Professional Documents
Culture Documents
DIFFERENTIAL PRESSURE
SWITCH
User Manual
Part No. 9A-C10766, Rev. 01
Model 581A-0 October 2011
Contents
Safety............................................................................................................. 2
Section 1—Introduction.................................................................................. 3
General.......................................................................................................... 3
Wiring......................................................................................................... 5
Differential Pressure Unit........................................................................... 5
Specifications . .............................................................................................. 6
General:..................................................................................................... 6
Nuclear Qualifications................................................................................ 6
Application Specifications ......................................................................... 7
Storage Specifications .............................................................................. 7
Section 2—Installation.................................................................................... 9
Mounting/Piping/DPU Installation.................................................................. 9
Electrical Connection..................................................................................... 9
Switch Use..................................................................................................... 9
Startup......................................................................................................... 10
Switch Wiring............................................................................................... 10
Section 3—Maintenance and Calibration.....................................................11
Tools.............................................................................................................11
Bezel/Lens Assembly or Cover Installation and Removal........................... 12
Pointer Installation and Removal................................................................. 13
Pointer Installation . ................................................................................. 13
Pointer Removal....................................................................................... 14
Scale Plate Installation and Removal.......................................................... 14
Indicator Calibration..................................................................................... 15
Drive Arm Tightness Test.......................................................................... 17
Drive Arm Stop Adjustment...................................................................... 17
Switch Calibration ...................................................................................... 17
Calibration Setup...................................................................................... 18
Linkage Calibration Procedure ................................................................ 18
Changing Switch Set Point ......................................................................... 20
Definitions of Terms.................................................................................. 20
Best Practices for Set Points.................................................................... 20
Changing Set Point of an In-Service Instrument...................................... 20
Changing Set Point of an Out-of-Service Instrument............................... 21
Range Changes....................................................................................... 22
Troubleshooting .......................................................................................... 22
Section 4—Assembly Drawing and Parts List............................................ 25
Section 5—Dimensional Drawings.............................................................. 33
Appendix A—Model 199 DPU......................................................................A-1
DPU Description.........................................................................................A-1
Specifications..........................................................................................A-2
DPU Components...................................................................................A-2
DPU Installation..........................................................................................A-5
Unpacking...............................................................................................A-5
Mounting.................................................................................................A-6
Piping—Standard Practices....................................................................A-6
Piping Diagrams......................................................................................A-8
Startup...................................................................................................A-14
DPU Maintenance.................................................................................A-14
Required Tools......................................................................................A-14
DPU Cleaning and Inspection...............................................................A-15
Calibration Setup...................................................................................A-16
Torque Tube Rotation Check (Replacement Units)...............................A-16
Changing the DPU Range.....................................................................A-17
Setting Bellows Travel...........................................................................A-21
Bellows Unit Assembly (BUA) Replacement.........................................A-21
Attaching Drive Arm to Torque Tube......................................................A-22
Drive Arm Tightness Test.......................................................................A-23
Adjusting Pulsation Dampener..............................................................A-23
Troubleshooting.....................................................................................A-24
DPU Assembly Drawing and Parts List.....................................................A-26
DPU Dimensional Drawings.....................................................................A-31
Safety
Before installing this product, become familiar with the installation instruc-
tions presented in Section 3, Appendix A and all safety notes throughout.
Section 1—Introduction
General
The 581A Differential Pressure Switch actuates single or dual signal circuits
when predetermined limits of flow or level are exceeded. The DPU is con-
nected to the process or vessel by tubing or piping. Changes in differential
pressure (DP) at the DPU produce a mechanical output which moves a pointer
and switch actuation controls.
The 581A Switches (indicating and blind) are qualified per IEEE 323-1974,
IEEE 344-1975, IEEE 381-1977and NUREG 0588, Rev. 1 for either Func-
tional and/or Pressure Boundary Integrity service in limited harsh environ-
ment applications.
The switch case is drawn stainless steel with a sealed glass header for the
electrical interface with the switches. No additional conduit seals are required
to ensure the integrity of the instrument.
3
Section 1 Model 581A Differential Pressure Switch
Right Switch
Left Switch
Lock Screw
Lock Screw
Drive Arm
Pointer
Stop Bracket
Switch Adjust Linkage
Movement Assembly
Left Switch
Switch Adjust Linkage
Adjustment
Assembly*
Terminal Strip
Right Switch
Adjustment
Assembly*
Range Adjust
* Switch Adjustment Assembly includes crank, index shaft, and index pointer.
Actuating Cam
Left Switch Right Switch
Right Switch
Left Switch
Lock Screw
Lock Screw
Cam Actuator
Left Switch
Adjustment Switch Adjust Linkage
Assembly*
Terminal Strip
Right Switch
Adjustment
Assembly*
* Switch Adjustment Assembly includes crank, index shaft, and index pointer.
The cam rotates counterclockwise with increased pressure. In Figure 1.3, page
5, the LOW switch (typically set to alarm at decreasing pressure) is set at 25%
differential pressure while the HIGH switch (typically set to alarm at increas-
ing pressure) is set at 75% differential pressure.
4
Model 581A Differential Pressure Switch Section 1
C AM 0 100
B LAC K
W HI T E
GREEN
B LU E
RED
T ERM I NAL
B LO C K I N
C AS E
0 100
0 100
CLOSED OPEN
S W I T C H "A" S W I T C H "B " SWITCH CONTACTS CONTACTS
(LEF T ) (RI GHT )
A (LEFT) YELLOW/BLUE YELLOW/RED
B (RIGHT) GREEN/BLACK GREEN/WHITE
Notes:
1. CAM ROTATES COUNTER-CLOCKWISE WITH INCREASED DIFFERENTIAL PRESSURE
2. EACH SWITCH MAY BE SET AT ANY POINT FROM 10% TO 90% OF DP RANGE
3. WIRES: No. 22 AWG (Internal) and No. 18 AWG (External)
Wiring
Each switch is wired internally to a glass-sealed header using No. 22 AWG
ethylene tetrafluoroethylene (Tefzel®) insulated wire for radiation service per
IEEE Standard 393-1974. Externally, from the header, permanently color
coded No. 18 AWG Tefzel® wires exit through a female conduit connection.
5
Section 1 Model 581A Differential Pressure Switch
Specifications
General:
Actuating Unit (DPU) ���������������������� Model 199 DPU
Dial Size ����������������������������������������� 6 inches (150 mm)
Service Conditions (Normal) ���������� 40°F (4.4°C) to +180°F (+57.2°C) @ <90% R.H.
Switch Repeatability ����������������������� ±1% of calibrated span
Switch Set-point Range ������������������ 10 to 90% of calibrated span
Switch Deadband ��������������������������� ±10% of calibrated span (maximum)
Switch Type ������������������������������������ Mechanical, Snap-Acting; both switches are SPDT
Activation ���������������������������������������� Increasing or decreasing scale
Switch Contact Rating:
5.0 Amps @ 125 VAC
5.0 Amps @ 250 VAC
3.0 Amps @ 30 VDC (resistive)
1.0 Amp @ 30 VDC (inductive w/arc suppression)
0.4 Amp @ 125 VDC (resistive)
0.2 Amp @ 125 VDC (inductive w/arc suppression)
1
Additional 2% is applicable near the switch setpoints (±10% span).
Nuclear Qualifications
The following nuclear qualification applications are based on Cameron Engi-
neering Report 9A-CR3-580A-29.
Design Basis Event (DBE) Performance Error1 (% of calibrated span):
Radiation (1 x 107 Rads) ���������������� ±10%
Seismic (12.5 Gs):
During Event ����������������������������������� ±10%
Following Event ������������������������������ ±5%
Harsh Environment (≤ 200°F):
During Event ����������������������������������� ±10%
Following Event ������������������������������ ±10%
1
Includes the cumulative effect of accelerated aging.
6
Model 581A Differential Pressure Switch Section 1
Application Specifications
Model 581A switches were subjected to IEEE 323-1974/344-1975 qualifi-
cation testing, which found the devices suitable for functional service in a
limited harsh environment. The service conditions associated with the limited
qualifications are:
• 22 year qualified service life1 normal conditions of 104°F and 0 psig
• Radiation exposure to 1 x 107 Rads (TID)
• 85 days @maximum DBE conditions of 200°F, 10 psig ambient pressure
and ≥ 95% R.H.
• Seismic qualifications:
–– OBE @ 6.25 Gs (series of 5)
–– SSE @ 12.5 Gs
–– 5% critical damping
–– No resonance in frequencies below 33 Hz
• Mechanical aging included:
–– 25,000 pressure cycles during accelerated aging
–– 12,500 electrical cycles during accelerated aging
–– Vibration cycling using 0.2 G sinusoidal sweeps over the 1 to 200
Hz range at 1.0 octave/min.
1
Qualified Life: This equipment contains some organic material. The qualified life will be
governed by factors such as environmental temperature, radiation exposure, and activa-
tion energy.
Storage Specifications
Storage per ANSI N45.2.2-1978 Level B at 70°F (21°C) ±20°F (±11°C) in
factory sealed package for 2.5 years maximum will not affect installed service
life.
7
Section 1 Model 581A Differential Pressure Switch
8
Model 581A Differential Pressure Switch Section 2
Section 2—Installation
The instrument should be inspected at time of unpacking to detect any dam-
age that may have occurred during shipment.
IMPORTANT: The DPU was checked for accuracy at the factory. Do not change any of
the settings during examination or accuracy could be affected.
Mounting/Piping/DPU Installation
Mount the instrument with four 5/16" (8mm) bolts, Grade 5 or greater,
torqued to 17 ft/lbs (see Figure 7.1). Mounting structures should be designed
to avoid resonance and/or minimize resulting amplification below 33 Hz.
Response spectra at mounting surface of the instrument shall not exceed those
in the 9A-CR3-580A-29 Qualification Report. Interfacing process tubing and
conduit shall be supported by the same mounting as the DP instrument base
in order to minimize relative motion of the DP instrument and connections.
See Appendix A for Model 199 DPU installation and maintenance informa-
tion.
Electrical Connection
Units are supplied with dual alarm switches. The direct-set switch contacts
are adjustable over 10% to 90% of the scale range.
The high switch and low switch set point adjustment procedures are covered
in Changing Switch Set Point, page 20.
For physical location of switches, see Figures 1.1 and 1.2, page 4.
Switch Use
Switch contact life is influenced by various application conditions such as
temperature, humidity, airborne contamination, vibration, amount of plunger
travel, cycling rate, and rate of plunger travel (and others), as well as by the
electrical (circuit) characteristics.
IMPORTANT: Arc suppression for inductive loads will prolong the life of the switch
contacts.
9
Section 2 Model 581A Differential Pressure Switch
Startup
For startup procedures, warnings, and other information, refer to Appendix A.
Switch Wiring
IMPORTANT: Figure 2.1 shows switch contacts in the relaxed (shelf) condition, the low
switch set to trip at a position below the pointer scale position, and the
high switch set to trip at a position above the pointer scale position. NO
= Normally Open in (shelf) condition. NC = Normally Closed in (shelf)
condition. C= Common.
NO C NC NC C NO
YELLOW
GREEN
BLACK
WHITE
BLUE
RED
1 2 3 4 5 6 7 8 9 10
YELLOW
GREEN
BLACK
WHITE
BLUE
RED
10
Model 581A Differential Pressure Switch Section 3
11
Section 3 Model 581A Differential Pressure Switch
Tools
The following tools are recommended for general maintenance of the Model
581A differential pressure switch.
Table 3.1—Tools
Equipment Purpose
Pointer Puller Pointer removal
Small Screwdriver Calibration adjustment
Medium Screwdriver Bezel removal and replacement
1/4" and 1/8" Open-end Wrenches Zero (1/4") and Range (1/8") adjustments
1/8 Hex Allen Wrench Switch set point adjustment
To reinstall the bezel/lens assembly (581A-0 only), install the O-ring in the
groove of the case, assemble the bezel, lens, and lens gasket, and realign the
screw holes with the holes in the case. Then, reinstall the screws in the bezel
to secure all components to the case. Torque the screws per Table 4.3, page 30.
To reinstall the cover (581A-2), install the O-ring in the groove of the case
and realign the screw holes in the cover with the holes in the case. Reinstall
the screws in the cover to secure it to the case. Torque the screws per Table
4.3, page 30.
IMPORTANT: Ensure the bezel gasket is properly oriented before placing the instru-
ment back in service. Incorrect gasket orientation will cause the instru-
ment indicator to jam, resulting in inaccurate readings.
Calibration Check
No. 9A-0288-1032B), hold the hexagon pointer hub fixed and rotate
the pointer with fingers until the pointer indicates zero on the scale.
See Figure 3.2, page 13.
c. Replace the bezel/lens assembly.
3. To test for reverse travel, connect the pressure source to the low-pressure
(LP) housing and vent the HP housing. Apply pressure to approximately
150% of the DP range. The pointer should move approximately 5% to
10% below zero.
4. To test for overtravel, connect the pressure source to the HP housing and
vent the LP housing. Apply pressure to approximately 150% of DP range.
The pointer should move approximately 5% to 10% above the full scale
mark.
5. Apply 0, 50, and 100% of full scale pressure. If the indication is within
specified limits, no adjustments are necessary. If the indication is not
within specified limits, perform a complete calibration (see Scale Plate
Installation and Removal, page 14).
6. Make sure the instrument zero indication is correct; otherwise, repeat
step 2.
7. Verify the switch set points (refer to Changing Switch Set Point, page 20).
Pointer Installation
1. Position the pointer on the movement shaft with the pointer set at zero
scale. Obtain 75% to 90% pointer hub engagement on the movement
pinion shaft. The pinion shaft must not extend through the pointer hub.It
may be necessary to enlarge the hub hole using a tapered broach (includ-
ed in the calibration toolkit, Part No. 9A-0288-1032B).
13
Section 3 Model 581A Differential Pressure Switch
2. Lightly tap the pointer hub with a hand-set or other flat-end tool. Use
perpendicular blows to avoid bending the shaft.
3. Check the calibration of the indicating switch over its entire range (refer
to Scale Plate Installation and Removal, page 14). If the indicating switch
is correctly calibrated, secure the pointer to the movement shaft by tap-
ping the hub with a hand-set or other flat-end tool.
4. Test the pointer for tightness by moving the pointer from the zero to the
50% position manually, and then letting the pointer return freely to zero.
If the pointer indicates a shift, tap the pointer hub to tighten it to the
shaft.
Pointer Removal
The pointer is removed with a pointer puller shown in Figure 3.3. This tool is
included in the calibration toolkit, Part No. 9A-0288-1032B.
HUB
POINTER
HUB
MOVEMENT SHAFT
To remove the scale plate, move the pointer to the 12 o'clock position, remove
the screws from the scale plate, and gently guide the scale plate halves from
beneath the bottom mounting tab. Screws are located at 12, 3 and 9 o'clock
positions and near the center of the plate which overlays a narrow bracket.
To reinstall the scale plate, repeat the installation steps in reverse order, as
follows. Holding the pointer in the 12 o'clock position, place the scale plate
halves against the top mounting tab and guide the bottom edges behind the
bottom tab. Replace the screws in the top tab and in the center of the scale
plate. No screw is used with the bottom tab.
Indicator Calibration
If the DPU assembly is replaced on a Model 581A differential pressure switch
with indicator, the indicator must be recalibrated. This procedure does not ap-
ply to the 581A-2 blind switch without indicator.
4. Check the pointer for zero indication. If necessary, set the pointer to zero
by slipping the pointer on the hub, per Calibration Check, step 2, page 12.
5. Apply 100% pressure. If the pointer exceeds full-scale, adjust the range
adjust screw with a 1/8-in. open-end wrench to lengthen the movement
range arm.
15
Section 3 Model 581A Differential Pressure Switch
6. Release pressure. Set the pointer to zero, by slipping the pointer on the
hub, per Calibration Check, step 2, page 12.
7. Repeat steps 4 and 5, as necessary, to obtain the correct zero/full-scale.
8. Apply 50% DP. If the pointer indicates 50%, proceed to step 9. If the
pointer does not indicate 50% scale, adjust linearity as follows.
a. Loosen the drive arm screw (Figure 3.5) and move the arm to shift
the pointer in the direction of the error a distance of about 10 times
the linearity error.
b. To prevent interference, check to see that the drive arm clears the
end of torque-tube housing by approximately 0.030 in. before re-
tightening the drive arm screw.
c. While supporting the block/shaft, tighten the clamp screw until it is
snug to the shaft.
d. Still supporting the block/shaft, tighten the clamp screw an addition-
al 1/3 to 1/2 turn. (This screw can normally turn one full revolution
before breaking.)
e. Perform the Drive Arm Tightness Test described on page 17.
LINK SINTERED
CLAMP BLOCK
.03 (approx)
DRIVE ARM
M199
1/8” WRENCH
(PN 9A-0163-0044C)
TO TIGHTEN
9. Release pressure and reset the pointer to zero. Check the span. If the gear
in the movement reaches a limit of travel as a result of linearity adjust-
ment (step 8), slip the range arm along the gear approximately 5 degrees
from the normal 37.5 degree angle to approximately 43 degrees (see
step 3). To slip the range arm, apply pressure to the range arm with the
thumb, while holding the gear firmly in place. Retest the pointer response
at 50%, 0%, and 100% of full-scale differential pressure, and adjust the
linkage until the readings are acceptable.
10. Apply 0%, 25%, 50%, 75%, 100%, 75%, 50%, 25%, and 0% of full-scale
differential pressure consecutively to the instrument without overshoot.
16
Model 581A Differential Pressure Switch Section 3
Lightly tap the indicator to overcome friction. The pointer should accu-
rately indicate each applied pressure.
11. Test the instrument repeatability by applying 0%, 50%, 0%, 50% of full-
scale differential pressure. The indicator should accurately indicate each
applied pressure.
12. Set the drive arm stop to prevent the pointer from striking the post lock
on the scale.
13. Test the pointer for tightness per the pointer installation instructions on
page 13.
14. If the drive arm screw has been loosened, check for tightness by perform-
ing the Drive Arm Tightness Test described below.
15. Reassemble the unit per the cover installation instructions on page 12.
Drive Arm Tightness Test
This procedure tests the tightness of the drive arm's connection to the torque
tube by applying torque developed by the DPU onto a fixed drive arm. Ap-
ply pressure slowly, as torque is being applied to the connection through the
torque tube drive shaft.
1. With the pointer at normal 0% torque tube rotation position (0% on a
normal 0 to 100% scale unit), adjust the drive arm stop bracket (or use
alternate means) to prevent the pointer from moving (the stop bracket
interferes with drive arm movement).
2. Pressurize the DPU to full scale DP (100% of full scale range) to apply 8
degrees of equivalent torque to the connection.
3. Observe the shift in the "zero" indication following DPU depressuriza-
tion (as required) and drive arm stop bracket readjusting (to allow free
movement of drive arm and pointer). A downscale (counter-clockwise)
shift in "zero" of greater than 1/2% indicates drive arm slippage, neces-
sitating further clamp block tightening.
Drive Arm Stop Adjustment
1. Apply sufficient pressure to the high-pressure housing to deflect the
pointer against the full range post lock on the scale plate.
2. Slide the upper drive arm stop bracket against the drive arm and tighten
the drive arm stop bracket screw.
3. Apply sufficient pressure to the low pressure housing to deflect the
pointer against the zero stop post lock on the scale plate.
17
Section 3 Model 581A Differential Pressure Switch
4. Bend the zero drive arm stop against the drive arm.
5. Verify calibration as applicable.
Switch Calibration
Before performing a complete calibration of the switch, perform a calibration
check (See Calibration Check, page 12). If the instrument is out of calibration,
remove the bezel/lens and scale plate and inspect the switch mechanism (Fig-
ure 3.6) to verify the following:
• The roller rotates without wobble or binding.
• The cam does not touch the roller side shields.
• The actuator arm moves freely on its pivot.
• All switch mounting screws are tight.
• Linkages are straight and do not bind at the pivots.
Correct any problems that are observed. If no problems are observed, proceed
with a complete calibration.
SWITCH
CAM ROLLER
ACTUATOR ARM
STOP
SCREW
CAM
PLUNGER
SCREW
Figure 3.6—Switch mechanism
Calibration Setup
1. Connect an output device (lamp or buzzer) to the switch output leads.
2. Connect a test voltage to the switch input terminals on the terminal strip.
(A low voltage is recommended for safety.)
3. Unlock the switch plate and move the plate until the roller is positioned
at the top of the cam.
4. Advance the plunger screw until the switch actuates, then advance the
plunger screw an additional 60° (one flat).
5. Exercise the switch roller across the top of the cam to verify steady
operation. Advance the stop screw to touch the switch, then back out the
screw 1.5 turns (9 flats).
18
Model 581A Differential Pressure Switch Section 3
To calibrate the switch linkage, perform the following steps. Refer to Figure
3.7 as needed.
1. Loosen the three linkage screws and turn the crank to the 12 o’clock
position.
2. Use a 1/8-inch Allen wrench to hold the index shaft and slip the index
pointer to 0 on the switch index. Tighten the screw on the crank to the
mid-slot position to set the crank radius.
3. Turn the index pointer to index number "1" on the outer edge of the
switch index.
4. Apply 10% differential pressure and adjust the switch plate until the
switch actuates. Lock the two linkage screws.
5. Rotate the index pointer to index number “9.” Apply 90% differential
pressure and adjust the crank radius until the switch actuates.
6. Recheck 10% and 90% set points. Adjust the crank radius and the index
pointer until both set points are accurate within a 2% tolerance (nominal).
19
Section 3 Model 581A Differential Pressure Switch
IMPORTANT: The high switch is usually set to actuate at increasing pressure. When
calibrating the high switch, apply test pressure in an increasing direction.
This amount of loading will prevent cam-runout of a similar condition.
IMPORTANT: Excessive plunger loading (more than 3 flats) may cause the roller to
drag on the cam. Cam friction will be apparent by excessive hysteresis,
erratic pointer readings, and inconsistent switch response.
Definitions of Terms
Set Point. The measured pressure at which the snap-switch actuates and
thereby changes the states of the N.O. and N.C. contacts. For example, the set
point of the low switch is 24 psid with decreasing pressure.
remove the scale plate and inspect the switch and the switch mechanism.
The scale plate is split to allow for removal without pulling the pointer.
Changing Set Point of an In-Service Instrument (Not Recommended for
Nuclear Qualified Units)
Use the following procedure to set a set point for an instrument that is in ser-
vice, when calibration pressures cannot be applied. See Figure 3.8 for specific
locations denoted by items A, B, C and D.
1. Remove the bezel/lens assembly or cover. Do not remove the pointer or
scale plate.
2. Insert hex wrench in the switch adjust post.
3. Loosen the switch lock-screw (item B) 1/2 to 1 turn.
4. With the hex wrench, move the index pointer (item C) to the new set
point as indicated on the switch index (item D).
5. If possible, check the set point by varying the process pressures and ob-
serving the pointer readings when the switch actuates. (Open the mani-
fold bypass valve slowly and watch for “pointer-jump” at the set point or
by electrical signal.) Adjust the setting, if necessary, and repeat the test
several times to verify stability.
IMPORTANT: The switch index has 10 divisions, marked 0, 5, and 10. These match the
markings on the outer edge of the scale.
21
Section 3 Model 581A Differential Pressure Switch
Example: Scale has range of 0-60 psid. Set point is 24 psid, with decreasing
pressure, (24/60 x 10 = 4).
1. Move the index pointer (item C) for the low switch from division 0 to
division 4 tick mark.
2. Tighten the switch lock (Item B) to a snug position, then turn an ad-
ditional 1/4 turn to place the set point within ±2% of full scale. Do not
overtighten.
Repeat steps 1 and 2 for the high switch, moving the index pointer from divi-
sion 10 to division 4.
Range Changes
The range of a Model 581A differential pressure switch can typically be
changed by replacing the range spring assembly inside the DPU. In some
cases, the replacement of the BUA may be required. See Changing the DPU
22
Model 581A Differential Pressure Switch Section 3
Range, page A-18, and Bellows Unit Assembly (BUA) Replacement, page A-22, for
details or contact the factory for assistance.
Troubleshooting
For switch troubleshooting tips, see Table 3.2 below. For information related
to the DPU, see Appendix A.
Table 3.2—Troubleshooting
Problem Possible Probable Cause Corrective Action
Source
Low or No Indica- Loose linkage or movement Tighten or replace
Indication tor, Alarm
Out of calibration Calibrate
Switch
Mechanism Pointer loose Tighten pointer
Dirty or corroded mechanism Clean or replace
Wiring interfering with move- Re-route wiring
ment
Dirty mechanism Clean mechanism
High Indicator, Loose linkage or movement Tighten or replace
Indication Alarm-
Switch Out of calibration Calibrate
Mechanism
Erratic Indica- Linkage dragging or dirty Adjust or clean
Indication tor, Alarm
Switch Pointer dragging on scale plate Adjust pointer position
Mechanism
Inaccurate Power Blown fuse Replace fuse
or No Supply
Broken or loose wire Repair
Electrical
Alarm Alarm Switch not properly adjusted Adjust switch
Switch
Dirty or burned contacts Replace switch
23
Section 3 Model 581A Differential Pressure Switch
Table 3.2—Troubleshooting
Problem Possible Probable Cause Corrective Action
Source
Switch Process Transients or surges cause Add time delay gages or add time
Drifts Changes switches to actuate prema- circuit
(set point turely
not repeat-
Set point and/or deadband are Specify DP range as low as
able)
too wide in pressure values practical; set point repeatability
and deadband are percentage of
full range
Electrical overloads affect the Examine circuits for voltage,
spring properties of the leaf amperes
actuator in the switch
DC inductive loads cause Consider arc-suppression devices
arcing and burning of contacts or relays
Accumulation of fluids in piping Vent gas or drain liquids from
generate artificial signal signal lines
Calibration Failure to check set point after Verify set point repeatability after
Techniques locking locking switch plate
Rapid pressure change or vent- During calibration, make pressure
ing system changes in slow, discrete steps
Pressure application in reverse Test low-alarm with decreasing
pressure and high-alarm with
increasing pressure
Reference gage inaccuracy Suitable pressure standard such
as manometer, dead-weight
tester, or Heise-type gage may be
required
Damage to switch contacts Adjust plunger screws carefully
to avoid damage to internal parts
of switch
24
Model 581A Differential Pressure Switch Section 4
Table 4.1 contains components for Model 581A-0 indicating switches and
Model 581A-2 blind switches with internal indicator. For parts list informa-
tion on Model 581A-2 blind switches without indicator, see Table 4.2, page A-16.
26
Model 581A Differential Pressure Switch Section 4
27
Section 4 Model 581A Differential Pressure Switch
Table 4.2 contains components for Model 581A-2 blind switches without
indicator. For parts list information on Model 581A-0 indicating switches and
Model 581A-2 blind switches with internal indicator, see Table 4.1, page 26.
29
Section 4 Model 581A Differential Pressure Switch
Table 4.3 presents torque requirements for the switch assembly fasteners.
30
Model 581A Differential Pressure Switch Section 4
31
Section 4 Model 581A Differential Pressure Switch
32
Model 581A Differential Pressure Switch Section 5
Dimension A
Model
(in., mm)
581A-0 11 7/16 (290.5)
581A-2 10 7/8 (276.2)
33
Section 5 Model 581A Differential Pressure Switch
34
Model 581A Differential Pressure Switch Appendix A
General
The Barton Model 199 Differential Pressure Unit (DPU) (see Figure A.1) is
a mechanical device which accurately measures differential pressure relative
to a gas or liquid flowing through a process system, or to the level of a liquid
contained in a process vessel.
Product Description
The Model 199 DPU is a dual bellows assembly enclosed within pressure
housings. The dual bellows assembly (Figure. A.2) consists of two opposing
internally connected liquid-filled bellows, a center plate, range springs, over-
range valves, and a torque tube assembly.
The pressure housings are connected by pipe or tubing to the primary device
located in the system piping. Variations in differential pressure within the
pressure housings cause the bellows to expand or contract in a linear direction
towards the side having the lowest pressure.
The linear movement of the bellows is converted into angular rotation when
transmitted to the torque tube shaft by the drive arm and this mechanical
motion actuates the mechanism of the process monitoring instrument. The
process monitoring instrument that is connected to the torque tube assem-
bly may be an indicator, a switch, a transmitter, a recorder, or other process
control device.
A-1
Appendix A Model 581A Differential Pressure Switch
Specifications
Housing Material ���������������������������� Forged 316 Stainless Steel
Torque Tube Rotation (full scale DP) 8° ±10%
Torque Tube Material ���������������������� Beryllium Copper (BeCu)
Temperature Limits ������������������������� -40°F (-40°C) to +180°F (+82°C)
Maximum Non-linearity: ������������������ 0-10" w.c. to 0-400" w.c.
(0-25 mbar to 0-993 mbar)
±0.5% of full scale with appropriate linkage
������������������������������������������������������� 0-401" w.c. to 0-100 psi
(0-996 mbar to 0-6.9 bar)
±0.75% of full scale with appropriate linkage
Repeatability ����������������������������������� 0.20% of full scale DP
Net Volume, L.P. Head 35 cu. in. (575 cc) 30 cu. in. (490 cc)
(cu. in.)
Net Volume, H.P. Head 31 cu. in. (510 cc) 26 cu. in. (425 cc)
(cu. in.)
Displacement (cu. in.) 0.5 cu. in. (8.2 cc) 1.5 cu. in. (25 cc)
for full-scale travel
NOTES: Zero center or split ranges available on special order (e.g., 0-50" w.c. (0-124
mbar) range may be ordered 25-0-25" w.c. (62-0-62 mbar) or 10-0-40" w.c. (25-0-99 mbar).
Intermediate DP ranges available from 0-20" w.c. to 0-100 psi (0-50 mbar ot 0-6.9 bar).
Standard pressure connections are 1/2" (top) and 1/4" (bottom) NPT. Range springs are not
interchangeable between the different size bellows. Metric conversions are approximate.
Dimensional drawings available on request.
A-2
Model 581A Differential Pressure Switch Appendix A
Theory of Operation
Basic Components
Torque Tube Center Plate
Shaft Low Pressure Bellows
Torque Tube
Range Spring
High Pressure
Bellows
High Pressure
Overrange Low Pressure
Valve Housing
Temperature
Compensator
Pulsation Dampener
Damping Valve Plug
Valve Stem
Pressure Housings
The two pressure housings of the Model 199 DPU are available in the various
safe working pressure ratings defined in Table A.1, page A-2, and in the dimen-
sional drawings in Section 6.
Each pressure housing has two tapped connection ports: one port is located in
the top of the housing, the other port is located in the bottom of the housing.
Bellows
The bellows of the Model 199 DPU are available in various materials and
sizes (refer to Table A.1, page A-2) to accommodate various calibration ranges.
The DPU has two bellows. One end of each bellows is sealed. The open end
of each bellows is attached and sealed to a side of the center plate (one bel-
lows on each side) as shown in Figure A.2.
A-3
Appendix A Model 581A Differential Pressure Switch
The bellows and center plate are filled with fill liquid via the drive arm hole
plug. An opening through the center plate provides a passageway for the
transfer of fill liquid between the two bellows. This opening also allows the
bellows to be connected internally by a valve stem.
Drive Arm
Hole Plug
(DO NOT LOOSEN)
Retainer
Torque Tube Gland Nut Screws
Range
(DO NOT LOOSEN)
Spring
Assembly
Push
Rod
Torque
Tube
Shaft Lock Nut
Range
Springs
2-1/8" Bellows Unit Assembly (BUA)
Drive Arm
Hole Plug
(DO NOT LOOSEN)
Retainer Nut
Range Springs
The range of the dual-bellows type DPU is determined by the force required
to move the bellows through their normal range of travel.
The range springs, which are available in various ranges (see Table A.1, page
A-2), act with the bellows and torque tube to balance the differential pressure
A-4
Model 581A Differential Pressure Switch Appendix A
applied to the unit. The number of springs used and their spring rate depends
on the individual differential pressure range requirement.
Torque Tube Assembly
As illustrated in Figure A.4, page A-5, the torque tube assembly consists of a
torque tube, a torque tube shaft, and the supporting members. The outboard
end of the torque tube shaft is attached to the center plate. The torque tube
shaft, located in the center of the torque tube, is welded to the inboard end of
the tube.
Needle Beaing
Center Plate
O-Ring Seal
Weld
Torque Tube
Lock Nut
Torque Tube
Torque Tube
Shaft
8° Drive Arm
Rotation
Valve Stem
Disc
Weld 0.200"
Bellows Travel
Jewel Bearing
Minature
Precision Valve Stem
Ball Bearing
Movement of the bellows is transmitted by the drive arm to the torque tube as
a rotary motion. Since the torque tube is attached to the center plate, the tube
must twist when subjected to torque. The torque tube shaft, which is freely
supported within the torque tube at its outer end, but connected to the torque
tube and drive arm at its inner end, rotates through the same angle as the dif-
ferential pressure unit.
Pulsation Dampener
The pulsation dampener (Figure A.2, page A-3) controls the flow of fill-liquid
between the high and low-pressure bellows with an externally adjustable pul-
sation dampener needle valve. Restriction of liquid flow reduces the effects of
pulsation. In applications where pulsation is not a problem, the needle valve is
set to the full-open position.
A-5
Appendix A Model 581A Differential Pressure Switch
Installation/Operation
Unpacking
Inspect the instrument after unpacking it to detect any damage that may have
occurred during shipment.
IMPORTANT: The DPU was checked for accuracy at the factory. Do not change any of
the settings during examination or accuracy could be affected.
IMPORTANT: Do not locate the instrument near vents or bleed holes that discharge
corrosive vapors or gases.
Mounting
See Mounting/Piping/DPU Installation, page 9.
Piping—Standard Practices
When installing piping, observe the following standard practices and refer to
the piping diagrams on pages A-8 through A-13.
1. Shorten the distance between the primary device and the DPU as much
as possible. Distances exceeding 100 feet are not recommended. For dis-
tances up to 50 feet, use 1/4-inch or 3/8-inch pipe or tubing. For runs of
50 to 100 feet, use 1/2-inch pipe or tubing. The recommended limitation
does not apply if an air purge or blow-back system is used.
2. Slope all piping at least one inch per linear foot to avoid liquid or gas
entrapment.
3. Provide 2 feet of non-insulated piping between the DPU and the primary
device for each 100°F (+37.8°C) in excess of +200°F (+93.3°C).
4. Assure that the temperature of the DPU never exceeds 180°F (+82°C).
When steam tracing is necessary, the steam pressure should not exceed
5 pounds per square inch and insulation should not be used. If pressure
must exceed 5 pounds per square inch, limit the length of tubing around
the DPU to two turns and do not insulate.
5. Install a suitable pulsation dampening device upstream of the DPU.
Where severe pulsation is present, the accuracy of the flow measurement
will be affected.
A-6
Model 581A Differential Pressure Switch Appendix A
Assure that the DPU high-pressure housing is connected to the upstream tap
of the primary device.
IMPORTANT: To prevent overheating the DPU during blowdown, monitor the tempera-
ture by placing hands on the pipe between the DPU and the manifold
pipe containing the vent valves.
The process media may be used as a reference leg seal fluid if it will condense
in the reference leg under all conditions.
If the process or process media characteristics are such that the above condi-
tions cannot be met, a special reference leg seal fluid is required. The special
seal fluid must not be volatile and must not be miscible with the process
media. Also, the difference in the densities of the special seal fluid and the
process media will require compensation in calculating the differential pres-
sure range of the DPU.
A-7
Appendix A Model 581A Differential Pressure Switch
Piping Diagrams
Diagrams for typical and special flow applications are presented on the fol-
lowing pages. Use the diagram best suited for the application as a guide for
piping configuration.
BYPASS
DPU VALVES
SHUT-OFF
STATIC PRESSURE VALVES
CONNECTION
(USE ONLY ONE)
VENT VALVE
BLOCK VALVES
PRIMARY DEVICE
BLOCK VENT
VALVES VALVE
PRIMARY
DEVICE SHUT-OFF
VALVES
BYPASS
VALVES
STATIC PRESSURE
CONNECTION DPU
(USE ONLY ONE)
DRIP POTS
A-8
Model 581A Differential Pressure Switch Appendix A
DPU
SHUT-OFF
VALVES
BYPASS
VALVE
PRIMARY
DEVICE
BLOCK VALVES
PRIMARY
DEVICE
PLUGS
BLOCK
VALVES
CONDENSING
RESERVOIR
SHUT-OFF
VALVES
VENT DPU
VALVES
(OPTIONAL)
VENT BYPASS
VALVE VALVES
PRIMARY DEVICE
PRIMARY
DEVICE FILL TEE AND
VENT VALVE
FOR HOT OR
GASSY LIQUIDS
BLOCK
VALVES
SHUT-OFF
VALVES
BYPASS
VALVE
DPU
BYPASS
VALVE
SEAL SHUT-OFF
POTS VALVES
DPU
PRIMARY
DEVICE
BLOCK
VALVES
BLOCK
VALVE
SHUT-OFF
VALVE
VENT
VALVE
BLOCK DPU
VALVE
HIGH-PRESSURE
SHUT-OFF DRAIN
VALVE VALVE
A-10
Model 581A Differential Pressure Switch Appendix A
BLOCK
VALVE
SHUT-OFF
VALVE
BLOCK
VALVE
DPU
SHUT-OFF
VALVE DRAIN
VALVE
HIGH-PRESSURE
Figure A.13—Cool Non-Condensing Liquid, DPU Below Tank
2" CROSS
BLOCK REFERENCE
VALVE LEG
VENT
VALVE SHUT-OFF
VALVE
DPU
BLOCK
VALVE
HIGH-PRESSURE
BYPASS
VALVE
Figure A.14—DPU Bottom Tank with Reference Leg
A-11
Appendix A Model 581A Differential Pressure Switch
BLOCK VALVE
INVERTED "U"
GAS TRAP
BLOCK
VALVE
MINIMUM LENGTH
DRAIN 6 INCHES
SHUT-OFF VALVE
VALVE
VAPOR GENERATOR
DPU
LOW-PRESSURE
INVERTED "U"
GAS TRAP
VAPOR GENERATOR
FLAT TUBULAR SPIRAL
IN OUTER 1/3 OF TANK
INSULATION KEEP
SHUT-OFF DEAD LEG BETWEEN
BLOCK TANK AND SPIRAL
VALVE VALVE 5 INCHES OR LESS
HIGH-PRESSURE
DPU
LOW-PRESSURE
A-12
Model 581A Differential Pressure Switch Appendix A
SIGHT FLOW
GAGE
BLOCK VALVES
INPUT GAS
REGULATOR
GAS
SUPPLY
SHUT-OFF VALVES
BYPASS VALVE
DPU
HIGH PRESSURE
GAS
SUPPLY SHUT-OFF VALVES
BYPASS VALVE
DPU
HIGH-PRESSURE
A-13
Appendix A Model 581A Differential Pressure Switch
IMPORTANT: For gas service, perform a zero check with both block valves closed. If
the gas flow is pulsating, a standing wave effect may occur in the pro-
cess line which can displace the indicator and appear as a zero error.
a. Open the bypass valve(s), then open one shutoff valve. This proce-
dure equalizes the pressure between both sides of the instrument.
The instrument should indicate zero.
b. If the instrument does not indicate zero, check for gas or liquid
entrapment in the DPU piping or in the DPU (depending on the
orientation of the piping layout and service).
c. If necessary, adjust the pen or pointer by turning the zero adjust knob
on the instrument until the pointer indicates zero.
3. Check the manifold and piping for leaks as follows.
a. Open the bypass valve(s), then open one shutoff valve to pressurize
the instrument.
b. Close the shutoff valve and the bypass valve.
c. Observe the instrument for pen or pointer movement, up or down the
scale, which could indicate leakage.
IMPORTANT: Do not subject the DPU to unnecessary shock or overrange pressure
during operations.
Table A.2—Tools
Description Purpose
7/16" Hex wrench Housing bolts
5/8" Hex wrench Housing bolts
1/8" Allen wrench Dampener Adjustment
1/2" Open ended wrench Push rod - bellows cup connection
Modified box/open-ended wrench (7/16") Kickoff spring lock nut
100 ft-lb Torque wrench Housing bolts
A-14
Model 581A Differential Pressure Switch Appendix A
Torque (lb/ft)
Rotation
(Note 1)
(Note 2)
(Note 3)
Housing Bolt
Lube
SWP Size Qty Part No.
Mat'l Mat'l
Calibration Setup
The test equipment required to calibrate the Model 199 DPU, when attached
to an indicator or switch is shown in Figure A.19, page A-17.
All Barton instruments using the Model 199 DPU as actuator are calibrated at
the factory.
All DPUs sold as replacement units or as actuators for other applications are
checked for accuracy prior to shipment.
A-16
Model 581A Differential Pressure Switch Appendix A
A-17
Appendix A Model 581A Differential Pressure Switch
IMPORTANT: The calibration range of the DPU was carefully set at the factory. The
following procedure maintains the original factory bellows travel setting
while the range spring change is performed. This procedure must be
strictly followed while changing the range spring assembly or altering the
adjustment.
To replace the range spring assembly with either a new assembly of the same
range or one with a different range value, proceed as follows (this procedure
assumes that the DPU is either connected to an indicating instrument or fitted
with a scale and pointer to measure torque tube shaft movement):
IMPORTANT: The pointer will shift from zero. This is a normal action and the pointer
should not be readjusted at this point.
4. Install the new range spring assembly onto the push rod, and replace the
spring retainer screw.
5. If the pointer is set above zero, rotate the spring adjustment clockwise
until the pointer is set at zero. If the pointer is below zero, rotate the
spring adjustment counterclockwise until the pointer is set at zero.
6. Replace and tighten the lock nut. If the pointer shifts from zero, loosen
the lock nut and reset the pointer as in step 5. Tighten the lock nut.
A-18
Model 581A Differential Pressure Switch Appendix A
7. Replace the low-pressure housing and install NEW bolts (new gaskets
are also recommended). Torque the bolts in accordance with Table A.3,
page A-16.
IMPORTANT: The pointer will shift from zero. This is a normal action and the pointer
should not be readjusted at this point.
4. Thread the new assembly onto the push rod, and align the holes in the
range spring assembly with the spring posts. Replace and tighten the
retainer nuts.
5. Using a spanner wrench, rotate the spring adjustment until the pointer
is set at exact zero. Replace and tighten the lock nut. The pointer must
remain at the zero setting. If the pointer shifts from zero, loosen the lock
nut and repeat this step.
6. Replace the low-pressure housing and install NEW bolts (new gaskets
are also recommended). Torque the bolts in accordance with Table A.3,
page A-16.
A-19
Appendix A Model 581A Differential Pressure Switch
IMPORTANT: The lock nut requires a modified 7/16", 12-point box wrench for tighten-
ing (Figure A.21).
1. Remove the instrument from service. Set the pointer at zero using the
instrument zero adjustment.
2. Remove the pressure housing bolts and remove the low-pressure housing.
3. With the pointer set at zero, remove the lock nut, spring retainer nuts,
washer and lock nut, and range spring assembly.
IMPORTANT: The pointer will shift from zero. This is a normal action and the pointer
should not be readjusted at this time.
A-20
Model 581A Differential Pressure Switch Appendix A
IMPORTANT: The range spring assembly consists of the range springs, kickoff spring
assembly, and the push rod. The assembly is furnished as a complete
and assembled unit.
5. Position range spring assembly over retainer posts and replace retainer
nuts.
6. Insert the tubing between the range springs and tighten inboard lock nut.
7. Zero the pointer by adjusting the spud on the push rod. Be sure that the
kickoff spring is not engaged during this operation.
8. Position the kickoff spring to prevent interference with the coil springs
and tighten the kickoff spring lock nut. If clearance is needed for the
wrench, note the position of the spud (measure with a scale or count
threads to the end of the push rod) and move the spud for the required
clearance. After the lock nut is tightened, return the spud to its original
position.
9. Check the back-clearance between the kickoff spring and the slot. Use a
wire feeler gage or comparable measuring device. The clearance between
slots will vary, but must be at least 0.005-in. to prevent interference with
instrument zero. Add or remove washers as required to obtain the proper
clearance. Tighten the lock nut securely.
10. Install the jam nut and lock the spud to the push rod. Hold with pliers.
If the spud was moved in step 8, return the spud to the proper position
before tightening the jam nut.
11. Apply 100 percent negative pressure to the high-pressure side of the DPU
and repeat the clearance check and adjustment described in step 9.
12. Apply 100 percent positive pressure to the high-pressure side of the DPU
and repeat the clearance check and adjustment described in step 9.
13. Release all pressure from the instrument, replace the low-pressure hous-
ing and install NEW bolts (new gaskets are also recommended). Torque
the bolts in accordance with Table A.3, page A-16.
IMPORTANT: Component Removal—See the manual for the actuated instrument for
particular components that must be removed to gain access to DPU case
mounting screws. Support DPU during disassembly. Do not loosen the
drive arm hold plug (located in the top of the BUA center plate) when
removing the mounting bracket. If the plug is loosened, bellows fill fluid
will be lost.
1. Disconnect the DPU piping and remove the instrument from service.
2. Loosen the actuated instrument drive arm and slide off of the torque tube
shaft. Do not disconnect the instrument linkage.
3. Remove the case-retaining screws and remove the case assembly.
4. Remove the indicator bracket screws and the indicator bracket.
5. Remove the pressure housings bolts and pressure housings.
6. Place the pressure housings on the replacement bellows unit assembly
and insert and start (only) the pressure housing bolts.
7. Attach the indicator bracket to the replacement assembly, using the indi-
cator bracket screws.
8. Attach the mounting bracket using the screws provided.
A-22
Model 581A Differential Pressure Switch Appendix A
9. Support the DPU and tighten and torque the pressure housing bolts in
accordance with Table A.3, page A-16.
10. Attach the case assembly using case retaining screws.
11. Slip the drive arm over the torque tube shaft as shown in Figure A.22,
page A-23; clear the end of the torque tube housing by approximately 0.030
inches before securing to prevent interference.
12. Tighten the drive arm assembly onto the torque tube shaft as follows:
a. Supporting the block and shaft, tighten the clamp screw until it is
snug to the shaft.
b. Still supporting the block/shaft, tighten the clamp screw an addi-
tional 1/3 to 1/2 turn. (The screw can normally turn one full revolu-
tion before breaking.)
LINK SINTERED
CLAMP BLOCK
.03 (approx)
DRIVE ARM
1/8” WRENCH
(PN 9A-0163-0044C)
TO TIGHTEN
13. Perform the drive arm tightness test (see Drive Arm Tightness Test, page
17) to ensure that the drive arm is adjusted properly and will not slip during
operation.
14. Replace any instrument components that were removed to allow removal
of the DPU, referring to Section 3 of this document as required.
15. Calibrate the instrument before returning it to service. See Indicator
Calibration, page 15.
Adjusting Pulsation Dampener
See Figure A.2, page A-3, for the location of the pulsation dampener valve.
1. Remove the dust cover and insert a 1/8-in. hex key into the needle valve.
2. Turn the valve clockwise to the closed position, approximately 3 turns.
3. Back out the valve 1/2 turn or as required to reduce pulsations/shock
pressures.
A-23
Appendix A Model 581A Differential Pressure Switch
Troubleshooting
See Table A.4 and refer to actuated instrument manual for more information.
A-24
Model 581A Differential Pressure Switch Appendix A
A-25
Appendix A Model 581A Differential Pressure Switch
Figure A.23—Model 199 DPU parts drawing (indicator and indicator switch mounting
brackets vary as shown; see 200A/289A and 581A insets)
A-26
Model 581A Differential Pressure Switch Appendix A
A-27
Appendix A Model 581A Differential Pressure Switch
A-28
Model 581A Differential Pressure Switch Appendix A
A-29
Appendix A Model 581A Differential Pressure Switch
A-30
Model 581A Differential Pressure Switch Appendix A
A-31
Appendix A Model 581A Differential Pressure Switch
A-32
Product Warranty
A. Warranty
Cameron International Corporation ("Cameron") warrants that at the time of shipment, the prod-
ucts manufactured by Cameron and sold hereunder will be free from defects in material and
workmanship, and will conform to the specifications furnished by or approved by Cameron.
B. Warranty Adjustment
(1) If any defect within this warranty appears, Buyer shall notify Cameron immediately.
(2) Cameron agrees to repair or furnish a replacement for, but not install, any product which
within one (1) year from the date of shipment by Cameron shall, upon test and examina-
tion by Cameron, prove defective within the above warranty.
(3) No product will be accepted for return or replacement without the written authorization of
Cameron. Upon such authorization, and in accordance with instructions by Cameron, the
product will be returned shipping charges prepaid by Buyer. Replacements made under
this warranty will be shipped prepaid.
C. Exclusions from Warranty
(1) THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER
EXPRESSED OR IMPLIED WARRANTIES OF MERCHANTABILITY, OR FITNESS FOR
A PARTICULAR PURPOSE, OR OTHERWISE.
(2) Components manufactured by any supplier other than Cameron shall bear only the war-
ranty made by the manufacturer of that product, and Cameron assumes no responsibility
for the performance or reliability of the unit as a whole.
(3) "In no event shall Cameron be liable for indirect, incidental, or consequential damages
nor shall the liability of Cameron arising in connection with any products sold hereunder
(whether such liability arises from a claim based on contract, warranty, tort, or otherwise)
exceed the actual amount paid by Buyer to Cameron for the products delivered hereun-
der."
(4) The warranty does not extend to any product manufactured by Cameron which has
been subjected to misuse, neglect, accident, improper installation or to use in violation of
instructions furnished by Cameron.
(5) The warranty does not extend to or apply to any unit which has been repaired or
altered at any place other than at Cameron's factory or service locations by persons not
expressly approved by Cameron.
Product Brand
Barton® is a registered trademark of Cameron International Corporation ("Cameron").
Tefzel® is a registered trademark of E.I. DuPont de Nemours & Co., Inc.
A-33
+603.5569.0501
ms-kl@c-a-m.com
RUSSIA