UROLITH+ with HF-49

EXTRA CORPOREAL SHOCK WAVE LITHOTRIPTER (E.S.W.L.)

TECHNICAL SPECIFICATIONS

Model: UROLITH+

General:

Operating Voltage 230V AC, 50 Hz, 15 Amps.

Operating Temperature 10-30 C

Humidity 20 80% Non-condensing.

Shock Wave Generation/Parameter:

Principal of Operation Under Water Spark Gap Discharge.

Reflector Opening 180 mm.

Spark Gaps Long Life Electrodes.

Shock Wave Coupling Flexible Latex Membrane.

Water Capacity 8 Liters.

Voltage Range 14 to 24 KV

Pressure Range at F2. 300 to 1200 Bars.

Digital Display KV & Shocks

Pre determined shocks 100, 500, 1000, 1500, 2000, 2500, 3000,
3500
Shocks/ Min. Selection 60/75/90 shocks per min.

Focal Distance 135 mm.

Focal Zone Diameter 4mm x 12 mm.

Mode Selection Normal/ECG/Respiration

Locating System Removable Focal Point Locator.

Treatment Table:
Treatment Table Motorized three-axis table with radiolucent
top, remote controlled.
Imaging:

ALLENGERS HF-49

ALLENGERS HF 49
(High Frequency (40 KHz) Mobile C-Arm / 9 Image Intensifier/ 32 Frame Image
Memory / 2 Nos. 17 Monitors.)

TECHNICAL SPECIFICATIONS

A) IMAGE INTENSIFIFER:
1. Image Intensifying Tube 9 Inches, Triple Field (Imported).

Normal = 9, Z1 = 6, Z2 = 4.5.
2. Nominal Entrance Field Size 23cm
3. Output Image diameter 20mm
4. CCD Camera Half Inch in Size, High Resolution Compact
CCD Camera.
5. Monitor 17, Monitor along with a trolley.
2 Nos.
B) C-ARM MOBILE STAND
1. Rotation 180 Degrees.
2. Motorized Up/Down 430 mm
3. Horizontal Travel 200 mm
4. Arc Orbital Movement 90 Degrees + 25 Degrees.
5. Wig Wag 25 Degrees total.
C) X-RAY GENERATOR
1. Tube Head 40 KHz. Stationery Anode Tube of focal spot
(Monoblock) 0.6mm (FLR) & 1.5mm (RAD).
(a)
Max. Voltage 110 KV
Fluoroscopic mA 0.1- 3mA
(b)
(c) Radiographic mA 30mA to 70mA
2. INVERTOR AC/DC Converter for H.F. X-Ray Generator.
(a) Mains Voltage 230V AC 10%, 50 Hz.
(b) Max. output power 3.5 KW
CONTROL
1. Digital Display of KV 40 to 110. User Selectable.
2. Technic Selection Fluoro & Rad. Mode Selection.
3. Radiographic Timer An inbuilt Rad. Timer enables to select mAs in
steps upto 200mAs max.
4. Fluoroscopic Timer 5 minutes cumulative Timer with Buzzer.
5. Fluoro Ma 0.1-3 mA, continuously variable.
6. Image Orientation on Monitor. Indicators for left/right, up/down Image
Rotation is provided on Control Panel.
7. Monoblock Temp. Sensor. Thermal Safety cut off provided.

8. Fault Status Various LEDs indicate the faults.


9. Power supply Single Phase 230 V, + 10% AC, 50 Hz
(Equipment Operatable on Standard 15
Ampere Power Socket)
10. Automatic Brightness Control Provided for hands free Imaging
D) DIGITAL IMAGE MEMORY SYSTEMS :
32 Frame Digital Memory with Last Image Hold, Recursive Filter and Negative Image
reproduction.