Thank you for choosing the 2132 or 2116 Temperature Controller. Supplied in 1/32 and 1/16 DIN panel sizes they
are designed for accurate, stable control of ovens, chillers, sterilisers and other heating and cooling processes.
Two outputs are configurable for heating, cooling and alarms.
The controller is supplied configured according to the order code given in section 5. Check this on the side labels
to determine the configuration of your particular controller.
103mm (4.01in)
48mm (1.89in)
45mm
1.77in
-0.0, +0.6
-0.0, +0.02
Panel cut-out
24mm
Latching ears
22mm
-0.0, +0.3
0.88in
-0.0, +0.10
Model 2116
45 x 45 mm
Panel cut-out
48mm
-0.0, + 0.6
1.77 x 1.77in
-0.00, +0.02
48mm (1.89in)
Issue 5.0
103mm (4.01in)
Aug 07
1.1
1.2
The controller can be unplugged from its sleeve by easing the latching ears outwards and pulling it forward out of
the sleeve. When plugging it back into its sleeve, ensure that the latching ears click back into place to maintain the
IP65 sealing.
1.3
Issue 5.0
Aug 07
2. Electrical Connections
Solid
State
Relay
(SSR)
Logic
I/O
1B
T/C
Pt100
OR
1B 1A
24
20-29 Vac/dc
Low voltage
supply
Model 2116
L
N
Model 2132
1A
V-
V+ AB
+ Output 2
Relay
2.49
mA
Sensor Inputs
2.1
24
External Relay
Module (Operated
by the logic
output)
Wire Sizes
L
Line 85-264Vac 50/60Hz
Neutral
24 24
20-29 Vac/dc
Low voltage supply
OR
Issue 5.0
Aug 07
Solid
State
Relay
(SSR)
AB
Logic
I/O
1A
V+
1B
V-
Line
85-264Vac
Neutral 50/60Hz
Output 2 Relay
+
-
2.49
T/C Pt100 mA
Sensor Inputs
Output ratings
Logic Output: 9Vdc, 12mA (non-isolated from
sensor input).
Used for: Heating, Cooling or Alarm.
Relay Output: 2A, 264V ac resistive.
Used for: Heating, Cooling or Alarm.
Contact Closure Input (replaces Logic Output).
Used for: Alarm Acknowledge or Timer start/reset
2.2
Line
Circuit
Breaker
Controller fuse
2A type T
Solid State
Relay
(e.g. TE10)
1B
1A
V- V+ A
Model 2132
A
Heater
Neutral
+
Thermocouple
Snubber*
Cooling or
alarm relay
Relay
output
fuse
2A type T
* When switching
inductive loads such as
contactors or solenoid
valves, wire the
22nF/100 snubber
supplied across relay
terminals AA & AB. This
will prolong contact life
and reduce interference.
! WARNING
Snubbers pass 0.6mA at
110V and 1.2mA at
230Vac, which may be
sufficient to hold on high
impedance loads. Do not
use in these installations.
Issue 5.0
Aug 07
3. Operation
3.1
Switch on the controller. Following a 3 second selftest sequence, you will see the display shown below.
It is called the HOME display.
Output 1
Output 2
OP1
OP2
20
Actual
Temperature
(or Process
Value PV)
OP1
OP1
OP2
OP1
OP2
20
60
Required
temperature
(Setpoint)
Actual temperature
**
Press and hold
OP2
Issue 5.0
Aug 07
3.2
C
20
button. The
C Deg Centigrade
F Deg Fahrenheit
*
K Deg Kelvin
Linear inputs - no
units displayed
*
*
and
together will
If you get lost, pressing
always return you to the HOME display.
If, at any time, no key is pressed within 45 seconds,
the display will always return to the HOME display.
Press
and
together. This will also reset any
latched alarms that are no longer true.
Alarm Messages
Display Units
or
3.3
3.4
Actual
temperature
1fsL
20
Alarm 1
Full Scale
Low
Possible messages
Alarm - Full Scale High
Alarm - Full Scale Low
Alarm - Deviation
Alarm - Deviation High
Alarm - Deviation Low
Sensor Break
Loop Break
Load Fail
End of Timing
In place of the dash the alarm number is shown Alarm 1 or 2 or 3.
-fsH
-FSL
-deV
-dHi
-dLo
Sbr
Lbr
Ldf
End
Issue 5.0
Aug 07
3.5
20.0
Press twice quickly
Warning!
OP 100.0
Press
or
Controller is
demanding 100% heat
Press
Issue 5.0
Aug 07
3.6
HOME display
20.0
X2
aL
Atun
Keep pressing
to select
more list headings, eventually
returning to the HOME display.
This is a continuous loop.
Issue 5.0
Aug 07
3.7
Press
Press
or
0.5 sec
20
2nd press
* Press
or
to
change the setpoint.
Press
Issue 5.0
Next list
Press
1st press
LiSt
AL
Aug 07
and
Alarm 1
*
Alarm 2
*
Alarm 3
*
1---
2---
3---
3.8
Parameter Lists
Home
List
X2
Alarm
List
20.0
AL
OP
1---(1)
w.SP
2---
m A
3---(1)
diSP
(1)
Autotune
List
Atun
PID
List(2)
Pid
Setpoint
List
Input
List
SP
Access List
Output
List(2)
On/Off
List
iP
oP
On.Of
ACCS
tunE
Pb
SP L
FiLt
OP.Lo
HYS.H
codE
AdC
ti
SP H
CJCO
OP.Hi
Hys.C
Goto
td
SPrr
mV
CYC.H
HC.db
Conf
rES
tm.OP
OFS
CYC.C
Lcb
tmr
CAL.P
Ont.H
Hcb
dwel
CAL
Ont.C
rEL.C
StAt
Pnt.L
HY
Lb t
OFS.L
Pnt.H
OFS.H
10
Issue 5.0
Aug 07
3.8.1
Summary
1.
Press
2.
Press
3.
Press
4.
Press
Issue 5.0
Aug 07
11
3.9
Parameter Tables
Home List
Adjustable Range
Op
Output
Power
w.SP
Working
Setpoint
Manual/
Auto
mAn
m-A
Auto Select
disp
Home
Display
Options
OP
the
AL.SP
pv.aL
Read only
Read only
Auto
PV
Customer setting
Std
NonE
Default setting
and
Std
buttons.
12
and
Issue 5.0
Aug 07
AL
1---
Alarm 1 Setpoint
2---
Alarm 2 Setpoint
3---
Alarm 3 Setpoint
-FSH
-dEv
Deviation
-dHi
Deviation High
-dLo
Deviation Low
Adjustable
Range
Default
Setting
Between low
and high
setpoint limits
HY
Alarm Hysteresis
Lb t
OFF
Issue 5.0
Aug 07
Customer setting
13
Atun
tunE
OFF or on
Off
Adc
man or
caLc
man
Adjustable
Range
Default Setting
Adjustable Range
Pb
Proportional Band
ti
Integral Time
360
td
Derivative Time
60
rES
-100 to 100.0 %
0.0
PiD
Default Setting
Customer setting
Customer setting
20
Low Cutback
Auto
Hcb
High Cutback
Auto
rEL.C
0.01 to 10.00
14
1.00
Issue 5.0
Aug 07
SP
Adjustable Range
SP L
-1999 to 999.9
As per order
SP H
-1999 to 999.9
As per order
sprr
tm.OP
Opt.1 to Opt.5
tmr
Time Remaining
0 to 9999 minutes
dwEl
Dwell Time
OFF
StAt
Timer Status
OFF or on
OFF
Issue 5.0
Aug 07
Default Setting
Customer setting
Off
OPt.1
0
15
iP
Adjustable Range
FiLt
CJC*
Read only
Read only
mV
Default Setting
OFS
CAL.P
Calibration Password
0 to 9999
CAL
Customer setting
1.6
0
3
FACt
OFS.L
Pnt.H
OFS.H
16
0
-1999 to 9999 display
units
0
100
0
Issue 5.0
Aug 07
oP
Output List
Adjustable Range
OP.Lo
-100 to 100.0 %
OP.Hi
-100 to 100.0 %
100.0
CYC.H
CYC.C
ont.H
Auto
ont.C
auto
onOF
Adjustable Range
Default Setting
hYS.H
Heating Hysteresis
hYS.C
Cooling Hysteresis
HC.db
Heat/Cool Deadband
ACCS
Adjustable Range
Default
Setting
codE
Goto
Conf
0 to 9999
Oper, Ful, Edit, conf
1
OPEr
2
Issue 5.0
0 to 9999
Aug 07
Default Setting
Customer setting
Customer setting
Customer setting
17
3.10
ACCS
Press
PASS
codE
Press
Edit
Goto
Press
reached.
Press
or
to enter the password.
The factory default is 1. PASS will be
displayed when the correct password has
been entered.
or
to select Edit level.
Press
Other options are:
OPEr Operator level - shows selected
parameters
FuLL Reveals the FULL set of
parameters
ConF Gives access to configuration
level.
Example:
HidE
2FSH
HidE
.
rEAd
Pro
alterable
ACCS
18
Press
and
in the normal way.
to select a parameter
Issue 5.0
Aug 07
OPer
Goto
Example:
Pro
tmr
or
1.
Press
until you reach the
ACCS list heading
2.
Press
Goto
3.
Press
Oper
or
4.
Press
level
to return to Operator
to choose Pro.
to select
Issue 5.0
Aug 07
19
3.11
In reset
Press
or
to select the timer operating
Press
mode, Opt.1 to Opt.5 as follows:
Temperature
Setpoint
Standby mode
Waiting to
reach
temperature
Reset
Press
Auto
mAn
End flashes
Timing
Running
20
m-A
End
Press
display
and
or
to select:
Automatic control
Standby mode. (the MAN
beacon below OP2 will
illuminate)
Issue 5.0
Aug 07
Waiting to
reach
temperature
Reset
End flashes
Timing
End
Running
This is the same as mode 1 except that at the end of the
timing period the controller will continue indefinitely in
automatic control.
When Reset
End will stop flashing. The controller will return to
reset in standby mode. It can be returned to automatic
control by setting the parameter m-A in the HOME list
to Auto.
Issue 5.0
Aug 07
21
Temperature
Temperature
Setpoint
Standby mode
Setpoint
Indefinite dwell
at setpoint
End flashes
End flashes
Reset
Timing
End
22
Reset
Timing
End
Issue 5.0
Aug 07
Temperature
Setpoint
Reset
Indefinite dwell
at setpoint
Timing
Reset
Issue 5.0
Aug 07
3.12
Method 1.
This is the simplest method to control the timer.
until you reach the SP list
Press
Press
until you reach the tMr parameter
(time remaining).
1234
Tmr
Press
or
to enter
Press
the required timing period in
minutes. (0 to 9999).
to return to the HOME display
and
together.
24
Issue 5.0
Aug 07
Method 2.
Use this method if you want to set a fixed time and use
the stat parameter to start and stop the timer.
Press
sp
1234
dwel
Dwell time
or
to enter a
Press
timing period in minutes
(0-9999).
Timer Status
run
stat
Press
and
Issue 5.0
Aug 07
25
4.1
ACCS
Press
codE
Press
inSt
PASS
Press
or
to enter the
password. The factory default is 1.
PASS will be displayed when the
correct password has been entered.
conF
Press
iP
AL
AA
Exit
pASS
1A
Press
Goto
or
to select conf
Press
or
to enter the
configuration level password. The
factory default is 2. PASS will be
PASS
ConF
displayed when the correct password
has been entered.
Press to enter configuration level.
Press
26
and
to select a
Issue 5.0
Aug 07
4.1.1
Instrument Configuration
inst
unit
Instr Conf
Display units
dEC.P
Decimal
places in
display
CtrL
Control type
Act
Control
action
Options
*C
*F
*K
nonE
nnnn
nnn.n
nn.nn
Pid
On.OF
AL
rEv
dir
Issue 5.0
Description
Centigrade
Fahrenheit
Kelvin
None
None
One
Two
PID Control
On/off Control
Convert to an
alarm unit
Reverse
(normal action
for
temperature
control)
Direct (output
decreases as
PV falls below
SP)
Aug 07
Inst
Pd.tr
Instr Conf
Manual
reset
tracking (PD
control)
Options
HoLd
trAc
Description
In Auto holds
manual reset
value
In Auto tracks
output for
bumpless A/M
transfer
27
4.1.2
Input Configuration
iP
Sensor
Input
Options
Meaning
inPt
Input
type
j.tc
k.tc
L.tc
r.tc
b.tc
n.tc
t.tc
S.tc
PL 2
rtd
mV
C.tc
J thermocouple
K thermocouple
L thermocouple
R thermocouple
B thermocouple
N thermocouple
T thermocouple
S thermocouple
Platinell II
CJC
(TC
only)
28
Cold
junction
compen
sation
100 PRT
Linear mV
Custom input
C=default
Auto
Automatic
0*C
0C external ref.
45*C
50*C
mV input
low
InP.H
mV input
high
Displayed value
VAL.H
VaL.L
Displayed
value low
VAL.H
Displayed
value high
ImP
Sensor
break
OFF
Off (Linear
inputs only)
input
impedance
Auto
1.5K
Hi
5K
HiHi
15K,
VAL.L
Inp.L
Issue 5.0
mV
Inp.H
Aug 07
4.1.3
Alarm Configuration
Alarm
Type
Meaning
AL 1
Alarm 1
OFF
The alarm is
disabled
Full Scale Low
alarm
Full Scale High
alarm
Deviation band
alarm
Deviation high
alarm
Deviation low
alarm
fsL
fsH
dEv
dHi
dLo
Issue 5.0
Aug 07
AL
Alarm
Type
Meaning
Ltch
Alarm
latching
no
YES
Non-latching
Latched with
automatic*
resetting.
Latched with
manual**
resetting.
No blocking
mAn
bLoc
Alarm
blocking
No
YES
Blocked until
first good
Alarm
setpoint
diS
limits
Con
Limited by
display range
Limited by
setpoint limits
4.1.4
Aa
1a
Relay
Logic I/O
Identity of
output
Func
Function
These
functions
only
appear for
the logic
I/O
diG.F
30
Digital
output
functions
See To
Options
Meaning
rELy
LOG
diG
Relay
Logic
Digital (alarm)
output
Heating output
Cooling output
PDSIO mode 1
Alarm
Acknowledge
Keylock digital
input
Run/reset timer
No change
Clear all alarms
Alarm 1 (Note 1)
Alarm 2 (Note 1)
HEAt
COOL
SSr.1
Ac.AL
Loc.b
rres
noch
CLr
1FSL
2FSH
SenS
Relay
Logic I/O
Operate
the Relay
or Logic
Output
from an
Alarm or
Digital
Function
section
4.1.5
Options
Meaning
3FSL
NW *
SBR*
LBR*
LDF*
MAn *
EnD*
TMG*
TMG*
(Note 2)
TMG3*
TMG4*
Sense of
the
output
nor
Inv
Alarm 3 (Note 1)
New alarm
Sensor break
Loop break
Load fail alarm
Man mode active
End of timing
Timer running
Timer counting
down
Timer running
Timer counting
down
Normal (Note 3)
Inverted (Note 3)
Issue 5.0
Aug 07
5.
6.
Issue 5.0
Aug 07
4.1.6
Press
until to reach diG.F
or
to select CLr
Press
Leave for 2 seconds. The display returns to
diG.F which disconnects all alarms from the
relay.
31
4.1.8
4.1.9
Passwords
PASS
Passwords
Range
Default
ACC.P
0-9999
CnF.P
Configuration
level password
0-9999
CAL.P
User
calibration
password
0-9999
32
Press
Exit
YES
Press
or
to select YES After 2 secs the display
will blink and return to the HOME display in Operator level.
Issue 5.0
Aug 07
Message
EE.Er
Err3
HHHH
Err1
Err4
Err5
Hardware error:
TU.Er
Issue 5.0
Aug 07
33
4.2
User Calibration
Press
Press
or
to enter the password. The
Press
factory default is 3. PASS will be displayed
when the correct has been entered.
Press
Press
or
, to select User (FAct will
restore the factory calibration)
34
Displayed Value
User calibration
OFS.H
Factory calibration
OFS.L
Pnt.L
Pnt.H
Factory
calibration
Issue 5.0
Aug 07
4.3
Automatic Tuning
Display
Pb
Integral time
ti
Meaning or Function
The bandwidth in C or f
over which the output
power is proportioned
between minimum and
maximum.
Determines the time taken
by the controller to
remove steady-state error
signals.
Issue 5.0
Aug 07
Parameter
Derivative
time
Display
td
Low cutback
Lcb
High
Cutback
Hcb
Relative cool
gain
rEL.C
Meaning or Function
Determines how strongly
the controller will react to
the rate-of-change of
temperature.
The number of C or f
below setpoint at which
the controller will cutback
the output power to
prevent overshoot on heat
up.
The number of C or f
above setpoint at which
the controller will increase
the output power to
prevent undershoot on
cool down.
Only present if cooling has
been configured. Sets the
cooling proportional band
by dividing the Pb value by
the rEL.C value.
35
Issue 5.0
Aug 07
4.3.3
Temperature
1.
2.
3.
4.
Time
4.3.4
4.4
Manual Tuning
5.
Type of
control
Proportional
band Pb
2xB
Integral
time
ti
OFF
Proportional
only
P+I
P+I+D
Derivative
time td
OFF
2.2xB
1.7xB
0.8xT
0.5xT
OFF
0.12xT
Issue 5.0
Aug 07
37
4.4.1
Temperature
Undershoot
Example (b)
Time
Manual reset
Example (a)
Time
38
Issue 5.0
Aug 07
5. Ordering Code
The controller is supplied configured according to the ordering code shown below.
Model number
Function
Model Number
2132
1/16 DIN
2116
1/8 DIN
CC
NF
TC
TN
VH
VL
Function
PID controller
On/Off
controller
PID controller
+ timer
On/Off
controller +
timer
XXX
ENG
FRA
GER
NED
SPA
SWE
ITA
Supply voltage
Manual
None
English
French
German
Dutch
Spanish
Swedish
Italian
Supply voltage
85-264Vac
20 -29Vdc or
ac
Issue 5.0
Aug 07
Manual
Output 1 (Logic)
Output 1: Logic
XX
Disabled
Logic output
LH
Heating
LC
Cooling
M1
PDSIO mode 1
FH
High alarm 1
FL
Low alarm 1
DB
Dev band alarm 1
DL
Dev. low alarm 1
DH
Dev. high alarm 1
NW
New alarm
Logic input
AC
Alarm ack/reset
KL
Keylock
TM
Timer Run/Reset
Output 2 (Relay)
Output 2: Relay
XX
RH
RC
FH
FL
AL
DB
DL
DH
NW
Disabled
Heating
Cooling
High alarm 2
Low alarm 2
High alarm 2 & low
alarm 3
Dev band alarm 2
Dev. low alarm 2
Dev. high alarm 2
New alarm
39
Sensor input
Sensor input
Setpoint min
Display range and
Setpoint min & max limits
Thermocouples
J
Type J
Type K
Type T
C
-210 to
1200
-200 to
1372
-200 to 400
Type L
-200 to 900
Type N
-200 to
1300
-50 to 1768
-50 to 1768
0 to 1820
0 to 1369
R
S
B
P
Type R
Type S
Type B
Platinell
II
Resistance thermometer
Z
Pt100
-200 to 850
40
F
-340 to
2192
-325 to
2500
-325 to
750
-325 to
1650
-325 to
2370
58 to 3200
-58 to 3200
32 to 3308
32 to 2496
-325 to
1562
Setpoint max
Units
Input adaptor
Range OC
Range OF
32 to 4200
32 to 4350
-325 to 1830
32 to 2550
32 to 3398
32 to 3632
32 to 3650
50 to 4172
32 to 3632
392 to 3272
-49 to 1202
-9.99 to +80mV
0 to 20mA
4 to 20mA
0 to 10Vdc (input adapter required)
Units
C
C
F
F
K
Kelvin
X
Linear I/P
Input Adaptor
XX
None
V1
0-10Vdc
A1
0-20mA sense
resistor (2.49.
0.1%)
Issue 5.0
Aug 07
6. Technical Specification
Panel sealing
Operating ambient
0 to 55oC. Ensure that the enclosure is adequately ventilated. 5 to 95%RH, non condensing
Storage temperature
Atmosphere
Power supply
High voltage unit: 100 to 240Vac -15%, +10%, 48-62Hz, 5Watts maximum consumption
Low voltage unit: 24Vdc/ac +/- 20%. DC to 62Hz, 5Watts maximum consumption
Wire sizes
Use independent 2A fuses for the supply and relay output. Suitable fuses are EN60127 (type T)
Electrical safety
Meets EN 61010 (Voltage transients on the power supply must not exceed 2.5kV). Pollution
degree 2.
Isolation:
All isolated inputs and outputs have reinforced insulation to protect against electric shock. (See live
sensor note)
Cold Junction
Compensation
Installation Category
Category II or CAT II
Issue 5.0
Aug 07
41
42
GENERAL
The information contained in these instructions is
subject to change without notice. While every effort
has been made to ensure the accuracy of the
information, Eurotherm shall not be held liable for
errors contained herein.
Unpacking and storage
The packaging should contain an instrument mounted
in its sleeve, two mounting brackets for panel
installation and an Installation & Operating guide.
Certain ranges are supplied with an input adapter.
If on receipt, the packaging or the instrument is
damaged, do not install the product but contact your
supplier. If the instrument is to be stored before use,
protect from humidity and dust in an ambient
temperature range of -30oC to +75oC.
Issue 5.0
Aug 07
Personnel
Installation must only be carried out by qualified
personnel in accordance with instructions given in this
handbook.
Part Number HA026270
Issue 5.0
Aug 07
Issue 5.0
Aug 07
Issue 5.0
Aug 07
45
8. RoHS Certificate
Restriction of Hazardous Substances (RoHS)
Product group
2100
2100
X
O
X
X
O
X
O
O
O
O
X
O
O
X
O
O
O
O
O
O
O
O
O
O
O
O
SJ/T11363-2006
SJ/T11363-2006
English
Pb
X
O
X
X
Hg
O
O
O
O
PBDE
O
O
O
O
Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for
this part is below the limit requirement in SJ/T11363-2006.
Indicates that this toxic or hazardous substance contained in at least one of the homogeneous
materials used for this part is above the limit requirement in SJ/T11363-2006.
Approval
Name:
Position:
Martin Greenhalgh
Quality Manager
Signature:
Date:
46