Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use
according to these instructions and regular calibration and maintenance by our factory-
trained service team ensure dependable and accurate operation to protect your investment.
Contact us about a ServiceXXL agreement tailored to your needs and budget.
www.mt.com/productregistration
so we can contact you about enhancements, updates and important notifications concern
ing your METTLER TOLEDO product.
1 Introduction..........................................................................................................4
1.1 Safety instructions.................................................................................................4
1.2 Presentation..........................................................................................................5
1.3 Commissioning................................................................................................... 11
2 Operation........................................................................................................... 15
2.1 Switching on and off ........................................................................................... 15
2.2 Simple weighing.................................................................................................. 15
2.3 Switching units.................................................................................................... 15
2.4 Zeroing / Zero point correction............................................................................... 16
2.5 Weighing with tare............................................................................................... 16
2.6 Displaying information......................................................................................... 18
2.7 Printing results.................................................................................................... 18
2.8 Average (dynamic) weighing................................................................................. 19
2.9 Working with identifications.................................................................................. 19
2.10 Cleaning............................................................................................................. 20
2.11 Verification test.................................................................................................... 20
6 Appendix...........................................................................................................54
6.1 Notice for verified instruments in EC countries.........................................................54
6.2 Tables of Geo Code values...................................................................................54
6.3 Disposal............................................................................................................. 56
6.4 Protocol printouts................................................................................................ 57
6.5 Index..................................................................................................................58
04/12 Order number 22023396A METTLER TOLEDO User manual ICS435 3
1 Introduction
Introduction
Compact scales
Avoid falling loads, shock loads as well as impacts from the side.
Depending on the connected weighing platform there are two basic versions of the ICS435
compact scales:
ICS435d-.../f ICS435 compact scale with strain gauge load cell
ICS435k-.../f ICS435 compact scale with MonoBloc load cell
The weighing platforms are available in three different sizes as shown below.
extra small weighing platform small weighing platform large weighing platform
Naming
The complete name of a compact scale also indicates the type, size and capacity of the
connected weighing platform. E.g., ICS435k-6SM/DR/M/f stands for
ICS435k ICS435 terminal, weighing platform with MonoBloc load cell
6 weighing capacity in kg
SM size of the weighing platform
DR if present: Delta Range weighing platform
M if present: weighing platform verified by the manufacturer
/f mechanical design: fixed connection
Interfaces
Default Optional
Further options
Exchangeable storage battery (IP54)
Built-in storage battery (IP65)
Bargraph
The device offers a bargraph indicating the scale capacity available.
The bargraph indicates roughly which part of the scale capacity is already occupied and
what capacity is still available.
In the example, approx. 3/4 of the scale capacity is occupied, although the applied net
weight isnt really high. The reason therefore could be a high tare weight.
The metrological data are stored in the weighing platform. The weighing terminal only
serves as indicator.
, Maximum capacity
l<>l 1 Weighing range For multi range or multi interval scales only
Scale number
1 2 3 4
1.3 Commissioning
Only weighing platforms that have been levelled precisely horizontally provide accurate
weighing results. Weights and Measures approved weighing platforms have a spirit level
to simplify levelling.
Turn the adjustable feet of the weighing platform until the spirit levels air bubble is
inside the inner circle.
CAUTION
Risk of electric shock!
Before connecting the power supply, check whether the voltage value printed on the
rating plate corresponds to your local system voltage.
Do not under any circumstances connect the device if the voltage value on the rating
plate deviates from the local system voltage.
Make sure the weighing platform has reached room temperature before switching on
the power supply.
CAUTION
No success in charging the storage battery due to low temperatures!
Do not charge the battery if the battery temperature is below 0 C (32 F). Charging
is not possible in this temperature range.
Do not operate the battery charger outside its temperature range of 0 C to 40 C
(32F to 104F).
The characteristics mentioned above are only valid if the following recommendations are
observed:
Change the battery as soon as the warning message "Lowbattery" appears and the
battery symbol starts flashing. When the message appeares you still have enough
time (at least 10minutes) to complete your current task.
For optimum battery performance operate the device with built-in storage battery at an
ambient temperature in the range of 10 C to 30 C (50 F to 86 F). This applies to
discharging as well as charging the battery.
If you plan to put the scale out of operation for a longer period, charge the battery
completely.
Even if you do not use the instrument, charge the battery at least every 3months to
avoid deep discharge.
On devices with removable battery, remove the battery prior to long term storage.
Changing battery
2.
1.
1. Unlock the battery by moving the slider to the outside and remove the discharged
battery.
2. Insert the fully charged battery and secure it by moving the slider to the inside.
With optional IP65 protection, the battery is not accessible from the outside. Please call
the METTLER TOLEDO service technician.
2.1.2 Resetting
Press and hold for approx. 5 seconds.
The device is reset and will operate with the last saved settings.
When in the Terminal menu Unit roll is set to On, the weight value can be
displayed in all available weight units by repeatedly pressing .
Zeroing corrects the influence of slight changes on the load plate or minor deviations from
the zero point.
Automatic In case of non-verified scales, the automatic zero point correction can be deactivated in
the menu or the zero range can be changed. Approved scales are set fixed to 0.5 d.
By default, the zero point of the scale is automatically corrected when the scale is unloaded.
2.5.1 Taring
Place the empty container on the scale and press .
The zero display and the symbol NET appear.
The tare weight remains stored until it is cleared.
If the symbol is displayed, i.e., the A-Clear Tare function is activated in the
menu under Scale -> Tare, the tare weight is automatically cleared as soon as the
scale is unloaded.
Prerequisite
99 The symbol is displayed, i.e., the tare function A-Clear Tare is activated in
the Scale menu.
Prerequisite
99 The symbol is displayed, i.e., the tare function Auto Tare is activated in the
Scale menu.
The weight to be tared automatically, e.g., packaging material, must be heavier than
9display steps of the scale.
Prerequisite
99 The tare function Chain tare is activated in the Scale menu.
1. Place the first container or packaging material on the scale and press .
The packaging weight is automatically saved as the tare weight, the zero display and
the symbol NET appear.
2. Weigh the weighing sample and read/print out the result.
3. Place the second container or packaging material on the scale and press
again.
The total weight on the scale is saved as the new tare weight. The zero display
appears.
4. Weigh the weighing sample in the second container and read/print the result.
5. Repeat steps 3 and 4 for other containers.
The entered tare weight is valid until a new tare weight is entered or the tare weight
is cleared.
The tare weight can be entered via barcode or SICS command as well.
Up to 5 different items for display can be configured in the menu for the info key. Depending
on the configuration in the Terminal menu, the following data can be assigned in a
free order, e.g.,
Date & Time
Weight values
Identifications
Device information
Serial numbers and software versions
On the second and third info page, system and contact information can be displayed.
1. Press .
The (first) info page is displayed.
2. Press again.
The next info screen is displayed.
3. To leave the info screens, press .
Press .
The defined data is printed or transferred to the computer.
With manual start 99 Application -> Average -> Mode > Print key (factory setting),
(factory setting) Info key or Switch key is selected in the menu.
99 Weighing sample heavier than 9 scale divisions.
With automatic start 99 Application -> Average -> Mode > Auto is selected in the menu.
99 Weighing sample heavier than 9 scale divisions.
Operator menu 1. Press and keep it pressed until Enter code appears.
2. Press again.
The menu item Terminal is displayed. Only parts of the submenu Device are
accessible.
Supervisor menu 1. Press and keep it pressed until Enter code appears.
2. Enter the password and confirm with
The first menu item Scale is highlighted.
No supervisor password has been defined when the device is first delivered. Therefore,
confirm the password inquiry with when you call up the menu for the first time.
If a password has still not been entred after a few seconds, the scale returns to the
weighing mode.
If all the settings of a menu item cannot be displayed on one page (e.g., all the info items),
just use to proceed to the hidden items.
3.2.1 Overview
After selecting SCALE 1, the following menu is available.
Level 1 Level 2 Level 3 Level 4
Identifica- Scale location
tion Scale ident.
Linear. & Linearisation
Calib.
Calibration
Auto print calib. On, Off
Disp. unit Display unit 1 g, kg, oz, lb, lb-oz, t
&res. Display unit 2 g, kg, oz, lb, lb-oz, t
Disp. resolution Off, ...
Unit roll On, Off
Zero AZM Off, 0.5d, 1d, 2d, 5d, 10d
Tare Auto tare On, Off
Chain tare On, Off
Auto clear tare On, Off
Restart On, Off
Filter Vibration Low, Medium, High
Process Universal, Dosing
Stability Fast, Standard, Precise
MinWeigh Function On, Off
FACT * Temperature Off, 1K, 2K, 3K
Time Time 1, Time 2, Time 3
Days Monday, ... Sunday Off, On
Reset Perform reset ?
3.2.7 Scale > Restart Automatic saving of zero point and tare value
Restart When the Restart function is activated, the last zero point and the tare value are saved.
After switching off/on or after a power interruption, the device continues to work with the
saved zero point and tare value.
3.2.8 Scale > Filter Adaptation of the ambient conditions and the weighing type
Vibration Adaptation to ambient conditions
Low Very steady and stable environment. The scale works very rapidly, but is very
sensitive to external influences.
Medium Normal environment. The scale operates at medium speed.
High Restless environment. The scale works more slowly, but is insensitive to external
influences.
Process Adaptation to the weighing process
Universal Universal setting for all weighing samples and normal weighing goods.
Dosing Dispensing liquid or powdery weighing samples.
Stability Adjusting the stability detector
Fast The scale operates very fast.
Standard The scale operates at medium speed.
Precise The scale operates with the greatest possible reproducibility.
The slower the scale works, the greater the reproducibility of the weighing results.
Function
On/Off Switching MinWeigh function on/off
When set to On and if the weight on the scale drops below the stored minimum
weight, will appear in the symbols and info line.
Overview
Description
Region Country specific settings
Language Selecting the language of the operator interface.
We will expand the available languages continuously.
Date format Selecting the date format
Set date
Set year Entering the year in the selected format
Set month Entering the month in the selected format
Set day Entering the day in the selected format
Time format Selecting the time format
Set time
Set hour Entering the hour in the selected format
Set minutes Entering the minutes
Time out Setting the behaviour when no action takes place on the terminal
Mode Setting the time out mode.
Off No time out setting.
Rental The scale can be used for a set time interval, e.g., when the scale is rented for a
special event like a fair or a market.
Screensaver When there is no action on the scale for the set time, the display will show the
screensaver.
Days Entering the number of days the scale will be active in rental mode.
The interfaces identify themselves. Therefore only those menu settings appear which are
relevant for the individual interface.
If no optional interface is installed, the COM2 menu will not appear.
Mode Print
Auto print
Continuous (Dialog) *
Dialog * Factory setting
External input
Toledo Cont.-weight
Second display
Digitol B
Digitol G
Printer
External input
STX
RS Type
Net Address
Load resistor
DHCP
IP address
Subnet mask
Gateway
TCP settings
Wireless settings
* for more information see Reference manual MT-SICS for ICS4xx / ICS6xx
Communication > COMx > Mode Operating mode of the serial interface
Preamble length The barcode may contain additional data ahead of the relevant data (preamble) and
Data length behind (postamble).
Enter the number of characters of preamble, (relevant) data and postamble
Postamble length
Termination char. Selecting the termination character which is used by the connected barcode scanner
Destination Selecting the item to be entered via barcode scanner
TCP Mode
Server ICS435 acting as server
Client ICS435 acting as client
Both ICS435 acting as both, server or client
Local Port Displaying/entering the local port
Remote IP Displaying/entering the remote IP address
Remote Port Displaying/entering the remote port
Connect timeout Setting timeout for connecting
Disconnect timeout Setting timeout for disconnecting
Configuring inputs
1. Select an input pin.
2. Assign an input signal to the selected input pin.
Configuring outputs
1. Select an output signal.
2. Assign an output pin.
Configuring setpoints
Enter values for the setpoints.
Configuring templates
1. Select a template.
2. Select a line.
3. Assign an item.
3.6.1 Overview
Level 1 Level 2 Level 3 Level 4
Scale test Scale Internal test Perform test?
External test Perform test?
Conf. ext. test Test weight
Weight name
Tolerance
Auto print On, Off
Keyboard test Perform test?
Display test Perform test?
Serial No. SNR scale
SNR terminal
Print setup Print menu settings?
Reset all Perform reset?
3.6.2 Description
Scale test Testing the scale
Internal test For scales with an internal test weight
Perform test? Press OK to start the test
The deviation of test weight value and actually weighed value is displayed.
External test For scales without an internal test weight
Perform test? 1. Press OK to start the test
Preload is displayed.
2. If applicable, load the preload, and press OK.
The test weight is blinking.
3. Load the requested test weight and press OK.
The deviation of test weight value and actually weighed value is displayed.
Conf. ext. test Configuring the external test weight
Test weight Setting the test weight value
Weight name Entering the test weight name
Tolerance Setting the test tolerance
Auto print When set to On, a protocol is printed for each scale test.
4.2.2 Warnings
Warnings are displayed briefly and then disappear automatically.
Example
This weighing instrument features several control functions to monitor the condition of the
device.
The METTLER TOLEDO service technician can setup and enable these functions.
This helps the user and the METTLER TOLEDO service technician to detemine how the
device is treated and what measures are needed to keep it in a good shape.
If the control functions triggers an alert, a message is shown.
You can confirm the message and continue to work with the weighing instrument. The
spanner icon lights up.
Service required
Overload problem
In case of an alert we strongly recommend calling the METTLER TOLEDO service technician
to replace parts which are at the end of lifetime,
to correct wrong settings,
to educate operators about proper handling,
to perform routine service work,
to reset the alert.
Applications
Weighing
Average weighing
The operating life during storage battery operation varies depending on the intensity of use,
the configuration and the connected scale.
The following approximate values apply with standard RS232 interface and the brightness
set to 5.
Impedance 80 W
Excitation 3.3 V
Sensitivity 2 to 3 mV/V
Max. resolution 7500 e (OIML
300,000 d (non approvable)
Min. verification interval 0.5 mV/e
ICS435d-.../f compact scales use a weighing platform with strain gauge load cell.
Approved resolution of 1 x 6,000 e (OIML, NTEP)
Non-approved resolutions up to 60,000 d
Capacity 3 kg 6 kg 15 kg 35 kg
6 lb 12 lb 30 lb 70 lb
Readability
Standard resolution: 6,000 d 0.5 g 1g 2g 5g
0.001lb 0.002 lb 0.005 lb 0.01 lb
Optional resolution: 30,000 d 0.1 g 0.2 g 0.5 g 1g
0.0002lb 0.0005lb 0.001lb 0.002 lb
Optional resolution: 60,000 d 0.05 g 0.1 g 0.2 g 0.5 g
0.0001 lb 0.0002lb 0.0005lb 0.001lb
Approved resolution: 6,000 e 0.5 g 1g 2g 5g e
0.001lb 0.002 lb 0.005 lb 0.01 lb
Repeatability (sd) 0.05 g 0.1 g 0.2 g 0.5 g
0.0001lb 0.0002lb 0.0005lb 0.001lb
Linearity 0.1 g 0.2 g 0.5 g 1g
0.0002lb 0.0005lb 0.001lb 0.002lb
Weight 5.5 kg 5.5 kg 7.7 kg 7.7 kg
12.1 lb 12.1 lb 17.0 lb 17.0 lb
Capacity 0.6 kg 3 kg 6 kg 6 kg 15 kg 35 kg
1.2 lb 6 lb 12 lb 12 lb 30 lb 70 lb
Readability
Standard resolution 0.001 g 0.01 g 0.01 g 0.1 g 0.1 g 0.1 g
0.000002lb 0.00002 lb 0.00002 lb 0.0002 lb 0.0002 lb 0.0002 lb
Approved resolution 0.01 g 0.1 g 0.1 g 1g 1g 1g
0.00002lb 0.0002 lb 0.0002 lb 0.002 lb 0.002 lb 0.002 lb
Repeatability (sd) 0.001 g 0.01 g 0.01 g 0.1 g 0.1 g 0.1 g
0.00002 lb 0.00002 lb 0.00002 lb 0.0002 lb 0.0002 lb 0.0002 lb
Linearity 0.002 g 0.02 g 0.02 g 0.2 g 0.2 g 0.3 g
0.000005 lb 0.00005lb 0.00005lb 0.0005lb 0.0005lb 0.0005lb
Weight 5.7 kg 5.7 kg 5.7 kg 5.7 kg 9.0 kg 9.0 kg
12.6 lb 12.6 lb 12.6 lb 12.6 lb 19.8 lb 19.8 lb
78 mm / 3.1"
33 mm / 1.3"
248 mm / 9.8" 164 mm / 6.5"
364 mm / 14.3"
130 mm / 5.1"
247 mm / 9.7"
160 mm / 6.3"
206 mm / 8.1"
110 - 120 mm
/ 4.3" - 4.7
/ 4.3" - 4.7
247 mm / 9.7"
Socket
Weighing instruments which are verified in two steps have no green "M" on the descriptive
plate and bear the preceding identification mark on the packing label. The second step of
the verification must be carried out by the approved METTLER TOLEDO service or Weights
and Measures authorities. Please contact your METTLER TOLEDO organisation. The first
step of the verification has been carried out at the manufacturing plant.
If national regulations in individual countries limit the period of validity of the verification, the
operator of such a weighing instrument is himself responsible for its timely re-verification.
* factory setting
Geographical latitude Geo Code value Geographical latitude Geo Code value
* factory setting
6.3 Disposal
In conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic
Equipment (WEEE), this device may not be disposed of with domestic waste. This also
applies to countries outside the EU, according to their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from
which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Straight weighing standard template Average weighing with header and identification data
Starline Date/time
Identifications
A I S
Accessories........................... 52 Identifications...... 19, 29, 31, 35 Safety instructions.............. 4, 20
Auxiliary line............................ 7 Info key..................... 10, 18, 30 Service information.......... 18, 46
Average weighing............. 19, 28 Interface settings SICS
Digital I/O......................... 40 Command set.................... 33
B Ethernet............................ 36 Identifications.................... 19
Battery RS232.............................. 35 Tare preset........................ 17
Changing.......................... 14 RS422 / RS485................. 36 Simple weighing.................... 15
Handling........................... 13 USB.................................. 37 Smart weighing counter.......... 46
Specification...................... 47 WLAN............................... 37 Spanner icon..................... 7, 46
Switching on and off............... 15
C K
Commissioning..................... 11 Keyboard........................ 10, 30 T
Compact scales....................... 5 Tare
M Automatic clearing............. 16
D Menu Automatic taring.......... 17, 26
Dimensions........................... 51 Application........................ 28 Chain tare......................... 17
Display Communication................. 33 Tare preset........................ 17
Auxiliary line..................... 31 Operation.......................... 21 Taring............................... 16
Metrological data................. 8 Password......................... 21 Templates....................... 28, 41
Symbols and info line........... 9 Scale................................ 24 Test
Weight display..................... 9 Terminal........................... 29 Display............................. 43
Disposal............................... 56 Messages............................. 44 Keyboard.......................... 43
Dynamic weighing........... 19, 28 MinWeigh......................... 9, 27 Scale................................ 42
Mode Verification........................ 20
E Dialog mode..................... 38
Environment.......................... 47 Digitol.............................. 38 V
Errors................................... 44 External input.................... 38 Verified instruments in EC
Error conditions................. 44 Print mode........................ 38 countries............................... 54
Error messages................. 45 Second Display................. 38
External input.................. 19, 39 Toledo Continuous............. 38 W
Warnings.............................. 45
G P
Geo value....................... 20, 54 Power supply.................. 13, 47 Z
Printing................................. 18 Zero
Printout examples.................. 57 Automatic zero update........ 26
Zeroing............................. 16
Zero point correction........... 16
www.mt.com/service
For more information
*22023396A*
Subject to technical changes
04/2012 Mettler-Toledo (Albstadt) GmbH
Printed in Germany
Order number 22023396A * 2 2 0 2 3 3 9 6 A *