Professional Documents
Culture Documents
We evaluated the Technicon RA-1 000#{174} random-access sensitive electrode and a silver/silver chloride reference
analyzer for the measurements of sodium, potassium, and electrode. For CO2 measurements, a carbonate-sensitive
carbon dioxide by an indirect potentiometnc method (ion- electrode (2,3) is used in the RA-1000 and a Pco2 electrode
selective electrode) and for chloride by a colorimetric method (measuring the rate of change of the pH differential) is used
(mercuric thiocyanate). For various concentrations of control in the ASTRA-8. The RA-1000 ISE module has three basic
materials the total precision (CV) rangedfrom0.9 to 1.2% for components: ISE assembly, sample assembly, and punip/e-
sodium, 1.1 to 1.3% for potassium, 1.0 to 1.2% for chloride, lectronic assembly. After dilution with a buffer in the
and 2.8 to 3.8% for carbon dioxide.The systemdemonstrat- sample assembly, the sample is aspirated through the ISE
assembly for analysis simultaneously with a counterfiow
ed acceptable performance in linearity and carryover. Pa-
reference reagent solution. The pump/electronic assembly
tients results from the RA-1000 correlated well with those
controls the movement of sample and reagent through the
from the Beckman ASTRA-8#{174}. In a study on potential inter-
ISE assembly and the electronic signals between the assem-
ferences, we found that high concentrations of salicylate and
blies. The RA-1000 is calibrated with a two-point calibra-
bromide significantly affected measurements of carbon diox-
tion, i.e., with high- and low-concentration calibrators. At
ide and chloride, respectively. The RA-1 000 requires only 30 the beginning of each worklist (run) and approximately
p.L of sample for all four tests and it offers a high throughput every 8 mm during the course of analysis, the ISE automati-
(30 specimens analyzed for the four tests in 25 mm). This cally analyzes a wash solution as a one-point calibration, to
precise, easy-to-use, random-access analyzer requires mini- adjust the baseline if necessary and to check for drift, and
mal maintenance. the voltage range.
To measure chloride in the RA-b000, a colorimetric meth-
AddItIonal Keyphrases: random-accessanalysis . ion-selective od with mercuric thiocyanate (4) is used; in the ASTRA-8 a
electrodes electrolyteprofile chloride electrode (coulometric-amperometric titration) is
used. In the HA-bOO chloride method-an equilibrium,
Ion-selective electrodes (ISEs) have gained increasing bichromatic assay-the assay mixture is measured at 500
popularity in recent years since simple, precise, and easy-to- nm after 3 mm of reaction; bichromatic blanking is made
use ISEs have become available. With a recently introduced with measurements at 600 nm.
ISE module, the RA-1000 analyzer (Technicon Instruments Calibration: The HA-bOO ISE was calibrated every 4 h
Corp., Tarrytown, NY 10591) now measures sodium, potas- with two solutions: a high calibrator (containing 159
sium, and carbon dioxide in addition to performing chemical mmol of Na, 10.3 mmol of K, and 48.0 mmol of CO2 per
and immunoassays. A bench-top, computer-controlled ana- liter) and a low calibrator (with respective concentrations
lyzer, the HA-bOO can perform, without batching, single or of 96, 2.2, and 12.0 mmol/L). We calibrated C1 every two
multiple tests in any combination or sequence (1). We weeks with a serum-based material containing 102 mmol of
evaluated the analytical performance of the RA-1000 ISE Cl per liter. The ASTRA-8 was calibrated every 75 mm
for Nat, K, and C02, and the colorimetric assay for with two serum-based calibrators: a high calibrator (per
chloride (Cl), together constituting the most common elec- liter, 172.2 mmol of Na, 8.8 mmol of K, 125.4 mmol of
trolyte profile in clinical laboratories. We attempted to Cl, and 31.24 mmol of C02) and a low calibrator containing
investigate all potential limitations and interferences and the respective analytes at 118, 3.7, 52.5, and 10.35 mmol/L;
report them as well as any favorable findings of the system. both contained ethylene glycol as preservative. After cali-
bration, we monitored the stability of RA-b000 performance
Materials and Methods by measuring the control or calibrator material both more
frequently than and also beyond the time recommended by
Instruments. Na, K, Cl and CO2 were analyzed with the manufacturer for a new calibration.
both the RA-1000 and the ASTRA-8 (Beckman Instruments, Precision. Precision was assessed according to the Nation-
Inc., Brea, CA 92621). Both systems involve a glass sodium- al Committee for Clinical Laboratory Standards (NCCLS)
sensitive electrode, a valinomycin-membrane potassium- protocol EP5-P by assaying quality-control materials and
plasma pools in duplicate in the morning and in the
afternoon for 11 days. The total SD and CV were calculated
Department of Laboratory Medicine, University Hospital, Boston from all results n each of three levels of quality-control
University Medical Center, Boston, MA 02118.
Presented in part at the 36th national meeting of the AACC, materials.
Washington, DC, July, 1984. Linearity. Technicon TQC Alert#{174} 1 and 2 lyophilized
Received October 9, 1984; accepted December 6, 1984. serum-based quality-control materials were reconstituted in
0
0
Co
120
I
S..
#{149}
tuted in solutions containing high concentrations of lithium, 0
0
calcium, ammonia, magnesium, and bromide. The Na and
K results for these preparations were compared with those
for the same control material reconstituted in de-ionized
water. To study salicylate interference with the CO2 elec- I i i
trode, we used patients specimens known to contain high 100 120 140 160
concentrations of the drug; we also added salicylate to drug- ASTRA-8 Sodium mmol/L
free pooled plasma. Concentrations of ammonia and saucy-
late were determined with an aca discrete analyzer (DuPont Ag. 1. Sodium in plasma as measured with the ASTRA-8 and the RA-
Instruments, Wilmington, DE 19898). To study interference 1000
from hemolysis and icterus, we determined the analytical
recoveries of Na, K, Cl-, and CO2 after adding standard RA-1 000 Values = 0.936 Astra Values +0.005
= 0.996
solutions of NaC1, KC1, or NaHCO3 to icteric or hemolyzed 9 Sy.x = 0.08
plasma pools. To evaluate the effect of lipemia, we analyzed n= 102
lipemic plasma or serum samples before and after ultracen-
trifugation. -l
o
Results E
E
For various concentrationsof quality-control materials,
the total CVs (11 days, 44 analyses) were no more than 1.3% E
..
#{149}
for Na, K, and C1, and no more than 3.8% for CO2 (Table a
1).
4
Table 1. PrecisIon of RA-1000 In Measuring Three
Concentrations of Electrolytes
Mean, Mean, Mean,
Na
K
Cl
CO2
mmol/L
115
3.2
89
13
CV, %
0.9
1.3
1.2
3.8
mmol/L
137
5.8
104
19
CV, %
1.3
1.2
1.0
2.8
mmol/L
148
8.9
115
25
CV, %
1.2
1.1
1.0
3.3
Il ASTRA-8 Potassium. mmol/L
n= 44 each.
Fig. 2. Same, for potassium in plasma
Discussion
ASTRA-8 Carbon Dioxide, mmoi/L
We have evaluated measurement of Na, K, C1, and
Fig. 4. Same, forcarbon dioxide in plasma
CO2 in the HA-bOO, with particular attention to potential
technical problems that may affect results. The precision
calibration. The mean values at different time intervals study conducted according to NCCLS protocol demonstrated
varied by no more than 1 mmol/L and 0.08 mmol/L for Na satisfactory performances for all four tests. The CVs were
and K, respectively (Table 3). For CO2, the mean values (n comparable to those exhibited by the ASTRA-8 system in
= 4) of the high calibrator varied by no more than 2 mmol/L our laboratory.
up to 8 h after a calibration. For C1, the mean values (n = Acceptable performance in linearity was confirmed with
4) of the quality-control serum varied by no more than 3 regression analysis of our data for the four assays. The
mmol/L for as long as four weeks after a calibration. The linear ranges are sufficient for clinical need.
HA-bOO rarely failed in calibration throughout our evalua- For method comparison, we compared many HA-bOO
tion. results for patients specimens with those obtained with the
ASTRA-8, finding a good correlation.
Table 3. RA-1000 Performance after Calibration Although the manufacturer recommends calibrating the
Mean (and SD), mmol/L ISE every 4 h, we found that a calibration can hold up to 8 h
Tim., Time, Cr, mean for Na, K, and CO2 measurements. For chloride, we found
h Ns CO2 days (and SD), mmol/L that the calibration curve was stable for at least four weeks.
0 125.7 (0.5) 5.47 (0.05) 50.5 (0.9) 0 113.2(0.5) Compared with the required calibration of the ASTRA,
2 124.7 (1.2) 5.45 (0.06) 51.0 (1.4) 7 112.2 (1.4) every 75 mm, the much less frequent calibrations needed
4 126.0 (0) 5.42 (0.05) 51.75 (0.5) 14 110.2 (0.5) with the HA-bOO save reagent, calibrator, and quality-
6 126.0 (1.1) 5.55 (0.06) 52.0 (0) 21 112.5(0.6)
8 126.0 (0) 5.50 (0) 51.75 (0.5) 28 113.2 (0.9) control materials as well as instrument time. This feature
becomes more important when rapid turnaround of results
n= 4 each.
is needed on stat requests and a calibration cycle (e.g.,
program on the HA-bOO so that it can detect any salicylate clinical evaluation of the random access analyzer HA-bOO. Clin
C/tern 30, 364-368 (1984).
interference exceeding approximately 4 mmol/L, and an
2. Diebler H, Adler H, Svenjak D, et al. A new approach for the
error code is printed (flagged) next to the CO2 result. The rapid, simultaneous determination of Na, K and total CO2 on the
revised software can distinguish an abnormal curve of CO2 Technicon RA-1000 system. Clin Chem 29, 1193 (1983). Abstract.
electrode response that does not plateau within the analysis 3. Scott W, Chapoteau E, Jensen M, et a!. Carbonate ion-selective
time (5.5 s), as in the case of salicylate interference. We have membrane electrode with improved characteristics. Clin C/tern 30,
confirmed the ability of the new software (RA-b000 Program 966 (1984). Abstract.
Disk, version 2.3) to detect the salicylate interference. We 4. Schoenfeld RG, Lewellen (11. A colorimetric method for deterrni-
(and the manufacturer) recommend that the four samples nation of serum chloride. Clin C/tern 10, 533-539 (1964).
following a flagged sample be reanalyzed to veri1 proper 5. Durst HA. Automated analyzer for the determination of potassi-
CO2 values. If the repeated results of the four specimens urnand sodium in whole blood. Clin Chim Acta 80,225-234(1977).
differ substantially from their original values, the rest of the 6. DeVivo DC, Keating JP. Reyes syndrome. Adu Pediatr 22, 175-
samples in the original run, excluding the flagged speci- 229 (1976).
men, should be repeated. The flagged specimen should be 7. Elm RJ, Robertson EA, Johnson E. Bromide interferes with
analyzed by an alternative method. In our evaluation, determination of chloride by each of four methods. Clin Chem 27,
778-779 (1981).
exposure to salicylate did not appear to shorten the useful
8. Steelman M, Smith CH, Menon A, et al. Interferences with
life of the CO2 membrane. potentiometry of CO2 in the Ektachem 400 analyzer. Clin C/tern 30,
Thus, unlike the Ektachem 400, the HA-b000 is pro- 562-565 (1984).
grammed to limit undetected interference of salicylate, if 9. Gilliland BC, Mannik M. Rheumatoid arthritis. In Harrisons
present, to an amount corresponding to no more than 4 Principles of Internal Medicine, KJ Isselbacher et al., Eds.,
mmol of CO2 per liter. In our method-comparison study, we McGraw-Hill Book Co., New York, NY, 1980, pp 1872-1880.