Professional Documents
Culture Documents
Service Manual
Legal & warranty..............................................................................4
Safety Information...........................................................................5
References to other documentation ....................................................................................... 6
Intended Use..........................................................................................................................7
References to other documents ............................................................................................. 8
Tools required ........................................................................................................................ 9
General description ........................................................................10
Block Diagrams (overview) ............................................................11
General Theory of operation...........................................................12
ID Module ............................................................................................................................12
Dispensing Module...............................................................................................................12
Wash and Soak Module ....................................................................................................... 13
Analyse module ................................................................................................................... 14
Turning plate module ........................................................................................................... 14
Climate module .................................................................................................................... 14
User interface module ..........................................................................................................15
Printer module...................................................................................................................... 16
Mechanics............................................................................................................................ 16
Teleca Instrument Platform ..................................................................................................17
Hardware .............................................................................................................................18
Service Software on PC .......................................................................................................18
Maintenance...................................................................................19
Preventive Maintenance....................................................................................................... 19
Basic Function test instruction and record............................................................................ 20
Annual preventive maintenance instruction and record ........................................................21
Service, Trouble shooting...............................................................22
Initial information:................................................................................................................. 22
Symptoms ......................................................................................23
Service & Adjustment/replacement procedures.................................................................... 25
Error codes and actions ....................................................................................................... 26
Service centers...............................................................................37
Instruction: how to remove the front air filter ...............................38
Instruction: how to exchange dispense needles............................. 46
Instruction: how to exchange wash needles ..................................50
Instruction: how to exchange filter in bottles ................................54
Instruction: how to replace rear ventilation filter .......................... 57
Instruction: how to replace the fuses.............................................60
Instruction: how to adjust the tension of the dispense unit belt .... 63
Instruction: how to adjust the tension of the turning table belt.....66
Instruction: how to replace the printer .......................................... 69
Instruction: how to cut and assemble fluid lines and assemble couplings
.......................................................................................................72
2(101)
Instruction: how to remove the covers ..........................................76
Instruction: how to replace the OD module.................................... 91
3(101)
LEGAL & WARRANTY
The content of this manual are the property of ORGENTEC Diagnostika GmbH. Any
reproduction in whole or in part is strictly prohibited.
At the time of printing, this manual correctly described the device and its functions.
However, as modifications may have been carried out since the production of this
manual, the system package contains this manual and a service manual and may
contain one or more amendments to the manuals. This manual including the service
manual and any amendments must be thoroughly read, before using the device.
ORGENTEC Diagnostika GmbH is only responsible for the reliability and performance of
the device if the following are strictly observed:
Authorized personnel (see back of manual for authorized service centres) carry out all
service, repairs and modifications.
The device must be used in accordance with the intended use and the instructions put
forth in the Safety Information section (instruction for use and this manual).
ORGENTEC Diagnostika GmbH offers one-year warranty, from the day of purchase, on
defective material and assembly.
The warranty does not cover damage resulting from incorrect use or user
maintenance or from authorized software modification.
ORGENTEC Diagnostika GmbH is only responsible for replacement of defective parts,
not wear of parts.
ORGENTEC Diagnostika GmbH is not responsible for any personal injury or any
damage resulting from incorrect use or incorrect performed service of the analyser.
If the above points are not strictly observed, the warranty will be considered invalid.
4(101)
SAFETY INFORMATION
This product is CE marked in conformity with the requirements of the IVD Directive
98/79/EC. Adhere to the following recommendations for safe and proper operation of the
device:
5(101)
References to other documentation
The below topics are described in Instructions for use (Use the Instructions for use
together with this service manual).
Basic Information
Packaging/un-packaging Procedure
Packaging, Consumables and spare part list
Symbols
Installation
General operation
Overview descriptions
Setup diagram
Software description
Shutdown
Replacement procedures
How to load and exchange sample plates/ samples
How to empty waste fluid and load buffers
How to load printer paper
How to remove Transport locking
Calibration
Software upgrade
Maintenance
Cleaning
Waste disposal
Technical specifications
Measures and weights
Classification
Operation conditions
Storage and transport conditions
6(101)
Intended Use
The Alegria analyser is intended for use by trained personnel to process and read IVD
test strips supplied by the ORGENTEC. The results obtained shall be used only in
conjunction with other medical evidence to diagnose conditions of the human body. The
Alegria analyser by it-self is not diagnostic. The instrument is to be used in medical
laboratories.
All service of the Alegria analyser is to be carried out by authorized service personnel.
This manual and the other manuals such as Intended Use shall be consulted for an
overview.
The intention of this service manual is to give an overview of the main parts and
systems in the instrument. References to other locations, such as manuals drawings or
part lists are to follow, if the referred document is not amended to the manual you may
need to contact you representative.
The manual will describe certain or common parts or procedures of interest and not all
details, i.e. the most common foreseeable issues that can be of interest.
7(101)
References to other documents
The below documents are of engineering type and require specific knowledge to
comprehend, if there are questions do not hesitate to contact your representative for
assistance.
Title Content
Instructions for Use Description of normal operation and technical data
Quick Guide A quick reference on how to get started using the
instrument
Service software manual Description of Service software
8(101)
Tools required
The table below gives an indication on the required tools for servicing the instrument.
9(101)
GENERAL DESCRIPTION
The Alegria analyser is built up by a large number of different modules. Most of the
modules
contains hardware, software and mechanics.
The general functionality of the instrument is best described as an instrument that can
be loaded with test strips containing different fluids that the instrument will:
Collect, Mix and Analyse according to a test protocol (the user defines the protocol). The
results are presented on display and printed. To accomplish these things there are a
need for different modules (see general description). These modules are interconnected
and controlled by the instruments software.
ID module
Dispensing module
Wash and soak module
Analysis module
Turning plate module
Climate control module
User Interface module
Printer module
Teleca Instrument Platform
Service PC module
Motorized Sample bay
10(101)
BLOCK DIAGRAMS (OVERVIEW)
All block diagrams are electronically appended:
Wiring diagram
Internal cables diagram
Note:
Click the link and the document will be shown.
11(101)
GENERAL THEORY OF OPERATION
The Alegria analyser different modules are described in the following sections (list of
modules).
ID Module
The ID module consists of two barcode readers, one internal and one external. The
external is optional for the end user to use.
The module is responsible for linking patient data with a sample strip by reading the bar
code attached on the strip (CODE 128) with an external bar code reader. The module
also handles the identification and positioning of the test strips in the instrument with an
internal bar code reader (fixed).
The internal barcode reader is mounted to the chassis in a fixed position. The external
barcode reader is mounted to a holder on the right side of the instrument. It is
connected to a COM port on the back of the instrument and can be removed for
customers that do not wish to use it.
Dispensing Module
The Dispensing module should dispense sample and the liquids that are already pre-
filled in the strip (conjugate, substrate, cut-off, sample buffer) into other wells in the
strip. It moves in radial direction compared to the turning plate and in vertical direction.
The Dispensing module consists of a precision pump and two separate needles, one for
use of dispensing conjugate and one for use of dispensing the other liquids
(programmed by process developer with the Alegria Developer SW).
The separate needle for conjugate is to ensure that there is no risk for cross
contamination of conjugate with the other liquids.
The precision pump is set to dispense volumes in a range of 10-200 µl and smallest step
is 1µl, (The volumes to dispense are programmed by process developer with the Alegria
Developer SW).
Each needle can move independently of the other down into a well and up above the
strip so the carriage can reposition itself over the next well. Radially the distance
between the two needles is fixed so when one needle has to be moved in this direction
the other one will follow.
The whole needle feed tubing is filled with hydraulic fluid (distilled water) to achieve as
high accuracy as possible. All tubing’s and needles for the dispense system can be
washed with either distilled water or wash buffer.
12(101)
Wash and Soak Module
This Wash and Soak module actually consists of three separate stations, two identical
“wash & soak” stations and one dry suction station.
The “wash & soak” stations main task is to put wash buffer solution inside the two
reaction wells on the strips. The stations do not perform the actual cleaning of the wells.
This is done chemically by the wash buffer solution.
The station only removes whatever liquid present in the wells and thereafter fills it with
wash buffer. The reason that there are two identical stations is that the cleaning
procedure should be performed twice. The turning plate positions a strip under the first
station that sucks the reaction wells dry and fills them with wash buffer.
Then the turning plate moves on until the strip ends up underneath the other station.
The wells are sucked dry and filled once more.
The “drier” station main task is to remove wash buffer solution inside the two reaction
wells on the strips. The turning plate positions a strip under the station that sucks the
reaction wells dry.
All tubing’s and needles for the wash and soak system can be washed with either
distilled water or wash buffer.
Placement of stations
The three stations are placed inside the instrument so that there are two “empty”
rotation steps between each station, this means that if each step takes 30 seconds it will
take a strip 1 minute to pass from the “end” of one station to the “start” of the next
station.
This is since when the strip is filled with wash buffer it shall react for at least 30 seconds
in the well.
13(101)
Analyse module
This Analyse module measures the optical density of the test samples. It has two
channels, positioned over two adjacent wells. The measurements are made at 650 nm
and the range is 0 to 2,0 OD (Optical Density). LED´s are used as light source and
before each measurement a reference voltage is measured and the same voltage is then
used when measuring. This leads to a very accurate and self-calibrating system.
The turning plate is a rotating table inside the climate module, on the turning plate three
individual sample plates can be loaded. Each sample plate can hold up to 10 test strips.
The three sample plates are individually keyed to the turning plate so the user cannot
put them in the wrong position.
Climate module
The functionality of the Climate module is to keep a steady and controlled environment
around the test strips heating or cooling dependent on what’s needed. It is designed as
a closed compartment with the inner volume as small as possible to facilitate an efficient
temperature control. The compartment is designed as round “thick” plate with enhanced
insulation on all sides. In the middle of the “roof” of the climate chamber a peltier
element with a heat sink and a small fan is placed to evenly spread the heat/cooling in
the climate chamber.
Below the heat sink there is a “collector” in case there will be humidity drops, this will
then be collected an lead out through a hole in the engine shaft of the turning plate
module.
The temperature in the controlled environment around the test strips shall be 26 2 °C
during the analysis run. The ambient conditions are specified to a temperature of 20-
32°C and a relative humidity of 25 – 75% this in combination that the dew point shall
not exceed 23°C.
14(101)
User interface module
The user interfaces are defined to parts that the user can interact with, such as:
Touch screen
A 10-inch colour touch screen is used for the instrument.
Level sensors
The fluid bottles used in the instrument are:
Two 1L bottles for wash buffer.
One 2,5 L bottle for distilled water.
One 5 L bottles for waste.
The wash buffer and distilled water bottles are each equipped with a level sensors that
will warn if there is not enough liquid in the bottle to perform a analyse with 30 test
strips. If these sensors are disconnected from the instrument the instrument will warn
for low level in the bottles.
The waste bottle is equipped with a level sensor that will warn if there is not enough
space in the bottle to perform a analyse with 30 test strips. If this sensor is disconnected
from the instrument the instrument will warn for high level in the bottle.
External connections
Connections for modem, service PC, external USB printer and external barcode reader
are located at the back of the instrument.
ON/OFF switch
The mains power ON/OFF switch is located at the back of the instrument in the lower
right corner when standing in front of the instrument.
15(101)
Printer module
The instrument is equipped with an internal thermo-printer with a paper width off 112
mm. After the analyse sequence is finished the results are automatically printed
The printer is placed in the right front corner of the instrument. To replace the printer
paper, open the lid and the printer compartment door. The printer paper is fed out
through the printer compartment door and the fluid bottle lid/door.
An external printer can be connected to the USB port of the instrument and the result
will then be printed on this.
Mechanics
The mechanics has been divided into two categories, external and internal:
External Mechanics
The covers consists of four items made from PU, integral skin foam, front lid, top
cover and two identical side covers
The front lid is mounted to a sheet metal structure, fastened to the bottom plate by
hinge bolts and held closed by a magnetic lock.
The side covers are mounted to a sheet metal structure, which is fastened to the
bottom plate.
The top cover is mounted to a sheet metal plate fastened to the bottom of the base
plate, the back plate and indexed to the side covers.
The robot is adjusted relative to the top cover to reduce the tolerance chain.
The display is screwed up against the top cover, lifting it free from the underlying
structure. The screws are covered by screw caps.
Internal Mechanics
Waste pump:
This pump disposes of unwanted fluids. The pump can be used simultaneously by
several components and hence runs until the last component turns it off.
Robot structure:
The main inner mechanics part is the robot structure where most off the modules are
built up. ID-Module (Internal Identification), Dispensing module, Wash and Soak
module, Analysis Module (Light source and Analysis), Turning Plate Module, Climate
Module. When built up the modules can be test run on the robot level before the robot
is actually built into the instrument. This is to facilitate easy mounting and module
verification.
Cabling and cable clips:
Routing of cables and tubes for the purpose to make sure that the cabling does not
disturb any of the movements.
Mounting mechanics for PCBs:
The PCB´s are mounted on sheet metal parts that also act as parts of the supporting
structure. The touch display is mounted on a “tower” that connects with the back
panel and stabilizes it.
16(101)
Teleca Instrument Platform
Teleca Instrument Platform (TIP) is designed to be a flexible foundation for a wide range
of products. The TIP is based on distributed processing, with a main processor
communicating with slave processors via a serial link, “Local Bus”. The Alegria version of
TIP consists of three different parts:
The motherboard
The I/O-board
The analysis board
17(101)
Hardware
The hardware architecture spans over the modules and each electronics board handles
functionality for several modules.
High power drivers for motors, pumps and valves are located on the stepper board
together with corresponding sensor inputs for home sensors, quadrature encoders etc.
The main board handles communication, display interface and high-level logic.
The analysis module has its own hardware that also measures the chamber temperature
for the climate module.
Service Software on PC
From an external PC the user can connect to the instrument to download and test run
the different modules on the instrument. There are three different types of PC SW for
three different users.
End user
The end user can download a copy of the patient archive
Service technician
Same as end user and test run the different modules
Process developer
Same as service technician and create and download new analyse files to the
instrument.
18(101)
MAINTENANCE
On a weekly basis, test the device for proper function (basic function test) and inspect
cables for cuts and other damage. If in doubt, replace the relevant parts. Additional
Preventive maintenance is required according to the Service manual (see preventive
maintenance).
Preventive Maintenance
The ORGENTEC Diagnostika instrument needs to be maintained properly. Preventive
maintenance and testing needs to be performed annually to ensure proper function of
the instrument and to ensure a long operational life.
Follow the annual preventive maintenance instruction and document the maintenance in
the record field. Check each item according to the instruction and make necessary
adjustments. If in doubt replace the relevant parts.
Also check to see that none of the instrument components have been removed or
modified. If so replace with original parts.
19(101)
Basic Function test instruction and record
OK Inspected/checked/tested ok RA Repaired
NOK Inspected/checked/tested Not ok RL Replaced
C Cleaned NA Not applicable
A Adjusted
20(101)
Institution: Date:
Model: Serial number:
Check washing Verify that washing of reaction wells and washing of needles
function function properly. Also check dispensing and suction motors,
dispensing and turning plate motor. Check the liquid level in cavities
after washing. Inspect the needles for wear, if broken or has low
performance then replace.
Replace dispense Check the dispensing (volume 100 µl) of both dispensing needles.
needle Inspect for wear, if broken or has low performance then replace
(standard needle once a year; conjugate needle every 3 years).
Temperature Double check the internal temperature inside the climate chamber
against a calibrated external thermometer.
Performance test / Perform a run with Positive Control (ORG 310) and optional
QC test Negative Control (ORG 311) test devices.
Date:
OK Inspected/checked/tested ok RA Repaired
NOK Inspected/checked/tested Not ok RL Replaced
C Cleaned NA Not applicable
A Adjusted
ORGENTEC Diagnostika GmbH Postfach 100352 D-55134 Mainz / Germany
FD4.25F Maintenance Checklist; Version 3 valid from 2012-01-02
Page 1/1
SERVICE, TROUBLE SHOOTING
Initial information:
Begin gathering information about the problem. Start by asking the operator about
the issue to make it easier to find the problem, and to verify the problem.
Note: After any corrective action the preventive maintenance instruction should be
performed.
WARNING
Only facility-authorized personnel should trouble shoot the ORGENTEC Diagnostika
Alegria Analyser. Troubleshooting by unauthorized personnel could result in personal
injury, equipment damage or property damage.
WARNING
Follow the removal procedure if replacing any part, otherwise serious injury, equipment
damage or property damage could occur.
If you have a Instrument or spare part you can interchange a part to verify the problem.
If the problem persists then see the symptom section and its corrective actions.
22(101)
SYMPTOMS
Below is a list of fault symptoms, find the one(s) that is most similar to the problem
experienced with the instrument and follow the instruction(s).
Note these symptoms are the most anticipated.
Symptom Instruction
Printer quality is degraded or printout How to replace the printer
fails
Fluid line leaks How to cut and assemble fluid lines
Fluid is not distributed. Check the fluid bottle filters and if saturated
–Replace, See How to replace fluid bottle(s)
filters
Air bubbles in system. Ensure that fluid bottle filters are in the right
position in bottles (close to bottom of bottle
and not on surface of fluid).
Check for air leakage. How to cut and
assemble fluid lines
The dispense unit will not move: Check that the transport locking is removed,
see instructions for use
Dispense needles are not moving Check that the needle can move freely as
descried in How to change dispense needles,
if the spring is worn out replace the needle
and spring
Dispense needles are jammed If or for some reason the needle has been
bent, remove it and replace, see How to
change dispense needles
Wash(and soak) needles are jammed If or for some reason the needle has been
bent, remove it and replace, see How to
change wash needles
The dispense unit will not move:
Correctly How to adjust the tension of dispense unit
belt
Check that home sensor works properly, in z
direction for needles and in x-direction for
carriage.
Not at all How to adjust the tension of dispense unit
belt
If belt drive is broken –replace and follow the
how to adjust the tension instruction
Check in error log if encoder works properly.
The turning table will not rotate:
Correctly How to adjust the tension of turning table belt
Check that home sensor works properly.
Not at all How to adjust the tension of turning table belt
If belt drive is broken –replace and follow the
how to adjust the tension instruction
Check in error log if encoder works properly.
Incorrect volumes when dispensing. Check that the needle can move freely in z-
direction as descried in How to change
dispense needles.
23(101)
Symptom Instruction
(continue) Incorrect volumes when Check if there are air bubbles in hoses or in
dispensing. precision pump.
Ensure that connections are correctly
tightened to prevent air leakage See,
How to cut and assemble fluid lines
Flush system to remove bubbles, use an
alcohol dilution if air bubbles in precision
pump remains.
Check if needles are clogged.
Flooding inside climate chamber. Check waste pump functionality, motor
functional, clogging in hoses.
Check wash and soak stations aspiration
needles that they aspirate in correct speed.
Check wash and soak stations feed needles
that position the yet inside the reaction strips.
If not check needles are bent? Check
calibration of carousel.
Could it be condense water from heat sink?
Check that drainage from heat sink is not
clogged. Also check that heat sink collector is
not bent or damaged.
Internal barcode reader fails to read Check that the sample barcode is of type
barcodes. CODE 128.
Check that barcode is not damaged or
destroyed.
Check if it’s possible to read the specified
barcode in another position in instrument.
Yes: Is the sample holder within correct
tolerances.
No: Barcode reader not assembled in correct
position.
External barcode reader fails to read Check what standards the barcode reader are
barcodes. programmed to read.
Unstable yet on wash and soak Check for air leakage. How to cut and
stations when feeding fluid into assemble fluid lines
reaction wells or wash cups.
Droplets on needles. Check for air leakage. How to cut and
assemble fluid lines
Check that back valves are working correctly
After any repair work on the Instrument the “Preventive maintenance instruction” must
be completed this to ensure the safe operation of the instrument.
If none of the fault symptoms corresponds to the problem experienced with the
Instrument then perform the “Preventive maintenance instruction” to locate the fault. If
problem still persists please contact your local ORGENTEC Diagnostika representative.
24(101)
Service & Adjustment/replacement procedures
The table below describes the adjustment and replacement procedures that may be
required.
Procedure
How to remove instrument covers
How to cut and assemble fluid lines
How to replace the printer
How to change the fuse
How to replace fluid bottle(s) filters
How to replace rear air filter
How to replace front Air filter
How to change dispense needles
How to replace wash needles
How to adjust the tension of dispense unit belt
How to adjust the tension of turning table belt
How to replace the Optical Density module
25(101)
Error codes and actions
The table below describes the error codes that the instrument can generate and
appropriate measures.
26(101)
xSpeed
xRamp
zFastSpeed
zSlowSpeed
zFastRamp
zSlowRamp
pumpSpeed
pumpRamp
stationPosition
ConOffset
zWashNeddlePos
wellZpos
Failed to read Shake from
performance description.ini file
Failed to read Timeslot from
performance description.ini file
5001 The Dispense process description has ERROR: To or from well >8 or <1 &&
wrong values or format != 9
5002 DispCRVAPIError The system cleaning procedure has
failed to:
get Digitalotput
set Digitalotput
get Digitalinput
getActive
The system dispense procedure has
failed to:
pumping with DC motor
get Disp_pip_choice Digitalotput
set Disp_pip_choice Digitalotput to
active.
dispense Wash Needle Radial-
>moveAbs
dispense Wash Needle, Zmotor-
>moveAbs(%d)
dispense Wash Needle,Pump-
>moveAbs(%d)
get Digitalotput (dispWaterFeed)
set DODispWaterFeed Digitaloutput
Dispensing Failed CRVAPI Error
The lid movement procedure has
failed to:
Radial to the door open position
The performance test and start test
run failed:
Make sure washing the needles with
distilled water Failed to get
Digitalotput
6000 TimeSlotError / AnalyteError One step has taken to long time.
7000 Wash1CRVAPIError The setting of the valves in position
wash failed to get an Digitalotput
7001 Wash1InitError Not available
8000 Wash2CRVAPIError The drier process is involed.
Drier station failed pumping with D
Failed going down with the drier z-
27(101)
motor
8001 Wash2InitError The drier process was failed to
initialize.
Failed Drier Init
The wash process was failed to
initialize.
Failed During Wash2 Init
9000 Wash2InitError The close valve “InDry” process was
failed.
Failed to get Digitalotput
The drier station process was failed.
Failed to open valve
The open, close, cleaning and drieing
process was failed.
Failed to get Digitalotput
The dry station process was failed.
Failed to set valve
The open valve “InDry” process was
failed.
Failed to get Digitalotput
The going down to rinse position
process including the going down with
drier Z motor failed.
The going up with the needle process
including the going up with the drier Z
motor to home position failed.
9001 DrierInitError Not available
10000 OpticalCRVAPIError These errors describe the full range of
the optical measurement procedure:
ActivateODError
Error setting
DO_OD_MEASURE_REF active
ODMeasError
Error CRV_TIME_OUT waiting on
DO_OD_MEASURE_REF
Error %d waiting on
DO_OD_MEASURE_REF
ReadLedAndRefValues
// Read reference measurements
from OD module
Error reading AI_OD_LED1 value
Error reading AI_OD_LED2 value
Error reading AI_OD_REF1 value
Error reading AI_OD_STD_REF1
value
Error reading AI_OD_STD_REF2
value
Error reading AI_OD_STD_REF2
value
ReadSensorValues
// Read Sensor measurements from
OD module
Error reading AI_OD_SENSOR1
value
28(101)
Error reading AI_OD_SENSOR2
value
Analyse2Completed
// Reading result from OD module
Error reading AI_OD_RESULT1
value
Error reading AI_OD_RESULT2
value
DoODMeasurement
Error setting
DO_OD_MEASURE_REF active
Starting OD Measurement (ref)
DoRefMeasurement
Starting OD Ref Measurement
Error setting
DO_OD_MEASURE_REF active
10001 OpticalInitError The optical initialization failed.
Could not find value in section %S in
file '%S'",
m_stationPosition
m_delayODMeasStart
m_timeoutRefMeasurement
m_timeoutODMeasurement
m_minLEDValue
m_maxLEDValue
m_minRefValue
m_maxRefValue
Failed during Optical Analyte Init
10002 OpticalQCFileError Failed during reading QC ini file
10003 OpticalLEDValueError Not available
10004 OpticalRefValueError Not available
11000 RotateCRVAPIError These errors describe the full range of
the motion procedure with including
motor movement:
StartFlushWash
Failed When rotating to the flush
position
StartWashWash1
Failed when rotating to the flush
position
StartCleanWash2
Failed when rotating to the flush 2
position
StartRinseWash1Drier
Failed When rotating to the flush
position
trStartRinseWash2
Failed when rotating to the flush 2
position
WashOfWash2Needles
Failed when rotating to the flush 2
position
inLifeTimeOpen
Failed Rotate plate sm-motor moveabs
29(101)
CRVApiError=%d
trCloseLid
Failed Sync rotate position
trOpenLid
Failed Rotate plate sm-motor moveabs
CRVApiError=%d
inLidRotate() {
// Move the rotating plate one strip
step
RotateOne Cake itsRotatorMotor-
>move(%d)
Error=%d",m_stepsBetweenStrips,
retval);
inMoveMicroSteps
// Move the rotating plate one strip
step
RotateOne Cake itsRotatorMotor-
>move(%d)
Error=%d",offset,retval);
inMoveNofStripSteps
"MoveOneStripStep
itsRotatorMotor->moveRel(%d)
Error=%d",m_stepsBetweenStrips,
retval);
inServiceReadBarcode
// Start Barcode reader
" ReadBarcode itsRotatorMotor-
>move(%d) Error=%d",2 *
// Move the rotating plate one strip
step.Reduce the step between the
strips with amount that are moved
with bar code scanner active
ReadBarcode itsRotatorMotor-
>move(%d) Error=%d"
inSyncPosition
Failed to sync encoder and Position
counter on the rotate motor");
inlLifeTimeMove
MoveOneStripStep itsRotatorMotor-
>moveRel(%d)
trMoveOneStripStep
MoveOneStripStep itsRotatorMotor-
>move(%d) Error=%d
trRotateOneCake
Failed to sync encoder and Position
counter on the rotate motor");
// Move the rotating plate one strip
step
RotateOne Cake itsRotatorMotor-
>move(%d)
Error=%d",m_stepsBetweenStrips,
retval);
trShake
Shake itsRotatorMotor->move(%d)
30(101)
Error=%d",m_shakeSteps,retval);
Shake itsRotatorMotor->STOP(%d)
Error=%d",m_stepsBetweenStrips,ret
val
Shake itsRotatorMotor->moveAbs()
Error=%d",retval);
ShakeAndMoveOneStripStep
Shake itsRotatorMotor->move(%d)
Error=%d",m_shakeSteps,retval);
Shake itsRotatorMotor->STOP(%d)
Error=%d",m_stepsBetweenStrips,ret
val);
MoveAbs itsRotatorMotor->moveAbs()
Error=%d"
11001 RotateInitError The movement initialisation failed:
m_flushStationPos
m_stripOnePos
m_stepsBetweenStrips
m_uniqueBarcode
m_servicePath
m_barcodeOffset
m_barcodeHalfAngle
m_shakeSteps
m_shakeSpeed
m_shakeRamp
Failed During Rotating Init
11002 RotateBarcodeWarning (internal This should happen if one or many
barcode reader) strips has an expired date
11003 Incompatible Performance numbers Failed to open or find " + barcode-
>inGetTestDefinition()+".ini" ;
11004 The Performance Description has inLifeTimeTestInit
wrong format Failed initialise procesdescription class
// m_firstStripPos is calculated in
prCalculateNoSteps()
11005 RotateBarcodeError trScanStrips
Barcode Error
11006 A strip has changed place between The strip has changed place according
different scans to first scan
11007 No strips found when scaning GotoPosition
Rotation plate: Failed to go
positio=%d
The else case is taken care of inside
prSetAnalyseProcessAttr method
No Strips found");
11008 RotatePerfFileError Failed reading test definition
calcMethod=%d",calcMethod);
"borderlineMin=%d",borderlineMin)
;
unitStr=%S",unitStr);
calibratorD=%d",calibratorD);
calibratorD=%d",calibratorA);
calibratorD=%d",calibratorB);
calibratorD=%d",calibratorC);
calibratorD=%d",calibratorD);
31(101)
12000 Log Not available
12001 Log: Corrupt or does not exist Not available
12002 Log: Write failed Not available
12003 Log: Bad file descriptor Not available
12004 Log: End of log entries Not available
12005 Log: Length error Not available
13000 Self monitoring errors Not available
13001 Sem: Intialisation error SemInitError
Failed during Temperature
Monitoring Init
Illegal value for cfg parameter
'ADCScaleP' in ini-file
Illegal value for cfg parameter
'ADCScaleI' in ini-file
Illegal value for cfg parameter
'RegulatorPeriod' in ini-file
Illegal value for cfg parameter
'TemperatureTolerance' in ini-file
Illegal value for cfg parameter
'TargetTemperature' in ini-file
Illegal value for cfg parameter
'LogInterval' in ini-file
Illegal value for cfg parameter
'TempArrayLength' in ini-file
13002 Sem Crv API Error SemCrvApiError
Error reading
AI_SENSOR_CLIMA_TEMP_LEVEL
Error writing
AO_PELTIER_CLIMA_TEMP_CONTR
OL
14000 Bar Code Reader: Initialization Error The internal and external
Could not find value in section %S
in file %S",
Illegal beeper intensity for the bar
code reader
Illegal start marker for the bar code
reader
Illegal stop marker for the bar code
reader
Illegal code marker for the bar code
reader
14001 Bar Code Reader: COM Port Error "Could not initialize COM port");
14002 Strip expired (external barcode Check strip ID first, since it has the
reader) hardest requirements on syntax.
Check if strip is valid and can be
used
// Legal strip ID
15000 Printer: Initialization Error Could not find value in section %S
in file %S",
Illegal character spacing for the
printer");
Illegal font size for the printer");
Illegal line spacing for the printer");
Illegal print density for the
32(101)
printer");
The format for the first line will be
long for the printer");
The format for the second line will
be long for the printer");
Could not initialize printer");
15001 Printer: COM Port Error Could not initialize external printer
port");
Could not initialize external printer
port");
Could not initialize external printer
port");
15002 Printer: Printing Error Failed to write to COM port");
Failed to write all data to COM
port");
Failed to open LPT1: with error
%d", err);
Failed to write ESC E to COM port");
Could not initialize external printer
port");
Failed to write to COM port");
Failed to write all data to COM
port");
Failed to write to COM port,
errorcode: %d", err);
Failed to write all data to COM port,
wrote %d of %d",
Failed to write all data to COM port,
wrote %d of %d",
Could not initialize external printer
port");
Failed to write to COM port");
Failed to write all data to COM
port");
15003 Printer: Out of Paper Error INTERNAL ONLY Printer out of paper
15004 Printer: Door open Error INTERNAL ONLY Printer door is open
16000 Start up, self test and monitoring Not available
16001 Startup: Error during application start Not available
up
17000 Lid sensor driver trCloseLid
trOpenLid
Faild open lid sm-motor moveabs
CRVApiError=%d",retval
LID: Failed reading
DI_SENSOR_DOOR_OPEN
LID: Failed turn climate fan on
LID: Failed turn climate fan off
Faild open lid sm-motor moveabs
CRVApiError=%d"
LID: Failed reading
DI_SENSOR_DOOR_OPEN
17001 Failed close or open the climate door Not available
17002 LidInitError Failed Lid Init
17003 LidOpenError Failed to open lid. The
33(101)
DI_SENSOR_DOOR_OPEN is not
Active
LID: Failed to open lid. The
DI_SENSOR_DOOR_OPEN is not
Active
17004 LidCloseError Failed close lid sm-motor moveabs
CRVApiError=%d
17005 Read from SSW Falied to read DigitalInput
(DI_SENSOR_DOOR_CLOSED) on
request from service PC
18000 Liquid levels To low level in WashBuffer 1 bottle
To little WashBuffer 1");
34(101)
25002 Failed to write permanent to flash Error setting AO_OD_CALIB_PRM
memory value
35(101)
Notes:
36(101)
SERVICE CENTERS
Germany
ORGENTEC Diagnostika GmbH
Carl-Zeiss-Straße 49
D-55129 Mainz
Germany
http://www.orgentec.com
Manufactured by:
Orgentec Diagnostica GmbH, Carl-Zeiss-Straße 49, D-55129 Mainz , Germany
Tel. +49 - (0)6131 / 92 58 - 0 Fax +49 - (0)6131 / 92 58 58
http://www.orgentec.com/
Printed in Germany, November 2011
37(101)
INSTRUCTION: HOW TO REMOVE THE
FRONT AIR FILTER
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step-by-step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
always be performed with protective gloves.
References:
Comments:
38(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order:
# Instruction/manoeuvre
1 If possible shut down the instrument, see instructions for use.
2 Remove the power cord.
3 Disconnect the fluid containers (cables and tubing) and remove the containers from
the device.
39(101)
Overview drawing (1)
The drawing below shows the general components of the external casing.
The following pages describe the parts that are to be removed or replaced.
# Part description
1 Top Cover
2 Rear panel (stainless steel)
3 Left side cover
4 Door Cover
5 Right side cover
40(101)
Removal sequence
The covers and parts shall be removed in the following order (re-assembled in the
reverse order).
Door cover
Inner dividing wall
Air filter
Caution:
When starting to remove covers locate a free surface to place the covers without
damaging them, preferably a desk with some protective cloth.
Note:
Reverse the order to assemble.
41(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
42(101)
Remove dividing wall/ cover
The table below describes how to remove and replace the part. See also the views on
the following pages for reference and identification.
Note:
Reverse the order to assemble.
43(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
44(101)
Drawing (3)
The Pictures below describe the parts that are to be removed or replaced.
45(101)
INSTRUCTION: HOW TO EXCHANGE
DISPENSE NEEDLES
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
Not holding the parts correctly may result in bending, kinking of the fluid lines/ tubes
and as a consequence leakage can be the result.
46(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove the instrument top cover, see How to remove covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
47(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
48(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
49(101)
INSTRUCTION: HOW TO EXCHANGE WASH
NEEDLES
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
50(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
Remove the instrument top cover See, How to remove covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
51(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
52(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
53(101)
INSTRUCTION: HOW TO EXCHANGE FILTER
IN BOTTLES
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
54(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Open the front cover and disconnect the appropriate bottle to replace the filter in,
see instruction for use (how to disconnect the bottle(s)).
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
55(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
56(101)
INSTRUCTION: HOW TO REPLACE REAR
VENTILATION FILTER
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
57(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
58(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
59(101)
INSTRUCTION: HOW TO REPLACE THE
FUSES
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
60(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 If possible shut down the instrument, see instructions for use
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
61(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
62(101)
INSTRUCTION: HOW TO ADJUST THE
TENSION OF THE DISPENSE UNIT BELT
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step-by-step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
63(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove the power cord
3 Remove the covers, according to How to remove the covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
64(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
65(101)
INSTRUCTION: HOW TO ADJUST THE
TENSION OF THE TURNING TABLE BELT
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
66(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove the power cord
3 Remove the covers, according to How to remove the covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
67(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
68(101)
INSTRUCTION: HOW TO REPLACE THE
PRINTER
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step-by-step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
69(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove instrument covers (top cover)
Remove/Replace
The table below describes how to remove and replace the part. See also the view’s on
the following pages for reference and identification.
70(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
71(101)
INSTRUCTION: HOW TO CUT AND
ASSEMBLE FLUID LINES AND ASSEMBLE
COUPLINGS
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step-by-step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is
handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
References:
Comments:
Never bend, kink or twist fluid lines, always ensure 90 degree cuts and ensure coupling
surface is flush with ferrule and fluid line
72(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove the covers, see How to remove covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
73(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
74(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
75(101)
INSTRUCTION: HOW TO REMOVE THE
COVERS
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step by step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”, It must be completed; this is to ensure the safe and proper
operation).
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed in an ESD safe environment (minimum requirement is
to use wrist strap and ESD conductive surface where the instrument is handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
always be performed with protective gloves.
References:
Comments:
76(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order:
# Instruction/manoeuvre
1 If possible shut down the instrument, see instructions for use.
2 Remove the power cord.
3 Disconnect the fluid containers (cables and tubing) and remove the containers from
the device.
77(101)
Overview drawing (1)
The drawing below shows the general components of the external casing.
The following pages describe the parts that are to be removed or replaced.
# Part description
1 Top Cover
2 Rear panel (stainless steel)
3 Left side cover
4 Door cover
5 Right side cover
78(101)
Removal sequence
The covers shall be removed in the following order (re-assembled in the reverse order).
Door cover
Inner dividing cover
Top cover
Left side cover
Right side cover
Caution!
When starting to remove covers locate a free surface to place the covers without
damaging them, preferably a desk with some protective cloth.
Note:
Reverse the order to assemble.
79(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
80(101)
Remove top cover (and inner dividing cover)
The table below describes how to remove and replace the part. See also the views on
the following pages for reference and identification.
Note:
There are to guiding pins at the front of the case.
Note:
Reverse the order to assemble.
81(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
82(101)
Drawing (3)
The Pictures below describe the parts that are to be removed or replaced.
83(101)
Drawing (4)
The Pictures below describe the parts that are to be removed or replaced.
84(101)
Drawing (5)
The Pictures below describe the parts that are to be removed or replaced.
85(101)
Drawing (6)
The Pictures below describe the parts that are to be removed or replaced.
86(101)
Drawing (7)
The Pictures below describe the parts that are to be removed or replaced.
87(101)
Remove side covers (left and right, right side shown)
The table below describes how to remove and replace the part. See also the views on
the following pages for reference and identification.
Note:
The same procedure is used for the left side cover.
Reverse the order to assemble.
88(101)
Drawing (8)
The Pictures below describe the parts that are to be removed or replaced.
89(101)
Drawing (9)
The Pictures below describe the parts that are to be removed or replaced.
90(101)
INSTRUCTION: HOW TO REPLACE THE OD
MODULE
What to do
This instruction is made up of four parts:
Start by reading the information section and if not performed, perform the pre requisite
instructions and then start the step-by-step procedure, use the drawings as a guide. At
last re-verse the procedure and test the instrument (according to the “Preventive
maintenance instruction”. It must be completed; this is to ensure the safe and proper
operation). Please also have a look into the technical bulletin TB 0907 – How to
exchange the Transmitterboard and check its functionality.
Note:
If none of the replace/ removal procedures solves the problem. Please contact your local
ORGENTEC Diagnostika representative.
Information
Safety precautions:
The device has connections to mains supply e.g. 110-240V~ and special caution shall
be taken when working with the instrument (lethal current is possible from most
electronics).
Unplug the power connector from its power source before cleaning or servicing.
Failure to do so could result in equipment damage or personal injury.
Use only ORGENTEC Diagnostika GmbH spare parts, accessories and consumables.
All service shall be performed by ORGENTEC Diagnostika authorized personnel.
All service shall be performed by in an ESD safe environment (minimum requirement
is to use wrist strap and ESD conductive surface where the instrument is handled).
Handling of samples shall follow laboratory infection risk procedures.
Handling of device parts that has the possible risk of sample contamination shall
Always be performed with protective gloves.
Comments:
Never bend, kink or twist fluid lines.
91(101)
Pre-requisites
Removal/replacement procedures (required to be done prior to this specific procedure)
listed in order
# Instruction/manoeuvre
1 Shut down the instrument, see instructions for use
2 Remove the covers, see How to remove covers
Remove/Replace
The table below describes how to remove and replace the part. See also the View’s on
the following pages for reference and identification.
5. Remove the Climate hose (from climate unit to rear filter) See drawing 3
6. Remove plane (Locate the three screws, remove them and, See drawing 4
disconnect any cabling/couplings to be able to lift of the
plane with the dispense/wash/soak stations)
Caution Note!
When lifting the plane, ensure to lift it straight up to avoid
needles bending.
7. Remove the trim by removing the screws around its See drawing 5
perimeter (8 screws).
8. Remove the rotating door / lid (lift it straight up). See drawing 6
9. Remove screws around the perimeter of the climate See drawing 7
module
(8 screws), Remove the climate module (straight up).
10. Remove the turning plate by removing the screws around See drawing 8
the center and the center of the turning plate.
11. Locate the OD module from beneath. Remove the 4 screws See drawing 9
(corners) to loosen the lid of the OD module.
12. Remove the 2 screws holding the OD module to the plane See drawing 9
(opposite diagonal position).
13. Remove the flat ribbon connector cable from the circuit -
board.
Note:
Reverse the order to assemble.
92(101)
Drawing (1)
The Pictures below describe the parts that are to be removed or replaced.
93(101)
Drawing (2)
The Pictures below describe the parts that are to be removed or replaced.
94(101)
Drawing (3)
The Pictures below describe the parts that are to be removed or replaced.
95(101)
Drawing (4)
The Pictures below describe the parts that are to be removed or replaced.
96(101)
Drawing (5)
The Pictures below describe the parts that are to be removed or replaced.
97(101)
Drawing (6)
The Pictures below describe the parts that are to be removed or replaced.
98(101)
Drawing (7)
The Pictures below describe the parts that are to be removed or replaced.
99(101)
Drawing (8)
The Pictures below describe the parts that are to be removed or replaced.
100(101)
Drawing (9)
The Pictures below describe the parts that are to be removed or replaced.
101(101)
DI_SENSOR_DOOR_OPEN
SM_MOTOR_DISP_Z_MOVE QE_SENSOR_DISP_RAD_MOVE_POS
SM_MOTOR_DISP_CON_Z_MOVE DI_SENSOR_DISP_RAD_MOVE_HOME
AO_FAN_CABINET
DI_SENSOR_WB_WB2
DC_PELTIER_FAN
DI_SENSOR_DOOR_CLOSED
DO_VALVE_WB_MEDIA_SELECTION_1 SM_PUMP_DISP_LARGE_DISP
DO_VALVE_WB_MEDIA_SELECTION_2
DO_VALVE_WB_MEDIA_SELECTION_3
DC_PUMP_WASTE
DI_SENSOR_DRAIN_WATER_LEVEL
DI_SENSOR_DISP_CON_Z_MOVE_HOME
DI_SENSOR_WB_WB1
DO_VALVE_WB_MEDIA_SELECTION_4 DI_SENSOR_HATCH_OPEN
DI_SENSOR_DISP_Z_MOVE_HOME DI_SENSOR_ROT_MOVE_HOME
DI_SENSOR_DRIER_Z_MOVE_HOME
DI_SENSOR_WS2_Z_MOVE_HOME
QE_ENCODER_ROT_MOVE_POS
DI_SENSOR_WB_DEST_WATER
DI_SENSOR_WS1_Z_MOVE_HOME
VALV_CLEAN_DRY_SUCTION_NEEDLE
DC_PUMP_WS2_WASH_BUFFER_FEED
VALVE_DRIER_SUCTION
DC_PUMP_DRIER_SUCTION VALVE_WS1_WASH
VALVE_WS2_WASH
Author: MS1
Date: 12.09.2011
Filename: 20100-3ME003-02 Connections.doc
OBS! Separate GND for 12V and 24V To Internal Printer To IO Board To Analysis Module
POWER TO IO BOARD
24V,12V,5V
To Bar Code To L CD
Mains in
Battery
AC/DC
24V
To
TouchScreen
AC/DC CPU Module
12V,5V,-12V SD Card To I2C
Thermometer
Modem
To Loudspeaker
Customer: Document:
Confidential DP1 Connections
Project: Doc. no. Date: Page:
20100 Carnival Confidential 2006-11-07 2(3)
Fluid handling
Dry Suction
2
Valve_Drier_Suction
3 1
Pump_Drier_Suction
Dispensing
1
Valve_WB_Media_Selection _4
Valve_WS1_Wash Pump_WS2_Wash_Buffer _Feed
3
Pump_Disp_Pip_Wash_Feed
1 2
2
1
2
3
3
1
2
Pump_WS1_Wash_Buffer _Feed
Valve_Clean_Dry_Suction_Needle
Pump_WS1_Suction Pump_WS2_Suction
3
Pipette Wash Cup
Valve_WB_Media_Selection _2
Needle Wash Cup
2 1
Needle Wash Cup
3 3
Valve_WB_Media_Selection _3 Valve_WB_Media_Selection _1
2 1 2 1
Pump_Waste
Wash Waste
Wash
Distilled Buffer
Buffer
Solution 2
Water Solution 1
Refill
Refill
5 4 3 2 1
D D
Revision history:
Rev 01: 2004-08-26 BPn First Release - Cable lengths are approximate.
Rev 02: 2004-11-30 BPn Error fixes - Cable lengths are still approximate.
Added fixes for FLEX-card.
C C
B B
A A
Title
Top
D D
To J3 on
Motherboard
J1 J2
1
3
2
4
Flat Cable 1
3
2
4
5 6 5 6
7
9
8
10
L = 1000 mm 7
9
8
10
11 12 11 12
13 14 13 14
To J2 on
Analysis
Processor
board
To J1 on
Analysis
Receiver
board
B J3 J4 B
1
3
2
4
Flat Cable 1
3
2
4
5 6 5 6
L = 120 mm
Mico MaTch 6 pole male Mico MaTch 6 pole male
To J2 on
Analysis
Processor
board
A A
Title
Analysis
D D
L = 100 mm
J5
Red 1 To J9 on Motherboard
C C
J6
2
2
B B
A A
Title
External scanner power
D D
J7
L = 600 mm
1
1 Push-On 2.8 mm
J8
+ MG1
Blue 1
DC_PUMP_WS1_SUCTION DC-MOTOR White 2 J27 on IO-board
A 3
-
Area = 0.22 mm2
J9 Molex KK 3 pole
2
Push-On 2.8 mm
C J10 C
L = 600 mm
1
Push-On 2.8 mm
1
J11
+
MG2 Blue 1
DC_PUMP_WS2_SUCTION White 2 J28 on IO-board
A DC-MOTOR 3
-
Area = 0.22 mm2
J12 Molex KK 3 pole
2
Push-On 2.8 mm
J13
L = 1200 mm
1
Push-On 2.8 mm
1
J14
+
DC_PUMP_WASTE MG3 Blue 1
White 2 J32 on IO-board
B A DC-MOTOR 3
B
-
Area = 0.22 mm2
J15 Molex KK 3 pole
2
Push-On 2.8 mm
A A
Title
DC-motors 1
D D
J16
L = 850 mm
1
Push-On 2.8 mm
1
J17
+
MG4 Blue 1
DC_PUMP_DISP_PIP_WASH_FEED White 2 J33 on IO-board
A 3
DC-MOTOR
-
Area = 0.22 mm2
J18 Molex KK 3 pole
2
1
Push-On 2.8 mm
J19
L = 600 mm
1
Push-On 2.8 mm
1
J20
+
MG5 Blue 1
C DC_PUMP_WS1_WASH_BUFFER_FEED White 2 J34 on IO-board C
A 3
DC-MOTOR
-
Area = 0.22 mm2
J21 Molex KK 3 pole
2
Push-On 2.8 mm
J22
L = 600 mm
1
Push-On 2.8 mm
1
J23
+
MG6 Blue 1
DC_PUMP_WS2_WASH_BUFFER_FEED White 2 J35 on IO-board
A 3
DC-MOTOR
-
Area = 0.22 mm2
J24 Molex KK 3 pole
2
Push-On 2.8 mm
B B
J25
L = 600 mm
1
Push-On 2.8 mm
1
J26
+
MG7 Blue 1
DC_PUMP_DRIER_SUCTION White 2 J36 on IO-board
A DC-MOTOR 3
-
Area = 0.22 mm2
J27 Molex KK 3 pole
2
Push-On 2.8 mm
A A
Title
DC-motors 2
L = 300 mm
J28 CON1
D D
1 R_7 1
2 VCC1 R_6 2
3 VCC2 R_5 3
4 R_4 4
5 PIXCLK R_3 5
6 DISPEN R_2 6
7 VSYNC GND1 7
8 HSYNC 8
9 GND1 G_7 9
10 R_2 G_6 10
11 R_3 G_5 11
12 R_4 G_4 12
13 R_5 G_3 13
R_6 G_2
To LCD 14
15 R_7 GND2
14
15
To CON8 on
16 GND2 16 Motherboard
17 G_2 B_7 17
18 G_3 B_6 18
19 G_4 B_5 19
20 G_5 B_4 20
21 G_6 B_3 21
22 G_7 B_2 22
23 GND3 GND3 23
24 B_2 24
25 B_3 VSYNC 25
26 B_4 HSYNC 26
27 B_5 PIXCLK 27
28 B_6 28
C 29 B_7 DISPEN 29 C
30 30
VCC1 31
VCC2 32
JAE-FI-X30M
L = 300 mm
CON2 CON3
Area = 0.14 mm2
1 1
2 2
3 3
To CN1 on 4
5
4
5
To CON9 on
Backlight 6 6 Motherboard
7 7
Power Supply 8 8
9 9
B B
10 10
L = 330 mm
J29 J30
Flat cable
1 1
To Touch Screen 2
3
2
3
To CON7 on
4 4 Motherboard
CON4 Male REC4 Female
A A
Title
Display
D D
L = 330 mm
C J31 J32 C
1
Area = 0.14 mm2 White 1
2 2
To CN6 3
4
3
4
To J7 on
on Printer 5 5 Motherboard
6 6
GND Black
AMP Quick 6 pole Molex KK 6 pole
L = 480 mm
B B
J33
To Power 4 +5V Red
terminal
1
2
To J8 on
on Printer 3 0V Black Motherboard
Molex KK 2 pole
Area = 0.22 mm2
A A
Title
Internal_printer
D D
C C
J34 J35
1 6 1 6
2
3
7
8
Flat Cable 2
3
7
8
4 9 4 9
5 10 L = 480 mm 5 10
B To J2 on To J5 on B
I/O-board Motherboard
A A
Title
Local Bus
D D
C
L = 200 mm C
J36 J37
White
1 1 To CON13 on
LS1 + 2
Motherboard
Push-On 4.8 mm
Black
Molex KK 2 pole
J38
SPEAKER 1
Area = 0.22 mm2
Push-On 2.8 mm
B B
A A
Title
Loudspeaker
D
L = 1600 mm D
J39
Red 1
2 J64 on IO-board
AO_PELTIER_CLIMA_TEMP_CONTROL Black 3
Molex KK 3 pole
Area = 0.75 mm2
L = 1600 mm
J40
Red 1
2
DC_PELTIER_FAN Black J3 on IO-board
C C
Molex KK 2 pole
Area = 0.22 mm2
L = 1600 mm
J41 J42
1 Red 1
2 2
AO_FAN_CLIMA_COOLING Black 3 J21 on IO-board
Terminal block
Molex KK 3 pole
Area = 0.22 mm2
B B
L = 600 mm
J43 J44
1 Red 1
AO_FAN_CABINET 2 2 J22 on IO-board
Black 3
Terminal block
Molex KK 3 pole
Area = 0.22 mm2
A A
Title
Peltier element and fans
L = 440 mm
CON4
Area = 0.75 mm2
CON5
D D
1
2 -12V 1
3 +5V 2
4 +5V 3
5 +5V 4
To CON1 on 6
GND
5
PSU1
7
8 GND
6
7
To CON2 on
9 GND
+12V
8 Motherboard
10 9
11 GND 10
12 GND
13
AMP_MTA156_3.96, 10P
GND = Black AMP_MTA156_3.96, 13P
+5V = Red
+12V = White L = 300 mm
-12V = Blue J45
Area = 0.75 mm2
+24V = Green
1 1
L = 350 mm Area = 0.75 mm2 J47
Fork 3.2 mm
J46 +24V 1
C +24V 2 C
GND
2 1
GND
3
4
To J1 on
To TB2 on +12V
GND
5 I/O-board
Fork 3.2 mm 6
PSU2 +5V
GND
7
J48 8
4 1
J49
5 1
Fork 3.2 mm
L = 130 mm
Area = 0.75 mm2
J51
J50
To L on 1
Blue 1
To L1 on
B B
PSU2 mains conn
Fork 3.2 mm Push-On 6.3 mm
J53
J52
Brown To L2 on
To N on 1 1
mains conn
PSU2
Fork 3.2 mm Push-On 6.3 mm
J55
J54
To E on 1
Green/Yellow 1
To PG on
PSU2 mains conn
Fork 3.2 mm Push-On 6.3 mm
L = 130 mm
J56
5
Blue
To P1 on 4
3
Brown
A PSU1 2 Green/Yellow A
1
Molex KK 3.96 mm
J57
Title
Green/Yellow Power
To chassis 1
Size Document Number Rev
EYELET 4,2 MM HOLE A3 20100-3SC009-02-Internal cables 02
L = 600 mm
J59
J58
1
1 White 2
DI_SENSOR_WB_WB1 Blue 3 J46 on IO-board
4
2
BNC
D Area = 0.22 mm2 Molex KK 4 pole
D
L = 1200 mm
J60
U1
1 3 Red 1
White 2
DI_SENSOR_HATCH_OPEN Black 3 J47 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 1200 mm
J61
U2
1 3 Red 1
White 2
DI_SENSOR_DRIER_Z_MOVE_HOME Black 3 J41 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
C C
L = 600 mm
J62
U3
1 3 Red 1
White 2
DI_SENSOR_ROT_MOVE_HOME Black 3 J43 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 1200 mm
J63
U4
1 3 Red 1
White 2
DI_SENSOR_WS1_Z_MOVE_HOME Black 3 J61 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 1200 mm
B B
J64
U5
1 3 Red 1
White 2
DI_SENSOR_WS2_Z_MOVE_HOME Black 3 J63 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 150 mm L = 850 mm
J65 J66 J67
U6
1 3 Red 4 4 Red 1
White 3 3 White 2
DI_SENSOR_DISP_Z_MOVE_HOME Black 2 2 Black 3 J56 on IO-board
2 4 Green 1 1 Green 4
OPTO-FORK
Molex KK 4 pole Molex KK 4 pole
Area = 0.14 mm2 Molex KK 4 pole
J3 on FLEX-board
A A
Title
Sensors 1
L = 1000 mm
J69
J68
1
DI_SENSOR_WB_DEST_WATER 1 White 2 J57 on IO-board
D Blue 3 D
4
2
BNC
Area = 0.22 mm2 Molex KK 4 pole
L = 1000 mm J71
J70
1
1 White 2
Blue 3
DI_SENSOR_DRAIN_WATER_LEVEL 4 J54 on IO-board
2
BNC
Area = 0.22 mm2 Molex KK 4 pole
L = 1200 mm
J72
U7
1 3 Red 1
DI_SENSOR_DOOR_OPEN White 2
Black 3 J49 on IO-board
2 4 Green 4
C OPTO-FORK C
Area = 0.14 mm2 Molex KK 4 pole
L = 1200 mm
J73
U8
1 3 Red 1
White 2
DI_SENSOR_DOOR_CLOSED Black 3 J53 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 1200 mm
J75
J74
1
1 White 2
Blue 3
DI_SENSOR_WB_WB2 4 J52 on IO-board
2
BNC
Area = 0.22 mm2 Molex KK 4 pole
B B
L = 600 mm
J76
U9
1 3 Red 1
White 2
DI_SENSOR_DISP_RAD_MOVE_HOME Black 3 J51 on IO-board
2 4 Green 4
OPTO-FORK
Area = 0.14 mm2 Molex KK 4 pole
L = 150 mm L = 850 mm
J77 J78 J79
U10
1 3 Red 4 4 Red 1
White 3 3 White 2
DI_SENSOR_DISP_CON_Z_MOVE_HOME Black 2 2 Black 3 J58 on IO-board
2 4 Green 1 1 Green 4
OPTO-FORK
Molex KK 4 pole Molex KK 4 pole
Area = 0.14 mm2 Molex KK 4 pole
A J4 on FLEX-board A
Title
Sensors 2
L = 1400 mm
MG8 J80
1 Red 4
SM_MOTOR_DRIER_Z_MOVE 2 Green 3
3 White 2 J10 on IO-board
4 Yellow 1
L = 600 mm
MG9 J81
1 Red 4
SM_MOTOR_DOOR_MOVE 2 Green 3 J12 on IO-board
3 White 2
4 Yellow 1
L = 1400 mm
MG10 J82
1 Red 4
SM_MOTOR_WS1_Z_MOVE 2 Green 3
3 White 2 J15 on IO-board
4 Yellow 1
C C
MOTOR STEPPER 2-PHASE
Area = 0.14 mm2 Molex KK 4 pole
L = 1400 mm
MG11 J83
1 Red 4
SM_MOTOR_WS2_Z_MOVE 2 Green 3 J16 on IO-board
3 White 2
4 Yellow 1
L = 150 mm L = 850 mm
MG12 J84 J85 J86
1 Red 4 4 Red 4
SM_MOTOR_DISP_Z_MOVE 2 Green 3 3 Green 3
3 White 2 2 White 2 J13 on IO-board
4 Yellow 1 1 Yellow 1
B B
L = 150 mm L = 850 mm
MG13 J87 J88 J89
1 Red 4 4 Red 4
SM_MOTOR_DISP_CON_Z_MOVE 2 Green 3 3 Green 3
3 White 2 2 White 2 J14 on IO-board
4 Yellow 1 1 Yellow 1
1 Black 4
SM_MOTOR_DISP_LARGE_DISP 2 Red 3
3 Green 2 J11 on IO-board
A 4 Blue 1 A
Title
2 phase stepper motors
L = 600 mm
D D
MG15 J91
1 Yellow 1
2 Blue 2 J37 on IO-board
3 Red 3
SM_MOTOR_ROT_MOVE
QE_ENCODER_ROT_MOVE_POS
J92 J93
5 Red 1
4 White 2
3 Green 3 J40 on IO-board
2 Black 4
1
Area = 0.14 mm2 Molex KK 2.54
AMP 103686-4/640442-5
C C
MG16 J94
R1 1R/5W R2 1R/5W
1 Yellow 1
2 Blue 2 J38 on IO-board
3 Red 3
R3 1R/5W
MOTOR STEPPER 3-PHASE
Area = 0.22 mm2 Molex KK 3.96
SM_MOTOR_DISP_RAD_MOVE
QE_ENCODER_DISP_RAD_MOVE_POS
J95 J96
5 Red 1
B B
4 White 2
3 Green 3 J50 on IO-board
2 Black 4
1
Area = 0.14 mm2 Molex KK 2.54
AMP 103686-4/640442-5
A A
Title
3 phase stepper motors
J97
L = 1200 mm
1
Push-On 2.8 mm
1
J98
+
MG17 Blue 1
VALVE_DRIER_SUCTION White 2 J19 on IO-board
A 3
DC-VALVE
D
-
Area = 0.22 mm2 D
2
1
Push-On 2.8 mm
J100
L = 1200 mm
1
Push-On 2.8 mm
1
J101
+
MG18 Blue 1
VALVE_CLEAN_DRY_SUCTION_NEEDLE White 2 J20 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J102 Molex KK 2.54
2
Push-On 2.8 mm
J103
C
L = 1200 mm C
1
Push-On 2.8 mm
1
J104
+
MG19 Blue 1
VALVE_WS1_WASH White 2 J17 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J105 Molex KK 2.54
2
Push-On 2.8 mm
J106
L = 1200 mm
1
Push-On 2.8 mm
1
J107
+
MG20 Blue 1
VALVE_WS2_WASH White 2 J18 on IO-board
A 3
DC-VALVE
B
-
Area = 0.22 mm2 B
J108 Molex KK 2.54
2
Push-On 2.8 mm
J109
L = 600 mm
1
Push-On 2.8 mm
1
J110
+
MG21 Blue 1
VALVE_DISP_PIP_WATER_FEED White 2 J25 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J111 Molex KK 2.54
2
Push-On 2.8 mm
A A
Title
Valves 1
J112
L = 500 mm
D 1 D
Push-On 2.8 mm
1
J113
+
MG22 Blue 1
VALVE_DISP_PIP_CHOICE White 2 J26 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J114 Molex KK 2.54
2
1
Push-On 2.8 mm
J115
L = 600 mm
1
Push-On 2.8 mm
1
J116
+
MG23 Blue 1
VALVE_WB_MEDIA_SELECTION_1 White 2 J29 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
C J117 Molex KK 2.54 C
2
Push-On 2.8 mm
J118
L = 600 mm
1
Push-On 2.8 mm
1
J119
+
MG24 Blue 1
VALVE_WB_MEDIA_SELECTION_2 White 2 J30 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J120 Molex KK 2.54
2
Push-On 2.8 mm
J121
L = 300 mm
1
B B
Push-On 2.8 mm
1
J122
+
MG25 Blue 1
VALVE_WB_MEDIA_SELECTION_3 White 2 J31 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
J123 Molex KK 2.54
2
Push-On 2.8 mm
J124
L = 600 mm
1
Push-On 2.8 mm
1
J125
+
MG26 Blue 1
VALVE_WB_MEDIA_SELECTION_4 White 2 J68 on IO-board
A 3
DC-VALVE
-
Area = 0.22 mm2
A J126 Molex KK 2.54 A
2
Push-On 2.8 mm
Title
Valves 2