Data Sheet
Introduction
The B1505A Power Device Analyzer/ Two independent analog-to-digital become productive quickly. Module
Curve Tracer is a single-box solution (A/D) converters on each channel selector and Quick Test feature
with next-generation curve tracer support a 2 µs sampling rate for accu- enable fully automated measurement
functionality that can accurately rate monitoring of the critical timings on multiple parameters without the
evaluate and characterize power that can affect device behavior. need to recable. The net result is
devices at up to 10 kV and 1500 improved ease of use, better data
amps. The B1505A is capable of It can also perform capacitance analysis and simplified data manage-
handling all types of power device measurements at high voltage biases ment for the measurement of power
evaluation, with features that include (up to 3000 V). The B1505A with devices and power circuitry.
a wide voltage and current range, EasyEXPERT software includes a
fast pulsing capability (10 µs), µΩ curve tracer mode that combines
level on-resistance measurement familiar curve tracer functionality
resolution and sub-pA level current with the convenience of a PC-based
measurement capability. In addition, instrument; this makes it easy for
an oscilloscope view permits visual traditional curve-tracer users to
verification of both current and volt-
age pulsed waveforms.
Basic features Speciication conditions Note: This document lists specifications
and supplemental characteristics for
• Performs wide range of IV The measurement and output the B1505A and its associated modules.
measurements accuracy are specified under the con- The specifications are the standards
• Up to 10 kV / 1500 A ditions listed below. Note: The SMU against which the B1505A and its
measurement and output accuracies associated modules are tested. When
• Large peak power : 22.5 kW
are specified at the SMU connector the B1505A or any of its associated
• Medium current measurement terminals, using the Zero Check modules are shipped from the factory,
with high voltage bias (e.g. terminal as a reference. they meet the specifications. The
500 mA at 1200 V,
“supplemental” characteristics
Peak power : 900 W) 1. Temperature: 23 °C ± 5 °C
described in the following specifica-
• µΩ resistance measurement 2. Humidity: 20% to 70% tions are not guaranteed, but provide
• sub-pA leakage measurement useful information about the functions
3. Self-calibration after a 40 minute
and performance of the instrument.
• Performs high bias voltage CV warm-up is required.
measurements Note: Module upgrades to existing
4. Ambient temperature change less
• Pulsed measurement (≥10 µs) than ±1 °C after self-calibration B1505A systems must be carried out
• Two independent A/D converters execution. (Note: This does not at an Agilent Technologies service
(22 bit equivalent) on each channel apply to the MFCMU). centre. In order to ensure system
enable the simultaneous high- specifications the new modules need
5. Measurement made within one to be installed and the complete
speed (2 µs) sampling of current
hour after self-calibration execution. unit calibrated. Contact your nearest
and voltage
(Note: This does not apply to the Agilent Technologies office to arrange
• Temperature measurement MFCMU). the installation and calibration of new
• Easy to use EasyEXPERT test B1505A modules.
6. Calibration period: 1 year
environment
• Curve tracer test mode with knob 7. SMU integration time setting:
sweep capability 1 PLC (1 nA to 1 A range, voltage
range), 200 µs (20 A range)
• Oscilloscope view
Averaging of high-speed ADC:
• Modular configuration with ten 128 samples per 1 PLC
module slots for supported modules
8. SMU filter: ON (for HPSMU and
• Multiple SMU types available:
MPSMU)
HPSMU, MPSMU, HCSMU,
MCSMU and HVSMU 9. SMU measurement terminal
• Support for high power devices connection: Kelvin connection
with up to 6 pins (for HPSMU, MPSMU, HCSMU and
MCSMU), non-Kelvin (for HVSMU)
• Fast high voltage/high current
switch for GaN current collapse
effect characterization
• Multi-frequency capacitance
measurement unit (MFCMU)
(1 kHz to 5 MHz) available
• Standard accessories for package
test and wafer test: test fixture,
module selector and high voltage
bias-tee
• 4.2-amp ground unit included
standard with the mainframe
• GPIB, USB, LAN interfaces and
VGA video output port
• Self-test, self-calibration, diagnostics
2
B1505A Speciications
3
HPSMU Module Speciications
4
HPSMU measurement and output range
Power consumption
Voltage source mode:
Voltage range Power Current (mA)
1000
2V 20 x Ic (W)
20 V 20 x Ic (W)
40 V 40 x Ic (W)
100 V 100 x Ic (W)
200 V 200 x Ic (W) 500
Where Ic is the current compliance setting.
125
Current source mode:
Voltage Power 50
compliance
Vc ≤ 20 20 x Io (W) -200 -100 -40 -20 20 40 100 200
20 < Vc ≤ 40 40 x Io (W)
-50 Voltage (V)
40 < Vc ≤ 100 100 x Io (W)
100 < Vc ≤ 200 200 x Io (W) -125
Where Vc is the voltage compliance
setting and Io is output current. -500
-1000
5
MPSMU Module Speciications
1. ± (% of reading value + ixed offset in A + proportional offset in A), Vo is the output voltage in V.)
2. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in Amps.
1. ±(% of reading value + offset value in mV). Averaging is 128 samples in 1 PLC.
2. 100 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in Volts.
1. ±(%of reading value + ixed offset in A + proportional offset in A), Vo is the output voltage in V.)
2. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in Amps.
6
Power consumption MPSMU measurement and output range
Voltage source mode:
Voltage range Power Current (mA)
0.5 V 20 x Ic (W) 100
2V 20 x Ic (W)
5V 20 x Ic (W)
20 V 20 x Ic (W) 50
40 V 40 x Ic (W)
100 V 100 x Ic (W)
Where Ic is the current compliance setting.
20
7
HCSMU Module Speciications
8
HVSMU Module Speciications
-8
9
MCSMU Module Speciications
10
SMU source measurement mode SMU pulse measurement For HVSMU:
For HPSMU and MPSMU: Pulse width, period, and delay: Pulse width: 500 µs to 2 s
VFIM, IFVM Pulse width resolution: 2 µs
For HCSMU, MCSMU and HVSMU: For HPSMU and MPSMU: Pulse period: 5 ms to 5 s
VFIM, VFVM, IFVM, IFIM Pulse width: 500 µs to 2 s Period ≥ delay + width + 2 ms
Output terminal/connection: Pulse width resolution: 100 µs (when delay + width ≤ 100 ms)
For HPSMU and MPSMU: Pulse period: 5 ms to 5 s Period ≥ delay + width + 10 ms
Dual triaxial connector, Period ≥ delay + width + 2 ms (when delay + width > 100 ms)
Kelvin (remote sensing) (when delay + width ≤ 100 ms) Pulse period resolution: 100 µs
For HCSMU: Period ≥ delay + width + 10 ms
Triaxial connector (for sense) and Pulse delay: 0 to (Period – width)
(when delay + width > 100 ms) Pulse output limitation:
coaxial connector (for force) Pulse period resolution: 100 µs
Kelvin (remote sensing) When the pulse voltage is more
Pulse delay: 0 s than 1500 volts, the peak and
For MCSMU:
Dual triaxial connector, Kelvin base of pulse should be same
For HCSMU: polarities.
(remote sensing)
Pulse width: Pulse measurement delay:
For HVSMU:
50 µs to 1 ms (20 A range) 2 µs to (Period – pulse
High voltage triaxial connector,
50 µs to 2 s (10 µA to 1 A range) measurement time – 2 m) s,
non-Kelvin
Pulse width resolution: 2 µs 2 µs resolution
Voltage/current compliance (limiting)
The SMU can limit output voltage Pulse period: 5 ms to 5 s
or current to prevent damaging the Pulse period resolution: 100 µs
device under test. Pulse duty:
For 20 A range: ≤ 1%
Voltage:
0 V to ±200 V (HPSMU) For 10 µA to 1 A range
0 V to ±100 V (MPSMU) Period ≥ delay + width + 2 ms
0 V to ±40 V (HCSMU) (when delay + width ≤ 100 ms)
0 V to ±30 V (MCSMU) Period ≥ delay + width + 10 ms
0 V to ±3000 V (HVSMU) (when delay + width > 100 ms)
Current: Pulse delay: 0 to (Period–width)
±1 pA to ±1 A (HPSMU)
±1pA to ±100 mA (MPSMU) For MCSMU:
±10 nA to ±20 A (HCSMU) Pulse width:
±10 nA to ±1 A (MCSMU) 10 μs to 100 ms (1 A range)
±1 pA to ±8 mA (HVSMU) 10 μs to 2 s (10 μA to 100 mA
Compliance accuracy: range)
Same as the current or voltage set Pulse width resolution: 2 μs
accuracy.
Pulse period: 5 ms to 5 s
Power compliance
Pulse period resolution: 100 μs
For HPSMU: Pulse duty:
Power: 0.001 W to 20 W For 1 A range: ≤ 5%
Resolution: 0.001 W For 10 μA to 100m A range
For MPSMU:
Period ≥ delay + width + 2 ms
Power: 0.001 W to 2 W
(when delay + width ≤ 100 ms)
Resolution: 0.001 W
For HCSMU: Period ≥ delay + width + 10 ms
Power: 0.001 W to 40 W (DC) (when delay + width > 100 ms)
0.001 W to 400 W (Pulse) Pulse delay: 0 to (Period–width)
Resolution: 0.001 W
For MCSMU:
Power: 0.001 W to 3 W (DC)
0.001 W to 30 W (Pulse)
Resolution: 0.001 W”
For HVSMU:
No power compliance
11
Supplemental Characteristics
Current compliance setting accuracy Current source output resistance: Noise characteristics:
(for opposite polarity): ≥ 1013 Ω (HPSMU, MPSMU) For HPSMU, MPSMU and HVSMU
≥ 109 Ω (HCSMU, MCSMU, ≤ 1 A), (Filter ON for HPSMU and MPSMU.)
For HPSMU and MPSMU: 80 kΩ (HCSMU, 20 A) Voltage source:
For 1 pA to 10 nA ranges: ≥ 1012 Ω (HVSMU, at 10 nA range) 0.01% of V range (rms.)
V/I setting accuracy ±12% of Current source:
range Maximum allowable cable resistance:
(Kelvin connection) 0.1% of I range (rms.)
For 100 nA to 1 A ranges: For HCSMU
V/I setting accuracy ±2.5% of For HPSMU and MPSMU:
Sense: 10 Ω Voltage/Current source:
range 100 mV (0 to peak) max
For HCSMU and MCSMU: Force: 10 Ω (≤ 100 mA),
1.5 Ω (>100 mA) For MCSMU
For 10 µA to 1 A ranges:
For HCSMU: Voltage / Current source:
V/I setting accuracy ±2.5% of
Sense: 10 Ω 200 mV (0 to peak) max
range
For 20 A range (HCSMU): Force: 0.6 Ω Overshoot:
V/I setting accuracy ±0.6% of (with Low Force) (Filter ON for all SMUs)
range For MCSMU For HPSMU and MPSMU
For HVSMU: Sense: 10 Ω Voltage source: 0.03% of V range
For 1 nA to 10 nA ranges: Force : 1 Ω Current source: 1% of I range
V/I setting accuracy ±12% of (with Low Force) For HCSMU and MCSMU (filter ON)
range Maximum allowable inductance: Voltage/Current source:
For 100 nA to 10 mA ranges: 10% of range
For HCSMU and MCSMU:
V/I setting accuracy ±2.5% of For HVSMU
Force 3 µH
range Voltage source: 1V (resistive load)
(with Low Force (shield))
Current source: 1% of I range
Maximum load capacitance:
SMU pulse setting accuracy Range switching transient noise:
For HPSMU and MPSMU:
(fixed measurement range): For HPSMU and MPSMU (filter ON):
1 pA to 10 nA ranges: 1000 pF
For HPSMU and MPSMU: Voltage ranging: 250 mV
100 nA to 10 mA ranges: 10 nF
Width: ±0.5% ± 50 µs Current ranging: 70 mV
100 mA and 1 A ranges: 100 µF
Period: ±0.5% ± 100 µs For HCSMU and MCSMU:
For HCSMU:
For HCSMU and MCSMU: 10 µA to 1 A ranges:
10 µA to 10 mA ranges: 12 nF
Width: ±0.1% ± 2 µs Voltage ranging: 250 mV
Period: ±0.1% ± 100 µs
100 mA to 20 A ranges: 100 µF
For MCSMU: Current ranging: 70 mV
For HVSMU:
10µA to 10 mA range : 12 nF 20 A ranges:
Width: ±0.1% ± 2 µs
100 mA to 1 A range : 100 µF Voltage ranging: 5 V max
Period: ±0.5% ± 100 µs
For HVSMU: For HVSMU:
Minimum pulse measurement time: 1 nA to 1 µA ranges: 1000 pF Voltage ranging: 300 mV
16 µs (HPSMU and MPSMU) 10 µA to 10 mA ranges: 10 nF Current ranging: 300 mV
2 µs (HCSMU and MCSMU) Maximum guard capacitance: Maximum guard offset voltage:
6 µs (HVSMU) 900 pF (HPSMU and MPSMU) ±1 mV (HPSMU)
Voltage source output resistance: 1500 pF (HVSMU) ±3 mV (MPSMU)
(Force line, non-Kelvin connection) Maximum shield capacitance: ±5 mV (HVSMU)
0.2 Ω (HPSMU) 5000 pF (HPSMU, MPSMU and Maximum slew rate:
0.3Ω (MPSMU) HVSMU) 0.2 V/µs (HPSMU and MPSMU)
3 Ω (HVSMU, at 10 mA range) 1 V/µs (HCSMU and MCSMU)
0.4 V/µs (HVSMU)
Voltage measurement input
resistance: Output settling time
≥ 1013 Ω (HPSMU, MPSMU) For HVSMU:
≥ 109 Ω (HCSMU, MCSMU, ≤ 1 A), Output settling time: 500 µs
80 kΩ (HCSMU, 20 A) To reach 0.01% of settling value.
≥ 1012 Ω (HVSMU) Conditions:
100 V step, 8 mA compliance,
1000 pF load capacitance
12
MFCMU (multi frequency capacitance measurement unit) module speciications
13
Parameters EPOSC ZOSC
Oscillator level EPOSC (%) ZOSC (mΩ)
125 mV < VOSC ≤ 250 mV 0.03 x (250/ VOSC – 1) 5 x (250/VOSC – 1)
64 mV < VOSC ≤ 125 mV 0.03 x (125/ VOSC – 1) 5 x (125/VOSC – 1)
32 mV < VOSC ≤ 64 mV 0.03 x (64/ VOSC – 1) 5 x (64/VOSC – 1)
VOSC ≤ 32 mV 0.03 x (32/ VOSC – 1) 5 x (64/VOSC – 1)
VOSC is oscillator level in mV.
Parameters YOSC YO EP ZS
Frequency YOSC (nS) YO (nS) EP (%) ZS (mΩ)
1 kHz ≤ f ≤ 200 kHz 1 x (125/ VOSC – 0.5) 1.5 0.095 5.0
200 kHz < f ≤ 1 MHz 2 x (125/ VOSC – 0.5) 3.0 0.095 5.0
1 MHz < f ≤ 2 MHz 2 x (125/ VOSC – 0.5) 3.0 0.28 5.0
2 MHz < f 20 x (125/ VOSC – 0.5) 30.0 0.28 5.0
f is frequency in Hz.
VOSC is oscillator level in mV.
14
UHC (Ultra High Current) Expander / Fixture (N1265A) Speciications
Speciications
Functions:
Fixture capability
Current expander capability
Expands the B1505A’s current capability up to 1500 A. Current expansion is made using the Ultra High Current Unit
(UHCU), which is comprised of an external module and either two MCSMUs, two HCSMUs or one MCSMU and one
HCSMU.
Selector capability
This allows the user to switch the output between the UHCU and other modules connected to the selector input ports. The
modules supported on the high-voltage input port are the HVSMU and HVMCU; the modules supported on the SMU input
port are the HPSMU and MPSMU.
Channels:
Channel Number Input Output
SMU 6 (When using non-Kelvin Triaxial1 Banana
connections)
3 (When using Kelvin
connections)
UHV 1 UHV coaxial (High), SHV (Low) UHV coaxial (High), SHV (Low)
Bias Tee 1 SHV x 2(High, Low) SHV x 2 (High, Low)
Gate 1 Triaxial x 2 (Force, Sense) Banana x 2 (High, Low)
control
Selector 12 HV Triaxial x 1 Banana x 6 (High Force/Sense, Low
Triaxial x 2 (Force, Sense) Force/Sense, Guard, Chassis)
1. Either the HCSMU or the Dual HCSMU can be connected to the SMU 3 port.
2. The UHCU or any module connected to one of the other two selector input terminals can be connected to the output terminal.
15
UHCU:
1. Maximum voltage compliance in current pulse mode is 63 V. Over 400 A at 500 A range and over 1200 A at 1500 A range are
supplemental characteristics.
2. Accuracy is defined with 1ms pulse width at 500 A range and with 500 μs pulse width at 1500 A range.
3. ±(% of reading value + fixed offset in A + proportional offset in A), Vo is the Output Voltage.
16
Other functionality UHC measurement and output range
Fiilter
Filter can be used for UHC output in Current (A)
current mode at 500 A range. 1500
Thermocouple input: 2ea
Two K-type thermocouple inputs I = (60 - V)/ 0.04
Leakage
Selector channel
HVSMU is applied at High Sense terminal: less than 1n A
HPSMU/MPSMU is applied at High Force terminal: less than 10 nA
UHVU channel
Less than 1nA
SMU channel
Less than 1nA
17
HVSMU Current Expander (N1266A) Speciications
Speciications
Functions:
Current expander capability
Expands HVSMU current up to 2.5 A. Current expansion is made using the High Voltage Medium Current Unit (HVMCU),
which is comprised of a module in the N1266A, HVSMU and two MCSMUs.
Selector capability
This allows the connections between the output terminal to be switched between the HVMCU and the HVSMU. The
HVSMU output can be routed either directly or through a 100 kΩ resistor.
Output Terminals:
High (HV Triaxial)
Low (BNC)
Maximum output:
HVSMU : ±3000 V/4 mA, ±1500 V/8 mA
HVMCU : Refer to HVMCU specification
HVMCU
18
HVMCU Pulse width and resolution
Output range Pulse width Resolution
1500 V / 2.5 A 10 μsec – 100 μsec 2 μsec
2200 V / 1.1 A 10 μsec – 100 μsec 2 μsec
2200 V / 110 mA 10 μsec – 1 msec 2 μsec
19
UHV (Ultra High Voltage) Expander (N1268A) Speciications
Speciications
Voltage range, resolution, and accuracy1
Voltage range Force resolution Measure resolution Setting accuracy2, 3 Measure accuracy2
±(% + V) ±(% + V)
± 10 kV 10 mV 10 mV ±(1.2 + 42) ±(1.2 + 42)
1. N1268A is controlled and makes measurement with two MCSMUs or a combination of a HCSMU and a MCSMU.
2. ±(% of reading value + fixed offset in V)
3. Setting accuracy is defined at open load.
DC and pulse
-0.02
20
N1267A High Voltage Source Monitor Unit / High Current Source
Monitor Unit Fast Switch
Features
The N1267A supports fast switching between the HVSMU and HCSMU to enable the measurement of the Gallium Nitride
current collapse effect.
The N1267A switch requires one MCSMU in the B1505A mainframe for control. The gate of the DUT (Device Under Test)
can be driven by either an MCSMU or an HCSMU.
Note #1: The N1267A can only be used with the B1513B HVSMU; it cannot be used with the B1513A HVSMU.
Note #2: The N1267A does not support the two HCSMU 40 A configuration.
Note #3: The N1267A does not support the N1265A test fixture/current expander.
Speciications
Input terminals:
HVSMU port, 1ea (HV triaxial)
HCSMU port, 1ea (Force: BNC, Sense: Triaxial)
MCSMU port, 1ea (Force/Sense: Triaxial)
GND port, 1ea (Triaxial)
Output terminals: High (HV triaxial), Low (BNC)
Maximum current: 20 A
Maximum voltage: 3000 V
Measurement mode
GaN Current collapse (Dynamic I-V) measure mode
1. I-V time domain measurement
2. I-V trace measurement
Static characteristics mode
1. Id-Vds, Vf-If measurement
2. Id(off)-Vds, Vr-Ir measurement
21
GaN current collapse measure mode
To make the GaN current collapse measurement, the HVSMU first applies high voltage stress to the DUT when the DUT is
in the OFF-state. Next the HVSMU performs voltage measurement and the HCSMU performs I-V measurement to monitor
the ON-state characteristics of the DUT. When making the ON-state measurement, the HVSMU is measuring voltage and
both the HVSMU and HCSMU are used to measure the total current.
HVSMU Source setting range for OFF-state
Voltage Current
+1 V - +3000 V 1) 4 mA (V > 1500 V), 8 mA (V ≤ 1500 V)
1
Setting value must be the ON state voltage plus 1 V or more.
The following information applies to measurement of the DUT OFF-state static characteristics. The N1267A ensures that
the DUT is in the OFF-state during these measurements. The HCSMU applies 0 V. At the same time, the HVSMU performs
I-V measurement and measures Vds or Vr. The Id(Off) or Ir is determined by adding together the total current measured by
both the HCSMU and the HVSMU.
23
N1259A-013 Curve Tracer test Supplemental characteristics N1259A-300 module selector
adapter socket module Leakage current: 10 pA at 100 V for test ixture
N1259A-022 100 kΩ
Specifications Specifications
Number of terminals:
resistor box
Input terminals:
Sockets, 6 ea. Specifications HPSMU port1, 1 ea.
(Ø4 mm jack (banana)) Input/output terminals: Force, sense (Triaxial)
Test adapter interface:* Ø4 mm jack (banana), 1 ea. HCSMU port, 1 ea.
Sockets, 6 ea. Resistance: 100 kΩ ±5% Force (BNC), sense (Triaxial)
(Ø4 mm jack (banana)) Maximum voltage: ±3000 V HVSMU port2, 1 ea. (HV triaxial)
Maximum voltage at terminals: Power rating: 6.4 W GNDU port, 1 ea. (Triaxial)
3000V Vdc Digital I/O port, 1 ea. (D-sub 25 pin)
Maximum current for terminals: Supplemental characteristics AC power line connector, 1 ea.
For Collector/Drain Force and Leakage current: 10 pA at 100 V 1. Either HPSMU or MPSMU can be connected
Emitter/Source Force to HPSMU port.
2. Either HVSMU or HVMCU can be connected
39 A (DC), 500 A (Pulse) N1259A-030 1 kΩ resistor box to HVSMU port.
For others Output terminal:
1A (DC), 20 A (Pulse)
for gate
High force and guard
*A test adapter for Tektronix curve Specifications High sense and guard
tracers (370B/371B) can be con- Input/output terminals: Low force
nected to this interface. Ø4 mm jack (banana), 1 ea. Low sense
Resistance: 1 kΩ ±10% (Ø4 mm jack (banana))
N1259A-020 high voltage Maximum voltage: ±200 V Protection:
bias-tee Maximum power: 1 W HPSMU, GNDU, HCSMU Low Force
Power indicator:
Specifications Supplemental characteristics LED turns yellow when AC power is
Input terminals: Leakage current: 10 pA at 100 V applied and turns green the module
DC bias input, 1 ea. selector is ready to use.
(Ø4 mm jack (banana)) N1259A-035 Universal resistor Status indicator:
MFCMU port, 1 ea. Green LED lights to indicate the
Hcur, Hpot, Lcur, Lpot, (BNC)
box
present connection path of module
Guard input, 1ea (Ø4 mm banana Specifications selector; Open, HCSMU, HPSMU, or
jack) Input/output terminals: HVSMU.
Output terminal: Ø4 mm banana jack, 1 ea. Maximum voltage/current:
MFCMU port Resistance: Installed by a user For HPSMU port:
High (SHV) Maximum voltage for terminals: ±200 V/1 A
Low (SHV) ±3000 V For HCSMU port:
External DC bias voltage: ±3000 V ±40 V/2 A, ±20 V/30 A
Frequency: (Pulse width 1 ms, duty 1%)
10 kHz to 1 MHz (150 Ω at 10 kHz) For HVSMU:
Series capacitance: 110 nF ±5% ±3000 V/4 mA,
Input resistance: 100 kΩ ±1% ±1500 V/2.5 A, ±2200 V/1.1 A
N1259A-021 1 MΩ
resistor box
Specifications
Input/output terminals:
Ø4 mm jack (banana), 1 ea.
Resistance: 1 MΩ ±5%
Maximum voltage: ±3000 V
Power rating: 9 W
24
Supplemental characteristics N1261A-002 protection adapter N1262A-002 100 kΩ resistor box
Leakage current: for GNDU (BNC output)
For HPSMU: Specifications
10 pA at 200 V Specifications Input terminals:
For HCSMU: Input terminals: HVSMU port, 1 ea. (HV triaxial)
100 pA at 10 V (High Force to Force/Sense (Triaxial) Output terminals:
Low Force, High Sense to Low Output terminals: SHV connector, 1 ea.
Sense) Force (BNC) Resistance: 100 kΩ ±5%
For HVSMU: Sense (BNC) Maximum voltage: ±3000 V
10 pA at 1500 V (humidity range: Maximum power: 6.4 W
20% to 70% RH) N1261A-003 protection adapter
30 pA at 3000 V (humidity range: for HPSMU (HV triaxial output)
Supplemental characteristics
20% to 50% RH) Leakage current: 10 pA at 100 V
Specifications
N1260A high voltage Input terminals1: N1262A-010 1 kΩ resistor box
Force (Triaxial) for gate (triaxial output)
bias-tee Sense (Triaxial)
Output terminals: Specifications
Specifications Force (HV triaxial) Input terminals:
Input terminals: Sense (HV triaxial) Triaxial connector, 1 ea.
HVSMU port, 1 ea. (HV triaxial) 1. Either the HPSMU or the MPSMU can be Output terminals:
MFCMU port, 1 ea. connected to HPSMU port. Triaxial connector, 1 ea.
(4 BNC, Hp, Hc, Lp, Hc) Resistance: 1 kΩ ±10%
Output terminal: Supplemental characteristics
Maximum voltage: ±200 V
H-AC Guard (SHV connector) Leakage current: 10 pA at 200 V
Maximum power: 1 W
L-AC Guard (SHV connector)
External DC bias voltage: ±3000 V N1261A-004 protection adapter Supplemental characteristics
Frequency: for GNDU (SHV output) Leakage current: 10 pA at 100 V
10 kHz to 1 MHz (150 Ω at 10 kHz)
Series capacitance: 110 nF ±5% Specifications
Input terminals: N1262A-011 1 kΩ resistor box
Input resistance: 100 kΩ ±1%
Force/Sense (Triaxial) for gate (SHV output)
Output terminals:
Specifications
Force (SHV)
N1261A protection Sense (SHV) Input terminals:
HV triaxial connector, 1 ea.
adapter Output terminals:
SHV connector, 1 ea.
N1261A-001 protection adapter N1262A Resistor Box Resistance: 1 kΩ ±10%
for HPSMU (triaxial output) Maximum voltage: ±3000 V
N1262A-001 1 MΩ resistor box Maximum power: 1 W
Specifications
Input terminals: Specifications Supplemental characteristics
Force (Triaxial) Input terminals: Leakage current: 10 pA at 100 V
Sense (Triaxial) HVSMU port, 1 ea. (HV triaxial)
Output terminals: Output terminals:
Force (Triaxial) SHV connector, 1 ea.
Sense (Triaxial) Resistance: 1 MΩ ±5%
1. Either the HPSMU or the MPSMU can be Maximum voltage: ±3000 V
connected to HPSMU port. Maximum power: 9 W
25
N1262A-020 Universal resistor Accessories for N1265A N1265A-011 Universal Socket
box, Triaxial Module
N1254A-524 500 A Ultra High
Specifications Specifications
Input terminals: Current Prober System Cable Number of terminal:
Triaxial connector, 1 ea. Specifications Sockets, 6 ea. (Ø4 mm jack
Output terminals: Input terminals: 8 ea. (Ø4 mm jack (banana))
Triaxial connector, 1 ea. (banana)) Maximum voltage for terminals:
Resistance: Installed by user Selector Output 3000 Vdc
Maximum voltage for terminals: ±200 V High Force Universal blank area :
High Sense 90 mm (W) x 81 mm (D)
N1262A-021 Universal resistor Low Force
box, HV Triaxial to SHV Low Sense N1265A-013 Curve Tracer Test
Guard Adapter Socket Module
Specifications Gate output
Input terminals: High Force Specifications
HVSMU port, 1 ea. (HV triaxial) Low Force Number of terminals: Sockets, 6 ea.
Output terminals: Chassis (Ø4 mm jack (banana))
SHV connector, 1 ea. Output terminals Test adapter interface:*
Resistance: Installed by user High Force (Ø4 mm jack Sockets, 6 ea. (Ø4 mm jack
Maximum voltage for terminals: (banana)) (banana))
±3000 V Low Force (Ø4 mm jack (banana)) Maximum voltage at terminals:
High Sense (HV triaxial) 3000V Vdc
N1262A-023 Universal resistor Low Sense (BNC) Maximum current for terminals:
box for Ultra High Voltage Gate (BNC) For Collector/Drain Force and
Maximum voltage / current Emitter/Source Force
Specifications For High Force 39 A (DC), 500 A (Pulse)
Input terminals: ±3000 V/39 A (DC), 500 A (Pulse) For others
UHV coaxial connector, 1 ea. For Low Force 1A (DC), 20 A (Pulse)
Output terminals: ±200 V/39 A (DC), 500 A (Pulse) *A test adapter for Tektronix curve
UHV coaxial connector, 1 ea. For High Sense tracers (370B/371B) can be con-
Resistance: Installed by user ±3000 V/1 A nected to this interface.
Maximum voltage for terminals: For Low Sense, Gate
±10 kV ±200 V/1 A N1265A-035 Universal R-Box
for N1265A
N1262A-036 50 Ohm N1265A-010 500 A Ultra High
Termination Adapter Specifications
Current 3-pin Inline Package Input: 4 ea. (Ø4 mm plug (banana))
Specifications Socket Module High (Force, Sense)
Input terminal (BNC) Low (Force, Sense)
Specifications
Output terminal (BNC) Output terminals: 2 ea. (Ø4 mm jack
Number of terminal:
Maximum power: 1 W (banana))
Sockets, 6 ea. (Ø4 mm jack (banana))
High, Low
DUT interface:
Resistance: Installed by a user
Inline package socket (3-pin)
Maximum voltage for terminals: ±200 V
Maximum voltage for terminals:
3000 Vdc
Maximum current for terminals:
For Force
39 A (DC), 500 A (Pulse)
For sense
1A (DC), 20 A (Pulse)
26
N1265A-040 10 kV Ultra N1269A Ultra High accessed using the B1505A’s 15-inch
touch screen, as well as through an
High Voltage Gate Protection
Adapter
Voltage Connection optional USB keyboard and mouse.
EasyEXPERT also allows you the
Adapter option of storing the test conditions
Specifications
Feature and measurement data automatically
Input: 4 ea. (Ø4 mm plug (banana)) after each measurement into unique
High (Force, Sense) To make the connection simple and to workspaces. This ensures that valu-
Low (Force, Sense) protect measurement resources from able information is not lost and that
Output terminals: 2 ea. (Ø4 mm jack unexpected surge when connecting measurements can be repeated at a
(banana)) UHVU to wafer prober. later date. Finally, EasyEXPERT has
High, Low Specifications built-in analysis capabilities and a
Maximum voltage: ±200 V Input terminals: graphical programming environment
Maximum surge voltage: ±10 kV Gate MCSMU Force, 1ea (Triaxial) that facilitate the development of
Gate MCSMU Sense, 1ea (Triaxial) complex testing algorithms.
Chuck MCSMU Force, 1ea (Triaxial) Key features:
N1265A-041 Thermocouple, Chuck MCSMU Sense, 1ea • Ready-to-use application test library
Type K, 2 ea (Triaxial) • Multiple measurement modes
UHV Low, 1ea (HV triaxial) (application test, classic test, tracer
Feature
Output terminals: 3ea (SHV) test, oscilloscope view and quick
N1265A-041 can be connected to
Gate, Chuck, Source test)
Thermocouple terminal inside the Maximum voltage: ±200 V • Multiple measurement functions
N1265A and enables B1505A to read Maximum surge voltage: ±10 kV (spot, sweep, time sampling, C-V,
out temperature at the top of the C-f, C-t, etc.)
thermocouple. • Data display, analysis and arithme
Specifications Agilent EasyEXPERT tic functions
Connector: Type K plug Software • Workspace and data management
Length: 3000 mm • External instrument control
Agilent EasyEXPERT, resident • Multiple programming methods
Temperature range: -50 °C to +180 °C GUI-based software running on (EasyEXPERT remote control and
the B1505A’s embedded Windows FLEX GPIB control)
N1265A-045 Container for 7 platform, supports efficient and • Multiple interface (USB, LAN, GPIB
repeatable device characterization
Protection Adapter and Bias ranging from interactive manual
and digital I/O)
Tee measurements all the way up to Application library
test automation across a wafer in
Feature EasyEXPERT comes with over 40
conjunction with an automatic wafer
N1265A-045 can accommodate prober. With hundreds of ready-to-use application tests conveniently
protection adapters and bias tee measurements (application tests) organized by device type, application,
which are used with N1265A to make furnished at no charge, EasyEXPERT and technology. You can easily edit
the measurement environment clean makes it easy to perform complex and customize the furnished applica-
and safe device characterization immedi- tion tests to fit your specific needs.
Specifications ately. The EasyEXPERT GUI can be Application tests are provided for the
Dimension: 420 mm W x 193 mm H x following categories; they are subject
565 mm D to change without notice.
Weight: 15 kg Device Type Application Tests
Maximum superimposed load: 50 kg
Power MOSFET Id-Vds, Rds-Id, Id-Vgs, Vth, Cgs, Cds, Cgd, Current col-
(Si, GaN) lapse, Breakdown, QSCV, etc.
IGBT Ic-Vce, Ic-Vge, Vth, Cge, Cce, Cgc, Breakdown, etc.
SiC Id-Vds, Rds-Id, Id-Vgs, Vth, Cgs, Cds, Cgd, Breakdown,
QSCV, etc.
Power BJT Ic-Vce, Vce(sat), Ic-Vcbo, Ic-Vceo, Ie-Vbeo, etc.
Power Diode If-Vf, Ir-Vr, Cj-Vr, etc.
Capacitor C-V, C-f, C-t, Ieak-V, Breakdown, TDDB, etc.
And more And more
27
Measurement modes and HCSMU, MCSMU, HVSMU, UHCU, Measurement modes:
functions HVMCU, and UHVU. The oscilloscope The Agilent B1505A supports the
view can display the pulsed waveform following measurement modes:
Operation mode:
timings at any (user specified) sweep
Application test mode •IV measurement
step of the sweep output.
The application test mode provides - Spot
Sampling interval: - Staircase sweep
application oriented point-and-
2 μs (HCSMU/MCSMU/UHCU/ - Pulsed spot
click test setup and execution. An
HVMCU/UHVU) - Pulsed sweep
application test can be selected from
6 μs (HVSMU) - Staircase sweep with pulsed bias
the library by device type and desired
measurement, and then executed Sampling points: - Sampling
after modifying the default input 2000 Sa (HCSMU/MCSMU/UHCU/ - Multi-channel sweep
parameters as needed. HVMCU/UHVU) - Multi-channel pulsed sweep
4000 Sa (HVSMU) - List sweep
Classic test mode
Marker function: - Linear search1
The classic test mode provides func-
Read-out for each data channel - Binary search1
tion oriented test setup and execu-
tion with the same look, feel, and Resolution: 2μs • C measurement
terminology of the 4155/4156 user Data saving: - Spot C
interface. In addition, it improves the Numeric: Text/CSV/XMLSS - CV (DC bias) sweep
4155/4156 user interface by taking Image: EMF/BMP/JPG/PNG - Pulsed spot C
full advantage of EasyEXPERT’s GUI - Pulsed sweep CV
Quick test mode - C-t sampling
features.
A GUI-based Quick Test mode enables - C-f sweep
Tracer test mode you to perform test sequencing without - CV (AC level) sweep
The tracer test mode offers intuitive programming. You can select, copy, - Quasi-Static CV (QSCV)
and interactive sweep control using a rearrange and cut-and-paste any appli- 1. Supported only by FLEX commands.
rotary knob similar to a curve tracer. cation tests with a few simple mouse
Just like an analog curve tracer, you clicks. Once you have selected and Sweep measurement
can sweep in only one direction (use- arranged your tests, simply click on the Number of steps: 1 to 10001 (SMU),
ful for R&D device analysis) or in both measurement button to begin running 1 to 1001 (CMU)
directions (useful in failure analysis an automated test sequence. Sweep mode: Linear or logarithmic
applications). Test set ups created in (log)
tracer test mode can be seamlessly Sweep direction: Single or double
and instantaneously transferred to sweep
classic test mode for further detailed Hold time:
measurement and analysis. 0 to 655.35 s, 10 ms resolution
Each SMU can sweep using VAR1 Delay time:
(primary sweep), VAR2 (secondary 0 to 65.535 s, 100 μs resolution
sweep), or VAR1’ (synchronous 0 to 655.35 s, 100 μs resolution
sweep). (CV (AC level) sweep, C-f sweep)
Step delay time:
Oscilloscope view
0 to 1 s, 100 μs resolution
The oscilloscope view (available in
Step output trigger delay time:
tracer test mode) displays measured
0 to (delay time) s, 100 μs
current or voltage data versus time.
resolution
The pulsed measurement waveforms
Step measurement trigger delay
appear in a separate window for easy
time:
verification of the measurement tim-
0 to 65.535 s, 100 μs resolution
ings. This function is useful for verifying
waveform timings and debugging
pulsed measurements. The following
modules are supported in this view:
28
Sampling (time domain) measurement Current offset cancel Data variable display
Displays the time sampled voltage/ This function subtracts the offset Up to 20 user-defined parameters
current data (by SMU) versus time. current from the current measure- can be displayed on the graphics
Sampling channels: Up to 10 ment raw data, and returns the screen.
Sampling mode: Linear, logarithmic result as the measurement data.
Automatic analysis function
(log) This function is used to compen-
Sampling points: sate the error factor (offset current) On a graphics plot, the markers
For linear sampling: caused by the measurement path and lines can be automatically
1 to 100,001/(number of channels) such as the measurement cables, located using the auto analysis
For log sampling: manipulators, or probe card. setup. Parameters can be auto
1 to 1+ (number of data for 11 matically determined using
Time stamp automatic analysis, user function,
decades) The B1505A supports a time stamp
Sampling interval range: and read out functions.
function utilizing an internal quartz
100 μs to 2ms, 10μs resolution clock.
Analysis functions
2 ms to 65.535 s, 1 ms resolution Up to 20 user-defined analysis
Resolution: 100 μs
For < 2ms, the interval is ≥ 100 μs functions can be defined using
+20 μs x (num. of channels – 1) Data display, analysis and arithmetic expressions.
Hold time, initial wait time: arithmetic functions Measured data, pre-defined
-90 ms to -100 μs, 100 μs resolu variables, and read out functions
tion Data Display
can be used in the computation.
0 to 655.35 s, 10 ms resolution X-Y graph plot The results can be displayed on
Measurement time resolution: 100 μs X-axis and up to eight Y-axes, the LCD.
Other measurement characteristics linear and log scale, real time
Read out functions
graph plotting. X-Y graph plot
Measurement control The read out functions are built-in
can be printed or stored as image
Single, repeat, append, and stop functions for reading various
data to clip board or mass stor
values related to the marker,
SMU setting capabilities age device. (File type: bmp, gif,
Limited auto ranging, voltage/ cursor, or line.
png, emf)
current compliance, power compli- Scale: Arithmetic functions
ance, automatic sweep abort func- Auto scale and zoom User functions
tions, self-test, and self-calibration Marker: Up to 20 user-defined functions can
Standby mode Marker to min/max, interpolation, be defined using arithmetic expres-
SMUs in “Standby” remain pro- direct marker, and marker skip sions.
grammed to their specified output Cursor: Measured data and pre-defined
value even as other units are reset Direct cursor variables can be used in the compu-
for the next measurement. Line: tation. The results can be displayed
Bias hold function Two lines, normal mode, grad on the LCD.
This function allows you to keep a mode, tangent mode, and regres Arithmetic operators
source active between measure- sion mode +, -, *, /, ^, abs (absolute value),
ments. The source module will Overlay graph comparison: at (arc tangent), avg (averaging),
apply the specified bias between Graphical plots can be overlaid. cond (conditional evaluation), delta,
measurements when running diff (differential), exp (exponent),
classic tests inside an application List display integ (integration), lgt (logarithm,
test, in quick test mode, or during a Measurement data and calculated base 10), log (logarithm, base e),
repeated measurement. The func- user function data are listed in mavg (moving average), max, min,
tion ceases as soon as these condi- conjunction with sweep step num- sqrt, trigonometric function, inverse
tions end or when a measurement ber or time domain sampling trigonometric function, and so on.
that does not use this function is step number. Up to 20 data sets
started. can be displayed.
29
Physical constants External instrument control Furnished software
Keyboard constants are stored in External instruments supported by • Prober control execution files
memory as follows: application tests:
q: Electron charge, 1.602177E-19 C • Desktop EasyEXPERT software
Agilent 4284A/E4980A, 81110A
k: Boltzmann’s constant, 1.380658E-23 • 4155/56 setup file converter tool
3458A
ε (e): Dielectric constant of vacuum, This tool can convert 4155 and
8.854188E-12 Prober control
4156 measurement setup files
Popular semi- or full-automatic
Engineering units (file extensions MES or DAT) into
wafer probers are supported by
The following unit symbols are also equivalent EasyEXPERT/Desktop
EasyEXPERT. You can define wafer,
available on the keyboard: EasyEXPERT classic test mode
die, and module information for prob-
a (10-18), f (10-15), p (10-12), n (10-9), setup files.
ing across an entire wafer. You can
u or μ (10-6), m (10-3), k (103), M (106), also combine wafer prober control • MDM file converter
G (109), T (1012) , P (1015) with either Quick Test mode or an This tool can convert data in the
application test based test sequence EasyEXPERT file formats (XTR/
Workspace and data to perform multiple testing on various ZTR) to Agilent IC-CAP MDM file
management devices across the wafer. format.
Workspace Only the following Classic Mode
Workspaces are separate work envi- Program and interface measurements made using
ronments residing on the B1505A’s capabilities EasyEXPERT are supported:
internal hard disk drive. Every - IV Sweep
workspace supports the following Data storage
Hard disk drive, DVD-R drive - Multi-channel IV Sweep
features:
Interfaces - CV Sweep
• Setup and execute the measure
ment GPIB, interlock, USB (USB 2.0, front 2, Supported operating systems:
rear 2), LAN (1000BASE-T/100BASE-
• Save/Recall “My Favorite Setups” TX/10BASE-T), trigger in/out, digital
Microsoft Windows XP Professional
• Save/Recall measurement data and (Service Pack 3 or later), Windows
I/O, VGA video output
settings Vista Business (Service Pack 2 or
Remote control capabilities later (32bit only)), and Windows 7
• Import/Export device definition, • FLEX commands (GPIB) Professional (Service Pack 1 or later
measurement settings, my favorite • EasyEXPERT remote control func (32bit and 64bit))
setup, measurement data, and tion (LAN)
application library Supported language: English (US))
Trigger I/O
• Test result data management Supported .NET Framework:
This feature is only available using
Microsoft .NET Framework 3.5 SP1
• Private/public accessibility setting GPIB FLEX commands.
Data auto record / auto export Trigger in/out synchronization pulses
EasyEXPERT has the ability to before and after setting and measur-
automatically store the measurement ing DC voltage and current. Arbitrary
setup and data within a workspace. It trigger events can be masked or
can also export measurement data in activated independently.
real time, in a variety of formats. You
can save data to any storage drive Recommended GPIB I/F
connected to the instrument’s PC. Interface B1505A
1
Import/export files 82350B PCI
File type: 82357A USB
2
Agilent
Agilent EasyEXPERT format, XML-SS 2
82357B USB
format, CSV format
2
National GPIB-USB-HS USB
Workspace management
Insturment
The EasyEXPERT has the ability to
1. An 82350B card is highly recommended because of stability and speed.
import/export a workspace for back- 2. USB GPIB interfaces might cause serial poll error intermittently due to the intrinsic communica-
up and portability. tion scheme differences.
30
Weight
General speciications B1505A (empty): 20 kg
B1511A: 1.1 kg
Temperature range Regulatory compliance B1510A: 2.0 kg
Operating: +5 °C to +40 °C EMC: B1512A: 2.1 kg
Storage: -20 °C to +60 °C IEC 61326-1 / EN 61326-1 B1513B: 2.0 kg
Humidity range1 Canada: ICES/NMB-001 B1514A: 1.3 kg
Operating: 20% to 70% RH, AS/NZS CISPR 11 B1520A: 1.3 kg
non-condensing Safety: N1258A: 5.0 kg
Storage: 10% to 90% RH, IEC61010-1 / EN 61010-1 N1259A: 12.0 kg
non-condensing CAN/CSA-C22.2 No. 61010-1 N1260A: 0.6 kg
Storage: 20% to 80% RH, N1261A: 0.3 kg
Certiication N1262A: 0.3 kg
non-condensing (N1268A) CE, cCSAus, C-Tick N1265A: 30 kg
Altitude N1266A: 10 kg
Operating: 0 m to 2,000 m (6,561 ft) Dimensions
B1505A: N1267A: 0.8 kg
Storage: 0 m to 4,600 m (15,092 ft) N1268A: 18 kg
0 m to 2,000 m (6,561 ft) (N1268A) 420 mm W x 330 mm H x 575 mm D
N1258A module selector: N1269A: 0.4 kg
Power requirement
330 mm W x 120 mm H x 410 mm D Furnished accessories
ac Voltage: 90 V to 264 V
N1259A test fixture: Measurement cables and adapter
Line Frequency: 47 Hz to 63 Hz
420 mm W x 272 mm H x 410 mm D Triaxial cable for HPSMU, MPSMU
Maximum volt-amps (VA) N1260A High voltage bias-tee:
B1505A: 900 VA and MCSMU, 2 ea.
164 mm W x 53 mm H x 125 mm D HCSMU cable, 1 ea.
N1258A: 65VA N1261A-001 HPSMU protection
N1259A-300: 35VA HCSMU Kelvin adapter, 1 ea.
adapter (Triaxial output): HVSMU cable, 1 ea.
N1265A: 400 VA 80 mm W x 40 mm H x 110 mm D
N1266A: 60 VA Interlock cable, 1 ea.
N1261A-002 GNDU protection adapter Ground unit cable, 1 ea.
N1268A: 350 VA (BNC output): Keyboard, 1 ea.
Acoustic Noise Emission 80 mm W x 40 mm H x 110 mm D Mouse, 1 ea.
Lpa < 70 dB N1261A-003 HPSMU protection Stylus pen, 1 ea.
About measurement accuracy adapter (HV triaxial output): Power cable, 1 ea.
RF electromagnetic ield and SMU 90 mm W x 40 mm H x 140 mm D Manual CD-ROM, 1 ea.
measurement accuracy: SMU voltage N1261A-004 GNDU protection adapter Desktop EasyEXPERT CD-ROM, 1 ea.
and current measurement accuracy (SHV output): License-to-use for EasyEXPERT and
can be affected by RF electromagnet- 80 mm W x 40 mm H x 125 mm D Desktop EasyEXPERT,
ic ield strengths greater than 3 V/m N1262A resister box: Software CD-ROM
in the frequency range of 80 MHz 50 mm W x 40 mm H x 125 mm D (including utility tools)
to 1 GHz. The extent of this effect N1265A UHC expander / fixture: Disk set for Agilent
depends upon how the instrument is 420 mm W x 285mm H x 575 mm D 4155B/4155C/4156B/4156C
positioned and shielded. N1266A HVSMU current expander: firmware update, 1 set
Induced RF ield noise and SMU 420 mm W x 75 mm H x 575 mm D SMU number label for the B1505A
measurement accuracy: N1267A HVSMU / HCSMU fast installed with SMU, 1 sheet
SMU voltage and current measure- switch: N1258A : Digital I/O cable, 1 ea.
ment accuracy can be affected by 202 mm W x 56 mm H x 175 mm D N1259A-300 : Digital I/O cable, 1 ea.
induced RF ield noise strengths N1268A UHV expander: N1265A : Digital I/O cable, 1 ea.
greater than 3 Vrms in the frequency 420 mm W x 222 mm H x 482 mm D N1266A : Digital I/O cable, 1 ea.
range of 150 kHz to 80 MHz. The N1269A Ultra High Voltage N1268A : Digital I/O cable, 1 ea.,
extent of this effect depends upon Connection Adapter: Interlock cable, 1 ea.
how the instrument is positioned and 134 mm W x 56 mm H x 150 mm D
shielded.
32
Order Information
33
www.agilent.com
www.agilent.com/find/b1505a