute to 1 L to give
a 1000 mg/L Ca solution.
A bso rban ce
0.8
0.6
0.4
0.2
0
0
0 .5
1
1.5
C o n c e n tr a tio n ( m g / L)
0.2
0.1
0
3
4.5
6
7.5
Fuel Flow Rate (L/min)
Interferences
0.3
Absorbance
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
15 mA
PMT Voltage:
240 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.0096 mg/L
Working Range:
0.005 - 2 mg/L
0 .3
0.25
0 .2
0
0 .2
0.4
S lit W i th ( n m )
58
0.6
0 .8
55
Cs
Analytical lines:
Cesium
852.1 nm
894.5 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
Standard Solutions
10
15
Concentration (mg/L)
20
Interferences
0.4
0.35
Absorbance
0.3
0.25
0.2
0.5
59
1
1.5
Fuel Flow Rate (L/min)
24
Cr
Analytical lines:
Chromium
357.9 nm
425.4 nm
air/acetylene
reducing
10 mA
348 V
0.2 nm
(3 5 7 . 8 7 n m , u p p e r ; 4 2 5 .4 4 n m lo w e r lin e )
1
0.8
Abso rban ce
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.04 mg/L
0.04 - 10 mg/L
Standard Solutions
0.6
0.4
0.2
0
0
Interferences
4
6
8
C o n c e n tr a tio n (m g / L)
10
60
A b s o rb a n c e
0 .3
0 .2
0 .1
0
1.5
2
2.5
F u e l F lo w R a te ( L/ m in )
24
Cr
Analytical lines:
Chromium
357.9 nm
425.4 nm
Parameter
nitrous oxide/acetylene
reducing
10 mA
321 V
0.2 nm
0.068 mg/L
0.1 - 15 mg/L
(3 5 7 . 8 7 n m , u p p e r ; 4 2 5 .4 4 n m lo w e r lin e )
1.2
Abso rban ce
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.8
0.4
Standard Solutions
10
15
C o n c e n tr a tio n (m g / L)
20
Interferences
Ab so r pt io n Sen s itivity vs . F ue l flo w Rat e
61
0 .4
A b so rb a nc e
0 .3
0 .2
0 .1
0
2
Fu e l F lo w Ra te (L /m in )
27
Co
Analytical lines:
Cobalt
240.7 nm
346.6 nm
air/acetylene
oxidizing
7 mA
425 V
0.2 nm
( 2 4 0 .7 3 n m u p p e r ; 3 4 6 .5 8 n m l o w e r lin e )
1
0.8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.04 mg/L
0.03 - 8 mg/L
Standard Solutions
0.6
0.4
0.2
0
0
2
4
6
C o n c e n tr a tio n ( m g / L)
A b s o rb a n c e
Interferences
0.4
0.3
0.2
0.5
1.5
Fu e l Flo w R a te ( L/ m in )
A b s o r p tio n S e n s i tiv i ty v s . La m p C u r r e n t
A b s o rb a n c e
0.5
0.4
0.3
0.2
0
62
10
15
La m p C u r r e n t ( m A )
20
29
Cu
Analytical lines:
Copper
324.7 nm
327.4 nm
Absorbance Sensitivity
1
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.03 mg/L
0.02 - 6 mg/L
0.8
0.6
0.4
0.2
0
0
2
4
Concentration (mg/L)
Standard Solutions
Interferences
No major interferences.
However, when Zn is present at high
concentrations, the Cu signal can be decreased.
This interference can be alleviated by using a lean
air/acetylene flame or a nitrous oxide/acetylene
flame.
63
0.6
Absorption
0.4
0.2
0
4
8
10
HCL current (mA)
12
66
Dy
Analytical lines:
Dysprosium
421.2 nm
404.6 nm
416.8 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
In the presence of H2SO4 and H3PO4, the Dy signal
is enhanced.
In the presence of Al, Si, HF, Dy signal is
decreased (especially when Na is also present), so
matrix matching is advised for such matrices.
Dy is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
64
0
0
50
100
Concentration (mg/L)
150
68
Er
Analytical lines:
Erbium
400.8 nm
389.3 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Standard Solutions
Dissolve 1.143 g of erbium oxide (Er2O3) in a
minimum volume of HCl. Dilute to 1 L with 1%
v/v HCl to give a 1000 ppm Er solution.
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Si, and HF can all decrease
the Er signal when Na is present, so matrix
matching is advised for such matrices.
Er is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
65
0
0
50
100
150
Concentration (mg/L)
200
63
Eu
Analytical lines:
Europium
459.4 nm
333.4 nm
321.3 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Si, and HF can all decrease
the Eu signal when Na is present, so matrix
matching is advised for such matrices.
Eu is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
66
0
0
25
50
75
Concentration (mg/L)
100
64
Gd
Analytical lines:
Gadolinium
368.4 nm
405.8 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Fe, Si, fluorides, and
phosphates can all significantly decrease the Gd
signal (even at moderate levels), so matrix
matching is advised for such matrices.
Gd is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
67
0
0
2000
4000
Concentration (mg/L)
6000
31
Ga
Analytical lines:
Gallium
294.4 nm
287.4 nm
403.3 nm
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
Absorption Sensitivity
(294.4nm upper; 287.4nm middle,
403.3 nm lower line)
Standard Solutions
0.6
0.4
0.2
0
50
100
150
C oncentration (mg/L)
200
Interferences
Absorption Sensitivity vs. Fuel flow Rate
Absorbance
0.3
0.2
0.1
0
0.5
68
1.5
2
2.5
Fuel Flow Rate (L/min)
32
Ge
Analytical lines:
Germanium
265.2 nm
259.3 nm
303.9 nm
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
5 mA
PMT Voltage:
350 V
Slit Width:
0.6 nm
Characteristic
Concentration:
1.26 mg/L
Working Range:
2-300 mg/L
0.6
0.4
0.2
0
Standard Solutions
Interferences
No interferences to report.
69
100
200
Concentration (mg/L)
300
79
Au
Analytical lines:
Gold
242.8 nm
267.6 nm
Flame:
air/acetylene
Flame Stoichiometry:
oxidizing
HCL Current:
10 mA
PMT Voltage:
323 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.085 mg/L
Working Range:
0.1 - 16 mg/L
Absorption Sensitivity
(242.80nm upper; 267.59nm lower line)
0.6
0.4
0.2
0
0
Standard Solutions
10
15
Concentration (mg/L)
20
Absorbance
0.25
0.2
0.15
0.1
Interferences
0.5
0.3
0.2
0.1
0
0
70
2.5
Absorbance
1
1.5
2
Fuel Flow Rate (L/min)
10
15
20
Lamp Current (mA)
25
72
Hf
Analytical lines:
Hafnium
286.6 nm
307.3 nm
377.8 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
At higher hafnium concentrations, oxides can
form. This can create calibration curve rollover.
To eliminate this effect, add 0.2% Al and 1% HF
to all solutions.
In general, the Hf signal is decreased by alkali and
alkaline earth metals, and H2SO4. Many transition
metals also cause interferences. The use of an
oxidizing nitrous oxide/acetylene flame can often
reduce these interferences.
71
0
0
500
1000
1500
2000
Concentration (mg/L)
2500
67
Ho
Analytical lines:
Holmium
410.4 nm
412.7 nm
425.4 nm
Standard Solutions
Dissolve 1.146 g of holmium oxide (Ho2O3) in a
minimum volume of HCl. Dilute to 1 L with 1%
v/v HCl to give a 1000 ppm Ho solution.
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Si, and HF can all decrease
the Ho signal, so matrix matching is advised for
such matrices.
72
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
0
0
50
100
150
Concentration (mg/L)
200
49
In
Analytical lines:
Indium
303.94 nm
325.61 nm
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
Standard Solutions
10
20
30
C o n c e n tr a tio n ( m g / L)
40
Absorbance
Interferences
0.4
0.3
0.2
0.5
73
1
1.5
Fuel Flow Rate (L/min)
77
Ir
Analytical lines:
Iridium
208.9 nm
264.0 nm
Absorption Sensitivity
1
0.8
0.8 mg/L
4-150 mg/L
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
Standard Solutions
Dissolve 2.294 g of (NH4)2IrCl6 (ammonium
hexachloroiridate) in a minimum volume of 1%
v/v HCl. Dilute to 1 L with 1% HCl to give a 1000
ppm Ir solution.
Interferences
There are many interferences to note. Some
metals (Al, Cu, K, La, Na, Pb, and Pt) increase the
signal, and other metals (Fe, Ni, Pd, Sn, and Ti)
decrease the signal.
The degree of interference depends largely on the
concentration ratio of Ir to interfering species.
Therefore, matrix matching of all solutions is
important.
Many interferences are eliminated in the nitrous
oxide/acetylene flame, but sensitivity is generally
decreased by 50%.
74
0.6
0.4
0.2
0
0
50
100
150
C oncentration (mg/L)
200
26
Fe
Analytical lines:
Iron
248.3 nm
372.0 nm
air/acetylene
oxidizing
5 mA
511 V
0.2 nm
( D o u b le t 2 4 8 .3 3 n m , u p p e r lin e ;
3 7 1 .9 9 n m , lo w e r lin e )
1
0 .8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.045 mg/L
0.03 - 8 mg/L
Standard Solutions
0 .6
0 .4
0 .2
0
0
3
6
9
C o n c e n tra tio n ( m g / L)
12
Interferences
A b s o r p tio n S e n s itiv ity v s . Fu e l f lo w
R a te
0 .5
A b s o rb a n c e
0 .4
0 .3
0 .2
0 .1
0.5
1
1.5
Fu e l Flo w R a te ( L/ m in )
A b s o rb a n c e
0.5
0.4
0.3
0.2
0
75
10
15
20
La m p C u r re n t (m A )
25
57
La
Analytical lines:
Lanthanum
550.1 nm
418.7 nm
403.7 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
Interferences
Ionization interference occurs in the nitrous
oxide/acetylene flame. Add an ionization
suppressant (K, Cs, or another alkali) to reduce
this effect.
Al, Fe, Si, fluorides, and phosphates all depress
the La signal.
Matrix matching of all solutions is recommended.
76
5000
10000
C oncentration (mg/L)
15000
82
Pb
Analytical lines:
Lead
217.0 nm
283.3 nm
261.4 nm
air/acetylene
oxidizing
5 mA
497 V
0.2 nm
0.08 mg/L
0.1-15 mg/L
Standard Solutions
Absorption Sensitivity
(upper line 217 nm, lower line 283.30nm)
Absorbance
0.8
OR
Dissolve 1.000 g of lead metal wire in
1:1 HNO3. Dilute to 1 L with deionized
water to give a 1000 ppm lead solution.
0.6
0.4
0.2
0
0
5
10
Concentration (mg/L)
15
77
0.15
0.1
0.05
0
1
1.5
2
Fuel Flow Rate (L/min)
2.5
Interferences
Absorbance
0.15
0.1
0.05
0
0
10
15
Lamp Current (mA)
20
Li
Analytical lines:
Lithium
670.8 nm
610.4 nm
Absorption Sensitivity
air/acetylene
oxidizing
5 mA
283 V
0.2 nm
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.01 mg/L
0.01 - 2.5 mg/L
0.6
0.4
0.2
0
Standard Solutions
0.5
1
1.5
2
2.5
Concentration (mg/L)
Absorbance
Interferences
0.3
0.2
0.1
0.5
1
1.5
2
Fuel Flow Rate (L/min)
2.5
Absorbance
0.4
0.3
0.2
0.1
0
78
10
15
Lamp Current (mA)
20
71
Lu
Analytical lines:
Lutetium
336.0 nm
356.8 nm
Standard Solutions
Dissolve 1.137 g of lutetium oxide (Lu2O3) in a
minimum volume of 1:1 HNO3. Dilute to 100 mL
to give a 10,000 ppm Lu solution.
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Fe, Si, fluorides, and
phosphates can all decrease the Lu signal, so
matrix matching is advised for such matrices.
Lu is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
79
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
0
0
500
1000
Concentration (mg/L)
1500
12
Mg
Analytical lines:
Magnesium
285.2 nm
202.6 nm
( 2 8 5 . 2 1 n m , u p p e r ; 2 0 2 .5 8 n m , lo w e r )
1
0.8
0.0024 mg/L
0.002 - 0.5 mg/L
A bso rban ce
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
Standard Solutions
0 .2
0.4
0 .6
C o n c e n tr a tio n ( m g / L)
0 .8
A b s o r p tio n S e n s itiv i ty v s . Fu e l f lo w R a te
0.3
80
A b s o rb a n c e
Interferences
0.2
0.1
0
0.5
1
1.5
2
F u e l F lo w R a te ( L/ m in )
2 .5
25
Mn
Analytical lines:
Manganese
279.5 nm
279.8 nm
280.1 nm
Absorption Sensitivity
(Doublet 279.8/280.1 nm, upper line;
279.48nm, lower line)
1.2
0.9
Absorbance
Parameter
0.6
0.3
0
0
4
6
Concentration (mg/L)
Standard Solutions
Interferences
0.4
0.3
0.2
0.1
0.5
1.5
2
2.5
Fuel Flow Rate (L/min)
Absorbance
0.5
0.4
0.3
0.2
0
81
10
15
20
Lamp Current (mA)
25
80
Hg
Analytical lines:
Mercury
253.7 nm
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
Absorption Sensitivity
3 mg/L
2 - 500 mg/L
0.6
0.4
0.2
0
0
Standard Solutions
Interferences
Mercury (I) is more sensitive than mercury (II).
Mercury (II) can be reduced to mercury (I) by
ascorbic acid, stannous chloride, and other
reductants, thus falsely increasing the signal.
High cobalt levels will interfere with the 253.7nm
mercury line.
It is advisable to check and correct for nonspecific absorption.
82
750
150
300
450
600
C oncentration (mg/L)
0.15
0.1
0.05
0
0.5
1
1.5
Fuel Flow Rate (L/min)
42
Mo
Analytical lines:
Molybdenum
313.3 nm
320.9 nm
Parameter
nitrous oxide/acetylene
reducing
10 mA
445 V
0.6 nm
0.64 mg/L
0.2 - 125 mg/L
1
0 .8
Abso rban ce
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0 .6
0 .4
0 .2
0
Standard Solutions
Interferences
The signal can be depressed in the presence of
Ca, Fe, Sr, and sulfate. The effects can be
reduced by the addition of 1-3 g/L aluminum.
When Mo is determined in an air/acetylene flame,
the sensitivities are 3-10 times poorer than those
obtained with a nitrous oxide/acetylene flame. As
well, there are far fewer interferences in the
nitrous oxide flame.
83
25
50
75
1 00
C o n c e n tr a tio n ( m g / L)
125
60
Nd
Analytical lines:
Neodymium
492.5 nm
463.4 nm
486.7 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Fe, Si, and HF can all decrease
the Nd signal, so matrix matching is advised for
such matrices.
Nd is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
The presence of Pr (another lanthanide element)
can cause interference with the overlap of its
492.5 nm line.
Neodymium, Nd
84
0
0
1000
2000
3000
Concentration (mg/L)
4000
28
Ni
Analytical lines:
Nickel
232.0 nm
341.5 nm
Parameter
air/acetylene
oxidizing
7 mA
449 V
0.2 nm
0 .8
A bso rban ce
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.042 mg/L
0.07 - 10 mg/L
0 .6
0 .4
0 .2
0
0
Standard Solutions
12
C o n c e n tra tio n ( m g / L)
Interferences
At the 232.0 nm absorption wavelength, it is
common for non-specific absorption to occur. This
is especially a problem when non-dissolved
species are present. To alleviate this interference,
a background correction method must be
employed. Non-specific absorption is not a
problem at the 341.5 nm wavelength.
Fe or Cr present at high levels can increase the Ni
signal.
The use of a nitrous oxide/acetylene flame will
remove most interferences, but signal sensitivity
will be sacrificed.
Absorbance
0.5
0.4
0.3
0.2
0.5
1
1.5
Fuel Flow Rate (L/min)
A b s o rb a n c e
0.5
0.4
0.3
0.2
0
85
10
15
La m p C u r re n t (m A )
20
41
Nb
Analytical lines:
Niobium
334.9 nm
408.0 nm
374.0 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
23 mg/L
20-6000 mg/L
Standard Solutions
0.6
0.4
0.2
Interferences
At higher concentrations of niobium, oxide
formation can occur. This results in calibration
curve roll-over. This effect can be countered by
adding 0.2% Al and 1% HF.
Ionization interference is common, but can be
reduced by the addition of an ionization
suppressant (an alkali salt such as K or Cs).
86
0
0
2000
4000
Concentration (mg/L)
6000
76
Os
Analytical lines:
Osmium
290.9 nm
305.9 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
Standard Solutions
Obtain a certified stock standard solution of 1000
ppm osmium from a chemicals supplier.
OR
Interferences
The nitrous oxide/acetylene flame is about 5 times
more sensitive than the air/acetylene flame.
87
0.6
0.4
0.2
0
0
100
200
C oncentration (mg/L)
300
46
Pd
Analytical lines:
Palladium
244.79 nm
247.64 nm
Absorption Sensitivity
(244.79 upper; 247.64 lower line)
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
0
5
10
Concentration (mg/L)
Standard Solutions
15
0.45
0.4
0.35
0.3
0.5
1
1.5
Fuel Flow Rate (L/min)
Interferences
A b s o r p tio n S e n s itiv ity v s . La m p C u r r e n t
A bso rb a nce
0 .5
0 .4
0 .3
0 .2
0
88
5
10
15
La m p C u r r e n t ( m A )
20
15
P
Analytical lines:
Phosphorous
213.6 nm
214.9 nm
Standard Solutions
Dissolve 37.138 g of dry NH4H2PO4 (ammonium
dihydrogen orthophosphate) in H2O. Dilute to 1 L
to give a 10,000 ppm P solution.
Interferences
There is little mention of interferences in the
literature.
However, matrix matching of solutions is advised
as a precaution.
89
Absorbance
1.2
0.8
0.4
0
0
10000
20000
30000
Concentration (mg/L)
40000
78
Pt
Analytical lines:
Platinum
265.9 nm
299.8 nm
Parameter
air/acetylene
oxidizing (lean, blue)
10 mA
384 V
0.2 nm
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6 mg/L
0.4 - 120 mg/L
0.6
0.4
0.2
0
0
50
100
C oncentration (mg/L)
150
Standard Solutions
Dissolve 2.275 g of ammonium chloroplatinate
((NH4)2PtCl6) in deionized water. Dilute to 1 L with
more water to give a 1000 ppm platinum solution.
Absorbance
Interferences
1
1.5
Fuel Flow Rate (L/min)
Absorbance
0.4
0.35
0.3
5
90
10
15
20
Lamp Current (mA)
25
19
K
Analytical lines:
Potassium
766.5 nm
769.9 nm
404.4 nm
( u p p e r lin e 7 6 6 . 4 9 n m , lo w e r lin e 7 69 .9 0
air/acetylene
oxidizing
5 mA
416 V
0.2 nm
nm )
1
0 .8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
A b s o r p ti o n S e n s iti v ity
0.0066 mg/L
0.02 - 1.5 mg/L
0 .6
0 .4
0 .2
0
0
Standard Solutions
Dissolve 1.907 g of dry potassium chloride (KCl)
in H2O. Dilute to 1 L to give a 1000 ppm K
solution.
0.5
1
1 .5
C o n c e n tr a tio n ( m g / L)
91
Absorbance
Interferences
0.3
0.2
0.1
0.9
1.2
1.5
Fuel Flow Rate (L/min)
1.8
59
Pr
Analytical lines:
Praseodymium
495.1 nm
513.3 nm
491.4 nm
Standard Solutions
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
Absorption Sensitivity
1
0.6
0.4
0.2
0
0
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
An excess of Si decreases the Pr signal.
Pr is one of the lanthanide elements, which are
known to display complex mutual interference
patterns.
92
1000
2000
3000
Concentration (mg/L)
4000
75
Re
Analytical lines:
Rhenium
346.0 nm
346.5 nm
345.2 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
14.7 mg/L
10-2000 mg/L
Standard Solutions
0.6
0.4
0.2
0
Interferences
The signal is depressed by the presence of Al, Ba,
Ca, Fe, K, Mg, Mn, Mo, Pb, and most transition
group metals. The use of a more oxidizing flame
can often alleviate these interferences somewhat.
Matrix matching of all solutions is recommended.
93
500
1000 1500
2000
C oncentration (mg/L)
2500
45
Rh
Analytical lines:
Rhodium
343.49 nm
339.69 nm
Parameter
air/acetylene
oxidizing (lean, blue)
8 mA
328 V
0.2 nm
0.11 mg/L
0.03 - 20 mg/L
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 4.120 g of ammonium hexachlororhodate
trihydrate ((NH4)3RhCl6 3H2O) in a minimum volume
of 10% (v/v) HCl. Dilute to 1 L with 10% (v/v) HCl to
give a 1000 ppm rhodium solution.
Interferences
10
15
20
C oncentration (mg/L)
25
0.45
0.4
0.5
94
1
1.5
Fuel Flow Rate (L/min)
37
Rb
Analytical lines:
Rubidium
780.0 nm
794.8 nm
Absorption Sensitivity
(780.0nm, upper line; 794.8 nm lower line)
air/acetylene
oxidizing (lean, blue)
8 mA
467 V
0.6 nm
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.015 mg/L
0.05 - 5 mg/L
0.6
0.4
0.2
0
Standard Solutions
1
2
3
C oncentration (mg/L)
Interferences
Rubidium is relatively easily ionized, but this can
be controlled by adding 0.1% alkali salt.
95
Absorbance
0.65
0.6
0.55
0.5
0.5
1
1.5
Fuel Flow Rate (L/min)
44
Ru
Analytical lines:
Ruthenium
349.9 nm
392.6 nm
Absorption Sensitivity
1
0.8
0.57 mg/L
1-150 mg/L
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
Standard Solutions
0.6
0.4
0.2
Interferences
Ruthenium sensitivity is decreased by
molybdenum, but increased by platinum, rhodium,
lanthanum, and HCl. In general, however, most
elements interfere unpredictably.
The nitrous oxide/acetylene flame removes most
interferences, and the addition of 0.5% lanthanum
improves sensitivity.
96
0
0
50
100
Concentration (mg/L)
150
62
Sm
Analytical lines:
Samarium
429.7 nm
476.0 nm
Standard Solutions
Dissolve 1.159 g of samarium oxide (Sm2O3) in a
minimum volume of HCl. Dilute to 100 mL with
10% v/v HCl to give a 10,000 ppm Sm solution.
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
Sm is one of the Lanthanide elements, which are
known to display complex mutual interference
patterns.
97
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
0
0
500
1000
1500
C oncentration (mg/L)
2000
21
Sc
Analytical lines:
Scandium
391.2 nm
327.4 nm
326.9 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.40 mg/L
0.5-80 mg/L
0.6
0.4
0.2
Standard Solutions
Interferences
Ionization in the nitrous oxide/acetylene flame can
be controlled by the addition of an alkali salt to all
solutions.
Significant signal depression occurs when fluoride,
sulfide, or sulfate anions are present. Many
cations are known to increase or decrease the
signal when present at high levels. To
compensate for these interferences, matrix
matching of all solutions is advised.
98
20
40
60
80
Concentration (mg/L)
100
34
Se
Analytical lines:
Selenium
196.0 nm
204.0 nm
Parameter
air/acetylene
oxidizing
10 mA
458 V
0.2 nm
0.8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.17 mg/L
0.8 - 40 mg/L
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 1.000 g of selenium metal pellets in a
minimum volume of 1:1 HNO3. Heat gently to
help the dissolution. Cool the solution and dilute
to 1 L to give a 1000 ppm Se solution.
Interferences
The analytical line of 196.03 nm is at the start of
UV range, where the flame gases can absorb and
scatter part of the radiation. Therefore, use of
background correction is recommended. The
absorption by the flame is reduced when a nitrous
oxide/acetylene flame is used, but sensitivity is
also reduced.
99
40
10
20
30
C o n c e n tr a tio n ( m g / L)
0.2
0.15
0.1
1
1.2
1.4
1.6
Fuel Flow Rate (L/min)
1.8
14
Si
Analytical lines:
Silicon
251.6 nm
221.1 nm
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
Absorption Sensitivity
(upper line 251.6nm, lower line 221.1nm)
0.6
0.4
0.2
0
0
100
200
Concentration (mg/L)
Standard Solutions
0.3
Absorbance
OR
Fuse 2.139 g of silicon dioxide with 20 g of Na2CO3
(sodium carbonate) in a platinum crucible. Cool the
melt and dissolve it with H2O. Dilute to 1 L to give a
1000 ppm Si solution.
300
0.2
0.1
0
5
Interferences
100
10
47
Ag
Analytical lines:
Silver
328.1 nm
338.3 nm
air/acetylene
oxidizing
5 mA
270 V
0.2 nm
0.016 mg/L
0.02 - 4 mg/L
be poor due to the low
in solutions.
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
Ag sensitivity can
dissociation of Ag
0.8
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 1.000 g of Ag metal strip in 20 mL of 1:1
nitric acid. Dilute to 1 L to give a 1000 ppm Ag
solution.
Interferences
2
3
Concentration (mg/L)
0.2
0.1
0
0
5
10
15
Lamp Current (mA)
20
Absorbance
0.15
0.1
0.05
0
0.5
101
1
1.5
2
Fuel Flow Rate (L/min)
2.5
11
Na
Analytical lines:
Sodium
589.0
589.6
330.2
330.3
nm
nm
nm
nm
Interferences
Na is easily ionized in the flame, so an ionization
suppressant must be added to eliminate this
interference (Cs, K, or some other alkali salt).
High levels of mineral acids can decrease the
sensitivity.
A b s o rb a n c e
0.8
0.6
0.4
0.2
0
0
0 .2
0 .4
0.6
0.8
C o n c e n tr a tio n ( m g / L)
0.35
0.3
0.25
0.5
1
1.5
2
Fuel Flow Rate (L/min)
102
0.4
Absorbance
Standard Solutions
( 5 8 9 .0 0 n m u p p e r ; 5 8 9 . 5 9 n m lo w e r li n e )
1
Absorbance
Flame:
air/acetylene
Flame Stoichiometry:
oxidizing
HCL Current:
10 mA
PMT Voltage:
244 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.0032 mg/L
Working Range:
0.002 - 1 mg/L
the 589.0/589.6 nm doublet can be used to
increase the S/N ratio, but calibration
linearity is sacrificed.
the 589.0 nm line is more sensitive than the
589.6 nm line.
the 330.2/330.3 nm lines are also a
doublet.
2.5
38
Sr
Analytical lines:
Strontium
460.7 nm
407.8 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
air/acetylene
Flame Stoichiometry:
oxidizing
HCL Current:
10 mA
PMT Voltage:
452 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.018 mg/L
Working Range:
0.007-3 mg/L
Data collected with potassium chloride added
(final concentration of 2g/L) to suppress
ionization.
0.6
0.4
0.2
0
Standard Solutions
Interferences
The air/acetylene flame is not as sensitive as the
nitrous oxide/acetylene flame, and more chemical
interferences are known in the air flame.
In the air flame, Al, Si, Ti, Zr, phosphate, and
sulfate all decrease the signal. To alleviate these
effects, add 1% La or the complexing agent EDTA
to all solutions.
Strontium undergoes ionization in both the nitrous
oxide and air flames. To reduce this interference,
add an ionization suppressant (eg. K, Cs, or other
alkali salt) to all solutions.
103
1
2
3
C oncentration (mg/L)
73
Ta
Analytical lines:
Tantalum
271.5 nm
277.6 nm
275.8 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
12 mg/L
20-3000 mg/L
Standard Solutions
Dissolve 1.000 g of tantalum metal strip in a
Teflon vessel by carefully adding 5 mL of conc. HF
dropwise.
Dissolve the remaining metal by slowly adding
conc. HNO3 dropwise. Dilute to 200 mL to give a
5,000 ppm Ta solution.
Store all solutions in polyethylene bottles.
WARNING: HF (hydrofluoric acid) is a toxic acid
and should be handled with extreme care.
Personal protective equipment is essential.
Interferences
Oxide formation in the flame is common and
results in calibration curve roll-over at higher
concentrations. To alleviate this problem and
improve calibration curve linearity, add 0.2% Al
and 1% HF to all solutions.
The signal is interfered with in varying ways when
alkali metals, sulfates, phosphates, and mineral
acids are present in various combinations. Matrix
matching of all solutions is recommended.
104
0.6
0.4
0.2
0
0
1000
2000
C oncentration (mg/L)
3000
52
Te
Analytical lines:
Tellurium
214.27 nm
225.9 nm
1
0.8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 1.000 g of tellurium metal granules in a
minimum volume of 1:1:1 HNO3:HCl:water. Heat to
aid dissolution, but do not boil. Dilute to 1 L with 1%
(v/v) HCl to give a 1000 ppm tellurium solution.
20
40
C o n c e n tr a tio n ( m g / L)
60
105
Absorbance
Interferences
0.4
0.35
0.3
0.5
1
1.5
Fuel Flow Rate (L/min)
81
Tl
Analytical lines:
Thallium
276.8 nm
258.0 nm
Parameter
( 2 7 6 . 8 n m u p p e r ; 3 7 7 . 6 n m lo w e r lin e )
air/acetylene
oxidizing
5 mA
484 V
0.2 nm
0.13 mg/L
0.1-20 mg/L
1
0 .8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
Standard Solutions
0 .6
0 .4
0 .2
0
0
Interferences
Many significant interferences exist.
Matrix matching of standards is advised for
samples with high levels of interfering elements.
Ionization occurs in the nitrous oxide/acetylene
flame.
106
10
15
20
25
C o n c e n tr a tio n ( m g / L)
30
0.3
0.25
0.2
0.5
1.5
2
2.5
Fuel Flow Rate (L/min)
50
Sn
Analytical lines:
Tin
224.6 nm
286.3 nm
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
nitrous oxide/acetylene
reducing
10 mA
536 V
0.2 nm
( 2 2 4 . 6 1 n m , u p p e r ; 2 8 6 .3 3 n m lo w e r lin e )
0.8
A bso rban ce
Parameter
1.5 mg/L
1 - 200 mg/L
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 1.000 g of tin metal granules in 100 mL
of conc. HCl. If necessary, heat gently to aid
dissolution. Cool and dilute to 1 L to give a 1000
ppm Sn solution.
Interferences
1 00
150
2 00
C o n c e n tr a ti o n ( m g / L)
250
A b s o r p tio n S e n s itiv i ty v s . Fu e l f lo w R a te
0.4
0.3
A b s o rb a n c e
50
0.2
0.1
0
A b s o rb a n c e
0.4
0.3
0.2
0.1
0
0
107
5
6
7
Fu e l Flo w R a te ( L/ m in )
0 .2
0.4
0 .6
S lit W i d th ( n m )
0.8
22
Ti
Analytical lines:
Titanium
364.3 nm
365.4 nm
399.0 nm
Standard Solutions
Dissolve 1.000 g of titanium metal in 10 mL of HF
and 10 mL H2O by slowly adding (dropwise) 20
mL of HNO3. After each addition, allow the
reaction to diminish before the next addition.
Dilute to 1 L to give a 1000 ppm Ti solution.
OR
Dissolve 1.000 g of titanium metal in 100 mL of
1:1 HCl. Cool and dilute to 1 L to give a 1000
ppm Ti solution.
Solutions made from this bulk standard should all
be made with 10% v/v HCl.
Interferences
Ionization interferences can be relieved by adding
an alkali salt as an ionization suppressant.
Most metals, as well as fluoride, chloride, and
ammonium, will increase the titanium signal.
108
Absorption Sensitivity
1
0.8
Absorbance
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
10 mA
PMT Voltage:
400 V
Slit Width:
0.2 nm
Characteristic conc.:
1.7 mg/L
Working Range:
1-300 mg/L
Data collected with potassium chloride added
(final concentration of 2g/L) to suppress
ionization.
0.6
0.4
0.2
0
0
100
200
300
C oncentration (mg/L)
400
74
W
Analytical lines:
Tungsten
255.1 nm
268.1 nm
400.9 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
13 mA
PMT Voltage:
450 V
Slit Width:
0.2 nm
Characteristic
Concentration:
7.9 mg/L
Working Range:
10-1500 mg/L
0.6
0.4
0.2
0
OR
Interferences
High levels of some mineral acids combined with
the presence of some elements (eg. Co, Cu, Fe,
and K) can cause interference. Matrix matching of
all solutions is recommended.
109
500
1000
1500
C oncentration (mg/L)
2000
92
U
Analytical lines:
Uranium
358.5 nm
351.5 nm
348.9 nm
Standard Solutions
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Interferences
Ionization interferences can be countered by
adding 0.1 % of an ionization suppressant (eg.
alkali salt).
In a complex matrix with excesses of other
elements (eg. Al, Co, Fe, Ni, Pb), the U signal can
be increased. Therefore, matrix matching is
required for such matrices.
110
0
0
5000
25000
23
V
Analytical lines:
Vanadium
318.4 nm
306.6 nm
370.4 nm
Standard Solutions
Dissolve 1.000 g of Vanadium metal granules in a
minimum volume of HNO3. Dilute to 1 L with 1%
v/v HNO3 to give a 1000 ppm V solution.
Interferences
Ionization interferences can be controlled by the
addition of an ionization suppressant (alkali salt)
to all solutions.
High levels of Al, Fe, Ti, ammonium and H2SO4 can
increase the signal. Some interference can be
reduced by adding an excess of Al. Matrix
matching of all solutions is advised.
111
Absorption Sensitivity
(318.4 nm upper; 306.6 nm lower line)
1
0.8
Absorbance
Flame:
nitrous oxide/acetylene
Flame Stoichiometry:
reducing
HCL Current:
10 mA
PMT Voltage:
308 V
Slit Width:
0.2 nm
Characteristic
Concentration:
0.44 mg/L
Working Range:
0.5-100 mg/L
Data collected with potassium chloride added
(final concentration of 2g/L) to suppress
ionization.
0.6
0.4
0.2
0
0
25
50
75
C oncentration (mg/L)
100
70
Yb
Analytical lines:
Ytterbium
398.8 nm
364.4 nm
246.5 nm
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
nitrous oxide/acetylene
reducing
(rich, red cone 1-2 cm)
5 mA
350 V
0.2 nm
0.07 mg/L
0.04-15 mg/L
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
0.6
0.4
0.2
0
0
Standard Solutions
Interferences
Ionization in the nitrous oxide flame can be
reduced by the addition of an ionization
suppressant (eg. 0.1% of an alkali salt).
The presence of Al, Fe, Si, and HF can all decrease
the Yb signal, so matrix matching is advised for
such matrices.
Yb is one of the Lanthanide elements, which are
known to display complex mutual interference
patterns.
112
15
0.6
Absorption
5
10
Concentration (mg/L)
0.4
0.2
0
4
8
10
HCL current (mA)
12
39
Y
Analytical lines:
Yttrium
410.2 nm
407.7 nm
414.3 nm
Parameter
nitrous oxide/acetylene
reducing
10 mA
400 V
0.2 nm
5.5 mg/L
2-500 mg/L
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.6
0.4
0.2
0
0
Standard Solutions
Dissolve 1.270 g of Y2O3 (yttrium oxide) in
minimum volume of HCl. Dilute to 1 L with 1%
v/v HCl to give a 1000 ppm Y solution.
Interferences
To reduce ionization interferences, add an
ionization suppressant (alkali salt) to all solutions.
The signal is depressed by the presence of Al, K
and some mineral acids. Matrix matching of all
solutions is recommended.
113
500
1000
1500
C oncentration (mg/L)
2000
30
Zn
Analytical lines:
Zinc
213.9 nm
307.6 nm
air/acetylene
oxidizing
5 mA
422 V
0.2 nm
1
0.8
A b s o rb a n c e
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic
Concentration:
Working Range:
0.008 mg/L
0.01 - 2 mg/L
0.6
0.4
0.2
0
Standard Solutions
1 .6
A b s o rb a n c e
Interferences
0.4
0 .8
1 .2
C o n c e n tr a tio n ( m g / L)
0.2
0.1
0
0.8
1
1 .2
1 .4
Fu e l Flo w R a te ( L/ m in )
1 .6
A b s o rb a n c e
0.3
0.2
0.1
0
0
114
10
15
La m p C u r r e n t ( m A )
20
40
Zr
Analytical lines:
Zirconium
360.1 nm
303.3 nm
352.0 nm
Absorption Sensitivity
1
0.8
Absorbance
Flame:
Flame Stoichiometry:
HCL Current:
PMT Voltage:
Slit Width:
Characteristic conc.:
Working Range:
0.6
0.4
0.2
Standard Solutions
Interferences
Reduce ionization interferences by adding an
ionization suppressant (an alkali salt) to all
solutions.
Oxide formation in the flame is common and
results in calibration curve roll-over at higher
concentrations. To alleviate this problem and
improve calibration curve linearity, add 0.2% Al
and 1% HF to all solutions. As well, the
interferences can also be reduced by using a more
oxidizing flame. However, sensitivity will then be
compromised.
115
500
1000
1500
Concentration (mg/L)
2000
116
117
118
Chapter 7:
Marine
Water
Biological
Food
Agricultural
Petrochemical
Miscellaneous
119
Application #AA-17
The Determination of Metal Cations in Sea Water by Flame AAS
Classification: Marine
Sample matrix: Sea water
Introduction
Reagents
Needed
Standard
Solutions
Sample
Preparation
Guideline
Instrument
Parameters
120
Application #AA-18
The Determination of Common Metals in Natural Water
by Flame AAS
Classification: Water
Sample matrix: Natural water
Introduction
Reagents
Needed
Standard
Solutions
Sample
Preparation
Guideline
Instrument
Parameters
121
Application #AA-19
The Determination of Copper and Zinc in Blood Serum and Plasma
by Flame AAS
Classification: Biological
Sample matrix: Blood serum, blood plasma
Introduction
Reagents
Needed
Standard
Solutions
Sample
Preparation
Guideline
Instrument
Parameters
122
Application #AA-20
The Determination of Iron and Nickel in Vegetable Oil
by Flame AAS
Classification: Food
Sample matrix: Vegetable oil
Introduction
Reagents
Needed
Standard
Solutions
Sample
Preparation
Guideline
Instrument
Parameters
123
Application #AA-16
The Determination of Potassium in Fertilizer by Flame AAS
Classification: Agricultural
Sample matrix: Fertilizer
Introduction
Elements of K
interest:
Reagents
Needed
Distilled water.
Standard
Solutions
Sample
Preparation
Guideline
Instrument
Parameters
Weigh out 2.5 g of the dried fertilizer sample, and transfer it to the
250 mL volumetric flask. Add 200 mL of distilled water to the
flask, and bring to a boil. Boil for at least half an hour. Cool the
mixture, and then make up to the mark with deionized water.
Remove the particulate matter by allowing the mixture to stand
overnight or by filtration. Transfer a 25mL aliquot of the liquid to a
100 mL volumetric flask, and make up to the mark. For very
concentrated samples, greater dilutions may be necessary.
124
Application #AA-16A
The Determination of various Metals in Plant by Flame AAS
Classification: Agricultural
Sample matrix: Plant Material
Introduction
Reagents
Needed
70% nitric
1.000 g Al
2.497 g of
1.000 g of
1.907 g of
1.000 g of
1.000 g of
2.542 g of
1.000 g of
Standard
Solutions
Sample
Preparation
Carry out the following acid digestion in a fume hood, and behind a
blast shield. Wear appropriate personal protective equipment,
including gloves and protective eyewear.
Accurately weigh about 0.25 g of the plant tissue in a 19 mm (27
mL) test tube. Put the tube in an aluminum test tube block on a
hot plate. Add 1 mL 70% HNO3 and 4mL of HClO4. Heat for 2
hours at 120 C. Ramp the temperature to 180 C over 3 hours to
drive off the nitric acid. The digestion is complete when white
fumes evolve (from the HClO4). It is very dangerous to boil the
HClO4 to dryness.
Quantitatively transfer the test tube contents to a 25 mL volumetric
flask and make up to the mark with water.
Guideline
Instrument
Parameters
125
Calculations
Standard #1
5
10
2
5
5
1
1
0.5
Standard #2
10
20
5
10
10
3
2
1
Standard #3
20
40
10
20
20
5
5
2
126
Application AA-7
Preparation of Petrochemical Samples for Atomic Absorption
Spectrometric Analysis
Introduction:
Atomic absorption spectrometry (AAS) has been used in three major areas for the
elemental analysis of petroleum products: crude oil, fuels and lubricants, and wear metal
analysis [13]. Flame AAS is recommended for the determination of higher concentration
metals and graphite furnace techniques for trace metals.
Sample Preparation:
Three sample preparation methods are currently employed for the determination of metals
in petroleum products:
1. Dilution of the petrochemical sample in an organic solvent such as methyl isobutyl
ketone (MIBK) or a mixed solvent system such as toluene and glacial acetic acid.
2. Ashing of the sample and dissolution with an appropriate acid.
3. Acid digestion.
Dilution
Some petrochemical samples can be directly nebulized into a flame or injected into a
graphite furnace for analysis. However, when the sample is too volatile, too viscous or the
content of the element of interest is too high in the sample, dilution is required.
The dissolution of oil/fuel samples in organic solvents is fast and simple. The degree of
dilution and choice of solvent will depend on the nature of the sample, the expected level of
the element of interest and the background signal appearing during atomization. The
volatility of the solvent plays an important role in the selection of drying conditions and the
use of an autosampler. Where evaporation is a problem, the sample container should be
sealed and the sample dispensed manually and/or a less volatile solvent employed. Mixed
solvent and emulsion systems allow the use of inorganic salts for the preparation of
standards and calibration curves.
Table 7.1 lists a number of organic solvents commonly used in oil/fuel analysis together
with their boiling points which can be used as a guide for setting the drying conditions for
GFAAS. With regular furnace technique (i.e. not a fast drying furnace technique), most
solvents can be dried at a temperature below their boiling point. For a 20 to 40 l sample,
a drying step at temperature setting of 80 to 90% of the boiling point for 30 to 60 seconds
has been found to be adequate in most analyses. The dispensing characteristics of the
solvent should be studied prior to analysis in order to determine the maximum injection
volume. The dispensed volume varies depending on the surface tension of the solvent.
Generally injection volumes between 20 and 40 l are used.
127
Solvent
Boiling point
(C)
Kerosene
175 - 325
Tetralin
207
Shellsol T
186 - 214
DIBK
166
Cyclohexanol
161
3-heptanone
148
m-xylene
139
Iso-amylalcohol
132
MIBK
118
Toluene
112
2-methy, propan-1-ol
108
Isopropanol
83
Carbon tetrachloride*
77
Tetrahydrofuran*
65
Chloroform*
62
* Not suitable for flame AAS.
Specific
gravity
0.78
0.97
0.75
0.81
0.96
0.82
0.80
0.81
0.79
0.86
0.78
0.78
1.59
0.88
1.47
Acid digestion
Acid digestion of oil samples permits the retention of volatile elements that may be lost
during ashing or masked by background interference. During the past years, microwave
oven assisted acid digestion has gained increased popularity. This technique makes acid
digestion much faster and simpler. References on oil sample digestion with acids can be
found in the literature.
128
129
Suggested Setting
0.2 nm
12 mA
251.6 nm
336 V
5 seconds
5.0 L/min.
preset
130
131
References
References
1.
2.
Lvov, B.V., Atomic Absorption Spectrochemical Analysis, Adam Hilger, London, 1970.
3.
Mavrodineanu, R. and Boiteux, H., Flame Spectroscopy, John Wiley and Sons, New
York, 1965.
4.
Price, W.J., Spectrochemical Analysis by Atomic Absorption, Heydon & Son Ltd,
London, 1979.
5.
6.
7.
Welz, B. and Sperling, M., Atomic Absorption Spectrometry, 3rd. edition, Wiley-VCH,
New York, 1999.
8.
9.
Holcombe, J.A. and Harnly, J.M., Anal. Chem., 57, 1983, 1985.
132
References
133