You are on page 1of 218

AUTO REFRACTOMETER, AUTO REF/KERATOMETER

AR-310A/AR-330A/AR-360A
ARK-510A/ARK-530A/ARK-560A

SERVICE MANUAL

September 2014
Pages in total: 218
XAR17ARDA001G/E
XAR17ARDA001G/E

Table of Contents

1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2. SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3 . 1 Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 . 2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

4. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
5. SUBTROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . .19
5 . 1 Initial Screen is not Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5 . 1 . 1 Power supply voltage check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 . 1 . 2 Backlight of color LCD does not light up . . . . . . . . . . . . . . . . . . . . . . 21
5 . 2 Error Message Appears . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 1 ERR001 (EEPROM error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 2 ERR002 (Clock error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 3 PD ERROR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5. 2. 3. 1 PD LED (30601-EA55) does not light up. . . . . . . . . . . . . . . 23
5 . 2 . 4 ERR011 (Time-out error in characters [OUT]) . . . . . . . . . . . . . . . . . . 24
5 . 2 . 5 ERR012 (Time-out error for receiving data [OUT]) . . . . . . . . . . . . . . 25
5 . 2 . 6 ERR013 (Receiving error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 7 ERR014 (Receiving code error [OUT]). . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 8 ERR015 (Reply time-out [OUT]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 9 ERR016 (DTR-DSR error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 10 ERR017 (Data error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 11 ERR018 (Command error [OUT]). . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 12 NO DAT (Data empty [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 13 ERR021 (Time-out error in characters [IN]) . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 14 ERR022 (Time-out error for receiving data [IN]) . . . . . . . . . . . . . . . . 25
5 . 2 . 15 ERR023 (Receiving error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 16 ERR024 (Receiving code error [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 17 ERR025 (Reply time-out [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 18 ERR026 (DTR-DSR error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 19 ERR027 (Data error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 20 ERR028 (Command error [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 21 ERR031 (Up-and-down tracking error). . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 22 ERR032 (Left-to-right tracking error) . . . . . . . . . . . . . . . . . . . . . . . . . 27
5 . 2 . 23 ERR033 (Back-and-forth tracking error) . . . . . . . . . . . . . . . . . . . . . . 28
5 . 2 . 24 ERR034 (Up/down chinrest error) . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5 . 2 . 25 NO PAPER. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3
XAR17ARDA001G/E

5 . 2 . 26 ERR043 (Printer connection error). . . . . . . . . . . . . . . . . . . . . . . . . . . 29


5 . 2 . 27 ERR101 (Sensor initialization error) . . . . . . . . . . . . . . . . . . . . . . . . . 30
5 . 2 . 28 ERR111 (Thermistor error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5 . 2 . 29 ERR112 (AR motor error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5 . 2 . 30 ERR121 (CYL 1 initialization error) . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5 . 2 . 31 ERR122 (CYL 2 initialization error) . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5 . 2 . 32 ERR123 (Chart initialization error) . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5 . 2 . 33 ERR501 (Power error of Eye Care card) . . . . . . . . . . . . . . . . . . . . . . 32
5 . 2 . 34 ERR502 (Writing error of Eye Care card) . . . . . . . . . . . . . . . . . . . . . 32
5 . 2 . 35 ERR503 (Power error of Eye Care card) . . . . . . . . . . . . . . . . . . . . . . 33
5 . 2 . 36 ERR601 (USB_A recognition error) . . . . . . . . . . . . . . . . . . . . . . . . . 33
5 . 2 . 37 ERR602 (USB_A class error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5 . 2 . 38 ERR700 (File sharing error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5 . 2 . 39 ERR703 (Hardware error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5 . 2 . 40 ERR704 (DHCP error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5 . 2 . 41 ERR750 (Network access error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5 . 2 . 42 ERR751 (Network writing error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5 . 2 . 43 ERR754 (PC name error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5 . 2 . 44 ERR755 (Read-only folder error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5 . 2 . 45 ERR756 (Log-on error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5 . 2 . 46 ERR757 (Shared folder error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5 . 2 . 47 ERR758 (Time-out error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5 . 2 . 48 ERR759 (Deletion error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5 . 2 . 49 ERR760 (Error during network initialization) . . . . . . . . . . . . . . . . . . . 38
5 . 2 . 50 ERR761 (Access right error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5 . 2 . 51 ERR762 (Account error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5 . 2 . 52 ERR771 (Cable connection error) . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5 . 2 . 53 ERR772 (Reply error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5 . 3 Function Buttons are not Operational . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5 . 4 Measuring Unit does not Move Smoothly along Base Unit . . . . . . . . . . . . . . . . 41
5 . 5 Rotating Joystick does not Move Measuring Unit Up or Down . . . . . . . . . . . . . . 42
5 . 6 Chinrest does not Move Up or Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5 . 7 Measurement cannot be Performed Properly . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5 . 7 . 1 Observed image is not displayed on LCD . . . . . . . . . . . . . . . . . . . . . 48
5 . 7 . 2 Mire ring is not displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5 . 7 . 3 Focus LED is not displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5 . 7 . 4 Chart is not visible . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5 . 7 . 5 Alignment is not automatically performed by 3D auto tracking function. . 51
5. 7. 5. 1 Alignment is not performed in side-to-side direction . . . . . . 51
5. 7. 5. 2 Alignment is not performed in up-and-down direction . . . . . 52
5. 7. 5. 3 Alignment is not performed in forward-and-backward direction . . 53
5 . 7 . 6 Measurement does not start even when start button is pressed . . . . 54
5 . 7 . 7 Measurement does not start even when alignment and focus are achieved. . 55
5 . 7 . 8 AR measurement data is not normal . . . . . . . . . . . . . . . . . . . . . . . . . 56
5. 7. 8. 1 BLK (blinking error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5 . 7 . 9 PD measurement data is not normal . . . . . . . . . . . . . . . . . . . . . . . . . 57
5 . 7 . 10 KM measurement data is not normal . . . . . . . . . . . . . . . . . . . . . . . . . 58

4
XAR17ARDA001G/E

5 . 7 . 11 KM periphery cannot be measured . . . . . . . . . . . . . . . . . . . . . . . . . . 58


5 . 7 . 12 CS measurement data is not normal . . . . . . . . . . . . . . . . . . . . . . . . . 59
5 . 7 . 13 PS measurement data is not normal . . . . . . . . . . . . . . . . . . . . . . . . . 59
5 . 8 Communication cannot be Performed Properly . . . . . . . . . . . . . . . . . . . . . . . . . 60
5 . 9 Printing cannot be Performed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5 . 10 LCD Auto Sleep Function does not Operate. . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

6. REMOVAL PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . .63


6 . 1 Measuring Unit Left Cover (30601-M707). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6 . 2 Measuring Unit Right Cover (30601-M708) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6 . 3 Measuring Unit Front Cover ASSY(30601-6300) . . . . . . . . . . . . . . . . . . . . . . . . 64
6 . 4 Front Body Cover (30601-M702) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6 . 5 Measuring Unit Rear Cover ASSY (30601-6000) . . . . . . . . . . . . . . . . . . . . . . . . 65
6 . 6 Rear Body Cover ASSY (30601-7000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6 . 7 Bottom Plate (30601-M002) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6 . 8 Dust Cover (30601-M304). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6 . 9 LCD Rear Cover (30601-M716) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6 . 10 LCD Front Cover (30601-M715) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

7. REPLACEMENT PROCEDURE . . . . . . . . . . . . . . . . . . . . . .69


7 . 1 ASSYs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7 . 1 . 1 Joystick ASSY (30601-2600) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7 . 1 . 2 Measuring ASSY (30601-9E00) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7 . 1 . 3 Phototransmitter ASSY (30601-3310) . . . . . . . . . . . . . . . . . . . . . . . . 70
7 . 1 . 4 Mire ASSY (30601-4000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7 . 1 . 5 LCD ASSY(30601-6100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7 . 1 . 6 Subjective cylinder ASSY (30703-3700) . . . . . . . . . . . . . . . . . . . . . . 72
7 . 1 . 7 Chart disk ASSY (30703-3800) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
7 . 2 Boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 . 2 . 1 Main board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 . 2 . 2 Tracking board (30601-BA02) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
7 . 2 . 3 Driver board (30601-BA03) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7 . 2 . 4 Base board (30601-BA05) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
7 . 2 . 5 LCD board (30601-BA06). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7 . 2 . 6 Left panel SW board (30601-BA07) . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7 . 2 . 7 U/D SW board (30601-BA09). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
7 . 2 . 8 Sensor carry board (30601-BA14) . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
7 . 2 . 9 Right panel SW board (30601-BA15). . . . . . . . . . . . . . . . . . . . . . . . . 79
7 . 2 . 10 PD boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
7 . 2 . 10. 1 BA08 PD CCD board (32105-BA08) . . . . . . . . . . . . . . . . . . 80
7 . 2 . 10. 2 PD CCD board (15601-BA08) . . . . . . . . . . . . . . . . . . . . . . . 80
7 . 2 . 11 DC-AC inverter (18535-E010) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
7 . 2 . 12 Printer interface board (80606-00019) . . . . . . . . . . . . . . . . . . . . . . . . 81
7 . 3 Electrical Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
7 . 3 . 1 R/L sensor (30602-EA20). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
7 . 3 . 2 F/B sensor (30602-EA21). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
7 . 3 . 3 U/D sensor (30601-EA22) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

5
XAR17ARDA001G/E

7 . 3 . 4 R/L motor (30602-EA26) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84


7 . 3 . 5 F/B motor (30602-EA27) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7 . 3 . 6 U/D motor (30601-EA28) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7 . 3 . 7 Origin sensor (30601-EA36). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7 . 3 . 8 CL motor (30703-EA37) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7 . 3 . 9 CL sensor (30703-EA38) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
7 . 3 . 10 Focus LED(30601-EA42) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
7 . 3 . 11 KM6 LED (30601-EA44). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
7 . 3 . 12 Chinrest (30601-EA53). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
7 . 3 . 13 PD LED (30601-EA55) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
7 . 3 . 14 Chart LED (30601-EA76) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
7 . 3 . 15 Chart (30703-EA74) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7 . 3 . 16 B/W CCD camera (30601-E001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
7 . 3 . 16. 1 Measurement CCD camera. . . . . . . . . . . . . . . . . . . . . . . . . 93
7 . 3 . 16. 2 Observation CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . 93
7 . 3 . 17 Color LCD (30601-E005) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
7 . 3 . 18 Brushless motor (30601-E011) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
7 . 3 . 19 Sensor motor (30601-E035). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
7 . 3 . 20 Switching power supply (80602-00102) . . . . . . . . . . . . . . . . . . . . . . . 96
7 . 3 . 21 Printer (80606-00018) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
7 . 4 Mechanical Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
7 . 4 . 1 R/L shaft (30601-M201) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
7 . 4 . 2 F/B shaft (30601-M103) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
7 . 4 . 3 Sliding plate (32105-M102) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
7 . 4 . 4 Wire (30601-M307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
7 . 4 . 5 Linear bearing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
7. 4. 5. 1 Linear bearing for F/B shafts (82001-LM028) . . . . . . . . . . 101
7. 4. 5. 2 Linear bearing for R/L shaft (82001-LM097) . . . . . . . . . . . 101

8. ADJUSTMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103
8 . 1 Entering Adjustment Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
8 . 1 . 1 MENU mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
8 . 1 . 2 CALIB MODE SELECT (for AR). . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 3 ALIGNMENT CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 4 Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 5 Parameter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
8 . 1 . 6 CALIB MODE SELECT (for KM) . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
8 . 1 . 7 EEPROM ERROR DISP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
8 . 2 Jig Attachment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
8 . 2 . 1 Service optical adjustment jig (30670-5300) . . . . . . . . . . . . . . . . . . 107
8. 2. 1. 1 Joint jig (30670-1100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
8. 2. 1. 2 Reticle illumination jig (30670-1200) . . . . . . . . . . . . . . . . . 109
8 . 2 . 2 0-D model eye 2 (32177-0618) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8 . 2 . 3 Dioptric tester (99104-G020) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8 . 2 . 4 Simple calibration model eye (32107-6200). . . . . . . . . . . . . . . . . . . 111
8 . 2 . 5 Chinrest attachment jig (32107-1100) . . . . . . . . . . . . . . . . . . . . . . . 111
8 . 2 . 6 22-model eye jig (32171-0600) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112

6
XAR17ARDA001G/E

8 . 2 . 7 Cylinder axis adjustment jig (32177-1500) . . . . . . . . . . . . . . . . . . . . 112


8 . 2 . 8 Chart test jig (32145-2100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
8 . 2 . 9 Chart test jig (30670-2300). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
8 . 2 . 10 Dioptric tester jig (30670-6000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
8 . 2 . 11 KM calibration jig (30670-0300) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
8 . 2 . 12 Aspheric model eye (32177-1400) . . . . . . . . . . . . . . . . . . . . . . . . . . 116
8 . 2 . 13 PD check jig (32545-1002). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
8 . 3 CCD Camera Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
8 . 3 . 1 Observation CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
8 . 3 . 2 Measurement CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
8 . 4 Phototransmitter Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
8 . 4 . 1 Internal reflection removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
8 . 4 . 2 AR SLD optical axis adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
8 . 5 Chart Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
8 . 5 . 1 AR-310A, AR-330A, ARK-510A, ARK-530A . . . . . . . . . . . . . . . . . . 124
8 . 5 . 2 AR-360A, ARK-560A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
8. 5. 2. 1 Subjective cylinder ASSY adjustment . . . . . . . . . . . . . . . . 125
8. 5. 2. 2 Chart position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . 127
8 . 6 ALIGNMENT CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
8 . 7 AR CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
8 . 7 . 1 AR CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
8 . 7 . 2 AR AXIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
8 . 7 . 3 AXIS OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
8 . 7 . 4 SPH OFFSET. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
8 . 7 . 5 CHART CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
8. 7. 5. 1 AR-310A/AR-330A/ARK-510A/ARK-530A . . . . . . . . . . . . 135
8. 7. 5. 1. 1 Chart test jig . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
8. 7. 5. 1. 2 Diopter tester jig ASSY . . . . . . . . . . . . . . . . . . . 138
8. 7. 5. 2 AR-360A/ARK-560A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8. 7. 5. 2. 1 Chart test jig . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8. 7. 5. 2. 2 Dioptric tester jig ASSY . . . . . . . . . . . . . . . . . . 148
8 . 8 KM CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
8 . 8 . 1 KM CALIB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
8 . 8 . 2 KM CALIB CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
8. 8. 2. 1 ADJUST GAIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
8. 8. 2. 2 ADJUST PHI6 GAIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
8 . 8 . 3 KM AXIS CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
8 . 9 PD CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
8 . 9 . 1 CENTER POSITION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
8 . 9 . 2 PD CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
8 . 10 CS/PS Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
8 . 11 AUTO TRC ADJUST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
8 . 12 Software Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169

7
XAR17ARDA001G/E

8 . 13 Time and Date Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171


8 . 14 CURSOR BOARD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
8 . 15 Rewriting EEPROM Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
8 . 16 Setting Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

9. SUPPLEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .177
9 . 1 Appearance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
9 . 1 . 1 Front view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
9 . 1 . 2 Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
9 . 2 Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
9 . 3 Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
9 . 3 . 1 Complete wiring diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
9. 3. 1. 1 30602-E000 (AR-330A and ARK-530A) . . . . . . . . . . . . . . 185
9. 3. 1. 2 30601-E000 (AR-310A and ARK-510A) . . . . . . . . . . . . . . 186
9. 3. 1. 3 30603-E000 (AR-360A) and 30703-E000 (ARK-560A) . . . 187
9 . 3 . 2 Partial writing diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
9 . 4 Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
9 . 5 Parameter List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
9 . 5 . 1 AR-310A/AR-330A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
9 . 5 . 2 AR-360A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
9 . 5 . 3 ARK-510A/ARK-530A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208
9 . 5 . 4 ARK-560A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
9 . 6 Jig List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 212
9 . 7 Adjustment Item List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
9 . 7 . 1 ASSYs, boards, and adjustment items. . . . . . . . . . . . . . . . . . . . . . . 213
9 . 7 . 2 Electrical components and adjustment items . . . . . . . . . . . . . . . . . . 214
9 . 7 . 3 Mechanical components and adjustment items . . . . . . . . . . . . . . . . 215
9 . 8 Screw List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216

8
XAR17ARDA001G/E

1 INTRODUCTION

This Service Manual is for the NIDEK AUTO REFRACTOMETER and AUTO REF/KERATOME-
TER models: AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A.
For correct service, thorough understanding of the contents in this manual is required prior to the
service.
Refer to the Operator’s Manual and Parts List for the AR-310A, AR-330A, AR-360A, ARK-510A,
ARK-530A, and ARK-560A.
In case the device cannot be repaired according to the procedures described in this manual,
please report the serial number of the device and details of the symptom or symptoms.
Refer to the Service Manual Annex for important changes.

9 Return to previous page


XAR17ARDA001G/E

This page is intentionally left blank.

10 Return to previous page


XAR17ARDA001G/E

2 SAFETY PRECAUTIONS

1 . Requirements for prior contact regarding repair


1 ) When conducting repairs described in the Service Manual or Service Manual Annex,
prior contact with NIDEK is not necessary.
2 ) When conducting repairs not covered in the Service Manual or Service Manual
Annex, prior contact with the NIDEK Service Department in writing is required.
3 ) When for some reason, repairs were conducted because they needed to be done
immediately, or other such circumstance, contact NIDEK as soon as possible in writ-
ing after conducting repairs to the device.

2 . General precautions
1 ) Repairs by the Service Manual or Service Manual Annex must be performed by per-
sonnel who have undergone training at NIDEK.
2 ) Conduct repairs according to the procedures described in the Service Manual or Ser-
vice Manual Annex.
3 ) Never wipe the covers using an organic solvent such as paint thinner.
Doing so may ruin the surface and impair the appearance of the device.
4 ) When conducting repairs, turn off the power switch and disconnect the power cord
from the wall outlet unless the power needs to be on.
5 ) A part emitting invisible laser beam (Class 3B laser) is used inside the device. (How-
ever, the device is classified as a Class 1 laser product indicating that the part does
not emit hazardous irradiation to the eye.)

3 . Cautions regarding power source and power cord


1 ) Disconnect the power cord by holding its plug. Do not place heavy objects on the cord
or coil the cord tightly.
A damaged power cord may cause fire or electric shock.
2 ) Occasionally clean the prongs of the power plug with a dry cloth.
If dust accumulates between the prongs, the dust could collect moisture, and short
circuit or fire may occur.
3 ) Fully insert the main power plug into an outlet. A fire may occur if the device is used
with improper connection.
4 ) Do not overload the electrical outlet. A fire may occur.
5 ) Be sure to use an outlet equipped with a ground terminal.
Electric shock may result in the event of device malfunction or power leakage.
6 ) Be sure to use a wall outlet that meets the power requirements.
If the line voltage is too high or too low, the device may not perform properly. Electric
shock, malfunction, or fire may result.

4 . Maintenance cautions
1 ) Never look directly at the beam from the phototransmitter ASSY (invisible laser: Class
3B laser) or block it with any part of the body.
2 ) Take antistatic precautions prior to repair.
3 ) Take proper care against electric shock during repair.
4 ) Tighten or loosen screws with proper tools.
5 ) Never drop parts or screws inside the device.
6 ) Prepare storage cases so as not to lose the removed screws or parts.

11 Return to previous page


XAR17ARDA001G/E

7 ) After loosening the screws fastened by a threadlocking adhesive, be sure to reapply


the threadlocking adhesive to the screws before retightening them.
8 ) After replacing parts, confirm that they are fastened in their original positions securely
before turning on the power.
9 ) See Wiring Diagram and Connecting Cable to check cable breaks as described in
TROUBLESHOOTING. In addition, check cables for the following:
a . Connectors and contact pins are connected securely.
b . Contact pins are crimped securely.
c . No contact failure occurs after re-connection of connectors.
d . Cables are soldered properly.
* Do not pull on the cables with excessive force. Doing so may result in cable
breakage. Never conduct repairs with wet hands. Electric shock or failure of the
device may result.
10) The labeled area indicates high voltage. Take proper precautions against electric
shock.
11 ) When a cable with any of the following marks is disconnected and reconnected, be
sure to confirm that the screws are tightened securely.
a . Protective ground:
b . Functional ground:
c . Equipotential:
12) Install the device where the outlet which the power plug is inserted into is easily
accessible during use. In addition, ensure that the power plug can be disconnected
without the use of a tool.
Otherwise, it may interfere with disconnection of the device from the input power
source in case of abnormality

5 . Adjustment precautions
1 ) Confirm that the following conditions are met before installation.
a . A place which is level and stable without vibration and shock
b . No exposure to direct sunlight or ultraviolet rays
c . A place with minimal external light sources
d . No exposure to water
e . A place where temperature and humidity meet the specifications for use
f . Dust-free and smoke-free
2 ) Never use the adjustment jigs for purposes other than those intended.

6 . Caution after repairs


3 ) Confirm that the device operates properly after repairs

12 Return to previous page


XAR17ARDA001G/E

3 SPECIFICATIONS

3.1 Classifications
1 . Type of protection against electrical shock: Class I
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as Class I devices.
A Class I device is one in which, in addition to basic insulation, protection against
electric shock is provided by connecting the device to a grounding source to prevent
accessible metal parts from becoming electrically charged, should the basic insulation
fail.
2 . Degree of protection against electrical shock: Type B applied part
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as devices with a Type B applied part.
A Type B Applied Part provides a particular degree of protection against electric
shock, particularly regarding the following:
- allowable leakage currents
- reliability of the protective ground connection (if applicable)
3 . Conformity to electromagnetic compatibility standard
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A conform
to the EMC standard of IEC60601-1-2: 2007.
4 . Degree of protection against harmful ingress of water: IPX0
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A provide
no protection against ingress of water with harmful effects.
Avoid splashing water or other liquids on or near these devices.
5 . Degree of safety for application in the presence of a flammable anesthetic mixture with air,
or with oxygen or nitrous oxide.
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are not
suitable for use in the presence of a flammable anesthetic mixture with air, or with
oxygen or nitrous oxide.
Do not operate the device near flammable type materials.
6 . Method(s) of sterilization or disinfection recommended by the manufacturer
The forehead rest and chinrest can be cleaned with a cloth dampened with rubbing
alcohol as necessary.
7 . Mode of operation
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as continuous operating devices.
8 . Installation and use
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as stationary devices.

13 Return to previous page


XAR17ARDA001G/E

3.2 Specifications
1 . Objective measurement of refractive error (AR measurement)
Spherical power (S) -30.00 to +25.00 D (VD = 12 mm)
0.01/0.12/0.25 D increments
Cylindrical power (C) -12.00 to +12.00 D
0.01/0.12/0.25 D
Cylinder axis (A) 0 to 180°
1°/5° increments
Vertex distance 0/10.5/12/13.75/15/16.5 mm
Minimum pupil diameter ø2 mm
measurable
Relaxation of Auto fogging system
accommodating eye
Chart Scenery chart
2 . PD measurement
Measurement range 30 to 85 mm
(Near PD = 28 to 80 mm, Near working distance = 40
cm)
1 mm increments
3 . Measurement of corneal curvature (KM measurement) (ARK-510A, ARK-530A, and ARK-
560A)
Corneal curvature radius 5.00 to 13.00 mm
(R1, R2, AVE) 0.01 mm increments
Corneal refractive power 25.96 to 67.50 D (n = 1.3375)
(R1, R2, AVE) 0.01/0.12/0.25 D increments
n = 1.3375/1.336/1.332
Corneal cylindrical power -12.00 to +12.00 D
(CYL) 0.01/0.12/0.25 D increments
Cylinder axis (A) 0 to 180°
1°/5°increments
KM measurable area ø3.3 mm (for mire ring/corneal curvature radius of 7.7
mm)
ø6.0 mm (for four points/corneal curvature radius of
7.7 mm)
Size of mire ring ø2 mm and ø3.3 mm on the cornea (for corneal curva-
ture of 7.7 mm)
Chart Scenery chart
4 . Sagittal radius measurement
Measurable angle Within 25° from center (SUP., INF, TEMP, and NAS.)
Sagittal curvature 5.00 to 13.00 mm
0.01 mm increments
Eccentricity (E) -4.10 to 2.05
5 . CS measurement (ARK-510A, ARK-530A, and ARK-560A)
Measurement range 10.0 to 14.0 mm
0.1 mm increments
6 . PS measurement (ARK-510A, ARK-530A, and ARK-560A)
Measurement range 1.0 to 10.0 mm
0.1 mm increments

14 Return to previous page


XAR17ARDA001G/E

7 . Subjective measurement of refractive error (AR-360A and ARK-560A)


Spherical power (S) -20.00 to +20.00 D (VD = 12 mm)
0.25 D increments
Cylindrical power (C) -8.00 to +8.00 D (AR measurement is used. The max-
imum value may be restricted depending on the SPH
value.)
0.25 D increments
Cylinder axis (A) 0 to 180°(AR measurement value is used.)
1°/5°increments
ADD power 0 to +9.75 D
0.25 D increments
Chart Alphanumeric charts
VA 20/200, 20/80, 20/60, 20/50, 20/40, 20/30, 20/25, 20/20,
or 20/200 to 20/20 (entire)
VA 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0, or 0.1 to 1.0
(entire)
Distant vision 20/200 or lower, 20/200, 20/80, 20/60, 20/50, 20/40, 20/30,
20/25, 20/20
0.1 or lower, 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0
Near vision 20/200 or lower, 20/200, 20/80, 20/60, 20/50, 20/40, 20/30,
20/25, 20/20
0.1 or lower, 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0
8 . Uncorrected visual acuity measurement (AR-360A, ARK-560A)
Measurement range 20/200 or lower, 20/200, 20/80, 20/60, 20/50, 20/40, 20/30,
20/25, 20/20
0.1 or lower, 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0
Chart VA 20/200, 20/80, 20/60, 20/50, 20/40, 20/30, 20/25, 20/20,
or 20/200 to 20/20 (entire)
VA 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0, or 0.1 to 1.0
(entire)
9 . Working range of auto tracking
Up and down ±16 mm or more
Right and left ±5 mm (except AR-310A and ARK-510A)
back and forth ±5 mm (except AR-310A and ARK-510A)
10. Movable range of horizontal
direction
Back and forth 36 mm or more
Right and left 85 mm or more

11 . Other functions
Auto shot The measurement automatically starts when the eye
is properly aligned.
Alignment/observation 5.7-inch color LCD panel
method
Printer Thermal line printer with auto cutter
Width 58 mm
Interface connectors RS-232C: Two ports (IN/OUT: two-phase connectable
system)
USB: One port
LAN: One port

15 Return to previous page


XAR17ARDA001G/E

12. Dimensions and mass


Dimensions 260 mm (W) × 481 mm (D) × 455 mm (H)
Mass 20 kg
Power source AC 100 to 240 V ±10% 50/60 Hz
Power consumption 100 VA
13. Environmental conditions (during use)
Temperature 10 to 35ºC (50 to 95°F)
Humidity 30 to 90ºC (No condensation)
Athmospheric pressure 800 to 1060 hPa
14. Environmental conditions (during transport and storage in packed condition)
Temperature -10 to 55ºC (14 to 131°F)
Humidity 10 to 95ºC (No condensation)
Athmospheric pressure 700 to 1060 hPa
15. Others
Service life 8 years from the date of initial operation (certified)
Regular check, maintenance, and consumable parts
replacements are necessary.
Unit per package 1 unit

16 Return to previous page


XAR17ARDA001G/E

4 TROUBLESHOOTING

Connect the power cord, and turn on the power


switch.
Yes

No
Is the initial screen displayed? 5.1 Initial Screen is not Displayed (p19)

Yes

Yes
Does an error message appear? 5.2 Error Message Appears (p22)

No

No
Can the function buttons be operated? 5.3 Function Buttons are not Operational (p40)

Yes

No 5.4 Measuring Unit does not Move Smoothly along


Does the main unit move horizontally?
Base Unit (p41)
Yes

Does rotating the joystick move the measuring unit No 5.5 Rotating Joystick does not Move Measuring
up and down? Unit Up or Down (p42)
Yes

No
Does the chinrest move up and down smoothly? 5.6 Chinrest does not Move Up or Down (p44)

Yes

No 5.7 Measurement cannot be Performed Properly


Can measurement be performed properly?
(p46)
Yes

17 Return to previous page


XAR17ARDA001G/E

No 5.8 Communication cannot be Performed Properly


Can communication be performed properly?
(p60)
Yes

No
Can printing be performed? 5.9 Printing cannot be Performed (p60)

Yes

No 5.10 LCD Auto Sleep Function does not Operate


Does the LCD auto sleep function operate?
(p61)
Yes

Turn off the power switch.

18 Return to previous page


XAR17ARDA001G/E

5 SUBTROUBLESHOOTING

5.1 Initial Screen is not Displayed


No
Is the power supply voltage proper? 5.1.1 Power supply voltage check (p20)

Yes

Are the voltages between the following TPs on the


tracking board (30601-BA02) as shown below (see
9.3 [p185])?
No Replace the Tracking board (30601-BA02) (see
TP6 and TP1: DC +3.3 V
7.2.2 [p74]).
TP5 and TP1: DC +5 V
TP4 and TP1: DC +12 V
TP3 and TP1: DC +15 V
Yes

Are there any breaks in the BA01-BA03 cable Yes


Replace the BA01-BA03 cable (30601-EA13).
(30601-EA13) (see 9.3 [p185])?
No

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E022).
(30601-E022) (see 9.3 [p185])?
No

Does the backlight of the color LCD (30601-E005) No 5.1.2 Backlight of color LCD does not light up
light up? (p21)
Yes

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E021).
(30601-E021) (see 9.3 [p185])?
No

Replace the LCD board (30601-BA06) (see 7.2.5


[p77]).

No Replace the Color LCD (30601-E005) (see 7.3.17


Is the problem resolved?
[p94]).
Yes

End

19 Return to previous page


XAR17ARDA001G/E

5.1.1 Power supply voltage check


No
Is the power supply voltage proper? Set a voltage regulator.

Yes

Is the voltage between the P001 1st and 3rd pins No Replace the primary power supply unit (30601-
on the primary power supply unit (30601-EA01) the
EA01).
same as the input voltage (see 9.3 [p185])?
Yes

Are the voltages between the J51 4th and 1st, 5th
and 2nd, and 6th and 3rd pins on the switching No Replace the Switching power supply (80602-
power supply (80602-00102) DC +15 V (see 9.3 00102) (see 7.3.20 [p96]).
[p185])?
Yes

Are there any breaks in the power supply output Yes Replace the power supply output cable (30601-
cable (30601-EA51) (see 9.3 [p185])? EA51).
No

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52).
EA52) (see 9.3 [p185])?
No

Are there any breaks in the BA02 power supply Yes Replace the BA02 power supply cable (30601-
cable (30601-EA11) (see 9.3 [p185])? EA11).
No

End

20 Return to previous page


XAR17ARDA001G/E

5.1.2 Backlight of color LCD does not light up


Are the voltages between the P103 (J3) A27 and
No
A28, and B27 and B28 pins on the main board Replace the Main board (see 7.2.1 [p73]).
(30601-BA01) DC +12 V (see 9.3 [p185])?
Yes

Are there any breaks in the BA01-BA03 cable Yes


Replace the BA01-BA03 cable (30601-EA13).
(30601-EA13) (see 9.3 [p185])?

No

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E022).
(30601-E022) (see 9.3 [p185])?

No

Are there any breaks in the FL inverter cable Yes


Replace the FL inverter cable (30601-EA63).
(30601-EA63) (see 9.3 [p185])?

No

Replace the DC-AC inverter (18535-E010) (see


7.2.11 [p81]).

No Replace the Color LCD (30601-E005) (see 7.3.17


Is the problem resolved?
[p94]).

Yes

End

21 Return to previous page


XAR17ARDA001G/E

5.2 Error Message Appears


5.2.1 ERR001 (EEPROM error)
Does an error appear in 17. EEPROM ERROR Yes Perform adjustment and calibration according to
DISP of MENU mode (see 8.1.1 [p103])? the error contents.
No

Rewrite the EEPROM data of the driver board


(30601-BA03) (see 8.15 [p173]).

Yes
Is the problem resolved? End

No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

5.2.2 ERR002 (Clock error)


Set the time and date (see 8.13 [p171]).

In about 30 minutes, turn off and on the power


switch.

No
Does ERR002 occur? End

Yes

Replace the Main board (see 7.2.1 [p73]).

22 Return to previous page


XAR17ARDA001G/E

5.2.3 PD ERROR
Yes
Is the PD phototransmitter soiled? Clean the PD phototransmitter.

No

Does PD LED (30601-EA55) light up?


No 5.2.3.1 PD LED (30601-EA55) does not light up
*Check it with a camera (a cellular phone camera
(p23)
or such).
Yes

Are there any breaks in the PDC CD cable (30601- Yes


Replace the PDC CD cable (30601-EA30).
EA30) (see 9.3 [p187])?
No

Does the oscilloscope show that the waveform for


the driver board (30601-BA03) IC20 1st pin and No Replace the BA08 PD CCD board (32105-BA08)
P310 (J1) 1st pin (trigger side) forms two peak (see (see 7.2.10.1 [p80]).
9.3 [p187])
Yes

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

5.2.3.1 PD LED (30601-EA55) does not light up


Are there any breaks in the BA01-BA03 cable Yes
Replace the BA01-BA03 cable (30601-EA13).
(30601-EA13) (see 9.3 [p187])?
No

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52.
EA52) (see 9.3 [p187])?
No

Does the oscilloscope show a square waveform of


No Replace the Driver board (30601-BA03) (see 7.2.3
DC +5 V between the P505 (J5) 1st and 2nd pins
[p75]).
on the base board (30601-BA05)?
Yes

Replace the PD LED (30601-EA55) (see 7.3.13


[p90]).

23 Return to previous page


XAR17ARDA001G/E

5.2.4 ERR011 (Time-out error in characters [OUT])


Correct the setting of the connected device.
No
Is the setting of the connected device correct? (Refer to the operator's manual of the connected
device.)
Yes

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52).
EA52) (see 9.3 [p185])?
No

Replace the Base board (30601-BA05) (see 7.2.4


[p76]).

No
Is the problem resolved? End

Yes

Replace the Main board (see 7.2.1 [p73]).

The connected device is defective.


No
Is the problem resolved? (Refer to the service manual for the connected
device.)
Yes

End

24 Return to previous page


XAR17ARDA001G/E

5.2.5 ERR012 (Time-out error for receiving data [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.6 ERR013 (Receiving error [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.7 ERR014 (Receiving code error [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.8 ERR015 (Reply time-out [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.9 ERR016 (DTR-DSR error [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.10 ERR017 (Data error [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.11 ERR018 (Command error [OUT])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.12 NO DAT (Data empty [OUT])


Perform the measurement, then confirm that the data displayed is the measurement data.

5.2.13 ERR021 (Time-out error in characters [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.14 ERR022 (Time-out error for receiving data [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.15 ERR023 (Receiving error [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.16 ERR024 (Receiving code error [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.17 ERR025 (Reply time-out [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

25 Return to previous page


XAR17ARDA001G/E

5.2.18 ERR026 (DTR-DSR error [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.19 ERR027 (Data error [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.20 ERR028 (Command error [IN])


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.2.21 ERR031 (Up-and-down tracking error)


Perform the procedure as in “5.5 Rotating Joystick does not Move Measuring Unit Up or
Down” (p42).

26 Return to previous page


XAR17ARDA001G/E

5.2.22 ERR032 (Left-to-right tracking error)


Are there any breaks in the R/L motor (30602- Yes Replace the R/L motor (30602-EA26) (see 7.3.4
EA26) (see 9.3 [p185])? [p84]).
No

Are the voltages between the following pins of


P209 (J9) on the tracking board (30601-BA02) as
shown below when each sensor of the R/L sensor
(30602-EA20) is shaded (see 9.3 [p185])?
• R/L position sensor: 2nd and 4th pins No Replace the R/L sensor (30602-EA20) (see 7.3.1
When shaded: DC 0 V
[p82]).
When not shaded: DC +5 V
• R/L limit sensor: 3rd and 4th pins
When shaded: DC 0 V
When not shaded: DC +5 V
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Replace the Tracking board (30601-BA02) (see


7.2.2 [p74]).

No Replace the Driver board (30601-BA03) (see 7.2.3


Is the problem resolved?
[p75]).
Yes

End

27 Return to previous page


XAR17ARDA001G/E

5.2.23 ERR033 (Back-and-forth tracking error)


Are there any breaks in the F/B motor (30602- Yes Replace the F/B motor (30602-EA27) (see 7.3.5
EA27) (see 9.3 [p185])? [p85]).
No

Are the voltages between the following pins of


P210 (J10) on the tracking board (30601-BA02) as
shown below when each sensor of the F/B sensor
(30602-EA21) is shaded (see 9.3 [p185])?
• F/B position sensor: 2nd and 4th pins No Replace the F/B sensor (30602-EA21) (see 7.3.2
When shaded: DC 0 V [p82]).
When not shaded: DC +5 V
• F/B limit sensor: 3rd and 4th pins
When shaded: DC 0 V
When not shaded: DC +5 V
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Replace the Tracking board (30601-BA02) (see


7.2.2 [p74]).

No Replace the Driver board (30601-BA03) (see 7.2.3


Is the problem resolved?
[p75]).
Yes

End

5.2.24 ERR034 (Up/down chinrest error)


Perform the procedure as in “5.6 Chinrest does not Move Up or Down” (p44).

5.2.25 NO PAPER
Replace the printing paper. (Refer to the operator's manual.)

28 Return to previous page


XAR17ARDA001G/E

5.2.26 ERR043 (Printer connection error)


No
Is the cover closed? Close the cover.

Yes

Open the cover once, then close it.

Yes
Is the problem resolved? End

No

Is the voltage between the P316 (J16) 1st and 2nd


No
pins on the driver board (30601-BA03) DC +5 V Replace the Driver board (30601-BA03) (see 7.2.3
(see 9.3 [p185])? [p75]).
Yes

Are there any breaks in the printer power supply Yes Replace the printer power supply cable (30601-
cable (30601-EA31) (see 9.3 [p185])? EA31).
No

Are there any breaks in the printer cable (30601- Yes


Replace the printer cable (30601-EA32).
EA32) (see 9.3 [p185])?
No

Replace the Printer interface board (80606-00019)


(see 7.2.12 [p81]).

Yes
Is the problem resolved? End

No

Replace the Printer (80606-00018) (see 7.3.21


[p97]).

29 Return to previous page


XAR17ARDA001G/E

Yes
Is the problem resolved? End

No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

5.2.27 ERR101 (Sensor initialization error)


Yes
Are there any breaks in the wire (30601-M307)? Replace the Wire (30601-M307) (see 7.4.4 [p100]).

No

Are there any breaks in the sensor motor (30601- Yes Replace the Sensor motor (30601-E035) (see
EA35) (see 9.3 [p185])? 7.3.19 [p95]).
No

Is the voltage between the P306 (J6) 3rd and 4th


pins on the driver board (30601-BA03) as shown
below when the sensor of the sensor origin (30601- No Replace the Origin sensor (30601-EA36) (see
EA36) is shaded? 7.3.7 [p86]).
• When shaded: DC +5 V
• When not shaded: DC 0 V
Yes

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

5.2.28 ERR111 (Thermistor error)


Replace the Driver board (30601-BA03) (see 7.2.3 [p75]).

30 Return to previous page


XAR17ARDA001G/E

5.2.29 ERR112 (AR motor error)


Are there any breaks in the prism motor cable No
Replace the prism motor cable (30601-EA34).
(30601-EA34) (see 9.3 [p185])?
Yes

Replace the Brushless motor (30601-E011) (see


7.3.18 [p94])..

Yes
Is the problem resolved? End

No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

5.2.30 ERR121 (CYL 1 initialization error)


Are there any breaks in the CL motor (30703- Yes Replace the CL motor (30703-EA37) (see 7.3.8
EA37) (see 9.3 [p185])? [p86]).
No

Are there any breaks in the CL sensor (30703- Yes Replace the CL sensor (30703-EA38) (see 7.3.9
EA38) (see 9.3 [p185])? [p87]).
No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

Yes
Is the problem resolved? End

No

Replace the Main board (see 7.2.1 [p73]).

5.2.31 ERR122 (CYL 2 initialization error)


Perform the procedure as in “5.2.30 ERR121 (CYL 1 initialization error)” (p31).

31 Return to previous page


XAR17ARDA001G/E

5.2.32 ERR123 (Chart initialization error)


Are there any breaks in the chart (30703-EA74) Yes Replace the Chart (30703-EA74) (see 7.3.15
(see 9.3 [p185])? [p92]).
No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

Yes
Is the problem resolved? End

No

Replace the Main board (see 7.2.1 [p73]).

5.2.33 ERR501 (Power error of Eye Care card)


Replace the Eye Care card.

No Replace the Base board (30601-BA05) (see 7.2.4


Is the problem resolved?
[p76]).
Yes

End

5.2.34 ERR502 (Writing error of Eye Care card)


Reinsert the Eye Care card for writing.

Yes
Does ERR502 appear? Replace the card.

No

End

32 Return to previous page


XAR17ARDA001G/E

5.2.35 ERR503 (Power error of Eye Care card)


Perform the procedure as in “5.2.34 ERR502 (Writing error of Eye Care card)” (p32).

5.2.36 ERR601 (USB_A recognition error)


Are there any USB devices other than the optional Yes Remove the connected USB device (except the
barcode scanner connected to the device?. USB flash drive for upgrade).
No

Replace the Main board (see 7.2.1 [p73]).

5.2.37 ERR602 (USB_A class error)


Reconnect the USB device, then restart the device.

Yes
Is the problem resolved? End

No

Connect other USB device.

No
Does ERR602 occur? Replace the USB device.

Yes

Replace the Base board (30601-BA05) (see 7.2.4


[p76]).

Yes
Is the problem resolved? End

No

Replace the Main board (see 7.2.1 [p73]).

33 Return to previous page


XAR17ARDA001G/E

5.2.38 ERR700 (File sharing error)


No
Is the LAN cable connected to the LAN port? Connect the LAN cable.

Yes

No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).

Yes

Contact the network administrator.

5.2.39 ERR703 (Hardware error)


Turn off the power switch of the device, then turn it
on again.

Yes
Is the problem resolved? End

No

Replace the Main board (see 7.2.1 [p73]).

5.2.40 ERR704 (DHCP error)


The IP address cannot be obtained.
Contact the network administrator as the DHCP server may not operate normally.
Or, set the DHCP connection to OFF to set the IP address and subnet mask (see 8.16
[p174]).

34 Return to previous page


XAR17ARDA001G/E

5.2.41 ERR750 (Network access error)


No
Is the LAN cable connected to the LAN port? Connect the LAN cable.

Yes

No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).

Yes

Yes
Are there any breaks in the LAN cable? Replace the LAN cable.

No

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52).
EA52) (see 9.3 [p185])?
No

Replace the Base board (30601-BA05) (see 7.2.4


[p76]).

Yes
Is the problem resolved? End

No

Replace the Main board (see 7.2.1 [p73]).

35 Return to previous page


XAR17ARDA001G/E

5.2.42 ERR751 (Network writing error)


Is the folder in the destination PC set to the shared No Set the folder in the destination PC to the shared
folder? folder.
Yes

Is the name of the folder in the destination PC the No


Correct the network setting (see 8.16 [p174]).
same as that of the folder in the device?
Yes

No
Is write access to the shared folder enabled? Enable write access to the shared folder.

Yes

Is there enough free space on the drive containing No


Free up space on the drive.
the shared folder?
Yes

Contact the network administrator.

36 Return to previous page


XAR17ARDA001G/E

5.2.43 ERR754 (PC name error)


No
Is the LAN cable connected to the LAN port? Connect the LAN cable.

Yes

No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).

Yes

Yes
Are there any breaks in the LAN cable? Replace the LAN cable.

No

No
Is the destination PC connected by LAN? Connect the destination PC by LAN.

Yes

Is the network setting of the destination PC No Correct the network setting of the destination PC
correct? (see 8.16 [p174]).
Yes

Replace the Main board (see 7.2.1 [p73]).

5.2.44 ERR755 (Read-only folder error)


The shared folder in the destination PC is write-protected.
Disable write protection of the shared folder in the destination PC.

37 Return to previous page


XAR17ARDA001G/E

5.2.45 ERR756 (Log-on error)


Is the user name entered in the user name field of No Enter the correct user name in the user name field
the network setting correct? of the network setting (see 8.16 [p174]).
Yes

Is the password entered in the password field of the No Enter the correct password in the password field of
network setting correct? the network setting (see 8.16 [p174]).
Yes

Is the user account setting of the destination PC No Correct the user account setting of the destination
correct? PC.
Yes

Contact the network administrator.

5.2.46 ERR757 (Shared folder error)


No Set the correct folder name in the network setting
Is the folder name in the network setting correct?
(see 8.16 [p174]).
Yes

Does a shared folder with the same name as the No Create in the destination PC a shared folder with
one set in the device exist in the destination PC? the same name as the one set in the device.
Yes

Contact the network administrator.

5.2.47 ERR758 (Time-out error)


Send the data again after a while.

5.2.48 ERR759 (Deletion error)


The shared folder in the destination PC is write-protected.
Disable write protection of the shared folder in the destination PC.

5.2.49 ERR760 (Error during network initialization)


Wait for some time then try again.
If the network initialization still fails after several tries, replace the main board (30601-BA01)
(see 7.2.1 [p73]).

38 Return to previous page


XAR17ARDA001G/E

5.2.50 ERR761 (Access right error)


Perform the procedure as in “5.2.45 ERR756 (Log-on error)” (p38).

5.2.51 ERR762 (Account error)


Perform the procedure as in “5.2.45 ERR756 (Log-on error)” (p38).

5.2.52 ERR771 (Cable connection error)


No
Is the LAN cable connected to the LAN port? Connect the LAN cable.

Yes

Yes
Are there any breaks in the LAN cable? Replace the LAN cable.

No

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52).
EA52) (see 9.3 [p185])?
No

Replace the Base board (30601-BA05) (see 7.2.4


[p76]).

5.2.53 ERR772 (Reply error)


The data reception software (NAVIS) in the PC could not delete the data within the specified
time (5 seconds after data reception).
Check the setting of the data reception software (NAVIS) in the PC.

39 Return to previous page


XAR17ARDA001G/E

5.3 Function Buttons are not Operational


Are there conductions between the following pins
of J1 on the left panel SW board (30601-BA07)
when each switch on the panel is pressed (see 9.3
[p185])? No Replace the Left panel SW board (30601-BA07)
SW1: 3rd and 4th pins (see 7.2.6 [p77]).
SW2: 3rd and 5th pins
SW3: 3rd and 6th pins
SW4: 3rd and 7th pins
Yes

Are there conductions between the following pins


of J1 on the right panel SW board (30601-BA15)
when each switch on the panel is pressed (see 9.3
[p185])? No Replace the Right panel SW board (30601-BA15)
SW1: 1st and 2nd pins (see 7.2.9 [p79]).
SW2: 1st and 3rd pins
SW3: 1st and 4th pins
SW4: 1st and 5th pins
Yes

Are there any breaks in the BA07 cable (30601- Yes


Replace the BA07 cable (30601-EA64).
EA64) (see 9.3 [p185])?
No

Are there any breaks in the BA15 cable (30601- Yes


Replace the BA15 cable (30601-EA65).
EA65) (see 9.3 [p185])?
No

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E022).
(30601-E022) (see 9.3 [p185])?
No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

40 Return to previous page


XAR17ARDA001G/E

5.4 Measuring Unit does not Move Smoothly along Base Unit
Clean the sliding plate (32105-M102).

Yes
Is the problem resolved? End

No

Does the measuring unit move smoothly to the No Replace the Sliding plate (32105-M102) (see 7.4.3
right, left, forward, and backward? [p99]).
Yes

Replace the R/L shaft (30601-M201) (see 7.4.1


Does the measuring unit move smoothly to the No [p98]).
right and left? Replace the Linear bearing for R/L shaft (82001-
LM097) (see 7.4.5.2 [p101]).
Yes

Replace the F/B shaft (30601-M103) (see 7.4.2


[p99]).
Replace the Linear bearing for F/B shafts (82001-
LM028) (see 7.4.5.1 [p101]).

41 Return to previous page


XAR17ARDA001G/E

5.5 Rotating Joystick does not Move Measuring Unit Up or Down


Are there any breaks in the U/D switch cable Yes
Replace the U/D switch cable (30601-EA25).
(30601-EA25) (see 9.3 [p185])?
No

Is the voltage between the P205 (J5) 8th and 9th


No Replace the Tracking board (30601-BA02) (see
pins on the tracking board (30601-BA02) DC +5 V
7.2.2 [p74]).
(see 9.3 [p185])?
Yes

Do the output voltages between the P205 (J5) 6th


and 9th, and 7th and 9th pins on the U/D SW board
(30601-BA09) show the following waveform when
the joystick is rotated (the frequency differs
No Replace the Joystick ASSY (30601-2600) (see
depending on the rotational speed of the joystick)
7.1.1 [p69]).
(see 9.3 [p185])?

0.8 V or more

Yes

Are there any breaks in the BA02 power supply Yes Replace the BA02 power supply cable (30601-
cable (30601-EA11) (see 9.3 [p185])? EA11).
No

Are the voltages between the following pins of


P211 (J11) on the tracking board (30601-BA02) as
shown below when each sensor of the U/D sensor
(30602-EA22) is shaded (see 9.3 [p185])?
• Up limit sensor: 3rd and 4th pins No Replace the U/D sensor (30601-EA22) (see 7.3.3
When shaded: DC 0 V [p83]).
When not shaded: DC +5 V
•Down limit sensor: 2nd and 4th pins
When shaded: DC 0 V
When not shaded: DC +5 V
Yes

Does the voltage between the P208 (J8) 1st and


2nd pins on the tracking board (30601-BA02)
Yes Replace the Tracking board (30601-BA02) (see
change in the range of 0 to 12 V (turning the
7.2.2 [p74]).
joystick CW or CCW inverts the + and - outputs)
when the joystick is turned (see 9.3 [p185])?
No

42 Return to previous page


XAR17ARDA001G/E

Replace the U/D motor (30601-EA28) (see 7.3.6


[p85]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

43 Return to previous page


XAR17ARDA001G/E

5.6 Chinrest does not Move Up or Down


Are there any breaks in the U/D switch cable Yes
Replace the U/D switch cable (30601-EA25).
(30601-EA25) (see 9.3 [p185])?
No

Are there conductions between the following pins


of P205 (J5) on the tracking board (30601-BA02)
when the up/down button is pressed (see 9.3 No Replace the U/D SW board (30601-BA09) (see
[p185])? 7.2.7 [p78]).
Up button: 1st and 3rd pins
Down button: 1st and 2nd pins
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Are there any breaks in the chinrest cable (30601- Yes


Replace the chinrest cable (30601-EA54).
EA54) (see 9.3 [p185])?
No

Is there any breaks in the chinrest (30601-EA53) Yes Replace the Chinrest (30601-EA53) (see 7.3.12
(see 9.3 [p185])? [p89]).
No

Are there any breaks in the BA01-BA03 cable Yes


Replace the BA01-BA03 cable (30601-EA13).
(30601-EA13) (see 9.3 [p185])?
No

Are there any breaks in the base cable (30601- Yes


Replace the base cable (30601-EA52).
EA52) (see 9.3 [p185])?
No

44 Return to previous page


XAR17ARDA001G/E

Are the voltages between the following pins of


P102 (J2) on the main board (30601-BA01) as
shown below when each sensor of the chinrest
(30601-EA53) is shaded (see 9.3 [p185])?
• Up limit sensor: 4th and 7th pins No Replace the Chinrest (30601-EA53) (see 7.3.12
When shaded: DC 0 V [p89]).
When not shaded: DC +5 V
•Down limit sensor: 6th and 7th pins
When shaded: DC 0 V
When not shaded: DC +5 V
Yes

Are the voltages between the following pins of


P102 (J2) on the main board (30601-BA01) as
shown below when the up/down button is pressed?
Up button:
No Replace the Driver board (30601-BA03) (see 7.2.3
26th and 33rd pins: DV +5 V
[p75]).
27th and 33rd pins: DC 0 V
Down button:
26th and 33rd pins: DV 0 V
27th and 33rd pins: DC +5 V
Yes

Is the voltage between the J004 1st and 2nd pins of


the chinrest (30601-EA53) as shown below when
No Replace the Base board (30601-BA05) (see 7.2.4
the up/down button is pressed (see 9.3 [p185])?
[p76]).
Up button: DC +12 V
Down button: DC -12 V
Yes

Replace the Chinrest (30601-EA53) (see 7.3.12


[p89]).

45 Return to previous page


XAR17ARDA001G/E

5.7 Measurement cannot be Performed Properly


Yes
Does an error appear? 5.2 Error Message Appears (p22)

No

No 5.7.1 Observed image is not displayed on LCD


Is the observed image displayed on the LCD?
(p48)
Yes

No
Is the mire ring displayed? 5.7.2 Mire ring is not displayed (p49)

Yes

No
Is the focus LED (30601-EA42) displayed? 5.7.3 Focus LED is not displayed (p49)

Yes

No
Is the chart visible? 5.7.4 Chart is not visible (p50)

Yes

Is alignment performed automatically by 3D auto No 5.7.5 Alignment is not automatically performed by


tracking function? 3D auto tracking function (p51)
Yes

Does the measurement start when the start button No 5.7.6 Measurement does not start even when start
is pressed in manual mode? button is pressed (p54)
Yes

No 5.7.7 Measurement does not start even when


Can the model eye be measured?
alignment and focus are achieved (p55)
Yes

Is the AR measurement data of the model eye No


5.7.8 AR measurement data is not normal (p56)
normal?
Yes

46 Return to previous page


XAR17ARDA001G/E

No
Is the PD measurement data normal? 5.7.9 PD measurement data is not normal (p57)

Yes

No
Is the KM measurement data normal? 5.7.10 KM measurement data is not normal (p58)

Yes

No
Can the KM periphery be measured? 5.7.11 KM periphery cannot be measured (p58)

Yes

No
Is the CS measurement data normal? 5.7.12 CS measurement data is not normal (p59)

Yes

No
Is the PS measurement data normal? 5.7.13 PS measurement data is not normal (p59)

Yes

End

47 Return to previous page


XAR17ARDA001G/E

5.7.1 Observed image is not displayed on LCD


Are the voltages between the following pins of
P105 (J5) on the main board (30601-BA01) as
No
shown below (see 9.3 [p185])? Replace the Main board (see 7.2.1 [p73]).
3rd and 2nd pins: DC +12 V
4th and 5th pins: DC +5 V
Yes

Are there any breaks in the ARCCD2 cable (30601- Yes


Replace the ARCCD2 cable (30601-EA12).
EA12) (see 9.3 [p185])?
No

Replace the Observation CCD camera (see


7.3.16.2 [p93]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

48 Return to previous page


XAR17ARDA001G/E

5.7.2 Mire ring is not displayed


Are the voltages between P309 (J9) 8th and 9th,
No Replace the Driver board (30601-BA03) (see 7.2.3
and 10th and 11th pins on the driver board (30601-
[p75]).
BA03) DC +15 V (see 9.3 [p185])?
Yes

Are there any breaks in the mire cable (300601- Yes


Replace the mire cable (30601-EA39).
EA39) (see 9.3 [p185])?
No

Replace the Mire ASSY (30601-4000) (see 7.1.4


[p71]).

No Replace the Driver board (30601-BA03) (see 7.2.3


Is the problem resolved?
[p75]).
Yes

End

5.7.3 Focus LED is not displayed


Are there any breaks in the mire cable (300601- Yes
Replace the mire cable (30601-EA39).
EA39) (see 9.3 [p185])?
No

Are the voltages between the J2 and J3 1st and


2nd pins DC +4 V or more when P402 and P403 on No Replace the Mire ASSY (30601-4000) (see 7.1.4
the mire board (30601-BA04) are disconnected [p71]).
(see 9.3 [p185])?
Yes

Are there any breaks in the focus LED (30601- Yes Replace the Focus LED(30601-EA42) (see 7.3.10
EA42) (see 9.3 [p185])? [p88]).
No

End

49 Return to previous page


XAR17ARDA001G/E

5.7.4 Chart is not visible


Are there any breaks in the sensor carry cable Yes
Replace the sensor carry cable (30601-EA14).
(30601-EA14) (see 9.3 [p185])?
No

Are there any breaks in the chart LED (30601- Yes Replace the Chart LED (30601-EA76) (see 7.3.14
EA76) (see 9.3 [p185])? [p91]).
No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

50 Return to previous page


XAR17ARDA001G/E

5.7.5 Alignment is not automatically performed by 3D auto tracking function


5.7.5.1 Alignment is not performed in side-to-side direction
* AR-330A, ARK-530A, AR-360A, and ARK-560A only
Does an error appear when the power is turned Yes
5.2 Error Message Appears (p22)
on?
No

No
Is auto tracking icon 3D displayed? Display auto tracking icon 3D.

Yes

Is the mire ring displayed? (Check it by using a No


5.7.2 Mire ring is not displayed (p49)
model eye.)
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Are there any breaks in the BA01-BA03 cable Yes


Replace the BA01-BA03 cable (30601-EA13).
(30601-EA13) (see 9.3 [p185])?
No

Does the voltage between the A3 and B3 pins of


P204 (J4) on the tracking board (30601-BA02)
change in the range of ±5 V (polarity is reversed No Replace the Tracking board (30601-BA02) (see
according to the directions) when the alignment is 7.2.2 [p74]).
moved in the side-to-side direction from the
focused position (see 9.3 [p185])?
Yes

Replace the R/L motor (30602-EA26) (see 7.3.4


[p84]).

51 Return to previous page


XAR17ARDA001G/E

5.7.5.2 Alignment is not performed in up-and-down direction


Does an error appear when the power is turned Yes
5.2 Error Message Appears (p22)
on?
No

No
Is the auto tracking icon displayed? Display the auto tracking icon.

Yes

Is the mire ring displayed? (Check it by using a No


5.7.2 Mire ring is not displayed (p49)
model eye.)
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Are there any breaks in the BA01-BA03 cable Yes


Replace cable BA01-BA03 (30601-EA13).
(30601-EA13) (see 9.3 [p185])?
No

Does the voltage between the A5 and B5 pins of


P204 (J4) on the tracking board (30601-BA02)
change in the range of DC ±5 V (polarity is No Replace the Tracking board (30601-BA02) (see
reversed according to the directions) when the 7.2.2 [p74]).
alignment is moved in the up-and-down direction
from the focused position?
Yes

Replace the U/D motor (30601-EA28) (see 7.3.6


[p85]).

52 Return to previous page


XAR17ARDA001G/E

5.7.5.3 Alignment is not performed in forward-and-backward direction


* AR-330A, ARK-530A, AR-360A, and ARK-560A only
Does an error appear when the power is turned Yes
5.2 Error Message Appears (p22)
on?
No

No
Is auto tracking icon 3D displayed? Display auto tracking icon 3D.

Yes

Is the mire ring displayed? (Check it by using a No


5.7.2 Mire ring is not displayed (p49)
model eye.)
Yes

Is the focus LED (30601-EA42) displayed? (Check No


5.7.3 Focus LED is not displayed (p49)
it by using a model eye.)
Yes

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Are there any breaks in cable BA01-BA03 (30601- Yes


Replace cable BA01-BA03 (30601-EA13).
EA13) (see 9.3 [p185])?
No

Does the voltage between the A4 and B4 pins of


P204 (J4) on the tracking board (30601-BA02)
change in the range of ±5 V (polar is reversed No Replace the Tracking board (30601-BA02) (see
according to the directions) when the alignment is 7.2.2 [p74]).
moved in the forward-and-backward direction from
the focused position?
Yes

Replace the F/B motor (30602-EA27) (see 7.3.5


[p85]).

53 Return to previous page


XAR17ARDA001G/E

5.7.6 Measurement does not start even when start button is pressed
Activate manual mode.

Is there conduction between the P902 1st and 2nd


No Replace the Joystick ASSY (30601-2600) (see
pins on the joystick (30601-EA92) when the start
7.1.1 [p69]).
button is pressed (see 9.3 [p185])?
Yes

Are there any breaks in the U/D switch cable Yes


Replace the U/D switch cable (30601-EA25).
(30601-EA25) (see 9.3 [p185])?
No

Are there any breaks in the BA02-BA03 cable Yes


Replace the BA02-BA03 cable (30601-EA24).
(30601-EA24) (see 9.3 [p185])?
No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

54 Return to previous page


XAR17ARDA001G/E

5.7.7 Measurement does not start even when alignment and focus are achieved
Yes
Does an error appear? 5.2 Error Message Appears (p22)

No

No Display the auto shot icon. (Refer to the operator’s


Is the auto shot icon displayed?
manual.)
Yes

Yes
Does the measurement start automatically? End

No

Replace the Driver board (30601-BA03) (see 7.2.3


[p75]).

55 Return to previous page


XAR17ARDA001G/E

5.7.8 AR measurement data is not normal


Does BLK (blinking error) appear when a model Yes
5.7.8.1 BLK (blinking error) (p57)
eye is measured?
No

Is the thumbnail result displayed properly on the Yes


Perform AR CALIBRATION (see 8.7 [p130]).
LCD after the AR measurement?
No

Are there any breaks in the ARCCD1 cable (30601- Yes


Replace the ARCCD1 cable (30601-EA72).
EA72) (see 9.3 [p185])?
No

Are there any breaks in the sensor carry cable Yes


Replace the sensor carry cable (30601-EA14).
(30601-EA14) (see 9.3 [p185])?
No

Does the phototransmitter ASSY (30601-3310) No Replace the Phototransmitter ASSY (30601-3310)
light up? (see 7.1.3 [p70]).
Yes

Replace the Measurement CCD camera (see


7.3.16.1 [p93]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

56 Return to previous page


XAR17ARDA001G/E

5.7.8.1 BLK (blinking error)


Is the result thumbnail displayed properly on the Yes
Perform AR CALIBRATION (see 8.7 [p130]).
LCD after the measurement?
No

Are there any breaks in the ARCCD1 cable (30601- Yes


Replace the ARCCD1 cable (30601-EA72).
EA72) (see 9.3 [p185])?
No

Are there any breaks in the sensor carry cable Yes


Replace the sensor carry cable (30601-EA14).
(30601-EA14) (see 9.3 [p185])?
No

Does the phototransmitter ASSY (30601-3310) No Replace the Phototransmitter ASSY (30601-3310)
light up? (see 7.1.3 [p70]).
Yes

Replace the Measurement CCD camera (see


7.3.16.1 [p93]).

Is the problem resolved? No


Replace the Main board (see 7.2.1 [p73]).

Yes

End

5.7.9 PD measurement data is not normal


Perform the procedure as in “5.2.3 PD ERROR” (p23).

57 Return to previous page


XAR17ARDA001G/E

5.7.10 KM measurement data is not normal


No Are there any breaks in the mire cable (30601-
Do the mire ring and focus LED light up?
EA39) (see 9.3 [p185])?
Yes
Yes No

Replace the Mire ASSY (30601-4000)


Replace the mire cable (30601-EA39).
(see 7.1.4 [p71]).

Perform KM CALIBRATION (see 8.8 [p156]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).

Yes

End

5.7.11 KM periphery cannot be measured


Do four bright spots of KM6 LED (30601-EA44)
No Are there any breaks in the mire cable (30601-
appear on the screen during the KM
EA39) (see 9.3 [p185])?
measurement?
Yes
Yes No

Replace the KM6 LED (30601-EA44) (see


Replace the mire cable (30601-EA39).
7.3.11 [p88]).

Perform KM CALIBRATION (see 8.8 [p156]).

No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73])..

Yes

End

58 Return to previous page


XAR17ARDA001G/E

5.7.12 CS measurement data is not normal


Perform the procedure as in “8.10 CS/PS Calibration” (p167).

5.7.13 PS measurement data is not normal


Perform the procedure as in “8.10 CS/PS Calibration” (p167).

59 Return to previous page


XAR17ARDA001G/E

5.8 Communication cannot be Performed Properly


Perform the procedure as in “5.2.4 ERR011 (Time-out error in characters [OUT])” (p24).

5.9 Printing cannot be Performed


Yes
Does an error appear? 5.2 Error Message Appears (p22)

No

Replace the printer cable (30601-EA32).

Yes
Is the problem resolved? End

No

Adjust the printing density of the printer by


parameter 34. PRINT DENCITY (see 8.1.5 [p105]).

Yes
Is the problem resolved? End

No

Replace the Printer (80606-00018) (see 7.3.21


[p97])..

60 Return to previous page


XAR17ARDA001G/E

5.10 LCD Auto Sleep Function does not Operate


No Activate sleep mode. (Refer to the operator's
Is sleep mode activated?
manual.)
Yes

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E021).
(30601-E021) (see 9.3 [p185])?
No

Are there any breaks in the flexible flat cable Yes


Replace the flexible flat cable (30601-E022).
(30601-E022) (see 9.3 [p185])?
No

Are there any breaks in the FL inverter cable Yes


Replace the FL inverter cable (30601-EA63).
(30601-EA63) (see 9.3 [p185])?
No

Replace the DC-AC inverter (18535-E010) (see


7.2.11 [p81]).

Yes
Is the problem resolved? End

No

Replace the LCD board (30601-BA06) (see 7.2.5


[p77]).

61 Return to previous page


XAR17ARDA001G/E

This page is intentionally left blank.

62 Return to previous page


XAR17ARDA001G/E

6 REMOVAL PROCEDURE

6.1 Measuring Unit Left Cover (30601-M707)


1 . Remove the caps [32105-M730 (n = 4)].
2 . Unscrew PC3 × 8 (n = 4). 
3 . Remove the measuring unit left cover
(30601-M707).
4 . Reassemble the parts in the reverse order.




2%˜

6.2 Measuring Unit Right Cover (30601-M708)


1 . Remove the caps [32105-M730 (n = 4)].
2 . Unscrew PC3 × 8 (n = 4).

3 . Remove the measuring unit right cover
(30601-M708).
4 . Reassemble the parts in the reverse order. 



2%˜

63 Return to previous page


XAR17ARDA001G/E

6.3 Measuring Unit Front Cover ASSY(30601-6300)


1 . Remove the measuring unit left cover (30601-M707) (see 6.1 [p63]).
2 . Remove the measuring unit right cover (30601-M708) (see 6.2 [p63]).
3 . Remove the caps [32909-M541 (n = 4)].
4 . Unscrew PC3 × 8 (n = 4).
5 . Remove the measuring unit front cover 
ASSY (30601-6300).
6 . Reassemble the parts in the reverse order.





2%˜

6.4 Front Body Cover (30601-M702)


1 . Remove the measuring unit front cover ASSY (30601-6300) (see 6.3 [p64]).
2 . Remove the caps [32105-M730 (n = 2)].
3 . Unscrew PC3 × 8 (n = 2).
4 . Remove the front body cover (30601-
M702). 
5 . Reassemble the parts in the reverse order.


2%˜




64 Return to previous page


XAR17ARDA001G/E

6.5 Measuring Unit Rear Cover ASSY (30601-6000)


1 . Remove the measuring unit left cover (30601-M707) (see 6.1 [p63]).
2 . Remove the measuring unit right cover (30601-M708) (see 6.2 [p63]).
3 . Disconnect P311 (J11), P312 (J12), and P316 (J16) on the driver board (30601-BA03).
4 . Remove the F.GND3 cable (30601-EA82).
5 . Remove the caps [32105-M730 (n = 4)].
6 . Unscrew PC3 × 8 (n = 4). 
7 . Remove the measuring unit rear cover
ASSY (30601-6000).
8 . Reassemble the parts in the reverse order.

2%˜




6.6 Rear Body Cover ASSY (30601-7000)


1 . Move the measuring unit to the top position.
2 . Remove the front body cover (30601-M702) (see 6.4 [p64]).
3 . Move the measuring unit to the right, and
unscrew SB3 × 8 (n = 2).


5$˜

4 . Move the measuring unit to the left, and


unscrew SB3 × 8 (n = 2). 
5 . Disconnect P901 (J1) and P902 (J2) on
the U/D SW board (30601-BA09).
6 . Remove the rear body cover ASSY
(30601-7000).
7 . Reassemble the parts in the reverse order.


5$˜

65 Return to previous page


XAR17ARDA001G/E

6.7 Bottom Plate (30601-M002)


1 . Unscrew BS3 × 6 (n = 4).
2 . Remove the bottom plate (30601-M002). 
3 . Reassemble the parts in the reverse order.

$5˜

6.8 Dust Cover (30601-M304)


1 . Remove the measuring unit rear cover ASSY (30601-6000) (see 6.5 [p65]).
2 . Unscrew AS3 × 6 (n = 2) on top of the dust

cover (30601-M304) and loosen AS3 × 6 #5˜

(n = 3) on the side of it.
3 . Remove the dust cover (30601-M304).
4 . Reassemble the parts in the reverse order.

66 Return to previous page


XAR17ARDA001G/E

6.9 LCD Rear Cover (30601-M716)


1 . Unscrew PC3 × 5 (n = 4).
2 . Remove the LCD rear cover (30601-
M716).
3 . Reassemble the parts in the reverse order.



3&™

6.10 LCD Front Cover (30601-M715)


1 . Remove the LCD ASSY (30601-6100) (see 7.1.5 [p71]).
2 . Unscrew PC3 × 5 (n = 4) to remove the LCD

rear cover (30601-M716). 3&™
3 . Disconnect P701 (J1) on the left panel SW
board (30601-BA07), and P1501 (J1) on the
right panel SW board (30601-BA15).
4 . Unscrew PT3 × 8 (n = 4) to remove the LCD
front cover (30601-M715) together with the left

panel SW board (30601-BA07) and right panel 
37™
SW board (30601-BA15).
5 . Remove the protective panel (30601-M602).
6 . Reassemble the parts in the reverse order.
7 . Perform CURSOR BOARD (see 8.14 [p172]).





67 Return to previous page


XAR17ARDA001G/E

This page is intentionally left blank.

68 Return to previous page


XAR17ARDA001G/E

7 REPLACEMENT PROCEDURE

7.1 ASSYs
7.1.1 Joystick ASSY (30601-2600)
Replacement part: 30601-2600
1 . Remove the rear body cover ASSY (30601-7000) (see 6.6 [p65]).
2 . Remove the GND cable of the joystick.
3 . Unscrew BS3 × 6 (n = 4) to replace the
joystick ASSY (30601-2600).
4 . Reassemble the parts in the reverse order. 


%6™

69 Return to previous page


XAR17ARDA001G/E

7.1.2 Measuring ASSY (30601-9E00)


Replacement part: 30601-9E00 (except the AR-360A and ARK-560A)
1 . Remove the covers (see Section 6 [p63]).
2 . Remove the mire ASSY (see 7.1.4 [p71]).
3 . Disconnect P105 (J5) on the main board (30601-BA01) and P304 (J4), P305 (J5), P306
(J6), P315 (J15), and the ground wire on the driver board (30601-BA03).
4 . Unscrew BS4 × 8 (n = 3) to remove the
measuring ASSY (30601-9E00). $5˜

'
5 . Reassemble the parts in the reverse order.
6 . Perform ALIGNMENT CALIBRATION (see
8.6 [p129]).
7 . Perform AR CALIBRATION (see 8.7
[p130]).
8 . Perform KM CALIBRATION (see 8.8
[p156]).
* The KM CALIBRATION applies only to
the ARK-510A and ARK-530A.
9 . Perform PD CALIBRATION (see 8.9
[p163]).
10. Perform CS/PS calibration (see 8.10
[p167]).
* The CS/PS calibration applies only to the ARK-510A, and ARK-530A.

7.1.3 Phototransmitter ASSY (30601-3310)


Replacement part: 30601-3310
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P1405 (J5) on the sensor carry board (30601-BA14).
3 . Unscrew HH3 × 4 to replace the pho-
totransmitter ASSY (30601-3310).
4 . Reassemble the parts in the reverse order.
5 . Adjust the phototransmitter (see 8.4
[p122]).
6 . Perform ALIGNMENT CALIBRATION **˜
(see 8.6 [p129]).
7 . Perform AR CALIBRATION (see 8.7
[p130]).



70 Return to previous page


XAR17ARDA001G/E

7.1.4 Mire ASSY (30601-4000)


Replacement part: 30601-4000
1 . Remove the measuring unit front cover ASSY (30601-6300) (see 6.3 [p64]).
2 . Disconnect P401 (J1) on the mire board (30601-BA04).
3 . Unscrew BS3 × 8 (n = 4) to remove the
mire ASSY (30601-4000).
4 . Reassemble the parts in the reverse order. 
5 . Perform ALIGNMENT CALIBRATION (see
8.6 [p129]).
6 . Perform KM CALIBRATION (see 8.8

[p156]). $5˜
* The KM CALIBRATION applies only to
the ARK-510A, ARK-530A, and ARK-
560A.

7.1.5 LCD ASSY(30601-6100)


Replacement part: 30601-6100
1 . Remove the measuring unit rear cover ASSY (30601-6000) (see 6.5 [p65]).
2 . Unscrew BS4 × 6 (n = 2) to replace the
LCD ASSY (30601-6100).
3 . Reassemble the parts in the reverse order.



$5˜

71 Return to previous page


XAR17ARDA001G/E

7.1.6 Subjective cylinder ASSY (30703-3700)


Replacement part: 30703-3700
* Applies only to the AR-360A and ARK-560.
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P307 (J7) and P308 (J8) on the driver board (30601-BA03).
3 . Unscrew BS3 × 8 (n = 2) to remove the

subjective cylinder ASSY (30703-3700). $5˜
4 . Reassemble the parts in the reverse order.
5 . Adjust the subjective cylinder ASSY (see 
8.5.2.1 [p125]).

7.1.7 Chart disk ASSY (30703-3800)


Replacement part: 30703-3800
* Applies only to the AR-360A and ARK-560A.
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P1404 (J4) on the sensor carry board (30601-BA14).
3 . Unscrew BS3 × 6 (n = 2) to remove the
chart disk ASSY (30703-3800).
4 . Reassemble the parts in the reverse order.
5 . Adjust the subjective cylinder ASSY (see
8.5.2.1 [p125]).

%6™
 

72 Return to previous page


XAR17ARDA001G/E

7.2 Boards
7.2.1 Main board
Replacement part: 30601-BA01P (for the AR-310A)
30602-BA01P (for the AR-330A)
30603-BA01P (for the AR-360A)
30701-BA01P (for the ARK-510A)
30702-BA01P (for the ARK-530A)
30703-BA01P (for the ARK-560A)
1 . Display the PARAMETER SETTING screen (see 8.1.5 [p105]).
2 . Press the print button to print out the parameter settings of the device.
3 . Remove the tracking board (30601-BA02) (see 7.2.2 [p74]).
4 . Remove the rear body cover ASSY (30601-7000) (see 6.6 [p65]).
5 . Unscrew AS3 × 6 (n = 4) to remove the
guard (30601-M225). $#
6 . Disconnect all connectors on the main
board (30601-BA01).

7 . Unscrew BS3 × 8 (n = 2) to remove the  
 #5˜
main board (30601-BA01) together with
the double locking spacers (80745-01103
[n = 4]). 
8 . Remove the double locking spacers $5˜
(80475-01103 [n = 4]).
9 . Replace the main board (30601-BA01).

10. Reassemble the parts in the reverse order.
11 . After the replacement, turning on power to
the device displays the following errors.
1 ) The AR-310A, AR-330A, AR-360A: ERR065
2 ) The ARK-510A, ARK-530A, ARK-560A: ERR066
* Do not turn off power to the device until the error is displayed.
* If the model of the replaced main board differs from that of the device, ERR062 to
ERR064 occur, and the device does not start up.
12. Turn the power off, then turn it on again.
13. Enter MENU mode, and select 9. PARAM-
2#4#/'6'45'66+0)
ETER SETTING (see 8.1.1 [p103]).
14. Press the up or down button to select the ,#2#0
(14'+)0
desired destination language. 810*1((
15. Press the set button, and confirm that OK 1%7.75
+0%
is displayed at the bottom of the screen. %*+0#
16. Press the exit button twice.
17. Turn the power off, and turn it on again.
18. Display the PARAMETER SETTING
screen (see 8.1.5 [p105]). 1-
19. Set the parameters based on the parame-
ter settings of the device that was printed
out earlier.

73 Return to previous page


XAR17ARDA001G/E

7.2.2 Tracking board (30601-BA02)


Replacement part: 30601-BA02
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect all connectors on the tracking
board (30601-BA02).
3 . Replace the tracking board (30601-BA02).
4 . Reassemble the parts in the reverse order.

$#

74 Return to previous page


XAR17ARDA001G/E

7.2.3 Driver board (30601-BA03)


Replacement part: 30601-BA03
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect all connectors on the driver
board (30601-BA03).
3 . Unscrew BS3 × 6 (n = 2). 
$5˜
4 . Replace the driver board (30601-BA03).
5 . Reassemble the parts in the reverse order.
* Reassemble the driver board (30601-
BA03) and PD board according to the
combinations below.
* Part and artwork numbers of each $#
board are described on the locations
as circled in the figures below.
Combination Driver board PD board
Part name Part number Artwork number Part name
Driver board 30601-BA03 PC5433 BA08 PD CCD board

Part name Part number Artwork number Part name


Driver board 30601-BA03 PC5780 PD CCD board

6 . After the replacement, turning on power to the device displays the following errors.
1 ) The AR-310A, AR-330A, AR-360A: ERR065
2 ) The ARK-510A, ARK-530A, ARK-560A: ERR066
* Do not turn off power to the device until the error is displayed.
* If the model of the replaced main board differs from that of the device, ERR062 to
ERR064 occur, and the device does not start up.
7 . Turn the power off, and turn it on again.
8 . Rewrite the EEPROM data (see 8.15 [p173]).

75 Return to previous page


XAR17ARDA001G/E

7.2.4 Base board (30601-BA05)


Replacement part: 30601-BA05
1 . Remove the bottom plate (30601-M002) (see 6.7 [p66]).
2 . Disconnect all connectors on the base board (30601-BA05).
3 . Unscrew BS3 × 8 (n = 3) and TW3 (n = 2), and disconnect the connector cable (30601-
EA01) and ground wire on the base board (30601-BA05).
4 . Unscrew BS3 × 8 (n = 3) to remove the 
inlet ASSY (30601-1100). $5˜



5 . Unscrew BS3 × 8 (n = 4), and replace the base


board (30601-BA05).
6 . Reassemble the parts in the reverse order.


$5˜
$#

76 Return to previous page


XAR17ARDA001G/E

7.2.5 LCD board (30601-BA06)


Replacement part: 30601-BA06
1 . Open the LCD ASSY (30601-6100) to remove the rear LCD cover (30601-M716) (see 6.9
[p67]).
2 . Disconnect all connectors on the LCD
board (30601-BA06).
3 . Unscrew PC2 × 4 (n = 2).
4 . Replace the LCD board (30601-BA06).
5 . Reassemble the parts in the reverse order. $#


2%˜

7.2.6 Left panel SW board (30601-BA07)


Replacement part: 30601-BA07
1 . Remove the LCD front cover (30601-M715) (see 6.10 [p67]).
2 . Unscrew PT2 × 5 (n = 2) to replace the left
panel SW board (30601-BA07).
3 . Reassemble the parts in the reverse order.

$#


26˜

77 Return to previous page


XAR17ARDA001G/E

7.2.7 U/D SW board (30601-BA09)


Replacement part: 30601-BA09
1 . Remove the rear body cover ASSY (30601-7000) (see 6.6 [p65]).
2 . Unscrew PT3 × 5 (n = 2).
3 . Replace the U/D SW board (30601-BA09). 26˜


4 . Reassemble the parts in the reverse order.


$#

7.2.8 Sensor carry board (30601-BA14)


Replacement part: 30601-BA14
1 . Remove the dust cover (see 6.8 [p66]).
2 . Disconnect all connectors on the sensor carry board (30601-BA14).
3 . Unscrew BS3 × 6 (n = 2).
4 . Replace the sensor carry board (30601-
BA14).
5 . Reassemble the parts in the reverse order.

$#


$5˜

78 Return to previous page


XAR17ARDA001G/E

7.2.9 Right panel SW board (30601-BA15)


Replacement part: 30601-BA15
1 . Remove the front LCD cover (30601-M715) (see 6.10 [p67]).
2 . Unscrew PT2 × 5 (n = 2) to replace the
right panel SW board (30601-BA15).
3 . Reassemble the parts in the reverse order.
$#


26˜

79 Return to previous page


XAR17ARDA001G/E

7.2.10 PD boards
7.2.10.1 BA08 PD CCD board (32105-BA08)
Replacement part: 32105-BA08
1 . Remove the measuring unit front cover ASSY (30601-6300) (see 6.3 [p64]).
2 . Disconnect all connectors on the BA08 PD
CCD board (32105-BA08).
3 . Unscrew BS3 × 6 (n = 2) to replace the
BA08 PD CCD board (32105-BA08).
* The BA08 PD CCD board (32105-
BA08) must be used in combination
with the driver board (30601-BA03),
artwork number PC5433 (see 7.2.3
[p75]).
4 . Reassemble the parts in the reverse order.
5 . Perform PD CALIBRATION (see 8.9
[p163]). $#

$5˜

7.2.10.2 PD CCD board (15601-BA08)


Replacement part: 15601-BA08
1 . Remove the measuring unit front cover ASSY (30601-6300) (see 6.3 [p64]).
2 . Disconnect all connectors on the PD CCD
board (15601-BA08).
3 . Unscrew BS3 × 6 (n = 2) to replace the PD
CCD board (15601-BA08).
* The PD CCD board (15601-BA08)
must be used in combination with the
driver board (30601-BA03), artwork
number PC5780 (see 7.2.3 [p75]).
4 . Reassemble the parts in the reverse order.
5 . Perform PD CALIBRATION (see 8.9
[p163]).
$#

$5˜

80 Return to previous page


XAR17ARDA001G/E

7.2.11 DC-AC inverter (18535-E010)


Replacement part: 18535-E010
1 . Open the LCD ASSY (30601-6100) to remove the rear LCD cover (30601-M716) (see 6.9
[p67]).
2 . Disconnect all connectors on the DC-AC
inverter (18535-E010).
3 . Unscrew PC2 × 4 (n = 2) to replace the
'
DC-AC inverter (18535-E010).
4 . Reassemble the parts in the reverse order.


2%˜

7.2.12 Printer interface board (80606-00019)


Replacement part: 80606-00019
1 . Remove the measuring unit rear cover ASSY (30601-6000) (see 6.5 [p65]).
2 . Unscrew PT3 × 8 (n = 2) to remove the
printer interface board (80606-00019)
together with the attachment plate (30601-
M609).
3 . Unscrew BS3 × 6 (n = 4) to remove the 
attachment plate (30601-M609). 
4 . Replace the printer interface board 26˜
(80606-00019).
5 . Reassemble the parts in the reverse order. 
$5˜



81 Return to previous page


XAR17ARDA001G/E

7.3 Electrical Components


7.3.1 R/L sensor (30602-EA20)
Replacement part: 30602-EA20
* Applies only to the AR-330A, AR-360A, ARK-530A, and ARK-560.
1 . Remove the covers (see Section 6 [p63]).
2 . Unscrew BS3 × 6 to replace the R/L sen-
sor (30602-EA20).
3 . Reassemble the parts in the reverse order.

'#
$5˜

7.3.2 F/B sensor (30602-EA21)


Replacement part: 30602-EA21
* Applies only to the AR-330A, AR-360A, ARK-530A, and ARK-560A.
1 . Unscrew BS3 × 6 to replace the F/B sen-
sor (30602-EA21).
2 . Reassemble the parts in the reverse order.

'# $5˜

82 Return to previous page


XAR17ARDA001G/E

7.3.3 U/D sensor (30601-EA22)


Replacement part: 30601-EA22
1 . Remove the covers (see Section 6 [p63]).
2 . Unscrew BS3 × 6 to replace the U/D sen-
sor (30601-EA22).
3 . Reassemble the parts in the reverse order.

$5˜

'#

83 Return to previous page


XAR17ARDA001G/E

7.3.4 R/L motor (30602-EA26)


Replacement part: 30602-EA26
* Applies only to the AR-330A, AR-360A, ARK-530A, and ARK-560A.
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P102 (J2) on the main board
(30601-BA01), and P201 (J1), P205 (J5),
P208 (J8), P211 (J11), and the ground
wire on the tracking board (30601-BA02).
3 . Unscrew BS4 × 8 (n = 6) to remove the '
tracking horizontal ASSY (30601-2500)
and measuring ASSY (30601-9E00)
together with the attachment plate (30601-

M226).



$5˜

4 . Unscrew BS4 × 6 (n = 2) to remove the R/


L motor (30602-EA26) together with the
attachment plate (30602-M261).
5 . Unscrew HH4 × 4 (n = 2) to remove the
holder bearing (32164-M216).
6 . Unscrew BS3 × 6 (n = 3) to remove the
attachment plate (30602-M261).
7 . Replace the R/L motor (30602-EA26).
8 . Reassemble the parts in the reverse order. 
9 . Perform AUTO TRC ADJUST (see 8.11 
**˜
[p168]). $5˜




'#

$5˜

84 Return to previous page


XAR17ARDA001G/E

7.3.5 F/B motor (30602-EA27)


Replacement part: 30602-EA27
* Applies only to the AR-330A, AR-360A, ARK-530A, and ARK-560A.
1 . Remove the covers (see Section 6 [p63]).
2 . Remove the measuring unit together with the tracking horizontal ASSY (30601-2500) (see
7.3.4 [p84]).
3 . Unscrew BS4 × 6 (n = 2) to remove the R/
L motor (30602-EA27) together with the
attachment plate (30602-M261).
4 . Unscrew HH4 × 4 (n = 2) to remove the
holder bearing (32164-M216).
5 . Unscrew BS3 × 6 (n = 3) to remove the
attachment plate (30602-M261).

6 . Replace the F/B motor (30602-EA27). 
**˜
7 . Reassemble the parts in the reverse order. 
$5˜
8 . Perform AUTO TRC ADJUST (see 8.11
[p168]).


'#


$5˜

7.3.6 U/D motor (30601-EA28)


Replacement part: 30601-EA28
1 . Remove the covers (see Section 6 [p63]).
2 . Unscrew AS4 × 6 (n = 2) to remove the U/
D motor (30601-EA28) together with the '#
bracket (30601-M211).
3 . Unscrew HH5 × 6 to remove the pulley 
(32907-M106).
4 . Unscrew FC3 × 6 (n = 3) to replace the U/ (%˜


D motor (30601-EA28).
5 . Reassemble the parts in the reverse order. 
 #5˜
6 . Adjust the position of the bracket (30601-
M211) by lightly pressing the belt (32156- **˜
M098) so that the belt bends about 3 - 4
mm.
7 . Adjust the tension of the belt (32156-
M098) so that the tension after the
replacement becomes almost the same as
before.
8 . Perform AUTO TRC ADJUST (see 8.11 [p168]).

85 Return to previous page


XAR17ARDA001G/E

7.3.7 Origin sensor (30601-EA36)


Replacement part: 30601-EA36
1 . Remove the covers (see Section 6 [p63]).
2 . Unscrew CK 2 × 4 (n = 2) to replace the

origin sensor (30601-EA36). %-˜
'#
3 . Reassemble the parts in the reverse order.
4 . Perform AR CALIBRATION (see 8.7
[p130]).

7.3.8 CL motor (30703-EA37)


Replacement part: 30703-EA37
* Applies only to the AR-360A and ARK-560A.
1 . Remove the covers (see Section 6 [p63]).
2 . Move the subjective focus base (30703-3300) to the operator’s side.
3 . Unscrew SB2 × 6 (n = 4) to remove the CL
motor (30703-EA37) together with the
spur gears (30703-M394 [n = 2]). 
5$˜
4 . Loosen FG2 × 4 (n = 4), and remove the
spur gears (30703-M394 [n = 2]) to
replace the CL motor (30703-EA37).
5 . Reassemble the parts in the reverse order.

6 . Adjust the subjective cylinder ASSY (see ()˜
8.5.2.1 [p125]). 


5$˜
'#

86 Return to previous page


XAR17ARDA001G/E

7.3.9 CL sensor (30703-EA38)


Replacement part: 30703-EA38
* Applies only to the AR-360A and ARK-560A.
1 . Remove the subjective cylinder ASSY (30703-3700) (see 7.1.6 [p72]).
2 . Unscrew CK2 × 3 (n = 3) to remove the
sensor retainer (30703-M395), then
remove the infrared LED from the CL sen-
sor (30703-EA38).

'#

 
%-˜

3 . Unscrew CK2 × 3 (n = 3) to remove the


sensor retainer (30703-M395), then
remove the phototransistor from the CL
sensor (30703-EA38).
4 . Reassemble the parts in the reverse order.
5 . Adjust the subjective cylinder ASSY (see '#
8.5.2.1 [p125]).

 
%-˜

87 Return to previous page


XAR17ARDA001G/E

7.3.10 Focus LED(30601-EA42)


Replacement part: 30601-EA42
1 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
2 . Unscrew CK2 × 3 (n = 4) to replace the
focus LEDs (30601-EA42 [n = 2]).
3 . Reassemble the parts in the reverse order.
4 . Perform ALIGNMENT CALIBRATION (see
8.6 [p129]).
5 . Perform KM CALIBRATION (see 8.8
[p156]).
* The KM CALIBRATION applies only to 
%-˜
the ARK-510A, ARK-530A, and ARK-
560.


'#

%-˜

7.3.11 KM6 LED (30601-EA44)


Replacement part: 30601-EA44
1 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
2 . Unscrew CK2 × 4 (n = 8) to replace KM6
LEDs (30601-EA44 [n = 4]).
3 . Reassemble the parts in the reverse order.
4 . Perform ALIGNMENT CALIBRATION (see
8.6 [p129]).
5 . Perform the KM CALIBRATION (see 8.8
[p156]).
* The KM CALIBRATION applies only to
the ARK-510A, ARK-530A, and ARK-
560A.

'#


%-˜

88 Return to previous page


XAR17ARDA001G/E

7.3.12 Chinrest (30601-EA53)


Replacement part: 30601-EA53
1 . Unscrew BS 3 × 6 (n = 2) to remove the lid (30601-M003).
2 . Disconnect P004 (J004).
3 . Unscrew SB3 × 10 to remove the chinrest
ASSY (30601-5000).

5$˜



$5˜

4 . Unscrew SB3 × 10 (n = 3) to remove the
chinrest cover (32105-M041).


5$˜



5 . Unscrew PC3 × 6.
6 . Unscrew HH3 × 5 (n = 3), and move the
bearing (32105-M048) upward. 2%˜
7 . Unscrew HH3 × 5 and CK 2 × 5 (n = 2) to
remove the motor brace (32105-M080).
8 . Unscrew SB3 × 6 (n = 2) and PC2 × 5 (n =
2) to replace the chinrest (30601-EA53).
9 . Reassemble the parts in the reverse order.


5$˜
 
**˜

**˜ %-˜ '#



 2%˜

89 Return to previous page


XAR17ARDA001G/E

7.3.13 PD LED (30601-EA55)


Replacement part: 30601-EA55
1 . Remove the bottom plate (30601-M002) (see 6.7 [p66]).
2 . Disconnect P505 (J5) on the base board (30601-BA05).
3 . Unscrew BS3 × 6 to replace PD LED
(30601-EA55).
4 . Reassemble the parts in the reverse order. $5˜
5 . Perform PD CALIBRATION (see 8.9
[p163]). '#

90 Return to previous page


XAR17ARDA001G/E

7.3.14 Chart LED (30601-EA76)


Replacement part: 30601-EA76
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P1406 (J6) on the sensor carry board (30601-BA14).
3 . Replace the chart LED (30601-EA76) with
a pair of tweezers.
4 . Reassemble the parts in the reverse order.
'#

5 . Use the steps below to adjust the LED intensity when the previous chart LED (30601-
EA76A) of the devices with serial numbers earlier than those shown in the following table is
replaced with the new one (30601-EA76B).
1 ) Serial numbers of post-design change devices
Countries other than
Model USA China
USA and China
AR-310A 120001— 150621— 136114—
AR-330A 220001— 250005— 230659—
AR-360A 820001— 850001— 830312—
ARK-510A 320342— 350157— 334477—
ARK-530A 421330— 450004— 431698—
ARK-560A 920348— 950001— 930622—

2 ) Identify the chart LED as follows:


a . The cable color of the previous
chart LED (30601-EA76A) is
black and white.
b . The cable color of the new
chart LED (30601-EA76B) is
black and red.

3 ) Enter MENU mode, and select 7. LED ADJUST MODE (see 8.1.1 [p103]).

4 ) Press the up or down but- /('$'-867


ton, and select the bottommost
numeric value from the first column.

5 ) Press the left - or right + but-


ton to change the numeric value
from 170 to 135.
6 ) Press the SAVE button to save the 
6/'$872
setting. 
* Do not change the other numeric val- 6$9( 

ues. 


91 Return to previous page


XAR17ARDA001G/E

7.3.15 Chart (30703-EA74)


Replacement part: 30703-EA74
* Applies only to the AR-360A and ARK-560A.
1 . Remove the chart disk ASSY (30703-3800) (see 7.1.7 [p72]).
2 . Unscrew CK2 × 4 (n = 2) to remove the
disk cover (30703-M404). 
3 . Unscrew FK2 × 3 (n = 2) to remove the

holder (30703-M408) and sensor stopper (-˜
(30703-M405). Then, remove the pho- '#
totransistor from the chart (30703-EA74).
* The wires from the phototransistor of
the chart (30703-EA74) are gray and
purple.
 
%-˜

4 . Loosen HH3 × 3 (n = 2) to remove the
chart disk (30703-G041) together with the )
disk spindle (30703-M403). 
5 . Unscrew SB2 × 6 (n = 2) to remove the
geared stepping motor from the chart
(30703-EA74). 
6 . Unscrew CK2 × 4 (n = 2) to remove the
LED retainer (30703-M406).  **˜

%-˜
7 . Remove the infrared LED from the chart
(30703-EA74) to replace the chart.
* The wires from the infrared LED of the
chart (30703-EA74) are green and 
'#
5$˜
blue.
8 . Reassemble the parts in the reverse order.
* Assemble the disk spindle (30703-
M403) and geared stepping motor of
the chart (30703-EA74) as shown to
the right.
9 . Adjust the chart (see 8.5 [p124]).
10. Adjust the subjective cylinder ASSY (see
8.5.2.1 [p125]).

 OO '#

92 Return to previous page


XAR17ARDA001G/E

7.3.16 B/W CCD camera (30601-E001)


7.3.16.1 Measurement CCD camera
Replacement part: 30601-E001
1 . Remove the covers (see Section 6 [p63]).
2 . Remove the cable clamp (80423-20153).
3 . Disconnect the connector on the B/W  '
CCD camera (30601-E001). 
4 . Unscrew CK2 × 3 (n = 4) to remove the
CCD cover (30601-M345).
5 . Unscrew CK2.6 × 6 (n = 2) to replace the
B/W CCD camera (30601-E001). 
%-˜
6 . Reassemble the parts in the reverse order. %-˜


7 . Adjust the measurement CCD camera 


(see 8.3.2 [p121]).
8 . Perform AR CALIBRATION (see 8.7 [p130]).

7.3.16.2 Observation CCD camera


Replacement part: 30601-E001
1 . Remove the covers (see Section 6 [p63]).
2 . Remove the cable clamp (80423-20153).
3 . Disconnect the connector on the B/W
CCD camera (30601-E001).
4 . Unscrew CK2 × 3 (n = 4) to remove the
CCD cover (30601-M345).
5 . Unscrew CK2.6 × 6 (n = 2) to replace the
B/W CCD camera (30601-E001).
6 . Reassemble the parts in the reverse order. '
7 . Adjust the observation CCD camera (see
8.3.1 [p117]).
8 . Perform ALIGNMENT CALIBRATION (see
8.6 [p129]). 
%-˜
9 . Perform KM CALIBRATION (see 8.8 
[p156]).   
%-˜
* The KM CALIBRATION applies only to
the ARK-510A, ARK-530A, and ARK-560A.
10. Perform CS/PS calibration (see 8.10 [p167]).
* The CS/PS calibration applies only to the ARK-510A, ARK-530A, and ARK-560A.

93 Return to previous page


XAR17ARDA001G/E

7.3.17 Color LCD (30601-E005)


Replacement part: 30601-E005
1 . Remove the LCD front cover (30601-M715) (see 6.10 [p67]).
2 . Disconnect P001 (CN1) on the color LCD (30601-E005).
3 . Disconnect P002 (CN2) on the DC-AC inverter (18535-E010).
4 . Unscrew PC3 × 10 and N3, and disconnect the F.GND cable (30601-EA83).
5 . Unscrew PT3 × 10 (n = 4) to replace the
color LCD (30601-E005). '
6 . Reassemble the parts in the reverse order.
7 . Perform CURSOR BOARD (see 8.14
[p172]).
2%˜

0
'#


26˜

7.3.18 Brushless motor (30601-E011)


Replacement part: 30601-E011
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Disconnect the connector of the brushless motor (30601-E011).
3 . Unscrew BS3 × 10 (n = 2) to remove the
brushless motor (30601-E011) together '
with the motor attachment plate (30601- 
M389). 
4 . Unscrew FG2 × 3 (n = 2) to remove the
wheel (30601-M368).  
%-˜ 
5 . Unscrew CK2.6 × 4 (n = 2) to replace the $5˜
brushless motor (30601-E011).

6 . Reassemble the parts in the reverse order. ()˜

94 Return to previous page


XAR17ARDA001G/E

7.3.19 Sensor motor (30601-E035)


Replacement part: 30601-E035
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Unscrew BS3 × 6 (n = 2) to remove the
pulley 2 ASSY (30601-3120).  
$5˜
3 . Remove the wire (30601-M307) from the
pulley (30601-M359) of the motor ASSY
(30601-3600).



$5˜




4 . Unscrew BS3 × 6 (n = 3) to remove the


motor ASSY (30601-3600).



$5˜

5 . Unscrew HH3 × 4 (n = 2) to remove the



pulley (30601-M359). **˜
6 . Unscrew BS3 × 6 (n = 3) to replace the
sensor motor (30601-E035).
7 . Reassemble the parts in the reverse order.
* Be sure to wind one turn of wire
(30601-M307) around the pulley
 '
(30601-M359) of the motor ASSY 
$5˜
(30601-3600).

* Loosen BS3 × 6 (n = 2) fastening the
pulley 2 ASSY (30601-3120).

95 Return to previous page


XAR17ARDA001G/E

8 . Attach the tension jig (30670-2100) to the


measuring unit ASSY (30601-3000). 
9 . Turn FC4 × 10 of the tension jig (30670-
2100) to adjust the tension of the wire
(30601-M307) so that the focus base
ASSY (30601-3300) moves smoothly.
10. Fasten the pulley 2 ASSY (30601-3120)
with BS3 × 6 (n = 2). 
11 . Perform AR CALIBRATION (see 8.7
[p130]).

(%˜


$5˜

7.3.20 Switching power supply (80602-00102)


Replacement part: 80602-00102
1 . Remove the bottom plate (30601-M002) (see 6.7 [p66]).
2 . Unscrew BS3 × 8 (n = 4) to replace the
switching power supply (80602-00102).  $5˜


3 . Reassemble the parts in the reverse order. 

96 Return to previous page


XAR17ARDA001G/E

7.3.21 Printer (80606-00018)


Replacement part: 80606-00018
1 . Remove the measuring unit rear cover (30601-6000) (see 6.5 [p65]).
2 . Unscrew PT3 × 8 (n = 2) to remove the
printer cover (30601-M711).


26˜


3 . Remove the printer door cover (30601-


M710).
4 . Unscrew FT2 × 6 (n = 2) to remove the
printer (80606-00018) from the printer
door cover (30601-M710). (6˜







5 . Unscrew PT2 × 5 (n = 2) to replace the


printer (80606-00018).
6 . Reassemble the parts in the reverse order.





26˜

97 Return to previous page


XAR17ARDA001G/E

7.4 Mechanical Components


7.4.1 R/L shaft (30601-M201)
Replacement part: 30601-M201
1 . Remove the covers (see Section 6 [p63]).
2 . Disconnect P201 (J1) on the tracking board (30601-BA02) and GND cable of the base
cable (30601-EA52).
3 . Unscrew SB6 × 12 (n = 2) to remove the chin-
rest ASSY (30601-5000) together with the base
ASSY (30601-1000).


5$˜



4 . Replace the R/L shaft (30601-M201).


5 . Reassemble the parts in the reverse order.


98 Return to previous page


XAR17ARDA001G/E

7.4.2 F/B shaft (30601-M103)


Replacement part: 30601-M103
1 . Remove the R/L shaft (30601-M201) (see 7.4.1 [p98]).
2 . Unscrew HH5 × 6 (n = 4) to replace the F/
B shafts (30601-M103 [n = 2]).
3 . Reassemble the parts in the reverse order.
* Reassemble the parts so that the side- 
to-side and forward-and-backward
movements become smooth.


**˜




7.4.3 Sliding plate (32105-M102)


Replacement part: 32105-M102
1 . Unscrew SB6 × 12 (n = 2) to remove the chinrest ASSY (30601-5000) together with the
base ASSY (30601-1000) (see 7.4.1 [p98]).
2 . Unscrew CK2.6 × 4 (n = 4) to replace the
sliding plate (32105-M102). %-˜


3 . Reassemble the parts in the reverse order.

99 Return to previous page


XAR17ARDA001G/E

7.4.4 Wire (30601-M307)


Replacement part: 30601-M307
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Unscrew BS3 × 6 (n = 2) to remove the
pulley 2 ASSY (30601-3120).  
$5˜
3 . Remove the wire (30601-M307) from the
pulley (30601-M359) of the motor ASSY
(30601-3600).
4 . Reassemble the parts in the reverse order.
* Be sure to wind one turn of wire
(30601-M307) around the pulley
(30601-M359) of the motor ASSY 
(30601-3600). $5˜





5 . Attach the tension jig (30670-2100) to the


measuring unit ASSY (30601-3000). 
6 . Turn FC4 × 10 of the tension jig (30670-
2100) to adjust the tension of the wire
(30601-M307) so that the focus base
ASSY (30601-3300) moves smoothly.
7 . Fasten the pulley 2 ASSY (30601-3120)
with BS3 × 6 (n = 2). 
8 . Perform AR CALIBRATION (see 8.7
[p130]).

(%˜


$5˜

100 Return to previous page


XAR17ARDA001G/E

7.4.5 Linear bearing


7.4.5.1 Linear bearing for F/B shafts (82001-LM028)
Replacement part: 82001-LM028
1 . Unscrew HH5 × 6 (n = 4) to remove the F/B shafts (30601-M103 [n = 2]) and sliding plate
(30601-M102) (see 7.4.2 [p99]).
2 . Unscrew TC3 × 5 (n = 4) to replace the linear 
bearings for F/B shafts (82001-LM028 [n = 4]). 
3 . Reassemble the parts in the reverse order. ./
* Reassemble the parts so that the side-to-side
and forward-and-backward movements
become smooth.


6%˜

7.4.5.2 Linear bearing for R/L shaft (82001-LM097)


Replacement part: 82001-LM097
1 . Remove the R/L shaft (see 7.4.1 [p98]).
2 . Unscrew TC3 × 5 (n = 2) to replace the linear 
bearings for R/L shaft (82001-LM097 [n = 2]). 
3 . Reassemble the parts in the reverse order. ./
* Reassemble the parts so that the side-to- 
side and forward-and-backward move-
ments become smooth.


6%˜

101 Return to previous page


XAR17ARDA001G/E

This page is intentionally left blank.

102 Return to previous page


XAR17ARDA001G/E

8 ADJUSTMENT

8.1 Entering Adjustment Mode


8.1.1 MENU mode
1 . Turn on the power switch while pressing the L-1 and R-1 function buttons.
* Continue pressing the function buttons until a beep sounds.
2 . The device starts, and the password entry screen appears.
3 . Press the function buttons in the order of
R-3 to R-1 to L-2 to enter the password.

. 4
. 4
. 4
. 4

4 . The MENU screen is displayed.


/'07
* The display of the MENU screen dif- 1$5'48#6+10%#/'4#
fers depending on the software ver- %*#46215+6+10
4+0)/#5-215+6+10
sion. 4'/18+0)+06'40#.4'(
5 . When the earlier versions below are #4%#/'4#
#2'4674'215+6+10
upgraded, checks may not be entered .'&#&,756/1&'
#.+)0/'06%#.+$4#6+10
properly. 2#4#/'6'45'66+0)
* There are no problems in operation or %*#46%#.+$4#6+10
64%#&,756/1&'
adjustment. #761%#.+$4#6+10
1 ) AR: Ver. 1.07 or earlier %#.+$4#6+10%*'%-
/1&'.';'%*'%-
2 ) ARK: Ver. 1.05 or earlier 2&#&,756/1&'
%525#&,756

6 . Press the up or down button to /'07


''241/'4414&+52
select each item. '4414*+5614;
* When changing pages, use the page 52'%+#.2#4#/#'6'4

up or page down button.

7 . Press the execute button to enter


each adjustment mode.
* 16. CS PS ADJUST only applies to
the ARK-510A, ARK-530A, and ARK-
560A.

103 Return to previous page


XAR17ARDA001G/E

8.1.2 CALIB MODE SELECT (for AR)


1 . Turn on the power switch while pressing
the L-1 and R-2 function buttons. &$/,%02'(6(/(&7
* Continue pressing the function but- 
&+$57
tons until a beep sounds. :,1'2:&+(&.
2 . The CALIB MODE SELECT screen is dis- $5
$5$;,6
played. $5&255(17
&85625%2$5'
3 . Press the up or down button to
select each item.

4 . Press the execute button to enter


each adjustment mode.

8.1.3 ALIGNMENT CALIBRATION


1 . Turn on the power switch while pressing
the L-1 and R-4 function buttons. ᵏ 
* Continue pressing the function but- #.+)0/'06%#.+$4#6+10
tons until a beep sounds.
2 . Enter ALIGNMENT CALIBRATION mode.

24'5556#46$76610

8.1.4 Measurement mode


1 . Turn on the power switch while pressing the L-3 and R-1 function buttons.
* Continue pressing the function buttons until a beep sounds.
2 . Perform alignment of the model eye (CL)
(32961-0700), then measure the model ᵏ &</ 㹊
eye.
* The screen as shown to the right is for
the AR-330. The screen differs for
other models.
3 . Confirm that the numeric values displayed
on the screen conform to those described
on the model eye (CL) (32961-0700) 5  / 
4 . After confirmation, turn off the power 66
switch. &&
$$
.

104 Return to previous page


XAR17ARDA001G/E

8.1.5 Parameter settings


1 . Display the PARAMETER SETTING screen following the steps below.
1 ) For the AR-310A, AR-330A, and AR-360A
* The icons on the AR-360A PARAMETER SETTING screen may be slightly differ-
ent.
a . Press the ring image enlarge-
ment/parameter button for ᵏ %;. 㧸
one second or more to display
the PARAMETER SETTING
screen.

4
 .

5 5 
% % 
##

2 ) For the ARK 510A, ARK-530A, and ARK-560A


a . Press the page button in the
order of to to , and
ᵏ %;. 㧸
select page 3.

4
 -
5 4
% 4
#

b . Press the parameter but- ᵏ %;. 㧸


ton for one second or more to
display the PARAMETER
SETTING screen.
<<
<

4
 -
5 4
% 4
#

105 Return to previous page


XAR17ARDA001G/E

2 . Press the up or down button to 2#4#/'6'45'66+0) .


select the parameter to be changed, and #4
-/
press the execute button. %1/2#4'
24+06
* Parameters vary according to each 24+06
model. 24+06
(70%6+10
%1//0+%#6+10
3 . Press the exit button to return to the 57$,'%6
measurement screen. 4'#&'4
0'6914-
%.1%-5'6
%1//'065'6
%..+565'6

8.1.6 CALIB MODE SELECT (for KM)


* The CALIB MODE SELECT screen for the KM can only be displayed in the ARK-510A,
ARK-530A, and ARK-560A.
1 . Turn on the power switch while pressing
the L-1 and R-3 function buttons. %#.+$/1&'5'.'%6
* Continue pressing the function but-
-/%#.+$
tons until a beep sounds. -/%#.+$%*'%-
2 . The CALIB MODE SELECT screen is dis- -/#:+5%*'%-
%7514$1#4&
played.

3 . Press the up or down button to


select each item.

4 . Press the execute button to enter


each adjustment mode.

8.1.7 EEPROM ERROR DISP


1 . Enter MENU mode, and select 17. EEPROM ERROR DISP (see 8.1.1 [p103]).
2 . Confirm that each item displays no error.
''241/'4414&+52
* When an error occurs, the back-
ground of the erroneous item turns 
(75'
pink. &#
.'&
64%8#.#4-A%*#46
#.+)0
4'6+%.'
5*+2/'06#4A%#.+$
#4A%144'
-/A%#.+$
-/A%144'
2&A%#.+$
%5A%#.+$
25A%#.+$

106 Return to previous page


XAR17ARDA001G/E

8.2 Jig Attachment Procedures


8.2.1 Service optical adjustment jig (30670-5300)
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Connect the flexible flat cable (30601-E002) of the measuring unit rear cover ASSY (30601-
6000) to J4 on the relay board (30670-BA01).
4 . Connect the flexible flat cable (30670-E001) to J1 on the relay board (30670-BA01).
5 . Connect the flexible flat cable (30670-E001) to J11 on the driver board (30601-BA03).

' $#


, , ,
#4&,$#':'$1#4&

,
,

,

'

, , , , , ,


,

,
,

, #4$# ,


&4+8'4$1#4&

, , , ,


,

$#

107 Return to previous page


XAR17ARDA001G/E

6 . Connect the connector of the EA39 extension cable (32557-EA11) to J1 of the mire ASSY.
7 . Connect the connector of the EA39 extension cable (32557-EA11) to J9 on the driver board
(30601-BA03).

,

,

,
,

,
, ,

'#

'# , , , , ,


,
,

,
$# ,

, #4$# ,


&4+8'4$1#4&

, , , ,


,

'#
6Q
,  6Q

'#
6Q
,  6Q$#
, 

108 Return to previous page


XAR17ARDA001G/E

8.2.1.1 Joint jig (30670-1100)


1 . Attach the service optical adjustment jig (30670-5300) (see 8.2.1 [p107]).
2 . Attach the joint jig (30670-1100) to the
measuring ASSY (30601-9E00) with FC3
× 15 (n = 2). '




(%˜

8.2.1.2 Reticle illumination jig (30670-1200)


1 . Attach the joint jig (30670-1100) to the measurement ASSY 30601-9E00 (see 8.2.1.1
[p109]).
2 . Align the key (31401-M357) of the reticle
illumination jig (30670-1200) to the notch
of the joint jig (30670-1100), and fasten it 
with the stepped knurled screw (82053- #$
A308B).






3 . Connect the connector of relay cable 3


,

(30670-EA06) to the connector of the reti- ,

cle illumination jig (30670-1200).



,
,

4 . Connect the connector of relay cable 3 ,


, ,

'#
(30670-EA06) to J3 of the mire ASSY
(30601-4000). '#
6Q
, 6Q

109 Return to previous page


XAR17ARDA001G/E

8.2.2 0-D model eye 2 (32177-0618)


1 . Attach the joint jig (30670-1100) to the measurement ASSY (30601-9E00) (see 8.2.1.1
[p109]).
2 . Attach 0-D model eye 2 (32177-0618) to
the joint jig (30670-1100), and fasten it
with the stepped knurled screw (82053- 
#$
A308B).





8.2.3 Dioptric tester (99104-G020)


1 . Attach the joint jig (30670-1100) to the measurement ASSY (30601-9E00) (see 8.2.1.1
[p109]).
2 . Attach any of the following jigs to the joint
jig (30670-1100), and fasten it with the 
stepped knurled screw (82053-A308B). #$
1 ) When the diopter is 0 D: Holder 
(30670-M031) 
2 ) When the diopter is -12 D: Chart 
lens M (30670-1800)
3 ) When the diopter is +12 D: Chart
lens P (30670-1900) 

3 . Attach the dioptric tester (99104-G020).






)

110 Return to previous page


XAR17ARDA001G/E

8.2.4 Simple calibration model eye (32107-6200)


1 . Attach the simple calibration model eye
(32107-6200) with its notches aligned to 
the pins (34131-M112 [n = 2]) on the chin-
rest ASSY (15601-1500).




8.2.5 Chinrest attachment jig (32107-1100)


1 . Attach the chinrest attachment jig (32107-
1100) to the chinrest ASSY (30601-5000).
2 . Fasten the chinrest attachment jig (32107-
1100) with the fastening pins (32177-M502
and 32177-M508) so the support board
(32177-M505) becomes parallel to the
measuring window.







111 Return to previous page


XAR17ARDA001G/E

8.2.6 22-model eye jig (32171-0600)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the 22-model eye jig (32171-0600)
to the chinrest attachment jig (32107-

1100).
* Fasten the jig with the fastening pin
(32177-M513).





8.2.7 Cylinder axis adjustment jig (32177-1500)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the cylinder axis adjustment jig
(32177-1500) to the chinrest attachment
jig (32107-1100).
* Fasten the jig with the fastening pin 
(32177-M513).





112 Return to previous page


XAR17ARDA001G/E

8.2.8 Chart test jig (32145-2100)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the chart test jig (32145-2100) to
the chinrest attachment jig (32107-1100).
* Fasten the jig with the fastening pin
(32177-M513). 





3 . Turn the turret plate (32145-M441) of the


chart test jig to select the model eye con-
vex lens (32527-G001).
4 . Perform alignment of the model eye con- )
vex lens (32527-G001) so that the focus- 
ing indicator displays the optimum
condition.
5 . Fasten the measuring unit with the locking
lever.
* Fasten the unit while the focusing indi- )
cator is in the optimum condition. )

113 Return to previous page


XAR17ARDA001G/E

8.2.9 Chart test jig (30670-2300)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the chart test jig (30670-2300) to
the chinrest attachment jig (32107-1100).
* Fasten the jig with the fastening pin 
(32177-M513).
3 . Display the CALIB MODE SELECT screen
(for AR), then select CHART (see 8.1.2
[p104]).
4 . Make the chart test jig (30670-2300) level
by aligning the marking-off line to the tar-
get using the dovetail goniostage (30670-

M075).

5 . Select STAR CALIB with the up or

down button, then press the execute

button.
6 . Turn the turret plate (30670-M071) of the
chart test jig (30670-2300) to select the
model eye convex lens (32527-G001).
7 . Perform alignment of the model eye con- )
)
vex lens (32527-G001) so that the focus-
ing indicator displays the optimum 
condition.
8 . Fasten the measuring unit with the locking
lever. 
9 . Fasten the unit while the focusing indica- )
tor is in the optimum condition.
* Maintain the alignment of the convex ) Marking-off line
lens during adjustment. ⟌ᦠ߈✢

114 Return to previous page


XAR17ARDA001G/E

8.2.10 Dioptric tester jig (30670-6000)


1 . Remove the measuring unit left cover (30601-M707) (see 6.1 [p63]).
2 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
3 . Attach the dioptric tester jig (30670-6000)
to the chinrest attachment jig (32107- 
1100).

* Fasten the jig with the fastening pin
(32177-M513).



4 . Disconnect P104 (J4) on the applicable


main board below, then connect the con-
nector of the diopter tester jig (30670-
6000) to P104 (J4) on the main board.
1 ) AR-310A: 30601-BA01P
2 ) AR-330A: 30602-BA01P 2
,
3 ) AR-360A: 30603-BA01P
4 ) ARK-510A: 30701-BA01P $#
5 ) ARK-530A: 30702-BA01P
6 ) ARK-560A: 30703-BA01P
5 . Enter MENU mode, and select 10. %*#46%#.+$4#6+10
CHART CALIBRATION (see 8.1.1 [p103]).

* The screen to the right differs from %*#46%#.+$
%*#46%*'%-
those in the AR-360A and ARK-560A.
6 . Make the chart test jig (30670-2300) level
by aligning the marking-off line to the tar-
get using the dovetail goniostage (30670-
M075).

7 . Turn the turret plate (30670-M071) of the


dioptric tester jig ASSY (30670-6000) to
select the model eye convex lens (32527-
G001).
8 . Perform alignment of the model eye con- )
vex lens (32527-G001) so that the focus- )
ing indicator displays the optimum 
condition. 
9 . Fasten the measuring unit with the locking ) Marking-off line
⟌ᦠ߈✢
lever.
* Fasten the unit while the focusing indi- )
cator is in the optimum condition.

115 Return to previous page


XAR17ARDA001G/E

8.2.11 KM calibration jig (30670-0300)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the KM calibration jig (30670-0300)
to the chinrest attachment jig (32107- 
1100).
* Fasten the jig with the fastening pin
(32177-M513).





8.2.12 Aspheric model eye (32177-1400)


1 . Attach the chinrest attachment jig (32107-1100) (see 8.2.5 [p111]).
2 . Attach the aspheric model eye (32177-
1400).

* Fasten the jig with the fastening pin
(32177-M513).





8.2.13 PD check jig (32545-1002)


1 . Attach the PD check jig (32545-1002) to
the chinrest. 

116 Return to previous page


XAR17ARDA001G/E

8.3 CCD Camera Adjustment


8.3.1 Observation CCD camera
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Remove the motor ASSY (30601-3600) (see 7.3.19 [p95]).
4 . Adjust the focus of the observation CCD camera as follows:
1 ) Attach the reticle illumination jig (30670-1200) to the joint jig (30670-1100) (see
8.2.1.2 [p109]).
2 ) Enter MENU mode, and select
)2&86
1.OBSERVATION CAMERA (see
8.1.1 [p103]). 581
3 ) Press the function button to display
RUN. 1
* Adjust the brightness of the reti-
cle illumination jig with the up
or down button so that
adjustment can be performed 
easily.

Reticle
ࣞࢳࢡࣝ༑Ꮠ⥺

4 ) Use the steps below to remove the reticle illumination jig (30670-1200) from the
devices with serial numbers earlier than those shown in the following table.
USA Countries other than USA
(S/N) (S/N)
AR-310A 120001 134438
AR-330A 220001 230544
AR-360A 820001 830108
ARK-510A 320182 333207
ARK-530A 420905 431269
ARK-560A 920153 930260

a . Loosen HH3 × 3 (n = 2).


b . Turn the adjustment ring
(30601-M355) with a preci-
sion screwdriver to focus the
reticle of the reticle illumination
jig (30670-1200).
c . Tighten HH3 × 3 (n = 2).
d . Remove the reticle illumination
jig (30670-1200) (see 8.2.1.2
[p109]).


**˜ 

117 Return to previous page


XAR17ARDA001G/E

5 ) Use the steps below to remove the reticle illumination jig (30670-1200) from the
devices with serial numbers shown in the following table and later.
USA Countries other than USA
(S/N) (S/N)
AR-310A 120001 134438
AR-330A 220001 230544
AR-360A 820001 830108
ARK-510A 320182 333207
ARK-530A 420905 431269
ARK-560A 920153 930260

1 ) Loosen HH3 × 3 (n = 2).


2 ) Move the camera ASSY (30601-
3340) back and forth so that the reti-
cle is properly focused.
3 ) Tighten HH3 × 3 (n = 2).
4 ) Remove the reticle illumination jig
(30670-1200) (see 8.2.1.2 [p109]).


++™ 
5 . Adjust the position of the observation CCD camera as follows:
1 ) Attach 0-D model eye 2 (32177-0618) to the joint jig (30670-1100) (see 8.2.2 [p110]).
0
2 ) Press the next button to display %%&215+6+10
1$5'4
the 1-2 CCD POSITION (OBSER.)
screen. 470
3 ) Press the function buttons to display
RUN and LED. .'& 0
* Adjust the brightness of the SLD
light with the up or down
button so that adjustment can 56#46595#8' 
be performed easily.


AR SLD light
#45.& శ 0-D
&model
ᮨဳ⌒ eye
2 SLD
5.&light brightness
శߩ᣿ࠆߐ

118 Return to previous page


XAR17ARDA001G/E

4 ) Adjust the SLD light using the



stepped knurled screws (82053- 
A308B [n = 4]) on 0-D model eye 2 #$
(32177-0618) so that the SLD light
forms a circle.



5 ) Loosen SB3 × 6 (n = 2) to adjust the
position of the camera ASSY
(30601-3340) so that the SLD light is
aligned to the centering box (  ) of
the observation CCD camera.


 5$˜

%%&215+6+10
1$5'4

470

.'& 0

56#46595#8' 


AR SLD light
#45.& శ

6 ) After the adjustment is complete, remove 0-D model eye 2 (32177-0618) (see 8.2.2
[p110]).
6 . Adjust the inclination of the observation CCD camera as follows:
1 ) Attach the reticle illumination jig (30670-1200) (see 8.2.1.2 [p109]).
0
2 ) Press the next button to display *14+<106#.&+4'%6+10
the 1-3 HORIZONAL DIRECTION
screen. 470
* Press the function buttons to
display RUN and LED, and .+0' 0
adjust the brightness of the reti-
cle illumination jig with the up
or down button so that 
adjustment can be performed 
easily.
3 ) Press the function buttons to display
RUN and LINE.

119 Return to previous page


XAR17ARDA001G/E

4 ) Loosen HH3 × 4 (n = 2) and adjust


the inclination of the camera ASSY
(30601-3340) so that the reticle hori- 
zontal line of the reticle illumination **˜

jig (30670-1200) becomes parallel to
those on the screen.
* Adjust the brightness of the hori-
zontal lines on the screen with
the up or down button so that adjustment can be performed easily.

5 ) Press the page down button to %%&215+6+10


1$5'4
display the 1-2 CCD POSITION
470
(OBSER.) screen.
6 ) Press the function buttons to display
RUN and LED. 537#4' 0
7 ) Confirm that the centering box (  )
of the observation CCD camera and
the SLD light are in alignment. If
56#46595#8' 
they are not, precisely align them as
follows: 
a . Press the function buttons on
the 1-2 CCD POSITION #45.&Center
AR SLD light శ of the
ⷰኤ↪observation CCD camera
%%& ࠞࡔ࡜ਛᔃ
(OBSER.) screen to display
RUN and SQUARE.

b . Perform adjustment with the up , down , right , or left button so


that the centering box (  ) of the observation CCD camera and the SLD light
are in alignment.
c . Press the start button to save the setting.
d . If the centering box (  ) of the observation CCD camera and the SLD light can-
not be precisely aligned, repeat adjustment starting from Step 5 again.

8 ) After the adjustment is complete, press the exit button and turn off power to the
device.
7 . Reassemble the removed units and covers.
8 . Perform ALIGNMENT CALIBRATION (see 8.6 [p129]).
9 . Perform KM calibration (see 8.8 [p156]).
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
10. Perform AUTO TRC ADJUST (see 8.11 [p168]).
11 . Perform CS/PS calibration (see 8.10 [p167]).
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.

120 Return to previous page


XAR17ARDA001G/E

8.3.2 Measurement CCD camera


1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Attach 0-D model eye 2 (32177-0618) (see 8.2.2 [p110]).
4 . Enter MENU mode, and select 5. AR
%%&215+6+10
#4 
CAMERA (see 8.1.1 [p103]).
5 . Press the function button to display RUN. 470

$4#-' 0



AR SLD
#45.& light
శ
6 . Loosen SB3 × 6 (n = 2).




5$˜

7 . Adjust the position of the camera ASSY


%%&215+6+10
#4 
(30601-3340) so that the center of the
SLD light is aligned to the scale (  and 470
alignment grid) displayed on the LCD.
$4#-' 0



AR SLDశ
#45.& light

0
8 . Press the next button to display the 5- %%&215+6+10
1$5'4
2 CCD POSITION (OBSER.) screen.
9 . Press the function buttons to display RUN 470
and LED.
10. Confirm that the centering box (  ) of the .'& 0
observation CCD camera and the SLD
light are in alignment.
11 . After the adjustment is complete, press
56#46595#8' 
the exit button and turn off power to the 
device.
12. Reassemble the removed units and cov- AR#45.&
SLD light
శ
ers.
13. Perform AR CALIBRATION (see 8.7 [p130]).

121 Return to previous page


XAR17ARDA001G/E

8.4 Phototransmitter Adjustment


1 . When this task can be performed in a darkened room, perform the internal reflection
removal (see 8.4.1 [p122]).
2 . If it is not possible, perform the AR SLD optical axis adjustment (see 8.4.2 [p123]).
8.4.1 Internal reflection removal
* Be sure to perform the internal reflection removal in a darkened room.
* Do not place any reflective object on the patient’s side.
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Enter MENU mode, and select 4. REMOVING INTERNAL REF (see 8.1.1 [p103])
0
3 . Press the next button to display the 4- 5.&215+6+10
2 SLD POSITION screen. 0)
4 . Press the function buttons to display RUN 470
and SLD.
5 . Move the focus base ASSY (30601-3300) 5.& 0
toward the operator’s side manually.
* The × marks appear where the inter-
nal reflections are detected, and NG is

displayed in the upper left of the
screen as shown to the right.

6 . Follow the steps below so that the internal


reflection marks disappear, and OK is dis-
played in the upper left of the screen.
1 ) Loosen HH3 × 3 to turn the SLD
**˜
holder (30601-M327) of the pho-
totransmitter ASSY. **˜
2 ) Loosen HH3 × 4 to turn the pho-
totransmitter ASSY (30601-3310). 
7 . Move the focus base ASSY (30601-3300)
toward the patient’s side, and confirm that 
the internal reflection marks (× marks) and
NG in the upper left of the screen disappear.
* If OK is not displayed in the upper left of the screen although the internal reflection
marks disappear within the movable range of the focus base ASSY (30601-3300),
repeat this task starting from Step 5 again.
* If a dead pixel occurs in the measurement CCD camera, a single × mark will appear
consistently.
* Should the occurrence of dead pixel in the measurement CCD cause two × marks or
more to appear consistently, replace the measurement CCD camera (30601-E001)
(see 7.3.16.1 [p93]).
8 . After the adjustment is complete, press the exit button.
9 . Adjust the measurement CCD camera (see 8.3.2 [p121]).
10. Reassemble the removed units and covers.
11 . Perform AR CALIBRATION (see 8.7 [p130]).

122 Return to previous page


XAR17ARDA001G/E

8.4.2 AR SLD optical axis adjustment


1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Attach 0-D model eye 2 (32177-0618) (see 8.2.2 [p110]).
4 . Enter MENU mode, and select 5. AR
%%&215+6+10
#4 
CAMERA (see 8.1.1 [p103]).
5 . Press the function button to display RUN. 470

$4#-' 0



AR SLD light
#45.& శ

6 . Follow the steps below so that the center


of the SLD light is aligned to the scale ( 
and alignment grid) displayed on the LCD.
1 ) Loosen HH3 × 3 to turn the SLD
**˜
holder (30601-M327) of the pho-
totransmitter ASSY. **˜
2 ) Loosen HH3 × 4 to turn the pho-
totransmitter ASSY (30601-3310). 



7 . After the adjustment is complete, press


%%&215+6+10
#4 
the exit button and turn off power to the
device. 470
8 . Reassemble the removed units and cov-
ers. $4#-' 0
9 . Perform AR CALIBRATION (see 8.7
[p130]).



#45.&
AR SLD light శ

123 Return to previous page


XAR17ARDA001G/E

8.5 Chart Adjustment


8.5.1 AR-310A, AR-330A, ARK-510A, ARK-530A
* Applies only to the AR-310A, AR-330A, ARK-510A, and ARK-530A.
* If the adjustment jigs are not available, return the device to the factory without perform-
ing adjustment.
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Attach any of the following jigs and dioptric tester (99104-G020) to the joint jig (30670-1100)
(see 8.2.3 [p110]).
1 ) When the diopter is 0 D: Holder (30670-M031)
2 ) When the diopter is -12 D: Chart lens M (30670-1800
3 ) When the diopter is +12 D: Chart lens M (30670-1900)
4 . Enter MENU mode, and select 2. CHART
%*#46215+6+10
POSITION (see 8.1.1 [p103])
5 . The 2-2 CHART POSITION screen is dis- 5612
played.
6 . Press the function button to display STOP. 0

7 . Press the function button to display ,


and turn on the chart LED.


8 . Follow the steps below to adjust the opti-


cal axis of the chart. 
1 ) Move the focus base ASSY (30601-
3300) to the position where the chart
can be seen through the dioptric tes-
ter.
2 ) Confirm that the chart does not shift
to the right or left at any position of -
12 D, 0 D, or +12 D.
3 ) If the chart shifts to the right or left, 
loosen BS3 × 6 (n = 2) to adjust the $5˜
position of the chart mirror ASSY
(30601-3270).
4 ) Follow the steps below so that the
center of the dioptric tester target
grid is inside the inner circle of the
clock dial chart (the balloon of the
scenery chart) at any position of -12
D, 0 D, and +12 D. **˜
a . Loosen HH3 × 3 to turn chart **˜

lens holder 2 (30601-M317) of
the chart lens 2 ASSY. 
b . Loosen HH3 × 4 to turn the
chart lens 2 ASSY (30601-3260).

9 . After the adjustment is complete, press the exit button and turn off power to the device.
10. Reassemble the removed units and covers.
11 . Perform CHART CALIBRATION (see 8.7.5 [p135]).

124 Return to previous page


XAR17ARDA001G/E

8.5.2 AR-360A, ARK-560A


8.5.2.1 Subjective cylinder ASSY adjustment
* The subjective cylinder ASSY must be adjusted for the emmetropic eyes.
* For astigmatic eyes, be sure to correct them before adjusting the subjective cylinder
ASSY.
1 . Enter MENU mode, and select 2. CHART POSITION (see 8.1.1 [p103]).
0 %*#46A24'%*'%-
2 . Press the next button to display 2-1
CHART_PRE-CHECK screen. 01+5'
3 . Press the function button to display
NOISE. 0



%
% 

4 . Follow the steps below to check the operating noise of the motor when the subjective cylin-
der ASSY is operated.
1 ) CL1 side of the subjective cylinder
ASSY.

a . Press the left- button to


rotate the cylinder to the CL1
side.
b . If the operating noise is loud,

loosen SB2 × 6 (n = 2) to 5$˜
adjust the position of the CL
motor (30703-EA37) to the '# %.
CL1 side.

2 ) CL2 side of the subjective cylinder


ASSY.

a . Press the right+ button to


rotate the cylinder to the CL2
side.
b . If the operating noise is loud,
loosen SB2 × 6 (n = 2) to
adjust the position of the CL
motor (30703-EA37) to the 
5$˜
CL2 side.
'# %.

125 Return to previous page


XAR17ARDA001G/E

5 . Press the function button to display CYL.


%*#46A24'%*'%-
* The chart changes to the clock dial
chart. %;.



6
15
6 . Move the subjective focus base ASSY
Measuring
(30703-3300) back and forth so that the window
chart can be seen through the measuring
window.



7 . Adjust the subjective cylinder ASSY

(30703-3700) with the right+ button or


left- button so that the clock dial chart
can be seen in uniform clarity.

Uniform

Not uniform Not uniform


8 . Press the function button to display
%*#46A24'%*'%-
CHART.
9 . Look through the measuring window, %*#46
press the reset button, then confirm
that there are no dust or scratches on all 0
charts displayed.
10. Adjust the chart position (see 8.5.2.2
[p127]).


 


126 Return to previous page


XAR17ARDA001G/E

8.5.2.2 Chart position adjustment


* Applies only to the AR-360 and ARK-560A.
* When adjustment jigs are not available, return the device to the factory without perform-
ing adjustment.
1 . Remove the dust cover (30601-M304) (see 6.8 [p66]).
2 . Remove the mire ASSY (30601-4000) (see 7.1.4 [p71]).
3 . Attach any of the following jigs and dioptric tester (99104-G020) to the joint jig (30670-1100)
(see 8.2.3 [p110]).
1 ) When the diopter is 0 D: Holder (30670-M031)
2 ) When the diopter is -12 D: Chart lens M (30670-1800)
3 ) When the diopter is +12 D: Chart lens P (30670-1900)
4 . Enter MENU mode, and select 2. CHART
%*#46215+6+10
POSITION
0 5612
5 . Press the next button to display 2-2
CHART POSITION screen.
0
6 . Press the function button to display STOP.
7 . Press the function button to display ,
and turn on the chart LED.


8 . Follow the steps below to adjust the optical axis of the chart.
1 ) Move the focus base ASSY (30601-
3300) to the position where the chart 
can be seen through the dioptric tes-
ter.
2 ) Confirm that the chart does not shift
to the right or left at any position of -
12 D, 0 D, or +12 D. 
$5˜
3 ) If the chart shifts to the right or left,
loosen BS3 × 6 (n = 2) to adjust the
position of the chart mirror ASSY 
(30601-3270).
4 ) Follow the steps below so that the center of the dioptric tester target grid is inside the
inner circle of the clock dial chart (the balloon of the scenery chart) at any position of -
12 D, 0 D, and +12 D.
a . Loosen HH3 × 3 to turn chart
lens holder 2 (30601-M317) of
the chart lens 2 ASSY.

**˜

 

127 Return to previous page


XAR17ARDA001G/E

b . Loosen HH3 × 4 to turn the


chart lens 2 ASSY (30601-

3260).

**˜



9 . After the adjustment is complete, press the exit button and turn off power to the device.
10. Reassemble the removed units and covers.
11 . Perform chart calibration (see 8.7.5 [p135]).

128 Return to previous page


XAR17ARDA001G/E

8.6 ALIGNMENT CALIBRATION


1 . Attach the 22-model eye jig (32171-0600) (see 8.2.6 [p112]).
2 . Enter MENU mode, and select 8. ALIGN-
MENT CALIBRATION (see 8.1.1 [p103]). ᵏ 
3 . Perform alignment of the 0-D model eye #.+)0/'06%#.+$4#6+10
as follows:
1 ) Align the mire ring that is slightly lit
to the center of the target mark.
2 ) Align so that the mire ring is
focused. 24'5556#46$76610

4 . Press the start button to perform ALIGNMENT CALIBRATION.


* A beep sounds when the start button is pressed.
5 . CALIBRATION FINISH 100 is displayed
when ALIGNMENT CALIBRATION is
complete. #.+)0/'06%#.+$4#6+10

%#.+$4#6+10(+0+5*

6 . The focusing indicator appears automati-


cally.
7 . Confirm that the alignment position (up- #.+)0/'06%#.+$4#6+10
and-down and side-to-side directions and
focusing back-and-forth direction) is 0
aligned to the focusing indicator.

%#.+$4#6+10(+0+5*

129 Return to previous page


XAR17ARDA001G/E

8.7 AR CALIBRATION
8.7.1 AR CALIBRATION
1 . Attach the 22-model eye jig (see 8.2.6 [p112]).
2 . Display the CALIB MODE SELECT
screen (for AR), and select AR (see 8.1.2 #4%#.+$4#6+10
[p104]). /1&'.';'07/$'4
%7561/
3 . Select CUSTOM with the up or down

button, and press the execute


button.

4 . Set MEAS COUNT to 9 with the up or


#4%#.+$4#6+10
/1&'.';'07/$'4
down button, and press the execute /'#5%1706
button.

5 . Perform AR CALIBRATION as follows:


1 ) Enter the correction value for each #4%#.+$4#6+10
diopter of model eye that is dis- /1&'.';'219'4
&

played on the screen, with the right

or left button.
2 ) Perform alignment of the model eye,
and press the start button to start
the AR CALIBRATION. 24'5556#46
3 ) Change the displayed diopter with

the up or down button.


4 ) Perform AR CALIBRATION for each
model eye following Steps 1) and 2).
* Be sure that the alignment is maintained during AR CALIBRATION.
* If misalignment occurs, display the numeric value of the misaligned model eye
and perform AR CALIBRATION again.

6 . After AR CALIBRATION is complete, press the exit button and turn off power to the
device.

130 Return to previous page


XAR17ARDA001G/E

7 . Enter measurement mode (see 8.1.4 [p104]).


1 ) For the AR-310A, AR-330A, and
AR-360A, the screen as shown to ᵏ %;. 㧸
the right appears.

4
 .

5 5 
% % 
##
2 ) For the ARK-510A, ARK-530A, and
ARK-560A, the screen as shown to ᵏ %;. 㧸
the right appears.

4
 -
5 4
% 4
#

8 . Press the auto button to select 3D auto tracking and auto shot .

* For the AR-310A and ARK-510A, the indication of auto tracking is .


* Icons vary depending on the model.

9 . Press the CYL mode button to select CYL-.


10. Measure the 22-model eye jig (32171-
0600), and confirm that the measured SPH CYL
Nominal value measurement measurement
value is within the specifications.
accuracy accuracy
+4 D ±0.10 D 0.10 D
+8 D ±0.11 D 0.11 D
+10 D ±0.12 D 0.12 D
+14 D ±0.18 D 0.15 D
+20 D ±0.30 D 0.23 D
0D ±0.10 D 0.10 D
-4 D ±0.10 D 0.10 D
-8 D ±0.11 D 0.11 D
-10 D ±0.12 D 0.12 D
-14 D ±0.18 D 0.15 D
-20 D ±0.30 D 0.23 D

131 Return to previous page


XAR17ARDA001G/E

8.7.2 AR AXIS
1 . Attach the cylinder axis adjustment jig (32177-1500) (see 8.2.7 [p112]).
2 . Display the CALIB MODE SELECT screen (for AR), and select AR AXIS (see 8.1.2 [p104]).
3 . Align the marking-off line to the three hori-
zontal lines of the cylinder axis adjustment #4#:+5%#.+$4#6+10
jig. 
&')
4 . Set the same numeric value as the nomi- 
nal value for 180º that is indicated on the 

base of the cylinder axis adjustment jig 

with the up or down or button. 
24'5556#46

5 . Press the lamp button to light up the


#4#:+5%#.+$4#6+10
mire ring. 
6 . Confirm that the centers of the mire ring &')
and target are in alignment, and the target
displays the optimum condition. Press the
start button.

* When the lamp button is pressed, 24'5556#46


the mire ring lights up and auto track-
ing becomes active.

* Pressing the auto button acti-


vates manual mode.
7 . After the calibration is complete, AR AXIS
CALIB OK is displayed. #4#:+5%#.+$4#6+10
8 . Turn off power to the device after the AR 
&')
AXIS calibration is complete.
9 . Enter measurement mode (see 8.1.4
[p104]).

10. Press the auto button to select 3D


#4#:+5%#.+$1-
auto tracking and auto shot .
* For the AR-310A and ARK-510A, the

indication of auto tracking is .


* Icons differ depending on the model.
11. Confirm that the measured value for each angle (0º, 45º, 90º, 135º) of the cylinder axis
adjustment jig is within ±3° of the intended angle.

12. After the adjustment is complete, press the exit button and turn off power to the device.
13. If the measured values are not within the specifications, perform AXIS OFFSET (see 8.7.3
[p133]).

132 Return to previous page


XAR17ARDA001G/E

8.7.3 AXIS OFFSET


1 . Attach the cylinder axis adjustment jig (32177-1500) (see 8.2.7 [p112]).
2 . Display the CALIB MODE SELECT screen (for AR), and select AR CORRECT (see 8.1.2
[p104]).

3 . Select AXIS OFFSET with the up or


%144'%6/1&'5'.'%6
down button, then press the execute 
*+)*&264)#+0
5*21((5'6
button. #:+51((5'6

4 . Change the numeric value with the up


#:+51((5'6
or down button. 

5 . Press the SAVE button to save the set- FGI
ting.

5#8'

6 . Turn off power to the device.


7 . Enter measurement mode (see 8.1.4 [p104]).
8 . Confirm that the measured value for each angle (0º, 45º, 90º, 135º) of the cylinder axis
adjustment jig is within ±3° of the intended angle.

9 . After the adjustment is complete, press the exit button twice and turn off power to the
device.

133 Return to previous page


XAR17ARDA001G/E

8.7.4 SPH OFFSET


* Change the SPH OFFSET values only when the AR measurement values differ by the
same amount.
1 . Display the CALIB MODE SELECT screen (for AR), and select AR CORRECT (see 8.1.2
[p104]).

2 . Select SPH OFFSET with the up or


%144'%6/1&'5'.'%6
down button, then, press the execute 
*+)*&264)#+0
5*21((5'6
button. #:+51((5'6

3 . Change the numeric value with the up


52*1((5'6
or down button. 

4 . Press the SAVE button to save the set- &
ting.

5#8'

5 . When the measured value for 0.00 D is -4.00 D, changing the numeric value results in as
below.
1 ) When the numeric value is changed to +0.50 D, the measured value becomes -3.50 D.
2 ) When the numeric value is changed to -0.50 D, the measured value becomes -4.50 D.

6 . After the adjustment is complete, press the exit button twice and turn off power to the
device.

134 Return to previous page


XAR17ARDA001G/E

8.7.5 CHART CALIBRATION


8.7.5.1 AR-310A/AR-330A/ARK-510A/ARK-530A
8.7.5.1.1 Chart test jig
* If the adjustment jig is not available, return the device to the factory without performing
adjustment.
1 . Attach the chart test jig (32145-2100 or 30670-2300) (see 8.2.8 [p113] and 8.2.9 [p114]).
2 . Display the CALIB MODE SELECT screen (for AR), and select CHART (see 8.1.2 [p104]).

3 . Select CHART CALIB with the up or


%*#46%#.+$4#6+10
down button, then press the execute %*#46%#.+$
%*#46%*'%-
button.

4 . Perform CHART CALIBRATION as follows:


1 ) Follow the steps below to adjust the
%*#46%#.+$4#6+10
0 D position.
a . Turn the turret plate to select a &
position without any lens.

b . Press the left- button until


the whole chart is focused for
the first time.
c . Press the SAVE button to
save the 0 D position.
5#8' 



2 ) Press the down button, and fol- %*#46%#.+$4#6+10


low the steps below to adjust the
 &
+12.12 D position.
a . Turn the turret plate to select
the +12.00 D lens (33510-
G033).

b . Press the left- button until


the whole chart is focused for
the first time. 5#8' 
c . Press the SAVE button to 

save the +12.12 D position.

135 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and fol- %*#46%#.+$4#6+10


low the steps below to adjust the - &
12.00 D position.
a . Turn the turret plate to select
0
the -12.00 D lens (33510-
G018).

b . Press the left- button until %#.+$%1/2.'6'&


the whole chart is focused for
the first time. 5#8' 
c . Press the SAVE button to 

save the -12.00 D position.
0
* Pressing the SAVE button displays the next button.
0
4 ) Press the next button to display the CHART CHECK screen.

5 . Perform CHART CHECK as follows:


1 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a
%*#46%*'%-
position without any lens.
&
b . Press the left- or right+

button several times.

c . Press the left- or right+

button until the whole


chart is focused for the first
time. 
d . Press the start button, and 

confirm that the displayed
numeric value is 0.00 ±0.25 D.

2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select
%*#46%*'%-
the +12.00 D lens (33510-
G033).  &

b . Press the left- or right+

button several times.

c . Press the right+ button


until the whole chart is focused
for the first time.
d . Press the start button, and 

confirm that the displayed 
numeric value is +12.12 ±0.25
D.

136 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select
%*#46%*'%-
the -12.00 D lens (33510-
G018). &

b . Press the left- or right+


button several times. 0

c . Press the left- or right+


button until the whole chart is
focused.
d . Press the start button, and con- 
firm that the displayed numeric 

value is -12.00 ±0.25 D.
6 . If the value is not within the specifications, repeat CHART CALIBRATION starting from Step
2.

7 . Press the exit button twice and turn off the power switch.
8 . Remove the chinrest attachment jig (32107-1100) and chart test jig (32145-2100 or 30670-
2300).

137 Return to previous page


XAR17ARDA001G/E

8.7.5.1.2 Diopter tester jig ASSY


* If the adjustment jig is not available, return the device to the factory without performing
adjustment.
1 . Attach the dioptric tester jig ASSY (30670-6000) (see 8.2.10 [p115]).
2 . Enter MENU mode, and select 10.
%*#46%#.+$4#6+10
CHART CALIBRATION (see 8.1.1 [p103]).

3 . Select CHART CALIB with the up or %*#46%#.+$
%*#46%*'%-
down button, then press the execute

button.

4 . Perform CHART CALIBRATION as follows:


1 ) Follow the steps below to adjust the 0.00 D position.
a . Turn the turret plate to select a
%*#46%#.+$4#6+10
position without any lens.
&
b . Press the left- button until
the whole chart is focused for
the first time.
c . Press the SAVE button to
save the 0.00 D position.

5#8' 



2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select
%*#46%#.+$4#6+10
the +12.00 D lens (33510-
G033).  &

b . Press the left- button until


the whole chart is focused for
the first time.
c . Press the SAVE button to
save the +12.12 D position.
5#8' 



138 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select
%*#46%#.+$4#6+10
the -12.00 D lens (33510-
G018). &

b . Press the left- button until


the whole chart is focused for 0
the first time.
c . Press the SAVE button to
%#.+$%1/2.'6'&
save the -12.00 D position.
* Pressing the SAVE but-
5#8' 
ton displays the next

0 
button.
0
4 ) Press the next button to display the CHART CHECK screen.
5 . Perform CHART CHECK as follows:
1 ) Follows the steps below to check the 0.00 D position.
a . Turn the turret plate to select a %*#46%*'%-
position without any lens.
&
b . Press the left- or right+

button several times.


c . Perform adjustment with the

left- or right+ button so


that the whole chart is
focused.

d . Press the start button, and 
confirm that the displayed 
numeric value is 0.00 ±0.25 D.

2 ) Press the down button, and follow the steps belows to check the +12.12 D posi-
tion.
a . Turn the turret plate to select %*#46%*'%-
the +12.00 D lens (33510-
G033).  &

b . Press the left- or right+

button several times.


c . Perform adjustment with the

left- or right+ button so


that the whole chart is

focused. 
d . Press the start button, and 
confirm that the displayed
numeric value is +12.12 ±0.25 D.

139 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select
%*#46%*'%-
the -12.00 D lens (33510-
G018). &

b . Press the left- or right+


0
button several times.
c . Perform adjustment with the

left- or right+ button so


that the whole chart is
focused. 

d . Press the start button, and 
confirm that the displayed
numeric value is -12.00 ±0.25 D.
0
4 ) Press the next button to display the CALIBRATION FINISH screen.
6 . When PLEASE SHUT OFF is displayed, turn off the power switch.
7 . Remove the chinrest attachment jig (32107-1100) and dioptric tester jig ASSY (30670-
6000) in the reverse order (see 8.2.10 [p115]).

140 Return to previous page


XAR17ARDA001G/E

8.7.5.2 AR-360A/ARK-560A
8.7.5.2.1 Chart test jig
* If the adjustment jig is not available, return the device to the factory without performing
adjustment.
1 . Attach the chart test jig (30670-2300) (see 8.2.9 [p114]).
2 . Display the CALIB MODE SELECT screen (for AR), and select CHART (see 8.1.2 [p104]).
3 . Perform STAR CALIB as follows:

1 ) Select STAR CALIB with the up


%*#46%#.+$4#6+10

or down button, then press the 56#4%#.+$
6*'6#%#.+$
execute button. 6*'6#%#.+$
6*'6#%*'%-
.+/+6%*'%-
56#4%*'%-
$#.110%#.+$
$#.110%*'%-

2 ) Follow the steps below to adjust the


56#4%#.+$4#6+10
0.00 D position.
a . Turn the turret plate to select a &
position without any lens.

b . Press the left- button until


the whole clock dial chart is
focused for the first time.
c . Press the SAVE button to
save the 0.00 D position.
5#8' 



3 ) Press the down button, and fol- 56#4%#.+$4#6+10


low the steps below to adjust the
 &
+12.12 D position.
a . Turn the turret plate to select
the +12.00 D lens (33510-
G033).

b . Press the left- button until


the whole clock dial chart is
focused for the first time. 5#8' 
c . Press the SAVE button to 

save the +12.12 D position.

141 Return to previous page


XAR17ARDA001G/E

4 ) Press the down button, and fol- 56#4%#.+$4#6+10


low the steps below to adjust the -
&
12.00 D position.
a . Turn the turret plate to select
the -12.00 D lens (33510-
0
G018).

b . Press the left- button until %#.+$%1/2.'6'&


the whole clock dial chart is
focused for the first time. 5#8' 
c . Press the SAVE button to save 

the -12.00 D position.
0
* Pressing the SAVE button displays the next button.
0
5 ) Press the next button to display the THETA1 CALIB screen.
4 . Perform THETA1 CALIB as follows:
1 ) Turn the turret plate to select a posi-
tion without any lens.

2 ) Perform adjustment with the left-

or right+ button so that the clock


dial chart can be seen in uniform
clarity. Uniform

Not uniform Not uniform


3 ) Press the SAVE button to save the
6*'6#
position.
* Pressing the SAVE button dis-
0
plays the next button.
0
4 ) Press the page down button to
confirm that the whole clock dial
chart is seen in uniform clarity at
every angle.
5#8'
* Pressing the page down 6 
button turns the subjective cylin- 15 
der ASSY up to 180º by incre-
ments of 45º.

* Pressing the page down button five times returns the subjective cylinder
ASSY to the original position.
5 ) If the chart is not focused at every angle, repeat adjustment starting from Step 2).
0
6 ) Press the NEXT button to display the THETA2 CALIB screen.

142 Return to previous page


XAR17ARDA001G/E

5 . Perform THETA2 CALIB as follows:


1 ) Turn the turret plate to select the -5.36D CYL lens (33510-G044).

2 ) Press the page down button to 6*'6#


select OFFSET.
* The indication of “OS” turns red.

3 ) Perform adjustment with the left- 0


or right+ button so that the whole
clock dial chart can be seen in uni-
form clarity.

4 ) Press the page down button to 5#8'


select THETA2. 6 
15 
* The indication of “T2” turns red.

5 ) Perform adjustment with the left- or right+ button so that the whole clock dial
chart can be seen in uniform clarity.
6 ) Perform Steps 2) to 5) twice.
7 ) Press the SAVE button to save the position.
0
* Pressing the SAVE button displays the NEXT button.
0
8 ) Press the NEXT button to display the THETA1 CHECK screen.
6 . Perform THETA1 CHECK as follows:
1 ) Turn the turret plate to select a position without any lens.

2 ) Press the left- or right+ but- 6*'6#%*'%-


ton several times.
 &
3 ) Perform adjustment with the left-

or right+ button so that the clock


dial chart can be seen in uniform
clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±0.25 D.
5 ) If the numeric value is not within

±0.25 D, press the exit button.


and repeat adjustment starting from Step 4 (THETA1 CALIB).

6 ) Press the down button to display the THETA2 CHECK screen.

143 Return to previous page


XAR17ARDA001G/E

7 . Perform THETA2 CHECK as follows:


1 ) Turn the turret plate to select the -5.36D CYL lens (33510-G044).

2 ) Press the left- or right+ but- 6*'6#%*'%-


ton several times.
 &
3 ) Perform adjustment with the left-

or right+ button so that the clock


dial chart can be seen in uniform
clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±0.25 D.
5 ) If the numeric value is not within

±0.25 D, press the exit button


and repeat adjustment starting from Step 5 (THETA2 CALIB).

6 ) Press the down button to display the OFFSET CHECK screen.


8 . Perform OFFSET CHECK as follows:
1 ) Turn the turret plate to select the -5.36 D CYL lens (33510-G044).

2 ) Press the left- button or right+ 1((5'6%*'%-


button several times. &')

3 ) Perform adjustment with the left-


0
or right+ button so that the clock
dial chart can be seen in uniform
clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±3º
5 ) If the numeric value is not within ±3º,

press the exit button and repeat


adjustment starting from Step 5 (THETA2 CALIB).
0
6 ) Press the next button to display the CHECK CYL LIMIT screen.
9 . Confirm that CHECK OK is displayed in the CHECK CYL LIMIT screen.
0
10. Press the next button to return to the STAR CALIBRATION screen.

144 Return to previous page


XAR17ARDA001G/E

11. Perform STAR CHECK as follows:

1 ) Select STAR CHECK with the up or down button, then press the execute
button.
2 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a position without any lens.
b . Perform adjustment with the 56#4%*'%-
left- or right+ button so &
that the whole clock dial chart
is focused.
c . Press the start button, and
confirm that the displayed
numeric value is within 0.00
±0.25 D.




3 ) Press the down button, and follow the steps below to check the +12.12 D posi-
tion.
a . Turn the turret plate to select 56#4%*'%-
the +12.00 D lens (33510-
G033).  &
b . Perform adjustment with the

left- or right+ button so


that the whole clock dial chart
is focused.
c . Press the start button, and
confirm that the displayed

numeric value is +12.12 ±0.25 
D. 

4 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select 56#4%*'%-
the -12.00 D lens (33510-
G018). &
b . Perform adjustment with the
0
left- or right+ button so
that the whole clock dial chart
is focused.
c . Press the start button, and
confirm that the displayed

numeric value is -12.00 ±0.25 
D. 

0
12. Press the next button to display the BALOON CALIB screen.

145 Return to previous page


XAR17ARDA001G/E

13. Perform BALOON CALIB as follows:


1 ) Follow the steps below to adjust the 0.00 D position.
a . Turn the turret plate to select a position without any lens.

b . Press the left- button until %*#46%#.+$4#6+10


the whole chart is focused for &
the first time.
c . Press the SAVE button to save
the 0.00 D position.

5#8' 



2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- button until %*#46%#.+$4#6+10


the whole scenery chart is  &
focused for the first time.
c . Press the SAVE button to save
the +12.12 D position.

5#8' 



3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).

b . Press the left- button until %*#46%#.+$4#6+10


the whole scenery chart is
&
focused for the first time.
c . Press the SAVE button to save
the -12.00 D position. 0
* Pressing the SAVE button
0
displays the next but- %#.+$%1/2.'6'&
ton.
5#8' 



0
4 ) Press the next button to display the BALOON CHECK screen.

146 Return to previous page


XAR17ARDA001G/E

14. Perform BALOON CHECK as follows:


1 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a position without any lens.

b . Press the left- or right+ %*#46%*'%-


button several times. &
c . Perform adjustment with the

left- or right+ button so


that the whole scenery chart is
focused.
d . Press the start button, and con-
firm that the displayed numeric
value is 0.00 ±0.25 D. 



2 ) Press the down button, and follow the steps below to check the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- or right+ %*#46%*'%-


button several times.  &
c . Perform adjustment with the

left- or right+ button so


that the whole scenery chart is
focused.
d . Press the start button and con-
firm that the displayed numeric
value is +12.12 ±0.25 D. 



3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).

b . Press the left- or right+ %*#46%*'%-


button several times. &
c . Perform adjustment with the

left- or right+ button so 0


that the whole scenery chart is
focused.
d . Press the start button, and con-
firm that the displayed numeric
value is -12.00 ±0.25 D. 

0 
15. Press the next button to display the
CALIBRATION FINISH screen.
16. When PLEASE SHUT OFF is displayed on the CALIBRATION FINISH screen, turn off the
power switch.
17. Remove the chinrest attachment jig (32107-1100) and chart test jig ASSY (30670-2300).

147 Return to previous page


XAR17ARDA001G/E

8.7.5.2.2 Dioptric tester jig ASSY


* If the adjustment jig is not available, return the device without performing adjustment.
1 . Attach the dioptric tester jig ASSY (30670-6000) (see 8.2.10 [p115]).
2 . Display the CALIB MODE SELECT screen (for AR), and select CHART (see 8.1.2 [p104]).
3 . Perform STAR CALIBRATION as follows.

1 ) Select STAR CALIB with the up


%*#46%#.+$4#6+10

or down button, then press the 56#4%#.+$
6*'6#%#.+$
execute button. 6*'6#%#.+$
6*'6#%*'%-
.+/+6%*'%-
2 ) Press the reset button to display 56#4%*'%-
the camera screen of the diopter $#.110%#.+$
$#.110%*'%-
tester jig ASSY.

3 ) Follow the steps below to adjust the 0.00 D position.


a . Turn the turret plate to select the position without any lens.

b . Press the left- button until 56#4%#.+$4#6+10


the whole clock dial chart is &
focused for the first time.
c . Press the SAVE button to save
the 0.00 D position.

5#8' 



4 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- button until 56#4%#.+$4#6+10


the whole clock dial chart is
 &
focused for the first time.
c . Press the SAVE button to
save the +12.12 D position.

5#8' 



148 Return to previous page


XAR17ARDA001G/E

5 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).

b . Press the left- button until 56#4%#.+$4#6+10


the whole clock dial chart is
&
focused for the first time.
c . Press the SAVE button to
save the -12.00 D position. 0
* Pressing the SAVE but-
0
ton displays the next %#.+$%1/2.'6'&
button.
0 5#8' 
6 ) Press the NEXT button to dis- 
play the THETA1 CALIB screen. 
4 . Perform THETA1 CALIB as follows:
1 ) Turn the turret plate to select the position without any lens.

2 ) Perform adjustment with the left-

and right+ button so that the


whole clock dial chart is focused.
3 ) Press the SAVE button to save the
position.
* Pressing the SAVE button dis- Uniform
0
plays the next button.

Not uniform Not uniform

4 ) Press the page down button to 6*'6#


confirm that the whole chart is
focused at every angle.

* Pressing the page down


button turns the subjective cylin-
der ASSY up to 180º by incre-
ments of 45º.

* Pressing the page down 5#8'


button five times returns the 6 
subjective cylinder ASSY to the 15 
original position.
5 ) If the chart is not focused at every angle, repeat adjustment starting from Step 2).
0
6 ) Press the NEXT button to display the THETA2 CALIB screen.

149 Return to previous page


XAR17ARDA001G/E

5 . Perform THETA2 CALIB as follows:


1 ) Turn the turret plate to select the -5.36 D CYL lens (33510-G044).

2 ) Press the page down button to 6*'6#


select OFFSET.
* The indication of “OS” turns red.

3 ) Perform adjustment with the left-

or right+ button so that the


whole clock dial chart can be seen in
uniform clarity.

4 ) Press the page down button to 5#8'


select THETA2. 6 
15 
* The indication of “T2” turns red.

5 ) Perform adjustment with the left- or right+ button so that the whole clock dial
chart can be seen in uniform clarity.
6 ) Perform Steps 2) to 5) again.
7 ) Press the SAVE button to save the position.
0
* Pressing the SAVE button displays the next button.
0
8 ) Press the NEXT button to display the THETA1 CHECK screen.
6 . Perform THETA1 CHECK as follows:
1 ) Turn the turret plate to select the position without any lens.

2 ) Press the left- button or right+ 6*'6#%*'%-


button several times.  &

3 ) Perform adjustment with the left-

or right+ button so that the


whole clock dial chart can be seen in
uniform clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±0.25 D.
5 ) If the numeric value is not within

±0.25 D, press the exit button


and repeat adjustment starting from Step 4 (THETA1 CALIB).

6 ) Press the down button to display the THETA2 CHECK screen.

150 Return to previous page


XAR17ARDA001G/E

7 . Perform THETA2 CHECK as follows:


1 ) Turn the turret plate to select the -5.36 D CYL lens (33510-G044).

2 ) Press the left- or right+ but- 6*'6#%*'%-


ton several times.
 &
3 ) Perform adjustment with the left-

or right+ button so that the


whole clock dial chart can be seen in
uniform clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±0.25 D
5 ) When the numeric value is not within

±0.25 D, press the exit button


and repeat adjustment starting from Step 5 (THETA2 CALIB).

6 ) Press the down button to display the OFFSET CHECK screen.


8 . Perform OFFSET CHECK as follows:
1 ) Turn the turret plate to select the position without any lens.

2 ) Press the left- or right+ but- 1((5'6%*'%-


ton several times.
&')
3 ) Perform adjustment with the left-

or right+ button so that the 0


whole clock dial chart can be seen in
uniform clarity.
4 ) Press the start button, and confirm
that the displayed numeric value is
within ±3º.
5 ) When the value is not within ±3º,

press the exit button and repeat


adjustment starting from Step 5 (THETA2 CALIB).
0
6 ) Press the next button to display the CHECK CYL LIMIT screen.
9 . Confirm that CHECK OK is displayed.
0
10. Press the next button to display the STAR CHECK screen.

151 Return to previous page


XAR17ARDA001G/E

11. Perform STAR CHECK as follows:


1 ) Select STAR CHECK with the up
%*#46%#.+$4#6+10
or down button, then press 
56#4%#.+$
the execute button. 6*'6#%#.+$
6*'6#%#.+$
6*'6#%*'%-
.+/+6%*'%-
56#4%*'%-
$#.110%#.+$
$#.110%*'%-

2 ) Follow the steps below to check the 0.00 D position.


a . Turn the turret plate to select a position without any lens.

b . Press the left- or right+ 56#4%*'%-

button several times. &


c . Perform adjustment with the

left- or right+ button so


that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed 
numeric value is 0.00 ±0.25 D. 

e . Press the down button.
3 ) Follow the steps below to check the +12.12 D position.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- or right+ 56#4%*'%-

button several times.  &


c . Perform adjustment with the

left- or right+ button


so that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed
numeric value is +12.12 ±0.25 

D. 

e . Press the down button.

152 Return to previous page


XAR17ARDA001G/E

4 ) Follow the steps below to check the -12.00 D position.


a . Turn the turret plate to select the -12.00 D lens (33510-G018).

b . Press the left- or right+ 56#4%*'%-

button several times. &


c . Perform adjustment with the
0
left- or right+ button so
that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed 
numeric value is -12.00 ±0.25 

D.

0
5 ) Press the next button to display the BALOON CALIB screen.
12. Perform BALOON CALIB as follows:
1 ) Follow the steps below to adjust the 0.00 D position.
a . Turn the turret plate to select a position without any lens.

b . Press the left- or right+ %*#46%#.+$4#6+10

button several times. &

c . Press the left- button until


the whole chart is focused for
the first time.
d . Press the SAVE button to
save the 0.00 D position.
5#8' 



2 ) Press the down button, and follow the steps below to adjust the +12.12D position.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- or right+ %*#46%#.+$4#6+10

button several times.  &

c . Press the left- button until


the whole chart is focused for
the first time.
d . Press the SAVE button to
save the +12.12 D position.
5#8' 



153 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).

b . Press the left- or right+ %*#46%#.+$4#6+10

button several times. &

c . Press the left- button until


the whole chart is focused for 0
the first time.
d . Press the SAVE button to
%#.+$%1/2.'6'&
save the -12.00 D position.
* Pressing the SAVE but-
5#8' 
0
ton displays the next 

button.

0
4 ) Press the next button to display the BALOON CHECK screen.
13. Perform BALOON CHECK as follows:
1 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a position without any lens.

b . Press the left- or right+ %*#46%*'%-

button several times. &


c . Perform adjustment with the

left- or right+ button so


that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed 
numeric value is 0.00 ±0.25 D. 

2 ) Press the down button, and fol-
low the steps below to check the +12.12 D position.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).

b . Press the left- or right+ %*#46%*'%-

button several times.  &


c . Perform adjustment with the

left- or right+ button so


that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed
numeric value is +12.12 ±0.25 

D. 

154 Return to previous page


XAR17ARDA001G/E

3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select the -12.00 D (33510-G018) position.

b . Press the left- or right+ %*#46%*'%-

button several times. &


c . Perform adjustment with the

left- or right+ button so 0


that the whole chart is
focused.
d . Press the start button, and
confirm that the displayed
numeric value is -12 ±0.25 D. 

0 
4 ) Press the next button to display
the CALIBRATION FINISH screen.
14. When PLEASE SHUT OFF is displayed, turn off the power switch.
15. Remove the chinrest attachment jig (32107-1100) and dioptric tester jig ASSY (30670-
6000) in the reverse order.

155 Return to previous page


XAR17ARDA001G/E

8.8 KM CALIBRATION
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
8.8.1 KM CALIB
1 . Attach the KM calibration jig (30670-0300) (see 8.2.11 [p116]).
2 . Display the CALIB MODE SELECT screen (for KM), and select KM CALIB (see 8.1.6
[p106]).
* When the software version is earlier than Ver. 2.03, press the reset button.
* When the software version is Ver. 2.03 or later, the reset button is not displayed.
3 . Perform KM CALIBRATION as follows:
1 ) Align the relevant steel ball of the
KM calibration jig (30670-0300) in -/%#.+$4#6+10
the order of the following numeric OO
values so that the focusing indicator
displays the optimum condition.
a . 7.94 mm
b . 5.95 mm
c . 9.13 mm %#.+$4#6+10(+0+5*
2 ) Press the start button to perform KM
CALIBRATION.
* Perform calibration of all three
steel balls.
3 ) Press the SAVE button to save the setting.

4 . Press the exit button.


5 . Perform KM CALIB CHECK (see 8.8.2 [p157]).
6 . Perform KM AXIS CHECK (see 8.8.3 [p161]).

156 Return to previous page


XAR17ARDA001G/E

8.8.2 KM CALIB CHECK


8.8.2.1 ADJUST GAIN
1 . Attach the KM calibration jig (30670-0300) (see 8.2.11 [p116]).
2 . Display the CALIB MODE SELECT screen (for KM), and select KM ADJUST (see 8.1.6
[p106]).

3 . Select ADJUST GAIN with the up or


-/#&,756
down button, then press the execute #&,756)#+0
#&,7562*+)#+0
button.

4 . Press the start button and measure each


steel ball of the KM calibration jig (30670- -/#&,756
0300). #&,756)#+0
4
5 . The center measurement values of each 4
steel ball of the KM calibration jig (30670- 4
4
0300) are displayed as follows:
1 ) K: the number of the KM measure-
ment
2 ) MAX: the maximum value among
the 10 measurements 5#8' -/#:/+0&+((
4
3 ) MIN: the minimum value among the 4
10 measurements
4 ) DIFF: the difference between R1 maximum value and R2 minimum value
5 ) R1: the measurement value of the major axis
6 ) R2: the measurement value of the minor axis
6 . Confirm that the center measurement values for each steel ball of the KM calibration jig are
within the following specifications.
1 ) The maximum value of R1 and mini-
mum value of R2 are within the tol- Corneal
Tolerance (mm) DIFF (mm)
erance range. curvature radius

2 ) Confirm that the DIFF value is within R 5.00 ±0.02 0.04

0.04 mm. R 5.95 ±0.02 0.04


R 7.94 ±0.02 0.04
R 9.13 ±0.02 0.04
R10.00 ±0.03 0.04
R13.00 ± 0.05 0.04

157 Return to previous page


XAR17ARDA001G/E

7 . If the center measurement values are not within the specifications, perform correction as
follows:
* When the measurement values for each steel ball of the KM calibration jig (30670-
0300) differ by the same amount, change all correction factors by the same numeric
value.
* When the measurement values for only some steel balls of the KM calibration jig
(30670-0300) differ by the same amount, change the correction factors close to the
radiuses of such steel balls.
1 ) Select the steel ball to be changed
-/#&,756
with the up or down button. #&,756)#+0
2 ) Change the numeric value with the 4
4
4
right or left button. 4
* Perform correction by taking the
median value of specifications
as a guide.
* Change the numeric value by
0.0010 (as a guide) for each 5#8' -/#:/+0&+((
4
correction of 0.01 mm. 4
3 ) Press the start button, and confirm
the measurement values.
4 ) Press the SAVE button to save the setting.
8 . Repeat Steps 5 to 6 until the center measurement values for each steel ball of the KM cali-
bration jig are within the specifications.

9 . After the adjustment is complete, press the exit button to return to the KM ADJUST
screen.
10. Perform ADJUST PHI6 GAIN (see 8.8.2.2 [p159]).

158 Return to previous page


XAR17ARDA001G/E

8.8.2.2 ADJUST PHI6 GAIN


1 . Attach the KM calibration jig (30670-0300) (see 8.2.11 [p116]).
2 . Display the CALIB MODE SELECT screen (for KM), and select KM ADJUST (see 8.1.6
[p106]).
3 . Select ADJUST PHI6 AGAIN with the up
-/#&,756
or down button, then press the
#&,756)#+0
execute button. #&,7562*+)#+0

4 . Press the start button and measure each


steel ball of the KM calibration jig (30670- -/#&,756
0300) #&,7562*+)#+0
4
5 . The peripheral measurement values of 4
each steel ball of the KM calibration jig 4
4
(30670-0300) are displayed as follows:
1 ) P: the number of the KM peripheral
measurement
2 ) MAX: the maximum value among
the 10 measurements 5#8' 2/#:/+0&+((
4
3 ) MIN: the minimum value among the 4
10 measurements
4 ) DIFF: the difference between R1 maximum value and R2 minimum value
5 ) R1: the measurement value of the major axis
6 ) R2: the measurement value of the minor axis
6 . Confirm that the peripheral measurement values for each steel ball of the KM calibration jig
are within the following specifications.
1 ) The maximum value of R1 and mini-
mum value of R2 are within the tol- Corneal
Tolerance (mm) DIFF (mm)
curvature radius
erance range
R 5.00 ± 0.05 0.05
2 ) Confirm that the DIFF value is within
R 5.95 ± 0.05 0.05
0.05 mm.
R 7.94 ± 0.05 0.05
R 9.13 ± 0.05 0.05
R10.00 ± 0.05 0.05
R13.00 ± 0.05 0.05

159 Return to previous page


XAR17ARDA001G/E

7 . If the peripheral measurement values are not within the specifications, perform correction
as follows:
* When the measurement values for each steel ball of the KM calibration jig (30670-
0300) differ by the same amount, change all correction factors by the same numeric
value.
* When the measurement values for only some steel balls of the KM calibration jig
(30670-0300) differ by the same amount, change the correction factors close to the
radiuses of such steel balls.
1 ) Select the steel ball to be changed
-/#&,756
with the up or down button. #&,756)#+0
#&,7562*+)#+0
2 ) Change the numeric value with the 4
4
right or left button. 4
4
* Perform correction by taking the
median value of specifications
as a guide.
* Change the numeric value by
-/#:/+0&+((
5#8' 2/#:/+0&+((
0.0010 (as a guide) for each 4
correction of 0.01 mm. 4
3 ) Press the start button, and confirm
the measurement values.
4 ) Press the SAVE button to save the setting.
8 . Repeat Steps 5 to 6 until the peripheral measurement values for each steel ball of the KM
calibration jig are within the specifications.

9 . Press the exit button three times to restart the device.

160 Return to previous page


XAR17ARDA001G/E

8.8.3 KM AXIS CHECK


1 . Attach the aspheric model eye (32177-1400) (see 8.2.12 [p116]).
2 . Display the CALIB MODE SELECT
screen (for KM), and select KM AXIS -/#:+51((5'6
CHECK (see 8.1.6 [p106]). 2*+
2*+ 
3 . Align the marking-off line to the three hori-
zontal lines of the aspheric model eye.
5#8'

4 . Press the lamp button to turn off the -/#:+51((5'6


three horizontal and violet square-shaped 2*+
lines. 2*+ 
5 . Perform alignment of the aspheric model
5#8'
eye (32177-1400).

6 . Press the start button to measure the cor-


neal cylinder axis angles (0º, 45º, 90º, -/#:+51((5'6
135º) of the aspheric model eye (32177- 2*+
2*+ 
1400).
1 ) A: Center KM measurement values
5#8'
a . MAX: the maximum value
among the 10 measurements
b . MIN: the minimum value -/#:/+0&+((
among the 10 measurements 4
4
2 ) B: Peripheral KM measurement val- 2/#:/+0&+((
4
ues 4
a . MAX: the maximum value
among the 10 measurements $ #
b . MIN: the minimum value
among the 10 measurements
7 . Confirm that the measurement values (maximum and minimum) for the corneal cylinder
axis angles (0º, 45º, 90º, 135º) of the aspheric model eye (32177-1400) are within the fol-
lowing specifications.
1 ) Center specifications: ±2º of each intended angle
2 ) Peripheral specifications: ±4º of each intended angle

161 Return to previous page


XAR17ARDA001G/E

8 . If the center KM measurement values are not within the specifications, perform the correc-
tion as follows:

1 ) Press the reset button to select PHI3.


2 ) Change the numerical value with the
-/#:+51((5'6
up and down buttons, 2*+
* The measured values can be 2*+ 
corrected within the range of
±5º. 5#8'
3 ) Press the start button to perform
measurement.
-/#:/+0&+((
4 ) Confirm that the center measure- 4
ment values (MAX and MIN) for the 4
2/#:/+0&+((
corneal cylinder axis angles (0º, 45º, 4
4
90º, 135º) of the aspheric model eye
are within ±2º of each intended $ #
angle.
* A: Center KM measurement values
9 . If the peripheral KM measurement values are not within the specifications, perform OFF-
SET PHI6 AXIS as follows:

1 ) Press the reset button to select PHI3.

2 ) Change the numerical value with the up and down button,


* The measured values can be corrected within the range of ±5º.
3 ) Press the start button to perform the measurement.
4 ) Confirm that the peripheral measurement values (MAX and MIN) for the corneal cylin-
der axis angles (0º, 45º, 90º, 135º) of the aspheric model eye are within ±4ºº of each
intended angle.
* B: Peripheral KM measurement values
10. Press the SAVE button to save the setting.

11 . Press the exit button and turn off the power switch.

162 Return to previous page


XAR17ARDA001G/E

8.9 PD CALIBRATION
8.9.1 CENTER POSITION
1 . Attach the PD checking jig (32545-1002) (see 8.2.13 [p116]).
* A scale can be used as substitute for the PD checking jig (32545-1002).
2 . Enter MENU mode, and select 15. PD ADJUST MODE (see 8.1.1 [p103]).
3 . Select CENTER POSITION with the up
2&#&,756/1&'
or down button, then press the %'06'4215+6+10
execute button. )#+01((5'6
%#.+$4#6+10

4 . Align the flap of the rear body cover


(30601-M701) to the groove of the base
(30601-M001).
5 . With the flap aligned to the groove, focus
the PD checking jig.


Flap
⓭⿠


Groove
┄ߺ

6 . Confirm that any shifting of the PD center


from 50 mm is within ±5 mm.
2&%'06'4#&,756
7 . Press the start button to save the PD cen-
ter position.
8 . Press the SAVE button to save the set-
ting.

9 . Press the exit button.


10. Perform PD CALIBRATION (see 8.9.2

㪌㪇
[p164]). 215+6+10
5#8' 1((5'6 

163 Return to previous page


XAR17ARDA001G/E

8.9.2 PD CALIBRATION
1 . Attach the PD checking jig (32545-1002) (see 8.2.13 [p116]).
* A scale can be used as substitute for the PD checking jig (32545-1002).
2 . Enter MENU mode, and select 15. PD ADJUST MODE (see 8.1.1 [p103]).

3 . Select PD CALIBRATION with the up 2&#&,756/1&'

or down button, then press the exe- %'06'4215+6+10


)#+01((5'6
cute button. %#.+$4#6+10

4 . Align the focusing indicator to the 20 mm


position of the jig, then press the start but- ᵏ 2&%#.+$4#6+10
OO/'#574'/'06
ton.

)#+0

㪉㪇
1((5'6

5 . Align the focusing indicator to the 84 mm


position of the jig, then press the start but- 2&%#4+$4#6+10
OO/'#574'/'06

ton.

6 . Press the right button.

)#+0


1((5'6

164 Return to previous page


XAR17ARDA001G/E

7 . Align the focusing indicator to the 90 mm


position of the jig, then press the start but- ᵏ 2&%#.+$4#6+10
OO/'#574'/'06

ton.

)#+0

㪐㪇
1((5'6

8 . Align the focusing indicator to the 10 mm


position of the jig, then press the start but- ᵏ 2&%#.+$4#6+10
OO/'#574'/'06

ton.

9 . Press the right button.

)#+0

㪈㪇
1((5'6

10. When CALIB COMPLETE is displayed,


press the SAVE button to save the setting. ᵏ  3'&$/,%5$7,21
&$/,%&203/(7(

11 . Press the right button, then the exit

button.

*$,1
6$9(

㻝㻜
2))6(7

12. .Select GAIN&OFFSET with the up or 2&#&,756/1&'

down button, then press the execute %'06'4215+6+10


)#+01((5'6
button. %#.+$4#6+10

165 Return to previous page


XAR17ARDA001G/E

13. Measure the PD values at the 64 mm and


80 mm positions. Confirm that the mea- ᵏ 2&)#+0#&,756 㧸
sured values are within the following spec-
ifications.
1 ) PD 64 mm: less than 64 ±1.0 mm
2 ) PD 80 mm: within 80 ±1.0 mm
14. If the measured values are not within the
specifications, perform PD CALIBRATION
again (see 8.9 [p163]).
)#+0
5#8'

㪐㪇
1((5'6
6*4'5



166 Return to previous page


XAR17ARDA001G/E

8.10 CS/PS Calibration


* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
1 . Attach the PD checking jig (32545-1002) (see 8.2.13 [p116]).
* A scale can be used as substitute for the PD checking jig (32545-1002).
2 . Enter MENU mode, and select 16. CS PS ADJUST (see 8.1.1 [p103]).
3 . Press the start button.
4 . The screen is captured, then the CS cali- ᵏ #&,756%5255%#.'
bration screen appears and the right cur- 
sor turns pink.

56#4659%#2674'
5#8'

5 . Perform CS calibration as follows:


1 ) Align the right cursor to the scale ᵏ #&,756%5255%#.'
(+0+5*
with the right or left button.
2 ) Select the left cursor with the down

button.
3 ) Align the left cursor to a position 12
mm away from the right cursor with

the right or left button. 5#8'


6 . Press the start button to display the PS 25%51((5'6 
25%5)#+0 
calibration screen.
7 . Perform PS calibration as follows:

1 ) Align the right cursor to the graduation of the scale with the right or left but-
ton.

2 ) Select the left cursor with the down button.

3 ) Align the left cursor to a position 5 mm away from the right cursor with the right or

left button.
8 . Press the SAVE button to save the setting.
9 . Measure the CS value at 12 mm and PS value at 5 mm, and confirm that the measured val-
ues are within the following the specifications.
1 ) CS: 12 mm ±0.2 mm
2 ) PS: 5 mm ±0.2 mm
10. If the CS or PS measurement values are not within the specifications, recalibrate the items
that are not within the specifications.

167 Return to previous page


XAR17ARDA001G/E

8.11 AUTO TRC ADJUST


* Perform adjustment only when the R/L motor (30602-EA26), F/B motor (30602-EA27),
or U/D motor (30601-EA28) is replaced.
1 . Attach the 22-model eye jig (32171-0600) (see 8.2.6 [p112]).
2 . Enter MENU mode, and select 11. TRC ADJUST MODE (see 8.1.1 [p103])

3 . Select AUTO TRACKING with the up 64%#&,756/1&'


.+/+6%'06'4
or down button, then press the exe- #76164#%-+0)
cute button.
0
4 . Confirm that the auto tracking operates as
follows:
1 ) Auto tracking moves smoothly in the
side-to-side, up-and-down, and for-
ward-and-backward directions.
2 ) Overshoot does not occur more than
four times.
5 . If the conditions described in Step 4 are
not met, change the four numeric values #76164%#&,756
in the second column from the left as
shown to the right according to the steps
below.
1 ) Select the numeric value with the up

or down button

a . The first row: upward 
b . The second row: downward 
5#8' 
c . The third row: left-and-right 

direction 
d . The fourth row: backward-
and-forward direction

2 ) Change the numeric value with the right , right + , left , or left - button.

* The right and left buttons increase or decrease the numeric value by 1

and the right + and left - buttons by 5.


* Increasing each numeric value accelerates the auto tracking speed and decreas-
ing the value slows down the speed.
* Increasing each numeric value too much destabilizes the movement of auto
tracking.

168 Return to previous page


XAR17ARDA001G/E

8.12 Software Upgrade


1 . Save the following upgrade files to the root of the USB flash drive.
1 ) For AR: ar17_softwrite_V***.mot
2 ) For ARK: ar17_softwrite_K***.mot
* *** indicates the version number.
* Do not create a folder to save the files
in it.
2 . Turn off power to the device.
3 . Connect the USB flash drive (32166-
E340) to the USB-A connector of the main
body.
* If the USB flash drive (32166-E340) is
not available, a USB flash drive made
by SanDisk is recommended.
* However, even if a SanDisk USB flash
drive is used, there is a possibility that
it may not be recognized by the
device.
4 . Turn on power to the device
5 . The screen for selecting file is displayed
when there are two or more upgrading
files in the USB flash drive (32166-E340).
* The screen for selecting file is dis-
8 %#0%'.
played only when two or more upgrad-
ing files are saved. 8
* When four or more upgrading files are
saved, no files from the fourth are rec- 8
ognized. Be sure to save three or less 5'.'%6(+.'
files.
6 . Select the file to be upgraded.

7 . The version upgrade confirmation screen


is displayed.
8 . Press the OK button.
8GT
1- %#0%'.

51(694+6'/1&'

169 Return to previous page


XAR17ARDA001G/E

9 . The screen as shown to the right appears


and upgrade starts.
8GT

4'#&+0) 94+6'

51(694+6'/1&'

10. When the upgrade is complete, the screen


as shown to the right appears. Turn off
power to the device.
9HU
11. Disconnect the USB flash drive.
12. Turn on power to the device, and confirm
that the device starts up properly. 5($',1* :5,7(
13. If upgrade fails, repeat it starting from
Step 1.
62)7:5,7(&203/(7(
32:(52))5(029(86%

170 Return to previous page


XAR17ARDA001G/E

8.13 Time and Date Settings


1 . Display the PARAMETER SETTING screen, and select CLOCK SET (see 8.1.5 [p105]).
2 . Move the cursor to the item to be changed
㧼㧭㧾㧭㧹㧱㨀㧱㧾‫ޓޓ‬㧿㧱㨀㨀㧵㧺㧳 .
with the up or down button. ‫ޓޓޓޓޓ‬㨇㧯㧸㧻㧯㧷㨉
3 . Change the numeric value with the right

or left button.

4 . Pressing the reset button changes the


㧞㧜㧜㧡㧛㧜㧥㧛㧝㧡‫ޓ‬㧝㧡㧦㧟㧜
format of the date in the order of “YY/MM/
DD” to “MM/DD/YY” to “DD/MM/YY”.
5 . Pressing 24H or 12H selects the 24-hour
or 12-hour indication.

6 . Press the exit button to return to the


PARAMETER SETTING screen.

7 . Press the exit button to return to the measurement screen.

171 Return to previous page


XAR17ARDA001G/E

8.14 CURSOR BOARD


1 . Display the CALIB MODE SELECT screen (for AR), and select CURSOR BOARD (see
8.1.2 [p104]).
2 . Confirm that pink lines are displayed on all
four edges of the color LCD (30601-
E005).
3 . If they are not displayed, adjust the posi- 
tion of the LCD front cover (30601-M715)
(see 6.10 [p67]).

172 Return to previous page


XAR17ARDA001G/E

8.15 Rewriting EEPROM Data


* Rewrite the EEPROM data only when the driver board (30601-BA03) is replaced.
* Improper operation may delete the adjustment data saved in the driver board (30601-
BA03).
1 . Enter MENU mode, and select 17. EEPROM ERROR DISP (see 8.1.1 [p103]).

2 . Press the reset button. ''241/'4414&+52


* When ERR001 occurs, the back- 
ground of the erroneous item turns (75'
&#
.'&
red. 64%8#.#4-A%*#46
#.+)0
4'6+%.'
5*+2/'06#4A%#.+$
#4A%144'
-/A%#.+$
-/A%144'
2&A%#.+$
%5A%#.+$
25A%#.+$

3 . Press the right or left button to ''241/18'494+6'


select Yes, then press the execute
button.
* The backup data is rewritten from the %#76+10
main board (30601-BA01) to the ''241/18'494+6'
driver board (30601-BA03). $#%-72&#6#!
;GU0Q

4 . When the screen as shown to the right is


''241/18'494+6'
displayed, press the exit button twice
to restart the device.
%#76+10
''241/94+6+0)019

94+6+0)1-

173 Return to previous page


XAR17ARDA001G/E

8.16 Setting Network


1 . Display the PARAMETER SETTING
screen, and select NETWORK (see 8.1.5
2#4#/'6'45'66+0)
=0'6914-?
.
[p105]). 0'6914-01
&*%201
2 . Press the up or down button to +2
/#5-
select the desired item. 75'4)7'56
2#55914&
3 . Set the network. &1/#+0914-)4172
1 ) Select 111. NETWORK. 2%0#/'2%
(1.&'4&#6#
+/#)'5'0&01
2 ) Press the right or left button
to switch to YES or NO. 6'56
a . YES: Establishes connection
to the network.
b . NO: Disconnects connection to the network.
4 . Set DHCP.
1 ) Select 112. DHCP.

2 ) Press the right or left button to switch to YES or NO.


a . ON: Establishes connection to DHCP.
b . OFF: Disconnects connection to DHCP.
5 . Set the IP address.
* When 112. DHCP is set to YES, 113. IP indicates “0.0.0.0” and it cannot be selected.
1 ) Select 113. IP.

* Selecting 113. IP displays the execute button.

2 ) Press the execute button to dis- 2#4#/'6'45'66+0) .


play the IP address entry screen. =0'6914-+2?
3 ) Change the IP address as follows:

a . Move the cursor with or

. +2#&&4'5
b . Change the numerical value 
with the right button, left

button, or joystick.

c . Press the exit button to 2#4#/'6'45'66+0) .


=0'6914-+2?
display the screen for confirm-
ing the changed IP address.
d . Select YES or NO with the %#76+10

right button, left but- 5'6+2#&&4'55!


ton, or joystick.
;'501
e . Press the execute button
or start button to confirm YES
or NO.

174 Return to previous page


XAR17ARDA001G/E

6 . Set the subnet mask.


* When 112. DHCP is set to YES, 114. MASK indicates “0.0.0.0” and it cannot be
selected.
1 ) Select 114. MASK.
2 ) Set the subnet mask in the same steps as those of the IP address.
7 . Set the user name
* Enter a user name that is registered in the connected PC.
1 ) Select 115. USER.

* Selecting 115. USER displays the execute button.

2 ) Press the execute button to dis- 2#4#/'6'45'66+0) .


=0'6914-75'4?
play the user name entry screen. ‫ޓ‬
3 ) Enter the user name as follows: #$%&'()*+,-
./0123456789:;<A
a . Move the cursor in the charac-

ter list field with , or , 75'40#/'


or joystick.
)7'56
b . Move the cursor in the entry

field with the right button,

left button, or joystick.

c . Press the OK button to


confirm the character to be entered. Then move the cursor in the entry field one
character to the right.

d . Press the delete button to delete a character in the entry field. Then move
the cursor in the entry field one character to the left.

e . Press the exit button to 2#4#/'6'45'66+0)


=0'6914-75'4?
.
display the screen for confirm-
ing the changed user name.
f . Select YES or NO with the %#76+10
5'675'40#/'!
right button, left but-
ton, or joystick.
;'501
g . Press the execute button
or start button to confirm YES
or NO.

8 . Set the password.


* Enter the login password for the user name of the connected PC.
1 ) Select 116. PASSWORD.

* Selecting 116. PASSWORD displays the execute button.


2 ) Set the password in the same steps as those of the user name.

175 Return to previous page


XAR17ARDA001G/E

9 . Set the domain name.


* Enter the domain name of the connected PC.
1 ) Select 117. DOMAIN.

* Selecting 117. DOMAIN displays the execute button.


2 ) Set the domain name in the same steps as those of the user name.
10. Set the PC name.
* Enter the name of the connected PC.
1 ) Select 118. PC NAME.

* Selecting 118. PC NAME displays the execute button.


2 ) Set the PC name in the same steps as those of the user name.
11 . Folder name setting
* Enter the shared folder name of the connected PC that the measurement data is trans-
mitted to.
1 ) Select 119. FOLDER.

* Selecting 119. FOLDER displays the execute button.


2 ) Set the folder name in the same steps as those of the user name.
12. Set the image to be sent.
1 ) Select 120. IMAGE SEND.

2 ) Select YES, NO, or LOW CONF with the ight button, left button, or joystick.
a . YES: Sends the measurement ring image.
b . NO: Does not send the measurement ring image.
c . LOW CONF: Sends the measurement ring image when the confidence index is
7 or less.
13. Check the network connection.
1 ) Press the TEST button.
2 ) The following message is displayed for about two seconds along with a beep.
a . When the LAN cable is connected prop-
erly, “CONNECTION OK” is displayed.
%100'%6+101-
b . If the LAN cable is not connected prop-
erly, “ERR771 NO NETWORK CABLE” 㧱㧾㧾㧣㧣㧝
is displayed. 㧺㧻‫ޓ‬㧺㧱㨀9㧻㧾㧷‫ޓ‬㧯㧭㧮㧸㧱

176 Return to previous page


XAR17ARDA001G/E

9 SUPPLEMENT

9.1 Appearance
9.1.1 Front view

Function buttons

LCD

Memory indicator

Cover open button

Printer cover

Start button

Joystick

Eye Care card slot

Locking lever

Power switch

177 Return to previous page


XAR17ARDA001G/E

9.1.2 Rear view

Forehead rest

Eye level
marker
Measuring window Chinrest
up/down
button

Chinrest

PD window

178 Return to previous page


XAR17ARDA001G/E

9.2 Labels
Front and side views

Destination Label Marco label


For countries other than
–––––– ––––––
USA, Canada, and China

USA

China ––––––

Portuguese regions –––––– ––––––

Canada ––––––

179 Return to previous page


XAR17ARDA001G/E

Rear view

72120(75(6$16&217$&7
72120(752$62)),2
121&217$&772120(7(5
721•0(752'(12&217$&72
0$

&/$66/('352'8&7

웢 웳월월웑웢
$8725()5$&720(7(5 For technical assistance please call:
,1387 9윟+]9$ 㻝㻙㻤㻜㻜㻙㻤㻣㻠㻙㻡㻞㻣㻠
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0'

180 Return to previous page


XAR17ARDA001G/E

Destination For countries other than USA, Canada, and China USA

&/$66/('352'8&7

웢 웳월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-310A
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30601-M803-C

&/$66/('352'8&7

웢 웳월월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-330A
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30602-M803-C

&/$66/('352'8&7

웢 웳월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-360A ––––––
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0&

ID
label
&/$66/('352'8&7

㸿 㹐㹉㸳㸯㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
ARK-510A
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30701-M803-C

&/$66/('352'8&7

㸿 㹐㹉㸳㸱㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
ARK-530A
6(512 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30702-M803-C

&/$66/('352'8&7

㸿 㹐㹉㸳㸴㸮㸿
$8725().(5$720(7(5
INPUT 9㹼+]9$
ARK-560A
SER.NO. 1111 ;;;;

0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0&

72120(75(6$16&217$&7
72120(752$62)),2
Multilingual label 121&217$&772120(7(5 ––––––
721•0(752'(12&217$&72
0$

181 Return to previous page


XAR17ARDA001G/E

Destination For countries other than USA, Canada, and China USA

Type B applied part label ––––––

For technical assistance please call:

MarcoFreeDial ––––––
㻝㻙㻤㻜㻜㻙㻤㻣㻠㻙㻡㻞㻣㻠

Chinese RoHS label 10 –––––– ––––––

182 Return to previous page


XAR17ARDA001G/E

Destination Canada China

㸿 㹐㸱㸯㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹐㸿㹁㹒㹍㹋㹃㹒㹃㹐
ൟো $5$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠‫ܝ‬Ҿ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$

AR-310A 6(512 1111 ;;;;


ᅝܼ㉏ൟ˖,㉏%ൟ
䇋࣓⫼࡯⺄ᩲ

䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼‫ݠ‬ᷛ‫ޚ‬㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼‫ݠ‬䆕к㓪ো˖೑亳㥃ⲥẄ 䖯 ᄫ㄀ো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⷇⬎ࠡ⌰⬾ഄ 0$

㸿 㹐㸱㸱㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹐㸿㹁㹒㹍㹋㹃㹒㹃㹐
රਧ $5$
ӗ૱਽〠˖‫ޘ‬㠚ࣘ⭥㝁傼‫ݹ‬Ԛ
,1387 9㹼+]9$
䗃‫⭥ޕ‬Ⓚ˖$&9+]9$

AR-330A 6(512 1111 ;;;;


ᆹ‫ޘ‬㊫ර˖,㊫%ර
䈧य⭘࣋⻠ᫎ

䘀㹼ᯩᔿ˖䘎㔝䘀㹼 1R1111
0$(+$0$ ⌘޼ḷ߶㕆ਧ˖<=%-$3
+,52,6+,&+2
*$0$*25,
५⯇ಘỠ⌘޼䇱Җ㕆ਧ˖ഭ伏㦟ⴁỠ 䘋 ᆇㅜਧ
$,&+,-$3$1 ⭏ӗԱъ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ൠ൰˖⡡⸕৯㫢䜑ᐲ᤮⸣⭪ࡽ⎌⮚ൠ 0$

㸿 㹐㸱㸴㸮㸿 &/$66/('352'8&7

㸿㹓㹒㹍 㹐㹃㹄㹐㸿㹁㹒㹍㹋㹃㹒㹃㹐 웢 웳월웑웢


$8725()5$&720(7(5
,1387 9㹼+]9$
,1387 9윟+]9$
AR-360A 6(512 1111 ;;;;
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2 0$(+$0$
*$0$*25, +,52,6+,&+2
$,&+,-$3$1 *$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0$ 0$'(,1-$3$1 0&

ID
label
㸿 㹐㹉㸳㸯㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐

ൟো $5.$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠‫ܝ‬Ҿ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$

ARK-510A 6(512 1111 ;;;;


ᅝܼ㉏ൟ˖,㉏%ൟ
䇋࣓⫼࡯⺄ᩲ

䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼‫ݠ‬ᷛ‫ޚ‬㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼‫ݠ‬䆕к㓪ো˖೑亳㥃ⲥẄ 䖯 ᄫ㄀ো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⷇⬎ࠡ⌰⬾ഄ 0$

㸿 㹐㹉㸳㸱㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐

ൟো $5.$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠‫ܝ‬Ҿ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$

ARK-530A 6(512 1111 ;;;;


ᅝܼ㉏ൟ˖,㉏%ൟ
䇋࣓⫼࡯⺄ᩲ

䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼‫ݠ‬ᷛ‫ޚ‬㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼‫ݠ‬䆕к㓪ো˖೑亳㥃ⲥẄ 䖯 ᄫ㄀ো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⷇⬎ࠡ⌰⬾ഄ 0$

㸿 㹐㹉㸳㸴㸮㸿 &/$66/('352'8&7

㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐

㸿 㹐㹉㸳㸴㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
INPUT 9㹼+]9$
ARK-560A 6(512 1111 ;;;;
SER.NO. 1111 ;;;;
0$(+$0$
+,52,6+,&+2 0$(+$0$
*$0$*25, +,52,6+,&+2
$,&+,-$3$1 *$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0$ 0$'(,1-$3$1 0&

Multilingual label –––––– ––––––


Type B applied part label –––––– ––––––
MarcoFreeDial –––––– ––––––

Chinese RoHS label 10 ––––––

183 Return to previous page


XAR17ARDA001G/E

Underside view
For USA, Canada, and China

For countries other than USA, Canada, and China

184 Return to previous page


XAR17ARDA001G/E

9.3 Wiring Diagram


9.3.1 Complete wiring diagram
9.3.1.1 30602-E000 (AR-330A and ARK-530A)

185 Return to previous page


XAR17ARDA001G/E

9.3.1.2 30601-E000 (AR-310A and ARK-510A)

186 Return to previous page


XAR17ARDA001G/E

9.3.1.3 30603-E000 (AR-360A) and 30703-E000 (ARK-560A)

187 Return to previous page


XAR17ARDA001G/E

9.3.2 Partial writing diagram


30601-EA01

30601-EA11

30601-EA12

188 Return to previous page


XAR17ARDA001G/E

30601-EA13

Cable 1: P301 A row side


Cable 2: P301 B row side

189 Return to previous page


XAR17ARDA001G/E

30601-EA14

190 Return to previous page


XAR17ARDA001G/E

30601-EA20 (AR-330A, AR-360A, ARK-530A, and ARK-560A only)

30601-EA21 (AR-330A, AR-360A, ARK-530A, and ARK-560A only)

30601-EA22 (AR-330A, AR-360A, ARK-530A, and ARK-560A only)

191 Return to previous page


XAR17ARDA001G/E

30601-EA23

30601-EA24

192 Return to previous page


XAR17ARDA001G/E

30601-EA25

30601-EA26 (AR-330A, AR-360A, ARK-530A, and ARK-560A only)

30602-EA27 (AR-330A, AR-360A, ARK-530A, and ARK-560A only)

30601-EA28

193 Return to previous page


XAR17ARDA001G/E

30601-EA30

30601-EA31

30601-EA32

194 Return to previous page


XAR17ARDA001G/E

30601-EA34

30601-EA36

30601-EA39

30601-EA42

30601-EA44

195 Return to previous page


XAR17ARDA001G/E

30601-EA51

30601-EA52

196 Return to previous page


XAR17ARDA001G/E

30601-EA53

(Motor is included.)

30601-EA54

30601-EA55

197 Return to previous page


XAR17ARDA001G/E

30601-EA63

30601-EA64

30601-EA65

30601-EA72

30601-EA75

198 Return to previous page


XAR17ARDA001G/E

30601-EA76

30601-EA80

30601-EA81

30601-EA82

30601-EA83

199 Return to previous page


XAR17ARDA001G/E

30601-EA93

type

30703-EA37(AR-360A and ARK-560A only)

(With motor)

(With motor)

200 Return to previous page


XAR17ARDA001G/E

30703-EA38 (AR-360A and ARK-560A only)

30703-EA74 (AR-360A and ARK-560A only)

(With motor)

201 Return to previous page


XAR17ARDA001G/E

9.4 Error Messages


Error codes Contents Details
ERR001 EEPROM error Defective EEPROM data
ERR002 Clock error Defective IC data of the clock
PD ERROR PD error Defective PD CCD, LED does not light up.
ERR011 Time-out error in characters (OUT) Communication breaks while receiving data.
ERR012 Time-out error for receiving data (OUT) Time-out after the DTR-DSR takes effect.
ERR013 Receiving error (OUT) Parity and flaming error
ERR014 Receiving code error (OUT) The received code is defective.
ERR015 Reply time-out (OUT) Time-out error between sending and receiving
ERR016 DTR-DSR error (OUT) Time-out between DTR and DSR
ERR017 Data error (OUT) Data error (Improper received data)
ERR018 Command error (OUT) Command error (Improper command)
NO DAT Data empty (OUT) No measurement data
ERR021 Time-out error in characters (IN) DSR is not flagged while receiving SD command.
ERR022 Time-out error for receiving data (IN) DSR is not flagged while sending SD command.
ERR023 Receiving error (IN) The received data does not come.
ERR024 Receiving code error (IN) The receiving buffer is full.
ERR025 Reply time-out (IN) An unreadable command was entered.
ERR026 DTR-DSR error (IN) -
ERR027 Data error (IN) -
ERR028 Command error (IN) -
ERR031 Up-and-down tracking error Time-out for up-and-down tracking
ERR032 Left-to-right tracking error Time-out for left-to-right tracking
ERR033 Back-and-forth tracking error Time-out for back-and-forth tracking
ERR034 Up/down chinrest error Time-out for up-and-down movement of the chinrest
The printer is short of paper, or the printer head is raised.
NO PAPER Printer paper shortage error
The printer head is raised.
ERR043 Printer connection error Defective connection of printer or malfunction
ERR061 Model code mismatching error -
The model of the main board or driver board assembled into the
ERR062 Model code mismatching error
AR series or ARK series differs from the model of the device.
The main board for AR series and driver board for ARK series
ERR063 Model code mismatching error
are assembled.
The driver board for AR series and main board for ARK series
ERR064 Model code mismatching error
are assembled.
ERR065 Model code mismatching error The driver boards for AR series are replaced.
ERR066 Model code mismatching error The driver boards for ARK series are replaced.
ERR067 Model code mismatching error The main boards for AR series are replaced.
ERR068 Model code mismatching error The main boards for ARK series are replaced.
ERR101 Sensor initialization error The sensor does not work.
ERR111 Thermister error The thermister is defective.
ERR112 AR motor error Improper numbers of rotation in AR motor
ERR121 CYL1 initialization error The initialization of subjective CYL lens 1 cannot be completed.
ERR122 CYL2 initialization error The initialization of subjective CYL lens 2 cannot be completed.
ERR123 Chart initialization error The initialization of subjective chart cannot be completed.
ERR501 Power error of Eye Care card The Eye Care card is defective.
ERR502 Writing error of Eye Care card The Eye Care card is removed while writing.
ERR503 Power error of Eye Care card The Eye Care card is removed after writing.
ERR601 USB_A recognition error The USB host fails to be recognized.
ERR602 USB_A class error The USB host class is improper.
ERR700 File sharing error Error related to Windows file sharing

202 Return to previous page


XAR17ARDA001G/E

Error codes Contents Details


ERR703 Hardware error Error related to IC board
ERR704 DHCP error The IP address cannot be obtained.
ERR750 Network access error The network cannot be accessed.
ERR751 Network writing error The data cannot be written to the PC.
ERR754 PC name error The PC with the specified name does not exist.
ERR755 Read-only folder error The shared folder in the computer is designated as read-only.
Logging onto the PC is not successful. (The user name or
ERR756 Log-on error
password is incorrect.)
The shared folder does not exist. (The name of the shared
ERR757 Shared folder error
folder is incorrect.)
Error related to time-out (The PC cannot complete the process
ERR758 Time-out error
within a time frame.)
ERR759 Deletion error The data in PC cannot be deleted.
The network initialization is in progress.(The initialization
ERR760 Error during network initialization
requires some time after the device starts up.)
ERR761 Access right error No permission to access the shared folder
ERR762 Account error The account is disabled. (The user setting is improper.)
ERR771 Cable connection error The LAN cable is not connected.
ERR772 Reply error No reply received from PC

*For the details of ERR061 to ERR068 caused by the combination of the main board and driver
board, see the table below.
Main board (30601-BA01)
-
Unwritten AR-310A AR-330A AR-360A ARK-510A ARK-530A ARK-560A
Unwritten ERR061 ERR065 ERR065 ERR065 ERR066 ERR066 ERR066
AR-310A ERR067 - ERR062 ERR062 ERR064 ERR064 ERR064

Driver board AR-330A ERR067 ERR062 - ERR062 ERR064 ERR064 ERR064


(30601- AR-360A ERR067 ERR062 ERR062 - ERR064 ERR064 ERR064
BA03) ARK-510A ERR068 ERR063 ERR063 ERR063 - ERR062 ERR062
ARK-530A ERR068 ERR063 ERR063 ERR063 ERR062 - ERR062
ARK-560A ERR068 ERR063 ERR063 ERR063 ERR062 ERR062 -

203 Return to previous page


XAR17ARDA001G/E

9.5 Parameter List


9.5.1 AR-310A/AR-330A
Countries other than
Parameter item No. Item USA China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
21 COMPARE SW MANUAL MANUAL MANUAL
COMPARE 22 COMPARE(AR) SE SE SE
(AR-330A only) 23 ADD SW YES YES YES
24 ADD SELECT 1.75 D 1.75 D 1.75 D
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
PRINT2 44 ERROR DATA YES YES YES
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY FINE FINE FINE
69 MAN FOCUS DISP YES YES YES

204 Return to previous page


XAR17ARDA001G/E

Parameter item No. Item USA Countries other than China


USA and China
71 I/F MODE NIDEK NIDEK NIDEK
72 I/F FORMAT SHORT SHORT SHORT
73 BAUD-RATE 9600 9600 9600
COMMUNICATION
74 BIT LENGTH 8 8 8
75 CR CODE NO NO NO
76 LM DATA PRINT NO NO NO
101 READER START 1 1 1
READER
102 READER LENGTH 14 14 14
111 NETWORK NO NO NO
112 DHCP NO NO NO
113 IP 192.168.0.20 192.168.0.20 192.168.0.20
114 MASK 255.255.255.0 255.255.255.0 255.255.255.0
115 USER GUEST GUEST GUEST
NETWORK
116 PASSWORD ---------- ---------- ----------
117 DOMAIN WORKGROUP WORKGROUP WORKGROUP
118 PC NAME PC PC PC
119 FOLDER DATA DATA DATA
120 IMAGE SEND NO NO NO
CLOCK SET CLOCK SET ---------- ---------- ----------
COMMENT SET COMMENT SET ---------- ---------- ----------
CL LIST SET CL LIST SET ---------- ---------- ----------

205 Return to previous page


XAR17ARDA001G/E

9.5.2 AR-360A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT OBJ->SBJ OBJ->SBJ OBJ->SBJ
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY FINE FINE FINE
69 MAN FOCUS DISP YES YES YES
71 I/F MODE NIDEK NIDEK NIDEK
72 I/F FORMAT SHORT SHORT SHORT
73 BAUD-RATE 9600 9600 9600
COMMUNICATION
74 BIT LENGTH 8 8 8
75 CR CODE NO NO NO
76 LM DATA PRINT NO NO NO

206 Return to previous page


XAR17ARDA001G/E

Parameter item No. Item USA Countries other than China


USA and China
91 AUTO SUBJECT NO NO NO
92 SUBJECT CHART 25 0.8 0.8
93 NEAR CHART 30 0.63 0.63
94 VA DISPLAY FRAC. DEC DEC
SUBJECT 95 SUBJECT START AR AR AR
96 NEAR RECALL NO NO NO
97 WD CHANGE NO NO NO
98 UCVA CHART ALL ALL ALL
99 ADD SELECT 1.75 D 1.75 D 1.75 D
101 READER START 1 1 1
READER
102 READER LENGTH 14 14 14
111 NETWORK NO NO NO
112 DHCP NO NO NO
113 IP 192.168.0.20 192.168.0.20 192.168.0.20
114 MASK 255.255.255.0 255.255.255.0 255.255.255.0
115 USER GUEST GUEST GUEST
NETWORK
116 PASSWORD ---------- ---------- ----------
117 DOMAIN WORKGROUP WORKGROUP WORKGROUP
118 PC NAME PC PC PC
119 FOLDER DATA DATA DATA
120 IMAGE SEND NO NO NO
CLOCK SET CLOCK SET ---------- ---------- ----------
COMMENT SET COMMENT SET ---------- ---------- ----------
CL LIST SET CL LIST SET ---------- ---------- ----------

207 Return to previous page


XAR17ARDA001G/E

9.5.3 ARK-510A/ARK-530A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
11 KM UNIT D mm mm
12 KM DISPLAY R1, R2 R1, R2 R1, R2
KM 13 REF. INDEX 1.3375 1.3375 1.3375
14 KM CONTINUE 3 3 3
15 PERIPHERAL NO NO NO
21 COMPARE SW MANUAL MANUAL MANUAL
22 COMPARE(AR) SE SE SE
COMPARE
23 ADD SW YES YES YES
24 ADD SELECT 1.75 D 1.75 D 1.75 D
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT R->L R->L R->L
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY NORMAL NORMAL NORMAL
69 MAN FOCUS DISP YES YES YES

208 Return to previous page


XAR17ARDA001G/E

Parameter item No. Item USA Countries other than China


USA and China
71 I/F MODE NIDEK NIDEK NIDEK
72 I/F FORMAT SHORT SHORT SHORT
73 BAUD-RATE 9600 9600 9600
COMMUNICATION
74 BIT LENGTH 8 8 8
75 CR CODE NO NO NO
76 LM DATA PRINT NO NO NO
81 SAGITTAL YES YES YES
SAGITTAL 82 SAGIT AXIS AXIS AXIS AXIS
83 SAGIT PRINT ALL ALL ALL
101 READER START 1 1 1
READER
102 READER LENGTH 14 14 14
111 NETWORK NO NO NO
112 DHCP NO NO NO
113 IP 192.168.0.30 192.168.0.30 192.168.0.30
114 MASK 255.255.255.0 255.255.255.0 255.255.255.0
115 USER GUEST GUEST GUEST
NETWORK
116 PASSWORD ---------- ---------- ----------
117 DOMAIN WORKGROUP WORKGROUP WORKGROUP
118 PC NAME PC PC PC
119 FOLDER DATA DATA DATA
120 IMAGE SEND NO NO NO
CLOCK SET CLOCK SET ---------- ---------- ----------
COMMENT SET COMMENT SET ---------- ---------- ----------
CL LIST SET CL LIST SET ---------- ---------- ----------

209 Return to previous page


XAR17ARDA001G/E

9.5.4 ARK-560A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
11 KM UNIT D mm mm
12 KM DISPLAY R1, R2 R1, R2 R1, R2
KM 13 REF. INDEX 1.3375 1.3375 1.3375
14 KM CONTINUE 3 3 3
15 PERIPHERAL NO NO NO
21 COMPARE SW MANUAL MANUAL MANUAL
22 COMPARE(AR) SE SE SE
COMPARE
23 ADD SW YES YES YES
24 ADD SELECT 1.75 D 1.75 D 1.75 D
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT AR->KM AR->KM AR->KM
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY NORMAL NORMAL NORMAL
69 MAN FOCUS DISP YES YES YES

210 Return to previous page


XAR17ARDA001G/E

Parameter item No. Item USA Countries other than China


USA and China
71 I/F MODE NIDEK NIDEK NIDEK
72 I/F FORMAT SHORT SHORT SHORT
73 BAUD-RATE 9600 9600 9600
COMMUNICATION
74 BIT LENGTH 8 8 8
75 CR CODE NO NO NO
76 LM DATA PRINT NO NO NO
81 SAGITTAL YES YES YES
SAGITTAL 82 SAGIT AXIS AXIS AXIS AXIS
83 SAGIT PRINT ALL ALL ALL
91 AUTO SUBJECT NO NO NO
92 SUBJECT CHART 25 0.8 0.8
93 NEAR CHART 30 0.63 0.63
94 VA DISPLAY FRAC. DEC DEC
SUBJECT 95 SUBJECT START AR AR AR
96 NEAR RECALL NO NO NO
97 WD CHANGE NO NO NO
98 UCVA CHART ALL ALL ALL
99 ADD SELECT 1.75D 1.75D 1.75D
101 READER START 1 1 NO
READER
102 READER LENGTH 14 14 14
111 NETWORK NO NO NO
112 DHCP NO NO NO
113 IP 192.168.0.30 192.168.0.30 192.168.0.30
114 MASK 255.255.255.0 255.255.255.0 255.255.255.0
115 USER GUEST GUEST GUEST
NETWORK
116 PASSWORD ---------- ---------- ----------
117 DOMAIN WORKGROUP WORKGROUP WORKGROUP
118 PC NAME PC PC PC
119 FOLDER DATA DATA DATA
120 IMAGE SEND NO NO NO
CLOCK SET CLOCK MODE ---------- ---------- ----------
COMMENT SET COMMENT SET ---------- ---------- ----------
CL SELECT CL SELECT ---------- ---------- ----------

211 Return to previous page


XAR17ARDA001G/E

9.6 Jig List


Part No. Name Part No. Name Part No. Name
High-power spherical
30670-0100 - -
model eye
30670-0200 Small pupil model eye - -
30670-0300 KM calibration jig - -
30670-1100 Joint jig
30670-1200 Reticle illumination jig
30670-1300 Reticle observation jig
30670-1400 Camera jig
30670-1500 Ring lens illumination jig
Optical adjustment jig 30670-1600 Diffuser
30670-1000
set 30670-1800 Chart lens P
30670-1900 Chart lens M
30670-M029 Reflector plate
30670-M031 Holder
32177-0618 0-D model eye 2
99104-G020 Dioptric tester
32171-0600 22-model eye jig
30670-0000 AR17DJ-1set 32107-1100 Chinrest attachment jig
32171-1200 Service model eye
30670-5200 Service calibration jig 32177-1400 Aspheric model eye
32177-1500 Cylinder axis adjustment jig
Attachment jig for service
32107-6300
model eye
32177-0618 0-D model eye 2
30670-1100 Joint jig
30670-1200 Reticle illumination jig
Service optical
30670-5300 30670-BA01 Relay board
adjustment jig
30670-E001 Flexible flat cable
30670-EA06 Relay cable 3
32557-EA11 EA39 extension cable
30670-BA01 Relay board
30670-E001 Flexible flat cable
30670-EA01 Relay cable ASSY
30670-EA04 Relay cable 1
30670-EA05 Relay cable 2
30601-E030 USB flash drive - -
Dioptric tester jig
30670-0600 30670-2300 Chart test jig - -
ASSY

212 Return to previous page


Adjustment item

8.16 Setting Network (p174)


8.5 Chart Adjustment (p124)
Replacement part

8.9 PD CALIBRATION (p163)


8.7 AR CALIBRATION (p130)
8.8 KM CALIBRATION (p156)

8.10 CS/PS Calibration (p167)

8.14 CURSOR BOARD (p172)


8.1.5 Parameter settings (p105)

8.11 AUTO TRC ADJUST (p168)


8.13 Time and Date Settings (p171)

8.15 Rewriting EEPROM Data (p173)


8.3.1 Observation CCD camera (p117)

8.4 Phototransmitter Adjustment (p122)

8.6 ALIGNMENT CALIBRATION (p129)


8.3.2 Measurement CCD camera (p121)
7.1.1 Joystick ASSY (30601-2600) (p69)
9.7 Adjustment Item List

o
o
o
o
o
7.1.2 Measuring ASSY (30601-9E00) (p70)

o
o
o
7.1.3 Phototransmitter ASSY (30601-3310) (p70)

o
o
7.1.4 Mire ASSY (30601-4000) (p71)
7.1.5 LCD ASSY(30601-6100) (p71)

213
o
7.1.6 Subjective cylinder ASSY (30703-3700) (p72)

o
7.1.7 Chart disk ASSY (30703-3800) (p72)
o

7.2.1 Main board (p73)


7.2.2 Tracking board (30601-BA02) (p74)
9.7.1 ASSYs, boards, and adjustment items

o
7.2.3 Driver board (30601-BA03) (p75)
7.2.4 Base board (30601-BA05) (p76)
7.2.5 LCD board (30601-BA06) (p77)
7.2.6 Left panel SW board (30601-BA07) (p77)
7.2.7 U/D SW board (30601-BA09) (p78)
7.2.8 Sensor carry board (30601-BA14) (p78)
7.2.9 Right panel SW board (30601-BA15) (p79)
o

7.2.10.1 BA08 PD CCD board (32105-BA08) (p80)


7.2.11 DC-AC inverter (18535-E010) (p81)
7.2.12 Printer interface board (80606-00019) (p81)

Return to previous page


XAR17ARDA001G/E
Adjustment item

(p129)
(p122)
(p121)
XAR17ARDA001G/E

Replacement part

8.16 Setting Network (p174)


8.5 Chart Adjustment (p124)

8.9 PD CALIBRATION (p163)


8.7 AR CALIBRATION (p130)
8.8 KM CALIBRATION (p156)

8.10 CS/PS Calibration (p167)

8.14 CURSOR BOARD (p172)


8.4 Phototransmitter Adjustment

8.6 ALIGNMENT CALIBRATION

8.11 AUTO TRC ADJUST (p168)


8.3.2 Measurement CCD camera

8.13 Time and Date Settings (p171)

8.15 Rewriting EEPROM Data (p173)


8.3.1 Observation CCD camera (p117)
7.3.1 R/L sensor (30602-EA20) (p82)
7.3.2 F/B sensor (30602-EA21) (p82)
7.3.3 U/D sensor (30601-EA22) (p83)

o
7.3.4 R/L motor (30602-EA26) (p84)

o
7.3.5 F/B motor (30602-EA27) (p85)

o
7.3.6 U/D motor (30601-EA28) (p85)

214
o
7.3.7 Origin sensor (30601-EA36) (p86)

o
7.3.8 CL motor (30703-EA37) (p86)

o
7.3.9 CL sensor (30703-EA38) (p87)

o
o
7.3.10 Focus LED(30601-EA42) (p88)

o
o
7.3.11 KM6 LED (30601-EA44) (p88)
7.3.12 Chinrest (30601-EA53) (p89)

o
7.3.13 PD LED (30601-EA55) (p90)
9.7.2 Electrical components and adjustment items

7.3.14 Chart LED (30601-EA76) (p91)


o

7.3.15 Chart (30703-EA74) (p92)


o
o

7.3.16.1 Measurement CCD camera (p93)


o
o
o
o

7.3.16.2 Observation CCD camera (p93)

o
7.3.17 Color LCD (30601-E005) (p94)
7.3.18 Brushless motor (30601-E011) (p94)
o

7.3.19 Sensor motor (30601-E035) (p95)

Return to previous page


7.3.20 Switching power supply (80602-00102) (p96)
7.3.21 Printer (80606-00018) (p97)
XAR17ARDA001G/E

9.7.3 Mechanical components and adjustment items

7.4.5.1 Linear bearing for F/B shafts (82001-LM028) (p101)


7.4.5.2 Linear bearing for R/L shaft (82001-LM097) (p101)
7.4.3 Sliding plate (32105-M102) (p99)
7.4.1 R/L shaft (30601-M201) (p98)
7.4.2 F/B shaft (30601-M103) (p99)

7.4.4 Wire (30601-M307) (p100)


Replacement part

8.3.1 Observation CCD camera (p117)


8.3.2 Measurement CCD camera (p121)
8.4 Phototransmitter Adjustment (p122)
8.5 Chart Adjustment (p124)
8.6 ALIGNMENT CALIBRATION (p129)
Adjustment item

8.7 AR CALIBRATION (p130) o


8.8 KM CALIBRATION (p156)
8.9 PD CALIBRATION (p163)
8.10 CS/PS Calibration (p167)
8.11 AUTO TRC ADJUST (p168)
8.13 Time and Date Settings (p171)
8.14 CURSOR BOARD (p172)
8.15 Rewriting EEPROM Data (p173)
8.16 Setting Network (p174)

215 Return to previous page


XAR17ARDA001G/E

9.8 Screw List

ࢿࢪ㢮グྕ⾲6&5(:6/,67

ฎ⌮ 㼀㼞㼑㼍㼠㼙㼑㼚㼠
㯮ࢡ࣮࣒࣓ࣟࢵ࢟ ࢫࢸࣥࣞࢫ ୕౯㯮ࢡ࣓࣮ࣟࢺ
%&U 㻯㼔㼞㼛㼙㼑㻌㼜㼘㼍㼠㼑㻦㼎㼘㼍㼏㼗 686 㻿㼠㼍㼕㼚㼘㼑㼟㼟 ͤ=Q%U
㼀㼞㼕㼢㼍㼘㼑㼚㼠㻌㼏㼔㼞㼛㼙㼍㼠㼑㻦㼎㼘㼍㼏㼗
ࢡ࣮࣒࣓ࣟࢵ࢟ ୕౯ࢡ࣓࣮ࣟࢺ 䈜㻔㼆㼚㻮㼞㻟䛿┬␎㻕 㻔㼆㼚㻮㼞㻟㻌㼛㼙㼕㼠㼠㼑㼐㻕
&U 㻯㼔㼞㼛㼙㼑㻌㼜㼘㼍㼠㼑 =Q&
㼀㼞㼕㼢㼍㼘㼑㼚㼠㻌㼏㼔㼞㼛㼙㼍㼠㼑

ࣆࣥᖹ⾜
1 ࢼࢵࢺභゅ ✀
+(;$*21187 %6 ࢭ࣒ࢫࢿࢪ 6:3:௜
6(06)$67(1(5 ZLWK6:3: +% භゅ࣎ࣝࢺ
+(;$*21+($'%2/7
63 675$,*+73,1

ࢼࢵࢺභゅ ✀ ᑠࢿࢪ ␒✀࡞࡭ ࣡ࢵࢩࣕࢫࣉࣜࣥࢢ


1 +(;$*21187 &. 0$&+,1(6&5(: 3DQKHDG
ᑠࢿࢪභゅ㢌ࣇࣛࣥࢪ௜ࡁ
+(;$*21+($'%2/7
6: 635,1*:$6+(5

࣡ࢵࢩࣕᖹ ✀ ᑠࢿࢪ ␒✀࡞࡭ ࢩࣥ࣊ࢵࢻࢿࢪ


3: :$6+(5 &. 0$&+,1(6&5(: 3DQKHDG ++ Ṇࡵࢿࢪභゅ✰௜
+(;$*2162&.(76(76&5(: 7% 0$&+,1(6&5(: 7KLQKHDG

࣡ࢵࢩࣕᖹ ≉Ṧᑍἲ Ṇࡵ㍯㹁ᆺ ᑠࢿࢪࢺࣛࢫ


: :$6+(5 6SHFLDOVL]H &7 5(7$,1,1*5,1*&7<3( 2& ᑠࢿࢪ୸─
0$&+,1(6&5(: 2YDOKHDG 7& 0$&+,1(6&5(: 7UXVVKHDG

࣡ࢵࢩࣕ ࣏ࣜࢫࣛ࢖ࢲ࣮ ࢭ࣒ࢫࢿࢪ 7:௜ Ṇࡵࢿࢪࡍࡾ๭௜


: :$6+(5 3RO\VOLGHU '6 6(06)$67(1(5 ZLWK7: 3& ᑠࢿࢪ࡞࡭
0$&+,1(6&5(: 3DQKHDG 7* 6/277('6(76&5(:

࣡ࢵࢩࣕ ࢼ࢖ࣟࣥ Ṇࡵ㍯㹃ᆺ Ṇࡵࢿࢪභゅ✰௜


: :$6+(5 1\ORQ (7 5(7$,1,1*5,1*(7<3( 37 ࢱࢵࣉࢱ࢖ࢺࢿࢪ ࡞࡭
7$33,1*6&5(: 3DQKHDG 7+ +(;$*2162&.(76(76&5(:

ࢭ࣒ࢫࢿࢪ 6:3:௜ ᑠࢿࢪ─ ࢸ࣮ࣃࣆࣥ


$6 6(06)$67(1(5 ZLWK6:3: )& )/$7+($'6&5(: 3: ࣡ࢵࢩࣕᖹᑠᆺ୸ ✀
:$6+(5 73 7$3(53,1

ࢱࢵࣆࣥࢿࢪ ᮌࢿࢪ⏝ Ṇࡵࢿࢪࡍࡾ๭௜ ᖹඛ ṑ௜ࡁᗙ㔠


$7 7$33,1*6&5(: :RRGVFUHZ )* 6/277('6(76&5(: )ODWSRLQW 53 ࣆࣥࢫࣉࣜࣥࢢ
635,1*3,1 7: 67$5:$6+(5

ᑠࢿࢪ ࣂ࢖ࣥࢻ ᑠࢿࢪ ␒✀─ ࣡ࢵࢩࣕἼ


%& 0$&+,1(6&5(: %LQGLQJKHDG ). 0$&+,1(6&5(: )ODWKHDG 6% ࣎ࣝࢺභゅ✰௜
+(;$*2162&.(7+($'&$36&5(: :: :$9(:$6+(5

216 Return to previous page


NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan
(Manufacturer) Telephone: +81-533-67-6611
Facsimile: +81-533-67-6610
NIDEK CO., LTD : 3F Sumitomo Fudosan Hongo Bldg., 3-22-5, Hongo,
(Tokyo Office) Bunkyo-Ku, Tokyo 113-0033, Japan
Telephone: +81-3-5844-2641
Facsimile: +81-3-5844-2642
NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.
(United States Agent) Telephone: +1-510-226-5700
Facsimile: +1-510-226-5750
NIDEK S.A. : Europarc 13, rue Auguste Perret, 94042 Créteil, France
(EU Authorized Representative) Telephone: +33-1-49 80 97 97
Facsimile: +33-1-49 80 32 08

218