controls

THN Digital thermocouple conditioner


Main Characteristics Types and ranges
Thermocouple type : K, J, T,
Compact and easy to install Ranges : -50 to 200C, -100 to +400C, 0 to 1250C
Works with any thermocouple Any other range may be achieved for improved resolution
Cold junction compensation from -40C to +125C ambient
type K, J, T... terminated with a temperature
miniature male connector
Standard Range from -100 to Specifications
1250C Supply voltage: 5 to 16 V or 12 to 33V

Document non contractuel, succeptible de modification sans pravis.


Supply current: 3mA
Supply 5 to 30V Sampling frequency: 5HZ
Digital output :
Both linear Output signals : 5HZ serial data 9600 Bauds ASCII format
- Digital Max error including nonlinearity and hysteresis: 0.5C
- Analog 0 to 5V or 0 to 10V Max error for ambient temperature : 0.5C
Analog output :
THN withstands -40 to +125 C 0-5V if supply is 5 to 16V,
ambient temperature allowing 0-10V if supply is 12 to 33V
a connection close to the probe Additional error including nonlinearity and hysteresis: 0.5C
Offset drift with temperature : < 0.25%
itself.
Non linearity and Hysteresis Operating Temperature: -40 to 125C
Storage Temperature: -40 to 125C
Max error 0.5C Protection : IP53

Connection

AXILANE
INSTRUMENTS
10 rue des Acacias

1/2
91620 Nozay
FRANCE
Tel. : +33 (0) 950 60 40 20
Fax : +33 (0) 955 60 40 20
controls

THN Digital thermocouple conditioner


Serial Data Data on Hyperterminal
Drivers Use HyperTerminal from Windows (up to XP*)
Download a VCP (Virtual Com Port) driver FT232R to 9600 bauds, 8 data bits, no parity, 1 bit stop, no
emulate a serial port with USB input. This driver must control, recommended font: Courrier New)
match the Windows version. At power up of the THN :
Downloading address: THN 3.6 Texys.fr software version
www.ftdichip.com/Drivers/VCP.htm T -50C +200C temperature range
Transfer the CDM.zip file and store it in a temporary S/N : 09011234 serial number
repertory, then unzip it in a local repertory. Table T - 50C / + 200C
Plug the Texense USB Cable and wait for the automatic Then, every 200msec :
installation window. Thermocouple Temperature Tc and Ambient
Choose to install the driver : choose no for internet Temperature TA
option, then give the browser the name of your local 30 Ascii characters + Line Feed + Return
repertory (with unzipped files) Tc = + 159.2C TA = + 22.6C
Attention, 2 successive installations are needed, with Tc = + 159.2C TA = + 22.6C
the same procedure.
More info @ www.ftdichip.com/Documents/ * For Windows Vista, HyperTerminal is available at:
www.hilgraeve.com/htpe/download.html.
InstallGuides.htm

Analog Output Signal Ordering Information

Supply Range Sensivity THN-t-r-s-o-h


t = type K,J,T
-50 to +200C 0V at -50C 1V at 0C 5V at +200C 20mV/C r = range : 200, 400, 1250
s = supply :12 (5 to 16 V) or
-100 to +400C 0V at -100C 1V at 0C 5V at +400C 10mV/C 24 (15 to 33 V)
12V o = output : 5 or 10
(if different from standard)
0 to +1250C 0V at 0C 1V at 250C 5V at +1250C 4mV/C
h = housing :
P for plastics
-50 to +200C 0V at -50C 2V at 0C 10V at +200C 40mV/C
A for aluminium

-100 to +400C 0V at -100C 2V at 0C 10V at +400C 20mV/C Default values


24V
o = 5 if supply is 12 V
0 to +1250C 0V at 0C 2V at 250C 10V at +1250C 8mV/C o = 10 if supply is 24 V
h = P (plastics)

AXILANE
INSTRUMENTS
10 rue des Acacias
91620 Nozay

2/2 FRANCE
Tel. : +33 (0) 950 60 40 20
Fax : +33 (0) 955 60 40 20