Manager:
Title Date
Head,
PERFORMER REVIEWER
NAME
SIGNATURE
DATE
QUALIFIERS IDENTIFICATION
The undersigned performer / reviewers certify that they are trained and qualified to perform
complete qualification procedure for the ARCTIKO Pharmaceutical Refrigerator PR 1400
for M/s. ARCTIKO ,DENMARK
Performed by:
Service Engineer
Reviewed by:
PERFORMER REVIEWER
NAME
SIGNATURE
DATE
v Installation Report
v Overview
v IQ Certificate
PERFORMER REVIEWER
NAME
SIGNATURE
DATE
IQ Installation Report 1 of 1
Module Pharmaceutical Refrigerator
Model PR 1400
Serial No.
Installation Report
No. Action Result
1. Check Packing list
2. Unpack
3. Check for completeness (compare with packing list)
4. Check for transport damage
5. Remove transport screws
6. Set up the Refrigerator at suitable location (See Instruction Manual)
7. Let the Refrigerator compartment air out with the doors open
8. Wire up the system
9. Connect the power cable to the power supply and turn on the switch.
10. Connect the battery cables to battery
11. Check whether display is coming in the temperature display that is in the
control panel showing actual temperature of the Refrigerator
compartment and recorder is recording the same temperature which is in
the display of the control Panel.
12. Set operating mode for desired application i.e., set the desired
temperature (see Instruction Manual)
13. Leave the Refrigerator ON to cool down (reach) to the desired
temperature
14. Instruct the customer in using the system.
Safety instructions
Operating Principle of Pharmaceutical Refrigerator & Recorder
Operating and function elements & Setting the desired temperature
Checking the instrument and the GLP functions
Simple maintenance procedures and replacement of parts & accessories
15. Handover the Instrument Documentation to the Customer
Instruction Manual
16. Comments:
Successfully Installed the Pharmaceutical Refrigerator as per the
procedure.
PERFORMER REVIEWER
SIGNATURE
DATE
Installation Qualification
Installation Documents
Instruction
Instrument/ Model Serial No. Report Handed
Performed
Prepared Over
Yes / No Yes / No Yes / No
Pharmaceutical
Refrigerator
Model: PR 1400
PERFORMER REVIEWER
SIGNATURE
DATE
v Test Reports
v Overview
v OQ Certificate
PERFORMER REVIEWER
SIGNATURE
DATE
OQ Test Report 1 of 5
Module Pharmaceutical Refrigerator
Model PR 1400
Serial No.
Function Test
No Item Tested Setting Specifications Result
1.
a. Display Buttons Press ON /OFF LCD Display will show
Power Supply Switch Button Once. the actual Temperature
of Pharmaceutical
Switch ON (move Refrigerator
to l position) the compartment, Date Time,
battery. Ambient temperature,
Battery level, Alarm
Indication, Alarm Icon
and compressor position.
Total Pharmaceutical
Press ON /OFF Refrigerator will turn off.
Button Once again.
.
PERFORMER REVIEWER
SIGNATURE
DATE
1.b. Up& Down Arrow keys Press key A beep sound will be heard
once and the display will show
the set numerical value,
icon.
Ex. 4.0
OQ Test Report 2 of 5
Module Pharmaceutical Refrigerator
Model PR 1400
Serial No.
Function Test
No. Item Tested Setting Specifications Result
1.c. Left & Right Arrow Press key A beep sound will come and
keys once the display left digit will blink
the numerical value, icon.
Ex. 4.0
1.d. Enter & accesses Press Display will show the Menu
the key once for mode.
menu Key access mode.
PERFORMER REVIEWER
SIGNATURE
DATE
OQ Test Report 3 of 5
PERFORMER REVIEWER
SIGNATURE
DATE
OQ Test Report 4 of 5
PERFORMER REVIEWER
SIGNATURE
DATE
SOP PR 1400 1 of 1
OQ
Objective To ensure that the analyst operates the instrument as per the
requirement
Operating Procedure
PERFORMER REVIEWER
SIGNATURE
DATE
By using arrow keys chose the custom setting icon. Then Press
Enter Key. Displays Enter password Icon
.
By using arrow keys chose the Set point icon. Then Press
Enter Key. Displays the previous set temperature value.
Set the desire temperature by using Arrow keys and press Enter
key.
Ex. 4.0
Now press the Esc key for exit.
set temperature will be memorized and actual temperature of the
Pharmaceutical Refrigerator compartment will be displayed.
Ex. 4.0
PERFORMER REVIEWER
SIGNATURE
DATE
Documents Necessary
Test Result
Handed Corrective
Instrument/ Model Serial No. Confirmation
Over Action
Pharmaceutical
Refrigerator
Model: PR 1400
PERFORMER REVIEWER
SIGNATURE
DATE
v Environment Conditions
v Materials Used
v PQ Procedures
v Test Reports
v Overview
v PQ Certificate
Aim
PERFORMER REVIEWER
SIGNATURE
DATE
Indoor use
Altitude up to 2000 meters.
Temperature 5C to 40C.
Maximum relative humidity 80% for temperature up to 31C decreasing linearly to
50% relative humidity at 40C.
Main supply voltage fluctuations not to exceed 10% of the nominal voltage.
Other supply voltage fluctuations as stated by the manufacturer.
Transient over voltage according to installation categories (Over voltage
categories) II, for mains supply the minimum and normal category is II.
POLLUTION DEGREE 2 in accordance with IEC 664.
PERFORMER REVIEWER
SIGNATURE
DATE
Materials Used
PERFORMER REVIEWER
SIGNATURE
DATE
5. Note down the time at different temperature say: initial temperature 15C,
10C and 5C as shown in Test Report1
7. See the Fig.1 of the refrigerator chamber. Total 6 points are shown to measure
the temperature.
8. Set the refrigerator temperature 5c and by keeping the calibrated data scanner
at all points shown in Fig.1. Note down the temperature values shown by the data
scanner and refrigerator display.
9. Data scanner connects to the printer for 24 hours data, with 1 Minute print interval.
Compare them whether they are within the tolerance limits specified by the
manufacturer.
PERFORMER REVIEWER
SIGNATURE
DATE
3. Close the door for the Refrigerator reached the set temperature.
PERFORMER REVIEWER
SIGNATURE
DATE
3 Now the Pharmaceutical Refrigerator temperature is too low, check the low alarm
temperature activated or not.
4 After reached the freezer 4C, set the high alarm temperature value 1C.
PERFORMER REVIEWER
SIGNATURE
DATE
PQ Test Report 1 1 of 1
Module Pharmaceutical Refrigerator
Model PR 1400
Serial No.
Pull down Performance of the Pharmaceutical Refrigerator
Time taken by the Time taken to
Pharmaceutical reach the Set
Temperature on
No. Time Refrigerator to Temperature Result
the Display
reach Set according to
Temperature manufacturer
1. Initial Temperature
2. 15C
3. 10C
4. 5C 4 Hours
5. Comments:
PERFORMER REVIEWER
SIGNATURE
DATE
1. 5C 1Min.
2. 5C 3Min.
3. 5C 5Min.
4. 5C 10Min.
5. Comments:
1. Initial
temperature
2. After 5 min.
PERFORMER REVIEWER
SIGNATURE
DATE
PQ Test Report 3 1 of 1
Module Pharmaceutical Refrigerator
Model PR 1400
Serial No.
Checking of the Temperature high and low alarm at +7C, +1C,
Pharmaceutical
No. Temperature Alarm Activation Result
Refrigerator display
1. Initial Temperature +5C
2. Set the Pharmaceutical +5C
Refrigerator
temperature
3. Set the low alarm +1C
temperature
4. Initial Temperature +5C
5. Set the high alarm +7C
temperature
6. Temperature of the +10C
Pharmaceutical
Refrigerator after doors
open
7. Comments:
PERFORMER REVIEWER
SIGNATURE
DATE
Documents Necessary
PQ Result
Instrument / Handed Corrective
Serial No. Confirmation
Model Over Action
Yes / No Yes / No Yes / No
Pharmaceutical
Refrigerator
Model: PR 1400
PERFORMER REVIEWER
SIGNATURE
DATE
4. Maintenance
Note: After guarantee period the customer should enter into Annual
Maintenance Contract (AMC) with the Indian Agent M/s Tec Pharma Systems
Inc.
a. Procedure for Cleaning
The appliance should be cleaned once a month.
Wipe clean the external surfaces and internal accessories of the appliance. Mild
detergent can be used if necessary.
Dry the appliance both inside with soft cloth after cleaning with water.
Do not use excessive water for cleaning the appliance to avoid damage to the
installation materials.
Do not use any lubrication inside the appliance since the compressor and other
mechanical accessories are all tightly sealed.
b. Lamp replacement
Unplug the appliance.
Hold lamp holder and remove the lampshade.
Rotate to remove the lamp, replace with a new lamp.
Connect the appliance to a power source.
Unplug the appliance and remove the starter.
Tightly installed the new one, connect the appliance to a power source.
PERFORMER REVIEWER
SIGNATURE
DATE
b. S A F E T Y P R E C A U T I O N S
5. Instrument Documents
Over View of the Documents Handed Over
MODEL: PR 1400
Serial No.:
PERFORMER REVIEWER
SIGNATURE
DATE
6. Spare Parts
PERFORMER REVIEWER
SIGNATURE
DATE