Professional Documents
Culture Documents
OPERATIONAL NOTE
The 3900 Sensor responds to changes in pH at a minimum conductivity of 0.1mS/cm in deionized water. Sample flow rate must be
controlled to 2 gph (7.6L/hr). The offset is approximately -0.2 pH after 3 months at ambient temperature.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is
lowered to a safe level!
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
CAUTION
The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear
the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immedi-
ately rinse with clean water.
Storage and 10.0 (pH buffers other than pH 4.0 and pH 10.0
can be used as long as the pH values are at least two pH
1. It is recommended that electrodes be stored in their units apart).
original shipping containers until needed. NOTE: A pH 7 buffer solution reads a mV value of approx.
2. Do not store at temperatures below -10C (14F). zero, and pH buffers read approximately 59.1 mV for
3. Electrodes should be stored with a protective cap each pH unit above or below pH 7. Check the pH buffer
containing KCl solution (PN 9210342). manufacturer specifications for millivolt values at various
4. For overnight storage, immerse the sensor in tap temperatures since it may affect the actual value of the
water or 4 pH buffer solution. buffer solution mV/pH value.
5. pH glass electrodes slowly deteriorate in storage. There 1. Immerse sensor in the first buffer solution. Allow
is no specific expiration date. However, the calibration sensor to equilibrate to the buffer temperature (to
procedure described below should be followed to avoid errors due to temperature differences between
determine that the sensor calibrates properly. the buffer solution and sensor temperature) and wait
for reading to stabilize. Value of buffer can now be
Electrode preparation acknowledged by analyzer/ transmitter.
2. Once the first buffer has been acknowledged by the
1. Remove electrode from shipping container. analyzer/transmitter, rinse the buffer solution off of
2. Remove the protective boot covering the electrode bulb. the sensor with distilled or deionized water.
3. Rinse away salt film with clean water, then shake the 3. Repeat steps 1 and 2 using the second buffer
electrode so that the internal solution fills the bulb, solution.
thus removing any air trapped there. 4. The theoretical slope value, according to the Nernst
equation for calculating pH, is approximately 59.1
Installation mV/pH. Over time the sensor will age, both in the
process and in storage, which will result in reduced
For sensor orientation, see Figure 2.
slope values. To ensure accurate readings, it is
For sensor dimensions, see Figure 3. recommended that the electrode be replaced when
For mounting options, see Figures 4-7. the slope value falls below 47 to 49 mV/pH.
For wiring, see Figures 8-41.
Recommended pH Sensor
For additional wiring information, please visit our website
at www.emersonprocess.com/raihome/liquid/wiring. Standardization
Wrap the sensor threads with six or seven (67) turns of For maximum accuracy, the sensor can be standardized
Teflon tape to prevent leakage. Do not over tighten the on-line or with a process grab sample after a buffer
sensor into its receptacle. Hand tighten the sensor, and calibration has been performed and the sensor has been
then tighten one or two (12) turns with a wrench. conditioned to the process. Standardization accounts
for the sensor junction potential and other interferences.
Two point buffer calibration Standardization will not change the sensors slope but
will simply adjust the analyzers reading to match that of
Select two stable buffer solutions, preferably pH 4.0 the known process pH.
2
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
ORP Calibration
1. After making an electrical connection between Maintenance for ORP Electrode
the sensor and the instrument, obtain a standard
Electrodes should respond rapidly. Sluggishness, offsets,
solution of saturated quinhydrone PN R508-8OZ
and erratic readings are indicators that the electrodes
(475mV). This can also be made quite simply by
may need cleaning or replacement.
adding a few crystals of quinhydrone to either pH 4
or pH 7 buffer. Quinhydrone is only slightly soluble, 1. To remove oil deposit, clean the electrode with a mild
therefore only a few crystals will be required. non-abrasive detergent.
2. Immerse the sensor in the standard solution. Allow 2. To remove scale deposits, soak electrodes for 1 to 5
1-2 minutes for the ORP sensor to stabilize. minutes in a 5% hydrochloric acid solution.
3. Standardize the instrument to the solution value 3. ORP (metallic) electrodes should be polished with
shown in the table, right. The resulting potentials, moistened baking soda.
measured with a clean platinum electrode and
TEMP COMP
SOLN GND
REFERENCE
JUNCTION
pH ELECTRODE
.92
2.33
6.91
8.63
3
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
TEMP COMP
SOLN GND
REFERENCE
JUNCTION
pH ELECTRODE
.92
2.33
6.91
8.63
Figure 4. Typical Flow Through insertion installation using PN 2002011 Pipe Tee (sensor must be installed at
least 10 above the horizon)
Horizontal Pipe Tee (PN 2002011)
Pressure/Temperature Ratings
psig (kPa [abs]) F (C)
150 (1136) 150 (65)
128 (984) 160 (71)
102 (805) 170 (77)
80 (653) 180 (82)
57 (494) 200 (93)
48 (432) 210 (99)
4
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
Figure 6. Jet Spray Cleaner PN 12707-00 (sensor must be installed at least 10 above the horizon)
5
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
NOTE
For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or
the Manual DVD enclosed with each product.
1056, 1057, 56, 5081, 6081, 54e, and XMT: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/
1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/Liquid/Sensors/Pages/ Wiring_Diagram.aspx
1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055
6
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
7
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
8
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
9
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
10
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
11
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
Figure 20. Wiring with Remote Junction Box and Preamp (PN 23555-00) to 3900-02 and 3900VP-02
Figure 21. 1055pH/pH Wiring through Remote Junction Boxes to 3900-02 and 3900VP-02
12
LIQ_MAN_ABR_3900
Non-preamp sensors
Blue Cable
ID (inner drain)
(No Connection
1 1 Connect to instrument 1
ground or as shown
Green ID (inner drain)
2
2 2
2 2
White White (RTD Return) 1
3
3 3
3 RTD RTN /RTN 3
White/Red White/Red (RTD Sense)
4
4 4
4 RTD SENSE /SNS 4 2
13
October 2014
General Purpose pH/ORP Sensor Sensors
14
-70, 3400HTVP-70,
-01
October 2014
Connect to instrument 1
Green Green ground or as shown
2
General Purpose pH/ORP Sensor
White/Gray (-5V) 9
-5VDC /-V SNSR 11
ID (inner drain) (+5V)
12 10
+5VDC / +V SNSR
54epH/ORP XMT
Rosemount Analytical
intruments: 56 / 1056 / 1057 / 5081
1066 / 6081
LIQ_MAN_ABR_3900
Instruction Sheet
3900 Sensors, Preamplifier in Sensor,
LIQ_MAN_ABR_3900
Preamplifier in sensor
Blue Cable
1 1 Connect to instrument 1
15
October 2014
General Purpose pH/ORP Sensor Sensors
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
16
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
Figure 10. Wiring for 3900-01 and 3900VP-01 to 56, 1056 & 1057 pH
Figure 11. Wiring for 3900-02 and 3900VP-02 to 56, 1056 & 1057 pH
17
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
18
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
19
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
Figure 30. pH/ORP Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order)
Figure 31. pH/ORP Sensor Wiring 1066/3900VP-02 (gray cable refer to manual for recommended cable order)
20
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
Figure 32. pH/ORP Sensor Wiring 1066/3900VP-01 (refer to manual for recommended cable order)
Figure 33. pH/ORP Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order)
21
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
Figure 34. pH/ORP Sensor Wiring 1066/3900VP-02 (refer to manual for recommended cable order)
Figure 35. pH/ORP Sensor Wiring 1066/3900VP-02 (blue cable refer to manual for recommended cable order)
22
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
23
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
24
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
25
General Purpose pH/ORP Sensor Instruction Sheet
October 2014 LIQ_MAN_ABR_3900
Ordering Information
The 3900/3900VP pH/ORP Sensor is housed in a PSS can be ordered separately if an analyzer/ transmitter with
plastic body with built-in solution ground for advanced an integral preamplifier is not located within 15 ft. of the
diagnostics. A Pt-100 RTD is offered for temperature sensor. SMART preamplifier (-01) is the standard pream-
compensator. The sensor is available with an integral cable plifier option, and is compatible with models XMT, 3081,
or VP8 connector. Junction box kits with preamplifiers 5081, 6081, 54e, 1055 and 1056 analyzers/transmitters.
EXAMPLE 3900 - 01 - 12
EXAMPLE 3900VP - 02 - 10
26
Instruction Sheet General Purpose pH/ORP Sensor Sensors
LIQ_MAN_ABR_3900 October 2014
Accessories
Part Number Description
27
LIQ_MAN_ABR_3900
Rev. E
October 2014
facebook.com/EmersonRosemountAnalytical
8
twitter.com/RAIhome
youtube.com/user/RosemountAnalytical
Emerson Process Management 2014 Rosemount Analytical, Inc. All rights reserved.
2400 Barranca Parkway The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one
Irvine, CA 92606 USA of the Emerson Process Management family of companies. All other marks are the property of their
Tel: (949) 757-8500 respective owners.
Fax: (949) 474-7250
The contents of this publication are presented for information purposes only, and while effort has been
RosemountAnalytical.com made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or
implied, regarding the products or services described herein or their use or applicability. All sales are
governed by our terms and conditions, which are available on request. We reserve the right to modify or
improve the designs or specifications of our products at any time without notice.