Gas Delivery Systems for Safety, Performance, and Cost-Efficiency in the Laboratory

Selecting Gas Delivery Systems for Safety, Performance, and Cost-Efficiency in Claim your Complimentary
the Laboratory Subscription

If you are a scientist,


Larry Gallagher you are eligible for a full

In almost any laboratory or research and scientific facility, there are numerous devices, instruments, or processes that subscription.

require gases to run the instrumentation or process or to calibrate the devices. Gas cylinders that are located in the laboratory Click here to get started »
area can present significant hazards, and the space they take up can be better used for other more appropriate purposes. Gas
delivery systems that are properly designed, sized, and located can improve safety in the laboratory. In addition, attention to
high-purity requirements—purity, compatibility, flow, materials of construction, and more—is vital for safety, performance,
and cost-efficiency. Related Products

There are a number of codes and standards that apply to the storage, use, and installation of gases and their delivery systems. CFH Hydrogen
This article is not meant to detail them all, but will reference portions of those standards and codes as they apply to the Generators
general principles of gas storage and delivery systems. The CFH Hydrogen
Generators are designed to
offer a pure source of
The codes and standards that are most significant are:
hydrogen for your GC
Application,…
National Fire Protection Association NFPA 55: Compressed Gases and Cryogenic Fluids Code
learn more
Compressed Gas Association CGA P-1: Safe Handling of Compressed Gases in Containers
OSHA Standards 29 CFR 1910.101: Compressed Gases General Requirements. Parker Balston PEM
Hydrogen Generators
Laboratories must understand and follow these codes and standards thoroughly in order to comply with the minimum Fuel-grade high purity
requirements of any gas system. hydrogen generators are
safer alternatives to
high-pressure gas cylinders.
Gas requirement audit …
learn more
The first step in properly designing any gas installation, whether it is for a new facility or a retrofit, is to conduct an audit or
Dual-Stage Ultra-High
survey of the gases required for each location or laboratory. Identifying what gases are required for each instrument, their
Purity Stainless Steel
purity level, required delivery pressure, and peak flow demand is essential in determining everything from the size of the Gas Regulators
piping required, to the storage area required, and even the mode or source in which the gas may be supplied. Overestimating UHP stainless steel
the pressure or flow requirements can result in higher installation costs and reduced gas savings resulting from residual regulators are the standard
for ultra-high-purity and
contents left in cylinders or higher monthly rental fees on cylinders that are not required. Underestimating, however, can
corrosion-resistant …
cause inefficient supply of gas to critical instruments or systems that impair their use or operation.
learn more

Using a simple form, like that shown in Figure 1, for each lab or use point allows the total requirement of each gas to be
determined for that area, and when compiled as a whole will determine the scope of the building’s needs. The total volume of Related Product
each gas required for an average week or month’s operation can then be compared to the options available for that gas to be Categories
delivered by the gas supplier.

Laboratory Gas Generators »

GC Gas Regulators / High Purity Gas


Regulator »

1 of 5 12/3/2013 2:42 PM
Selecting Gas Delivery Systems for Safety, Performance, and Cost-Efficie... http://www.americanlaboratory.com/914-Application-Notes/118381-Sele...

Gas Flow Meters / Gas Flow Monitor »

Subscribe to eNewsletters and


Email Alerts

Keep up with our latest articles, news and events. Plus,


get special offers and more delivered to your inbox.

Your Email Address

Find us Follow us

advertisement

Figure 1 - Gas requirement audit form.

While some gases or purities can only be supplied in high-pressure cylinders such as helium with a purity of 99.999% used
for gas chromatographs, others—such as argon with the same purity for use by an inductively coupled plasma analyzer—can
be supplied in high-pressure cylinders or cryogenic liquid cylinders that equal the volume of as many as 18 high-pressure
cylinders. The total number of cylinders required to supply the laboratory for no less than an average week’s usage will then
be the ideal number of cylinders connected as the primary supply, with an equal number of cylinders connected as the
reserve supply and an additional number as backup. This will then assist in determining the space and location required for
the gas delivery system and cylinder storage area.
Product Guide

Analytical Lasers and Optical


Selecting a location for storage and gas delivery systems Instruments Components
Chemical Analysis Mass Spectrometry
There are some very specific storage and separation requirements for the areas in which compressed or cryogenic gases are
Equipment
stored and/or connected to a building’s gas delivery system. They may vary by local code requirements, but at a minimum Materials Research
Clinical Diagnostics
must meet those of NFPA codes and OSHA standards. As a minimum, gas cylinders should be stored and secured in an Microscopy
upright position using brackets, chains, or straps in a well-lit and ventilated area away from combustible materials and Environmental Test
Equipment Pharmaceutical Lab
sources of heat or ignition. The gases should be separated by hazard type as detailed in NFPA 55 Table 2-1.5, “Separation of Equipment
Gas Cylinders by Hazard.” Space should be allowed to segregate full from empty cylinders with the appropriate hazard Food Testing
Equipment Spectroscopy
notification signage as required under Chapter 4 of NFPA 55.
Forensic Laboratory Test and
Equipment Measurement
Regarding the storage requirements for hydrogen, there are additional distances from other hazards that must be met. These
are now listed under the current edition of NFPA 55, as included in chapters as shown in Figure 2. They were previously General Laboratory
listed separately under NFPA 50-A, “Standards for Gaseous Hydrogen Systems at Consumer Sites.” Of note is the separation Equipment
of hydrogen storage or use areas from other flammable or combustible liquids of at least 20 ft from ignition sources or
more products from labcompare »
ordinary electrical equipment of at least 25 ft, and 50 ft from air-compressor intakes, ventilation intakes, or air-conditioning
intakes. This also limits the total amount of gas allowed based on where it is located. Where that allows inside storage,
systems containing less than 3000 ft3 must be located in “special rooms” that are of sturdy construction and made of
Industry News
noncombustible material. Walls must have a minimum 2-hr fire rating, with at least one wall in an outside location. In
general, the preferred location of hydrogen systems is outside the building where there are no wall openings above the Targeted Analyses of Secondary
system, with any wall openings separated from the storage area by at least 25 ft. Metabolites in Herbs, Spices, and
Beverages
If dealing with oxygen storage where it is defined as a “bulk oxygen installation” with a volume of more than 20,000 ft3, there Tuesday, December 03, 2013

are additional restrictions that must be followed, as detailed in NFPA 55. These are now included as chapters, though at one
time they were listed separately under NFPA 50, “Standards for Bulk Oxygen Systems at Consumer Sites.” In general, bulk New Whitepaper Highlights Impact of
oxygen systems are not permitted to be located inside buildings; thus when the total volume of oxygen for a system exceeds NGS Target Enrichment Assay Choice
20,000 ft3, the storage location must be outside the building. Tuesday, December 03, 2013

Pipe sizing and flow considerations Magritek announce Spinsolve Carbon,


the world's first Carbon-13 capable
When determining what size pipe or tubing to use for a specific application, the main consideration is to determine what the benchtop NMR spectroscopy system
maximum required flow for that specific gas would be if all application or use points were flowing to their maximum at the Thursday, November 28, 2013
same time. This can be found by totaling the individual use points and applying a conservative safety factor to allow for

2 of 5 12/3/2013 2:42 PM
Selecting Gas Delivery Systems for Safety, Performance, and Cost-Efficie... http://www.americanlaboratory.com/914-Application-Notes/118381-Sele...

growth of requirements of at least 20–50%, depending on what the future outlook is for that gas. Texture Pro Software 21 CFR
Compliance
In addition, when sizing the piping, the total distance to the very last use point in the piping run must be considered. If the Thursday, November 28, 2013
total flow is measured over distance, then there will be a pressure drop from what the initial inlet pressure is to what the
pressure is over that distance. In general, the maximum pressure drop should be no more than 10% of the primary inlet more news »

pressure. The maximum pipeline pressure would be equal to the highest required pressure for the use point of that specific
gas or instrument, plus a minimum of either 20% or 25 psig, depending on the flow requirement for that use point. It may be Most Viewed Articles Most Recent Articles
higher, depending on how far the piping run is.
Measurement of Viscosity in a
Vertical Falling Ball Viscometer
Monday, October 27, 2008

Hyphenation: The Next Step in


Thermal Analysis
Friday, January 01, 2010

Synthesis, Characterization, and


Evaluation of a Fluid Loss Agent
for High-Temperature Drilling
Figure 2 - Hazard distance requirements.
Fluid
Thursday, November 07, 2013

A useful tool is the pipe size chart shown in Figure 3. It gives the maximum flow of air for various pipe sizes with either a 10
or 5% pressure drop. The chart can be found under the resources category at www.concoa.com. As an example of how to use Analyzing Volatile Organic
this information, if flowing at an initial inlet pressure of 100 psig, through a ½” diameter pipe, one can expect to flow 3240 Chemicals in Food: Emerging
ft3 with over 100 ft of piping with a 10% drop in pressure at that point. In other words, 90 psig can be delivered to that use Trends and Recent Examples
point with that flow rate, if it was the only connection on the system. One would, however, specify the gas delivery system to Thursday, February 14, 2013

be delivering a minimum of 120–125 psig into the pipeline to allow for individual control of the delivered pressure at each
point of use, including the one requiring 90 psig. Since piping use points should also have reverse flow check valves,
Determination of Association
especially those for flammable gases like hydrogen or oxygen, the cracking pressure of these check valves also needs to be (kon) and Dissociation (koff) Rates
considered. of Spiperone on the Dopamine D2
Receptor Using a Platform for
GPCR Applications
Saturday, October 01, 2011

Figure 3 - Pipe size chart.

Choosing the right gas delivery system


For almost all laboratory gases, maintaining gas purity is a critical requirement. To that end, the choice of materials of
construction and their compatibility with that specific gas and its purity level must be considered. It is not enough that the
materials are compatible with the specific gas, but that the design ensures and retains the purity of the gas. From the inlet to
the outlet, a system that is designed with either bar stock brass or 316L stainless steel components is desired over forged
components. Any diaphragms should be made of 316L stainless steel, and the diaphragm seals should be of a metal-to-metal
design.

The system should have minimum leak integrity of 1 × 10–8 standard cm3/sec of helium, which equals the volume of a cubic
centimeter escaping or entering over 3.1 years. This not only ensures that gas will not leak out of the system, but also that
impurities will not be allowed to enter the system and compromise the purity of the gas. For the inlet of the system, the inlet
connections should have reverse flow check valves in the cylinder connections such that once the system has been put into
service, the amount of any ambient air that may enter is limited to a very small amount. Two such systems are shown in

3 of 5 12/3/2013 2:42 PM
Selecting Gas Delivery Systems for Safety, Performance, and Cost-Efficie... http://www.americanlaboratory.com/914-Application-Notes/118381-Sele...

Figure 4.

The gas delivery system should be capable of delivering the


maximum required pressure and flow as described above
and the information determined from the gas audit.

Though there are many other areas that must be


considered, the primary areas of concern are where to
locate the gas storage area outside the laboratory, how
much gas will be required, and how the gas will be
delivered to the end use points. Users who follow these
guidelines are well on their way to selecting a safe and
cost-effective gas delivery system.

Figure 4 - Laboratory switchovers with the highest leak


Larry Gallagher is Specialty Gas Products Manager,
integrity.
CONCOA, 1501 Harpers Rd., Virginia Beach, VA 23454,
U.S.A.; tel.: 800-225-0473; e-mail:
lgallagher@concoa.com.

Comments

0 comments ★ 0

Best Community Share ⤤ Login

Happy Anniversary October 2013


• •
— My involvement with — It's the cocktail "Mint
analytical instrumentation started in 1967, Julip" - another beautiful image from
and I have followed the American … Michael Davidson's Cocktail Collection!

June/July 2013 Field-Flow Fractionation Supporting


• Consumer Safety Evaluation of Silver …
— Saccharomyces •
cerevisiae — A complete bibliography of every
(known) paper published on FFF may be
found at Wyatt.com.

Sections Company Clients


Home About Us Advertise With Us
Application Notes Contact Us Media Kit American Lab is Powered by Labcompare, the Buyer's
Guide for Laboratory Professionals
Technical Articles Press Room

4 of 5 12/3/2013 2:42 PM
Selecting Gas Delivery Systems for Safety, Performance, and Cost-Efficie... http://www.americanlaboratory.com/914-Application-Notes/118381-Sele...

Statistics in Analytical Chemistry Privacy Policy

Featured Products Disclaimer


Videos
Events
News
Article Archive & Filter
Newsletter Archive
See our other sites »
Newsletter Signup
Manage Your Print Subscription Copyright © 2013 CompareNetworks, Inc. All rights reserved.

5 of 5 12/3/2013 2:42 PM

You might also like