You are on page 1of 3

Proline XC

Digital X-ray Imaging


Hardware Diagrams and
System Requirements

Proline XC
Digital X-Ray
CPU Specifications Network Server CPU Specifications
Digital Image Workstation Specifications Additional Workstations
For Panoramic, Tru Profile TMJ and Pentium 4 or Higher For Panoramic, Tru Profile TMJ and
Cephalometric 2D images 2 Hard Drives, 80 GB each min. Cephalometric images
Pentium 4 or Higher 2 GB or more RAM Refer to Digital Image
60 GB or better HD WINDOWS XP / Server 2003 OS / Acquisition Workstation (2D)
2 GB or more RAM Vista (Business or Ultimate
WINDOWS XP / Server 2003 OS / versions 32-bit OS only) Additional Workstation
Vista (Business or Ultimate Ethernet 10/100 Mbit For Panoramic,
versions 32-bit OS only) / MAC Tru Profile TMJ and
OSX 10.4 and above (Intel-Based only) Network Server Cephalometric images (2D)
Ethernet 10/100 Mbit Stores and archives all data Romexis Software
24-32 bit graphic card to MS SQL / SQL Express.
Monitor 1280 x 1024 res. min. The data server should have
High Speed Internet, DSL or better large hard drive disk space,
hardware redundancy and a
Image Acquisition Workstation (2D) logical backup system.
Romexis Software acquires the
Panoramic, Tru Profile TMJ and
Cephalometric 2D images (2D).
Note: An Uninterrupted Power Romexis Imaging Software is used
Supply (UPS) / Power Conditioner to enhance, filter, measure,
is recommended for the X-Ray. diagnose, archive and export
Install a new Planmeca Power the images.
Conditioner with a newly
purchased Planmeca Imaging
System and Planmeca USA, Inc.
will extend the the product
warranty from two years to three.

Prolines Require
Network Interface Box
Local Area Network Connection
Required for sharing printer or adding
additional workstations.

NIB Standard 10/100 Switch Optional


Required Optional Printer

PLANMECA USA, Inc. 100 N. Gary Ave., Suite A, Roselle, Illinois 60172 U.S.A. Phone: (630) 529-2300 www.planmecausa.com PMUSA1209
Generator Constant potential,
Minimum operational microprocessor controlled,
space requirements operating frequency 80kHz

Width Depth Height X-ray tube D-052SB

XC Pan Only 153 cm 103.5 cm 220 cm Focal spot size 0.5 x 0.5 mm (IEC 336)

60 in. 40.5 in. 86.5 in. Total filtration min. 2.5 mm Al

Anode voltage 60 - 80 kV 5 %

XC Pan with 163 cm 103.5 cm 220 cm Anode current 4 - 12 mA DC ( 5 %+ 0.5mA)


Autoprint
64 in. 40.5 in. 86.5 in
Line voltage 100-132 VAC
60 Hz
Regulation 10 % (automatic)
XC Pan 223 cm 103.5 cm 220 cm
with Ceph Line current max. 8A at 230V, 15A at 100V
88 in. 40.5 in. 86.5 in.
Digital Sensor CCD Technology

Pixel Size 33 microns

Image Pixel Size 130 m normal resolution


99 m enhanced resolution

Active Sensor Pan 9 x 136 mm


Surface Ceph 9 x 230 mm

Digital Image Pan 14 x 30 cm (5.5 x 12)


Field Ceph 22.5 x 30 cm (8.9 x 12 in.)

Resolution 6-9 lp/mm pan; 3.8-5 lp/mm ceph

Exposure time Pan 2.5 - 18 sec. as indicated 8 %


Ceph: 0.2 - 23 sec. 10 %

SID Pan 480 mm (19 in.)


Ceph 163 - 170 cm (64 - 67 in.)

Cassette Flat

Film Size Pan 15 x 30 cm


Ceph 8 x 10 in.

Magnification Pan constant 1.2 (20%)


Ceph 1.08 - 1.13 (8-13%)

Exterior color RAL 9016 (white)

Weight Panoramic Model


108 kg (237 lbs.)
Panoramic with Autoprint
112 kg (249 lbs.)
Panoramic with Ceph
126 kg (278 lbs.)

14 PLANMECA
Proline XC
Technical specifications
Digital Proline XC Film-based Proline XC
with Digital Ceph with Film-based Ceph

80.5" 84.5"
80.5" 84.5"

86.5"
86.5"

88" Recommended (with Ceph) 88" Recommended (with Ceph)


73" with Cephalostat 77"
47
45"
64" with Autoprint
21" 43"
21" 41" Rec.
64" with Autoprint 17" 32" 11"

9" 29" 14"


28.5"
28.5"

56" Recommended

56" Recommended

Optional
37.5"

37.5"

Admark
Keyboard

32.5"
32.5"
39" Pan Only
41"
62" Recommended (Pan Only)
64" Recommended (Pan Only)