You are on page 1of 3

JOURNAL OF CLINICAL MICROBIOLOGY, Mar. 1991, p. 592-594 Vol. 29, No.

3
0095-1137/91/030592-03$02.00/0
Copyright 1991, American Society for Microbiology

Comparison of Microbiologic Assay Methods for


Hemodialysis Fluids
MATTHEW J. ARDUINO,* LEE A. BLAND, SONIA M. AGUERO, LORETTA CARSON,
MITCHELL RIDGEWAY, AND MARTIN S. FAVERO
Nosocomial Infections Laboratory Branch, Hospital Infections Program, Center for
Infectious Diseases, Centers for Disease Control, Atlanta, Georgia 30333
Received 5 October 1990/Accepted 20 December 1990

To help prevent pyrogenic reactions and bacteremia in hemodialysis patients, the Association for the
Advancement of Medical Instrumentation and the Centers for Disease Control recommend microbiologic assay

Downloaded from http://jcm.asm.org/ on November 9, 2016 by guest


of hemodialysis fluids at least monthly. Five commercially available assay systems were evaluated by using the
membrane filtration technique with standard methods agar and trypticase soy agar as the standards for
comparison. Each assay system was challenged with dialysate and reverse-osmosis water from local dialysis
centers, aqueous suspensions of eight laboratory strains of gram-negative bacilli and nontuberculous
mycobacteria, and a mixed microbial flora inoculated into reverse-osmosis water and laboratory-prepared
dialysate. Mean viable counts from triplicate samples were obtained after incubation at 37C for up to 72 h. The
efficiency of recovery varied with the specific type of microbial challenge. The SPC water sampler (Millipore
Corp., Bedford, Mass.) was the most consistent in obtaining the highest viable counts. Other commercial
systems were comparable to each other in overall performance. All assay systems tested provided an acceptable
balance between microbial recovery and required sampling time, equipment, and expertise.

Hemodialysis patients are at risk of developing endotoxin- soy agar (TSA) and standard methods agar (SMA) (2). Each
mediated pyrogenic reactions (11, 14) and bacteremia caused assay system was subjected to the same microbial challenges
by gram-negative bacilli during hemodialysis (8). These with all samples run in triplicate.
complications are believed to be a result of high bacterial Challenge organisms. Each assay system was challenged
contamination in dialysis fluids primarily caused by gram- with an aqueous suspension of one of the following bacteria:
negative water bacteria (9) or inadequately reprocessed Mycobacterium fortuitum, M. chelonae, M. chelonae-like
hemodialyzers (1, 4, 10, 12). The Association for the Ad- organisms, Flavobacterium meningosepticum, Pseudomo-
vancement of Medical Instrumentation (AAMI) and the nas paucimobilus, CDC group Ve-1, and two strains of
Centers for Disease Control (CDC) have developed micro- Methylobacterium mesophilicum. Each strain was first
biologic guidelines for dialysis fluids to protect the hemodi- grown on blood-agar plates and then isolated for purity,
alysis patients from these adverse reactions. These guide- harvested, and resuspended in an aqueous suspension to a
lines recommend that water used to prepare dialysate should final concentration of approximately 107 CFU/ml. The titer
not have viable colony counts that exceed 200 CFU/ml and of this suspension was determined, and a working stock
that water used to reprocess hemodialyzers should not solution of each test organism was made to contain 100 to
contain either bacteria >200 CFU/ml or endotoxin .1 ng/ml 1,000 CFU/ml.
(8). These recommendations suggest that dialysate should Two of the challenge solutions contained a mixture of
not have total viable colony counts which exceed 2,000 Acinetobacter calcoaceticus subsp. anitratus, Pseudomonas
CFU/ml and that dialysate should be cultured at least aeruginosa, P. cepacia, and P. putida. These challenge
monthly (3, 7). AAMI and CDC guidelines list a variety of solutions were made by first making an aqueous suspension
different sampling methods for quantitative microbiology of each organism containing approximately 107 CFU/ml.
that can be used, such as conventional laboratory proce- Titers of these suspensions were then determined, and equal
dures (i.e., spread plate and membrane filtration [MF]). In proportions of each suspension were added to 250 ml of
this study, we evaluated microbial recovery by using several sterile reverse-osmosis water to achieve a final concentra-
commercially available samplers and the MF technique. tion of 100 to 1,000 CFU of each of the four organisms per
ml. The second was prepared in the same way except that
MATERIALS AND METHODS 250 ml of filter-sterilized (0.2-,um-pore-size Nalgene dispos-
Microbial assay systems. Five commercially available mi- able filtration device) acetate dialysate was used.
crobial assay systems were evaluated: (i) the SPC sampler Hemodialysis fluids. Samples of water, acetate dialysate,
(Millipore Corp., Bedford, Mass.), (ii) the total-count water and bicarbonate dialysate were collected from three local
tester (TCWT) (Millipore Corp.), (iii) the nutrient pad kit dialysis centers to run as unknowns.
with standard media (NP) (Nalgene, Rochester, N.Y.), (iv) Assay procedure. The SPC and TCWT were tested by
NP, with triphenyltetrazolium chloride (NP+) (Nalgene), immersing the paddle portion into a 100-ml beaker contain-
and (v) aerobic count plates (ACP), formerly PetriFilm SM ing 90 ml of sample fluid for 30 s. The paddle was then
plates (3M Corp., St. Paul, Minn.). Each assay system was withdrawn from the beaker, and excess fluid was removed
compared with the standard MF technique using trypticase by gently shaking. The paddle was then replaced into its case
for incubation.
The NP and NP+ were rehydrated with 3.5 ml of sterile
*
Corresponding author. water for injection (Abbott Laboratories, Chicago, Ill.). One
592
VOL. 29, 1991 MICROBIOLOGIC ASSAY METHODS FOR HEMODIALYSIS FLUIDS 593

TABLE 1. Recovery of laboratory challenge organisms by seven assay procedures tested


Mean bacterial count (CFU/ml) by assaya
Organism
SMA TSA SPC TCWT NP NP+ ACP
Mycobacterium fortuitumb 425 307 567 558 71 143 527
Mycobacterium chelonae 111 148 83 76 73 76 121
Mycobacterium chelonae-like organism 106 81 113 107 107 120 NTC
Methylobacterium mesophilicum 1 22 0 12 9 15 26 NT
M. mesophilicum 2 36 35 39 39 43 45 NT
Pseudomonas paucimobilus 85 49 106 98 104 105 NT
GroupVe-1 145 149 212 189 112 145 NT
Flavobacterium meningosepticum 26 22 21 20 23 25 26
Reverse-osmosis water miXd 126 335 348 119 220 210 159
Dialysate miXd 182 335 150 77 199 255 205
" Assays: SMA, standard methods agar; TSA, trypticase soy agar; SPC, SPC sampler (Millipore Corp.); TCWT, Total count water tester (Millipore Corp.); NP,
nutrient pad (Nalgene Corp.); NP+, nutrient pad plus triphenyltetrazolium chloride (Nalgene Corp.); ACP, aerobic count plate (3M Corp.).

Downloaded from http://jcm.asm.org/ on November 9, 2016 by guest


b Nontuberculous mycobacteria incubated for up to 7 days.
c
NT, Not tested.
d The reverse-osmosis and dialysate mixes contain the following organisms: A. calcoaceticus, P. aeruginosa, P. cepacia, and P. paucimobilus.

milliliter of sample fluid was filtered through a 0.45-,um-pore- tested. Recovery rates varied from system to system de-
size sterile cellulose nitrate filter (Micro Filtration Systems, pending on the microorganism used in the challenge solution
Dublin, Calif.) and aseptically placed on the rehydrated pads (Table 1). SPC had the highest recovery in 40% of the
for incubation. challenges, followed by TSA (MF), which had the highest
The ACP were tested by placing 1 ml of sample fluid recovery in 20% of the challenges. ACP and NP+ both had
directly between the two pieces of medium-coated polypro- the highest recoveries in 10% of the challenges.
pylene and spreading the sample to a fixed diameter with a When dialysate from three local dialysis centers (one using
special tool supplied with the plates. bicarbonate dialysate) was used as unknowns, center-to-
MF of samples was done by previously described methods center variation was noted. Microbial recovery from acetate
(2). The 0.45-p.m-pore-size sterile cellulose nitrate filters dialysate in centers A and B was best achieved using the MF
were placed on either SMA or TSA plates. method employing SMA, although no significant difference
All assay systems were incubated at 37C for 72 h and up was seen between the standard procedure MF and the NP
to 7 days for test solutions containing mycobacteria. AAMI commercial system in samples from center A. The ACP
and CDC both recommend incubation of cultures at 37C for plates were superior for bacterial recovery from bicarbonate
48 h. Colonies were counted by using a dissecting micro- dialysate samples from center C (P < 0.03). When all 27
scope (15x objective). dialysate samples were analyzed, the SPC and ACP systems
Statistical analysis. Statistical evaluation of the different were the best overall recovery systems for bicarbonate
assay methods was done by using Waller-Duncan grouping dialysate and SMA was best for acetate dialysate (Table 2).
based on the Waller-Duncan K-ratio t test and the Student t Overall bacterial recovery from 21 water samples from
test for paired samples. centers A, B, and C was best achieved by the SPC system.
There was no significant difference between the overall
RESULTS bacterial recovery rates of SMA, TCWT, and NP+. When
analyzed separately by center, the SPC and TCWT systems
The performances of five different commercial bacterial had the highest recovery rates, followed by either SMA or
recovery systems were compared against the MF technique. the NP+ system (Table 3).
Each system was challenged with a variety of organisms
commonly found in hemodialysis fluids and with water and DISCUSSION
dialysate from three dialysis centers. All challenge solutions
were assayed in triplicate by each bacterial recovery Hemodialysis patients can be exposed to 400 to 600 liters
method. There was no significant difference in the overall of dialysis fluids per week during dialysis (6). The quality of
recovery of specific challenge organisms in all methods the water and the dialysate is important because of the
nature of the contact between the dialysate and the patient's

TABLE 2. Recovery of microorganisms from the dialysate of


three dialysis centers by six assay methods TABLE 3. Recovery of microorganisms from the reverse-osmosis
water of three dialysis centers by six assay methods
Mean bacterial count (CFU/ml) by assay"
Center Mean bacterial count (CFU/ml) by assaya
SMA SPC TCWT NP NP+ ACP Center
SMA SPC TCWT NP NP+ ACP
A (n = 3) 56 34 23 48 42 25
B (n = 3) 42 6 1 17 10 0 A (n = 9) 168 178 98 110 130 68
Cb (n = 3) 206 691 130 136 57 1,526 B (n = 9) 207 334 229 173 254 86
Meanc 101 244 51 67 36 517 C (n = 3) 2,303 3,630 2,267 1,107 1,020 703
" Assays are described in Table 1, footnote a.
Meanb 490 738 464 280 310 166
b Bicarbonate dialysate used. "Assays are described in Table 1, footnote a.
' Means based on loglo CFU per milliliter from each center. b Mean based on A+B+C (n = 21).
594 ARDUINO ET AL. J. CLIN. MICROBIOL.

blood system. High concentrations of bacteria can pose risks microbiology. Two of the tests (SPC and TCWT) involved
of bacteremia or endotoxemia to hemodialysis patients be- simply dipping a paddle containing dehydrated media into
cause of possible passage of bacterial endotoxin across the the test solution for 30 s, while the other four tests involved
membrane or because of transmembrane stimulation of rehydration of media with 1 ml of test solution. The costs per
macrophages and subsequent cytokine production by endo- test range from approximately $0.60 for ACP to approxi-
toxin or bacteria. Early studies have shown that the attack mately $3.00 for SPC and TCWT. The only additional
rates for pyrogenic reactions can be directly proportional to equipment that would be needed by the dialysis center would
the level of bacterial contamination in the dialysis fluid (7, 9). be a 37C incubator and a dissecting microscope. We con-
As a result of these studies, AAMI and CDC made the clude that all of these commercial assay systems provide an
following recommendations regarding the bacterial content acceptable balance between microbial recovery and required
of hemodialysis fluids: (i) water used to make dialysate sampling time, equipment, and expertise.
should contain <200 CFU/ml; (ii) dialysate should not con- REFERENCES
tain >2,000 CFU/ml; and (iii) water used to prepare germi-
cide for the reprocessing of dialyzers should contain <200 1. Alter, M. J., M. S. Favero, J. K. Miller, P. Coleman, and L. A.
Bland. 1988. Reuse of hemodialyzers: results of nationwide
CFU of bacteria per ml or <1 ng of endotoxin per ml (3, 12). surveillance for adverse effects. JAMA 260:2073-2076.
Testing of water and dialysate should be done at least

Downloaded from http://jcm.asm.org/ on November 9, 2016 by guest


2. American Public Health Association. 1985. Membrane filtration
monthly (3, 13). method, p. 869-870. In A. E. Greenberg, R. R. Trussell, and
A variety of bacteriologic samplers designed primarily for L. S. Clesceri (ed.), Standard methods for the examination of
water testing are commercially available. We evaluated five water and wastewater, 16th ed. American Public Health Asso-
of these sampling systems for their possible use in the ciation, Washington, D.C.
sampling of hemodialysis fluids compared with the recom- 3. Association for the Advancement of Medicine Instrumentation.
mended MF technique using TSA and SMA. The bacterio- 1981. American national standard for hemodialysis systems.
logic samplers are readily available to the dialysis facility Association for the Advancement of Medical Instrumentation,
and easy to use by dialysis personnel with little laboratory Arlington, Va.
4. Beck-Sague, C., W. R. Jarvis, L. A. Bland, M. J. Arduino, S. M.
training or equipment. All tests however, require an incuba- Aguero, and G. Verosec. 1990. Outbreak of Gram-negative
tion of culture at 37C as per AAMI guidelines (3), and a bacteremia and pyrogenic reactions in a hemodialysis center.
dissecting microscope to count viable bacterial colonies is Am. J. Nephrol. 10:397-403.
strongly recommended. In our evaluation, we found perfor- 5. Bland, L. A., M. R. Ridgeway, S. M. Aguero, L. A. Carson, and
mance of each system varied depending on the microbial M. S. Favero. 1987. Potential bacteriologic and endotoxin haz-
flora of the solution being tested. This would also explain the ards associated with liquid bicarbonate concentrate. Trans. Am.
center-to-center variation noted in the bacterial recoveries Soc. Artif. Intern. Organs 33:542-545.
from water and dialysate samples from these dialysis cen- 6. Favero, M. S. 1987. Good, clean water. How dialysis centers
treat the vital fluid. Renalife 2:14-16.
ters. 7. Favero, M. S., and N. J. Petersen. 1977. Microbiologic guide-
The SPC sampler consistently had higher microbial recov- lines for hemodialysis systems. Dial. Transplant. 6:34-36.
ery rates than the four other assay systems included in this 8. Favero, M. S., N. J. Petersen, K. M. Boyer, L. A. Carson, and
study. In overall performance, when compared to MF with W. W. Bond. 1974. Microbial contamination of renal dialysis
SMA or TSA, all of the commercial systems that were tested systems and associated health risks. Trans. Am. Soc. Artif.
produced an acceptable range of microbial recovery. Differ- Intern. Organs 20-A:175-183.
ences in the ability of each method in recovering organisms 9. Favero, M. S., N. J. Petersen, L. A. Carson, W. W. Bond, and
may be partly due to the matrix (i.e., water, acetate, or S. H. Hindman. 1975. Gram-negative water bacteria in hemodi-
bicarbonate dialysate) in which the organisms had adapted. alysis systems. Health Lab. Sci. 12:321-333.
10. Gordon, S. M., M. Tipple, L. A. Bland, and W. R. Jarvis.
For example, recovery of organisms isolated from bicarbon- Pyrogenic reactions associated with the reuse of disposable
ate dialysate will depend on the salt content of the medium hollow-fiber hemodialyzers. JAMA 260:2077-2081.
because of the presence of haloduric or halophilic organisms 11. Hindman, S. H., M. S. Favero, L. A. Carson, N. J. Petersen,
(5). In this case, ACP, which are inoculated by adding a 1-ml L. B. Schonberger, and J. T. Solano. 1975. Pyrogenic reactions
sample to the system, had the best recovery. This inocula- during hemodialysis caused by extramural endotoxin. Lancet
tion ensures that the salt content in the recovery medium ii:732-734.
remains close to that of the original matrix. 12. Lowry, P. W., C. Beck-Sague, L. A. Bland, S. M. Aguero, M. J.
However, differences in the numbers of organisms recov- Arduino, A. N. Minuth, R. A. Murray, J. M. Swenson, and
ered from system to system were generally close when W. R. Jarvis. 1990. Mycobacterium chelonae infection among
logarithmic levels were compared. Thus, any of the assay patients receiving high-flux dialysis in a hemodialysis clinic in
California. J. Infect. Dis. 161:85-90.
systems could be easily used by dialysis system personnel 13. National Kidney Foundation. 1983. Revised standards for reuse
for microbiologic monitoring of dialysis fluids. All of these of hemodialyzers. National Kidney Foundation, New York.
commercial assay systems are relatively easy to use and do 14. Raij, L., F. L. Shapiro, and A. F. Michael. 1973. Endotoxemia in
not require extensive knowledge or expertise in the field of febrile reactions during hemodialysis. Kidney Int. 4:57-60.

You might also like