Wire, Cable and Accessories

General Information and Selection


Different strain gage installation conditions and test
specifications often necessitate the use of different
types or sizes of leadwires. For accurate, reliable strain
measurements, it is important to use an appropriate type
of leadwire for each installation. Micro-Measurements
stocks a wide variety of wires and cables, cataloged in
tabular form on the following pages. All wires and cables
listed in the tables have been proven in the field to give
excellent sensor performance when properly used in the
specified environments. Special gage wiring problems may
require the use of wires not listed here. In such cases, our
Applications Engineering Department can recommend
appropriate wire types and can suggest suppliers.

WIRE AND CABLE CODING SYSTEM

3 26 D F V
Number of Conductors
AWG (American Wire Gauge) Wire Size

Types of Wire Construction Insulation


A: Solid copper F: Flat cable E: Etched TFE Teflon
B: Stranded copper J: Twisted cable with jacket F: Fiberglass braid
D: Tinned stranded copper S: Shielded/twisted with jacket K: Kapton (polyimide) wrap
F: Silver-plated stranded T: Twisted cable without jacket N: Nylon/polyurethane enamel
G: Nickel-clad solid copper W: Round single wire P: Polyurethane enamel
H: Solid Manganin Q: Polyimide enamel
J: Solid Balco T: TFE Teflon
V: Vinyl (PVC)
Diameter Diameter
AWG (nominal) AWG (nominal)
in (mm) in (mm)
22 0.0253 0.643 34 0.0063 0.160
Balco is a Registered Trademark of W.B. Driver Company.
26 0.0159 0.404 36 0.0050 0.127 Teflon and Kapton are Registered Trademarks of DuPont.
27 0.0142 0.361 37 0.0045 0.114
30 0.0100 0.254 42 0.0025 0.064
*Solid core wire

RIBBON WIRE CODING SYSTEM


1 G L 64 001
Number of Conductors Thickness, in mils
Alloy Width
G: NiClad copper 64: 1/64 in (0.4 mm)
K: Nichrome V L: Uninsulated ribbon 16: 1/16 in (1.6 mm)
08: 1/8 in (3.2 mm)

The Wire and Cable Coding System shown above gives according to the number of conductors. All wires and
the unique designation of each wire type for ordering cables are supplied on spools for user convenience.
purposes. The leadwire and cabling selection charts Some styles may not be continuous length.
presented on the next three pages are organized

References:
Application Note: TT-601, Techniques for Bonding Leadwires to Surfaces Experiencing High Centrifugal Forces.
Application Note: TT-604, Leadwire Attachment Techniques for Obtaining Maximum Fatigue Life of Strain Gages.
Application Note: TT-608, Techniques for Attaching Leadwires to Unbonded Strain Gages.

Document No.: 11024 For technical questions, contact www.micro-measurements.com


Revision: 08-Dec-2014 micro-measurements@vpgsensors.com 1
Wire, Cable and Accessories

General Information and Selection

SINGLE-CONDUCTOR TYPES: SOLID WIRE


Type Packaging
Description
(See Note 1) Foot (Meter)*

Solid copper wire, polyurethane enamel: General-purpose intragage


134-AWP 500 ft (150 m)
hookup wire. Useful from 100 to +300F (75 to +150C). Enamel coating
136-AWP 500 ft (150 m)
easily removed by applying heat from soldering iron.

AWP AWN Solid copper wire, nylon/polyurethane enamel: Identical in use and
127-AWN 500 ft (150 m)
specifications to Type AWP above, but with superior abrasion resistance
130-AWN 500 ft (150 m)
and slightly reduced insulation resistance at elevated temperatures.
134-AWN 500 ft (150 m)
134-AWN and 136-AWN are available in four colors; specify: R (red),
136-AWN 500 ft (150 m)
W (white), B (black), G (green).

Solid copper wire, polyimide enamel: Intragage hookup wire. Temperature


127-AWQ 500 ft (150 m)
range 452 to +428F (269 to +220C) short term. Enamel is extremely
130-AWQ 500 ft (150 m)
tough and abrasion resistant, with excellent electrical properties; generally
134-AWQ 500 ft (150 m)
removed by mechanical scraping or sanding.

Solid nickel-clad copper wire, fiberglass braid insulation: Useful from


AWQ GWF 126-GWF 100 ft (30 m)
452 to +482F (269 to +250C). Recommended for use with WK-Series
126-GWF 1000 ft (300 m)
gages when silver solder is used for lead attachment.

Solid manganin wire, nylon/polyurethane enamel: Used for bridge


137-HWN 200 ft (60 m) balance and span set in transducer circuits. Nominal resistance: 14 ohms/ft
(50 ohms/m). Temperature range: +10 to +125F (10 to +50C).

Solid Balco wire, nylon/polyurethane enamel: Used for bridge


HWN JWN temperature compensation of zero shift or span. Nominal resistance: 19
142-JWN 500 ft (150 m)
ohms/ft (65 ohms/m). Temperature coefficient of resistance: +0.25%/F
(+0.45%/C). Temperature range: +10 to +300F (10 to +150C).

SINGLE-CONDUCTOR TYPES: STRANDED WIRE


Type Packaging
Description
(See Note 1) Foot (Meter)*
Stranded tinned-copper wire, vinyl insulation: General-purpose
leadwire. Useful to +180F (+80C). Vinyl insulation becomes brittle at low
126-DWV 100 ft (30 m)
temperature; not normally used below 60F (50C). Specify red, white,
black, or green.

Stranded silver-plated copper wire, Kapton polyimide insulation: High-


performance. Recommended for unusually severe service from 452 to
126-FWK 25 ft (7.5 m)
over +600F (269 to +315C) short term. Excellent resistance to abrasion,
radiation, and outgassing in high vacuum. Treated for bondability.

DWV FWK FWT Stranded silver-plated copper wire, Teflon insulation: Wide
temperature range. Useful from 452 to +500F (269 to +260C). When
130-FWT 100 ft (30 m) bonding to Teflon-insulated wire, insulation must be treated with Tetra-Etch
compound (see Special-Purpose Materials.) Specify red, white, black, or
green.

*Some types may not be continuous length.

Note 1: Products shown in bold are RoHS compliant.

Balco is a Registered Trademark of W.B. Driver Company.


Kapton and Teflon are Registered Trademarks of DuPont.
TetraEtch is a Registered Trademark of W.L. Gore.

www.micro-measurements.com For technical questions, contact Document No.: 11024


2 micro-measurements@vpgsensors.com Revision: 08-Dec-2014
Wire, Cable and Accessories

General Information and Selection

THREE-CONDUCTOR CABLE
Type Packaging
Description
(See Note 1) Foot (Meter)*

Stranded tinned-copper wire, 3-conductor twisted cable, chrome


322-DJV 500 ft (150 m) PVC vinyl jacket, vinyl insulation: Good choice for use with EGP-Series
Embedment Strain Gages. Color-coded red/white/black.

326-DFV 100 ft (30 m) Stranded tinned-copper wire, 3-conductor flat cable, vinyl insulation:
DJV DFV 326-DFV 1000 ft (300 m) Convenient general-purpose cable. For use from 60 to +180F (50 to
330-DFV 100 ft (30 m) +80C). Flat construction requires minimum space. Color-coded red/white/
330-DFV 1000 ft (300 m) black.

326-BSV 100 ft (30 m) Stranded copper wire, 3-conductor twisted cable, PVC insulated,
326-BSV 1000 ft (300 m) braided shield: For use from 60 to 180F (50 to +80C).

Stranded tinned-copper wire, 3-conductor twisted cable, vinyl


326-DTV 100 ft (30 m) insulation: Convenient general-purpose cable for low electrical noise
326-DTV 1000 ft (300 m) pickup. For use from 60 to +180F (50 to +80C). Color-coded red/
white/black.

BSV DTV DSV Stranded tinned-copper wire, 3-conductor twisted cable, vinyl
326-DSV 100 ft (30 m) insulation, braided shield, vinyl jacket: Special-purpose cable to minimize
326-DSV 1000 ft (300 m) electrical noise interference. Useful from 60 to +180F (50 to +80C).
Color-coded red/white/ black.

Stranded silver-plated copper wire, 3-conductor flat cable, etched


330-FFE 100 ft (30 m)
Teflon insulation: For use from 85F to +392F (65C to +200C). Color-
330-FFE 1000 ft (300 m)
coded red/white/black. Insulation treated for bonding.

FFE
Stranded silver-plated copper wire, 3-conductor twisted cable, Teflon
insulation, Teflon jacket: Small, flexible. For use from 452 to +500F
330-FJT 100 ft (30 m)
(269 to +260C). Color-coded red/white/black. When bonding Teflon-
330-FJT 1000 ft (300 m)
insulated wire, insulation must be treated with Tetra-Etch compound (see
Special-Purpose Materials.)
Stranded silver-plated copper wire, 3-conductor twisted cable, etched
Teflon insulation: Small, flexible cable. For use from 452 to +500F
336-FTE 50 ft (15 m)
(269 to +260C). Color-coded red/white/black. Insulation treated for
bonding.

Stranded silver-plated copper wire, 3-conductor twisted cable, etched


330-FTE 100 ft (30 m)
Teflon insulation: For use from 452 to +500F (269 to +260C). Color-
330-FTE 500 ft (150 m)
coded red/white/black. Insulation treated for bonding.
FJT FTE GJF
Solid nickel-clad copper wire, 3-conductor twisted cable, fiberglass
326-GJF 100 ft (30 m) braid insulation and jacket: For use from 452 to +900F (269 to
326-GJF 1000 ft (300 m) +480C). Recommended for use with WK-Series gages when silver solder is
used for lead attachment. Color-coded red/white/black.

*Some types may not be continuous length.

Note 1: Products shown in bold are RoHS compliant.

Teflon is a Registered Trademark of DuPont.


TetraEtch is a Registered Trademark of W.L. Gore.

Document No.: 11024 For technical questions, contact www.micro-measurements.com


Revision: 08-Dec-2014 micro-measurements@vpgsensors.com 3
Wire, Cable and Accessories

General Information and Selection

FOUR-CONDUCTOR CABLE
Type Packaging
Description
(See Note 1) Foot (Meter)*

426-DFV 100 ft (30 m)


Stranded tinned-copper wire, 4-conductor flat cable, vinyl insulation:
426-DFV 1000 ft (300 m)
For use from 60 to +180F (50 to +80C). Conductors easily separated
430-DFV 100 ft (30 m)
for stripping and wiring. Color-coded red/white/black/green.
430-DFV 1000 ft (300 m)
DFV DSV
Stranded tinned-copper wire, 4-conductor polypropylene insulated:
422-DSV 100 ft (30 m)
Twisted shielded pairs (red/black and white/green) with a drain wire, PVC
422-DSV 1000 ft (300 m)
jacket. For use from 60 to +180F (30 to +60C).

426-BSV 100 ft (30 m) Stranded copper wire, 4-conductor twisted cable, PVC insulated
426-BSV 1000 ft (300 m) braided shield: For use from 60 to +180F (50C to +80C).

Stranded tinned-copper wire, 4-conductor twisted cable, vinyl


426-DTV 100 ft (30 m)
BSV DTV insulation: For use from 60 to +180F (50 to +80C). Color-coded red/
426-DTV 1000 ft (300 m)
white/black/green.

Stranded silver-plated copper wire, 4-conductor twisted cable, Teflon


insulation, braided shield, Teflon jacket: Small, flexible cable. For use
430-FST 100 ft (30 m) from 452 to +500F (269 to +260C). Color-coded red/white/black/
430-FST 1000 ft (300 m) green. When bonding Teflon-insulated wire, insulation must be treated with
Tetra-Etch compound (see Special-Purpose Materials, document number
11008).

FST FTT Stranded silver-plated copper wire, 4-conductor twisted cable, Teflon
insulation: Small, flexible cable. For use from 452 to +500F (269
436-FTT 100 ft (30 m)
to +260C). Color coded red, white, black, green. When bonding Teflon
436-FTT 500 ft (150 m)
insulated wire, insulation must be treated with Teflon etchant, such as
TEC-1 (see Special- Purpose Materials, document number 11008).

Stranded silver-plated copper wire, 4-conductor flat cable, Teflon


insulation: For use from 452 to +500F (269 to +260C). Color coded
426-FFT 100 ft (30 m)
red, white, black, green. When bonding Teflon insulated wire, insulation
426-FFT 500 ft (150 m)
must be treated with a Teflon etchant, such as TEC-1 (see Special-Purpose
Materials, document number 11008).
FFT

FLAT RIBBON LEAD (UNINSULATED)


Packaging
Type Description
Foot (Meter)*

Uninsulated flat ni-clad copper ribbon: 1/64 in wide x 0.001in thick (0.4
1-GL-64-001 50 ft (15 m) x 0.025 mm). For use from 452 to 900F (269 to +480C). Can be easily
soldered or spot welded.

Uninsulated Nichrome V: 1/16 in wide x 0.002 in thick (1.6 x 0.05 mm). For
1-KL-16-002 50 ft (15 m)
use from 452 to + 2000F (269 to +1100C).

Uninsulated Nichrome V: 1/8 in wide x 0.003 in thick (3.2 x 0.08 mm). For
1-KL-08-003 50 ft (15 m)
use from 452 to +2000F (269 to +1100C).

Uninsulated Nichrome V: 1/8 in wide x 0.005 in thick (3.2 x 0.127 mm). For
1-KL-08-005 50 ft (15 m)
use from 452 to +2000F (269 to +1100C).

*Some types may not be continuous length.

Note 1: Products shown in bold are RoHS compliant.

Teflon is a Registered Trademark of DuPont.


TetraEtch is a Registered Trademark of W.L. Gore.

www.micro-measurements.com For technical questions, contact Document No.: 11024


4 micro-measurements@vpgsensors.com Revision: 08-Dec-2014
Wire, Cable and Accessories

General Information and Selection

HST-1 HEAT-SHRINKABLE WIRE SPLICE SEALANT

Fast, easy-to-use method for protecting wire splice connections.


Constructed of irradiated polyolefin plastic tubing with a heat-
flowable inner liner sealant. Forms an immediate and tight seal
to splice connection at a shrink temperature of +275F (+135C).
Inside diameter before heating is 0.125 in (3.2 mm); after heating,
0.023 in (0.6 mm). Large range of shrinkage allows use with
leadwire insulation diameters from 0.03 to 0.11 in (0.75 to 2.8
mm). The operating temperature range is 65 to +230F (55 to
+110C). Package of eight 6-in (150-mm) lengths.

THERMAL WIRE STRIPPER

The ease and simplicity of operation of the Thermal Wire Stripper


make it ideal for most strain gage leadwire stripping. The variable
heat control allows stripping of all thermoplastic insulations,
including Teflon, in sizes No. 18 to No. 36 AWG (1 to 0.1 mm
diameter). The foot switch and tweezer hand- piece give excellent
operator control over the stripping operation. Includes power
unit and foot switch, both with 3-wire NEMA plugs, and tweezer
handpiece.
WTS-1: 110Vac
WTS-2: 220Vac
WTS-A Replacement Elements: Set of two.

Teflon is a Registered Trademark of DuPont.

Document No.: 11024 For technical questions, contact www.micro-measurements.com


Revision: 08-Dec-2014 micro-measurements@vpgsensors.com 5
Legal Disclaimer Notice
Vishay Precision Group, Inc.

Disclaimer
ALL PRODUCTS, PRODUCT SPECIFICATIONS AND DATA ARE SUBJECT TO CHANGE WITHOUT NOTICE.

Vishay Precision Group, Inc., its affiliates, agents, and employees, and all persons acting on its or their behalf
(collectively, VPG), disclaim any and all liability for any errors, inaccuracies or incompleteness contained herein or in
any other disclosure relating to any product.

The product specifications do not expand or otherwise modify VPGs terms and conditions of purchase, including but
not limited to, the warranty expressed therein.

VPG makes no warranty, representation or guarantee other than as set forth in the terms and conditions of purchase.
To the maximum extent permitted by applicable law, VPG disclaims (i) any and all liability arising out of the
application or use of any product, (ii) any and all liability, including without limitation special, consequential or
incidental damages, and (iii) any and all implied warranties, including warranties of fitness for particular purpose,
non-infringement and merchantability.

Information provided in datasheets and/or specifications may vary from actual results in different applications and
performance may vary over time. Statements regarding the suitability of products for certain types of applications
are based on VPGs knowledge of typical requirements that are often placed on VPG products. It is the customers
responsibility to validate that a particular product with the properties described in the product specification is suitable for
use in a particular application. You should ensure you have the current version of the relevant information by contacting
VPG prior to performing installation or use of the product, such as on our website at vpgsensors.com.

No license, express, implied, or otherwise, to any intellectual property rights is granted by this document, or by any
conduct of VPG.

The products shown herein are not designed for use in life-saving or life-sustaining applications unless otherwise
expressly indicated. Customers using or selling VPG products not expressly indicated for use in such applications do
so entirely at their own risk and agree to fully indemnify VPG for any damages arising or resulting from such use or sale.
Please contact authorized VPG personnel to obtain written terms and conditions regarding products designed for such
applications.

Product names and markings noted herein may be trademarks of their respective owners.

Copyright Vishay Precision Group, Inc., 2014. All rights reserved.

Document No.: 63999 www.vpgsensors.com


Revision: 15-Jul-2014 1
Micro-Measurements

Strain Gages and Instruments Application Note TT-609

Strain Gage Soldering Techniques

Introduction temperature to suit varying installation conditions and


requirements. The temperature must be adjusted, of
T he most com mon method of mak i ng ele ctr ical course, to accommodate the melting points of the different
connections in strain gage circuits is by means of soft solders commonly used for strain gage connections, but
solders, in wire form. Other methods, such as spot also to allow for environmental conditions such as drafts
welding, brazing, compression bonding, paste solders, or outdoor soldering in cold weather. Moreover, the
and conductive epoxies, are also available, but find only temperature controller should be carefully designed to
limited application. Solders have many advantages for ensure that it does not generate electrical noise that could
strain gage use they are low in cost, readily available in adversely affect nearby measuring instruments when both
various alloy compositions to provide a range of melting are in use.
temperatures, and are easily obtained in the form of either
solid wire or wire with a core of flux. They are convenient Design of the soldering pencil also requires special
to use, and offer an excellent combination of electrical and consideration. It should be light in weight, with a very
mechanical properties. flexible power cord, and with the gripping area thermally
insulated from the heating element. These characteristics
Although soldering is basically a simple procedure, contribute to the comfort, ease, and precision of soldering,
it must be done with appropriate tools, supplies, and and minimize operator fatigue during long periods of
techniques to assure accurate strain measurement. This is use. The soldering tip itself should be of the flat, chisel, or
particularly true when test requirements are severe in the screwdriver type. Pointed tips should not be used, because
sense of approaching the limits of the strain gage circuit they tend to draw solder away from the work area, and thus
capabilities; e.g., long-term stability, high-elongation make it more difficult to achieve a proper joint. In contrast,
measurements, fatigue endurance, etc. Use of improper flat tips act to confine the solder, while offering greater
materials or techniques can significantly degrade strain surface area for better heat transfer and more effective
gage performance. soldering, generally.
The purpose of this Application Note is to outline Micro-Measurements soldering units incorporate all of the
recommended procedures and materials for attaching above features and a number of others, designed to help the
leadwires to strain gage solder tabs or to bonded user easily make consistent, reliable solder joints. These
printed-circuit terminals. These reliable, experience- soldering units are widely used by professional strain gage
proven methods are based on the use of a professional installers everywhere, in both stress analysis laboratories
quality soldering station, in conjunction with Micro- and in transducer manufacture.
Measurements solders and installation accessories.
Solder Selection
Soldering Station and Pencil
The Micro-Measurements Division stocks a broad range
For precision soldering of strain gages, it is always necessary of solder types to meet various installation and test
to use a temperature- or power-controlled soldering station requirem ents.While solders are sometimes selected to
that provides low voltage and adjustable temperature to provide specific electrical or mechanical properties, the
the soldering iron tip. An unregulated soldering iron, most common basis for selection is simply the melting-
connected directly to the power line, is not ordinarily temperature range. Low-melting-point solders, for
Tech Note

suitable for strain gage use because the tip temperature is example, are generally used for strain gage installations
apt to be far too high. This tends to oxidize the tip, and to on nonmetallic test parts to avoid damaging the gage,
instantly vaporize the flux, making soldering much more bonding adhesive, or test material due to overheating. In
difficult. In addition, the unnecessarily high temperature contrast, high-temperature solders are normally selected
may damage the strain gage, the bonding adhesive, or even only when necessary to satisfy elevated-temp erature
the test specimen. For these reasons, the soldering station testing requirements. These solders are somewhat more
should incorporate provision for adjusting the soldering difficult to handle because the higher working temp erature

Document Number: 11089 For technical support, contact www.micro-measurements.com


Revision 13-Nov-2010 micro-measurements@vishaypg.com 207
TT-609
Micro-Measurements

Strain Gage Soldering Techniques


rapidly vaporizes the f lux, and oxidizes the soldering When tinning bare (without soldering options) solder
tip, both of which tend to impede the soldering process. tabs of Micro-Measurements K- and D-alloy strain
Specially designed soldering tips are recommended for gages, a liquid acid flux (M-Flux SS) is recommended.
high-temperature use. After the tinning operation, the residual flux must be
completely neutralized within one to two minutes;
For routine applications, where test conditions do not
and then the leadwire joint can be completed using the
dictate the use of either a low- or high-temperature solder,
same solder and M-Flux AR rosin flux or a rosin-cored
an alloy with an intermediate melting temperature is the
solder.
normal selection. The 63/37 tin-lead alloy ( Type 361A-
20R) is an excellent choice for general-purpose strain
gage soldering. As an eutectic alloy, it has a sharply Preparation of the Soldering Tip
defined melting temp erature a characteristic that New soldering tips should always be tinned with solder prior
largely eliminates cold solder joints. The addition of a to initial use. This is easily accomplished by wrapping one
trace of antimony provides superior performance when to two in (25 to 50 mm) of solder wire around the working
the soldered connections will be exposed to very low portion of the tip while the soldering iron is cold, before
(cryogenic) temperatures for long periods of time. applying power to the soldering station. If rosin-core
The general-purpose solders are supplied with a core of solder is used, no external flux is required. With solid-wire
activated rosin f lux.This makes soldering much more solder, however, the wrapped tip should be dipped into
convenient, and is particularly useful in field applications liquid rosin flux (M-Flux AR) to provide sufficient flux
where accessory liquid rosin flux (M-Flux AR) may not be for initial tinning. Set the control on the soldering station
available. Solid-wire solder, with externally applied acid to the appropriate temperature range for the solder, and
flux (M-Flux SS), is recommended for making soldered apply power to the unit. Allow the soldering pencil to heat
connections to Micro-Measurements K- and D-alloy until the solder wrapped around the tip melts completely.
(modified Karma and isoelastic) strain gages. Rosin-core Remove excess melted solder from the tip with a dry gauze
solders should not be used in conjunction with acid flux. sponge. Never knock the heated soldering pencil against
any object to remove excess solder, since this may result in
Silver solder ( Type 1240-FPA) is available for applica personal injury or damage to the soldering pencil.
tions where leadwire connections will be exposed to
temperatures above about +550F (+290C). This solder,  ote: Cross-alloying of solders can change the electrical,
N
in paste form, is not suitable for attaching wires directly chemical, thermal and mechanical properties of the solder
to strain gage solder tabs or to bondable terminals, but is being used. To prevent cross-alloying, it is recommended
intended for connecting instrument leads to preattached that only one type of solder be used with each soldering
strain gage leads, as with WK-Series gages using a special tip. Of course, if one type of solder is incorporated in a
resistance soldering unit. Techniques for making leadwire gage with solder dots and another type is added, a mixture
connections with silver solder are described in Micro- is produced. This mixture cannot be expected to have
Measurements Application Note TT-602, Silver Soldering melting and strength properties any better than those of
Technique for Attachment of Leads to Strain Gages. the lower temperature component.
Oxidation of the soldering tip seriously hinders the
Soldering Flux soldering operation. The tendency for oxidation can be
minimized by ensuring that excess melted solder remains
The function of a soldering flux is to remove oxidation
on the tip at all times when it is not actually in use.
from the members being joined (solder tabs, terminals,
Negligent maintenance practices, or wiping the hot tip
leadwires), and to prevent further oxidation during
with materials that char on the surface, will produce a
soldering. For making leadwire splices, or soldering
buildup of oxide that prevents proper soldering. If the tip
directly to constantan foil or copper terminals, the flux
does become oxidized, the following procedure is effective
contained in a rosin-core solder is usually sufficient.
for cleaning and re-tinning:
With higher temperature solders, however, it may be
necessary to supply additional flux. A liquid activated- 1. Set the soldering station to the appropriate temperature
Tech Note

rosin flux such as M-Flux AR is recommended for this range for the solder in use.
purpose.
2. Place several drops of M-Flux SS on a glass plate.
Acid fluxes should never be used on constantan strain Re-tin the soldering surface by holding the heated tip in
gages or copper terminals, or for splicing copper leadwires; the SS flux while feeding solder onto the tip. A generous
and paste f luxes, containing chlorides, should not be amount of solder is essential for proper tinning.
used under any circumstances for strain gage soldering.

www.micro-measurements.com For technical questions, contact Document Number: 11089


208 micro-measurements@vishaypg.com Revision: 13-Nov-2010
TT-609
Micro-Measurements

Strain Gage Soldering Techniques


3. Wipe the excess solder from the tinned tip with a dry M-Flux AR to the tab or terminal (this step can be omitted
gauze sponge. For severely oxidized tips, it may be if a rosin-core solder is used). When soldering directly to
necessary to repeat this operation several times to bare Karma or isoelastic foil, use M-Flux SS on the gage
obtain a properly tinned surface. The soldering tip tabs only. Hold the soldering pencil in a nearly horizontal
should never be filed or sanded, since this may remove position (<30), with the flat surface of the tip parallel to
the plating on the tip, accelerating the oxidation and the solder tab or terminal. Place the solder wire flat on the
leading to the early deterioration of the tip. After the gage tab, and press firmly with the tinned hot soldering tip
cleaning operation, remove excess solder, re-tin and for about one to two seconds, while adding approximately
clean the tip several times, using rosin-core solder, or 1/8 in (3 mm) of fresh solder at the edge of the tip. This
solid-wire solder withM-Flux AR. procedure assures that there is sufficient solder and flux
for effective tinning. Simultaneously lift both the soldering
Tinning Solder Tabs and pencil and solder wire from the tab area.
Bondable Terminals  OTE: Lifting the soldering iron before lifting the solder
N
may result in the end of the solder wire becoming attached
All strain gage solder tabs, terminals, and leadwires must
to the tab; lifting them in the reverse order can leave
be properly tinned before making soldered connections.
a jagged (spike) solder deposit on the tab. When the
This helps ensure active surface wetting and good heat
operation is performed properly, it will produce a small,
transfer during the soldering operation. Tinning stranded
smoothly tinned area on the tab or terminal.
leadwires to produce a formable solid conductor will also
greatly simplify the leadwire attachment procedure. If M-Flux AR or a rosin-core solder is used in the tinning,
it is not necessary to remove the residual soldering flux
Before tinning the solder tabs on open-face (unencapsulat-
at this time. However, when M-Flux SS is employed to
ed) strain gages, the measuring grid should be protected
tin the bare solder tabs of K- or D-alloy gages, the acidic
with PDT-1 drafting tape.The drafting tape is positioned
flux residue must be removed immediately following the
to cover the entire grid and the upper portion of the solder
tinning operation. To remove the residue, apply M-Prep
tabs, as shown in Figure 1. This not only shields the grid
Conditioner A liberally, and wash the area with a soft
from soldering flux and inadvertent solder splash, but
brush; then blot dry with a clean gauze sponge. Next, wash
also restricts the flow of solder on the tabs. The tinned
again with freely applied M-Prep Neutralizer 5A, and blot
area on the solder tabs should be only large enough to
dry with a clean gauze sponge.
easily accommodate the leadwire size in use. The latter
consideration is particularly important when making  OTE: Special procedures for tinning and wiring strain
N
installations for dynamic applications or large-strain gages supplied with preattached solder dots are described
measurement. in Micro-Measurements Application Note TT-606,
Soldering Techniques for Lead Attachment to Strain Gages
The tinning procedure for strain gage tabs and terminals
with Solder Dots.
consists of first cleaning and reapplying a small amount of
solder to the hot soldering iron tip. Next, apply a drop of
Tinning and Attaching Leadwires
Of course, leadwire ends must be stripped of insulation
before tinning, and this should be done with a thermal
wire stripper to avoid the damage to the wire that often
occurs when mechanical wire strippers are used. After
the wires are stripped, the ends of stranded conductors
should be twisted tightly together before tinning. The bare
leadwire ends can then be tinned easily with the following
procedure:
1. Remove excess solder from the soldering tip, using a
Tech Note

dry gauze sponge. Then melt fresh solder on the hot tip
to form a hemisphere of molten solder about twice the
diameter of the wire to be tinned.
2. If rosin-core solder is used, slowly draw the bare wire
through the molten solder while continuously adding
Figure 1 Gage grid and upper portion of fresh solder to the interface of the wire and soldering
solder tabs masked with drafting tape. tip. With solid-wire solder, apply M-Flux AR to the

Document Number: 11089 For technical questions, contact www.micro-measurements.com


Revision: 13-Nov-2010 micro-measurements@vishaypg.com 209
TT-609
Micro-Measurements

Strain Gage Soldering Techniques

Figure 2 Trimming leadwire ends before taping in place. Figure 3 Leadwire end taped to surface
in preparation for soldering.

wire end before starting to tin, and proceed in the same gage, terminal, or bonding adhesive. Apply a small amount
manner. This will produce a smooth, shiny coating of of M-Flux AR to the joint area and, holding the soldering
solder over the bare wire. pencil nearly horizontal, firmly press the flat surface of the
tip on the junction for about one second; then lift the tip
For applications employing bondable terminal strips and
from the soldered joint. If needed, additional flux can be
stranded instrumentation wire, it may be convenient to use
provided during the joining operation by feeding a little
a single strand of the wire as a jumper between the terminal
fresh solder into the joint from a spool of rosin-core solder.
and the strain gage solder tab. In such cases, the single
This procedure should result in a smooth, hemispherical
wire strand should be separated out before twisting and
solder joint, without any peaks or jagged areas. If the
tinning the remaining strands (see Micro-Measurements
solder joints are not smooth and uniform in size, repeat the
Application Note TT-603, The Proper Use of Bondable
soldering procedure, using additional flux and/or solder as
Terminals in Strain Gage Applications).
necessary.
Leadwires should be formed and routed to the strain
gage or terminal strip, then firmly anchored to the Cleanup and Inspection of
test-part surface with drafting tape before making the Soldered Joints
soldered connection. Attempting to route the leadwires
after completing the solder joint will often result After completing the soldering operation, it is imperative
in damage to the gage or terminals. Routing into the that all traces of residual f lux be completely removed
connection area should be along a minimum strain with RSK Rosin Solvent. The same solvent is used to
direction (such as the Poisson direction in a uniaxial soften the mastic of the drafting tape, permitting its easy
stress field) particularly for high elongation or dynamic removal. Do not try to pull away the tape with tweezers
tests. The tinned leadwire end should be trimmed short or other tools, because this may result in damage to the
enough so that it will not protrude through the connection soldered connections or the strain gage grid. Thoroughly
area, and cannot inadvertently make electrical contact clean the entire installation area with generously applied
with the test-part surface or adjacent solder connections. rosin solvent and a soft-bristled brush. Clean the solder
Figure 2 illustrates this stage in the procedure. In the final connection area until no visible signs of residual f lux
preparatory step, bend the leadwire end slightly to form remain, and blot the area dry with a clean gauze sponge.
a spring-like loop, and tape the wire firmly in place over Any traces of residual f lux can cause gage instability
the connection area, using PDT-1 drafting tape. The tape and drift, and will inhibit bonding of the installations
Tech Note

should be within about 1/8 in (3 mm) of the connection protective coating. Incompletely removed soldering flux is
area, as shown in Figure 3. the most common cause of degraded performance in strain
gage installations. Residual flux mixed with a protective
Clean and re-tin the soldering iron tip with fresh solder. coating application can completely destroy the coating
The temperature of the iron should be adjusted so that the objective.
solder is easily melted, without rapidly vaporizing the flux.
If the iron temperature is either too low or too high, it may Visually inspect the soldered joints for any gritty or jagged
cause poor solder connections, or it may damage the strain joint surfaces, and for traces of flux. Solder connections

www.micro-measurements.com For technical questions, contact Document Number: 11089


210 micro-measurements@vishaypg.com Revision: 13-Nov-2010
TT-609
Micro-Measurements

Strain Gage Soldering Techniques


should be smooth, shiny, and uniform in appearance. professional soldering equipment which is specially
Any soldered joints that look questionable should be designed for making soldered connections in strain gage
re-soldered, and flux removed. Check the resistance-to- circuits. The soldering pencil should be lightweight, with
ground of the completed gage installation, using the Model a flat chisel or screwdriver tip, and it should be connected
1300 Gage Installation Tester. Low or marginal resistance to the soldering station with a very flexible power cord.
readings suggest a leakage path between the soldered Requirements for the soldering station include low-voltage
connections and the test-part surface. This condition operation of the soldering pencil, and provision for
usually results from residual soldering flux, or from bare temperature adjustment to suit the type of solder and the
leadwire conductors partially shorting the gage tabs or application conditions. The equipment should not generate
terminals to the test part. Soldered joints should not be electrical interference that could affect sensitive measuring
tested by pulling on the leadwire, or by probing at the joint instrumentation. Solder selection is based primarily on the
area. These practices frequently cause lifting or tearing of expected operating temperature range of the strain gage
the solder tab from the gage backing material. installation; and all solder tabs, bondable terminals, and
leadwire ends should be tinned before soldering the joints.
Summary Soldered joints should always be smooth and shiny, with
no jagged or irregular edges, and all traces of residual flux
The ability to make consistently good soldered joints is must be thoroughly removed prior to the application of
essential for precision strain gage measurements. The protective coating. Use of the recommended materials and
techniques described here are straightforward and easily techniques, with careful attention to detail, will result in
mastered, but they are most effective when used with consistently proper and reliable soldered connections.

Tech Note

Document Number: 11089 For technical questions, contact www.micro-measurements.com


Revision: 13-Nov-2010 micro-measurements@vishaypg.com 211
MICRO-MEASUREMENTS

Strain Gages and Instruments Application Note TT-604

Leadwire Attachment Techniques for


Obtaining Maximum Fatigue Life of Strain Gages
Introduction
Micro-Measurements WA-, WK-, and WD-Series Strain
Gages are manufactured with integral beryllium-copper
lead ribbons. To maximize the inherently good fatigue
life of this gage/lead combination, the gage-to-instrument
leadwire connection must also be highly resistant to
fatigue damage. The techniques described here are
recommended for applications where the cyclic strain
endurance requirements approach the fatigue limits of the Figure 1
gage in use.
Wir ing procedures for gages with integral solder
dots instead of lead ribbons are described in Micro-
Measurements Application Note TT-606, and general gage
Lead Preparation and Attachment
soldering techniques are discussed in detail in Application
Note TT-609. Step 1
Carefully lift the gage lead
Bonding ribbons from the specimen
M-Bond 600 and M-Bond 610 are the strain gage surface by laying the wood
adhesives generally recommended for high-performance extension of a cotton-
applications. However, any M-Bond adhesive can be used tipped applicator firmly
where temperature conditions and other factors permit. across the gage at the lead
Although Micro-Measurements strain gage adhesives are exit point, grasping each
not particularly fatigue sensitive, some potential problem lead end with tweezers
areas are: and raising the lead at a
shallow angle (Figure 2).
A
 dhesive failure due to improper surface preparation Avoid introducing kinks Figure 2
or environmental protection. or sharp corners.
V
 oids or air bubbles in the glueline, causing hot
spots in the gage and subsequent premature failure. Step 2
Bonding instructions for M-Bond 600 and 610 are detailed With tweezers, guide each
in Micro-Measurements Instruction Bulletin B-130. lead ribbon upward and
Wherever possible, the use of bondable terminals is around the tip of a dental
recommended to provide an anchor for the instrument probe to form a small
leadwire, preventing damage to the gage when the leadwire loop above the surface
is subjected to any type of force. Micro-Measurements betwe en the lead ex it
Type CPF-75C terminals are particularly suitable for this point and the edge of the
gage backing (Figure 3).
TECH NOTE

application within their operating temperature range.


If bonded in a high strain field, terminals should be cut in Do not twist or kink the Figure 3
half and/or bonded 90 to the maximum strain direction. ribbons. Once kinked, a
Figure 1 is representative of typical installations. lead will easily break.

Document Number: 11084 For technical support, contact www.micro-measurements.com


Revision 20-Feb-2012 micro-measurements@vishaypg.com 189
TT-604
Micro-Measurements

Leadwire Attachment Techniques for


Obtaining Maximum Fatigue Life of Strain Gages
Step 3
Insulated, three-conductor, stranded tinned-copper
leadwire cable is normally selected for static or static/
dynamic measurements. Refer to Micro-Measurements
Strain Gage Accessories Data Book for the selection of a
suitable leadwire. Typically recommended cable types are
Micro-Measurements 326-DFV (flat, and most convenient
for routine applications) or 326-DSV (twisted, and shielded
for use when electrical noise fields are present). Both cables Figure 6 Figure 7
are color-coded red, white, and black. The 326-DFV is
used to illustrate the following step-by-step procedures.
Step 6
Cut the leadwire cable to the appropriate length. At the
gage end of the leadwire, separate the three color-coded With tweezers, form the single wire strands and insert
conductors, and strip 0.5 in (13 mm) of insulation from through the gage lead loops as shown in Figure 6. Tighten
each. A thermal wire stripper is recommended to help the gage lead loops by pulling gently with a pair of tweezers
prevent nicking the wires. (Figure 7). To avoid damage to the gage joint, lightly press
the wood extension of a cotton-tipped applicator over the
area as indicated in the figure.
Step 4
Solder the joints and cut off excess lead ribbons and wire
Separate a single wire strand from both the red conductor strands.
and the black conductor.
Twist the remaining strands of the red conductor together, Step 7
leaving the single strand separated. Twist the remaining
st r a nd s of t h e bl a ck Remove solder flux with rosin solvent, gently blotting dry
conductor together with with a gauze sponge. If any wire strands come in contact
all strands of the white with the specimen, gently lift with a dental probe.
c o n du c t o r, l e av i ng
the single strand from Protective Coating Application
t h e b l a c k c o n du c t o r
Refer to Micro-Measurements Strain Gage Accessories
separated.
Figure 4
Data Book for the selection and application of a suitable
Ti n t h e t wo t w i s t e d protective coating.
bundles of stranded wire
with solder. With diagonal cutters, trim each bundle to
approximately 0.1 in (2.5 mm) from the insulation, being Note: WD-Series Strain Gages are supplied with single
careful not to trim the two single strands of wire extending beryllium-copper leads. The WA- and WK-Series gages, as
from the red conductor and the black/white conductor shown throughout this Application Note, are supplied with
(Figure 4). dual leads. The WD gages should be handled in the same
manner as the dual lead connections.
Step 5
Tin the bonded terminals
w ith solder. Plac e the
tinned, trimmed sections
of the leadwires on the
terminals, and anchor the
TECH NOTE

cable in this position with


drafting tape. Solder the
leads in place (Figure 5),
Figure 5 avoiding large masses of
solder.

www.micro-measurements.com For technical questions, contact Document Number: 11084


190 micro-measurements@vishaypg.com Revision: 20-Feb-2012
Micro-Measurements

Strain Gages and Instruments Application Note TT-605

High-Elongation Strain Measurements

Introduction EP-Series
Experimental stress analysis of structural materials EP-Series gages are recommended when measurement
sometimes requires testing to complete failure. In such requirements are beyond the 3% 5% elongation capability
cases, particularly with ductile materials, failure is often of the EA-Series. P-alloy is a fully annealed constantan foil
preceded by large local strains, the magnitudes of which processed for very high ductility. A properly bonded
are of interest to the test engineer. and wired EP-Series strain gage is capable of strain
measurements to 20% (200 000 microstrain) or greater.
High-elongation strain measurements place severe demands As in the EA-Series, smaller gages will exhibit a lower
on the gage installation, and necessitate special gage and maximum elongation capability.
adhesive selection and surface preparation procedures.
This Application Note outlines recommendations for gage
and adhesive selection, surface preparation, gage bonding Options
and wiring, and protective coating selection for high- Any leadwire attachment option, such as Option W
elongation strain measurements. (encapsulated gages with integral terminals) or Option
High elongations are usually expressed as percentages: 1% LE (encapsulated gages with leadwires), will lower the
elongation is equal to 10 000 microstrain; 5% elongation maximum elongation capability of EP-Series gages. But
is equal to 50 0 0 0 microstrain. The maximum rated because the elongation limits of EA-Series strain gages
elongation of any bonded strain gage is based on a single are more moderate (3% 5%), and because options will
tensile measurement. Cyclic strains of high levels must normally withstand at least the lower end of this range,
be limited to a value much lower than maximum due to they are not generally a limitation in the selection of
fatigue limitations on the strain gage foil. Consult Micro- EA-Series gages.
Measurements Tech Note TN-508, Fatigue Characteristics To obtain maximum possible elongation, select EP-Series
of Micro-Measurements Strain Gages, for details on strain gages without options.
gage performance under repetitive loading conditions.
Adhesive Selection
Gage Selection
High-elongation strain measurements place severe
Selection of proper strain gages for use in high-elongation demands on the adhesive system. The adhesive must be
testing is based on both anticipated strain levels and test rigid enough to prevent gage relaxation (creep), yet flexible
temperature. Polyimide (E) backing is normally selected enough to permit large deformations without cracking.
for this type of service because it has superior elongation Recommended adhesives are listed in the following
capabilities and an operating temperature range suitable selection chart.
for most high-elongation testing. The E-backing is available
with constantan (A) foil or fully annealed constantan
Surface Preparation
(P) foil.
The selection and implementation of proper surface
EA-Series preparation procedures is equally as important as using
proper gages and adhesives. High-elongation measurements
Properly installed and wired EA-Series (polyimide-backed
Tech Note

demand closer attention to recommended procedures than


constantan foil) strain gages are capable of measuring do normal elastic strain measurements.
maximum elongations in the range of 3% 5%. While gage
lengths of 1/8 in (3 mm) or longer will normally achieve Micro-Measurements Instruction Bulletin B-129, Surface
5%, shorter gage lengths may be limited to 3%. Since most Preparation for Strain Gage Bonding, outlines steps for
structural materials (e.g., metals) yield well below this surface preparation on a variety of materials. These
limit, the EA-Series is a popular choice for use in obtaining procedures produce a smooth, clean surface, usually
yield point information on these materials. having been abraded in a single direction during

Document Number: 11085 For technical support, contact www.micro-measurements.com


Revision 12-Nov-2010 micro-measurements@vishaypg.com 191
TT-605
Micro-Measurements

High-Elongation Strain Measurements


ELONGATION CAPABILITIES
ADHESIVE PREFERRED
GENERAL DESCRIPTION 320F +75F +200F
TYPE CURE
(195C) (+24C) (+95C)

Special pretested grade of cyanoacrylate, 6%


M-BOND certified for strain gage use. Fast room- 3 min at Not Up to 15% for Not
200 temperature cure. +75F (+24C ) applicable installations less applicable
than 30 minutes old
Two-component, 100%-solids epoxy
M-BOND 6 hr at
system. Capable of room-temperature cure. 1% 6 10% 10 15%
AE-10 +75F (+24C )
Transparent, medium-viscosity adhesive.
Two-component, 100%-solids epoxy
M-BOND 2 hr at
system. Moderately elevated temperature cure. 2% 15% 15%
AE-15 +150F (+65C )
Transparent, medium-viscosity adhesive.
Two-component, partially filled, 100%-solids 6 min at
M-BOND epoxy system. May be cured at room +75F (+24C )
4% 10 15% 15 20%
GA-2 temperature. Higher viscosity than AE system. followed by
1 hr at +125F (+50C )
Two-component, filled, 100%-solids epoxy
M-BOND system. Paste-like consistency. Specifically 2 hr at Not
15 20% 15 20%
A-12 recommended for elongations above 10% on +165F (+75C ) applicable
steel.
For additional specifications, consult Micro-Measurements Strain Gage Accessories Data Book
and specific adhesive instruction bulletin.

preparation. Because of higher bondline forces involved Many materials oxidize readily in air. If allowed to
in high-elongation measurements, the surface should be form on the surface, oxidation will greatly reduce
altered further for greater bond strength, as indicated in bond strength and elongation capabilities. It is strongly
the following procedures: recommended that the gage bonding operation be
completed as soon as possible after the surface is
1. Completely prepare the surface of the test specimen as
prepared.
described in Bulletin B-129.
2. Abrade the specimen surface in a direction 45 to the GageBonding
intended axis of strain measurement. On soft materials,
such as aluminum, use 320-grit silicon-carbide paper; Follow the standard bonding procedures outlined in the
on harder materials, such as steel, use 60-grit. respective instruction bulletin for the adhesive selected.
3. Lightly abrade, with the appropriate grit paper as
Gage Wiring
indicated in Step 2, in a direction 90 to the first abrasion.
This will produce a cross-hatched abraded surface. Relatively high displacements occur with high-elongation
Typical surface roughness desired approximates 250 in strain measurements, suggesting the use of an external
(6.4 m) rms. ter-minal strip. The strain gage should be wired to the
4. Repeat the degreasing steps outlined in Bulletin B-129. terminal strip with a small-diameter, single-conductor
wire no larger than 30 AWG (0.25 mm diameter) as shown
5. Condition the surface with Micro-Measurements in Figure 1. Drafting tape (Micro-Measurements PDT-1)
Tech Note

M-Prep Conditioner A (if appropriate for the specimen applied over the gage during soldering is recommended to
materials as detailed in B-129), scrubbing with cotton- restrict the flow of solder to the tab ends.
tipped applicators. Wipe dry with a gauze sponge.
Adequate strain relief loops, as shown in the following
6. Neutralize the surface with Micro-Measurements illustration, are of particular importance in high-elongation
M-Prep Neutralizer 5A (if appropr iate for the measurements. Note also that the leads attached to the
s p e c i m e n m at e r i a l), s c r ub b i ng w it h c ot t o n - strain gage tabs approach in a direction 90 from the
tipped applicators. Wipe dry with a gauze sponge. maximum principal strain axis.

www.micro-measurements.com For technical questions, contact Document Number: 11085


192 micro-measurements@vishaypg.com Revision: 12-Nov-2010
TT-605
Micro-Measurements
Tape Coverage

High-Elongation Strain Measurements


Strain relief loop
Tin entire copper Style C terminal
Half-Terminal Technique
terminal,for
leavingUsing
neat Bonded Terminals in High Strain Fields
cut in half lengthwise
dome of solder

Note: When the main leadwire is stranded, it is


often convenient to cut all strands but one to fit
Tape Coverage
the size of the copper terminal. The long strand
can then be used as the jumper wire. Soldering
is made considerably easier by this method.

Strain relief loop


Tin entire copper Style C terminal
terminal, leaving neat cut in half lengthwise
dome of solder

Single Strand

Twist and tin remaining strands before


cutting and soldering to the terminal

Figure 1

After soldering, all tape and soldering f lux should be Instrumentation


removed with several brush applications of rosin solvent
(Micro-Measurements RSK-1). Wheatstone bridge
Single Strandcircuits are commonly used in strain
gage measurements. These circuits are inherently nonlinear
Gage Linearity when large gage resistance changes are encountered.
Consult Micro-Measurements Tech Note TN-507, Errors
A discussion of high-elongation strain gage measurements Due to Wheatstone Bridge Nonlinearity, for correction
Twist and tin remaining strands before
is not complete without including the cutting
subject of gage procedures.
and soldering to the terminal
linearity under plastic straining conditions.
Although constantan alloy is one of the most linear of all
Common Installation Problems
strain gage alloys, it does exhibit small changes in gage
factor during high elongation. Unbonding of the Gage
Without treating this subject in detail, it can be shown in Assuming the proper adhesive has been selected (see
theory that the gage factor of the strain gage approaches a Chart), loss of bond before maximum gage elongation
value of 2+ when the gage alloy is strained beyond its elas- is reached can usually be traced to a contaminated or
tic range. The elastic range for constantan is approximately improperly prepared surface. A visual examination of the
0.5%. bond area will usually detect this. Cured adhesive on the
back of the strain gage but not on the specimen generally
Using the 2+ guideline, the gage factor at a strain level of
indicates improper or incomplete surface preparation.
10% would be close to 2.1 in tension and 1.9 in compression.
Tech Note

At a 20% strain level, the gage factor approaches 2.2 in It is important to note that some materials, like polyethylene,
tension and 1.8 in compression. are difficult to bond; and it will not be possible to reach
maximum gage elongation before bond failure.
While these numbers are correct in theory, they have
not, to our knowledge, been substantiated by actual test
results. Unbonding of the Gage Tabs
Gage tab unbonding is often due to an excessive amount
of solder, which reinforces the gage tabs, and causes them

Document Number: 11085 For technical questions, contact www.micro-measurements.com


Revision: 12-Nov-2010 micro-measurements@vishaypg.com 193
TT-605
Micro-Measurements

High-Elongation Strain Measurements


to unbond as a result of their inflexibility. Drafting tape
is recommended for restricting the amount of solder (see Thin Metal Strip Hardened
Gage Wiring). to Retain Shape

Excessive soldering temperatures, or terminals placed too


close to gage tabs, can also contribute to this problem.

Gage Grid Failure (Open Circuit) FOOT

Assuming that the strain gage readings were within the STRAIN GAGES
strain range capability of the gage, and that the backing TOP AND BOTTOM
remains bonded, premature grid failures are often the
result of high local strains within the area covered by the
gage. Since the strain gage is an averaging device, it will
indicate average strain along its grid length. Steep strain COMPRESSION
gradients can cause localized excessive strain damage
while the gage may have been indicating a strain that was
well within its elongation limits.
TENSION
Grid failures may also result from strain concentrations
caused by inclusions in the adhesive layer (unmixed
adhesive particles, dirt, etc.) or by uneven gluelines, often
caused by irregular or pitted specimen surfaces or uneven
clamping pressures.

Elongations Exceeding 20% BRIDGE


EXCITATION
Strain measurements, particularly on materials such as
plastic or rubber, sometimes surpass the capability of
the EP-Series strain gage. Measurements in this range
can be obtained through simple flexure devices designed
to reduce the strain level on the gages. These devices,
commonly referred to as clip gages, are shown in Figures Figure 2A Clip Gage (Half Bridge)
2A and 2B. Strain gages are installed on the top and
bottom of the clip, and the clip mounted to the specimen
by bonding or spot welding, depending on the specimen
material. The displacement of the feet under strain causes
corresponding resistance changes in the strain gages. With
the gages wired into either a half or full Wheatstone bridge,
their resistance changes are additive, thereby increasing
the amount of available signal. Wherever space permits,
full bridges are recommended because the effective bridge STRAIN GAGES
output is twice the half-bridge signal for the same foot TOP AND BOTTOM
displacement.
Since the clip gage is normally a nonlinear device, the
gage output is calibrated against known displacements
before installation. The gage zero reading is monitored
during the calibration, since a permanent zero offset after
Tech Note

BRIDGE
calibration can usually be traced to localized yielding of EXCITATION
the metal strip. If this is indicated, the calibration should
be repeated several times to check for reproducibility.
A clip gage is particularly useful in measuring very large
specimen strains or actual displacements occurring
between two bodies, as with expansion joints or crack
Figure 2B Clip Gage (Full Bridge)
openings.

www.micro-measurements.com For technical questions, contact Document Number: 11085


194 micro-measurements@vishaypg.com Revision: 12-Nov-2010
Micro-Measurements

Strain Gages and Instruments Application Note TT-606

Soldering Techniques for Lead Attachment to


Strain Gages with Solder Dots
Introduction
Many gages are now purchased with integral solder dots.
These offer advantages of accurately controlled solder
joint size, convenience when working in confined areas,
and of having pretinned hard-to-solder alloys such as
isoelastic and Karma. Controlled joint size, in turn, helps
to increase gage fatigue life, and minimizes thermocouple
Correct
effects.
Shape
If proper installation techniques are used, there should
be no difficulty experienced in obtaining well-soldered
connections; but care must be exercised. Open-faced gages
with Option S (solder dots only) should be coated with
adhesive at the time of installation. Such encapsulation Incorrect
effectively prevents dot enlargement and undesirable Shape
spreading during subsequent soldering steps. To expose
the dots, lightly abrade with 400-grit silicon-carbide paper
or simply solder through this overcoat. Encapsulated gages
Figure 1
do not require additional coating with adhesive.
The following are important guidelines:

Soldering Temperature
T backed gages with Option S, and S Series gages,
have dots composed of +570F (+300C) lead-tin-silver
solder. Micro-Measurements wire solders have melti ng
Iron Direction
points ranging from +361 to +570F (+183 to +300C);
the most common are the +361 and +570F (+183 and
+300C) solder wire with rosin-core flux. Solid wires are
also available, including a popular +430F (+220C) type. Flat Surface
A temperature-controlled soldering station, such as the
Correct
Mark V or Mark VIII Soldering Units, is recommended.
Tip temperature should be high enough to assure good
wetting of the solder but not so high as to remove the
dots, vaporize flux, or hinder keeping the iron properly
tinned and clean. Unfortunately, most uncontrolled irons
are quite capable of tip temperatures in excess of +900F
(+480C), which is much too high for general strain gage
Tech Note

soldering.
Iron Direction
Soldering TipDesign
Incorrect
Never solder with a sharply pointed tip (Figure 1). A hot
point applied to a solder dot usually draws out the solder.
Use a clean, flat 1/16-in (1.5-mm) wide chisel or screwdriver Figure 2
type tip held flat against the work, as in Figure 2.

Document Number: 11086 For technical support, contact www.micro-measurements.com


Revision 12-Nov-2010 micro-measurements@vishaypg.com 195
TT-606
Micro-Measurements

Soldering Techniques for Lead Attachment to


Strain Gages with Solder Dots
Fluxing Hold-down
Proper fluxing is essential. Enough M-Flux AR, either Each connecting wire must be carefully and firmly
from melting of flux-core solder or adding M-Flux AR anc hored in place with spring loading against the dot
separately, must be introduced to assure complete wetting bead, as illustrated in Figure 5, before soldering the
action. This is particularly important with +570F (+300C) connection. When this is properly done, a one-second
solder as there is an increased tendency for the flux to be touch of a hot iron and fresh solder with f lux should
boiled away at higher temperatures. complete the process.
If spikes are produced instead of smooth beads, it is a
sign of inadequate fluxing, dwelling too long with the iron, Solder Compatibilities
and/or an improper iron temperature. All Micro-Measurements solders are compatible with each
Rosin flux should always be used in making connections other. Of course, if one type is incorporated in the dot and
to Option S gages. All flux residue should be thoroughly another added from wire, a mixture is produced. This new
removed with RSK Rosin Solvent. alloy cannot be expected to have melting and strength
properties any better than those of the lower temperature
BeadFormation component. In special cases of mixing +430 and +570F
(+220 and +300C) solders, a final alloy similar to +361F
It is important to first build up a well-formed bead on (+183C) solder can be encountered.
each dot as in Figure 3. This is done by laying rosin-core
Figure 6 illustrates a typical installation. Note that solder
solder wire across each dot, firmly applying the iron for
dot (Option S or SE) gages are always connected to bonded
one second, then simultaneously lifting both wire and iron.
term inals for strength. Refer to Application Note TT-603,
See Figure 4. If the first attempt fails, simply repeat this
The Proper Use of Bondable Terminals in Strain Gage
procedure, making certain to use enough flux. Feeding a
Applications, for other arrangements.
cored solder into the joint area during heat application will
increase the available flux.

Before
After Soldering

After
Before Soldering
Figure 3 Figure 5

Iron Direction
Tech Note

Solder

Flat Surface

Figure 4 Figure 6

www.micro-measurements.com For technical questions, contact Document Number: 11086


196 micro-measurements@vishaypg.com Revision: 12-Nov-2010
TT-606
Micro-Measurements

Soldering Techniques for Lead Attachment to


Strain Gages with Solder Dots
The various installation accessories referred to throughout Application Note TT-609, Strain Gage Soldering Techniques,
th is Appl ic at ion Not e are Mic ro -Me a su re ment s provides a detailed discussion on general strain gage
Accessories, listed in our Strain Gage Accessories Data soldering techniques.
Book and available directly from Micro-Measurements.

Tech Note

Document Number: 11086 For technical questions, contact www.micro-measurements.com


Revision: 12-Nov-2010 micro-measurements@vishaypg.com 197