Hoffer Mini-Flowmeter

Engineering Guide
MF SERIES FOR LOW FLOW range, corrosion resistance and
APPLICATIONS operating life for a particular
New devices with low flow application.
measurement capability are Bearing Selection
increasingly in demand by Bearings are chosen based on
industrial and laboratory users as media compatibility, desired turndown
requirements for accurate low flow range, service life, serviceability, cost,
measurement devices proliferate. and accuracy considerations. Several
The Hoffer MF Series has been bearing types are available including
developed to meet the need for a self-lubricating, ceramic hybrid ball
low flow measurement device for bearings and sleeve constructions in
service on clean, low and moderate tungsten carbide and carbon
viscosity liquids, and for gas composites.
measurement applications. The side access port permits for
The MF Series is a family of rapid inspection and servicing of the
low flow rate measurement devices flowmeter internals including the
based on a pelton wheel like rotor. bearings.
The measured fluid is directed When selecting among several
tangentially thru a velocity nozzle chemically compatible bearings, the
against the pelton rotor causing it to ball bearing design offers the highest
rotate. The motion of the rotor is accuracy and widest usable range and
sensed by the pickup coil and may be considered the standard. It is
converted to a pulsing output signal required for gas measurement
where the frequency is related to applications.
the flow rate, and the accumulated Field bearing
pulses are related to the total replacement may
volume passing thru the flowmeter. be performed
Materials Selection and without a
Construction significant shift
Selection of the materials of in the original
construction is usually dictated by calibration.
the requirements of performance, Tungsten carbide offers the most
durability, media compatibility, durable bearing material in service
availability, and cost conditions where it is compatible, and
considerations. will offer the longest life in rugged
The materials of construction for industrial liquid service.
the MF Series are intended to Hard carbon composite bearings
provide a standard flowmeter offer properties which are nearly
configuration which will suit the chemically inert to many corrosive
requirements of a broad base of fluids and are utilized for this reason in
industrial users. Our applications demanding applications.
group can assist you in the selection
of options that will provide the
construction which offers the
optimum combination of useful
Flowmeter Pickup Selection conditioned pulse output without the The K-Factor is the number of
The flowmeter pickup coil senses need of a separate signal conditioner pulses per unit volume produced by
the motion of the rotor and converts and my be transmitted up to 5,000 the flowmeter under a given set of
feet. Since the flowmeter signal is conditions. Repeatability is a
it to a pulsing electrical signal which measure of the stability of the output
is of a discrete digital nature. Two conditioned directly at the under a given set of flowing
basic pickup types are offered for use flowmeter, this feature offers the conditions. The repeatability is
with the MF Series which will satisfy benefit of greater noise immunity defined as the allowable percentage
standard electrical interface when transmitting through a noisy deviation from the stated K-factor.
requirements for turbine flowmeters. environment. When size or weight is Other performance characteristics
an issue, the Redi-Pulse eliminates include the pressure drop characteristic
The standard flowmeter pickup is the need for conventional signal and output frequency range and are
a low drag magnetic type whose self- conditioners that are meter mounted
addressed in later sections.
generating output signal voltage is in a junction box directly on top of
approximately sinusoidal. The the flowmeter. Redi-Pulse coils are ELECTRONIC LINEARIZATION
amplitude of the output signal is a certified with the CE mark for ACCESSORIES FOR THE MF
function of flow rate, with a compliance with EMC Directive SERIES
minimum signal strength which is 89/336/EEC. See Redi-Pulse Several methods are available to
greater than 30 mVp-p. The standard technical data sheet for more reduce the inherent nonlinearity of
maximum frequency ranges at the information and detailed the MF Series flowmeter. The most
flowmeter's full scale capacity are economical method of linearization
specifications on this coil. is by offset frequency injection. In
offered at 1000, 2000, and 3000 Hz.
This signal may be transmitted up to this method, a signal equal to the
200 feet without preamplification Performance Characteristics offset frequency of the
Mini-Flowmeter is injected
and is directly compatible with many The basic performance electronically to the actual output of
instruments. A flow range of 10:1 is characteristics of the the flowmeter. The result effectively
common with this pickup type on Mini-Flowmeter are depicted in shifts the output characteristic to that
liquid applications. Figures 1 and 2. The MF Series of the ideal. A number of low cost
establishes a linear response after an linearizers are available for Hoffer
The Redi-Pulse pickup coil, initial offset correction when Signal Conditioners/Converters and
offered in a magnetic type for the operated at a constant viscosity. Indicator Instrument Series.
MF Series, includes a self-contained Over the linear flow range, the
preamplifier within the pickup coil Factory calibrations normally
input/output characteristic takes the provide the user with a precalibrated
housing. The pickup coil produces a following form: Mini-Flowmeter and Converter
calibrated to the users specifications.
Equation 1
Frequency = C1 X Flow rate - C2 Hazardous and Weatherproof Coil
Enclosures
The Mini-Flowmeter requires The explosion-proof requirements
of UL Class I, Div 1 are
the use of a linearization implemented by enclosing the pickup
conditioner available in all Hoffer coil in a suitably rated housing.
flow measurement instrumentation. Signal Conditioner/Converters may
Accuracies of ±1% of reading also be mounted within this
typical after initial correction for enclosure. Addition of an "O" Ring
offset. Better accuracies provides an additional NEMA-4
approaching ±0.2% are possible rating. Intrinsically safe meter
using flow computers which can systems are also available.
store the entire characteristics of
the Mini- Flowmeter.
Installation Requirements
The MF Series is capable of
sensing fluid flow in one
direction only. The flowmeter
may be installed with the
pickup coil in either a vertical
or horizontal position.
Filtration to at least 100
microns is recommended to
prolong the life of the
flowmeter and prevent damage
to the bearings.
Flow pulsation should be
minimized through the
application of pulsation
dampeners before the
flowmeter.
Standard Specifications Figure 4
End Fitting Type: 1/2" MS-33656-8 (Std.
others available upon request) Flow Rate Capability with Ball Bearings
Physical: Flow- Linear Range Repeatable Repeatable
Operating Pressure: meter MIN MAX MIN MAX
1000 PSI (standard) Size GPM CC/M GPM CC/M GPM CC/M GPM CC/M
6000 PSI (optional)
MF20 .015 57 .049 185 .007 26.5 .065 250
MF30 .017 65 .120 450 .0085 32.0 .150 575
Operating Temperature:
Mag. Pickup: -450 to +450º F MF40 .020 75 .150 575 .010 38.0 .200 750
High Temp. Pickup: -450 to +850º F MF50 .023 85 .200 750 .013 50.0 .350 1325
MF60 .025 95 .250 950 .015 57.0 .500 1900
Materials: MF70 .050 190 .500 1900 .030 115 .750 2850
Housing: 316 Stainless Steel MF80 .075 285 .750 2850 .040 150 1.00 3750
Bearing: Ceramic Hybrid Ball Bearings MF90 .085 325 .850 3200 .050 200 1.25 4750
(self-lubricating) MF100 .100 375 1.00 3750 .060 225 1.60 6000
Tungsten Carbide Sleeve
MF125 .130 490 1.30 4900 .075 275 2.10 8000
(optional)*
MF150 .200 750 2.00 7500 .080 300 3.00 11500
Carbon Composite Sleeve
(optional)* MF175 .300 1150 3.00 11500 .100 375 3.50 13250
Rotor: 17.4 PH (standard) * When tungsten carbide or carbon composite sleeve bearings are specified, only the linear
430 Stainless Steel (optional) ranges are available as the minimum flow rate is affected due to bearing drag. The MF20 and
Ni 200 (optional) MF30 are not available with sleeve bearings.
Seal: Rulon-J (standard, others
available) Figure 5

Gross Pressure Drop Characteristic on Water for the


MF SERIES FOR LIQUID
APPLICATIONS MF Series on Liquid Service
The liquid MF Series offers
measurement capability from .007
to 3.5 GPM in twelve overlapping
ranges for service on clean, low
viscosity liquids.
Size selection may be
accomplished by referring to
Figure 4 after giving consideration
to the service, construction, and
the maximum allowable pressure
drop.
Fluid viscosity effects should
be considered when the liquid's
temperature will be varying
significantly in actual service.
Adequate back pressure on the
flowmeter should also be
maintained to avoid cavitation
which may result in metering
errors.
It should be observed that with
the standard magnetic pickup, no
preamplifier is required in many
flow indicator applications and
can transmit up to 200 feet. The
Redi-pulse mag pickups are
available and can transmit up to
5000 feet.
field with no special tools or repair correction.
skills. Sleeve bearing designs are
Selection Considerations
The widest flow turndown range available for the MF 40 to 175 sizes.
The standard liquid MF Series and best performance on viscous The ruggedness of the tungsten carbide
flowmeter consists of the basic device service is achieved with a ball bearing configurations makes it attractive for
equipped with a magnetic pickup and design . The sizing chart lists the industrial applications requiring long
self-lubricating ball bearings. The usable range for this construction. The life. Carbon composite bearing designs
flow rate capability of this linear range listed is that range over may be used in corrosive applications.
configuration is detailed in the size which the flowmeter frequency Such designs will offer greater
selection chart. This configuration characteristic will maintain the ±1% operating life if sized to operate
was designed for serviceability. The linearity limits to a straight line at a toward the bottom of the flowmeter
ball bearings may be replaced in the constant viscosity after offset range.
The pressure drop characteristics are given based on water at Special Configurations Figure 6
a viscosity of 1 Cpse and a specific gravity of 1.0. For other
fluids, the following equation may be used to estimate the The physical configurations VISCOSITY EFFECTS ON MF175
pressure drop across the Mini-Flowmeter given the pressure for the MF Series is intended to
drop on water at the maximum flow rate and the fluids suit the diverse requirements of 1 CSTK
viscosity and density. the industrial users. Requests 5 CSTK
Equation 2 for other configurations,
20 CSTK
PSID=(Cpse)¼ X (SpGr)3/4 x PSID(H2O) intended to satisfy special 1000

requirements, may be directed to


where: the applications group at Hoffer

Output Frequency (Hz)


for review and
Cpse is viscosity in centipoise recommendations.
SpGr is the specific gravity 500

PSID(H2O) is pressure drop from Figure 5


Installation and Adapter Kits
Specifications for Liquid Applications Installation kits serve a two
fold purpose, the first of which
Output Frequency Range: 50 to 1000 Hz (3333 Hz
Maximum) is to ease customer installation. 0
Flow Rate (GPM) 1.5 3.0
The second is to facilitate
Linearity: ±1.0% reading (over 10:1 range) adapting from the MS 33656-8 fitting on the flowmeter to the
Linear Range: 10:1 turndown with magnetic pickups and desired user end fitting. Flow straighteners are not required.
ball bearings
The kit consists of two lengths of 1/2" stainless steel
Usable Flow Range: 25:1 with ball bearings tubing with one end flared and the length cut to an
Repeatability: ±0.1% typ. (±0.25% worst case) appropriate size. The mating sleeves and nuts are also
Reference conditions: 1 cstk. test fluid at 70° F provided.
The kits may be conveniently butt welded into the user's
Viscosity Effects piping. Alternately, the kits may be provided with NPT or
flanged connections to facilitate installation into existing
An ideal flowmeter may be defined as one in which the lines.
output is solely a function of the fluid flow being measured.
Real flowmeters display dependencies on secondary fluid
properties, such as viscosity, temperature, and/or pressure. Strainers/Filters
These effects tend to obscure or degrade the precision of the Installation of a suitable strainer / filter reduces or
flow measurement. eliminates the potential of bearing damage by a foreign body.
In selecting a MF Series flowmeter for operation on a Such devices also maximize the life of the flowmeter since
viscous fluid, it is generally preferable to size the flowmeter bearing life is maximized when contaminating particles are
so it will be operated in the higher portion of its range to removed prior to passage through the flowmeter. The
minimize viscosity effects in the measurement. Some loss in suggested filtration recommendation is 100 microns or better.
flow turndown range may be expected.

MF SERIES FOR GAS APPLICATIONS Figure 7

MF Series meters may be used for gas flow measurement Mini-Flow Gas Meter Capacity
and come in 6 overlapping sizes detailed in Figure 7. Specific
criteria must be met when using the MF Series on gas flow
Flowmeter Size Maximum Flow Rate
applications. Those criteria include constant gas pressure and ACFM
constant gas temperature. There is no effective method for MF30 0.05
compensating the output of the MF Series due to changing MF50 0.10
pressure and temperature. Additionally, MF Series meters MF80 0.20
must be air or gas calibrated in order to produce accurate
MF90 0.30
results. Consult the factory for specific sizing and calibration
recommendations. MF125 0.50
MF175 1.00

107 Kitty Hawk Lane, P.O. Box 2145, Elizabeth City, NC 27906-2145
800-628-4584 252-331-1997 FAX 252-331-2886
www.hofferflow.com email: info@hofferflow.com
The quality system covering the design,
The specifications contained herein are subject to change without notice and any user of said specifications should manufacture and testing of our products is
verify from the manufacturer that the specifications are currently in effect. Otherwise, the manufacturer assumes no
responsibility for the use of specifications which may have been changed and are no longer in effect. MF-103F certified to International Standard ISO 9001.