Professional Documents
Culture Documents
Atomic Molecular
Electrochemical
spectrometry spectrometry
• AAS • UV-vis • ISE
• ICP-AES • IR
• MS
Separation Others
• Chromatography • Statistics
• Lab management
1
Question
2
Department of Chemical and
Environmental Engineering
CHEE 2023 Analytical Measurement
Lecture 2:
UV-Visible Spectroscopy
3
Learning Outcomes (LOs)
1. Understand the principles and applications of atomic and
molecular spectroscopy, then compare and contrast them.
2. Define and identify the types of errors that can arise from
measurements, apply statistical equations to analyse these
errors, then suggest ways to overcome them.
4
Overview
5
Review: The Electromagnetic spectrum
Introduction
• Visible light absorption is known to all of us, because this is what causes objects to be coloured.
• E.g. a blue dye appears blue because the light at the red end of the spectrum is absorbed, leaving the
blue light to be transmitted (refer to diagram).
7
Regions of the electromagnetic spectrum
When light is absorbed by a material, valence electrons are promoted from their normal
(ground) states to higher energy (excited) states. 8
UV-Visible Spectrometer
• Because only small numbers of absorbing molecules are required, it is convenient to
have the sample in solution
ideally the solvent should not absorb in the ultraviolet/visible range however, this is rarely
the case.
• Radiation across the whole of the ultraviolet/visible range is scanned over a period of
approximately 30 s
and radiation of the same frequency and intensity is simultaneously passed through a
reference cell containing only the solvent.
• Photocells then detect the radiation transmitted and the spectrometer records the
absorption by comparing the difference between the intensity of the radiation passing
through the sample and the reference cells (Fig. next page).
9
UV-Visible Spectrometer (2)
Double-beam:
• Light source
a combination of tungsten/halogen and deuterium lamps.
The output from the light source is focused onto the diffraction grating which splits the incoming light into its
component colours of different wavelengths, like a prism (shown) but more efficiently.
10
UV-Visible Spectrometer (3)
• No single lamp provides radiation across the whole of the range
required, so two are used.
A hydrogen or deuterium discharge lamp covers the ultraviolet range,
and a tungsten filament (usually a tungsten/halogen lamp) covers the
visible range.
• Wavelength checks are made by passing the sample beam through glass
samples (containing holmium oxide) that have precise absorption peaks,
The absorption is calibrated by passing the sample beam through either a
series of filters, each with a specific and known absorption, or a series of
standard solutions.
Holmium oxide: have a series of sharp optical absorption peaks in the visible
range. Used as a convenient calibration standard for optical spectrometers.
12
UV-Visible Spectrometer (5)
• For liquids, the sample is held in an optically flat, transparent container called a cell or cuvette.
• The reference cell/cuvette contains the solvent in which the sample is dissolved, commonly
referred to as the blank.
• For each wavelength the intensity of light passing through both a reference cell (I0) and the
sample cell (I) is measured.
If I is less than Io, then the sample has absorbed some of the light.
The absorbance (A) of the sample is related to I and Io according to the following equation:
I0
A = log10
I
• The detector converts the incoming light into a current.
the higher the current, the greater the intensity.
• The chart recorder usually plots the absorbance against wavelength (nm) in the UV and visible
section of the electromagnetic spectrum. (Note: absorbance does not have any units).
13
The Beer-Lambert Law
• Absorbance is proportional to concentration of the substance in solution.
ONLY true for DILUTE solutions!
I0
A = log10 = ε C l
I
• Where:
A : absorbance
l : optical path length, i.e. dimension of the cell/cuvette (cm)
C : concentration of solution (mol L-1)
ε : a constant for each absorbing material, known as the molar absorption
coefficient , (unit: L mol-1 cm-1, but by convention the units are not quoted).
14
Calibration graph
15
Example: UV-Vis Spectrum: buta-1,3-diene
• One can readily see what
wavelengths of light are absorbed
(peaks), and what wavelengths of
light are transmitted (troughs).
The higher the value, the more of a
particular wavelength is being
absorbed.
16
UV-Vis Spectrum: blue copper complex
17
Quantification by UV-Vis
18
Quantification by UV-Vis (2)
• Conditions for Beer-Lambert law to be obeyed:
Light must be monochromatic - measure at λmax
Conc. of solution must be weak
Solution must not fluoresce, must be homogenous
Solute must not be photoreactive
Solute must not react with solvent
19
Applications
Used extensively for the quantitative analysis of all molecules that absorb UV and visible
electromagnetic radiation.
Clinical chemistry: study of enzyme kinetics
• Enzymes cannot be studied directly, but their activity can be studied by analysing the speed of the
reactions which they catalyse.
• Reagents/labels can be attached to molecules to permit indirect detection and measurement of
enzyme activity.
• Indicator of tissue damage:
• When cells are damaged by disease, enzymes leak into the bloodstream – the amount present
indicates the severity of the tissue damage.
• Relative proportions of different enzymes can be used to diagnose disease, e.g. of the liver, pancreas
or other organs which otherwise exhibit similar symptoms.
21
Question 1:
1. Show that the Beer-Lambert’s law is additive.
22
Questions 2:
A 1.28x10-4 M solution of potassium permanganate has a
transmittance of 0.5 when measured in a 1 cm cell at 525 nm.
(a)Calculate the molar absorptivity coefficient for the permanganate at
this wavelength.
23