ADJUSTMENTS
ASSEMBLY SCHEMATIC
CPU SCHEMATIC
PRINTER SCHEMATIC & PCBs
BICHANNEL SCHEMATIC & PCBs
KEYBOARD SCHEMATIC & PCBs
(./0)1234)5467)8529:
2/6
$*;"-(#,+(Once all components of the instrument have been assembled, it is necessary to perform
some adjustments for properly functioning.
In the CPU board (Figure 1) there are some potentiometers which will help to calibrate the
tension of control.
Below, the required adjustments for every instrument are described:
!
Mixer: After 60 seconds of mixing, the tension should be adjusted to 1285 Volts.
The motor 1 should be adjusted by the RP2 potentiometer placed on the CPU board and
the tension should be read in the indicated points in the Bichannel Board (Figure 3).
The motor 2 should be adjusted with the P2 potentiometer placed on the Bichannel Board
and measure the tension in the indicated points. (Figure 3).
Printer: The printing intensity should be adjusted by means of the RP1 potentiometer, to
read a 6 Volts tension at the C2 condensator bornes of the Printer Board (Figure 2).
3/6
RP2:
Adjustment
of
the
RP1: Adjustment of
intensity.
printing intensity.
4/6
Figure 1
printing intensity.
Figure 2
5/6
In Figure 3, it is indicated the pins where the tension of the motors must be read, when acting
on RP2 potentiometer of Motor 1 and on P2 for Motor 2.
BOTTOM
Tension of Motor 2
Tension of Motor 1
TOP
Figure 3
P2: Adjustment of
stirring speed of Motor 2
6/6