Charpter1 using This Manual


Introduction
This chapter explains how to use your GE-200 service manual, which provides the
reference information and procedures needed for servicing your GE-200 analyzer. Read
this manual carefully before servicing your analyzer and service your analyzer strictly as
instructed in this manual.
This manual is to be used in conjunction with the GE-200 analyzer operation manual and
does not contain information and procedures already covered in the operation manual.

Who Should Read This Manual


This service manual is written for people who
*

have a thorough understanding of electronic and fluidic principles.

* have a thorough understanding of reagent systems.


*

have a thorough understanding of quality control.

have a thorough understanding of troubleshooting concepts.

have an operator's knowledge of the analyzer.

have the ability to use basic mechanical tools and understand related
terminology.

have the ability to use a digital voltmeter (DVM) and an oscilloscope.

have the ability to read electronic and fluidic schematics and understand
related terminology.

trial version

Charpter2 Hardware
1. CPU board

trial version
Figure 1: Schematic Circuit Diagram of CPU Board
The CPU, program chip and memory chip are the major components on the board.
The CPU carries out the instructions and functions as the core of the board. The program
chip is software chip.
The Clock is a real time clock. System configurations are stored in the EEPROM.

2. interface board
The whole block diagram

trial version

PCB diagram

The important connector

trial version

The below table shows the connection of Interface Bd and Interface2 Bd..
Designator

Connecting to

Function

JA1

Electrode Na

signal collection

JA2

Electrode K

signal collection

JA3

Electrode Cl

signal collection

JA4

Electrode Ca

JA5

Electrode pH

signal collection
pH value signal collection

J3

RS-232

Serial Port

J7

Sensor

Control probe

J9

Pump

Aspirating pump

J10

Std Reagent A
aspirating pump
Probe pump

Aspirate Std. Reagent A

Std Reagent B
aspirating pump

Aspirate Std. Reagent B

J11
J12

Move the probe

trial version
J20

J23

J4
of
Power Bd.
J5
of Power Bd.
J1-1
of Interface2 Bd.
LCD

J24

Key

Keyboard control

J28

J3 of
Printer Retainer Bd
J22 of Interface Bd

Transmit data of result

J21
J22

J1-1
J6-1

Get power supply from Power Bd.


Get analog power supply from Power Bd.
Show if the power supply is on
LCD drive

Show if the power is on

LCD

LCD backlight

3. Power supply

trial version

4. Analog board

trial version

Charpter3 Important Parts


1. electrode
1.1 install electrode
Open front Cover: Open front Cover of instrument. You can see the
K.Na.Cl.etc. electrodes.
Take out the electrodes: Release the electrodes nut(at the right side of
the electrodes) Uplift the bottom of the electrodes slightly, and then you can
take out the electrodes
Installation of electrodes:
Before Installing the electrodes,use fingers to flip the electrodes bottom gently
to eliminate air bubble inside the electrodes(The top of the electrodes is the
side with copper cobbra).First, put the top of the electrodes under the point of
contact, and push in the bottom of electrodes, then the electrodes are fixed.

turn nut

locknut

electrodes

cover

trial version
Figure : How to install electrodes
After the electrodes are fixed, you should screw on the nut at the right side
of the electrodes.
Do the same operation if your instrument type is GE200 F.

1.2 Input the electrodes filler


The electrodes must be input the filler liquid before using it .It has to use the
special injector to input the filler liquid (the length must be less than l0mm), in
case that poke the ion selective theca, the filler liquid must be filled 2/3 ~3/4 of
Electrode port

electrode hull

Ag/AgCl inner
side electric electrode

airproof loop

electrodes
filler liquid
Note
Fill the electrodes
antrum with filler liquid
to about 2/3 ~3/4

Ion selective
biomembrane

Figure How to fill the electrodes filler


the electrodes antrum. Then screw off the inner-side electric electrode,
between the filler liquid and electrodes biomembrane, the existence of air
bubble is prohibited, then wipe the top of the electrodes, and fix them back to
place.

1.3 activate the electrode


Slope value will fall and the stability will be worse after using electrode for
a period of time, now carry out this function, use DEPROTEINIZED
SOLUTION to make electrode active to increase slope value and stability.
Generally we suggest you activate electrodes every week.
Press ACTIVITY key on keyboard or press 4 in MAINTAIN submenu
to active electrode when you need to activate electrode.
Carry out this function, the probe uplifts automatically, the screen will
display:
ACTIVE ELECTRODE
PUT ACTIVE LIQUID
ENTER ASP MENU
10 - RETURN

trial version

Put DEPROTEINIZED SOLUTION follow the screens suggestion, press


ENTER key, system will aspirate serum, it will show ACTIVE ELECTRODE.

ACTIVE ELECTRODES
ACTIVING ELECTRODE
10%
PASSED TIME 06:14
LEFT TIME 54:46
And when the activity of PRESSMENUTO
the electrodes haveQUIT
finished, the display will show :
ACTIVE FINISHED
1 START CALIBRATE
2 - RESUME ACTIVE
30Spress
CALIBRATE
Press 1 to start calibration, and
2 to return to the ACTIVE interface.
MENU - RETURN

2. printer
install the paper
Open the printer cover interface, press FEED, press it again to stop it
after output the residual paper, get the surplus paper out;
Put the new paper in paper groove, the thermal printer paper are different
in two sides, so pull the end of paper out when the thermal side downwards;
Take the roll axis out upwards, put the thermal side under the roll axis as
the chart, the paper s direction should be vertically with the one of the roll
axis, it avoids the printer paper locked so that it cant print normally;

Paper Cover

11

trial version

The non-thermal
face
paper
groove

output
paper

roll-axis

paper

Figure : How to install printer-paper


Install the print roll axis, put the gear in the right side of the printer, then
put another side and press it to lock it tight;
Press FEED, adjust it if the paper cant get out or its direction is not
vertically with the one of roll axis, or it cant work; press FEED to stop it after
print a segment(When paper get out of the paper cover) ;

3. Install pump tube


Screw off the pump plate, slide the pump plate for a distance from the pump
head;

12

trial version

Figure: How to change the pump pipe


Take the pump tube out of the gap, Pull out the white ties connecting with
pump tubes two sides;
Get a new pump tube; connect the 2 soft tubes ties with the correspondent
pump tube;
Install new pump tube, put one side in the left flute, push another side in
the right flute according to the middle section;
Close the pump plate with proper strength, make sure the pump can
aspirate reagent; then screw on the plate tight. When the pump works, assure
it can absorb the right quantity of reagent, if not, regulate the place of the
baffle plate again.

4. Clean the rinse cup


If there is blockage or dirty thing in the rinse cup, you have to clean it. Switch
the power off, screw off the back cover, pull out of it, find the rinse cup, screw
off it, pull out the ties connected with the rinse cup, get the rinse cup out and
clean it with distilled water. Then install it back to place again.

13

trial version

Figure 2: How to clean the rinse cup

14

trial version

Charpter4 Fluidic System


1. GE200F

Fluid system of GE200 F Analyzer

2. GE200 A-E

Fluid system of GE200 A-E Analyzer

15

trial version

Charpter5 Maintenance
Daily maintenance:
1. clean the probe with tampon
2. clean the tubing with cleaner:
maintain---clean tube, then
aspirate the cleaner two second , wait then second; repeat this
step five times.
3. clean the surface of instrument

Weekly maintenance:
1. Activate the electrode with DEPROTEINIZED SOLUTION.
2. clean the wast bottle
3. check the reagent bottle

Charpter6 Troubleshooting
Problem
description
Calibration
failure

Analysis

Solution

fluidic system block


pump D can not supply
power to aspirate

16

Follow the fluidic system,


solve the block
1. fasten the pump tube
2. change the pump tube

trial version
no reagent in rinse
cpu

static
interferenceMV value
fluctuation
Reagent Expiry
Calibration setting
one point

Electrode problem

REF electrode problem

When you test


sample ,there are some
bubble in electrode
tubing

The problem in the fluidic


system from Reagent A and B
to rinse cup: 1. solve the block
2. fasten the pump tube of
reagent A and B
3. change the pump tube of
reagent A and B
1. connect the ground wire
2. clean
the
electrode
contactor
Change the reagent
Change the setting to two
point:
enter
utilitycalibration setting
---two point
1. change
the
filler
of
electrode
2. change the electrode
1. change
the
filler
of
electrode
2. change the electrode
1. Follow the fluidic system,
check the leakage
2. check
the
aspirate
volume :
enter into
utility---set ASP press
feed key to aspirate the
sample , then modify the
real volume.
3. adjust the moving volume:
enter
into
utility---set
ASP

Flick the electrode


Bubble at the bottom
of electrode
Activate the software: 1.
MV value normal, but
delete the sample result
calibration failure
2. modify the unit and
referance

17

trial version
Result too
Modify the factor
high or too low

Result
fluctuation

MV value :240

Rinse cup
leakage

Sample probe
shake

Recover the QC calibrate


factors back to original value
1.00.
then
enter
QC
calibrate---input the target
value of calibrator---test the
calibratorOK.

1. connect ground wire, the


1. ground wire
voltage of GND and N is
2. Sample isnt in
less than 5
electrode tubing
2. adjust the moving volume:
3. reagent problem
enter
into
utility---set
4. electrode problem
ASP
3. when reagent have been
opened, it can be use one
month.
4. change the filler or
activate the electrode,
replace the electrode
Short
connect
the 1. check voltage of the
analog board: if it is ok,
REF
contactor
and
change
the
interface
electrode contactor:
board, if the voltage is
1. MV value is 240
abnormal, change the
2. MV value is not 240
analog board
2. replace a working REF
1. fluidic
system 1. Follow the fluidic system,
block
from sample solve the block
2. Fasten the pump tube
probe to waste
change the pump tube
2. pump D can not
supply
power
to
aspirate
The position of sheet regulate the sheet metal of
metal
the optical baffle. Adjust
the sheet metal to the
proper
place
until
it
works.

18

trial version
Language
change

All the pump


can not move

No display

Display fuzzy

Switch on machine , then


press menu to main
menu,
then utility--press 000000---password
:123456---select
2english

check
the
power supply

out

of 1. If the output +12V is


ok,
change
the
interface board
2. if the output +12V is
abnormal,
change
the
power supply
Check other function, 1. Power
cant
get
EXP: pump, keyboard
through; Reconnect power
plug
2. Display signal line is
loose; Connect display
cable
3. if other function is OK,
replace the display or
interface board
4. there is no function ,
replace the CPU board
with software chip
The contrast of the Adjust
the
contrast
display
is
changed adjustor at the back of
with the environment instrument
temperature