You are on page 1of 11

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/3090844

New Digital Multimeter for Accurate Measurement of Synchronously Sampled


AC Signals

Article  in  IEEE Transactions on Instrumentation and Measurement · July 2004


DOI: 10.1109/TIM.2004.827313 · Source: IEEE Xplore

CITATIONS READS

31 288

1 author:

Predrag Petrovic
University of Kragujevac
74 PUBLICATIONS   213 CITATIONS   

SEE PROFILE

All content following this page was uploaded by Predrag Petrovic on 10 June 2016.

The user has requested enhancement of the downloaded file.


716 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 3, JUNE 2004

New Digital Multimeter for Accurate Measurement


of Synchronously Sampled AC Signals
P. Petrovic, Member, IEEE

Abstract—An approach to the design of a digital multimeter between the time of the second phase of integration and the
based on the use of dual-slope ADC is described in this paper. measuring time (first phase).
Based on this approach, a new technique for algorithm de-
The measurement method proposed in this paper is based on
sign is developed. The main advantage of this approach is the
determination of all electric values in electric utilities by the selecting samples (the original assumption) of the input variable
same algorithm. The algorithm has high accuracy and a regular in a large number of periods in which the system (in this case,
structure. Measurements of voltage and current are made in the electric utility) is considered to be stationary. The stationary
successive periods by a stroboscopic technique (synchronous condition can be proved by values obtained from measurements
undersampling). The assumed stationarity of the electric utilities
is validated by measurements with an experimental setup, con- of rms values of system voltage [1]. From that finding, the math-
sisting of a fast high-precision sigma-delta ADC. The necessary ematical presentation was concluded, i.e., confirmation that the
synchronization is reached by software measurements of the bandwidth of that system is very limited. That fact enabled ap-
frequency of the measured signal. Assuming stationarity of the plication of the Nyquist criterion.
observed system (electric utilities) in the stated 1-s interval, it is
proved that precise digital processing can be achieved without By stationarity of the system, we understand that slowly
using a sample-and-hold circuit. The suggested measuring system changing quantities, like current and voltage and their harmonic
was simulated, realized in a practical setup, and tested. The content, are constant within the measuring interval. In this case,
obtained results completely confirmed the starting postulates. A undersampling contrary to the Nyquist criteria is possible. The
processing precision of 0.01% was achieved.
measuring time is about 1 s. The algorithm is of an adaptable
Index Terms—Adaptable algorithm, digital multimeter, dual- type and depends on the harmonic content of the input signal
slope ADC, high accuracy, synchronous undersampling.
and network frequency (50 or 60 Hz) [1]. The current makes
this system nonlinear due to the fact that one cannot predict
I. INTRODUCTION in advance the type of load which will be used and when it
will be connected to the investigated system. However, after
T HE CONVENTIONAL algorithms for the measurement
of active power and effective values of voltage and current
employ integration or summation over an abruptly limited time
a certain number of periods, the current can be considered
a slowly changing variable during processing. This is why
very slow, low-cost, but very accurate ADCs [3], such as a
interval. Consequently, they operate correctly at a periodic input
dual-slope type, were used in the proposed measuring system.
signal only. The voltage and current waveforms of the power
Integrating ADCs work best in high-accuracy, fine resolution
network, however, are not strictly periodic, due to their nonhar-
systems because they remove any power-line frequency noise
monic components and stochastic variation. The outputs of both
from the input signal. Voltage and current from real electric
the conventional and new algorithms do not represent the input’s
utilities were used as input variables. The sampling procedure
instantaneous value, but the input observed over a certain period
is initiated arbitrarily (Fig. 1).
of time. The measurement method described in this paper can be
The distance between two consecutive samples is given by
applied for the accurate determination of periodic signals. The
method relies on adaptation of the measurement time to the pe-
riod of the measurement signal.
(1)
The synchronous method is based on dual-slope integration
performed according to the synchronization signal (input
signal). The first phase of integration is fully synchronized
where is the number of periods between sampling, is the
with the input signal. The time of the first phase is also mea-
period of the input voltage, and is the delay determined
sured. The output voltage of the integrator depends directly
by the delay of elements in the processing circuit. depends
on the input signal and the time of the first integration phase.
on the harmonic content of the input signal, ,
Measurement values are the results of arithmetic operations
, where is the number of the highest
harmonic of the voltage, is the number of the highest har-
monic of the current signal, and is the period of the input
Manuscript received March 19, 2003; revised January 12, 2004. signal. All conditions, which have to satisfy both and to
The author is with the Cacak College of Engineering, University of Kragu-
jevac, 32000 Cacak, Yugoslavia. get an accurate result of measuring, can be derived [1]. For that
Digital Object Identifier 10.1109/TIM.2004.827313 reason, they cannot be arbitrary.
0018-9456/04$20.00 © 2004 IEEE

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
PETROVIC: NEW DIGITAL MULTIMETER FOR ACCURATE MEASUREMENT 717

Fig. 1. Proposed method of sampling.

Calculation by the proposed method is carried out on the basis voltage signals have different harmonic content, the conditions
of the following: for delay would determine a signal with “richer” harmonic
content (which is usually the current signal because of its greater
dynamics). On the basis of (38) in the Appendix of [1]

(2)
(3)

where is the number of samples needed for accurate pro- where is the angular frequency of the fundamental
cessing of observed values (in our case , , harmonic. It is necessary to take the samples equidistantly on
s), and are the rms values of the interval of one period. In that way, on the basis of established
current and voltage, respectively, and is the active power. The limits in processing [1], it is possible to calculate accurately the
reason for taking samples is that, even the most com- observed electrical values in electric utilities.
plex harmonic content processed by the proposed method de- One of the basic principles or postulates in the processing of
mands more than voltage and current samples [1]. the continuous signals (analog signal, which is continual per am-
Index provides the jump from period to period of the mea- plitude and time) by the dual-slope ADC is that at its input there
sured voltage and current and delay for samples at the next has to be a sample-and-hold circuit, which is a possible source
processing. of systematic errors. In [5], we show that use of this circuit is
According to the suggested measurement method, it is as- not necessary.
sumed that, during the measuring interval, the harmonics con-
tent of the measured signal does not change very much. The II. SIMULATION OF THE SUGGESTED MEASURING METHOD
algorithm deals with arbitrary harmonics content. In the case
of a current with high harmonic content, as in a system sup- Additional testing of the suggested digital measuring system
plying phase-switched loads (such as thyristor controlled), the was carried out by simulation in the program package Matlab
(version 5.1), module Simulink.
accuracy of the method is preserved, provided the system is sta-
In Fig. 2, a block diagram of the suggested digital measuring
tionary within the measuring interval. The possible nonlinear
system is shown. The construction is made of ready-made
distortions in transition processes do not last very long. Thus,
Simulink models. The special advantage of such a program
they can be avoided when the functioning of this multimeter de-
environment or surrounding is that we are able to give an
signed for measuring periodic variables is considered.
arbitrary input signal, which is further processed. In the sense
The obtained conditions for the value of delay ( of its harmonic content, white noise and different irregularities
in [1]) are completely equal to the Nyquist condition. In other may occur and simulate jump functions, with changeable
words, as it was described in the introduction [1], we are not phase position between processed voltage and current signals.
able to carry out real spectral reconstruction of the processed Thus, both voltage and current signals are introduced into the
signal by the proposed method of measurement because of the simulation, while a completely arbitrary phase relation and
extremely low speed of sampling. It is only possible to carry amplitude value are given to them. In addition, an absolutely
out “virtual” (delay in time) spectral reconstruction. Since the arbitrary noise power is superimposed on them.
delay is responsible for the movement forward of the mo- After forming a complex harmonic input signal, the signal
ment of sampling from period to period (or more periods de- was taken into the circuit for the sample-and-hold (unit delay),
pending on the parameter value), the delay must satisfy the which is located in front of the real ADC [2]. The signal
Nyquist criterion. In other words, it must be in accordance with was taken from the output sample-and-hold circuit into the
a basic postulate of synchronous sampling, where the proposed D flip-flop as a delay element. It is clocked from the special
measurement method conditionally belongs. If the current and signal generator (rectangular series of impulses) for which the

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
718 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 3, JUNE 2004

Fig. 2. Block-diagram of simulation model for measuring active power, based on the suggested measuring concept with dual-slope ADC and the sample-and-hold
circuit.

arbitrary duty ratio can be given. In this way, the continual in Simulink has been constructed which corresponds to the pro-
signal was measured, and that sample was held constant up to posed measurement method.
the moment of the next measurement or sampling, which was In the model (Fig. 3), special selectors of analog signals
given on the basis of the previously proposed form. The next (Switch 1 and 2), passing through complex periodic signals
sample was taken from one of the next periods of the input of current and voltage in the intervals, correspond to the
signal, which can be adjusted by the choice of parameters of the work speed of the dual-slope ADC. The pulse generator is
simulation model. Voltage and current signals were multiplied used, and its duty-ratio and period can be given. The selector
and then integrated in time, thus obtaining the data of active passes through the signal, from the first input, during the
power. time when the signal, from the second input, is bigger than
All model parameters are adjusted to the real dual-slope ADC the given threshold. Otherwise, it would pass through or let
and its speed and conversion precision. Because of this, we can out a signal from the third input. This is the reason why we
claim that this kind of simulation completely presents the real use the constant block, with the value 0. We use the constant
conditions of actual use. The only difference is that by simula- on the final multiplier (Figs. 2 and 3) only as scaling factor,
tion, the work is tested in more extreme conditions that the ones to produce the active power of processing signals. From the
which may be expected in practice. The obtained results fully Figs. 2 and 3, it can be noticed that voltage/current signals
confirmed all the suppositions and conclusions that were made with three harmonics (besides the basic one, the third, and
before. the fifth, either even or odd) were used because they are the
In the simulation model designed in this way, ideal synchro- most dominant harmonics in practice. However, it is not a
nization in processing with the frequency of the processed signal limitation factor in such a simulation. In other words, we are
was achieved. This is very difficult in practice, but together with able to introduce higher harmonics, and the conclusions are
planned hardware resources (precise comparator and micropro- still valid. All of this was worked out in detail in paper [1] when
cessor), we will be able to determine accurately the frequency of arbitrary harmonic content was accepted. After mathematical
the processed signals. Since it has such an input block, Simulink analyses for that case, certain conclusions were made about the
gives the possibility for introducing a sinus signal whose fre- necessary number of samples and delay time during processing
quency varies completely arbitrarily. Thus, this block was also [1], in order to get absolutely accurate, theoretical results of
used during the simulation, but the results were still beyond our measurement. Since it has such an input block, Simulink gives
expectations (to the third decimal). the possibility for introducing a sinus signal whose frequency
In order to estimate by simulation the power of digital pro- varies completely arbitrarily. Thus, this block was also used
cessing without using the circuit for sample-and-hold, a model during the simulation.

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
PETROVIC: NEW DIGITAL MULTIMETER FOR ACCURATE MEASUREMENT 719

Fig. 3. Circuit model, which carries out the processing of active power without using the sample-and-hold circuit.

TABLE I
VALUES OF ACTIVE POWER WHEN PROCESSING SIGNALS WITH THE BASIC
HARMONIC WITH AMPLITUDE OF 1 V, THIRD HARMONIC WITH
AMPLITUDE OF 0.3 V, AND FIFTH HARMONIC WITH AMPLITUDE
OF 0.1 V, FOR DIFFERENT METHODS OF POWER PROCESSING

Fig. 4. Voltage signal to which an extremely powerful noise signal is


superpenetrated (with power of 0.004) which is used for testing the processing
concept with and without the sample-and-hold circuit.

have to catch. In practical realizations, this circuit introduces a


In the above-described processing, signals of small amplitude significant error when processing very dynamic signals. In the
were used because such amplitudes can be expected on the plate real system, it will be rare to encounter such a strong noise,
after scaling and introduction of signals into the ADC and be- which can be prevented by additional filtering, as already ex-
cause of clearance. The results obtained when calculating active plained in paper [1].
power followed, when the product of the two compound period- Based on the suggested concept of processing, the complete
ical signals polluted by very strong noise is processed. This is realization of the dual-slope ADC can be achieved through the
most critical from the digital processing point of view. Fig. 4 is resources of the microprocessor itself, as well as through its
an illustration of the signal obtained in this way. counting resources. Only the most modern type of processor
In Table I, the results of processing over three models of dig- is satisfactory here, however, with a considerably high perfor-
ital processing are given. In the real environment, such a strong mance. All these elements will lead to reduced cost of the sug-
noise will rarely be encountered. Because of that, this model gested solution, without any changes to the basic conclusion
with a circuit for sampling should be accepted as very accu- [1], dealing with the adaptability and the necessary number of
rate. The model without the sample-and-hold circuit gives much recorded measures. It is assumed that, during these observa-
better results at great distortion because processing also includes tions, the system such as an electric utility remains stationary,
those distortions, which the sample-and-hold circuit does not and that the harmonious content is preserved, within the time

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
720 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 3, JUNE 2004

Fig. 5. Block scheme of the realized digital multimeter.

interval necessary to conduct the measurement according to the the integrator during the fixed-time portion of the cycle. Any
suggested method. In the same way, we can process other slowly changes in the analog signal during that period of time have
changing values such as pressure, temperature, and so on. a cumulative effect on the digital output at the end of that
The suggested method gives very accurate results when mea- cycle. Other ADC strategies merely “capture” the analog signal
suring time-average power, regardless of the harmonic content level at a single point in time every cycle. If the analog signal
of the input signal. After detection of the signal harmonic con- is “noisy” (contains significant levels of spurious voltage
tent, the necessary number of samples can be specified in the spikes/dips), one of the other ADC converter technologies may
form of an adaptive algorithm [1]. The suggested method of pro- occasionally convert a spike or dip because it captures the
cessing ac values in electric utilities theoretically gives an abso- signal repeatedly at a signal point time. A dual-slope ADC, on
lutely accurate result of measurement, if the observed system the other hand, averages together all the spikes and dips within
can be considered stationary during a sufficiently long period the integration period, thus providing an output with greater
of time [1]. The proposed approach is suitable for real-time pro- noise immunity. Delta-sigma (or more accurately, sigma-delta)
cessing and is characterized by a low computational burden in ADCs are the newest architecture and are used in systems
comparison to the described algorithms in [3], [4]. This gives demanding high-resolution data acquisition; but when they are
an opportunity for developing a measurement system with very used in instrumentation, their filter delays prevent multiplexing
simple and inexpensive hardware, in contrast to the highly so- and loop stability. Otherwise, they have poor step response
phisticated and expensive hardware described in [4], [6]. contrary to dual-slope ADCs. The realized algorithm is com-
Another important advantage of this method (integrating parable in accuracy with [6]–[8], but yet it has a considerably
ADC) is that the input signal becomes averaged as it drives modest structure.

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
PETROVIC: NEW DIGITAL MULTIMETER FOR ACCURATE MEASUREMENT 721

III. PRACTICAL REALIZATION OF THE PROPOSED MULTIMETER

The block diagram of the proposed multimeter is shown in


the Fig. 5.
After being adjusted to the measuring range of a converter,
both voltage and current signals are brought into the acquisi-
tion board. The voltage signal has been adopted from a precise
resistance network (in the range 0–400 V). The transducer con-
sists of a resistive voltage divider buffered with a low-distortion,
low-noise, and wideband operation amplifier. Precision resis-
tors (tolerance: 0.01%, nominal power: 0.6 W, and tempera-
ture coefficient: ppm C) have been used for the voltage
divider. When 230 V are applied to the transducer, the power
consumption of each resistor is lower than 10% of the nominal
power. We have analyzed some error sources and their effects, Fig. 6. Circuit for detecting the frequency of the measured signal (R = 330 k,
C = 220 pF, R = 10 k, and R = 100 k, R = 1 k).
which might be commonly encountered in using the series-re-
sistor-type voltage dividers. For calibration of the voltage trans-
ducers we have used the procedure described in [13]. of both sinusoidal and distorted current waveforms (provided
The current signal from an accurate current transformer has the dc components are not present). The uncertainties due to the
been taken for the input of the dual-slope ADC (Linear Tech- exciting current, which is the main source of errors in instru-
nology ADC TC530) (in the range 0–10 A). In this case, cur- ment current transformers, can be strongly reduced, provided
rent transducers consist of the traditional current transformers that an accurate preliminary identification of the transformer
(CT) with a magnetic core. Under sinusoidal conditions the ac- has been performed. On this basis, a compensation technique
curacy of the passive components can be very high, provided for CT has been set up that allows obtaining, simply by mea-
that the load applied to their output is close to the nominal suring the secondary current, a much more accurate estimate of
burden. When distorted waveforms are dealt with (typically be- the primary quantity than that usually achievable using the CTs
cause of the presence of power electronic components), these nominal ratio. The proposed procedure needs at least one pe-
devices can be inadequate, owing to their nonlinear behavior riod of the signal to be acquired to perform the compensation.
caused by the saturation and hysteresis phenomena. In any case, The proposed technique [10], [11] reduces the so-called com-
in order to comply with safety requirements, current transducers posed error (for 1%), which includes amplitude errors, time shift
must guarantee galvanic separation between the power system between primary and secondary current, and harmonic content.
and the measurement system. In this case, the overall behavior The current transformer has a secondary circuit operation ampli-
of the current-to-voltage transducer can be examined, offering fier, which provides practically zero resistance and, thus, much
high accuracy specifications (0.01% at 20 C) and residual in- better linearity of the transfer function of all transformers. To
ductance so low as to be negligible in all performed tests. minimize common-mode errors, the voltage developed across
It is interesting to observe that the phase difference, unlike is isolated and converted to a ground-referenced signal using a
that which happens for the active devices, decreases for in- unity-gain differential amplifier. Feedback amplifiers have been
creasing frequency. This is due in practice to the magnetizing employed to increase, in effect, the permeability of transformer
current, the main factor responsible for inaccuracies of CTs cores. Such techniques have been quite successful in reducing
at low frequencies. The magnetizing current decreases as the low-frequency errors [12], [15].
frequency increases. Hall-effect probes may represent the right solution in many
On the other hand, traditional current transformers are intrin- cases, since they can measure dc components and their band-
sically unsuitable when currents with a dc component have to width extends up to hundreds of kilohertz. Moreover, they are
be measured. In addition, the presence of dc components, su- practical to use. On the other hand, the two main drawbacks of
perimposed on alternating components, can make it impossible these transducers are the fact that measurement results depend
to use the transformer correctly, owing to the polarization of its on the position of the primary wire with respect to the core axis
magnetic core. As for the problems concerning the nonlinearity, and the Hall sensor, and that results are influenced by magnetic
in traditional current transformers, the effects of saturation and fields generated by nearby power wires. This is the reason why
hysteresis, more or less evident but always present in the mag- we did not use this type of current transducer.
netic core, cause distortion in the secondary waveform. Thus, The TC530 consists of a dual-slope integrating ADC, nega-
they limit the above-observed advantages on phase shifts. tive power supply generator and three-wire serial interface port.
Suitable compensation methods [10], [11] could reduce this The key ADC operating parameters (auto zero and integration
problem. A digital technique improved the accuracy of instru- time) are programmable, allowing the user to trade off conver-
ment current transformers. A simple scalar model for CTs mag- sion time for resolution. Integration time must be chosen as a
netic core, taking into account saturation as well as hysteresis multiple of the period of the line frequency. The ADC conver-
and eddy current phenomena, has been implemented in a soft- sion time must be less than the distance between two consecu-
ware compensation routine. This allows us to improve the ac- tive samples (1). Data conversion is initiated when the RESET
curacy in the reproduction of the primary current, in the case input is brought low. After conversion, data is loaded into the

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
722 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 3, JUNE 2004

Fig. 7. Flow chart of the program running on the microcontroller.

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
PETROVIC: NEW DIGITAL MULTIMETER FOR ACCURATE MEASUREMENT 723

output shift register, and EOC is asserted, indicating new data is


available. The converted data (plus over-range and polarity bits)
are held in the output shift register until read by the processor,
or until the next conversion is completed, allowing the user to
access data at any time. The TC530 requires a single 5-V power
supply and features a 5-V, 10-mA output, which can be used to
supply negative bias to other components in the system. We use
two ADCs, one for the voltage and one for the current signal.
The voltage and current signals must be adapted to the range of
2 V to activate the maximum linearity of the ADC used, and
over the special resistant circuit specify the reference signal on
1.025 V.
As sample-and-hold circuit, we use the Analog Devices
circuit AD684. The AD684 is ideal for high-performance,
multichannel data acquisition systems. Independent inputs,
outputs, and sample-and-hold controls allow user flexibility
in the system architecture. Each SHA channel can acquire a
signal in less than 1 ms and retain the held value with a droop
rate of less than 0.01 mV/ms. The AD684 has a self-correcting
architecture that minimizes hold-mode errors and insures
accuracy over temperature, with a fast acquisition time (1 s) Fig. 8. Photo of the realized digital multimeter.
and low aperture jitter (75 ps). Each channel of the AD684 is
capable of sourcing 5 mA and incorporates output short-circuit
protection. Low droop (0.01 mV/ms) and internally compen-
sated hold-mode error result in superior system accuracy. DSP for control, and consequently the price of the final instru-
A special circuit detects the passing of the signal through zero ments will be higher.
with a comparator (Fig. 6), in this way achieving synchroniza- In the existing electric utilities, the system frequency swings
tion of the measuring cycle with the electric utility frequency. In in the range of 49.06–50.02 Hz (allowed by existing regula-
order to avoid multiple false detections of zero crossing, caused tions). This can certainly influence the accuracy of the proposed
by noise, the comparator is replaced by Schmitt-triggers. The measuring concept, due to the error made when determining the
microcontroller generates sampling intervals and is able to per- sampling period by zero crossing detection. The error can be as
form the necessary calculations, based on sampling values of the high as 2%. When the period is read using an internal counter in
measured signals. The analyses of operation of the zero crossing the microprocessor, a zero crossing is required. The least expen-
detector proved that satisfactory accuracy could be achieved. sive comparators have a slew-rate of 50 s. As the system
The number of zero crossings of the signal is evaluated by a test voltage is scaled to about 2 V on the board (ratio of 1:150),
on the sign of consecutive samples. The program corrects for this comparator triggers at about 2.5 mV, so the error is about
multiple transitions due to noise and checks that the distance 20 ns. This error can be ignored, since there is no accumulation.
between zero crossings is approximately compatible with the The accuracy of period reading then depends only on the pro-
frequency expected by the program. An odd integer number of cessing power of the microprocessor. The processing demands
valid zero crossings are taken, and the frequency of the signal described above can be satisfied using a wide range of low-cost
is calculated from the period of time between the first and the microprocessors. The microprocessors are also capable of car-
last zero crossing, divided by the number of periods. For better rying out the DFT to detect the harmonic content of an input
accuracy, the values of the samples around these two zero cross- signal, automatically adjusting the algorithm to the real condi-
ings are interpolated by means of a least squares procedure. In tions in electric utilities (special subroutine in Fig. 7).
this way, we eliminate possible errors if the input periodic signal The following constraints must hold to attain the lowest un-
has more than two zero crossings per period. In real electric util- certainties. 1) must hold at all times for a mul-
ities, the possible dc component is never higher than its ampli- tiple of two. This is the condition for synchronous sampling of
tude, and this is the reason why we did not consider this special a signal with frequency generated from a common
problem. In the case of such high dc components, we must trans- clock-reference. 2) The number of sampled periods must be
late the input voltage signal or adapt the level of comparison. an integer multiple of power-line cycles in order to reduce power
The usage of PLL multiplier circuits offers much better syn- line interferences. Conditions 1) and 2) prevent artificial spectral
chronization with the input signal frequency, avoiding errors components (leakage) from appearing when doing a DFT on the
caused by frequency measurements. The majority of all PLL sampled data. The realized instrument was checked in the fre-
design problems can be approached using the Laplace Trans- quency range from 45–65 Hz. The methods for estimating the
form technique, and relatively complex mathematical tools for frequency of a signal under noisy conditions [9] are also consid-
their realization. One project demand was simplicity in realiza- ered. The proposed algorithm [9] is suitable for real-time appli-
tion and usage of very simple microprocessors such as Motorola cations, especially when the frequency changes are abrupt and
68HC11. The existing PLL circuit needs a very sophisticated the signal is corrupted with noise and other disturbances due

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
724 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 3, JUNE 2004

TABLE II TABLE III


RESULTS OF MEASURING IN THE NATIONAL INSTITUTE FOR MEASUREMENTS IN EXPERIMENTAL RESULTS OF MEASURING IN THE REAL ENVIRONMENT
BELGRADE WITH THE PROPOSED DIGITAL MULTIMETER; IN THE FIRST WITH THE PROPOSED DIGITAL MULTIMETER WITHOUT
CASE, WE USE 40 SAMPLES TO CALCULATE THE OBSERVED USING A SAMPLE-AND-HOLD CIRCUIT
VALUES, AND IN THE SECOND CASE, WE USE 80 SAMPLES

to harmonics. It is highly iterative and, therefore, needs a fast In order to test the accuracy of the proposed method of mea-
processor. surement without the sample-and-hold circuit, after adopting the
Complete control of the measuring process and all necessary proposed algorithm for this purpose [5], parallel measurements
calculations are performed by the microcontroller (Motorola of rms voltage with an extremely precise multimeter, Hewlett
68HC11) (a flow chart in Fig. 7). The microcontroller works by Packard HP 3475 A were carried out. These parallel measure-
a program, which is stored in EEPROM, and complete calcula- ments (Table III) provide verification of the suggested mea-
tions were realized using (1) and (2), with the specified , suring concept, and give more accurate measurement results. In
, s. Thus, the obtained results are sent the future, it is necessary to check this “new” instrument in one
to a display so that the process of measuring can be followed vi- of the authorized laboratories in the country.
sually, too. An interface circuit (RS-232) is designed to enable
a connection between the meter and PC with the possibility to
choose the number of samples needed to calculate observed ac IV. CONCLUSION
values. Over the installed display, we can visually follow the re-
The main advantage of the digital multimeter described in this
sults of calculation (rms values of voltage, current, active power,
paper is that the measurements are performed exactly according
and frequency of the basic harmonics of voltage).
to the determination of power and true rms voltage. The multi-
The installed keyboard, which is used to set up the measuring meter works with an algorithm that is suitable for on-line mea-
range and the start of measurement, is not presented in the Fig. 5. surements, without the high computational burden as in [3], [4],
We can also choose the type of measurement. For example, this [6]. The algorithm is of an adaptable type and depends on the
multimeter can be used for measuring the rms values of voltage harmonic content of the input signal and network frequency (50
and current, as well as average and reactive power. or 60 Hz) [1]. The dual-slope ADC converter with very simple
The realized multimeter (Fig. 8) was checked in the National and inexpensive hardware, in contrast to the highly sophisti-
Institute for Measurements in Belgrade as the official laboratory cated and expensive hardware described in [4], [6], [7], meets all
for this purpose. The obtained results completely proved our as- price and accuracy requirements for the design of the measure-
sumptions [1] (Table II). The instruments for calibration in the ment system. This reduces the price of the entire device, while
Institute during the measurement with the proposed multimeter the high level of accuracy in processing ac values is preserved,
show that rms values of voltage, current and active power are: better than in some other solutions [7], [14], with excellent noise
, , and rejection.
. These parallel measurements showed the same re- The necessary synchronization is achieved by software
sults up to the third decimal. This provides verification for using measurements of the frequency of the measured signal. The
the suggested measuring concept in extremely precise reference prototype of a digital multimeter is being tested in the National
and laboratory measurements. As can be seen from Table II, Institute for Measurement in Belgrade. The precision limit of
if we use more samples, then the results of calculation will be the instrument was investigated theoretically, experimentally
better. and by simulation. The real limit of precision was found
The multimeter can be used as a watt-hour meter in real elec- experimentally, and it was 0.01%. The most accurate watt-hour
tric utilities, where there is the essential monitoring of basic ac meters used in the Yugoslav electric power net are three-phase
values (voltage, current, and frequency). For this reason, we can induction watt-hour meters of class 0.2%. Field experiments
use ADCs with low resolution (up to 12 bits) for realization. The gave excellent confirmation of the good application properties
results show that the accuracy of the proposed algorithm is ac- of this instrument. By eliminating the sample-and-hold circuit
ceptable under all conditions. It is independent of the harmonic from the final design of the measuring system, a possible source
content of the voltage and current signals and has good accuracy of systematic errors is eliminated and hardware requirements
even if the frequency deviates from the nominal value. are significantly simplified.

Authorized licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.
PETROVIC: NEW DIGITAL MULTIMETER FOR ACCURATE MEASUREMENT 725

REFERENCES [11] X. Qifeng and A. Refsum, “A new method to compensate current


transformer using microprocessors,” in Proc. Inst. Elect. Eng. Int. Conf.
[1] P. Petrovic, S. Marjanovic, and M. Stevanovic, “Digital method for Advance in Power System Control, Operation and Management, Hong
power frequency measurement using synchronous sampling,” Proc. Kong, Nov. 1991, pp. 355–358.
Inst. Elect. Eng., vol. 146, no. 4, pp. 383–390, July 1999. [12] E. So, “Harmonic Measurements: Current and Voltage Transducers,”
[2] E. R. Hnatek, A User’s Handbook of D/A and A/D Converters. New IEEE, Rep. no. 0-7803-5569-5/99, 1999.
York: Wiley, 1976. [13] K.-Y. Kim, S. I. Park, Y. M. Cho, and S.-W. Kwon, “Common error
[3] J. Xi and J. F. Chicharo, “A new algorithm for improving the accuracy sources in using a series-resistor-type voltage divider below the level
of periodic signal analysis,” IEEE Trans. Instrum. Meas., vol. 45, pp. of 0.01 ppm,” IEEE Trans. Instrum. Meas., vol. 46, pp. 776–780, Aug.
827–831, Aug. 1996. 1997.
[4] L. Toivonen and J. Morsky, “Digital multirate algorithms for measure- [14] C.-P. Young and M. J. Devaney, “Digital power metering manifold,”
ment of voltage, current, power and flicker,” IEEE Trans. Power De- IEEE Trans. Instrum. Meas., vol. 47, pp. 224–228, Feb. 1998.
livery, vol. 10, Jan. 1995. [15] N. Locci and C. Muscas, “Comparative analysis between active and pas-
[5] P. Petrovic, S. Marjanovic, and M. Stevanovic, “Measuring of slowly sive current transducers in sinusoidal and distorted conditions,” IEEE
changing AC signals without sample and hold circuit,” IEEE Trans. In- Trans. Instrum. Meas., vol. 50, pp. 123–128, Feb. 2001.
strum. Meas., vol. 49, pp. 1245–1248, Dec. 2000.
[6] H. Germer, “High-precision AC measurements using the Monte Carlo
method,” IEEE Trans. Instrum. Meas., vol. 50, pp. 457–460, Apr. 2001.
[7] B. C. Waltrip and N. M. Oldham, “Wideband wattmeter based on P. Petrovic was born in Cacak, Yugoslavia, on Jan-
RMS voltage measurements,” IEEE Trans. Instrum. Meas., vol. 46, pp. uary 26, 1967. He received the B.S.E.E. and M.Sc.
781–783, Aug. 1997. degrees in electrical engineering from the University
[8] T. M. Sounders, B. C. Waltrip, and O. B. Laug, “A wideband sampling of Belgrade, Belgrade, Yugoslavia, in 1991 and 1994,
voltmeter,” IEEE Trans. Instrum. Meas., vol. 46, pp. 947–953, Aug. respectively. He is currently pursuing the Ph.D. de-
1997. gree in the field of digital signal processing at the
[9] A. Routray, A. K. Pradham, and K. P. Rao, “A novel Kalman filter for fre- University of Novi Sad, Novi Sad, Yugoslavia.
quency estimation of distorted signals in power systems,” IEEE Trans. Since 1991, he has been a Teaching and Research
Instrum. Meas., vol. 51, pp. 469–479, June 2002. Assistant, Cacak College of Engineering, University
[10] N. Locci and C. Muscas, “Hysteresis and eddy currents compensation of Kragujevac, Cacak, Yugoslavia, where his main in-
in current transformers,” IEEE Trans. Power Delivery, vol. 16, pp. terests are digital signal processing, microcontroller
154–159, Apr. 2001. programming, AD conversion, mathematics, and cryptology.

Authorized
View publication stats licensed use limited to: Predrag B Petrovic. Downloaded on September 3, 2009 at 12:20 from IEEE Xplore. Restrictions apply.

You might also like