Professional Documents
Culture Documents
7-1
ACCURACY AND
RELEVANT EFFECTS
7-2
Synopsis
Definitions of accuracy and point accuracy precede a detailed discussion of relevant
terminology necessary for interpreting instrument accuracy specifications: zero and
span, systematic error, turndown (rangeability), precision, reproducibility, repeatability
and sensitivity.
The environment in which an instrument operates effects its accuracy. Consequently,
terminology relating to functional specifications is introduced.
A system for comparing transmitter specifications of different manufacturers is outlined.
The calibration of transmitters requires an understanding of other error sources: linearity,
hysteresis, repeatability, reproducibility, dead band.
Methods, including MTBF, are examined as means of calculating system errors.
7-3
Accuracy
1.
The degree to which an indicated value matches the actual value of a measured variable.
2.
Quantitatively, the difference between the measured value and the most probable value for
the same quantity, when the latter is determined from all available data, critically adjusted for
sources of error.
3.
Accuracy, Measured
The maximum positive and negative deviation observed in testing a device under specified conditions
and by a specified procedure.
Speaking more fundamentally, an accuracy of 99% is an inaccuracy of 1%. For convenience,
however, the latter inaccuracy has always been referred to as accuracy.
Sometimes an accuracy statement is spoken of in point accuracy terms.
Point Accuracy
The limits of error of an instrument may be expressed in a number of ways. In some cases the
point accuracy is given. This is the accuracy of the instrument at one point on its scale only, and
does not give any information on the general accuracy of the instrument.
Before we delve into specifications, lets take a look at some common terminology.
Terminology
RANGE: The region between the limits within which a quantity is measured, received, or
transmitted.
UPPER RANGE LIMIT (URL):
LOWER RANGE LIMIT (LRL): The lowest quantity that a device can be adjusted to measure.
UPPER RANGE VALUE (URV): The highest quantity that a device is adjusted to measure.
LOWER RANGE VALUE (LRV): The lowest quantity that a device is adjusted to measure.
SPAN: The algebraic difference between the upper and lower range values.
A typical instrument specification could thus look:
Example: Model 1151DP4 Calibrated 0 to 100 in. H20
Range = 0/25 to 0/150 Inches H20
URL = 150
Inches H20
LRL = -150
Inches H20
URV = 100
Inches H20
LRV = 0
Inches H20
Span = 100
Inches H20
7-4
7-5
7-6
A high degree of precision is, however, no indication that the value of the measured variable has
been accurately determined. A manometer may give a reading of 10 pounds per square inch for a
certain pressure to within 0.01 pounds per square inch, but it may be several pounds per square inch
in error, and the true value of the pressure will be unknown until the instrument has been calibrated.
It is accurate calibration that makes accurate measurement possible.
Fig. 7.4
1.
2.
3.
Bias error is small and precision is good (device 3); this is an accurate device.
Sensitivity. The sensitivity of an instrument is usually taken to be the size of the deflection produced
by the instrument for a given change in the measured variable. It is, however, quite frequently
used to denote the smallest change in the measured quantity to which the instrument responds.
The largest change in the measured variable to which the instrument does not respond is called the
dead zone.
Rangeability. The rangeability of a measuring instrument is usually taken to mean the ratio of the
maximum meter reading to the minimum meter reading for which the error is less than a stated
value. For example: In a positive displacement flow meter a certain quantity of liquid passes the
fixed and moving parts. The maximum quantity which can be measured by the meter is usually
fixed by the meter size. Increasing the flow above the meter maximum will shorten the life of the
meter owing to greatly increased wear and is therefore highly undesirable. It will be seen from the
graph below that the minimum flow for which the meter accuracy is plus or minus 0.5 per cent is 10
per cent of the maximum reading. The rangeability for an accuracy of plus or minus 0.5 per cent of
true value is therefore 10 : 1.
Fig. 7.5
7-7
7-8
Vibration Effect
The effect upon output is solely due to the vibratory environment to which the transmitter is subject.
Mounting Position Effect
Difference in output when a transmitter is mounted in a position different to which it was calibrated.
Shows itself as a zero shift or systematic error which can be recalibrated away.
Load Effect
If the total loop should alter, then the output of the transmitter (for the same process input) may be
effected.
EMI/RFI Effect
An output change solely due to interference of an electromagnetic or radio frequency radiator.
7-9
Specifications
Transmitter A
Transmitter B
300 in H2O
300 in H2O
Accuracy
0.2% of span
0.1% of URL
Temperature Effect
Zero
Span
Total*
Static Pressure Effect
Zero
Span
Total*
Table 7.1
0 to 100 in H20
500 F
500 psig
75 in H20
Table 7.2
7 - 10
Transmitter A
Transmitter B
Accuracy
Temperature Effect
Zero
Span
Total*
Span
0.91 in H2O
0.19 in H2O
0.25% x 75 x 500/1000=
0.94 in H2O
0.12 in H2O
Total*
Table 7.3
TRANSMITTER A
TRANSMITTER B
Manufacturers may use any or all of the above terms when citing transmitter specifications; thus it
is imperative the entire specification is read and understood when evaluating transmitters.
7 - 11
7 - 12
Hysteresis:
The maximum difference for the same input between the up-scale and the downscale output values during a full range traverse in each direction.
Fig. 7.9
Repeatability:
Fig. 7.10
Reproducibility:
7 - 13
Dead Band:
The range through which an input can be varied without initiating response.
Load Limitations
The maximum load that can be present in the loop for the transmitter to operate
over its full output range for a given power supply voltage.
Fig. 7.11
This is calculated by
Max Load (RL)
Typical Lift Off Voltage would be 12V so for a Power Supply of 24 Volts and
20mA signal the maximum logs resistance would be:
RL
=
=
24 - 12
0.02
600 ohms
7 - 14
Power Supply: Describe the power that is required to operate the instrument. EX: 12-45 volts dc
Load Limitation: The maximum loop resistance that can be present with a specific power supply.
Indication: Addresses the use of indicating meters.
Hazardous Locations: Describes the types of hazardous location that the instrument is certified for
use within.
e.g.:
Damping: A discussion involving time constants will be useful before defining damping. A time
constant is the time required for an instruments output to complete 63.2% of a total input step
change. Time constants are designated by the Greek letter tau () and defined by the equation 1-ne
where:
e = natural logarithmic base (2.718)
The Table summarises the present changes for a duration up to 5 time constants.
Time constant ()
1
2
3
4
5
Equation
1- e
1- 2e
1- 3e
1- 4e
1- 5e
% Output change
63.2%
86.5%
95.0%
98.2%
99.3%
Since the logarithmic function never reaches 100%, it is common to use 4 or 5 time constants to
approximate the total response of the instrument. This total response is commonly referred to as
Response Time. For example, if an instrument has a 0.2 second time constant, total response time
can be calculated by multiplying the time constant (0.2 sec) by the number of desired time constants
(e.g. 4 or 5).
Damping is the ability to electronically adjust an instruments time constant.
Turn-on Time: The maximum time required, during power-up, which allows instrument operation.
B. Performance Specifications
Temperature Effect
The temperature effect, unless otherwise stated, is assumed to include both the zero error and the
total effect. It can be expressed a number of ways, i.e.,
0.15% of span / 100F
1.0% of max. range/100F between 50F and 150F
0.01 F/F between -15 and 185F
Overpressure Effect
On a differential pressure transmitter, overpressure may be either on the high side or the low side.
It can be expressed a number of ways, i.e.,
7 - 15
2.
3.
4.
7 - 16
5.
As you can see, this calculation is a tool for design and reliability engineers to pinpoint a weak area
of design, not an implicit guarantee of any type.
D. Sigma Conformance
Some manufacturers are now stating that their instruments are manufactured to a certain sigma ()
conformance. This means that the manufacturer uses data drawn from a small sample to represent
the complete population using a normal distribution curve to determine the standard Deviation
from specification. This is a means of predicting instrument errors and is vital in setting specifications.
The area under the normal distribution curve is used to determine the probabilty of picking an
instrument with a given error from the population of instruments manufactured. For example, a
manufacturer specifies a product to have a maximum error of 0.75%. If the manufacturer designs to
a three sigma deviation, then 99.73% of the instruments they manufacture would meet or exceed
the 0.75% specification. In turn, this implies that 95.45% (two sigma deviations) of the manufactured
instruments would meet or exceed an error of 0.50%. Finally, the area under the one sigma deviation
represents the 68.27% of instruments with errors less than or equal to 0.25%. Therefore, to publish
a specification based on a three sigma standard deviation ensures that 99.73% of the instruments
manufactured meet or exceed the published specifications.
Fig. 7.12
Spec.Limit
Defective ppm
1 Sigma
68.27
317300
2 Sigma
95.45
45500
3 Sigma
99.73
2700
4 Sigma
99.9937
5 Sigma
99.999943
0.57
6 Sigma
99.9999998
0.002
63
7 - 17
Case Study
Fig. 7.14
7 - 18
The digitalised signal from the sensor is corrected for non-linearity and
temperature effects. The length of time for the microprocessor to perform
the polynomial correction is the Signal Processing Time.
Protocol Processing:
Fig. 7.15
Update Time
Definition: How frequently an output is provided by the transmitter.
The joint effect of A/D and P determine the Update Time.
7 - 19
This effects the frequency of sampling the process in the measurement situation.
Fig. 7.16
Resolution
Definition:
Fig. 7.17
7 - 20
Deadtime
Definition: The time between the initial input change and the output response.
Example: Shows the effect of dead time when considering a step input change in pressure.
Fig. 7.18
The following graphs show how transmitter deadtime destroys measurement accuracy. It manifests
itself as a lateral shift on the time axis.
Fig. 7.19
This lateral shift of the graph represents the age of the process measurement since sampled.
Digital Sampling times should ideally be integral multiples of the Input measurement frequency. If
not, Beat Frequency Error or Steady State Error can result.
7 - 21
Input/Output Relationship
fs
sampling frequency
fi
input frequency
The faster the sampling frequency, the more closely the output will follow the input.
For example:
If we have a sine wave pressure input cycling at 2 Hz (2 cycles per second), and a sampling rate of
1 Hz (1 cycle per second), then the output will not follow the input as accurately as if the sampling
and processing rate was 4 Hz (4 cycles per second).
Fig. 7.21
Fig. 7.20
STEADY STATE ERROR
If the sampling rate equals the input rate, a steady state error will result.
The following example illustrates worst case. If the sampling point was lower on the curve,
obviously the error would be less.
Fig. 7.22
Are we saying that the digital output cant be used to control a process? No, it can be used on slow
loops such as temperature or tank level, but unless the loop is lethargic, the analog output will allow
better control of the process.
7 - 22
Summary
Accuracy may be quoted as a percentage of a specified range or referenced against a single
measurement on its scale (point accuracy).
Range terminology includes: URL, LRL, URV, LRV and Span. Span is the actual, calibrated
measurement limits of a specific device (URV - LRV).
Errors may be classified as:
1.
2.
3.
4.
Systematic - predictable, repeatable error. Examples include temperature and static pressure
effects.
5.
Turndown - arises when a span less than the instruments maximum is used. The smaller the
calibrated span, the greater the errors over that span.
2.
3.
Rangeability - limits upon the instrument spans wherein errors are of an acceptable value.
4.
Linearity - the deviation of actual instrument output from ideal linear calibration.
5.
6.
Repeatability - Deviation in measurements for same input approaching from one direction.
7.
8.
7 - 23
Discussion
a.
From the point of view of accuracy, compare the relative merits of analogue and digital
instruments.
b.
c.
Indicate any industry trends towards accuracy of which you are aware.
7 - 24
Test
1.
2.
3.
A control loop consists of 3 instruments each having a load of 250 ohms. For successful
operation over 4 - 20 mA, what is the minimum power supply required? Assume the transmitter
has a 12V lift-off.
___________________________________________________