Oscilloscope

HM303-6

Manual
English
Contents

Oscilloscope
HM 303-6

General information regarding the CE marking ........... 3


4 in DUAL mode ............................................................... 13
Phase difference measurement in DUAL mode ........... 14
General Information ......................................................... 6 Measurement of an amplitude modulation .................. 14
Symbols ........................................................................... 6
General Information ........................................................ 6 Triggering and time base .............................................. 14
Use of tilt handle ............................................................. 6 Automatic Peak (value) - Triggering .............................. 15
Safety .............................................................................. 6 Normal Triggering .......................................................... 15
Intended purpose and operating conditions ................... 6 Slope ...................................................................... 15
Warranty .......................................................................... 7 Trigger coupling ............................................................. 15
Maintenance .................................................................... 7 Line triggering (~) .......................................................... 16
Protective Switch-Off ...................................................... 7 Alternate triggering ....................................................... 16
Power supply ................................................................... 7 External triggering ......................................................... 16
Trigger indicator ............................................................ 16
Type of signal voltage ...................................................... 8 Holdoff-time adjustment ............................................... 16
Amplitude Measurements .............................................. 8
Voltage values of a sine curve ........................................ 8 Component Tester .......................................................... 17
Total value of input voltage ............................................. 9 General .......................................................................... 17
Time Measurements ....................................................... 9 Using the Component Tester ........................................ 17
Connection of Test Signal ............................................. 10 Test Procedure .............................................................. 17
Test Pattern Displays .................................................... 17
First Time Operation ...................................................... 11 Testing Resistors .......................................................... 17
Trace Rotation TR .......................................................... 11 Testing Capacitors and Inductors ................................. 17
Probe compensation and use ....................................... 11 Testing Semiconductors ............................................... 18
Adjustment at 1kHz ....................................................... 11 Testing Diodes .............................................................. 18
Adjustment at 1MHz ..................................................... 12 Testing Transistors ........................................................ 18
In-Circuit Tests .............................................................. 18
Operating modes of
the vertical amplifiers .................................................... 12 Front Panel Elements HM 303-6
X-Y Operation ................................................................ 13 (Brief Description - Front View) ............................... 20
Phase comparison with Lissajous figures .................... 13
Phase difference measurement .................................... 13 Short Instruction for HM303-6 ...................................... 22

2 Subject to change without notice


General information regarding the CE marking

KONFORMITTSERKLRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
Hersteller HAMEG Instruments GmbH Angewendete harmonisierte Normen / Harmonized standards applied / Normes harmonises
Manufacturer Industriestrae 6 utilises
Fabricant D-63533 Mainhausen Sicherheit / Safety / Scurit
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994
Die HAMEG GmbH bescheinigt die Konformitt fr das Produkt EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05
The HAMEG GmbH herewith declares conformity of the product berspannungskategorie / Overvoltage category / Catgorie de surtension: II
HAMEG GmbH dclare la conformite du produit Verschmutzungsgrad / Degree of pollution / Degr de pollution: 2

Bezeichnung / Product name / Designation: Elektromagnetische Vertrglichkeit / Electromagnetic compatibility /


Compatibilit lectromagntique
Oszilloskop/Oscilloscope/Oscilloscope
EN 61326-1/A1
Typ / Type / Type: HM303-6 Straussendung / Radiation / Emission: Tabelle / table / tableau 4; Klasse / Class / Classe B.
Strfestigkeit / Immunity / Imunitee: Tabelle / table / tableau A1.
mit / with / avec:
EN 61000-3-2/A14
Optionen / Options / Options: Oberschwingungsstrme / Harmonic current emissions / missions de courant harmoni-
que: Klasse / Class / Classe D.
mit den folgenden Bestimmungen / with applicable regulations / avec les directives
suivantes EN 61000-3-3
Spannungsschwankungen u. Flicker / Voltage fluctuations and flicker /
EMV Richtlinie 89/336/EWG ergnzt durch 91/263/EWG, 92/31/EWG Fluctuations de tension et du flicker.
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amende par 91/263/EWG, 92/31/CEE Datum /Date /Date Unterschrift / Signature /Signatur

Niederspannungsrichtlinie 73/23/EWG ergnzt durch 93/68/EWG


15.01.2001
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amende par 93/68/CEE
E. Baumgartner
Technical Manager /Directeur Technique

General information regarding the CE marking


HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product stan-
dards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for residential, commercial and
light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been used.

The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the acceptance limits. For
different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding emission
and immunity should be observed:

1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened cables must be used.
Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3 meters and not be
used outside buildings. If an interface has several connectors only one connector must have a connection to a cable.

Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG are
suitable.

2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for
a shorter length, signal lines must be less than 3 meters and not be used outside buildings.

Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened
cables (RG223/U, RG214/U) must be used.

3. Influence on measuring instruments.


Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an influence of such
signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments spe-
cifications may result from such conditions in individual cases.

4. RF immunity of oscilloscopes.

4.1 Electromagnetic RF field


The influence of electric and magnetic RF fields may become visible (e.g. RF superimposed), if the field intensity is high. In most cases the coupling
into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/or radiation. The device under test as well
as the oscilloscope may be effected by such fields.

Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each amplifier stage
is higher than the total 3dB bandwidth of the oscilloscope, the influence RF fields of even higher frequencies may be noticeable.

4.2 Electrical fast transients / electrostatic discharge


Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via test leads and/or control
cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals may effect the trigger unit and/or may become
visible on the CRT, which is unavoidable. These effects can also be caused by direct or indirect electrostatic discharge.

HAMEG Instruments GmbH

Subject to change without notice 3


HM303-6

35 MHz Analog Oscilloscope


HM303-6

Maximum signal fidelity with minimum overshoot


No signal distortion
resulting from overshoot
2 Channels with deflection coefficients of 1 mV 20 V/cm,
Low Noise Amplifiers

Time Base 0.2 s 100 ns/cm, with X Magnification to 10 ns/cm

Triggering from 0 to 50 MHz from 5 mm signal level


Line triggered composite
(100 MHz 8 mm)
video signal

Analog mode provides unexcelled signal presentation at high


resolution and up to 500,000 signal displays/sec.

Yt, XY and component-test modes

Full screen display of


35 MHz sine wave signal

4 Subject to change without notice


Specifications

Horizontal Deflection
35 MHz Analog Oscilloscope HM303-6 Time Base: 0.2 s/div. 0.1 s/div. (1-2-5 Sequence)
Accuracy: 3%
Valid at 23 C after a 30 minute warm-up period
Variabel (uncalibrated): 2.5:1 to 0.5 s/div.
X Magnification x 10: up to 10 ns/div.
Accuracy: 5%
Vertical Deflection
Hold-Off Time: variable to approx. 10 : 1
Operating Modes: Channel I or II only
XY
Channels I and II (alternate or chopped)
Bandwidth X Amplifier: 0 2.5 MHz (-3 dB)
Sum or Difference of CH I and CH II
XY Phase shift 3: 120 kHz
Invert: CH II
XY Mode: via CH I (X) and CH II (Y)
Component Tester
Bandwidth: 2 x 0 to 35 MHz (-3 dB)
Test Voltage: approx. 7 Vrms (open circuit)
Rise Time: 10 ns
Test Current: max. 7 mArms (short-circuit)
Overshoot: max. 1%
Test Frequency: approx. 50 Hz
Deflection Coefficients: 1-2-5 Sequence
Test Connection: 2 banana jacks 4 mm
1 mV/div. 2 mV/div.: 5 % (Bandwidth 0 10 MHz (-3 dB))
One test circuit lead is grounded via protective earth (PE)
5 mV/div. 20 V/div.: 3 % (Bandwidth 0 35 MHz (-3 dB))
Variable (uncalibrated): 2.5 : 1 to 50 V/div.
Miscellaneous
Input Impedance: 1 M II 20 pF
CRT: D14-363GY, 8 x 10 cm with internal graticule
Input Coupling: DC, AC, GND (ground)
Acceleration Voltage: approx. 2 kV
Max. Input Voltage: 400 V (DC + peak AC)
Trace Rotation: adjustable on front panel
Calibrator Signal (Square Wave): 0.2 V 1 %, 1 kHz/1 MHz (tr 4 ns)
Triggering
Power Supply (Mains): 105 253 V, 50/60 Hz 10 %, CAT II
Automatic (Peak to Peak): 20 Hz 50 MHz ( 5 mm)
Power Consumption: approx. 36 Watt at 230 V/50 Hz
50 MHz - 100 MHz ( 8 mm)
Ambient temperature: 0 C...+ 40 C
Normal with Level Control: 0 - 50 MHz ( 5 mm)
Safety class: Safety class I (EN61010-1)
50 MHz 100 MHz ( 8 mm)
Weight: approx. 5.4 kg
Trigger Indicator: LED
Dimensions (W x H x D): 285 x 125 x 380 mm
Slope: positive or negative
Sources: Channel I or II, CH I / CH II alternate
( 8 mm), Line and External
Coupling: AC: 10 Hz 100 MHz
DC: 0 100 MHz
LF: 0 1.5 kHz
Trigger Indicator: LED
External Trigger Signal: 0.3 Vpp (30 Hz 50 MHz)
Accessories supplied: Line Cord, operators manual, 2 Probes 1:1 / 10:1 (HZ154)
Active TV sync. separator: pos. and neg.

w w w. h a m e g . co m
HM303-6E/260107/ce Subject to alterations HAMEG Instruments GmbH Registered Trademark DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH Industriestr. 6 D-63533 Mainhausen Tel +49 (0) 6182 800 0 Fax +49 (0) 6182 800 100 www.hameg.com info@hameg.com

Subject to change without notice 5


General information
B
B
Please check the instrument for mechanical damage or loose C
parts immediately after unpacking. In case of damage we advise T
to contact the sender. Do not operate. A

C
List of symbols used
D F
E
Consult the manual High voltage

Important note Ground D

Positioning
STOP the instrument

As can be seen from the figures, the handle can be set into diffe-
rent positions: E
A = carrying
B = handle removal and horizontal carrying
C = horizontal operating A
D and E = operating at different angles
F = handle removal
T = shipping (handle unlocked) PUOPFGkT PUOPFGkT PUOPFGkT PUOGkT PUOPFGkT PUOPFGkT PUOPFGkT

HM507
PUOPFGkT PUOPFGkT

PUOPFGkT PUOPFGkT PUOPFGkT PUOPFGkT

Attention! PUk
PUOPFGkT

PUk PUk
PUOPFGkT

PUk PUk
PUOPFGkT

PUk

B
PUkT
HGOPFFD

PUOPFGkT

When changing the handle position, the instru- PUkT

INPUT CHI
OPK
HJ
PUkT
PUkT
HGOFFD

PUkT

PUkT
PUkT
PUkT

PUkT

INPUT CHI
OPK
HJ
PUOPFGkT

HAMEG
PUOPFGkT PUOPFGkT

INPUT CHI
OPK
HJ

ment must be placed so that it can not fall (e.g.


VBN VBN VBN
PUOPFGkT PUOPFGkT
HJKL HJKL HJKL

placed on a table). Then the handle locking knobs


STOP must be simultaneously pulled outwards and
rotated to the required position. Without pulling
the locking knobs they will latch in into the next
locking position. T

Handle mounting/dismounting
T
The handle can be removed by pulling it out further, depending on
the instrument model in position B or F.

Safety

The instrument fulfils the VDE 0411 part 1 regulations for


electrical measuring, control and laboratory instruments and Safe operation may be endangered if any of the following
was manufactured and tested accordingly. It left the factory in was noticed:
perfect safe condition. Hence it also corresponds to European in case of visible damage.
Standard EN 61010-1 resp. International Standard IEC 1010-1. in case loose parts were noticed
In order to maintain this condition and to ensure safe operation if it does not function any more.
the user is required to observe the warnings and other directions after prolonged storage under unfavourable conditions (e.g.
for use in this manual. Housing, chassis as well as all measu- like in the open or in moist atmosphere).
ring terminals are connected to safety ground of the mains. after any improper transport (e.g. insufficient packing not
All accessible metal parts were tested against the mains with conforming to the minimum standards of post, rail or transport
200 VDC. The instrument conforms to safety class I. company)

The oscilloscope may only be operated from mains outlets with a Proper operation
safety ground connector. The plug has to be installed prior to con-
necting any signals. It is prohibited to separate the safety ground Please note: This instrument is only destined for use by personnel
connection. well instructed and familiar with the dangers of electrical mea-
surements.
Most electron tubes generate X-rays; the ion dose rate of this
instrument remains well below the 36 pA/kg permitted by law. For safety reasons the oscilloscope may only be operated from
mains outlets with safety ground connector. It is prohibited to
In case safe operation may not be guaranteed do not use the in- separate the safety ground connection. The plug must be inserted
strument any more and lock it away in a secure place. prior to connecting any signals.

6 Subject to change without notice


CAT I burn in, a final functional and quality test is performed to check
all operating modes and fulfilment of specifications. The latter is
This oscilloscope is destined for measurements in circuits not performed with test equipment traceable to national measurement
connected to the mains or only indirectly. Direct measurements, standards.
i.e. with a galvanic connection to circuits corresponding to the
categories II, III, or IV are prohibited! Statutory warranty regulations apply in the country where the
HAMEG product was purchased. In case of complaints please
The measuring circuits are considered not connected to the mains contact the dealer who supplied your HAMEG product.
if a suitable isolation transformer fulfilling safety class II is used.
Measurements on the mains are also possible if suitable probes Maintenance
like current probes are used which fulfil the safety class II. The
measurement category of such probes must be checked and Clean the outer shell using a dust brush in regular intervals. Dirt can
observed. be removed from housing, handle, all metal and plastic parts using
a cloth moistened with water and 1 % detergent. Greasy dirt may
Measurement categories be removed with benzene (petroleum ether) or alcohol, there after
The measurement categories were derived corresponding to the wipe the surfaces with a dry cloth. Plastic parts should be treated
distance from the power station and the transients to be expected with an antistatic solution destined for such parts. No fluid may
hence. Transients are short, very fast voltage or current excursions enter the instrument. Do not use other cleansing agents as they
which may be periodic or not. may adversely affect the plastic or lacquered surfaces.

Measurement CAT IV: Line voltage


Measurements close to the power station, e.g. on electricity
meters The instrument has a wide range power supply from 105 to 253 V,
50 or 60 Hz 10%. There is hence no line voltage selector.
Measurement CAT III:
Measurements in the interior of buildings (power distribution instal- The line fuse is accessible on the rear panel and part of the line input
lations, mains outlets, motors which are permanently installed). connector. Prior to exchanging a fuse the line cord must be pulled
out. Exchange is only allowed if the fuse holder is undamaged, it
Measurement CAT II: can be taken out using a screwdriver put into the slot. The fuse
Measurements in circuits directly connected to the mains (house- can be pushed out of its holder and exchanged.
hold appliances, power tools etc).
The holder with the new fuse can then be pushed back in place
Environment of use. against the spring. It is prohibited to repair blown fuses or to
bridge the fuse. Any damages incurred by such measures will
The oscilloscope is destined for operation in industrial, business, void the warranty.
manufacturing, and living sites.

Environmental conditions Type of fuse:


Size 5 x 20 mm; 250V~, C;
Operating ambient temperature: 0 to + 40 degrees C. During trans-
IEC 127, Bl. III; DIN 41 662
port or storage the temperature may be 20 to +55 degrees C.
(or DIN 41 571, Bl. 3).
Cut off: slow blow (T) 0,8A.
Please note that after exposure to such temperatures or in case
of condensation proper time must be allowed until the instrument
has reached the permissible range of 0 to + 40 degrees resp. until
the condensation has evaporated before it may be turned on!
Ordinarily this will be the case after 2 hours. The oscilloscope is
destined for use in clean and dry environments. Do not operate in
dusty or chemically aggressive atmosphere or if there is danger
of explosion.

The operating position may be any, however, sufficient ventilation


must be ensured (convection cooling). Prolonged operation requires
the horizontal or inclined position.

Do not obstruct the ventilation holes!

Specifications are valid after a 20 minute warm-up period between


15 and 30 degr. C. Specifications without tolerances are average
values.
STOP

Warranty and repair

HAMEG instruments are subjected to a rigorous quality control.


Prior to shipment each instrument will be burnt in for 10 hours.
Intermittent operation will produce nearly all early failures. After

Subject to change without notice 7


Type of signal voltage

Type of signal voltage The minimum signal voltage which must be applied to the Y
input for a trace of 1div. height is 1mVpp when the Y-MAG. x5
With the HM303-6, most repetitive signals in the frequency pushbutton is depressed, the VOLTS/DIV. switch is set to
range up to at least 35MHz (-3dB) can be examined. 5mV/div., and the vernier is set to CAL by turning the fine
adjustment knob of the VOLTS/DIV. switch fully clockwise.
Sinewave signals of 50MHz are displayed with a height of However, smaller signals than this may also be displayed. The
approx. 50% (-6dB). However when examining square or pulse deflection coefficients on the input attenuators are indicated
type waveforms, attention must be paid to the harmonic in mV/div. or V/div. (peak-to-peak value).
content of such signals. The repetition frequency (fundamen-
tal frequency) of the signal must therefore be significantly The magnitude of the applied voltage is ascertained by
smaller than the upper limit frequency of the vertical amplifier. multiplying the selected deflection coefficient by the vertical
Displaying composite signals can be difficult, especially if they display height in div.
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance. To If an attenuator probe x10 is used, a further multiplication by a
obtain a well-triggered display in this case, the assistance of the factor of 10 is required to ascertain the correct voltage value.
variable holdoff and/or variable time control may be required.
Television video signals are relatively easy to trigger using the For exact amplitude measurements, the variable control on the
built-in TV-Sync-Separator (TV). For optional operation as a attenuator switch must be set to its calibrated detent CAL.
DC or AC voltage amplifier, the vertical amplifier input is When turning the variable control ccw, the sensitivity will be
provided with a DC/AC switch. The DC position should only be reduced by a factor of 2.5.
used with a series-connected attenuator probe or at very low
frequencies or if the measurement of the DC voltage content Therefore every intermediate value is possible within the 1-2-
of the signal is absolutely necessary. 5 sequence.

When displaying very low frequency pulses, the flat tops may be With direct connection to the vertical input, signals up to
sloping with AC coupling of the vertical amplifier (AC limit 400Vpp may be displayed (attenuator set to 20V/div., variable
frequency approx. 1.6 Hz for 3dB). In this case, DC operation is control to left stop).
preferred, provided the signal voltage is not superimposed on a
too high DC level. Otherwise a capacitor of adequate capacitance With the designations
must be connected to the input of the vertical amplifier with DC
coupling. This capacitor must have a sufficiently high breakdown H = display height in div.,
voltage rating. DC coupling is also recommended for the display U = signal voltage in Vpp at the vertical input,
of logic and pulse signals, especially if the pulse duty factor D = deflection coefficient in V/div. at attenuator switch,
changes constantly. Otherwise the display will move upwards or
downwards at each change. Pure direct voltages can only be the required value can be calculated from the two given
measured with DC-coupling. quantities:

Amplitude Measurements
In general electrical engineering, alternating voltage data However, these three values are not freely selectable. They
normally refers to effective values (rms = root-mean-square have to be within the following limits (trigger threshold, accuracy
value). However, for signal magnitudes and voltage designations of reading):
in oscilloscope measurements, the peak-to-peak voltage (Vpp)
value is applied. The latter corresponds to the real potential H between 0.5 and 8div., if possible 3.2 to 8div.,
difference between the most positive and most negative U between 0.5mVpp and 160Vpp,
points of a signal waveform. D between 1mV/div. and 20V/div. in 1-2-5 sequence.

If a sinusoidal waveform, displayed on the oscilloscope screen, Examples:


is to be converted into an effective (rms) value, the resulting Set deflection coefficient D = 50mV/div. 0.05V/div.,
peak-to-peak value must be divided by 2x2 = 2.83. Conversely, observed display height H = 4.6div.,
it should be observed that sinusoidal voltages indicated in Vrms required voltage U = 0.054.6 = 0.23Vpp.
(Veff) have 2.83 times the potential difference in Vpp. The
relationship between the different voltage magnitudes can be Input voltage U = 5Vpp,
seen from the following figure. set deflection coefficient D = 1V/div.,
required display height H = 5:1 = 5div.

Signal voltage U = 230Vrms22 = 651Vpp


(voltage > 160Vpp, with probe 10:1: U = 65.1Vpp),
desired display height H = min. 3.2div., max. 8div.,
max. deflection coefficient D = 65.1:3.2 = 20.3V/div.,
min. deflection coefficient D = 65.1:8 = 8.1V/div.,
adjusted deflection coefficient D = 10V/div.

The input voltage must not exceed 400V, independent


from the polarity. If an AC voltage which is superimposed on
a DC voltage is applied, the maximum peak value of both
voltages must not exceed + or 400V. So for AC voltages with
a mean value of zero volt the maximum peak to peak value
Voltage values of a sine curve is 800Vpp. If attenuator probes with higher limits are used,
Vrms = effective value; Vp= simple peak or crest value; the probes limits are valid only if the oscilloscope is set to DC
Vpp = peak-to-peak value; Vmom = momentary value. input coupling.

8 Subject to change without notice


Type of signal voltage

If DC voltages are applied under AC input coupling conditions The variable time control (identified with an arrow knob
the oscilloscope maximum input voltage value remains 400V. cap) must be in its calibrated position CAL. (arrow pointing
The attenuator consists of a resistor in the probe and the 1M horizontally to the right).
input resistor of the oscilloscope, which are disabled by the AC
input coupling capacity when AC coupling is selected. This also With the designations
applies to DC voltages with superimposed AC voltages. It also
must be noted that due to the capacitive resistance of the AC L = displayed wave length in div. of one period,
input coupling capacitor, the attenuation ratio depends on the T = time in seconds for one period,
signal frequency. For sinewave signals with frequencies higher F = recurrence frequency in Hz of the signal,
than 40Hz this influence is negligible. Tc = time coefficient in s/div. on time base switch and
the relation F = 1/T, the following equations
In the GD (ground coupling) setting, the signal path is interrupted can be stated:
directly beyond the input. This causes the attenuator to be
disabled again, but now for both DC and AC voltages.

With the above listed exceptions HAMEG 10:1 probes can be


used for DC measurements up to 600V or AC voltages (with a
mean value of zero volt) of 1200Vpp. The 100:1 probe HZ53
allows for 1200V DC or 2400Vpp for AC.
With depressed X-MAG. (x10) pushbutton the Tc value must be
It should be noted that its ACpeak value is derated at higher divided by 10.
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer However, these four values are not freely selectable. They
bridging the attenuator series resistor will break down causing have to be within the following limits:
damage to the input of the oscilloscope. However, if for
example only the residual ripple of a high voltage is to be L between 0.2 and 10div., if possible 4 to 10div.,
displayed on the oscilloscope, a normal x10 probe is sufficient. T between 0.01s and 2s,
In this case, an appropriate high voltage capacitor (approx. 22- F between 0.5Hz and 30MHz,
68nF) must be connected in series with the input tip of the Tc between 0.1s/div. and 0.2s/div. in 1-2-5 sequence
probe. (with X-MAG. (x10) in out position), and
Tc between 10ns/div. and 20ms/div. in 1-2-5 sequence
(with pushed X-MAG. (x10) pushbutton).

Examples:
Displayed wavelength L = 7div.,
set time coefficient Tc = 0.1s/div.,
required period T = 7x0.1x10-6 = 0.7s
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.

Signal period T = 1s,


set time coefficient Tc = 0.2s/div.,
required wavelength L = 1:0.2 = 5div..
Total value of input voltage
Displayed ripple wavelength L = 1div.,
The dotted line shows a voltage alternating at zero volt level. set time coefficient Tc = 10ms/div.,
If superimposed on a DC voltage, the addition of the positive required ripple freq. F = 1:(1x10x10 -3) = 100Hz.
peak and the DC voltage results in the max. voltage (DC +
ACpeak). TV-line frequency F = 15625Hz,
set time coefficient Tc = 10s/div.,
With Y-POS. control (input coupling to GD) it is possible to use required wavelength L = 1:(15 625x10-5) = 6.4div..
a horizontal graticule line as reference line for ground
potential before the measurement. It can lie below or above Sine wavelength L = min. 4div., max. 10div.,
the horizontal central line according to whether positive and/ Frequency F = 1kHz,
or negative deviations from the ground potential are to be max. time coefficient Tc = 1:(4x103) = 0.25ms/div.,
measured. min. time coefficient Tc = 1:(10x103) = 0.1ms/div.,
set time coefficient Tc = 0.2ms/div.,
Time Measurements required wavelength L = 1:(103x0.2x10-3) = 5div.

As a rule, most signals to be displayed are periodically repeating Displayed wavelength L = 0.8div.,
processes, also called periods. The number of periods per set time coefficient Tc = 0.5s/div.,
second is the repetition frequency. Depending on the time pressed X-MAG. (x10) button: Tc = 0.05s/div.,
base setting of the TIME/DIV. switch, one or several signal required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
periods or only a part of a period can be displayed. The time required period T = 1:(25x10-6) = 40ns.
coefficients are stated in s/div., ms/div. and s/div. on the
TIME/DIV.-switch. The scale is accordingly divided into three If the time is relatively short as compared with the complete
fields. signal period, an expanded time scale should always be
applied (X-MAG. x10 button pressed). In this case, the
The duration of a signal period or a part of it is determined by ascertained time values have to be divided by 10. The time
multiplying the relevant time (horizontal distance in div.) by the interval of interest can be shifted to the screen center using
time coefficient set on the TIME/DIV.-switch. the X-POS. control.

Subject to change without notice 9


Type of signal voltage

When investigating pulse or square waveforms, the critical


feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly influence
the measuring accuracy, the risetime is generally measured Connection of Test Signal
between 10% and 90% of the vertical pulse height. For
measurement adjust the Y attenuator switch with its variable Caution:
control together with the Y-POS. control so that the pulse height When connecting unknown signals to the oscilloscope
is precisely aligned with the 0% and 100% lines of the internal input, always use automatic triggering and set the DC-AC
graticule. The 10% and 90% points of the signal will now coincide input coupling switch to AC. The attenuator switch should
with the 10% and 90% graticule lines. The risetime is given by initially be set to 20V/div.
the product of the horizontal distance in div. between these
two coincidence points and the time coefficient setting. If X Sometimes the trace will disappear after an input signal has
x10 magnification is used, this product must be divided by 10. The been applied. The attenuator switch must then be turned back
fall time of a pulse can also be measured by using this method. to the left, until the vertical signal height is only 3-8div. With
a signal amplitude greater than 160Vpp, an attenuator probe
The following figure shows correct positioning of the oscilloscope must be inserted before the vertical input. If, after applying
trace for accurate risetime measurement. the signal, the trace is nearly blanked, the period of the signal
is probably substantially longer than the set value on the
TIME/DIV. switch. It should be turned to the left to an
adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-


input of the oscilloscope with a shielded test cable such as HZ
32 or HZ 34, or reduced through a x10 or x100 attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to approx.
50kHz). For higher frequencies, the signal source must be of
low impedance, i.e. matched to the characteristic resistance
of the cable (as a rule 50). Especially when transmitting
square and pulse signals, a resistor equal to the characteristic
impedance of the cable must also be connected across the
With a time coefficient of 0.2s/div. and pushed X-MAG x10 cable directly at the Y-input of the oscilloscope. When using
button the example shown in the above figure results in a a 50 cable such as the HZ 34, a 50 through termination type
measured total risetime of HZ22 is available from HAMEG.

ttot = 1.6div x 0.2s/div. / 10 = 32ns When transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become
When very fast risetimes are being measured, the risetimes of visible if the correct termination is not used. A terminating
the oscilloscope amplifier and of the attenuator probe has to be resistance is sometimes recommended with sine signals as
deducted from the measured time value. The risetime of the well. Certain amplifiers, generators or their attenuators maintain
the nominal output voltage independent of frequency only if
their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor
signal can be calculated using the following formula. HZ22 will only dissipate a maximum of 2 Watts. This power is
reached with 10 Vrms or at 28.3 Vpp with sine signal.
In this ttot is the total measured risetime, tosc is the risetime of the
oscilloscope amplifier (approx. 10ns), and tp the risetime of the If a x10 or x100 attenuator probe is used, no termination is
probe (e.g. = 2ns). If ttot is greater than 100ns, then ttot can be necessary. In this case, the connecting cable is matched
taken as the risetime of the pulse, and calculation is unnecessary. directly to the high impedance input of the oscilloscope. When
using attenuators probes, even high internal impedance sources
Calculation of the example in the figure above results in a signal are only slightly loaded (approx. 10M II 16pF or 100M II 9pF
risetime with HZ53). Therefore, if the voltage loss due to the attenuation
of the probe can be compensated by a higher amplitude setting,
tr= 322 - 102 - 22 = 30,3ns the probe should always be used. The series impedance of the
probe provides a certain amount of protection for the input of
The measurement of the rise or fall time is not limited to the the vertical amplifier. Because of their separate manufacture,
trace dimensions shown in the above diagram. It is only all attenuator probes are only partially compensated, therefore
particularly simple in this way. In principle it is possible to accurate compensation must be performed on the oscilloscope
measure in any display position and at any signal amplitude. It (see Probe compensation ).
is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness and Standard attenuator probes on the oscilloscope normally
that the horizontal distance at 10% and 90% of the amplitude reduce its bandwidth and increase the rise time. In all cases
is measured. If the edge shows rounding or overshooting, the where the oscilloscope bandwidth must be fully utilized (e.g.
100% should not be related to the peak values but to the mean for pulses with steep edges) we strongly advise using the
pulse heights. Breaks or peaks (glitches) next to the edge are probes HZ51 (x10) HZ52 (x10 HF) and HZ54 (x1 and x10). This
also not taken into account. With very severe transient can save the purchase of an oscilloscope with larger bandwidth
distortions, the rise and fall time measurement has little meaning. and has the advantage that defective components can be
For amplifiers with approximately constant group delay ordered from HAMEG and replaced by oneself. The probes
(therefore good pulse transmission performance) the following mentioned have a HF-calibration in addition to low frequency
numerical relationship between rise time tr (in ns) and bandwidth calibration adjustment. Thus a group delay correction to the
B (in MHz) applies:

10 Subject to change without notice


First Time Operation

upper limit frequency of the oscilloscope is possible with the If only a spot appears (CAUTION! CRT phosphor can be damaged),
aid of an 1MHz calibrator, e.g. HZ60. reduce the intensity immediately and check that the XY
pushbutton is in the released (out) position. If the trace is not
In fact the bandwidth and rise time of the oscilloscope are not visible, check the correct positions of all knobs and switches
noticeably changed with these probe types and the waveform (particularly AT/NM button in out position).
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscopes individual pulse response. To obtain the maximum life from the cathode-ray tube, the
minimum intensity setting necessary for the measurement in
If a x10 or x100 attenuator probe is used, DC input hand and the ambient light conditions should be used.
coupling must always be used at voltages above 400V.
With AC coupling of low frequency signals, the attenuation is Particular care is required when a single spot is displayed, as
no longer independent of frequency, pulses can show pulse a very high intensity setting may cause damage to the fluorescent
tilts. Direct voltages are suppressed but load the oscilloscope screen of the CRT. Switching the oscilloscope off and on at short
input coupling capacitor concerned. Its voltage rating is max. intervals stresses the cathode of the CRT and should therefore
400V (DC + peak AC). DC input coupling is therefore of quite be avoided.
special importance with a x100 attenuation probe which
usually has a voltage rating of max. 1200 V (DC + peak AC). A The instrument is so designed that even incorrect operation will
capacitor of corresponding capacitance and voltage rating not cause serious damage. The pushbuttons control only minor
may be connected in series with the attenuator probe input functions, and it is recommended that before commencement
for blocking DC voltage (e.g. for hum voltage measurement). of operation all pushbuttons are in the out position. After this
the pushbuttons can be operated depending upon the mode of
With all attenuator probes, the maximum AC input voltage operation required.
must be derated with frequency usually above 20kHz.
Therefore the derating curve of the attenuator probe type Trace Rotation TR
concerned must be taken into account.
In spite of Mumetal-shielding of the CRT, effects of the earths
The selection of the ground point on the test object is important magnetic field on the horizontal trace position cannot be
when displaying small signal voltages. It should always be as completely avoided. This is dependent upon the orientation of
close as possible to the measuring point. If this is not done, the oscilloscope on the place of work. A centerd trace may not
serious signal distortion may result from spurious currents align exactly with the horizontal center line of the graticule. A few
through the ground leads or chassis parts. The ground leads on degrees of misalignment can be corrected by a potentiometer
attenuator probes are also particularly critical. They should be accessible through an opening on the front panel marked TRACE
as short and thick as possible. When the attenuator probe is ROTATION.
connected to a BNC-socket, a BNC-adapter, which is often
supplied as probe accessory, should be used. In this way Probe compensation and use
ground and matching problems are eliminated.
To display an undistorted waveform on an oscilloscope, the
Hum or interference appearing in the measuring circuit probe must be matched to the individual input impedance of the
(especially when a small deflection coefficient is used) is vertical amplifier.
possibly caused by multiple grounding because equalizing
currents can flow in the shielding of the test cables (voltage For this purpose a square wave signal with a very fast rise time
drop between the protective conductor connections, caused and minimum overshoot should be used, as the sinusoidal
by external equipment connected to the mains/line, e.g. contents cover a wide frequency range. The frequency accuracy
signal generators with interference protection capacitors). and the pulse duty factor are not of such importance.

First Time Operation The built-in calibration generator provides a square wave signal
with a very fast risetime (<4ns), and switch-selectable frequencies
Before applying power to the oscilloscope it is recommended of approx. 1kHz and 1MHz from the output socket below the CRT
that the following simple procedures are performed: screen.

Check that all pushbuttons are in the out position, i.e. This signal should not be used for frequency calibration!
released.
Rotate the variable controls with arrows, i.e. TIME/DIV. The output provides 0.2Vpp 1% (tr <4ns) for x10 probes. When
variable control, CH I and CH II attenuator variable controls, the Y deflection coefficient is set to 5mV/div, the calibration
and HOLDOFF control to their calibrated detent. voltage corresponds to a vertical display of 4 divisions (x10
Set all controls with marker lines to their midrange position probe).
(marker lines pointing vertically).
The TRIG. MODE selector switch should be set to the The output sockets have an internal diameter of 4.9mm to
position uppermost (AC). accommodate the internationally accepted shielding tube
Both GD input coupling pushbutton switches for CH I and diameter of modern Probes and F-series slimline probes. Only
CH II should be set to the GD position. this type of construction ensures the extremely short ground
connections which are essential for an undistorted waveform
Switch on the oscilloscope by depressing the red POWER reproduction of non-sinusoidal high frequency signals.
pushbutton. An LED will illuminate to indicate working order.
The trace, displaying one baseline, should be visible after a Adjustment at 1kHz
short warm-up period of approx. 10 seconds. Adjust Y-POS.I
and X-POS. controls to center the baseline. Adjust INTENS The C-trimmer adjustment (low frequency) compensates the
(intensity) and FOCUS controls for medium brightness and capacitive loading on the oscilloscope input (approx. 20 pF for
optimum sharpness of the trace. The oscilloscope is now the HM303-6). By this adjustment, the capacitive division
ready for use. assumes the same ratio as the ohmic voltage divider to
ensure the same division ratio for high and low frequencies,

Subject to change without notice 11


Operating modes of the vertical amplifiers

as for DC. (For x1 probes or switchable probes set to x1, this slightly more difficult, but causes a better result. The rising
adjustment is neither required nor possible). A baseline exactly edge should be as steep as possible, with a pulse top remaining
parallel to the horizontal graticule lines is a major condition for as straight and horizontal as possible.
accurate probe adjustments. (See also Trace rotation).

Connect the probes (Types HZ51, 52, 54, or HZ36) to the CH I


input. All pushbuttons should be released (in the out position).
Set input coupling to DC, the attenuator to 5 mV/div., and TIME/
DIV. switch to 0.2 ms/div., and all variable controls to CAL.
position. Plug the probe tip into the calibrator output socket.

After completion of the HF-adjustment, the signal amplitude


displayed on the CRT screen should have the same value as
during the 1kHz adjustment.

Probes other than those mentioned above, normally have a


larger tip diameter and may not fit into the calibrator outputs.
Whilst it is not difficult for an experienced operator to build a
Approximately 2 complete waveform periods are displayed on suitable adapter, it should be pointed out that most of these
the CRT screen. Now the compensation trimmer has to be probes have a slower risetime with the effect that the total
adjusted. The location of the low frequency compensation bandwidth of scope together with probe may fall far below
trimmer can be found in the probe information sheet. Adjust that of the HM303-6. Furthermore, the HF-adjustment feature
the trimmer with the insulating screw driver provided until the is nearly always missing so that waveform distortion can not
tops of the square wave signal are exactly parallel to the be entirely excluded.
horizontal graticule lines (see 1 kHz diagram). The signal
height should then be 4 div. 0.16div. (= 4 % (oscilloscope 3% The adjustment sequence must be followed in the order
and probe 1%) ). During this adjustment, the signal edges will described, i.e. first at 1kHz, then at 1MHz. The calibrator
remain invisible. frequencies should not be used for time base calibration. The
pulse duty cycle deviates from 1:1 ratio. Prerequisites for
Adjustment at 1MHz precise and easy probe adjustments, as well as checks of
deflection coefficients, are straight horizontal pulse tops,
Probes HZ51, 52 and 54 can also be HF-compensated. They calibrated pulse amplitude, and zero-potential at the pulse
incorporate resonance de-emphasing networks (R-trimmer in base. Frequency and duty cycle are relatively uncritical. For
conjunction with capacitors) which permit probe compensation interpretation of transient response, fast pulse risetimes and
in the range of the upper frequency limit of the vertical low-impedance generator outputs are of particular importance.
oscilloscope amplifier. Only this compensative adjustment Providing these essential features, as well as switch-selectable
ensures optimum utilisation of the full bandwidth, together output-frequencies, the calibrator of the HM303-6 can, under
with constant group delay at the high frequency end, thereby certain conditions, replace expensive squarewave generators
reducing characteristic transient distortion near the leading when testing or compensating wideband-attenuators or -
edge (e.g. overshoot, rounding, ringing, holes or bumps) to an amplifiers. In such a case, the input of an appropriate circuit will
absolute minimum. be connected to the CALIBRATOR-output via a suitable probe.

Using the probes HZ51, 52 and 54, the full bandwidth of the The voltage provided at a high-impedance input (1M II 15-
HM303-6 can be utilized without risk of unwanted waveform 30pF) will correspond to the division ratio of the probe used
distortion. (x10 = 20mVpp). Suitable probes are HZ51, 52, and 54.

Prerequisite for this HF compensation is a square wave Operating modes of


generator with fast risetime (typically 4 ns), and low output the vertical amplifiers
impedance (approx. 50), providing 0.2V at a frequency of
approx. 1MHz. The calibrator output of the HM303-6 meets The vertical amplifier is set to the desired operating mode by
these requirements when the CAL. pushbutton is depressed. using the 3 pushbuttons (CH I/II, DUAL and ADD) underneath
the attenuator switches. For Mono mode all 3 buttons must be
Connect the probe to CH I input. Depress the CAL. pushbutton in their released positions; only channel I can then be operated.
for 1MHz. All other pushbuttons should be released (out The button CH I/II-TRIG.I/II must be depressed in mono mode
position). Set the CH I input coupling to DC, attenuator switch for Channel II. The internal triggering is simultaneously switched
to 5mV/div, and TIME/DIV. switch to 0.2s/div. Set all over to Channel II with this button. If the DUAL button is
variable controls to CAL. position. depressed, both channels are working. Two signals can be
displayed together in this button position (alternate mode) if the
Insert the probe tip into the output socket. A waveform will be time-base setting and the repetition frequency of the signal are
displayed on the CRT screen, with leading and trailing edges suited. This mode is not suitable for displaying very slow-
clearly visible. For the HF-adjustment now to be performed, it running processes. The display then flickers too much or it
will be necessary to observe the rising edge as well as the appears to jump. If the ADD button is depressed in addition to
upper left corner of the pulse top. The location of the high DUAL, both channels are switched over constantly at a high
frequency compensation trimmer(s) can also be found in the frequency within a sweep period (CHOP mode). Low frequency
probe information sheet. These R-trimmer(s) have to be signals below 1kHz, or with periods longer than 1ms are
adjusted such that the beginning of the pulse is as straight as then also displayed without flicker. CHOP mode is not
possible. Overshoot or excessive rounding are unacceptable. recommended for signals with higher repetition frequencies.

The adjustment is relatively easy if only one adjusting point is If only the ADD button is depressed, the signals of both
present. In case of several adjusting points the adjustment is channels are algebraically added (+I II). Whether the resulting

12 Subject to change without notice


Operating modes of the vertical amplifiers

display shows the sum or difference is dependent on the Calculation of the phase angle or the phase shift between the
phase relationship or the polarity of the signals and on the X and Y input voltages (after measuring the distances a and b
positions of the INV. (invert) button. on the screen) is quite simple with the following formula, and
a pocket calculator with trigonometric functions. Apart from
In-phase input voltages: the reading accuracy, the signal height has no influence on the
INV. button released = sum. result.
INV. button depressed = difference.

Antiphase input voltages:


INV. button released = difference.
INV. button depressed = sum.

In the ADD mode the vertical display position is dependent


upon the Y-POS. setting of both channels. The same attenuator
switch position is normally used for both channels with
algebraic addition. The following must be noted here:

Please note that the Y-POS. settings are added too but are not - Because of the periodic nature of the trigonometric
affected by the INV. pushbutton. functions, the calculation should be limited to angles 90.
However here is the advantage of the method.
Differential measurement techniques allow direct measure-
ment of the voltage drop across floating components (both - Do not use a too high test frequency. The phase shift of the
ends above ground). Two identical probes should be used for two oscilloscope amplifiers of the HM303-6 in the X-Y
both vertical inputs. In order to avoid ground loops, use a mode can exceed an angle of 3 above120 kHz.
separate ground connection and do not use the probe ground
leads or cable shields. - It cannot be seen as a matter of course from the screen
display if the test voltage leads or lags the reference
X-Y Operation voltage. A CR network before the test voltage input of the
oscilloscope can help here. The 1 M input resistance can
For X-Y operation, the pushbutton underneath the TIME/ equally serve as R here, so that only a suitable capacitor C
DIV.-knob marked XY must be depressed. The X signal is then needs to be connected in series. If the aperture width of
derived from the INPUT CH I (X). The calibration of the X the ellipse is increased (compared with C short-circuited),
signal during X-Y operation is determined by the setting then the test voltage leads the reference voltage and vice
of the Channel I input attenuator and variable control. versa. This applies only in the region up to 90 phase shift.
This means that the sensitivity ranges and input impedances Therefore C should be sufficiently large and produce only
are identical for both the X and Y axes. However, the Y-POS.I a relatively small just observable phase shift.
control is disconnected in this mode. Its function is taken over
by the X-POS. control. It is important to note that the X-MAG. Should both input voltages be missing or fail in the X-Y mode,
x10 facility, normally used for expanding the sweep, is a very bright light dot is displayed on the screen. This dot can
inoperative in the X-Y mode. burn into the phosphor at a too high brightness setting
(INTENS knob) which causes either a lasting loss of brightness,
The bandwidth of the X amplifier, is lower than the Y amplifier or in the extreme case, complete destruction of the phosphor
and the phase angle which increases with higher frequencies, at this point.
must be taken into account (please note Specifications).
Phase difference measurement
The inversion of the Y-input signal (CH II) using the INV. in DUAL mode
button is possible.
A larger phase difference between two input signals of the
Lissajous figures can be displayed in the X-Y mode for certain same frequency and shape can be measured very simply on
measuring tasks: the screen in Dual mode (DUAL button depressed). The time
base should be triggered by the reference signal (phase
- Comparing two signals of different frequency or bringing position 0).
one frequency up to the frequency of the other signal. This
also applies for whole number multiples or fractions of the The other signal can then have a leading or lagging phase
one signal frequency. angle. Alternate mode should be selected for frequencies 1
- Phase comparison between two signals of the same kHz; the Chop mode is more suitable for frequencies <1 kHz
frequency. (less flickering).

Phase comparison with Lissajous figures For greatest accuracy adjust not much more than one period
and approximately the same height of both signals on the
The following diagrams show two sine signals of the same screen. The variable controls for amplitude and time base and
frequency and amplitude with different phase angles. the LEVEL knob can also be used for this adjustment without
influence on the result.

Both base lines are set onto the horizontal graticule center line
with the Y-POS knobs before the measurement. With sinusoidal
signals, observe the zero (crossover point) transitions; the sine
peaks are less accurate. If a sine signal is noticeably distorted
by even harmonics, or if a D.C. voltage is present, AC coupling
is recommended for both channels. If it is a question of pulses
of the same shape, read off at steep edges.

Subject to change without notice 13


Operating modes of the vertical amplifiers Triggering and time base

Phase difference measurement in DUAL mode Triggering: NM (NORMAL);


with LEVEL-setting;
internal (or external) triggering.

t = horizontal spacing of the zero transitions in div.


T = horizontal spacing for one period in div.

In the example illustrated, t = 3div. and T = 10div. The phase


difference in degrees is calculated from Figure 2.
Amplitude modulated oscillation:
F = 1 MHz; f = 1 kHz;
m = 50 %; UT = 28.3 mVrms.
or expressed in radians
If the two values a and b are read from the screen, the
modulation factor is calculated from

Relatively small phase angles at not too high frequencies can be


measured more accurately in the X-Y mode with Lissajous figures.

Measurement of an amplitude modulation where

The momentary amplitude u at time t of a HF-carrier voltage, which a = UT (1+m) and b = UT (1-m).
is amplitude modulated without distortion by a sinusoidal AF
voltage, is in accordance with the equation The variable controls for amplitude and time can be set arbitrarily
in the modulation factor measurement. Their position does not
influence the result.

where Triggering and time base


UT = unmodulated carrier amplitude Time related amplitude changes on a measuring signal (AC
F = angular carrier frequency
= 2 voltage) are displayable in Yt-mode. In this mode the signal
f = modulation angular frequency
= 2 voltage deflects the beam in vertical direction while the time
m = modulation factor (i.a. 1 100%). base generator moves the beam from the left to the right of the
screen (time deflection).
The lower side frequency F-f and the upper side frequency F+f arise
because of the modulation apart from the carrier frequency F. Normally there are periodically repeating waveforms to be
Figure 1. displayed. Therefore the time base must repeat the time
deflection periodically too. To produce a stationary display, the
time base must only be triggered if the signal height and slope
condition coincide with the former time base start conditions.
A DC voltage signal can not be triggered as it is a constant signal
with no slope.

Triggering can be performed by the measuring signal itself


(internal triggering) or by an external supplied but synchronous
Amplitude and frequency spectrum for AM display (m = 50%) voltage (external triggering).

The display of the amplitude-modulated HF oscillation can be The trigger voltage should have a certain minimum amplitude.
evaluated with the oscilloscope provided the frequency spectrum This value is called the trigger threshold. It is measured with a
is inside the oscilloscope bandwidth. The time base is set so that sine signal. When the trigger voltage is taken internally from the
several cycles of the modulation frequency are visible. Strictly test signal, the trigger threshold can be stated as vertical
speaking, triggering should be external with modulation frequency display height in div., through which the time base generator
(from the AF generator or a demodulator). However, internal starts, the display is stable, and the trigger LED lights.
triggering is frequently possible with normal triggering (AT/NM
button depressed) using a suitable LEVEL setting and possibly The internal trigger threshold of the HM303-6 is given as
also using the time variable adjustment. 0.5div. When the trigger voltage is externally supplied, it can be
measured in Vpp at the TRIG. EXT. socket. Normally, the trigger
Oscilloscope setting for a signal according to figure 2: threshold may be exceeded up to a maximum factor of 20.
Depress no buttons.
Y: CH I; 20mV/div.; AC. The HM303-6 has two trigger modes, which are characterized
TIME/DIV.: 0.2ms/div. in the following.

14 Subject to change without notice


Triggering and time base

Automatic Peak (value) - Triggering However the trigger point may be varied within certain limits on
the chosen edge using the LEVEL control. The slope direction is
If the AT/NM pushbutton is in the out position AT, the sweep always related to the input signal and the non inverted display.
generator is running without test signal or external trigger voltage. .
A base line is always displayed even without a signal applied. Trigger coupling
In automatic trigger mode the sweep generator can run without The coupling mode and accordingly the frequency range of the
test signal or external trigger voltage. A base line will always be trigger signal can be changed using the TRIG. MODE selector
displayed even with no signal. With an applied AC signal the switch.
peak value triggering enables the user to select the voltage
point on the trigger signal (trigger point), by the adjustment of AC: Trigger range <20Hz to 100MHz.
the trigger LEVEL control. The control range depends on the This is the most frequently used trigger mode. The trigger
peak to peak value of the signal. This trigger mode is therefore threshold is increasing below 20Hz and above 100MHz.
called Automatic Peak (Value)- Triggering. Operation of the
scope needs only correct amplitude and time base settings, for DC: Trigger range DC to 100MHz.
a constantly visible trace. Automatic mode is recommended for DC triggering is recommended, if the signal is to be triggered
all uncomplicated measuring tasks. However, automatic with quite slow processes or if pulse signals with constantly
triggering is also the appropriate operation mode for the entry changing pulse duty factors have to be displayed.
into difficult measuring problems, e.g. when the test signal is
unknown relating to amplitude, frequency or shape. Presenting With DC- or LF-trigger coupling, always work with normal
of all parameters is now possible with automatic triggering; the triggering and LEVEL adjustment.
change to normal triggering can follow thereafter.
LF: Trigger range DC to 1.5kHz (low-pass filter).
The automatic triggering works above 20Hz. The failure of The LF position is often more suited for low-frequency
automatic triggering at frequencies below 20Hz is abrupt. signals than the DC position, because the (white) noise in
However, it is not signified by the trigger indicator LED this is still the trigger voltage is strongly suppressed. So jitter or
blinking. Break down of triggering is best recognizable at the left double-triggering of complex signals is avoidable or at least
screen edge (the start of the trace in differing display height). reduced, in particular with very low input voltages. The
trigger threshold increases above 1.5kHz.
The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20Hz. However, if the TV: The built-in active TV-Sync-Separator enables the
pulse duty factor of a square-wave signal exceeds a ratio of separation of sync pulses from the video signal. Even
100:1, switching over to normal triggering will be necessary. distorted video signals are triggered and displayed in a
Automatic triggering is practicable with internal and external stable manner.
trigger voltage.
Video signals are triggered in the automatic mode. The
Normal Triggering internal triggering is virtually independent of the display
height, but the sync pulse must exceed 0.5div. height. For
With normal triggering (AT/NM button depressed) the sweep TV sync pulse separation the TRIG. MODE switch must be
can be started by AC signals within the frequency range set to TV. The TIME/DIV.-switch selects between field
defined by the TRIG. MODE (trigger coupling) setting. (0.2s/div. - 1ms/div.) and line (0.5ms/div. - 0.1s/div.).

In the absence of an adequate trigger signal or when the trigger The slope of the leading edge of the synchronization pulse is
controls (particularly the LEVEL control) are misadjusted, no critical for the SLOPE pushbutton setting. If the displayed sync
trace is visible, i.e. the screen blanked completely. pulses are above the picture (field) contents, then the SLOPE
pushbutton ( ) must be in / position (out). In the case of sync
When using the internal normal triggering mode, it is possible pulses below the field/line, the leading edge is negative and the
to trigger at any amplitude point of a signal edge, even with very SLOPE pushbutton must therefore be depressed (to \). Since
complex signal shapes, by adjusting the LEVEL control. Its the INV. (invert) function may cause a misleading display, it must
adjusting range is directly dependent on the display height, not be activated until after correct triggering is achieved.
which should be at least 0.5div. If it is smaller than 1div., the On the 2ms/div setting field TV triggering is selected and 1 field
LEVEL adjustment needs to be operated with a sensitive is visible if a 50 fields/s signal is applied. If the holdoff control is
touch. In the external normal triggering mode, the same applies in fully ccw position, it triggers without line interlacing affects
to approx. 0.3Vpp external trigger voltage amplitude. caused by the consecutive field. More details in the video signal
become visible if the X-MAG. x10 pushbutton is depressed (in).
Other measures for triggering of very complex signals are the The X-POS. control allows to display any part of the expanded
use of the time base variable control and HOLDOFF time control, signal. The influence of the integrating network which forms a
hereinafter mentioned. trigger pulse from the vertical sync pulses may become visible
under certain conditions.
Slope
Disconnecting the trigger circuit (e.g. by rapidly pressing and
The time base generator can be started by a rising or falling edge releasing the TRIG. EXT. button) can result in triggering the
of the test signal. This is valid with automatic and with normal consecutive (odd or even) field.
triggering. The selected slope is set with the SLOPE ( )
pushbutton. The / sign (button released) means an edge, which On the 10s/div setting line TV triggering is selected and approx.
is coming from a negative potential and rising to a positive 1 lines are visible. Those lines originate randomly from the odd
potential. That has nothing to do with zero or ground potential and even fields.
and absolute voltage values. The positive slope may also lie in a
negative part of a signal. A falling edge ( \ sign) triggers, when the The sync-separator-circuit also operates with external triggering.
SLOPE ( ) pushbutton is depressed. It is important that the voltage range (0.3Vpp to 3Vpp) for external

Subject to change without notice 15


Triggering and time base

triggering should be noted. Again the correct slope setting is External triggering
critical, because the external trigger signal may not have the
same polarity or pulse edge as the test signal. This can be The internal triggering is disconnected by depressing the TRIG.
checked, if the external trigger voltage itself is displayed first EXT. button. The time base can be triggered externally via the
(with internal triggering). TRIG. EXT. socket using a 0.3Vpp to 3Vpp voltage, which is in
synchronism with the test signal. This trigger voltage may have
In most cases, the composite video signal has a high DC completely different form from the test signal voltage. Triggering
content. With constant video information (e.g. test pattern or is even possible in certain limits with whole number multiples
color bar generator), the DC content can be suppressed easily or fractions of the test frequency, but only with synchronous
by AC input coupling of the oscilloscope amplifier. With a signals.
changing picture content (e.g. normal program), DC input
coupling is recommended, because the display varies its The input impedance of the TRIG. EXT. socket is approx.
vertical position on screen with AC input coupling at each 100k II 10pF. The maximum input voltage of the input circuit
change of the picture content. The DC content can be is 100V (DC+peak AC).
compensated using the Y-POS. control so that the signal
display lies in the graticule area. Then the composite video It must be noted that a different phase angle between the
signal should not exceed a vertical height of 6div. measuring and the triggering signal may cause a display not
coinciding with the SLOPE pushbutton setting. The trigger
Line triggering (~) coupling selection can also be used in external triggering mode.
Unlike internal triggering, the lowest frequency for external
A voltage originating from mains/line (50 to 60Hz) is used for triggering is 20Hz in all trigger coupling conditions.
triggering purposes if the AT/NM and ALT pushbuttons are
depressed ( ~ symbol). When the AT/NM pushbutton is Trigger indicator
depressed the instrument is automatically set to normal
triggering. An LED on condition (above the SLOPE switch) indicates that
the trigger signal has a sufficient amplitude and the LEVEL
This trigger mode is independent of amplitude and frequency control setting is correct. This is valid with automatic and with
of the Y signal and is recommended for all mains/line normal triggering. The indication of trigger action facilitates a
synchronous signals. This also applies within certain limits to sensitive LEVEL adjustment, particularly at very low signal
whole number multiples or fractions of the line frequency. frequencies. The indication pulses are of only 100ms duration.
Line triggering can also be useful to display signals below the Thus for fast signals the TR-LED appears to glow continuously,
trigger threshold (less than 0.5div). It is therefore particularly for low repetition rate signals, the LED flashes at the repetition
suitable for measuring small ripple voltages of mains/line rate or at a display of several signal periods not only at the start
rectifiers or stray magnetic field in a circuit. In this trigger of the sweep at the left screen edge, but also at each signal
mode the SLOPE pushbutton selects the positive or negative period.
portion of the line sinewave. The LEVEL control is operative
for trigger point adjustment. In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If the
Magnetic leakage (e.g. from a power transformer) can be trigger signal frequency is <20Hz the sweep generator starts
investigated for direction and amplitude using a search or without awaiting the trigger pulse. This causes an untriggered
pick-up coil. The coil should be wound on a small former with display and a flashing trigger LED (TR).
a maximum of turns of a thin lacquered wire and connected to
a BNC connector (for scope input) via a shielded cable. Holdoff-time adjustment
Between cable and BNC center conductor a resistor of at least
100 should be series-connected (RF decoupling). Often it is If it is found that a trigger point cannot be located on extremely
advisable to shield statically the surface of the coil. However, complex signals even after repeated and careful adjustment of
no shorted turns are permissible. Maximum, minimum, and the LEVEL control, a stable display may be obtained using the
direction to the magnetic source are detectable at the HOLDOFF control. This facility varies the hold-off time between
measuring point by turning and shifting the coil. two sweep periods approx. up to the ratio 10:1. Pulses or other
signal waveforms appearing during this off period cannot
Alternate triggering trigger the time base. Particularly with burst signals or aperiodic
pulse trains of the same amplitude, the start of the sweep can
With alternate triggering (ALT button depressed) it is possible be delayed until the optimum or required moment.
to trigger two signals which are different in frequency
(asynchronous). In this case the oscilloscope must be operated A very noisy signal or a signal with a higher interfering frequency
in alternate DUAL mode with signals of sufficient height at is at times displayed double. It is possible that LEVEL adjustment
each input. To avoid trigger problems due to different DC only controls the mutual phase shift, but not the double display.
voltage components, AC input coupling for both channels is The stable single display of the signal, required for evaluation,
recommended. is easily obtainable by expanding the holdoff time. To this end
the HOLD OFF knob is slowly turned to the right, until one
The peak (value) detection which is normally present in signal is displayed. A double display is possible with certain
automatic triggering mode (AT), is automatically switched off; pulse signals, where the pulses alternately show a small
the LEVEL control becomes inoperative and the trigger difference of the peak amplitudes. Only a very exact LEVEL
threshold is set to 0 Volt. adjustment makes a single display possible. The use of the
HOLD OFF knob simplifies the right adjustment.
The internal trigger source is switched in the same way as the
channel switching after each time base sweep. After specific use the HOLD OFF control should be reset into
its calibration detent (fully ccw), otherwise the brightness of
Phase difference measurement is not possible in this trigger the display is reduced drastically. The function is shown in the
mode. following figures.

16 Subject to change without notice


Component Tester

For the component connection, two simple test leads with


4mm banana plugs, and with test prod, alligator clip or
sprung hook, are required. The test leads are connected to the
insulated socket and the adjacent ground socket beneath the
screen. The component can be connected to the test leads
either way round.

After use, to return the oscilloscope to normal operation,


release the COMP. TESTER pushbutton (off).

Test Procedure
Caution!
Do not test any component in live circuitry - remove all
grounds, power and signals connected to the component
under test. Set up Component Tester as stated above.
Connect test leads across component to be tested.
Observe oscilloscope display.

Only discharged capacitors should be tested!

Fig. 1 shows a case where the HOLD OFF knob is in the minimum Test Pattern Displays
position and various different waveforms are overlapped on the
screen, making the signal observation unsuccessful. Open circuit is indicated by a straight horizontal line.
Fig. 2 shows a case where only the desired parts of the signal are Short circuit is shown by a straight vertical line.
stably displayed.
Testing Resistors
Component Tester
If the test object has a linear ohmic resistance, both deflecting
General voltages are in the same phase. The test pattern expected from
a resistor is therefore a sloping straight line. The angle of slope
The HM303-6 has a built-in electronic Component Tester is determined by the resistance of the resistor under test. With
(COMP. TESTER), which is used for instant display of a test high values of resistance, the slope will tend towards the
pattern to indicate whether or not components are faulty. The horizontal axis, and with low values, the slope will move
COMP. TESTER can be used for quick checks of se- towards the vertical axis.
miconductors (e.g. diodes and transistors), resistors,
capacitors, and inductors. Certain tests can also be made to to 4.7k
Values of resistance from 20 can be approximately
integrated circuits. All these components can be tested in and evaluated. The determination of actual values will come with
out of circuit. experience, or by direct comparison with a component of a
known value.
The test principle is fascinatingly simple. A built-in generator
delivers a sine voltage, which is applied across the component Testing Capacitors and Inductors
under test and a built-in fixed resistor. The sine voltage across the
test object is used for the horizontal deflection, and the voltage Capacitors and inductors cause a phase difference between
drop across the resistor (i.e. current through test object) is used current and voltage, and therefore between the X and Y
for vertical deflection of the oscilloscope. The test pattern shows deflection, giving an ellipse-shaped display. The position and
a current-voltage characteristic of the test object. opening width of the ellipse will vary according to the impedance
value (at 50Hz) of the component under test.
Since this circuit operates with a frequency of 50Hz (10%)
and a voltage of approx. 7Vrms (open circuit), the indicating A horizontal ellipse indicates a high impedance or a relatively
range of the component tester is limited. The impedance of small capacitance or a relatively high inductance.
the component under test is limited to a range from 20 to
4.7k. Below and above these values, the test pattern shows A vertical ellipse indicates a small impedance or a relatively
only short-circuit or open-circuit. For the interpretation of the large capacitance or a relatively small inductance.
displayed test pattern, these limits should always be borne in
mind. However, most electronic components can normally be A sloping ellipse means that the component has a
tested without any restriction. considerable ohmic resistance in addition to its reac-
tance.
Using the Component Tester
The values of capacitance of normal or electrolytic capacitors
The component tester is switched on by depressing the from 0.1F to 1000F can be displayed and approximate values
COMP. TESTER pushbutton (on) beneath the screen. This obtained. More precise measurement can be obtained in a
makes the vertical preamplifier and the time base generator smaller range by comparing the capacitor under test with a
inoperative. A shortened horizontal trace will be observed. It capacitor of known value. Inductive components (coils,
is not necessary to disconnect scope input cables unless in- transformers) can also be tested. The determination of the
circuit measurements are to be carried out. In the COMP. value of inductance needs some experience, because inductors
TESTER mode, the only controls which can be operated are have usually a higher ohmic series resistance. However, the
INTENS, FOCUS, X-POS. and X-MAG. X10 pushbutton impedance value (at 50Hz) of an inductor in the range from 20
(must be released). All other controls and settings have no to 4.7k can easily be obtained or compared.
influence on the test operation.

Subject to change without notice 17


Component Tester

Testing Semiconductors For a transistor the figures b-e and b-c are important. The
figure e-c can vary; but a vertical line only shows short circuit
Most semiconductor devices, such as diodes, Z-diodes, condition.
transistors, FETs can be tested. The test pattern displays vary
according to the component type as shown in the figures These transistor test patterns are valid in most cases, but
below. there are exceptions (e.g. Darlington, FETs). With the COMP.
TESTER, the distinction between a P-N-P and an N-P-N
The main characteristic displayed during semiconductor testing transistor is discernible. In case of doubt, comparison with a
is the voltage dependent knee caused by the junction changing known type is helpful. It should be noted that the same socket
from the conducting state to the non conducting state. It connection (COMP. TESTER or ground) for the same terminal
should be noted that both the forward and the reverse is then absolutely necessary. A connection inversion effects
characteristic are displayed simultaneously. This is a two- a rotation of the test pattern by 180 degrees round about the
terminal test, therefore testing of transistor amplification is center point of the scope graticule.
not possible, but testing of a single junction is easily and
quickly possible. Since the test voltage applied is only very Pay attention to the usual caution with single MOS-
low, all sections of most semiconductors can be tested components relating to static discharge or frictional
without damage. However, checking the breakdown or reverse electricity!
voltage of high voltage semiconductors is not possible. More
important is testing components for open or short-circuit, In-Circuit Tests
which from experience is most frequently needed.
The test patterns show some typical displays for in-circuit
Testing Diodes tests.

Diodes normally show at least their knee in the forward


characteristic. This is not valid for some high voltage diode
types, because they contain a series connection of several
diodes. Possibly only a small portion of the knee is visible. Z-
diodes always show their forward knee and, up to approx. 9V,
their Z-breakdown, forms a second knee in the opposite
direction. A Z-breakdown voltage of more than approx. 9V can
not be displayed.

The polarity of an unknown diode can be identified by


comparison with a known diode.

Testing Transistors
Three different tests can be made to transistors: base-emitter,
base-collector and emitter-collector. The resulting test patterns
are shown below.

The basic equivalent circuit of a transistor is a Z-diode between


base and emitter and a normal diode with reverse polarity
between base and collector in series connection. There are
three different test patterns:

Caution!
During in-circuit tests make sure the circuit is dead. No
power from mains/line or battery and no signal inputs
are permitted. Remove all ground connections including
Safety Earth (pull out power plug from outlet). Remove
all measuring cables including probes between
oscilloscope and circuit under test. Otherwise both
COMP. TESTER leads are not isolated against the circuit
under test.

18 Subject to change without notice


Component Tester

In-circuit tests are possible in many cases. However, they are


not well defined. This is caused by a shunt connection of real
or complex impedances - especially if they are of relatively
low impedance at 50Hz - to the component under test, often
results differ greatly when compared with single components.
In case of doubt, one component terminal may be unsoldered.
This terminal should then be connected to the insulated
COMP. TESTER socket avoiding hum distortion of the test
pattern.

Another way is a test pattern comparison to an identical circuit


which is known to be operational (likewise without power and
any external connections). Using the test prods, identical test
points in each circuit can be checked, and a defect can be
determined quickly and easily. Possibly the device itself under
test contains a reference circuit (e.g. a second stereo channel,
push-pull amplifier, symmetrical bridge circuit), which is not
defective.

Subject to change without notice 19


Front Panel Elements HM 303-6 (Brief Description - Front View)

Element Function Element Function

POWER Turns scope on and off. SLOPE / = rising edge;


(pushbutton + LED) LED indicates operating (pushbutton switch) \ = falling edge.
condition.
TR LED lights, if sweep is triggered.
INTENS Intensity control (LED)
(knob) for trace brightness
LEVEL Adjustment of trigger level.
TRACE ROTATION To align trace with horizontal (knob)
(potentiometer; graticule line. Compensates
adjustment with influence of earths magnetic X-POS. Controls horizontal
screwdriver) field. (knob) position of trace.

FOCUS Focus control for X-MAG. x10 10:1 expansion in the X direction.
(knob) trace sharpness. (pushbutton switch) Max. resolution 10ns/div.
(inoperative in XY-mode).
Y-POS. I Controls vertical position
(knob) of channel I display. VOLTS/DIV. Channel I input attenuator.
Inoperative in X-Y mode. (12 position Selects Y input sensitivity
rotary switch) in mV/div. or V/div.
Y MAG. x5 When depressed, increasing of in 1-2-5 sequence.
(pushbutton switch) Y-sensitivity CH I 5 fold (max.
1mV/div.) VAR. GAIN Fine adjustment of Y amplitude
(knob) CH I. Increases attenuation factor
Y MAG. x5 When depressed, increasing of min. by 2.5 (left hand stop).
(pushbutton switch) Y-sensitivity CH II 5 fold (max. For amplitude measurement
1mV/div.) must be in CAL. position
(right hand stop).
Y-POS. II Controls vertical position
(knob) of channel II display. CH I/II-TRIG. I/II No button depressed: CH I only
(pushbutton switch) and triggering from channel I.
SLOPE Selects the slope of the trigger When depressed, channel II only
signal. and triggering from channel II.
(Trigger selection in DUAL
mode).

20 Subject to change without notice


Front Panel Elements HM 303-6 (Brief Description - Front View)

Element Function Element Function

DUAL Button released: XY Selects X-Y operation,


one channel only. (pushbutton switch) stops sweep.
(pushbutton Button depressed: channel I X signal via CH I.
switch) and channel II in alternating Attention!
mode. Phosphor burn-in without signal.

CHOP. DUAL and ADD buttons depressed: TRIG. EXT. Button released = internal trig.
CH I and CH II in chopped mode. (pushbutton switch) Button depressed = external
triggering, trigger signal via TRIG.
EXT. BNC socket.
ADD ADD depressed only:
(pushbutton switch) algebr. addition. INPUT CH I Channel I signal input and input
In combination with INV.: (BNC socket) for horizontal deflection in X-Y
difference. mode.
Input impedance 1M II 20pF.
VOLTS/DIV. Channel II input attenuator.
(12 position Selects Y input sensitivity AC-DC Selects input coupling of CH I
rotary switch) in mV/div. or V/div. (pushbutton switch) vertical amplifier.
in 1-2-5 sequence. DC = direct coupling
AC = coupling via capacitor.
VAR. GAIN Fine adjustment of Y amplitude
(knob) CH II. Increases attenuation GD GD = signal disconnected
factor min. by 2.5 (left hand stop). (pushbutton switch) from input, Y amplifier grounded.
For amplitude measurement
must be in CAL. position (right Connector for reference potential
hand stop). (4mm socket) (galvanically connected to earth).

TRIG. MODE Trigger selector: INPUT CH II CH II signal input.


(switch) AC: 10Hz-100MHz. (BNC socket) Input impedance 1M II 20pF.
AC-DC-LF-TV DC: DC-100MHz.
LF: DC-1.5kHz. AC-DC Selects input coupling of the CH II
TV: Triggering for frame and line. (pushbutton switch) vertical amplifier. Specs see (29).

AT/NM Button released = autom. trig. GD GD = signal disconnected


(pushbutton trace visible without input signal. (pushbutton switch) from input, Y amplifier grounded.
switch) Button depressed = normal trig.
with LEVEL adjustment. INV. Inversion of CH II display.
(pushbutton switch) In combination with ADD button
~ AT/NM and ALT pushbuttons = difference CH I, CH II.
depressed:
Internal line triggering in combi- TRIG. EXT. Input for external trigger signal.
nation with normal triggering. (BNC socket) (Pushbutton TRIG. EXT. depressed.)

ALT Triggering alternates between COMP. TESTER Switch to convert oscilloscope


(pushbutton switch) CH I and CH II in alternating (pushbutton switch) to component tester mode.
DUAL Channel mode only. Release X-MAG. X10 pushbutton.

HOLD OFF Controls holdoff-time COMP. TESTER Connectors for test leads
(knob) between sweeps. (4mm sockets) of the Component tester.
Normal position = full ccw.
0.2Vpp Calibrator square wave output
TIME/DIV. Selects time coefficients (test socket) 0,2Vpp.
(rotary switch) (speeds) of time base,
from 0.2s/div. to 0.1s/div. CALIBRATOR
1kHz/1MHz Selects calibrator frequency.
Variable Variable adjustment of time base. (pushbutton switch) Button released: approx. 1kHz,
time base control Decreases time deflection speed Button depressed: approx. 1MHz.
(center knob) at least 2.5 fold.
For time measurements
turn to right hand stop.

Subject to change without notice 21


Short Instruction for HM303-6

Switching on and initial setting

Connect instrument to power outlet, depress red POWER button. LED indicates operating condition.
Case, chassis and all measuring terminals are connected to the safety earth conductor (Safety Class I).
Do not depress any further button. TRIG. MODE selector switch to AC.
AT/NM button released, CH I input coupling switch to GD, set TIME/DIV. switch to 50s/div.
Adjust INTENS control for average brightness.
Center trace on screen using X-POS. and Y-POS.I controls. Then focus trace using FOCUS control.

Vertical amplifier mode

Channel I: CH I/II, DUAL and ADD pushbuttons in out position.


Channel II: CH I/II pushbutton depressed.
Channel I and II: DUAL pushbutton depressed. Alternate channel switching: ADD (CHOP.) pushbutton in out position.
Signals <1kHz or time coefficient 1ms/div: DUAL and ADD (CHOP.) buttons depressed.
Channel I+II (sum): depress only ADD button.
Channel I-II (difference): depress ADD and INV. pushbuttons.

Triggering mode

Select trigger mode with AT/NM pushbutton:


AT = Automatic Peak (value) Triggering <20Hz to 60MHz (out position). NM = Normal Triggering (depressed).
Trigger edge direction: select slope with SLOPE ( ) pushbutton.
Internal triggering: select channel with TRIG. I/II (CH I/II) pushbutton.
Alternating triggering (internal): DUAL and ALT. pushbuttons depressed, ADD (CHOP.) pushbutton in the out position.
External triggering: TRIG. EXT. pushbutton depressed; sync signal (0.3Vpp to 3Vpp) to TRIG. EXT. socket.
Line triggering (normal triggering): AT/NM and ALT pushbuttons depressed (~).
Select trigger coupling with TRIG. MODE selector switch. Trigger frequency ranges:
AC: >20Hz to 100MHz; DC: DC to 100MHz; LF: DC to 1.5kHz.
TV: Composite video signal with line or horizontal frequency.
TIME/DIV. 0.2s/div. - 1ms/div. = field frequency
TIME/DIV. 0.5ms/div. - 0.1s/div. = line frequency
Select edge direction with SLOPE ( ) pushbutton (sync. pulse above /, below \).
Pay attention to trigger indicator: TR LED above the SLOPE pushbutton.

Measurements

Apply test signal to the vertical input connectors of CH I and/or CH II.


Before use, calibrate attenuator probe with square wave CALIBRATOR-signal.
Switch input coupling to AC or DC.
Adjust signal to desired display height with attenuator switch (VOLTS/DIV.).
Select time coefficient on the TIME/DIV. switch.
Set trigger point with LEVEL knob for normal triggering.
Trigger complex or aperiodic signals with longer HOLD OFF-time.
Amplitude measurement with Y fine control at right stop (CAL.).
Time measurement with time fine control at right stop (CAL.).
Horizontal expansion 10 fold with X-MAG. x10 button depressed.
External horizontal deflection (X-Y mode) with XY pushbutton depressed (X input: CH I).

Component Tester

Press COMP. TESTER pushbutton (on). Connect both component terminals to COMP. TESTER jacks.
In-circuit test: Circuit under test must be disconnected from battery or power (pull out power plug), signals and ground (earth).
Remove all signal connections to HM303-6 (cable, probe), release X-MAG. X10 pushbutton, then start testing.

22 Subject to change without notice


MANUALHANDBUCHMANUEL
Oscilloscopes

Spectrum Analyzer

Power Supplies

Modular System
8000 Series

Programmable Instruments
8100 Series

authorized dealer
41-0303-06E0

www.hameg.com
Subject to change without notice