You are on page 1of 28

Instruction Sheet

General Purpose pH/ORP Sensor Sensors


LIQ_MAN_ABR_3900 October 2014

General Purpose pH/ORP Sensor


Specifications Figure 1. 3900 Insertion/Submersion Sensor

Sensor Type: General purpose 3900


Measured Range:
pH AccuGlass 0-14
ORP -1500 to +1500 mV
Percent Linearity Over pH Range:
Range Linearity
0-7 97%
1-7 98%
4-7 98%
7 - 10 99%
7 - 12 97%
7 - 13 96%
Materials of ORP: Platinum
7 - 14 95%
Process Connections:
Operating Temperature: -10 to 100C (14 to 212F) Front facing: 34 and 1 MNPT
Automatic temperature compensation -10 to 100C (14
Rear facing: 1 MNPT
to 212F)
Integral Cable: 32 ft (10m) cable with integral SMART
Maximum Pressure: 790 kPa [abs] (100 psig) at 100C
preamp; 15 ft (4.7m) cable without preamp sensor
(212F) CRN rating: 60 psig up to 100C (212F).
VP8 Cable: use 24281-XX, 2.5 ft (.8m) to 100 ft (31m)
Materials of Construction: stainless steel, glass, Teflon,
polyphenylene sulfide (PPS), and EPDM Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)

OPERATIONAL NOTE
The 3900 Sensor responds to changes in pH at a minimum conductivity of 0.1mS/cm in deionized water. Sample flow rate must be
controlled to 2 gph (7.6L/hr). The offset is approximately -0.2 pH after 3 months at ambient temperature.

CAUTION: SENSOR/PROCESS APPLICATION COMPATIBILITY


The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility
is entirely the responsibility of the user.

WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is
lowered to a safe level!
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

CAUTION
The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear
the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immedi-
ately rinse with clean water.

ATEX DIRECTIVE: Special Conditions for safe use


1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of
an electrostatic charge.
2. All pH/ORP sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth.
This must be taken into consideration at installation.

Storage and 10.0 (pH buffers other than pH 4.0 and pH 10.0
can be used as long as the pH values are at least two pH
1. It is recommended that electrodes be stored in their units apart).
original shipping containers until needed. NOTE: A pH 7 buffer solution reads a mV value of approx.
2. Do not store at temperatures below -10C (14F). zero, and pH buffers read approximately 59.1 mV for
3. Electrodes should be stored with a protective cap each pH unit above or below pH 7. Check the pH buffer
containing KCl solution (PN 9210342). manufacturer specifications for millivolt values at various
4. For overnight storage, immerse the sensor in tap temperatures since it may affect the actual value of the
water or 4 pH buffer solution. buffer solution mV/pH value.
5. pH glass electrodes slowly deteriorate in storage. There 1. Immerse sensor in the first buffer solution. Allow
is no specific expiration date. However, the calibration sensor to equilibrate to the buffer temperature (to
procedure described below should be followed to avoid errors due to temperature differences between
determine that the sensor calibrates properly. the buffer solution and sensor temperature) and wait
for reading to stabilize. Value of buffer can now be
Electrode preparation acknowledged by analyzer/ transmitter.
2. Once the first buffer has been acknowledged by the
1. Remove electrode from shipping container. analyzer/transmitter, rinse the buffer solution off of
2. Remove the protective boot covering the electrode bulb. the sensor with distilled or deionized water.
3. Rinse away salt film with clean water, then shake the 3. Repeat steps 1 and 2 using the second buffer
electrode so that the internal solution fills the bulb, solution.
thus removing any air trapped there. 4. The theoretical slope value, according to the Nernst
equation for calculating pH, is approximately 59.1
Installation mV/pH. Over time the sensor will age, both in the
process and in storage, which will result in reduced
For sensor orientation, see Figure 2.
slope values. To ensure accurate readings, it is
For sensor dimensions, see Figure 3. recommended that the electrode be replaced when
For mounting options, see Figures 4-7. the slope value falls below 47 to 49 mV/pH.
For wiring, see Figures 8-41.
Recommended pH Sensor
For additional wiring information, please visit our website
at www.emersonprocess.com/raihome/liquid/wiring. Standardization
Wrap the sensor threads with six or seven (67) turns of For maximum accuracy, the sensor can be standardized
Teflon tape to prevent leakage. Do not over tighten the on-line or with a process grab sample after a buffer
sensor into its receptacle. Hand tighten the sensor, and calibration has been performed and the sensor has been
then tighten one or two (12) turns with a wrench. conditioned to the process. Standardization accounts
for the sensor junction potential and other interferences.
Two point buffer calibration Standardization will not change the sensors slope but
will simply adjust the analyzers reading to match that of
Select two stable buffer solutions, preferably pH 4.0 the known process pH.

2
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Maintenance for pH electrode Figure 2. Sensor Orientation

Electrodes should respond rapidly. Sluggishness, offsets,


and erratic readings are indicators that the electrodes
may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild
non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 1 to 5
minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass
electrode life expectancy is 100% at 25C (77F), then
it will be approximately 25% at 80C (176F), and saturated KCl/AgCl reference electrode, should be
approximately 10% at 100C (212F). within +/- 20 millivolts of the value shown in the
table below. Solution temperature must be noted to
CAUTION ensure accurate interpretation of results. The ORP
Hydrochloric acid is toxic and highly corrosive. Avoid skin value of saturated quinhydrone solution is not stable
contact, wear protective gloves. Use only in a well venti- over long periods of time. Therefore, these standards
lated area. Do not inhale fumes. In case of an accident, should be made up fresh each time they are used.
consult a doctor immediately. 4. Remove the sensor from the buffer, rinse, and install
in the process.

ORP Calibration
1. After making an electrical connection between Maintenance for ORP Electrode
the sensor and the instrument, obtain a standard
Electrodes should respond rapidly. Sluggishness, offsets,
solution of saturated quinhydrone PN R508-8OZ
and erratic readings are indicators that the electrodes
(475mV). This can also be made quite simply by
may need cleaning or replacement.
adding a few crystals of quinhydrone to either pH 4
or pH 7 buffer. Quinhydrone is only slightly soluble, 1. To remove oil deposit, clean the electrode with a mild
therefore only a few crystals will be required. non-abrasive detergent.
2. Immerse the sensor in the standard solution. Allow 2. To remove scale deposits, soak electrodes for 1 to 5
1-2 minutes for the ORP sensor to stabilize. minutes in a 5% hydrochloric acid solution.
3. Standardize the instrument to the solution value 3. ORP (metallic) electrodes should be polished with
shown in the table, right. The resulting potentials, moistened baking soda.
measured with a clean platinum electrode and

Figure 3. Sensor Dimensions


1 IN MNPT WRENCH FLATS
1.30 IN ACROSS
3/4 IN MNPT 1 IN MNPT

TEMP COMP
SOLN GND

REFERENCE
JUNCTION

pH ELECTRODE

.92
2.33
6.91
8.63

3
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 3. Sensor Dimensions


1 IN MNPT WRENCH FLATS
1.30 IN ACROSS
3/4 IN MNPT 1 IN MNPT

TEMP COMP
SOLN GND

REFERENCE
JUNCTION

pH ELECTRODE

.92
2.33
6.91
8.63

Figure 4. Typical Flow Through insertion installation using PN 2002011 Pipe Tee (sensor must be installed at
least 10 above the horizon)
Horizontal Pipe Tee (PN 2002011)
Pressure/Temperature Ratings
psig (kPa [abs]) F (C)
150 (1136) 150 (65)
128 (984) 160 (71)
102 (805) 170 (77)
80 (653) 180 (82)
57 (494) 200 (93)
48 (432) 210 (99)

Figure 5. Low Flow Cell PN 24091-00


INCH
MILLIMETER

Inlet and outlet connections are stainless steel and


take 1/4-inch OD tubing. Flow cell is polycarbonate
with 1/4-inch FNPT fittings.
Wetted materials:
Body & Nut Polyester/Polycarbonate
Fittings 316 SST
Seals Silicone
Flow cell ratings:
Temperature: 32-158F (0-70C)
Max. Pressure: 90 PSIG (721 kPa [abs])
Flow rate: 2-5 GPH (7.6-18.9 LPH)

4
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 6. Jet Spray Cleaner PN 12707-00 (sensor must be installed at least 10 above the horizon)

The Jet Spray Cleaner eliminates routine,


manual sensor maintenance by cleaning the
sensor with water or compressed air. Flow
through the cleaner can be controlled by a
solenoid valve.
NOTE: The Jet Spray Cleaner can be used with
Handrail Mounting Assembly (PN 11275-01,
not shown) or can be mounted through
conduit as shown below.

Jet Spray Cleaner shown with


3900VP pH sensor

Figure 7. Low Flow Panel PN SQP10077-LQD

Low Flow Panel Specifications


Inlet flow 3-80 gph (11.4-304 L/hr)
Inlet pressure 3 65 psig (122 549 kPa abs)*
Temperature 32 - 122F (0 - 50C)
*The minimum inlet pressure is required to open a check valve, which pre-
vents the flow cell from draining if sample flow is lost. Removing the check
valve lowers the inlet pressure requirement to a few feet of water head.

5
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

NOTE
For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or
the Manual DVD enclosed with each product.
1056, 1057, 56, 5081, 6081, 54e, and XMT: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/
1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/Liquid/Sensors/Pages/ Wiring_Diagram.aspx
1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055

Figure 8. 3900-01 and 3900VP-01 to 54epH/ORP

Figure 9. 3900-02 and 3900VP-02 to 54epH/ORP

6
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 10. Sensor Wiring 3900VP-01 gray cable to 54epH/ORP

Figure 11. Sensor Wiring 3900VP-02 gray cable to 54epH/ORP

7
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 12. 3900-01 and 3900VP-01 to Xmt

Figure 13. 3900-02 and 3900VP-02 to Xmt

8
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 14. Sensor Wiring 3900VP-01 with gray cable to XMT

Figure 15. Sensor Wiring 3900VP-02 with gray cable to Xmt

9
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 16. 3900-01 and 3900VP-01 to 1055

Figure 17. Two-Input 3900-01 and 3900VP-01 to 1055

10
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 18. 3900-02 and 3900VP-02 to 1055

Figure 19. Two-Input 3900-02 and 3900VP-02 to 1055

11
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 20. Wiring with Remote Junction Box and Preamp (PN 23555-00) to 3900-02 and 3900VP-02

Figure 21. 1055pH/pH Wiring through Remote Junction Boxes to 3900-02 and 3900VP-02

12
LIQ_MAN_ABR_3900

3900 Sensors, Preamplifier in Junction Box


For 3900-02 and 3900VP-02 sensors
Instruction Sheet

Non-preamp sensors
Blue Cable
ID (inner drain)
(No Connection

1 1 Connect to instrument 1
ground or as shown
Green ID (inner drain)
2
2 2
2 2
White White (RTD Return) 1
3
3 3
3 RTD RTN /RTN 3
White/Red White/Red (RTD Sense)
4
4 4
4 RTD SENSE /SNS 4 2

Red 5 5 Red (RTD in) 3


5 5 RTD IN 5
White/Gray
Figure 22. 3900 Sensors, Preamplifier in Junction Box

6 6 White/Gray (Reference Shield) 4


6 6 Ref SHLD / SHLD /GUARD 6

Gray 7 7 Gray (Reference in) 5


7 7 REF / REF IN 7
Blue 8 8 Blue (Solution Ground) 6
8 8 GND/SOL GND 8
Clear White/Black ( mV/pH Shield) 7
9
9 9
9 pH SHLD / SHLD / GUARD 9
Orange 10 10 Black ( mV/pH in) 8
10 10 pH/pH in 10

11 Brown (-5VDC) -5VDC /-V SNSR 9


TB 1 11 11
Green (+5 VDC) 10
12 +5VDC / +V SNSR 12

TB 2 Extension Cable PN 23646-01 or


9200273 shown Rosemount Analytical 54epH/ORP XMT
BNC intruments: 5081
Extension Cable PN 9200348 can also 56 / 1056 / 1057 /
No Connection be used; extend color scheme from
1066 / 6081
sensor or VP8 cable instead of using
shown wiring scheme.
Remote Junction Box PN 23555-00 or Sensor Head Junction Box PN 23709-00
(Both J-boxes include Preamplifier Board PN 23557-00)

13
October 2014
General Purpose pH/ORP Sensor Sensors
14
-70, 3400HTVP-70,
-01
October 2014

3900 Sensors, Preamplifier in Sensor


For 3900-01 and 3900VP-01 sensors
Preamplifier in sensor
Blue Cable

Connect to instrument 1
Green Green ground or as shown
2
General Purpose pH/ORP Sensor

White White (RTD Return) RTD RTN /RTN 3 1

White/Red White/Red (RTD Sense) RTD SENSE /SNS 4 2


Figure 23. 3900 Sensors, Preamplifier in Sensor

Red Red (RTD in)


RTD IN 5 3
White/Gray
Ref SHLD / SHLD /GUARD 6 4

Gray Gray (Reference in) 5


7
REF / REF IN
Blue Blue (Solution Ground) 8 6
GND/SOL GND
Clear Clear ( mV/pH Shield) 9 7
pH SHLD / SHLD / GUARD
Orange Orange ( mV/pH in)
pH/pH in 10 8

White/Gray (-5V) 9
-5VDC /-V SNSR 11
ID (inner drain) (+5V)
12 10
+5VDC / +V SNSR

54epH/ORP XMT
Rosemount Analytical
intruments: 56 / 1056 / 1057 / 5081
1066 / 6081
LIQ_MAN_ABR_3900
Instruction Sheet
3900 Sensors, Preamplifier in Sensor,
LIQ_MAN_ABR_3900

Junction Box used for extending cable


For 3900-01 and 3900VP-02 sensors
Instruction Sheet

Preamplifier in sensor
Blue Cable

1 1 Connect to instrument 1

Green Inner Drain ground or as shown


2
2 2
2 2
White 3 3 White (RTD Return) 1
3 3 RTD RTN /RTN 3
White/Red White/Red (RTD Sense) 2
4
4 4
4 RTD SENSE /SNS 4
Figure 23. 3900 Sensors, Preamplifier in Sensor

Red 5 5 Red (RTD in) RTD IN 3


5 5 5

6 6 White/Gray (No connection) 4


6 6 Ref SHLD / SHLD /GUARD 6
Gray
7 7 Gray (Reference in) 5
7 7 REF / REF IN 7
Blue 8 8 Blue (Solution Ground) 6
8 8 GND/SOL GND 8
Clear (Orange shield) 9 9 White/Black (mV/pH Sheild) 7
9 9 pH SHLD / SHLD / GUARD 9
Orange 10 10 Black ( mV/pH in) 8
10 10 pH/pH in 10
White/Gray Brown (-5VDC) 9
11 11
11 -5VDC /-V SNSR 11
ID (inner drain)
12 12 Green (+5 VDC) 12 10
+5VDC / +V SNSR
TB 1 TB 2
Extension Cable PN 23646-01 54epH/ORP XMT
Rosemount Analytical
or 9200273 shown. 5081
intruments: 56 / 1056 / 1057 /
Remote Junction Box PN 23550-00 (with Extension Cable PN 9200348 1066 / 6081
Extension board for point-to-point wiring) can also be used; White/gray
is replacement wire color for Brown wire.

15
October 2014
General Purpose pH/ORP Sensor Sensors
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 22. 3900-01 and 3900VP-01 to Models 5081, 4081 or 3081

Figure 23. 3900-02 and 3900VP-02 to Models 5081, 4081 or 3081

16
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 10. Wiring for 3900-01 and 3900VP-01 to 56, 1056 & 1057 pH

Figure 11. Wiring for 3900-02 and 3900VP-02 to 56, 1056 & 1057 pH

17
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 26. Sensor Wiring 3900VP-01 with gray cable to 1056/56

Figure 27. Sensor Wiring 3900VP-02 with gray cable to 1056/56

18
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 28. Sensor Wiring 3900VP-01 with gray cable to 1057

Figure 29. Sensor Wiring 3900VP-01 with gray cable to 1057

19
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 30. pH/ORP Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order)

Figure 31. pH/ORP Sensor Wiring 1066/3900VP-02 (gray cable refer to manual for recommended cable order)

20
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 32. pH/ORP Sensor Wiring 1066/3900VP-01 (refer to manual for recommended cable order)

Figure 33. pH/ORP Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order)

21
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 34. pH/ORP Sensor Wiring 1066/3900VP-02 (refer to manual for recommended cable order)

Figure 35. pH/ORP Sensor Wiring 1066/3900VP-02 (blue cable refer to manual for recommended cable order)

22
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 36. Sensor Wiring 3900VP-01 with gray cable to 5081

Figure 37. Sensor Wiring 3900VP-01 with gray cable to 5081

23
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Figure 38. 3900-01 and 3900VP-01 to 6081-P

Figure 39. 3900-02 and 3900VP-02 to 6081-P

24
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Figure 40. Sensor Wiring 3900VP-01 with gray cable to 6081

Figure 41. Sensor Wiring 3900VP-02 with gray cable to 5081

25
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900

Ordering Information
The 3900/3900VP pH/ORP Sensor is housed in a PSS can be ordered separately if an analyzer/ transmitter with
plastic body with built-in solution ground for advanced an integral preamplifier is not located within 15 ft. of the
diagnostics. A Pt-100 RTD is offered for temperature sensor. SMART preamplifier (-01) is the standard pream-
compensator. The sensor is available with an integral cable plifier option, and is compatible with models XMT, 3081,
or VP8 connector. Junction box kits with preamplifiers 5081, 6081, 54e, 1055 and 1056 analyzers/transmitters.

3900 pH/ORPSensor - New Generation general purpose sensor


Code Preamplifier Option
01 Preamplifier
02 No preamplifier

Code Measuring Electrode


10 General Purpose Low resistivity GPLR pH Glass (0 - 14pH)
12 Platinum ORP

EXAMPLE 3900 - 01 - 12

3900VP pH/ORPSensor - New Generation general purpose sensor with VP

Code Preamplifier Option


01 Preamplifier
02 No preamplifier

Code Measuring Electrode


10 General Purpose Low resistivity GPLR pH Glass (0 - 14pH)
12 Platinum ORP

EXAMPLE 3900VP - 02 - 10

26
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014

Accessories
Part Number Description

24281-00 15 ft. cable with mating VP8 connector


24281-01 25 ft. cable with mating VP8 connector
24281-02 2.5 ft. cable with mating VP8 connector
24281-03 50 ft. cable with mating VP8 connector
24281-04 100 ft. cable with mating VP8 connector
24281-05 4 ft. cable with mating VP8 connector
24281-06 10 ft. cable with mating VP8 connector
24281-07 20 ft. cable with mating VP8 connector
24281-08 30 ft. cable with mating VP8 connector
12707-00 Jet cleaner
23242-02 Insertion adapter
24091-00 Low flow cell
23555-00 Junction Box with Preamplifier for Models 54e, 3081, 4081, 5081, XMT, 1055, 1056
2002011 CPVC flow through Tee, 1-1/2 in. NPT process connection
11275-01 Sensor handrail assembly
9210012 Buffer solution, 4.01 pH, 16oz
9210013 Buffer solution, 6.86 pH, 16oz
9210014 Buffer solution, 9.18 pH, 16oz
9200273 Extension cable, gray
9200348 Extension cable, blue
SQP10077-LQD Low Flow Control Panel

27
LIQ_MAN_ABR_3900
Rev. E
October 2014

facebook.com/EmersonRosemountAnalytical
8

AnalyticExpert.com Credit Cards for U.S. Purchases Only.

twitter.com/RAIhome

youtube.com/user/RosemountAnalytical

Emerson Process Management 2014 Rosemount Analytical, Inc. All rights reserved.

2400 Barranca Parkway The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one
Irvine, CA 92606 USA of the Emerson Process Management family of companies. All other marks are the property of their
Tel: (949) 757-8500 respective owners.
Fax: (949) 474-7250
The contents of this publication are presented for information purposes only, and while effort has been
RosemountAnalytical.com made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or
implied, regarding the products or services described herein or their use or applicability. All sales are
governed by our terms and conditions, which are available on request. We reserve the right to modify or
improve the designs or specifications of our products at any time without notice.

You might also like