You are on page 1of 10

Product Information PI

ROTANODE
E7239X
E7239FX
E7239GX
Rotating Anode X-ray Tube Assembly

Rotating anode X-ray tube assembly for the


purpose of general diagnostic X-ray
procedures.

Specially processed Rhenium-tungsten faced


molybdenum target of 74 mm diameter.

These tubes have foci 2.0 and 1.0, and are


available for a maximum tube voltage 125 kV.

Accommodated with IEC60526 type


high-voltage cable receptacles.

General Data

IEC Classification (IEC60601-1:2005) ............................................ Class ME EQUIPMET

Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 125 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................. 2.0
Small Focus ............................................................................................................................. 1.0
Nominal Anode Input Power (at 0.1s):
60 Hz 50 Hz
Large Focus ..................................................................................................... 47 kW 42.5 kW
Small Focus .................................................................................................. 22.5 kW 21 kW
Nominal Radiographic Anode Input Power (IEC60613:2010):
60 Hz 50 Hz
Large Focus ..................................................................................................... 47 kW 42.5 kW
Small Focus .................................................................................................. 22.5 kW 21 kW

The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Toshiba Electron
Tubes & Devices Co., Ltd. (TETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
The information contained herein may be changed without prior notice. It is therefore, advisable to contact to TETD before processing with the design of
equipment incorporating this product.
E7239X / E7239FX / E7239GX

Motor Ratings:
Stator: XS-AV
Starting Running
Driven Frequency [Hz] 50/60 50/60
Input Power [W] 1050 270 43
1) 3)
Voltage [V] 200 100 40
2)
Current [A] 6.0 3.0 1.2
4)
Min. Speed Up [s] 0.8 1.5 -
Capacitor [F] 24 24 24
Note: 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
60 Hz ............................................................................................................ Minimum 3200 min-1
50 Hz ............................................................................................................ Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ...................................................................................................... 27.5
Common-Auxiliary Winding ................................................................................................ 58.0
Resistance Between Housing and Low Voltage Terminals ........................................ Minimum 2 M
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 C
Mode of Operation ........................................................................................................... Intermittent

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 479 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................... 16 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Filtration:
Permanent Filtration ............................................................. 0.9 mm Al / 75 kV IEC60522:1999
Available Additional Filter combination (0.4 - 1.5 mm) ..................... Maximum 2.4 mm Al / 75 kV
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor ..................................................................................... 125 kV, 4 mA
X-ray Coverage .................................................................................. 354 354 mm at SID 750 mm
Weight (Approx.) ........................................................................................................................ 16 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number:
E7239X ............................................................................................................................. XH-121
E7239FX ........................................................................................................................... XH-126
E7239GX .......................................................................................................................... XH-150

-2-
E7239X / E7239FX / E7239GX

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage (IEC60613:2010):


Radiographic ...................................................................................................................... 125 kV
Between Anode (or Cathode) and Ground ............................................................................. 62.5 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current: (IEC60613:2010) ................................................... See rating charts
Large Focus ...................................................................................................................... 570 mA
Small Focus ...................................................................................................................... 340 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 5.1 A
Small Focus .......................................................................................................................... 5.1 A
Filament Voltage:
Large Focus (At maximum filament current 5.1 A) ..................................................... 7.7 ~ 10.4 V
Small Focus (At maximum filament current 5.1 A) ....................................................... 5.8 ~ 7.8 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power (IEC60613:2010) ..................................................... 60 W (85HU/s)
Thermal Characteristics:
Anode Heat Content ......................................................................................... 100 kJ (140 kHU)
Maximum Anode Heat Dissipation ................................................................... 475 W (667 HU/s)
X-ray Tube Assembly Heat Content ................................................................ 900 kJ (1250 kHU)
Nominal Continuous Input Power (IEC60613:2010):
Without Air-circulator ................................................................................ 180 W (15 kHU/min)

Environmental Limits

Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 C
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature ............................................................................................................... -20 ~ 70 C
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-3-
E7239X / E7239FX / E7239GX

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 60Hz

Nominal Focal Spot Value: 2.0 Nominal Focal Spot Value: 1.0
570 100kV 90kV 80kV 70kV 60kV 50kV
340
70kV 60kV 50kV
550 80kV
110kV
40kV 90kV 40kV
500 300
125kV
450 100kV
250 110kV
TUBE CURRENT [mA]

TUBE CURRENT [mA]


400
125kV
350
200
300

250 150

200
100
150

100
50
50

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Refer to IEC60613:2010

Conditions: Tube Voltage


Constant Potential High-Voltage Generator
Stator Power Frequency 50Hz

Nominal Focal Spot Value: 2.0 Nominal Focal Spot Value: 1.0
570 90kV 80kV 70kV 60kV 50kV 70kV 60kV 50kV
340
550
100kV 80kV
40kV 300 40kV
500
110kV
90kV
450
125kV 250 100kV
TUBE CURRENT [mA]
TUBE CURRENT [mA]

400
110kV
350
200 125kV
300

250 150

200
100
150

100
50
50

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7

EXPOSURE TIME [s] EXPOSURE TIME [s]


Refer to IEC60613:2010

-4-
E7239X / E7239FX / E7239GX

Emission & Filament Characteristics


Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 2.0


600 18

80kV
500 16
100kV

FILAMENT VOLTAGE [V]


TUBE CURRENT [mA]

400 125kV 14
60kV

300 12
50kV

40kV
200 10

Ef
100 8

0 6
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1

FILAMENT CURRENT [A]

Note1) For Reference Only


Note2) Refer to IEC60613:2010

Nominal Focal Spot Value: 1.0

350 16
80kV

300 100kV 14
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

250 125kV 12
60kV
200 10
50kV

150 40kV 8

100 6

50 Ef 4

0 2
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1

FILAMENT CURRENT [A]

Note1) For Reference Only


Note2) Refer to IEC60613:2010

-5-
E7239X / E7239FX / E7239GX

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve

900
180W
800

700
HEAT STORAGE [kJ]

600

500
COOLING
400

300

200
HEATING
100

0
0 20 40 60 80 100 120 140 160 180 200

TIME [min]

Anode Heating / Cooling Curve

100

COOLING
80
HEAT STORAGE [kJ]

60

40

60W

20

HEATING
0
0 1 2 3 4 5 6 7 8
TIME [min]
The heating curves are showing example of average
input power to anode in operation.

-6-
E7239X / E7239FX / E7239GX

Dimensional Outline of E7239X


Unit mm
140

ANODE
4-M6 8 DEEP

CENTRAL RAY AND


REFERENCE AXIS
152.4
11.1

170


16
41.27

125.4

479
92

90


45
11.1
4-M5 8 DEEP

90
156
11.1

L S

CATHODE

C
2-M63.5, P=1.27

TERMINAL CONNECTIONS
11.1

121
9 8 7 6 5 1 2 3 4 E
53

NC NC NC NC
81

C1 M A
69

ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and No.5 or 6
FOCAL SPOT in series circuit.

CENTER OF FOCAL SPOT


CATHODE TERMINAL
A

C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS B

-1.5mmA1.5mm
TERMINAL CONNECTIONS
-1.5mmB1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE

-7-
E7239X / E7239FX / E7239GX

Dimensional Outline of E7239FX


Unit mm
140

72
ANODE

4-M6 8 DEEP
CENTRAL RAY AND REFERENCE AXIS
152.4

11.1
170
16

41.27
125.4
479

92
90



45

90
156

4-M5 8 DEEP
11.1

L S

CATHODE
81

C
2-M63.5, P=1.27

11.1

TERMINAL CONNECTIONS 121

9 8 7 6 5 1 2 3 4 E
53

81

NCNC NC NC
C1 M A
69

ET
TEMPERATURE RELAY
(NORMALLY CLOSED)

Note) Do not connect terminal No.1 and FOCAL SPOT


No.5 or 6 in series circuit.

CENTER OF FOCAL SPOT


CATHODE TERMINAL
A

C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS B

-1.5mmA1.5mm
TERMINAL CONNECTIONS
-1.5mmB1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE

-8-
E7239X / E7239FX / E7239GX

Dimensional Outline of E7239GX


Unit mm

140

CENTRAL RAY AND REFERENCE AXIS


ANODE
4-M6 8 DEEP

11.1

170
16
41.27

125.4

479
90
4-M5 8 DEEP

90 152.4 156
11.1

L S

CATHODE

C
2-M63.5, P=1.27

81 121
TERMINAL CONNECTIONS
CENTER OF
FOCAL SPOT 9 8 7 6 5 1 2 3 4 E
A

NCNC NC NC
C1 M A
69

B
ET
-1.5mmA1.5mm
TEMPERATURE RELAY
-1.5mmB1.5mm (NORMALLY CLOSED)

Note) Do not connect terminal No.1 and


No.5 or 6 in series circuit.
53 FOCAL SPOT

EXPLANATION OF SYMBOLS
TERMINAL CONNECTIONS : CENTRAL X-RAY
CATHODE TERMINAL
C : COMMON C1 : COMMON ANODE & CATHODE TERMINAL
L : LARGE FOCUS M : MAIN WINDING OF THE STATOR : IEC60526 TYPE
S : SMALL FOCUS A : AUX. WINDING OF THE STATOR
NC : NON-CONNECTION
ET : EARTH TERMINAL

-9-
Marketing Engineering Group, Sales Department
1385,Shimoishigami,Otawara-shi,Tochigi 324-8550,Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

The head office of Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. http://www.toshiba-tetd.co.jp/tetd/eng/company/quality.htm.
PE-E7239X/FX/GX 2017-02