Liquisys M

BA 200C/07/en/12.01
51500274

Valid from software version:


2.20 CUM 223 / 253
Transmitter for Turbidity and
Solids Content

Operating Instructions

ENDRESS+HAUSER
LIQUISYS S

ENDRESS+HAUSER
LIQUISYS S

Endress + Hauser
The Power of Know How
Need information on the instrument?
Please read the following chapters:

General information

Safety

You wish to install the instrument.


The required steps are described in this chapter:

Installation

You wish to operate or reconfigure the instrument.


The operating concept is explained in these chapters:

Operation

Instrument configuration

Interfaces

Need help with problems or maintenance? Please refer


to these chapters:

Maintenance and troubleshooting

Diagnostics and corrective maintenance

% °C
Accessories Ω Technical data

Appendix Index
Liquisys M CUM 223 / 253 Table of contents
UM253E00.CHP

Table of contents
1 General information
1.1 Symbols used . . . . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Storage and transport . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Unpacking. . . . . . . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4 Dismantling, packaging and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Product structure . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2 Safety instructions
2.1 Intended application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 General safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Installation, start-up, operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4 Monitoring and safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 Immunity to interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

3 Installation
3.1 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Sensor installation and measuring cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

4 Operation
4.1 Operator interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3 Function of the operating keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4 Auto / manual mode of operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.5 Operating concept. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

5 Instrument configuration
5.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2 System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.3 Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.4 Monitoring functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.5 Relay contact configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.6 Concentration measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.7 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.8 E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.10 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.11 Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

6 Interfaces

7 Maintenance and troubleshooting


7.1 Troubleshooting common problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.2 Troubleshooting using the error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

8 Diagnosis and corrective maintenance


8.1 Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.2 Corrective maintenance of Liquisys M CUM 223 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.3 Corrective maintenance of Liquisys M CUM 253 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.4 Spare parts orders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.5 Service aids “Optoscope” with “Scopeware” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.6 Corrective maintenance of measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

11 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

12 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

Endress+Hauser 1
General information Liquisys M CUM 223 / 253

1 General information

1.1 Symbols used


Warning: Double insulation

This symbol alerts to hazards which Equipment protected by double


could cause serious injuries as well insulation.
as damage to the equipment if
ignored. Alarm relay

Caution:

This symbol alerts to possible faults Input


which could arise from incorrect
operation. They could cause
damage to the equiment if ignored.
Output
Note:

This symbol indicates important


items of information.

1.2 Storage and transport


The packaging material used to store or Conformance with the ambient conditions
transport the transmitter must provide shock (see Technical data) must be assured.
protection. Optimal protection is provided by
the original packaging materials.

1.3 Unpacking
Verify that the packaging and the contents Keep the original packaging materials for
are undamaged! Inform the post office or future storage or shipping of the instrument.
freight carrier of any damage. Damaged
merchandise must be retained until the matter If you have any questions, consult your
has been settled. supplier or the Endress+Hauser sales agency
in your area (see back cover of these
Check that the delivery is complete and operating instructions for addresses).
agrees with the shipping documents and your
order (refer to nameplate for type and
version).

The delivery includes:

• Transmitter CUM 223 (panel-mounted • Field instrument:


instrument) or CUM 253 (field device) – 1 plug-in screw terminal
• Operating instructions BA 200C/07/en – 1 × cable gland Pg 7
• Panel-mounted instrument: – 1 × cable gland Pg 16, reduced
– 1 set of plug-in screw terminals – 2 × cable glands Pg 13.5
– 2 clamping screws for panel mounting – 1 x NPT adapter set
(optional for CSA versions)

1.4 Dismantling, packaging and disposal


Package the assembly properly for reuse at a provided by the original packaging materials.
later point in time. Optimal protection is Observe local regulations for disposal.

2 Endress+Hauser
Liquisys M CUM 223 / 253 General information

UM253E01.CHP

1.5 Product structure


You can identify the instrument variant by the upgrade for Chemoclean (left of diagonal line)
order code on the nameplate. Under “Codes”, or Plus package (right of diagonal line).
you can find the release code for Software

ENDRESS+HAUSER ENDRESS+HAUSER
LIQUISYS-M turbidity / Trübung
LIQUISYS-M turbidity / Trübung
order code / Best.Nr. : CUM 253-TS0115 order code / Best.Nr. : CUM 223-TU0110
serial no. / Ser.-Nr. : 123456 Codes: 1234 / 5678 serial no. / Ser.-Nr. : 123456 Codes: 1234 / 5678

131085-4B

131085-4B
measuring range / Messbereich : FNU, ppm g/l, % measuring range / Messbereich : FNU, ppm g/l, %
temperature / Temperatur : –5 ... 70 °C temperature / Temperatur : –5 ... 70 °C

output 1 / Ausgang 1 : 0/4 ... 20 mA output 1 / Ausgang 1 : 0/4 ... 20 mA


output 2 / Ausgang 2 : 0/4 ... 20 mA output 2 / Ausgang 2 : 0/4 ... 20 mA Fig. 1.1 Nameplate CUM 253
mains / Netz : 230 VAC 50/60 Hz 7,5 VA mains / Netz : 230 VAC 50/60 Hz 7,5 VA (left)
prot. class / Schutzart : IP 65 prot. class / Schutzart : IP 54 / IP 30
ambient temp. / Umgebungstemperatur : –10 ... +55 °C ambient temp. / Umgebungstemperatur : –10 ... +55 °C Fig. 1.2 Nameplate CUM 223
253-TYP.CDR 223-TYP.CDR (right)

Liquisys M CUM 223 / 253

Version
TB Solids concentration measurement with initial setting for residual concrete water
TU Turbidity and solids concentration measurement
TS Turbidity and solids concentration measurement, with additional functions
(Plus package)
Power supply
0 Power supply 230 V AC
1 Power supply 115 V AC
2 Power supply 230 V AC, CSA Gen. Purp.
3 Power supply 115 V AC, CSA Gen. Purp.
5 Power supply 100 AC
7 Power supply 24 AC, CSA Gen. Purp.
8 Power supply 24 V AC/DC

Measurement output
0 Turbidity / DM
1 Turbidity / DM and temperature
3 PROFIBUS-PA
4 PROFIBUS-DP
5 Turbidity / DM with HART®
6 Turbidity / DM and temperature with HART®

Contacts
05 No additional contacts
10 2 contacts (limits / PID / timer)
15 4 contacts (limits / PID / timer / Chemoclean)
16 4 contacts (limits / PID / timer)

Additional equipment
RM1 Initial setting for residual concrete water

CUM223-
complete order code
CUM253-

Additional functions of the Plus package (TB and TS versions).


• Current output table, fields O23x • Automatic start of cleaning function,
• Monitoring for sensor and process, Field F8
function group P • Concentration measurement, function
• Cleaning (Chemoclean), fields R25x group K

Endress+Hauser 3
Safety instructions Liquisys M CUM 223 / 253

2 Safety instructions

2.1 Intended application


Liquisys M CUM 223 / 253 is a field-tested • Chemicals
and reliable transmitter to determine the • Pharmaceuticals
turbidity and the solids concentration of fluid • Drinking water treatment
media. • Condensate treatment
• Municipal sewage treatment plants
The Liquisys M CUM 223 / 253 is particularly • Water conditioning
suitable for use in the following areas:

2.2 General safety notes


This device has been manufactured for safe Warning:
operation according to the state of the art in
engineering and conforms to the applicable
regulations and European standards (see • Operating this instrument in any
Technical data). It has been designed way other than as described in
according to EN 61010-1 and has left the these instructions may
manufacturer’s works in perfect condition. compromise the safety and
function of the measuring system
However, if used improperly or for purposes and is therefore impermissible.
other than the intended purpose, it may be
dangerous, e.g. due to incorrect connection. • The notes and warnings in these
installation and operating
instructions must be strictly
adhered to!

2.3 Installation, start-up, operation


Warning:

• This device may only be installed, • Check that all connections have
connected electrically, been properly made before
commissioned, operated and powering up the system!
serviced by properly trained
personnel authorized by the • Damaged equipment that may be
system operator. dangerous must not be operated
and should be clearly identified as
• The personnel must be familiar being defective.
with these operating instructions
and must adhere to the • Any troubleshooting of the
instructions described therein. measuring system is to be
performed exclusively by
• Make sure that the power supply authorized, trained personnel.
ratings match the data specified
on the nameplate before you • If faults cannot be remedied, the
connect the instrument to a powre instrument must be removed from
source. service and secured to prevent
accidental start-up.
• A clearly identified mains
disconnecting device must be • Repairs not described in these
installed closed to the instrument. operating instructions may only be
performed at the manufacturer’s
• Live components can be touched works or by the Endress+Hauser
through the vent slots in the Service Organization.
housing and the openings on the
rear of the housing. Do not insert
any tools, wires, etc., in these slots
(CUM 223 only).

4 Endress+Hauser
Liquisys M CUM 223 / 253 Safety instructions

UM253E02.CHP

2.4 Monitoring and safety features


Safety features Monitoring features

The transmitter is protected against external In the event of a system error or power failure,
influences and damage by the following an alarm condition is signalled via a fail-safe
design measures: fault signalling contact.

• Rugged housing
• Degree of protection provided by
enclosure: IP 65 (CUM 253)
• UV resistance

2.5 Immunity to interference


This instrument has been tested according to Warning:
the applicable European standards for
industrial applications with regard to
The specified immunity to
electromagnetic compatibility. It is carefully
interference only applies for devices
protected against electromagnetic
connected as outlined in these
interference by constructive design measures.
operating instructions.

2.6 Declaration of conformity


The transmitter CUM 223/253 has been
developed and manufactured in accordance
with currently valid European standards and
directives..
E+H certifies the compliance with the
standards by using the 4 sign.

Endress+Hauser 5
Installation Liquisys M CUM 223 / 253

3 Installation
The following procedure should be followed
for a complete measuring system installation:

• Installation or attachment transmitter


(see chap. 3.3)
• Selection and connection of cable and
sensor (see chap. 3.4, 3.5 and 9)
• Installation is followed by start-up
(see chap. 5).

3.1 Measuring system


The complete measuring system comprises: Optional:

• The Liquisys M CUM 223 or CUM 253 • Extension cable CYK 8


transmitter • Junction box VBM or RM
• A sensor with integrated temperature
sensor and a permanently attachedcable
• An installation assembly

Flow assembly CUA 250

Liquisys M CUM 253

ENDRESS+HAUSER
LIQUISYS S

REL1
ALARM
REL2
CAL REL

REL1

E REL2

Sensor cable

Retractable Liquisys M CUM 223


assembly ENDRESS+HAUSER

CUA 461 LIQUISYS S

REL1

Complete measuring CAL REL


REL2
ALARM

system consisting of E
REL1

REL2

Liquisys M CUM 223 /


253 with measuring cable, Immersion assembly
assembly and turbidity CYA611
Fig. 3.1 sensor UM2X3SYS.CDR

6 Endress+Hauser
Liquisys M CUM 223 / 253 Installation

UM253E03.CHP

3.2 Dimensions

Pg 13.5

Pg 7

Pg 16

11

ENDRESS+HAUSER
LIQUISYS S

157
REL1
ALARM
REL2
CAL REL

REL1

E REL2

247
∅6

70
115 70

154

170

Dimensions of Liquisys M
FELD1.CDR Fig. 3.2 CUM 253

4
2

Inside of housing of
Liquisys M CUM 253
1 Removable
3 electronics box
2 Partition plate
3 Terminal blocks
FELD2.CDR Fig. 3.3 4 Fuse

Endress+Hauser 7
Installation Liquisys M CUM 223 / 253

96
90 6

139

Liquisys M CUM 223,


EINBAU2.CDR
Fig. 3.4 panel-mounted version

3.3 Mounting

3.3.1 Field instrument


Several mounting options are available for the Weather protection cover CYY 101 can be
Liquisys M in the fieldinstrument version: used for outdoor installation in conjunction
with all mounting variants.
• Post mounting on cylindrical pipes
• Post mounting on a square post Note:
• Wall mounting with fastening screws
If installed outdoors, weather
protection cover required.

Weather protection cover CYY 101

Weather protection cover for outdoors For post or pipe installation, the post
installation, to be mounted on field instrument; mounting kit is additionally required.
material: stainless steel SS 304
Order no: CYY 101-A

270
320

300

Weather protection cover


Fig. 3.5 for field instruments DACH.CDR

8 Endress+Hauser
Liquisys M CUM 223 / 253 Installation

UM253E03.CHP

Post mounting kit Universal mounting post CYY 102

Mounting kit for installation of field housing on Square tube for mounting of measuring
horizontal and vertical pipes (max. ∅ 60 mm); transmitters;
also aids fixing of weather protection cover. material: stainless steel SS 304
Material: stainless steel SS 304 Order no: CYY 102-A
Order no.: 50086842

90
70

∅ max. 60
1495

60 x 60

M6
90

60

∅6
70

90
150

left:
Mounting kit
for post mounting on
cylindrical pipes
120
right:
MONT5.CDR Fig. 3.6 Square mounting post

Endress+Hauser 9
Installation Liquisys M CUM 223 / 253

3.3.2 Mounting examples

Pipe mounting of Liquisys


MONT10.CDR
Fig. 3.7 M field instrument

Liquisys M field instrument

left:
Wall mounting

right:
Mounting with
universal post and
MONT2.CDR
Fig. 3.8 weather protection cover

10 Endress+Hauser
Liquisys M CUM 223 / 253 Installation

UM253E03.CHP

3.3.3 Panel-mounted instrument


The instrument is attached using the supplied tensioning screws (see Fig. 3.9).
The required overall installation depth is approx. 165 mm.

92+0.5 92+0.5

2
92

3
90

96
57

3
Attachment of
ap- panel-mounted device
max. 45
prox. 1 Wall of control
25 139 6
cabinet
2 Gasket
MONT9.CDR Fig. 3.9 3 Tensioning screws

Endress+Hauser 11
Installation Liquisys M CUM 223 / 253

3.4 Electrical connection


Connection diagram

The connection diagram depicted in Fig. 3.10 shows the instrument connections for turbidity
sensors CUS 31 or CUS 41. The connection of the varrious sensors is shown in more detail in
Figs. 3.13 and 3.14.

S
BN 87 41

42
Alarm
WH 88 43 (current-free
Sensor contact postion)

GN 96 47

48
97 Relay 1
YE 49 (current-free
contact postion)
57
31
mA 58
Relay 2
Signal output 1 32 59
TU (current-free
turbidity
contact postion)

51
33
mA
52
Signal output 2 Relay 3
34 53 (current-free
temperature
contact postion)
for Chemoclean:
81 54 Water
Digital
input 1 55
10-50 V Relay 4
(Hold) 82 56 (current-free
contact postion)
for Chemoclean:
93 Cleaner
Digital
input 2 10-50 V
(Chemoclean) 94

∼ =
85 L L+
Auxiliary voltage
15 V
86 N L-
output Mains

Electrical connection of
Liquisys M CUM 223/253
Fig. 3.10 with full wiring ANU01.CDR

Note:

• The instrument has protection • 24V AC/DC models must be


supplied from an energy limiting
class II and is generally operated
SELV source in accordance with
without protective earth connection.
dir. IEC 1010.1 Annex H.

• Mains supply voltage fluctuations


should not exceed ten percent of
the nominal supply voltage.

12 Endress+Hauser
Liquisys M CUM 223 / 253 Installation

UM253E03.CHP

Connections of field instrument

For connection, the measuring cables are introduced through the cable glands on the field
instrument and connected according to the diagrams infigures 3.10 and 3.11.

Note: Caution!

Please label the sensor terminal • Terminal designated as NC


block with the enclosed sticker. may not be switched.

• Undesignated terminals
may not be switched.

SENSOR REL 4 REL 3 REL 2 REL 1 Contacts:


max. 2A
AC: 250V/500VA
DC: 30V/60W

WH YE 54 55 56 51 52 53 57 58 59 47 48 49 41 42 43
B Mains
Hilfsenergie
NC 88 97 NC
NC

NC 31 32 33 34 NC 85 86 93 94 81 82 NC NC NC L1 N (AC)
+ - + - + - + - + - + - (DC)
Digital 2

Digital 1

NC 87 96 S
A
(opt.)
Temp.

10mA
+15V
TU

BN GN

Connection compartment
87, 88 CUM 253 sticker for field
131096-4E instruments CUM 253;
Instrument status:
131096E.CDR Fig. 3.11 de-energised

Connections of panel-mounted instrument

131084-4B 131083-4C
Contacts: max. 2A, AC:250V/500VA, DC:30V/60W

49
Sensor
REL1 47
31 + NC
TU
COM 48 32 - 97 B YE
59 33 + 96 A GN
Temp.
REL2 57 (opt.)
34 - 88 - WH
COM 58 NC 87 + BN
53 85 + NC
+15V
REL3 51 86 - 10mA NC

COM 52 93 + S
Digital 2
56 94 - 42

REL4 54 81 + 43
Digital 1
COM 55 82 - 41

NC NC AC DC
Mains

L1 +
NC NC N -
Connection compartment
sticker for panel-mounted
131083C.CDR Fig. 3.12 instruments CUM 223

Endress+Hauser 13
Installation Liquisys M CUM 223 / 253

3.5 Sensor installation and measuring cable connection


Measuring cable connection

The terminal block is located in a separate Note:


connection compartment which is accessible
by opening the plastic cover. Remove the
pre-pressed knock-outs for cable entry. • Protect cable ends and terminals
against moisture to prevent
The sensor is connected via a multiple-core, inaccurate measurement!
special screened cable which is completely
terminated and permanently connected. Use • For further information on cables
a junction box, e.g. VBM to extend the and junction boxes refer to chap.
measuring cable. Termination instructions are 9 Accessories.
supplied with the measuring cables.

Measuring cable requirements

Sensor type Cable Extension


Non-detachable cable
Turbidity sensor CUS 31, CUS 41 VBM box + CYK 8
on sensor

Maximum cable length


CUS 31 / CUS 41 max. 200 m using CYK 8

Structure and termination of the measuring cable

Sensor Pin CUM 2x3

Screen S

GN
Com A 96

YE
Com B 97

CYK 8 WH
–UB 88

+UB BN
Structure of the special +UB 87
Fig. 3.13 measuring cable CYK 8 U_AN03E.CDR

14 Endress+Hauser
Liquisys M CUM 223 / 253 Installation

UM253E03.CHP

Connection examples

S 87 88 96 97

BN WH GN YE

Cable CYK 8

Sensor

Connection of turbidity
sensors CUS 31 and
U2X3AN.CDR Fig. 3.14 CUS 41

Endress+Hauser 15
Operation Liquisys M CUM 223 / 253

4 Operation

4.1 Operator interface

ENDRESS+HAUSER
LIQUISYS M Display of active
Liquid crystal display
contact and key for
for display measured
relay switchover
values and
in manual mode
configurations data NTU
185 R262

Setpoint
LED indicator for
Field for user labelling alarm function

LEDs indicators for


4 main operating keys limit contactor relay
for calibration and (switching state)
instrument configuration
Key for switching
between automatic /
Operating elements of manual operation
Fig. 4.1 Liquisys M DISPLAY1E.CDR

4.2 Display
LED Display

Indication of current operating mode “Auto” (green LED) or “Manual” (yellow LED)

Indication of controlled relay in “Manual” mode (red LED)

Indicates the state of relays 1 and 2


LED green: measured value within permissible limits, relay inactive
LED red: measured value outside permissible limits, relay active

Alarm indication e.g. for continuous limit violation, temperature sensor failure or system error
(see error list chap. 7)

Liquid crystal display

Setup mode “Hold” mode indicator


Measuring mode Calibration indicator (outputs reflect last For instruments with
indicator (normal mode indicator (configuration) current state) communication:
operation) Display to receive
messages.
Constantly visible in
Burst mode (with HART)
Rel 3/4:
NTU
185
Sensor symbol state indicators
flashes during inactive active
calibration R262
Function coding display

Setpoint In measuring mode:


Main measured value.
In setup / calibr. mode
Error: In measuring mode: e.g. parameter
error display Secondary measured value.
In setup / calibr. mode
e.g. setting value
Fig. 4.2 Liquid crystal display DISPLAY3.CDR

16 Endress+Hauser
Liquisys M CUM 223 / 253 Operation

UM253ED4.CHP

4.3 Function of the operating keys


CAL key
When the CAL key is pressed, the instrument prompts for the calibration
access code (fixed setting: 22 for calibration; any setting for calibration
data checking). Acknowledge with the CAL key to proceed. Calibration
uses the calibration data set in Function Group C.

ENTER key
The ENTER key has the following functions:
Opens the Setup menu in Measurement mode
Stores (acknowledges) entered data in Setup mode.
Moving on within the function groups.

PLUS key and MINUS key


The PLUS and MINUS keys have the following functions:
Selection of function groups
Setting of parameters and numeric values
Relay operation in manual mode (see chap. 4.4).
Pressing the PLUS key allows you to switch between the current
output in % and mA.
Repeatedly pressing the PLUS key displays the following settings in
sequence as secondary measured values:
1. Temperature display in °F
2. Hiding the temperature display
3. Measured value display in FNU
4. Back to basic setting
Repeatedly pressing the MINUS key displays the errors:
1. The current errors are displayed one after the other (max. 10)
2. After all the errors are displayed, the standard display is unhidden.
In function group F, you can define an alarm for each error code
separately.

REL key
The REL key is used in manual mode to switch between the relays and
the manual start cleaning function.

AUTO key
The AUTO key is used to toggle between the automatic and manual
modes of operation.

Escape function
Press the PLUS and MINUS keys simultaneously to return to the main
menu. Press the PLUS and MINUS keys again to return to measuring
mode.

Locking the keypad


Press the PLUS and ENTER keys simultaneously to lock the keypad.
The code prompt displays the code 9999.

Unlocking the keypad


Press the CAL and MINUS keys simultaneously to unlock the keypad.
The code prompt displays the code 0:

Endress+Hauser 17
Operation Liquisys M CUM 223 / 253

4.4 Auto / manual mode of operation

Auto mode
In this mode of operation, the relays are controlled by the transmitter.

REL key
The REL key is used to select one of the relays present in the instrument.

Switch to manual mode


The instrument is switched to manual mode by pressing the following
keys:

Press AUTO key.

Enter code 22.

Select relay or function. Press the REL key to toggle between the relays.
The display shows the selected relay in the second line.

Set the relays. Switch on with PLUS, switch off with MINUS. The relay
state remains in effects until it is actively reset.

Press AUTO key for returning to the measuring mode.

Note:

• Enable the manual mode by • Hardware locking in the manual


entering setup code “22”. mode is not possible.

• The operating mode is retained • The manual settings remain in


even after a power failure. effect until they are actively reset.

• The manual mode takes • Error code E102 is signalled in the


precedence over any other manual mode.
automatic function (hold).

18 Endress+Hauser
Liquisys M CUM 223 / 253 Operation

UM253ED4.CHP

4.5 Operating concept

4.5.1 Operating modes

Measuring mode:
Standard mode of
operation, display of current
measured values

Calibration mode: Setup mode:


Execution of Access to all
calibration routine configuration
settings

Code

The function groups


are selected with the
PLUS or MINUS keys

Description of possible
SET-MOD.CDR
Fig. 4.3 operating modes

Note:

• Remains in setup mode for


approx. 15 mins., the system
automatically returns to measuring
mode. An active hold function
(Hold at Setup) is then reset.

Endress+Hauser 19
Operation Liquisys M CUM 223 / 253

4.5.2 Access codes


All instrument access codes are fixed, i.e.
they cannot be modified. When the instrument
requests the access codes, it recognises the
difference between codes (cf. Fig. 5.3):

• CAL key + Code 22: Access to Calibration


and Offset menus. • PLUS/ENTER keys + Code 9999: Locks
• ENTER key + Code 22: Access to the the keypad.
Configuration menus, allowing • CAL/MINUS keys + Code 0: Unlocks the
configuration and user-specific settings. keypad.
• CAL or ENTER key + any Code: Access
to Read mode, i.e. all settings can be read
but not changed.

4.5.3 Menu structure


The configuration and calibration functions Note:
are arranged in a menu structure by function
groups.
• If a change is made but not
The function groups are selected in the setup confirmed by pressing the ENTER
mode with the PLUS and MINUS keys. key, the previous setting is
The ENTER key is used to move from one retained.
function to the next within a function group.
The PLUS and MINUS keys are used for • See the appendix of these
option selection and editing. Selections must operating instructions for an
be confirmed by pressing the ENTER key. overview of the Liquisys M
This also moves the cursor to the next menu structure.
function.
Pressing the PLUS and MINUS keys at the
same time terminates programming (return to
main menu).

Functions
(parameter selection, value entry)

Function groups
Move forward and backward with
the PLUS and MINUS keys

Move from onee function to the


next with the ENTER key

E E E E
Schematic represemtation
of Liquisys M menu
Fig. 4.4 structure BEDSCHE.CDR

20 Endress+Hauser
Liquisys M CUM 223 / 253 Operation

UM253ED4.CHP

4.5.4 Hold function: “Freezes” the outputs


The current output is “frozen” in the setup
mode and during calibration, i.e. the last
current value is constantly output. HOLD
appears on the display.

Note:

• Hold settings can be found in


chap. 5.6, function S2.
• A possibly accumulated alarm
delay is reset to “0".

• During automatic operation, all


contacts will go to their normal
• The hold function can also be
activated externally via the hold
positions.
input (see connection diagram in
Fig. 3.10; digital input 1).
• An active hold has priority over all
other automatic functions.
• The manual hold (field S3) remains
active even after a power failure.
• With every hold, the I component
of the controller is set to zero.

Endress+Hauser 21
Instrument configuration Liquisys M CUM 223 / 253

5 Instrument configuration
After power-up the instrument performs a Setup mode
self-test and then enters the measuring mode.
➜ SETUP 1 (A) see chap. 5.2.1
Now it can be configured and calibrated for ➜ SETUP 2 (B) see chap. 5.2.2
the first time. The values set by the user are ➜ CURRENT OUTPUT (O)
kept even in the event of a power failure. see chap. 5.3
➜ ALARM (F) see chap. 5.4.1
The following function groups are available on ➜ CHECK (P) see chap. 5.4.2
the Liquisys M (the groups that are only
➜ RELAY (R) see chap. 5.5
available in the Plus package are marked
➜ CONCENTRATION (K)
accordingly in the functional descriptions):
see chap. 5.6
➜ SERVICE (S) see chap. 5.7
➜ E+H SERVICE (E) see chap. 5.8
➜ INTERFACE (I) see chap. 5.9

Calibration and offset mode

➜ CALIBRATION (C) see chap. 5.10


➜ OFFSET (V) see chap. 5.11

Function coding.
The displayed code indicates
the function position in the
s
2 F3
function group.

Example of display
Err.Delay Additional information
DISPLAY4.CDR
Fig. 5.1 in setup mode

MATRIX.CDR Selecting and locating functions is facilitated


by a code displayed for each function in a
A131 A132 A133 special display field. The structure of this
coding is given in Fig. 5.2.
The first column indicates the function group
A121
as a letter (see group designations). The
functions in the individual groups are counted
A A1 A111 from the top to the bottom and from the left to
Fig. 5.2 Function coding the right.

Factory settings

When the instrument is switched on for the Please refer to the descriptions of the
first time, the factory settings are in effect. The individual functions in chapter 5 for all other
following table provides an overview of all factory settings (the factory setting is printed
major settings. in bold type).

Function Factory setting


Type of measurement Turbidity in FNU, temperature in °C
Temperature offset / turbidity 0 °C
offset 0 FNU
Limit 1 9999 FNU
Contact function of limit MAX contact without delay
contactor 1
Limit 2 9999 FNU

22 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Contact function of limit MAX contact without delay


contactor 2
Current outputs 1 and 2* 4 ... 20 mA
Current output 1: measured value 0 FNU
for 4 mA signal current
Current output 1: measured value 9999 FNU
for 20 mA signal current
Current output 2: temperature
–5.0 C
value at 4 mA signal current*
Current output 2: temperature
70.0 °C
value for 20 mA signal current*
Measured value damping 10
Calibration data set no. 3
Wiper controller Off

* if equipped accpordingly

Alarm contacts

Normal operating state Alarm state

41 41

42 42

43 43

Recommended fail-safe
circuit for the alarm
RELAIS2.CDR Fig. 5.3 contact

Normal operating state: Alarm state:

• Instrument in operation • Error message available (Alarm LED red)


• No error message available or
(Alarm LED off) • Device defective or voltage-free
(Alarm LED off)
→ Relay picked up
→ Contact 42/43 closed → Relay dropped out
→ Contact 41/42 closed

Endress+Hauser 23
Instrument configuration Liquisys M CUM 223 / 253

5.1 Start-up
After power-up the user must make the following selections in the specified function groups:

• Function group SERVICE (S) • Function group SETUP 2 (B)


S1: Select language and exit function Make all settings in this group, see chap.
group (not required for English). 5.2.2.
• Function group SETUP 1 (A)
Adjust all parameter in this group, see Other configuration options are explained in
chap. 5.2.1. the chapters to follow for each menu.

5.2 System configuration


The system is configured using the function groups SETUP 1 and SETUP 2. Here, select the
measurement type, the sensor and the settings for the temperature measurement. All the
parameters in these two function groups are to be configured to avoid measuring errors or
failure to measure at all.

5.2.1 Setup 1

Selection or range
Coding Field Display Info
Factory setting

Function group A
Initial display in function group
A SETUP 1.
SETUP 1
SETUP 1
FNU Any change in operating mode
ppm causes an automatic reset of
A1
Operation mode
selection
mg/l FNU A1 userdefined settings. The offsets
g/l for turbidity and temperature are
% Oper.Mode reset to zero.
spec.

Selection of unit kg/l


A2 for display (only % kg/l A2 A2 only active, if A1 = spec.
with Plus package) t/m3
none Conc.Unit
Selection of
display format for XX.xx
A3 concentration X.xxx XX.xx A3 A3 only active, if A1 = conc.
unit (only with Plus XXX.x
package)
XXXX Format

Factory settings are printed in bold face;


base version does not include functions in italic.

24 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Selection or range
Coding Field Display Info
Factory setting

A4
Connected sensor
is displayed
CUS31
CUS41
CUS31 A4 Liquisys M automatically detects
which sensor is connected.
Sensor
Measured value damping causes
averaging over the specified
number of individual measured
values.It is used, for example, to
A5
Enter measured
value damping
10 10 A5 stabilise the display with
1 ... 60 applications that fluctuate a great
Damping deal.
There is no damping if “1” is
entered.

5.2.2 Setup 2

Selection or range
Coding Field Display Info
Factory setting

Function group B
Initial display in function group
B SETUP 2.
SETUP 2
SETUP 2
If “auto” is selected, the wiper is
operated in combination with a
off
B1
Switch wiper
controller on or off on off B1 cleaning function via timer /
Chemoclean (“wipe and clean”).
auto
wiper In this case B2 and B3 are not
applicable.

s
B2
Enter operation
period of wiper
30 s 30 B2
3 ... 999 s
CleanTime

Enter pause time min


B3 between two 120 min 120 B3
wiping cycles 1 ... 7200 min
PauseTime

There are 3 calibration data sets


Selection of
B4 calibration data 3 3 B4 stored in every operation mode
(A1). Data set 1 cannot be
set to be used 1 ... 3
data set changed (see chap. 5.10).

The factory settings are printed in bold; functions printed in


italics are not available in the basic version.

Endress+Hauser 25
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

no Data set 1 cannot be changed


1 -> 2
B5 Copy data sets 1 -> 3 no B5 (factory setting). However, it can
be used as a basic for a
2 -> 3
3 -> 2 copy data customer calibration data set.

With the CUS 31 / 41: displays


Display measured
the measured value with or
B6
value with
reflection
yes yes B6 without assembly adaptation
no (reflection). Only effective in
adaptation? refl.adj. FNU, ppm, mg/l ranges).

Enter corrected
Current measured °C This entry can be used to
B7
temperature of the
temperature value 0.0 B7 calibrate the temperature sensor
–5.0 ... 100.0 °C to an external measurement.
sensor RealTemp

Enter temperature °C The difference between the


B8 difference
Current offset 0.0 B8 temperatures measured and
(offset) –5,0 ... 5,0 °C entered is displayed.
Temp.Offs

5.3 Current outputs


The function group CURRENT OUTPUT is used to configure the individual outputs. Either a
linear (O2 (1)) or, in conjunction with the Plus package, a user-defined current output
characteristic (O2 (3)) can be entered. Furthermore, a current output value can be simulated to
check the current outputs (O2 ( 2)).

Selection or range
Coding Field Display Info
Factory setting

Function group
O
Initial display in function group
O CURRENT CURRENT OUTPUT.
OUTPUT
OUTPUT

Select the operating mode /


O1
Selection of
current output
Out1 Out1 O1 characteristic separately for each
Out2 output.
Sel. Out
The characteristic can have a
positive or negative slope.
lin = linear (1)
O2 (1)
Enter linear
characteristic
sim = simulation (2) lin O2 See O2 (2), O2 (3) for the
Tab = Table (3)
Sel.Type simulation and table
characteristics.

Factory settings are printed in bold face;


base version does not include functions in italic.

26 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Selection or range
Coding Field Display Info
Factory setting

O211
Selection of
current range
4–20 mA 4-20 O211
0–20 mA
Sel.Range
0.0 FNU
0.0 ppm
0/4 mA value; 0.0 mg/l
Enter the measured value
enter 0.0 g/l
FNU corresponding to the minimum
O212
corresponding
turbidity or
0.0 kg/l 0.0 O212 current value (0/4 mA) at the
0.0 t/m3 transmitter output.
temperature
0.0 % 0/4 mA Display format from A3.
value
0.0 °C
entire measuring
range

9999 FNU
Enter the measured value
9999 ppm / 3000 ppm corresponding to the minimum
20 mA value; 9999 mg/l / 3000 mg/l current value (0/4 mA) at the
enter 300.0 g/l / 3 g/l transmitter output.
FNU
O213
corresponding
turbidity or
99.99 kg/l 9999 O213 Display format from A3.
99.99 t/m3
temperature
10.0 % 20 mA If two factory settings are given,
value then the left-hand one is valid for
100.0 °C CUS 41, the right-hand one for
entire measuring CUS 31.
range

lin = linear (1) The simulation is terminated by


O2 (2)
Current output is
simulated sim = simulation (2) lin O2 selecting (1) or (3).
See O2 (1), O2 (3) for other
Tab = Table (3)
Sel.Type characteristics.

mA
O221
Enter simulation
value
Current value 4.00 O221 The current value entered here is
output through the current output.
0.00 ... 22.00 mA
Simulat.
Values may also be added or
changed at a later point in time.
Enter current lin = linear (1) The values entered are auto-
O2 (3)
output table
(only with Plus
sim = simulation (2) table O2 matically sorted in ascending
order by current value The
Tab = Table (3)
package) Sel.Type display, however, is not in sorted
order. See O2 (1), O2 (2) for
other characteristics.

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 27
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

O231
Selection of table
option
Read read O231 Check (Read) or enter (Edit) a
characteristic.
Edit
Sel.Table

This is where the number of pairs


1
O232
Enter number of
table value pairs
1 O232 of x and y value (measured value
and associated current value) is
1 ... 10
No. Elem. entered.

The function chain O233 ... O235


1 will run though as many times as
O233
Select table value
pair
1 ... Number of table 1 O233 correspond to the value in O232.
“Finished” appears as the last
value pairs
Finished Sel.Elem. step. After confirmation the
system jumps to O236.

0.0 FNU
0.0 ppm
0.0 mg/l
0.0 g/l
FNU
O234
Enter x value
(measured value)
0.0 kg/l 0.0 O234 x value = measured value
determined by user.
0.0 t/m3
0.0 % Meas.val.
0.0 °C
entire measuring
range

y value = current value


mA
O235
Enter y value
(current value)
4.00 mA 0.00 O235 determined by user to be
associated wht measured value
0.00 ... 20.00 mA
mA value entered in O234.

Return to O2.
Message to say If no, then set table correctly
O236 whether table
yes yes O236 (all previous settings are kept) or
status is OK no back to measurement mode
Status ok (this makes the table invalid).

5.4 Monitoring functions


The monitoring functions are used to define various alarms and to set output contacts. Each
individual error can be defined to be effective or not (at the contact or as an error current). An
alarm condition can be defined to activate a cleaning function (F8).

Factory settings are printed in bold face;


base version does not include functions in italic.

28 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

5.4.1 Alarm

Selection or range
Coding Field Display Info
Factory setting

Function group F
F Alarm function settings.
ALARM
ALARM

Stead = steady
F1
Selection of
contact type
contact Stead F1 The contact type selected here
only applies to the alarm contact.
Fleet = fleeting
contact Cont.Type

F2
Selection of time
unit
s s F2
min
Time Unit

s Depending on the unit selected


F3 Enter alarm delay 0 s (min) 0 F3 in F2, the alarm delay in entered
0 ... 2000 s (min) in s or min.
Err.Delay
This selection must be made
even if all errors are suppressed
F4
Selection of error
current
22 mA 22mA F4 in F5.
2.4 mA If you did not select “0–20 mA” in
Err.Curr O211, you may not use “2.4 mA”.

All errors can be selected for


which an alarm message should
occur. Select using the fault
F5
Selection of error
number
1 1 F5 numbers. For individual error
numbers, please refer to the
1 ... 255
Sel.Error table in chap. 7. The factory
settings remain in effect for all
errors not edited.

yes If set to “no”, all the other alarm


no settings (e.g. alarm delay) are
Make alarm
F6 contact effective
From E080 factory
yes F6 also deactivated. The settings
themselves are retained.
for selected errors
setting Rel.Assg This setting only applies to the
no error selected in F5.

The error current selected in


Set error current F4 becomes effective or is
F7 to be effective for
no no F7 suppressed in case of error.
selected error yes
This setting only applies to the
Curr.Assg faults selected in F5.

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 29
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

Automatic start
of cleaning
F8 function? no no F8 This field does not exist for some
errors, see chap. 7.
(Only with Plus yes
package) CleanTrig

next = next error If <—- is selected. the display


F9
Return to menu or
select next error <—-R next F9 returns to F. If next is selected, the
display returns to F5.
Select

Factory settings are printed in bold face;


base version does not include functions in italic.

5.4.2 Check

PCS alarm (Process Check System)

This function is used to examine the


Constant measuring signal
= alarm is triggered measuring signal for deviations. If the
measuring signal is constant for a specific
period of time (several measured values),an
alarm is issued. This type of sensor behaviour
may be caused by soiling or similar.

Note:
t

Fig. 5.4 PCS alarm (live-check) LIVE.CDR


A current PCS alarm is automatically
deleted as soon as the sensor signal
changes.

Selection or range
Coding Field Display Info
Factory setting

Function group P
Settings for sensor and process
P monitoring.
CHECK
CHECK
This function is used to examine
the measuring signal. An alarm is
signalled if it does not change for
off the period selected here.
P1
Set PCS alarm
(live check)
1h off P1 Monitoring limit:
2h 0.3 % of the mean value across
4h PCS Alarm the set time period.
(Error no.: E152.)

Factory settings are printed in bold face;


base version does not include functions in italic.

30 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

5.5 Relay contact configuration


The relay contacts described below can be selected and configured as required (up to fouer
times with four contacts depending on options installed).

• Limit contactor for measured turbidity value R2 (1)


• Limit contactor for temperature: R2 (2)
• P(ID) controller: R2 (3)
• Timer for cleaning function: R2 (4)
• Chemoclean function R2 (5)

5.5.1 Limit contactor for measured turbidity value and temperature


The relay contacts in the Liquisys M can be When the measured value increases (max.
assigned different functions. function), the relay contact is closed at time t2
Switch-on and switch-off points and pickup when the switch-on point has been exceeded
and dropout delays can be defined for the (t1) and the pickup delay (t2 – t1) has expired.
limit contactor. Moreover, an alarm threshold When the alarm threshold (t3) is reached and
can be set to issue an error message and to the alarm delay (t4 – t3) also has expired, the
start a cleaning function. alarm contact is switched.
These functions may be used for turbidity and When the measured value decreases, the
temperature measurement. alarm contact is reopened when the
measured value drops below the alarm
Please refer to Fig. 5.5 for a graphic threshold (t5), and the relay contact also
representation of the alarm contact states of opens (t7, after the dropout delay t7 – t6 ).
any relay or alarm contact. When the pickup and dropout delays are set
to 0 s, the switch-on and switch-off points are
identical to the contact switching points.
Settings analogous to the max function can
also be made for a min function.

Switch-on point > switch-off point: maximum function


Measured value
Alarm ON
Alarm OFF
Alarm threshold
Contact ON

Switch-on point
Switch-off point
Contact OFF

t1 t2 t3 t4 t5 t6 t7 t

Switch-on point < switch-off point: minimum function

Contact OFF
Switch-on point
Switch-off point
Contact ON
Alarm OFF
Alarm threshold
Alarm ON

t1 t2 t3 t4 t5 t6 t7 t
Depiction of alarm
GRENZ-E.CDR Fig. 5.5 functions

Endress+Hauser 31
Instrument configuration Liquisys M CUM 223 / 253

5.5.2 P(ID) controller


The Liquisys M supports the definition of various controller functions. On the basis of the PID
controller, P, PI, PD and PID controllers can be implemented. To obtain the best control
response, use the controller best suited to the application in question.

• P controller: Used for simple linear control purposes with small system deviations. Where
major changes are to be controlled, overshooting may occur. A control offset is to be
expected.
• PI controller: Used for processes where overshooting is to be avoided and permanent
offsets are not allowed.
• PD controller: Used for processes that require quick response and where peaks are to be
corrected.
• PID controller: Used for processes for which the type of control provided by a P, PI or PD
controller is inadequate.

Start-up

• If there is no previous experience for the control parameter settings, use the values which
guarantee maximum control loop stability (see table).
• For optimisation, reduce the controller gain Kp until the control variable is slightly overswung.
• Increase Kp and then change Tn so that the shortest possible recovery time without
overswing is achieved. Adjust Tv in the case of fast correction times.

Adjustments of P(ID) controller


Three parameters can be adjusted in the
case of a PID controller:
y
100 %
On
• the control gain Kp (P influence)
yh
• the integral action time Tn (I influence)
• the derivative action time Tv (D influence)
0%
Off
t Step response of process
Xmax y = set value
[%]
[°C] yh = control range
Tu = delay time [s]
Tg = recovery time [s]
X max ∆x
Vmax = =
Tg ∆t
∆x = max. rate of rise
the control parameter [K/s]
Xh ∆t Xmax = maximum process value
Xh = adjustment range of controller

Controller characteristics
Tu Tg
Vmax
K= ⋅Tu
Xh
Fig. 5.6 Control characteristic REGEL01.CDR

Recommendable settings for all types


Contr. response Kp Tv [min] Tn [min]

P 1·K 0* 0*

PI 2.6·K 0* 6 Tu

PD 0.5 K· Tu 0*

PID 1.7 K· 2 Tu 2 Tu

Tv = 0: component not calculated


Tn = 0: component not calculated

32 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Checking and fine tuning of parameter settings using a recorder

Actual Actual
value

Tn too big
Tn too small

Time Time

Actual Actual
value

Kp too big
Kp too small

Time Time

Actual

Optimal setting

Time Setting optimisation


REGEL02.CDR Fig. 5.7 Tn and Kp

Actuating signal outputs (R237 ... R2310)

The control contact in question outputs a switched signal. The intensity of this signal is
proportional to the controller’s output. A distinction is made according to the type of signal
clock:

• Pulse length modulation • Pulse frequency modulation


The greater the calculated control output, The greater the calculated control output,
the longer the contact in question remains the higher the switching frequency of the
picked up. The period can be adjusted contact. The maximum switching
between 0.5 and 99 s. Pulse-length frequency 1/T can be adjusted between 60
modulated outputs are used to control and 180 min–1. The ON period tON is
solenoid valves. constant. Pulse frequency-modulated
outputs are intended for controlling
solenoid-operated metering pump s.

Contact IMPULS.CDR
Contact
tON
tOFF tON
On On

Off Off Signal of a pulse length-


modulated (left) and a
Period T Time [s] T1 T2 Time pulse frequency-
modulated control contact
Fig. 5.8 (right)

Endress+Hauser 33
Instrument configuration Liquisys M CUM 223 / 253

Control characteristic for direct and inverse control action

Field R236 offers two control characteristics for selection which have the effects shown in the
following diagram.

100 %

direct =max
function

– Xw Xw

Y
100 %
inverse =min
function

Control characteristic of a
proportional-action
controller with direct and –Xw Xw
Fig. 5.9 inverse control action REGEL03.CDR

5.5.3 Timer for cleaning function


This function can be used to implement a Note:
simple cleaning routine. The user can specify
a time interval after which cleaning is to start; The timer and Chemoclean do not
i.e. only constant intervals can be defined. work independently of each other.
Whilst one of the functions is active,
There are other cleaning functions in the other cannot be started.
conjunction with the Chemoclean function
(version with four contacts, see chap. 5.5.4).

Wiper and / or spray cleaning


system Cleaning time Pause time between
(0 ... 999 s) 2 cleaning intervals
active (1 ... 7200 min)

inactive

Hold function Clean-Hold


delay time
(0 ... 999 s)
active

inactive
Relationship between Time
cleaning time, pause time
Fig. 5.10 and hold delay time DIAGR-TI.CDR

34 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

5.5.4 Chemoclean function


Just like the timer function, Chemoclean can Note:
also be used to start a cleaning cycle.
However, Chemoclean supports different
cleaning and rinse intervals. • The Chemoclean function is only
available with relays 3 (water) and
Thus, irregular cleaning with different repeat 4 (cleaner).
cycles is possible, and cleaning times with
post-rinse times can be individually defined. • Abortion of the cleaning process is
always followed by a post-rinse
time.

• When “Economy” is selected,


cleaning is performed with water
only.

Hold

t1 Cleaning trigger
Water t2 t1 Pre-rinse time
(relay 3) t3 t2 Cleaning time
t4 t3 Post-rinse time
t5 t4 Hold dwell period

Cleaner
(relay 4)

t0 t1 t2 t3 t4 t5 Sequence of a cleaning
CHEMO.CDR Fig. 5.11 cycle

Selection or range
Coding Field Display Info
Factory setting

Function group R
Relay contacts can be selected
R and adjusted.
RELAY
RELAY
Rel3 (water) and Rel4 (cleaner)
Rel1 are only available on transmitter
Selection of
R1 contact to be Rel2 Rel1 R1 equipped accordingly.
If Chemoclean is selected as the
configured Rel3
Rel4 Sel.Relay cleaning type, Rel4 is not
available.
PV = Process value
LC PV = Limit
contactor TU (1) If Rel4 is selected in Field R1,
Configuration limit
LC °C = Limit Clean = Chemoclean cannot be
R2 (1)
contactor for
turbidity contactor T (2) LC PV R2 selected.
PID controller (3) Confirmation with ENTER
measurement
Timer (4) Sel.Type switches off a different, already
Clean = Chemoclean (5) switched-on function and its
settings are reset to the default.

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 35
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

R211
Switch function of
R2 (1) off or on
off off R211 All settings are retained.
on
Function
9999 FNU
Never set the switch-on point and
9999 ppm / 3000 ppm switch-off point to the same value.
9999 mg/l / 3000 mg/l (Comp. chap. 5.5.1.)
300.0 g/l / 3 g/l FNU (Only the operating mode
R212
Enter switch-on
point of contact. 99.99 kg/l 9999 R212 selected in A1 appears.)
If two factory settings are given,
99.99 t/m3 On value then the left-hand one is valid for
200.0 % CUS 41, the right-hand one for
entire measuring CUS 31.
range
Entering the switch-off point
9999 FNU
selects either a max contact
9999 ppm / 3000 ppm (switch-off point < switch-on
9999 mg/l / 3000 mg/l point) or a min contact (switch-off
300.0 g/l / 3 g/l FNU point > switch-on point) and
R213
Enter switch-off
point of contact 99.99 kg/l 9999 R213 implements a constantly required
hysteresis (see Fig. 5.5).
99.99 t/m3 Off value If two factory settings are given,
200.0 % then the left-hand one is valid for
entire measuring CUS 41, the right-hand one for
range CUS 31.

s
R214 Enter pickup delay 0s 0 R214
0 ... 2000 s
On Delay

s
R215
Enter dropout
delay
0s 0 R215
0 ... 2000 s
Off Delay
When the alarm threshold is
exceeded/undershot, the
measuring transmitter issues an
9999 FNU
alarm with an error message and
9999 ppm / 3000 ppm error current (note alarm delay).
9999 mg/l / 3000 mg/l
Enter alarm
300.0 g/l / 3 g/l FNU When defining the min contact,
R216
threshold
(as an absolute 99.99 kg/l 9999 R216 the alarm threshold must be set
to a lower value than the
value) 99.99 t/m3 A.Thresh switch-off point.
200.0 %
entire measuring If two factory settings are given,
range then the left-hand one is valid for
CUS 41, the right-hand one for
CUS 31.

The factory seFactory settings are printed in bold face;


base version does not include functions in italic.

36 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Selection or range
Coding Field Display Info
Factory setting
LC PV = Limit
contactor TU (1)
Configure limit LC °C = Limit Confirmation with ENTER
R2 (2)
contactor for
temperature
contactor T (2) LC °C R2 switches off a different, already
switched-on function and its
PID = PID controller
measurement (3) Sel.Type settings are reset to the default.
Timer (4)
Clean = Chemoclean (5)

R221
Switch function of
R2 (2) off or on
off off R221
on
Function

100.0 R222
°C
Never set switch-on point and
R222 80C6144%5 ... 10.0 °C80C6144%5 switch-off point to the same value!
On value 80C
(Comp. chap. 5.5.1.)80C6144%5

6144%580C6144%5

Entering the switch-off point


selects either a max contact
(switch-off point < switch-on
100.0 R223
°C
Enter switch-off 100.0 C
R223 point) or a min contact (switch-off
temperature 5.0 ... 100.0 C point > switch-on point) and a
Off value constantly required hysteresis is
implemented (see Fig. 5.5.)

0 R224
s
R224 Enter pickup delay 0s
0 ... 2000 s
On Delay

0 R225
s
Enter dropout 0s
R225
delay 0 ... 2000 s
Off Delay
When the alarm threshold is
exceeded/undershot, the
transmitter issues an alarm with
Enter alarm
an error message and error
100.0 R226
°C
threshold 150.0 °C
R226 current (note alarm delay).
(as an absolute –20.0 ... 150.0 °C
When defining the min contact,
value)
A.Thresh the alarm threshold must be set
to a lower value than the
switch-off point.
LC PV = Limit
contactor TU (1)
Confirmation with ENTER
LC °C = Limit
R2 (3)
P(ID) controller
configuration
contactor T (2) PID R2 switches off a different, already
switched-on function and its
PID controller (3)
Timer (4) Sel.Type settings are reset to the default.
Clean = Chemoclean (5)

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 37
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

R231
Switch function of
R2 (3) off or on
off off R231
on
Function
0 FNU
0 ppm
0 mg/l The set point is the value to be
0.0 g/l FNU maintained by the control. This
R232 Enter set point 0.0 kg/l 0 R232 value is to be re-establish by the
control in the event of any
0.0 t/m3 Setpoint deviation (up or down).
0.0 %
entire measuring
range

R233
Enter control gain
Kp
1.00 1.00 R233 See chap. 5.5.2, p. 32.
0.01 ... 20.00
Kp

See chap. 5.5.2.


Enter integral
min Each hold resets the I
R234
action time Tn
(0.0 = no I
0.0 min 0.0 R234 component to zero. Hold can be
0.0 ... 999.9 min deactivated in S2, but not for
component)
Time Tn Chemoclean and timer or wiper!

Enter derivative
min
R235
action time Tv
(0.0 = no D
0.0 min 0.0 R235 See chap. 5.5.2.
0.0 ... 999.9 min
component)
Time TD

Setting may or may not be


Selection of
R236 controller
dir = direct dir R236
required depending on control
deviation (up or down deviation,
characteristic inv = inverted
Direction see chap. 5.5.2).

Pulse length e.g. for solenoid


Selection of pulse
R237 length or pulse
len = Pulse length len R237
valve, pulse frequency e.g. for
solenoid-operated metering
frequency freq = Pulse frequency
Oper.Mode pump (see chap. 5.5.2).

This field only appears if pulse


s

R238 Enter pulse interval 10.0 s 10.0 R238 length is selected in R237. When
pulse frequency is selected,
0.5 ... 999.9 s
PulsePer. R238 is skipped and input
continues in R239.

The factory settingFactory settings are printed in bold face;


base version does not include functions in italic.

38 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5A.CHP

Selection or range
Coding Field Display Info
Factory setting

This field only appears if pulse


Enter maximum 1/min frequency is selected in R237.
R239 pulse frequency
120 min–1 120 R239 When pulse length is selected,
of actuator 60 ... 180 min–1 R239 is skipped, and input
Max.PFreq continues in R2310.

s
R2310
Minimum ON time
tON
0.3 s 0.3 R2310
0.1 ... 5.0 s
Min.PTime
Cleaning with only one cleaning
LC PV = Limit agent (usually water) and poss.
contactor TU (1) as Wipe-Wash function together
Configure LC °C = Limit with the wiper (see Fig. 5.10 and
R2 (4) cleaning function contactor T (2) Timer R2 setting in B1).
(Timer) PID controller (3) Confirmation with ENTER
Timer (4) Sel.Type switches off a different, already
Clean = Chemoclean (5) switched-on function and its
settings are reset to the default.

R241
Switch function of
R2 (4) off or on
off off R241
on
Function

s The hold and relay settings are


R242
Enter rinse /
cleaning time
30 s 30 R242 activated for the period of time
0 ... 999 s specified here.
RinseTime

min The pause time is the time


R243 Enter pause time 360 min 360 R343 between two cleaning cycles
1 ... 7200 min (see chap. 5.5.4).
PauseTime

The minimum pause time


min
R244
Enter minimum
pause time
120 min 120 R244 prevents continuous cleaning
when the cleaning trigger is
1 ... 3600 min
Min.Pause present.

LC PV = Limit
Configure contactor cond. (1) See chap. 5.5.4.
Chemoclean LC °C = Limit
function (on contactor T (2) Contact 3 = Water,
R2 (5)
version with four
contacts and
PID = PID controller Clean R2 Contact 4 = Cleaner.
Confirmation with ENTER
(3)
appropriate
assignment of
Timer (4) Sel.Type switches off a different, already
switched-on function and its
Clean = Chemoclean
contacts 3 and 4) settings are reset to the default.
(5)

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 39
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

R251
Switch function of
R2 (5) off or on
off off R251
on
Function
int = internal
(timer-contr.)
ext = external There is no real-time clock.
R252
Selection of start
pulse
(digital input 2) int R252
External suppression is required
for irregular time intervals (e.g.
i+ext = intern. + extern.
i+stp = internal, CleanTrig weekends).
suppressed
by external

s
R253
Enter pre-rinse
time
20 s 20 R253 Water is used for rinsing.
0 ... 999 s
PreRinse

s
R254
Enter cleaning
time
10 s 10 R254 Cleaning agent and water are
used for cleaning.
0 ... 999 s
CleanTime

s
R255
Enter post-rinse
time
20 s 20 R255 Water is used for rinsing.
0 ... 999 s
PostRinse

R256
Enter number of
repeat cycles
0 0 R256 R253 ... R255 is repeated.
0 ... 5
Rep.Rate

min
R257 Enter pause time 360 min 360 R257 The pause time is the time
between two cleaning cycles.
1 ... 7200 min
PauseTime

The minimum pause time


min
R258
Enter minimum
pause time
120 min 120 R258 prevents continuous cleaning
when the external cleaning
1 ... R 257
Min.Pause trigger is present.

Enter number of Cleaning with cleaner can be


cleaning cycles followed by aup to 9 cleaning
R259 without cleaning 0 0 R259 cycles that use only water until
agent (economy 0 ... 9 the next cleanig cycle with
function) EconomyCl cleaner is performed.

The factory settings are functions in italic. face;


base version does not include functions in italic.

40 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

5.6 Concentration measurement


The Liquisys M can convert turbidity values to
concentration values as standard. Since not
all possible units or the corresponding [kg/l]
conversions can be made available, the user
can create a custom conversion table.

Any assignment of a measured value to a 1,4


converted value is possible (see Fig. 5.12).
The defined values are connected by a curve
based on linear interpolation, i.e. the
dinstances between the points should not be
too large.
1,0
0 500 [g/l]
Note:
Assignment of measured
U-UMRECH.CDR Fig. 5.12 value to a converted value
The conversion is exclusively related
to the calibration operating mode %.

Selection or range
Coding Field Display Info
Factory setting

Four different concentration


Function group
K K curves can be entered into this
CONCENTRATION function group.
CONCENTRA
Selection of
concentration
The curves are independent of
K1
curve, which will
be the basis
1 1 K1 each other. Therefore, you can
1 ... 4 define four different curves.
of the display act.Curve
value calculation

The value pairs entered is


independent of the curve used
K2
Selection of table,
you want to edit
1 1 K2 (K1). Only when entry is
1 ... 4 complete is the saved curve
editCurve overwritten.

This selection is valid for the


K3
Selection of table
option
Read read K3 concentration curve selected in
Edit K2.
Table

K4
Enter number of
triplets
1 1 K4 Each triplet consists of a numeric
pair.
1 ... 10
No. Elem.

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 41
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

1
K5 Selection of triplet 1 ... Number of support 1 K5 Any triplet can be edited.
points from K4
Fin Sel.Elem.
The function chain K5 ... K7 is
run through as many times as is
0.0 % %
K6
Enter turbidity
value
Entire measuring 0 K6 equal to the value in K4.
“Finished” appears as the last
range
MeasValue step. After confirmation the
system jumps to K8.

Enter kg/l
K7 concentration
Entire measuring
range
0 K7 Measurement unit as in A2.
value
concentr.
Return to K.
Enter whether or
K9 not table status is yes yes K8 If not, then set table correctly (all
previous settings are kept) or
no
okay
Status ok back to measurement mode (this
makes the table invalid).

5.7 Service
Selection or range
Coding Field Display Info
Factory setting

Function group S
S
SERVICE
SERVICE
ENG = English
GER = Deutsch This field must be set once
S1
Selection of
language
FRA = Francais ENG S1 during device configuration. After
ITA = Italiano confirmation with ENTER
NEL = Nederlands Language
ESP = Espanol

S+C = during
setup and

S2 Hold configuration
calibration
CAL = during S+C S2
S = setup,
C = calibration.
calibration
Setup = during setup Auto HOLD
none = no hold

Factory settings are printed in bold face;


base version does not include functions in italic.

42 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

Selection or range
Coding Field Display Info
Factory setting

S3 Manual hold off aus S3


The setting remains active even
after a power failure.
on
Man. HOLD

s
S4
Enter hold dwell
period
10 s 10 S4
0 ... 999 s
Cont.Time
The code is located on the
nameplate (see Figs. 1.1 1 and
Enter SW upgrade 1.2). Entry of an incorrect code
S5 release code
0000 0000 S5 returns you to the measurement
(Plus package) 0000 ... 9999 menu. The number is edited with
PlusCode the PLUS or MINUS key and
confirmed with the ENTER key.
The code is located on the
nameplate (see Figs. 1.1 1 and
Enter SW upgrade 1.2). Entry of an incorrect code
S6 release code
0000 0000 S6 returns you to the measurement
Chemoclean 0000 ... 9999 menu. The number is edited with
CleanCode the PLUS or MINUS key and
confirmed with the ENTER key.

S7
Order number is
displayed
order S7 When you upgrade the device

TU0005

S8
Serial number is
displayed
SerNo S8

12345678
Reset of
instrument Facty= All data except
(reset to default no language (field S1)
S9 values) Sens = Sensor data no S9
are erased and
reset to the
Facty= Factory
settings S.Default factory setting!
Sens = no function

S10
Perform
instrument test
no no S10
displ = display test
Test

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 43
Instrument configuration Liquisys M CUM 223 / 253

5.8 E+H Service


Selection or range
Coding Field Display Info
Factory setting

Function group
E E
E+H SERVICE
E+H SERV

Contr = controller (1)


E1
Selection of
module
Trans = transmitter (2) Contr E1
Main = power unit (3)
Rel = relay (4) Select

E111
E121
E131
Software version
is displayed
xx.xx E111 This field cannot be edited.
E141
SW-Vers.

E112
E122
E132
Hardware version
is displayed
xx.xx E112 This field cannot be edited.
E142
HW-Vers.

E113
E123
E133
Serial number is
displayed
SerNo E113 This field cannot be edited.
E143
12345678

E114
E124
E134
Module name is
displayed
LSG E114 This field cannot be edited.
E144
Modul-ID

Factory settings are printed in bold face;


base version does not include functions in italic.

44 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

5.9 Interfaces
Selection or range
Coding Field Display Info
Factory setting

Function group I
I
INTERFACE
INTERFACE

Address

I1 Enter address
HART: 0 ... 15
or
126 I1 For communication only.

PROFIBUS: 1 ... 126 Address

The factory settings are printed in bold; functions printed in


italics are not available in the basic version.

Endress+Hauser 45
Instrument configuration Liquisys M CUM 223 / 253

5.10 Calibration
This function group is used to calibrate the Note:
transmitter.

Calibration options
• If the calibration procedure is
aborted by pressing the PLUS and
• Three-point calibration MINUS keys at the same time
• Correction of an existing calibration with (return to C115, C123, C135, C143
the laboratory value or C153) or if the calibration is
• Modification of individual values of the faulty, then the previous
three-point calibration calibration data are reinstated. A
• Installation adjustment calibration error is indicated by the
• One-point calibration “ERR” message and flashing of
the electrode symbol on the
display. Repeat calibration!

• After calibration, the system


returns to Measuring mode.
During the hold delay time (Field
S4), the Hold symbol appears in
the display.

Overview of calibration

The calibration of the meeasuring chain is concentration graduation for the three-point
three-point calibration, i.e., a turbidity calibration. This method also has the
ascertained from three samples or from a advantage the only the turbidity or solids
known solids content of the medium is used concentration value of the original laboratory
to calculate the complete calibration sample need be defined exactly.
characteristic of the measuring chain for the
medium of the transmitter CUM 223 / 253. Note:

It is wise to undertake calibration in the


In particular, the concentration
turbidity or solids concentration range in
graduation must changed for
which you want to make measurements.
samples with a very high
concentration or a very dark colour.
To simplify the three-point calibration, we
Very good results are achieved with
suggest to produce three samples required
a graduation of 3%, 10% and 100%.
for calibration by thinning a medium sample.
The original sample must also be
Usually, very good calibration results are
measured in the laboratory for
achieved with the concentration graduations
results correction after the laboratory
10 %, 33 % and 100 % of the original
measurement.
concentration. The transmitter suggests this

Original sample

Calibration value 3
9 parts zero 2 parts zero
solution solution
+ +
1 part original 1 part original
sample sample
Sample 3
original sample
Sample 1 Sample 2
Creation of samples for a original sample × 1/10 original sample × 1/3
Fig. 5.13 three-point calibration UM_3PROB.CDR

46 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

However, this procedure is not a mandatory • Calibration must be carried out relating to
prerequisite for calibration. You can also take the concentration in increasing order.
three different samples of the process • The three samples must differ by at least
medium and determine their turbidity or solids 10 % relating to the measured value.
concentration values. The following conditions • At higher solids concentrations, clean
must be met: water is sufficient as a zero solution.

Note:

• Sludge samples tend to sediment. • If the calculated relative


It is therefore important to stir the calibration results deviate by more
sample well, possible even during than –50 % / +100 % from the
calibration. When stirring, avoid reference value 100 %, a warning
forming a cyclone by not stirring in is output (E084). The calibration
a circle. can be transferred nonetheless.

• The characteristic calculated • If the calculated results lie outside


during the three-point calibration the permitted limits, then the
is stored in the currently selected calibration error is displayed
calibration data record (see Field (E045). The calibration is not
B4). If you select the transferred.
unchangeable (read only)
calibration data record no. 1, no
calibration is possible.

Management of the transmitter calibration data records

There are three calibration data records The other two are stored at the factory as
available for each application (FNU, ppm, g/l, standard characteristics. You can change
%). Of these three data records (sample and adapt these (three-point calibration,
characteristics), one has been permanently one-point adjustment, installation calibration).
programmed at the factory (read only).

FNU ppm or mg/l g/l %


Formazine SiO2 Activated sludge Residual
Data set no. 1 concrete water
(read only) (read only) (read only) (read only)
Formazine Kaolin Activated sludge Residual
Data set no. 2 concrete water
(Changeable) (Changeable) (Changeable) (Changeable)
Formazine SiO2 Activated sludge Residual
Data set no. 3 concrete water
(Changeable) (Changeable) (Changeable) (Changeable)

Note:

The data records are saved in the


sensor. If you replace the
transmitter, this data is not lost. It is
loaded automatically (according to
the selection) into the transmitter.

Endress+Hauser 47
Instrument configuration Liquisys M CUM 223 / 253

Version TB: Initial settings for residual concrete water

The version Liquisys M CUM 223-TB and However, should you reset the device to the
CUM 253-TB contains the Plus package with original factory settings (set default), then you
additional setting options, as for the TS can find the values which are valid for this
version. application.

On delivery, different values from the Note:


standard version are set in the individual
fields of the menus to make commissioning as
Should you wish to use a voltage
simple as possible.
output instead of the current output
intended by the factory, then please
These initial settings are so selected so that
connected the supplied shunt
no additional settings need be made for
resistor according to the instructions
applications in residual concrete water. on the sheet enclosed with the
transmitter. You will then obtain a
0 ... 10 V current output.

Factory settings

Menu field Setting


Mode of operation A1 spec.
Engineering unit A2 kg/l
Display format A3 XX.xx
Measured value damping A5 10
Calibration data record B4 3
Current output O1 Out1
Characteristic O2 lin
Current range O211 0–20 mA
Measured value 0 mA O212 1.00
Measured value 20 mA O213 1.30
Active characteristic K1 1
Edited characteristic K2 1
Table option K3 edit
Number of value pairs K4 2
Support point K5 1 ... 2
Measured value / display value K6 / K7 1: 0 % / 1.00 2: 50 % / 1.50
Language version S1 GER

Note:

Calibration is carried out in the %


measuring range (the transmitter
switches automatically). With the
above setting, a density of e.g. 1.12
kg/l = 12 % must be selected.
In individual cases, it may be
necessary to adapt the calibration to
a real sample. For this, a “single-
point calibration” is sufficient.

48 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

Selection or range
Coding Field Display Info
Factory setting

Function group C
C
CALIBRATION
CALIBRAT
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit For data set 1 (B4), only the
calibration (3)
C1 (1)
Selection of
calibration Refl = Fitting 3-Pt C1 “Data” function is accessible.
The offset is reset with 3 Pt
with reflection
effects (4) Calibrat and Edit.

1-Pt = Single-point
calibration (5)
Data = Calibration
data (6)

Immerse the sensor so that there


Immerse sensor in the calibration solution (sample 1). is sufficient distance to the tank
wall (no reflection).

Enter concentra- FNU


C111 tion of the first
Value from last
100.0 C111
calibration solution calibration
Concentr1

Immerse the sensor so that there


Immerse sensor in the calibration solution (sample 2). is sufficient distance to the tank
wall (no reflection).

Enter concentra- FNU


C112 tion of second
Value from last
calibration
330.0 C112 C112 ≥ 1.1 × C111
calibration solution
Concentr2

Immerse the sensor so that there


Immerse sensor in the calibration solution (sample 3 =
is sufficient distance to the tank
original sample) .
wall (no reflection).

Enter concentra-
FNU
C113
tion of the third
calibration
Value from last
calibration
1000.0 C113 C113 ≥ 1.1 × C112
solution Concentr3

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 49
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting
Cancel Warning: Warning: Cancel
E045 E045
C114
Calibration status
is displayed
o.k.
E xxx
o.k. C114 E084 E084

Status
20 % 50 % 200 % 500 %
C161 ... C163

If C114 = E xxx, then only no or


new (Exception: calibration
yes
C115
Store calibration
results no yes C115 warning E 84).
If new, return to C.
new
Store If yes/no, return to
“Measurement”.
3 Pt = Three-point
calibration (1)
Corr= Three-point
correction (2)
Edit = Edit

calibration (3)

C1 (2)
Selection of
calibration
Refl = Fitting Corr C1
with reflection

effects (4)
Calibrat
1-Pt = Single-point

calibration (5)

Data =Calibration

data (6)

Enter correct
Current value from In case of an unknown probe
concentration of FNU
C121 the third C113 1000.0 C121 concentration, but with a definite
dilution (1/10; 1/3; 1), the
calibration entire measuring
solution range Concentr3 laboratory value is to be entered .

C122
Calibration status
is displayed
o.k.
E xxx
o.k. C122
Status
If C122 = E xxx, then only no or
new (Exception: calibration
yes
C123
Store calibration
results no yes C123 warning E 84).
If new, return to C.
new
Store If yes/no, return to
“Measurement”.

Factory settings are printed in bold face;


base version does not include functions in italic.

50 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

Selection or range
Coding Field Display Info
Factory setting
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
If the individual samples were
C1 (3)
Selection of
calibration
Refl = Fitting Edit C1 defined in the laboratory, you can
with reflection correct the calibration here.
effects (4) Calibrat
1-Pt = Single-point

calibration (5)

Data =Calibration

data (6)

Enter concentra- Current value from


FNU
C131
tion of the first
calibration
C111 100.0 C131
entire measuring
solution range Concentr1

Enter concentra- Current value from FNU


C132 tion of the second C112 330.0 C132
calibration C132 ≥ 1.1 ⋅ C131
Concentr2

Enter concentra- Current value from FNU


C133
tion of the third
calibration
C113 1000.0 C133
C133 ≥ 1.1 ⋅ C132
solution Concentr3

C134
Calibration status
is displayed
o.k.
E xxx
o.k. C134
Status
If C134 = E xxx, then only no or
new (Exception: calibration
yes
C135
Store calibration
results no yes C135 warning E 84).
If new, return to C.
new
Store If yes/no, return to
“Measurement”.
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3) Only for solutions ≤ 2 FNU /
C1 (4)
Selection of
calibration
Refl = Fitting Refl C1 5 ppm! There is no effect of the
with reflection installation in clear media
effects (4)
Calibrat
1 Pt = Single-point
calibration (5)
Data =Calibration
data (6)
Factory settings are printed in bold face;
base version does not include functions in italic.

Endress+Hauser 51
Instrument configuration Liquisys M CUM 223 / 253

Selection or range
Coding Field Display Info
Factory setting

0.0 FNU
0.0 ... 2.0 FNU
FNU
C141
Enter correct
measured value
0.0 ppm 0.0 C141 Only for the ranges
FNU, ppm, mg/l.
0.0 ... 5.0 ppm
0.0 mg/l Real PV
0.0 ... 5.0 mg/l

C142
Calibration status
is displayed
o.k.
E xxx
o.k. C142
Status
If C142 = E xxx, then only no or
new (Exception: calibration
yes
C143
Store calibration
results no yes C143 warning E 84).
If new, return to C.
new
Store If yes/no, return to
“Measurement”.
3 Pt = Three-point
calibration (1) For FNU:
Corr = Three-point Adapter C164, C165.
correction (2)
Edit = Edit For ppm, mg/l:
calibration (3) up to 500 adapter C164, C165,
C1 (5) Select calibration Refl = Fitting 1-Pt C1 then adapter C166.
with reflection
effects (4) Calibrat For g/l, %:
adapter C166. An existing basic
1-Pt = Single-point
calibration (three-point) is
calibration (5) corrected by the single-point
Data= Calibration calibration.
data (6)

Current measured
FNU
C151
Enter current
calibration value
value 0.0 C151
Complete measuring
range Real PV

C152
Calibration status
is displayed
o.k.
E xxx
o.k. C152
Status
If C152 = E xxx, then only no or
new (Exception: calibration
yes
C153
Store calibration
results no yes C153 warning E 84).
If new, return to C.
new
Store If yes/no, return to
“Measurement”.

Factory settings are printed in bold face;


base version does not include functions in italic.

52 Endress+Hauser
Liquisys M CUM 223 / 253 Instrument configuration

UM253E5B.CHP

Selection or range
Coding Field Display Info
Factory setting
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
C1 (6)
Selection of
calibration
Refl = Fitting Data C1
with reflection
effects (4) Calibrat
1-Pt = Single-point
calibration (5)
Data =Calibration
data (6)

%
C161
Calibration point 1
is displayed
Comparison value 101.4 C161 Deviation relative to standard
sensor (= 100 %)
Concentr1

%
C161
Calibration point 2
is displayed
Comparison value 99.3 C162 Deviation relative to standard
sensor (= 100 %)
Concentr2

%
C161
Calibration point 3
is displayed
Comparison value 98.7 C163 Deviation relative to standard
sensor (= 100 %)
Concentr3

C164
Slope 1
is displayed
Current value 230 C164 Slope of characteristic 1
of the sensor
Slope 1

C164
Slope 2
is displayed
Current value 375 C165 Slope of characteristic 2
of the sensor
Slope 2

Conversion factor of
C166
Conversion factor
is displayed
Current value 1 C166 internal turbidity units
into the displayed unit
ConvFact

Factory settings are printed in bold face;


base version does not include functions in italic.

Endress+Hauser 53
Instrument configuration Liquisys M CUM 223 / 253

5.11 Offset
The settings in the OFFSET function group
can be used to calibrate the measurement to

Measured value
Calibrated value a reference measurement. This requires a
linear shift of all measured values, i.e., the
adjustment is determined for one measured
value, and all the others are calculated using
the same offset.

Current measured value

Time t

OFFSET.CDR
Fig. 5.14 Offset

Selection or range
Coding Field Display Info
Factory setting

Function group V
V
OFFSET
OFFSET

0.0 V1
FNU
Enter absolute Current measured
V1
value value
Real PV

0.0 V2
FNU
V2 Enter offset Current offset

PV Offset

V3
Calibration status
is displayed
o.k.
E xxx
o.k. V3
Status

If V3 = E xxx, then only no or new.


yes
V4 Store offset no yes V4 If new, return to V.
If yes/no, return to
new
Store “Measurement”.

Factory settings are printed in bold face;


base version does not include functions in italic.

54 Endress+Hauser
Liquisys M CUM 223 / 253 Interfaces

UM253E06.CHP

6 Interfaces
For instruments equipped with a
communication interface, please refer to
separate operating instructions
BA 208C/07/en (HART®) or BA 209C/07/en
(PROFIBUS®).

Endress+Hauser 55
Maintenance and troubleshooting Liquisys M CUM 223 / 253

7 Maintenance and troubleshooting


Maintenance:

Please take all measures in due time which will guarantee the safety of operation and reliability
of the entire measuring system.

Maintenance of CLM 223 / 253 comprises:


• Calibration (see chap. 5.10)
• Cleaning of assembly and sensor
• Checking of cables and connections

Troubleshooting
Determination and elimination of the cause of the problem in the case of an operating fault.
Troubleshooting refers to measures that can be performed without intervention in the instrument
(for instrument defects, see chap. 8, Repairs). Troubleshooting of the CLM 223/253 and the
measuring system is performed with the aid of the troubleshooting table in chap. 7.1.

Warning: Note:

• When working on the device, bear • Please contact you E+H support
in mind the possible office if you have any queries. You
repercussions on the process also send your queries to the E+H
controller or the process itself. Service Organization via the
Internet. www.endress.com
• When removing the sensor during
maintenance or calibration, please
consider potential hazards due to
pressure, high temperatures and
contamination.

7.1 Troubleshooting common problems


Equipment need, spare parts,
Error Possible cause Remedy
information
Instrument cannot be – Operation locked Press simultaneously the CAL and See chap. 4.3
operated, value 9999 MINUS keys
– Sensor / sensor cable defective Test with new or with different CUS 31 or 41 sensor (either type is
functional sensor suitable for rough function test)
– Instrument input defective Replace MKT1 module for testing S. spare parts list in chap. 8.2.4/8.3.4
Display value 0.0,
and error message – Break in sensor extension line Check junction boxes and line. Measuring point test in chap. 8.6.1
E008 “Sensor” Check sensor supply.
and display —— in A4 – Data transfer error Replace LSGA (AC power supply See spare parts list in chap. 8.2.4 /
unit or L2GD (DC power supply unit) 8.3.4
module for testing
– Incorrect sensor connection Check connection Connection see chap. 3.4 and 3.5
Display value 0.0 – Sensor completely blocked Clean optics Use spray cleaning or wiper
– Impermissible instrument Switch instrument off and back on EMV problem: check line routing if
Fixed, incorrect
operating state (no response to problem persists, check for
measured value
key actuation) possible sources of interference.
– Measuring cable interference Connect cable screen acc. to See chap. 3.4 and 3.5
connection diagram (not
grounding)
– Signal output line interference – check line routing, Separate signal output, meas.
Measured value – try separate line routing, input and supply lines
fluctuates grounding screen to SPS/PLS
– Irregular flow rate / turbulence / Choose better place of installation
air bubbles / big solids particles or eliminate turbulences.
Possibly use large measured value Measured value damping see
damping factor. field A5

56 Endress+Hauser
Liquisys M CUM 223 / 253 Maintenance and troubleshooting

UM253E07.CHP

Equipment need, spare parts,


Error Possible cause Remedy
information
– No or incorrect sensor calibration Calibration with original sample See chap. 5.10
required for concentration or solids
concentration
– Sensor soiled – Clean sensor – Remove coarse coats mit brush
– Remove carbonates and similar
coats with 3% hydrochlorid acid
– Remove organic coats and
Display value grease with oxidation agents
implausible / no or and / or grease solvents.
creeping change of – Use spray cleaning See instructions of assembly used
display for spray cleaning
– User wiper version Retrofitting possible: “wiper
retrofitting” service kit 50089254
– Wiper rubber defective Replace wiper arm “Wiper arm” service kit 50089252
– Sensor installed in “dead” zone Check installation conditions, move
or air cushion in assembly or sensor in area with optional flow
flange conditions. Caution if installed in
horizontal lines!
– Incorrect sensor orientation Orient sensor:
– Measuring surface should face Frontal “bombardment” of
flow in normal media. measuring surface with highly
– Orient meas. surfacre at 90° to viscous solids may result in an
flow in media with high solids adhering coating.
concentration
Incorrect temperature – Temperature sensor defective If temperature display is required: Turbidity measurement itself does
value replace sensor not require temperature meas.
Controller or timer – No relay module installed Install LSR1-2 or LSR1-4 module See chap. 8.2 and 8.3
cannot be activated
– Controller switched off Activate controller See chap. 5.5 or fields R2xx
– Controller in “Manual / Off” mode Choose “Auto” or “Manual/On” mode Keyboard, REL key
Controller / limit – Pickup delay setting too long Disable or shorten pickup delay See fields R2xx
contactor does not work
– “Hold” function active “Autom. hold” during calibration, See fields S2 to S4
“hold” input activated;
“hold” via keyboard active
– Controller in “Manual / On” mode Set controller to “Manual / Off” or Keyboard, REL and AUTO keys
“Auto”
Controller / limit
contactor works – Dropout delay setting too long Shorten dropout delay See fields R2xx
continuously – Control loop interruption Check measured value, current
output or relay contacts, actuators,
chemical supply
– Line open or short-circuited Disconnect line and measure mA meter 0–20 mA DC
directly on instrument
– Total load in current loop Disconnect line and measure Ohmmeter
No turbidity current
excessive (> 500 Ω) directly on instrument
output signal
– Output defective See diagnosis table in chap. 8.1
– Instrument with PA-/DP instruments have no
PROFIBUS-PA/-DP current output!
– Current simulation active Switch off simulation See field O2(2)
Fixed current output
signal – Processor system out of sync Switch instrument off and back on EMV problem: check installation if
problem persists
– Incorrect current assignment Check current assignment: Field O2(1), O2(3) for table
Incorrect current output 0–20 mA or 4–20 mA?
signal – Total load in current loop Disconnect output and measure Ohmmeter
excessive (> 500 Ω) directly on instrument
Current output table not – Value interval too small Use sensible intervals
accepted

Endress+Hauser 57
Maintenance and troubleshooting Liquisys M CUM 223 / 253

Equipment need, spare parts,


Error Possible cause Remedy
information
– Instrument only has one current Refer to nameplate for variant, LSCH-x2 module,
No temperature output output change LSCH-x1 module if nec. see chap. 8.2.4 and 8.3.4
signal – Instrument with PA-/DP instruments have no
PROFIBUS-PA/-DP current output!
Install LSR1-4 module.
Chemoclean function – No relay module (LSR1-x) or nur Chemoclean is enabled via release LSR1-4 module,
not available LSR1-2 installed code supplied by E+H with see chap. 8.2.4 and 8.3.4
Chemoclean upgrade
– Plus package upgrade:
– Plus package not enabled code received from E+H ⇒ enter
(enable with code that depends – Following replacement of Detailed description see chap.
Plus package functions
on serial number and is received defective LSCH/LSCP module: 8.3.5
not available
from E+H with order or Plus first enter instrument serial
package) number (see nameplate)
manually, then enter code
– No central HART module Verify by looking at nameplate: Upgrade to LSCH-H1 / -H2 module
HART = -xxx5xx und -xxx6xx
– No or wrong DD
device description)
– HART interface missing
– Instrument not registered with
No HART® HART server For further information, see
communication – Load < 230 Ω additional Operating Instructions
BA 208C, “HART® – Field
– HART receiver (e.g. FXA 191) not
communication with Liquisys M
connected via load
CxM 223 / 253”
– Incorrect instrument address
(Adr. = 0 for single operation,
Adr. > 0 for multidrop operation)
– Line capacity too high
– Line interference
– No central PA-/DP module Verify by looking at nameplate: Upgrade to LSCP module
PA = -xxx3xx / DP = -xxx4xx
– Wrong SW version
(without PROFIBUS)
– Commuwin (CW) II:
CW II version and instrument
SW version incompatible
– No or wrong DD/DLL
For further information, see
– Incorrect baud rate setting for additional Operating Instructions
segment coupler in DPV-1 server BA 209C, “PROFIBUS-PA/-DP –
No PROFIBUS®
communication – Incorrect station (master) Field communication with
address or duplicate address Liquisys M CxM 223 / 253”
– Incorrect station (slaves) address
– Bus line not terminated
– Line problems
(too long, cross section too
small; not shielded, screen not
grounded, wires not twisted)
– Bus voltage too low The voltage at the PA-/DP
(bus supply voltage connection of the instrument must
typ. 24 V DC for non-Ex, be at least 9 V.
13,5 V DC for Ex)
No HART or PROFIBUS – Several instruments on the same Check addresses and reset if
communication address necessary

58 Endress+Hauser
Liquisys M CUM 223 / 253 Maintenance and troubleshooting

UM253E07.CHP

7.2 Troubleshooting using the error messages


Display and select error messages by pressing the MINUS key.

Error Contact Error current Cleaning trigger


Display Measures
no.
Fact User Fact User Fact User

E001 EEPROM memory error Switch instrument off and back on, yes no — —*
return instrument to your local
Endress+Hauser sales agency for repair yes no — —*
or replace instrument.
Instrument not calibrated,
calibration data invalid,
E002 Load software compatible with hardware.
no user data invalid
(EEPROM error).
Load measuring parameter specific
instrument software.
Invalid configuration. Repeat download, yes no — —*
E003 Download error
check optoscope.
Instrument software Load software compatible with hardware. yes no — —*
version not compatible
E004
with module hardware Load measuring parameter specific
version instrument software.
Transmitter malfunction yes no — —*
E007 Software not compatible
with hardware (transmitter)
Sensor or sensor Check sensor and sensor connection yes no no
E008
connection defective (E+H Sevice)
Check wiper and test function using yes no no
E026 Wiper error
manual control if necessary

E045 Calibration aborted Repeat calibration no no — —*

Check measurement and connections; yes no no


Below main parameter
E055 check instrument and measuring cable
measuring range
with simulator if necessary.
Main parameter Check measurement, control and yes no no
E057 measuring range connections
exceeded
Below temperature yes no no
E059
measuring range
Temperature measuring yes no no
E061
range exceeded
Below current output Check configuration yes no no
E063
range 1
Current output range 1 Check measured value and current yes no no
E064
exceeded assignment
Below current output yes no no
E065
range 2
Current output range 2 yes no no
E066
exceeded
Set point exceeded Check configuration. yes no no
E067
controller 1
Set point exceeded yes no no
E068
controller 2
Set point exceeded Check configuration yes no no
E069
controller 3
Set point exceeded yes no no
E070
controller 4

Endress+Hauser 59
Maintenance and troubleshooting Liquisys M CUM 223 / 253

Error Contact Error current Cleaning trigger


Display Measures
no.
Fact User Fact User Fact User
Measured value outside Clean sensor; check tables yes no no
E079
concentration table
Parameter range for Decrease current output spreading no no — —*
E080 current output 1 range too
small
Parameter range for Decrease current output spreading no no — —*
E081 current output 2 range too
small
Calibration warning Calibration data are within limits but no no no
E084 deviate from standard values by a factor of
more than two

E100 Current simulation yes Switch off current simulation no no — —*

E101 Service function yes Switch instrument off and back on no no — —*

E102 Relais simulation yes Check relay configuration no no — —*

E106 Download yes Wait for download to end no no — —*

E116 Download error Repeat download no no — —*

Check sensor and connection, maintain or no no no


E152 PCS alarm
replace

E153 Offset Adjustment range exceeded no no no

Below lower alarm If necessary perform manual reference yes no no


E154 threshold for period measurement. Repair sensor and
exceeding alarm delay recalibrate
Lower alarm threshold yes no no
E155 exceeded for period
exceeding alarm delay
Actual value undershoots yes no no
set point for longer than
E156
the set permissible
maximum period
Actual value exceeds set yes no no
point for longer than the
E157
set permissible maximum
period

* If this error is present, the cleaning function cannot be started.


(Field F8 does not exist for this error.)

60 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8 Diagnosis and corrective maintenance


Diagnosis

• Diagnosis refers to the identification of instrument malfunctions and defects.

Corrective maintenance:
• replacement of parts diagnosed to be defective;
• testing of instrument and measuring system function;
• restoration of complete functionality.

Diagnosis based on the error table below and depending on difficulty and measuring
equipment at hand is to be performed by:
• trained operator personnel
• operator’s electricians
• company responsible for system installation/ operation
• E+H Service.
Select the required spare parts using the tables in chap. 8.2 and 8.3.

Warning: Caution: ESD!

• Disconnect the instrument from the


• Electronic components are
power source before opening it
sensitive to electrostatic
up. Work with live lines may only
discharges. Personal protective
be performed by trained
measures, such as discharge via
electricians.
PE or permanent grounding using a
wrist strap, are to be taken.
• Switched contacts may be
supplied from separate circuits.
• For your own safety, use only
These circuits must also be
original spare parts. Original parts
de-energized before work on the
will guarantee functionality,
terminals is performed.
accuracy and reliability after
repairs.

8.1 Diagnosis
The table below will help you diagnose problems and provides information about spare parts
required. Note on the required spare parts. You can find information concerning the exact
spare part designations and the installation of these parts in chap. 8.2.3 and 8.3.3.

Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Display dark, no LEDs – No mains voltage Check if mains voltage is available Electrician / e.g. multimeter
active – Wrong supply voltage / voltage Compare mains voltage and rating Operator (utility company
too low on nameplate specification or multimeter)
– Connection fault Terminal not tightened; Electrician
insulation clamped in terminal;
wrong terminal used
– Instrument fuse defective Replace fuse, Electrician /
first compare mains voltage and see drawings in chap. 8.2.1 and
rating on nameplate 8.3.1 for corrected fuse
– Power supply unit defective Replace power supply unit, On-site diagnosis by E+H Service
using correct variant (test module LSGx Lrequired)
– Controller defective Replace controller, On-site diagnosis by E+H Service
using correct variant (test module LSCxx Lrequired)
– CUM 253: ribbon cable Check ribbon cable, See spare parts for CUM 253
Pos. 310 loose or defetive replace if necessary
Display dark, but Controller defective Replace controller On-site diagnosis by E+H Service
LED(s) active (Module: LSCH/LSCP) (test module LSCxx required)

Endress+Hauser 61
Diagnosis and corrective maintenance Liquisys M CUM 223 / 253

Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Display shows – Instrument or module not CUM 223: Reinstall module Refer to assembly drawings in
measured values but properly installed in instrument CUM 253: Reinstall display module chapters 8.2.1 and 8.3.1.
– value does not change – Impermissible operating system Switch instrument off and back on Possible EMC problem: if problem
and / or state persists, call E+H Service to have
– instrument cannot be installation checked.
operated
Instrument gets hot – Incorrect voltage / too high Compare mains voltage and rating Correct voltage setting,
of nameplate see chap. 8.4.1 / 8.5.1.
– Power supply unit defective Replace power supply unit Can only be diagnosed by E+H
Service.
Incorrect measured – Transmitter module defective Test measuring inputs: If test fails: replace module (using
turbidity and / or (module: MKT1). Sensor simulation is not possible, correct variant), refer to exploded
temperature value Please perform tests and take check with new or other sensor views in chap. 8.2.1 and 8.3.1.
measures according to chap. 7.1
to eliminate the sensor or wiring If module tests out OK: check
as the cause of the problem peripherals once more.
– Data transfer fault Replace LSGA module (AC power Spare parts list see chap. 8.4.4 or
supply unit) or LSGD (DC power 8.5.4.
supply unit)
– Data transfer problem (EMC) Check line routing. Separate Connect sensor line screen to “S”
sensor line from supply lines terminal, do not ground.
– Wrong sensor cable / cable too Max. line length with extension is
long 200 m; use only calbe type CYK 8
Current outputs / – Not calibrated correctly Test with built-in current simulation, If simulation value is incorrect:
incorrect current – Load excessive connecting mA meter directly to recalibration at factory or new
current output LSCxx module required.
– Shunt / short-circuit to frame in
current loop
If simulation value is correct:
check current loop for load and
shunts.
– Incorrect mode of operation Check whether 0–20 mA or 4–20
mA has been selected
No current output signal – Current output stage defective Test with built-in current simulation, If test fails:
(module: LSCH/LSCP) connecting mA meter directly to Replace controller
current output. First disconnect (using correct variant)
current output line completely for
safety
Additional relays do not – CUM 253: ribbon cable Make sure ribbon cable is properly See Spare parts CUM 253 F
work Pos. 320 loose or defective connected, replace cable if nec.
Only 2 additional relays – Relay module LSR1-2 with 2 Convert to LSR1-4 with 4 relays Operator or E+H Service
can be used relays installed
Enhanced functions – No or incorrect release code If upgraded: verify that correct Handled by E+H Sales
(Plus package) not used serial number has been used to
available order Plus package
– Incorrect serial number of Check whether serial number on Serial number stored in
instrument stored in LSCH/LSCP nameplate matches with stored LSCH/LSCP module is required to
module serial number (can be read in field enable Plus package.
S8).
Enhanced functions – Plus package or Chemoclean For new LSCH / LSCP, the A detailed description can be
(Plus package and / or has not been enabled with release instrument serial number can be found in chap. 8.3.5.
Chemoclean) are not codes. entered once in fields E114 to
available after E116. Next step is to enter release
replacement of codes for Plus package and/or
LSCH-/LSCP module Chemoclean.
No HART or PROFIBUS – Wrong controller HART: LSCH-H1 or -H2 Replace controller;
interface function PROFIBUS-PA/-DP: LSCP-PA/-DP operator or E+H Service
module, see field E112
– Wrong software SW version see field E111
– Bus problem Remove some instruments and Contact E-H Service for support
repeat testing

62 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8.2 Corrective maintenance of Liquisys M CUM 223

8.2.1 Exploded view


223_EXO2.CDR

Pos. 426a
Terminal for signal
inputs/outputs (external)
Pos. 340
End frame
Pos. 400 Pos. 426b
Special Terminal for alarm relay
screw and power supply (interior)

Pos. 270
Turbidity transmitter MKT1

Pos. 10 / 20
Power supply unit LSGA / LSGD

Pos. 40 – 90
Controller LSCH-x /
LSCP
End frame
latches Pos. 330
Pos. 320 Housing
Back plate assembly

Pos. 310
Side wall for
instruments
without relay
module

Pos. 430
Push-on
terminal strip for
LSR1-x module

Pos. 10
Pos. 290 / 300 Power supply unit
Optional relay 100 – 230 V AC
module LSR1-x

Module rotated
180° for
representation!

Jumper
Fuse

Mains voltage 230V


selection
115V

100V

Endress+Hauser 63
Diagnosis and corrective maintenance Liquisys M CUM 223 / 253

8.2.2 Disassembly of CUM 223


• Consider potential effects on process
when removing the instrument from service!
• Before opening, make sure the instrument
is voltage-free.
• First pull off the terminal block (Pos. 426b)
on the rear of the instrument.
• First pull off the terminal blocks
(Pos. 426a and if necessary 430) on the
rear of the instrument. You can now
dismantle the instrument.
• Press the latches of the end frame (Pos.
340) inwards and pull off the frame towards
the rear.
• Loosen the special screw (Pos. 400) by
turning it counterclockwise.
• Remove the complete electronics block
from the housing. The modules are
plugged together mechanically and can be
easily separated:
– Pull the controller / display module
forwards
– Gently pull out the backplane latches
(Pos. 320) outwards slightly to remove
the side modules.
• Removing the turbidity transmitter
(Pos. 270):
Pull off the module towards the top.

8.2.3 Assembly of CUM 223


• Reverse the dismantling sequence for
assembly.
• Hand-tighten the special screw without
using a tool.
• Incorrect assembly is not possible!
An module block which has been plugged
together incorrectly cannot be inserted into
the housing.

64 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8.2.4 Spare parts for CUM 223

Position Designation Name Function or content Order number

Sections printed in italics are identical for CUM 223 and CUM 253 (Chap. 8.3.4).

10 Power supply unit LSGA 100/115/230 V AC 51500317

20 Power supply unit LSGD 24 V AC + DC 51500318

290 Relay module LSR1-2 2 relays 51500320

300 Relay module LSR1-4 4 relays 51500321

40 Controller LSCH-S1 1 current output 51501228

50 Controller LSCH-S2 2 current outputs 51501229

60 Controller LSCH-H1 1 current output + HART 51501230

70 Controller LSCH-H2 2 current outputs + HART 51501231

PROFIBUS-PA /
80 Controller LSCP-PA 51501232
no current output.

PROFIBUS-DP /
90 Comntroller LSCP-DP 51502499
no current output

270 Turbidity transmitter MKT1 Turbidity + temperature 51501209

310 Side panel Kit with 10 side panels 51502124

End frame with D-Submin


340 End frame PROFIBUS-DP 51502513
plug-in connector

Housing with front


membrane, Sensor plungers,
330, 400 Housing module Seal, Special screw, Clamp 51501075
tappets, Connection and
nameplates

310, 320, Backplate, Side wall, End


Mechanical parts Housing 51501076
340, 400 frame, Special screw

Terminal strip set


Complete terminal strip set, Standard + HART,
426a, 426b 51501205
Standard + HART Inputs/outputs, power supply,
alarm relay

Terminal strip set


Complete terminal strip set, PROFIBUS-PA,
426a, 426b 51502128
PROFIBUS-PA Inputs/outputs, Power supply,
Alarm relay

Terminal strip set


Complete terminal strip set, PROFIBUS-DP,
426a, 426b 51502491
PROFIBUS-DP Inputs/outputs, Power supply,
Alarm relay

430 Terminal strip Terminal strip for relay module 51501078

Endress+Hauser 65
Diagnosis and corrective maintenance Liquisys M CUM 223 / 253

8.3 Corrective maintenance of Liquisys M CUM 253

8.3.1 Exploded view


253_EXO2.CDR

Electronics box with Fuse is accessible with


relay module LSR1-x (bottom) electronics box installed
and power supply unit
LSGA/LSGD (top)
Pos. 10
Power supply unit
100 – 230 V AC

Jumper
Pos. 410 Fuse
Display lid
assy. Pos. 370
Front membrane Mains voltage 230V
Pos. 40 ... 90 selection
Controller 115V
LSCH-x / LSCP
Pos. 450b
Torx screw for 100V
Controller
Pos. 460b
Cover Pos. 330
screws Electronics box

Pos. 450a
Pos. 310 / 320 Torx screws K4x10
Ribbon cables
Pos. 460a Pos. 400
Cover Bottom of housing
screws
Pos. 270
Transmitter Top
module MKT1 Bottom
Pos. 430
Hinge

Pos. 420
Connection
compartment
cover Pos. 10 / 20
Power supply
module Pos. 290 / 300
LSGA / LSGD Relay module
LSR1-x

Pos. 340
Docking module
LSK1
Pos. 480
Cable glands
Pos. 450c Pos. 470
Torx screws K4x10 Plug-in terminal for
power connection

66 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8.3.2 Disassembly of CUM 253


• Before opening, make sure the instrument
is voltage-free.
• Open and remove the connection
compartment cover (Pos. 420).
• Remove the network terminal (Pos. 70).
• Open the display cover (Pos. 410) and
loosen the ribbon cable (Pos. 310 / 320) on
the electronics box side (Pos. 330).
• Removal of electronics box (Pos. 330):
Loosen screws (Pos. 450a) in the bottom of
the housing 2 revolutions, then slide entire
box backward and remove towards the top.
Make sure that the module catches do not
come unlatched while you push the
electronics box back!
• Now bend the module latches outward and
remove the module(s).
• Dismantling the controller (Pos. 40):
Loosen screw (Pos. 450b) in the display lid.
• Expansion of the docking module (Pos.
340):
Loosen screws (Pos. 450c) in the bottom of
the housing and remove entire module
towards the top.
• Removing the transmitter (Pos. 270):
Simply pull off the module towards the top.

8.3.3 Assembly of CUM 253


• Insert the module(s) in the electronics box
guide rails carefully and latch into the
lateral lugs in the box.
• You cannot assemble the instrument
incorrectly!
Modules inserted in the electronics box
incorrectly are not operable since the
ribbon cables cannot be inserted.
• Make sure that the cover gaskets are intact
since they are required to guarantee
protection class IP 65.

Endress+Hauser 67
Diagnosis and corrective maintenance Liquisys M CUM 223 / 253

8.3.4 Spare parts for CUM 253

Position Designation Name Function or content Order number

Sections printed in italics are identical for CUM 223 and CUM 253 (Chap. 8.2.4).

10 Power supply unit LSGA 100/115/230 V AC 51500317

20 Power supply unit LSGD 24 V AC + DC 51500318

290 Relay module LSR1-2 2 relays 51500320

300 Relay module LSR1-4 4 relays 51500321

40 Controller LSCH-S1 1 current output 51501228

50 Controller LSCH-S2 2 current outputs 51501229

60 Controller LSCH-H1 1 current output + HART 51501230

70 Controller LSCH-H2 2 current outputs + HART 51501231

PROFIBUS-PA /
80 Controller LSCP-PA 51501232
no current output.

PROFIBUS-DP /
90 Controller LSCP-DP 51502499
no current output

270 Turbidity transmitter MKT1 Turbidity + temperature 51501209

Display cover, connection


370, 410,
compartment cover, front
420, 430, Housing cover 51501068
membrane, Hinge, Cover
460
screws, Small parts

400, 480 Bottom of housing Bottom, cable glands 51501072

330, 340, Docking module, empty


Internal housing parts 51501073
450 electronics box, small parts

310, 320 Ribbon cables 2 ribbon cables 51501074

430 Hinges 2 pairs of hinges 51501069

Terminal strip for power


470 Terminal strip 51501079
connection

68 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8.3.5 Special case: Replacement of controller

Note: • Disassemble the instrument as described


in chapters 8.2.2 and 8.3.2..
• Refer to the part no. of the central module
• After controller replacement all the to determine whether the new module has
editable data are reset to the the same part no. as the old one.
factory settings. • Reassemble the instrument with the new
module as described in chap. 8.2.3 and
• The serial number can only be 8.3.3.
entered – and only once – in the • Start up the instrument and test its basic
case of a new module from the functions (e.g. meas. value and
factory with serial number 0000! temperature display, operation via
Make sure that your entry is keyboard).
correct before confirming with
ENTER! Entry of an incorrect code • Enter the instrument serial number:
will prevent the enhanced – Read the serial number (“ser-no.”) of the
functions from being enabled. An instrument from the nameplate.
incorrect serial number can only be – Enter this number in the fields E115
corrected at the factory. (year), E116 (month), E117 (sequence
number, four-digit).
Proceed as described below after controller – Field E118 displays the complete number
replacement: for verification; acknowledge with ENTER
• If possible, record the user settings of the or abort and re-enter.
instrument, e.g.: • Verify that the Plus package is enabled
– Measuring range and configuration (e.g. by accessing function group CHECK
setting / Code P) or the Chemoclean function.
– Turbidity and temperature current • Check the Plus package release (e.g. by
assignments opening the function group CHECK /
– relay functions selected Code P) or the Chemoclean function.
– limits/controller settings • Restore the user settings of the instrument.
– cleaning settings
– monitoring functions
– interface parameters

8.4 Spare parts orders


Spare parts are to be ordered from your local • instrument order code
E+H Sales Agency. See the back cover of • serial number (ser-no.)
these operating instructions for addresses. • software version, if possible
For this, refer to the order numbers in
chapters 8.2.4 and 8.3.4. The order code and serial number are on the
nameplate.
To be on the safe side, you should always The software version is displayed in field
specify the following data with spare parts E111 when the instrument processor system
orders: is functional.

8.5 Service aids “Optoscope” with “Scopeware”


In connection with the “Scopeware” software, Handling and operation are described in the
the optoscope offers the following optoscope operating instructions. The
possibilities, with the need to disassemble or user-friendly Windows software required for
open the Liquisys S and without a galvanic the PC or laptop is contained in the scope of
connection to the device: supply.
• Documentation of the instrument settings in
conjunction with Commuwin II The optoscope is supplied in a sturdy case
• Software update by the service technician with all the accessories required.
• Upload/download a hex dump to duplicate
configurations. Order number of optoscope: 51500650

The optoscope serves as an interface


between the Liquisys M and PC / laptop.
Information is exchanged on the instrument-
side using the optical interfaces of the
Liquisys M and to the PC / laptop using the
interface RS 232.

Endress+Hauser 69
Diagnosis and corrective maintenance Liquisys M CUM 223 / 253

8.6 Corrective maintenance of measuring system

8.6.1 Turbidity transmitter


The sensors CUS 31 and CUS 41 cannot be • Measure the sensor operating voltage:
simulated as they contain the complete data approx. 10 ... 16 V at terminals 87 (+) and
processing and all the measured values are 88 (–).
transmitted using the digital interface RS 485 • The cause for a defective voltage may be
to CUM 223 / 253. Therefore a functional present either at the device or at the
sensor is required for the measuring point test. sensor. Therefore, the next thing you
should do is
Method for testing a measuring point: • connect a new or a different sensor.
– If the function is now o.k. -> Sensor
• Check that device is operable and that the defective.
display reacts appropriate, e.g. by – If the sensor operating voltage is
pressing the PLUS key. still to low -> replace power supply
• Check the current outputs by carrying out module LSGA/LSGD (Pos. 10 / 20,
a current simulation (Field O22). check the required version).
• Sensor voltage o.k., but no measured
turbidity value even with a new sensor:
-> Replace transmitter module MKT1.

Cleaning the front panel

To clean the front panel, only use standard Caution!


cleaning agents.
Do not use any concentrated mineral
The front panel is resistant to the following acids or alkalis, benzyl-alcohol,
cleaning agents (acc. methylene chloride or high-pressure
• Isopropanol water vapour at over 100 °C.
• Thinned acids (3 %)
• Thinned alkalis (5 %)
• Ester
• Hydrocarbons
• Ketone
• Household cleaners

8.6.2 Turbidity sensors CUS 31 / 41


The sensors CUS 31 / 41 contain their own When replacing a sensor, check the following:
digital signal processor and communicate
with the turbidity measuring instrument via an • Replacing the sensor CUS 31-xxA or
interface RS 485. All the sensor data (factory CUS 41
calibration data and customer calibration All the calibration data is saved in the
data) are permanently saved in the sensor. sensor. When using the original data
records (“read only”), no calibration is
You can find detailed information on these necessary after sensor replacement.
sensors in the “Technical Information Matter-specific calibrations must be
Turbidity sensor TurbiMax W CUS 31 / CUS repeated.
31-W” TI 176C/07/en or in “Technical
Information Solid matter sensor TurbiMax W • Replacing the sensor CUS 31-xxE or
CUS 41 / CUS 41-W” TI 177C/07/en. CUS 31-xxS
All the factory calibration data is saved in
the sensor. The sensor and the assembly
are calibrated together. No additional
calibration is required for applications with
pure or ultrapure water, if the sensor and
the assembly are replaced together. The
calibration data of the sensor is transferred
automatically to the measuring instrument.

70 Endress+Hauser
Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance

UM253E08.CHP

8.6.3 Assembly
Refer to the assembly operating instructions Flow
for maintenance and troubleshooting! For the • Flow assembly FlowFit W CUA 250
CUS 31/41, you can use: Operating manual BA 096C/07/en
• Welding socket with adapter CUA 120-B
Basin and channel Retractable assembly
• Suspended assembly holder CYH 101 and • ProbFit CUA 461
basin wall bracket CYY 106 Operating manual BA 134C/07/en
Technical Information TI 092C/07/en Pure and ultrapure water applications
• Assembly DipFit W CYA 611 • CUS 31-xxE: contains special flow
Operating manual BA 166C/07/en assembly without gas bubble trap
• Sensor fixing bracket RM • CUS 31-xxS: contains special flow
Operating manual BA 205C/07/en assembly with gas bubble trap

Endress+Hauser 71
Accessories Liquisys M CUM 223 / 253

9 Accessories
Mounting accessories Connection accessories

• Weather protection cover CYY 101 • Extension cable CYK 8, metre goods
For mounting on field housing, for outdoor Unterminated measuring cable for turbidity
installation. sensors CUS 31 and CUS 41
Dimensions (H × W × D): Order no.: 50089633
320 × 300 × 270 mm
Material: Stainless steel SS 304 • Extension cable CYK 8, 50 m
Order no.: CYY 101-A Terminated measuring cable for turbidity
sensors CUS 31 and CUS 41.
• Universal upright post CYY 102 Order no.: 51500830
Square tube for mounting of field housing.
Dimensions (H × W × D): • Junction box VBM
1495 × 60 × 60 mm Junction box for extending the measuring
Material: Stainless steel SS 304 cable connection between sensor and
Order no.: CYY 102-A instrument. Pg 13.5 cable entries.
Material: cast aluminium;
Sensors Ingress protection: IP 65
Order no.: 50003987
• TurbiMax W CUS 31 / 31 W
pH/ redox combination electrode • Junction box RM
with PTFE diaphragm, with gel filling, Junction box for extending the measuring
usable at up to 6 bar. cable connection between sensor and
Technical Information TI 028C/07/en instrument. Pg 13.5 cable entries.
Order no.: 50088400 Material: cast aluminium;
Ingress protection: IP 65
• TurbiMax W CUS 41 / 41 W Order no.: 51500832
Sterilisable pH /redox electrode with
double-chamber reference and integrated Assemblies
bridge electrolyte.
Technical Information TI 245C/07/en • DipFit W CYA 611
Order no.: 50088402 Immersion assembly for turbidity sensors,
unpressurised.
Spare parts for CUS 31 and CUS 41 Technical Information TI 166C/07/en
Order no.: 50085985
• Service kit for wiper operation cpl.
(upgrade or replacement) • FlowFit W CUA 250
Order no.: 50089254 Flow assembly for turbidity sensors,
applicable at up to 6 bar
• Service Kit for seal Technical Information TI 096C/07/en
(for wiper extension, contains 5 sets) Order no.: 50063827
Order no.: 50089253
• ProbFit CUA 461
• Service Kit for wiper arm Manually replacable assembly for turbidity
(with rubber and mounting material, sensors, applicable at up to 1 bar
contains 3 sets) Technical Information TI 134C/07/en
Order no.: 50089252 Order no.: 50075623

• Measuring cable 7 m terminated, Device upgrade


with housing upper section and screw
union (Order only possible with specification of the
Order no.: 50088167 serial number of the appropriate device)

• O-ring 28 × 3.5 mm • Plus package


(Seal for G 1 pipe installation) Order no.: 51500385
Order no.: 50051753
• Chemoclean
Accessories for CUM 253 CSA version Order no.: 51500963

• NPT adapter set • Two-relay card


Order no.: Order no.: 51500320

• Four-relay card
Order no.: 51500321

72 Endress+Hauser
Liquisys M CUM 223 / 253 Technical data % °C

UM253E10.CHP

10 Technical data

General specifications Manufacturer Endress+Hauser


Product designation Liquisys M CUM 223
Liquisys M CUM 253

Function and system design Measuring principle A sensor CUS 31 or CUS 41 is connected to the transmitter
Liquisys M CUM 223 / 253 using the digital sensor interface.
The sensors supply a standardised signal for the turbidity
values and the temperature value.

Input Measured variables Turbidity, Solids content, Temperature

Turbidity measurement with CUS 31


Measuring range 0.000 ... 9999 FNU, 0.00 ... 3000 ppm,
0.0 ... 3.0 g/l, 0.0 ... 200.0 %
Turbidity offset range ±99.99 FNU, ±99.99 ppm, ±99.9 g/l, ±99.9 %

Solids contentration measurement with CUS 41


Measuring range 0.00 ... 9999 FNU, 0.00 ... 9999 ppm,
0.0 ... 300.0 g/l, 0.0 ... 200.0 %
Solids concentration offset range ±99.99 FNU, ±99.99 ppm, ±99.9 g/l, ±99.9 %

Temperature measurement
Temperature sensor NTC, 30 kΩ at 25 °C
Sensor measuring range –5.0 ... +70.0 °C
Temperature offset range ±5 °C

Signal input for turbidity / solids concentration / temperature


Sensor interface digital
Maximum cable length to sensor 200 m
Terminal cross-section 2.5 mm2
Required conductor cross-section CUM 223 0.75 mm2 / AWG 18
CUM 253 1.50 mm2 / AWG 14

Digital inputs 1 and 2


Voltage 10 ... 50 V
Current consumption max. 10 mA

Output Signal output for turbidity / solids concentration


Current range 0 / 4 ... 20 mA, galvanically isolated;
error current 2.4 / 22 mA
Max. resolution 700 digits / mA
Load max. 500 Ω
Transmission range adjustable, min. ∆ 0.1 FNU, ∆ 0.1 ppm, ∆ 1 g/l, ∆ 0.1 %
Separation voltage max. 350 Vrms / 500 V DC
Overvoltage (lightning) protection acc. to EN 61000-4-5:1995

Endress+Hauser 73
% °C Technical data Liquisys M CUM 223 / 253

Output (continued) Signal output temperature (optional)


Current range 0 / 4 ... 20 mA, galvanically isolated
Max. resolution 700 digits / mA
Load max. 500 Ω
Transfer range adjustable, ∆ 10 ... ∆ 100 % from measuring range
Separation voltage max. 350 Vrms / 500 V DC
Overvoltage (lightning) protection acc. to EN 61000-4-5:1995

Auxiliary voltage output


Output voltage 15 V ± 0.6 V
Output current max. 10 mA

Contact outputs (potential-free changeover contacts)


Switching current with ohmic load (cos ϕ = 1) max. 2 A
Switching current with ohmic load (cos ϕ = 0.4) max. 2 A
Switching voltage max. 250 V AC, 30 V DC
Switching power with ohmic load (cos ϕ = 1) max. 500 VA AC, 60 W DC
Switching power with inductive load (cos ϕ = 0.4) max. 500 VA AC, 60 W DC

Limit contactor
Pickup / dropout delay 0 ... 7200 s

Controller
Function (adjustable) Pulse length / pulse frequency controller
Controller response P, PI, PD, PID
Controller gain Kp 0,01 ... 20.00
Integral action time Tn 0.0 ... 999.9 min
Derivative action time Tv 0.0 ... 999.9 min
Period for pulse length controller 0.5 ... 999.9 s
Max. frequency for pulse frequency controller 60 ... 180 min–1

Alarm
Function (switchable) Steady / fleeting contact
Alarm delay 0 ... 2000 s (min)

74 Endress+Hauser
Liquisys M CUM 223 / 253 Technical data % °C

UM253E10.CHP

Accuracy Turbidity measurement with CUS 31


Measured value resolution 0.001 FNU, 0.01 ppm, 0.01 g/l, 0.01 %
Display deviation1 ±2 % of measured value (at least 0.02 FNU)
Repeatability1 ±1 % of measured value (at least 0.01 FNU)
Measurement deviation1 turbidity signal output 1 % of current output range (at least 0.02 FNU)

Solids content measurement with CUS 41


Measured value resolution 0.01 FNU, 0.01 ppm, 0.01 g/l, 0.01 %
1
Display deviatoin ±2 % of measured value (at least 0.02 FNU)
Repeatability1 ±1 % of measured value (at least 0.01 FNU)
Measurement deviation1, solids concentration 1 % of current output range (at least 0.02 FNU)
signal output

Temperature measurement
Measured value resolution 0,1 °C
Display deviation1 max. 1.0 % of measuring range
Measurement deviation1 temperature signal output max. 1.25% of current output range

Ambient conditions Ambient temperature (nominal operating conditions) –10 ... +55 °C
Ambient temperature (limit operating conditions) –20 ... +60 °C
Storage and transportation temperature –25 ... +65 °C
Relative humidity (nominal operating conditions) 10 ... 95 %, non-condensing
Ingress protection of panel-mounted instrument IP 54 (front panel), IP 30 (enclosure)
Ingress protection of field instrument IP 65
Pollution degree 2 acc. to IEC 61010-1
Installation category II
Maximum altitude 2000 m above sea level
Electromagnetic compatibility Interference emission and immunity acc. to
EN 61326: 1997/ A1: 1998

Mechanical construction Dimensions of panel-mounted instrument (H × W × D) 96 × 96 × 139 mm


Installation depth approx. 165 mm
Dimensions of field instrument (H × W × D) 247 × 170 × 115 mm
Weight of panel-mounted unit max. 0.7 kg
Weight of field instrument max. 2.3 kg
Measured value display LC display, 2-line, 5- and 9-digit with status displays

Materials
Housing of panel-mounted unit Polycarbonate
Front membrane Polyester, UV-resistant
Field housing standard version ABS PC Fr
CSA GP version Polycarbonate

Power supply Supply voltage 100 / 115 / 230 V AC +10 / –15 %, 48 ... 62 Hz
24 V AC/DC +20 / –15 %
Power consumption max. 7.5 VA
Mains fuse Microfuse, medium slow 250 V / 3.15 A

1
According to IEC 746-1, for nominal operating conditions
Subject to modifications.

Endress+Hauser 75
Appendix Liquisys M CUM 223 / 253

11 Appendix
Function group Entry of absolute value Entry offset Calibration status Store
OFFSET is displayed offset
current measured value
o.k. yes; no; new
V V1 V2 V3 V4
E---

Calibration selection Display of cal. point 1 Display of cal. point 2 Display of cal. point 3 Display of Display of
slope 1 slope 2
Data Comparative value Comparative value Comparative value
= calibration data; C1 (6) current value current value
only C1 (6) is available
for data set 1
C161 C162 C163 C164 C165

Entry of current Calibration status Store calibration


1-pt = 1-point calibration value is displayed results
calibration C1 (5)
current measured value o.k. yes; no; new
entire measuring range E---
C151 C152 C153

Refl
= reflection adaptation Entry of correct Calibration status Store calibration
C1 (4) measured value is displayed results

0; 0 ... 2.0 FNU o.k. yes; no; new


0; 0 ... 5.0 ppm (mg/l) E---
C141 C142 C143

Edit = edit
calibration Entry of concentration Entry of concentration Entry of concentration Store calibration
Calibration status
of first calibration solution of second cal. solution of third calibration solution is displayed result

current value in C111 current value in C112 current value in C113 o.k. yes; non; new
C1 (3)
entire measuring range > value in C131 × 11 > value in C132 × 1.1 E---
C131 C132 C133 C134 C135

Corr
= 3-point correction Entry of correct Store calibration
Calibration status
concentration of is displayed results
C1 (2) calibration solution 3
o.k. yes; no; new
current value in C113 E---
entire meas. range
C121 C122 C123

3-pt = 3-point
calibration
Function group Entry of concentration Entry of concentration Entry of concentration Calibration status Store calibration
CALIBRATION of first calibration solution fo second cal. solution of third calibration solution is displayed result
C1 (1) .
100.0 FNU; .

100.0 ppm (mg/l); o.k. yes; non; new


last calibration value last calibration value E---
10.0 g/l; 10.0 %
entire meas. range C112 ≥ C111 × 1.1 C113 ≥ C112 × 1.1
C C111 C112 C113 C114 C115
any code
Read mode:
Code 22
Edit mode:

Temperature display Temperature display Measured value display


in °F suppressed in FNU
MEAS. VALUE DISPLAY

with

1st error is displayed Other errors are


TEMPERATURE DISPLAY
displayed
in °C
(if present)
(up to 10 errors)
any code
Read mode:
Code 22
Edit mode:

Function group Selection of oper. mode Selection of unit Display format selection Display of sensor Entry of damping
displayed (if A1=spec) (if A1=spec) connected (1=no damping)
SETUP 1
FNU; ppm; mg/l; g/l;
%; spec. 3 XX.xx; X.xxx; XXX.x; XXXX CUS 31; CUS 41 10
kg/l; t/m
1-60
%; none
A A1 A2 A3 A4 A5

Function group Switch wiper control Set duration of Set pause time between Selection of calibration Copy data sets Display of measured value
on or off wiper operation wiper cycles data set used with reflection adaptation
SETUP 2
no
off 30 120 min 3 1 → 2; 1 → 3 yes; no
on 3 ... 999 s 1 ... 7200 min 1 ... 3 2 → 3; 3 → 2
B B1 B2 B3 B4 B5 B6

Characteristic selection Tablen option selection Set number of Selection of Entry of x value
table value pairs table value pairs (measured value)
read .
edit 1 1 0.00 FNU / ppm (mg/l)
O2 (3)
1 ... 10 1 ... number of g/l / %
Tab = table table value pairs; entire meas. range
O231 O232 asign O233 O234

Entry of simulation value

current value
O2 (2) 0 ... 22.00 mA

sim = simulation
O221

Function group Currect output selection Current range selection 0/4 mA value 20 mA value
CURRENT OUTPUT entry entry
Out1; Out2 4-20 mA; 0-20 mA
O2 (1) 0.0 FNU; 0.0 ppm (mg/l); 9999 FNU; 9999 ppm(mg/l);
0.0 %; 0.0 °C 300.0 g/l; 10.0 %; 100.0 °C
lin = linear entire meas. range entire meas. range
O O1 O211 O212 O213

Function group Select contact type Select alarm Alarm delay Error current Error number Set alarm contact
ALARM delay unit setting selection to be effective
Steady = steady contact 0s (min)
Fleet = fleeting contact F1 s; min 0 ... 2000 s (min) 22 mA 1 yes; no
F F2 (depends on F2) F3 2.4 mA F4 1 ... 255 F5 F6

76 Endress+Hauser
Liquisys M CUM 223 / 253 Appendix

UM253E11.CHP

Display of
conversion factor

current value

C166

Entry of correct Entry of


process temperature temperature difference
(offset)
current meas. value
–5.0 ... 100.0°C current offset
–5.0...5.0°C
B7 B8

Entry of y value Table status o.k.


(current value)
yes; no
4.00 mA
0.0 ... 20.0 mA
O235 O236

Field for customer setting

Activate error current Automatic start of Select “next error”


for previously set cleaning function or return to menu
error no; yes
(not always displayed, next=next error
no; yes F7 see error messages) F8 ←R F9

Endress+Hauser 77
Appendix Liquisys M CUM 223 / 253

Function group PCS alarm setting


CHECK (live-check)

off / 1h / 2h / 4h
P P1

Limit contactor Function R2 (5) Start pulse selecction Entry of pre-rinse time Entry of cleaning time
configuration Switch off or on int = internal;
ext = external; 20 s 10 s
Clean = R2 (5) off; on i+ext = internal+external; 0 ... 999 s 0 ... 999 s
Chemoclean i+stp = internal
(only with rel3 and rel4) suppressed by ext
R251 R252 R253 R254

Function R2 (4) Rinse time setting Pause time setting Set minimum
Switch off or on pause time
30 s 360 min
off; on 3 ... 999 s 1 ... 7200 min 120 min
Timer R2 (4) 1 ... 3600 min
R241 R242 R243 R244

Function R2 (3) Entry of set value Entry of Entry of integral action time
Switch off or on control gain Kp Tn (0.0 = no I component)
0 FNU; 0 ppm (mg/l);
off; on 0 g/l; 0 % 1.00 0.0 min
PID controller R2 (3)
entire measuring range 0.01 ... 20.00 0.0 ... 999.9 min
R231 R232 R233 R234

Function R2 (2) Entry of switch-on Entry of switch-off Pickup delay


Switch off or on temperature temperature setting

off; on 100.0 °C 100 °C 0s


LC °C = R2 (2) –5.0 ... +100.0 °C –5.0 ... +100.0 °C 0 ... 2000 s
T limit contactor
R221 R222 R223 R224

Function group Select contact to be Function R2 (1) Select contact Select contact Pickup delay
CONTACTS configuredl Switch off or on switch-on point switch-off point setting
LC PV = R2 (1)
TU / TS 9999 FNU; 9999 ppm (mg/l);
Rel1; Rel2; Rel3; Rel4 off; on 9999 FNU; 9999 ppm (mg/l); 0s
Limit contactor 300.0 g/l; 200.0 % 300 g/l; 200 % 0 ... 2000 s
entire measuring range entire measuring range
R R1 R211 R212 R213 R214

Function group Selection of concentration Select table to be Table option Set number of Value pair selection Entry of turbidity
CONCENTRATION curve for calibration of edited selection value pairs value
display value 1
MEASUREMENT read 1 1 ... number of 0 FNU / ppm (mg/l) /
1
Curve 1 ... 4 1 ... 4 edit 1 ... 10 value pairs in K4 g/l / %
K3 entire measuring range K6
K K1 K2 K4 K5

Function group Language selection Hold configuration Manual hold Entry hold dwell Entry of SW upgrade Entry of SW upgrade
SERVICE s+c=during setup period release code release code
ENG; GER and calibration off; on (Plus package) Chemoclean
ITA; FRA CAL =during calibration 10 s
ESP; NEL Setup =during setup 0...999 s 0000 0000
S S1 none =no hold S2 S3 S4 0000...9999 S5 0000...9999 S6

Module selection Software Hardware Serial number Module name


version version is displayed is displayed
Rel = relay E1(4)
SW version HW version

E141 E142 E143 E144

Software Hardware Serial number Module name


version version is displayed is displayed
MainB E1(3)
= mainboard SW version HW version

E131 E132 E133 E134

Software Hardware Serial number Module name


version version is displayed is displayed
Trans E1(2)
= transmitter SW version HW version

E121 E122 E123 E124

Function group Software Hardware Serial number Module name


E + H SERVICE version version is displayed is displayed
Contr E1(1)
= controller SW version HW version

E E111 E112 E113 E114

Function group Entry of address


INTERFACE
HART: 0 ... 15
PROFIBUS: 1 ... 126
I I1

78 Endress+Hauser
Liquisys M CUM 223 / 253 Appendix

UM253E11.CHP

Entry of post-rinse time Number of repeat cycles Set interval between Set minimum Number of cleaning
two cleaning cycles pause time cycles without
20 s 0 (pause time) cleaning agents
0 ... 999 s 0 ... 5 120 min
360 min 1 ... R357 min 0
R255 R256 1 ... 7200 min 0 ... 9
R257 R258 R259

Entry of derivate Selection of Selection Entry of Entry of max. Entry of minimum


action time Tv controller characteristic pulse interval pulse frequency ON time ton
Ien = pulse length
0.0 min dir = direct freq = pulse frequency 10.0 s 120 1/min 0.3 s
0.0...999.9 min inv = inverted 0.5...999.9 s 60 ... 180 1/min 0.1 ... 5.0 s
R235 R236 R237 R238 R239 R2310

Dropout delay Setting of alarm threshold


setting (as an absolute value)

0s 100.0 °C
0 ... 2000 s –5.0 ... +100.0 °C
R225 R226

Dropout delay Setting of alarm threshold


setting (as an absolute value)

0s 9999 FNU; 9999 ppm (mg/l);


0 ... 2000 s 300.0 g/l; 200.0 %
entire measuring range
R215 R216

Entry of concentration Table status o.k.

entire measuring range yes; no

K7 K8

Order number Serial number Reset of instrument Perform instrument test


is displayed is displayed to basic values
no no
Sens = sensor data; Displ = display test
Facty = factory settings
S7 S8 S9 S10

Endress+Hauser 79
Index Liquisys M CUM 223 / 253

12 Index
! D
0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 20 D component . . . . . . . . . . . . . . . . . . . . . . . . 38
22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
9999 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 20 Declaration of conformity . . . . . . . . . . . . . . . . 5
Definition of terms . . . . . . . . . . . . . . . . . . . . 61
A Derivative action time. . . . . . . . . . . . . . . . . . 38
A-Function group . . . . . . . . . . . . . . . . . . 24, 25 Device upgrade . . . . . . . . . . . . . . . . . . . . . . 72
Access codes . . . . . . . . . . . . . . . . . . . . . . . . 20 Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . 61-71
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . 72 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Actuating signal outputs . . . . . . . . . . . . . . . . 33 Disassembly of CUM 223 . . . . . . . . . . . . . . 64
ALARM (F) . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Disassembly of CUM 253 . . . . . . . . . . . . . . 67
Alarm contact . . . . . . . . . . . . . . . . . . . . . 23, 29 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Alarm threshold. . . . . . . . . . . . . . . . . . . . 36, 37 Dropout delay . . . . . . . . . . . . . . . . . . . . 36, 37
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Assembly of CUM 223 . . . . . . . . . . . . . . . . . 64 E
Assembly of CUM 253 . . . . . . . . . . . . . . . . . 67 E+H SERVICE (E). . . . . . . . . . . . . . . . . . . . . 44
AUTO key . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 E-Function group . . . . . . . . . . . . . . . . . . . . . 44
Economy function . . . . . . . . . . . . . . . . . . . . 40
B Electrical connection. . . . . . . . . . . . . . . . . . 12
B-Function group . . . . . . . . . . . . . . . . . . 25, 26 Electronics box . . . . . . . . . . . . . . . . . . . . . . . 7
Error codes . . . . . . . . . . . . . . . . . . . . . . . . . 59
C Error current . . . . . . . . . . . . . . . . . . . . . . . . . 29
C-Function group . . . . . . . . . . . . . . . . . . 49-53 Error log . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Cable length . . . . . . . . . . . . . . . . . . . . . . . . . 14 Extension cable CYK 8 . . . . . . . . . . . . . . . . 72
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Extension of measuring cable . . . . . . . . . . . 14
CALIBRATION (C) . . . . . . . . . . . . . . . . . 22, 49
Calibration data records . . . . . . . . . . . . . . . . 47 F
Calibration mode . . . . . . . . . . . . . . . . . . 19, 22 F-Function group . . . . . . . . . . . . . . . . . . 29, 30
CHECK (P) . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Factory settings . . . . . . . . . . . . . . . . . . . . . . 22
Chemoclean function . . . . . . . . . . . . . . . 31, 35 Freezing of outputs . . . . . . . . . . . . . . . . . . . 21
Cleaning function . . . . . . . . . . . . . . . . . . . . . 39 Function coding . . . . . . . . . . . . . . . . . . . . . . 22
Cleaning time . . . . . . . . . . . . . . . . . . . . . 35, 40 Function group. . . . . . . . . . . . . . . . . . . . . . . 20
Code 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Function of keys . . . . . . . . . . . . . . . . . . . . . . 17
Code 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Code 9999 . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Coding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 G
Coding structure . . . . . . . . . . . . . . . . . . . . . . 22 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 3
Complete Installation . . . . . . . . . . . . . . . . . . . 6 General safety notes . . . . . . . . . . . . . . . . . . . 4
Complete measuring system . . . . . . . . . . . . . 6
CONCENTRATION (K) . . . . . . . . . . . . . . 22, 41 H
Concentration measurement . . . . . . . . . . . . 41 Hardware version. . . . . . . . . . . . . . . . . . . . . 44
Configure hold . . . . . . . . . . . . . . . . . . . . . . . 42 HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Connection diagram . . . . . . . . . . . . . . . . . . . 12 Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Connection examples . . . . . . . . . . . . . . . . . . 15 Hold dwell period. . . . . . . . . . . . . . . . . . 35, 43
Connections of field instrument . . . . . . . . . . 13 Hold function . . . . . . . . . . . . . . . . . . . . . . . . 21
Connections of panel-mounted instrument . 13
Contact selection . . . . . . . . . . . . . . . . . . . . . 29 I
Control characteristic . . . . . . . . . . . . . . . . . . 34 I component . . . . . . . . . . . . . . . . . . . . . . . . . 38
Controller characteristic . . . . . . . . . . . . . . . . 38 I-Function group. . . . . . . . . . . . . . . . . . . . . . 45
Corrective maintenance . . . . . . . . . . . . . 61-71 Immunity to interference . . . . . . . . . . . . . . . . 5
Corrective maintenance CUM 223 . . . . . . . . 63 Installation . . . . . . . . . . . . . . . . . . . . . . 4, 6-15
Corrective maintenance CUM 253 .22, . . .23,
. . .24,
. 66
..............
25,............................................................
26,
Instrument
27, 28,............................................................
29,
configuration
30,............................................................
31, 32, ..........................................................
33, 34, 35, 36, 37, 40
Corrective maintenance Integral action time . . . . . . . . . . . . . . . . . . . 38
of measuring system . . . . . . . . . . . . . . . . . . 70 Intended application . . . . . . . . . . . . . . . . . . . 4
CURRENT OUTPUT (O) . . . . . . . . . . . . . 22, 26 INTERFACE (I) . . . . . . . . . . . . . . . . . . . . . . . 45
Current output table . . . . . . . . . . . . . . . . . . . 27 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Current outputs . . . . . . . . . . . . . . . . . . . . . . . 26
CYK 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 K
K-Function group . . . . . . . . . . . . . . . . . . 41, 42
Kp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

80 Endress+Hauser
Liquisys M CUM 223 / 253 Index

UM253E12.CHP

L P
LED Display . . . . . . . . . . . . . . . . . . . . . . . . . 16 P controller . . . . . . . . . . . . . . . . . . . . . . . . . 32
Limit contactor . . . . . . . . . . . . . . . . . . . . 31, 35 P(ID) controller . . . . . . . . . . . . . . . . . . . 31, 32
Limit contactor for turbidity measured value 31 P-Function group. . . . . . . . . . . . . . . . . . . . . 30
Limit contactor for temperature . . . . . . . 31, 37 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Linear characteristic . . . . . . . . . . . . . . . . . . 26 Panel-mounted instrument . . . . . . . . . . . . . 11
Liquid crystal display . . . . . . . . . . . . . . . . . . 16 Partition plate. . . . . . . . . . . . . . . . . . . . . . . . . 7
Locking the keypad . . . . . . . . . . . . . . . . 17, 20 Pause time. . . . . . . . . . . . . . . . . . . . 34, 39, 40
PCS alarm . . . . . . . . . . . . . . . . . . . . . . . . . . 30
M PD controller . . . . . . . . . . . . . . . . . . . . . . . . 32
Maintenance . . . . . . . . . . . . . . . . . . . . . 56-60 PI controller . . . . . . . . . . . . . . . . . . . . . . . . . 32
Management of calibration data records . . 47 Pickup delay . . . . . . . . . . . . . . . . . . . . . 36, 37
Measurement mode . . . . . . . . . . . . . . . . . . . 22 PID controller. . . . . . . . . . . . . . . . . . . . . 32, 37
Measuring cable . . . . . . . . . . . . . . . . . . . . . 14 PLUS key . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Measuring cable connection . . . . . . . . . . . . 14 Post mounting kit. . . . . . . . . . . . . . . . . . . . . . 9
Measuring system . . . . . . . . . . . . . . . . . . . . . 6 Post mounting on a square post . . . . . . . . . . 8
Menu structure . . . . . . . . . . . . . . . . . . . . . . . 20 Post mounting on cylindrical pipes. . . . . . . . 8
Minimum pause time . . . . . . . . . . . . . . . 39, 40 Post-rinse time. . . . . . . . . . . . . . . . . . . . 35, 40
MINUS key . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Pre-rinse time . . . . . . . . . . . . . . . . . . . . 35, 40
Module name . . . . . . . . . . . . . . . . . . . . . . . . 44 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Monitoring features . . . . . . . . . . . . . . . . . . . . 5 Product structure. . . . . . . . . . . . . . . . . . . . . . 3
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Mounting examples . . . . . . . . . . . . . . . . . . . 10 Programming. . . . . . . . . . . . . . . . . . . . . . . . 20
Mounting of field instrument . . . . . . . . . . . . . 8 Pulse frequency . . . . . . . . . . . . . . . . . . . . . 38
Pulse frequency modulation . . . . . . . . . . . . 33
O Pulse length. . . . . . . . . . . . . . . . . . . . . . . . . 38
O-Function group. . . . . . . . . . . . . . . . . . 26-28 Pulse length modulation . . . . . . . . . . . . . . . 33
Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
OFFSET (V) . . . . . . . . . . . . . . . . . . . . . . . . . 54 Q
Offset mode . . . . . . . . . . . . . . . . . . . . . . . . . 22 Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
ON time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Operating concept. . . . . . . . . . . . . . . . . . . . 19 R
Operating keys. . . . . . . . . . . . . . . . . . . . . . . 17 R-Function group . . . . . . . . . . . . . . . . . . 35-40
Operating modes . . . . . . . . . . . . . . . . . . . . . 19 REL key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Operation . . . . . . . . . . . . . . . . . . . . . . 4, 16-21 RELAY (R) . . . . . . . . . . . . . . . . . . . . . . . 22, 35
Operator interface . . . . . . . . . . . . . . . . . . . . 16 Relay contact configuration . . . . . . . . . . . . 31
Optoscope . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Repeat cycles . . . . . . . . . . . . . . . . . . . . . . . 40
Order code. . . . . . . . . . . . . . . . . . . . . . . . 3, 43 Replacement of processor module . . . . . . 69
Residual concrete water . . . . . . . . . . . . . . . 48
Rest period between wiping cycles . . . . . . 25
Rinse time . . . . . . . . . . . . . . . . . . . . . . . . . . 39
RM junction box . . . . . . . . . . . . . . . . . . . . . 72

Endress+Hauser 81
Index Liquisys M CUM 223 / 253

S T
S-Function group . . . . . . . . . . . . . . . . . . 42, 43 Technical data . . . . . . . . . . . . . . . . . . . . 73-75
Safety features . . . . . . . . . . . . . . . . . . . . . . . . 5 Terminal blocks . . . . . . . . . . . . . . . . . . . . . . . 7
Safety instructions. . . . . . . . . . . . . . . . . . . . 4, 5 Termination of measuring cablel . . . . . . . . . 14
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Timer for cleaning function . . . . . . . . . . 31, 34
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . 2 Tn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Scopeware . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Selection of error. . . . . . . . . . . . . . . . . . . . . . 29 Troubleshooting . . . . . . . . . . . . . . . . . . . 56-60
Selection of language . . . . . . . . . . . . . . . . . . 42 Tv . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Sensor installation. . . . . . . . . . . . . . . . . . . . . 14 U
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Universal mounting post CYY 102 . . . . . . . . . 9
Serial number . . . . . . . . . . . . . . . . . . . . . . . . 44 Unlocking the keypad . . . . . . . . . . . . . . 17, 20
Service . . . . . . . . . . . . . . . . . . . . . . . . . . 42, 61 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
SERVICE (S) . . . . . . . . . . . . . . . . . . . . . . 24, 42 Upright post CYY 102 . . . . . . . . . . . . . . . . . 72
SETUP 1 (A) . . . . . . . . . . . . . . . . . . . . . . . . . 24 USP 23 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
SETUP 2 (B) . . . . . . . . . . . . . . . . . . . . . . 24, 25
Setup mode. . . . . . . . . . . . . . . . . . . . . . . . . . 22 V
Setup-Modus . . . . . . . . . . . . . . . . . . . . . . . . 19 V-Function group . . . . . . . . . . . . . . . . . . . . . 54
Shipping documents. . . . . . . . . . . . . . . . . . . . 2 VBM junction box . . . . . . . . . . . . . . . . . . 14, 72
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Version TB . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Software version . . . . . . . . . . . . . . . . . . . . . . 44
Spare parts for CUM 223 . . . . . . . . . . . . . . . 65 W
Spare parts for CUM 253 . . . . . . . . . . . . . . . 68 Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . 8
Spare parts orders . . . . . . . . . . . . . . . . . . . . 69 Weather protection cover CYY 101 . . . . . 8, 72
Start pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Wiper controller . . . . . . . . . . . . . . . . . . . . . . 25
start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Wiping cycle. . . . . . . . . . . . . . . . . . . . . . . . . 25
Start-up menu . . . . . . . . . . . . . . . . . . . . . . . . 24
Stch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Structure of measuring cable . . . . . . . . . . . . 14
SW upgrade . . . . . . . . . . . . . . . . . . . . . . . . . 43
Switch-off point of contact . . . . . . . . . . . . . . 36
Switch-off temperature . . . . . . . . . . . . . . . . . 37
Switch-on point of contact . . . . . . . . . . . . . . 36
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . 2
System configuration . . . . . . . . . . . . . . . . . . 24

82 Endress+Hauser
Declaration of contamination
Dear costumer,
Because of legal determinations and for the safety of our employes and operating equipment we need this
“Declaration of contamination” with your signature before your order can be handled. Please put the completely
filled in declaration to the instrument and to the shipping documents in any case. Add also safety sheets and/or
specific handling instructions if necessary.

type of instrument / sensor: __________________________________ serial number: _______________________

medium / concentration: __________________________________ temperature: ______ pressure: _______

cleaned with: __________________________________ conductivity: ______ viscosity: _______

:DUQLQJKLQWVIRUPHGLXPXVHG

SAFE

o o o o o o o o
radioactive explosive caustic poisonous harmful of biological inflammable safe
health hazardous
Please mark the appropriate warning hints.

5HDVRQIRUUHWXUQ
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

&RPSDQ\GDWD

company: ______________________________ contact person: _________________________


______________________________ _________________________
______________________________ department: _________________________
address: ______________________________ phone number: _________________________
______________________________ Fax/E-Mail: _________________________
______________________________ your order no.: _________________________

I hereby certify that the returned equipment has been cleaned and decontaminated acc. to good industrial prac-
tices and is in compliance with all regulations. This equipment poses no health or safety risks due to contamination.

_______________________________ ___________________________________
(Date) (company stamp and legally binding signature)

More information about services and repairs:


www.services.endress.com
Norway Brazil Papua New Guinea
Europe ❑ Endress+Hauser A/S ❑ Samson Endress+Hauser Ltda. SBS Electrical Pty Limited
Tranby Sao Paulo Port Moresby
Austria Tel. ++47 (32) 85 98 50, Fax (32) 85 98 51 Tel. ++55 (11) 5 0 31 34 55, Tel. ++6 75 (3) 25 11 88, Fax (3) 25 95 56
❑ Endress+Hauser Ges.m.b.H. Fax (11) 5 0 31 30 67
Wien Poland Philippines
Tel. ++43 (1) 8 80 56-0, Fax (1) 8 80 56-35 ❑ Endress+Hauser Polska Sp. z o.o. Canada Brenton Industries Inc.
Warszawy ❑ Endress+Hauser Ltd. Makati Metro Manila
Belarus Tel. ++48 (22) 7 20 10 90, Fax (22) 7 20 10 85 Burlington, Ontario Tel. ++63 (2) 6 38 80 41, Fax (2) 6 38 80 42
Belorgsintez Tel. ++1 (9 05) 6 81 92 92,
Minsk Portugal Fax (9 05) 6 81 94 44 Singapore
Tel. ++3 75 (1 72) 26 31 66, Fax (1 72) 26 31 11 Tecnisis - Tecnica de Sistemas Industriais ❑ Endress+Hauser (S.E.A.) Pte., Ltd.
Linda-a-Velha Chile Singapore
Belgium / Luxembourg Tel. ++3 51 (1) 4 17 26 37, Fax (1) 4 18 52 78 DIN Instrumentos Ltda. Tel.++65 (5) 66 82 22, Fax (2) 66 68 48
❑ Endress+Hauser S.A./N.V. Santiago
Brussels Romania Tel. ++56 (2) 2 05 01 00, Fax (2) 2 25 81 39 South Korea
Tel. ++32 (2) 2 48 06 00, Fax (2) 2 48 05 53 Romconseng SRL
❑ Endress+Hauser (Korea) Co., Ltd.
Bucharest Colombia Seoul
Bulgaria Tel. ++40 (1) 4 10 16 34, Fax (1) 4 10 16 34 Colsein Ltd.Bogota D.C. Tel. ++82 (2) 6 58 72 00, Fax (2) 6 59 28 38
INTERTECH-AUTOMATION
Sofia Russia Tel. ++57 (1) 2 36 76 59, Fax (1) 6 10 78 68
❑ Endress+Hauser Moscow Office Taiwan
Tel. ++3 59 (2) 66 48 69, Fax (2) 9 63 13 89 Kingjarl Corporation
Moscow Costa Rica
Croatia Tel.++7 09 (5) 1 58 75 71, Fax (5) 1 58 98 64 Taipei R.O.C.
EURO-TEC S.A. Tel. ++886 (2) 27 18 39 38, Fax (2) 27 13 41 90
❑ Endress+Hauser GmbH+Co. San Jose
Zagreb Slovak Republic Tel. ++5 06 (2) 96 15 42, Fax ( 2 ) 96 15 42
Tel. ++3 85 (1) 6 63 77 85, Fax (1) 6 63 78 23 Transcom Technik s.r.o. Thailand
Bratislava ❑ Endress+Hauser Ltd.
Ecuador Bangkok
Cyprus Tel. ++4 21 (74) 4 88 86 84, Fax (74) 4 88 71 12 Insetec Cia. Ltda.
I+G Electrical Services Co. Ltd. Tel. ++66 (2) 9 96 78 11 20, Fax (2) 9 96 78 10
Slovenia Quito
Nicosia Tel. ++5 93 (2) 26 91 48, Fax (02) 46 18 33
Tel. ++3 57 (2) 48 47 88, Fax (2) 48 46 90 ❑ Endress+Hauser D.O.O. Vietnam
Ljubljana Guatemala Tan Viet Bao Co. Ltd.
Czech Republic Tel. ++3 86 (61) 1 59 22 17, Fax (61) 1 59 22 98 ACISA Automatizacion Y Control Industrial S.A. Ho Chi Minh City
❑ Endress+Hauser GmbH+Co. Ciudad de Guatemala, C.A. Tel. ++84 (8) 8 33 52 25, Fax (8) 8 33 52 27
Praha Spain
❑ Endress+Hauser S.A. Tel. ++5 02 (3) 34 59 85, Fax (2) 32 74 31
Tel. ++4 20 (26) 6 78 42 00, Fax (26) 6 78 41 79
Barcelona
Mexico Iran
Denmark Tel. ++34 (93) 4 80 33 66, Fax (93) 4 73 38 39
❑ Endress+Hauser I.I.
❑ Endress+Hauser A/S Telephone Technical Services Co. Ltd.
Sweden Mexico City Tehran
Søborg
❑ Endress+Hauser AB Tel. ++52 (5) 56 89 65, Fax (5) 56 84 18 Tel. ++98 (21) 8 74 67 50, Fax (21) 8 73 72 95
Tel. ++45 (70) 13 11 32, Fax (70) 13 21 33
Sollentuna
Estonia Tel. ++46 (8) 55 51 16 00, Fax (8) 55 51 16 00 Paraguay
Incoel S.R.L. Israel
Elvi-Aqua Instrumetrics Industrial Control Ltd.
Tartu Switzerland Asuncion
Tel. ++5 95 (21) 21 39 89, Fax (21) 22 65 83 Tel-Aviv
Tel. ++3 72 (7) 42 27 26, Fax (7) 42 27 27 ❑ Endress+Hauser AG Tel. ++9 72 (3) 6 48 02 05, Fax (3) 6 47 19 92
Reinach/BL 1
Finland Tel. ++41 (61) 7 15 75 75, Fax (61) 7 11 16 50 Uruguay
Circular S.A. Jordan
❑ Endress+Hauser Oy A.P. Parpas Engineering S.A.
Espoo Turkey Montevideo
Intek Endstriyel lc ve Kontrol Sistemleri Tel. ++5 98 (2) 92 57 85, Fax (2) 92 91 51 Amman
Tel. ++3 58 (9) 8 59 61 55, Fax (9) 8 59 60 55 Tel. ++9 62 (6) 4 64 32 46, Fax (6) 4 64 57 07
Istanbul
France Tel. ++90 (2 12) 2 75 13 55, USA
❑ Endress+Hauser ❑ Endress+Hauser Inc. Kingdom of Saudi Arabia
Fax (2 12) 2 66 27 75
Huningue Greenwood, Indiana Anasia
Tel. ++33 (3) 89 69 67 68, Fax (3) 89 69 48 02 Ukraine Tel. ++1 (3 17) 5 35 71 38, Jeddah
Industria Ukraïna Fax (3 17) 5 35 84 89 Tel. ++9 66 (2) 6 71 00 14, Fax (2) 6 72 59 29
Germany Kiev
❑ Endress+Hauser Metechnik GmbH+Co. Tel. ++3 80 (44) 2 68 81, Fax (44) 2 69 08 Venezuela Kuwait
Weil am Rhein H. Z. Instrumentos C.A. Kuwait Maritime & Mercantile Co. K.S.C.
Tel. ++49 (76 21) 9 75 01, Fax (76 21) 9 7 55 55 Yugoslavia Caracas Safat
Meris d.o.o. Tel. ++58 (2) 9 44 09 66, Fax (2) 9 44 45 54 Tel.++9 65 (2) 44 14 81, Fax (2) 44 14 86
Great Britain Beograd
❑ Endress+Hauser Ltd. Tel. ++3 81 (11) 4 44 61 64, Fax (11) 4 44 19 66 Lebanon
Manchester
Tel. ++44 (1 61) 2 86 50 00, Asia Nabil Ibrahim
Jbeil
Fax (1 61) 9 98 18 41
Africa China
Tel. ++9 61 (3) 25 40 52, Fax (9) 54 80 38
Greece ❑ Endress+Hauser Shanghai Sultanate of Oman
I & G Building Services Automation S.A. Egypt Instrumentation Co. Ltd. Mustafa & Jawad Science & Industry Co.
Athens Anasia Shanghai L.L.C.
Tel. ++30 (1) 9 24 15 00, Fax (1) 9 22 17 14 Heliopolis/Cairo Tel. ++86 (21) 54 90 23 00, Ruwi
Tel. ++20 (2) 4 1 79 00, Fax (2) 4 1 79 00 Fax (21) 54 90 23 03 Tel. ++9 68 (60) 20 09, Fax (60) 70 66
Hungary
Mile Ipari-Elektro Morocco ❑ Endress+Hauser Beijing Office United Arab Emirates
Budapest Oussama S.A. Beijing Descon Trading EST.
Tel. ++36 (1) 2 61 55 35, Fax (1) 2 61 55 35 Casablanca Tel. ++86 (10) 68 34 40 58, Dubai
Tel. ++2 12 (2) 24 13 38, Fax (2) 40 26 57 Fax (10) 68 34 40 68 Tel. ++9 71 (4) 65 36 51, Fax (4) 65 32 64
Iceland
Vatnshreinsun HF Nigeria
Reykjavik J F Technical Invest. Nig. Ltd. ❑ Endress+Hauser (H.K.) Ltd. Yemen
Lagos Hong Kong Yemen Company for Ghee and Soap Industry
Tel. ++3 54 (5) 61 96 16, Fax (5) 61 96 17
Tel. ++2 34 (1) 62 23 45 46, Fax (1) 62 23 45 48 Tel. ++8 52 (2) 5 28 31 20, Fax (2) 8 65 41 71 Taiz
Ireland Tel. ++9 76 (4) 23 06 64, Fax (4) 21 23 38
Flomeaco Company Ltd. South Africa India
Kildare ❑ Endress+Hauser Pty. Ltd. ❑ Endress+Hauser India Branch Office
Tel. ++3 53 (45) 86 86 15, Fax (45) 86 81 82 Sandton
Tel. ++27 (11) 4 44 13 86, Fax (11) 4 44 19 77
Mumbai
Tel. ++91 (22) 8 52 14 58, Fax (22) 8 52 19 27 Australia + New Zealand
Italy
❑ Endress+Hauser Italia S.p.A. Tunisia Indonesia Australia
Cernusco s/N Milano Controle, Maintenance et Regulation PT Grama Bazita ALSTOM Australia Ltd.
Tel. ++39 (02) 92 10 64 21, Tunis Jakarta Sydney
Fax (02) 92 10 71 53 Tel. ++2 16 (1) 79 30 77, Fax (1) 78 85 95 Tel. ++62 (21) 7 97 50 83, Fax (21) 7 97 50 89 Tel. ++61 (2) 97 22 47 77, Fax (2) 97 22 48 88

Latvia Japan New Zealand


Raita Ltd.
Riga America ❑ Sakura Endress Co., Ltd.
Tokyo
EMC Industrial Instrumentation
Auckland
Tel. ++3 71 (7) 31 28 97, Fax (7) 31 28 94 Tel. ++81 (4 22) 54 06 11, Fax (4 22) 55 02 75 Tel. ++64 (9) 4 15 51 10, Fax (9) 4 15 51 15
Argentina
Lithuania ❑ Endress+Hauser Argentina S.A. Malaysia
Agava Ltd.
Kaunas
Buenos Aires
Tel. ++54 (1) 1 45 22 79 70,
❑ Endress+Hauser (M) Sdn. Bhd.
Petaling Jaya, Selangor Darul Ehsan All other countries
Tel. ++3 70 (7) 20 24 10, Fax (7) 20 74 14 Fax (1) 1 45 22 79 09 Tel. ++60 (3) 7 33 48 48, Fax (3) 7 33 88 00
❑ Endress+Hauser GmbH+Co.
Netherlands Bolivia Pakistan Instruments International
❑ Endress+Hauser B.V. Tritec S.R.L. Speedy Automation D-Weil am Rhein
Naarden Cochabamba Karachi Germany
Tel. ++31 (35) 6 95 86 11, Fax (35) 6 95 88 25 Tel. ++5 91 (42) 5 69 93, Fax (42) 5 09 81 Tel. ++92 (21) 7 72 29 53, Fax (21) 7 73 68 84 Tel. ++49 (76 21) 9 75 02, Fax (7 6 21) 97 53 45

❑ Members of the Endress+Hauser group

Endress Hauser
51500274 The Power of Know How

BA 200C/07/en/12.01
Printed in Germany / CV5 / DT