You are on page 1of 5

Analytica Chimica Acta 689 (2011) 34–38

Contents lists available at ScienceDirect

Analytica Chimica Acta


journal homepage: www.elsevier.com/locate/aca

Investigation of iodine liberation process in redox titration of potassium iodate


with sodium thiosulfate
Toshiaki Asakai ∗ , Akiharu Hioki
National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 3-9, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563,
Japan

a r t i c l e i n f o a b s t r a c t

Article history: Potassium iodate is often used as a reference material to standardize a sodium thiosulfate solution which
Received 15 November 2010 is a familiar titrant for redox titrations. In the standardization, iodine (triiodide) liberated by potassium
Received in revised form 12 January 2011 iodate in an acidic potassium iodide solution is titrated with a sodium thiosulfate solution. The iodine
Accepted 12 January 2011
liberation process is significantly affected by the amount of acid, that of potassium iodide added, the
Available online 19 January 2011
waiting time for the liberation, and light; therefore, the process plays a key role for the accuracy of
the titration results. Constant-voltage biamperometry with a modified dual platinum-chip electrode
Keywords:
was utilized to monitor the amount of liberated iodine under several liberation conditions. Coulometric
Coulometric titration
Oxidation–reduction titration
titration was utilized to determine the concentration of a sodium thiosulfate solution on an absolute basis.
Iodine liberation Potassium iodate was assayed by gravimetric titration with the sodium thiosulfate solution under several
Certified reference material iodine liberation conditions. The liberation process was discussed from the changes in the apparent assay
Sodium thiosulfate of potassium iodate. The information of the appropriate titration procedure obtained in the present study
Potassium iodate is useful for any analysts utilizing potassium iodate to standardize a thiosulfate solution.
© 2011 Elsevier B.V. All rights reserved.

1. Introduction Germany according to the International Database for Certified Ref-


erence Materials (COMAR) as of September 2010 [6–8], few reports
Sodium thiosulfate (Na2 S2 O3 ) is widely used in titrimetric anal- on the KIO3 assay [9–12] and on the Na2 S2 O3 assay by coulo-
ysis and is often used as a central material for iodometry [reduction metric titration [10–14]. In those past reports [10–12], KIO3 was
of iodine (I2 )] [1–4]. In general, the concentration of Na2 S2 O3 in assayed with Na2 S2 O3 whose concentration was determined by
a redox titration is determined based on the assay of potassium coulometric titration with electrogenerated I2 ; that is reverse to
iodate (KIO3 ) or potassium dichromate (K2 Cr2 O7 ) according to the a usual redox volumetric titration in which KIO3 as a reference
following reactions [reactions (1)–(3)] [1–4]. material is used to standardize Na2 S2 O3 . The accuracy of the coulo-
metric titration for Na2 S2 O3 was discussed in the reports and the
KIO3 + 5KI + 3H2 SO4 → 3K2 SO4 + 3H2 O + 3I2 (1) repeatability for the assay of KIO3 was sufficient; however, the
K2 Cr2 O7 + 6KI + 7H2 SO4 → Cr2 (SO4 )3 + 4K2 SO4 + 7H2 O + 3I2 iodine liberation processes were insufficiently investigated. Ma did
not mention the validity of the adopted weak acidic condition for
(2) the liberation in his paper [10]. One of the authors and Tanaka
et al. [11,12] adopted the condition similar to the Japanese Indus-
2Na2 S2 O3 + I2 → 2NaI + Na2 S4 O6 (3) trial Standard (JIS) [3,15]. Various conditions for the liberation of
I2 have been used as shown in Table 1 [1–3,10–12,15]. The lib-
In the reaction (1) or (2), the I2 (triiodide, I3 −)
liberated by eration of I2 is significantly affected by the amount of acid, that
KIO3 or K2 Cr2 O7 in an acidic potassium iodide (KI) solution can of KI added, the waiting time for liberation and light; therefore,
be titrated with a Na2 S2 O3 solution to determine the concen- the liberation process is essential to the accuracy of the titration
tration of the Na2 S2 O3 solution. In literature, the availability of results.
KIO3 for the titration was discussed and it was recommended In the present study, we utilized constant-voltage biamper-
as a primary standard [1,5]; however, there are only two certi- ometry (a dead stop method) [16] to monitor the amount of I2
fied reference materials (CRMs) of high purity KIO3 in Japan and generation under several liberation conditions and to examine the
sufficient conditions for I2 generation. This method was often used
for an electrical end-point determination in the reactions involv-
∗ Corresponding author. Tel.: +81 29 861 6881; fax: +81 29 861 6890. ing I2 [10–13,16–20]. To investigate the I2 liberation process more
E-mail address: t-asakai@aist.go.jp (T. Asakai). accurately, we assayed KIO3 under several liberation conditions

0003-2670/$ – see front matter © 2011 Elsevier B.V. All rights reserved.
doi:10.1016/j.aca.2011.01.021
T. Asakai, A. Hioki / Analytica Chimica Acta 689 (2011) 34–38 35

Table 1
Conditions for the liberation of iodine.

Reference KIO3 Water to dissolve KI Acid I2 liberation time


−1
[1] 0.12–0.17 g 25 mL 2g 10 mL of 1 mol L HCl 0 min
[2] 0.14 g 50 mL 3g 1 mL of H2 SO4 3 min
[3] 0.4–0.5 g 100 mL 5g 5 mL of 9 mol L−1 H2 SO4 5 min
[3] 0.09–0.11 g as an aliquot of a 100 mL 2g 2 mL of 9 mol L−1 H2 SO4 5 min
KIO3 concentrated solution
[3] 0.0075–0.01 g as an aliquot of a 100 mL 1g 2 mL of 9 mol L−1 H2 SO4 5 min
KIO3 concentrated solution
[10] 0.15 g 90 mL of 0.5 mol L−1 CH3 COOH– 2g 15 min
CH3 COONa (pH 4.5)
−1
[11] 0.15 g 150 mL 2g 5 mL of 3 mol L H2 SO4 10 min
[12] 0.15 g 25 mL 2g 5 mL of 3 mol L−1 H2 SO4 5 min
[15] 0.14–0.16 g 100 mL 2.00 g ± 0.05 g 5 mL of 3 mol L−1 H2 SO4 10 min

based on the procedure which was used for the assay of KIO3 in 3. Results and discussion
the past reports [10–12]. The liberation processes are clearly dis-
cussed with high accuracy from the changes in the apparent assay 3.1. Determination of Na2 S2 O3 by coulometric titration and
of KIO3 . uncertainty estimation

2. Experimental The major uncertainty sources for Na2 S2 O3 by coulomet-


ric titration are the current efficiency and the diffusion loss of
2.1. Reagents, apparatus and procedure of constant-voltage Na2 S2 O3 from the working-electrode chamber to the counter-
biamperometry electrode one. Marinenko et al. investigated the current efficiency
of I2 electrogeneration in a phosphate buffer containing KI
High purity KIO3 was obtained from Wako Pure Chemical Indus-
[18]. We adopted the conditions under which the 100% cur-
tries, Ltd., Osaka, Japan. Analytical reagent grade KI and H2 SO4 were
rent efficiency was substantially achieved according to them.
used for I2 liberation.
To examine the current efficiency and the sample diffusion
A DC current source (type 7651, Yokogawa Electric Corpora-
loss using our own coulometric titration system, we measured
tion, Tokyo, Japan) and a digital multimeter (type 7562, Yokogawa
ca. 200 mmol kg−1 Na2 S2 O3 with different applied electroly-
Electric Corporation) with a modified dual platinum-chip electrode
sis currents (50 mA and 100 mA) and different sample masses
(each platinum chip: 0.5 mm  × 0.5 mm long, placed 2 cm apart)
(3 g, 6 g and 9 g). The measured concentrations of the Na2 S2 O3
were used for constant-voltage biamperometry. Constant voltage
solution (ca. 6 g) electrolyzed with 50 mA and 100 mA were
was applied between the platinum pair, and the current was moni-
195.737 mmol kg−1 (RSD 0.005%, n = 3) and 195.743 mmol kg−1
tored. Using the method, the amounts of liberated I2 under several
(RSD 0.005%, n = 3), respectively (‘RSD’ means ‘relative stan-
conditions were measured.
dard deviation’, and ‘n’ is number of measurements under
2.2. Reagents, apparatus and procedure of coulometric titration a repeating condition). In addition, the results for different
and gravimetric titration amounts of the Na2 S2 O3 solution electrolyzed with 100 mA were
195.739 mmol kg−1 (ca. 3 g, RSD 0.004%, n = 4), 195.741 mmol kg−1
The KIO3 here was the same as that in the constant- (ca. 6 g, RSD 0.005%, n = 4) and 195.739 mmol kg−1 (ca. 9 g, RSD
voltage biamperometry (vide supra). The following constants 0.002%, n = 4). There were no significant dependencies of the con-
were used for the material: molar mass 214.001 g mol−1 [21] centration on the electrolysis current and the mass of sample
and density 3.89 g cm−3 [22]. The Faraday constant used was solution; therefore, we concluded that the assays had no significant
96 485.339 9 C mol−1 [23]. Buoyancy corrections were always biases.
applied. Applying short pulses of a constant current (typically 10 mA,
The instruments used for coulometric titration were the same 1 s) and measuring an indicator current were repeated to draw a
as those described previously [24]. The coulometric titration of titration curve near the end point. The measured concentrations
Na2 S2 O3 and the gravimetric titration of KIO3 with Na2 S2 O3 were of Na2 S2 O3 under applying different number of the short constant
also presented in detail in the previous report [11]. Reagents and current (4–304 times) were in good agreement with each other
electrodes used for the coulometric titration are briefly described within the repeatability (RSD 0.002%, n = 15; the difference between
here. The salt bridge was 3% (mass fraction) agar gel contain- maximum and minimum values was 0.01%). This fact confirmed
ing 3% (mass fraction) sodium sulfate. The catholyte consisted that there was no influence of non-Faradaic current and that the
of 42.6 g L−1 disodium hydrogen phosphate and 27.2 g L−1 potas- current efficiency of the short pulses of constant current was sub-
sium dihydrogen phosphate, and the anolyte consisted of the stantially 100%.
catholyte containing 16 g L−1 KI. The electrodes used were a plat- The contributions to the combined standard uncertainty of
inum plate (40 mm wide × 25 mm long, supporting rod: 1.5 mm the Na2 S2 O3 solution (ca. 6 g of ca. 200 mmol kg−1 ) were as
 × 150 mm long) as a working electrode and a platinum wire follows (the number in each parenthesis indicates the rela-
(1.5 mm  × 300 mm long) as a counter electrode. A weighed plas- tive standard uncertainty): repeatability (0.0014%, RSD of the
tic syringe with a PEEK (polyetheretherketone) needle was used mean, n = 7), Faraday constant (0.000 002 5%), standard resistor
for gravimetric titration and also for solution-sample introduction (0.000 22%), timer (0.000 05%), switching performance (0.000 12%),
to the coulometric titration cell. The end-point determination used voltmeter (0.000 15%) and weighing (sensitivity and linear-
in both coulometric titration and gravimetric one was constant- ity: 0.000 96%, and other possible bias: 0.000 84%). Finally, the
voltage biamperometry (applied voltage: 150 mV) using a dual combined standard uncertainty was 0.0019% relative and the
platinum-plate electrode (each 5 mm × 5 mm, type M-512, Kyoto expanded uncertainty with a coverage factor k = 2 was 0.004% rel-
Electronics Manufacturing Co., Ltd., Kyoto, Japan). ative.
36 T. Asakai, A. Hioki / Analytica Chimica Acta 689 (2011) 34–38

Fig. 2. I2 liberation process monitored by constant-voltage biamperometry for


0.15 g of KIO3 using a dual Pt-chip electrode (applied constant-voltage 500 mV), stir-
ring at 900 rpm. (a) 100 mL of water, 1 g, 2 g, 3 g, 4 g and 5 g of KI, 1 mL of 9 mol L−1
Fig. 1. Linearity of constant-voltage biamperometric detection. : a dual Pt-chip H2 SO4 , (b) 100 mL of water, 3 g of KI, 0.15 mL, 0.25 mL, 0.5 mL, 1 mL, 2 mL and 3 mL
electrode (each 0.5 mm  × 0.5 mm long); 100 mL of water, 5 g of KI, 2 mL of 9 mol L−1 of 9 mol L−1 H2 SO4 .
H2 SO4 , constant-voltage 500 mV, stirring at 900 rpm. : a dual Pt-plate electrode
(each 5 mm × 5 mm); the same condition as for the dual Pt-chip electrode.
initial increase. Using 1 mL of 9 mol L−1 H2 SO4 and 2 g of KI, the cur-
3.2. Linearity of constant-voltage biamperometry detection rent gradually decreased with the waiting time; from this fact, it
was inferred that the generated I2 evaporated because the concen-
In constant-voltage biamperometry, a constant voltage is tration of I− was not enough to keep all I2 in the solution. Brown
applied between twin platinum plates of the electrode and the cur- color was observed on the lid over the beaker after several runs.
rent change is measured. The current under a sufficiently small These conditions were apparently inappropriate; however, those
voltage (e.g., 150 mV, where water is not electrolyzed) is nearly could be appropriate because I− is resupplied through the reaction
zero in the presence of K+ , I− , and IO3 − in the absence of I3 − . But (3) between Na2 S2 O3 and I2 . Regarding the amount of 9 mol L−1
the current easily flows in the presence of both I- and I3 - through H2 SO4 , the liberation of I2 was not complete under the condition
the reaction 3I− ↔ I3 − + 2e− on each platinum surface. Constant- using 0.15 mL of 9 mol L−1 H2 SO4 because the volume of 9 mol L−1
voltage biamperometry is a sensitive method for the detection of I2 ; H2 SO4 was theoretically short [reaction (1)] for 0.15 g of KIO3 . There
however, the current under a high concentration of I2 could be not were apparently no problems under the other conditions in Fig. 2.
proportional to the amount of liberated I2 because the electrochem- The indicator current instantaneously rose (within 1–2 s) using
ical reaction 3I− ↔ I3 − + 2e− is diffusion-controlled. Therefore, we sufficient amounts of 9 mol L−1 H2 SO4 and KI. Judging from Fig. 2,
used a small platinum–chip pair and obtained wider linearity the titration was expected to completely proceed using both more
instead of higher sensitivity. than 3 g of KI with 1 mL of 9 mol L−1 H2 SO4 and 3 g of KI with more
The linearity between the amount of I2 and the current was than 0.25 mL of 9 mol L−1 H2 SO4 . In addition, it was recommended
examined by adding I2 as a solid to a KI solution as shown in Fig. 1. to immediately titrate after I2 generation to minimize both the
While the linearity using a dual platinum-plate electrode (each vaporization of I2 and the influence of light.
5 mm × 5 mm) was poor, a wide linearity of the indicator current
was obtained using a small dual platinum–chip pair (each 0.5 mm 3.4. Background caused by oxidation of iodide ions
 × 0.5 mm long). To improve the linearity, relatively higher volt-
age (500 mV) and higher stirring rotation rate [900 rpm] were used. End-point determinations in both coulometric titration and
The current had good proportionality to the amount of liberated I2 gravimetric one were done by constant-voltage biamperometry,
up to ca. 0.8 g at least. Therefore, the chip pair made it possible to whose instrument was the same as that used in the investiga-
monitor the amount of generated I2 corresponding to 0.15 g of KIO3 tion of I2 liberation except for the dual platinum-plate electrode
through the reaction (1). (each 5 mm × 5 mm) and the voltage of 150 mV. The electrode
showed higher sensitivity than the modified dual platinum–chip
3.3. Examination of I2 liberation by constant-voltage electrode. In gravimetric titration, the indicator current proportion-
biamperometry ally decreased with the titration proceeding near the end point and
the current was kept at zero amperes after the end point. The end
In the reaction (1), 0.15 g of KIO3 require a minimum of ca. 0.58 g point was the intersection between the linear regression curve of
of KI and ca. 0.23 mL of 9 mol L−1 H2 SO4 , and liberate ca. 0.53 g of I2 . indicator currents vs. the amounts of Na2 S2 O3 used and the baseline
Fig. 2 shows changes in the indicator current by constant-voltage (0 ␮A).
biamperometry after acid addition; the indicator current is pro- Using the constant-voltage biamperometry to detect I2 , the
portional to the amount of I2 liberation. Under the condition using influence of air oxidation of iodide ions was examined under sev-
1 mL of 9 mol L−1 H2 SO4 and 1 g of KI, the lack of I− in the solu- eral conditions and its contribution to the uncertainty of the KIO3
tion led to the deposition of solid I2 on the electrodes and in the assay was estimated. Fig. 3 illustrates changes in background cur-
solution, and the indicator current drastically decreased after the rents resulting from air oxidation of iodide ions. The changes in the
T. Asakai, A. Hioki / Analytica Chimica Acta 689 (2011) 34–38 37

Fig. 3. Increase of background current after acid addition 100 mL of water, 1–5 mL Fig. 4. Dependency of the apparent KIO3 assay on the I2 liberation time (100 mL of
of 9 mol L−1 H2 SO4 , 3 g of KI, without KIO3 . The currents were substantially at zero water, 0.15 g of KIO3 , 3 g of KI and 1 mL of 9 mol L−1 H2 SO4 ). The assay was based on
amperes using 1 mL of 9 mol L−1 H2 SO4 under both conditions of bright and dark. the concentration of Na2 S2 O3 standardized by coulometric titration. For the equa-
tions in the Figure, x indicates the liberation time (min) and y indicates the apparent
purity (mass fraction, %). Each data point under the same condition was measured
indicator currents were monitored under the conditions of some on independent runs to obtain the repeatability. : stoppered beaker during the
different 9 mol L−1 H2 SO4 concentrations with ca. 3 g of KI in both liberation, : unstoppered beaker during the liberation.
the presence and the absence of room light. The current exponen-
tially increased in all cases in a bright room and proportionally KI. There is no contradiction between the results of Figs. 2 and 5.
increased in a dark room. The larger amount of 9 mol L−1 H2 SO4 Fig. 2 apparently indicates that using 1–2 g of KI was not enough to
used, the larger increasing rate of the current under each condition completely generate I2 ; however, the results shown in Fig. 5 were
of bright and dark. The currents were below 0.06 ␮A using 1 mL of relatively close to 99.97%. The unexpected good results were caused
9 mol L−1 H2 SO4 at 5 min under both conditions of bright and dark. probably because iodide ions were sufficiently resupplied by the
The current 0.6 ␮A corresponds to 0.005% of 0.15 g of KIO3 under reaction between Na2 S2 O3 and I2 generated by KIO3 [reaction (3)].
the adopted measurement condition. The titration was practically In the case using 1–2 g of KI, the apparent assays of KIO3 titrated at
completed in 2 min after I2 generation; therefore, the background 5 min after starting I2 generation were lower than those (Fig. 5) of
current did not cause any error. The larger background current was KIO3 titrated immediately; probably this fact was caused by loss of
caused by the larger amount of KI used. To be concluded, the most
appropriate way was to finish a titration within a couple of minutes
in a dark room using smaller amounts of KI and 9 mol L−1 H2 SO4 if
the amounts are enough to completely generate I2 from KIO3 .

3.5. Examination of I2 liberation by coulometric titration and


gravimetric titration

Gravimetric titration was performed to assay KIO3 in a dark


room using a Na2 S2 O3 solution standardized by coulometric titra-
tion. The dependencies of the apparent KIO3 assay on the liberation
time of I2 are shown in Fig. 4, where 100 mL of water, 0.15 g
of KIO3 , 3 g of KI, 1 mL of 9 mol L−1 H2 SO4 and a 200 mL tall
beaker were used. The apparent assay of KIO3 with an unstop-
pered beaker proportionally decreased during the I2 liberation due
to the volatilization of generated I2 . Shorter times taken for the I2
liberation and for the titration with a Na2 S2 O3 solution were better.
The dependencies of the apparent assay of KIO3 on the amounts
of 9 mol L−1 H2 SO4 and KI used are shown in Fig. 5, where the gravi-
metric titrations were carried out immediately after I2 liberation
starting. The apparent assays of KIO3 were in the 99.96–99.97%
(mass fraction) range using 0.25–5 mL of 9 mol L−1 H2 SO4 and 3 g
of KI. The apparent assay using 0.15 mL of 9 mol L−1 H2 SO4 and 3 g
of KI was, however, about 66%. It agreed with the observation in
Fig. 2 and was natural because 0.15 mL of 9 mol L−1 H2 SO4 was Fig. 5. Dependency of the apparent KIO3 assay on the amount of KI and 9 mol L−1
H2 SO4 used. Each data point under the same condition was measured on indepen-
stoichiometrically short in the reaction (1) for 0.15 g of KIO3 . The
dent runs to obtain the repeatability. : dependency on the amount of 9 mol L−1
apparent assays of KIO3 were in the 99.96–99.97% range using 1 mL H2 SO4 with 3 g of KI, ×: dependency on the amount of KI with 1 mL of 9 mol L−1
of 9 mol L−1 H2 SO4 and 3–5 g of KI, but were lower using 1–2 g of H2 SO4 .
38 T. Asakai, A. Hioki / Analytica Chimica Acta 689 (2011) 34–38

slowly generated I2 from the solution. In conclusion, the liberation eration in a dark room gave no significant difference from the
of I2 from 0.15 g of KIO3 dissolved in 100 mL of water should be coulometric titration (within ±0.04% relative). The results for the
carried out using up to 5 mL more than 0.5 mL of 9 mol L−1 H2 SO4 other two concentrations by the modified JIS method under the
at least and 3 g of KI. The 0.25 mL aliquot of 9 mol L−1 H2 SO4 prac- condition of no delay after I2 generation in a dark room were consis-
tically raised no problem but the condition is not recommended tent with that determined by coulometric titration within ±0.02%
because the theoretical amount required is 0.23 mL. The titration relative, and the concentrations by the JIS method were within
of KIO3 with a Na2 S2 O3 solution should be done immediately after ±0.1% relative.
I2 liberation to avoid the influence of light and the volatilization of
I2 and air-oxidation of I− . 4. Conclusions
The uncertainty for the KIO3 assay can be calculated from the
uncertainties of the repeatability, the weighings of both KIO3 and a Coulometric titration and gravimetric titration were performed
Na2 S2 O3 solution, the determination of the Na2 S2 O3 concentration, to investigate the I2 liberation process in the redox titration of KIO3
the molar mass of KIO3 and the possible bias caused by I2 liber- with a Na2 S2 O3 solution. Constant-voltage biamperometry with a
ation. The approximate figures of relative standard uncertainties modified dual platinum–chip electrode was used to detect a wide-
are as follows: repeatability 0.005% (RSD based on the 20 measure- range concentration of liberated I2 under several concentrations
ments from 10 different bottles, 2 measurements in each bottle), of KI and H2 SO4 , and the appropriate liberation conditions were
the both weighings 0.003%, the concentration of Na2 S2 O3 0.004% found. The I2 liberation process was evaluated more accurately
and the molar mass of KIO3 0.001%. An estimation of the bias caused from changes in the apparent assay of KIO3 by gravimetric titraion
by I2 liberation was especially important. Fig. 4 indicates that the with a Na2 S2 O3 solution standardized by coulometric titration on
amount of I2 loss with an unstoppered beaker was 0.0174% relative an absolute basis. In the most appropriate procedure, KIO3 (0.15 g)
a minute. Most of the liberated I2 was titrated within 30 s, and a was dissolved into 100 mL of water, then more than 3 g of KI and
whole titration was completed in 2 min; therefore, the bias for I2 0.5 mL of 9 mol L−1 H2 SO4 were added, and immediately the solu-
loss fell much smaller than half of 0.0174%. On the other hand, the tion was titrated with a Na2 S2 O3 solution. No waiting time for the I2
influence of light was negligible since the increase of background liberation was necessary and a room light influence was negligible
caused by light was less than 0.0005% in 2 min (Section 3.4 and under quick operation. In addition, the JIS method was evaluated by
Fig. 3). comparing the concentrations of a Na2 S2 O3 solution determined by
the JIS method with that by coulometric titration. The information
3.6. Drying conditions for KIO3 of the appropriate titration procedure obtained in the present study
is useful for any analysts utilizing KIO3 to standardize a Na2 S2 O3
A change in the mass of KIO3 (ca. 3 g) by weighing before and solution.
after drying was examined to investigate the instablity of the mate-
rial. The change in each mass under drying at 50 ◦ C, 130 ◦ C and
References
180 ◦ C for 6 h was within 0.003%. A change in the mass of KIO3 (ca.
3 g) placed under a humid condition for 6 h was examined to inves- [1] I.M. Kolthoff, R. Belcher, Volumetric Analysis Volume III Titration methods:
tigate the hygroscopicity. The observed change was within 0.001%. Oxidation–Reduction Reactions, Interscience Publications Inc., New York, 1957.
In addition, there was no dependency of the assays of KIO3 on the [2] American Chemical Society, ACS—Specifications Reagent Chemicals, 9th ed.,
Oxford University Press, New York, 2000.
drying temperatures from room temperature to 180 ◦ C, judging [3] JIS K 8001, General rule for test methods of reagents, Japanese Industrial Stan-
from the results of the gravimetric titration with Na2 S2 O3 stan- dards Committee, Tokyo, 2009 (in Japanese).
dardized with coulometric titration. Therefore, it was concluded [4] T. Asakai, M. Murayama, Accred. Qual. Assur. 13 (2008) 351–360.
[5] V.A. Stenger, Anal. Chem. 23 (1951) 1540–1543.
that KIO3 was very stable and did not have hygroscopicity. [6] The International Database for Certified Reference Materials (COMAR) provided
by the Federal Institute for Materials Research and Testing (BAM).
3.7. Consideration for procedures in Japanese Industrial Standard [7] R. Pradel, T. Steiger, H. Klich, Accred. Qual. Assur. 8 (2003) 317–318.
[8] R. Pradel, T. Steiger, Accred. Qual. Assur. 12 (2007) 265–268.
[9] T. Asakai, Y. Kakihara, Y. Kozuka, S. Hossaka, M. Murayama, T. Tanaka, Anal.
The methods to prepare three different concentrations Chim. Acta 567 (2006) 269–276.
of Na2 S2 O3 (ca. 10 mmol kg−1 , ca. 100 mmol kg−1 and ca. [10] L. Ma, Accred. Qual. Assur. 7 (2002) 163–167.
[11] T. Asakai, M. Murayama, T. Tanaka, Talanta 73 (2007) 346–351.
500 mmol kg−1 ), which are standardized with KIO3 (Table 1), are
[12] T. Tanaka, H. Hayashi, Y. Komiya, H. Nabekawa, H. Hayashi, Bunseki Kagaku 56
defined in JIS K 8001 [3]. We assayed KIO3 by gravimetric titration (2007) 327–332 (in Japanese).
with a Na2 S2 O3 solution whose concentration was determined by [13] K. Rowley, E.H. Swift, Anal. Chem. 26 (1954) 373–375.
coulometric titration under the appropriate condition described [14] T.R. Markova, Y.I. Tur’yan, Sov. Electrochem. 22 (1986) 505.
[15] JIS K 8005, Reference Materials for Volumetric Analysis, Japanese Industrial
above. The JIS method using the KIO3 as a standard was vali- Standards Committee, Tokyo, 2006 (in Japanese).
dated by comparing the concentration of Na2 S2 O3 by the method [16] C.W. Foulk, A.T. Bawden, J. Amer. Chem. Soc. 48 (1926) 2045–2051.
with that directly determined by coulometric titration. Conse- [17] J.J. Lingane, A.J. Bard, Anal. Chim. Acta 16 (1957) 271–273.
[18] G. Marinenko, J.K. Taylor, Anal. Chem. 39 (1967) 1568–1571.
quently, the results indicated that the concentration of Na2 S2 O3 [19] A. Hioki, M. Kubota, A. Kawase, Analyst 117 (1992) 997–1001.
(ca. 10 mmol kg−1 ) by the JIS method under room light was ca. 1.7% [20] W.J. Ramsey, P.S. Farrington, E.H. Swift, Anal. Chem. 22 (1950) 332–335.
lower than the concentration determined by coulometric titration. [21] IUPAC, Pure Appl. Chem. 78 (2006) 2051–2066.
[22] D.R. Lide, CRC Handbook of Chemistry and Physics 2005–2006, 86th ed., CRC
This difference resulted from the I2 liberation by the light dur- Press, Boca Raton, 2005.
ing the titration of the JIS method. On the other hand, regarding [23] P.J. Mohr, B.N. Taylor, D.B. Newell, Rev. Mod. Phys. 80 (2008) 633–730.
the Na2 S2 O3 solution (ca. 10 mmol kg−1 ), the modified JIS method [24] T. Asakai, A. Hioki, Accred. Qual. Assur. 15 (2010) 391–399.
under the appropriate condition of immidiate titration after I2 gen-

You might also like