Atlas Monitor

Service Manual

Atlas Patient Monitor

Welch Allyn
8500 SW Creekside Pl
Beaverton, Oregon 97008
6200-43E Revision D Copyright 2007
Part No. Rev. Description ECN# Date Approved
6200-43E A New release of Atlas Service Manual 5-40429 10/99 RS/LP
6200-43E B Updated calibration procedures 5-44782 07/02 DK
Updated performance verification procedures
Updated drawing 620150 Rev B to Rev C
Updated drawing 620201 Rev A to Rev B
Updated drawing 620524 Rev A to Rev B
Added complete repair parts lists.
Added calibration date table
6200-43E C Updated to reflect changes in hardware and software 5-45730 03/03 DK

Drawings and/or illustrations and/or part numbers contained in this


document are for reference purposes only. For current revisions call the
Welch Allyn Customer Service phone number listed in Section 1 page 2.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor i


CONTENTS
List of Figures ...................................................................................................................vi
List of Tables .....................................................................................................................ix

Section 1: General Information


About the Atlas Monitor........................................................................................................ 1
Help Information ................................................................................................................... 2
Agency Approvals ................................................................................................................. 3
Warning, Cautions and Notes................................................................................................ 3
Product Model Number Structure......................................................................................... 5
623xP Main Menu Architecture............................................................................................ 6
622xP Main Menu Architecture............................................................................................ 7
621xP Main Menu Architecture............................................................................................ 8
Atlas System Block Diagram ................................................................................................. 9

Section 2: Service
Incoming Inspection ............................................................................................................ 11
Calibration and Maintenance Schedule.............................................................................. 12
Setting Date and Time ......................................................................................................... 14
Pangea Communication Protocol ........................................................................................ 14
BP Calibration ...................................................................................................................... 14
50mmHg Calibration............................................................................................................ 17
250mmHg Calibration.......................................................................................................... 17
CO2 Reset (623xx Models Only).......................................................................................... 18
ET CO2 Calibration .............................................................................................................. 19
No-Load Battery Voltage Calibration .................................................................................. 20
Battery Voltage Calibration.................................................................................................. 21
Printer Print Adjustment ..................................................................................................... 22
Temperature Calibration...................................................................................................... 23
Calibration Date Set ............................................................................................................. 24
Explanation of an Atlas Service Screen.............................................................................. 25
Software Upgrade Procedure............................................................................................... 27
Down Loading NVRAM Files.............................................................................................. 28

ii Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Section 3: Functional tests and Trouble Shooting
Power on Self Test ...............................................................................................................37
Menu Functional Test..........................................................................................................38
BP Test..................................................................................................................................40
Static Manometer Accuracy Test.........................................................................................41
BP Leak Test ........................................................................................................................42
Over 15 mmHg Test .............................................................................................................42
BP Dump Verification Test ..................................................................................................43
Hardware Fail Safe Tests .....................................................................................................43
ECG/Respiration Test...........................................................................................................44
SpO2 Test..............................................................................................................................45
Temperature Test .................................................................................................................45
CO2 Functional Test.............................................................................................................45
Battery Functional Test........................................................................................................46
Printer Functional Test ........................................................................................................47
Print on Alarm Test..............................................................................................................47
Software/Firmware ..............................................................................................................47
ECG Alarms Test ..................................................................................................................48
Respiration Alarms Test ......................................................................................................50
SpO2 Alarm Test ..................................................................................................................51
Silence Alarm Test...............................................................................................................52
CO2/RESP Test .....................................................................................................................52
Blood Pressure Alarm Test ..................................................................................................53
Trouble Shooting..................................................................................................................56

Section 4: Disassembly and Repair


About Section 4 ...................................................................................................................63
Battery Door Removal..........................................................................................................64
Battery Removal...................................................................................................................64
Rear Housing Removal ........................................................................................................65
Printer Cable and Display Cable Removal ..........................................................................66
Power Supply Ground Wire Removal.................................................................................66
Tie Wrap Removal................................................................................................................67
CPU Removal .......................................................................................................................67

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor iii


Main PCB Removal .............................................................................................................. 67
Main PCB Side Connector Removal.................................................................................... 68
Main PCB Hose and Wire Removal..................................................................................... 68
Pump Removal..................................................................................................................... 68
CO2 Exhaust Removal.......................................................................................................... 69
Pryon PCB Removal............................................................................................................. 69
E-Pac Foam Removal (Top). ................................................................................................ 69
Power Supply Removal. ...................................................................................................... 70
Power Supply Fuse Location .............................................................................................. 70
Fan Removal ........................................................................................................................ 70
E-Pac Removal (Lower)........................................................................................................ 71
Deflection PCB Removal...................................................................................................... 71
CRT Removal ....................................................................................................................... 71
Display PCB Removal .......................................................................................................... 72
Printer Door Button Removal .............................................................................................. 73
Printer Assembly Locking Tab Release. .............................................................................. 73
Printer Positioning Slot Removal........................................................................................ 74
Printer Assembly Removal. ................................................................................................. 74
Printer Lip Assembly. .......................................................................................................... 74
Printer Cable Routing. ......................................................................................................... 75

Appendix A: Theory of Operation


Power Supply Circuits......................................................................................................... 77
Model 622 and Model 623................................................................................................... 80
On/Off Circuits .................................................................................................................... 81
Battery Charger .................................................................................................................... 84
NIBP Circuits ...................................................................................................................... 87
CRT Deflection Board .......................................................................................................... 92
Vertical Amplifier: ............................................................................................................... 92
Recorder Electronics............................................................................................................ 96
Power Supply 24 Switcher............................................................................................... 96
Motor Driver:........................................................................................................................ 97
Patient Isolated Circuits ...................................................................................................... 98
Linear Regulators and Filtering Model 621 ..................................................................... 99

iv Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Linear Regulators and Filtering Model 622 and 623 .....................................................100
A/D Circuits .......................................................................................................................101
A/D Multiplexer.................................................................................................................102
Serial Communication.......................................................................................................103
Signal Isolation ..................................................................................................................104
ECG Circuits.......................................................................................................................104
Impedance Respiration (Model 622/623 only) .................................................................109
SpO2 Circuits .....................................................................................................................110

Appendix B: Repair Parts Lists


621N0 .................................................................................................................................113
621NP .................................................................................................................................116
621S0..................................................................................................................................119
621SP..................................................................................................................................121
622N0 .................................................................................................................................124
622NP .................................................................................................................................127
622S0..................................................................................................................................130
622SP..................................................................................................................................133
623NP .................................................................................................................................136
623SP..................................................................................................................................142

Appendix C: Interconnect Diagram


Interconnect diagram 620396 Rev A.................................................................................145

Appendix D: Atlas Drawings and Electrical Schematics


See List of Figures on page v for individual drawings.....................................................146

Appendix E: Calibration Date Table


Calibration Date Index Table. ............................................................................................187

Appendix F: Serial Number Cut in History. ...............................................211

Appendix G: Atlas Calibration and Functional Test Form........................213

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor v


List of Figures
Figure 1-1. 623xP Main Menu Architecture. ........................................................................ 6
Figure 1-2. 622xP Main Menu Architecture. ........................................................................ 7
Figure 1-3. 621xP Main Menu Architecture. ........................................................................ 8
Figure 1-4. Atlas System Block Diagram .............................................................................. 9
Figure 2-1. Example of a new HyperTerminal connection. ............................................... 15
Figure 2-2. Choosing COM1 in HyperTerminal.................................................................. 15
Figure 2-3. HyperTerminal Port Settings. ........................................................................... 16
Figure 2-4. Photo of Atlas BP calibration setup. ................................................................ 17
Figure 2-5. Example of CO2 calibration setup. ................................................................... 21
Figure 2-6. Example and Explanation of an Atlas Service Screen .................................... 25
Figure 2-7. Pangea prompt in HyperTerminal .................................................................... 29
Figure 2-8. Choosing Send Text File in HyperTerminal................................................. 29
Figure 2-9. Choosing directory where NVRAM files are stored. ....................................... 30
Figure 2-10. Choosing the nvram_cal_init.txt file. ............................................................. 30
Figure 2-11. Pangea screen after opening the nvram_cal_init.txt file. .............................. 31
Figure 2-12. Pangea screen after opening the nvram_common.txt file. ............................ 31
Figure 2-13. Pangea prompt after opening the model# file................................................ 32
Figure 2-14. Pangea screen after loading a language file. .................................................. 33
Figure 2-15. Pangea screen after downloading the nvram_printer.txt file. ....................... 34
Figure 2-16. Pangea screen nvram set serial command. .................................................... 35
Figure A-1. Model 621 DC/DC circuit block diagram. ....................................................... 77
Figure A-2. On/Off Control Circuits. .................................................................................. 78
Figure A-3. Model 621 Buck converter. .............................................................................. 79
Figure A-4. Model 621 linear regulation. ........................................................................... 80
Figure A-5. DC/DC circuit diagram..................................................................................... 81
Figure A-6. Boost converter and off/on switch. ................................................................. 82
Figure A-7. Boost converter................................................................................................. 83
Figure A-8. Battery charging. .............................................................................................. 85
Figure A-9. Linear regulator. ............................................................................................... 86
Figure A-10. NIBP PWM A/D converter. ............................................................................ 86
Figure A-11. PWM timing. .................................................................................................. 87
Figure A-12. Primary transducer amplifier......................................................................... 89
Figure A-13. Safety transducer amplifier............................................................................ 90

vi Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Figure A-14. Over pressure fault circuit. ............................................................................91
Figure A-15. Pump relay circuit..........................................................................................91
Figure A-16. Ramp generator circuit...................................................................................92
Figure A-17. Vertical amplifier circuit................................................................................93
Figure A-18. Horizontal amplifier grid voltage circuit. .....................................................95
Figure A-19. Video amplifier circuit. ..................................................................................96
Figure A-20. Recorder power supply circuit. .....................................................................97
Figure A-22. Print head temperature circuit. .....................................................................98
Figure A-23. Patient isolation circuit..................................................................................99
Figure A-24. Model 621 line regulation and filtering. .......................................................99
Figure A-25. Model 622/623 line regulation and filtering. ..............................................100
Figure A-26. A/D converter circuit. ..................................................................................101
Figure A-27. A/D multiplexer circuit. ..............................................................................102
Figure A-28. Serial communication circuit. .....................................................................104
Figure A-29. Defib protect and RFI filtering circuit. ........................................................105
Figure A-30. Input buffer circuit.......................................................................................105
Figure A-31. Lead select circuit. .......................................................................................106
Figure A-32. Differential amp/slew rate amp circuit. ......................................................107
Figure A-33. Right leg drive circuit. .................................................................................107
Figure A-34. High pass and gain stage circuit. .................................................................108
Figure A-35. Temperature amplifier circuit......................................................................108
Figure A-36. Respiration drive circuit. .............................................................................109
Figure A-37. Respiration diff amp circuit.........................................................................110
Figure A-38. Gain and filter circuit...................................................................................110
Figure C-1. Atlas Interconnect Diagram............................................................................145
Figure D-1. Atlas model 200 main board..........................................................................147
Figure D-2. Atlas model 200 NIBP electronics. ................................................................148
Figure D-3. Atlas model 200 power supply......................................................................149
Figure D-4. Atlas Power isolation /SpO2 interface...........................................................150
Figure D-5. Atlas model 200 ECG amp. ............................................................................151
Figure D-6. Atlas ECG A/D interface.................................................................................152
Figure D-7. Atlas CPU subsystem. ....................................................................................153
Figure D-8. Atlas FPGA. ....................................................................................................154
Figure D-9. Atlas 200/210/220 I/O filters. ........................................................................155

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor vii


Figure D-10. Flash, SDRAM HRESET config word. ......................................................... 156
Figure D-11. Power PC MP C823 CPU. ............................................................................. 157
Figure D-12. Power distribution........................................................................................ 158
Figure D-13. Front panel display. ..................................................................................... 159
Figure D-14. Key board scanner. ....................................................................................... 160
Figure D-15. Temp/pulse LED drivers. ............................................................................. 161
Figure D-16. NIBP LED drivers. ........................................................................................ 162
Figure D-17. SpO2 LED drivers. ........................................................................................ 163
Figure D-18. CRT deflection board. .................................................................................. 164
Figure D-19. Printer electronics. ....................................................................................... 165
Figure D-20. Model 220 main board. ................................................................................ 166
Figure D-21. 210/220 NIBP electronics............................................................................. 167
Figure D-22. DC/DC power supply.................................................................................... 168
Figure D-23. Power isolation/SpO2 interface. .................................................................. 169
Figure D-24. 210/220 ECG amp......................................................................................... 170
Figure D-25. 210/220 ECG A/D interface.......................................................................... 171
Figure D-26. 210/220 Respiration circuit. ........................................................................ 172
Figure D-27. 210/220 serial communication. ................................................................... 173
Figure D-28. 200/210/220 transformer isolation. ............................................................. 174
Figure D-29. 200/210/220 ECG patient cable. .................................................................. 175
Figure D-30. 200/210/220 power supply sub assy. .......................................................... 176
Figure D-31. 220 ETCO2 PCB assembly. ........................................................................... 177
Figure D-32. 200/210/220 Power supply. ......................................................................... 178
Figure D-33. Nonin SpO2 Board........................................................................................ 179
Figure D-34. Motor Stepper. .............................................................................................. 180
Figure D-35. Pump Pneumatic. ......................................................................................... 181
Figure D-36. 200/210/220 ECG cable assembly. ............................................................... 182
Figure D-37. CO2 cable assembly. ..................................................................................... 183
Figure D-38. Nellcor sensor cable assy. ............................................................................ 184
Figure D-39. 200/210/220 Pneumatic sub assy................................................................. 185
Figure D-40. 220 CO2 cable assy. ...................................................................................... 186

viii Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


List of Tables
Table 1-1. Operator manual part numbers. ...........................................................................1
Table 2-1. Atlas Packing List. ..............................................................................................11
Table 2-2. Tools Required for Service. ................................................................................13
Table 2-3. Software Revision Table .....................................................................................26
Table 2-4. Files needed to download language(s)...............................................................33
Table 3-1. Transducer pressure table...................................................................................41
Table 3-2 Resistor and temperature reference. ...................................................................45
Table 3-3. Atlas failure trouble shooting.............................................................................56
Table 3-4. Atlas failure trouble shooting.............................................................................57
Table 3-5. Atlas failure trouble shooting.............................................................................58
Table 3-6. Atlas failure trouble shooting.............................................................................59
Table 3-7. Atlas failure trouble shooting.............................................................................60
Table 3-8. Atlas failure trouble shooting.............................................................................61
Table A-1. Temperature coefficient. ....................................................................................84
Table B-1. Atlas 621N0 repair parts ..................................................................................113
Table B-2. Atlas 621NP repair parts. .................................................................................116
Table B-3. Atlas 621S0 repair parts...................................................................................119
Table B-4. Atlas 621SP repair parts...................................................................................121
Table B-5. Atlas 622N0 repair parts. .................................................................................124
Table B-6. Atlas 622NP repair parts. .................................................................................127
Table B-7. Atlas 622S0 repair parts...................................................................................130
Table B-8. Atlas 622SP repair parts...................................................................................133
Table B-9. Atlas 622NP repair parts. .................................................................................136
Table B-10. Atlas 623SP repair parts.................................................................................139
Table B-11. Atlas 623NP repair parts. ...............................................................................142
Table D-1. Atlas drawings revision page...........................................................................146
Table F-1. Changed to a new BP Valve..............................................................................211
Table F-2. Changed E-Pac and Changed to class B power supply. ..................................211
Table F-3. Upgraded printer cable to a cable with foam tape. .........................................211
Table F-4. Upgraded Revision level of CPU from 501 to 502...........................................211
Table F-5. Upgraded Revision level of Nellcor SpO2 to 506............................................212

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor ix


x Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D
General Information Section 1

About the Atlas Monitor

The Atlas Monitor combines in one unit all the necessary measurements for patients under
anesthesia, for surgical recovery, or bed side monitoring. See Section 1, page 5 for a complete
listing of product models and options.

According to the standards of care for Nurse Anesthetists and Anesthesiologists, all patients
receiving conscious sedation are to be continuously monitored throughout the procedure and
recovery phase by ECG, SpO2, and NIBP. CO2 monitoring is a requirement during gas
anesthesia (when patient is ventilated).
The Atlas combines a CRT to display ECG, CO2, SpO2, and respiration waveforms. It utilizes
LEDs for the other numeric values to maximize visibility and viewing angle. Although not
designed to be a transport product, the monitor has an integral handle and it is small and
light enough at 13 lbs. to be easily moved.

A battery was added to enable the monitor to be used if there is a power outage or to be
unplugged momentarily when moved with the patient from the surgery room to the recovery
room. The battery was not designed to make the Atlas into a transport monitor. The Atlas
monitor should be plugged into AC as much as possible to give you the maximum battery
backup time when there is a power outage. It will maintain unit operation for up to an hour
when power is interrupted if the battery is fully charged.

WARNING: Discharging the battery frequently will shorten the battery


life and will shorten the battery backup time.

IMPORTANT: For a complete description on the function and use of the Atlas, as well as user
safety warnings, cautions, and warranty information, read and understand the Atlas
Operators Manual. See the Table 1-1 below for correct Operators Manual part #.

Table 1-1. Operator manual part numbers.


Language Part#
English 6200-42E
French 6200-42F
German 6200-42G
Italian 6200-42I
Spanish 6200-42S
Portuguese 6200-42P
Japanese 6200-42J
Chinese 6200-42C

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 1


Section 1 General Information

Help Information

To assure correct operation and performance all service and repairs must be performed by
fully trained and properly equipped personnel, using genuine replacement parts and correct
procedures. Failure to do so will also invalidate the product warranty.

Welch Allyn Inc. MD International


State Street Plant /Corporate Headquarters 11300 N.W. 41st Street
4341 State Street Road Miami, FL 33178 USA
Skaneateles Falls, NY 13153-0220, USA
tel: 305-669-9003
fax: 305-669-1971
tel: 1-800-535-6663
315-685-4100
fax: 315-685-3361
www.welchallyn.com/medical/support/tech

Welch Allyn GMBH Welch Allyn France SARL


Zollerstrasse 2-4 814 rue Charles de Gaulle
72417 Jungingen, Germany 77100 Mareuil les Meaux
France
tel: (011) 49-7477-927186
fax: (011) 49-7477-927193 tel: ( 011) 33 1.6009.3366
Fax: (011) 33 1 .6009.6797
Welch Allyn UK Ltd Welch Allyn Australia Pty
St. Georges House The Metro Centre Unit 5
Vernon Gate, Derby 38-46 South Street
England DE1 1UQ Rydalmere NSW 2116, Australia
Postal address:
tel: (011) 44-01332-206208 PO Box 132
fax: (011) 44-01332-206209 Rydalmere NSW 1701, Australia
mob: (011) 44-0467-301814
tel: (011) 612 9638-3000
fax: (011) 612 9638-3500

Welch Allyn Singapore Ltd China Service Center


300 Beach Road, #25-08 Room 708 Central Plaza
The Concourse No. 277 Huang Pi Bei Rd.
Singapore 199589, Singapore Huang Pi District
Shanghai 200003
tel: (011) 65-291-0882 China
fax: (011) 65-291-5780
tel: (011) 86-21-63279631
fax: (011) 86-21-63279632
Welch Allyn Japan K.K.
Bon Marusan 8F, 3-5-1 Kanda
Jinbo-Cho, Chiyoda-Ku
Tokyo 101-0051
Japan

tel: (011) 813-5212-7391


fax: (011) 813-3261-7372

2 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


General Information Section 1

Agency Approvals

ETL Listed
U L 2 6 01- 1 ,
C S A C 2 2 . 2 N o . 6 01. 1
I E C 6 0 6 01- 1 , A S 3 2 0 0 . 1
I E C 6 0 6 01- 1- 2
C LI
STED US
74227

The CE mark on this product indicates that it


has been tested to and conforms with the
provisions noted within the 89/336/ECC
Electromagnetic Compatibility Directive.

Australia EMC Framework Compliance

N344
Warning, Cautions and Notes
All operating and service personnel should be familiar with the general safety information
in this summary. Specific warnings and cautions will also be found throughout the opera-
tors manual. Such specific warnings and cautions may not appear here in the summary.

Electrical Shock Hazard

CAUTION

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 3


Section 1 General Information

ATTENTION. Consult accompanying document

Handle With Care

95%
Storage Humidity. Refer to technical specifications in operator's
for more details.
MAX

Storage temperature. Refer to technical specifications in operator's


manual for more details.

Lead Acid Battery. For disposal see maintenance


section in opertor's manual
PB

4 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


General Information Section 1

Product Model Number Structure

621S0-E1 ECG, Nonin SpO2, NIBP

621SP-E1 ECG, Nonin SpO2, NIBP, Printer

621N0-E1 ECG,Nellcor SpO2,NIBP

621NP-E1 ECG,Nellcor SpO2,NIBP, Printer

622S0-E1 ECG, Nonin SpO2, NIBP, Temp, Respiration, Battery, RS232

622SP-E1 ECG, Nonin SpO2, NIBP, Temp, Respiration, Battery, RS232, Printer

622N0-E1 ECG, Nellcor SpO2, NIBP, Temp, Respiration, Battery, RS232

622NP-E1 ECG, Nellcor SpO2, NIBP, Temp, Respiration, Battery, RS232, Printer

623SP-E1 ECG, Nonin SpO2, NIBP, ETCO2, Temp, Respiration, Battery, RS232, Printer

623NP-E1 ECG, Nellcor SpO2, NIBP, ETCO2, Temp, Respiration, Battery, RS232, Printer

Product Structure Meaning:

The first three digits in the product structure sequence designates the model number. The
fourth character in the sequence designates the SpO2. The fifth character in the sequence
designates if it has a printer or not. The first suffix designates the country language when
shipped. The second suffix designates the power cord shipped with the Atlas.

First Three Characters: Model number, 621, 622, 623

Fourth Character: N=Nellcor SpO2, S=Nonin SpO2

Fifth Character: P=Printer, 0=No Printer

First Suffix: E = English, F= French, G= German, I= Italian, S= Spanish, P= Portuguese C =


Chinese,
J= Japanese

Second Suffix:

1 = US, Canada, Japan


2 = European
4 = United Kingdom
6 = Australian

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 5


Section 1 General Information

623xP Main Menu Architecture

Figure 1-1. 623xP Main Menu Architecture.

623xP

Date/Time
Waveform Screen
screen
Trend Advanced Configuration Screen
Screen Language
Silence duration
ECG gain
ECG lead set
ECG speed
ECG bandwidth
Set date & time
Initial pressure
MAP
Second trace selection
Advanced Temperature units
Configuration Respiration speed
CO 2 units
Date/Time
Co2 Print on alarm (only with software Rev.
Screen BB.2.2000 or higher)
Reset
Service Mode Screen
Save Verify manometer
Settings Reset to factory defaults
Battery test
Calibrate Co 2
Service Mode
Screen CRT test pattern
Printer test pattern (Models 621SP and
LED test 621NP only)
Button test
Display A/D channels *

NOTE: The Print on alarm function in Advanced configuration mode screen is only
available with software version BB.2.2000 and higher.

* NOTE: The Display A/D channels function in the Service Mode Screen only allows
you to view system information by pressing the SET button. It does not allow
you to change system settings.

6 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


General Information Section 1

622xP Main Menu Architecture

Figure 1-2. 622xP Main Menu Architecture.

622xP

Date/Time
Waveform Screen
screen
Trend Advanced Configuration Screen
Screen Language
Silence duration
ECG gain
ECG lead set
ECG speed
ECG bandwidth
Set date & time
Initial pressure
MAP
Second trace selection
Advanced Temperature units
Configuration Respiration speed
Print on Alarm (Only with software Rev.
Date/Time BB.2.2000 and higher.)
Screen

Service Mode Screen


Save Verify manometer
Settings Reset to factory defaults
Battery test
Service Mode CRT test pattern
Screen Printer test pattern (Models 622SP and
622NP only)
LED test
Button test
Display A/D channels *

NOTE: The Print on alarm function in Advanced Configuration Screen is only


available with software version BB.2.2000 and higher.

NOTE: On models 622S0 and 622N0 delete the Print test pattern function in the
Service Mode Screen. Model 622S0 and 622N0 do not come with printers.

* NOTE: The Display A/D channels function in the Service Mode Screen only allows
you to view system information by pressing the SET button. It does not allow
you to change system settings.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 7


Section 1 General Information

621xP Main Menu Architecture

Figure 1-3. 621xP Main Menu Architecture.

621xP

Date/Time
Waveform Screen
screen
Trend Advanced Configuration Screen
Screen Language
Silence duration
ECG gain
ECG lead set
ECG speed
ECG bandwidth
Set date & time
Initial pressure
MAP
Second trace selection
Advanced Print on Alarm (Only with software Rev.
Configuration BB.2.2000 and higher.)

Date/Time
Screen

Service Mode Screen


Save Verify manometer
Settings Reset to factory defaults
CRT test pattern
Printer test pattern (Models 621SP and
Service Mode 621NP only)
Screen LED test
Button test
Display A/D channels *

NOTE: The Print on alarm function in Advanced configuration mode screen is only
available with software version BB.2.2000 and higher.

NOTE: On models 621S0 and 621N0 delete the Print test pattern function in the
Service Mode Screen. Model 622S0 and 622N0 do not come with printers.

* NOTE: The Display A/D channels function in the Service Mode Screen only allows
you to view system information by pressing the SET button. It does not allow
you to change system settings.

8 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


5 Volt Video Signals:
4M X 16 H-Sync, V Sync, Video CRT SubSystem
SDRAM

CO2 Sensor
CO2 SubSystem
512 X 16 Flash
Local System Bus
Address,Data,Control
Configuration Motor
EEPROM Driver Motor
General Information

Service Manual 6200-43E Rev. D


Bus Config.
Register AT90A4414
8 bit
Print Head
Atlas System Block Diagram

Controller
Clock, Load
Strobe
SPI

Motorola CPU Printer


MPC823Z3 GPIO Subsystem

3-wire SPI + nine


Timer 2
chip select
UART-1 UART-2 GPIO GPIO
P
ON,OFF W
TTL TTL
2 ACON, PW
RS-232 RS-232
SYNC

Front End
Figure 1-4. Atlas System Block Diagram

RS-232 SPO2 NIBP Power Front Panel


Front End
Drivers Sub-System Pnuematic Supply EKG +
(Isolated) (Isolated) Control Temp
NIBP
Drivers (Isolated) Key LED
Board Drivers
Battery
Rs-232 SPO2 Pump Valve
Test Sensor
Port AC Primary EKG 15 X 7
Temp 24
Line Sensor X5 Keys 31 LEDs
Section 1

Welch Allyn Atlas Monitor 9


Service Section 2

Incoming Inspection
NOTE: Use the following guidelines when unpacking the monitor from its shipping carton.

1. Before opening the monitor shipping carton, check for damage.

2. If damage is apparent, stop unpacking the carton and contract the shipping company for
further instructions. If the carton is intact, unpack the monitor.

3. With the monitor out of its carton, check to see that all the items listed on the packing
slip are in the shipping carton. See table 2-1 below.

Table 2-1. Atlas Packing List.


Qty. Description
1 Warranty card
1 Adult Durable 1pc. Cuff
1 Large Adult Durable 1pc. Cuff
1 5ft. Straight Hose
1 ECG Patient Cable (Three Lead)
3 ECG Lead Wires
1 SpO 2 Sensor/ Finger Clip
1 ETCO 2 Water Trap 623 only
1 ETCO 2 Scrubber 623 only
1 Nasal Canula 623 only
1 Atlas Monitor
1 Detachable Power Cord
1 Printer Paper Roll
1 Operation Manual
1 Skin Temperature Probe 622/623 only

4. If an item is missing, first check the carton, then check with your receiving department. If necessary
contact Welch Allyn at the address and phone number shown on Section 1.

5. Clean and disinfect by following the instructions printed in the Operator Manual.

NOTE: Perform all functional tests as listed in Section 3 before and after servicing.
Operate the Atlas to verify the customer complaint before making any changes to the unit.
Call the customer if the complaint is unclear.

If the unit has caused or is suspected of having caused an injury of any type: DO NOT
DISASSEMBLE OR REPAIR THE UNIT IN ANY WAY. Contact Welch Allyn Customer
Service immediately.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 11


Section 2 Service

Calibration and Maintenance Schedule


The Atlas Monitor must be serviced by authorized Welch Allyn personnel or agents at 6
month intervals. Maintenance requirements are specified for 6 month and 12 month service
intervals. The monthly CO2 Reset operation can be performed by the user.

Service Interval Maintenance Requirements

Every 6 months: CO2 Calibration

Every 12 months: BP calibration, CO2 reset, battery voltage calibration,


printer adjustment,and temperature calibration.
Complete functional test

12 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

Table 2-2. Tools Required for Service.

Description Part# Company


100cc Test Volume T-112189 Welch Allyn
500cc Test Volume T-112854 Welch Allyn
T-10 TORX screwdriver XTD-10 Xcelite brand
7/16" deep socket for Temperature Port nut Generic 1/4" drive
Squeeze Bulb and Valve 5088-01 Welch Allyn
Calibrated Manometer (0-10PSIG) Digimano 1000 Netech
Bio-Tek (NIBP TESTER) BP Pump Bio-Tek
Pneumatic Tubing 5200-19 Welch Allyn
Pneumatic Tubing (coiled) 5200-19M Welch Allyn
"Y" Fitting "Optional" 9586TPK4 Welch Allyn
"T" Fitting (3) 9858TPK4 Welch Allyn
Nonin Patient Simulator 8000S Nonin
Nonin Cable 5200-52 Nonin
Calibrated Thermometer for 90F to 115F range 1002-3FC ERTCO
LG, Adult Cuff/Bag 5200-02 Welch Allyn
Nellcor Patient Simulator For atlas with "MP
204/205 SpO2 PCB ONLY " SRC-2 Nellcor
Nellcor Patient Simulator "MP 506 PCB ONLY " SRC-MAX Nellcor
Nellcor Sensor Cable "Purple Connectors" will work
on all Atlas units with 204/205/506 SpO2 PCB DEC-8 Nellcor
Nellcor Sensor Cable "Gray Connectors" "For Atlas
units with MP 204/205 SpO2 PCB only" EC-8 Nellcor
ECG Simulator with Impedance Respiration 214B DNI Nevada
ECG Patient Cable (5 lead AHA) 6200-02 Welch Allyn
ECG Patient Cable (5 lead IEC) 6200-04 Welch Allyn
ECG leads, 5 Lead (IEC) 6200-08 Welch Allyn
ECG leads, 5 Lead AHA 6200-06 Welch Allyn
Certified Gas "10% CO2,10% O2, balance N2" 0304724SRBD Scott Medical Products
ETCO2 Water Trap (package of 5) 6200-20 Welch Allyn
ETCO2 Scrubber 6200-21 Welch Allyn
ETCO2 Adult Nasal Sample Line 6200-22 Welch Allyn
Surface Sensor, Temperature 6200-15 Welch Allyn
Atlas Repair and Calibration Software 620538 Welch Allyn
Printer Paper - Case 6200-40 Welch Allyn
Digital Multimeter with 10mV accuracy on a 10V
scale; 10A Range
Adjustable DC power supply 5A @ 7V
Battery cable assembly 620174-1 Welch Allyn
Atlas interface cable to PC 6200-60 Welch Allyn
PC with Windows 95 and above and with Part of Windows 95 &
HyperTerminal serial port software above
1/4" Mono Phono Jack
1000 Ohm precision resistor 1%
1200 Ohm precision resistor 1%
1350 Ohm precision resistor 1%
1540 Ohm precision resistor 1%
1870 Ohm precision resistor 1%

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 13


Section 2 Service

Setting Date and Time


NOTE: Check date and time before doing any calibration. Set date and time if incorrect. Set
time and date as follows;

1. Turn Atlas on.

2. Press CLOCK button next to power on button to check date and time. Use the far right
SELECT button to scroll. Highlight the date or time that needs changed.

3. Press the SET button to adjust date and time.

4. Press CLOCK button to exit.

Pangea Communication Protocol


The Pangea Communication Protocol allows interaction of the Atlas with the computer
through the serial interface port. A prompt is emitted at the computer screen when the
instrument powers up and is ready to accept commands. The prompt is Pangea>. A prompt
is emitted after the completion of each pangea command. Pangea commands are case sensi-
tive. They are in English only.

BP Calibration
Required material.
1. 500cc vessel 4. Calibrated digital manometer
2. Squeeze bulb with one-way valve 5. Tubing and T fittings
3. PC with HyperTerminal 6. Serial cable

NOTE: To start HyperTerminal you must have Windows7 95 or higher installed on your
computer.

1. From the Windows main screen follow the following sequence:


Start !Programs ! Accessories ! Communication ! HyperTerminal

2. HyperTerminal setting are:


9600 baud rate, 8 bit word, 1 stop bit
no parity, no flow control
ANSI character set

14 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

3. When you open HyperTerminal you will see a screen similar to the example in Figure
2-1. You will be prompted for a connection description. Choose any name. After you
type in a name click OK.

Figure 2-1. Example of a new HyperTerminal connection.

4. The next window you will see will be the window as shown in Figure 2-2. Click on
Connect using then click on COM1.

Figure 2-2. Choosing COM1 in HyperTerminal.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 15


Section 2 Service

5. Set the port settings as shown in the example in Figure 2-3. Now click OK.

Figure 2-3. HyperTerminal Port Settings.

6. Connect the pressure meter, bulb, and 500cc vessel to BP port with T connectors as
shown in photograph Figure 2-4.

7. Connect the Atlas to PC with serial cable.

8. Turn the Atlas on. Start HyperTerminal on PC. Press the <Enter> key and you should see
a Pangea> prompt.

NOTE: Take no more than 3 minutes for the 50mmHg calibration nor more than 3 minutes
for the 250mmHg calibration as the Atlas will automatically, as a safety feature, open the
blood pressure valve. If this happens you will have to turn the Atlas off then back on again
and restart the calibration again.

16 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

T112854

Pressure
Meter T Fittings Calibrated 500cc Volume

Squeeze Bulb

Figure 2-4. Photo of Atlas BP calibration setup.

50mmHg Calibration
1. Enter the following commands at the Pangea prompt.
Pangea> bp valve close <ENTER>
Pangea> bp safety off <ENTER>
Pangea> bp cal 5000 Do not press <ENTER> yet!

2. Raise the pressure with bulb to as close to 50.00mmHg as possible. Now press <ENTER>.

3. Release the pressure.

250mmHg Calibration
1. Enter the following command
Pangea> bp cal 25000 Do not press <ENTER> yet!

2. Raise the pressure with bulb as close to 250.00 mmHg as possible. Now press < ENTER >

3. Enter the following command to save the calibration in the Atlas.


Pangea> nvram write <ENTER>

4. Release the pressure.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 17


Section 2 Service

CO2 Reset (623xx Models Only)


Required material. 1. Watertrap
2. Scrubber
NOTE: The Scrubber looks similar to a watertrap, but it is filled with white granules. The
scrubber is included with the 623XX models only.

NOTE: Make sure date and time are correct before performing the CO2 reset.

1. Turn Atlas on. Make sure the watertrap and scrubber are NOT attached to the Atlas.

2. Press the DATE/TIME button on the lower right of the monitor. The Set Date and Time
and Other Options menu will be displayed.

3. Press the CO2/RESP ALARMS Off button. The CO2 Reset screen will appear.

4. You will see the following messages on the CRT.


CO2 Reset
Install CO2 scrubber
Press Trend to abort

5. Install the watertrap to the Atlas. Install the scrubber to the watertrap.

6. You will see the following instructions on the screen.


Warming up will be flashing on CRT.
May take up to five minutes on CRT.
Press Trend to abort on CRT.

7. After about 5 minutes you will see on the CRT.


CO2 Reset
Reset complete
Remove CO2 scrubber
Press the trend button to exit

8. Remove the CO2 watertrap and scrubber.

9. Press TREND button to return to idle screen.

Replace watertrap after every six hours of use. Treat watertrap and used CO2
sample lines as bio hazard material!

18 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

ET CO2 Calibration
Required material:
1. Tank of approximately 10% CO2, balance N2 (certified) Blood Gas Mixture.
2. Tubing and T connectors.
3. Watertrap and scrubber.

1. Make sure the watertrap and scrubber are not attached to the Atlas. Turn the Atlas on.

2. Place the instrument into the Service Mode by pressing the DATE/TIME button. Make
sure date and time are correct. Press the LEAD SELECT button.

3. Press SELECT button and scroll down to Calibrate CO2.


The message Install CO2 Scrubber will appear on the right side of screen.

4. Attach the scrubber to the water trap.

5. Insert the scrubber/water trap assembly into water trap socket. The message
Enter span gas value using Set button 10% will appear.

6. Press the SET button to change the value of span gas being used. The factory default
value is 10%. Calibrate with a 8% to 12% certified CO2 concentration known to be
0.01%).

7. The message press BP Start/Cancel will appear at the bottom right of CRT. Press the
BP/Start/Cancel button.

8. If you receive a Calibration Failed message at this point, check the date. If date is
2022 or above it will fail CO2 calibration.

9. Next you will see a message Warming up. After the Atlas warms up you will see a
message Attach CO2 gas. Remove the scrubber from the CO2 water trap. Do not
remove water trap.

10. Attach the certified source of CO2 gas to the CO2 side-stream sampling tube as per
Figure 2-5 below.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 19


Section 2 Service

Adjust Regulator to
Approximately 2 psi

Water
Trap
Vent
CO2

Figure 2-5. Example of CO2 calibration setup.

CAUTION: IMPROPER USE,STORAGE OR HANDLING OF COMPRESSED


GAS VESSELS CAN CAUSE DEATH OR INJURY. FOLLOW GAS MANU-
FACTURES SAFETY PROCEDURES!

11. Adjust the CO2 regulator just enough to allow a small amount of gas to flow out of the
vent (approximately 2 psi).

12. Press the BP START/CANCEL button. The message: Sampling will appear on CRT.
CO2 calibration successful or CO2 calibration failed will appear on the CRT
display.

13. Press the TREND button to exit.

No-Load Battery Voltage Calibration


NOTE: No-Load Battery Voltage Calibration procedure is for models 622 and 623 only. If
you disconnect the battery you must reset time and date after you reconnect the battery.

Specifications: No load battery charge 6.85 VDC

Required materials: DVM

1. Remove the battery and disconnect the battery leads from the Atlas. Use the DVM to
measure across the connectors, red+ and black-.

2. Adjust the voltage to 6.85 VDC by turning potentiometer R338, located behind the right
battery jack. Turning clockwise will increase the no-load voltage and counter clockwise
will decrease the no-load voltage.

20 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

Battery Voltage Calibration


Required materials.
1. DC power supply rated: 7 VDC at 5A
2. DMM / DVM with 10mV resolution on a 10 DC volt scale.
3. PC with HyperTerminal
4. Serial interface cable

1. Connect the serial cable to the PC and Atlas.

2. Hyper Terminal Settings are:


9600 Baud, 8 bit word, 1 stop bit
no parity, no flow control
ANSI character set
Find HyperTerminal in Windows 95 or higher
Start !Programs!Accessories ! Communication! HyperTerminal

NOTE: Make sure the Atlas IS NOT plugged into AC for this calibration procedure.

3. Remove and disconnect the battery from Atlas.

4. Set the power supply to 6.8VDC " 200mV and connect the power supply to the battery
connector on the Atlas.

5. Turn Atlas on..

6. Reduce the power supply to 6.0VDC and measure the voltage at the battery connector
(at the Atlas) to the nearest 10mV.

NOTE: Do not measure at the power supply, since cable resistance will introduce error.

7. At the HyperTerminal prompt type:


Pangea> power cal XXXX <ENTER>

NOTE: XXXX represents the measured voltage in millivolts no decimal point. For
example, if you measured 6.010VDC at the battery connector, use the command power cal
6010 <ENTER>.

8. The Atlas will respond: raw = ZZZZ mV true = 6010 mV OK

NOTE: ZZZZ is the raw uncalibrated reading that the instrument made.

9. Reduce the power supply to 5.6 VDC. You should soon hear the low battery alarm.

10. Measure the voltage at the battery connector to the nearest 10mV.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 21


Section 2 Service

11. At the HyperTerminal prompt type:


Pangea> power cal XXXX <ENTER>

NOTE: XXXX represents the measured voltage in millivolts with no decimal point. For
example if you measured 5.590 volts at the battery connector, then you would enter the
command power cal 5590 <ENTER>.

12. The Atlas will respond:


raw = ZZZZmV true = 5590 OK

NOTE: ZZZZ is the raw uncalibrated reading that the instrument made.

13. Finish the calibration by typing:


Pangea> hw reset <ENTER>

NOTE: This will re-boot the Atlas.

14. Turn the Atlas off and reinstall the battery.

15. Reset time and date.

Printer Print Adjustment

1. Install new paper.

2. Turn Atlas on.

3. Attach an ECG simulator to Atlas and set simulator for a heart rate of 60 bpm, normal
sinus rhythm.

4. Press PRINT button. Evaluate the darkness of waveform and text printout.

5. If either need to be changed press DATE/TIME button then press LEAD SELECT button
to access Advanced Configuration menu.

6. Press SELECT button and scroll down to Printer Test Pattern. Then Press
HR ALARMS OFF button

7. Two lines will be displayed:


1. Waveform+128
2. Text+ 78

NOTE: These two numbers are the factory defaults and are a good starting point if the sys-
tem is printing poorly or not at all.

22 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

8. The left SET button controls the waveform darkness and the right SET button controls
the text darkness.

NOTE: Pressing the SET button up will increase the number and darken the waveform
while pressing the SET button down will decrease the number and will lighten the wave-
form.

9. Make changes to the printout as needed.

10. Press Trend button to return to waveform screen.

Temperature Calibration
Required Material:
1. PC with Windows7 95 or higher 4. 1/4O mono phono jack
2. Atlas serial cable 5. Soldering iron and solder
3. 1k Ohm, 1/2 watt precision resistor 6. Ohm Meter

1. Solder the 1 K Ohm resistor to 1/4 mono phono jack.

2. Measure the resistance at the tip of the phono jack. Record that resistance reading, to two
decimal points.

3. Plug the phono jack into the Atlas.

NOTE: The Atlas will show a temperature reading in the temperature display.

4. Start HyperTerminal on PC.

5. Hyper Terminal Settings are:


9600 Baud, 8 bit word, 1 stop bit
no parity, no flow control
ANSI character set
Find HyperTerminal in Windows7 95 or higher
Start !Programs ! Accessories ! Communication ! HyperTerminal

6. At the Pangea prompt Type: PANGEA> temp cal XXXXXX <ENTER>

NOTE: XXXXXX is the resistance reading you measured and recorded in milliohms at the
tip of the mono phono jack.

Example: If you measure 1000.40 ohms at the tip of the phono jack then you would type
temp cal 100040 <Enter>.

7. Wait four seconds then type: PANGEA> temp state <ENTER>. You will see a value
returned at the Pangea prompt.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 23


Section 2 Service

8. Verify that the resistance given by the above command returns a value " 0.5 ohms.

9. Verify temperature accuracy as outlined in Chapter 3, Table 3-2.

Calibration Date Set

NOTE: After calibrating the Atlas you must reset the calibration date. The calibration date
is the date you performed the calibration. The calibration date appears in the Service Mode
menu. See Figure 2-6 for an explanation of the Service screen.

1. At the Pangea prompt type.

Pangea> nvram set cal_date XXXX <ENTER>

NOTE: XXXX is the number of days from January 1, 1998 until the present date. See
APPENDIX E for that number or calculate manually.

24 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

Explanation of an Atlas Service Screen


Service Mode
Select a test
Verify
f manomete
Reset to factory default 0.00
Battery test
Calibrate CO2
CRT test pattern
Printer test pattern
LED test
Button test
Display A/D channels
A
Model 220 with printer Cal:27 Nov 2002
B Sn: 62304064 Sw: BB.025000 Boot: AA.01.0000
C SpO2: NELLCOR MP506 V1.8.0 6/21/02
CO2 1.0 V3.00 #2542 Cal: 27 Nov 2002 14 Nov 2002
D
Press the Trend button to exit
Figure 2-6. Example and Explanation of an Atlas Service Screen

Line A Indicates that the Atlas is a model 220 and that the Atlas has a printer. If the Atlas
did not have a printer then the text with printer would be absent. The date field following
indicates the date the Atlas was last calibrated.

Line B starts with a numeric sequence. The first three digits are the model number of the
Atlas. The next five numbers are the Atlas serial number. The next sequence,alphanumeric,
indicates what software version is currently loaded into the Atlas. The third sequence,
alphanumeric, indicates what version boot software is currently loaded into the Atlas.

Line C starts with the SpO2 OEM board used in the Atlas. There are two SpO2 OEM boards
used in the Atlas. One is Nellcor and the other is Nonin. The next sequence indicates the
model of the SpO2 board. The next sequence, that starts with a letter V, is the version soft-
ware used with the current SpO2 board. The date following the SpO2 software is the date
the OEM loaded the software into the SpO2 board. SpO2 OEM software can not be upgraded.
If the most current software is needed, you will need to replace the SpO2 board.

Line D is the CO2 information if your Atlas has CO2 installed. Only models 623 have CO2
installed. If your Atlas does not have CO2 installed then line D will be absent. If your Atlas
has CO2 installed then line D will start with CO2 followed by a numeric number. That
numeric value is the software loaded into the CO2 board. The next alphanumeric sequence,
starting with a V, is the version of that software. The next numeric sequence starting with
a # is the serial number of the CO2 board. The next sequence, a date, is the date the CO2
was last calibrated. The last sequence, a date, is the date the CO2 was reset. CO2 OEM soft-
ware can not be upgraded. If the current software is needed, you will have to replace the
CO2 board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 25


Section 2 Service

Table 2-3. Software Revision Table.

Model Operating System Boot loader Nellcor Nonin Pryon


621S0 AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V7
621S0 BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V7
621S0 BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
621S0 BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V7

621N0 BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V1.8.1.0, 10/14/02

621SP AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V7


621SP BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V7
621SP BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
621SP BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V7

621NP BB.02.5000, 10/14/02 AA.01.0000, 6/20/02 V1.8.1.0, 10/14/02

622S0 AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V7


622S0 BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V7
622S0 BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
622S0 BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
622S0 BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V7

622SP AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V7


622SP BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V7
622SP BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
622SP BB.02.5000, 6/14/02 AA.01.0000, 6/20/99 V7

622N0 AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97


622N0 BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97
622No BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97
622N0 BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V 1.8.1.0 10/14/02

622NP AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97


622NP BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97
622NP BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V 1.2.0.0 10/14/02
622NP BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V 1.8.1.0 10/14/02

623SP AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V7 0.E V1.00


623SP BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V7 0.E V1.00
623SP BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V7
622SP BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V7

623NP AA.01.4000, 9/8/99 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97 0.E V1.00


623NP BB.02.0000, 4/4/01 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97 0.E V1.00
623NP BB.02.2000, 6/4/02 AA.01.0000, 6/20/99 V 1.2.0.0 12/17/97
623NP BB.02.5000, 10/14/02 AA.01.0000, 6/20/99 V 1.8.1.0 10/14/02

26 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

Software Upgrade Procedure


NOTE: The following procedures are required to upgrade the software on a fully function-
ing Atlas or to reload software after replacing the CPU board. The download utility
atlas_dl.exe loads the following files automatically.
1. atlas.out.gz
2. nvram_common.txt
3. nvram_(model#).txt
4. nvram_(language).txt

Atlas_dl.exe will also query the Atlas to determine what model number the Atlas is and
what language to download.

Equipment or supplies required:


1. PC with Windows7 95 or higher
2. Atlas serial cable
3. File: atlas_dl.exe (Included in the Atlas Repair Software listed in Table2-2.)

NOTE: Make sure you have HyperTerminal turned off or the following utility download will
not work!

1. Run the program atlas_dl.exe from the CD or copy the file to your hard drive and run the
program from there.

2. Connect the serial cable between the Atlas and the PC sCOM1 port.

3. Turn your computer on and locate the atlas_dl.exe file.

4. Double left click on the atlas_dl.exe file.

5. When the file starts to download, the CRT will go blank on the Atlas.

NOTE: Do not use the computer while the program is downloading.

6. After down loading is complete, check all alarm settings and all user advanced
configuration settings since these are RESET by this utility software download
procedure.

NOTE: Stop here if you are just upgrading software on a fully functional Atlas monitor or if
you have replaced the CPU board.

NOTE: If you have replaced the MAIN BOARD then you MUST continue with the next
procedure (DOWNLOADING NVRAM FILES).

NOTE: Perform the following NVRAM downloading procedure if you have replaced the
MAIN BOARD. The NVRAM resides on Main Board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 27


Section 2 Service

Down Loading NVRAM Files


Required Materials.
1. Computer with Windows7 95 or higher.
2. The latest Atlas repair software. Call your local Welch Allyn
representative.

1. Once you have procured the latest Atlas Repair Software, you will be able
to run the programs straight from the CD or copy the files to your hard drive.

NOTE: Through HyperTerminal you will need to load the following files:
1. nvram_cal_init.txt
2. nvram_common.txt
3. model# of Atlas
3.1.You will load nvram200.txt if your Atlas is a model 621xx.
3.2.You will load nvram210.txt if your Atlas is a model 622xx.
3.3 You will load nvram220.txt if your Atlas is a model 623xx.
4. language.txtAny combination of or all of the following files
4.1 nvram_english.txt for the English language
4.2 nvram_french.txt for the French language
4.3 nvram_german.txt for the German language
4.4 nvram_spanish.txt for the Spanish language
4.5 nvram_potuguese.txt for the Portuguese language
4.6 nvram_italian.txt for the Italian language
4.7 nvram_chinese.txt for the Chinese language
4.8 nvram_japanese.txt for the Japanese language
5. printer.txt (if fitted with a printer), or no_printer.txt. (if not fitted with a printer)

NOTE: After loading the preceding files you, must then type the following commands at
the Pangea prompt.
1. nvram set serial xxxxx where xxxxx is the serial number of the Atlas monitor.
2. nvram write writes the information to memory.
3. hw reset performs a hardware reset.

2. Connect the serial cable between the Atlas and the PC COM1 port.

3. Open HyperTerminal program on PC.


Hyper Terminal Settings are:
9600 Baud, 8 bit word, 1 stop bit
no parity, no flow control
ANSI character set
Find HyperTerminal in Windows7 95 or higher
Start !Programs ! Accessories ! Communication ! HyperTerminal

28 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

4. Turn Atlas on. You should see the Pangea prompt. See example in Figure 2-7.

Figure 2-7. Pangea prompt in HyperTerminal.

5. Scroll over to Transfer and then scroll down and choose Send Text File.
See example in Figure 2-8.

Figure 2-8. Choosing Send Text File in HyperTerminal.

6. Another window will then appear and prompt you for the location of the files. Double
left click on that folder to open that directory. You can also run these programs from the
CD. See example in Figure 2-9.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 29


Section 2 Service

Figure 2-9. Choosing directory where NVRAM files are stored.

7. Once the folder is open, open the file nvram_cal_init.txt file by double left clicking on
that file or by high lighting the file and then click on the Open button. See example in
Figure 2-10.

Remember that after you have download the files in this section and you must
complete this entire Downloading NVRAM files section, that you must perform
a complete calibration on the Atlas monitor.

Figure 2-10. Choosing the nvram_cal_init.txt file.

30 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

NOTE: After the nvram_cal_init.txt has executed you should see a Pangea screen similar to
window as shown in Figure 2-11.

Figure 2-11. Pangea screen after opening the nvram_cal_init.txt file.

8. From your pangea window choose Transfer then choose Send Text File.

9. Open the directory where the file nvram_common.txt is located or open from the CD.

10. Double left click on nvram_comm.txt file or highlight the file and then choose Open.

NOTE: After opening the nvram_common.txt file you should see a screen similar to the
window shown in Figure 2-12.

Figure 2-12. Pangea screen after opening the nvram_common.txt file.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 31


Section 2 Service

NOTE: Next you will download the model# of the Atlas by choosing only one of the three
following files.

1. If you have an Atlas model 621 then you will only download the file nvram200.txt.
2. If you have an Atlas model 622 then you will only download the file nvram210.txt.
3. If you have an Atlas model 623 then you will only download the file nvram220.txt.

11. Open the directory where the model number files are located or open from CD.

12. Open the file by double left clicking on the file that represents the model number of the
Atlas you are working on or highlight that file then choosing Open.

NOTE: After opening the Atlas model# file you should see a Pangea screen similar to the
window as shown in Figure 2-13.

Figure 2-13. Pangea prompt after opening the model# file.

32 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

NOTE: Next you will download the language(s) that you want the Atlas to store in its
Advanced Configuration menu for languages. You can have one or all the languages loaded
in the Atlas. Listed in Table 2-4 are the language(s) choices and the files you will need to
download to have the language(s) loaded in the Atlas. Each language you want loaded in
the Atlas will require that the file associated with that language be loaded.

Table 2-4. Files needed to download language(s).

Language File needed to download


English nvram_english.txt
French nvram_french,txt
Spanish nvram_spanish.txt
German nvram_german.txt
Portuguese nvram_portuguese.txt
Italian nvram_italian.txt
Chinese nvram_chinese.txt
Japanese nvram_japanese.txt

13. Open the directory where the language files are located or open from CD.

14. Double left click on language.txt file or highlight the file and then choose Open.

NOTE: After you download a language.txt file you should see a Pangea screen similar to
Figure 2-14.

Figure 2-14. Pangea screen after loading a language file.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 33


Section 2 Service

NOTE: Next you tell the Atlas if it does or does not have a printer.
1. If the Atlas has a printer you will download the file nvram_printer.txt.
2. If the Atlas does not have a printer then you will download the file
nvram_no_printer.txt.

15. Open the directory where the printer.txt files are located or open from CD.

16. Double left click on the printer file or the no_printer file or highlight the printer file or
the no_printer file and then choose Open.

NOTE: After you have downloaded the printer or no_printer file you should see a Pangea
screen similar to the window as shown in Figure 2-15.

Figure 2-15. Pangea screen after downloading the nvram_printer.txt file.

NOTE: The next three commands will require that you actually type the command at the
Pangea prompt.

NOTE: XXXXX on line 17 is the Atlas 5 digit serial number. The serial number is located on
the bottom of the Atlas. Make sure that you have a space between the command serial and
the five numbers you are entering.

17. At the Pangea prompt type the following command nvram set serial XXXXX<ENTER>

NOTE: After you have typed the nvram set serial XXXXX command and hit ENTER you
should see a Pangea screen similar to the example in Figure 2-16.

34 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Service Section 2

Figure 2-16. Pangea screen after you have downloaded the nvram set serial command.

NOTE: To finish downloading the NVRAM files, type the next two commands at the Pangea
prompt.

18. At the Pangea prompt type the command nvram write <ENTER>.

19. At the Pangea prompt type the command hw reset <ENTER>.

NOTE: The hw reset command will reboot the Atlas.

Remember that after you have download the files in this section and you must
complete this entire Downloading NVRAM files section, that you must perform
a complete calibration on the Atlas monitor.

NOTE: The firmware is not up grade able on the OEM boards (SpO2 and CO2) boards. If a
higher version software is needed, then you will need to replace the OEM board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 35


Troubleshooting Section 3

NOTE: Review customer complaint and determine if it is safe to plug in the Atlas monitor
into AC power and if it is safe to turn on!

Power on Self Test


1. Plug in Atlas into AC power with no sensors attached. Verify that the AC~ LED indictor
is lit. Install paper in printer if fitted.

2. Turn on power and verify the following:


1. Green light in power button.
2. Loud beep when Power On button pressed.
3. Three dashes in SYSTOLIC.
4. Three dashes in DIASTOLIC.
5. Two dashes in SpO2 (takes several seconds after BP dashes come on).
6. Three dashes in PULSE (takes several seconds after BP dashes come on).
7. Single bar at the bottom of Plethysmograph.
8. No lights in TEMP.
9. No lights in ALARMS OFF buttons.
10. X lit on AUTO.
11. CRT display may comes on slowly if cold or quickly if still warm from last use.
12. May see version string in center if comes on quickly. It is not a problem if not seen
because it comes on slowly.

3. On the CRT you should see:


1. Three dashes for Heart Rate
2. Heart symbol
3. Lead select symbol
4. Scale bar
5. One or two lines of dashes for waveforms depends upon settings.
6. Three dashes for MAP or blank, depending upon settings.
7. Error message(s) at bottom of screen?

4. On models 622 or 623, disconnect AC cable. You should see no change except the
AC~ LED indicator will not be lit. If you see warnings of a low, very low or depleted
battery, or if the system turns off by its self, plug the Atlas back into the AC outlet and
let the unit charge with unit turned off. Repeat test in 2 hours.

5. Plug the monitor back into AC.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 37


Section 3 Troubleshooting

Menu Functional Test


1. Press DATE/TIME button to enter menu page. Press TREND button to access the
Advanced Configuration menu.

2. Write down all configuration settings so that they can be reset to customer preferences
later. If you have a printer press the PRINTER button to print out customer configuration
settings.

3. Set language to your native tongue if necessary to allow you to write down the other
settings. The top item is always the language, press either SET buttons to step through
list.

4. Press TREND button to return to idle screen. Press DATE/TIME button and verify that
date and time are correct. Set date and time if necessary.

NOTE: An incorrect date may indicate a battery problem. If date was incorrect, turn off
unit, disconnect AC power cable, wait 5 minutes then reconnect the power cable. Turn Atlas
on. Check date again. If date comes back incorrect on model 621 then replace main board.
On model 622 and 623, if battery was not dead then replace main board.

5. Press DATE/TIME button to return to waveform screen if needed.

6. Press DATE/TIME button then press LEAD SELECT button to access service mode

7. Version/configuration data is in lower half of menu, write it all down.

8. Press SELECT button to highlight reset to factory defaults.

9. Press BP START/CANCEL button to reset configuration.

10. Press TREND button to return to idle screen.

11. Press DATE/TIME button then press TREND button to access the Advanced
Configuration menu.

12. Set language to your native tongue if necessary

NOTE: We have just reset the monitor to factory defaults. Compare settings to factory
defaults appropriate for the country in Operator manual. If the factory defaults do not
match then that may indicate a memory problem. Changing only the language should not
change any of the other factory default settings.

13. Press DATE/TIME button then press LEAD SELECT button to access service mode.

14. Press SELECT button to highlight CRT test pattern.

38 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

15. Press BP START/CANCEL button to show test pattern. Examine display. Press any key to
end display.

16. Press Select button to highlight Printer test pattern (if fitted). Press BP START/CANCEL
button to start test pattern. Printer should print a test pattern. Press any key to end
display. Examine printout.

17. Press SELECT button to highlight LED test. Press AUTO button to turn on all LEDs.
Press BP START/CANCEL button to show automatic test pattern. Watch for a while,
look for glitches in pattern. Press SET button to go to manual mode and step through
individual segments if needed to observe a problem.

18. Press SELECT button to highlight Button test. Press BP START/CANCEL button to
start test. Press every button on system, press BP START/CANCEL button last.
Verify that buttons match up with their names and that all buttons are functional.

NOTE: If name does not correspond to the button pushed then this may indicate memory
corruption. You may need to replace the main board.

19. Press SELECT button to highlight Display A/D channels. Press SET button and write
down all values for each of the four screens for later review. Press TREND button to
return to Idle screen.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 39


Section 3 Troubleshooting

BP Test
1. Press DATE/TIME button then press TREND button to access the Advanced
Configuration menu. Press SELECT button to highlight Initial pressure. Press SET button
to change Initial pressure to 280 mmHg. Press SELECT button to highlight MAP. Press
SET button to change MAP to Yes. Press TREND button to return to idle screen.

2. Connect the BP port to the BP simulator. Set the simulator for a normal reading
(140/80, 100BPM, NSR). Press BP START/CANCEL button. Atlas should start to
pump and display a manometer value in SYSTOLIC LED. This value should track
and be very close to pressure displayed by manometer on BP simulator (if fitted).
Largest number shown in SYSTOLIC LED should be very close to the initial pressure
setting recorded above from Advanced Configuration. System should start stepping
down pressure, showing step values in SYSTOLIC LED display. It then should display
the correct SYSTOLIC and DIASTOLIC values. System may show MAP value depending
upon country language setting.

3. Press BP START/CANCEL button. System should start pump, display manometer


value in SYSTOLIC LED. This value should track and be very close to pressure
displayed by manometer in BP simulator (if fitted). Largest number shown in
SYSTOLIC should be very close to the Initial pressure setting of 280 mmHg. If
pressure shown exceeds 300 mmHg then re calibrate the Atlas BP.

4. System should start to step down pressure, showing step values in SYSTOLIC LED and
then display correct SYSTOLIC and DIASTOLIC values. System should show MAP
value. MAP value should match what is shown by simulator.

5. Set simulator to (120/70, 60 (beats per minute)bpm, NSR).

6. Press the AUTO button. The X manual BP cycle indicator will go out and the 1 minute
auto BP cycle indicator will flash for 10 seconds. 20 seconds after the 1 minute auto BP
cycle indicator stops flashing, BP measurement starts.

7. Verify the BP reading. If BP measurement incorrect recalibrate.

8. No less than 30 seconds after completing the measurement another measurement


should start. While it is pumping up, press the BP START/CANCEL button.
Measurement should stop immediately and pressure should be dumped (as seen on
the simulator manometer).

9. Press AUTO button and X manual BP cycle indicator should light (not flashing).

10. Disconnect the tubing from the BP port on the Atlas. Press BP START/CANCEL
button and note the time (to the second) when the Atlas aborts the BP cycle and gives
you an alarm. It should take no longer than 1 minute. Turn Atlas off.

40 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Static Manometer Accuracy Test


Required materials.
1. Serial interface cable
2. PC with HyperTerminal (Windows 95 or above)
3. 500cc vessel with inflation system.
4. Calibrated manometer

1. Turn computer on and activate Hyper terminal.


Hyper Terminal Settings are:
9600 Baud, 8 bit word, 1 stop bit
no parity, no flow control
ANSI character set
Start !Programs ! Accessories ! Communication ! HyperTerminal

2. Turn Atlas on. Refer to Figure 2.1 in Section 2 for inflation setup.

3. Enter the following commands at the PANGEA> prompt.


PANGEA> bp safety off <ENTER>
PANGEA> bp valve close <ENTER>

4. Pressurize the Atlas monitor to the applied pressures in Table 3-1.

5. To compare the applied pressure to the transducer readings use the following command.
PANGEA> bp press <ENTER>

Table 3-1. Transducer pressure table.

Applied Primary Safety


Pressure Transducer Transducer
0 mmHg "0.75 mmHg "1.0 mmHg
50 mmHg "1.5 mmHg "1.5 mmHg
150 mmHg "1.5 mmHg "4.5 mmHg
250 mmHg "1.5 mmHg "7.5 mmHg

6. Verify that the primary transducer and secondary transducer are within the
specifications listed in Table 3-1.

NOTE: The Atlas monitor will return a response to the computer screen as follows.
If you tested at the 50mmHg pressure you would receive a message similar to: px = 5279,
sx = 5305 x 0.01. Where px is the primary transducer and 5279 = 52.79 mmHg and sx is the
secondary transducer and 5305 = 53.05 mmHg.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 41


Section 3 Troubleshooting

BP Leak Test
NOTE: Unit must not leak more than 5 mmHg in a 15 second interval while attached to a
100cc test cavity pressurized at 50 mmHg, 150 mmHg and 250 mmHg.

Required Materials: 100cc test vessel (+10cc/-0cc), stopwatch, squeeze bulb, calibrated
manometer.

1. Connect Atlas to 100cc test vessel and calibrated manometer as shown in Section 2,
Figure 2.1.

2. Turn Atlas on.

3. Press DATE/TIME button then press LEAD SELECT button to access the Service Mode.

4. Press SELECT button and choose Verify manometer.

5. Pressurize the Atlas with squeeze bulb to 50mmHg and allow reading to stabilize.

6. Observe pressure for 15 seconds.

6. Unit should not leak more than 5mmHg during this time.

7. Repeat this process at 150mmHg and 200mmHg.

Over 15 mmHg Test


NOTE: Atlas hardware must detect if the pneumatic system has been pressurized greater
than 15 mmHg for more than 155 seconds but less than 180 seconds.

Required Materials: 500cc test vessel (+10cc/-0cc), stopwatch, squeeze bulb, calibrated
manometer.

1. Connect Atlas to 500cc test vessel and calibrated manometer as shown in Section 2,
Figure 2.1.

2. Turn Atlas on. Press DATE/TIME button then press LEAD SELECT button to access the
Service Mode. Press SELECT button and choose Verify manometer.

3. Pressurize the Atlas with squeeze bulb to any pressure above 15 mmHg and maintain
that pressure during the test. Start the stop watch.

4. Verify that the valve opens and pressure drops between 155 seconds and 180 seconds.

5. Turn Atlas off.

42 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

BP Dump Verification Test


1. Connect the Atlas to a 500 cc vessel as shown in Section 2, Figure 2.1.

2. Turn Atlas on.

3. Press DATE/TIME button then press LEAD SELECT button to access the Service Mode.

4. Select Verify manometer.

5. Pressurize the Atlas to just over 260 mmHg.

6. Allow reading to stabilize for 15 seconds.

7. Ensure the reading is still over 260 mmHg.

8. Start the stop watch when you press the SELECT button. When the SELECT button is
pressed the Atlas will start a pressure dump.

9. Verify that the pressure drops to below 15 mmHg within 10 seconds.

Hardware Fail Safe Tests


NOTE: Atlas hardware must detect over pressure on unit pneumatic system between 296.0
mmHg and 330.0 mmHg.

1. Connect the Atlas for testing as shown in Section 2, Figure 2.1

2. Turn Atlas on.

3. Press CLOCK/DATE button then press LEAD SELECT button to access Service Mode.

4. Press SET button to highlight the Manometer.

5. Pump system up and record at what pressure the Atlas dumps.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 43


Section 3 Troubleshooting

ECG/Respiration Test
NOTE: Simulator should support ECG impedance respiration.
1. Cycle power on Atlas, connect a 5 lead ECG cable set to the simulator. Configure the
simulator for NSR, 100 bpm and set the impedance respiration rate to 20.

2. Press DATE/TIME button then press TREND button to access Advanced Configuration
menu.

3. Press SELECT button to highlight ECG lead set and press SET button to select 5 wire.

4. Set ECG gain to Automatic, ECG speed to 25mm/s, ECG bandwidth to Monitor, Second
trace selection to ECG.

5. Press TREND button to return to idle screen.

6. Verify you see: 1. ECG cascading onto second line


2. Scale bar on top left of CRT
3. Heart rate displayed as set on simulator
4. Lead Selected = II
5. Pulse tone high pitched

7. Press LEAD SELECT button and step through each of the lead settings.
(I,II,III,aVR,aVL,aVF,V)

8. Verify that you see a different ECG waveforms for each lead selected.
Heart Rate will go to dashes and alarms on some leads.

9. Set lead selected to II.

10. Press DATE/TIME button then press TREND button to access Advanced Configuration
menu.

11. Change Second trace selection to Respiration. Press TREND button to return to idle
screen.

12. Verify that you see:


1. ECG on top line
2. Scale bar on left of top line
3. Heart rate displayed as set on simulator
4. Lead Selected = II
5. Pulse tone high pitched
6. Respiration trace on second line and rate as set on simulator
7. Respiration rate displayed on CRT as set on simulator.

13. Disconnect ECG simulator.

44 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

SpO2 Test
1. Turn SpO2 alarms on.

2. Connect SpO2 cable to simulator.

3. Set simulator pulse rate to 60 beats/minute and verify that pulse rate is 60
3 beats/minute or 3%, whichever is greater.

4. Set SpO2 between 70% and 100% and verify that SpO2 is accurate 2 digits.

Temperature Test
1. Without anything connected to the temperature connector on the Atlas, verify that the
temperature LED is blank.

2. Using Table 3-2 below, connect the resistors to a 1/4 inch male mono plug and verify
temperature readings.
Table 3-2 Resistor and temperature reference.

3. Connect the temperature probe to the Atlas monitor. You should see the display read
room ambient temperature. Disconnect the probe and verify you see dashes is the
display and that there are no alarms.

CO2 Functional Test


1. Press DATE/TIME button then press TREND button to access Advanced Configuration
menu

2. Set Second trace selection to CO2.

3. Press TREND button to return to idle screen.

4. Insert watertrap with tubing attached.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 45


Section 3 Troubleshooting

5. Verify that you: 1. Hear pump motor start.


2. See solid line waveform on lower trace.
3. See dashes in Respiration Rate.
4. See dashes in mmHg (or % or kPa, as configured).

6. Breathe gently and repeatedly over end of tubing.

7. Verify that you:


1. See waveform within seconds of breathing.
2. See respiration rate non zero within one minute.
3. See CO2 concentration non zero within one minute.

Battery Functional Test


1. System must have been plugged in for 24 hours to guarantee that the battery is fully
charged. But, for now we can do a functional test as follows.

2. Press DATE/TIME button then press LEAD SELECT button to access Service Mode
menu.

3. Press SELECT button to highlight Battery test.

4. Verify that the menu reports: 1. Battery Low Time XXX


2. Battery Dead Time YYY

NOTE: These are the results from the last battery test. The Battery Low Time is the time in
hours and minutes that the battery ran in the last test until the Low Battery alarm started,
and the Battery Dead Time is the time from the beginning of the Low Battery Alarm until the
system turned itself off when the battery voltage reached the cutoff level.

NOTE: 2:08 means 128 minutes which is the default setting indicating a battery test has
never been made before.

5. Write down the Battery Low Time and Battery Dead Time.

6. Unplug AC cord to start battery test.

7. The timers will begin the moment you unplug the AC cord. Leave the system on until it
powers down by itself automatically. Plug Atlas into AC and turn the Atlas on, enter the
Service Mode menu, select Battery Test again, and write down the new values. Compare
these to the previous values, and to the minimum specification:
1. Battery Low Time = 1 Hour
2. Battery Dead Time = 10 Minute minimums

8. Replace the battery if performance falls below specification.

46 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Printer Functional Test


NOTE: Configuration settings for printing are different for text pages than for waveforms.

1. Connect ECG simulator to generate a sample waveform.

2. Press PRINT button and look at waveform printout. Look for darkness, thickness of
lines, legibility of text, blurring, blooming of text.

3. If waveform or printed text needs to be darker or lighter, go to Section 2, page 10 for


printer calibration.

Print on Alarm Test


1. Turn the Atlas on.

2. Press the DATE/TIME button then press the TREND button to access Advanced
Configuration menu. Press the SELECT button and scroll down to Print on Alarm. Press
the SET button and set Print on alarm to No.

3. Connect an ECG simulator to the Atlas. Set the ECG rate on the simulator to 60 bpm,
normal sinus rhythm.

4. Press left SELECT button and the HI heart rate alarm parameter will flash. Set the high
heart rate alarm to 110 bpm on the Atlas and press the SELECT button again and the low
heart parameter will flash. Press SELECT button one more time to return to the default
screen.

5. Set the ECG simulator to heart rate to 120 bpm.

6. Verify that you hear the audio alarm and that the printer is not printing.

7. Set the ECG simulator to 60 bpm and make sure that the alarm condition has cleared.

8. Press the DATE/TIME button then press the TREND button to access Advanced
Configuration menu. Press the SELECT button and scroll down to Print on Alarm. Press
the SET button and set Print on alarm to Yes.

9. Set the ECG simulator to 50 bpm.

10. Verify that the printer printed a 15 second strip.

Software/Firmware
1. Review versions written down earlier and compare to latest available. Make sure that all
components are compatible with each other. See section 2, page 26, Table 2-3.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 47


Section 3 Troubleshooting

ECG Alarms Test


1. Turn the Atlas on. Verify that you hear the power on tone.

2. Press DATE/TIME button then press TREND button to access Advanced Configuration
menu. Press SELECT button and scroll down to Second Trace. Press SET button and set
the second trace to ECG.

3. Connect ECG simulator to the Atlas. Set the ECG rate on simulator to 60 bpm,
normal sinus rhythm.

4. Verify you hear the heart rate beep at a constant high pitch.

5. Press SpO2 - + button - 8 times. Should get quieter and finally silent.

6. Press DATE/TIME button then press PRINT button to save settings.

7. Turn system off and back on.

8. Verify pulse tone is silent even though heart rate is shown.

9. Press SpO2 volume button + 8 times. Should get audible tone and then louder.

10. Disconnect ECG cable.

11. Verify you hear audio alarm and see error message ECG Lead fault and that you see the
three LED bars flash where the heart rate was on the CRT.

12. Press HR ALARMS Off button.

13. Verify that the audio alarm stops and that the visual alarm on the CRT remains.

14. Press HR ALARMS OFF button. Connect ECG simulator to the Atlas and set simulator
to normal sinus, 60 bpm.

15. Press left SELECT button and the HI heart rate alarm parameter will flash. Set the high
heart rate alarm to 55 bpm on the Atlas and press the SELECT button again and the low
heart parameter will flash. Press SELECT button one more time to return to the default
screen.

16. Verify that you hear the audio alarm and that you see the numeric value flashing in
the upper right portion of the monitor.

17. Return the high alarm limit to 120 bpm and press SELECT button. Verify the
audio alarm should is silent and the numeric visual alarm is flashing.

18. Press SELECT button twice. Verify the numeric low alarm parameter is flashing.

48 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

19. Set the LOW alarm for 65 bpm by pressing the up SET button. Press SELECT
button.

20. Verify that you hear the audio alarm and that you see the numeric value is flashing in
the upper right portion of the monitor.

21. Turn the Atlas off.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 49


Section 3 Troubleshooting

Respiration Alarms Test


1. Turn Atlas on.

2. Press DATE/TIME button then press TREND button to access Advanced Configuration
menu. Press SELECT button to scroll down to Second Trace. Press SET button and set
second trace to respiration. Press TREND button to return to default screen.

3. Set respiration on simulator to 20 respirations/minute and connect to the Atlas.

4. Press left SELECT button three times and set the hi respiration alarm to less than 20.
Wait 10 seconds and the CRT will return to the default screen.

5. Verify that you hear the audio alarm and that the respiration numeric value on the CRT
is flashing.

6. Press CO2/RESP ALARMS off button.

7. Verify that the audio alarms are silent, the visual numeric alarm is flashing and
that the red LED in the CO2/RESP ALARM off button is lit.

9. Press left SELECT button three times to access hi respiration alarms and set hi alarm to
30.

10. Wait ten seconds and the CRT will return to the default screen.

11. Press CO2/RESP ALARM off BUTTON then press left SELECT button four times to
access low respiration alarms and set low respiration alarms to 25.

12. Wait 10 seconds and CRT will return to the default screen.

13. Verify that you hear the audio alarm and that the respiration numeric value on the CRT
is flashing.

14. Press CO2/RESP ALARMS off button.

15. Verify that the audio alarms are silent, the visual numeric alarm is flashing and
that the red LED in the CO2/RESP ALARM off button is lit.

16. Turn Atlas off.

50 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

SpO2 Alarm Test


1. Turn Atlas on.

2. Attach SpO2 to a simulator and set for 95% SpO2.

3. Verify that you hear a pulse rate beep and that it is a different pitched tone than the
ECG pulse beep.

4. Change SpO2% setting on the simulator. Verify that tone changes pitch up or down
tracking simulator setting.

5. Press lower far right SELECT button until SpO2 LO is flashing, press right SET button
UP to change SpO2 LO setting to 99.

6. Wait until the SpO2 LO LED stops flashing.

7. Verify that you hear the SpO2 LO limit alarm.

8. Press Alarm Volume button - eight times. Verify tones gets quieter but not silent.

9. Press Alarm Volume button + eight times. Verify the tone gets louder.

10. Disconnect SpO2 cable from simulator and verify that the visual alarm SpO2 cable
not detected is displayed on the bottom left side of CRT. Verify that you hear the
audio alarm.

11. Press SpO2 ALARMS Off button.

12. Verify that the audio alarm sound is silent, that you do not see the visual alarm
message SpO2 cable not detected on the CRT and that the LED in the SpO2
ALARMS OFF button is on.

13. Turn Atlas off.

14. Disconnect SpO2 cable from Atlas.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 51


Section 3 Troubleshooting

Silence Alarm Test


1. Turn Atlas on.

2. Press DATE/TIME button then press TREND button to access Advanced


Configuration menu. Press SELECT button until Silence Duration is highlighted.
Press SET button until 90 seconds is highlighted. Press TREND button to go to
default screen.

3. Connect ECG simulator to the Atlas. Set the ECG rate on simulator to 60 bpm, normal
sinus wave.

4. Press left SELECT button once and set the HI alarm to 55 bpm.

5. Wait until you hear the audio alarm.

6. Press SILENCE button and start stopwatch. Verify that the alarm returns at 90
seconds.

7. Turn Atlas off.

CO2/RESP Test
1. Turn Atlas on.

2. Press DATE/TIME button then press TREND button to access the Advanced
Configuration menu.

3. Set Second trace selection to CO2.

4. Press TREND button to return to idle.

5. Insert CO2 watertrap with hose.

6. Breath into hose until a waveform is displayed.

7. Remove CO2 watertrap .

8. Verify you hear the audio alarm and that you see in the visual alarm CO2 watertrap not
detected in the bottom left portion of CRT.

9. Press the CO2/RESP ALARMS off button. Verify that the audio alarm goes silent, the
visual alarm goes out and the red LED in the CO2/RESP ALARMS off button is lit.

10. Turn Atlas off.

52 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Blood Pressure Alarm Test


1. Turn Atlas on.

2. With no hose connected to the BP port, press BP Start/Cancel button.

3. Place finger over BP port. This will block the flow causing the Atlas to detect an over
pressure.

4. Verify that you hear the audio alarm, that the visual alarm Check Blood Pressure Cuff
appears in the lower left portion of the CRT and that the SYSTOLIC and DIASTOLIC
LED displays are flashing.

5. Press the BP ALARMS OFF button.

6. Verify the audio alarm is silent, the visual alarms have turned off and that the red LED in
the BP ALARMS OFF button is lit.

7. Press the BP ALARMS OFF button then press the BP START/CANCEL button twice.

8. Press BP START/CANCEL button and let Atlas run without anything attached to the BP
port.

9. Verify at 60 seconds that you hear the audio alarm and that the visual alarm Check
Blood Pressure Cuff appears in the lower left portion of the CRT and that the SYSOLIC
and DIASTOLIC LED displays are flashing.

10. Press the BP ALARMS OFF button.

11. Verify the audio alarm is silent, the visual alarms are still off and that the red LED in
the BP ALARMS OFF button is lit.

12. Press the BP ALARMS OFF button then press the BP START/CANCEL button twice.

13. Connect a BP simulator to the Atlas BP port. Set the simulator for a normal reading
(140/80, 100 bpm, normal sinus rythym).

14. Press the right SELECT button twice and the current SYSTOLIC HI alarm value will be
flashing.

15. Press the SET button to set SYSTOLIC HI alarm to 130.

16. Press BP START/CANCEL button.

17. Verify that you hear the audio alarm and that the numeric value in the SYSTOLIC LED
display is flashing.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 53


Section 3 Troubleshooting

18. Press the BP ALARMS OFF button.

19. Verify the audio alarm is silent, the numeric value in the SYSTOLIC LED display is still
flashing and that the red LED in the BP ALARMS OFF button is lit.

20. Reset the SYSTOLIC HI alarm to 200 (factory default). Press BP ALARMS OFF button.

21. Press the right SELECT button three times and the numeric value in the SYSTOLIC LO
alarm LED display will be flashing.

22. Press the SET button to set the SYSTOLIC BP LO alarm to 160.

23. Press BP START/CANCEL button.

24. Verify that you hear the audio alarm and that the numeric value in the SYSTOLIC LED
display is flashing.

25. Press the BP ALARMS OFF button.

26. Verify the audio alarm is silent, the numeric value in the SYSTOLIC LED display is still
flashing and that the red LED in the BP ALARMS OFF button is lit.

27. Reset SYSTOLIC LO alarm to 70 (Factory default). Press BP ALARMS OFF button.

28. Press right SELECT button four times and the numeric value in DIASTOLIC HI alarm
LED display will be flashing.

29. Press the SET button and set DIASTOLIC HI alarm to 70.

30. Press BP START/CANCEL button.

31. Verify that you hear the audio alarm and that the numeric value in the DIASTOLIC HI
LED display is flashing.

32. Press the BP ALARMS OFF button.

33. Verify the audio alarm is silent, the numeric value in the DIASTOLIC LED display is still
flashing and that the red LED in the BP ALARMS OFF button is lit.

34. Reset DIASTOLIC hi alarm to 155 (Factory Default).

35. Press right SET button five times and the numeric value in the DIASTOLIC LO alarm
LED display will be flashing.

36. Press SET button and set DIASTOLIC LO alarm to 90. Press BP ALARMS OFF button.

37. Press BP START/CANCEL button.

54 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

38. Verify that you hear the audio alarm and that the numeric value in DIASTOLIC LED
display is flashing.

39. Press the BP ALARMS OFF button.

40. Verify the audio alarm is silent, the numeric value in the DIASTOLIC display is still
flashing and that the red LED in the BP ALARMS BUTTON is lit.

41. Reset DIASTOLIC LO alarm to 50 (Factory Default).

42. Turn Atlas off.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 55


Section 3 Troubleshooting

Table 3-3. Atlas trouble shooting.

Failures Possible Solutions


Power on:
Unit does not power up, no AC on LED, 1 Check AC power cord.
no Power on LED.
2 Verify connection from power supply to main
board.
3 Verify +12V output from power supply.
4 Replace Power Supply.
5 Check Cable from main board to display board.
6 Replace Display board.

Unit does not power up - AC and Power 1 Check that the CPU board is fully seated.
on LED on.
2 Replace CPU board.
3 Check for conformal coating in J4 header - replace
board if coating is found.
4 Replace Main board.
5 Replace Display board.

Unit powers up intermittently or Power 1 Remove flashing on Bezel at power button.


button sticks or cannot duplicate power
on problem But unit passes burn in.
2 Check for severely tipped power button LED on
display board.

Unit powers up - but no image on CRT. 1 Check for conformal coating in J4 header - replace
board if coating is found.
2 Replace CPU board.
3 Check for correct installation of connection from
main board to deflection board. If installed with
pin 1 in the pin 2 socket deflection board will be
damaged. If installed with pin 1 out of socket
deflection board will be OK.
4 Replace CPU assembly.

56 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Table 3-4. Atlas trouble shooting.

Failures Possible Solutions


Display:
Segment out, LED out, Buttons do not 1 Replace Display board.
work.

All or most segments stay on at power 1 Replace CPU board.


up.
2 Check cable from Display board to Main board.
3 Replace Power Supply.

CRT:
CRT image rotated or poor alignment. 1 Rotate yoke and/or adjust magnets.

Only see a dot in the center of the 1 Check cable from yoke to CRT board.
screen.

Only see a horizontal line in the center 1 Check cable from yoke to CRT board.
of the screen.
2 Replace Deflection board.

No image - CRT to Main board 1 Check flex-cable from Deflection board to tube.
connection OK.
2 Check high voltage anode connection.
3 Replace Deflection board.
4 Replace tube.
5 Replace CPU board.

No image - Glow at bottom of screen. 1 Check for conformal coating in J4 header - replace
board if coating is found.
2 Replace CPU board.

Dark vertical scan lines in image. 1 Replace CPU board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 57


Section 3 Troubleshooting

Table 3-5. Atlas trouble shooting.

Failures Possible Solutions


Printer:
Motor does not run. 1 Check that the unit is configured for printer. (CRT
image will freeze for 10 seconds if configured for
no printer.)
2 Check cable from printer PCB to motor.
3 Check cable from printer PCB to Main board.
4 Check motor gear alignment to first plastic gear.
5 Check for paper jam.
6 Replace printer PCB.
7 Replace CPU.

Motor runs or chatters - paper does not 1 Check for missing dowel pin in roller.
advance.
2 Check for missing bearings/gears.
3 Replace printer PCB.
4 Replace CPU.

Light print or no print. 1 Check that paper is properly installed (thermal


side visible).
2 Check cable from printer PCB to print head.
3 Check printer-strobe-width in NVRAM (could
indicate error in printer thermistor circuit). If low,
check printer PCB, Main board, and CPU board.
4 Check mechanical assembly for proper force of
printhead on paper.

Prints gibberish. 1 Reseat CPU board.


2 Replace CPU board.
3 Replace Printer to Main board cable.
4 Replace printer PCB.

58 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Table 3-6. Atlas trouble shooting.

Failures Possible Solutions


Safety:
Earth resistance over limit. 1 Check ground wire from motor to power supply
can.
2 Tighten screws that attach ground tabs to power
supply can.
3 Tighten screws holding power supply PCB to can.
4 Tighten quick connects on AC harness to power
supply.
5 Replace power supply PCB.

Blood Pressure:
Leak test - Fails Leak Specification 1 Run individual test. If unit passes repeat all tests.
2 Check seals @ tubing in BP assembly.
3 Check for lifted pins on the CPU connector JP8.
4 Replace Check valve.

Leak test - Cant pump beyond xxx.xx. 1 Verify unit can pump beyond 300 mmHg then:
Retest.
2 If pump will not reach 300 mmHg, then:
A. Check for correct direction on check valve.
B. Check for possible leaks.
C. Check for pinched tubing.
D. Replace pump.

Manometer Accuracy - Cant pump 1 Verify unit can pump beyond 300 mmHg then:
beyond xxx.xx. Retest.
2 If pump will not reach 300 mmHg, then:
A. Check for correct direction on check valve.
B. Check for possible leaks.
C. Check for pinched tubing.
D. Replace pump.
3 Check for lifted pins on the CPU connector JP8.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 59


Section 3 Troubleshooting

Table 3-7. Atlas trouble shooting.

Failures Possible Solutions


Manometer Accuracy - Fails Accuracy 1 Replace Main board. Faulty transducer or amp.
Specification.

Deflation Profile. 1 Replace dump valve.


2 Replace Main board. Valve drive circuitry fault.

Overpressure Limit - Cant pump 1 Verify unit can pump beyond 300 mmHg then:
beyond xxx.xx. Retest.
2 If pump will not reach 300 mmHg then:
A. Check for correct direction on check valve.
B. Check for possible leaks.
C. Check for pinched tubing.
D. Replace pump.

Overpressure Limit - Fails Accuracy 1 Replace Main board. Fault with pressure
Specification. transducer, amplifier, or comparator.
2 Replace CPU board.

Dwell time. 1 Replace Main board. - Fault with 15 mmHg detect


circuit.
2 Replace CPU board. - Fault with timer.

Dwell Pressure. 1 Replace Main board- If fault with 15 mmHg detect


circuit.
2 Replace CPU board.

60 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Troubleshooting Section 3

Table 3-8. Atlas failure trouble shooting.

Failures Possible Solutions


SpO2:
No Signal - SpO2 fault message on CRT. 1 Plug in finger cuff, connect to your finger & check
for function. If function is normal retest. If not
continue to step 2.
2 Check cables - SpO2 board to Main board, SpO2
board to Bezel.
3 Replace Main board.
4 Replace SpO2 board.
5 Replace CPU board.

SpO2 signal accuracy. 1 If Nellcor, check ribbon cable and harness. Replace
as needed.
2 Replace SpO2 board.

ETCO2:
Specification limits fail - Flow rate or 1 Perform calibration. If that does not fix then go to
Percentage. step 2.
2 Replace Pryon board.

ETCO2 failure. 1 Check for broken switch in trap receptacle.


2 Verify board ID in service module.
3 Perform functional test of ETCO2. Verify pump
comes on when switch in water trap receptacle is
activated. If switch is intermittent replace the
receptacle.
4 Replace Pryon board.
5 Replace CPU board.

Temperature:
1 Check connection - main board to temperature
jack.
2 Check pins in Temperature harness for correct
seating.
3 Replace Main board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 61


Section 3 Troubleshooting

62 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

About Section 4

This section is a guide for disassembly and reassembly of the Atlas Monitor. Always refer to
current revision schematics, diagrams and final safety test procedures before attempting to
service this device.

Do not attempt to service this instrument unless you have received Service Training from
Welch Allyn or an authorized Training Agent, and are equipped with approved processes
and test equipment. For more information about training call the Welch Allyn Customer
Service phone number listed in Section 1 of this manual.

General:
The outside housing is removable in order to provide full access to all internal printed cir-
cuit boards and other components. Most of these are held in position with the surrounding
E-Pac foam. E-Pac foam provides shock absorption, ventilation channels, and spaces for
the components and boards. Pneumatic tubing and cables must be placed correctly in E-Pac
to avoid problems with pinched tubing.

Problems During Service:


A Technical Support phone number is listed in Section 1 of this manual to answer questions
you may have during the servicing of the Atlas Monitor.

Printer:
Print-head, printer motor, and printer roller can be replaced as necessary.

Printed Circuit Boards:


There are no component level repairs for the Printed Circuit Boards(PCB). These are
replaced if found to be defective. During the warranty period, boards that are found to be
defective should be returned to Welch Allyn.

EXERCISE EXTREME CAUTION WHEN


SERVICING THE ATLAS MONITOR! THE
Electrical Shock Hazard CRT ASSEMBLY AND CRT DEFLECTION
BOARD OPERATE ON APPROXIMATELY
8,000 VOLTS.
CAUTION

CRT:
The CRT and CRT Deflection Board are aligned at the factory and as such are replaced as a
matched set if one or the other should fail in service.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 63


Section 4 Disassembly and Repair

ESD: Circuit Boards are


sensitive to static electric-
ity. Use wrist straps, ESD
mats and ESD storage
bags.

Model 220 (623NP)

1. Battery Door Removal


Remove the two screws holding the bat-
tery door. Remove the battery door.

2. Battery Removal
Remove the battery connector from the
main PCB. Remove battery.

64 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

3. Rear Housing Removal


3.1 Remove the four Torx head machine
screws. Use a #10 Torx driver.

3.2 Pull the rear housing away from the


front panel. Take special care not to dam-
age any cables.

3.3 Flip the housing forward. Take special


care not to damage any cables.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 65


Section 4 Disassembly and Repair

NOTE: Observe how hose fits in E-Pac


Foam. The hose fits over wiring keeping
wiring secure.

Printer Display
Cable PCB Cable
4. Printer cable and Display Cable
Removal
Remove both Printer and Display PCB
cables from the Main PCB.

5. Power Supply Ground Wire


Removal
Remove the two ground connectors from
power supply chassis.

66 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

6. Tie Wrap Removal


Cut the tie wrap that secures the ground
wires to the E-pac foam.

7. CPU Removal
Remove the CPU from the Main PCB.

8. Main PCB Removal


Use a tool T-16654 to pry the main Board
away from the Deflection Board. Pull the
E-pac tab to the right as you pry. Support
the Main PCB while titling it away.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 67


Section 4 Disassembly and Repair

9. Main PCB Side Connector Removal.


Remove the four connectors(power sup-
ply, fan, speakers, BP pump) from the side
of the main PCB.

10. Main PCB Hose and Wire


Removal.
Disconnect hoses and connectors. BP
hose, ecg cable, temp cable (modles 622
and 623 only, SpO2 cable, CO2 cable
(model 623 only)and ground wire (Nelcor
SpO2 only) from main PCB.

NOTE: Notice how hoses and connectors


fit on Main PCB.

11. Pump Removal.


Remove the pump assembly from the E-
Pac foam.

NOTE: Notice how pump fits in E-Pac


Foam and how wires are routed out of
pump.

68 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

12. CO2 Exhaust Removal.


Remove the CO2 exhaust cover and
remove the CO2 connector from the E-Pac
foam.

13. Pryon PCB Removal.


Remove connectors from Pryon PCB. Pull
Pryon PCB from E-Pac foam.

14. E-Pac Foam Removal (Top).


Remove the top E-Pac foam.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 69


Section 4 Disassembly and Repair

15. Power Supply Removal.


Slide the power supply out from the E-Pac
Foam.

16. Power Supply Fuse Location


NOTE: Notice power supply and power
supply fuse location. Fuse

17. Fan Removal


Slide fan out from E-Pac foam.

70 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

18. E-Pac Removal (Lower)


Remove lower E-Pac foam from unit.

19. Deflection PCB Removal


Carefully remove the small deflection PCB
from the CRT.

20. CRT Removal


Remove the two machine screws holding
CRT to the front panel.

NOTE: Notice how the ground wire is


attached and is located along the side of
CRT.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 71


Section 4 Disassembly and Repair

NOTE: The CRT and deflection PCB are


factory tuned as a unit and are sold only
together as a unit.

21. Display PCB Removal


Remove all 9 self tapping screws that hold
the Display PCB to the Bezel.

NOTE: Notice the correct orientation of


CRT screw and washer.

72 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

22. Printer Door Button Removal


Use a small screwdriver to push the door
button tab up and away from the pin that
locks it onto the printer frame.

NOTE: Notice how the printer button is


being removed.

23. Printer Assembly Locking Tab


Release.
Locking Tab
Pull the locking tab down and towards the
front of the rear housing assembly.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 73


Section 4 Disassembly and Repair

24. Printer Positioning Slot Removal


Make sure printer door is open. Begin to
pull the positioning slots towards the front
of the rear housing and then out from the
rear housing. Positioning
Slots

25. Printer Assembly Removal.


After printer positioning slots have
cleared the front rear housing, pull the
printer down and out of the rear housing.

26. Printer Lip Assembly.

Observe how the printer assembly fits into


the lip of the rear housing.

74 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Disassembly and Repair Section 4

27. Printer Cable Routing.

Observe how the printer cable lies over the


printer PCB.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 75


Theory of Operation Appendix A

Power Supply Circuits


Model 621
Atlas Model 621 uses a 50W, medical grade, offline switcher which provides 12VDC from a
universal AC input (85VAC to 264VAC 50/60Hz). The model 621 can only be powered
from external AC.

The following supplies are generated on the Atlas Main Board:


+12V DC CRT, input to isolated power supply, misc. analog circuits.
6V DC NIBP Pump and Valve, input to 5V and 3V regulators (5.7V nom).
+5V DC Front panel LEDs, misc. logic.
+3.3V DC CPU board.
Vbackup: On/Off circuits.

Note: Additional electrically isolated supplies are generated for patient connected circuits.

Model 621 Lithium


Battery
DC/DC Circuits
Vbackup
Block Diagram

Q301
AC/DC
Converter AC+12V
+12V

Backup On/Off
Regulator Vb
Control On/Off
Circuits

+6V

5.7V Buck +5VDC


Converter Regulator +5V

AC Present

+3.3VDC
Regulator +3.3V

Figure A-1. Model 621 DC/DC circuit block diagram.

On/Off Circuits Backup Regulator and Micro-Controller


A 3.3V regulator, U304, provides power to the On/Off circuits. This voltage is diode ORd
(D304) with the lithium backup battery to provide power to the real time clock on the CPU
board. The On/Off circuitry is controlled by the micro controller U305. This controller per-
forms the following functions:

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 77


Appendix A Theory of Operation

1. Monitors the status of the front panel ON-OFF key. If the unit is off and the ON-OFF key
is pressed, the controller will drive U10-6 high, which will enable power to the remainder
of the instrument.

2. At power up, the micro controller will drive the beeper for about 1 second.

3. At power up, the micro controller will reset Shift Register U2. This will cause the follow-
ing:
a. The front panel LED's are blanked.
b. The NIBP pump drive is placed in the off state.

4. When the unit is ON the micro controller will communicate with the system CPU. When
the front panel ON-OFF key is pressed, the CPU will store away current operating condi-
tions, then issues a command to the micro controller to shut instrument power off. U10-6 is
driven low which will remove power from the remainder of the instrument.

The instrument is powered on as follows:


U305-6 is driven high, which turns transistor Q302 on (output low)
Q301 (P-channel MOSFET) ON when the gate is pulled low, then +12V active.
Q306 is turned off, allowing the 5.7V buck converter to power on.
As 5.7V becomes active, the 5V and 3.3V supplies also become active.

R314
100
d304-1

1
Vb
3
U304 VBackup
C325 R316
8 1 .047uF Lithium 100 2
u304-fb 7 VIN VOUT 5 d304-2
FDBK ERR BackupBattery
3 C322
+ C323 SHTDN + C324 .047uF
D
4.7uF 6 4.7uF R318 D304
14
GND

TAP
1

2 10K U305 MMBD354LT1 D


SENS
Vdd

Vss

D
LP2951CD-3.3 2 13
4

D PIC-Data ClkIn/Osc1/RB5 RB0


D
3 12
On/Off-Key Clkout/Osc2/RB4 RB1
4 11
Vpp/MCLR/RB3 RB2 PIC-Clk
5 10
TC5/TOCK1 RC0
On-Off 6 9 PIC-audio
filt+12V RC4 RC1 Speaker-Audio
7 8 R320
RC3 RC2 1.62K

D620361-MX10000

U2-Reset

On/Off Switch
Q301
MTP23P06V
L301
47uH
3 2
AC+12V +12V
4

R303
+ + C302 C303 + C304 C305 10K
C301 1uF .01uF 330uF .01uF
1

1000uF

R301
Q301-G

51.1K
U301-comp
Switcher
3

Shutdown
3

Q306
Q302-B 1 2N7002
Q302 1
C328 + R302 MMBT3904L
1uF 100K
2

Figure A-2. On/Off Control Circuits.

Buck Converter
A buck converter is used to generate 6V from the 12VDC output of the offline switcher. The
UC3843A (U301), normally a current mode controller, is configured for voltage mode feed-
back. The UC3843A has an under-voltage lockout for Vcc<8.5V. The controller operates
such that with Vcc less than 8.5V, the reference out is 0V, and will be at 5V with Vcc>8.5V.

78 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Then, the reference out (pin 14) can be used as an AC-ON detection signal.
The switching transistor for the buck converter is a P-Channel MOSFET (Q305). The output
drive of the controller is the wrong polarity for driving a P-Channel MOSFET in a step down
mode. Therefore, transistor Q303 is added to invert the PWM out signal. Fast turn-on of
Q305 is provided when Q303 is low, fast turn-off is though Q304 (configured as an emitter
follower). The PWM controller includes an internal 2.5V reference. Voltage regulation is
controlled with sampling resistor R304 and R305 such that Vout = 2.5V*(1+R304/R305) =
5.68V. At power down pin U301-1 (comp pin) is pulled low. This will cause U301-10 to go
low, which turns Q305 off and disables power to +6V.

Q305
MTP23P06V L302
47uH
3 2 Q305-s
filt+12V +6V
4

LC Filtered C326
.047uF D303 + C314
AC/DC converter 330uF C327

1
output Q305-g MBRS340 .047uF
R304
127K
R313
10

U301-fb Q304
MMBT3904L Q304-e
3 2 +6V Buck
R305 C313
1

100K R306 .047uF (5.68V nominal)


10K D301
BAV99L
1

R312 3
C311-1 1.96K

U301
12

11

C311 UC3843A
2200pF 3 Q303-c
2
Vc
Vcc

VFB
U301-out

Q303-b

U301-comp 1 R310
COMP
3

1.96K
14 10 1 Q303
VREF OUT MMBT3904L
PWR-GND

C309 R307 R311


2

.047uF 7.5K 7 5 10K


GND

RT/CT ISENSE
8

9
U301-osc

C310
2200pF
R308
27.4K
U301-5

R309
5Vref
10K
ACON

Figure A-3. Model 621 Buck converter.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 79


Appendix A Theory of Operation

Linear Regulators 5VDC and 3.3VDC


Three terminal linear regulators, U302 and U303, are used to develop regulated 5V and 3V.
The regulators are powered by the output of the 6V PWM converter. These regulators do not
include a power down mode, but the voltage will collapse as the input voltage shuts down.
U302
LM2940CT-5.0
1 3
+6V IN OUT +5V

GND
+ C316 C317
2 10uF .047uF

+ C315 D D
1uF D

U303
LF33CV
1 3
IN OUT +3.3V
GND

+ C319 C320
100uF .047uF
2

D D
D

Figure A-4. Model 621 linear regulation.

Model 622 and Model 623


Atlas Model 622 and 623 can be powered either from AC or battery. A 50W, medical grade,
offline switcher provides 12VDC from a universal AC input (85VAC to 264VAC 50/60Hz).
The battery is a rechargeable 6-Volt, 6.5 Amp-Hour, Sealed Lead Acid Battery, and will pro-
vide about 1.5 hours on a fully charged battery (Battery Life depends on usage, especially
printer usage, NIBP cycles, and CO2 operation).

The unit will operate from AC when the unit is plugged in, and switches to battery opera-
tion when AC is removed. The battery is automatically charged whenever AC is connected.

The following supplies are generated on the Atlas Main Board:

+12V DC CRT, CO2, input to isolated power supply, misc. analog circuits.
+5V DC Front panel LEDs, misc. logic.
+3.3V DC CPU board.
3.3V Backup: Real time clock and On/Off circuits.
Battery Charger: Charges SLA battery.

Note: Additional electrically isolated supplies are generated for patient connected circuits.

80 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

DC/DC Circuits
Block Diagram Q303
+12VDC
Regulator +12V

AC/DC
Converter AC+12V

Backup On/Off
Regulator Control On/Off
Circuits

Battery

Q311

AC Not Present
+5VDC
Regulator +5V

Battery D310
Charger

+3.3VDC
Regulator +3.3V
AC Present

Figure A-5. DC/DC circuit diagram.

On/Off Circuits
Backup Regulator and Micro-Controller
A 3.3V low current regulator, U307, provides power to the On/Off circuits and to the real
time clock on the CPU board. The On/Off circuitry is controlled by the micro controller
U10.

This controller performs the following functions:

1. Monitors the status of the front panel ON-OFF key. If the unit is off and the ON-OFF key
is pressed, the controller will drive U10-6 high, which will enable power to the remainder
of the instrument.

2. At power up, the micro controller will drive the beeper for about 1 second.

3. At power up, the micro controller will reset Shift Register U2. This will cause the follow-
ing:
a. The front panel LED's are blanked.
b. The NIBP pump drive is placed in the off state.

4. When the unit is powered on the micro controller will communicate with the system
CPU. When the front panel ON-OFF key is pressed, the CPU will store away present operat-
ing conditions, then issues a command to the micro controller to shut instrument power off.
U10-6 is driven low which will remove power from the remainder of the instrument.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 81


Appendix A Theory of Operation

18V boost converter and FET ON/OFF switch


The PWM controller, U301, is configured as an 18V boost-converter. An N-Channel MOS-
FET transistor, Q303, is used to switch power to the 12V regulator. As Q303 is configured as
a high side switch, it is necessary to develop a gate voltage of proper magnitude to turn
Q303 on. Then, to enable power to the 12V regulator, the following takes place:

1. The micro controller drives U10-6 high, turning transistor Q302 on.

2. Transistor Q301 is switched on, supplying power to the PWM controller, U301.

3. The boost regulator develops 18VDC, which will switch transistor Q303 On.

On/Off Switch
4
2 3
V
Q303
U307 MTP30N06VL
8 1

1
VIN VOUT VBackup
u307-6 7 5
+ C331 3 FDBK ERR + C332
4.7uF SHTDN 4.7uF
6
2 TAP
4 SENS
D GND D

LP2951CD-3.3
D
On/Off Circuit
18V Boost - Mosfet Gate Voltage
U301-7
+Vfuse
HS301
Heatsink L301
1mH
for D301 F303 R305 U301
1 MC34063AD
6078 4Amp D302
8 U301-sw BAV99L
Q301 7 IDC
1 MPSW51A IPK 1 1 2
AC+12V ISWC +18V
2 1 3 U301-pwr 6
3 4 Vcc 2
Battery ISWE R306

3
R301 61.9K
GND

D301 + C301 10K C302 + U301-ct 3 5 U301-comp


2

MBR2535CTL 330uF Q301-b 1uF TCAP COMP

+ C304
4

C303 R307 100uF


R302 470pf 4.64K
1.96K

Q302-c
R303 R304
51.1K 100K 3
Q302-B 1 Q302
2 MMBT3904L
+ C339
1uF

C42
.047uF
Vbackup

R22
14
1

10K U10
Vdd

Vss

(pg 1)
2 13
PIC-Data ClkIn/Osc1/RB5 RB0
3 12
On/Off-Key Clkout/Osc2/RB4 RB1
(Front Panel Key) 4 11
Vpp/MCLR/RB3 RB2 PIC-Clk
5 10
TC5/TOCK1 RC0
6 9
On/Off RC4 RC1 PIC-audio
(pg 3)
7 8
U2-Reset RC3 RC2
(To U2)
D620361-MX10000

Figure A-6. Boost converter and off/on switch.

82 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

+12V Boost Converter

The +12V boost circuit provides regulated 12VDC. The input to the 12V converter is either
battery or 12V from the AC/DC converter. The circuit is configured as a boost PWM using
current mode feedback. The PWM controller is a UC3843A. The controller includes an
internal 2.5V, 2% reference, and an external 5V, 2% reference. Nominal output voltage for
the boost converter is Vout = [2.5V*(1 + 173.8K/4.64K)] = 12.1V.

The UC3843A requires 8.5V minimum to power on. The maximum voltage allowed on the
switching transistor, Q305, is 15V. To meet both these requirements, the UC3843A is pow-
ered from 12.9V (+18V 5.1V zener diode D303)

When operating from battery, the converter will run at duty cycles over 50%, which
requires slope compensation for a current mode controller. Slope compensation is achieved
by summing in part of the oscillator signal (pin 4) with the current sense line.

The PWM controller is current limited on a cycle to cycle basis by monitoring the voltage on
the Isense line, U302-5. Current limit is activated when the voltage at the sense line reaches
1V. The nominal DC voltage at Isense is about 0.5V when operating from battery only, and
0.7V when operating from AC. Then, current limit is set to 5Amps when Atlas in operating
on battery and 3Amps when operating on AC.

Q303-s
+V
L303
C335 Bead
.047uF
L302-1

+12V Boost
(12.1V nominal)

3 1 U302-pwr
+18V
R308 D303
17.8K 5.1V + C306 + C310 L302
U302-fb 1uF 1000uF 22uH

R309 R310 D304


4.64K 10K MBRS340T3 L305
Bead
C305-1 Q305-d D304-C
+12V
U302
12

11

C305 UC3843A
5600pF 3 + C311 + C312 + C313
Vc
Vcc

VFB
2
4

1000uF 1000uF C336 330uF


U302-out

Q305-g

U302-comp 1 R314 Q305 .047uF


COMP 10
14 10 1 MTP30N06VL
5VRef VREF OUT
PWR-GND

C307
3

.047uF R311 U302-ct 7 5 U302-sense


GND

10K RT/CT ISENSE


Q305-s

R315
196
8

9
1

R312
C308 3.83K
2200pF 3 2 Q304-e

Q304 R313 C309


MMBT3904L 1.96K 5600pF R316
0.1

D312
BAV99L R317
7.5K
1 2 D312-c
AC+12V-fuse
3

Figure A-7. Boost converter.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 83


Appendix A Theory of Operation

Battery Charger
The battery charger is a PWM buck converter. The input to the battery charger is 12VDC
from the offline switcher. The UC3843A, normally a current mode controller, is configured
for voltage mode feedback. The UC3843A has an under-voltage lockout for Vcc<8.5V. The
controller operates such that with Vcc less than 8.5V, the reference out is 0V, and will be at
5V with Vcc>8.5V. Then, the reference out (pin 14) can be used as an AC-ON detect signal.

The switching transistor for the buck converter is a P-Channel MOSFET (Q308). The output
drive of the controller is the wrong polarity for driving a P-Channel MOSFET in a step down
mode. Therefore, transistor Q306 is added to invert the PWM out signal. Fast turn-on of
Q308 is provided when Q306 is low, fast turn-off is though Q307 (configured as an emitter
follower).

The battery charger is a current limited - temperature compensated charger. Current limit is
set to 1.5Amps. Current through the 0.1Ohm sense resistor R327 is measured with Diff-Amp
U305B. When the charger current is at 1.5Amps, feedback is controlled by Op-Amp U304A.
When the current drops below 1.5Amps, the output of U304A goes low, reverse biasing
Diode D308, and feedback will be controlled by Op-amp U304B.

Battery charge voltage is temperature compensated using Thermistor RT301, a 10K negative
temperature coefficient resistor. Voltage over temperature follows the following charge pro-
file:
Table A-1. Temperature coefficient.
Temperature Voltage Thermistor
0C 7.05V 26.9K
10C 7.0V 20.7K
25C 6.85V 10K
40C 6.7V 5.17K
50C 6.65V 3.45K

For optimum battery life, the float voltage (25C, full charge) should be set to 6.85V +/- 50mV
(6.85V +/- 0.7%). To accomplish this tight tolerance, charge voltage at room temperature
will be adjusted with potentiometer R328. Nominal charge voltage at room temperature is:
Vcharge = 5V * (1 + RA/RB) where,
5V is the reference in U303.
RA is the series/parallel combination of R335, R336, and RT301 (nominal 3.465K @ 25C).
RB is the series combination of R337and R338 (nominal 936 Ohms)

It is necessary to minimize current out of the battery when the unit is off. To reduce off cur-
rent, transistor Q309 disconnects the battery from the battery sense resistors when AC in off.
In addition, diode D307 is added to prevent current from flowing from the battery into the
battery current sense circuit, and to the output of the AC/DC converter.

84 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

F301
4Amp
Bat-fuse
Battery

To Linear Regulators
Battery
Charger Q308 L304 D307 Input to 5V and
F302 MTP23P06V 47uH R327 MBRS340T3
4Amp 0.1 3.3V linear
AC+12V
F302-2 AC+12V-fuse 3 2 Q308-d C319-1 D307-A regulators.
4
L306
C340 Bead C337 + C318
.047uF .047uF 330uF C319 + C338
Q308-g D306 1000uF .047uF

1
MBRS340T3

R326
Q307 10 R328 R329
MMBT3904L 1K 1K
3 2 Q307-e

1
R318 D305
10K C314 BAV99L R331
U303-fb .047uF U305-6

1
R325 3
1.96K 33.2K

R319
19.6K U303 R330
12

11

4
UC3843A 33.2K

2
3 6
Vc
Vcc

-
VFB Q306-c 7
U303-out

Q306-b
U303-cp 1 U305-5 5 +
COMP R323 1.96K 3 U305B
14 10 1 Q306 MC34072
VREF OUT 2MMBT3904L

8
PWR-GND

C315 R320 C320 R332


.047uF 7.5K 7 5 U303-5 .01uF 19.6K U305-pwr
GND

RT/CT ISENSE U304-2


Ref-Batcharge
R324 Current Limit
10K (1.5 Amp)

U304-1
8

4
C316 R321 D308 BAV99L
2200pF 27.4K - 2 R333
U303-ct 2 1 1 19.6K
+ 3 U304-3
U304A
R322 MC34072 C322
3

10K .047uF

8
ACON +12V
R341
10K
Ref-Batcharge
AC+12V-fuse
R342 R343
Batt-fb R339 C321 R334 2 19.6K 10K
19.6K .01uF 19.6K Q309 1

Q309-b
R339-2 MMBT3906L3
Ref-Batcharge
R343-1
R335
4.32K
4

2
R340
10K - 6 U304-6
U304-7 7 D309
+ 5 U304-5 Q309-3 BAV70L
U304B R336 RT301
MC34072 7.5K 10K
R337
8

3
RT301-1
Float Voltage 8.87K R344

t
10K 3
AC+12V-fuse Q310-b 1 Q310
ACON
Bat-Adj 2MMBT3904L
1

R338
2
1K R345
10K
3

Figure A-8. Battery charging.

Linear Regulators 5VDC and 3.3VDC


A series pass 5V regulator is built using transistor Q312 and op amp U305A. A 5V reference
from U302 PWM IC) is used as the control voltage for the regulator. The output of the 5V
regulator is turned off when the unit is turned off. When the 18V-boost circuit is shut down,
power is removed to U302, then the U302 reference is driven to 0V. Setting the reference to
0V will cause op amp U305 to turn the series pass transistor Q312 off.

When operating from AC, transistor Q310 is turned on, which will turn transistor Q311 off.
Power to the 5V regulator is then provided through Diode D310 from the output of the bat-
tery charger circuit. Note that this voltage tap is before the current sense resistor, then load
current on +5V does not affect the battery charger current limit circuit. When AC is
removed, the AC-On signal goes low, and Q310 turns off. The gate of Q311 is then pulled up
to 11.4V (12V Vdiode). Transistor Q311 then turns on, and the 5V regulator is powered
from the battery.

3.3VDC is derived using a three terminal regulator. The output of the 5V regulator is used to
power the 3.3V regulator. The 3.3V regulator does not have an independent shutdown, but
powers down as the 5V regulator shuts down.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 85


Appendix A Theory of Operation

+Vpump
D310
MBRS340T3 HS302
Heatsink
Batt Charger input Q312 forQ312
From Batery MTP30N06VL
Charger Circuit. 4 6078
2 3
+5V
4
3 2 + C323 + C325
Battery
100uF Q312-g 100uF

1
Q311
MTP30N06VL R346 U306
100 R348 LF33CV

1
10K 1 3
IN OUT +3.3V
C324 R347
+12V 1000pF 10K

GND
R341 C324-2 C328 + C329 C330
10K .047uF 100uF .047uF

R342 R343

2
19.6K 10K

U305-1

U305-2
To Q309 +5V Linear D D D
D

4
R343-1 Low
2 R349
Dropout 1
-
10K
1

+ 3 U305-3
5VRef
U305A
D309 MC34072
BAV70L C326 R350

8
.01uF
10K
D311
3

R344 BAV70L
10K 3 1 C327
+12V
Q310-b 1 Q310 .047uF
ACON
2MMBT3904L 3

2
AC+12V-fuse
R345
10K

Figure A-9. Linear regulator.

A/D grounded circuits

The A/D converter is designed by building a pulse width modulator (PWM) and a timer cir-
cuit. The PWM runs at a 1.2KHz rate, synchronized by the A/D sync signal (NIBP-ADC-
Clock). A/D sync is low for 52.1uSec, high for 781.25uSec. Component values are selected
such that the integrator will ramp down 4.7V, and ramp up 7.83V. The voltage at the integra-
tor output (U601-1) is limited to about 5V [5V*(73.2/83.2) + Vdiode)]. Then, the integrator
starts at 5V and ramps linearly down to 0.3V.

The analog input voltage to be digitized and the integrator output are the inputs to compar-
ator U210. The output of the comparator is low at the start of an A/D cycle, and switches
high as the integrator ramp drops below the input voltage being digitized (see the timing
diagram below). A/D conversion is accomplished by measuring the width of the PWM out-
put signal. The A/D timer runs at 25.175MHz, then the A/D resolution is about 21000
counts (over 14 bits). Note that since the output of the comparator is low at the start of the
A/D cycle, a resistor divider is formed at the comparator input. This divider reduces the
Analog-In signal by 0.75% [464K/(464K+3.48K)].

R235 R237
NIBP 3.48K 464K
PWM A/D Analog-Input
U209-Out U210-2

Converter P+5V C225


.047uF
P+5V
R230 P+12V
10K C221
U208-3 .047uF P+5V
5
4

1 A
2 14
+
R231 C219 C224 7 U210-7 4
73.2K .047uF A 1000pF 3 - 6
NIBP-PWM-ADC
8

R234 6 U210 5
3 3.48K MAX941 7
+
1 U208-1 U210-3
A A 2 U207B
-
C218 + 74HCT00
4.7uF U208A A A
R232 MC34072
4

U207A 10K
A 74HCT00 U208-2
14 A C220
1 PWM A/D
3 Converter
2 R233 .01uF
NIBP-ADC-Clock
3

7 3.48K
D201-C 2 1

A D201
MMBD1503A
U207-3

Figure A-10. NIBP PWM A/D converter.

86 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Integrator 52uSec
Input - f=1.2KHz
(U207-3) 781uSec

+5V
Comparator
Inputs

Analog In

0.31V

PWM-Out
(U207-6)

PWM-timer

Figure A-11. PWM timing.

A/D Mux grounded circuits


Analog signals are switched to the A/D converter through MUX U209. Control of the multi-
plexer is through a serial communication channel from the main CPU. The following signals
are digitized:

+5V/2 Sampled version of the 5V supply. Used to verify A/D operation.


SafetyPres Safety Pressure Transducer
PrimaryPres Primary Pressure Transducer
P.75V Reference voltage for A/D calibration.
0mmHg pressure for Primary Transducer.
P4.25V Reference voltage for A/D calibration.
300mmHg pressure for Primary Transducer.
Print-Temp Printhead Temperature, used to compensate printer strobe width.
Model 622/623: Vbatt Battery Voltage, used to warn of low battery condition
Model 621: Not Used.
Model 622/623 BattCurrent: Battery charger current, used in service mode verification.
Model 621 LithMeas Backup Battery voltage (Lithium Battery)
Used to warn user of low backup battery condition.

NIBP Circuits
Overview - Safety:
Two pressure transducers are used, a primary and safety transducer. The primary is used to
make all BP measurements. Hardware circuits monitor the output of the primary transducer,
looking for overpressure faults. In addition, Software monitors the digitized outputs of the
primary transducer, and detects overpressure faults. The following overpressure faults are
detected in software (monitored once per second):

>=10mmHg pressure for 295 seconds


>=15mmHg pressure for 175 seconds
>295mmHg pressure for 0.5 seconds.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 87


Appendix A Theory of Operation

Software detected overpressure faults are considered application faults. The user is warned
of a fault with an audible alarm and a Check blood pressure cuff message on the CRT dis-
play. NIBP is not disabled for this type of fault. In the event of a fault, the drive signals to the
NIBP pump and valve are opened.

Two hardware faults are detected, pressure over 330mmHg (nominal trip point 314mmHg,
and pressure over 15mmHg for three minutes (13.3mmHg nominal trip point). These faults
are considered more serious (since software should have detected and corrected this condi-
tion). The user is notified with a BP SYSTEM FAULT message, and NIBP is disabled. A
redundant safety transistor is opened to ensure the NIBP pump is off and the valve is open.

The primary and safety transducer outputs are continuously digitized. The outputs of the
transducers are checked vs. each other, and if they disagree, an NIBP Fault Message is
declared and the NIBP system is disabled. The outputs of the transducers are linearly pro-
portional to the supply voltage (supply current for the safety transducer). The transducers
use unique reference voltages to ensure that a fault in one reference will not cause an equiv-
alent gain error in both transducers.

The A/D also has redundant checks. Two reference voltages (derived from the primary
transducer reference supply) are measured, and the A/D gain and zero is checked. In addi-
tion, a unique reference is digitized, and compared vs. expected results. An error in any of
these A/D measurements will again cause an NIBP Fault Message, and the NIBP system
will be disabled.

Primary Transducer - Amplifier:


The primary pressure transducer is a fully calibrated and compensated transducer with
built in voltage amplification. The output of the transducer is proportional to the supply
voltage. With a 5V supply, the output of the transducer is:
0mmHg = 0.5V.
300mmHg = 4V.
Op amp U204A is used to level shift the output of the transducer such that the nominal volt-
age for 0mmHg is set to 0.75V. The CPU monitors the digitized zero pressure voltage, and
any offset is corrected. This correction comes from the summation of an error correction sig-
nal through op amp U204B. The CPU will output a pulse width modulated signal at
76.8KHz. This signal is RC filtered to provide ad DC voltage at U204-5. This signal can
adjust the offset seen at the A/D converter by +/-20mmHg.

88 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

P+5V

C226 L202
.047uF Primary
Bead
Transducer
A
0mmHG = 0.5V
P201-pwr
300mmHG = 4.0V

C232
.047uF
R236
A 3.83K
3

R212
56.2K
2 P201-Out
Gain P+12V C209
and
C206 .047uF
Comp 6 P201-6 .047uF

P201 C227
XFPM-050KPGR-P1 680pF A

8
A R219
U204-3 3 1.96K
1

+
1 U204-1
PrimaryPres
A A 2
-
U204A
MC34072 C228

4
+3.3V .047uF
R213
Zero Adjust P+12V 237K R217
A
5.62K
R244
+/-20mmHg range U204-2
A

21.5K
R214 R216 C231
8

10K 19.6K .01uF


C207-1 U204-5 5
NIBP-Offset-DAC +
7
PWM D/A 6
-
R215 R218
76.8KHz 100K + C207C208 + U204B 23.7K
1uF 1uF MC34072
4

A A A A
U204-7 P+.75V

Figure A-12. Primary transducer amplifier.

Safety Transducer - Amplifier


The safety transducer is compensated for temperature drift, but gain and zero are not cali-
brated. The safety transducer does not include built in voltage amplification. The safety
transducer output is a differential voltage, proportional to the supply current through the
device.

The initial accuracy of the safety transducer is very loose, in the order of +/-50%. However,
the drift over time and temperature is very good. Then, it is necessary to calibrate the output
of the safety transducer. This is done my measuring a know pressure, measuring the output
of the safety transducer, and storing calibration constants in NVRAM. A two-point calibra-
tion procedure is used. Calibration is done at the factory, and can be recalibrated in the field
if necessary.

Current Source:
Op amp U205A is configured as a current source for the Pressure Transducer, with the cur-
rent through the transducer set to 1.5mA. Nominal gain for the transducer is 300mmHg =
75mV.

Differential Amplifier and Offset Centering:


The output of the safety transducer is a differential voltage, with a nominal gain of 0.25mV/
mmHg, and a zero pressure offset voltage of +/-25mV. The A/D converter has an input volt-
age range of 0.5V to 5V. It is necessary to both add both signal gain and offset centering to
the transducer output before digitalization.

Op amp U206A/B is configured as a differential amplifier, with a voltage gain of 22.5. The
output of the differential amplifier is offset by 1.2V (U205B).

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 89


Appendix A Theory of Operation

P+12V
P+12V C216
Safety .047uF
C211 Transducer
A
.047uF TP226

8
A R227
8

P202-Out+ 3 5.11K
+ U206-1
U205-3 3 1
+ P202-2 SafetyPres
1 2
-
2
-
U206A R223 C214 C217
U205A MC34072 21.5K .047uF

4
MC34072 .047uF
4

P202

2
FPNS-07PGR A A
A C212 U206-2
680pF
1
P+12V
3 R224
+ -
6 1K

8
P202-Out- 5
+
7 U206-7
P202-5 6
5

-
U206B R225
R220 MC34072 1K

4
806

A
5VRef
A
P+12V

Zero R221 R226 C215


3.16K 21.5K .047uF
Offset
8

5
+
7 U205-7
6
-
C213 R222 U205B
.01uF 1K MC34072
4

A A A

Figure A-13. Safety transducer amplifier.

Hardware Overpressure:
The output of the primary transducer is monitored for two overpressure conditions; pres-
sures in excess of 13.3mmHg (nominal) and 314mmHg (nominal) are detected. These error
conditions are transmitted to the gate array on the CPU board, and if the error conditions are
present for a long enough time period, a fault message is displayed, and NIBP is disabled
(see above for safety performance operation).

The output of the Primary Transducer drives the two comparators U203A/B. The compari-
son voltage is derived through a resistor divider chain from a 5V regulator (U201). This reg-
ulator is the supply voltage for the primary transducer, and sense the primary output is
proportional to the supply voltage, tolerance errors in the regulator are not critical.

90 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

R207
464K +3.3V

L201 C205
220uH
P+12V
R208
+12V P+12V P+5V PrimaryPres
.047uF 1.96K
R206 A

8
C229 + 1K
10uF U201 R201 U203-3 3
+
78L05 576 1
NIBP-Overpressure
8 1 U203-2 2
IN OUT -
A U203A
C203 C222 LM393

GND

GND

GND

GND
R202 .01uF

4
P+12V
C201 + C202 174
.047uF 4.7uF .047uF
R210 +3.3V
2

7
A A A

8
A 464K
A U202-3 3
+
1
P+4.25V
2
- P+12V
A R203 R211
3.32K U202A 1.96K
MC34072 R209

8
1K
A U203-5 5
+
7
NIBP-15mmHg
U203-6 6
-
U203B
C223 LM393
.01uF Over Pressure

4
P+12V
R204
150
Fault Circuit
A A

8
U202-5 5
+
7
P+.75V
6
-
C204 R205
.047uF 750 U202B
MC34072

4
A
A A

Figure A-14. Over pressure fault circuit.

Pump and Relay Drive:


Two independent dual transistor switches, Q202 and Q203 control the pump and valve.
Q202 is controlled by logic circuits on the CPU board, and is normally in the on state. Q202
is only opened in a fault condition (over pressure, A/D calibration error, or transducer mis-
match). Once a fault is detected, Q202 will remain open until power is cycled. Under nor-
mal operation, dual transistor Q203 is used to switch the pump on and close the valve. The
pump-on and valve-close commands are controlled by software. Note that the valve is nor-
mally open. Then, in the case of no power, the valve will be in the open state.

+5V +Vpump
7
8

5
6

+12V
Q202
MMDF2N02E

R242 2 4
10K
1

Q202-G
R240
10K 3
Q201-B 1 R243
NIBP-Safety
2

2 100K J201
Q201 D202 Valve+
R241 MMBT3904L BAV99L Valve- 1
10K 3 2
Pump+ 3
Pump- 4
5
D201-A

CON5
1

R246

31.6
7
8

5
6

Q203
MMDF2N02E

2 4
NIBP-Valve-Close NIBP-Pump-On
1

R247 R248
1K 1K

Figure A-15. Pump relay circuit.

Note: Model 621 uses a single Mosfet for transistor Q202, with the drain connected to +6V,
and the source connected to nets Pump+ and Valve+ through a series diode.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 91


Appendix A Theory of Operation

CRT Deflection Board


Overview:

Atlas uses a 5-inch monochrome CRT display. This CRT will display Waveform Data (ECG
and Respiration or SpO2 or ETCO2), plus Text Data (Heart Rate, Alarm Values, Trend Data,
setup, and service menus). The CRT Deflection board performs the following functions:
1. Vertical Deflection.
2. Horizontal Deflection.
3. CRT Grid Voltages.
4. Video Amplifier.
CRT Deflection is magnetic, vertical and horizontal deflection is controlled by regulating
current through the vertical and horizontal coils of the CRT Yoke. The Deflection board is
designed to the following specifications:

Resolution: VGA (640X480)


Dot Clock: 39.7nSec (1/25.175Mhz)
Display Size: 100mm (Horizontal) X 68mm (Vertical)

Vertical: Scan Rate: 60Hz


Reset Time: 750uSec
Blanking Time:1.2mSec

Horizontal: Scan Rate: 31.5Khz


Reset Time: 5uSec
Blanking Time: 5.7uSec

Vertical Amplifier:
Ramp Generator:
Vertical Sync Ramp Generator
R1
1K
U1.3 +12V
+5V

R2
7.50K
C3
750uSec .047uF
16.7mSec
8

U4A R3 3
+
23.7K 1
1 2 U4.2 2 Ramp
Vsync -
U1A
TLC272
R52 Vertical
4

74HCT04
1.96K R43 Sync 5.0V
C1
5.11K .1uF
U1.2
3.95Vpp

d6.c 2 1 2 1

D6 D1 750uSec
1N4148 1N4148
16.7mSec

Figure A-16. Ramp generator circuit.

A ramp generator is built from the integrator (U1A, C1, and R3). The slope of the integrator

92 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

is: V = (I*T)/C. The integrator is designed so that it ramps up 6.97V in 750uS and ramps
down 3.95V in 15.9mS. Diode D1 clips the output voltage at about 5.0V (U1-pin3 + Vdiode).
Then the ramp resets at 5.0V each cycle, and integrates down 3.95V.

Vertical Amplifier
+12V
+12V

C4 + C5 +12V
.047uF R12 22uF R19
562 3 3 562
q1.b 2 Q1 Q3 2 q3.b +5V
8 1 MPSW01A MPSW01A 1 8
R13 R18 R20
3 + 169 R15 38.3 562 169 + 5 R22
Ramp
1 q1.e 7 10K
2 - - 6
U2A R14 R21 U2B
MC34072 562 1 1 562 MC34072
4 q2.b 2 vadj 2 q4.b 4
3 3
C2 Q2 R16 10 R17 10 Q4
100pf MPSW51A MPSW51A
u2.2
Size Adjust u2.7 u2.6

R4 R23
10K 10K
R5 R6
42.2K 41.2K
u1.6

L
Vertical Yoke
4
- 6 R7
u1.7 7 41.2K
+ 5 u1.5
U1B
TLC272
8
R8
41.2K

+12V
Zero Adjust
vcadj
R9 R11
vcadj+ vcadj-
+5V
1.47K R10 1K 1.96K

Figure A-17. Vertical amplifier circuit

Vertical Amplifier:

The Vertical amplifier will generate a linear current ramp of +/-200mA. The vertical ampli-
fier is an H-Bridge type driver. Positive current flow (deflecting the beam above the center-
line) is defined as current from +12V to Q1 to Rsense through the coil to Q4 to Ground. The
negative current path is from +12V to Q3 through the Coil to Rsense to Q2 to Ground.

The input to the vertical amplifier is the ramp voltage generated above. The objective of the
vertical amp is to match the current through the vertical coil with the input ramp control
voltage. Current through the vertical coil is monitored through the sense resistor, formed
from R15, R16, and R17. Voltage across the sense resistor is measured with the differential
amplifier U1B. This voltage is then used as the feedback voltage to the control opamp, U2A.

Zero Adjust:
The output voltage from the ramp generator is a ramp from 5.0V to 1.05V (nominal). The
center of this ramp is 3.0V. Then, the output of the current sense diff amp must be offset by
3.0V. This is accomplished with the Zero Adjust Network, Resistors R9, R10, and R11. Verti-
cal centering is then accomplished by writing a pattern to the CRT, and adjusting R10 to
center the display.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 93


Appendix A Theory of Operation

Size Adjust:

Adjusting the current through the Vertical Coil changes vertical deflection. The voltage
across the sense resistor is:
Vsense = (Vramp-Voffset) / 1.02 (1.02 is the gain of the current sense diff amp)
Current through the coil is equal to current through the sense resistor network.
coil = Isense = Vsense / Rsense
Then, adjusting the value of the sense resistor will change the current through the vertical
coil. Vertical gain is then accomplished by writing a pattern to the CRT, and adjusting R16 to
set vertical deflection.

Horizontal Amplifier:

Current through the transformer increases linearly as Transistor Q6 is On (I = VdT/L). When


the transistor opens, the drain voltage kicks up, and the current through the Transformer
coil flows through C9, L2, L3, and the Horizontal Coil to Ground The transformer quickly
loses flux (reset time < 5uSec). Current continues to flow in load inductance, from Ground
through Q6-diode, C9, L2, L3, and the Horizontal Coil to Ground. This current flow charges
capacitor C9. The current decreases linearly to 0, then changes directions. This is due to the
AC coupling capacitor C9 being charged to a negative voltage. Current flow is then from
Ground through the Horizontal Coil, L2, L3, C9, and Q6 to Ground.

The drive to MOSFET Q6 is AC coupled. This will prevent Q6 to be driven high in the event
of a faulty driver on the CPU board.

In order to get adequate deflection current (about +/-2.3Amps), 18.5V across the transformer
coil is necessary. A boost winding is added to the transformer, then when the voltage on
the transistor drain flys up, current flows into capacitor C11. C11 charges to a voltage deter-
mined by the turns ratio in the transformer.

Horizontal Gain:
Horizontal deflection is adjusted by changing the current through the Horizontal coil.
Changing the series inductance in the Horizontal Deflection Path modifies the current.
Increasing Horizontal gain is then accomplished by writing a pattern to the CRT, and adjust-
ing the width coil, L3.

Horizontal Centering:
Horizontal centering is accomplished by rotating magnets mounted on the CRT
Yoke assembly.

94 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Grid Voltages:

The following voltages are developed to bias the grids on the CRT:
Grid 1: 10V to -50V DC (Brightness Adjust)
Grid 2: 350VDC
Grid 4: 0V to 350VDC (Focus Adjust)
Anode: 7.5KV

These voltages are derived from additional windings on the FBT. In addition, the supply
voltage for the video amplifier (+36VDC) is generated from a tap on the FBT.

To CRT
T1 Anode
CRT-FBT

Horizontal Amplifier
Boost
Grid Voltages
C11 +
1000uF

9
+12V 8 D3
D2 1N4935 Grid1
+Vvid Intensity
L1
+12filt 1 2 +B 10 2 Vd 1 2

47uH C15
+ MUR120 + C12 .01uF
C10 Hor 7 5 22uF R29 100K
1000uF D4 R28
1N4935 Brightness 19.6K

3 Vg1 2 1 d4.a Adjust


1 Vg2
+ C13
R30
Horizontal Sync 22uF 100K
Grid 2
D5
1N4937 C16
.01uF
1 2 d5.c
5.5uSec
C14
.01uF Focus
31.7uSec
R32
+12V Adjust 2.5Meg
Focus
C9 C17
+12V 6.8uF .01uF
R26 hlin hwide
4
2

1K
R24 R27 Q6 C8 L2 L3
10K 10 IRF640 .022uF 15uH 15- 35uH
q5.c q6.d 1
U4B C22 R25
.022uF 215 3
3 4 C22-2 q5.b
2 Q5 L
3

Hsync
12N3904 Horizontal Yoke

R50 74HCT04
1.96K

Figure A-18. Horizontal amplifier grid voltage circuit.

Video Amplifier:

The CRT tube turns a dot on when video out is low (near 0V), and off when video out is high
(+Vvid = 28V). The input to the video amplifier is a digital signal (3.3V logic level) from the
uProcessor. An input of 0V turns the dot off, an input of 3.3V turns the dot on. Transistor Q7
amplifies and inverts this signal. Video out is driven low through D7, and driven high
through emitter follower Q8. .

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 95


Appendix A Theory of Operation

+Vvid

Video Amplifier
C19
R36 R37 .01uF
1.96K 1.96K

3
+5V q8-b 2 Q8
1 2N3904

2 1 R41
R51 681

1
19.6K D7 q8-e
R48 1N4148 VideoOut
U4E
100
11 10 u4-6 q7-g 2 Q7
VideoIn BS170

3
74HCT04

Figure A-19. Video amplifier circuit.

Recorder Electronics
Overview:
Atlas includes a thermal strip chart printer (optional on Model 200 and 210, standard on
model 220). The user can print either annotated waveform data or Patient trend informa-
tion.

The printer specifications are:


Paper Size: 56 or 58mm (2 1/4 inches)
Printhead Width: 54mm (2 1/8 inches)
Resolution: Vertical: 8 dots/mm, (200 dots/inch)
Horizontal: 12 dots/mm (300 dots/inch)
Chart Speed: 25mm/sec (1 inch/second)

The main CPU controls the printer. Data timing, clock signals, and strobe widths are all gen-
erated by the FPGA on the CPU board. These signals are buffered on the recorder board
(inverter U5), before transmission to the printhead.

Power Supply 24 Switcher


It is necessary to generate 24VDC for the print head and motor. The input to the 24V
switcher is either 12V from the AC/DC converter, or Battery voltage on the model 210/220
when AC is not present. The circuit is configured as a boost PWM using current mode
feedback. The PWM controller is a UC3843A. The controller includes an internal 2.5V, 1%
reference, and an external 5V, 1% reference. Nominal output voltage for the boost converter
is Vout = [2.5V*(1 + 84.5K/10K)] = 23.6V.

The converter will run at duty cycles over 50%, which requires slope compensation for a
current mode controller. Slope compensation is added by summing in part of the oscillator
signal with the current sense line.

96 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

The PWM current limits on a cycle to cycle basis. The supply will be in current limit when
the Isense line reaches 1V. Current limit is set to about 3.5 Amps from 12VDC or 5Amps
from battery.

Two control signals exist for the 24V switcher, n24Von and Rec-Supply-On. The 24V
switcher is disabled when n24Von is high. This signal is controlled by the on board PIC pro-
cessor. The switcher is held off at power up, and allowed to start after 50mSec. This is done
to reduce inrush current at power up. The signal Rec-Supply-On enables power to the print
head, and is controlled by the main CPU. Power is only applied to the print head
when the recorder is running.
RecVdc

+24V
To
Recorder R7
Motor
Power 84.5K +12V
TP10 + C4 L1
Supply 1000uF 22uH
U2-fb

+ C6 D2 Q1
R12 1uF A MBRS340T3 MTP30P06V
10K C7
.01uF TP14 Q4-D 3 2
+24V-switch
U2 4
12

11

A
UC3843A To
+5V 3 C8 R16
U2-out

TP8 + Printhead
Vc
Vcc

A VFB

2
4
1800uF R3 19.6K

1
U2-comp 1 R13 Q4-G Q4 10K
COMP 10 MTP30N06VL
TP17 U2-Ref 14 10 1 A
3

R6 VREF OUT
PWR-GND

10K TP19 R15


C11 R14 147 3 Printhead A
1 .047uF 10K U2-RT 7 5 Q4-s
GND

n24VOn
Q3 RT/CT ISENSE Power
2N7002 Switch
U2-Isense
2

Switcher A
TP21 C13
8

Disable 5600pF R17 R18 R9 TP9


A
1

C14 A A 0.1 no-load 10K 3


2200pF A
Rec-Supply-On Q2-b 1 Q2
3 2 Q6-e 2 MMBT3904L
A
Q6 TP23 R19 R10
MMBT3904L 3.16K 10K A

A
A
R22
4.22K

RecVdc
R11
5.11K

Figure A-20. Recorder power supply circuit.

Motor Driver:
Atlas uses a stepper motor to drive the paper. The microcontroller (U3) is programmed to
apply the appropriate phased signal to the motor. A quad darlington switch (U6) amplifies
the signal from the controller to signal levels needed to drive the motor. Motor speed timing
is derived from the main CPU, and transmitted to U3 on signal line U3-clk (U2-pin2).

+24V +24V
1

TP26 R33 D6
+5V C18 1.47K 27V +
.047uF U3-10 C19 C20
.047uF 47uF
3

TP1
d6-c
1

D
14

A A
1

U3 D1
27V
Vdd

Vss

U6
2 13 TP27 R34
3

ULN2065 J3
U3-clk ClkIn/Osc1/RB5 RB0 1.47K u6-3 3 2 J3-1
3 12 U3-9 u6-6 6 1B 1C 7 1
Clkout/Osc2/RB4 RB1 u6-11 11 2B 2C 9 J3-3 2
TP67 3B 3C 3
4 11 n24Von u6-14 14 16 J3-4
Vpp/MCLR/RB3 RB2 4B 4C 4
5 10 1 J3-6 5
RC5/TOCK1 RC0 CLMP1 8 u6-clmp 6
6 9 CLMP2
CON6
RC4 RC1
7 8
RC3 RC2 R35
1.47K Connector Pin #
PIC16C505 U3-8
A 1 3 4 6
TP28 phase 1 + +
2 + +
TP29 R36
3 + +
1.47K 4 + +
U3-7

Figure A-21. Printer motor drive circuit.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 97


Appendix A Theory of Operation

Temperature Amplifier:
The printer will print darker as temperature is increased. Print darkness can adjusted by
controlling the time a dot is turned on. A thermistor is included on the printhead. This ther-
mistor is nominally 30K, and decreases as temperature increases. The output of the temper-
ature amplifier is a function of the thermistor voltage, Temp = .755*(1+R8/Rtherm). This
voltage is digitized (on the main board), and the CPU can compensate dot width in order to
maintain consistent printing over temperature.

Printhead
Temperature +5V
Amplifier
+5V

C1
.047uF
R1
51.1K
TP3 A
8

R2
U1-3 3 1K
+
1 Temp
C2 U1-2 2
-
R4 .01uF U1A
9.09K MC34072
4

TP54

A A TP6
A
U1-1

C3
R5 .01uF
1K

R8 23.7K

C5 .01uF

+5V
2

Thermistor2
D3

Thermistor 3 BAV99L
Input
1

Thermistor1

Figure A-22. Print head temperature circuit.

Patient Isolated Circuits


Isolated Power
Isolated power is provided to the ECG, SpO2, Temperature, and Impedance Respiration cir-
cuits using an Isolation Transformer T401 (temperature and respiration on models 622 and
623 only). The power supply is a PWM controlled flyback converter. The PWM controller is
an LM3524. Output voltage f+V is sampled and compared with a 5V reference voltage. An
error voltage is generated, and this voltage returned via opto U410. Voltage f+V is regulated
to 7.4V. A second tap on transformer T401 provides f-V (-7.4V). The switcher operates at
76.8KHz, controlled by a sync signal from the CPU board. The power supply is synchro-
nized with the A/D converter used to digitize ECG, Temperature and Respiration signals.

98 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Isolation
Barrier

Patient +12V
Isolated Grounded
Circuits Circuits

D401
MBRS130T3 + C412
10uF
f+V T401-1
f+V
C409
470pf
C405 + R419 T401 +12V
C402 100uF 215 C409-1 TP449
.047uF 1

S S
TP450 5 R405
215
R419-2 2 C414 U402-ref TP451
C433 .047uF
470pf 8 T401-8 TP466

15
TP453 U402 C420
3 D405 16 .047uF

VIN
VREF

4
2
BAV99L
TP454

Q401-g

Q402-e
R407 10
SHTDWN FE-Pwr-Sync

3
D E E S
Q401 10 12
4 MTD3055V 1 2 1 U402-e 11 CA 3
13 EA OSCOUT 6 U402-rt TP455
D403 14 CB RT 7 U402-ct
IsoXfmr TP456 2 R409 EB CT TP457

3
T401-4 Q402 1 5.11K 2
f-V V+
MMBT3906L 3 U402-sen 4 1

GND
+SENSE V-

U402-1
R406 5 9
C407 MBRS130T3 0.2 -SENSE COMP C411 R410
C404 100uF
+
R408 TP458 C410 LM3524DM 2200pF 7.5K
.047uF 100 .01uF

8
Q401-s
S S

f+V

R422 +5V
10K
TP459 U410
8

R403 4N25
U401-3 3 1K R411
+
1 U401-1 iso401-a 1 6 iso401-b 5.11K
U401-2 2
-
R423 U401A 5 opto-fb
5.11K LM358
4
4

TP462 2
R404
S S
196K

R401 C408 R402 S


10K .01uf 10K
C408-2
s+5V

TP463 TP464
R417
10K

Figure A-23. Patient isolation circuit.

Linear Regulators and Filtering Model 621


From the raw f+V and f-V voltages (labeled fVcc and fVee on model 621), the following sup-
plies are generated for the various patient isolated circuits:
iso+5dig: Regulated 5V supply. This voltage is used to power the opto-isolators and digital
control logic. This is also the reference voltage used in the regulation of f+V.
f+5V: Regulated +5V supply: Filtered and buffered from iso+5dig, this voltage is used for the
A/D reference.
fVcc: This voltage is used to power the Op-Amps used in the ECG and A/D circuits. This
voltage also powers the Nonin SpO2 electronics.
fVee: This voltage is used to power the Op-Amps used in the ECG and A/D circuits.

8 1
fVcc U403-7 VIN VOUT iso+5Vdig
7 5
3 FDBK ERR
SHTDN C415 + C406 + C403
C401 6 1uF 100uF .047uF
GND

.047uF 2 TAP
SENS

S S S S
U403
4

LP2951C

fVcc

R416
8

10K R415
U401-5 5 100
+ U401-7
7
f+5V
6
-
C418 +
4.7uF U401B C416
MC34072 .047uF
4

S
S
C417 S
680pf

Figure A-24. Model 621 line regulation and filtering.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 99


Appendix A Theory of Operation

Linear Regulators and Filtering Model 622 and 623


From the raw f+V and f-V voltages, the following supplies are generated for the various
patient isolated circuits:
iso+5dig: Regulated 5V supply. This voltage is used to power the opto-isolators, digital con-
trol logic, and 5V digital supply for the Nellcor SpO2 board.
s+5V: Regulated +5V supply: Analog 5V supply for the Nellcor SpO2 board. This is also the
reference voltage used in the regulation of f+V.
f+5V: Regulated +5V supply: Filtered and buffered from s+5V, this voltage is used for the A/
D reference.
s-5V: Regulated 5V supply: Analog 5V supply used for the Nellcor SpO2 board.
fVcc: LC Filtered voltage, derived from f+V. This voltage is used to power the Op-Amps
used in the ECG, Respiration, Temperature, and A/D circuits.
fVee: LC Filtered voltage, derived from f-V. This voltage is used to power the Op-Amps used
in the ECG, Respiration, Temperature, and A/D circuits.

U407
78L05
8 1
f+V IN OUT iso+5Vdig
Isolated Supply Voltages
Regulation and Filtering
GND

GND

GND

GND

C430 C403 + C406


.047uF .047uF 100uF
2

S S S

8 1
VIN VOUT s+5V
7 5
3 FDBK ERR
SHTDN f+V
U403-fb

C401 + C415
.047uF 6 1uF R416 TP460 U401B
TAP
8

2 10K LM358 R415 TP461


4 SENS U401-5 5 100
GND +
7 U401-7
S S f+5V
U403 6
-
S S
LP2951C C418 +
4.7uF C416
.047uF
4

2 1 C417
f-V IN OUT s-5V
3 S S
680pF S
6 IN C421 TP465
GND

7 IN
C424 IN +
1uF
.047uF
U404
5

S
79L05 S

S TP452
L401 220uH
fVcc

+ C434 + C431
10uF 10uF

S S TP468
L402 220uH
fVee

+ C435 + C432
10uF 10uF

S S

Figure A-25. Model 622/623 line regulation and filtering.

100 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

A/D Circuits
M A/D

A pulse width modulator is used as an A/D converter. The PWM runs at a 1.2KHz rate, syn-
chronized by the A/D sync signal. A/D sync is low for 52.1uSec, high for 781.25uSec. Using
the values shown, the integrator will ramp down 10.01V, and ramp up 25V. The voltage at
the integrator output (U601-1) is limited to about 5V [5V*(73.2/83.2) + Vdiode)]. Then, the
integrator starts at 5V and ramps linearly down to 5V.

The analog input voltage to be digitized and the integrator output are the inputs to a com-
parator. The output of the comparator is low at the start of an A/D cycle, and switches high
as the integrator ramp drops below the input voltage being digitized (see the timing diagram
below). A/D conversion is accomplished by measuring the width of the PWM output signal.
The A/D timer runs at 25.175MHz, then the A/D resolution is about 21000 counts (over 14
bits). Hysteresis is added to the comparator to avoid oscillations during switching. Note that
since the output of the comparator is low at the start of the A/D cycle, a resistor divider is
formed at the comparator input. This divider reduces the Analog-In signal by 0.75% [464K/
(464K+3.48K)].

R605 U602-2 R607


3.48K 464K
Analog In
U602-5 iso+5Vdig
fVcc

U602-7
R601 fVcc f+5V
10K R608
U601-3 C603 1K
f+5V
5
6
8

680pF 14
8

2
+
R602 3 7 9 8
+ U602-3 PWM-Out
73.2K 1 U601-1 3
-
2 U602 7
-
R606 LM311 U604D
f+5V U601A 3.48K 74HCT04
4
1

R603 MC34072
4

E E
46.4K
U601-2
14 C601 fVee
PWM fVee E
11 10 U604-10
A/D-sync
R604 1000pF
A/D
3

7 10K Converter
U604E D601-2 2 1
74HCT04
E
D601
MMBD1503A

Integrator 52uSec
Input
833uSec
(U604-10)

+5V
Comparator
Inputs

Analog In

-5V

PWM-Out
(U604-8)

PWM-timer

Figure A-26. A/D converter circuit.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 101


Appendix A Theory of Operation

A/D Multiplexer
The model 621 uses an 8-channel multiplexer, the model 622/623 uses a 16-channel multi-
plexer, to select the analog signal to be digitized. Control of the multiplexer is through a
serial communication channel from the main CPU. The following signals are digitized:

Ground Ground Reference for digitized signals, used in calibrating the A/D converter
ECG Amplified ECG signal
V Buffer Output of the V-lead buffer, used to determine leads off
LL Buffer Output of the LL-lead buffer, used to determine leads off
LA Buffer Output of the LA-lead buffer, used to determine leads off
RA Buffer Output of the RA-lead buffer, used to determine leads off
f1.24V Reference Voltage, used in calibrating the A/D converter
RL Output Output of the RL amplifier
Patient Temp Analog voltage representing patient Temperature (model 622/623 only)
Respiration Amplified Respiration signal (model 622/623 only)
Resp Leads Off DC impedance for respiration, used to determine respiration Leads off
(model 622/623 only).

fVee
8

E
7
9 VEE
VSS

Mux-A/D-C C
10
Mux-A/D-B B
11
Mux-A/D-A A
6
Mux B INH U609
4 4051
Resp-LdsOff X7
2 MuxB
An-Resp X6
5
1 X5
12 X4
15 X3
14 X2
VDD

13 X1 3
Patient-Temp X0 X
E
16

fVcc
fVee
8

E
7
9 VEE
VSS

10 C
11 B
6 A
Mux A INH U603
4 4051
vRL X7
2 MuxA
f+1.24V X6
5
RA X5
1
LA X4
12
LL X3
15
V+ X2
14
VDD

Analog ECG X1
13 3
X0 X Analog In
16

E
fVcc

Figure A-27. A/D multiplexer circuit.

Note: U609 not used in model 621.

102 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Serial Communication
Serial data is transmitted to the isolated circuits through Optical Isolators. The following
signals are transmitted from the CPU board to the isolated circuits:

1. FE-Serial-Data: Serial Data transmitted from the CPU board to the Isolated circuits.
2. FE-Data-Clk: Serial Data Clock.
3. A/D Clk: Serial Data latch. Also used as clock for PMW A/D converter.

Serial Data is converted to a parallel format using Shift Register U605 and U606. The follow-
ing Data is transmitted from the CPU board to the isolated circuits:

1. Mux-A/D-A Control bit A for A/D Mux


2. Mux-A/D-B Control bit B for A/D Mux
3. Mux-A/D-C Control bit C for A/D Mux
4. Mux-Lds-A Control bit A for ECG Lead Select Mux
5. Mux-Lds-B Control bit B for ECG Lead Select Mux
6. Mux-Lds-C Control bit C for ECG Lead Select Mux
7. Sw-RLD-RA Switch Reference Electrode to RA
8. Sw-RLD-LA Switch Reference Electrode to LA
9. Sw-RLD-LL Switch Reference Electrode to LL
10. Sw-RLD-RL Switch Reference Electrode to RL
11. MonBW Select 0.5Hz ECG high pass pole (used to restore ECG baseline).
12. MuxB 0 Select A/D Mux B; 1 - Select A/D Mux A
13. Sw-RLD-V Switch Reference Electrode to V
14. Sw-RespReset Switch Respiration High Pass Pole, used to restore baseline
(model 622/623)
15. SpO2Reset Reset Nellcor SpO2 board (model 622/623 only)
16. Resp-Off Turn respiration off when respiration not used (shuts down drive signal)
(model 622/623 only).

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 103


Appendix A Theory of Operation

iso+5Vdig

U605 16
15 VCC 14
Resp-Off QA SER FE-Serial-Data
1
SpO2-Reset QB
2 11
Sw-RespReset QC SRCLK FE-Data-Clk
3 10
Sw-RL-V QD SRCLR
4
MuxB QE
5 12
MonBW QF RCLK A/D-Clk
6 13
SW-RLD-RL QG G
7
SW-RLD-LL QH
9
QH 8
GND
74HCT595
D E
U605-out

U606 16
15 VCC 14
SW-RLD-LA QA SER
1
SW-RLD-RA QB
2 11
Mux-Lds-C QC SRCLK
3 10
Mux-Lds-B QD SRCLR
4
Mux-Lds-A QE
5 12
Mux-A/D-C QF RCLK
6 13
Mux-A/D-B QG G
7
Mux-A/D-A QH
9
QH 8
GND
74HCT595
D E

Figure A-28. Serial communication circuit.

Signal Isolation
Optocouplers are used to electrically isolate signals. The following signals are transmitted
across the isolation barrier through the optocouplers:
1. SpO2 out: SpO2 serial data, waveform and status information.
Opto U411 Data from the SpO2 board to the CPU board.
2. Serial Data: Serial Control data for isolated circuits.
Opto U613 Data from the CPU board to isolated circuits.
3. Serial Data Clock:Data clock for serial control data.
Opto U612 Clock from the CPU board to isolated circuits.
4. ADC Clock Clock for PWM A/D converter, also used to latch control shift registers.
Opto U611 Clock from the CPU board to isolated circuits.
5. PWM A/D dataA/D Pulse width data.
Opto U610 Pulse width data from isolated circuits to CPU board.
6. Respiration ClockClock for Respiration drive circuit (same signal as power supply sync)
U710 Clock from the CPU board to isolated circuits (model 622/623 only).

ECG Circuits
Overview
Atlas provides a 5-wire front end, and will be compatible with both a 3-wire and 5-wire
cable. Monitor (0.5Hz to 40Hz) Extended (0.05Hz to 100Hz) bandwidth are provided. The
ECG amplifier always transmits 0.05Hz to 100Hz data to the CPU board (unless in baseline
restore mode), additional filtering for Monitor Bandwidth is implemented in software.

104 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Defib Protect, RFI Filtering


The ECG lead set includes a 1K resistor for current limiting. The neon bulbs act as a voltage
clamp, limiting voltage to about 100V. Neons are chosen for their low capacitance and high
DC impedance when off. Series resistors and diode clamps are used as a second set of pro-
tection circuits for the front-end amplifiers. Resistors R506, R507 and Diodes D506, D507,
are used to reverse bias the lead clamp diodes (D501 to D505). Two passive RC filters are
used to reduce susceptibility to RFI and ESU.

Defib Protect
RFI Filtering
R506 D506
10K 5.1V
+clamp 3 1
fVcc

R507
10K
-clamp 1 3
fVee E
D507
D501 5.1V
MMBD1503A

1 2

R501 R508
51.1K 10K
3

D501-3
RA To Buffer Amp

N501
LAMP NEON C501 C506
220pF 220pF

S E E

Figure A-29. Defib protect and RFI filtering circuit.


Note: Typical protection for each lead wire. Model 621 uses a dual silicon series diode in
place of zener diodes D506 and D507.

Front End Buffer - 1st Gain Stage


Gain of the first stage is 9.26 (1+10K/1.21K). Bias current for each input buffer is set to
1.25V/44Meg = 28nA (VldsOff = 1.25V). On a leads-off condition, the two 22Meg resistor
drives the input of the buffer to 1.25V (VldsOff), and the amplifier output saturates high (at
about 5.5V). The voltage out of each buffer is monitored to detect a leads-off condition.

Input Buffers
fVcc Gain = 9.26
8

U501-3 3
In +
1
Out
2
-
R509
22Meg U501A R517
MC34002 10K C510
4

0.1% 220pF
R509-1
fVee
R557 U501-2
22Meg

R518
1.21K
VLdsOff 0.1%

Figure A-30. Input buffer circuit.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 105


Appendix A Theory of Operation

Wilson Network - Lead Select Mux


A 5 wire front end must be able to render the following vectors:
I LA - RA
II LL - RA
III LL - LA
aVR RA - 1/2*(LA + LL)
aVL LA - 1/2*(RA + LL)
aVF LL - 1/2*(LA + RA)
V V - 1/3*(RA + LA + LL)
A resistor divider network (Wilson Network) is used to create the terms above. These vec-
tors are then switched though the two Lead Select Muxs and into the differential amplifier.

Lead Select
fVcc
Wilson 16 U505
Network 4051
RA 13 3 mux-
VDD

RA RA X0 X
14
LA 15 X1
R529 10K aVR 12 X2
aVL 1 X3
aVF 5 X4
R530 10K V- 2 X5
4 X6
LA X7
6
R531 10K 11 INH
10 A
E
VSS

9 B
R532 10K C 7
VEE
LL
8

fVee
E
R533 10K
fVcc
U506
16

4051
R534 10K LA 13 3 mux+
VDD

14 X0 X
LL 15 X1
RA 12 X2
V+ LA X3
1
R535 10K LL 5 X4
V+ 2 X5
4 X6
R536 10K X7
6
11 INH
R537 10K 10 A
VSS

9 B
E C 7
VEE
8

fVee
E

Figure A-31. Lead select circuit.

Differential Amplifier and Slew rate limit


A traditional 3-opamp gain stage (U507 and U508) is used to construct the differential
amplifier. The topology is modified slightly with the addition of slew rate limiting (U509).
The ECG signal is slew rate limited to about 100V/sec. This is done to minimize distortion
of the ECG trace in the presence or large pace pulses. The differential outputs of this ampli-
fier stage are used as input to the Right Leg Drive amplifier.

106 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Diff Amp and Slew Rate Limit


fVcc
DC Gain = 1

C544
1000pF

8
3
mux- + diff-
1
2 R544 R546
-
10K 0.1% 10K 0.1%
U507A U508-3
MC34002
4

C543
fVee .047uF
fVcc
fVcc

u509-4
fVcc

8
R565
8

2
R545 3 121K
+
5 10K 0.1% 1 4
mux+ + -
7 diff+ 2 1
- SingEnd Out
6 U508A 3
- +
AD712 U509
U507B LMC7101

4
MC34002

u501-8
4

5
E
C542 fVee
fVee 220pF
fVee u508-2

R547
10K
0.1% C545
1000pF

E E

Figure A-32. Differential amp/slew rate amp circuit.

Right Leg Drive Amplifier


The right leg amplifier performs two functions. First is to provide bias current to the input
buffer amplifiers. Second, to reduce 60Hz interference (increase CMR). Patient Common
mode signals are amplified and inverted, and this resultant signal used to drive the patient.
Gain of the RLD amplifier at 60Hz is maximized, while keeping the system gain stable. The
RLD output is muxd to the appropriate wire depending on the chosen input vector. For
example, when looking at Lead I (LA-RA), the RLD signal is switched to the LL buffer.

RLD Mux
U504
3 2
To RL S1 D1
1
Sw-RLD-RL IN1
RLD Amp
14 15 fVcc
To LL S2 D2
16
Sw-RLD-LL IN2
11 10 U503A R525
To LA S3 D3
8

9 R516 MC34002 10K


Sw-RLD-LA IN3 10K 3
+ Diff+
6 7 RL-mux 1
To RA S4 D4
8 2 E
R526
Sw-RLD-RA IN4 -
10K
13 TP551 U503-2
fVcc V+ Diff-
4
4

fVee 5 V-
GND

E DG202CSE R527 fVee C514


1K 0.1uF
C514-2

R528 TP552
51.1K

Figure A-33. Right leg drive circuit.

High Pass and Gain Stage


The signal from the output of the differential amplifier is AC coupled before a final stage of
amplification. The High Pass Pole is set to 0.05Hz (a 3.3 second time constant). Large DC
signals are sometimes present, for example, following a Leads Off condition. In order to
quickly restore the ECG baseline, a transistor switch (Q504, Q505) is used to change the AC
pole time constant to 0.33 seconds. Op Amp U508 provides the final amplification prior to
the A/D converter.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 107


Appendix A Theory of Operation

High Pass (0.05Hz or 0.5Hz) Gain Stage


2-pole low pass fVcc
Gain = 74

C515 R550

8
3.3uF 19.6K
ecg-hp U508-5 5
Diff-out +
7
An-ECG
6
-
C516
R548 R549 .047uF U508B
1Meg 110K AD712 R551 C517

4
73.2K .01uF
Q504-d E
fVee
E
U508-6
3

R552
1K

Q504 1
2N7002
E
Q504-s R553
22

100K
1 Q504-g
Sw-MonBw
Q505
2N7002
3

Figure A-34. High pass and gain stage circuit.

Temperature Amplifier (Model 622/623 only)


The temperature option of Atlas is designed to operate with YSI-400 series probes. This
probe has a negative temperature coefficient, R@30C = 1815Ohms, R@40C = 1200Ohms. The
temperature amplifier is configured as a non-inverting gain stage. The positive input to the
amplifier is a 1.25V reference. The output of the temperature amplifier is V = Ref*(1+Rfb/
Rthermistor).

The A/D converter will digitize the An-Temp input, along with the reference voltage and
ground. The gain of the A/D is calibrated from the reference voltage measurement, and any
drifts in the reference voltage or A/D gain are continuously compensated during monitoring.

fVcc Temperature
Amplifier
R624
8

51.1K R626
U608-3 3 1K
f+1.24V +
1 U608-1
Patient-Temp
C633 2
-
.01uF
U608A
AD712
4

E
fVee
C634 .01uF

R625
-Clamp +Clamp 1K
D602 R627 1.21K
BAV99L
1 2
0.1%

C635 .01uF
3

J601
Temp-In
2
1 C636
Temperature
CON2 Thermistor 2200pF

Figure A-35. Temperature amplifier circuit.

108 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Impedance Respiration (Model 622/623 only)


Overview
Impedance respiration is accomplished by measuring a change of resistance across the
patients chest during breathing. The RA and LA electrodes (Lead I) are uses to acquire the
respiration signal. A current source is driven through the patient, and the voltage developed
across the patient is measured. The change in resistance caused by respiration is small, in
the order of 1 Ohm. This small change must be measured on top of a large baseline imped-
ance, typically 100 to 1000 Ohms.

Signal Source Current Drive


An AC current source is used as the input signal, a 34.8KHz square wave. The Power supply
sync signal, at 76.8KHz, is transmitted across opto isolator U710. Flip Flop U701, config-
ured as a divide by 2, converts this signal to a 34.8KHz. Current through the patient is about
150uA Pk-Pk, set by the output of U710 (5V) and the circuit series impedance, 69K includ-
ing cable resistance. The impedance of the patient is small (<2K) compared to the series
impedance and has minimal affect on the magnitude of the drive current. Diode clamps are
used to protect the respiration circuits in the event of a Defibrillation pulse.

f+5V

-Clamp +Clamp Respiration


Drive + C719
Circuit 4.7uF
D701
MMBD1503A
E
1 2
R701 C701 R703 U701A
4

1.96K 1000pF 31.6K 14 74HC74


D701-3 C701-1 RespDr+ 5 VCC 2
3

PR

Right-Arm Q D
3
CLK Resp-Clk
6
CL

Q 7
D702 GND
MMBD1503A
1

1 2 E
R702 C702 R704
Resp-Off
1.96K 1000pF 31.6K
D702-3 C702-1 RespDr-
3

Left-Arm

Figure A-36. Respiration drive circuit.

Differential Amplifier and Peak Detector


The AC current driven though the patient develops a voltage across the RA to LA electrodes.
This signal is measured and amplified with the Differential Amplifier formed by U702A/B.
This amplifier converts the differential voltage across the patient to single ended signal,
with a gain of 16.8. Inverting this signal (U703A), and summing the original signal and
inverted signal through the dual Diode D705 then forms a synchronous peak detector. Then,
the voltage on C711 is the DC resistance seen across the RA to LA buffer. This baseline resis-
tance is digitized, and if the signal is too large (>2.5K), a Respiration Lead Fault message is
displayed. Note that this circuit measurement includes the two 1K resistors in the EKG
cable set. This 2K resistance is subtracted in software before determining a leads off condi-
tion.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 109


Appendix A Theory of Operation

-Clamp +Clamp

D703 R724
MMBD1503A fVcc Respiration 196K

1 2
Diff Amp Resp-DC
R707 C706 C720

8
1.96K 330pF .047uF
D703-3 U702-3 3

3
Right-Arm + U702-1
1 D705
2 MMBD354LT1 E
-
U702A
R709 AD712 R711 R715 R716 1
464K 15.8K 3.16K 3.16K

4
3
D605-C
fVee U702-2 2
fVee
E

U703-2
R717
R712 3.16K C711

D703-1
D704 1K 0.1uF

4
MMBD1503A
2 -
1 2 fVcc 1 E E
3 +
U702B U703A
R708 C707 AD712
3

8
1.96K 330pF

8
AD712
D704-3 U702-5 5
Left-Arm +
7 U702-7
6 fVcc
-

R713

4
1K
R710
464K fVee

U702-6
E
R714
15.8K

Figure A-37. Respiration diff amp circuit.

Gain and Filtering


The DC content (baseline resistance) of the respiration signal is not needed, so the signal is
next AC coupled, and additional gain applied before digitization. Low pass filtering is per-
formed at this stage to reduce high frequency noise outside the respiration signal band-
width. A baseline-reset circuit (transistor switch Q701) is used to quickly restore the high
pass pole if excessive DC voltage is present, for example following a Leads-Off condition.
B.W. = .06Hz to 16 Hz fVcc Gain =1000
B.W. = 10Hz
C712 R720
8

3.3uF 100K
Resp-AC U703-5 5
D605-C +
7
An-Resp
6
-
U703B
R718 R719 C713 AD712
825K 51.1K 0.1uF
4

R721
Q701-d E
fVee 316K
E
U703-6

2 R722
Q701 3 316
MMBF4393L 1 C714
.047uF
Q701-g

E E

D706
MMBD1503A

1 2
Resp-Reset
3

R723
1Meg

Figure A-38. Gain and filter circuit.

SpO2 Circuits
The SpO2 transducer senses oxygen content of functional arteriolar hemoglobin through the
use of light (red and infrared) passed through the sensor. The reflective characteristics of
hemoglobin at the wavelengths used allow the pulse oximetry circuits to obtain changing
saturation levels. This data is then processed to obtain the oxygen saturation percentage and
pulse rate.

110 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Theory of Operation Appendix A

Nellcor or Nonin Medical provides the SpO2 board (Nonin only for the model 621). The
SpO2 board includes amplifiers and processing, and transmits serial data to the CPU board
(Waveform data, SpO2%, and pulse rate). The Atlas monitor provides electrical isolation
(power and data) to the SpO2 board. Note that you must use Nonin probes with the Nonin
SpO2 board, and Nellcor probes with Nellcor SpO2 board.

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 111


Repair Parts Lists Appendix B

Table B-1. Atlas 621N0 repair parts.


Part# Description
620001-503 MAIN PCB ASSY - LOW END
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620089 BATTERY, 3V LITHIUM
620105-3 BATTERY DOOR
620119 LATCH SPO2
620125-502 MAIN CASE/HANDLE SET W/PANEL
620126-4 BEZEL SET - HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620178 CABLE ASSY SP02
620181-1 LABEL,BOTTOM,LOW END
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 113


Appendix B Repair Parts List

Table B-1. Atlas 621N0 repair parts.


Part# Description
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-2C LOW LABEL W/FREEZE (CHINESE)
620384-2E DISPLAY LABEL-LOW/FREEZE/ENG
620384-2F DISPLAY LABEL-LOW/FREEZE/FR.
620384-2G LOW LABEL W/FREEZE (GERMAN)
620384-2I LOW LABEL W/PRINTER (ITALIAN)
620384-2J LOW LABEL W/FREEZE (JAPANESE)
620384-2P LOW LABEL W/FREEZE (PORT)
620384-2S LOW LABEL W/FREEZE (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED

114 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-1. Atlas 621N0 repair parts.


Part# Description
620403 FLAT TIE HOLDER
620432 ATLAS MONITOR CALIBRATION CERT
620494-501 ATLAS FAN KIT
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 SEN.
620532 NELLCOR MP506 SPO2 PCB
761077-1 WIRE TIE

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 115


Appendix B Repair Parts List

Table B-2. Atlas 621NP repair parts.


Part# Description
620001-503 MAIN PCB ASSY - LOW END
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620089 BATTERY, 3V LITHIUM
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-4 BEZEL SET - HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER

116 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-2. Atlas 621NP repair parts.


Part# Description
620172-2 CABLE ASSY AC
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620178 CABLE ASSY SP02
620181-1 LABEL,BOTTOM,LOW END
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-1C LOW LABEL W/PRINTER (CHINESE)
620384-1E DISPLAY LABEL/LOW/PRINTER-ENG
620384-1F LOW LABEL W/PRINTER (FRENCH)

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 117


Appendix B Repair Parts List

Table B-2. Atlas 621NP repair parts.


Part# Description
620384-1G LOW LABEL W/PRINTER (GERMAN)
620384-1I LOW LABEL W/PRINTER (ITALIAN)
620384-1J LOW LABEL W/PRINTER (JAPANESE)
620384-1P LOW LABEL W/PRINTER (PORT)
620384-1S LOW LABEL W/PRINTER (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620403 FLAT TIE HOLDER
620494-501 ATLAS FAN KIT
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 Sen.
620532 NELLCOR MP506 SPO2 PCB
761077-1 WIRE TIE

118 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-3. Atlas 621S0 repair parts.


Part# Description
620001-503 MAIN PCB ASSY - LOW END
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
6200-43E SERVICE MANUAL (ATLAS)
620089 BATTERY, 3V LITHIUM
620105-3 BATTERY DOOR
620119 LATCH SPO2
620125-502 MAIN CASE/HANDLE SET W/PANEL
620126-4 BEZEL SET - HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620181-1 LABEL,BOTTOM,LOW END
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 119


Appendix B Repair Parts List

Table B-3. Atlas 621S0 repair parts.


Part# Description
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620373-501 NONIN SUB ASSEMBLY
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-2C LOW LABEL W/FREEZE (CHINESE)
620384-2E DISPLAY LABEL-LOW/FREEZE/ENG
620384-2F DISPLAY LABEL-LOW/FREEZE/FR.
620384-2G LOW LABEL W/FREEZE (GERMAN)
620384-2I LOW LABEL W/PRINTER (ITALIAN)
620384-2J LOW LABEL W/FREEZE (JAPANESE)
620384-2P LOW LABEL W/FREEZE (PORT)
620384-2S LOW LABEL W/FREEZE (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-1 LABEL, NONIN SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620403 FLAT TIE HOLDER
620494-501 ATLAS FAN KIT
761077-1 WIRE TIE

120 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-4. Atlas 621SP repair parts.


Part# Description
620001-503 MAIN PCB ASSY - LOW END
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
6200-43E SERVICE MANUAL (ATLAS)
620089 BATTERY, 3V LITHIUM
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-4 BEZEL SET - HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620176 CABLE ASSY - PRINTER

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 121


Appendix B Repair Parts List

Table B-4. Atlas 621SP repair parts.


Part# Description
620177 CABLE ASSY - PRINT HEAD
620181-1 LABEL,BOTTOM,LOW END
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620373-501 NONIN SUB ASSEMBLY
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-1C LOW LABEL W/PRINTER (CHINESE)
620384-1E DISPLAY LABEL/LOW/PRINTER-ENG
620384-1F LOW LABEL W/PRINTER (FRENCH)
620384-1G LOW LABEL W/PRINTER (GERMAN)
620384-1I LOW LABEL W/PRINTER (ITALIAN)
620384-1J LOW LABEL W/PRINTER (JAPANESE)
620384-1P LOW LABEL W/PRINTER (PORT)

122 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-4. Atlas 621SP repair parts.


Part# Description
620384-1S LOW LABEL W/PRINTER (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-1 LABEL, NONIN SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620403 FLAT TIE HOLDER
620494-501 ATLAS FAN KIT
761077-1 WIRE TIE

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 123


Appendix B Repair Parts List

Table B-5. Atlas 622N0 repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620016-503 MAIN PCB ASSY-MID/HIGH END
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620105-1 BATTERY DOOR
620105-3 BATTERY DOOR
620119 LATCH SPO2
620125-502 MAIN CASE/HANDLE SET W/PANEL
620126-5 BEZEL SET/MID PRINTED
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620178 CABLE ASSY SP02
620181-2 LABEL,BOTTOM,MID END
620182 LABEL - RS232

124 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-5. Atlas 622N0 repair parts.


Part# Description
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-4C MID LABEL W/FREEZE (CHINESE)
620384-4E DISPLAY LABEL-MID/FREEZE/ENG
620384-4F MID LABEL W/FREEZE (FRENCH)
620384-4G MID LABEL W/FREEZE (GERMAN)
620384-4I LOW LABEL W/FREEZE (ITALIAN)
620384-4J LOW LABEL W/FREEZE (JAPANESE)
620384-4P LOW LABEL W/FREEZE (PORT)
620384-4S LOW LABEL W/FREEZE (SPANISH)
620385 SUB-LABEL DISPLAY

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 125


Appendix B Repair Parts List

Table B-5. Atlas 622N0 repair parts.


Part# Description
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 SEN.
620532 NELLCOR MP506 SPO2 PCB
761077-1 WIRE TIE

126 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-6. Atlas 622NP repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620016-503 MAIN PCB ASSY-MID/HIGH END
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-5 BEZEL SET/MID PRINTED
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 127


Appendix B Repair Parts List

Table B-6. Atlas 622NP repair parts.


Part# Description
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620178 CABLE ASSY SP02
620181-2 LABEL,BOTTOM,MID END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS

128 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-6. Atlas 622NP repair parts.


Part# Description
620384-3C MID LABEL W/PRINTER(CHINESE)
620384-3E DISPLAY LABEL-MID/PRINTER/ENG
620384-3F MID LABEL W/PRINTER (FRENCH)
620384-3G MID LABEL W/PRINTER (GERMAN)
620384-3I MID LABEL W/PRINTER (ITALIAN)
620384-3J MID LABEL W/PRINTER (JAPANESE)
620384-3P MID LABEL W/PRINTER (PORT)
620384-3S LOW LABEL W/PRINTER (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 SEN.
620532 NELLCOR MP506 SPO2 PCB
761077-1 WIRE TIE

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 129


Appendix B Repair Parts List

Table B-7. Atlas 622S0 repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620016-503 MAIN PCB ASSY-MID/HIGH END
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620119 LATCH SPO2
620125-502 MAIN CASE/HANDLE SET W/PANEL
620126-5 BEZEL SET/MID PRINTED
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620181-2 LABEL,BOTTOM,MID END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM

130 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-7. Atlas 622S0 repair parts.


Part# Description
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620214-1 TUBING, SILICONE, 1/16.50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620373-501 NONIN SUB ASSEMBLY
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-4C MID LABEL W/FREEZE (CHINESE)
620384-4E DISPLAY LABEL-MID/FREEZE/ENG
620384-4F MID LABEL W/FREEZE (FRENCH)
620384-4G MID LABEL W/FREEZE (GERMAN)
620384-4I LOW LABEL W/FREEZE (ITALIAN)
620384-4J LOW LABEL W/FREEZE (JAPANESE)
620384-4P LOW LABEL W/FREEZE (PORT)
620384-4S LOW LABEL W/FREEZE (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-1 LABEL, NONIN SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 131


Appendix B Repair Parts List

Table B-7. Atlas 622S0 repair parts.


Part# Description
620403 FLAT TIE HOLDER
761077-1 WIRE TIE

132 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-8. Atlas 622SP repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620016-503 MAIN PCB ASSY-MID/HIGH END
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-5 BEZEL SET/MID PRINTED
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 133


Appendix B Repair Parts List

Table B-8. Atlas 622SP repair parts.


Part# Description
620175 CABLE ASSY - FAN
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620181-2 LABEL,BOTTOM,MID END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620373-501 NONIN SUB ASSEMBLY
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-3C MID LABEL W/PRINTER(CHINESE)
620384-3E DISPLAY LABEL-MID/PRINTER/ENG
620384-3F MID LABEL W/PRINTER (FRENCH)
620384-3G MID LABEL W/PRINTER (GERMAN)

134 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-8. Atlas 622SP repair parts.


Part# Description
620384-3I MID LABEL W/PRINTER (ITALIAN)
620384-3J MID LABEL W/PRINTER (JAPANESE)
620384-3P MID LABEL W/PRINTER (PORT)
620384-3S LOW LABEL W/PRINTER (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-1 LABEL, NONIN SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
761077-1 WIRE TIE

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 135


Appendix E
Appendix B Repair Parts List

Table B-9. Atlas 622NP repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620016-503 MAIN PCB ASSY-MID/HIGH END
620032-1 ETCO2 PCB ASSY
620034 ETCO2 CONNECTOR & CABLE
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-6 BEZEL SET/HIGH HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620148 GASKET CO2 EXHAUST
620149 FITTING,1/16 X 1/8-27 NPT
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG

136 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-9. Atlas 622NP repair parts.


Part# Description
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620178 CABLE ASSY SP02
620181-3 LABEL,BOTTOM,HIGH END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 137


Appendix B Repair Parts List

Table B-9. Atlas 622NP repair parts.


Part# Description
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-5C HIGH LABEL (CHINESE)
620384-5E DISPLAY LABEL-HIGH/PRINTER/ENG
620384-5F HIGH LABEL (FRENCH)
620384-5G HIGH LABEL (GERMAN)
620384-5I HIGH LABEL (ITALIAN)
620384-5J HIGH LABEL (JAPANESE)
620384-5P HIGH LABEL (PORT)
620384-5S HIGH LABEL (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
620411-1C PRYON CO2 CAUTION LABEL (Chinese)
620411-1E PRYON CO2 CAUTION LABEL (English)
620411-1F PRYON CO2 CAUTION LABEL (French)
620411-1G PRYON CO2 CAUTION LABEL (German)
620411-1J PRYON CO2 CAUTION LABEL (Japanese)
620411-1P PRYON CO2 CAUTION LABEL (Port)
620411-1S PRYON CO2 CAUTION LABEL (Spanish)
620524 CABLE-CO2 TO MAIN BD
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 SEN.
620532 NELLCOR MP506 SPO2 PCB
623NP-ACCKT ATLAS ACC KIT/HIGH/NEL-W/PRNT
761077-1 WIRE TIE

138 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-10. Atlas 623SP repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620016-503 MAIN PCB ASSY-MID/HIGH END
620032-1 ETCO2 PCB ASSY
620034 ETCO2 CONNECTOR & CABLE
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-6 BEZEL SET/HIGH HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620148 GASKET CO2 EXHAUST
620149 FITTING,1/16 X 1/8-27 NPT
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG
620167 CABLE-MAIN BD TO DISPLAY BD

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 139


Appendix B Repair Parts List

Table B-10. Atlas 623SP repair parts.


Part# Description
620168-2 CABLE ASSY POWER SUPPLY
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620181-3 LABEL,BOTTOM,HIGH END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16 .50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620373-501 NONIN SUB ASSEMBLY
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS

140 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-10. Atlas 623SP repair parts.


Part# Description
620384-5C HIGH LABEL
620384-5E DISPLAY LABEL-HIGH/PRINTER/ENG
620384-5F HIGH LABEL (FRENCH)
620384-5G HIGH LABEL (GERMAN)
620384-5I HIGH LABEL (ITALIAN)
620384-5J HIGH LABEL (JAPANESE)
620384-5P HIGH LABEL (PORT)
620384-5S HIGH LABEL (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-1 LABEL, NONIN SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
620411-1C PRYON CO2 CAUTION LABEL (Chinese)
620411-1E PRYON CO2 CAUTION LABEL (English)
620411-1F PRYON CO2 CAUTION LABEL (French)
620411-1G PRYON CO2 CAUTION LABEL (German)
620411-1I PRYON CO2 CAUTION LABEL (Italian)
620411-1J PRYON CO2 CAUTION LABEL (Japanese)
620411-1P PRYON CO2 CAUTION LABEL (Port)
620411-1S PRYON CO2 CAUTION LABEL (Spanish)
620524 CABLE-CO2 TO MAIN BD
761077-1 WIRE TIE

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 141


Appendix B Repair Parts List

Table B-11. Atlas 623NP repair parts.


Part# Description
620004-502 ATLAS CPU PCB ASSY
620007-501 FRONT PANEL DISPLAY PCB ASSY.
620013-502 PRINTER PCB
620016-503 MAIN PCB ASSY-MID/HIGH END
620032-1 ETCO2 PCB ASSY
620034 ETCO2 CONNECTOR & CABLE
620038 NELLCOR LABEL (CARTON)
6200-43E SERVICE MANUAL (ATLAS)
620105-3 BATTERY DOOR
620117 PRINTER DOOR BUTTON
620119 LATCH SPO2
620125-501 MAIN CASE/HANDLE SET
620126-6 BEZEL SET/HIGH HEAT TRANSFER
620131-1 MAIN FOAM
620132-2 TOP FOAM
620133 BP FOAM
620134-501 GEAR ASSY
620138-501 PRINTER DOOR ASSY
620139-501 PRINTER FRAME ASSE
620140-501 MAIN PRINTER ASSY
620140-502 PRINTER ASSY W/STATIC BAG
620141 METAL HOUSING - TOP
620142-1 METAL HOUSING-BOTTOM (LG)
620143 NUT TEMPERATURE
620144 BRACKET SPO2
620148 GASKET CO2 EXHAUST
620149 FITTING,1/16 X 1/8-27 NPT
620150-1 AC/DC POWER SUPPLY
620156 PUMP
620157 STEPPER VALVE
620158 CHECK VALVE
620163 CABLE ASSY -PUMP/PNEUMATIC VLV
620165 CABLE-CONN TO MAIN BD,ECG

142 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Repair Parts Lists Appendix B

Table B-11. Atlas 623NP repair parts.


Part# Description
620167 CABLE-MAIN BD TO DISPLAY BD
620168-2 CABLE ASSY POWER SUPPLY
620170 CABLE ASSY NELLCOR
620171 CABLE ASSY SPEAKER
620172-2 CABLE ASSY AC
620173 CABLE-TEMP SENSOR TO MAIN BD
620175 CABLE ASSY - FAN
620176 CABLE ASSY - PRINTER
620177 CABLE ASSY - PRINT HEAD
620178 CABLE ASSY SP02
620181-3 LABEL,BOTTOM,HIGH END
620182 LABEL - RS232
620187-501 PNEUMATIC SUB ASSY
620189 4-20 X 5/16 SLT/TORX PN PL ZIN
620192-1 PNEUMATIC CONNECTOR
620193 NUT, HEX 5/16-32 X.095 PNEUM
620194 WASHER,FLT.33 X.62 X.049 PNEUM
620197 BUSHING, SPLIT NYLON
620198 FOOT
620200-501 CRT SUB ASY
620205 4-40 X .31 TX PN MC ST ZN
620207 NUT, M3-.5 HEX KEPS ST ZN
620214-1 TUBING, SILICONE, 1/16.50
620215-1 TUBING, SILICONE 1/8 X1.00
620215-2 TUBING, SILICONE 1/8 X 2.125
620215-3 TUBING, SILICONE 1/8 X 7.50
620215-4 TUBING, SILICONE 1/8 X.750
620216 FITTING Y 1/8 X 1/8 X 1/8
620217 FITTING, T, 1/8 X 1/16X 1/8
620377-1 NELLCOR WORKS HERE LABEL
620377-2 NELLCOR PATENT LABEL
620378-1 ATLAS KEYPAD(BP START/CANCEL)
620378-2 ATLAS KEYPAD(TREND/LEAD/PRINT)
620378-3 ATLAS KEYPAD(SP02/CLOCK/POWER)

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 143


Appendix B Repair Parts List

Table B-11. Atlas 623NP repair parts.


Part# Description
620378-4 ATLAS KEYPAD(4 ALARMS)
620378-5 ATLAS KEYPAD(SILENCE ALARM)
620378-6 ATLAS KEYPAD(SILENCE BELL)
620379-501 MONITOR BOX W/INSERTS
620384-5C HIGH LABEL (CHINESE)
620384-5E DISPLAY LABEL-HIGH/PRINTER/ENG
620384-5F HIGH LABEL (FRENCH)
620384-5G HIGH LABEL (GERMAN)
620384-5I HIGH LABEL (ITALIAN)
620384-5J HIGH LABEL (JAPANESE)
620384-5P HIGH LABEL (PORT)
620384-5S HIGH LABEL (SPANISH)
620385 SUB-LABEL DISPLAY
620386 FASTON TAB
620388-2 LABEL, NELLCOR SENSOR
620393 PWR SUPPLY INSULATOR LABEL
620394-1 WASHER, SHOULDER PLATED
620395-501 REPLACEMENT BATTERY SUB ASSY
620403 FLAT TIE HOLDER
620411-1C PRYON CO2 CAUTION LABEL (Chinese)
620411-1E PRYON CO2 CAUTION LABEL (English)
620411-1F PRYON CO2 CAUTION LABEL (French)
620411-1G PRYON CO2 CAUTION LABEL (German)
620411-1J PRYON CO2 CAUTION LABEL (Japanese)
620411-1P PRYON CO2 CAUTION LABEL (Port)
620411-1S PRYON CO2 CAUTION LABEL (Spanish)
620524 CABLE-CO2 TO MAIN BD
620530 D CONN.LOCATOR\NELLCOR MP506
620531 CABLE ASSY,NELLCOR MP506 SEN.
620532 NELLCOR MP506 SPO2 PCB
623NP-ACCKT ATLAS ACC KIT/HIGH/NEL-W/PRNT
761077-1 WIRE TIE

144 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Figure C-1. Atlas Interconnect Diagram.
Welch Allyn Atlas Monitor, Service Manual 6200-43E Rev D 145
A B C D E

REV Description ECN/ECO Init Date Ckd


AC/DC Power +5V

A Release to Production 5-39118 JAC 3/15/99 LPP


Connector AC+12V
EKG-PWM-ADC 6 5
FE-PWM-ADC
J1 U1C B See ECN worksheet 5-39456 LPP 4/27/99 LPP
+5V
R7 AC 74HCT04
1
150 TP7
2 Rec+12V minimum D Improvements for ESD
C 5-39822 LPP 6/23/99 LPP
3
load +12V EKG-Data-Clock 9 8 and Fast Transients
4 Rec-GND FE-Data-Clock
5 +5V Mods to U2 clock line,
U1D D 5-39934 LPP 7/9/99 LPP
6 Fan 74HCT04 TP8
NVRAM pullup
7
+5V Add Nellcor SpO2 cabability.
C22 Connector EKG-ADC-Clock
D
11 10
FE-ADC-Clock
F Modify RLD to reduce V-lead noise. 5-44460 LPP 3/26/02 JK
CON7 .047uF J5 Improve ESU rejection.
+5V
C39 U1E Add 330uF bypass for Nellcor 506
1
.047uF 74HCT04 TP9 G board. 5-44857 LPP 7/18/02 JK
A D 2
4 D D 4
CON2 FE-Data 13 12
FE-Data-In
Backup U2-Reset Serial TP1 TP2 TP3 TP4 U1F
BackupBattery +5V
74HCT04
Battery Data Out D
TP10
TP5 Q4 TP6 14 15 LED-On
SER QA
J2 R22 1 Power-Sync 3 4
2N7002 QB FE-Pwr-Sync
10K 11 2
1
q4-d 3 2 10
SRCLK QC
3
NIBP - Page 2 U1B
2 LithMeas SRCLR QD
4 74HCT04
QE NIBP-A/D-SelC
CON2 Page2 12 5 Audio-Shutdown
RCLK QF NIBP-A/D-SelB D
BT1 13 6 EROM-wr
G QG NIBP-A/D-SelA
BR-2/3A R23 7
QH NIBP-Pump-On

1
J3 3V Lithium 10K
NIBP-ADC-Clock
9
1 +12V QH NIBP-Safety +5V
2 NIBP-15mmHg
U2 C23
D NIBP-Overpressure D
CON2 74HCT595 .047uF
D NIBP-Valve-Close
SPO2-Rx 2 1
D NIBP-PWM-ADC CRT Board SpO2Out
NIBP-Offset-DAC
U1A
Connector 74HCT04
+12V
D

J6
Serial I/O
1
Video-VSync
Video-HSync
2 connector
3
+5V C30 +5V
4
Video-Data 10uF
5
+3.3V

+
6 J10
1
Main PCA CON6 D 2
3
RS232-RX 12 13 Model 200 Serial I/O
C6 + Connector 9
R1OUT
R2OUT
R1IN
R2IN
8 Rx
4
5
100uF 11 14 Tx (manufacturing test
3 T1IN T1OUT 6 only) 3
J4 10 7
T2IN T2OUT 7
1 2
1 2 TP11 8
NIBP-Safety 3 4 U8-1 1
D 3 4 C+
+3.3V 5 6 NIBP-PWM-ADC TP17 U8-3 3 RJ-45
5 6 C1-
7 8 U8-4 4
7 8 C2+ D
9 10 RS232-Tx TP13 U8-5 5 U8
9 10 C2-
C2 C3 R20 NIBP-OP 11 12 PIC-Data +3.3V C29 U8-2 2 SP232ACN
11 12 PIC-Data V+
.047uF 220pF 215 13 14 RS232-Rx U7 .047uF TP15 U8-6 6
13 14 (pg 3) V-
ETCO2-Tx 15 16 SPO2-Tx AT24C02N-10SC-2.7 TP16
ETCO2-Rx 17
15 16
18 SPO2-Rx R15 TP12 TP14
Front Panel
D D 17 18
Spare-2 19 20 ADC-Clock SCD 5 1 1.96K
19 20 ADC-Clock SDA A0
NIBP-Data
NIBP-15mmHg
21
23
21 22
22
24
FE-Clock
SPI-Data-Out (pg 3)
SCL 6
SCL A1
2
3
D
C31 C32 C33 C34
D
C5 .047uF Display
0.1uF 0.1uF 0.1uF
LED-Latch 25
23 24
26 SPI-Clock
A2
TP18
0.1uF +5V Connector +5V
25 26
Video-HSync 27 28 Batt-ID 7 U7-wc +5V
27 28 WC D D
Spare-3 29 30 FE-Data 1 J8
29 30
Power-Sync 31 32 Rec-Motor-Step 3 CON14AP
31 32
Vbackup 33 34 Key-Clock 2 Key-Clock-b 2 1
Vbackup 33 34 + +
NIBP-Valve-Close 35 36 PIC-Clk R16 Key-Latch-n-b 4 3
35 36 PIC-Clk D D + + On/Off-Key
Video-VSync 37 38 Speaker-Audio R5 1.21K 3.48K U3A Key-Data 6 5
37 38 Speaker-Audio + +
NIBP-Offset-DAC 39 40 74HCT08 LED-Data-b 8 7
Page3
39 40 (pg 3) +3.3V D + +
41 42 NIBP2-PW-ADC LED-On 10 9 + C4
ac-on 41 42 + +
43 44 R6 1.21K 12 11 330uF
43 44 +5V + +
45 46 Video-Data LED-Latch-b 14 13
45 46 D + +
Key-Latch-n 47 48
47 48
49 50 Nurse-Call 4
49 50
Rec-Clock 51 52 Rec-Data-Latch 6
51 52
Rec-Data 53 54 Rec-Strobe Key-Latch-n 5
53 54 D D

ACON-LED
HRESET 55 56
55 56
SPI-data-in 57 58 SCD U3B filt+12V
57 58 +5V
59 60 74HCT08 Q1
59 60 D
MMBT3906L
Socket 30x2 9 2 3
8
LED-Data 10 R1
D D
10K

1
U3C
2 +5V 2
74HCT08 Q1-e TP19
D

12 +3.3V
11 R2
LED-Latch 13 1.96K
Q2-c TP20
U3D R3 TP21 R21

3
74HCT08 10K Q2 10K
D
Q2-b 1 MMBT3904L
ACON

ac-on

2
R4
10K Page1
C35
Speaker Drive

3
+5V 4.7uF
TP22

+
Q3
1 2N7002
PIC-Vprog
Audio-Shutdown

Rec-clock

2
TP24 TP25
TP23 4Vpp max.
Rec-Data

1
6
Rec+12V C36 J11
4.7uF R17 3 5 SP
+ 1
19.6K
2
J7 R17-1 U9-4 4 8 SM
- 3

+
1 13
Rec-Supply-On + C7 C8 U9 CON3
2 14

2
7
330uF .047uF TPA301
3 15
Rec-Motor-Step
4 16 U9-3
R18
+5V 5 17
619 C37
+12V 6 18 Print-Temp
Rec-Data-Latch 0.1uF
7 19
Rec-Strobe
8 20
Rec-Clock R19
9 21
Rec-Data 19.6K
10 22
1 11 23 1
12 24

C14 C20
.047uF .047uF CON24B C38
D
220pF

Rec-GND Initial Date


Welch Allyn Inc.
D D
Recorder Drawn Schematic: 620002.dsn
Rick Myers
Board 3/05/99

Designed
Rick Myers 3/05/99
Title
Connector Checked
Main Board - Model 200
L.Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620002 G
Release For Production
J. Bello 3/17/99
Date: Monday, August 05, 2002 Sheet 1 of 6
A B C D E

Figure D-1. Atlas model 200 main board.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 147


A B C D E

REV Description ECN/ECO Init Date Ckd


L201 TP201
220uH
TP202 R207 A Release to Production 5-39118 JAC 3/15/99 LPP
+12V P+12V +5V
464K

C229 + B See ECN worksheet 5-39456 LPP 4/27/99 LPP


10uF C205

P+12V Improvements for ESD


R208 C 5-39822 LPP 6/23/99 LPP
P+5V
.047uF 1.96K and Fast Transients
A
R206
A Mods to U2 clock line,

8
TP206 1K D 5-39934 LPP 7/9/99 LPP
NVRAM pullup
U201 R201 U203-3 3
+ 1 Add Nellcor SpO2 cabability.
78L05 576
NIBP-Overpressure
8 1 U203-2 2 F Modify RLD to reduce V-lead noise. 5-44460 LPP 3/26/02 JK
IN OUT -
U203A Improve ESU rejection.
C203 C222 LM393 Add 330uF bypass for Nellcor 506

GND

GND

GND

GND

4
C201 R202 .01uF G board. 5-44857 LPP 7/18/02 JK
P+12V
.047uF + C202 174
4 4.7uF .047uF 4
TP207

7
TP208 TP209 R210 +5V
A A A

8
464K
A
U202-3 3 P+12V
A + 1
2 P+5V
- P+12V P+5V
R203 R211 TP203 TP204
A

8
3.32K U202A 1.96K

4
MC34072 R209 5
+

8
1K 7 u207-9 9
U203-5 5 6 8
A + -
P+5V TP210 7 10
NIBP-15mmHg
U203-6 6 U208B

4
-
TP240 U203B MC34072
C223 LM393 U207C
Over Pressure A

4
.01uF 74HCT00
P+12V
R204
TP211 150
P+.75V
Fault Circuit A
U208-7 P+5V
TP212
A A

8
C226
.047uF L202 U202-5 5 12
+ 7 11
Bead
6 13
-
C204 R205
A
.047uF 750 U202B +12V +6V

4
P201-pwr MC34072 R239 U207D
A
1K 74HCT00

A
C232 R242
Primary A A

2
4
.047uF R236 10K TP213
3.83K TP214 Q202
Transducer A
3

R212 MMFT2N02EL
A
56.2K TP215 Q202-G 1
0mmHG = 0.5V 2 P201-Out R240

3
10K
Gain

3
300mmHG = 4.0V Q201-B 1 Q201 R243
NIBP-Safety
and C206 TP216 MMBT3904L 100K Q202-D
3 6 P201-6 .022uF P+12V C209 3
Comp

2
.047uF R241
C227 10K
680pF
P201
A
1

XFPM-050KPGR-P1 TP217
A

8
R219
U204-3 3 1.96K
A A + 1 U204-1 TP218
2
-

3
J201
P+5V U204A 1 2 Valve+
1

4
Zero Adjust R213 MC34072 C228 Valve-
2
P+12V 237K .047uF
3

D201-A
+/-20mmHg range D202 Pump+
4
R244 R217 BAV99L Pump-
A 5
38.3K TP219 TP220 5.62K
A
R214 R216 U204-2 CON5
8

10K 19.6K TP221 TP222 R246


C207-1 U204-5 5 C231 31.6
NIBP-Offset-DAC + 7 .01uF
NIBP-A/D-SelC
PWM D/A 6
-
R215
NIBP-A/D-SelB
76.8KHz 100K + C207 C208 + U204B
4

1uF 1uF MC34072 R218


NIBP-A/D-SelA

7
8

5
6
23.7K
Q203
A/D Mux MMDF2N02E
A A A
U204-7
A
2 4
NIBP-Valve-Close NIBP-Pump-On
P+.75V

3
A
9 R248 R249
C
10 1K 1K
B
11
A
6 U209
2 INH 2
4051
4
LithMeas X7
P+12V 2
X6
P+12V C216 5
Safety .047uF TP224
Print-Temp
P4.25V 1
X5
X4
C211 P.75V 12 TP225
Transducer PrimaryPres 15
X3
X2
R235
SafetyPres 14 3.48K
.047uF
A
TP226 +5V/2 13
X1
3 U209-Out U210-2
NIBP PWM
X0 X
8

R227 R228
A
A/D Converter
8

P202-Out+ 3 5.11K TP227


+ U206-1 +5V
U205-3 3 1 C230
+ 1 P202-2 2 1.96K .047uF
2 -
R237 TP228
- P+5V
U206A R223 C214 C217 C210 R229 464K
4

U205A MC34072 21.5K .047uF .047uF 1.96K


4

.047uF
MC34072
A
P202 TP229
2

FPNS-07PGR P+5V C225


A A A A
C212 U206-2 .047uF
A
680pF P+5V TP231
1 R230 TP230 P+12V
P+12V
3 R224 10K C221
+ - 6 U208-3 P+5V
1K .047uF

5
4
TP232 A
8

2
P202-Out- 5 + 7 U210-7 4
R231 C219 C224
+ 7 U206-7 3 6
73.2K .047uF 1000pF
A - NIBP-PWM-ADC
5

8
P202-5 6 R234 U210 5
- 3 3.48K MAX941

6
+ 1 U208-1 U210-3
U206B R225
4

R220 MC34072 1K 2 U207B


A A -
806 C218 + 74HCT00
TP233 4.7uF U208A
A A

4
R232 MC34072 TP235
A
5VRef U206-6 U207A 10K
74HCT00 U208-2
1 A A 1
C220
A
P+12V 1
3 TP236
Zero R221 R226 C215 2 R233 .01uF
NIBP-ADC-Clock
3

3.16K TP237 21.5K .047uF 3.48K


Offset D201-C 2 1
8

5
TP238
D201 Initial Date
Welch Allyn Inc.
+ A
7 U205-7 MMBD1503A
6 U207-3 Drawn Schematic: 620002.dsn
- Rick Myers 3/05/99
TP234
C213 R222 U205B Designed
3/05/99
4

Rick Myers Title


.01uF 1K MC34072
Checked
NIBP Electronics
L.Phillips 3/16/99

Approved Size Document Number Rev


A A A Z. Psenicnik 3/16/99
C 620002 G
Release For Production
J. Bello 3/17/99
Date: Monday, August 05, 2002 Sheet 2 of 6
A B C D E

Figure D-2. Atlas model 200 NIBP electronics.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 148


A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39118 JAC 3/15/99 LPP


PIC-Vprog

Rec-Data
B See ECN worksheet 5-39456 LPP 4/27/99 LPP

Rec-Clock
Improvements for ESD
C 5-39822 LPP 6/23/99 LPP
and Fast Transients
filt+12V
Mods to U2 clock line,
D 5-39934 LPP 7/9/99 LPP
NVRAM pullup

Add Nellcor SpO2 cabability.


TP315 F Modify RLD to reduce V-lead noise. 5-44460 LPP 3/26/02 JK
U304
Improve ESU rejection.
8 1 Add 330uF bypass for Nellcor 506
VIN VOUT Vb
u304-fb 7 5 G board. 5-44857 LPP 7/18/02 JK
FDBK ERR
3
4 SHTDN 4
+ C323 + C324 TP26 C325
4.7uF 6 4.7uF .047uF
TAP
2 TP27
SENS

R318
D
LP2951CD-3.3 10K

14
D D

1
U305
D

Vdd

Vss
2 13
PIC-Data ClkIn/Osc1/RB5 RB0

3 12 Page 1
On/Off-Key Clkout/Osc2/RB4 RB1

4 11
Vpp/MCLR/RB3 RB2 PIC-Clk

TP313 5 10
TC5/TOCK1 RC0

On-Off 6 9 PIC-audio
RC4 RC1
R320 1.62K
7 8
RC3 RC2 Speaker-Audio

D620361-MX10000 TP320
R319 19.6K
U2-Clk
ADC-Clock

On/Off Switch
TP316 TP317

Q301
U2-Reset
MTP23P06V
L301
47uH
3 2
AC+12V
4
TP318
R303 R314 D304
3 C302 C303 C304 C305 10K 100 MMBD354LT1 3
+ + +

1
C301 1uF .01uF 330uF .01uF d304-1 1
1000uF
3

TP319 2
VBackup
R316

Q301-G
100
d304-2 C322
BackupBattery
.047uF

+12V
R301
51.1K

+ C306 C307
330uF .047uF
3

C328-1 Q302-B 1
TP301 Q302
R302 MMBT3904L
2

C328 + 100K
TP321
1uF
Q305
MTP23P06V L302
47uH
3 2 Q305-s
+6V
4
+6V Buck
(5.68V nominal) C326 C321
.047uF D303 + C314 .047uF

1
2 330uF C327 2
Q305-g MBRS340
.047uF
R304
127K
R313
10

TP302 U302
U301-fb Q304 LM2940CT-5.0
MMBT3904L Q304-e TP303 1 3
IN OUT +5V
3 2
R305 C313

GND
1
100K R306 .047uF + C316 C317
10K D301 10uF .047uF

1
BAV99L

2
R312 3
TP304
C311-1 1.96K
+ C315
D D
U301 TP305 1uF
12

11

D
TP306 C311 UC3843A TP307
2200pF 3 Q303-c 2
VFB
Vc
Vcc

D
U301-comp 1
U301-out

R310 TP308
COMP
Q303-b

1.96K
3

14 10 1 Q303 U303
VREF OUT
Q306 MMBT3904L LF33CV
2N7002 1 3
IN OUT +3.3V
2

1
PWR-GND

C309 R307 R311


.047uF 7.5K 7 5 10K
RT/CT ISENSE
GND

GND
+ C319 C320
2

100uF .047uF
8

2
U301-osc

C310
D D
2200pF
D
R308
27.4K
1 TP310 1
U301-5

TP309
R309
10K

Initial Date
Welch Allyn Inc.
TP312 Drawn Schematic: 620002.dsn
Rick Myers 3/05/99
5Vref
Designed
Rick Myers 3/05/99
Title
ACON
Checked
Power Supply - DC/DC
L.Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620002 G
Release For Production
J. Bello 3/17/99
Date: Monday, August 05, 2002 Sheet 3 of 6
A B C D E

Figure D-3. Atlas model 200 power supply


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 149
A B C D E

REV Description ECN/ECO Init Date Ckd

U407
78L05 A Release to Production 5-39118 JAC 3/15/99 LPP
1 8
iso+5Vdig OUT IN

B See ECN worksheet 5-39456 LPP 4/27/99 LPP

GND

GND

GND

GND
C406 +
100uF C403 C430
Improvements for ESD
.047uF .047uF C 5-39822 LPP 6/23/99 LPP

2
and Fast Transients
S S S Mods to U2 clock line,
D 5-39934 LPP 7/9/99 LPP
NVRAM pullup

Isolation Add Nellcor SpO2 cabability.


F Modify RLD to reduce V-lead noise. 5-44460 LPP 3/26/02 JK
S
Barrier Improve ESU rejection.
Add 330uF bypass for Nellcor 506
G board. 5-44857 LPP 7/18/02 JK
filt+12V
4 1 8
Isolated SpO2 Grounded SpO2 4
s+5V VOUT VIN
5 7 U403-7 Circuits Circuits
ERR FDBK
3
SHTDN
C415 + TP401
1uF 6 C401
TAP
U403 2 .047uF + C412
SENS D401
LP2951C 10uF
f+V T401-1
S S
C409
470pf
C405 + MBRS130T3 T401 +12V
C402 100uF R419 C409-1
S
.047uF 215 1 TP402
1 2
s-5V OUT IN
3 TP403 5 R405 TP404
IN S S
U404 6 C421-1 215
IN

GND
C421 79L05 7 2 C414 U402-ref
+ IN
1uF C424 C433 TP406 .047uF
.047uF 470pf 8 T401-8

15
5
TP408 U402 C420
3 D405 16 .047uF
S S VREF

4
2

VIN
BAV99L TP409
R407 10
S SHTDWN

Q401-g

Q402-e

3
Q401 10 12
CA
TP452 L401 4 MTD3055V 1 2 1 U402-e 11 3
EA OSCOUT FE-Pwr-Sync
220uH 13 6 U402-rt
D E E S CB RT
14 7 U402-ct
fVcc EB CT

2
D403 IsoXfmr R409
Q402 1 5.11K 2
V+
+ C431 + C434 f-V T401-4 MMBT3906L U402-sen 4 1 TP412 TP413
+SENSE V-

GND

U402-1
10uF 10uF R406 5 9
-SENSE COMP

3
0.2
C407 MBRS130T3 TP414 TP415 R408 TP416 C410 LM3524DM

8
+
C404 100uF 100 .01uF
S S TP417
.047uF Q401-s
C411 R410
2200pF 10K
3 S S 3
TP468 L402
220uH fVcc
R417 +5V
fVee
10K
TP419
+ C432 + C435 TP420 U410

8
10uF 10uF R403 4N25
U401-3 3 1K R411
+ 1 U401-1 iso401-a 1 6 iso401-b 5.11K
U401-2 2
S S -
R418 5 opto-fb
5.11K U401A

4
MC34072 4
2
fVcc
R404
S S S
316K
TP405 TP421 TP422
TP407 R416 R401 C408 R402
8

R415 10K 10K .01uf 10K


100 5 U401-5 C408-2
+ s+5V s+5V
U401-7 7
f+5V
6
-
U401B + C418
C416 MC34072 R420
4

4.7uF
.047uF 10K

S
C417
S S
680pf
TP411 S

2 2

Nonin SpO2
Connector
J401

1
SpO2 ESD
fVcc
J403 Ground
2
sSpO2tx iso+5Vdig
3 1
+ C419
4 2
10uF C413
Tab-.187 .047uF
CON4

S
D E
+5V
S

3
Q403
sSpO2tx 1 MMBT3904L
U411
R412 HCNW4503

2
511 R414
TP423 TP424
Q403-e iso402-a 2 8 6.19K
R421
Nellcor SpO2 10K
6 u411-6
Connector SpO2Out
3
5
C422 C423
J402 .047uF 1000pF
D E
1 2
+ +
3 4
+ + SpO2-Reset
5 6
+ +
7 8
s-5V + +
9 10
+ + s+5V
11 12
+ +
13 14
1 iso+5Vdig + + 1
C429 + C437
CON14AP .047uF 330uF

S S

C427 C428
.047uF
+
10uF
S S
Initial Date
Welch Allyn Inc.
Drawn Schematic: 620002.dsn
Rick Myers 3/05/99
S S
Designed
Rick Myers 3/05/99
Title
Checked
Isolated Power and SpO2 interface
L.Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620002 G
Release For Production
J. Bello 3/17/99
Date: Monday, August 05, 2002 Sheet 4 of 6
A B C D E

Figure D-4. Atlas Power isolation /SpO2 interface.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 150


A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39118 JAC 3/15/99 LPP


U501 U502 U503 U504 U505 U506 U507 U508

fVcc
B See ECN worksheet 5-39456 LPP 4/27/99 LPP

+ C534 C518 C520 C522 C524 C526 C528 C530 C532


Improvements for ESD
4.7uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF C 5-39822 LPP 6/23/99 LPP
and Fast Transients

Mods to U2 clock line,


D 5-39934 LPP 7/9/99 LPP
E E E E E E E E E NVRAM pullup

Add Nellcor SpO2 cabability.


fVee
F Modify RLD to reduce V-lead noise. 5-44460 LPP 3/26/02 JK
Improve ESU rejection.
C535 C519 C521 C523 C525 C527 C529 C531 C533 Add 330uF bypass for Nellcor 506 board.
+
4.7uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF G 5-44857 LPP 7/18/02 JK

4 4

E E E E E E E E E

RA

Input Buffers LA A/D Inputs

ESU filter Gain = 9.26 Leads Off


LL

fVcc
V+
J501 TP501 TP503 Lead Select
CON6 R501 R561 R508 TP504 TP505 fVcc

8
19.6K 73.2K 10K
Right-Arm C501-1 C537-1 U501-3 3 TP506
1 + fVcc
1 U505
2

8
2 Wilson 4051
3 -
R509 13 3 mux- 3
4 X0 X +
N501 C501 C537 C506 TP507 22Meg U501A R517 Network RA 14 1 diff-
5 X1

4
330pF 220pF 100pF TL072 10K C510 TP508 LA 15 2 C544
6 X2 -
LAMP NEON TP509 0.1% 220pF R529 10K aVR 12 Diff Amp 1000pF
X3
R509-1 aVL 1 U507A
X4

4
fVee aVF 5 TL072 Gain = 1
S E E E X5
R557 U501-2 R530 10K V- 2
E X6
Shield 22Meg 4 R544 R546
X7
fVee 10K 0.1% 10K 0.1%
R518 TP510 6 U508-3
INH
fVcc 1.21K R531 10K 11
A
VLdsOff 0.1% 10
E B
TP534 TP511 9
C
R502 R562 R510 TP512 TP513 R532 10K C543

8
19.6K 73.2K 10K TP514 fVcc TP557 TP558 .047uF
E
Left-Arm C502-1 C538-1 U501-5 5 fVcc u509-4
+ fVee
7 TP515 E
6 R533 10K TP516 fVcc
3 - fVcc 3

8
R511 U506 R565

2
N502 C502 C538 C507 22Meg U501B R519 4051 R545 3 121K

4
LA 13 3 mux+ 5 + 1 4
330pF 220pF 100pF TP519 TL072 10K C511 R534 10K 10K 0.1%
X0 X + -
LAMP NEON TP520 0.1% 220pF LL 14 7 diff+ U508-2 2 1 diff-out
X1 -
R511-1 LL 15 6 U508A 3
X2 - +
fVee TP521 RA 12 AD712 U509
S E E E X3

4
U501-6 LA 1 TP522

U508-1
R558 R535 10K U507B LMC7101
X4

5
22Meg LL 5 TL072
X5 E
V+ 2 C542 fVee
X6
R520 R536 10K 4 fVee 220pF
X7
fVcc 1.21K fVee
VLdsOff 0.1% 6
INH
TP541 TP523 R537 10K 11
A
R503 R563 R512 TP524 TP525 10 R547
B
8

19.6K 73.2K 10K 9 10K C545


E E C
Left-Leg C503-1 C539-1 U502-3 3 0.1% 1000pF
+
High Pass (0.05Hz or 0.5Hz) Gain Stage
1 fVcc
2 Low Pass (150Hz) Gain = 74
- fVee
R513 TP517
E E E
N503 C503 C539 C508 22Meg U502A R521 TP528 TP529 C515 R550 TP518
4

8
330pF 220pF 100pF TP531 TL072 10K C512 3.3uF 19.6K
LAMP NEON TP533 0.1% 220pF ecg-hp U508-5 5
R513-1 + 7
An-ECG
fVee 6
S E E E -
R559 U502-2 C516
22Meg R548 R549 .047uF U508B

4
TP535 TP536 TP537 1Meg 110K AD712 R551 C517
R522 73.2K .01uF
fVcc 1.21K
E
VLdsOff 0.1% Q504-d fVee
Mux-LdS-A
TP544 TP538 U508-6
E
R504 R564 R514 TP539 TP540
Mux-LdS-B
8

19.6K 73.2K 10K TP526 TP527


E
V-Lead C504-1 C540-1 U502-5 5 R552
+ Mux-LdS-C
7 1K
6
-

3
R515 TP530
2 N504 C504 C540 C509 22Meg U502B R523 2
4

330pF 220pF 100pF TP542 TL072 10K C513


E
LAMP NEON 0.1% 220pF TP532 Q504 1
R515-1 TP543 2N7002
fVee
S E E E

2
R560 U502-6 Q504-s R553

2
22Meg 100K
1 Q504-g
Sw-MonBw
R524 Q506
1.21K 2N7002
VLdsOff 0.1%
fVcc

3
U504
3 2 TP547
S1 D1 E E
1 TP548 U503B R555
Sw-RLD-RA IN1
8

TL072 30.1K
14 15 5 U503-5
S2 D2 + f+5V
16 7
Sw-RLD-LL IN2 VLdsOff
6
TP550 11 10 -
TP545
S3 D3
R505 R566 9 C536 R556
Sw-RLD-LA IN3 f+1.24V
4

19.6K 73.2K .01uF


Right-Leg C505-1 C541-1 6 7 10K
S4 D4
8
Sw-RLD-RL IN4
fVee
13
fVcc V+ E E
N505 C505 C541 4
fVee V-
330pF 220pF 5
GND
LAMP NEON
DG202CSE
S E E E

fVcc TP549 fVcc

TP559
2

Q501 Q502 U503A R525


vRL
8

NL NL R516 TL072 10K


Q501-S

1 D505 10K 3 1
3 3 2 2 3 RL-mux 1 +
MMBD1503A
2 R526
- E
V-Lead Defib Protect, 10K
U503-2
TP562
4

Diodes internal to U504 RLD Amp


1

R567
protect other inputs.
NL Initial Date
Welch Allyn Inc.
fVee Q501-G TP560 R527 fVee C514
Sw-RLD-V
1K 0.1uF Drawn Schematic: 620002.dsn
Rick Myers 3/05/99
C514-2
C546 No Load, Designed
Rick Myers 3/05/99
NL R528 TP552 Title
Allows RLD switch TP551 51.1K Checked
ECG Amp
L.Phillips 3/16/99
to V-Lead, not used. Rev
Approved Size Document Number
E Z. Psenicnik 3/16/99
C 620002 G
Release For Production
J. Bello 3/17/99
Date: Monday, August 05, 2002 Sheet 5 of 6
A B C D E

Figure D-5. Atlas model 200 ECG amp.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 151


A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production
U601 U602 U603 5-39118 JAC 3/15/99 LPP
fVcc

C622 + C609 C623 C624 B See ECN worksheet 5-39456 LPP 4/27/99 LPP
.047uF 4.7uF .047uF .047uF

Improvements for ESD


C 5-39822 LPP 6/23/99 LPP
and Fast Transients
E E E E
Mods to U2 clock line,
D 5-39934 LPP 7/9/99 LPP
NVRAM pullup

U601 U602 U603 Add Nellcor SpO2 cabability.


fVee F 5-44460 LPP 3/26/02 JK
Modify RLD to reduce V-lead noise.
fVcc Improve ESU rejection.
C625 C614 C626 C627 Isolation Add 330uF bypass for Nellcor 506
+ board.
.047uF 4.7uF .047uF .047uF fVcc G 5-44857 LPP 7/18/02 JK
4 Barrier 4

R623
E E E E
46.4K TP625

8
5
+
Sw-RLD-V 7
6 f+1.24V
-
U607B iso+5Vdig
LM393 iso+5Vdig

4
iso+5Vdig
C629
U604 .047uF
f+5V +5V
U613
fVee Serial/Parallel TP627 R613 C641 HCNW4503
+ C631 f+5V 6.19K 0.1uF
D E
4.7uF fVcc Control Data TP602 U604A 8 2
U605 74HCT04 R625
D E
fVcc 74HCT595 10K R619

U604-1
15 14 6 619
E U604-2 2 1 U613-6
QA SER
SpO2-Reset 1 3 iso604-a FE-Data-In
R622 TP601 QB
2 11 5
46.4K QC SRCLK C634
SW-RL-V 3 10
QD SRCLR

8
4 220pF
3 QE
Sw-DiagBw 5 12 E D E
+ QF RCLK
1 6 13 TP603
Sw-MonBw Sw-RLD-RL QG G D E
2 7
- f+1.24V Sw-RLD-LL QH
U607A
LM393 9

4
TP604 QH
TP621 TP622 iso+5Vdig
TP605

D E
fVee
U605-out
U612 +5V
TP623 TP624 TP628 R615 C642 HCNW4503
U606 f+5V 6.19K 0.1uF
3 74HCT595 U604B 3
8 2
74HCT04 R626
Sw-RLD-LA 15 14 D E
QA SER 10K R620

u604-3
Sw-RLD-RA 1
QB 4 3 6 619
Mux-LdS-C 2 11 U604-4 U612-6
QC SRCLK 3
Mux-LdS-B 3 10 iso603-a FE-Data-Clock
QD SRCLR 5
Mux-LdS-A 4
QE C635
Mux-A/D-C 5 12
QF RCLK 220pF
Mux-A/D-B 6 13
QG G
Mux-A/D-A 7 E D E
QH TP606 iso+5Vdig
U605-12 D E
9
QH C637
fVee
1000pF
C628 U611 +5V
TP607 9 R628 TP629 R617 0.1uF HCNW4503
C D E
TP608 10 10K f+5V 2.15K
B D E
11 U604C
A 8 2
TP609 6 A/D Mux 74HCT04 R627
INH D E
U603 5.11K R621

U604-5
4 4051 6 5 6 215
vRL X7 U604-6 U611-6
2 3
f+1.24V X6 iso602-a FE-ADC-Clock
5 TP610 TP611 TP612 5
RA X5
1 C636
LA X4
12 100pF
LL X3
15 R605 R607
V+ X2 E D E

U603out
14 3.48K 464K

U602-2
An-ECG X1 D E
13 3
X0 X
iso+5Vdig
iso+5Vdig
TP613 fVcc
E
fVcc

U602-5
TP614
R601 fVcc f+5V

U602-7
10K R608 C632
U601-3 C603 1K TP615 .047uF
f+5V
680pF +5V

8
6
5
8

U604-8

3
TP616 2
+
R602 3 7 9 8 1 Q601
2 + E 2
73.2K 1 U601-1 U602-3 3 MMBT3904L
-
2 U602
-

2
R606 LM311 U604D R611 C621

1
4
f+5V U601A 3.48K 74HCT04 6.19K 0.1uF
4

TP617 R603 MC34072 R610 U610


E E

Q601-e
46.4K 511 HCNW4503
TP630 U601-2 iso601-a 2 8
C601 fVee TP618 TP626
fVee R624
11 10
PWM E
6
U604-10 TP619 U610-6 FE-PWM-ADC
fVcc R604 1000pF 3
A/D
3

10K 5 5.11K
U604E 2 1 Converter C633
D601-2
74HCT04 100pF
D E
8

TP620 E
5 D601
+ 7 MMBD1503A
6
-
E
U601B
4

MC34072

fVee
U601-7
52uSec
Integrator
f+5V Input
781uSec
(U604-10)

+5V
Comparator
13 12
Inputs

Analog In
U604F
74HCT04 -5V
1 1
E E

PWM-Out
(U604-8)

Initial Date
Welch Allyn Inc.
PWM-timer
Drawn Schematic: 620002.dsn
Rick Myers 3/05/99
Designed
Rick Myers 3/05/99 Title
Checked
ECG A/D and Interface
L.Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620002 F
Release For Production
J. Bello 3/17/99 Date: Monday, August 05, 2002 Sheet 6 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 152


Figure D-6. Atlas ECG A/D interface
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

Component change for


B 5-39353 JJC 4/15/99 ZIP
frequency modulation

Change Reset IC U3,


C 5-39830 LPP 6/24/99 LPP
No Load C40

D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK

4 4

CPU PCA SW Debug Port


(aka BDM Connector)
Main PCA Connectors VDDH
JP5

VFLS0 1 2 SRESET_n
JP8 3 4 DSCK
GND
5 6 VFLS1
GND 1 2
1 2 EKG-PWM-ADC HRESET_n 7 8 DSDI
3 4 GND
NIBP-Fault 3 4 9 10 DSDO
5 6
+3.3V_unfiltered 5 6 NIBP-PWM-ADC
7 8 GND HEADER 5X2
+3.3V_unfiltered 7 8
GND 9 10
9 10 RS423-Tx
11 12
NIBP-Overpressure 11 12 Spare-1
GND 13 14
13 14 RS423-Rx
15 16 1 - Visual History Buffer Flush Status ('823 output)
ETCO2-Tx 15 16 SpO2-Tx
17 18
ETCO2-Rx 17 18 SpO2-Rx 2 - Reset Output ('823 input)
19 20
Spare-2 19 20 ADC-Clock 3 - Ground
21 22
NIBP-Data 21 22 FE-Clock 4 - BDM Clock ('823 input)
23 24
NIBP-15mmHg 23 24 SPI-data-out
25 26 5 - Ground
3 LED-Latch 25 26 SPI-clock 3
27 28 6 - Visual History Buffer Flush Status ('823 output)
Video-Horizontal-Sync 27 28 Batt-ID
29 30
Spare-3 29 30 FE-Data 7 - Reset Input ('823 Output)
31 32
Main-Power-Sync 31 32 Recorder-Motor-Step 8 - BDM Data Input ('823 Input)
33 34 GND
Vbackup 33 34
35 36 9 - Vcc
NIBP-Valve 35 36 On-Standby-Key
37 38 10 - BDM Data Output ('823 Output)
Video-Vertical-Sync 37 38 Speaker-Audio
39 40
NIBP-Offset-DAC 39 40 Main-Supply-On
41 42
AC-On 41 42 NIBP2-PW-ADC
43 44 GND
Rec-Paper-Out 43 44
GND 45 46
45 46 Video-Data
47 48 GND
Key-Latch 47 48
49 50
RS423-CTS 49 50 Nurse-Call
51 52
Recorder-Clock 51 52 Recorder-Data-Latch
53 54
Recorder-Data 53 54 Recorder-Strobe
55 56
HRESET 55 56 SCL
57 58
SPI-data-in 57 58 SCD
GND 59 60 GND
59 60

HEADER 30x2

Test Points

1 TP58 TP59 1
TS_n PA4

1 TP60 TP61 1
TA_n PA5

1 TP62 TP63 1
2 Burst_n PA6 2
1 TP64 TP65 1
PA14 PA7

1 TP66 TP67 1
PC6 PA15

1 TP68 TP69 1
PC13 RD/WR_n

1 TP70 TP71 1
CLKOUT SPIMOSI

1 TP72 TP73 1
PC10 SPIMISO

1 TP74 TP75 1
PC12 SPISEL_n

1 TP76 TP77 1
PB16 SPICLK

TP78 1
Video-Clock-3

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked
L. Phillips 3/19/99 Atlas CPU Subsystem
Approved Size Document Number Rev
Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99 Date: Monday, August 05, 2002 Sheet 1 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 153


Figure D-7. Atlas CPU subsystem.
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

B Component change for 5-39353 JJC 4/15/99 ZIP


frequency modulation
Change Reset IC U3,
C 5-39830 LPP 6/24/99 LPP
No Load C40

D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK

4 4

IRQ_n[0:7]
A[6:31] GPL[0:7] CS_n[0:7]

A25
IRQ_n2 1 100 A26
P1 P100
IRQ_n1 2 99 A27
P2 P99
3 98 A28
TIN1 P3 P98
7 97 A29
TIN2 P7 P97
8 96 GPL1 D[0:31]
TIN3 P8 P96
9 95
NIBP-Fault-x P9 P95
10 U7 94 CS_n2
PA6 TDI-10 P94
11 93
PA7 P11 P93 RD/WR_n
12 EPF6016A 92 D0
Field GCLK-12 P92
13 91 D1
SPICLK GCLK-13 DEV_CLRn-91
14 90 D2 VIDEO[0:7]
ADC-Clock-x P14 P90
15 85 D3
76.8KHz P15 DEV_OE-85
16 84 D4
FE-Clock-x P16 P84
17 83 D5
FE-Data-x P17 nRS-83
18 82 D6
PB16 TMS-18 P82
19 81 D7
NIBP-Data-x P19 nWS-81
23 80 VIDEO0
PC10 TCK-23 P80
24 79 VIDEO1
3 PA5 P24 P79 3
25 78 VIDEO2
DREQ2_n P25 Cs-78
26 77 VIDEO3
DREQ1_n P26 nCS-77
27 76 VIDEO4
SDACK2_n P27 P76
28 75 VIDEO5
SDACK1_n P28 P75
29 74 VIDEO6
Recorder-Data-x P29 P74
30 73 VIDEO7
Recorder-Data-Latch-x P30 P73
31 69
Recorder-Clock-x P31 CLKUSR-69 VSYNC
32 67
Recorder-Strobe-x P32 RDYnBSY-67 HSYNC
33 66
Recorder-Motor-Step-x P33 P66 Blank
34 65
NIBP-Offset-DAC-x P34 P65 Video-Clock-3
35 64
NIBP-Overpressure-x P35 INIT_DONE-64 SPIMISO
40 63
NIBP-15mmHg-x P40 GCLK-63 CLKOUT
S7 41 62
P41 GCLK-62 Video-Clock-25
S6 42 61
P42 P61 SPIMOSI
S5 43 60
P43 P60 SPISEL_n
S4 44 59
P44 P59 Video-Data-x
S3 45 58
P45 P58 HSYNC-x
S2 46 68
P46 P68 VSYNC-x
S1 47 57
P47 P57 SPI-data-in-x
S0 48 56
P48 P56 SPI-clock-x
49 55
PA14 nCEO-49 P55 SPI-data-out-x
50 52
Key-Latch-x P50 P52 LED-Latch-x

CONFIG_DONE
51
PA15 TDO-51

nCONFIG
nSTATUS

GND-20
GND-37
GND-53
GND-70
GND-87
VCC-21
VCC-38
VCC-54
VCC-71
VCC-88

GND-5
VCC-6
DATA
MSEL

DCLK
nCE
22
39
36
72
86
89

21
38
54
71
88

20
37
53
70
87
4

5
Audio DAC Resistors
S0 R66 100K
Speaker-Audio
R125

2
1.58K S1 R67 49.9K
NSTATUS C121
PC12
VDDH S2 R68 24.9K

1
330pF
2 PC6 2
SPICLK S3 R69 12.4K
CONFIG_DONE
PC11
SPIMOSI S4 R70 6.19K
33
1

S5 R71 3.09K
R126
S6 R72 1.58K
TP80

S7 R73 787

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked
L. Phillips 3/19/99 FPGA
Approved Size Document Number Rev
Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99
Date: Monday, August 05, 2002 Sheet 2 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 154


Figure D-8. Atlas FPGA.
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

B Component change for 5-39353 JJC 4/15/99 ZIP


frequency modulation
Change Reset IC U3,
C 5-39830 LPP 6/24/99 LPP
No Load C40

D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK

4 4

R74
VDDH R76 1 8
SCD I2CSDA
1 2 7
SCL SCL I2CSCL
5 2 3 6
SCD SpO2-Rx SMRXD2
3 4 5
SPI-data-in SpO2-Tx SMTXD2
4
VFLS0

1
10K
100
6 C54 C55 C56 C57
VFLS1 VDDH BR_n
7 330pF 330pF 330pF 330pF
DSDI

2
10 8 6.19K R127
SRESET_n R77
9
RS423-CTS
C 1 8
ETCO2-Rx SMRXD1 VDDH BI_n
2 7
ETCO2-Tx SMTXD1
3 6 6.19K R128
RS423-Rx RXD2
VDDH R103 4 5
RS423-Tx TXD2
1
PA4 VDDH BB_n

1
5 2
PA5 100
3 C62 C63 C64 C65 3.09K R129
PC12
4 330pF 330pF 330pF 330pF
PA7

2
10K
R80 VDDH TA_n
6
PC10
7 1 8 3.09K R130
NIBP2-PW-ADC RS423-CTS PC9
10 8 GPL5 2 7
Nurse-Call PB18
9 GPL4 3 6
Key-Latch Key-Latch-x VDDH TEA_n
C 4 5 MODCK (0,0) Selects 32KHz as main oscillator
NIBP-Valve PA6
3.09K R131
RSTCONF = 0 allows '823 to sample data bus during HRESET

1
100
C70 C71 C72 C73
GPL[0:7] DSCK
330pF 330pF 330pF 330pF VDDH

2
1.58K R132
IRQ_n[0:7] R82
R83
VDDH 1 8
SPI-clock SPI-clock-x ALE_B
1 IRQ_n7 2 7
SPI-data-out SPI-data-out-x
5 2 IRQ_n6 3 6 NL R133 R104 R105
LED-Latch LED-Latch-x
3 IRQ_n5 4 5
3 SPI-data-in SPI-data-in-x NL 1.58K 3
4 IRQ_n4
RSTCONF_n

1
10K
100
6 IRQ_n3 C78 C79 C80 C81 1.58K R134
MODCK1
7 IRQ_n2 330pF 330pF 330pF 330pF

2
10 8 IRQ_n1
R85 TRST_n MODCK2
9 IRQ_n0
C 1 8 1.58K R152
Recorder-Data Recorder-Data-x
2 7
Recorder-Strobe Recorder-Strobe-x
VDDH R49 3 6 R107 R108
Recorder-Clock Recorder-Clock-x
1 4 5 R151
RS423-Rx Recorder-Data-Latch Recorder-Data-Latch-x NL 1.58K
5 2 215
NIBP-Fault

1
3
SpO2-Rx 100 Video-Data Video-Data-x
4 C86 C87 C88 C89
NIBP-15mmHg

1
10K 330pF 330pF 330pF 330pF
2

2
6 C122
NIBP-Overpressure R87
7 22pF VDDH
AC-On

2
10 8 1 8
PA15 Video-Horizontal-Sync HSYNC-x
9 2 7
ETCO2-Rx Video-Vertical-Sync VSYNC-x
C 3 6
Batt-ID PC8
4 5
NIBP-Data NIBP-Data-x
VDDH R86
1

1
1 R78
NIBP-PWM-ADC 100
5 2 C90 C91 C92 C93 1 8
Rec-Paper-Out Recorder-Motor-Step Recorder-Motor-Step-x
3 330pF 330pF 330pF 330pF 2 7
PC11 Main-Power-Sync 76.8KHz
2

2
4 3 6
TMS R88 Main-Supply-On TEXP
10K 4 5
On-Standby-Key PC4
6 1 8
EKG-PWM-ADC NIBP-Offset-DAC NIBP-Offset-DAC-x

1
7 2 7 NL
Spare-3 Rec-Paper-Out PC7 100
10 8 3 6 C66 C67 C68 C69 IRQ_n0
Batt-ID
9 4 5 330pF 330pF 330pF 330pF
PB16 AC-On PC5

2
C R120
1

1 R81
100
C94 C95 C96 C97 1 8
EKG-PWM-ADC TIN1
330pF 330pF 330pF 330pF 2 7
NIBP-15mmHg NIBP-15mmHg-x
2

VDDH R16 3 6
R90 NIBP-Overpressure NIBP-Overpressure-x
1 4 5
SPKROUT NIBP-PWM-ADC TIN2
5 2 1 8
2 TS_n FE-Clock FE-Clock-x 2

1
3 2 7
BG_n FE-Data FE-Data-x 100
4 3 6 C74 C75 C76 C77
Burst_n ADC-Clock ADC-Clock-x
10K 4 5 330pF 330pF 330pF 330pF PWM Input Timer Signals
NIBP-Fault NIBP-Fault-x

2
6
Spare-1
1

7
WAIT_B_n 100
10 8 IP_B2 C102 C103 C104 C105
9 IP_B3 330pF 330pF 330pF 330pF
2

R116
VDDH R91 10K 1 8
IP_B4 Spare-1 PB19
1 2 7
IP_B5 Spare-2 PC13
5 2 3 6
IP_B6 Spare-3 PB17
3 4 5
IP_B7 NIBP2-PW-ADC TIN3
4
1

100
6 C123 C124 C125 C126
FRZ IP_B[2:7]
7 330pF 330pF 330pF 330pF
DREQ1_n
2

10 8
DREQ2_n
9
On-Standby-Key
C

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked
L. Phillips 3/19/99
I/O Filters
Approved Size Document Number Rev
Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99
Date: Monday, August 05, 2002 Sheet 3 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 155


Figure D-9. Atlas 200/210/220 I/O filters.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

B Component change for 5-39353 JJC 4/15/99 ZIP


frequency modulation
Change Reset IC U3,
C 5-39830 LPP 6/24/99 LPP
No Load C40

D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK

4 4

HRESET_n

D[0:31]

A[6:31]

A10 A9 VDDH VDDH

1
U5
TP53 TP54
SDRAM-4Mx16

43
49

14
27
CS_n[0:7]

3
9

1
A30 23 2 D15
A0 DQ0

VccQ1
VccQ2
VccQ3
VccQ4

Vcc1
Vcc2
Vcc3
A29 24 4 D14
A1 DQ1
A28 25 5 D13
A2 DQ2
A27 26 7 D12 VDDH
GPL[0:7] A3 DQ3
A26 29 8 D11
A4 DQ4
A25 30 10 D10
A5 DQ5
A24 31 11 D9
A6 DQ6

2
GPL0 A23 32 13 D8 Boot Configuration Pull-Up: 0 1 0 0 1 0 0 1 1 1 1 0 0 0 0 0
A7 DQ7
U5-25 A22 33 42 D7 R57
A8 DQ8 (see page 4-10)
U5-35 A21 34 44 D6
A9 DQ9 100K
22 45 D5
A10 DQ10 0: internal arbitration
R153 R154 35 47 D4 19
A11 DQ11 2G

1
3 48 D3 1 1: MSR(ip) = 0, interrupt prefix is 0x0000 3
6.19K 6.19K DQ12 1G
A11 20 50 D2
BA0 DQ13 0: reserved
21 51 D1 D1 3 17 dx1
BA1 DQ14 2Y4 2A4 0: memory controller is active after reset
53 D0 D4 5 15 dx4
DQ15 2Y3 2A3 1,0: 16-bit boot device
CS_n1 19 D7 7 13 dx7
CS# 2Y2 2A2
GPL1 18 15 D8 9 11 dx8 0: reserved
RAS# DQML 2Y1 2A1
GPL3 16 39 D9 12 8 dx9 1,1: IMMR base is 0xfff0-0000
WE# DQMH 1Y4 1A4
GPL2 17 D10 14 6 dx10
CAS# 1Y3 1A3 1,1: DBGC mode
40 16 4 dx11
NC2 1Y2 1A2 0,0: DBPC mode
38 36 18 2 dx12
CLKOUT CLK NC/A12 1Y1 1A1

U5-36
37 0,0: External bus division factor
CKE
U4

VssQ1
VssQ2
VssQ3
VssQ4

Vss1
Vss2
Vss3
74AHC244
T1
R155 Boot Configuration Latch

12
46
52

28
41
54
6
VDDH 6.19K
Active during power-on
BS_AB[0:3]
VDDH

BS_AB0

BS_AB1

VDDH

FLASH-512Kx16

U6
13

37
A30 25 29 D15
A0 D0
NC3

Vcc
A29 24 31 D14
2 A1 D1 2
A28 23 33 D13
A2 D2
A27 22 35 D12
A3 D3
A26 21 38 D11
A4 D4
A25 20 40 D10
A5 D5
A24 19 42 D9
A6 D6
A23 18 44 D8
A7 D7
A22 8 30 D7 AMD N/C: 9, 10, 13, 14
A8 D8
A21 7 32 D6
A9 D9 AMD RY/BY#: 15
A20 6 34 D5
A10 D10 ----------------
A19 5 36 D4
A11 D11
A18 4 39 D3 MICRON N/C: 9, 10, 15
A12 D12
A17 3 41 D2 MICRON Vpp: 13
A13 D13
A16 2 43 D1
A14 D14
A15 1 45 D0
T2 A15 D15
A14 48
A16
A13 17 47
A17 Byte# VDDH
A12 16 14
A18 NC4

CS_n0 26 9
CE# NC1
GPL1 28 10
OE# NC2
11 15
WE# RY/BY#
12
Gnd-b

Gnd-a

SRESET_n RESET#
1
27

46

TP55

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked
Flash, SDRAM and HRESET Config Word
L. Phillips 3/19/99

Approved Size Document Number Rev


Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99
Date: Monday, August 05, 2002 Sheet 4 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 156


Figure D-10. Flash, SDRAM HRESET config word.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

B Component change for 5-39353 JJC 4/15/99 ZIP


frequency modulation
Change Reset IC U3,
C 5-39830 LPP 6/24/99 LPP
No Load C40

D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK

4 4

BS_AB[0:3]
Burst_n BS_AB[0:3]
RD/WR_n
GPL[0:7]
TSIZ[0:1] GPL[0:7]
TSIZ[0:1] IRQ_n[0:7]
D[0:31] IRQ_n[0:7]

TEA_n
TA_n
TS_n

BS_AB0
BS_AB1
BS_AB2
BS_AB3
D[0:31] CS_n[0:7]

GPL7

GPL0
GPL1
GPL2
GPL3
GPL4
GPL5
GPL6
A[6:31] CS_n[0:7]

BG_n
BB_n
BR_n
BI_n
A[6:31]

IRQ_n0
IRQ_n1
IRQ_n2
IRQ_n3

IRQ_n4
IRQ_n5
IRQ_n6

IRQ_n7
TSIZ0
TSIZ1
D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23
D24
D25
D26
D27
D28
D29
D30
D31

CS_n0
CS_n1
CS_n2
CS_n3
CS_n4
CS_n5
CS_n6
CS_n7
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9

SPKROUT
A10
A11
A12
A13
A14
A15
A16
A17
A18
A19
A20
A21
A22
A23
A24
A25
A26
A27
A28
A29
A30
A31
TS_n
A6
A7
A8
A9

TA_n

FRZ
TEA_n
BI_n
BR_n
M13

M15

M16
M14

D10

D16

D15

D14
D11

D12

D13
N15
N16

H15
H13
H14

H16
G13

G14

G16

G15

A13

A12

A11

A10

A14

A15
C13

C11

C10

C16
C15

C12

C14
K14
K13

K15

K16

B10

B12
B11

B13

B14

B16

B15
E15

E16

E13
F14

F16

F15

F13
L13

L14
L15
L16

J15
J14

M1

M2

M4
M3

D2

D9

D4
D3
N1
N2

N3
H1

H2

H3

H4
G2
G3

G4

C3

C2
K2
K3
K1

K4

B7
E1

E4

E2

E3
F1

F2

F3

F4
L1

L2

L4

L3
J2
J1

J3
J4
BG_n
BB_n

D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
A6
A7
A8
A9

D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23
D24
D25
D26
D27
D28
D29
D30
D31

BG_n
A10
A11
A12
A13
A14
A15
A16
A17
A18
A19
A20
A21
A22
A23
A24
A25
A26
A27
A28
A29
A30
A31

TA_n

BB_n
BR_n
TEA_n
TS_n

CS0_n
CS1_n
CS2_n
CS3_n
CS4_n
CS5_n
GPL_A5_n

IRQ0_n
IRQ1_n

IRQ7_n
BI_n
TSIZ1
RD / WR_n
Burst_n
SPKROUT

DP0 / IRQ3_n

DP1 / IRQ4_n
DP2 / IRQ5_n
DP3 / IRQ6_n
FRZ / IRQ6_n
UPWAITB / GPL_B4_n

CS6_n / CE1_B_n
CS7_n / CE2_B_n
RSV_n / IRQ2_n
TSIZ0 / REG_n

GPL_A0_n / GPL_B0_n
A16

BDIP_n / GPL_B5_n

WE3_n / BS_AB3_n / PCWE_n


NC1

WE1_n / BS_AB1_n / IOWR_n


WE2_n / BS_AB2_n / PCOE_n
FRZ

UPWAITA / GPL_A4_n / AS_n


WE0_n / BS_AB0_n / IORD_n

GPL_A1_n / GPL_B1_n / OE_n


GPL_A2_n / GPL_B2_n / CS2_n
GPL_A3_n / GPL_B3_n / CS3_n
VDDH C1
NC2

KR_n / IRQ4_n / RETRY / SPKROUT


C6
NC3

E5
VDDH1
E6
VDDH2
E7
VDDH3
E8 P16
VDDH4 PA15 / USBRXD PA15
E9 R15
VDDH5 PA14 / USBOE_n PA14
E10 R14
VDDH6 PA13 / RXD2 RXD2
E11 R13
VDDH7 PA12 / TXD2 TXD2
E12 N10
VDDH8 PA9 / L1TXDA / SMRXD2 SMRXD2
F5 T9
VDDH9 PA8 / L1RXDA / SMTXD2 SMTXD2
Place VSSSYN F12 T8
3 VDDH10 PA7 / CLK1 / TIN1 / L1RCLKA / BRGO1 PA7 3
on a split G5 P8
VDDH11 PA6 / CLK2 / TOUT1_n / TIN3 PA6
G12 T6
plane beneath VDDH12 PA5 / CLK3 / TIN2 / L1TCLKA / BRGO2 PA5
H5 R6
the MPC823 PLL VDDH13 PA4 / CLK4 / TOUT2_n / TIN4 PA4
H12
section VDDH14
N14
PB31 / SPISEL_n / LCD_A SPISEL_n
P15
PB30 / SPICLK SPICLK
J5 P14
VDDH17 PB29 / SPIMOSI SPIMOSI
Attach GND and J12 T15
VSSSYN at a K5
VDDH18 MPC823 PB28 / SPIMISO
T14
SPIMISO
VDDH19 PB27 / I2CSDA / BRGO1 I2CSDA
K12 P12
single point. VDDH20 (the monster) PB26 / I2CSCL / BRGO2 I2CSCL
L5 N11
VDDH21 PB25 / SMTXD1 SMTXD1
L12 T11
VDDH22 PB24 / SMRXD1 SMRXD1
M5 T10
VDDH23 PB23 / SMSYN1_n / SDACK1_n SDACK1_n
M6 MPC823ZC R9
VDDH24 PB22 / SMSYN2_n / SDACK2_n SDACK2_n
M7 R7
VDDH25 PB19 / L1ST1 / LCD_B PB19
M8 U1 P7
VDDH26 PB18 / RTS2_n / L1ST2 PB18
M9 N7
VDDH27 PB17 / L1ST3 / LCD_C PB17
M10 R5
VDDH28 PB16 / L1RQA / L1ST4 PB16
VDDL M11
VDDH29
M12 R16
VDDH30 PC15 / DREQ1_n / L1ST5 DREQ1_n
T16
PC14 / DREQ2_n / RTS2_n / L1ST6 DREQ2_n
A7 P13
VDDL1 PC13 / L1ST7 PC13
G1 T13
VDDL2 PC12 / L1RQA / L1ST8 PC12
J16 R10
VDDL3 PC11 / USBRXP PC11
T7 P9
VDDL4 PC10 / TGATE1_n / USBRXN PC10

BLANK / LOE / LCD_AC / PD6


R8
PC9 / CTS2_n PC9

IP_B2 / IOIS16_B_n / AT2


N8

OP2 / MODCK1 / STS_n


OP3 / MODCK2 / DSDO
PC8 / CD2_n / TGATE1_n PC8

VSYNC / FRAME / PD5


E14 T5

HSYNC / LOAD / PD4


IP_B0 / IWP0 / VFLS0
IP_B1 / IWP1 / VFLS1

IP_B7 / PTR_n / AT3


ALE_B / DSCK / AT1
NC4 PC7 / USBTXP PC7

IP_B4 / LWP0 / VF0


IP_B5 / LWP1 / VF1
IP_B3 / IWP2 / VF2

IP_B6 / DSDI / AT0


J13 N6

FIELD / LD0 / PD7


VD7 / LD8 / PD15
VD6 / LD7 / PD14
VD5 / LD6 / PD13
VD4 / LD5 / PD12
VD3 / LD4 / PD11
VD2 / LD3 / PD10
CLK-SHIFT / PD3

VD1 / LD2 / PD9


VD0 / LD1 / PD8
NC5 PC6 / USBTXN PC6
KAPWR VDDSYN P6

TDO / DSDO
RSTCONF_n

PC5 / L1TSYNCA / SDACK1_n PC5


PORESET_n

TCK / DSCK

WAIT_B_n
T4

TDI / DSDI
HRESET_n
SRESET_n
VSSSYN1
VDDSYN

PC4 / L1RSYNCA
CLKOUT
PC4
VSSSYN

EXTCLK

TRST_n
KAPWR
GND10
GND11
GND12
GND13
GND14
GND15
GND16
GND17
GND18

GND25
GND26
GND27
GND28
GND29
GND30
GND31
GND32
GND33
GND34
GND35
GND36
GND37
GND38
GND39
GND40
GND41
GND42

EXTAL
GND1
GND2
GND3
GND4
GND5
GND6
GND7
GND8
GND9

XTAL

TEXP

TMS
NC6
NC7
NC8
NC9

XFC
N13

N12
H10
H11
G10
G11

K10
K11

R12
R11
P10

P11
T12
F10
F11

L10
L11
J10
J11

D1

D5

D7

D8

D6
N9

N4

N5
H6
H7
H8
H9
G6
G7
G8
G9

A1
A2

A3

A4
A5

A6

A8

A9
C5

C4

C8

C9
C7
K6
K7
K8
K9

B1

B3

B5
B4

B2

B8

B9

B6

R1
R2
R4

R3
P1

P3
P2

P5

P4
T3

T2

T1
F6
F7
F8
F9

L6
L7
L8
L9
J6
J7
J8
J9
VSSSYN

VIDEO[0:7]
GND

2 Field 2

VIDEO7
VIDEO6
VIDEO5
VIDEO4
VIDEO3
VIDEO2
VIDEO1
VIDEO0
Blank

IP_B2
IP_B3
IP_B4
IP_B5
IP_B6
IP_B7
VSYNC
PORESET_n
RSTCONF_n HSYNC
HRESET_n
SRESET_n Video-Clock-3
XTAL
EXTAL MODCK1
XFC
CLKOUT MODCK2
EXTCLK
TEXP
DSCK IP_B[2:7]
DSDI TMS IP_B[2:7]
DSDO
VFLS0 ALE_B
VFLS1 TRST_n WAIT_B_n

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked
L. Phillips 3/19/99
PowerPC MPC823 CPU
Approved Size Document Number Rev
Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99
Date: Monday, August 05, 2002 Sheet 5 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 157


Figure D-11. Power PC MP C823 CPU.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39173 JAC 3/18/99 LPP

B Component change for 5-39353 JJC 4/15/99 ZIP


MPC823 PLL frequency modulation
Supply Change Reset IC U3,
C 5-39830 LPP 6/24/99 LPP
No Load C40
FM Section for spreading RF
emissions. R2 selects the
to CPU PLL VDDSYN deviation. D Upgrade to larger SDRAM 5-44759 LPP 6/20/02 JK
pin B1
1 2
+3.3V_unfiltered

L2 Ferrite-120

1
C2 R1
100nF 100
4 4

1
C18 R2 From FPGA

2
C17 C117 C107 1K
100nF 330pF 6.8uF NL XFCX
PA14

1
C1
1.5uF

2
VSSSYN PLL Filter
Capacitor
System Clock
to CPU PLL pins A1
XFC
and A2
R50 200K VDDH
xtalp
XTAL
Place decoupling capacitors next to CPU VDDSYN pin B1 and VSSSYN pin A1 and A2
3.9M R51
FPGA Clock
Use separate PPL VDDSYN power plane from 3.3V plain
EXTAL

Split ground plane under these components


R52
Y1
Video-Clock-25
1536*830 < XFC < 1536*1470 pf Value selected for the final mulitplication ratio 1 2 G1
33
R94

1
Load C107 w/1.5uF and unload C1 for non-FM clocking 32768Hz 1 4
OE VDD EXTCLK
C108 C109
NL
22pF 22pF 2 3 VCLK
GND OUT

2
R117
Place these components
200K
as close to the SG-636PCE-25.175M
processor as possible

VDDH
RF Suppression
L1 VDDH VDDL FLASH decoupling caps '244
3 Ferrite-120 to CPU U2 3
decoupling cap
VDDH
System 3.0V

1
1 2 2 1
+3.3V_unfiltered Input Reset_n
C110 C4 C5 C6 C7 C106
6.8uF 100nF 100nF 100nF 100nF 100nF

2
Gnd
3
C14 C115 330pF 6.8uF
1

1
C12 C13 C114 6.8uF 330pF C116 C16
DRAM decoupling caps VDDH
6.8uF 100nF 330pF
VDDH
2

2
Oscillator

1
decoupling
C111 C19 C20 C21 C22 6.8uF cap

1
6.8uF 100nF 100nF 100nF 100nF C120

2
C128
100nF

2
MPC823 decoupling caps
R118
VDDH
NL 3

1
Q1-Base 1 Q1
HRESET
NL C112 C28 C29 C30 C31 C32
6.8uF 100nF 100nF 100nF 100nF 100nF
2

2
1

C130 R119
NL NL
VDDH
2

FPGA decoupling caps


R124
VDDH
1.58K

1
KAPWR
2 to CPU 32KHz C113 C41 C42 C43 C44 C15 2
R121 6.8uF 100nF 100nF 100nF 100nF 100nF
Supply RF/EMI Clock

2
HRESET_n
Filters and Reset
NL
Circuits 12.4K 100K
PORESET-Input

RF Suppression R122 R123


L3
Ferrite-120 U3
2.0V
1 2 2 1
Vbackup Input Reset_n PORESET_n
Gnd

C40
1

C38 C39 C118 C119 100nF


3

C36 C37
6.8uF 100nF 330pF 100nF 6.8uF 330pF
2

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620005.dsn
Jim Belesiu Mar 5, 1999

Designed
Jim Belesiu Mar 5, 1999 Title
Checked Power Distribution, Clocks and Reset
L. Phillips 3/19/99

Approved Size Document Number Rev


Z. Psenicnik 3/19/99
C 620005 D
Release For Production
J. Bello 3/19/99
Date: Monday, August 05, 2002 Sheet 6 of 6
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 158


Figure D-12. Power distribution.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39251 JAC 3/23/99 LPP

B No Load D16 5-39485 LPP 4/28/99 LPP

C D13 to Low Profile, R51 to 316 5-39994 LPP 7/23/99

Break-Away PCB Section


4 4
Flag Pole Area
Trend SW9
Trend-sw 2 1

Lead Select SW11


Lead-Select-sw 2 1

Print SW13
Print-sw 2 1

ACON-LED

Keyboard Circuits
R65 1.21K
1 2
On/Off-sw 5x5 Green
On/Off-sw Trend J2A
J2xA D1
Clock1 Trend 1 2 ACLED-1 1 2
Place capacitors Key-Clock-b Lead-Select 1 1
Lead-Select
2 2
by connector Key-Latch-b Print
Print
3 3
R66 1.21K
VCC VCC AC-On-x AC On LEDs
4 4
Key-data-b
Key-Data-b 5 5
Silence-Switch
1 2 ACLED-2 1 2
+ C9 + C27 flex-cable-5 flex-cable-5
R67 1.21K D2
4.7uF 4.7uF Front Panel - A
VCC 1 2 5x5 Green
TP182
R68 1.21K
R71
J1 U26 100
Key-Clock 2 18 u26-18 FP-Common
1 2 1A1 1Y1
Key-Latch-n 4 16 Key-Latch-b
3 3 4 1A2 1Y2 3
Key-Data 6 14
5 6 1A3 1Y3
LED-Data 8 12 LED-Data-b
7 8 1A4 1Y4
LED-Enable 11 9 u26-9
9 10 2A1 2Y1
13 7 u26-7
11 12 2A2 2Y2
LED-Latch-n 15 5 LED-Latch-1
13 14 2A3 2Y3
R51 17 3 LED-Latch-2
2A4 2Y4
316 HEADER 7X2
1
1G
19
2G
1
4
3
2

On-Led R1 74HC244
2.2K
1

D13 I705144
8
5
6
7

T1 (3mm) Green

LED_TP_Enable
On/Standby
TP183 R72
LED_TP_Data_In
2

VCC 100
Clock2 Silence Button
LED_TP_Clock

LED_TP_Latch Break-Away PCB


R2
LED_TP_Data_Out Section
U1D 2.15K D10-Vcc
U1A Front Panel - B
9 8
D10
1 2 LED_TP_enable TP2 VCC D10-C 2 1
u19-out
74HC04
74HC04 T1 (3mm) Red
J3A
NIBP LED Circuits J3xA Silence LED
U1E 1 1
U1B Silence-LED
LED_NIBP_Enable Silence-LED 2 2
11 10 Silence-Switch Sil-Sw-a
3 3
3 4 LED_NIBP_enable TP3 TP184 R73
LED_NIBP_Data_In 4 4
100 SW22
2 5 5 2
74HC04 Clock3 Sil-Sw-b 1 2
LED_NIBP_Clock
74HC04
flex-cable-5 flex-cable-5
LED_NIBP_Latch
Silence
LED_NIBP_Data_Out
U1F U1C
Switch

13 12 5 6 LED_SPO2_enable TP4 1 2
Front Panel - C

u13-out SW28
74HC04 74HC04

I705124

LED_SpO2_Enable

VCC VCC VCC VCC VCC


LED_SpO2_Data_In

LED_SpO2_Clock

+ C1 + C8 + C10 + C11 + C28 Place one tantalum cap near LED_SpO2_Latch


4.7uF 4.7uF 4.7uF 4.7uF 4.7uF each seven segment LED
group
Front Panel - D

VCC VCC VCC VCC VCC VCC VCC VCC VCC VCC

C2 C3 C4 C5 C6 C7 C12 C13 C14 C15

0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF
1 1

Place one 9 8 7 6

ceramic cap
near each IC
VCC VCC VCC VCC VCC VCC VCC VCC VCC VCC VCC 10 5
Initial Date
Welch Allyn Inc.
Drawn Schematic:
Jim Belesiu 3/15/99
C16 C17 C18 C19 C20 C21 C22 C23 C24 C25 C26
1 2 3 4 Designed
0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF 0.1uF Jim Belesiu 3/15/99 Title
Pull-up Resistor Network Detail
Checked
Front Panel Display Board
L. Phillips 3/24/99

Approved Size Document Number Rev


Z. Psenicnik 3/24/99
C 620008 C
Release For Production
J. Bello 3/24/99
Date: Friday, July 23, 1999 Sheet 1 of 5
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 159


Figure D-13. Front panel display.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39251 JAC 3/23/99 LPP

B No Load D16 5-39485 LPP 4/28/99 LPP

C D13 to Low Profile, R51 to 316 5-39994 LPP 7/23/99

4 4

VCC
TP66

SW23
1 2 9 8 7 6
VCC

10

5
C
ETCO2 Alarm On-Off SW24 R5
1 2 10K 10 5
TP67 TP68 TP69 TP70 TP71
U2

10
6
7
8
9

1
2
3
4

5
C
SW1 u2-10 10
SER
1 2 etco2-sw 11 1 2 3 4 R6
A
HRalarm-sw 12 Pull-up Resistor Network Detail 10K
B
HR Alarm On-Off LHselect-sw 13
C

9
8
7
6

4
3
2
1
SW2 LHAdjust+sw 14 TP72 TP73 TP74 TP75 TP76
D
1 2 LHAdjust-sw 3 U3
E
4 SW4 10
F SER
5 QRS Volume (+) 1 2 QRSvol+sw 11
G TP77 A
SW3 6 9 u2-9 QRSvol-sw 12
H QH B
LH Select 1 2 Alarmvol+sw 13
C
2 7 SW6 Alarmvol-sw 14
CLK QH D
15 QRS Volume (-) 1 2 Time-sw 3
INH E
SW5 1 U3-4 4
SH/LD F
LH Adjust (+) 1 2 5
G
SW8 6 9
3 H QH 3
74HC165 Alarm Volume (+) 1 2 TP178
SW7 2 7
CLK QH
LH Adjust (-) 1 2 15
INH
SW10 1
SH/LD
Alarm Volume (-) 1 2

74HC165
SW12
Trend Time 1 2
Trend

Lead-Select
Lead-Select D48

1
Print VCC
Print BAV99L

SW25

2
1 2
10

5
C

NIBP Alarm On-Off SW26 R7 On / Standby


1 2 10K
TP79 TP80 TP81 TP82
SW14
9
8
7
6

4
3
2
1

U4 1 2
SW15 On/Off-sw
10
SER
1 2 NIBPalarm-sw 11
A
SpO2alarm-sw 12 SW27
B
RHselect-sw 13 1 2
C
SpO2 Alarm On-Off SW16 RHadjust+sw 14
D
1 2 RHadjust-sw 3 TP83
E
BPauto-sw 4
F
BPstart/stop-sw 5
G
SW17 6 9
2 H QH Key-Data-b 2
1 2
RH Select 2 7
CLK QH
15
INH
SW18 TP84 TP85 TP86 1
SH/LD
1 2 TP87
RH Adjust (+)
74HC165
SW19
Key-Latch-b
1 2
RH Adjust (-) TP88

SW20
BP Auto 1 2 Clock1
Key-Clock-b

SW21
BP Start-Stop 1 2
Silence-Switch

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620008.dsn
Jim Belesiu 3/15/99

Designed
Jim Belesiu 3/15/99 Title
Keyboard Scanner
Checked
L. Phillips 3/23/99

Approved Size Document Number Rev


Z. Psenicnik 3/23/99
C 620008 A
C
Release For Production
J. Bello 3/23/99
Date: Friday, July 23, 1999 Sheet 2 of 5
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 160


Figure D-14. Key board scanner.
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39251 JAC 3/23/99 LPP

B No Load D16 5-39485 LPP 4/28/99 LPP

C D13 to Low Profile, R51 to 316 5-39994 LPP 7/23/99

4 4

VCC

TP10 TP12
TP14 R26 Temperature Tens VCC R27 VCC R28
150 150 Temperature Ones 150 Temperature Tenths
U14 5 4 D34 U15 u15-15 8 1 D35 U16 u16-15 8 1 D36
LED-Data-b 14 15 u14-15 6 3 d34-g 3 14 15 u15-1 5 4 d35-g 3 14 15 u16-1 5 4 d36-g 3
LED_TP_Data_In SER QA "G" SER QA "G" SER QA "G"
1 u14-1 7 2 d34-f 2 1 u15-2 6 3 d35-f 2 1 u16-2 6 3 d36-f 2
QB "F" A QB "F" A QB "F" A
11 2 u14-2 8 1 d34-e 4 11 2 u15-3 7 2 d35-e 4 11 2 u16-3 7 2 d36-e 4
SRCLK QC "E" F B SRCLK QC "E" F B SRCLK QC "E" F B
10 3 u14-3 d34-d 5 10 3 d35-d 5 10 3 d36-d 5
SRCLR QD "D" G SRCLR QD "D" G SRCLR QD "D" G
TP5 4 u14-4 d34-c 8 4 d35-c 8 4 d36-c 8
QE "C" E C QE "C" E C QE "C" E C
12 5 u14-5 5 4 d34-b 9 12 5 u15-4 5 4 d35-b 9 12 5 u16-4 5 4 d36-b 9
RCLK QF "B" D RCLK QF "B" D RCLK QF "B" D
13 6 u14-6 6 3 d34-a 10 13 6 u15-5 6 3 d35-a 10 13 6 u16-5 6 3 d36-a 10
G QG "A" G QG "A" G QG "A"
7 u14-7 7 2 7 7 u15-6 7 2 d35-dp 7 7 u16-6 7 2 7
QH Dp QH Dp QH Dp
8 1 u15-7 8 1 8 1
9 1 9 1 9 1
QH Anode-1 QH Anode-1 QH Anode-1
TP21 R29 6 R30 6 R31 6
Anode-2 Anode-2 Anode-2
74HC595A 150 74HC595A 150 74HC595A 150
7-SEG LED 0.36 TP25 7-SEG LED 0.36 TP30 7-SEG LED 0.36

3 VCC VCC 3
VCC

R53 Degree C
215 u15-9 R52 Degree F
D47
d47-c 2 1 215
VCC D46
u16-7 d46-c 2 1
VCC
TP34
TP22 T1.75 (5mm) Green
TP35 T1.75 (5mm) Green

u14-9

Temperature Hundreds
TP33 D33
3
"G"
2
"F" A
4
"E" F B
5
"D" G
d33-c 8
"C" E C
d33-b 9
"B" D
10
"A"
7
Dp

1
Anode-1
6
TP37 VCC Anode-2
u16-9
7-SEG LED 0.36

TP38 TP40 TP42


VCC R32 VCC R33 VCC R34 SpO2 Pulse Rate Ones
150 SpO2 Pulse Rate Hundreds 150 SpO2 Pulse Rate Tens 150
U17 u17-15 8 1 D37 U18 u18-15 8 1 D38 U19 u19-15 8 1 D39
2 14 15 u17-1 5 4 d37-g 3 14 15 u18-1 5 4 d38-g 3 14 15 u19-1 5 4 d39-g 3 2
SER QA "G" SER QA "G" SER QA "G"
1 u17-2 6 3 d37-f 2 1 u18-2 6 3 d38-f 2 1 u19-2 6 3 d39-f 2
QB "F" A QB "F" A QB "F" A
11 2 u17-3 7 2 d37-e 4 11 2 u18-3 7 2 d38-e 4 11 2 u19-3 7 2 d39-e 4
SRCLK QC "E" F B SRCLK QC "E" F B SRCLK QC "E" F B
10 3 d37-d 5 10 3 d38-d 5 10 3 d39-d 5
SRCLR QD "D" G SRCLR QD "D" G SRCLR QD "D" G
4 d37-c 8 4 d38-c 8 4 d39-c 8
QE "C" E C QE "C" E C QE "C" E C
12 5 u17-4 5 4 d37-b 9 12 5 u18-4 5 4 d38-b 9 12 5 u19-4 5 4 d39-b 9
RCLK QF "B" D RCLK QF "B" D RCLK QF "B" D
13 6 u17-5 6 3 d37-a 10 13 6 u18-5 6 3 d38-a 10 13 6 u19-5 6 3 d39-a 10
G QG "A" G QG "A" G QG "A"
7 u17-6 7 2 7 7 u18-6 7 2 7 7 u19-6 7 2 7
QH Dp QH Dp QH Dp
8 1 u18-7 8 1 u19-7 8 1
9 1 9 1 9 1
QH Anode-1 QH Anode-1 QH Anode-1
R35 6 R36 6 R37 6
Anode-2 Anode-2 Anode-2
74HC595A TP50 150 74HC595A 150 74HC595A 150
TP52 TP64 7-SEG LED 0.36
7-SEG LED 0.36 7-SEG LED 0.36
VCC VCC VCC
u17-9
TP60 u18-9
TP61 TP62 TP63
TP59

LED-TP-Enable
LED_TP_Enable
Clock2
LED_TP_Clock
LED-Latch-1
LED_TP_Latch

u19-out
LED_TP_Data_Out

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620008.dsn
Jim Belesiu 3/15/99

Designed
Jim Belesiu 3/15/99 Title
Checked
Temp/Pulse LED Drivers
L. Phillips 3/24/99

Approved Size Document Number Rev


Z. Psenicnik 3/24/99
C 620008 C
Release For Production
J. Bello 3/24/99
Date: Friday, July 23, 1999 Sheet 3 of 5
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 161


Figure D-15. Temp/pulse LED drivers.
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5.-39251 JAC 3/23/99 LPP

B No Load D16 5-39485 LPP 4/28/99 LPP

C D13 to Low Profile, R51 to 316 5-39994 LPP 7/23/99

VCC
TP89 VCC VCC
4 R38 Systolic Hundreds TP91 R39 Systolic Tens TP93 R40 Systolic Ones 4
680 680 680
U20 u20-15 8 1 D40 U21 8 1 D41 U22 8 1 D42
u19-out 14 15 u20-1 5 4 d40-g 10 14 15 u21-15 5 4 d41-g 10 14 15 U22-15 5 4 d42-g 10
LED_NIBP_Data_In SER QA "G" SER QA "G" SER QA "G"
1 u20-2 6 3 d40-f 9 1 u21-1 6 3 d41-f 9 1 u22-1 6 3 d42-f 9
QB "F" A QB "F" A QB "F" A
11 2 u20-3 7 2 d40-e 1 11 2 u21-2 7 2 d41-e 1 11 2 u22-2 7 2 d42-e 1
SRCLK QC "E" F B SRCLK QC "E" F B SRCLK QC "E" F B
10 3 d40-d 2 10 3 u21-3 d41-d 2 10 3 u22-3 d42-d 2
SRCLR QD "D" G SRCLR QD "D" G SRCLR QD "D" G
4 d40-c 4 4 u21-4 d41-c 4 4 u22-4 d42-c 4
QE "C" E C QE "C" E C QE "C" E C
12 5 u20-4 5 4 d40-b 6 12 5 u21-5 5 4 d41-b 6 12 5 u22-5 5 4 d42-b 6
RCLK QF "B" D RCLK QF "B" D RCLK QF "B" D
13 6 u20-5 6 3 d40-a 7 13 6 u21-6 6 3 d41-a 7 13 6 u22-6 6 3 d42-a 7
G QG "A" G QG "A" G QG "A"
7 u20-6 7 2 5 7 u21-7 7 2 5 7 u22-7 7 2 5
QH Dp QH Dp QH Dp
8 1 8 1 8 1
9 3 9 3 9 3
QH Anode-1 QH Anode-1 QH Anode-1
R41 8 TP101 R42 8 TP103 R43 8
Anode-2 Anode-2 Anode-2
74HC595A TP105 680 74HC595A 680 74HC595A 680
7-SEG LED 0.5 7-SEG LED 0.5 7-SEG LED 0.5

TP180

R54 NIBP Alarm On/Off R55 BP Sys Hi R56 BP Sys Lo


2.15K 215 215
D15 D6 D8
u20-7 d15-c 2 1 d6-c 2 1 d8-c 2 1
VCC VCC VCC

TP110 TP111
T1 (3mm) Red T1.75 (5mm) Green T1.75 (5mm) Green
TP112 u20-out TP113 u21-out

TP181

u22-out

3 3

VCC
VCC VCC 680
TP114 R44 Diastolic Hundreds TP116 R45 Diastolic Tens TP118 Diastolic Ones
R46
680 680
U23 8 1 D43 U24 8 1 D44 U25 8 1 D45
14 15 u23-15 5 4 d43-g 10 14 15 u24-15 5 4 d44-g 10 14 15 u25-15 5 4 d45-g 10
SER QA "G" SER QA "G" SER QA "G"
1 u23-1 6 3 d43-f 9 1 u24-1 6 3 d44-f 9 1 u25-1 6 3 d45-f 9
QB "F" A QB "F" A QB "F" A
11 2 u23-2 7 2 d43-e 1 11 2 u24-2 7 2 d44-e 1 11 2 u25-2 7 2 d45-e 1
SRCLK QC "E" F B SRCLK QC "E" F B SRCLK QC "E" F B
10 3 u23-3 d43-d 2 10 3 u24-3 d44-d 2 10 3 u25-3 d45-d 2
SRCLR QD "D" G SRCLR QD "D" G SRCLR QD "D" G
4 u23-4 d43-c 4 4 u24-4 d44-c 4 4 u25-4 d45-c 4
QE "C" E C QE "C" E C QE "C" E C
12 5 u23-5 5 4 d43-b 6 12 5 u24-5 5 4 d44-b 6 12 5 u25-5 5 4 d45-b 6
RCLK QF "B" D RCLK QF "B" D RCLK QF "B" D
13 6 u23-6 6 3 d43-a 7 13 6 u24-6 6 3 d44-a 7 13 6 u25-6 6 3 d45-a 7
G QG "A" G QG "A" G QG "A"
7 u23-7 7 2 5 7 u24-7 7 2 5 7 u25-7 7 2 5
QH Dp QH Dp QH Dp
8 1 8 1 8 1
9 3 9 3 9 3
QH Anode-1 QH Anode-1 QH Anode-1
TP126 R47 8 TP128 R48 8 TP130 8
Anode-2 Anode-2 Anode-2
74HC595A 680 74HC595A 680 74HC595A 680
TP132 7-SEG LED 0.5 7-SEG LED 0.5 R49 7-SEG LED 0.5

R58 BP Dia Hi BP Dia Lo R57


VCC
215 215
D11 D14
d11-c 2 1 d14-c 2 1
VCC VCC

TP136 TP137 T1.75 (5mm) Green


TP138 TP141 T1.75 (5mm) Green
TP139 u23-out TP140 u24-out

LED-Latch-2
LED_NIBP_Latch

Clock3
LED_NIBP_Clock

LED-NIBP-Enable
2 LED_NIBP_Enable 2

TP142

u25-out

T1.75 (5mm) Green

T1.75 (5mm) Green


T1.75 (5mm) Green

T1.75 (5mm) Green

T1.75 (5mm) Green

T1.75 (5mm) Green

T1.75 (5mm) Green

T1.75 (5mm) Green


R59
1.21K

Silence-LED VCC

1
D31 D30 D29 D28 D25 D26 D24 D22
Q60 Q30 Q15 Q10 Q5 Q3 Q1 Qx
TP179

2
u13-out
LED_NIBP_Data_Out

VCC TP1 R24


150
U13 5 4 d31-a
14 15 u13-15 6 3 d30-a
SER QA
1 u13-1 7 2 d29-a
QB
11 2 u13-2 8 1 d28-a
SRCLK QC
10 3 u13-3
SRCLR QD
4 u13-4 d25-a
QE
12 5 u13-5 5 4 d26-a
RCLK QF
13 6 u13-6 6 3 d24-a
G QG
7 u13-7 7 2 d22-a
QH
8 1
9
QH
R25
74HC595A TP8 150

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620008.dsn
Jim Belesiu 3/15/99

Designed
Jim Belesiu 3/15/99 Title
Checked
NIBP LED Drivers
L. Phillips 3/24/99

Approved Size Document Number Rev


Z. Psenicnik 3/24/99
C 620008 C
Release For Production
J. Bello 3/24/99
Date: Friday, July 23, 1999 Sheet 4 of 5
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 162


Figure D-16. NIBP LED drivers.
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39251 JAC 3/23/99 LPP

B No Load D16 5-39485 LPP 4/28/99 LPP

C D13 to Low Profile, R51 to 316 5-39994 LPP 7/23/99

4 4

VCC

TP143 R21 SpO2 Bargraph VCC


680
U12 5 4 D27
u13-out 14 15 u12-15 6 3 d27-20 20 1
LED_SpO2_Data_In SER QA
1 u12-1 7 2 d27-19 19 2
QB
11 2 u12-2 8 1 d27-18 18 3
SRCLK QC
10 3 u12-3 d27-17 17 4
SRCLR QD
4 u12-4 d27-16 16 5
QE
12 5 u12-5 5 4 d27-15 15 6
RCLK QF
13 6 u12-6 6 3 d27-14 14 7
G QG
7 u12-7 7 2 d27-13 13 8
QH
8 1 d27-12 12 9
9 d27-11 11 10
QH
R23
74HC595A TP149 680
TP177 10 LED Bargraph

u12-out TP151 R69


681

VCC R70
TP152 681

3 3
1

1
D7 D9 D12
T1 (3mm) Red T1 (3mm) Red T1 (3mm) Red

HR Alarm ETCO2 Alarm SpO2 Alarm


2

2
On/Off On/Off On/Off
d9-c
VCC
d7-c d12-c VCC VCC
SpO2 % Hundreds TP153 R9 TP155 R10
680 SpO2 % Tens 680 SpO2 % Ones
U5 TP162 D3 U6 8 1 D4 U7 8 1 D5
14 15 u5-15 R62 10 14 15 u6-15 7 2 d4-g 10 14 15 u7-15 6 2 d5-g 10
SER QA "G" SER QA "G" SER QA "G"
1 u5-1 2.2K 9 1 u6-1 6 3 d4-f 9 2 u7-1 7 3 d5-f 9
QB "F" A QB "F" A QC "F" A
11 2 u5-2 8 1 1 11 2 u6-2 5 4 d4-e 1 11 1 u7-2 5 4 d5-e 1
SRCLK QC "E" F B SRCLK QC "E" F B SRCLK QB "E" F B
10 3 u5-3 7 2 2 10 3 u6-3 d4-d 2 10 3 u7-3 d5-d 2
SRCLR QD "D" G SRCLR QD "D" G SRCLR QD "D" G
4 u5-4 6 3 d3-c 4 4 u6-4 d4-c 4 4 u7-4 d5-c 4
QE "C" E C QE "C" E C QE "C" E C
12 5 u5-5 5 4 d3-b 6 12 5 u6-5 8 1 d4-b 6 12 5 u7-5 8 1 d5-b 6
RCLK QF "B" D RCLK QF "B" D RCLK QF "B" D
13 6 u5-6 7 13 6 u6-6 7 2 d4-a 7 13 6 u7-6 7 2 d5-a 7
G QG "A" G QG "A" G QG "A"
7 TP175 5 7 u6-7 6 3 5 7 u7-7 6 3 5
QH Dp QH Dp QH Dp
5 4 5 4
9 u5-out 3 9 u6-out 3 9 3
QH Anode-1 QH Anode-1 QH Anode-1
8 R12 8 R13 8
Anode-2 Anode-2 Anode-2
74HC595A R63 74HC595A TP168 680 74HC595A TP170 680
215 7-SEG LED 0.5 7-SEG LED 0.5 7-SEG LED 0.5
D16
d16-c 2 1 VCC
VCC
u7-out VCC
SpO2 Hi VCC TP173
TP172 No Load
R64
215
D18
d18-c 2 1
VCC

SpO2 Lo TP176
TP174 T1.75 (5mm) Green

2 Clock1 2
LED_SpO2_Clock

LED-Latch-2
LED_SpO2_Latch

LED-SPO2-Enable
LED_SpO2_Enable

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620008.dsn
Jim Belesiu 3/15/99

Designed
Jim Belesiu 3/15/99 Title
Checked
SPO2 LED Drivers
L. Phillips 3/24/99

Approved Size Document Number Rev


Z. Psenicnik 3/24/99
C 620008 C
Release For Production
J. Bello 3/24/99
Date: Friday, July 23, 1999 Sheet 5 of 5
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 163


Figure D-17. SpO2 LED drivers.

A B C D E
Width Coil (L3) inductance range changed
REV Description ECN/ECO Init Date Ckd

Vertical Amplifier A Release to Production 5-39097 JAC 3/9/99 LPP

Vertical Sync R1 Ramp Generator


1K u2.1 B Modify Video Amplifier 5-39339 JJC 4/15/99 ZIP
U1.3 +12V +12V
+5V Reduce RF Emissions
+12V Increase Flex Cable to
C 5-39831 LPP 6/24/99 LPP
3 inches
R2
7.50K Width Coil Inductance
C3 C4 + C5 +12V D 5-40006 LPP 7/23/99
(L3) increased
.047uF .047uF R12 22uF R19
750uSec

3
562 562
q1.b 2 Q1 Q3 2 q3.b +5V
16.7mSec

8
MPSW01A MPSW01A

8
U4A R3 3 R13 R18 R20

1
+ 1 ramp 3 5
23.7K 169 R15 38.3 562 169 R22
1 2 U4.2 2 + 1 q1.e 7 +
10K
4 - 2 6 4
U1A
- -
TLC272 U2A R14 R21 U2B

1
R52 74HCT04 MC34072 562 562 MC34072

4
1.96K R43 q2.b 2 vadj 2 q4.b
C1
5.11K
.1uF C2 Q2 R16 10 R17 10 Q4

3
U1.2 100pf MPSW51A MPSW51A
u2.2 Size Adjust
u2.7 u2.6
d6.c 2 1 2 1
R4 R23 To Yoke
D6 D1 10K 10K
1N4148 1N4148 R5 R6 J2
42.2K 41.2K vert-
5
u1.6 vert+
4
j2-3
3
To Main hor+
2
hor-
Board +12V
1

4
U3 CON5
J1 MC78L05ACP - 6 R7
3 1 u1.7 7 41.2K R31
1 VIN VOUT +5V
Vsync + 5 u1.5 10K
2
Hsync U1B
3
GND
TLC272
4

8
Video C6 C7
5
.047uF .047uF R8
6
2

41.2K
Zero Adjust
CON6 +12V

vcadj
R9 R11
vcadj+ vcadj-
+5V

1.47K R10 1K 1.96K


6 Position flat ribbon cable
3 3

Connects Main CRT to CRT


socket Board
Horizontal Sync +12V C9 Break Away CRT socket Board
6.8uF
+12V hlin hwide
1/2W Carbon Comp
R26 L2 L3

4
2
1K C8 15uH 15- 35uH +12V R44 330
R24 R27 Q6 .022uF J3.1
10K 10 IRF640
5.5uSec
q5.c q6.d 1 J3 J3x J4
R45 330
U4B C22 R25 tube.g1
31.7uSec 1 1 1
3

.022uF 215 tube.vd


2 2 2

3
3 4 u4.4 C22-2 q5.b 2 Q5 J3.3 tube.12V
3 3 3
2N3904 To CRT tube.gnd
4 4 4
T1 J3.5 tube.g1
Anode 5 5 R46 47K 5
1

74HCT04 CRT-FBT J3.6 tube.g2


6 6 6
R50 tube.g4
1.96K Horizontal Amplifier Flex6 Flex6
R47 47K
7

CRT-Socket

Boost Grid Voltages

1
E5

E4

E3

E2

E1
C11 +
1000uF

2
9
+12V 8 D3
D2 1N4935 Grid1
+Vvid
L1
+12filt 1 2 +B 10 2 Vd 1 2

2 47uH C15 2

+ MUR120 + C12 .01uF


C10 Hor 7 5 22uF R29 100K
1000uF D4 R28
1N4935 Brightness 19.6K

3 Vg1 2 1 d4.a Adjust


Unused
1 Vg2

+ C13
+Vvid R30
Video Amplifier 22uF
100K
Grid2
U4C

5 6
D5
1N4937 C16
.01uF 74HCT04
1 2 d5.c
U4D
C14
C19 .01uF 9 8
Focus
+5V R36 R37 .01uF R32
1.96K 1.96K Adjust 2.5Meg Focus
74HCT04
C17
3

+5V C18 .01uF U4F


.047uF q8-b 2 Q8
2N3904 13 12
1

R51 2 1 R41
19.6K 681 74HCT04
1

D7 q8-e vidout
U4E R48 1N4148
100
11 10 u4-6 q7-g 2 Q7
BS170 C20
No Load
3

1 1
74HCT04

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620011.dsn
Rick Myers Mar 4 1999

Designed
Rick Myers Mar 4 1999
Title
Checked
CRT Deflection Board
L.Phillips 3/9/99

Approved Size Document Number Rev


Z.Psenicnik 3/9/99
C 620011 D
Release For Production J.Bello 3/9/99
Date: Friday, July 23, 1999 Sheet 1 of 1
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 164


Figure D-18. CRT deflection board.

A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39179 JAC 3/18/99 LPP

Add EMI bead to Q4, improve


B 5-41395 JB 5/9/00
radiated emissions

4 4

+5V
Temperature
+5V Amplifier

C1
.047uF
R1
51.1K
RecVdc
TP3
A

8
R2
U1-3 3 1K
+ 1
Rec-Temp
C2 U1-2 2
-
R4 .01uF U1A
R9 TP9 9.09K MC34072

4
3
10K
TP54
Q2-b 1
+24V
Q2
A A TP6
MMBT3904L U1-1
+24V Boost A

2
R10
10K R7 (23.6 V nominal) C3
A
84.5K +12V R5 .01uF
C4 L1 q1-g 1K
TP10 +
1000uF 22uH
Printhead
A
U2-fb R8 23.7K
On/Off Switch
R3

1
+ C6 D2 10K
R12 1uF MBRS340T3 C5 .01uF
A
10K C7 4
.01uF TP14 Q4-D 3 2
U2

12

11
A
UC3843A +5V
Thermistor
+5V TP8 3 + C8
A VFB Q1 Input

U2-out
3 3

Vc

2
4
Vcc
1800uF R16

2
U2-comp 1 R13 Q4 19.6K
COMP MTP30P06V
+5V 10 MTP30N06VL D3

Q4-G
J1 TP17 U2-Ref 14 10 1
VREF OUT A

3
CON24A R6 3 BAV99L
Vpp Rec-Supply-On 10K TP19 R15
1 2

3
Rec-Motor-Step

PWR-GND
C11 R14 147 E1
3 4 A
+12V 1 .047uF 10K U2-RT 7 5 Q4-s 2643000301
5 6 RT/CT ISENSE

GND
+12V
Rec-Data-Latch Rec-Strobe Q3
7 8 Place Bead on

1
Rec-Clock Rec-Data 2N7002
9 10
Pin 2 of Q4
2

11 12 A

U2-Isense
C21 TP21 C13
13 14 A
.047uF 5600pF R17 R18
15 16 A

1
C14 0.1 no-load
17 18 A A
2200pF
19 20 A A
C12 3 2 Q6-e Printhead
21 22
.047uF
23 24 A
Q6 TP23 R19
Connector
MMBT3904L 3.16K +24V-switch J2
1
A +5V 2
Thermistor1
3
R22 Thermistor2
D D 4
4.22K 3 4 Strobe
A 5
6
U5B
RecVdc 7
74HCT14 +5V
8
R11
+5V D 9
5.11K
10
Strobe
11
5 6 Latch
12
Clock
13
U5C Data
+5V 14
74HCT14
15
D 16
1 2 C10
+24V .1uF CON16
2 U5A 2
+5V

1
74HCT14

D A A
TP26 R33 D6 13 12
+5V +5V C18 1.47K 27V
.047uF U3-10 U5F

3
TP24 74HCT14
TP1
d6-c
D

1
11 10 u3-clk
D
14
1

U5E U3 D1
R24 74HCT14 27V
Vdd

Vss

1.96K U6
D

3
2 13 data-IO TP27 R34 ULN2065
ClkIn/Osc1/RB5 RB0
1.47K u6-3 3 2 +24V
TP66 1B 1C
U3-3 3 12 clk U3-9 u6-6 6 7
Clkout/Osc2/RB4 RB1 2B 2C
u6-11 11 9
D TP67 3B 3C
Vpp 4 11 n24Von u6-14 14 16
Vpp/MCLR/RB3 RB2 4B 4C

TP68 U3-5 5 10 1 +
RC5/TOCK1 RC0 CLMP1
8 u6-clmp C19 C20
CLMP2
U3-6 6 9 .047uF 47uF
RC4 RC1 Connector Pin #
7 8
RC3 RC2 1 3 4 6
+5V R35
A A
1.47K
PIC16C505 U3-8
phase 1 + +
A
8

5 J3-1
J3 2 + +
+ 1
7 TP28
6 J3-3
2 3 + +
- 3
U1B J3-4
A 4 4 + +
MC34072
5
4

J3-6
6
TP29 R36
1.47K CON6
1 A 1
u1-7 U3-7

+5V

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620014.dsn
9 8 Rick Myers Mar 5, 1999

Designed
Rick Myers Mar 5, 1999 Title
U5D
74HCT14 Checked
Printer Electronics
L.Phillips 3/22/99

Approved Size Document Number Rev


D Z. Psenicnik 3/19/99
C 620014 B
Release For Production
J. Bello 3/19/99
Date: Thursday, May 11, 2000 Sheet 1 of 1
A B C D E

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 165


Figure D-19. Printer electronics.
A B C D E

REV Description ECN/ECO Init Date Ckd

+5V EKG - Page6 A Release to Production 5-39121 JAC 3/15/99 LPP

EKG-PWM-ADC 6 5
FE-PWM-ADC
B See ECN worksheet 5-39455 LPP 4/27/99 LPP
J1 U1C
AC+12V +5V
74HCT04 Improvements for ESD, RFI,
1
TP7 D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
2 D
Respiration Diff-Amp Topology
3
EKG-Data-Clock 9 8 Mods to U2 clock line, Resp
4 FE-Data-Clock
E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
5 +5V
U1D NVRAM pullup
6
74HCT04 TP8 Buffer for ETCO2, Nurse call relay
7
+5V F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
D
C22 EKG-ADC-Clock 11 10 Improve ESU rejection
Rec-GND FE-ADC-Clock
CON7 .047uF SpO2 bypass cap, RS232 PS mods
+5V
U1E G 5-44857 LPP 7/18/02 JK
74HCT04 4KV nurse call relay,Improve CMRR for Resp Topology
A D
TP9
4 4
TP11 D D
+3.3V FE-Data 13 12
FE-Data-In
U2-Reset Serial TP1 TP2 TP3 TP4 U1F
+5V
74HCT04
R20
14
Data 15 LED-On
D
TP10
EKG - Page4
10K
Out
SER QA
1 Power-Sync 3 4
J2 QB FE-Pwr-Sync
11 2
Battery
10
SRCLK QC
3
NIBP - Page 2 U1B
1 SRCLR QD
4 74HCT04
2 Batt-ID QE NIBP-A/D-SelC
12 5 Audio-Shutdown
3 RCLK QF NIBP-A/D-SelB D
13 6 EROM-wr
G QG NIBP-A/D-SelA
7
QH NIBP-Pump-On
CON3
NIBP-ADC-Clock +5V +5V
9
QH NIBP-Safety
C23
SpO2 - Page4
NIBP-15mmHg D
U2 .047uF
NIBP-Overpressure
74HCT595 SPO2-Rx 2 1 13 12
NIBP-Valve-Close SpO2Out
D NIBP-PWM-ADC CRT Board U1A U8F
NIBP-Offset-DAC
74HCT04 74HCT04
Connector D
+12V
D D
C30
J6
+5V
1
Video-VSync .047uF
2 D
Video-HSync +5V
+3.3V
3 Serial IO - Page8
4
Video-Data
5
2 1 +12V
6 RS423Rx

R6 U8A
Main PCA CON6
1.96K 74HCT04
Fan
+5V
C6 + C3
100uF 220pF Connector D
TP5 Connector
3 3
J5
J4 3 4 C39
RS423Tx 1
1 2 .047uF
D D 1 2 2
NIBP-Safety 3 4 U8B
3 4 +5V
+3.3V 5 6 NIBP-PWM-ADC TP17 74HCT04 CON2
5 6
7 8 TP6
7 8
9 10 RS423-Tx
9 10 D
C2 NIBP-OP 11 12 PIC-Data +3.3V C29 5 6
11 12 PIC-Data
.047uF 13 14 RS423-Rx U7 .047uF NurseCall
13 14
ETCO2-Tx 15 16 SPO2-Tx AT24C02N-10SC-2.7 U8C
15 16
ETCO2-Rx 17 18 SPO2-Rx R15 74HCT04
D 17 18 +12V
Spare-2 19 20 ADC-Clock 5 1 1.96K
19 20 SDA A0
NIBP-Data 21 22 FE-Clock 6 2 +5V
NIBP-15mmHg 23
21 22
24 SPI-Data-Out
SCL A1
3
D D
C40 + C41
CO2 Board
23 24 A2
LED-Latch 25 26 SPI-Clock TP18 .047uF 1uF
Vbackup Video-HSync 27
25
27
26
28
28 Batt-ID
WC
7 U7-wc Connector
Spare-3 29 30 FE-Data R23 R24 R8 J3
29 30 +5V +5V
Power-Sync 31 32 Rec-Motor-Step 1.21K 1.21K 1.96K
31 32 1
Vbackup 33 34
33 34 2
NIBP-Valve-Close 35 36 PIC-Clk R16
35 36 D D 3
Video-VSync 37 38 Speaker-Audio 3.48K 11 10 U8-10 9 8 ETCO2-TX-B
37 38 4
C1 NIBP-Offset-DAC 39 40
39 40
.047uF 41 42 NIBP2-PW-ADC U8E U8D CON4
ac-on 41 42
43 44 +3.3V +5V 74HCT04 74HCT04
43 44 D
45 46 Video-Data
D 45 46
Key-Latch-n 47 48
47 48 D D
RS232-CTS 49 50 Nurse-Call 1
Rec-Clock 51
49 50
52 Rec-Data-Latch 3
Front Panel
51 52 To PIC
53 54 Rec-Strobe Key-clock 2
hReset 55
53
55
54
56
56 SCL Display Connector On/Off-Key
SPI-data-in 57 58 SCD U3A J8 +5V
57 58
59 60 74HCT08 CON14AP
59 60
+5V Key-Clock-b 2 1
D + +
Socket 30x2 Key-Latch-n-b 4 3
+ +
Key-Data 6 5 C5 + C4
+ +
LED-Data-b 8 7 .047uF 330uF
D D + +
4 LED-On 10 9
2 + + 2
6 12 11
+ +
5 LED-Latch-b 14 13
+ +

U3B
74HCT08
+5V D D
D
9
8
LED-Data 10

U3C AC+12V-fuse
+5V
74HCT08 Q1
MMBT3906L
D
2 3 ACON-LED
12
11 R1
13 10K

1
Q1-e
U3D
74HCT08
TP19
Speaker Drive D
R2
Recorder Board C35 1.96K TP20 +3.3V
TP22 +5V 4.7uF q2-c
+5V Connector TP13 R3 TP21

3
Audio-Shutdown

+
10K
J7 Q2-b 1 Q2 R21
TP14 TP15 ACON
PIC-Vprog Rec-Supply-On MMBT3904L 10K
1 2 Rec-Supply-On TP25
Rec-Motor-Step
3 4

2
C42 R4
5 6 +12V

1
6
.047uF Rec-Data-Latch Rec-Strobe C36 J11 10K
7 8 ac-on
Rec-Clock Rec-Data 4.7uF R17 3 5 SP
Vbackup 9 10 + 1

3
C14 19.6K
11 12 2
.047uF R17-1 4 8 SM Page1
D Rec+DCV 13 14 Rec+DCV - 3
+

C20 Q3
15 16
R22 .047uF U9 CON3 1 2N7002
14

17 18 Print-Temp

2
7
1

10K U10 TPA301 4Vpp max.


1 19 20 D TP23 1
R18 U9-3
21 22

2
Vdd

Vss

(pg 1) 1.21K
Rec-GND 23 24 Rec-GND
2 13 C37
PIC-Data ClkIn/Osc1/RB5 RB0 D
CON24A 0.1uF
U9-4

3 12
On/Off-Key Clkout/Osc2/RB4 RB1
R25
D D
(Front Panel Key) 4 11 1.21K R19
Vpp/MCLR/RB3 RB2 PIC-Clk
19.6K Initial Date
Welch Allyn Inc.
Rec+DCV
TP12 5 10 TP24
TC5/TOCK1 RC0 TP26 Drawn Schematic: 620017.dsn
6 9 PIC-audio Rick Myers Mar 5, 1999
R5 + C7 C8
On/Off RC4 RC1
19.6K 330uF .047uF Designed
(pg 3) Rick Myers Mar 5, 1999
7 8 U2-Clk Title
U2-Reset RC3 RC2
C38 Checked
Main Board - Model 220
(To U2) L. Phillips 3/16/99
220pF
D620361-MX10000 Approved Size Document Number Rev
Rec-GND Z. Psenicnik 3/16/99
TP16 C 620017 G
Release For Production
J. Bello 3/16/99
Date: Monday, August 05, 2002 Sheet 1 of 8
A B C D E

Figure D-20. Model 220 main board.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 166
A B C D E

REV Description ECN/ECO Init Date Ckd


L201 TP201
220uH
TP202 R207 A Release to Production 5-39121 JAC 3/15/99 LPP
+12V P+12V +3.3V
464K

C229 + B See ECN worksheet 5-39455 LPP 4/27/99 LPP


10uF C205
Improvements for ESD, RFI,
P+12V
R208 D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
P+5V
.047uF 1.96K Respiration Diff-Amp Topology
A
R206 Mods to U2 clock line, Resp
A

8
TP206 1K E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
U201 R201 U203-3 3 NVRAM pullup
+ 1 Buffer for ETCO2, Nurse call relay
78L05 576
NIBP-Overpressure
8 1 U203-2 2 F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
IN OUT -
U203A Improve ESU rejection
C203 C222 LM393 SpO2 bypass cap, RS232 PS mods

GND

GND

GND

GND

4
R202 .01uF G 5-44857 LPP 7/18/02 JK
P+12V 4KV nurse call relay,Improve CMRR for Resp Topology
C201 + C202 174
4 .047uF 4.7uF .047uF 4
TP207

7
TP208 TP209 R210 +3.3V
A A A

8
464K
A
U202-3 3
A + 1 P+12V
2
- P+12V P+5V P+5V
R203 R211 TP239
A
3.32K U202A 1.96K TP203

8
MC34072 R209

8
1K 5
U203-5 5 + 7 u207-9 9
A +
TP210 7 6 8
NIBP-15mmHg -
U203-6 6 10
-
U203B U208B

4
TP240 P+5V C223 LM393 MC34072
Over Pressure

4
.01uF U207C
P+12V
R204 74HCT00
TP211 150
P+.75V
Fault Circuit A
A

TP212 U208-7 P+5V


A A

8
C226 L202
.047uF Bead U202-5 5
+ 7
6 +12V +5V +Vpump 12
A -
C204 R205 11
P201-pwr .047uF 750 U202B 13

4
MC34072

7
8

5
6
R242
C232 10K TP213 Q202 R239 U207D
A
.047uF MMDF2N02E 1K 74HCT00
Primary R236
A A A

3.83K TP214 TP215 Q202-G 2 4


Transducer A
3

R212 R240

3
56.2K 10K
A

3
0mmHG = 0.5V 2 P201-Out Q201-B 1 Q201 R243
NIBP-Safety
MMBT3904L 100K
Gain
300mmHG = 4.0V

2
and C206 TP216 R241
3 6 P201-6 .047uF P+12V C209 10K 3
Comp
.047uF
C227
680pF
P201
A
1

2
XFPM-050KPGR-P1 TP217
A

8
R219 D202
U204-3 3 1.96K BAV99L
A A + 1 U204-1 3
2 J201
- Valve+
1
+3.3V Valve-
Zero Adjust U204A
2

D201-A
4
R213 MC34072 C228
3

1
+/-20mmHg range P+12V 237K .047uF Pump+
4
Pump-
5
R244 R217
A
21.5K TP219 TP220 5.62K CON5
A
R214 R216 U204-2 R246
TP204
8

10K 19.6K TP221 TP222 31.6


C207-1 U204-5 5 C231
NIBP-Offset-DAC + 7 .01uF
NIBP-A/D-SelC
PWM D/A 6
-
R215
NIBP-A/D-SelB
76.8KHz 100K + C207 C208 + U204B
4

7
8

5
6
1uF 1uF MC34072 R218
NIBP-A/D-SelA
23.7K Q203
MMDF2N02E

A A A A A/D Mux
U204-7 2 4
NIBP-Valve-Close NIBP-Pump-On

P+.75V

3
R247 R248
A
9 1K 1K
C
10
B
11
A
6 U209
2 INH 2
4051
4
BattCurrent X7
P+12V 2
BattVoltage X6
P+12V C216 5
Safety .047uF TP224
Print-Temp
P4.25V 1
X5
X4
C211 P.75V 12
Transducer PrimaryPres 15
X3 TP225
R235
NIBP
X2
SafetyPres 14 3.48K
.047uF
A
TP226 +5V/2 13
X1
X0 X
3 U209-Out U210-2 PWM A/D
8

R227 R228
A
Converter
8

P202-Out+ 3 5.11K TP227


+ U206-1 +5V
U205-3 3 1 C230
+ 1 P202-2 2 1.96K .047uF
2 -
R237 TP228
- P+5V
U206A R223 C214 C217 C210 R229 464K
4

U205A MC34072 21.5K .047uF .047uF 1.96K


4

.047uF
MC34072
A
P202 TP229
2

FPNS-07PGR P+5V C225


A A A A
C212 U206-2 .047uF
A
680pF P+5V TP231
1 R230 TP230 P+12V
P+12V
3 R224 10K C221
+ - 6 U208-3 P+5V
1K .047uF

5
4
TP232 A
8

2
P202-Out- 5 + 7 U210-7 4
R231 C219 C224
+ 7 U206-7 3 6
73.2K .047uF 1000pF
A - NIBP-PWM-ADC
5

8
P202-5 6 R234 U210 5
- 3 3.48K MAX941

6
+ 1 U208-1 U210-3
U206B R225
4

R220 MC34072 1K 2 U207B


A A -
806 C218 + 74HCT00
TP233 4.7uF U208A
A A

4
R232 MC34072 TP235
A
5VRef U206-6 U207A 10K
74HCT00 U208-2
1 A A 1
C220
A
P+12V 1 PWM A/D
3 TP236
R221 C215 2 R233 .01uF
Converter
Zero R226
NIBP-ADC-Clock
3

3.16K TP237 21.5K .047uF 3.48K


Offset D201-C 2 1
8

5
TP238
D201 Initial Date
Welch Allyn Inc.
+ A
7 U205-7 MMBD1503A
6 U207-3 Drawn Schematic: 620017.dsn
- Rick Myers Mar 5, 1999
TP234
C213 R222 U205B Designed
4

Rick Myers Mar 5, 1999 Title


.01uF 1K MC34072
Checked
NIBP ELECTRONICS
L. Phillips 3/16/99

Approved Size Document Number Rev


A A A Z. Psenicnik 3/16/99
C 620017 G
Release For Production
J. Bello 3/16/99
Date: Monday, August 05, 2002 Sheet 2 of 8
A B C D E

Figure D-21. 210/220 NIBP electronics.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 167
A B C D E

Q303 REV Description ECN/ECO Init Date Ckd

Rec+DCV MTP30N06VL
4 A Release to Production
2 3 5-39121 JAC 3/15/99 LPP
Q303-s
L303
U307 C335 Bead B See ECN worksheet 5-39455 LPP 4/27/99 LPP
On/Off Switch

1
8 1 .047uF
VIN VOUT VBackup
Improvements for ESD, RFI,
7 5 +12V Boost

L302-1
u307-6 FDBK ERR D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
3
C331 SHTDN C332
+
4.7uF
+
4.7uF
(12.1V nominal) Respiration Diff-Amp Topology
6 Mods to U2 clock line, Resp
TAP E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
2
SENS NVRAM pullup
V+18V 3 1 U302-pwr Buffer for ETCO2, Nurse call relay
LP2951CD-3.3 R308 TP302 D303 F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
D D
17.8K 5.1V + C306 + C310 L302 Improve ESU rejection
1uF 1000uF 22uH SpO2 bypass cap, RS232 PS mods
D
U302-fb TP304 G 4KV nurse call relay,
4 5-44857 LPP 7/18/02 JK 4
Improve CMR for Resp Topology
+Vfuse On/Off Circuit R309 TP305 R310 D304
4.64K 10K MBRS340T3 L305
18V Boost U301-7
- Mosfet Gate Voltage Bead

C305-1 Q305-d D304-C +12V


HS301 U302

11
12
L301 C305 UC3843A
Heatsink F303 1mH 5600pF + C311 + C312 + C313
3
R305 U301 VFB 1000uF 1000uF C336 330uF

Vc
Vcc

4
2
for D301 4Amp
1 MC34063AD TP307 TP306 R314 Q305 .047uF
1
U302-comp

U302-out
6078 D302 COMP 10

Q305-g
MTP30N06VL
BAV99L 14 10
AC+12V 8 U301-sw 5VRef 1
Q301 IDC VREF OUT
7
MPSW51A IPK TP308
1 1 1 2
1 3 ISWC C307

3
PWR-GND
2 U301-pwr 6
Vcc .047uF R311 7 5
Battery 3 4 2 U302-ct U302-sense

GND
ISWE R306 10K RT/CT ISENSE

3
R301 TP309 61.9K

2
D301 C301 10K C302 R315

GND
+ +

Q305-s
U301-ct 3 5 U301-comp

9
8
MBR2535CTL 330uF 1uF TCAP COMP 196
Q301-b

1
R312
C304 C308 3.83K

4
+
TP310 TP311
C303 R307 100uF 2200pF 3 2
Q304-e
R302 470pf 4.64K
TP314 1.96K Q304 R313 C309
MMBT3904L 1.96K 5600pF R316
TP312
TP313 0.1
Q302-c
R303 R304
3

51.1K 100K D312


C339-1

1 Q302 BAV99L R317


On/Off Q302-B
MMBT3904L 7.5K
1 2 TP315
AC+12V-fuse
2

D312-c
+ C339
TP354
1uF

3
TP316

3 3

F301
+Vpump
4Amp D310
MBRS340T3 HS302
Bat-fuse
Heatsink
Q312
forQ312
MTP30N06VL
6078
4
Battery 2 3 +5V
TP349 Charger Q308
4
3 2
L304 TP353 D307 + C323 + C325
MTP23P06V
F302 47uH R327 MBRS340T3 Q311 100uF 100uF

1
Q312-g
4Amp 0.1 MTP30N06VL

1
3 2 R346 U306
F302-2 AC+12V-fuse Q308-d C319-1 D307-A
4 100 R348 LF33CV
L306 10K
1 3 +3.3V
C340 Bead C337 + C318 C324 R347 IN OUT
.047uF .047uF 330uF C319 + C338 1000pF 10K
1

GND
D306 1000uF .047uF TP317 C328 + C329 C330
Q308-g C324-2
MBRS340T3 TP318 .047uF 100uF .047uF

2
R326 TP319
Q307 10 R328 R329
MMBT3904L 1K 1K
D D D
3 2 TP320 +5V Linear

U305-1

U305-2
Q307-e D

4
TP324 TP323
1

R318 D305 TP321 TP322 Low -


2 R349
10K C314 BAV99L R331 10K
Dropout 1
1

.047uF +
3
U303-fb U305-6 U305-3 5VRef
R325 3 U305A
1.96K 33.2K MC34072

8
C326 R350
2 .01uF 2
10K
R319 U303 R330
11
12

19.6K UC3843A TP326 33.2K D311


TP351 4
3 6
- BAV70L
VFB
Vc
Vcc

Q306-c 7 1 C327
+12V
U303-cp 1 U305-5 5
+ .047uF
U303-out

COMP
3
Q306-b

R323 1.96K U305B 3


14 10 1 Q306 MC34072
VREF OUT
8

MMBT3904L 2
AC+12V-fuse
TP328
2

C315 R320 TP330 C320 R332


PWR-GND

.047uF 7.5K 7 5 U303-5 TP329 .01uF 19.6K U305-pwr


GND

RT/CT ISENSE
U304-2 TP331
Ref-Batcharge
R324
10K
Current Limit
9
8

TP332 Q313
(1.5 Amp) TP334
U304-1

TP333 R351
2N7002
4

C316 R321 D308 BAV99L 5.11K


2200pF 27.4K - 2 TP336 R333 TP337 Q313-d 3 2
BattVoltage
U303-ct TP335 2 1 1 19.6K
+ 3 U304-3 I-Batt
U304A R352 C333
TP338
3

R322 MC34072 C322 5.11K .047uF

1
8

10K .047uF
ACON +12V +12V

1
R341
10K R353 TP340
Ref-Batcharge
AC+12V-fuse TP339 5.11K
R342 R343 Q314-d 3 2
BattCurrent
2

Batt-fb TP341 TP342 R339 C321 R334 19.6K 10K


1
D309-a

19.6K .01uF 19.6K Q309


Q309-b

R339-2 MMBT3906L
Ref-Batcharge TP343
R343-1 Q314 R354 R355
3

TP345 R335 TP346 5.11K 14.7K C334


4.32K TP355
2N7002
.047uF
Ibatt=0: BattCurr=0.74V
TP344
4

2
1

R340
10K - 6 U304-6 TP352
TP347 Ibatt=1.5: BattCurr=2.87V
1 7 D309 1
U304-7
+ 5 U304-5 Q309-3 BAV70L +5V
U304B R336 TP348 RT301
MC34072 7.5K 10K
8

R337 Q310-c
TP350
RT301-1

8.87K R344
Float Voltage Welch Allyn Inc.
3

10K Initial Date


AC+12V-fuse 1 Q310
t

ACON
Bat-Adj MMBT3904L Drawn Schematic: 620017.dsn
Q310-b

Rick Myers Mar 5, 1999


1

R338 Designed
2 Rick Myers Mar 5, 1999 Title
1K
R345 Checked
DC/DC Power Supplies
L. Phillips 3/16/99
10K
Approved Size Document Number Rev
Z. Psenicnik 3/16/99
C 620017 G
3

Release For Production


J. Bello 3/16/99 Date: Monday, November 04, 2002 Sheet of
3 8

A B C D E

Figure D-22. DC/DC power supply.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 168
A B C D E

REV Description ECN/ECO Init Date Ckd

U407 A Release to Production 5-39121 JAC 3/15/99 LPP


78L05
1 8
iso+5Vdig OUT IN
B See ECN worksheet 5-39455 LPP 4/27/99 LPP

GND

GND

GND

GND
C406 + Improvements for ESD, RFI,
100uF C403 C430 D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
.047uF .047uF Respiration Diff-Amp Topology

2
Mods to U2 clock line, Resp
E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
S S S
NVRAM pullup
Buffer for ETCO2, Nurse call relay
F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
S
Improve ESU rejection
SpO2 bypass cap, RS232 PS mods
1 8 G 4KV nurse call relay,
s+5V VOUT VIN
5 7 Improve CMR for Resp Toplogy 5-44857 LPP 7/18/02 JK
4 ERR FDBK 4
3
SHTDN

U403-fb
C415 + Isolation
1uF U403 6 C401
TAP
LP2951C
SENS
2 .047uF Barrier

S S
+12V
Isolated SpO2 Grounded SpO2
S S
Circuits Circuits

1 2 D401
s-5V OUT IN
3 MBRS130T3 + C412
IN
U404 6 10uF
IN

GND
C421 79L05 7 f+V T401-1
+ IN
C424 C409
1uF
.047uF 470pf

5
C405 + R419 T401 +12V
C402 100uF 215 C409-1 TP449
S S
.047uF 1

S
TP450 5 R405
S S
215
TP452 L401 R419-2 2 C414 U402-ref TP451
220uH C433 .047uF
470pf 8 T401-8 TP466

15
fVcc TP453 U402 C420
3 D405 16 .047uF
VREF

4
2

VIN
+ C431 + C434 BAV99L
10uF 10uF R407 TP454 10
SHTDWN FE-Pwr-Sync

Q401-g

Q402-e

3
Q401 10 12
D E E S CA
4 MTD3055V 1 2 1 U402-e 11 3
EA OSCOUT
13 6 U402-rt TP455
S S CB RT
D403 14 7 U402-ct
EB CT

2
IsoXfmr TP456 R409 TP457
f-V T401-4 Q402 1 5.11K 2
V+
MMBT3906L U402-sen 4 1
+SENSE V-

GND

U402-1
3 R406 5 9 3
-SENSE COMP

3
C407 + MBRS130T3 0.2 C411 R410
TP468 L402 C404 100uF R408 TP458 C410 LM3524DM 2200pF 7.5K

8
220uH .047uF 100 .01uF
Q401-s
fVee
S S

+ C432 + C435
10uF 10uF

f+V

S S
R422 +5V
10K
TP459 U410

8
R403 4N25
U401-3 3 1K R411
+ 1 U401-1 iso401-a 1 6 iso401-b 5.11K
U401-2 2
f+V -
R423 U401A 5 opto-fb
TP460 R416 5.11K LM358

4
8

TP461 R415 10K 4


100 5 U401-5 TP462 2
+ s+5V
U401-7 7 R404
f+5V
6 196K
- S S
U401B + C418
C416 LM358 4.7uF R401 C408 R402
S
4

.047uF 10K .01uf 10K


C408-2
s+5V

C417
S S S
TP465 680pF
TP463 TP464
R417
10K

2 2
S

Nonin SpO2
Connector

J401
SpO2 ESD
J403 Ground
1 f+V iso+5Vdig
2 1
sSpO2tx
3 2
+ C419 C413
4
10uF Tab-.187 .047uF
+5V
CON4
S
S S D E
3

Q403
sSpO2tx 1 MMBT3904L C422 R414
.047uF 1.96K
R412 U411
2

511 HCNW4503 TP469


Nellcor SpO2 Q403-e iso402-a 2 8
Connector R424
10K
6 U411-6
J402 TP467 SpO2Out
3
1 2 5
+ +
3 4 C436
+ + SpO2-Reset
5 6 1000pF
+ +
7 8
s-5V + + D E
9 10
+ + s+5V
11 12
+ +
13 14
1 iso+5Vdig + + 1
+ C437
C429 330uF
CON14AP .047uF

S S

S S
C427 C428
.047uF
+
10uF Initial Date
Welch Allyn Inc.
Drawn Schematic: 620017.dsn
Rick Myers Mar 5, 1999
S S
Designed
Rick Myers Mar 5, 1999 Title
Checked
Power Isolation and SpO2 Interface
L. Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620017 G
Release For Production
J. Bello 3/16/99
Date: Monday, August 05, 2002 Sheet 4 of 8
A B C D E

Figure D-23. Power isolation/SpO2 interface.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 169


A B C D E

REV Description ECN/ECO Init Date Ckd

Defib Protect A Release to Production 5-39121 JAC 3/15/99 LPP

Clamp voltage for Temp U501 U502 U503 U504 U505 U506 U507 U508
B See ECN worksheet 5-39455 LPP 4/27/99 LPP
and Resp circuits fVcc

Improvements for ESD, RFI,


TP561 + C534 C518 C520 C522 C524 C526 C528 C530 C532 D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
+clamp
R506 D506 4.7uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF Respiration Diff-Amp Topology
10K 5.1V Mods to U2 clock line, Resp
+clamp 3 1 E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
fVcc NVRAM pullup
E E E E E E E E E
Buffer for ETCO2, Nurse call relay
TP502 F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
fVee
R507 Improve ESU rejection
10K SpO2 bypass cap, RS232 PS mods
-clamp 1 3 C535 C519 C521 C523 C525 C527 C529 C531 C533 G 4KV nurse call relay,
fVee E +
4.7uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF .047uF Improve CMR for Resp Toplogy 5-44857 LPP 7/18/02 JK
4 D507 4
5.1V
-clamp
E E E E E E E E E

RA

Input Buffers LA A/D Inputs


Right-Arm ESU filter
Gain = 9.26 Leads Off
LL
Left-Arm
fVcc Lead Select
V+
J501 TP501 TP503
CON6 R501 R561 R508 TP504 TP505 fVcc

8
19.6K 73.2K 10K
Right-Arm C501-1 C537-1 U501-3 3 TP506
1 + fVcc
1 U505
2

8
2 Wilson 4051
3 -
R509 13 3 mux- 3
4 X0 X +
N501 C501 C537 C506 TP507 22Meg U501A R517 Network RA 14 1 diff-
5 X1

4
330pF 220pF 100pF TL072 10K C510 TP508 LA 15 2 C544
6 X2 -
LAMP NEON TP509 0.1% 220pF R529 10K aVR 12 Diff Amp 1000pF
X3
R509-1 aVL 1 U507A
X4

4
fVee aVF 5 TL072 Gain = 1
S E E E X5
R557 U501-2 R530 10K V- 2
E X6
Shield 22Meg 4 R544 R546
X7
fVee 10K 0.1% 10K 0.1%
R518 TP510 6 U508-3
INH
fVcc 1.21K R531 10K 11
A
VLdsOff 0.1% 10
E B
TP534 TP511 9
C
R502 R562 R510 TP512 TP513 R532 10K C543

8
19.6K 73.2K 10K TP514 fVcc TP557 TP558 .047uF
E
Left-Arm C502-1 C538-1 U501-5 5 fVcc u509-4
+ fVee
7 TP515
3 E 3
6 R533 10K TP516 fVcc
- fVcc

8
R511 U506 R565

2
N502 C502 C538 C507 22Meg U501B R519 4051 R545 3 121K

4
LA 13 3 mux+ 5 + 1 4
330pF 220pF 100pF TP519 TL072 10K C511 R534 10K 10K 0.1%
X0 X + -
LAMP NEON TP520 0.1% 220pF LL 14 7 diff+ U508-2 2 1 diff-out
X1 -
R511-1 LL 15 6 U508A 3
X2 - +
fVee TP521 RA 12 AD712 U509
S E E E X3

4
U501-6 LA 1 TP522

U508-1
R558 R535 10K U507B LMC7101
X4

5
22Meg LL 5 TL072
X5 E
V+ 2 C542 fVee
X6
R520 R536 10K 4 fVee 220pF
X7
fVcc 1.21K fVee
VLdsOff 0.1% 6
INH
TP541 TP523 R537 10K 11
A
R503 R563 R512 TP524 TP525 10 R547
B
8

19.6K 73.2K 10K 9 10K C545


E E C
Left-Leg C503-1 C539-1 U502-3 3 0.1% 1000pF
+
High Pass (0.05Hz or 0.5Hz) Gain Stage
1 fVcc
2 Low Pass (150Hz) Gain = 74
- fVee
R513 TP517
E E E
N503 C503 C539 C508 22Meg U502A R521 TP528 TP529 C515 R550 TP518
4

8
330pF 220pF 100pF TP531 TL072 10K C512 3.3uF 19.6K
LAMP NEON TP533 0.1% 220pF ecg-hp U508-5 5
R513-1 + 7
An-ECG
fVee 6
S E E E -
R559 U502-2 C516
22Meg R548 R549 .047uF U508B

4
TP535 TP536 TP537 1Meg 110K AD712 R551 C517
R522 73.2K .01uF
fVcc 1.21K
E
VLdsOff 0.1% Q504-d fVee
Mux-LdS-A
TP544 TP538 U508-6
E
R504 R564 R514 TP539 TP540
Mux-LdS-B
8

19.6K 73.2K 10K TP526 TP527


E
V-Lead C504-1 C540-1 U502-5 5 R552
+ Mux-LdS-C
7 1K
6
2 - 2

3
R515 TP530
N504 C504 C540 C509 22Meg U502B R523
4

330pF 220pF 100pF TP542 TL072 10K C513


E
LAMP NEON 0.1% 220pF TP532 Q504 1
R515-1 TP543 2N7002
fVee
S E E E

2
R560 U502-6 Q504-s R553

2
22Meg 100K
1 Q504-g
Sw-MonBw
R524 Q505
1.21K 2N7002
VLdsOff 0.1%
fVcc

3
U504
3 2 TP547
S1 D1 E E
1 TP548 U503B R555
Sw-RLD-RA IN1
8

TL072 30.1K
14 15 5 U503-5
S2 D2 + f+5V
16 7
Sw-RLD-LL IN2 VLdsOff
6
TP550 11 10 -
TP545
S3 D3
R505 R566 9 C536 R556
Sw-RLD-LA IN3 f+1.24V
4

19.6K 73.2K .01uF


Right-Leg C505-1 C541-1 6 7 10K
S4 D4
8
Sw-RLD-RL IN4
fVee
13
fVcc V+ E E
N505 C505 C541 4
fVee V-
330pF 220pF 5
GND
LAMP NEON
DG202CSE
S E E E

fVcc TP549 fVcc

TP559
2

Q501 Q502 U503A R525


1 vRL 1
8

NL NL R516 TL072 10K


Q501-S

D505 10K 3
3 3 2 2 3 RL-mux 1 +
MMBD1503A
2 R526
- E
V-Lead Defib Protect, 10K
U503-2
TP562
4

Diodes internal to U504 RLD Amp Welch Allyn Inc.


1

R567 Initial Date


protect other inputs.
NL
fVee Q501-G TP560 R527 fVee C514 Drawn Schematic: 620017.dsn
Sw-RLD-V Rick Myers Mar 5, 1999
1K 0.1uF
C514-2 Designed
Rick Myers Mar 5, 1999
C546 No Load, Title
NL
Allows RLD switch TP551
R528
51.1K
TP552 Checked
L. Phillips 3/16/99 ECG Amp
Approved Size Document Number Rev
to V-Lead, not used. Z. Psenicnik 3/16/99
C 620017 G
E
Release For Production
J. Bello 3/16/99
Date: Monday, August 05, 2002 Sheet 5 of 8
A B C D E

Figure D-24. 210/220 ECG amp.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 170


A B C D E

REV Description ECN/ECO Init Date Ckd


U601 U602 U603 U608 U609
5-39121 JAC 3/15/99 LPP
A Release to Production
fVcc

fVcc B See ECN worksheet 5-39455 LPP 4/27/99 LPP


C622 + C609 C623 C624 C637 C639
.047uF 4.7uF .047uF .047uF .047uF .047uF Improvements for ESD, RFI,
R623 D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
46.4K fVcc Respiration Diff-Amp Topology
TP604 Mods to U2 clock line, Resp
E E E E E E
U605 U606

8
TP603 E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
5 NVRAM pullup
U601 U602 U603 U608 U609
7 + iso+5Vdig
Buffer for ETCO2, Nurse call relay
Resp-Reset 6 Resp-Off
F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
fVee -
U607B Improve ESU rejection
SpO2-Reset
LM393 C629 C631 SpO2 bypass cap, RS232 PS mods

4
TP601
C625 + C614 C626 C627 C638 C640 .047uF .047uF G 4KV nurse call relay,
.047uF 4.7uF .047uF .047uF .047uF .047uF Sw-RespReset 5-44857 LPP 7/18/02 JK
fVee Improve CMR for Resp Toplogy
fVcc iso+5Vdig
4 D E D E 4
Isolation
E E E E E E fVcc iso+5Vdig
Serial/Parallel Barrier
R622 TP605 C641 +5V

8
46.4K f+5V TP639 R613 0.1uF
3
Control Data
6.19K 2
+ 8
1 TP638 U604A
Sw-MonBw 2

U604-1
U605 74HCT04 R628
- D E
U607A 74HCT595 U604-2 2 1 10K U613-6 R619
TP606 15 14 6
LM393 QA SER iso604-a 619

4
TP609 1 3
QB 5
2 11 FE-Data-In
QC SRCLK
fVee 3 10
MuxB QD SRCLR TP636 C643 U613
4
MonBw QE 220pF HCNW4503
5 12
QF RCLK
fVcc 6 13 E D E
QG G
SW-RLD-RL 7 D E
QH
SW-RLD-LL

8
SW-RL-V
5 9
QH
7 +
SW-RLD-V
6 iso+5Vdig
- f+1.24V TP610 TP611 TP607 TP612
AD712
U608B U605-out D E

4
C642 U612 +5V
fVee TP613 TP614 TP640 0.1uF HCNW4503
U606 f+5V R615
8 2
15 74HCT595 14 U604B 6.19K
QA SER

U604-3
1 74HCT04 R629
TP616 TP617 TP618 SW-RLD-LA D E
QB 10K R620
SW-RLD-RA 2 11 U604-4 U612-6 6
C630 QC SRCLK 4 3 619
3 10 iso603-a
Mux-LdS-C SRCLR 3
f+5V 4.7uF QD
Mux-LdS-B 4 5 FE-Data-Clock
QE
Mux-LdS-A Mux-A/D-C 5 12
QF RCLK

+
fVcc Mux-A/D-B 6 13 C644
f+5V G
QG TP637 220pF
Mux-A/D-A 7
QH
TP615 E TP635 E D E
9 D E
8

3 fVee QH 3
13 12 U605-12
5
Resp-DC + 7
9
6 C C646
- 10
U604F B 1000pF iso+5Vdig
11 D E
U601B 74HCT04 A
6
4

E
MC34072 INH
Resp-LdsOff U609
4 E
X7 4051 R630 C628 U611 +5V
fVee
2
X6
MuxB 10K TP641 0.1uF HCNW4503
U601-7 Temperature An-Resp 5
X5 f+5V R617
1 8 2
X4 U604C 2.15K
Amplifier 12
X3 74HCT04 R631

U604-5
15 D E
fVcc X2 TP642 5.11K R621
14 U604-6 6 5 U611-6 6
X1 iso602-a 215
13
X0 X
3 A/D Mux 3
TP620 5 FE-ADC-Clock
R624
8

51.1K R626 TP621


U608-3 3 1K C645
f+1.24V + 1
U608-1 Patient-Temp
fVcc 100pF
U608-2 2 E D E
C633 fVee
-
.01uF
D E
U608A
4

AD712 9
C
10
E B
11
MuxA A
fVee 6
TP624 INH
TP623 C634 .01uF U603
4 4051
vRL X7 MuxA
2
f+1.24V X6
5
RA X5
R625 1
LA X4
-Clamp +Clamp 1K 12
LL X3

U603out
15 R605 R607

U602-2
V+ X2
D602 R627 1.21K 14 3.48K 464K
An-ECG X1
BAV99L 13 3
X0 X
1 2 iso+5Vdig
2 0.1% 2
iso+5Vdig
TP626 fVcc

U602-5
3

E TP625
C635 .01uF fVcc
R601 TP627 fVcc f+5V

U602-7
J601 10K R608 C632
Temp-In U601-3
2 C603 1K .047uF

5
6
8
f+5V +5V
1 680pF TP628

U604-8

3
Temperature
8

2
C636 R606
3 + 7 9 8 1
CON2 2200pF R602 3.48K Q601
Thermistor + 1 U601-1 U602-3 3 E
73.2K MMBT3904L
2 -
U602

2
-
LM311 U604D C621

4
1
TP629
f+5V U601A 74HCT04
4

TP631 TP630 0.1uF


R603 MC34072 R610 U610 R611

Q601-e
E E E
46.4K 511 HCNW4503 6.19K
U601-2 iso601-a 2 8 TP643
C601 fVee R632
PWM fVee 5.11K
U604-10 E U610-6
11 10 6
TP632 3 FE-PWM-ADC
R604 1000pF A/D
3

10K Converter 5
D601-2 2 1
U604E
74HCT04 C647
D E
100pF
E
D601
TP633
MMBD1503A

52uSec
Integrator
Input
833uSec
(U604-10)
1 1

+5V
Comparator
Inputs

Analog In

Initial Date
Welch Allyn Inc.
-5V

Drawn Schematic: 620017.dsn


Rick Myers Mar 5, 1999

PWM-Out Designed
Rick Myers Mar 5, 1999
(U604-8) Title
Checked
ECG A/D and Interface
L. Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620017 G
PWM-timer Release For Production
J. Bello 3/16/99
Date: Monday, November 04, 2002 Sheet 6 of 8
A B C D E

Figure D-25. 210/220 ECG A/D interface.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 171


A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39121 JAC 3/15/99 LPP

B See ECN worksheet 5-39455 LPP 4/27/99 LPP

U702 U703 Improvements for ESD, RFI,


D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
fVcc
Respiration Diff-Amp Topology
Mods to U2 clock line, Resp
C715 C716 fVcc E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
.047uF .047uF f+5V NVRAM pullup
Buffer for ETCO2, Nurse call relay
F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK

8
E E Improve ESU rejection
3 SpO2 bypass cap, RS232 PS
+ mods
1 u703-1 G 5-44857 LPP 7/18/02 JK

10
fVee 4KV nurse call relay,Improve
2 CMRR for Resp Topology
4 12 9 - 4
U703A
D Q E

PR
C717 C718 AD712

4
.047uF .047uF 11
CLK

CL
Q
fVee
E E E

13
U701B
74HC74

Isolation
Respiration
Barrier
Drive
Circuit
f+5V

-Clamp +Clamp

+ C719 iso+5Vdig
4.7uF iso+5Vdig
D701
TP701 MMBD1503A
C705
E
1 2 0.1uF +5V
R701 C701 TP702 R703 TP703 U701A R705
TP723

4
31.6K 1000pF 1.96K 74HC74 2.15K U710
3

C701-2 D701-3 RespDr+ 5 2 R725 HCNW4503


Right-Arm Q D E

PR
2.15K
TP704 3 resp-clk u710-6 8 2
CLK
C723
3 22pF 6 R706 3

CL
Q
6 215
D702 C721 3 iso701-a
E FE-Pwr-Sync

1
MMBD1503A 100pF 5
TP706
1 2
TP705 E D E D E
R702 C702 R704
Resp-Off
31.6K 1000pF 1.96K
3

C702-2 D702-3 RespDr-


Left-Arm

C724 R724 TP707


22pF 196K
d705-c
Resp-DC
E

C720
.047uF

D703
E
MMBD1503A Respiration
fVcc
1 2 Diff Amp
C706 TP710 R707 TP708 D705 fVcc
B.W. = .06Hz to 16 Hz Gain =1000
TP709
8

330pF 1.96K R726 TP727 R727 TP725 MMBD354LT1


3

c706.1 U702-3 3 3.16K 3.16K


+ TP712
B.W. =
1 U702-1 D705-1 1 C712 R720 10Hz TP713

8
2 3.3uF 100K
- TP726 3 Resp-AC U703-5 5
U702A
+ 7
TL072 R711
An-Resp
4

R709 178K 2 6
-
464K U703B
fVee
R717 R718 R719 C713 AD712

4
U702-2 fVee
3.16K C711 825K 51.1K 0.1uF

D705-2
0.1uF
E
U704-2

R715 TP724 R716 R721


2 R712 2 - 3.16K 3.16K Q701-d fVee 316K 2
TP714 E
19.6K 1 U703-6
E E E
3 + TP718 TP717
TP711 c722-1 U704A
fVee
U704-3

2
AD712 R722
8

C722 Q701 3 TP719 316


U704-6

4
0 MMBF4393L C714
fVcc 6 - .047uF

1
U702-6 TP722 7
D704 5 +
E E

Q701-g
MMBD1503A
R713 U704B
8

1 2 fVee
178K AD712 D706
E
MMBD1503A
fVcc
4

R728
3

C707 TP721 R708 TP720 6 - 3.16K 1 2


Resp-Reset
330pF 1.96K 7 U702-7 TP728
c707.1 U702-5 5 +

3
U702B
TL072 R729 R723
TP715
8

3.16K 1Meg

fVcc
R710
464K

E E

1 1

Initial Date
Welch Allyn Inc.
Drawn Schematic: 620017.dsn
Rick Myers Mar 5, 1999

Designed
Rick Myers Mar 5, 1999
Title
Checked
Respiration
L. Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620017 G
Release For Production
J. Bello 3/16/99 Date: Monday, August 05, 2002 Sheet 7 of 8
A B C D E

Figure D-26. 210/220 Respiration circuit.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 172
A B C D E

REV Description ECN/ECO Init Date Ckd

A Release to Production 5-39121 JAC 3/15/99 LPP

B See ECN worksheet 5-39455 LPP 4/27/99 LPP


Isolation
Barrier Improvements for ESD, RFI,
D and Fast Transients, Correct 5-39821 LPP 6/24/99 LPP
Respiration Diff-Amp Topology
Mods to U2 clock line, Resp
E gain amp, ECG supply and 5-39936 LPP 7/9/99 LPP
NVRAM pullup
D801 Buffer for ETCO2, Nurse call relay
BAV99L F Modify RLD to reduce V-Lead noise 5-44461 LPP 3/26/02 JK
Improve ESU rejection
T801-1 1 2 SpO2 bypass cap, RS232 PS mods
RVcc
G 4KV nurse call relay,
Improve CMR for Resp Toplogy 5-44857 LPP 7/18/02 JK

3
4 R812 C803 4
Isolated Supply +
215 100uF
T801
Serial 1 TP802
Communication

R812-1
D802
T801-5 R
5 BAV99L

2 T801-2 1 2
R+5V
R816
215 8

3
C813 + C804
3 470pf 100uF
U801 TP801 R815-1
MC34063AD
+5V R801
R
1 8 C807 4
IDC R
7 470pf D803
IPK
1 U801-1 BAV99L
ISWC
6 IsoXfmr
Vcc
2 T801-4 2 1
TP803 ISWE RVee

C801 +

GND

3
1uF 3 5 U801-5 C805
TCAP COMP +
100uF

U801-ct

4
R804
+5V U805 511
U805-6 R
C802 6 1 U805-1
220pf
5
R803 C806 R805

R803-2
4 10K .01uF 3.16K
2 U805-2

R802 R821 4N25

1
5.11K 1Meg
TP805
8 ISO801-4 TP804
3 3
U806
TL431/SO

2
3
6
7
R806
3.16K

+Vpump

TP810 R

R813
3

100
q802-b 1 Q802 K801
NurseCall
MMBT3904L relay-JQ1-5V

4
2

q802-e 1
Coil+

N.C.
COM1

N.O.
5
Coil-

3
R+5V RVcc
2 2

C808
C809 .047uF
.047uF

R R

R+5V R+5V TP806 Serial I/O

16
TP807

1
C815 connector
.047uF R819 215

Vdd
U802-2

Vcc
15 2 J801
1RY 1RA
R810 U803 rly-NO
1
619 HCNW4503 13 4 rly-NC
R 2RY 2RA TP808 2
iso803-a 2 8 R811
RS423Tx 3
6.19K 11 6
3RY 3RA R 4
R817 215 Rx
5
6 iso803-c 14 3 U802-3 Tx
1DA 1DY 6
3
7
5 12 5 rly-COM
2DA 2DY 8

10 7 RJ-45
3DA 3DY

GND

Vss
R

U802

8
TL145406 C814 C812
R
680pF 680pF
R+5V
RVee
R R

+5V +5V
R R
3

C816 C810
.047uF 1 U802-15 .047uF

U804 R808 Q801


2

R807 HCNW4503 619 MMBT3904L


1 R 1
6.19K 8 2 iso802-a q801-e

6
RS423Rx TP809
3
5

Initial Date
Welch Allyn Inc.
R

Drawn Schematic: 620017.dsn


Rick Myers Mar 5, 1999

Designed
Rick Myers Mar 5, 1999
Title
Checked
Serial Communication
L. Phillips 3/16/99

Approved Size Document Number Rev


Z. Psenicnik 3/16/99
C 620017 G
Release For Production
J. Bello 3/16/99
Date: Monday, August 05, 2002 Sheet 8 of 8
A B C D E

Figure D-27. 210/220 serial communication.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 173
4 3 2 1
REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PROD. ENG. 5-39342 JRS 4/08/99 T.W.

NOTE 5 WAS
B 5-43326 RLB 10-12-01 T.W.
"INDUCTANCE : 380 uH p 10%"

D D

.150
.514 .750 MAX
.710 MAX

.600 .815 MAX

.200
.200
PCB
C C
HOLE PATTERN

NOTES: 1
7.5V .5V 40MA
+12V
1. MANUFACTURER: ZMAN MAGNETIC
BEAVERTON OREGON
5
2. TRANSFORMER MUST MEET UL CONSTRUCTION TECHNIQUES.
2
3. LABELING: THE TRANSFORMER SHALL BE LABELED, AND + 5 V (REGULATED) @ 100 mA
INCLUDE THE FOLLOWING:

- MANUFACTURER'S NAME
- WELLCH ALLYN PART NUMBER I.E. 620020 3
B - DATE CODE: YYWW (OR APPROVED ACCEPTABLE FORM) B

4. SCHEMATIC AS NOTED.

5. SPECIFICATIONS:

ISOLATION CAPACITANCE: < 7.5 pF 8


VOLTAGE BREAKDOWN : > 8 KV DC 4
INPUT : +12 V DC 7.5V .5V 40MA
OUTPUT 1 : +7.5 V 0.5 V @ 40 mA
OUTPUT 2 : +5 V @ 100 mA (REGULATED) PWM
OUTPUT 3 : -7.5 V 0.5 V @ 40 mA

SWITCHING FREQUENCY : 76.8 kHZ


- -
OPERATING MODE : DISCONTINUOUS
PART NO. DESCRIPTION

PRIMARY WINDING : PIN 5 TO PIN 8 FEEDBACK +5V THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.
INDUCTANCE : 135 uH 10% AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE

RESISTANCE : 0.15 OHMS 15% OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

DRAWN DATE
SECONDARY WINDING : PIN 1 TO PIN 4 R

INDUCTANCE : 380 uH +10%/-15% B WJM 01-Apr-99


RESISTANCE : 0.25 OHMS 15% APPROVED DATE MATERIAL:
A ZIP 4/08/99 - A
CIRCUIT SCHEMATIC
6. THE FOLLOWING TESTS TO BE PERFORMED ON 100% OF THE PARTS AND REL TO PROD DATE FINISH:

WRITTEN CERTIFICATION TO BE SUPPLIED WITH EACH LOT: J. BELLO 4/08/99 -


TITLE

A. PRIMARY INDUCTANCE (PIN 5 TO PIN 8). - - TRANSFORMER


B. SECONDARY INDUCTANCE (PIN 1 TO PIN 4). CAD SOFTWARE: PROE ISOLATION
C. TURNS RATIOl 1-4 TO 5-8. TRANSLATED FROM: ME10
D. VOLTAGE BREAKDOWN. DRAWING NO. REV

C
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN 620020 B
7. PART TO BE A RECOGNIZED COMPONENT UNDER UL2601 OR EN60601.1 CONSTRUCTION. .XXX =
ANGLES
.005
2
INCHES
SCALE FULL SHEET 1 of 1
4 3 2 1

Figure D-28. 200/210/220 transformer isolation.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 174
4 3 2 1
REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PROD. ENG. 5-39391 JRS 4/16/99 T.W.

ADD 620027-3,-4 (IEC)


B 5-39935 ASK 7-8-99 S.S.
CHANGE NOTES 3,4

120.00 3.00
120.00 3.00
PATIENT END

B RA
PATIENT END

RA
RA (WHT)

PART#:1340PS LOT#: XXXXXX

RA
KENDALL
P/N:1540PS LOT#XXXXXX
RL (GRN)
D LL
D

LL

RL
C (BRN)

C
LA

LL
LL (RED)
LA
620027-2

LA
B A LA (BLK)

620027-1 3
1 RA WHITE 3
WHITE (RA, 1) 1 RA WHT
6 SHIELD 1 WHT (RA)
3 LL RED 5 RL GRN
2 LA BLACK 4 C BRN
RED (5) BLACK (LA, 2) JUMPER 3 LL RED
5
PATIENT END 5 GRN (RL) 2 BLK (LA) 2 LA BLK
4 NA
SHIELD
6
RED (LL, 3) 6 NA
4 BRN (C) 3 RED (LL)
CONNECTOR END CONNECTOR END PATIENT END

VIEW B-B CABLE WIRING DIAGRAM VIEW A-A CABLE WIRING DIAGRAM

120.00 3.00 120.00 3.00


PATIENT END
PATIENT END

C D R C R (RED)
C

R
R

KENDALL
P/N:1540PS-I LOT#XXXXXX
N (BLK)

PART#:1340PS-I LOT#: XXXXXX

N
F C (WHT)

C
F
F (GRN)

L
L C L (YEL)
D 620027-4
620027-3 3
3
RED (R, 1) 1 R RED
1 R RED
6 SHIELD 1 RED (R)
3 F GREEN 5 N BLK
2 L YELLOW 4 C WHT
RED (5) GREEN (F, 2) JUMPER 3 F GRN
5 B
PATIENT END 5 BLK (N) 2 YEL (L) 2 L YEL
4 NA B SHIELD
6
YELLOW (L, 3) 6 NA
4 WHT (C) 3 GRN (L)
CONNECTOR END CONNECTOR END PATIENT END

VIEW D-D CABLE WIRING DIAGRAM VIEW C-C CABLE WIRING DIAGRAM

CATALOG REF DESCRIPTION


P/N P/N
B 620027-1 1340PS MW02500 3-LEAD; DIN STYLE SAFETY PATIENT CABLE; B

PACKAGING
NOTES: COLOR CODE:W,R,BL 5
1. CONNECTOR AND TRUNK CABLE SHALL BE 414C GRAY 620027-2 1540PS MW03775 5-LEAD; DIN STYLE SAFETY PATIENT CABLE;
AND THE PATIENT END SHALL BE BLACK. COLOR CODE:W,G,BR,R,BL 6

2. ELECTRICAL RESISTANCE SHALL BE BETWEEN 900 AND 1100 OHMS. 620027-3 1340PS-I 3-LEAD, DIN STYLE SAFETY PATIENT CABLE
IEC COLOR CODE: R,G,Y 8
3. CABLE NAMEPLATE SHALL CONTAIN:
KENDALL 620027-4 1540PS-I 5-LEAD, DIN STYLE SAFETY PATIENT CABLE
B PART NUMBER: SEE TABLE - CATALOG P/N IEC COLOR CODE: R,B,W,G,Y 9
LOT NUMBER: XXXXXX

4. PACKAGING BAG LABEL SHALL CONTAIN: 4 - -


KENDALL
B CATALOG NUMBER: SEE TABLE
PART NO. DESCRIPTION

PART NUMBER: SEE TABLE - REF P/N THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.
DESCRIPTION: SEE TABLE AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE

QUANTITY: 1 EACH 7. CABLE MUST MEET THE PERFORMANCE STANDARD FOR ELECTRODE LEAD WIRES OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

LOT NUMBER: XXXXXX AND PATIENT CABLES FOUND IN TITLE 21 CFR, PART 898 DRAWN DATE
R

JB ENGEL 09-Apr-99
5. CHANNEL DESIGNATIONS: RA=WHITE, LL=RED, LA=BLACK; 8. CHANNEL DESIGNATIONS: R=RED, F=GREEN, L=YELLOW; APPROVED DATE MATERIAL:
A WITH 1 K OHM RESISTORS; WITH 1 K OHM RESISTORS; ZIP 4/16/99 - A
B WIRING PER ANSI/AAMI 1983 ECG CONNECTOR STANDARD;
WIRING PER ANSI/AAMI 1983 ECG CONNECTOR STANDARD; REL TO PROD DATE FINISH:

(COMPLIANT WITH ANSI/AAMI EC53-1995 CABLE AND LEAD STANDARD) (COMPLIANT WITH ANSI/AAMI EC53-1995 CABLE AND LEAD STANDARD) J. BELLO 4/16/99 -
TITLE

6. CHANNEL DESIGNATIONS: RA=WHITE, RL= GREEN, C=BROWN; 9. CHANNEL DESIGNATIONS: R=RED, N=BLACK, C=WHITE; - - ECG PATIENT
LL=RED, LA=BLACK; WITH 1 K OHM RESISTORS; F=GREEN, L=YELLOW; WITH 1 K OHM RESISTORS;
B WIRING PER ANSI/AAMI 1983 ECG CONNECTOR STANDARD; CAD SOFTWARE: PRO E CABLE
WIRING PER ANSI/AAMI 1983 ECG CONNECTOR STANDARD; TRANSLATED FROM:
(COMPLIANT WITH ANSI/AAMI EC53-1995 CABLE AND LEAD STANDARD) (COMPLIANT WITH ANSI/AAMI EC53-1995 CABLE AND LEAD STANDARD) DRAWING NO. REV

C
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN 620027 B
.XXX = .005 INCHES
ANGLES 2 SCALE NONE SHEET 1 of 1
4 3 2 1

Figure D-29. 200/210/220 ECG patient cable.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 175
4 3 2 1
REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PROD. ENG. 5-39387 JRS 4/15/99 T.W.

REM'D FAN,TOP HSG & VIEWS W/FAN;


B 5-44568 CRW 5/8/02 JPK
DELETED WIRE TWIST NOTES

D D

5-6 IN-LB TORQUE


GROUND WIRE
TYPICAL
NOTE FEED DIRECTION

C C

4
PARTIAL ASSEMBLY
2

5
EXPLODED VIEW

B B

B 620201-502 SUB-ASSEMBLY

620201-501 SUB-ASSEMBLY WITHOUT FAN

PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.


AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

DRAWN DATE
R

JB ENGEL 13-Apr-99
APPROVED DATE MATERIAL:
A ZIP 4/15/99 - A
REL TO PROD DATE FINISH:

J. BELLO 4/15/99 -
TITLE

- - POWER SUPPLY\
CAD SOFTWARE: PRO E SUB-ASSEMBLY
TRANSLATED FROM: ME-10
DRAWING NO. REV

C
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN 620201 B
620201X1.MI .XXX =
ANGLES
.005
2
INCHES
SCALE NONE SHEET 1 of 1
4 3 2 1

Figure D-30. 200/210/220 power supply sub assy.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 176
REV DESCRIPTION ECN/ECO INIT DATE CKD

3.150 A RELEASE TO PROD. ENG. 5-39363 JRS 4/14/99 T.W.

B CHANGE PRYON P/N TO 000.10100 5-39549 JRS 5/5/99 T.W.

C ADDED NOTE 6, 620032 -1 5-40470 TJW 11/2/99 ASK


Seri
al#
XXX
XX PRY
ON
C-CA
P

3.900

1 3
2

POWER AND COMMUNICATION


PUMP EXHAUST
MOLEX 22-11-2042 (1/16 I.D.)
SENSOR INLET
(FROM WATER TRAP RECPTACLE)
4
WATER TRAP SENSE
MOLEX 22-29-2021

.832

NOTES: THIS DWG -1 PRYON SERIAL #400 AND HIGHER


B
1 PRYON LC-101 GENERIC SUBASSEMBLY P/N 000.10100 PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.


2 NOMINAL LENGTH (FROM PUMP BODY T} O TUBING EXHAUST) AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
13.4 INCHES. OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

DRAWN DATE
R

3 NOMINAL LENGTH (FROM SENSOR BODY TO FEMALE LUER INLET) WJM 01-Apr-99
9.8 INCHES. APPROVED DATE MATERIAL:

ZIP 4/14/99 -
4 WATERTRAP AND WATERTRAP RECEPTAC} LE REQUIRED FOR OPERATION. REL TO PROD DATE FINISH:

J. BELLO 4/14/99 -
5. PACKAGING TO BE APPROVED BY WELC H ALLYN. HANDLING AND PACKAGING TO TITLE
PREVENT DAMAGE DUE TO ESD, BENDI}} NG AND OTHER DAMAGE. - - ETCO2 PCB
CAD SOFTWARE: PROE ASSEMBLY
6. ASSY TO BE MARKED WITH MANUFACTURER'S SERIAL NUMBER. TRANSLATED FROM: ME10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620032 C
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Figure D-31. 220 ETCO2 PCB assembly.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 177


REV DESCRIPTION ECN/ECO INIT DATE CKD

SEE ECN
B 5-39954 ASK 7-12-99 T.W.
(REV A REL. AS EXT ITM MSTR)

PERFORMANCE SPECIFICATIONS J2 OUTPUT J1 AC INPUT


C REVISED SUPPLIER P.N. 5-41401 JRS 5/18/00 T.W.

INPUT VOLTAGE 90 - 264 Vac 1) OUTPUT 1 (+) 1) LINE

INPUT FREQUENCY 47 - 63 Hz 2) OUTPUT 1 (+) 2)

INPUT CURRENT (FULL LOAD) 1.6 A 3) OUTPUT 1 (+) 3) NEUTRAL

OVERVOLTAGE PROTECTION 14 1.1 V 4) RETURN

OUTPUT VOLTAGE 12 VDC @4.2 A 5) RETURN

OUTPUT POWER-CONTINUOUS 50 W 6) RETURN

OUTPUT POWER - PEAK 55 W

OPERATING TEMPERATURE 0-50 C

CISPRII CLASS B
EMI COMPLIANCE
(EN55011)

OUTPUT NOISE P-P 1%

620150-1 REV B OR HIGHER

NOTES: PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.


1. SUPPLIER: CONDOR POWER SUPPLIES INC. AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE

2311 STATHAM PKWY. OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

OXNARD, CA. 93033 DRAWN DATE


R

PART NO: GLM50-12


C ASK 12-Jul-99
APPROVED DATE MATERIAL:

2. SUPPLY MUST BE APPROVED TO ZIP 7-12-99 -


UL 2601 AND IEC 601-1 REL TO PROD DATE FINISH:

J. BELLO 7-12-99 -
3. SUPPLIER REV. C OR HIGHER TITLE

- - POWER SUPPLY, 50W


CAD SOFTWARE: PRO/E
TRANSLATED FROM:
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620150 C
.XXX = .005 INCHES
ANGLES 2
SCALE SHEET 1 of 1

Figure D-32. 200/210/220 Power supply.


Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 178
Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 179
Figure D-33. Nonin SpO2 Board with Shield.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 180


Figure D-34. Motor Stepper.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 181


Figure D-35. Pump Pneumatic.

REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PRODUCTION (X2) 5-39025 JRS 2/18/99 T.W.

A
1.75 .25

.38

9 4
2
10
8
ITEM 1 ITEM 3
6 1
1 1
5 2
2 2
3 3
4 3
PIN 1 4 4
11 5 5
1 5 7 3
SECTION A-A 6 6
6
3.50 .25
1 11 SHRINK TUBE
1 10 WIRE, 22GA, GREEN
A
1 9 WIRE, 22GA, BROWN
1 8 WIRE, 22GA, BLUE
1 7 WIRE, 22GA, RED
1 6 WIRE, 22GA, BLACK
1 5 WIRE, 22GA, WHITE
NOTES:
6 4 CONTACTS, MOLEX 08-5 0-0113

1. ALL CONNECTORS ARE TO BE UL RECOG NIZED COMPONENTS AND CSA CERTIFIED. 1 3 6 PIN CONNECTOR, MOL EX 22-01-3067
CONNECTOR MATERIAL SHALL HAVE FLA MMABILITY RATINGS OF 94V-2 OR BETTER . 6 2 ECG PINS (FEMALE), A MP-66105-3
1 1 ECG CONNECTOR WA PAR T # 620102
QTY ITEM DESCRIPTION
2. WIRE SPECIFIATIONS
PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.


-TYPE U.L. 1061, CSA AWM I A/B FT1 AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.
-COLOR CODED .009 THK. (NOM) PVC INS ULATION DRAWN DATE
R

-TEMP RANGE: -10 C TO +80 C JB ENGEL 13-Jan-99


APPROVED DATE MATERIAL:
-VOLTAGE RATING: 300V
ZIP 2/19/99 -
-STRANDED TINNED COPPER CONDUCTOR REL TO PROD DATE FINISH:

-RECOGNIZED COMPONENT MARK RU AND CS A MUST J. BELLO 2/19/99 -


TITLE
APPEAR ON SPOOL OR BE STAMPED ON WI RE
- - CABLE ASSY
CAD SOFTWARE: PE ECG
TRANSLATED FROM: ME10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620165 A
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 182


Figure D-36. 200/210/220 ECG cable assembly.

REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PRODUCTION (X2) 5-39025 JRS 2/18/99 T.W.

B CHANGE WIRE SPEC IN NOTE 2 5-39543 JRS 5/5/99 T.W.

ADD .75" IN LENGTH


C C ADD FERRITE 5-39879 JRS 6/24/99 T.W.
CHANGED PART # 620166-1
4.75 .12
3.75 .12
.75 .12 .378 .12

SINGLE LOOP, ALL WIRES 2


PIN 1 THROUGH FERRITE

1 1
2 2
3 3
3 PIN 1 4 4
2 5
1
8 4 6
7 1
C C 1 8 FAIR RITE P/N 2643801902
1 7 SHRINK TUBE
1 6 WIRE, 22GA, YELLOW
1 5 WIRE, 22GA, ORANGE
1 4 WIRE, 22GA, RED
1 3 WIRE, 22GA, BROWN
8 2 CONTACTS, MOLEX 08-5 0-0113
NOTES:
2 1 4 PIN CONNECTOR, MOL EX 22-01-3047
1. ALL CONNECTORS ARE TO BE UL RECOG NIZED COMPONENTS AND CSA CERTIFIED. QTY ITEM DESCRIPTION
CONNECTOR MATERIAL SHALL HAVE FLA MMABILITY RATINGS OF 94V-2 OR BETTER . 620166-1 THIS DRAWING REV C AND ABOVE

620166
2. WIRE SPECIFICATIONS
PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.


B -TYPE U.L. 1061, CSA AWM I A/B FT1 AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.
-COLOR CODED .009 THK. (NOM) PVC INS ULATION DRAWN DATE
R

-TEMP RANGE: -10 C TO +80 C JB ENGEL 13-Jan-99


APPROVED DATE MATERIAL:
-VOLTAGE RATING: 300V
ZIP 2/19/99 -
-STRANDED TINNED COPPER CONDUCTOR REL TO PROD DATE FINISH:

-RECOGNIZED COMPONENT MARK RU AND CS A MUST J. BELLO 2/19/99 -


TITLE
APPEAR ON SPOOL OR BE STAMPED ON WI RE
- - CABLE ASSY
CAD SOFTWARE: PE CO2
TRANSLATED FROM: ME10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620166 C
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 183


Figure D.37. CO2 cable assembly.
REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PRODUCTION (X2) 5-39025 JRS 2/18/99 T.W.

ADD FERRITE, ITEM 6


B CHANGED PART NO. TO 620169-1
5-39627 JRS 5/14/99 T.W.

CHANGE CONN. TO METAL SHELL


C CHANGED PART NO. TO 620169-2
5-39885 JRS 6/29/99 T.W.

.250 MAX FOR SHIELDED CABLE


.130 MAX. FOR COAX. CABLE SHIELD
4.00 .25
.00 /.06 HEAT SHRINK REF.
B
BOTH CABLES THROUGH CABLE REF.
FERRITE - LOCATE AT MID POINT

1 A SEE DETAIL B

DRAINWIRE YELLOW 2
3
SHRINKWRAPPED (BOTH ENDS)

CONNECTOR REF.
PIN 1 .250 STRIP

SHIELD OVER SHIELD DETAIL B


(OVERLAP SEAM) 4 PLCS

B 1 6 FAIR-RITE 2643625102
PIN 1
4 4 5 HEAT SHRINK
5
A 1 4 COAX CABLE, BELDON#821 6
SECTION A-A 6
1 3 SHIELDED CABLE, CAROL# CO742 W/ DRAIN
BRAID SHIELD CONNECTED TO 26 AWG ORANGE 1 2 3M CONNECTOR, 89110-0101
JUMPER WIRE VIA SOLDER BRIDGE, COVER WITH
C 1 1 SPO2 CONNECTOR, AMP #7 45491-2
LOW TEMPERATURE SHRINK TUBING BOTHE ENDS
QTY ITEM DESCRIPTION
C 620169-2 THIS PART REV C AND ABOVE

MIDDLE COAX 620169-1 THIS PART REV B AND ABOVE


5 1 B
ORANGE JUMPER FROM COAX SHIELD
9 4 620169 THIS PART
6 GREEN 5 PART NO. DESCRIPTION
1 WHITE 6
THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.
BLACK
3 7 AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.
2 RED 9 DRAWN DATE
7 CABLE DRAIN/SHIELD (YELLOW) 10
R

JB ENGEL 14-May-99
4 2 APPROVED DATE MATERIAL:

8 3 ZIP 2/18/99 -
REL TO PROD DATE FINISH:
DB9
8
NOTES:
SCHEMATIC
J. BELLO 2/18/99 -
10 PIN TITLE

1. ALL CONNECTORS ARE TO BE UL RECOG NIZED COMPONENTS AND CSA CERTIFIED. - - CABLE ASSY
CONNECTOR MATERIAL SHALL HAVE FLA MMABILITY RATINGS OF 94V-2 OR BETTER . CAD SOFTWARE: PE NELLCOR SENSOR
TRANSLATED FROM: ME10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
2. CABLE ASSEMBLY TO BE 100% TESTED FOR POINT TO POINT
AND SHIELD OVER SHIELD CONTINUITY.
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620169 C
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 184


Figure D-38. Nellcor sensor cable assy.

REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PROD. ENG. 5-39387 JRS 4/15/99 T.W.

B REVISED ITEM NUMBERS 5-39513 TJW 5/3/99 T.W.


4
ORANGE
WHITE

8 5 10 6 11 6
2
2

ON 8
ECTI
DIR 10
FLOW 7
3
9 2 REF
5
13
2
+ 4
- 1
RED 1 REF

BLACK 620187-501 PNEUMATIC SUB-ASSEMBLY

3 REF PART NO. DESCRIPTION

THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.

4 AND SHALL NOT BE REPRODUCED, OR COPI


OR SALE OF EQUIPMENT OR DEVICES WITH
ED, OR USED AS A BASIS FOR MANUFACTU
OUT WRITTEN PERMISSION.
RE

DRAWN DATE
NOTES:
12 JB ENGEL 13-Apr-99
R

1. ENDS OF TUBING TO BE FLUSH TO BAR B FLANGES


APPROVED DATE MATERIAL:

ZIP 4/15/99 -
2 SOLDER WIRE FROM ITEM 4 TO PUMP TERMINALS USING SOLDER ITEM 13.
REL TO PROD DATE FINISH:

J. BELLO 4/15/99 -
3 ATTACH FAST-ONS FROM ITEM 4. TITLE

- - PNEUMATIC SUB-ASSY
4 MANIFOLD ORIENTATION - HOLES FACI NG UP AS NOTED CAD SOFTWARE: PRO E
TRANSLATED FROM: ME-10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
5 PUMP INTAKE ORIENTATION - TUBE ON TOP SIDE AS NOTED. TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620187 B
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 185


Figure D-39. 200/210/220 Pneumatic sub assy.
REV DESCRIPTION ECN/ECO INIT DATE CKD

A RELEASE TO PRODUCTION 5-39928 JRS 7/2/99 T.W.

B CHANGE DIODE VENDOR 5-44091 RLB 1/22/02

4.75.12

3.75.12

.75.12 .378.12
NOTE CATHODE
SINGLE LOOP, ALL WIRES ORIENTATION
THROUGH FERRITE 2
PIN 1 1 1
2 2
3 3
4 4
10 9 3 PIN 1
2 5
1
4 6 1 10 CLEAR HEAT SHRINK
8 7 1 1 9 DIODE,VISHAY TELEFUNKEN P/N IN4148
1 8 FAIR RITE P/N 2643801902
1 7 SHRINK TUBE
1 6 WIRE, 22GA, YELLOW
1 5 WIRE, 22GA, ORANGE
1 4 WIRE, 22GA, RED
1 3 WIRE, 22GA, BROWN
8 2 CONTACTS, MOLEX 08-5 0-0113
NOTES: 2 1 4 PIN CONNECTOR, MOL EX 22-01-3047
QTY ITEM DESCRIPTION
1. ALL CONNECTORS ARE TO BE UL RECOG NIZED COMPONENTS AND CSA CERTIFIED.
CONNECTOR MATERIAL SHALL HAVE FLA MMABILITY RATINGS OF 94V-2 OR BETTER .

2. WIRE SPECIFICATIONS

PART NO. DESCRIPTION

-TYPE U.L. 1061, CSA AWM I A/B FT1 THESE DRAWINGS AND SPECIFICATIONS AR E THE PROPERTY OF WELCH ALLYN, INC.
AND SHALL NOT BE REPRODUCED, OR COPI ED, OR USED AS A BASIS FOR MANUFACTU RE
-COLOR CODED .009 THK. (NOM) PVC INSULATION OR SALE OF EQUIPMENT OR DEVICES WITH OUT WRITTEN PERMISSION.

DRAWN DATE
-TEMP RANGE: -10 C TO +80 C R

JB ENGEL 13-Jan-99
-VOLTAGE RATING: 300V APPROVED DATE MATERIAL:

-STRANDED TINNED COPPER CONDUCTOR CDM 7/2/99 -


REL TO PROD DATE FINISH:
-RECOGNIZED COMPONENT MARK RU AND CS A MUST J. BELLO 7/2/99 -
APPEAR ON SPOOL OR BE STAMPED ON WI RE TITLE

- - CABLE ASSY
CAD SOFTWARE: PE CO2
TRANSLATED FROM: ME10
DRAWING NO. REV

B
UNLESS OTHERWISE SPECIFIED
TOLERANCES
.XX = .02
DIMENSIONS
ARE IN
620524 B
.XXX = .005 INCHES
ANGLES 2
SCALE NONE SHEET 1 of 1

Figure D-40. 220 CO2 cable assy.

Welch Allyn Atlas Monitor Service Manual 6200-43E Rev D 186


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
Table E-1. Calibration Date Table. 2/10/2003 1866
1/1/2003 1826 2/11/2003 1867
1/2/2003 1827 2/12/2003 1868
1/3/2003 1828 2/13/2003 1869
1/4/2003 1829 2/14/2003 1870
1/5/2003 1830 2/15/2003 1871
1/6/2003 1831 2/16/2003 1872
1/7/2003 1832 2/17/2003 1873
1/8/2003 1833 2/18/2003 1874
1/9/2003 1834 2/19/2003 1875
1/10/2003 1835 2/20/2003 1876
1/11/2003 1836 2/21/2003 1877
1/12/2003 1837 2/22/2003 1878
1/13/2003 1838 2/23/2003 1879
1/14/2003 1839 2/24/2003 1880
1/15/2003 1840 2/25/2003 1881
1/16/2003 1841 2/26/2003 1882
1/17/2003 1842 2/27/2003 1883
1/18/2003 1843 2/28/2003 1884
1/19/2003 1844 3/1/2003 1885
1/20/2003 1845 3/2/2003 1886
1/21/2003 1846 3/3/2003 1887
1/22/2003 1847 3/4/2003 1888
1/23/2003 1848 3/5/2003 1889
1/24/2003 1849 3/6/2003 1890
1/25/2003 1850 3/7/2003 1891
1/26/2003 1851 3/8/2003 1892
1/27/2003 1852 3/9/2003 1893
1/28/2003 1853 3/10/2003 1894
1/29/2003 1854 3/11/2003 1895
1/30/2003 1855 3/12/2003 1896
1/31/2003 1856 3/13/2003 1897
2/1/2003 1857 3/14/2003 1898
2/2/2003 1858 3/15/2003 1899
2/3/2003 1859 3/16/2003 1900
2/4/2003 1860 3/17/2003 1901
2/5/2003 1861 3/18/2003 1902
2/6/2003 1862 3/19/2003 1903
2/7/2003 1863 3/20/2003 1904
2/8/2003 1864 3/21/2003 1905
2/9/2003 1865 3/22/2003 1906

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 187


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
3/23/2003 1907 5/3/2003 1948
3/24/2003 1908 5/4/2003 1949
3/25/2003 1909 5/5/2003 1950
3/26/2003 1910 5/6/2003 1951
3/27/2003 1911 5/7/2003 1952
3/28/2003 1912 5/8/2003 1953
3/29/2003 1913 5/9/2003 1954
3/30/2003 1914 5/10/2003 1955
3/31/2003 1915 5/11/2003 1956
4/1/2003 1916 5/12/2003 1957
4/2/2003 1917 5/13/2003 1958
4/3/2003 1918 5/14/2003 1959
4/4/2003 1919 5/15/2003 1960
4/5/2003 1920 5/16/2003 1961
4/6/2003 1921 5/17/2003 1962
4/7/2003 1922 5/18/2003 1963
4/8/2003 1923 5/19/2003 1964
4/9/2003 1924 5/20/2003 1965
4/10/2003 1925 5/21/2003 1966
4/11/2003 1926 5/22/2003 1967
4/12/2003 1927 5/23/2003 1968
4/13/2003 1928 5/24/2003 1969
4/14/2003 1929 5/25/2003 1970
4/15/2003 1930 5/26/2003 1971
4/16/2003 1931 5/27/2003 1972
4/17/2003 1932 5/28/2003 1973
4/18/2003 1933 5/29/2003 1974
4/19/2003 1934 5/30/2003 1975
4/20/2003 1935 5/31/2003 1976
4/21/2003 1936 6/1/2003 1977
4/22/2003 1937 6/2/2003 1978
4/23/2003 1938 6/3/2003 1979
4/24/2003 1939 6/4/2003 1980
4/25/2003 1940 6/5/2003 1981
4/26/2003 1941 6/6/2003 1982
4/27/2003 1942 6/7/2003 1983
4/28/2003 1943 6/8/2003 1984
4/29/2003 1944 6/9/2003 1985
4/30/2003 1945 6/10/2003 1986
5/1/2003 1946 6/11/2003 1987
5/2/2003 1947 6/12/2003 1988

188 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
6/13/2003 1989 7/24/2003 2030
6/14/2003 1990 7/25/2003 2031
6/15/2003 1991 7/26/2003 2032
6/16/2003 1992 7/27/2003 2033
6/17/2003 1993 7/28/2003 2034
6/18/2003 1994 7/29/2003 2035
6/19/2003 1995 7/30/2003 2036
6/20/2003 1996 7/31/2003 2037
6/21/2003 1997 8/1/2003 2038
6/22/2003 1998 8/2/2003 2039
6/23/2003 1999 8/3/2003 2040
6/24/2003 2000 8/4/2003 2041
6/25/2003 2001 8/5/2003 2042
6/26/2003 2002 8/6/2003 2043
6/27/2003 2003 8/7/2003 2044
6/28/2003 2004 8/8/2003 2045
6/29/2003 2005 8/9/2003 2046
6/30/2003 2006 8/10/2003 2047
7/1/2003 2007 8/11/2003 2048
7/2/2003 2008 8/12/2003 2049
7/3/2003 2009 8/13/2003 2050
7/4/2003 2010 8/14/2003 2051
7/5/2003 2011 8/15/2003 2052
7/6/2003 2012 8/16/2003 2053
7/7/2003 2013 8/17/2003 2054
7/8/2003 2014 8/18/2003 2055
7/9/2003 2015 8/19/2003 2056
7/10/2003 2016 8/20/2003 2057
7/11/2003 2017 8/21/2003 2058
7/12/2003 2018 8/22/2003 2059
7/13/2003 2019 8/23/2003 2060
7/14/2003 2020 8/24/2003 2061
7/15/2003 2021 8/25/2003 2062
7/16/2003 2022 8/26/2003 2063
7/17/2003 2023 8/27/2003 2064
7/18/2003 2024 8/28/2003 2065
7/19/2003 2025 8/29/2003 2066
7/20/2003 2026 8/30/2003 2067
7/21/2003 2027 8/31/2003 2068
7/22/2003 2028 9/1/2003 2069
7/23/2003 2029 9/2/2003 2070

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 189


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
9/3/2003 2071 10/14/2003 2112
9/4/2003 2072 10/15/2003 2113
9/5/2003 2073 10/16/2003 2114
9/6/2003 2074 10/17/2003 2115
9/7/2003 2075 10/18/2003 2116
9/8/2003 2076 10/19/2003 2117
9/9/2003 2077 10/20/2003 2118
9/10/2003 2078 10/21/2003 2119
9/11/2003 2079 10/22/2003 2120
9/12/2003 2080 10/23/2003 2121
9/13/2003 2081 10/24/2003 2122
9/14/2003 2082 10/25/2003 2123
9/15/2003 2083 10/26/2003 2124
9/16/2003 2084 10/27/2003 2125
9/17/2003 2085 10/28/2003 2126
9/18/2003 2086 10/29/2003 2127
9/19/2003 2087 10/30/2003 2128
9/20/2003 2088 10/31/2003 2129
9/21/2003 2089 11/1/2003 2130
9/22/2003 2090 11/2/2003 2131
9/23/2003 2091 11/3/2003 2132
9/24/2003 2092 11/4/2003 2133
9/25/2003 2093 11/5/2003 2134
9/26/2003 2094 11/6/2003 2135
9/27/2003 2095 11/7/2003 2136
9/28/2003 2096 11/8/2003 2137
9/29/2003 2097 11/9/2003 2138
9/30/2003 2098 11/10/2003 2139
10/1/2003 2099 11/11/2003 2140
10/2/2003 2100 11/12/2003 2141
10/3/2003 2101 11/13/2003 2142
10/4/2003 2102 11/14/2003 2143
10/5/2003 2103 11/15/2003 2144
10/6/2003 2104 11/16/2003 2145
10/7/2003 2105 11/17/2003 2146
10/8/2003 2106 11/18/2003 2147
10/9/2003 2107 11/19/2003 2148
10/10/2003 2108 11/20/2003 2149
10/11/2003 2109 11/21/2003 2150
10/12/2003 2110 11/22/2003 2151
10/13/2003 2111 11/23/2003 2152

190 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
11/24/2003 2153 1/4/2004 2194
11/25/2003 2154 1/5/2004 2195
11/26/2003 2155 1/6/2004 2196
11/27/2003 2156 1/7/2004 2197
11/28/2003 2157 1/8/2004 2198
11/29/2003 2158 1/9/2004 2199
11/30/2003 2159 1/10/2004 2200
12/1/2003 2160 1/11/2004 2201
12/2/2003 2161 1/12/2004 2202
12/3/2003 2162 1/13/2004 2203
12/4/2003 2163 1/14/2004 2204
12/5/2003 2164 1/15/2004 2205
12/6/2003 2165 1/16/2004 2206
12/7/2003 2166 1/17/2004 2207
12/8/2003 2167 1/18/2004 2208
12/9/2003 2168 1/19/2004 2209
12/10/2003 2169 1/20/2004 2210
12/11/2003 2170 1/21/2004 2211
12/12/2003 2171 1/22/2004 2212
12/13/2003 2172 1/23/2004 2213
12/14/2003 2173 1/24/2004 2214
12/15/2003 2174 1/25/2004 2215
12/16/2003 2175 1/26/2004 2216
12/17/2003 2176 1/27/2004 2217
12/18/2003 2177 1/28/2004 2218
12/19/2003 2178 1/29/2004 2219
12/20/2003 2179 1/30/2004 2220
12/21/2003 2180 1/31/2004 2221
12/22/2003 2181 2/1/2004 2222
12/23/2003 2182 2/2/2004 2223
12/24/2003 2183 2/3/2004 2224
12/25/2003 2184 2/4/2004 2225
12/26/2003 2185 2/5/2004 2226
12/27/2003 2186 2/6/2004 2227
12/28/2003 2187 2/7/2004 2228
12/29/2003 2188 2/8/2004 2229
12/30/2003 2189 2/9/2004 2230
12/31/2003 2190 2/10/2004 2231
1/1/2004 2191 2/11/2004 2232
1/2/2004 2192 2/12/2004 2233
1/3/2004 2193 2/13/2004 2234

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 191


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
2/14/2004 2235 3/26/2004 2276
2/15/2004 2236 3/27/2004 2277
2/16/2004 2237 3/28/2004 2278
2/17/2004 2238 3/29/2004 2279
2/18/2004 2239 3/30/2004 2280
2/19/2004 2240 3/31/2004 2281
2/20/2004 2241 4/1/2004 2282
2/21/2004 2242 4/2/2004 2283
2/22/2004 2243 4/3/2004 2284
2/23/2004 2244 4/4/2004 2285
2/24/2004 2245 4/5/2004 2286
2/25/2004 2246 4/6/2004 2287
2/26/2004 2247 4/7/2004 2288
2/27/2004 2248 4/8/2004 2289
2/28/2004 2249 4/9/2004 2290
2/29/2004 2250 4/10/2004 2291
3/1/2004 2251 4/11/2004 2292
3/2/2004 2252 4/12/2004 2293
3/3/2004 2253 4/13/2004 2294
3/4/2004 2254 4/14/2004 2295
3/5/2004 2255 4/15/2004 2296
3/6/2004 2256 4/16/2004 2297
3/7/2004 2257 4/17/2004 2298
3/8/2004 2258 4/18/2004 2299
3/9/2004 2259 4/19/2004 2300
3/10/2004 2260 4/20/2004 2301
3/11/2004 2261 4/21/2004 2302
3/12/2004 2262 4/22/2004 2303
3/13/2004 2263 4/23/2004 2304
3/14/2004 2264 4/24/2004 2305
3/15/2004 2265 4/25/2004 2306
3/16/2004 2266 4/26/2004 2307
3/17/2004 2267 4/27/2004 2308
3/18/2004 2268 4/28/2004 2309
3/19/2004 2269 4/29/2004 2310
3/20/2004 2270 4/30/2004 2311
3/21/2004 2271 5/1/2004 2312
3/22/2004 2272 5/2/2004 2313
3/23/2004 2273 5/3/2004 2314
3/24/2004 2274 5/4/2004 2315
3/25/2004 2275 5/5/2004 2316

192 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
5/6/2004 2317 6/16/2004 2358
5/7/2004 2318 6/17/2004 2359
5/8/2004 2319 6/18/2004 2360
5/9/2004 2320 6/19/2004 2361
5/10/2004 2321 6/20/2004 2362
5/11/2004 2322 6/21/2004 2363
5/12/2004 2323 6/22/2004 2364
5/13/2004 2324 6/23/2004 2365
5/14/2004 2325 6/24/2004 2366
5/15/2004 2326 6/25/2004 2367
5/16/2004 2327 6/26/2004 2368
5/17/2004 2328 6/27/2004 2369
5/18/2004 2329 6/28/2004 2370
5/19/2004 2330 6/29/2004 2371
5/20/2004 2331 6/30/2004 2372
5/21/2004 2332 7/1/2004 2373
5/22/2004 2333 7/2/2004 2374
5/23/2004 2334 7/3/2004 2375
5/24/2004 2335 7/4/2004 2376
5/25/2004 2336 7/5/2004 2377
5/26/2004 2337 7/6/2004 2378
5/27/2004 2338 7/7/2004 2379
5/28/2004 2339 7/8/2004 2380
5/29/2004 2340 7/9/2004 2381
5/30/2004 2341 7/10/2004 2382
5/31/2004 2342 7/11/2004 2383
6/1/2004 2343 7/12/2004 2384
6/2/2004 2344 7/13/2004 2385
6/3/2004 2345 7/14/2004 2386
6/4/2004 2346 7/15/2004 2387
6/5/2004 2347 7/16/2004 2388
6/6/2004 2348 7/17/2004 2389
6/7/2004 2349 7/18/2004 2390
6/8/2004 2350 7/19/2004 2391
6/9/2004 2351 7/20/2004 2392
6/10/2004 2352 7/21/2004 2393
6/11/2004 2353 7/22/2004 2394
6/12/2004 2354 7/23/2004 2395
6/13/2004 2355 7/24/2004 2396
6/14/2004 2356 7/25/2004 2397
6/15/2004 2357 7/26/2004 2398

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 193


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
7/27/2004 2399 9/6/2004 2440
7/28/2004 2400 9/7/2004 2441
7/29/2004 2401 9/8/2004 2442
7/30/2004 2402 9/9/2004 2443
7/31/2004 2403 9/10/2004 2444
8/1/2004 2404 9/11/2004 2445
8/2/2004 2405 9/12/2004 2446
8/3/2004 2406 9/13/2004 2447
8/4/2004 2407 9/14/2004 2448
8/5/2004 2408 9/15/2004 2449
8/6/2004 2409 9/16/2004 2450
8/7/2004 2410 9/17/2004 2451
8/8/2004 2411 9/18/2004 2452
8/9/2004 2412 9/19/2004 2453
8/10/2004 2413 9/20/2004 2454
8/11/2004 2414 9/21/2004 2455
8/12/2004 2415 9/22/2004 2456
8/13/2004 2416 9/23/2004 2457
8/14/2004 2417 9/24/2004 2458
8/15/2004 2418 9/25/2004 2459
8/16/2004 2419 9/26/2004 2460
8/17/2004 2420 9/27/2004 2461
8/18/2004 2421 9/28/2004 2462
8/19/2004 2422 9/29/2004 2463
8/20/2004 2423 9/30/2004 2464
8/21/2004 2424 10/1/2004 2465
8/22/2004 2425 10/2/2004 2466
8/23/2004 2426 10/3/2004 2467
8/24/2004 2427 10/4/2004 2468
8/25/2004 2428 10/5/2004 2469
8/26/2004 2429 10/6/2004 2470
8/27/2004 2430 10/7/2004 2471
8/28/2004 2431 10/8/2004 2472
8/29/2004 2432 10/9/2004 2473
8/30/2004 2433 10/10/2004 2474
8/31/2004 2434 10/11/2004 2475
9/1/2004 2435 10/12/2004 2476
9/2/2004 2436 10/13/2004 2477
9/3/2004 2437 10/14/2004 2478
9/4/2004 2438 10/15/2004 2479
9/5/2004 2439 10/16/2004 2480

194 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
10/17/2004 2481 11/27/2004 2522
10/18/2004 2482 11/28/2004 2523
10/19/2004 2483 11/29/2004 2524
10/20/2004 2484 11/30/2004 2525
10/21/2004 2485 12/1/2004 2526
10/22/2004 2486 12/2/2004 2527
10/23/2004 2487 12/3/2004 2528
10/24/2004 2488 12/4/2004 2529
10/25/2004 2489 12/5/2004 2530
10/26/2004 2490 12/6/2004 2531
10/27/2004 2491 12/7/2004 2532
10/28/2004 2492 12/8/2004 2533
10/29/2004 2493 12/9/2004 2534
10/30/2004 2494 12/10/2004 2535
10/31/2004 2495 12/11/2004 2536
11/1/2004 2496 12/12/2004 2537
11/2/2004 2497 12/13/2004 2538
11/3/2004 2498 12/14/2004 2539
11/4/2004 2499 12/15/2004 2540
11/5/2004 2500 12/16/2004 2541
11/6/2004 2501 12/17/2004 2542
11/7/2004 2502 12/18/2004 2543
11/8/2004 2503 12/19/2004 2544
11/9/2004 2504 12/20/2004 2545
11/10/2004 2505 12/21/2004 2546
11/11/2004 2506 12/22/2004 2547
11/12/2004 2507 12/23/2004 2548
11/13/2004 2508 12/24/2004 2549
11/14/2004 2509 12/25/2004 2550
11/15/2004 2510 12/26/2004 2551
11/16/2004 2511 12/27/2004 2552
11/17/2004 2512 12/28/2004 2553
11/18/2004 2513 12/29/2004 2554
11/19/2004 2514 12/30/2004 2555
11/20/2004 2515 12/31/2004 2556
11/21/2004 2516 1/1/2005 2557
11/22/2004 2517 1/2/2005 2558
11/23/2004 2518 1/3/2005 2559
11/24/2004 2519 1/4/2005 2560
11/25/2004 2520 1/5/2005 2561
11/26/2004 2521 1/6/2005 2562

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 195


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
1/7/2005 2563 2/17/2005 2604
1/8/2005 2564 2/18/2005 2605
1/9/2005 2565 2/19/2005 2606
1/10/2005 2566 2/20/2005 2607
1/11/2005 2567 2/21/2005 2608
1/12/2005 2568 2/22/2005 2609
1/13/2005 2569 2/23/2005 2610
1/14/2005 2570 2/24/2005 2611
1/15/2005 2571 2/25/2005 2612
1/16/2005 2572 2/26/2005 2613
1/17/2005 2573 2/27/2005 2614
1/18/2005 2574 2/28/2005 2615
1/19/2005 2575 3/1/2005 2616
1/20/2005 2576 3/2/2005 2617
1/21/2005 2577 3/3/2005 2618
1/22/2005 2578 3/4/2005 2619
1/23/2005 2579 3/5/2005 2620
1/24/2005 2580 3/6/2005 2621
1/25/2005 2581 3/7/2005 2622
1/26/2005 2582 3/8/2005 2623
1/27/2005 2583 3/9/2005 2624
1/28/2005 2584 3/10/2005 2625
1/29/2005 2585 3/11/2005 2626
1/30/2005 2586 3/12/2005 2627
1/31/2005 2587 3/13/2005 2628
2/1/2005 2588 3/14/2005 2629
2/2/2005 2589 3/15/2005 2630
2/3/2005 2590 3/16/2005 2631
2/4/2005 2591 3/17/2005 2632
2/5/2005 2592 3/18/2005 2633
2/6/2005 2593 3/19/2005 2634
2/7/2005 2594 3/20/2005 2635
2/8/2005 2595 3/21/2005 2636
2/9/2005 2596 3/22/2005 2637
2/10/2005 2597 3/23/2005 2638
2/11/2005 2598 3/24/2005 2639
2/12/2005 2599 3/25/2005 2640
2/13/2005 2600 3/26/2005 2641
2/14/2005 2601 3/27/2005 2642
2/15/2005 2602 3/28/2005 2643
2/16/2005 2603 3/29/2005 2644

196 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
3/30/2005 2645 5/10/2005 2686
3/31/2005 2646 5/11/2005 2687
4/1/2005 2647 5/12/2005 2688
4/2/2005 2648 5/13/2005 2689
4/3/2005 2649 5/14/2005 2690
4/4/2005 2650 5/15/2005 2691
4/5/2005 2651 5/16/2005 2692
4/6/2005 2652 5/17/2005 2693
4/7/2005 2653 5/18/2005 2694
4/8/2005 2654 5/19/2005 2695
4/9/2005 2655 5/20/2005 2696
4/10/2005 2656 5/21/2005 2697
4/11/2005 2657 5/22/2005 2698
4/12/2005 2658 5/23/2005 2699
4/13/2005 2659 5/24/2005 2700
4/14/2005 2660 5/25/2005 2701
4/15/2005 2661 5/26/2005 2702
4/16/2005 2662 5/27/2005 2703
4/17/2005 2663 5/28/2005 2704
4/18/2005 2664 5/29/2005 2705
4/19/2005 2665 5/30/2005 2706
4/20/2005 2666 5/31/2005 2707
4/21/2005 2667 6/1/2005 2708
4/22/2005 2668 6/2/2005 2709
4/23/2005 2669 6/3/2005 2710
4/24/2005 2670 6/4/2005 2711
4/25/2005 2671 6/5/2005 2712
4/26/2005 2672 6/6/2005 2713
4/27/2005 2673 6/7/2005 2714
4/28/2005 2674 6/8/2005 2715
4/29/2005 2675 6/9/2005 2716
4/30/2005 2676 6/10/2005 2717
5/1/2005 2677 6/11/2005 2718
5/2/2005 2678 6/12/2005 2719
5/3/2005 2679 6/13/2005 2720
5/4/2005 2680 6/14/2005 2721
5/5/2005 2681 6/15/2005 2722
5/6/2005 2682 6/16/2005 2723
5/7/2005 2683 6/17/2005 2724
5/8/2005 2684 6/18/2005 2725
5/9/2005 2685 6/19/2005 2726

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 197


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
6/20/2005 2727 7/31/2005 2768
6/21/2005 2728 8/1/2005 2769
6/22/2005 2729 8/2/2005 2770
6/23/2005 2730 8/3/2005 2771
6/24/2005 2731 8/4/2005 2772
6/25/2005 2732 8/5/2005 2773
6/26/2005 2733 8/6/2005 2774
6/27/2005 2734 8/7/2005 2775
6/28/2005 2735 8/8/2005 2776
6/29/2005 2736 8/9/2005 2777
6/30/2005 2737 8/10/2005 2778
7/1/2005 2738 8/11/2005 2779
7/2/2005 2739 8/12/2005 2780
7/3/2005 2740 8/13/2005 2781
7/4/2005 2741 8/14/2005 2782
7/5/2005 2742 8/15/2005 2783
7/6/2005 2743 8/16/2005 2784
7/7/2005 2744 8/17/2005 2785
7/8/2005 2745 8/18/2005 2786
7/9/2005 2746 8/19/2005 2787
7/10/2005 2747 8/20/2005 2788
7/11/2005 2748 8/21/2005 2789
7/12/2005 2749 8/22/2005 2790
7/13/2005 2750 8/23/2005 2791
7/14/2005 2751 8/24/2005 2792
7/15/2005 2752 8/25/2005 2793
7/16/2005 2753 8/26/2005 2794
7/17/2005 2754 8/27/2005 2795
7/18/2005 2755 8/28/2005 2796
7/19/2005 2756 8/29/2005 2797
7/20/2005 2757 8/30/2005 2798
7/21/2005 2758 8/31/2005 2799
7/22/2005 2759 9/1/2005 2800
7/23/2005 2760 9/2/2005 2801
7/24/2005 2761 9/3/2005 2802
7/25/2005 2762 9/4/2005 2803
7/26/2005 2763 9/5/2005 2804
7/27/2005 2764 9/6/2005 2805
7/28/2005 2765 9/7/2005 2806
7/29/2005 2766 9/8/2005 2807
7/30/2005 2767 9/9/2005 2808

198 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
9/10/2005 2809 10/21/2005 2850
9/11/2005 2810 10/22/2005 2851
9/12/2005 2811 10/23/2005 2852
9/13/2005 2812 10/24/2005 2853
9/14/2005 2813 10/25/2005 2854
9/15/2005 2814 10/26/2005 2855
9/16/2005 2815 10/27/2005 2856
9/17/2005 2816 10/28/2005 2857
9/18/2005 2817 10/29/2005 2858
9/19/2005 2818 10/30/2005 2859
9/20/2005 2819 10/31/2005 2860
9/21/2005 2820 11/1/2005 2861
9/22/2005 2821 11/2/2005 2862
9/23/2005 2822 11/3/2005 2863
9/24/2005 2823 11/4/2005 2864
9/25/2005 2824 11/5/2005 2865
9/26/2005 2825 11/6/2005 2866
9/27/2005 2826 11/7/2005 2867
9/28/2005 2827 11/8/2005 2868
9/29/2005 2828 11/9/2005 2869
9/30/2005 2829 11/10/2005 2870
10/1/2005 2830 11/11/2005 2871
10/2/2005 2831 11/12/2005 2872
10/3/2005 2832 11/13/2005 2873
10/4/2005 2833 11/14/2005 2874
10/5/2005 2834 11/15/2005 2875
10/6/2005 2835 11/16/2005 2876
10/7/2005 2836 11/17/2005 2877
10/8/2005 2837 11/18/2005 2878
10/9/2005 2838 11/19/2005 2879
10/10/2005 2839 11/20/2005 2880
10/11/2005 2840 11/21/2005 2881
10/12/2005 2841 11/22/2005 2882
10/13/2005 2842 11/23/2005 2883
10/14/2005 2843 11/24/2005 2884
10/15/2005 2844 11/25/2005 2885
10/16/2005 2845 11/26/2005 2886
10/17/2005 2846 11/27/2005 2887
10/18/2005 2847 11/28/2005 2888
10/19/2005 2848 11/29/2005 2889
10/20/2005 2849 11/30/2005 2890

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 199


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
12/1/2005 2891 1/11/2006 2932
12/2/2005 2892 1/12/2006 2933
12/3/2005 2893 1/13/2006 2934
12/4/2005 2894 1/14/2006 2935
12/5/2005 2895 1/15/2006 2936
12/6/2005 2896 1/16/2006 2937
12/7/2005 2897 1/17/2006 2938
12/8/2005 2898 1/18/2006 2939
12/9/2005 2899 1/19/2006 2940
12/10/2005 2900 1/20/2006 2941
12/11/2005 2901 1/21/2006 2942
12/12/2005 2902 1/22/2006 2943
12/13/2005 2903 1/23/2006 2944
12/14/2005 2904 1/24/2006 2945
12/15/2005 2905 1/25/2006 2946
12/16/2005 2906 1/26/2006 2947
12/17/2005 2907 1/27/2006 2948
12/18/2005 2908 1/28/2006 2949
12/19/2005 2909 1/29/2006 2950
12/20/2005 2910 1/30/2006 2951
12/21/2005 2911 1/31/2006 2952
12/22/2005 2912 2/1/2006 2953
12/23/2005 2913 2/2/2006 2954
12/24/2005 2914 2/3/2006 2955
12/25/2005 2915 2/4/2006 2956
12/26/2005 2916 2/5/2006 2957
12/27/2005 2917 2/6/2006 2958
12/28/2005 2918 2/7/2006 2959
12/29/2005 2919 2/8/2006 2960
12/30/2005 2920 2/9/2006 2961
12/31/2005 2921 2/10/2006 2962
1/1/2006 2922 2/11/2006 2963
1/2/2006 2923 2/12/2006 2964
1/3/2006 2924 2/13/2006 2965
1/4/2006 2925 2/14/2006 2966
1/5/2006 2926 2/15/2006 2967
1/6/2006 2927 2/16/2006 2968
1/7/2006 2928 2/17/2006 2969
1/8/2006 2929 2/18/2006 2970
1/9/2006 2930 2/19/2006 2971
1/10/2006 2931 2/20/2006 2972

200 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
2/21/2006 2973 4/3/2006 3014
2/22/2006 2974 4/4/2006 3015
2/23/2006 2975 4/5/2006 3016
2/24/2006 2976 4/6/2006 3017
2/25/2006 2977 4/7/2006 3018
2/26/2006 2978 4/8/2006 3019
2/27/2006 2979 4/9/2006 3020
2/28/2006 2980 4/10/2006 3021
3/1/2006 2981 4/11/2006 3022
3/2/2006 2982 4/12/2006 3023
3/3/2006 2983 4/13/2006 3024
3/4/2006 2984 4/14/2006 3025
3/5/2006 2985 4/15/2006 3026
3/6/2006 2986 4/16/2006 3027
3/7/2006 2987 4/17/2006 3028
3/8/2006 2988 4/18/2006 3029
3/9/2006 2989 4/19/2006 3030
3/10/2006 2990 4/20/2006 3031
3/11/2006 2991 4/21/2006 3032
3/12/2006 2992 4/22/2006 3033
3/13/2006 2993 4/23/2006 3034
3/14/2006 2994 4/24/2006 3035
3/15/2006 2995 4/25/2006 3036
3/16/2006 2996 4/26/2006 3037
3/17/2006 2997 4/27/2006 3038
3/18/2006 2998 4/28/2006 3039
3/19/2006 2999 4/29/2006 3040
3/20/2006 3000 4/30/2006 3041
3/21/2006 3001 5/1/2006 3042
3/22/2006 3002 5/2/2006 3043
3/23/2006 3003 5/3/2006 3044
3/24/2006 3004 5/4/2006 3045
3/25/2006 3005 5/5/2006 3046
3/26/2006 3006 5/6/2006 3047
3/27/2006 3007 5/7/2006 3048
3/28/2006 3008 5/8/2006 3049
3/29/2006 3009 5/9/2006 3050
3/30/2006 3010 5/10/2006 3051
3/31/2006 3011 5/11/2006 3052
4/1/2006 3012 5/12/2006 3053
4/2/2006 3013 5/13/2006 3054

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 201


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
5/14/2006 3055 6/24/2006 3096
5/15/2006 3056 6/25/2006 3097
5/16/2006 3057 6/26/2006 3098
5/17/2006 3058 6/27/2006 3099
5/18/2006 3059 6/28/2006 3100
5/19/2006 3060 6/29/2006 3101
5/20/2006 3061 6/30/2006 3102
5/21/2006 3062 7/1/2006 3103
5/22/2006 3063 7/2/2006 3104
5/23/2006 3064 7/3/2006 3105
5/24/2006 3065 7/4/2006 3106
5/25/2006 3066 7/5/2006 3107
5/26/2006 3067 7/6/2006 3108
5/27/2006 3068 7/7/2006 3109
5/28/2006 3069 7/8/2006 3110
5/29/2006 3070 7/9/2006 3111
5/30/2006 3071 7/10/2006 3112
5/31/2006 3072 7/11/2006 3113
6/1/2006 3073 7/12/2006 3114
6/2/2006 3074 7/13/2006 3115
6/3/2006 3075 7/14/2006 3116
6/4/2006 3076 7/15/2006 3117
6/5/2006 3077 7/16/2006 3118
6/6/2006 3078 7/17/2006 3119
6/7/2006 3079 7/18/2006 3120
6/8/2006 3080 7/19/2006 3121
6/9/2006 3081 7/20/2006 3122
6/10/2006 3082 7/21/2006 3123
6/11/2006 3083 7/22/2006 3124
6/12/2006 3084 7/23/2006 3125
6/13/2006 3085 7/24/2006 3126
6/14/2006 3086 7/25/2006 3127
6/15/2006 3087 7/26/2006 3128
6/16/2006 3088 7/27/2006 3129
6/17/2006 3089 7/28/2006 3130
6/18/2006 3090 7/29/2006 3131
6/19/2006 3091 7/30/2006 3132
6/20/2006 3092 7/31/2006 3133
6/21/2006 3093 8/1/2006 3134
6/22/2006 3094 8/2/2006 3135
6/23/2006 3095 8/3/2006 3136

202 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
8/4/2006 3137 9/14/2006 3178
8/5/2006 3138 9/15/2006 3179
8/6/2006 3139 9/16/2006 3180
8/7/2006 3140 9/17/2006 3181
8/8/2006 3141 9/18/2006 3182
8/9/2006 3142 9/19/2006 3183
8/10/2006 3143 9/20/2006 3184
8/11/2006 3144 9/21/2006 3185
8/12/2006 3145 9/22/2006 3186
8/13/2006 3146 9/23/2006 3187
8/14/2006 3147 9/24/2006 3188
8/15/2006 3148 9/25/2006 3189
8/16/2006 3149 9/26/2006 3190
8/17/2006 3150 9/27/2006 3191
8/18/2006 3151 9/28/2006 3192
8/19/2006 3152 9/29/2006 3193
8/20/2006 3153 9/30/2006 3194
8/21/2006 3154 10/1/2006 3195
8/22/2006 3155 10/2/2006 3196
8/23/2006 3156 10/3/2006 3197
8/24/2006 3157 10/4/2006 3198
8/25/2006 3158 10/5/2006 3199
8/26/2006 3159 10/6/2006 3200
8/27/2006 3160 10/7/2006 3201
8/28/2006 3161 10/8/2006 3202
8/29/2006 3162 10/9/2006 3203
8/30/2006 3163 10/10/2006 3204
8/31/2006 3164 10/11/2006 3205
9/1/2006 3165 10/12/2006 3206
9/2/2006 3166 10/13/2006 3207
9/3/2006 3167 10/14/2006 3208
9/4/2006 3168 10/15/2006 3209
9/5/2006 3169 10/16/2006 3210
9/6/2006 3170 10/17/2006 3211
9/7/2006 3171 10/18/2006 3212
9/8/2006 3172 10/19/2006 3213
9/9/2006 3173 10/20/2006 3214
9/10/2006 3174 10/21/2006 3215
9/11/2006 3175 10/22/2006 3216
9/12/2006 3176 10/23/2006 3217
9/13/2006 3177 10/24/2006 3218

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 203


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
10/25/2006 3219 12/5/2006 3260
10/26/2006 3220 12/6/2006 3261
10/27/2006 3221 12/7/2006 3262
10/28/2006 3222 12/8/2006 3263
10/29/2006 3223 12/9/2006 3264
10/30/2006 3224 12/10/2006 3265
10/31/2006 3225 12/11/2006 3266
11/1/2006 3226 12/12/2006 3267
11/2/2006 3227 12/13/2006 3268
11/3/2006 3228 12/14/2006 3269
11/4/2006 3229 12/15/2006 3270
11/5/2006 3230 12/16/2006 3271
11/6/2006 3231 12/17/2006 3272
11/7/2006 3232 12/18/2006 3273
11/8/2006 3233 12/19/2006 3274
11/9/2006 3234 12/20/2006 3275
11/10/2006 3235 12/21/2006 3276
11/11/2006 3236 12/22/2006 3277
11/12/2006 3237 12/23/2006 3278
11/13/2006 3238 12/24/2006 3279
11/14/2006 3239 12/25/2006 3280
11/15/2006 3240 12/26/2006 3281
11/16/2006 3241 12/27/2006 3282
11/17/2006 3242 12/28/2006 3283
11/18/2006 3243 12/29/2006 3284
11/19/2006 3244 12/30/2006 3285
11/20/2006 3245 12/31/2006 3286
11/21/2006 3246 1/1/2007 3287
11/22/2006 3247 1/2/2007 3288
11/23/2006 3248 1/3/2007 3289
11/24/2006 3249 1/4/2007 3290
11/25/2006 3250 1/5/2007 3291
11/26/2006 3251 1/6/2007 3292
11/27/2006 3252 1/7/2007 3293
11/28/2006 3253 1/8/2007 3294
11/29/2006 3254 1/9/2007 3295
11/30/2006 3255 1/10/2007 3296
12/1/2006 3256 1/11/2007 3297
12/2/2006 3257 1/12/2007 3298
12/3/2006 3258 1/13/2007 3299
12/4/2006 3259 1/14/2007 3300

204 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
1/15/2007 3301 2/25/2007 3342
1/16/2007 3302 2/26/2007 3343
1/17/2007 3303 2/27/2007 3344
1/18/2007 3304 2/28/2007 3345
1/19/2007 3305 3/1/2007 3346
1/20/2007 3306 3/2/2007 3347
1/21/2007 3307 3/3/2007 3348
1/22/2007 3308 3/4/2007 3349
1/23/2007 3309 3/5/2007 3350
1/24/2007 3310 3/6/2007 3351
1/25/2007 3311 3/7/2007 3352
1/26/2007 3312 3/8/2007 3353
1/27/2007 3313 3/9/2007 3354
1/28/2007 3314 3/10/2007 3355
1/29/2007 3315 3/11/2007 3356
1/30/2007 3316 3/12/2007 3357
1/31/2007 3317 3/13/2007 3358
2/1/2007 3318 3/14/2007 3359
2/2/2007 3319 3/15/2007 3360
2/3/2007 3320 3/16/2007 3361
2/4/2007 3321 3/17/2007 3362
2/5/2007 3322 3/18/2007 3363
2/6/2007 3323 3/19/2007 3364
2/7/2007 3324 3/20/2007 3365
2/8/2007 3325 3/21/2007 3366
2/9/2007 3326 3/22/2007 3367
2/10/2007 3327 3/23/2007 3368
2/11/2007 3328 3/24/2007 3369
2/12/2007 3329 3/25/2007 3370
2/13/2007 3330 3/26/2007 3371
2/14/2007 3331 3/27/2007 3372
2/15/2007 3332 3/28/2007 3373
2/16/2007 3333 3/29/2007 3374
2/17/2007 3334 3/30/2007 3375
2/18/2007 3335 3/31/2007 3376
2/19/2007 3336 4/1/2007 3377
2/20/2007 3337 4/2/2007 3378
2/21/2007 3338 4/3/2007 3379
2/22/2007 3339 4/4/2007 3380
2/23/2007 3340 4/5/2007 3381
2/24/2007 3341 4/6/2007 3382

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 205


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
4/7/2007 3383 5/18/2007 3424
4/8/2007 3384 5/19/2007 3425
4/9/2007 3385 5/20/2007 3426
4/10/2007 3386 5/21/2007 3427
4/11/2007 3387 5/22/2007 3428
4/12/2007 3388 5/23/2007 3429
4/13/2007 3389 5/24/2007 3430
4/14/2007 3390 5/25/2007 3431
4/15/2007 3391 5/26/2007 3432
4/16/2007 3392 5/27/2007 3433
4/17/2007 3393 5/28/2007 3434
4/18/2007 3394 5/29/2007 3435
4/19/2007 3395 5/30/2007 3436
4/20/2007 3396 5/31/2007 3437
4/21/2007 3397 6/1/2007 3438
4/22/2007 3398 6/2/2007 3439
4/23/2007 3399 6/3/2007 3440
4/24/2007 3400 6/4/2007 3441
4/25/2007 3401 6/5/2007 3442
4/26/2007 3402 6/6/2007 3443
4/27/2007 3403 6/7/2007 3444
4/28/2007 3404 6/8/2007 3445
4/29/2007 3405 6/9/2007 3446
4/30/2007 3406 6/10/2007 3447
5/1/2007 3407 6/11/2007 3448
5/2/2007 3408 6/12/2007 3449
5/3/2007 3409 6/13/2007 3450
5/4/2007 3410 6/14/2007 3451
5/5/2007 3411 6/15/2007 3452
5/6/2007 3412 6/16/2007 3453
5/7/2007 3413 6/17/2007 3454
5/8/2007 3414 6/18/2007 3455
5/9/2007 3415 6/19/2007 3456
5/10/2007 3416 6/20/2007 3457
5/11/2007 3417 6/21/2007 3458
5/12/2007 3418 6/22/2007 3459
5/13/2007 3419 6/23/2007 3460
5/14/2007 3420 6/24/2007 3461
5/15/2007 3421 6/25/2007 3462
5/16/2007 3422 6/26/2007 3463
5/17/2007 3423 6/27/2007 3464

206 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
6/28/2007 3465 8/8/2007 3506
6/29/2007 3466 8/9/2007 3507
6/30/2007 3467 8/10/2007 3508
7/1/2007 3468 8/11/2007 3509
7/2/2007 3469 8/12/2007 3510
7/3/2007 3470 8/13/2007 3511
7/4/2007 3471 8/14/2007 3512
7/5/2007 3472 8/15/2007 3513
7/6/2007 3473 8/16/2007 3514
7/7/2007 3474 8/17/2007 3515
7/8/2007 3475 8/18/2007 3516
7/9/2007 3476 8/19/2007 3517
7/10/2007 3477 8/20/2007 3518
7/11/2007 3478 8/21/2007 3519
7/12/2007 3479 8/22/2007 3520
7/13/2007 3480 8/23/2007 3521
7/14/2007 3481 8/24/2007 3522
7/15/2007 3482 8/25/2007 3523
7/16/2007 3483 8/26/2007 3524
7/17/2007 3484 8/27/2007 3525
7/18/2007 3485 8/28/2007 3526
7/19/2007 3486 8/29/2007 3527
7/20/2007 3487 8/30/2007 3528
7/21/2007 3488 8/31/2007 3529
7/22/2007 3489 9/1/2007 3530
7/23/2007 3490 9/2/2007 3531
7/24/2007 3491 9/3/2007 3532
7/25/2007 3492 9/4/2007 3533
7/26/2007 3493 9/5/2007 3534
7/27/2007 3494 9/6/2007 3535
7/28/2007 3495 9/7/2007 3536
7/29/2007 3496 9/8/2007 3537
7/30/2007 3497 9/9/2007 3538
7/31/2007 3498 9/10/2007 3539
8/1/2007 3499 9/11/2007 3540
8/2/2007 3500 9/12/2007 3541
8/3/2007 3501 9/13/2007 3542
8/4/2007 3502 9/14/2007 3543
8/5/2007 3503 9/15/2007 3544
8/6/2007 3504 9/16/2007 3545
8/7/2007 3505 9/17/2007 3546

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 207


Appendix E Calibration Date Table
Calibration Date Number of Days Calibration Date Number of Days
9/18/2007 3547 10/29/2007 3588
9/19/2007 3548 10/30/2007 3589
9/20/2007 3549 10/31/2007 3590
9/21/2007 3550 11/1/2007 3591
9/22/2007 3551 11/2/2007 3592
9/23/2007 3552 11/3/2007 3593
9/24/2007 3553 11/4/2007 3594
9/25/2007 3554 11/5/2007 3595
9/26/2007 3555 11/6/2007 3596
9/27/2007 3556 11/7/2007 3597
9/28/2007 3557 11/8/2007 3598
9/29/2007 3558 11/9/2007 3599
9/30/2007 3559 11/10/2007 3600
10/1/2007 3560 11/11/2007 3601
10/2/2007 3561 11/12/2007 3602
10/3/2007 3562 11/13/2007 3603
10/4/2007 3563 11/14/2007 3604
10/5/2007 3564 11/15/2007 3605
10/6/2007 3565 11/16/2007 3606
10/7/2007 3566 11/17/2007 3607
10/8/2007 3567 11/18/2007 3608
10/9/2007 3568 11/19/2007 3609
10/10/2007 3569 11/20/2007 3610
10/11/2007 3570 11/21/2007 3611
10/12/2007 3571 11/22/2007 3612
10/13/2007 3572 11/23/2007 3613
10/14/2007 3573 11/24/2007 3614
10/15/2007 3574 11/25/2007 3615
10/16/2007 3575 11/26/2007 3616
10/17/2007 3576 11/27/2007 3617
10/18/2007 3577 11/28/2007 3618
10/19/2007 3578 11/29/2007 3619
10/20/2007 3579 11/30/2007 3620
10/21/2007 3580 12/1/2007 3621
10/22/2007 3581 12/2/2007 3622
10/23/2007 3582 12/3/2007 3623
10/24/2007 3583 12/4/2007 3624
10/25/2007 3584 12/5/2007 3625
10/26/2007 3585 12/6/2007 3626
10/27/2007 3586 12/7/2007 3627
10/28/2007 3587 12/8/2007 3628

208 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Calibration Date Table Appendix E
Calibration Date Number of Days Calibration Date Number of Days
12/9/2007 3629
12/10/2007 3630
12/11/2007 3631
12/12/2007 3632
12/13/2007 3633
12/14/2007 3634
12/15/2007 3635
12/16/2007 3636
12/17/2007 3637
12/18/2007 3638
12/19/2007 3639
12/20/2007 3640
12/21/2007 3641
12/22/2007 3642
12/23/2007 3643
12/24/2007 3644
12/25/2007 3645
12/26/2007 3646
12/27/2007 3647
12/28/2007 3648
12/29/2007 3649
12/30/2007 3650
12/31/2007 3651

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 209


Serial Number Cut in History Appendix F

Table F-1. Changed to a new BP Valve

Model Serial#
621N0 62201361
621SP 62100948
622N0 62201361
622NP 62201357
623SP 62300410
623NP 62300412

Table F-2. Changed E-Pac. Changed to class B power supply.

Model Serial#
621S0 62100935
621SP 62100941
622S0 62101305
622SP 62200135
622N0 62201305
622NP 62201312
623SP 62300405
623NP 62300408

Table F-3. Upgraded printer cable to a cable with foam tape attached.

Model Serial#
621xx 62104069
622xx 62208111
623xx 62304018
The foam tape ads extra safety into the printer cable.

Table F-4. Upgraded the Main PCB from 501 to 502.

Model Serial#
621xx 62103000
622xx 62206000
623xx 62303000
Upgraded the Main PCB to make it ESU tolerant. Added user feature to print on alarm or not
print on alarm. Software changed to 2.2

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 211


Appendix F Serial Number Cut in History

Table F-5. Upgraded the Nellcor SpO2 PCB from 204/205 to 506.

Model Serial#
621xx 62104000
622xx 62208000
623xx 62304000

NOTE: The new Nellcor 506 SpO2 PCB is backward compatibly. The new 506 SpO2 sensor
extension cable DEC-8, that can be identified by the purple connector, is backward com-
patible with the older 204/205 SpO2 PCB. The older 204/205 SpO2 sensor extension cable
EC-8, that can be identified by the gray connector, will not work it will not physically
plug into the Atlas connector with the new Nellcor 506 SpO2 PCB. The software was
changed to version 2.5.

The older sensors, that are indentified by the gray connector, will not work with the newer
sensor extension cables even though the sensor will physically plug into the newer sensor
extension cable DEC-8. The newer patient sensors have a purple connector and will work
on the older sensor extension cables.

212 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D


Atlas Calibration and Function Test Form
Atlas Model Number______________ Atlas Serial Number_______________
Calibration date__________________ Last Calibration date_______________
Software Version_________________ Boot Revision_____________________
SpO2 Manufacture_______________ SpO2 PCB________________________
SpO2 Software Revision___________ CO2 Revision_____________________
Last CO2 Reset Date:_______________ Last CO2 Calibration______________

N/A=Not Applicable Calibration Results


PASS FAIL Place a check mark by each of the following data items

____ ____ Date and time accuracy


____ ____ BP calibration (50mmHg and 250mmHd)
____ ____ CO2 reset
____ ____ CO2 calibration
____ ____ No load battery voltage calibration
____ ____ Battery voltage calibration
____ ____ Printer adjustment
____ ____ Calibration date set
____ ____ Software upgrade
____ ____ Download NVRAM files

Functional test Results


____ ____ Power on self test
____ ____ Menu functional test
____ ____ BP dymanic test
____ ____ BP statc test
____ ____ Over 15mmHg test
____ ____ BP Dump test
____ ____ Hardware functional test
____ ____ ECG/Respiration test
____ ____ SpO2 test
____ ____ Temperature test
____ ____ CO2 functional test
____ ____ Battery functional tests
____ ____ Printer functional test
____ ____ Print on alarm test
____ ____ ECG alarm test
____ ____ Respiration alarm test
____ ____ Silence alarm test
____ ____ CO2/Respiration test
____ ____ Blood Pressure alarm test

Date Completed__________ Technician_____________________

Service Manual 6200-43E Rev. D Welch Allyn Atlas Monitor 213


214 Welch Allyn Atlas Monitor Service Manual 6200-43E Rev. D

You might also like