You are on page 1of 4

SM

NALCO 72100,
ANALYTICAL PROCEDURE 22130
DR/2010 Procedure

AP-60E-1/2010
SPECIFICATIONS
Method: Spectrophotometric (DR/2010)
Range: 100 to 6000 50 ppb as NALCO
72100 or 22130
Testing Time: 15 minutes

DESCRIPTION
An accurate, simple method of colourimetric
analysis, the portable DR/2010 spectrophotometer
is utilised to quickly analyse NALCO 72100 or
22130 concentration for monitoring proper applica-
tion levels. The test is based on iron reduction
Nalco DR/2010 Spectrophotometer
chemistry and is simple to perform. The NALCO
72100 or 22130 in the sample reduces ferric iron to
ferrous iron in direct proportion to the NALCO
72100 or 22130 concentration. This solution then Description Part No.
reacts with Reagent 1 to form a purple colour with
the ferrous iron. Nalco DR/2010 Spectrophotometer P1310A
Thermometer, lab, 0 to 230F P2028
Pipette, disposable, 1 mL P2107
ORDERING INFORMATION Sample cells, 25 mL, matched set of two P2502
Order all replacement parts and reagents by their Flask, Erlenmeyer, glass, 250 mL P1501
part numbers. To be sure this product is compatible Cap, plastic (for P1501) P3037
with your treatment programme, contact your local
Nalco representative. Items marked with an * are Optional Parts and Reagents
included with the kit or set.
Description Part No.
To place your order, please contact your local
Customer Service Department. Pour-Through cell assembly kit P2517
Sample cells, 25 mL, matched pair P2502
Thermometer, -40 to 450F, digital P2098
REPLACEMENT PARTS AND REAGENTS Thermometer, 50 to 300F, dial P2019
Required Parts and Reagents Thermometer, 0 to 230F, glass P2028
Pocket pH Meter P1326A
Description Part No. pH (3-colour) test strips (0-14 pH)
ELIMIN-OX Reagent Set (100 Tests) P3115 (100/box) P3204
pH (3-colour) test strips (510 pH)
Includes: (100/box) P3207
ELIMIN-OX Reagent 1* 100/pkg (2 included) Sample Cooler, Threaded (Inconel/SS) P4625
ELIMIN-OX Reagent 2* 100 mL Sample Cooler II (316 SS/MS) P4627
Description Size Part No.
ELIMIN-OX Reagent 1 100/pkg S0195-87
ELIMIN-OX Reagent 2 100 mL S0196-72

NALCO APPLIED SERVICES OF EUROPE B . V.


I R. G. TJALMAW EG 1 2342 BV OEGSTGEEST P.O. BOX 627 2300 AP LEIDEN THE NETHERLANDS

Registered Trademarks of Nalco Chemical Company 1995,1997 Nalco Chemical Company


All Rights Reserved Printed in U.S.A. 1-97
Description Size Part No. SAFETY PRECAUTIONS
1:1 HCl 50 mL S0726-71 Refer to Material Safety Data Sheet (MSDS) for
1:1 HCl 100 mL S0726-72 specific reagent personal safety and disposal
1:1 HCl 500 mL S0726-74 information.
1:1 HCl 1L S0726-75
1:1 HCl 4L S0726-08
Fe-4 50 mL S0734-71 SUPPORT
Fe-4 500 mL S0734-74 If you have any questions regarding this procedure,
please contact your local Nalco representative.
INTERFERENCES
Substances that reduce ferric iron (such as other SAMPLING
oxygen scavengers) can interfere. Substances that 1. This is a low level test and is very susceptible to
strongly complex iron may also interfere. The contamination. Care must be taken to eliminate
following substances may interfere when present sources of air dissolution and other interfering
in excess of the stated amount. species. A cooler equipped with a copper cooling
Ammonia 100 ppm coil should NOT be used. Trace levels of copper
Borate (as Na2B4O7) 500 ppm catalyse the reaction of ELIMIN-OX and oxygen.
Cobalt 25 ppb As a result of copper contamination, any
Copper 50 ppb ELIMIN-OX present in the system may be
Cyclohexylamine 3 ppm consumed prior to analysis.
Hardness (as CaCO3) 1000 ppm 2. Reaction glassware should be scrupulously clean.
Hydrazine 250 ppb For best results, isolate this glassware to be used
Iron (Ferrous) *See Note only for this test. If glassware is contaminated, it
Lignosulphonates 50 ppb can be cleaned by rinsing with 1:1 HCl (S0726)
Manganese 0.8 ppm followed by several iron-free distilled water
Molybdenum 80 ppm rinses. Commercially available distilled water is
Morpholine 50 ppm acceptable for use.
N7200, N7205 25 ppm
Nickel 0.8 ppm 3. Periodically clean the sample cells with a warm
Phosphate, phosphonates 10 ppm solution of distilled water containing
Sulphate 1000 ppm 1.0 mL of Fe-4 (S0734) per 25 mL water. After
Zinc 50 ppm soaking, rinse the clean cells with a few mL of 1:1
HCl (S0726) followed by at least 3 rinses
Light can also interfere with colour development. Keep with iron-free distilled water.
reacted samples in a dark area during the waiting step.

Note: Any Ferrous iron will cause a positive interference


and can be blanked out by performing the procedure on
the sample omitting Step #6. Subtract this result from
the samples EOx result to determine the samples true
ELIMIN-OX concentration. Do not use this iron blanking
procedure to quantify Fe+2.
ELIMIN-OX DR/2010 PROCEDURE
Special Notes: Analyse sample as soon as it is collected. The temperature of the sample should be 22-28 C (72-82 F). Do
not use a copper-based sample cooler as copper interferes with this reaction. If the temperature of the sample is not in the
proper range, correct the result as follows:
EOx [9 X (C*1.8 - 45)] = EOx TC OR EOx [9 X (F - 77)] = EOx TC

562 nm
TIMER PROG EDIT

7 3 2
READ
ENTER

1 2 43 54
Enter the stored programme Dial the wavelength until the Collect a feedwater sample Fill one of two matched 25
number for NALCO 72100 or small display shows after the NALCO 72100 or mL sample cells to the mark
22130: 22130 feed point. Stopper with sample. Fill the second
562 nm the flask with a plastic cell with demineralised or
Press: 732 ENTER. stopper (P3037). distilled water. This second
When the correct wavelength
The display will show: is dialed in, the display will cell will be the blank.
quickly show: Notes: Do not use a
DIAL nm to 562
Zero Sample rubber stopper. Notes: Matched cells must
be used for this procedure.
then: ppb ELIMIN-OX Sampling technique is
extremely important. Refer The sample temperature
to Sampling Section of this should be 22-28 C
procedure. (72-82 F).

TIMER
SHIFT
7

65 76 87 98
Add the contents of one (1) Add exactly 0.5 mL ELIMIN- Immediately after the When the timer beeps, the
ELIMIN-OX Reagent 1 OX Reagent 2 to each cell. addition of the ELIMIN-OX display will show:
pillows to each sample cell. Mix. Place both cells in the Reagent 2, press SHIFT
dark. TIMER and a 10 minute ppb ELIMIN-OX
Swirl the cells to dissolve. reaction period will begin. After wiping any moisture or
Note: A purple colour fingerprints from the outside
Note: For accurate results will slowly develop if Notes: Keep both cells in of the cell, place the blank
use pillows from the same ELIMIN-OX is present. the dark during this into the cell holder and close
manufacturing lot. period. the light shield.
Temperature and
reaction time affect the Note: ppb ELIMIN-OX is
results. Be sure these equal to ppb 22130 or
factors are controlled as 72100.
described.
CLEAR READ
ZERO ENTER

10
9 11
10 12
11
Press: ZERO. Immediately remove the Press: READ.
blank cell and insert the
The display will show: sample cell. Close the light The display will show:
ZEROING shield. READING
then: 0 ppb ELIMIN-OX The result in ppb ELIMIN-OX
will be displayed.

Note: ppb ELIMIN-OX is


equal to ppb 22130 or
72100.