ASTM Host Interface Simulator

User Manual
Version 5.6
Supported Instruments:
Elecsys 2010/1010,
STA, STA-C, STA-R,
LSM, PSM,
CC Modular, Hybrid Modular
Cardiac Reader,
Aliquoter VS II,
Amplilink, Urisys 2400, AVL OMNI,OMNI-C
ASTM+

ASTM Host Simulator User Manual V5-6.doc/C. Mahlmann, LG-I, 14.02.2003

1 of 23

ASTM Host Simulator Tool

Operators Manual

Contents:
1. PURPOSE OF THE ASTM HOST INTERFACE SIMULATOR

2. SYSTEM REQUIREMENTS

3. INSTALLATION OF THE TOOL

4. MAIN SCREEN OF THE TEST TOOL

5. MAIN MENU SYSTEM


5.1. Multifunctional text editor

6
9

6. TECHNICAL DETAILS
6.1. Record editor rules
6.2. TCP/IP Communication

12
12
12

7. DATA FLOW OF ASTM HOST INTERFACE SIMULATOR


7.1. ASTM+
7.2. Amplilink
7.3. AVL OMNI
7.4. Roche OMNI-C
7.5. Cardiac Reader
7.6. ELECSYS 2010
7.7. ELECSYS 1010
7.8. LSM
7.9. Modular CC
7.10. Modular Hybrid
7.11. PSM
7.12. STA, STA-C, STA-R
7.13. UriSys 2400
7.14. VS II

13
13
13
14
14
15
16
18
19
20
21
22
22
23
23

2 of 23

ASTM Host Simulator Tool

Operators Manual

1. Purpose of the ASTM Host Interface Simulator


The ASTM Host Interface Simulator allows to simulate the host communication for all Roche analyzers that are
equipped with an ASTM host protocol.
The tool may simulate the analyzer or the host. It was developed with the programming language Visual Basic 6.0
and requires a 32-bit operating system (Windows 95 or NT).
The tool has instrument-independant features ...
low-level communication
trace recording
messages may be created with a record editor, saved as so-called record files (*.DAT) and sent to the
connected device
error simulation by sending wrong checksums
free timer setting
and instrument specific features ...
the extracted info of received messages is stored in instrument-specific MS ACCESS databases
realtime request mode may be simulated with this database
the meaning of the record fields may be displayed in the lower part of the record editor

2.System Requirements
Hardware:
IBM-compatible computer, capable of running Windows NT, 2000, XP
Software:
32-bit Windows Operating System (NT, 2000, XP)

3. Installation of the Tool


The ASTM Host Interface Simulator is to be installed on the hard disk by the SETUP.EXE routine.
The SETUP routine creates a new directory and copies the program files and library files to the hard disk.
Be sure that no other application is open during setup !

3 of 23

ASTM Host Simulator Tool

Operators Manual

4. Main Screen of the Test Tool

The Main Screen contains the following items (top down):


Title line (containing the selected instrument, the programs name and version)
Pull-down menu system (see chapter 2.1)
Local Device setting (HOST or INST) and their names
The Operating Mode setting screen can be accessed by clicking on this field
Symbolic indication of the local device (blue frame)
A click on the corresponding icon selects the local device
actual Date and Time
actual RS232-Parameter setting
The RS232 Parameter setting screen can be accessed by clicking on this field
Step-Mode status (visible or invisible means enabled or disabled)
Last Sent/Received messages
if activated on the Mode of Operation screen an info line is appended to the list for each sent and received
message. The list may be cleared with the Clear List button.
actually opened Trace file name
The file selection mask for trace files can be accessed by clicking on this field
Size of the actually opened trace file in bytes.

4 of 23

ASTM Host Simulator Tool

Operators Manual

actually opened Record file name


The file selection mask for record files can be accessed by clicking on this field
Number of messages to send
Communication Trace Window
Sent messages are displayed in green, received messages in blue and error messages in red color. If the space
character (code 20hex) within records is replaced by another character, this is indicated above the trace
window.
The icon right to the Communication Trace Window allows to enter a comment line which is written to the
trace file.
Status line
Indication of the actual transition state (Idle, Transfer, Receive) and the active keys with time counter, if
STEP mode is enabled)

5 of 23

ASTM Host Simulator Tool

Operators Manual

5. Main Menu System


Record File
Delete

Record files (*.DAT) which are no longer needed can be deleted from the
harddisk

Display/Edit/Parser/Create

A record file (*.DAT) can be displayed, printed, searched, ... with a multifunctional text editor (see 3.1 for a description of the editor options)
For starting the record editor you need to open an already existing DAT
file

Send

Any previously created record file can be sent via the serial interface

Exit

This option quits the program

Trace File
Open for Recording

A file can be opened where the complete communication in both


directions is stored with time and sender information

Close

The above opened trace file can be closed with this option

Delete

Trace files which are no longer needed can be deleted from the harddisk

Display/Edit

Any trace file can be displayed, printed, searched, ... with a multifunctional text editor (see 3.1 for a description of the editor options)

Extract Records

All records of any trace file are extracted and written to a record file by
this option.

Insert Comment

This option allows to enter a comment which is written to the actually


open trace file.

Clear Window

This option clears the trace window.

6 of 23

ASTM Host Simulator Tool

Operators Manual

Options
Interface Settings

The tool may communicate via RS232 or TCP/IP.


For the RS232 interface the baudrate, data-, stopbits, parity, COM port and
kind of handshaking can be set on this screen.
The TCP/IP communication is not yet evaluated !
This screen can also be opened by clicking on the RS232 parameter field on
the main screen

Mode of Operation

The following settings can be done on this screen:


the connected instrument
enter the operators name
the local device (HOST or INSTRUMENT) incl. the device names
the colors for the trace messages
enable/disable step mode
enable/disable acustical error signal
enable/disable storage of received records in the instrument-specific
database
display one line for each sent or received message in a list control on
the main screen
definition what to send in case of host if there is no TS available in the
database for a received request
definition of a space replacement character for the trace display
For Elecsys 1010/2010, the following settings can be done
decision which sample info items shall be echoed on a test request
enable/disable storage of received manufacturer records in the database
enable/disable storage of manufacturer records in the trace file
For Modular/Hybrid, the following settings can be done
decision which sample info items shall be echoed on a test request
activation of rules for a reflex testing if working as host
For PSM, the following settings can be done
single record or single frame
For OMNI, the following settings can be done
<CR> or <CR><LF> as record delimiter

Timer Settings

All timing and retry adjustments can be done on this screen

Delimiters

The four delimiter characters of the ASTM protocol can be selected with
this option

Low-Level Errors

Different kind of error simulation can be activated for testing the interface
software of the connected device

continued on next page

7 of 23

ASTM Host Simulator Tool

Operators Manual

Options menu continued


DB Access

This screen allows to take a look at the different database tables which
contain the received data. Test selections and test selection requests can
be generated and sent to the connected device.
There are two options for the database in a pull-down menu available:
Compress DB eliminates holes in the database file and decreases
the file size
Store Column Width stores the actual column width of the
displayed table after a manual adaptation.
Create Report creates result reports, if the result table is displayed

Desktop

The programs desktop language can be selected (german or english)

Communication Test

This option initiates a short [ENQ] - [ACK] - [EOT] sequence for testing
the communication, if the test program is connected to itself.
This test is supported by the STA Compact, it is not supported by the
Elecsys 2010 interface !

?
Help on ASTM Simulator

This option gives access to the integrated online help for the ASTM
Simulator itself

Help on ASTM 1381 (Low Level


Specification)

The ASTM specification document as helpsystem

Help on ASTM 1394 (High Level


Specification)

The ASTM specification document as helpsystem

Software Changes History

Displays the software changes of previous releases of the tool

System Info

Displays system information

Save Screen Shot

Allows to save a previously captured screen shot (via [PrtSc] key to the
clipboard) as bitmap file. As an additional option this bitmap file may be
compressed with PKZIP.

About

An info window with version number and programmers address

Clear Send Buffer

Clears all message that are to be send

8 of 23

ASTM Host Simulator Tool

Operators Manual

5.1. Multifunctional text editor


For displaying record and trace files there is a text editor window which looks as follows:

The upper window displays the contents of the record or trace file. The lower window is an option of the record
editor where the meaning and contents of the fields is displayed

9 of 23

ASTM Host Simulator Tool

Operators Manual

The record/trace editor has a pull-down menu with the following options:
File
Open

Another record/trace file can be opened

Save as

The modified file or a selected part of it can be stored in


another file

Print

The whole file, a selected part of it or the parser information


(only for record files) can be written to the printer, a text
file or the clipboard.

Send

(only available, if record files are displayed)


The contents of the text window is sent to the other device

Exit

This option quits the text editor

Edit
Cut

the selected part of the text is cut and copied to the


clipboard

Copy

the selected part of the text is copied to the clipboard

Paste

the contents of the clipboard is inserted at the cursor


position

Delete

the selected part of the text is deleted

Insert Trace Comment

enters !C at the actual cursor position.


the text which may be entered after that expression is not
sent but displayed in the trace file

Insert Delay

enters !D at the actual cursor position.


the time in seconds which may be entered after that
expression causes a delay during sending

Provoke Wrong Checksum

enters !S at the actual cursor position.


this expression causes a checksum error in the following
record during sending

Provoke Wrong Frame Number

enters !F at the actual cursor position.


this expression causes a wrong frame number in the
following record during sending

10 of 23

ASTM Host Simulator Tool

Operators Manual

text editor menu system (cont)


Search
Search text

any expression can be searched in the text file

Search again

the above specified search is repeated

Options
Parser

(only available, if record files are displayed)


if a record file is displayed the text window is divided into
two parts. In the upper part, the record file is displayed, in
the lower part the field information of the selected record is
displayed. If the Parser option is deactivated, the text
window increases to full height.

Instrument Messages

(only visible, if record files are displayed and the program


works as instrument simulator)
All messages that the instrument can send during operation
are offered in a option box. The selected message is
displayed with default values in the text editor window and
can be modified.

Host Messages

(only visible, if record files are displayed and the program


works as host simulator)
All messages that the host sends during operation are
offered in a option box. The selected message is displayed
with default values in the text editor window and can be
modified.
On this screen the test selection default profile can be
displayed, modified and stored in a separate file.

Small/Large Window

with this option, the text editor window can be reduced to


the half height. So it is possible to observe the trace window
on the main screen while the text editor is open. If a record
file is displayed and the small window is selected, the parser
option is disabled

?
Help

This option gives access the integrated online help and the
System Information window

Test List

If working as Elecsys or Modular, the list of tests with test


codes is displayed in a small list control

11 of 23

ASTM Host Simulator Tool

Operators Manual

6. Technical Details
6.1. Record editor rules
The record text file can contain more than one message (a message starts with the Header record, contains several
data records and ends with the Termination record). Each message has to start with a Header record; otherwise it
is not sent.
The time delay between messages can be specified on the Timer Settings screen.
6.2. TCP/IP Communication
Beside the communication via the serial interface RS232 the simulator may also communicate via TCP/IP
network.
The corresponding setting is done on the Options/Interface Settings screen.
Here you select between RS232 and Network interface.
If RS232 is selected, the baud rate, port number, data-, stopbits and parity check may be selected.
If Network is selected, there are three further options:
Device works as ...

TCP/IP Server (the other device needs to be TCP/IP Client


TCP/IP Client (the other device needs to be TCP/IP Server
TCP/IP UDP = User Datagram Protocol (the other device needs the same setting)

According to the above listed options, you have to enter additional settings.
If Server is selected, the local port number needs to be entered. This must be the same number than the remote
port number of the client.
After re-starting the simulator, a dialog is displayed which indicates the network options and the simulator as
server waits for a client to connect before he gives access to the program features.
If Client is selected, the remote port number needs to be entered. This must be the same number than the local
port number of the server.
In addition the servers IP address (format: aaa.bbb.ccc.ddd) must be entered.
After re-starting the simulator, a dialog is displayed which indicates the network options and the simulator as
client needs to be connected to a listening server by pressing the [Connect] buttons.
The User Datagram Protocol (UDP) is a connectionless protocol. Unlike TCP operations, computers do not
establish a connection. Also, a UDP application can be either a client or a server.
You have to enter the other devices IP address, the local port number (which must be entered as remote port
number at the other device) and the remote port number (which must be entered as local port number at the
other device).
After re-starting the simulator, a dialog is displayed whcih indicates the network options and both devices need
to be connected before you get access to the program features.
When selecting the option 'RS232 <> TCP/IP Conversion' the ASTM Simulator passes messages received via the
RS232 port to the TPC/IP port and vice versa. In this case both setting frames are visible.

12 of 23

ASTM Host Simulator Tool

Operators Manual

7. Data Flow of ASTM Host Interface Simulator


7.1. ASTM+
The table on the following pages lists the data flow options of the test program for the ASTM+ protocol
specification.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor

Send out of DB

Test Selection

P,O

Test Requests
Test Results
Manufacturer
Records

Q
P,O,R,C
M

Edit and send


Edit and send
Edit and send

Edit and send


Edit and send
Edit and send

Messages

Records

Send out of DB

Test Selection

P,O

Send out of
Record-Editor
Edit and send

Test Requests
Test Results
Manufacturer
Records

Q
P,O,R,C
M

answer requests

DB write
Store received data
Store received data
Store received data
Store received data

HOST Mode:
answer requests

Edit and send

Store received data


Send TS if available

Edit and send

DB write

Edit and send

Store received data


Store received data
Store received data

7.2. Amplilink
The table on the following pages lists the data flow options of the test program for the Amplilink.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Test Selection

P,O

Result Request
Test Results

Q
P,O,R

Send out of
Record-Editor

Send out of DB

answer requests

DB write
Store received data
Store received data

Edit and send

Edit and send

As answer to result request

Send out of
Record-Editor
Edit and send

Send out of DB

answer requests

Edit and send

Edit and send

HOST Mode:
Messages

Records

Test Selection

P,O

Result Request
Test Results

Q
P,O,R

DB write

Edit and send


Store received data

13 of 23

ASTM Host Simulator Tool

Operators Manual

7.3. AVL OMNI


The table on the following pages lists the data flow options of the test program for the AVL Omni.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of DB

Send out of
Record-Editor
Edit and send

Patient Query

answer requests

DB write

Edit and send

Store received data

Test Result

P,O,R

Edit and send

Edit and send

Store received data

Calibration
Report
Maintenance
Report
QC Report

P,O,R

Edit and send

Edit

Store received data

Edit and send

Edit

Store received data

P,O,R

Edit and send

Edit

Store received data

Messages

Records

Send out of DB

answer requests

DB write

Patient Info

P,O

Send out of
Record-Editor
Edit and send

Edit and send

send, if query is
received

Store received data

HOST Mode:

7.4. Roche OMNI-C


The table on the following pages lists the data flow options of the test program for the Roche OMNI-C.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of DB

Send out of
Record-Editor
Edit and send

Patient Query

answer requests

DB write

Edit and send

Store received data

Test Result

P,O,R

Edit and send

Edit and send

Store received data

Calibration
Report
Maintenance
Report
QC Report

Edit and send

Edit

Store received data

Edit and send

Edit

Store received data

O,R

Edit and send

Edit

Store received data

Messages

Records

Send out of DB

answer requests

DB write

Patient Info

P,O

Send out of
Record-Editor
Edit and send

Edit and send

send, if query is
received

Store received data

HOST Mode:

14 of 23

ASTM Host Simulator Tool

Operators Manual

7.5. Cardiac Reader


The table on the following pages lists the data flow options of the test program for the Cardiac Reader.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

TS Request

Test Selection

P,O

Result Request
Result

Q
P,O,R,C

Action Request
Action
Confirmation
Inst Config
Request
Inst
Configuration
Inst Status
Request
Inst Status

AR
AC

Proc Message
Request
Proc Message

PMQ

Subst Data
Request
Substance Data

XTQ

Send out of
Record-Editor

Send out of DB

Store received data

Edit and send

Edit and send

Store received data

Edit and send

Edit and send

Send confirmation

PM

Store received data


Store received data

Send first DB entry if


available
Edit and send

Edit and send

ISQ
IS

DB write

Edit and send

ICQ
IC

answer requests

Store received data


Send first DB entry if
available

Edit and send

Edit and send

Store received data


Send first DB entry if
available

Edit and send

Edit and send

Store received data


Send DB entry if
available

XT

Edit and send

Messages

Records

Send out of
Record-Editor

TS Request

Test Selection

P,O

Result Request
Result

Q
P,O,R,C

Edit and send

Action Request
Action
Confirmation
Inst Config
Request
Inst
Configuration
Inst Status
Request
Inst Status

AR
AC

Edit and send

ICQ

Edit and send

IC

Edit and send

ISQ

Edit and send

Proc Message
Request
Proc Message

PMQ

Subst Data
Request
Substance Data

XTQ

Edit and send

XT

Edit and send

Store received data

HOST Mode:
Send out of DB

answer requests

DB write

Edit and send

Store received data


Edit and send

Send confirmation
Store received data

IS

Edit and send

Store received data

Store received data


Edit and send

PM

Store received data

Store received data

15 of 23

ASTM Host Simulator Tool

Operators Manual

7.6. ELECSYS 2010


The table on the following pages lists the data flow options of the test program for the ELECSYS 2010 analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
(mm)
(mr)
(sr)

each selected DB entry is sent in a separate message


each selected DB entry is sent in a separate record within one message
if more than one entry is selected only the first will be sent

INSTrument Mode:
Messages

Records

Record Editor
(Requests)

Calibration
Result
Calibration
Parameters
Control
Parameters
Diluent
Parameters
Instrument
Config
Instrument
Status
Log File

CR

Record-Editor
(Data)
Edit and send

XT, CP

Edit and send

Send out of DB

answer requests

DB write
Store received data
Store received data

XT, QP

Store received data

XT, DP

Store received data

IC

Edit and send

Select DB and
send (sr)
Select DB and
send (sr)
Select DB and
send (lf)

IS

Edit and send

LF

Edit and send

Processing
Message
Request Cancel
Sample Status

PM

Edit and send

Select DB and
send (sr)

SS

Edit and send

Service Data

SD

Edit and send

Select DB and
send (sr)
Select DB and
send (sr)

Substance Data
Test Application

XT
TA

Test Conditions

TC

Edit and send

Test Results

P, O, R

Edit and send

Result Context
Raw Result

RC
RR

Test Selection

Q, P, O

Select DB and
send (sr)
Select DB and
send (mr)
Select DB and
send (mr)

send 1. entry of DB
on request
send 1. entry of DB
on request
send on request with
date and category
check

Store received data


Store received data
Store received data
Store received data

send 1. entry of DB
on request
send 1. entry of DB
on request

Store received data


Store received data
Store received data
Store received data

send 1. Entry or all of


DB on request

Store received data


Store received data
Store received data
Store received data

edit and send TS


Requests

Select Request
DB and send
(mm)

store received TS
data
create no. of
requests

16 of 23

ASTM Host Simulator Tool

Operators Manual

HOST Mode:
Messages

Records

Calibration
Result
Calibration
Parameters
Control
Parameters
Diluent
Parameters
Instrument
Config
Instrument
Status
Log File
Processing
Message
Request Cancel
Sample Status
Service Data

CR

Substance Data
Test Application
Test Conditions
Test Results
Result Context
Raw Result'

XT
TA
TC
P, O, R
RC
RR

Test Selection

Q, P, O

Record Editor
(Requests)

Record-Editor
(Data)

DB read

DB write
Store received data

XT, CP

Edit and send request

Edit and send

Store received data

XT, QP

Edit and send request

Edit and send

Store received data

XT, DP

Edit and send request

IC

Edit and send request

IS

Edit and send request

Store received data

LF
PM

Edit and send request

Store received data


Store received data

SS
SD

Store received data


Edit and send

Store received data

Edit and send request


Edit and send request
Edit and send request

Store received data


Store received data

Edit and send request


Edit and send request
Edit and send request

Store received data


Store received data
Store received data
Store received data
Store received data
Store received data

Edit and send

Edit and send

select DB and send


(mm)
send on request, if
available, or default TS if
enabled

store received
request
create no. of TS

17 of 23

ASTM Host Simulator Tool

Operators Manual

7.7. ELECSYS 1010


The table on the following pages lists the data flow options of the test program for the ELECSYS 1010 analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
(mm)
(mr)
(sr)

each selected DB entry is sent in a separate message


each selected DB entry is sent in a separate record within one message
if more than one entry is selected only the first will be sent

INSTrument Mode:
Messages

Records

Calibration
Parameters
Control
Parameters
Diluent
Parameters
Instrument
Config
Instrument
Status
Log File

XT, CP

Record Editor
(Requests)

Record-Editor
(Data)
Edit and send

Send out of DB

answer requests

DB write
Store received data

XT, QP

Store received data

XT, DP

Store received data

IC

Edit and send

IS

Edit and send

LF

Edit and send

Processing
Message
Service Data

PM

Edit and send

SD

Edit and send

Substance Data
Test Application

XT
TA

Test Conditions

TC

Edit and send

Test Results

P, O, R

Edit and send

Result Context
Raw Result

RC
RR

Edit and send


Edit and send

Test Selection

Q, P, O

edit and send TS


Requests

Select DB and
send (sr)
Select DB and
send (sr)
Select DB and
send (lf)
Select DB and
send (sr)
Select DB and
send (sr)
Select DB and
send (sr)
Select DB and
send (mr)
Select DB and
send (mr)

send 1. entry of DB
on request
send 1. entry of DB
on request
send on request with
date and category
check

Store received data


Store received data
Store received data
Store received data

send 1. entry of DB
on request

Store received data


Store received data
Store received data

send 1. Entry or all of


DB on request

Store received data


Store received data
Store received data
Store received data

Select Request
DB and send
(mm)

store received TS
data
create no. of
requests

18 of 23

ASTM Host Simulator Tool

Operators Manual

HOST Mode:
Messages

Records

Record Editor
(Requests)

Record-Editor
(Data)

DB read

DB write

Calibration
Parameters
Control
Parameters
Diluent
Parameters
Instrument
Config
Instrument
Status
Log File
Processing
Message
Service Data

XT, CP

Store received data

XT, QP

Store received data

XT, DP

Store received data

IC

Edit and send request

Edit and send

Store received data

IS

Edit and send request

Store received data

LF
PM

Edit and send request

Store received data


Store received data

SD

Edit and send request

Store received data

Substance Data
Test Application
Test Conditions
Test Results
Result Context
Raw Result'

XT
TA
TC
P, O, R
RC
RR

Edit and send request

Store received data


Store received data
Store received data
Store received data
Store received data
Store received data

Test Selection

Q, P, O

Edit and send request


Edit and send request

Edit and send

select DB and send


(mm)
send on request, if
available, or default TS if
enabled

store received
request
create no. of TS

7.8. LSM
The table on the following pages lists the data flow options of the test program for the LSM.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor

Send out of DB

Test Selection

P,O

TS Request

Edit and send

Edit and send

Test Results

P,O,R,C

Edit and send

Edit and send

Messages

Records

Send out of DB

Test Selection

P,O

Send out of
Record-Editor
Edit and send

TS Request

Test Results

P,O,R,C

answer requests

DB write
Store received data

HOST Mode:
answer requests

DB write

Store received data

Edit and send


Send all TS if
ALL request

Store received data

19 of 23

ASTM Host Simulator Tool

Operators Manual

7.9. Modular CC
The table on the following pages lists the data flow options of the test program for the Modular analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor
Edit and send

Send out of DB

answer requests

DB write

Realtime TS
Request
Routine Test
Selection
Result Request

P,Q

Routine
Result
Photo Calib
Result
ISE Calib Result
Absorbance
Result

P,O,R,C

Edit and send

Edit and send

Store received data

Edit and send

Edit and send

Store received data

M
P,O,R,C,M

Edit and send


Edit and send

Edit and send


Edit and send

Store received data


Store received data

Messages

Records

Send out of
Record-Editor

Send out of DB

Realtime TS
Request
Routine Test
Selection
Result Request
Routine
Result
Photo Calib
Result
ISE Calib Result
Absorbance
Result

P,Q

Edit and send

P,O

Store received data

P,Q

Send result if
available

Store received data

HOST Mode:

P,O

Edit and send

P,Q
P,O,R,C

Edit and send

answer requests

DB write

Send TS if available

Store received data

Edit, send and


create
Store received data

Store received data

M
P,O,R,C,M

Store received data


Store received data

20 of 23

ASTM Host Simulator Tool

Operators Manual

7.10. Modular Hybrid


The table on the following pages lists the data flow options of the test program for the Hybrid Modular analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Realtime TS
Request
Routine Test
Selection
Result Request

Routine
Result
Photo Calib
Result
ISE Calib Result
E Calib Result
CC Abs. Result

P,O,C,R,C

Edit and send

Edit and send

Store received data

Edit and send

Edit and send

Store received data

M
M
P,O,C,R,C,
M
P,O,C,R,C,
M

Edit and send


Edit and send
Edit and send

Edit and send


Edit and send
Edit and send

Store received data


Store received data
Store received data

Edit and send

Edit and send

Store received data

Messages

Records

Send out of
Record-Editor

Send out of DB

Realtime TS
Request
Routine Test
Selection
Result Request
Routine
Result
Photo Calib
Result
ISE Calib Result
E Calib Result
CC Abs. Result

E Absorbance
Result

Send out of
Record-Editor
Edit and send

Send out of DB

answer requests

DB write

Edit and send

P,O

Store received data

Send result if
available

Store received data

HOST Mode:

E Absorbance
Result

P,O

Edit and send

Q
P,O,C,R,C

Edit and send

answer requests

DB write

Send TS if available

Store received data

Edit, send and


create
Store received data

Store received data

M
M
P,O,C,R,C,
M
P,O,C,R,C,
M

Store received data


Store received data
Store received data
Store received data

21 of 23

ASTM Host Simulator Tool

Operators Manual

7.11. PSM
The table on the following pages lists the data flow options of the test program for the PSM.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor

Send out of DB

Test Selection

P,O

TS Request

Edit and send

Edit and send

Test Results

P,O,R

Edit and send

Edit and send

Messages

Records

Send out of DB

Test Selection

P,O

Send out of
Record-Editor
Edit and send

TS Request

Test Results

P,O,R

answer requests

DB write
Store received data

HOST Mode:
answer requests

DB write

Store received data

Edit and send


Send all TS if ALL request
search sampleID
search patientID, if sampleID is
empty

Store received data

7.12. STA, STA-C, STA-R


The table on the following pages lists the data flow options of the test program for the STA analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor

Send out of DB

Worklist

P,O

Worklist
Request
Result

Edit and send

Edit and send

P,O,R,M

Edit and send

Edit and send

Messages

Records

Send out of DB

Worklist

P,O

Send out of
Record-Editor
Edit and send

Worklist
Request
Result

answer requests

DB write
Store received data

HOST Mode:

P,O,R,M

answer requests

DB write

Send worklist if
available

Store received data

Edit and send

Store received data

22 of 23

ASTM Host Simulator Tool

Operators Manual

7.13. UriSys 2400


The table on the following pages lists the data flow options of the test program for the UriSys analyzer.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Result

P,O,R,C,M

Send out of
Record-Editor
Edit and send

Send out of DB

Send out of
Record-Editor

Send out of DB

answer requests

DB write

answer requests

DB write

Edit and send

HOST Mode:
Messages

Records

Result

P,O,R,C,M

Store received data

7.14. VS II
The table on the following pages lists the data flow options of the test program for the Aliquoter VS II.
Depending on wether the program works as Host or Instrument simulator it is shown which messages can be
generated, modified and send with which option.
INSTrument Mode:
Messages

Records

Send out of
Record-Editor

Send out of DB

Test Selection

P,O

TS Request

Edit and send

Edit and send

Sort List

LI

Edit and send

Edit and send

Messages

Records

Send out of DB

Test Selection

P,O

Send out of
Record-Editor
Edit and send

TS Request
Sort List

Q
LI

answer requests

DB write
Store received data

HOST Mode:
answer requests

DB write

Send TS if available

Store received data


Store received data

Edit and send

23 of 23